- Industrial & lab equipment
- Measuring, testing & control
- Endres+Hauser
- Stamolys CA71SI
- Operating instructions
Endres+Hauser Stamolys CA71SI Operating Instruction
Add to my manuals
64 Pages
Endres+Hauser Stamolys CA71SI is a robust analyzer designed for continuous monitoring of silicate in ultrapure water. It features a user-friendly interface, low maintenance requirements, and reliable measurements. The analyzer is suitable for various applications, including power plants, semiconductor manufacturing, and pharmaceutical industries.
advertisement
Operating Instructions
Stamolys CA71SI
Analyzer for the photometric determination of silicate
BA364C/07/EN/14.11
71133178
Valid of:
Software version 5.9
Brief overview
5
6
7
3
4
1
2
Reagent pump (P2), supply from canister
Display
Serial interface RS 232
Photometer cell
Static mixer (depending on version)
Valve V4 (only SI-A)
Reaction coil
8
9
10
11
12
13
Outlet for sample or reagent mixture (SI-A)
Valve V2
Outlet for sample or reagent mixture (SI-B)
Valve V1
Valves V3: top ch. 1, bottom ch. 2
1)
Sample pump P1, supply see below
1) Only for two-channel version
Inlet to the sample pump:
• Valve V1
– hose in front: sample inlet
– hose at the back: inlet from valve V2 (cleaner or standard solution)
• Valve V2
– hose in front: inlet from canister with standard solution
– hose at the back: inlet from canister with cleaner
a0001352
Endress+Hauser
Stamolys CA71SI
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Installation, commissioning and operation . . . . . . . . 4
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Safety messages and their meaning . . . . . . . . . . . . . 5
2 Identification . . . . . . . . . . . . . . . . . . . . 6
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8
3 Installation . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . 12
3.4 Installation examples . . . . . . . . . . . . . . . . . . . . . . . 14
3.5 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 14
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3 Switching contacts . . . . . . . . . . . . . . . . . . . . . . . . 19
4.4 Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 21
5 Operation . . . . . . . . . . . . . . . . . . . . . . 22
5.1 Display and operating elements . . . . . . . . . . . . . . . 22
5.2 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6 Commissioning. . . . . . . . . . . . . . . . . . 31
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2 Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7 Maintenance. . . . . . . . . . . . . . . . . . . . 35
7.1 Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . 35
7.2 Replacing reagents . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.3 Replacing pump hoses . . . . . . . . . . . . . . . . . . . . . . 37
7.4 Replacing valve hoses . . . . . . . . . . . . . . . . . . . . . . 38
7.5 Replacing the static mixer . . . . . . . . . . . . . . . . . . . 39
7.6 Replacing the photometer optical cell . . . . . . . . . . . 39
7.7 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.8 Placing out of service . . . . . . . . . . . . . . . . . . . . . . . 41
8 Accessories. . . . . . . . . . . . . . . . . . . . . 42
8.1 Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.2 Reagents, cleaner, standard solutions . . . . . . . . . . . 42
8.3 Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.4 Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.5 Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.6 Additional accessories . . . . . . . . . . . . . . . . . . . . . . 43
9 Trouble-shooting . . . . . . . . . . . . . . . . . 44
9.1 Trouble-shooting instructions . . . . . . . . . . . . . . . . . 44
9.2 System error messages . . . . . . . . . . . . . . . . . . . . . . 44
9.3 Process errors without messages . . . . . . . . . . . . . . 45
9.4 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
9.5 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
9.6 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.7 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
10 Technical data . . . . . . . . . . . . . . . . . . . 50
10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
10.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
10.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
10.4 Performance characteristics . . . . . . . . . . . . . . . . . . 51
10.5 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 52
10.6 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.7 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.8 Mechanical construction . . . . . . . . . . . . . . . . . . . . 52
11 Appendix . . . . . . . . . . . . . . . . . . . . . . . 54
11.1 Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . . 54
11.2 Analyzer settings . . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.3 Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . 59
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Endress+Hauser 3
4
Safety instructions Stamolys CA71SI
1 Safety instructions
1.1
Designated use
The analyzer is a compact photometric analysis system.
It is designed for the monitoring of the silicate content in ultrapure water and boiler feed water.
In particular, CA71 is designated for:
• boiler feed water cycles
• ultrapure water
• steam and condensate analysis
• reversed osmosis
• demineralizers.
Any other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2
Installation, commissioning and operation
Please note the following items:
• Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions and must adhere to them.
• Before commissioning the entire measuring point, check all the connections. Ensure that electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning.
Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorized and specially trained personnel.
• If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the manufacturer’s or by the service organization.
1.3
Operational safety
The analyzer has been designed and tested to the highest standards and left the factory in perfect functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
1.4
Return
Before returning observe the following items:
• Contact your regional Endress+Hauser sales center to receive detailed shipping conditions.
• Enclose the duly completed "Declaration of de-contamination" with the packaging.
You can find the form as master at the end of these instructions.
Endress+Hauser
Stamolys CA71SI Safety instructions
1.5
Safety messages and their meaning
The structure, signal words and safety colors of the signs comply with the specifications of
ANSI Z535.6 ("Product safety information in product manuals, instructions and other collateral materials").
Safety message structure Meaning
!
DANGER
Cause (/consequences)
Consequences if safety message is not heeded
► Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid the situation will result in a fatal or serious injury.
!
WARNING
Cause (/consequences)
Consequences if safety message is not heeded
► Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid the situation can result in a fatal or serious injury.
!
CAUTION
Cause (/consequences)
Consequences if safety message is not heeded
► Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or medium injury.
NOTICE
Cause/situation
Consequences if safety message is not heeded
► Action/note
This symbol alerts you to situations that can result in damage to property and equipment.
Endress+Hauser 5
6
Identification Stamolys CA71SI
2 Identification
2.1
Device designation
2.1.1
Nameplate
Compare the (extended) order code on the nameplate (on the analyzer) to the product structure and your order.
You can read the following information from the nameplate:
• Order code
• Extended order code (=order code from the product structure)
• Serial number
• Measuring range
• Outputs and communication
• Power supply
• Ingress protection
• Environment conditions
2.1.2
Product structure
CA71SI-
A
B
Y
Measuring range
1.0 to 200 μg/l SiO
2
50 to 5000 μg/l SiO
2
Special version acc. to customer’s specification
Sample transfer
1
2
From one measuring point (one channel version)
From two measuring points (two channel version)
Power supply
0
1
2
3
230 V AC / 50 Hz
115 V AC / 60 Hz
115 V AC / 50 Hz
230 V AC / 60 Hz
Collecting vessel (for up to 3 analyzers)
A
B
C
D
Not selected (without collecting vessel)
Collecting vessel without level measurement
Collecting vessel with level measurement (one-channel version only)
Two collecting vessels without level measurement (two-channel version)
Housing
1
2
3
Open design
GFR housing
Stainless steel (1.4301 / AISI 304) + purge
Output
A 0/4 to 20 mA, RS 232
Reagents
1
2
3
To order separately
One set, inactive
Three sets, inactive
complete order code
Endress+Hauser
Stamolys CA71SI Identification
2.2
Scope of delivery
Order reagents separately with analyzer version CA71XX-XXXXXX1.
With all other versions, inactive reagents are included in the scope of delivery. You have to mix the reagents before using them. Please, read the instructions attached to the reagents.
2.2.1
Version CA71SI-B
The scope of delivery comprises:
• an analyzer with mains plug
• a cleaning injector
• a tube of silicone grease
• a Norprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)
• two hose fittings of each size:
– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)
– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)
– 6.4 mm x 3.2 mm (0.25 inch x 0.13 inch)
• two T-hose fittings of each size:
– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)
– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)
– 6.4 mm x 6.4 mm x 6.4 mm (0.25 inch x 0.25 inch x 0.25 inch)
• an interference suppressor for the current output
• 4 edge covers (version with GFR housing only)
• PTFE strip
• a quality certificate
• Operating Instructions (English).
2.2.2
Version CA71SI-A
The scope of delivery comprises:
• an analyzer with mains plug
• a cleaning injector
• a tube of silicone grease
• a Norprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)
• a Grifflex hose, length 2.0 m (6.6 ft), ID 19 mm (0.75 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)
• two hose fittings of each size:
– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)
– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)
– 6.4 mm x 6.4 mm (0.25 inch x 0.25 inch, version without housing only)
• two T-hose fittings of each size:
– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)
– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)
• an interference suppressor for the current output
• a hose clamp
• 2 pipe clamps (version without housing only)
• a screwed socket for the outlet pipe
• 4 edge covers
• PTFE strip
• a quality certificate
• Operating Instructions (English).
Endress+Hauser 7
Identification Stamolys CA71SI
2.3
Certificates and approvals
2.3.1
4 approval
Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the
4 symbol.
2.3.2
Manufacturer certificate
Quality certificate
With the certificate the manufacturer confirms compliance with all technical regulations and the successful testing individually for your product.
8 Endress+Hauser
Stamolys CA71SI Installation
3 Installation
3.1
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the contents. Keep the damaged products until the matter has been settled.
• Check that the order is complete and agrees with your shipping documents.
• The packaging material used to store or to transport the product must provide shock protection and humidity protection. The original packaging offers the best protection. Also, keep to the approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your local sales center.
3.2
Installation conditions
3.2.1
Design, dimensions
Version with stainless steel housing
Endress+Hauser
Fig. 2: Stainless steel version
a0001361
9
Installation
10
Stamolys CA71SI
Version with GFR housing, CA71SI-B
163 (6.42)
330 (13.0)
7-11/
0.28 - 0.43
(M 6)
200 (7.87)
326 (12.8)
360 (14.2)
436 (17.2)
465 (18.3)
500 (19.7)
Version with GFR housing, CA71SI-A
38 (1.5)
60 (2.36)
250 (9.84)
160 (6.30)
140 (5.51)
100(3.94)
Ø 26 (1.02)
Ø 16 (0.63)
120(4.72)
Ø 27 (1.06) mm(inch) a0001354-en
60 (2.36)
250 (9.84)
160 (6.30)
140 (5.51)
100(3.94)
330 (13.0)
7-11/
0.28 - 0.43
(M 6)
200 (7.87)
326 (12.8)
360 (14.2)
436 (17.2)
465 (18.3)
500 (19.7)
Ø 26(1.02)
Ø 16 (0.63)
120(4.72)
Ø 27 (1.06) mm (inch) a0001355
Endress+Hauser
Stamolys CA71SI
Open version
381 (15.0)
351 (13.8)
19 (0.75)
197 (7.76)
Installation
325 (12.8)
9 (0.35)
Fig. 5: Open version (without housing)
8 (0.31) mm (inch) a0001356
With the open version, you need an additional platform for the reagents. Mount this platform max.
35 cm (13.8 inch) below the pumps. The reagent bottels have the following dimensions: 90 x 90 x
215 mm (3.54 x 5.54 x 8.46 inch). The number of bottles varies from 2 to 5 depending on the analyzer version.
For these versions, the outlet pipe must be installed right of the analyzer.
The outlet pipe must be mounted to a wall so that the sample outlet hoses from the photometer have a gradient of 5 to 10 %. If neccessary, extend the hoses.
Endress+Hauser 11
Installation
12
Stamolys CA71SI
3.2.2
Connecting the sample line
27 (1.06)
145 (5.71)
60 (2.36)
4 x Ø 6.6 (0.26)
10 (0.39)
115 (4.53)
3 x Ø 28/20 (1.1/0.79)
Ø 34/25 (1.34/0.98)
4 x Ø 5.5/10.4 (0.22/0.41)
Abb. 6:
3
4
1
2
5
Collecting vessel at analyzer (optional)
Ventilation
Sample inlet from sampling
Collecting vessel
Electrical connections
Analyzer sample inlet
Ø 56 (2.2)
Ø 50H7 (1.97H0.28)
*
* a0001363
M20x1.5
Ø 24 (0.94) mm (inch)
*
Abb. 7:
7
8
*
6
Collecting vessel dimensions variable, freely adjustable dimensions
Sampling for analyzer
Sample overflow
Analyzer outlet
70 (2.76) a0001346
One-channel version
Collecting vessel (at analyzer, with or without level measurement)
Connection hose ID 3.2 mm (1/8 inch)
Customer collecting vessel
Connection
Max. distance from collecting vessel to analyzer
Max. height difference from collecting vessel to analyzer hose ID 1.6 mm (1/16 inch)
1 m (3.3 ft)
0.5 m (1.6 ft)
Two-channel version
• Depending on the ordered version, one or two collecting vessels (with or without level measurement) are included in the scope of delivery.
• Only one collecting vessel can be mounted at the housing. The second is to be placed nearby the analyzer.
3.3
Installation instructions
To install the analyzer at the intended location, proceed as follows:
1.
Mount the analyzer and secure it to a vibration-free wall using screws (Ø6 mm / 0.24"). The fastening clips on the housing make sure that the necessary distance to the wall is observed for ventilation purposes.
For the installation dimensions, please see the previous chapter.
2.
Using a spirit level, check that the cabinet is hanging straight. Only in this way can any air bubbles present escape from the cell.
3.
Fit the edge covers (with GFR housing only).
Endress+Hauser
Stamolys CA71SI Installation
4.
Lay the drain pipe for the reaction products. Where possible, use solid pipes (PE, internal diameter 1" with 3% incline).
5.
Screw the screw-in connector ID 16 into the outlet pipe from below. Secure the Grifflex hose
ID 19 to the nozzle with a hose clip (only CA71SI-A).
6.
Insert the valve hoses accordingly. These were partly removed from the valves for transportation. This prevents the hoses from sticking or pressure being applied to a point in the hose over an extended period.
1 2 3
V3 V1 V2
Fig. 8:
1
7
Valves and valve hoses CA71SI-B
V3 V1
6
2
V2
5
3
V4
a0006362
7 6 5 4
a0001730
1
2
Fig. 9:
V1-4
3
Valves and valve hoses CA71SI-A
Valves
Channel switchover
To sample pump
Connecting hose to valve 1, at rear
6
7
4
5
Outlet hose
Hose, valve 2 at front, standard
Hose, valve 2 at rear, cleaner
Hose, valve 1 at front, sample
7.
Secure the hose boxes in the appropriate pump holders:
Sample pump, left, reagent pump, right. The direction of flow of the sample and reagent must be counterclockwise.
R3
R2
R1
Fig. 10:
R1
R2
R3
Reagent pump, top view
Reagent 1
Reagent 2 (if present)
Reagent 3 (if present)
8.
Connect the sample supply.
a0002206
Endress+Hauser 13
Installation Stamolys CA71SI
The sample can be obtained as follows:
• Directly or via a reversible flow filter or a cross-flow filter by means of a small pump (rating approx. 300 ml/min), suitable for clear media
For questions regarding sample conditioning and its automation, please contact Service or the sales center responsible for your region.
9.
Connect the tubes from canisters containing reagents, standards and cleaner to the following nozzles:
Canister
Sample
Reagent 1
Reagent 2
Reagent 3
Standard solution
Cleaner
Tube designation (mark)
S
R
P
SI-1
SI-2
SI-3
3.4
Installation examples
• Permeate can contain air bubbles (CAT430) or is free of bubbles (customer-supplied ultrafiltration)
• Distance between the analysers as short as possible: sampling line between T-piece and the second analyser (
shorter than 1.5 m
• Cross-section of sampling line ID 3.2 - 4 mm
• Only one sample receiver required
NOTICE
Measuring errors due to missing sample
► Ensure that there is always sufficient sample available for both analysers.
► Observe this when selecting maintenance intervals for CAT430 and when setting the buffer volume on the collecting vessel.
3
4
1
2
5
1
CA 71
2
Sample from CAT430
Sampling line
Sample receiver overflow
T-piece
Collecting vessel
3
5
4
3.5
Post-installation check
• After installation, check that all connections are fitted tightly and are leakage resistant.
• Ensure that the hoses cannot be removed without effort.
• Check all hoses for damage.
CA 71 a0001728
14 Endress+Hauser
Stamolys CA71SI Wiring
4 Wiring
!
WARNING
Device is energized
Improper connection can cause injury or death.
► The electrical connection must only be carried out by a certified electrician.
► Technical personnel must have read and understood the instructions in this manual and must adhere to them.
► Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
4.1
Electrical connection
4.1.1
Quick wiring guide
NOTICE
Reaching the terminal block and avoiding damage
► You have to fold out the frame of the analyzer to reach the terminal block.
► Beforehand, release the hoses from the outlet pipe. Otherwise there is a danger of flooding.
► Connect the hoses again after folding the frame back in.
Fold out the frame as follows:
1.
Release the two bottom Allen screws (AF 6) by 3-4 rotations (
2.
Unscrew the two upper Allen screws completely so that the frame folds out. In this way, you reach the terminal block (item 2).
2
M E K RS 232
1 a0001370
Fig. 12:
1
2
Folding out the frame
Allen screws AF 6
Terminal strip
NOTICE
The device does not have a mains switch
► Install the device near to a fused socket which is easy to reach.
Endress+Hauser 15
Wiring Stamolys CA71SI
4.1.2
Terminal assignment
!
CAUTION
Shown diagram (
The terminal assignment and cable colors can deviate from the actual assignment and colors!
► Only use the terminal assignment of the sticker in the device (
analyzer!
16
2
1
Fig. 14:
1
2
3
Analyzer from top (open version resp. folded out)
Connection department sticker
Printed circuit board with terminal strip
Backside of the analyzer
Schirm
Screen
Schirm
Shield n.b.
n.c.
n.b.
n.c.
Schirm
Shield n.b.
n.c.
WH keine Probe
No sample
Mess 2
Meas 2
Analog Out
0/4-20mA
Kanal 1
Channel 1
COM
NC
Kanal 2
Channel 2 n.b.
n.c.
n.b.
n.c.
NO
COM
NC
NO
7
6
5
N max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC
GN/YE 4
AC Motor
T 0.1A
Elektronik
Electronics
T 0.5A
3
2
1
L
N
3 a0001369 a0001372
Endress+Hauser
Stamolys CA71SI
Endress+Hauser
Wiring
Function Designation Terminal one channel
Mains
Alarm value 1, channel 1
Alarm value 2, channel 1
Alarm value 1, channel 2
Alarm value 2, channel 2
Fault
Reserve (unassigned terminals)
Analog output 1
0/4 ... 20 mA
Analog output 2
0/4 ... 20 mA
Sample conditioning remote control
Channel switch-over
1)
+
–
Screen
Input
0 V
Input
0 V
NC
NO
COM
NC
NO
+
–
Screen
NO
COM
NC
NO
COM
NC
NO
COM
L
N
PE
COM
NC
NO
COM
NC
Brass screw with bolt top right in the connection compartment (marked with
*
)
53
–
–
–
57
–
–
20
36
35
PE
1
18
17
22
21
–
19
–
–
–
–
26
–
24
23
28
27
1
25
3
2
38
PE
1)
57
53
55
53
22
21
20
36
35
PE
1
39
14
19
18
17
12
11
16
15
28
27
26
13
Terminal two channels
3
2
1
25
24
23
NOTICE
Terms for special terminals
► Limit switches 1 and 2 do not need to be connected if the PLC sets its own alarm values at the analog output.
► When using a sample conditioning system: Connect terminals 57 and 53 on the analyzer to the corresponding terminals on the sample conditioning system. For the allocation of these terminals see the sample conditioning system operating instructions.
► If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement (sample not ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.
► All external relay contacts have to be connected to 230 V.
17
Wiring Stamolys CA71SI
4.2
Signal connection
4.2.1
Screening of the analog outputs
The interference suppressor attenuates electromagnetic effects on control, power and signal lines.
After the connection of the data transfer cables clip the interference suppressor (in scope of delivery) on the cable cores (not on the outer insulation of the cable!). Place the cable screen out of the interference suppressor and connect it to PE (brass screw with bolt, top right in the connection
4 3 a0001746
Fig. 15:
3
4
1
2
Interference protection of the signal cable
Cable screen (to PE )
Signal cable
Interference suppressor
Cable cores of the signal cable
With the two-channel version, place the cable cores of all cables (data cables to analog output 1 and to analog output 2) through the interference suppressor.
4.2.2
One-channel version
Connection
Signal inputs
Signal outputs
Designation
Leak
No sample
AV 1
AV 2
Fault
Measurement end l-1 channel 1
Analog output
Function
Liquid has collected in the drip pan
No sample available, measurement is not started, display flashes
Alarm value 1 exceeded or undershot
Alarm value 2 exceeded or undershot
Retrieves error message using operation menu
Displays "Measurement finished" (5 s)
0 or 4 mA = measuring range start
20 mA = measuring range end
18 Endress+Hauser
Stamolys CA71SI
4.2.3
Two-channel version
Connection
Signal inputs
Signal outputs
Analog output
Channel selection
Designation
Leak
No sample
AV 1-1
AV 2-1
AV 1-2
AV 2-2
Function
Liquid has collected in the drip pan
No sample available, measurement is not started, display flashes
Alarm value 1, channel 1 exceeded or undershot
Alarm value 2, channel 1 exceeded or undershot
Alarm value 1, channel 2 exceeded or undershot
Alarm value 2, channel 2 exceeded or undershot
Fault Retrieves error message using operation menu
Channel 1/2 or measurement end
1)
Displays active channel
Displays "Measurement finished" (5 s) l-1 channel 1 l-2 channel 2
0 or 4 mA = measuring range start
20 mA = measuring range end
0 or 4 mA = measuring range start
20 mA = measuring range end
Meas. 2 0 V = channel 1
24 V = channel 2
Wiring
4.3
Switching contacts
One-channel version
Connection
AV 1
AV 2
Fault
Unassigned
A:
R:
A:
R:
A:
R:
Terminal connection for condition fulfilled
25
25
-
-
23
24
28
28
19
19
-
-
-
-
26
27
17
18
22
16
13
-
-
-
20
14
11
A = NO current configured
R = NC current configured
A:
R:
A:
R:
A:
R:
Terminal connection for condition not fulfilled
25
25
-
-
24
23
28
28
19
19
-
-
-
-
27
26
18
17
22
16
13
-
-
-
21
15
12
Terminal connection for power off
25 24
28
19
22
16
13
-
-
-
-
-
27
18
21
15
12
Endress+Hauser 19
Wiring
20
Stamolys CA71SI
Two-channel version
Connection
AV 1 - 1
AV 1 - 2
AV 2 - 1
AV 2 - 2
Fault
Channel 1/2 measurement end
A:
R:
A = NO current configured
R = NC current configured
A:
R:
A:
R:
A:
R:
A:
R:
A:
R:
Terminal connection for condition fulfilled
25
25
-
-
23
24
13
13
28
28
-
-
-
-
11
12
26
27
16
16
19
19
22
22
-
-
-
-
-
-
14
15
17
18
20
21
A:
R:
A:
R:
A:
R:
A:
R:
A:
R:
A:
R:
Terminal connection for condition not fulfilled
25
25
-
-
24
23
13
13
28
28
-
-
-
-
12
11
27
26
16
16
19
19
22
22
-
-
-
-
-
-
15
14
18
17
21
20
Terminal connection for power off
25 24
13
28
16
19
22
-
-
-
-
-
12
27
15
18
21
Condition fulfilled means:
• AV 1: concentration > Alarm value 1
• AV 2: concentration > Alarm value 2
• Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
4.4
Serial interface
RS 232 of CA71
SUB-D, nine-pin Function
3
2
8
TxD
RxD
CTS
5
Software protocol:
Output format:
Function
RxD
TxD
COM 1/2 at PC
RTS
CTS
GND GND
9600 baud, 8 data bits, 1 stop bit, no parity (9600, N, 8, 1)
ASCII
SUB-D, nine-pin
2
3
7
8
5
The results (measured value+unit of measure) are output in the "Data memory Measured values" menu.
The calibration results (measured value+unit of measure) are output in the "Data memory-
Calibration factors" menu.
A null modem cable is required (not a crossed one).
The analyzer does not have to be configured for the interface.
The following commands can be sent from the PC, in order to read out data:
• "D" = Data memory-Measured values
• "C" = Data memory-Calibration factors
• "S" = Setup (parameter entry, configuration...)
• "F" = Frequency (current)
Endress+Hauser
Stamolys CA71SI Wiring
4.5
Post-connection check
Carry out the following checks after electrical connection:
Device status and specifications
Is the analyser or cable externally undamaged?
Note
Visual inspection
Electrical connection
Does the supply voltage correspond to the data on the nameplate?
Note
230 V AC / 50 Hz
115 V AC / 60 Hz
Are current outputs screened and connected?
Are the mounted cables relieved of tension?
Cable type properly disconnected?
Guide power supply and signal lines separately over the entire travel distance.
Separate cable channels are ideal.
Cable routing without scuffing or cross-overs?
Are power supply and signal lines connected correctly according to wiring diagram?
Are all screw terminals tightened?
Are all cable entries mounted, tightened and leak-resistant?
Interference suppressor at the analogue output?
Current output simulation See procedure below
Current output simulation:
1.
Hold both arrow keys down (see "Display and operating elements" chapter) and connect the analyser to the mains or switch the mains switch on (if available). Wait until the display "0 mA" appears.
2.
Check on your PLC, PCS or you data logger whether the current value is the same.
3.
Press the
F
key. Browse to the next current values (4, 12, 20 mA, depending on the setting).
4.
Check that the respective current values are also on your PLC, PCS or your data logger.
5.
If the values are not there: a. Check the terminal assignment for analogue output 1 or 2.
b. Disconnect the analogue outputs from your PLC, PLS or your data logger and repeat steps
14 with the restriction that you measure the current values at the terminals of the analyser and not at the PLC, SPL or data logger.
If these current values are correct, please check your PLC, SPL, the data logger or the electrical cables.
Endress+Hauser 21
Operation Stamolys CA71SI
5
5.1
Operation
Display and operating elements
1 2
RS 232
M CE E K
4
Fig. 16: Display operating elements
3
4
1
2
LED (measured value)
LC display (measured value and status)
Serial interface RS 232
Operating keys and control LEDs
3
5.2
Local operation
The operating keys and the integrated indicator LEDs have the following functions:
Key Key function
H
– “Auto measuring” option
– back to the main menu from all sub-menus
K
– backwards in the sub-menu
(horizontal, see Appendix)
V
– backwards in the main menu (vertical)
– increase value
W
– forwards in the main menu (vertical)
– reduce value
F
– select option
– adopt value, forwards in the sub-menu (horizontal)
1)
J
– selection in the sub-menu
Indicator LED function
Alarm value 1 exceeded
Alarm value 2 exceeded
Measuring range exceeded
Measuring range undershot
Retrieve error message
Unassigned
V
or
W
and the
F
key simultaneously, you can set the digit after the decimal point.
a0001679
22 Endress+Hauser
Stamolys CA71SI Operation
5.2.1
Main menu
Access the main menu by holding down the
H
key until “AUTO MEASURING” is displayed.
For the main menu options and information about them, please see the following table.
Selection Display
AUTO MEASURING
Info
Calibration, measurement, flushing time-controlled actions
PARAMETER ENTRY
CONFIGURATION
LANGUAGE
Default settings for measuring ranges, alarm values, calibration, rinsing
Basic settings such as parameters, measuring units, arrangement of analog outputs and alarm values, date, time, offset values
Selecting menu language
ERROR DISPLAY
SERVICE
DATA MEMORY 1
DATA MEMORY 2
(Two-channel version only)
Displaying error messages
Manually switching valves and pumps
Last 1024 measured values channel 1
Last 1024 measured values channel 2
5.2.2
AUTO MEASURING
The actions "calibration", "measuring" and "flushing" are triggered by time-control.
The respective action is displayed in the LC display. The most recently registered concentration value is displayed until the end of the next measurement.
Otherwise, "wait" is displayed when
• the time of the first measurement has not yet been reached or
• the measuring interval has not yet expired.
"Measuring" flashes when the analyzer is ready for the next measurement but has not yet received the enable signal from the sample collector or the sample conditioning unit.
Endress+Hauser 23
Operation
Menu item
Code number
Photometer
Default settings
Measuring unit
Calibration factor
Concentration offset
Dilution
Delay to sample
Stamolys CA71SI
Depending on specification
SI-A
SI-B
Yes / no
μg/l / ppb
20 to 300 s
SI-A: 120 s
SI-B: 160 s
5.2.3
CONFIGURATION
NOTICE
Some settings that can be made in this menu affect the defaults in the PARAMETER
ENTRY menu.
► Complete the CONFIGURATION menu first during initial start-up.
Display Range of adjustment
(factory settings in bold)
03
0.10 to 10.00
0.00 to 50.0 mg/l
0.10 to 10.00
1.00
Info
Input 03.
If an incorrect code is entered the program exits the submenu.
The setting displays the parameter that is being measured.
This is defined by the product specification and set in this menu item at the factory. Do not change the value.
Otherwise, you will receive an "Incorrect photometer" error message.
If "yes" is selected, all the settings are reset to the factory settings.
You must set the current date and time (scroll through the menu, third-last point).
In the event of a reset, the date for the 1st calibration and the 1st flush is set to the following day.
The measuring unit selected depends on the type of photometer. This setting also affects the scope of the measuring range.
The calibration factor is the ratio of the measured concentration of the calibration standard to the predefined concentration of the standard (see
"PARAMETER ENTRY", calibration solution). The deviation results from factors such as reagent aging, aging of constructive components, etc.
The calibration factor compensates for these effects. CA
71 checks the registered calibration factor logically. If the factor is outside of the error tolerance, the calibration is automatically repeated. If the value is still outside the tolerance range after repeating calibration, an error message appears and the analyzer continues to work with the most recently registered, logically correct factor.
The last 100 calibration factors are stored in the memory with the date and time and can be retrieved by pressing the
J
key.
The calibration factor can be changed manually.
Optional dilution modules are taken into consideration in the calibration factor.
The offset gives users the option of adapting to an external comparison measurement.
(Change the sign with the
J
key.)
If the sample is diluted externally between taking the sample and the analyzer, the dilution factor has to be entered here.
If an optional dilution module is used, the factor remains
1. The dilution is taken into consideration by means of the calibration factor.
Dosing time for sample or standard. During this time, the entire system is flushed with sample or standard, so when the reagent is added there is definitely only fresh sample in the mixer. If sufficient sample is available, or in the event of an external collecting vessel with a long supply line, select the highest possible value.
24 Endress+Hauser
Stamolys CA71SI Operation
Menu item
Analog output 1
Analog output 2
Alarm value AV 1-1
Alarm value AV 2-1
Alarm value AV 1-2
Alarm value AV 2-2
Error contact
Current date/time
Calibrate offset
Range of adjustment
(factory settings in bold)
0 to 20 mA / 4 to 20 mA
NO
NC
Changes only take effect after a reset (power off/on)!
01.01.96 00:00 to 31.12.95 23:59
Yes / no
Display Info
Selection for the measuring range scope for channel 1. If the concentration measuring range is 0 to x mg/l, either
4 mA or 0 mA corresponds to 0 mg/l. The end of the measuring range is the same in both cases at 20 mA.
Only two-channel version!Selection for the measuring range scope for channel 2. The scopes of measuring range are independent of each other for channel 1 and channel
2 and are determined by the start of measuring range setting (channel 1 / channel 2) or end of measuring range setting (channel 1 / channel 2) in the PARAMETER
ENTRY menu.
Setting as to whether the contact for alarm value 1, channel 1 works as an NO or NC contact.
Setting as to whether the contact for alarm value 2, channel 1 works as an NO or NC contact.
Two-channel version only!
Setting as to whether the contact for alarm value 1, channel 2 works as an NO or NC contact.
Two-channel version only!
Setting as to whether the contact for alarm value 2, channel 2 works as an NO or NC contact.
Setting as to whether the contact for the fault message works as an NO or NC contact.
Setting the system clock.
Format DD.MM.YY hh:mm.
Frequency offset
1)
Pressing the
J
key starts blank value measurement for compensating the reagent's inherent color.
Manually changing the frequency offset.
1)
Frequency offset
1)
-5000 to +5000
0
1) Redetermine the frequency offset and the calibration factor every time the reagents are changed or the photometer is replaced. To do so, instead of sample connect deionized water to the sample inlet to determine the frequency offset (blank value). The value is generally between 0 and 10 Hz.
Endress+Hauser 25
Operation
Menu item
Measuring range start 1
Measuring range start 2
Measuring range end 1
Measuring range end 2
Alarm value
AV 1 - 1
Alarm value
AV 2 - 1
Alarm value
AV 1 - 2
Alarm value
AV 2 - 2
Time
1. measurement
5.2.4
PARAMETER ENTRY
Range of adjustment
(factory settings in bold)
SI-A: 1 to 200 μg/l / 0 μg/l
SI-B: 50 to 5000 μg/l / 0 μg/l
SI-A: 1 to 200 μg/l / 200 μg/l
SI-B: 50 to 5000 μg/l / 5000 μg/l
SI-A: 1 to 200 μg/l / 100 μg/l
SI-B: 50 to 5000 μg/l / 2500 μg/l
SI-A: 1 to 200 μg/l / 200 μg/l
SI-B: 50 to 5000 μg/l / 5000 μg/l
SI-A: 1 to 200 μg/l / 100 μg/l
SI-B: 50 to 5000 μg/l / 2500 μg/l
SI-A: 1 to 200 μg/l / 200 μg/l
SI-B: 50 to 5000 μg/l / 5000 μg/l
01.01.96 00:00 to 31.12.95 23:59
Measuring interval
Frequency of measurement
Channel 1
Frequency of measurement
Channel 2
2 to 120 minutes
10 min
0 to 9
1
2)
0 to 9
1
Display
Time
1. calibration
01.01.96 00:00 to 31.12.95 23:59
Calibration interval
0 to 720 h
48 h
Calibration solution
SI-A: 1 to 200 μg/l / 500 μg/l
SI-B: 50 to 5000 μg/l / 500 μg/l
26
Stamolys CA71SI
Info
The specified concentration is allocated a value of 0 or 4 mA at analog output 1
1)
.
Only two-channel version!The specified concentration is allocated a value of 0 or 4 mA at analog output 2.
The specified concentration is allocated a value of 20 mA at analog output 1.
Only two-channel version!The specified concentration is allocated a value of 20 mA at analog output 2.
Concentration threshold value for limit relay 1, channel 1
(switching hysteresis 2% of alarm value).
Concentration threshold value for limit relay 2, channel 1
(switching hysteresis 2% of alarm value).
Only two-channel version!Concentration threshold value limit relay 1, channel 2 (switching hysteresis 2% of alarm value).
Only two-channel version!Concentration threshold value limit relay 2, channel 2 (switching hysteresis 2% of alarm value).
Date format DD.MM.YY, time hh.mm. After each change the instrument does not wait for the measuring interval. If the measurement is to start immediately, set the time in the past.
Time between two measurements.
If the setting is 2 minutes, the measurements take place without any pauses.
Two-channel version only!
Number of measurements at channel 1 before switching to channel 2.
Two-channel version only!
Number of measurements at channel 2 before switching to channel 1.
Time of 1st calibration (DD.MM.YY, time hh.mm). After each change the instrument does not wait for the calibration interval. If the calibration is to start immediately, set the time in the past.
Analyzers are delivered pre-calibrated.
– Start the 1st calibration 2 hours after the initial startup at the earliest (warm-up phase)
– Set the time to 8:00 to track the effects of calibration in the curve.
– If you have started a calibration manually, you should redefine the time of the 1st calibration because the interval is calculated starting from the last calibration.
Time between two calibrations.
The "0 h" setting means that calibration does not take place.
Recommended: calibration interval of 48 to 72 h.
Concentration of the calibration standard.
Select a standard whose concentration is in the top third of the measuring range.
Endress+Hauser
Stamolys CA71SI Operation
Menu item Range of adjustment
(factory settings in bold)
Display Info
1. flushing
Flushing interval
Flushing hold on
01.01.96 00:00 to 31.12.95 23:59
0 to 720 h
0 h (SI-A)
48 h (SI-B)
0 to 60 s
0 s (SI-A)
60 s (SI-B)
Time of 1st flushing (DD.MM.YY, time hh.mm). After each change the instrument does not wait for the flushing interval. If the flushing is to start immediately, set the time in the past.
– Set the time to 4:00 to track the effects of flushing in the curve.
– If you have started flushing manually, you should redefine the time of the 1st flushing because the interval is calculated starting from the last flushing.
Time between two flushings.
The "0 h" setting means that cleaning no longer takes place.
Dwell time of flushing solution in the pump-mixerphotometer line.
Recommended: 30 to 60 s.
1)
2)
Option 0 or 4 mA: see "CONFIGURATION".
All channels set to 0 means that the channel is selected externally. All channels set to 1 means alternating, beginning with channel 1.
NOTICE
Measuring errors due to contamination by cleaning agents
► Always coordinate calibration and flushing.
► Carry out flushing with standard cleaners approx. 3-4 hours before the next calibration.
► Flushing with special cleaners (e.g. hydrochloric acid or ammonia) has a lasting effect on calibration. Therefore, carry out this cleaning 3-4 hours before calibration. Then rinse with deionized water afterwards.
5.2.5
LANGUAGE
The following languages are available:
• Deutsch
• English
• Français
• Suomi
• Polski
• Italiano.
5.2.6
ERROR DISPLAY
• This menu is a "read-only menu".
• You can find the individual error messages, their meaning and solutions to problems in chapter
"Trouble-shooting instructions".
• If there is at least one error message, the signal output is set to "fault".
• Causes of faults are requested for every measurement. If an error which occurred previously no longer exists, it is automatically cancelled.
Endress+Hauser 27
Operation
28
Stamolys CA71SI
5.2.7
SERVICE
NOTICE
In this menu, you can trigger direct function controls. If these controls are incorrectly combined, however, this could damage the analyzer.
► Using the Service menu requires detailed knowledge of how the analyzer works and operates.
► Exit the Service menu with the "M" key. All functions will be reset to the delivery state then.
Option Display
Pumps and valves
Signal frequency
Info
"Virtual switching board"
Various valve and pump combinations can be selected.
The setting options are:
– Valve 1:
P (sample) or S (standard)
– Valve 2:
S (standard) or C (cleaning agent)
– Valve 3 (Two-channel version only):
1 (channel 1) or 2 (channel 2)
– Valve 4 (for optical cell outlet, improves cleaning and avoids memory effects): s (stop) or g (go)
– Pump 1 (sample or standard/cleaner) and pump 2
(reagents): s (stop) or g (go)
– Mixture
The reagent and sample pumps can be switched on together, so that they run in the same ratio as for filling the sample-reagent mixture in measuring mode.
s (stop) and g (go)
P1 and P2 are inactiv, when G is at go. If P1 or
P2 is at go, G is not available.
The following valve combinations are possible:(applies to one-channel and two-channel version, whereby where the latter is concerned selection is made by positioning valve 3 between channel 1 and 2)
– V1: P, V2: S
Passage for the sample. This combination is automatically reset on leaving the service menu.
– V1: S, V2: S
Passage for standard solution
– V1: S, V2: R
Passage for cleaning agent
Signal frequency of the photometer
5.2.8
DATA STORAGE-Measured values
Two menus, "DATA MEMORY 1" and "DATA MEMORY 2", are only for the two-channel
version. In the one-channel version there is only one menu "DATA MEMORY".
Selection Display
Measured values
Serial output
Info
The data memory contains the last 1024 concentration measured values with date and time. If there are no values available, "Empty set" appears.
Browse through the data sets by pressing the
V
and
W keys.
You can output all data sets (in ASCII-format) via the serial interface. For this, the receiving end (PC) must be configured like this: 9600, N, 8, 1.
To send data, the receiving end (PC) must send the ASCII character 81 ("Shift", "D").
Endress+Hauser
Stamolys CA71SI
Endress+Hauser
Operation
Selection
Clear data
Display Info
This deletes all data sets.
5.2.9
DATA STORAGE-Calibration data
To enter this menu select the CONFIGURATION menu, browse to the "Calibration factor" option and press the
J
key.
Selection Display
Calibration factor
Serial output
only available via PC!
no display
Info
This data memory contains the last 100 calibration factors with date and time. If there are no values available,
"Empty set" appears.
Browse through the data sets by pressing the
V
and
W keys.
You can output all data sets (in ASCII-format) via the serial interface. For this, the receiving end (PC) must be configured like this: 9600, N, 8, 1.
To send data, the receiving end (PC) must send the ASCII character 81 ("Shift", "C").
Clear data This deletes all data sets.
5.3
Calibration
5.3.1
Standard calibration data
The signal strength is processed device-internally as a frequency.
The following table provides an overview of the standard calibration data.
Compare these values to your own data.
After changes in the CONFIGURATION menu and in case of software updates, you can check and, if necessary, change the calibration data in the sub-menu.
Silicate, lower measuring range
SI-A
Silicate, upper measuring range
SI-B
Measuring range
1 to 200 μg/l
Concentration [mg/l]
0.000
0.020
0.040
0.060
0.080
0.100
0.120
0.140
0.160
0.200
50 to 5000 μg/l 0.00
0.10
0.50
1.00
1.50
2.00
2.50
3.00
4.00
5.00
Frequency [Hz]
0
68
142
217
289
369
443
515
591
742
0
107
211
311
396
515
555
636
671
858
29
Operation Stamolys CA71SI
5.3.2
Calibration example
Complete calibration, consisting of two offset measurements and one standard measurement, is triggered via the automatic calibration mode. In addition, you can start offset and standard calibration manually via the CONFIGURATION menu and view the current calibration data.
Offset calibration with the CAY642-V10C00AAE standard
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the CONFIGURE menu and press
F
.
3.
Use
V
to enter a 3 in "Code No.?".
4.
Use
F
to go to the "Calibrate offset" option.
5.
Now use
J
to start the offset calibration. Wait until the offset value is displayed.
6.
Press
F again. This takes you back to measuring mode.
Calibration
Ensure that you have changed the reagents, filled the hoses again (no air bubbles) and the analyzer is in measuring mode.
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the PARAMETER ENTRY menu and press
F
.
3.
Use
F
to go to the "1st calibration" option.
4.
Now use the
W
or
V
and
F
keys to set a time which lies in the past.
5.
Press
F
to accept the value and then press
H
twice to return to the main menu (AUTO
MEASURING).
6.
Press
F again. This takes you back to measuring mode.
The calibration is now carried out automatically.
NOTICE
After the calibration has finished, the analyzer automatically switches to the measuring mode.
► You now have to set the time of the 1st calibration back into the future in order to align the calibration and rinsing times to each other. The rinse must be performed 3-4 hours before the next calibration.
► Proceed as described above to change the setting for the time of the 1st calibration. After changing to measuring mode, the analyzer automatically begins measuring, rinsing and calibrating at the defined times.
30 Endress+Hauser
Stamolys CA71SI Commissioning
6 Commissioning
6.1
Function check
!
CAUTION
Avoid damage due to incorrect power supply and inaccurate or insecure hose connections
► Check that all connections are secure. In particular, ensure that all hose connections are secure, so that no leaks occur.
► Ensure that the power supply voltage corresponds to the voltage specified on the nameplate.
6.2
Switch-on
6.2.1
Dry commissioning
When the analyzer has been configured and calibrated, the measuring cycle starts automatically.
Configuration is no longer necessary.
To perform initial commissioning or to readjust the device parameters, proceed as follows:
1.
Plug in the analyzer's plug with grounding contact into the socket.
2.
Press the
H
key until AUTO MEASURING is displayed.
• If possible, let the analyzer warm up in standby mode before commissioning ("Auto measuring" display). The time can be defined via the "1. measurement" option in the PARAMETER ENTRY menu.
• At the start of measurement with a cold analyzer, the first measuring results will be incorrect. The reaction is temperature-dependent and if the temperature is too low the predefined reaction time is insufficient for a complete reaction. For this reason, never carry out calibration with a cold analyzer. Wait at least two hours before carrying out calibration.
3.
Select the CONFIGURATION menu and configure up to and including "Current date/time".
With
H
you can return to the main menu.
4.
Now work through the PARAMETER ENTRY and SERVICE menus.
With
H
you can return to the main menu.
5.
Select CONFIGURATION again and use
F
to go to the "Calibrate offset" menu item
1)
.
6.
Connect a vessel containing the 0 μg/l standard CAY642-V10C00AAE or a plastic vessel containing ultrapure water to the "Sample" connection and start the frequency offset (
J
key).
The value determined is displayed and saved. Repeat the frequency offset measurement several times.
7.
Then reconnect the sample line.
With
H
you can return to the main menu.
The analyzer starts automatically (triggered by a control signal or integrated timer) with the
"Calibration", "Measurement" and "Flushing" procedures in accordance with the device parameters that you have set (1st calibration, 1st measurement, 1st flushing times and the respective intervals control the procedure).
1) Steps 5 and 6 are not required for version CA71SI-A.
Endress+Hauser 31
Commissioning
32
Stamolys CA71SI
Settings for CA71SI-B
Function
Measurement Flushing (sample)
Delay to sample
Stabilization
1. measurement (basis)
Fill mixture
Reaction
2nd measurement (plateau)
Flushing (sample)
Calibration Flushing (standard)
Delay to standard
Stabilization
1. measurement (basis)
Fill mixture
Reaction
2nd measurement (plateau)
Flushing (sample)
Flushing Pump cleaning solution
Allow to react
Pump cleaning solution
Settings for CA71SI-A
Function
Measurement Flushing (sample)
Delay to sample
Stabilization
1. measurement (basis)
Flushing (reag. line)
Fill mixture
Reaction
2nd measurement (plateau)
Empty cell
Flushing (sample)
Calibration Flushing (standard)
Delay to standard
Stabilization
1. measurement (basis)
Flushing (reag. line)
Fill mixture
Reaction
2nd measurement (plateau)
Empty cell
Flushing (sample)
Duration [s] Range of adjustment
120
20 to 999
8
CONFIGURATION / "Delay to sample"
SERVICE / "Pumps and valves"
17
120
30
120
20 to 999
8
17
120
CONFIGURATION / "Delay to sample"
SERVICE / "Pumps and valves"
30
½ Flush hold on
5
½ Flush hold on
PARAMETER ENTRY / "Flushing hold on"
Duration [s] Range of adjustment
160
20 to 999
8
CONFIGURATION / "Delay to sample"
2
24
90
SERVICE / "Pumps and valves"
30
160
20 to 999
8
2
24
90
CONFIGURATION / "Delay to sample"
SERVICE / "Pumps and valves"
30
6.2.2
Wet commissioning
Wet commissioning is different to dry commissioning in that in wet commissioning you fill the reagent lines before the automatic cycles are started.
Proceed as follows:
1.
Plug in the analyzer's plug with grounding contact into the socket.
2.
Press the
H
key until AUTO MEASURING is displayed.
3.
Select the SERVICE menu.
4.
Switch the P2 reagent pump "on" (select P2 with
F
and set to "g" with
V
) and leave it running until you can tell that there are reagents at the T-hose connector. After this, switch P2 "off" (s) again with
W
.
Endress+Hauser
Stamolys CA71SI Commissioning
5.
Now switch the valves to passage for standard (select V1: S, V2: S; with
F
or
K
switch to
"S" with
V
) and then switch the P1 sample pump "on". Leave the pump running until you can tell that there is standard at the T-hose connector. Switch P1 back "off".
6.
CA71SI-B only:
Now switch the valves to passage for cleaning agent (select V1: S, V2: R; with to "R" with
F
or
K
, switch
V
) and then switch the P1 sample pump "on". Leave the pump running until you can tell that there is cleaning agent at the T-hose connector. Switch P1 back "off".
7.
Now switch the valves to passage for sample (select V1: P, V2: S; with and "S" with
F
or
K
, switch to "P"
V
) and then switch the P1 sample pump "on". If you notice that sample is present at the T-hose connector, let the pump run for another 2 minutes. This removes any standard or cleaner residues.
Then switch P1 "off".
8.
Now proceed as with dry commissioning (from step 2).
6.2.3
Using a dilution module
You can extend the measuring range of your analyzer with the sample dilution module.
How it works:
• The analyzer determines the absorption of the sample at the specified wavelength.
It does not matter whether the sample was diluted previously or not.
• The characteristic of the photometer (--> standard calibration data) refers to the concentrations of the diluted sample and not to those of the (undiluted) original sample.
• The dilution factor is not taken into account until the measured value is calculated.
• The concentration information you specify for the measuring range and the alarm values in
PARAMETER ENTRY refer to the concentrations of the original sample.
You achieve maximum measuring accuracy if you also dilute the standard with the dilution module.
In the event of a 1:10 dilution, you have to use a 10 mg/l standard solution instead of a 1.0 mg/ l standard solution (without a dilution module) for example.
Balance changes to the dilution ratio (e.g. through hose aging) with the calibration. The calibration factor determined takes into account all deviations from the ideal state through reagents, the photometer and dilution.
Installing the dilution module
The dilution module is supplied ready to use.
Installation:
1.
Release the sample pump hose from the hose connector to valve 1 (
2.
Connect the sample hose of the dilution module (black/black or yellow/blue, depending on the version) to the hose connector that is now free.
3.
Connect the T-section (item 1) of the dilution module to the input of the sample pump (hose box K1). Use the original sample pump hose for this purpose.
4.
Fit the dilution module onto the free hose box positions of the sample pump.
5.
Connect an outflow hose (O) to the free outlet of the T-piece (item 1).
Route this hose in the device in such a way that is guided at least 20 cm upwards and then into the open outlet. This prevents the hose from running dry.
NOTICE
Measuring errors due to malfunctioning outflow from photometer
Backwater, siphon effects or similar influence the sample outflow and cause measuring errors
► Do not connect the outflow hoses of the dilution module and the photometer.
Endress+Hauser 33
Commissioning Stamolys CA71SI
34 a0006622
2
3
4
O
P
W
1
K1
K2
K3
A
Hose box of the original sample pump
Hose box for additional sample (standard)
Hose box for dilution water
Sample supply to photometer
Overflow (in outlet of analyzer)
Sample hose (sample or standard)
Dilution water inlet (to be provided by the client)
T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)
Static mixer 122-012
T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)
Hose connector 3.2 x 1.6 mm (0.13 x 0.06 inch)
Changing parameters
The following section only lists the parameters where you have to make changes to the standard settings.
Start with CONFIGURATION.
CONFIGURATION
Delay to sample Increase the value of the standard setting by 40 seconds since additional sample volume is drawn in.
PARAMETER ENTRY
Measuring range start 1 / measuring range start 2
Measuring range end 1 / measuring range end 2
Alarm values AV
1-1 / 1-2 / 2-1 / 2-2
Calibration solution
Enter values that refer to the concentration of the undiluted sample.
If you are using a dilution module with the factor 10, for example, multiply the settings to date (without the dilution module) by 10.
Enter the original concentration (undiluted) of the standard here.
Endress+Hauser
Stamolys CA71SI Maintenance
7 Maintenance
!
CAUTION
You must not carry out any procedures not listed in the following chapters, yourself.
► Not listed works must only be carried out by the service.
If you use components for maintenance or modification, which are not approved by
Endress+Hauser, the warranty and the certified conformity of the device will become void.
7.1
Maintenance schedule
All maintenance duties that have to be carried out during normal operation of the analyzer are explained below.
If you are using a sample conditioning unit, e.g. CAT430, coordinate the maintenance work required for it with that of the analyzer. For this, read the maintenance chapter in the respective operating instructions.
Period of time
Weekly
Monthly
Every 3 months
Every 6 months
Duty Note
– Check and note calibration factor (for service purposes)
– Move valve hoses into their position and spray with silicone (extends the service life).
CONFIGURATION
– Flush sample line hose system with pressurized water (disposable syringe), check and replace reagents if necessary
– Remove the cassette of the sample pump.
– Grease pump hoses with silicone grease
– Check sample collector for fouling and clean it if necessary
– Connect the one-way syringe instead of the sample inlet.
– Cleaning the drain lines:
Flush all hoses and then rinse with sample for at least 30 minutes
– Rotate pump hoses
– Cleaning of the filter mat of the housing fan
– Replace pump hoses
– Replace valve hoses
– s. chapter "Cleaning the filter mat of the housing fan"
– see chapter "Replacing pump hoses"
– see chapter "Replacing valve hoses"
Whenever working on the reagent hoses, the hoses must be disconnected from the canisters, in order to prevent contamination of the reagents. Empty the reagent hoses via the service menu.
Endress+Hauser 35
Maintenance Stamolys CA71SI
7.2
Replacing reagents
!
CAUTION
Corrosive chemicals and other risks
Eye or skin injuries or crushes possible
► There is a danger of crushing limbs at doors, inserts and pump heads.
► Refer to the warning instructions in the safety data sheets when handling reagents or cleaning solutions. Wear protective clothing, gloves and goggles.
► If reagents come into contact with the skin or eyes, carefully rinse with copious amounts of water and consult a physician immediately.
► Never add water to reagents. Reagents containing acids may splash and heat may build up.
If you store the reagents correctly (in the dark, not over 20 ˚C) they will be stable for minimum
12 weeks from the date of manufacture (batch number). When this period of time has expired, the reagents must be replaced. Shelf life can be prolonged by keeping the reagents in a dark, cool storage place. The reagents absolutely must be replaced when:
• The reagents have been contaminated by sample (see "Trouble-shooting instructions")
• The reagents are too old
• The reagents have been spoilt by incorrect storage conditions or environmental influences.
7.2.1
Checking reagents
1.
Check the concentration of the standard in the laboratory. Adapt the values (PARAMETER
ENTRY, calibration solution") or replace the standard solution.
2.
Mix 10 ml of standard (c=500 μg/l or higher) and 5 ml of reagent SI 1 in a beaker.
3.
After approx. 5 minutes, add 5 ml each of reagent SI 2 and SI 3.
The finished mixture must be free of particles.
The reagents must be replaced if no visual coloration (blue) occurs after max. 2 minutes or if the mixture is not free of particles.
7.2.2
Replacing reagents
1.
Carefully remove the hoses from the canisters and wipe them with a clean and dry (paper) towel. Wear protective gloves when doing this.
2.
Switch on the reagent pump for about 5 seconds for the hoses to be drained.
3.
Replace the reagent canisters and feed the hoses into the new canisters.
4.
Fill the reagent hose with the new reagents (SERVICE). Switch all pumps to "g" for this purpose. Do not switch the pumps off ("s") until no more air bubbles are to be seen in the hoses.
5.
Connect standard CAY642-V10C00AAE to the sample flow and rinse the system (V1: P, P1: g and after 5 minutes P1: s).
6.
Exit the Service menu and carry out an offset calibration with CAY642-V10C00AAE and a calibration.
36 Endress+Hauser
Stamolys CA71SI Maintenance
7.3
Replacing pump hoses
!
CAUTION
Risk of splashing reagents when removing hoses from the hose nozzles
► Wear protective clothing, gloves and goggles.
The peristaltic pumps used for the analyzer convey the medium in a combination of vacuum and displacement pump. The pump rate is dependent on the elasticity of the pump hoses. Elasticity decreases and the pump rate drops as mechanical stress increases. Wear depends on mechanical stress (measuring interval, pump pressure). The wear effect can be compensated by periodical calibration. If the hose elasticity becomes too low and the pump rate is no longer reproducible, it is necessary to replace the hoses.
Removing the old hoses
1.
Remove the reagent hoses from the reagent canisters in order to prevent contamination of the reagents.
2.
Flush the old hoses with ultrapure water first and then with air to empty them (see SERVICE).
3.
Remove the hoses from the nipples of the hose boxes (item 5).
4.
Release the hose box(es) (1 to 3 per pump):
– Press against the lower retainer (item 3).
– You can now remove the hose box along with the pump hose.
– Remove the old hose from the box any dispose of it.
– Clean the hose box and the roller head (item 1) with a tissue.
1 2
3
4
1
2
5
Roller head
Upper retainer of hose box
Lower retainer of hose box
Guide at pump hose
Nipple with guide
5
3
4
a0001652
Installing the new hoses
1.
Grease the new pump hoses and the pump heads with silicone grease.
NOTICE
Danger of mix-up and resulting malfunction
► Make sure you connect the new pump hoses to the correct connections at the T-hose connector.
► For the order numbers of the pump hoses, see the chapter "Troubleshooting"/"Spare parts".
► The orange/orange Fluran hose must be installed with reagent R3.
2.
Fit the new hose on the hose box.
3.
First pull the hose downwards at both ends and then push the guide on the hose into that of the nipple on the hose box. Make sure it is seated correctly.
4.
First place the hose box into the upper retainer (item 2) of the pump and then press the box into the lower retainer (item 3).
Make sure the hose boxes are in the correct order in the pump (
Endress+Hauser 37
Maintenance Stamolys CA71SI
R3
R2
R1 a0002206
Fig. 19:
R1
R2
R3
Reagent pump, top view
Reagent 1
Reagent 2 (if present)
Reagent 3 (if present)
5.
Reconnect the reagent hoses to the canisters.
6.
After installation, refill the hoses with sample, standard or cleaner (SERVICE)
2)
.
7.
Carry out an offset calibration and a calibration.
7.4
Replacing valve hoses
To replace the hoses, proceed as follows:
1.
Rinse the old hoses first with water and then with air to empty them (see SERVICE).
2.
Remove the hoses from the valves: a. You can disconnect the front hoses directly because the valves are open when deenergised.
b. To remove the back hoses, press the black button on the valve and disconnect the hoses.
3.
Grease the new hoses with silicone grease before inserting them.
4.
Install the new valve hoses in the reverse sequence of operations.
Ensure that the hoses are connected correctly.
5.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
6.
Carry out an offset calibration and a calibration.
2) Make sure that the hoses are bubble-free when filled. If not, check the sample flow rate (see below).
38 Endress+Hauser
Stamolys CA71SI
Endress+Hauser
1
Maintenance
V1
2
V2
3
V4
V3
7
Fig. 20: Valves and valve hoses
5
6
3
4
7
1
2
V1-4 Valves 1, 2 and 4 (if present)
V3 Valve 3, two channel switch (optional)
Sample
To the pump
Y-piece, connecting hose to valve 1, behind it
Outlet hose
Hose valve 2, front, standard
Hose valve 2 back, cleaning agent
Hose valve 1 front, sample
6 5
7.5
Replacing the static mixer
To replace the mixer, proceed as follows:
1.
Rinse first with water and then with air (see SERVICE).
2.
Unscrew the four screws on the photometer housing and remove it.
3.
Disconnect the mixer from the photometer and from the T-piece below the photometer housing or release the mixer from the holder.
4.
Remove the old mixer and insert the new one.
5.
Connect the new mixer to the photometer and the T-piece again.
6.
Attach the photometer housing and screw it down.
7.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
8.
Carry out an offset calibration and a calibration.
4 a0001730
7.6
Replacing the photometer optical cell
NOTICE
Electrostatic discharge (ESD) can cause damage to electronic assemblies
► Discharge yourself at a protective earth before handling the assemblies.
Removing the old cell
1.
Rinse first with water and then with air (see SERVICE).
2.
Switch the analyzer off.
3.
Unscrew the four screws on the photometer housing and remove it.
4.
Release the nuts from the guide screws and remove the photometer completely.
5.
Unscrew the four screws on the side of the photometer on which there is no ribbon cable.
6.
Take the photometer's electronics apart.
7.
Take out the cell and remove the hoses.
NOTICE
Traces of grease on the optical surfaces can cause corrupted measured values
► Do not under any circumstances touch the optical window of the cell with your fingers.
39
Maintenance Stamolys CA71SI
Installing the new cell
1.
Insert the new cell.
2.
Connect the cell to the hoses such that the sample is fed in from below.
3.
Secure the hoses with the supplied cable connectors to stop the cell from slipping.
4.
Reassemble the photometer and tighten the screws and nuts.
5.
Attach the photometer housing and screw it down.
6.
After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).
7.
Carry out an calibration ("Calibration").
7.7
Cleaning
7.7.1
Housing
NOTICE
When cleaning, ensure that you do not damage the nameplate on the analyzer.
► Do not use any solvent-based cleaning agents.
To clean the analyzer housing, proceed as follows:
• Stainless steel housing (stainless steel SS 1.4301 (AISI 304)): with a lint-free cloth and Glittol RG 10.51
• GFR housing: with a damp cloth or with tenside-based (alkaline) cleaning agent.
7.7.2
Cleaning of the filter mat of the housing fan
The cover of the filter mat is at the backside of the analyzer (center, top).
1.
Remove the cover to replace the filter mat.
2.
To clean the filter mat you have various options: a. Wash the filter mat in water (approx. 40 ˚C) with a mild detergent if necessary.
b. Alternatively you can beat or vacuum the mat or blow it with compressed air.
c. In case of greasy substances: Wash the filter mat in warm water with grease solvent added.
NOTICE
Damage due to residues of organic solvents and incorrectly installed filter mat
Lost of the guaranteed protection degree due to missing, damaged or incorrectly installed filter mat
► If organic solvents are used, the mat has to be completely dried on air before re-assembled.
► Always take care to install the filter mat correctly.
40 Endress+Hauser
Stamolys CA71SI Maintenance
7.8
Placing out of service
You must place the analyzer out of service before shipping or before longer operation breaks (more than 5 days).
!
CAUTION
Residues of chemicals can cause injuries
► Thoroughly rinse all of the lines of the measuring system with ultrapure water.
To place the analyzer out of service, proceed as follows:
1.
Remove the reagent and standard hoses from the canisters and immerse them in a tank containing ultrapure water.
2.
Switch valve 1 to "Standard" and switch pumps 1 and 2 on for one minute (see SERVICE).
3.
Remove the hoses from the water and allow the pumps to run until the hoses are completely dry.
4.
If you are using a continuous sample supply, disconnect sampling line.
5.
Flush sampling hoses with ultrapure water and then with air, in order to completely empty the hoses.
6.
Remove the valve hoses from the valves.
7.
Remove the load from the pump hoses by removing the hose cassette from the bracket below.
Keep opened reagents and standards in a refrigerator. Observe the shelf-life.
Endress+Hauser 41
Accessories
42
Stamolys CA71SI
8 Accessories
In the following sections, you find the accessories available at the time of issue of this documentation.
For information on accessories that are not listed here, please contact your local service or sales representation.
8.1
Collecting vessel
– for sampling from pressurized systems
– results in an unpressurised continuous sample stream
• Collecting vessel without level measurement; order no. 51512088
• Collecting vessel with level monitoring for pure and ultrapure water, level monitoring with polypropylene float lever; Order no. C-A061019-50
8.2
Reagents, cleaner, standard solutions
• Reagent set active, 1 l SI1+SI2+SI3 each; order no. CAY643-V10AAE
• Reagent set, inactive, 1 l SI1+SI2+SI3 each; order no. CAY643-V10AAH
• Cleaning agent, 1 l; order no. CAY641-V10AAE
• Standard solution 0.0 μg/l SiO
2
• Standard solution 50 μg/l SiO
2
; order no. CAY642-V10C00AAE
; order no. CAY642-V10C50AAE
• Standard solution 100 μg/l SiO; order no. CAY642-V10C01AAE
• Standard solution 500 μg/l SiO
2
• Standard solution 1000 μg/l SiO
; order no. CAY642-V10C05AAE
2
; order no. CAY642-V10C10AAE
NOTICE
Reagents can cause environmental damage
► Note the disposal instructions in the safety data sheets of the reagents.
8.3
Maintenance kit
CAV740, maintenance kit for CA71
• Pump hoses
• Valve hoses
• Hose connectors
• Ordering acc. to product structure
CAV740-
For CA71 parameter
4
6
SI-B
SI-A
B
C
Inlet and outlet hoses
A not selected selected, for CA71SI-B selected, for CA71SI-A
complete order code
8.4
Hose cleaner
• Cleaning agent, alkaline, 250 ml (8.5 fl.oz.); order no. CAY746-V02AAE
• Cleaning agent, acidic, 250 ml (8.5 fl.oz.); order no. CAY747-V02AAE
Endress+Hauser
Stamolys CA71SI
8.5
Dilution module
Dilution module
1 set of pump hoses, 2 hose boxes, 1 set of hose connectors, static mixer
• Dilution 1:3
Order no. C-A030228-10
• Dilution 1:10
Order no. C-A030228-11
8.6
Additional accessories
• Interference suppressor for control, power and signal lines order no. 51512800
• Silicon grease, tube, 35 g order no. 71017654
• Valve set, 2 pieces, for two-channel version order no. 51512234
• Upgrade kit for upgrading from one-channel to two-channel version order no. 51512640
Accessories
Endress+Hauser 43
Trouble-shooting Stamolys CA71SI
9 Trouble-shooting
9.1
Trouble-shooting instructions
Although the analyser is not very prone to faults due to its simple assembly, problems can, of course, not be completely ruled out.
Possible errors, their causes and their possible remedies are listed below.
9.2
System error messages
Error message
Calibration failed
Cell dirty
Incorrect photometer
No sample
Spillage error
No measuring signal
Possible cause
Air bubbles in system Start calibration manually (PARAMETER ENTRY, "1. calibration", change the date accordingly, start measurement) or enter a new calibration factor.
Incorrect concentration of standard Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER
ENTRY", "Calibration solution") or replace the standard.
Reagents contaminated or aged Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.
Standard dosing defective
Tests and / or corrective measures
If a calibration fails, you can enter a new calibration factor manually (CONFIGURATION menu,
"Calibration factor"). Cancel the error message by switching the analyzer off and on briefly.
Check the valves for contamination, obstructions (visual inspection). Replace valve hoses if necessary.
Check the setting in the CONFIGURATION menu.
Incorrect photometer
Not enough light intensity at receiver
Incorrect photometer
No sample
Level measurement defective
Leak at canisters or hoses
– Flush with 12.5% bleaching lye.
– When using a CAT430:
Check the filter.
Check the setting in the CONFIGURATION menu, "Photometer".
Establish sample transfer.
Check level measurement at the sample collector.
Replace defective components and clean and dry the analyzer or those components affected by the leak.
Cell filled with air
Photometer defective
Electrical connection
Fuse defective
Pump sample for 1 minute (SERVICE).
Inform Service
Check all electrical connections and make sure the fuses are firmly seated.
Replace fuse F4 or F5 (semi-delay 0.2 A)
44 Endress+Hauser
Stamolys CA71SI Trouble-shooting
9.3
Process errors without messages
Error Possible cause Tests and / or corrective measures
Reagents contaminated or aged
Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.
Measured values always the same
No sample, no reagents Make sure sample and reagents are supplied, check level monitor and clean if necessary
System blocked Flush with 12.5% bleaching lye.
Check the hose from valve 4.
Incorrect concentration of standard
Reagents contaminated or aged
Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER ENTRY",
"Calibration solution") or replace the standard.
Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.
Measured values inaccurate
Reagent blank value too high
After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,
"Offset calibration")
Wrong dimension
Wrong cell
Check the setting in the CONFIGURATION menu, "Measuring unit" (e.g. ion instead of element).
Check the setting in the CONFIGURATION menu, "Photometer"
Sample suction time too short
Increase the suction time (CONFIGURATION, "Delay to sample")
Matrix effects
(substances that interfere with the photometric method)
Detect interfering substances (see Technical Information, "Measuring principle"), possibly use sample conditioning
Filter life too long
System blocked or contaminated
Take a counter sample at the analyzer inlet and check the concentration in the laboratory. If there is no deviation to the analyzer measured values, replace the ultrafiltration modules or reversible flow filters more frequently.
Flush with 12.5% bleaching lye.
Dosing
Cell dirty
Replace pump hoses.
Flush with 12.5% bleaching lye first and then with 5% hydrochloric acid
Measured values inaccurate
Counter sample in the laboratory delivers deviating measured values
Standard is dosed to the sample
Sample aging
Wrong size for analog output
Measured value transfer defective
Incorrect measuring range
Check the valves and valve settings. Replace valve hoses if necessary.
Shorten time between sampling and analysis.
Check the setting (CONFIGURATION, "Analog output 1" or " 2").
Adjust measuring range (PARAMETER ENTRY, "Measuring range")
Analyzer will not switch on
Analyzer is running but display is incomplete or off
Pumps fail to run
Measurement does not start
Background noise
No power
Fuse
Initialization failed
Leak
Leak sensor bypassed
Fuse
Pump defective
Leak at photometer
Check electrical cables for interference from strong sources of induction.
Check the electrical connection and ensure power supply.
Replace fuse F1 (slow-blow 0.5 A)
Switch the analyzer off and after approx. 30 seconds back on.
See "Spillage error" error message
Interrupt the contact between the two leak sensors (terminals 67-66)
Check all fuses and replace if necessary.
Service
Service
Endress+Hauser 45
Trouble-shooting
Error
"Measurement" display flashes
Possible cause
Time of 1st measurement not reached
Interval not expired
Time of 1st calibration not reached
Tests and / or corrective measures
The date must be between 01.01.1996 and the current date.
Change parameter setting.
The date must be between 01.01.1996 and the current date.
Calibration does not start Interval not expired or
0 h
Flushing does not start
Change parameter setting.
Leak at photometer
Time of 1st flushing not reached
Service
The date must be between 01.01.1996 and the current date.
Change parameter setting.
Leak at photometer
Blockage, deposits in device
Interval not expired or
0 h
Device or discharge blocked
Not enough sample conditioning
Remove blockage.
Service
Shorten the cleaning intervals of sample conditioning.
9.4
Spare parts
9.4.1
Overview for CA71SI-B
Stamolys CA71SI
46 a0006414
A
B
C
P
H
F
Sample input two-channel version
Standards and reagent canisters
Collecting vessel
Sample
Holder for dosage loop
Fluran hose, orange/orange (Kit CAV740)
R
R1
R2
R3
S
G
Cleaner canister
Reagent 1 canister
Reagent 2 canister
Reagent 3 canister
Standard solution canister
Coded pump hose (see Kit CAV740)
Endress+Hauser
Stamolys CA71SI
9.4.2
Overview for CA71SI-A
Trouble-shooting
Endress+Hauser a0006420
Fig. 22:
A
B
C
P
H
Overview of spare parts
Sample input two-channel version
Standards and reagent canisters
Collecting vessel
Sample
Holder for dosage loop
F
R1
R3
S
G
Fluran hose orange/orange (in Kit CAV740)
Reagent 1 canister
Reagent 3 canister
Standard solution canister
Coded pump hose (see Kit CAV740)
9.4.3
Spare parts for sample and reagent transportation
142
143
154
155
135
136
140
141
160
161
Item
120
121
122
123
130
131
134
Spare part
Norprene hose, 1.6 mm
C-Flex hose, 3.2 mm (permeate inlet and overflow with sample collector.)
C-Flex hose ID 6.4 mm
C-Flex hose, 1.5 mm
Hose connector 1.6 mm x 1.6 mm (10 pieces)
T-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)
Y-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)
Connection nipples for sample collector (10 piece)
T-hose connector 3.2 mm x 3.2 mm x 3.2 mm (10 pieces)
Mixer set, complete (2 pieces)
Collecting tank
Outlet pipe with hose nipple (2 pieces)
Kit CA71 Leak sensor for collecting tank (2 pieces)
Kit CA71 Valve, complete (1 piece for one-channel version)
Valve set for two-channel version
Roller head with holder for hose pump
Kit CA71 Hose box for pump (1 piece)
Order number
51504116
51504114
51504115
51512535
51506495
51506490
51512096
51512099
51516166
51512101
51512102
51512104
51512103
51512100
51512234
51512085
51512086
47
Trouble-shooting Stamolys CA71SI
Item
171
182
183
184
185
300
301
302
Spare part
Collecting vessel without level monitoring
Outlet pipe with hose nipple
Leak sensor for collecting tank
Mixer assembly
Drain valve
Kit CA T-hose connectors 1.6 mm x 0.8 mm x 1.6 mm
Kit CA hose connectors 3.2 mm x 1.6 mm
Kit CA hose connectors 1.6 mm x 0.8 mm
9.4.4
Analyzer spare parts
Item
H
144
200
Spare part
Dosing loop (H), complete
Photometer cell, 2 pieces
– SI-B
– SI-A
Photometer
– SI-A
– SI-B
Order number
51512088
51515578
51515581
51515579
51515580
71039848
51506491
71039845
Order number
51512463
51505778
51512409
51512078
51512079
9.5
Software history
Date Version Changes in the software
08/2006 6.0
Extension
• New parameter FE-D
06/2006 5.9
09/2005
06/2005
5.8
5.7
Extension
• New parameter NO-D
• Stepping motor position is also reset in a reset
• New stepping motor setting for CU-A/B (50 rpm)
• Spanish replaces Dutch and Hungarian
Extension
• New parameter PH-E
• Altered factory settings for SI-A and AM-B
• Characteristics for FE-B, CU-A and CR-A extended
• Measuring range limitation increased to 50000 for all parameters and engineering units
• Calibration value can be adjusted to end of measuring range
Improvement
• Stack overflow eliminated
• FE-C valve control corrected
• RTC properly dealt with when setting/reading the time
07/2004 5.6
Extension
• Range of adjustment for delay to sample to 999s
• Range of adjustment for cleaning duration to 300s
• Range of adjustment for frequencies to 5800 Hz
• Factory settings accepted as specified
• Altered measuring cycles for AM-D, AL-A (like hardness)
Documentation
BA360C/07/xx/07.06
BA355C/07/xx/07.06
BA356C/07/xx/07.06
-
BA353C/07/xx/09.04
BA357C/07/xx/10.04
48 Endress+Hauser
Stamolys CA71SI Trouble-shooting
Date Version
05/2004 5.5
Changes in the software
Extension/improvement
• Longer initialization
• Altered measuring cycle for SI
• Shortened reaction time, AM-C (90 s)
• SI-A: 0 Hz signal error only after 90 s
• Reason for crashing during frequency display in SERVICE eliminated
• Restart after power failure
03/2004 5.4
Extension/improvement
• New process: HA
• Menu extension: Cell drain time
• Calibration valve off during reaction
• Hungarian and Spanish no longer available
• New hardness engineering units
• New standard parameters/factory settings
• Stepping motor activation corrected
08/2003 5.3
08/2003
05/2003
12/2002
5.2
5.1
5.0
Improvement
• Fault elimination for measuring range switchover, HA
Extension/improvement
• Factory settings for SI changed
• CL characteristic changed
• New startup message
• Fault elimination for output for current output, two-channel
Extension/improvement
• New parameter CL
• Memory correction
• Time for "Delay to sample" is used for zero measurement
• Factory settings for MN, SI, HA changed
Original software
Documentation
BA364C/07/xx/06.04
BA361C/07/xx/02.04
-
-
BA354C/07/07.03
BA353C/07/xx/01.03
BA355C/07/xx/01.03
BA356C/07/xx/01.03
BA357C/07/xx/01.03
BA358C/07/xx/01.03
BA359C/07/xx/01.03
BA360C/07/xx/01.03
BA361C/07/xx/01.03
BA362C/07/xx/01.03
BA363C/07/xx/01.03
BA364C/07/xx/07.03
9.6
Return
Before returning observe the following items:
• Contact your regional Endress+Hauser sales center to receive detailed shipping conditions.
• Enclose the duly completed "Declaration of de-contamination" with the packaging.
You can find the form as master at the end of these instructions.
9.7
Disposal
The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.
Please observe local regulations.
Endress+Hauser 49
Technical data
Measured variable
Measuring range
Wavelength
Reference wavelength
Output signal
Signal on alarm
Load
Serial interface
Data logger
Load capacity
Supply voltage
Power consumption
Current consumption
Fuses
Stamolys CA71SI
10 Technical data
10.1
Input
SIO
2
[μg/l, mg/l]
SI-A
1 to 200 μg/l
SI-B
50 to 5000 μg/l
810 nm
565 nm (SI-B only)
10.2
Output
0/4 to 20 mA
Contacts: 2 limit contacts (per channel), 1 system alarm contact optional: end of measurement (with two channel version display of channel no. available) max. 500
Ω
RS 232 C
1024 data pairs per channel with date, time and measured value
100 data pairs with date, time and measured value for calibration factor determination (diagnostic tool)
230 V / 115 V AC max. 2 A, 30 V DC max. 1 A
10.3
Power supply
115 V AC / 230 V AC ±10%, 50/60 Hz approx. 50 VA approx. 0.2 A at 230 V approx. 0.5 A at 115 V
1 x slow-blow 0.5 A for electronics
2 x semi-delay 0.2 A for photometer
1 x slow-blow 0.1 A for motors
1 x slow-blow 1 A for fan
50 Endress+Hauser
Stamolys CA71SI Technical data
10.4
Performance characteristics
Time between two measurements
t mes
= reaction time + rinse time + waiting time + rinse again time + filling time + sampling time + reagent refusal time (min. waiting time = 0 min)
Maximum measured error 1 to 200 μg/l:
±2 % of measuring range end there of 5 to 20 μg/l:
±2 μg/l
50 to 5000 μg/l:
±2 % of measuring range end
Measuring interval
t mes
to 120 min
Reaction time
Sample requirement
6 minutes
• SI-A:
90 s
• SI-B:
120 s
20 ml (0.68 fl.oz.) per measurement
Reagent requirement
Calibration interval
SI-B: 3 x 0.18 ml (0.006 fl.oz.)
0.86 l (0.23 US.gal) per reagent per month with 15 minute measuring interval
SI-B: 3 x 0.18 ml (0.006 fl.oz.)
0.52 l (0.14 US.gal) per reagent per month with 15 minute measuring interval
0 to 720 h
Rinse interval (SI-B only)
0 to 720 h
Rinse again time
Filling time
30 s
SI-A
24 s
SI-B
17 s
Empy optical cell
Maintenance interval
Servicing requirement
t refusal
= 30 s (SI-A) t refusal
= 0 s (SI-B)
6 months (typical)
15 minutes per week (typical)
Endress+Hauser 51
Technical data
Mounting location
10.5
Installation conditions
Installation on a vibration-free wall
Ambient temperature
Humidity
Ingress protection
Design, dimensions
Weight
Material
10.6
Environment
5 to 35 ˚C (40 to 95 ˚F), avoid strong fluctuations below the condensation limit, installation in usual, clean rooms outdoor installation only possible with protective devices (customer supplied)
IP 54
10.7
Process
5 to 50 ˚C (40 to 120 ˚F)
Sample temperature
Sample flow rate
min. 5 ml (0.17 fl.oz.) per min
Consistency of the sample
low solid content (< 50 ppm)
Sample inlet
Unpressurized
10.8
Mechanical construction
see chapter "Installation"
GFR housing
Stainless steel housing
Without housing
Housing:
Front windows:
Endless hose:
Pump hose:
Valves: approx. 28 kg (62 lbs) approx. 33 kg (73 lbs) approx. 25 kg (55 lbs)
Stainless steel 1.4301 (AISI 304) or glass-fibre reinforced carbon(GFR)
Polycarbonate
®
C-Flex
®
, Norprene
®
Tygon
®
, Viton
®
Tygon
®
, silicone
Stamolys CA71SI
52 Endress+Hauser
Stamolys CA71SI Technical data
Endress+Hauser 53
Appendix
11 Appendix
11.1
Operating matrix
One-channel version
Stamolys CA71SI
54 a0001907-en
Endress+Hauser
Stamolys CA71SI Appendix
Endress+Hauser a0001908-en
55
Appendix
Two-channel version
Stamolys CA71SI
56 a0001909-en
Endress+Hauser
Stamolys CA71SI Appendix
Endress+Hauser a0001910-en
57
Appendix
11.2
Analyzer settings
Place:
Type:
Serial no. analyzer:
Serial no. Photometer:
Software version:
Date:
Photometer Type:
Measuring unit:
Calibration factor: c-offset:
Dilution:
Delay to sample:
Analog output:
AV 1:
AV 2:
Fault signal:
Frequency offset:
Ground line: (demineralised water without reagent)
• 0-20 mA
• normally closed
• normally closed
• normally closed
Measuring range start:
Measuring range end:
AV 1:
AV 2:
1st measuring:
Measuring interval:
1st calibration:
Calibration interval:
Calibration solution:
1st rinsing:
Rinse interval:
Rinse time:
• 4-20 mA
• normally open
• normally open
• normally open
• mg/l s h s
Hz
Hz
• mg/l
• mg/l
• mg/l
• mg/l min h
• mg/l
• μg/l
• μg/l
• μg/l
• μg/l
• μg/l
• μg/l
Stamolys CA71SI
58 Endress+Hauser
Stamolys CA71SI
11.3
Maintenance schedule
Appendix
Endress+Hauser a0001911-en
59
Stamolys CA71SI
Index
A
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Alarm values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–26
Analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
AUTO MEASURING . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Calibration factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Calibration interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Certificates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Checking
Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Collecting vessel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 31
Dry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Wet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Concentration offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Connection
Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
D
DATA STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33, 43
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
ERROR DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Process errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
System errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
F
Filter mat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Flushing interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Frequency offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
G
GFR housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
60
H
Housing
GFR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Stainless steel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Without . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Housing fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
I
Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 9, 12
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Interference suppressor . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
L
LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
M
Main menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Menu
Auto Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Error display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Parameter entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Offset
Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Open version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 22
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
P
PARAMETER ENTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Performance characteristics. . . . . . . . . . . . . . . . . . . . . . . . 51
Photometer optical cell . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Placing out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Endress+Hauser
Stamolys CA71SI
Pump hoses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Q
Quality certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
R
Reaching the terminal block . . . . . . . . . . . . . . . . . . . . . . . 15
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36, 42
Replacing
Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Photometer optical cell . . . . . . . . . . . . . . . . . . . . . . . . 39
Pump hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 49
S
Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Screening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
SERVICE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Stainless steel housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Standard solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Switching contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
T
Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50–52
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
V
Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Endress+Hauser 61
Stamolys CA71SI
62 Endress+Hauser
Declaration of Hazardous Material and De-Contamination
Erklärung zur Kontamination und Reinigung
RA No.
Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung.
Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the packaging.
Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene
"Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der
Verpackung an.
Type of instrument / sensor
Geräte-/Sensortyp ____________________________________________
Serial number
Seriennummer ________________________
Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen
Process data/ Prozessdaten
Temperature /
Conductivity /
Temperatur
Leitfähigkeit
_____ [°F]
_____
[°C]
________ [µS/cm]
Pressure
Viscosity
/
/
Druck
Viskosität
_____ [psi]
_____ [cp]
_____
_____
[ Pa ]
Medium and warnings
Warnhinweise zum Medium
Medium /concentration
Medium /Konzentration
Identification
CAS No.
flammable
entzündlich
toxic
giftig
corrosive
ätzend
harmful/ irritant
gesundheitsschädlich/ reizend
other *
sonstiges*
harmless
unbedenklich
Process medium
Medium im
Prozess
Medium for process cleaning
Medium zur
Prozessreinigung
Returned part cleaned with
Medium zur
Endreinigung
* explosive; oxidising; dangerous for the environment; biological risk; radioactive
* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv
Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.
Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.
Description of failure / Fehlerbeschreibung __________________________________________________________________________
______________________________________________________________________________________________________________
______________________________________________________________________________________________________________
Company data / Angaben zum Absender
Company /
Firma ___
________________________________
_________________________________________________
Address /
Adresse
_________________________________________________
_________________________________________________
Phone number of contact person /
Telefon-Nr. Ansprechpartner:
____________________________________________
Fax / E-Mail ____________________________________________
Your order No. /
Ihre Auftragsnr.
____________________________
“We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returned parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.”
“Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen w eiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringender Menge sind.”
( place, date / Ort, Datum) Name, dept./Abt.
(please print /
bitte Druckschrift
)
Signature / Unterschrift
BA364C/07/EN/14.11
FM9
www.endress.com/worldwide
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Key Features
- - Measures silicate concentration in ultrapure water
- - Photometric measurement principle
- - Measuring range: 0.01 - 1000 µg/l SiO2
- - Compact and wall-mountable design
- - Easy-to-use interface with clear display
- - Automatic calibration and cleaning functions
- - Low maintenance and operating costs
Related manuals
Frequently Answers and Questions
What is the measuring range of the analyzer?
How often does the analyzer need to be calibrated?
What is the power supply requirement of the analyzer?
Can the analyzer be used in hazardous areas?
advertisement
Table of contents
- 4 Safety instructions
- 4 Designated use
- 4 Installation, commissioning and operation
- 4 Operational safety
- 4 Return
- 5 Safety messages and their meaning
- 6 Identification
- 6 Device designation
- 7 Scope of delivery
- 8 Certificates and approvals
- 9 Installation
- 9 Installation conditions
- 12 Installation instructions
- 14 Installation examples
- 14 Post-installation check
- 15 Wiring
- 15 Electrical connection
- 18 Signal connection
- 19 Switching contacts
- 20 Serial interface
- 21 Post-connection check
- 22 Operation
- 22 Display and operating elements
- 22 Local operation
- 29 Calibration
- 31 Commissioning
- 31 Function check
- 31 Switch-on
- 35 Maintenance
- 35 Maintenance schedule
- 36 Replacing reagents
- 37 Replacing pump hoses
- 38 Replacing valve hoses
- 39 Replacing the static mixer
- 39 Replacing the photometer optical cell
- 40 Cleaning
- 41 Placing out of service
- 42 Accessories
- 42 Collecting vessel
- 42 Reagents, cleaner, standard solutions
- 42 Maintenance kit
- 42 Hose cleaner
- 43 Dilution module
- 43 Additional accessories
- 44 Trouble-shooting
- 44 Trouble-shooting instructions
- 44 System error messages
- 45 Process errors without messages
- 46 Spare parts
- 48 Software history
- 49 Return
- 49 Disposal
- 50 Technical data
- 50 10.1 Input
- 50 10.2 Output
- 50 10.3 Power supply
- 51 10.4 Performance characteristics
- 52 10.5 Installation conditions
- 52 10.6 Environment
- 52 10.7 Process
- 52 10.8 Mechanical construction
- 54 Appendix
- 54 11.1 Operating matrix
- 58 11.2 Analyzer settings
- 59 11.3 Maintenance schedule
- 60 Index