4 Technical Description
4.10 RS 232 interface
Device address 1
9600 bauds
Parity none
4.11 Cold start delay
This function is used to suppress false measurement values while the probe is warming up. A cold start delay is always activated after “Power off” and probe replacement. The cold start delay can be interrupted at any time:
• using the multi-function starter
• using the display and operating unit
• using remote display software, see separate publication
During the cold start delay or a fault:
• a default value (factory setting)
O
2
→
0 vol. % (P361), COe Æ 0 ppm (P371) can be output.
• In P362 for O
2 in P372 for CO
,
e
the “Type of default value” can be set:
OFF: No default value is output.
ON: The default value set in the parameter in front is output.
+Maintenance (factory setting): The default value set in the parameter in front is output during “MAINTENANCE” as well.
+Maint.froz.: With this setting, the default value set in the parameter in front is output during cold start and fault, as before, but in addition the previous measurement value is frozen as long as Maintenance mode is active.
During cold start/fault, the default value has priority over the freezing of the measurement value during maintenance.
After a heating phase of 10 minutes, the probe voltage U-O2 stabilises to values between -20 mV…+20mV and the internal cell resistance Ri settles at values below
100
Ω
.
34
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Compliance with the German law on device safety
- 7 Proper use, conditions for use
- 8 Permitted users
- 9 Safety equipment / protective measures
- 10 Environmentally responsible behaviour, disposal instructions
- 11 System overview
- 15 Theoretical principles of measurement
- 15 -sensitive electrode
- 18 Advantages of the LAMTEC measurement principle
- 18 Necessary components
- 19 Basic design of the LT 2/KS1-DK
- 20 LT 2/KS1-DK
- 22 Internal display and control elements of the LT 2/KS1-DK lambda transmitter
- 24 Display and operating unit
- 32 Analogue output
- 33 Digital inputs
- 35 LAMTEC SYSTEM BUS
- 36 RS 232 interface
- 36 Cold start delay
- 37 Optional extras
- 45 Limit values
- 50 Analogue inputs
- 51 Device configuration and factory settings
- 53 General Information
- 53 Installing the electronic evaluation unit of the LT 2/KS1-DK lambda transmitter
- 57 Installing the probe
- 59 Switch on measurement
- 61 Setting the reference air quantity
- 68 Ending calibration
- 69 Test report
- 70 Probe certificate
- 71 Setting service warnings
- 71 Practical information for operation
- 71 Taking out of operation
- 72 Checking/calibrating the KS1-DK combination probe
- 73 Checking the LT 2/KS1-DK
- 74 Maintenance
- 77 Fault indicator via LED line in the LT 2/KS1-DK
- 80 Warning indicator via LED line in the LT 2/KS1-DK
- 84 Resetting faults/warnings
- 85 Wearing parts
- 86 Spare parts
- 88 Technical data of the LT 2/KS1-DK lambda transmitter
- 91 Technical data of the KS1-DK combination probe
- 92 Electrical connection
- 95 Pneumatic connection
- 96 Electrical connection to field bus module
- 97 Fuses
- 98 Jumpers
- 99 Dimension drawing of KS1-DK combination probe
- 100 Dimension drawing of LT 2/KS1-DK in wall-mounted housing
- 101 Dimension drawing of LT 2/KS1-DK on mounting plate
- 102 Dimension drawing of display and operating unit for panel installation
- 103 Wet/dry measurement – deviation, conversion table
- 104 LT 2/KS1-DK lambda transmitter