Replacing valve hoses. Endres+Hauser Stamolys CA71SI

Add to My manuals
64 Pages

advertisement

Replacing valve hoses. Endres+Hauser Stamolys CA71SI | Manualzz

Maintenance Stamolys CA71SI

R3

R2

R1 a0002206

Fig. 19:

R1

R2

R3

Reagent pump, top view

Reagent 1

Reagent 2 (if present)

Reagent 3 (if present)

5.

Reconnect the reagent hoses to the canisters.

6.

After installation, refill the hoses with sample, standard or cleaner (SERVICE)

2)

.

7.

Carry out an offset calibration and a calibration.

7.4

Replacing valve hoses

To replace the hoses, proceed as follows:

1.

Rinse the old hoses first with water and then with air to empty them (see SERVICE).

2.

Remove the hoses from the valves: a. You can disconnect the front hoses directly because the valves are open when deenergised.

b. To remove the back hoses, press the black button on the valve and disconnect the hoses.

3.

Grease the new hoses with silicone grease before inserting them.

4.

Install the new valve hoses in the reverse sequence of operations.

Ensure that the hoses are connected correctly.

5.

After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).

6.

Carry out an offset calibration and a calibration.

2) Make sure that the hoses are bubble-free when filled. If not, check the sample flow rate (see below).

38 Endress+Hauser

advertisement

Key Features

  • - Measures silicate concentration in ultrapure water
  • - Photometric measurement principle
  • - Measuring range: 0.01 - 1000 µg/l SiO2
  • - Compact and wall-mountable design
  • - Easy-to-use interface with clear display
  • - Automatic calibration and cleaning functions
  • - Low maintenance and operating costs

Related manuals

Frequently Answers and Questions

What is the measuring range of the analyzer?
The measuring range is 0.01 - 1000 µg/l SiO2.
How often does the analyzer need to be calibrated?
The analyzer has automatic calibration functions, but the frequency of calibration may vary depending on the application.
What is the power supply requirement of the analyzer?
The power supply requirement is specified in the technical data section of the manual.
Can the analyzer be used in hazardous areas?
The manual does not provide information on the suitability of the analyzer for hazardous areas.