advertisement
7.11.3 Resetting Care Group Alarms
You can reset the alarm of the currently viewed bed in the View Other Patient window. To enable this function:
Setup>>].
2. In the [Alarm Setup] menu, set [Reset Other Bed’s Alarms] to [On].
When your monitor is viewed by other monitors and [Alarm Reset By Other Bed] is switched on, alarms on your monitor can be reset by other monitors.
WARNING
Resetting care group alarms may cause a potential hazard.
T1 Patient Monitor Operator’s Manual 7-15
FOR YOUR NOTES
7-16 T1 Patient Monitor Operator’s Manual
8
Monitoring ECG
8.1 Introduction
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on the monitor as a waveform and a numeric. ECG monitoring provides the following algorithms:
Advanced ECG algorithm
The Advanced ECG algorithm provides 3-, 5- and 12-lead ECG monitoring, ST-segment analysis, and extended arrhythmia analysis.
Mortara ECG algorithm
The Mortara ECG algorithm provides 3-, 5- and 12-lead ECG monitoring, ST-segment analysis, and arrhythmia analysis.
Glasgow algorithm
Glasgow algorithm provides an interpretation of the resting 12-lead ECG.
You can select algorithms at the time of purchase or as upgrades post sale:
The equipment incorporating the Advanced ECG algorithm is labeled with a ST/ARR label.
The equipment incorporating the Mortara ECG algorithm is labeled with the Mortara logo.
The equipment incorporating the Glasgow algorithm is labeled with the Glasgow logo.
8.2 Safety
WARNING
This equipment is not intended for direct cardiac application.
Use only ECG electrodes and cables specified by Mindray.
Make sure the conductive parts of electrodes and associated connectors for applied parts, including the neutral electrode, should not contact any other conductive parts including earth.
Periodically inspect the electrode application site to ensure skin quality. If the skin quality changes, replace the electrodes or change the application site.
Use defibrillation-proof ECG cables during defibrillation.
Do not touch the patient or metal devices connected to the patient during defibrillation.
To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s cables and transducers never come into contact with the electrosurgery unit (ESU).
The neutral electrode of the HF surgical unit shall properly contact the patient. Otherwise, burns may result.
T1 Patient Monitor Operator’s Manual 8-1
CAUTION
Interference from a non-grounded instrument near the patient and electrosurgery interference can cause problems with the waveform.
NOTE
After defibrillation, the screen display recovers within 10 seconds if the correct electrodes are used and applied in accordance with the instructions for use.
8.3 Preparing to Monitor ECG
8.3.1 Preparing the Patient and Placing the Electrodes
To prepare the patient and place the electrodes:
1. Prepare the patient’s skin. Proper skin preparation is necessary for good signal quality at the electrode site, as the skin is a poor conductor of electricity. To properly prepare the skin, choose flat areas and then follow this procedure:
Shave hair from skin at chosen sites.
Gently rub skin surface at sites to remove dead skin cells.
Thoroughly cleanse the site with a mild soap and water solution. We do not recommend using ether or pure alcohol, because this dries the skin and increases the resistance.
Dry the skin completely before applying the electrodes.
2. Attach the clips or snaps to the electrodes before placing them.
3. Place the electrodes on the patient.
4. Attach the electrode cable to the patient cable and then plug the patient cable into the ECG connector.
8.3.2 Choosing AHA or IEC Lead Placement
To choose AHA or IEC lead placement:
1. Select the ECG parameter window or waveform area to enter the [ECG Setup] menu. electrodes.
8-2 T1 Patient Monitor Operator’s Manual
8.3.3 ECG Lead Placements
The electrode placement illustrations in this chapter adopt the AHA standard.
3-Leadwire Electrode Placement
Following is an electrode configuration when using 3 leadwires:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
LL placement: on the left lower abdomen.
5-Leadwire Electrode Placement
Following is an electrode configuration when using 5 leadwires:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
RL placement: on the right lower abdomen.
LL placement: on the left lower abdomen.
V placement: on the chest.
The chest (V) electrode can be placed on one of the following positions:
V1 placement: on the fourth intercostal space at the right sternal border.
V2 placement: on the fourth intercostal space at the left sternal border.
V3 placement: midway between the V2 and V4 electrode positions.
V4 placement: on the fifth intercostal space at the left midclavicular line.
V5 placement: on the left anterior axillary line, horizontal with the V4 electrode position.
V6 placement: on the left midaxillary line, horizontal with the V4 electrode position.
V3R-V6R placement: on the right side of the chest in positions corresponding to those on the left.
VE placement: over the xiphoid process.
V7 placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.
V7R placement: on posterior chest at the right posterior axillary line in the fifth intercostal space.
T1 Patient Monitor Operator’s Manual 8-3
12-Leadwire Electrode Placement
12-lead ECG uses 10 electrodes, which are placed on the patient’s four limbs and chest. The limb electrodes should be placed on the soft skin and the chest electrodes placed according to the physician’s preference.
Lead Placement for Surgical Patients
The surgical site should be taken into consideration when placing electrodes on a surgical patient. e.g. for open-chest surgery, the chest electrodes can be placed on the lateral chest or back. To reduce artifacts and interference from electrosurgical units, you can place the limb electrodes close to the shoulders and lower abdomen and the chest electrodes on the left side of the mid-chest.
Do not place the electrodes on the upper arm. Otherwise, the ECG waveform will be very small.
WARNING
When using electrosurgical units (ESU), patient leads should be placed in a position that is equal distance from the electrosurgery electrotome and the grounding plate to avoid burns to the patient. Never entangle the ESU cable and the ECG cable together.
When using electrosurgical units (ESU), never place ECG electrodes near to the grounding plate of the ESU, as this can cause a lot of interference on the ECG signal.
8.3.4 Checking Paced Status
It is important to set the paced status correctly when you start monitoring ECG. The paced symbol is displayed when the [Paced] is set to [Yes]. The pace pulse markers “|” are shown on the ECG wave when the patient has a paced signal. If [Paced] is set to [No] or the patient’s paced status is not selected, the symbol will be shown in the patient information area.
To change the paced status, you can select either:
the patient information area, or
[Main Menu]→[Patient Setup]→[Patient Demographics], or,
the ECG parameter window or waveform area→[Others >>], and then, select [Paced] from the pop-up menu and then select [Yes] or [No].
If you do not set the paced status, the monitor issues a prompt tone when pace pulse is detected. At the same time, the paced symbol flashes and the message [Please confirm the pace of patient] appears in the ECG waveform area. Then, please check and set the paced status of the patient.
8-4 T1 Patient Monitor Operator’s Manual
WARNING
For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the monitor could mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak.
On ventricular paced patients, episodes of Ventricular Tachycardia may not always be detected.
False low heart rate indicators or false Asystole calls may result with certain pacemakers because of pacemaker artifact such as electrical overshoot of the pacemaker overlapping the true QRS complexes.
Do not rely entirely upon the system’s automated arrhythmia detection algorithm. Keep pacemaker patients under close surveillance.
For non-paced patients, you must set [Paced] to [No].
The auto pacer recognition function is not applicable to pediatric and neonatal patients.
8.4 Understanding the ECG Display
Your display may be configured to look slightly different.
1 2 3 4
1. Lead label of the displayed wave
3. ECG filter label
4. Notch filter status
Additionally, when a paced signal has been detected, the pace pulse marks “|” are shown on the ECG wave if the [Paced] setting has been configured to [Yes].
2
3
1
1. Current heart rate alarm limits
2. Heart beat symbol
3. Current heart rate
For 12-lead ECG display screen, refer to section 8.10 12-Lead ECG Monitoring.
T1 Patient Monitor Operator’s Manual 8-5
8.5 Changing ECG Settings
8.5.1 Accessing ECG Menus
By selecting the ECG parameter window or waveform area, you can access the [ECG Setup] menu.
8.5.2 Setting Pacemaker Rate (For Mortara ECG Algorithm only)
Some pacemaker pulses can be difficult to reject. When this happens, the pulses are counted as a QRS complex and could result in an incorrect HR and failure to detect some arrhythmias. You can set [Pacemaker Rate] to the pacemaker’s rate in the [ECG Setup] menu. In this way, the monitor can calculate HR and detect arrhythmias more accurately. When
[Paced] is set to [No], the pacemaker rate cannot be set.
8.5.3 Choosing the Alarm Source
In most cases the HR and PR numerics are identical. In order to avoid simultaneous alarms on HR and PR, the monitor uses either HR or PR as its active alarm source. To change the alarm source, select [Alm Source] in the [ECG Setup] menu and then select either:
[HR]: if you want the HR to be the alarm source for HR/PR.
[PR]: if you want the PR to be the alarm source for HR/PR.
[Auto]: If the [Alm Source] is set to [Auto], the monitor will use the heart rate from the ECG measurements as the alarm source whenever a valid heart rate is available. If the heart rate becomes unavailable, for example the ECG module is turned off or becomes disconnected, the monitor will automatically switch to PR as the alarm source.
8.5.4 Changing ECG Wave Settings
In the [ECG Setup] menu:
You can select [ECG], [ECG1], or [ECG2] to select a lead to view. The ECG lead (s) selected is used for analysis (beat detection, beat classification, and V-Fib detection). The waveform of selected lead should have the following characteristics:
The QRS should be either completely above or below the baseline and it should not be biphasic.
The QRS should be tall and narrow.
The P-waves and T-waves should be less than 0.2mV.
If the wave is too small or clipped, you can change its size by selecting an appropriate [Gain] setting. If you select
[Auto] from [Gain], the monitor will automatically adjust the size of the ECG waves. In normal screen, only the selected ECG wave can be adjusted. In other screens, all ECG waves can be adjusted simultaneously.
You can change the wave sweep speed by selecting [Sweep] and then selecting the appropriate setting.
NOTE
Changing the display gain on the monitor does not affect the amplitude of signal that is used by the algorithm for beat detection.
8-6 T1 Patient Monitor Operator’s Manual
8.5.5 Changing the ECG Filter Settings
The ECG filter setting defines how ECG waves are smoothed. To change the filter setting, select [Filter] from [ECG Setup] and then select the appropriate setting.
[Mon]: Use under normal measurement conditions.
[Diag]: Use when diagnostic quality is required. The unfiltered ECG wave is displayed so that changes such as
R-wave notching or discrete elevation or depression of the ST segment are visible.
[Surg]: Use when the signal is distorted by high frequency or low frequency interference. High frequency interference usually results in large amplitude spikes making the ECG signal look irregular. Low frequency interference usually leads to wandering or rough baseline. In the operating room, the surgery filter reduces artifacts and interference from electrosurgical units. Under normal measurement conditions, selecting [Surgery] may suppress the QRS complexes too much and then interfere with ECG analysis.
[ST]: Use when ST monitoring is applied.
CAUTION
The [Diagnostic] filter is recommended when monitoring a patient in an environment with slight interference only.
8.5.6 Setting the ECG Lead Set
You can set the [Lead Set] by selecting [ECG Setup]
→[Others >>]. You can set the [Lead Set] as [Auto] if the auto lead detection function is available.
8.5.7 Choosing an ECG Display Screen
When monitoring with a 5-lead or 12-lead set, you can select [Main Menu]→[Screens], or [Others>>]→[ECG Display] in the [ECG Setup] menu to choose the screen type as:
[Normal]: The ECG waveform area shows 2 ECG waveforms.
[Full-Screen]: The whole waveform area shows 7 ECG waveforms only.
[12-Lead]: The whole waveform ares displays 12 ECG waveforms (for 12-lead set only).
When monitoring with a 3-lead set, the screen type can only be set to [Normal].
When the screen type is set to [Normal], cascaded ECG waveforms can be displayed.
To cascade ECG waveforms:
1. Select the [Main Menu]→[Screens]→[Screen Setup].
T1 Patient Monitor Operator’s Manual 8-7
8.5.8 Setting the Notch Filter
The notch filter removes the line frequency interference. Only when [Filter] is set to [Diagnostic], the [Notch Filter] is adjustable.
To set the notch filter:
1. Select the ECG parameter window or waveform area to enter its setup menu. Then select [Others >>].
[Strong]: when waveform interference is strong (such as spikes).
[Weak]: when waveform interference is weak.
[Off]: to turn the notch filter off.
Set notch frequency according to the electric power frequency of your country. To set notch filter frequency: required password.
8.5.9 Changing the Pacer Reject Settings
To change the pacer reject settings, select [ECG Setup]→[Others>>]→[Pacer Reject], and then select [On] or [Off].
When [Paced] is set to [Yes]:
When [Pacer Reject] is switched on, the pace pulses are not counted as extra QRS complexes.
The pace pulse marks “|” are shown on the ECG wave when pace pulses are detected.
When [Paced] is set to [No], the pace markers are not shown on the ECG wave, and the options of [Pacer Reject] are inactivated.
8.5.10 Enabling Smart Lead Off
When the smart lead off function is enabled and there is a “lead off” in the lead of the first ECG wave, if another lead is available, the monitor will automatically select the available lead. The system will re-calculate HR and analyze and detect arrhythmia. When the “lead off” condition is corrected, the leads are automatically switched back.
To switch on/off the smart lead off function, select [Others >>] from the [ECG Setup] menu; select [Smart Lead Off] and then select [On] or [Off] from the pop-up menu.
8.5.11 Setting the Alarm Level for ECG Lead Off Alarms
Select [Alarm Setup >>] from the [User Maintenance] menu. You can set [ECGLeadOff Lev.] from the pop-up menu.
8-8 T1 Patient Monitor Operator’s Manual
8.5.12 Adjusting QRS Volume
QRS sounds are produced based on the alarm source.
To adjust the QRS volume,
When a valid SpO
2
measured value is available, the system adjusts the pitch tone of QRS sound based on the SpO
2
value.
8.5.13 Adjusting the Minimum QRS Detection Threshold (For Advanced ECG Algorithm
Only)
To avoid false asystole alarms when the R wave amplitude is low and missed asystole alarms during ventricular standstill
(tall P waves, but no QRS), a means to manually adjust the minimum QRS detection threshold is provided.
To adjust the QRS threshold,
1. In the [ECG Setup] menu, set [Filter] to [Monitor].
2. Select [Others >>]→[Minimum QRS Threshold >>] to enter the [Minimum QRS Threshold] menu.
3. Select the up or down arrow to adjust the QRS threshold. Selecting [Default] resets the QRS threshold to the default value (0.16 mV).
CAUTION
The setting of QRS threshold can affect the sensitivity of arrhythmia, ST, QT/QTc detection, and heart rate calculation.
If QRS amplitude is low, the monitor might not be able to calculate heart rate and false asystole may occur.
NOTE
The minimum QRS detection threshold can only be adjusted when the ECG filter is set to [Monitor].
T1 Patient Monitor Operator’s Manual 8-9
8.5.14 About the Defibrillator Synchronization
A synchronization pulse (100 ms, +5 V) is output through the multifunctional connector each time the monitor detects an R-wave.
To use the defibrillator synchronization function, connect the monitor and the defibrillator with a synchronization cable.
WARNING
Improper use of a defibrillator may cause injury to the patient. The user should determine whether to perform defibrillation or not according to the patient’s condition.
8.6 About ST Monitoring
Mortara ST segment analysis is not intended for neonatal patients.
ST segment analysis calculates ST segment elevations and depressions for individual leads and then displays them as numerics in the ST1 and ST2 areas.
A positive value indicates ST segment elevation; a negative value indicates ST segment depression.
Measurement unit of the ST segment deviation: mV or mm. You can set the unit in the [Unit Setup] menu from the
[User Maintenance] menu.
Measurement range of the ST segment deviation: -2.0 mV to +2.0 mV.
WARNING
The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment changes need to be determined by a clinician.
8-10 T1 Patient Monitor Operator’s Manual
8.6.1 Switching ST Monitoring On and Off
To switch ST monitoring on or off:
1. In the [ECG Setup] menu, select [ST Analysis >>].
Reliable ST monitoring cannot be ensured if:
You are unable to get a lead that is not noisy.
Arrhythmias such as atrial fib/flutter cause irregular baseline.
The patient is continuously ventricularly paced.
The patient has left bundle branch block.
In these cases, you may consider switching ST monitoring off.
8.6.2 Changing ST Filter Settings
ST-segment analysis can be carried out only when the filter mode is set to [Diagnostic] or [ST]. When ST-segment analysis is switched on, [Filter] will automatically switch to [ST] if it is not [Diagnostic] or [ST]. When ST-segment analysis is switched off, the filter mode automatically switches to previous manual setting.
However, if you switch [Filter] to [Monitor] or [Surgery], ST-segment analysis will turn off automatically. If you change
[Monitor] or [Surgery] to [Diagnostic] or [ST], ST-segment analysis remains off, you can turn it on manually.
8.6.3 Understanding the ST Display
8.6.3.1 ST Numerics
This example shows ST numerics with 5-lead ECG. Your monitor screen may look slightly different from the illustration.
8.6.3.2 ST Segment
ST segment shows a QRS segment for each measured ST lead. The current ST segment is drawn in the same color as the
ECG wave, usually green, superimposed over the stored reference segment, drawn in a different color. The information is updated once every ten seconds.
To display the ST segment on normal screen:
1. Enter the [ST Analysis] menu. Set [ST Analysis] to [On].
2. Enter the [Screen Setup] window from the [Screens] menu. Set [ST Segment] to be displayed.
T1 Patient Monitor Operator’s Manual 8-11
advertisement
Related manuals
advertisement
Table of contents
- 42 1 Safety
- 42 1.1 Safety Information
- 42 1.1.1 Warnings
- 42 1.1.2 Cautions
- 42 1.1.3 Notes
- 42 1.2 Equipment Symbols
- 43 2 The Basics
- 43 2.1 Monitor Description
- 43 2.1.1 Intended Use
- 43 2.1.2 Equipment Features
- 43 2.1.3 Applied Parts
- 43 2.2 Equipment Description
- 43 2.2.1 Main Unit
- 43 2.2.2 T1 handle
- 43 2.2.3 T1 Docking Station
- 43 2.2.4 External Parameter Modules
- 43 2.3 Installation
- 43 2.3.1 T1 in Use with the T1 Handle
- 43 2.3.2 T1 Handle in Use with the T1 Docking Station
- 43 2.3.3 T1 in Use with the T1 Docking Station
- 43 2.4 Display Screen
- 44 3 Basic Operations
- 44 3.1 Installation
- 44 3.1.1 Unpacking and Checking
- 44 3.1.2 Environmental Requirements
- 44 3.2 Getting Started
- 44 3.2.1 Connecting to Power Source
- 44 3.2.2 Turning Power On
- 44 3.2.3 Starting Monitoring
- 44 3.3 Turning Off the Monitor
- 44 3.4 Using the Touchscreen
- 44 3.5 Using the On-Screen Keyboard
- 44 3.6 Using the External Display
- 44 3.7 Using the Mouse and Keyboard
- 44 3.8 Using the Main Menu
- 44 3.9 Changing General Settings
- 44 3.9.1 Setting Up a Monitor
- 44 3.9.2 Changing Language
- 44 3.9.3 Setting the Date and Time
- 44 3.9.4 Setting the Docking Station
- 44 3.10 Setting Parameters
- 61 3.10.1 Switching the Parameters On/Off
- 61 3.10.2 Accessing the Parameters Menu
- 61 3.11 Operating Modes
- 61 3.11.1 Monitoring Mode
- 61 3.11.2 Night Mode
- 61 3.11.3 Privacy Mode
- 61 3.11.4 Outdoor Mode
- 61 3.11.5 Module Mode
- 61 3.11.6 Standby Mode
- 62 4 Managing Patients
- 62 4.1 Admitting a Patient
- 62 4.2 Quick Admitting a Patient
- 62 4.3 Querying and Obtaining Patient Information
- 62 4.4 Associating Patient Information
- 62 4.5 Editing Patient Information
- 62 4.6 Discharging a Patient
- 62 4.7 Transferring Patient Data
- 62 4.7.1 Transferring Patient Data via a USB Drive
- 62 4.7.2 Transferring Patient Data via the T1 to a Host Monitor
- 62 4.8 Auto Deleting History Data from the SD Card
- 62 4.9 Connecting to a CMS
- 63 5 Managing Configurations
- 63 5.1 Introduction
- 63 5.2 Accessing the [Manage Configuration] Menu
- 63 5.3 Changing Department
- 63 5.4 Setting Default Configuration
- 63 5.5 Saving Current Settings
- 63 5.6 Editing Configurations
- 63 5.7 Deleting a Configuration
- 63 5.8 Transferring a Configuration
- 63 5.9 Loading a Configuration
- 63 5.10 Restoring the Latest Configuration Automatically
- 64 6 User Screens
- 64 6.1 Adjusting the Screen Brightness
- 64 6.2 Adjusting Volume
- 64 6.3 Configuring Your Screens
- 64 6.3.1 Changing the Waveform Line Size
- 64 6.3.2 Changing Measurement Colors
- 64 6.3.3 Choosing a Screen
- 64 6.3.4 Changing the T1 Screen Layout
- 64 6.3.5 Changing Screen Layout on the External Display
- 64 6.4 Understanding the Big Numerics Screen
- 64 6.5 Viewing Minitrends (Only Available with the External Display)
- 85 6.5.1 Having a Split-Screen View of Minitrends
- 85 6.5.2 Setting Minitrends
- 85 6.6 Viewing OxyCRG (only available with the external display)
- 85 6.7 Viewing Other Patients (Only Available with the External Display)
- 85 6.7.1 Care Group
- 85 6.7.2 Viewing the Care Group Overview Bar
- 85 6.7.3 Understanding the View Other Patient Window
- 86 7 Alarms
- 86 7.1 Alarm Categories
- 86 7.2 Alarm Levels
- 86 7.3 Alarm Indicators
- 86 7.3.1 Alarm Lamp
- 86 7.3.2 Alarm Messages
- 86 7.3.3 Flashing Numeric
- 86 7.3.4 Audible Alarm Tones
- 86 7.3.5 Alarm Status Symbols
- 86 7.4 Alarm Tone Configuration
- 86 7.4.1 Setting the Minimum Alarm Volume
- 86 7.4.2 Changing the Alarm Volume
- 86 7.4.3 Setting the Interval between Alarm Sounds
- 86 7.4.4 Changing the Alarm Tone Pattern
- 86 7.4.5 Setting the Reminder Tones
- 86 7.5 Understanding the Alarm Setup Menu
- 86 7.5.1 Setting Alarm Properties for All Parameters
- 86 7.5.2 Adjusting Alarm Limits Automatically
- 86 7.5.3 Setting Alarm Delay Time
- 86 Technical Alarm Delay
- 86 7.5.5 Setting Recording Length
- 86 7.5.6 Entering CPB Mode (Cardiopulmonary Bypass Mode)
- 86 7.6 Pausing Alarms
- 86 7.7 Switching Off All Alarms
- 86 7.8 Resetting Alarms
- 86 7.9 Latching Alarms
- 86 7.10 Testing Alarms
- 86 7.11 Using Care Group Alarms (Only Available with the External Display)
- 86 7.11.1 Care Group Auto Alarms
- 86 7.11.2 Setting Care Group Alert Tone
- 86 7.11.3 Resetting Care Group Alarms
- 87 8 Monitoring ECG
- 87 8.1 Introduction
- 87 8.2 Safety
- 87 8.3 Preparing to Monitor ECG
- 87 8.3.1 Preparing the Patient and Placing the Electrodes
- 87 8.3.2 Choosing AHA or IEC Lead Placement
- 100 8.3.3 ECG Lead Placements
- 100 8.3.4 Checking Paced Status
- 100 8.4 Understanding the ECG Display
- 100 8.5 Changing ECG Settings
- 100 8.5.1 Accessing ECG Menus
- 100 8.5.2 Setting Pacemaker Rate (For Mortara ECG Algorithm only)
- 100 8.5.3 Choosing the Alarm Source
- 100 8.5.4 Changing ECG Wave Settings
- 100 8.5.5 Changing the ECG Filter Settings
- 100 8.5.6 Setting the ECG Lead Set
- 100 8.5.7 Choosing an ECG Display Screen
- 100 8.5.8 Setting the Notch Filter
- 100 8.5.9 Changing the Pacer Reject Settings
- 100 8.5.10 Enabling Smart Lead Off
- 100 8.5.11 Setting the Alarm Level for ECG Lead Off Alarms
- 100 8.5.12 Adjusting QRS Volume
- 100 8.5.13 Adjusting the Minimum QRS Detection Threshold (For Advanced ECG Algorithm Only)
- 100 8.5.14 About the Defibrillator Synchronization
- 100 8.6 About ST Monitoring
- 100 8.6.1 Switching ST Monitoring On and Off
- 100 8.6.2 Changing ST Filter Settings
- 100 8.6.3 Understanding the ST Display
- 100 8.6.4 Saving the Current ST Segment as Reference
- 100 8.6.5 Changing the Reference Segment
- 100 8.6.6 Deleting a Reference Segment
- 100 8.6.7 Changing the ST Alarm Limits
- 100 8.6.8 Setting the ST Alarm Delay Time
- 100 8.6.9 Adjusting ST Measurement Points
- 100 8.7 QT/QTc Interval Monitoring (For Advanced ECG Algorithm Only)
- 100 8.7.1 QT/QTc Monitoring Limitations
- 100 8.7.2 Enabling QT/QTc Monitoring
- 100 8.7.3 Displaying QT/QTc Parameters and Waveform
- 100 8.7.4 Accessing the QT View
- 100 8.7.5 Changing QT Settings
- 100 8.7.6 Changing QTc Formula
- 100 8.8 About Arrhythmia Monitoring
- 100 8.8.1 Understanding the Arrhythmia Events
- 100 8.8.2 Changing Arrhythmia Alarm Settings
- 100 8.8.3 Changing Arrhythmia Threshold Settings
- 100 8.8.4 Setting the Extended Arrhythmia (For Advanced ECG Algorithm Only)
- 100 8.8.5 Reviewing Arrhythmia Events
- 100 8.9 ECG Relearning
- 100 8.9.1 Initiating an ECG Relearning Manually
- 100 8.9.2 Automatic ECG Relearning
- 100 8.10 12-Lead ECG Monitoring
- 100 8.10.1 Setting ECG Waveform Sequence
- 128 8.10.2 Extending the rhythm lead waveform area
- 128 8.11 Resting 12-Lead ECG Analysis
- 128 8.11.1 Accessing the 12-Lead Screen
- 128 8.11.2 Entering Patient Information
- 128 8.11.3 12-Lead Setup
- 128 8.11.4 Resting 12-lead ECG Analysis
- 128 8.11.5 12-lead ECG Report
- 128 8.12 Troubleshooting
- 129 9 Monitoring Respiration (Resp)
- 129 9.1 Introduction
- 129 9.2 Safety Information
- 129 9.3 Understanding the Resp Display
- 129 9.4 Placing Resp Electrodes
- 129 9.4.1 Optimizing Lead Placement for Resp
- 129 9.4.2 Cardiac Overlay
- 129 9.4.3 Abdominal Breathing
- 129 9.4.4 Lateral Chest Expansion
- 129 9.5 Choosing the Respiration Lead
- 129 9.6 Changing the Apnea Alarm Delay
- 129 9.7 Changing Resp Detection Mode
- 129 9.8 Changing Resp Wave Settings
- 129 9.9 Setting Respiration Rate (RR) Source
- 129 9.10 Setting Alarm Properties
- 130 10 Monitoring PR
- 130 10.1 Introduction
- 130 10.2 Setting the PR Source
- 130 10.3 Selecting the Active Alarm Source
- 130 10.4 QRS Tone
- 131 11.1 Introduction
- 131 11.2 Safety
- 131 Modules
- 131 11.4 Applying the Sensor
- 131 Settings
- 131 11.5.1 Accessing SpO2 Menus
- 131 11.5.2 Adjusting the Desat Alarm
- 131 11.5.3 Setting SpO2 Sensitivity
- 131 11.5.4 Changing Averaging Time
- 131 11.5.5 Monitoring SpO2 and NIBP Simultaneously
- 131 11.5.6 Sat-Seconds Alarm Management
- 131 11.5.7 Changing the Speed of the Pleth Wave
- 131 11.5.8 Zooming PI Value (Only Available with the External Display)
- 131 11.5.9 Setting the Alarm Level for SpO2 Sensor Off Alarm
- 141 11.5.10 Setting the SpO2 Tone Mode
- 141 11.6 Measurement Limitations
- 141 11.7 Masimo Information
- 141 11.8 Nellcor Information
- 141 11.9 Troubleshooting
- 142 12 Monitoring NIBP
- 142 12.1 Introduction
- 142 12.2 Safety
- 142 12.3 Measurement Limitations
- 142 12.4 Measurement Methods
- 142 12.5 Setting Up the NIBP Measurement
- 142 12.5.1 Preparing the Patient
- 142 12.5.2 Preparing to Measure NIBP
- 142 12.5.3 Starting and Stopping Measurements
- 142 12.5.4 Correcting the Measurement if Limb is not at Heart Level
- 142 12.5.5 Enabling NIBP Auto Cycling and Setting the Interval
- 142 12.5.6 Enabling Measurement on Clock
- 142 12.5.7 Starting a STAT Measurement
- 142 12.6 Understanding the NIBP Numerics
- 142 12.7 Changing NIBP Settings
- 142 12.7.1 Setting the Initial Cuff Inflation Pressure
- 142 12.7.2 Setting NIBP Alarm Properties
- 142 12.7.3 Displaying NIBP List
- 142 12.7.4 Setting the Pressure Unit
- 142 12.7.5 Switching On NIBP End Tone
- 142 12.8 Assisting Venous Puncture
- 143 13 Monitoring Temp
- 143 13.1 Introduction
- 143 13.2 Safety
- 143 13.3 Taking a Temp Measurement
- 143 13.4 Understanding the Temp Display
- 143 13.5 Setting the Temperature Unit
- 144 14 Monitoring IBP
- 144 14.1 Introduction
- 144 14.2 Safety
- 144 14.3 Zeroing the Transducer
- 144 14.4 Setting Up the Pressure Measurement
- 144 14.5 Understanding the IBP Display
- 144 14.6 Changing IBP Settings
- 144 14.6.1 Changing a Pressure for Monitoring
- 144 14.6.2 Setting Alarm Properties
- 144 14.6.3 Changing Averaging Time
- 144 14.6.4 Setting Up the IBP Wave
- 165 14.6.5 Setting the Pressure Unit
- 165 14.6.6 Enabling PPV Measurement and Setting PPV Source
- 165 14.7 Overlapping IBP Waveforms
- 165 14.8 Measuring PAWP (Only Available with the External Display)
- 165 14.8.1 Preparing to Measure PAWP
- 165 14.8.2 Setting Up the PAWP Measurement
- 165 14.8.3 Understanding the PAWP Setup Menu
- 165 14.9 Troubleshooting
- 166 15.1 Introduction
- 166 Modules
- 166 Measurement
- 166 Settings
- 166 Menus
- 166 Unit
- 166 15.4.3 Setting Up Gas Compensations
- 166 15.4.4 Setting Up Humidity Compensation
- 166 15.4.5 Setting the Apnea Alarm Delay
- 166 15.4.6 Choosing a Time Interval for Peak-Picking
- 166 15.4.7 Setting the Flow Rate
- 166 15.4.8 Setting Up the CO2 Waveform
- 166 15.4.9 Setting RR Source
- 166 15.4.10 Barometric Pressure Compensation
- 166 15.4.11 Entering the Standby Mode
- 166 15.5 Measurement Limitations
- 166 15.6 Leakage test
- 166 15.7 Troubleshooting the Sidestream CO2 Sampling System
- 166 15.8 Removing Exhaust Gases from the System
- 166 15.9 Zeroing the Sensor
- 166 15.10 Calibrating the Sensor
- 166 15.11 Oridion Information
- 167 16 Review
- 167 16.1 Accessing Respective Review Windows
- 167 16.2 Reviewing Graphic Trends
- 167 16.3 Reviewing Tabular Trends
- 167 16.4 Reviewing Events
- 167 16.4.1 Marking Events
- 167 16.4.2 Reviewing Events
- 167 16.5 Reviewing Waveforms
- 167 16.6 Reviewing OxyCRG (Only Available with the External Display)
- 168 17 Calculations (Only Available with the External Display)
- 185 17.1 Introduction
- 185 17.2 Dose Calculations
- 185 17.2.1 Performing Calculations
- 185 17.2.2 Selecting the Proper Drug Unit
- 185 17.2.3 Titration Table
- 185 17.2.4 Drug Calculation Formulas
- 185 17.3 Oxygenation Calculations
- 185 17.3.1 Performing Calculations
- 185 17.3.2 Entered Parameters
- 185 17.3.3 Calculated Parameters and Formulas
- 185 17.4 Ventilation Calculations
- 185 17.4.1 Performing Calculations
- 185 17.4.2 Entered Parameters
- 185 17.4.3 Calculated Parameters and Formulas
- 185 17.5 Hemodynamic Calculations
- 185 17.5.1 Performing Calculations
- 185 17.5.2 Entered Parameters
- 185 17.5.3 Calculated Parameters and Formulas
- 185 17.6 Renal Calculations
- 185 17.6.1 Performing Calculations
- 185 17.6.2 Entered Parameters
- 185 17.6.3 Calculated Parameters and Formulas
- 185 17.7 Understanding the Review Window
- 186 18 Printing
- 186 18.1 Printer
- 186 18.2 Connecting a printer
- 186 18.3 Setting the Printer
- 186 18.4 Starting Report Printouts
- 186 18.5 Stopping Report Printouts
- 186 18.6 Setting Up Reports
- 186 18.6.1 Setting Up ECG Reports
- 186 18.6.2 Setting Up Tabular Trends Reports
- 186 18.6.3 Setting Up Graphic Trends Reports
- 186 18.6.4 Setting Up Realtime Reports
- 186 18.7 End Case Reports
- 186 18.8 Printer Statuses
- 186 18.8.1 Printer Out of Paper
- 186 18.8.2 Printer Unavailable
- 187 19 Other Functions
- 187 19.1 Analog Output
- 187 19.2 Exporting the Log
- 187 19.3 Using the Storage Card
- 187 19.4 Transferring Data
- 187 19.4.1 Data Export System
- 192 19.4.2 Transferring Data by Different Means
- 192 19.5 Network Setup
- 192 19.5.1 Setting the Network Type
- 192 19.5.2 Wireless Network Connection
- 192 19.5.3 Certificates Maintenance
- 192 19.5.4 Setting the Central Stations
- 192 19.5.5 Setting the Network Service Quality Level
- 192 19.5.6 Setting the Multicast Parameters
- 193 20 Battery
- 193 20.1 Overview
- 193 20.2 Safety
- 193 20.3 Battery Guidelines
- 193 20.4 Installing or Replacing a Battery
- 193 20.5 Charging a Battery
- 193 20.6 Conditioning a Battery
- 193 20.7 Checking Battery Performance
- 193 20.8 Storing a Battery
- 193 20.9 Recycling Batteries
- 194 21 Care and Cleaning
- 194 21.1 General Points
- 194 21.2 Cleaning
- 194 21.3 Disinfecting
- 194 21.4 Sterilization
- 195 22 User Maintenance
- 195 22.1 General Inspection
- 195 22.2 Maintenance and Testing Schedule
- 195 22.3 Checking Monitor and Module Information
- 195 22.4 ECG Verification
- 195 22.5 NIBP Leakage Test
- 195 Leakage Test
- 195 22.7 Calibrating the Touchscreen
- 196 23 Accessories
- 196 23.1 ECG Accessories
- 196 Accessories
- 196 23.3 NIBP Accessories
- 196 23.4 Temp Accessories
- 196 23.5 IBP Accessories
- 196 Accessories
- 196 23.7 Mount and Mounting Accessories
- 196 23.8 Miscellaneous Accessories