Endres+Hauser Stamolys CA71SI Operating Instruction


Add to my manuals
64 Pages

advertisement

Endres+Hauser Stamolys CA71SI Operating Instruction | Manualzz

Operating Instructions

Stamolys CA71SI

Analyzer for the photometric determination of silicate

BA364C/07/EN/14.11

71133178

Valid of:

Software version 5.9

Brief overview

5

6

7

3

4

1

2

Reagent pump (P2), supply from canister

Display

Serial interface RS 232

Photometer cell

Static mixer (depending on version)

Valve V4 (only SI-A)

Reaction coil

8

9

10

11

12

13

Outlet for sample or reagent mixture (SI-A)

Valve V2

Outlet for sample or reagent mixture (SI-B)

Valve V1

Valves V3: top ch. 1, bottom ch. 2

1)

Sample pump P1, supply see below

1) Only for two-channel version

Inlet to the sample pump:

• Valve V1

– hose in front: sample inlet

– hose at the back: inlet from valve V2 (cleaner or standard solution)

• Valve V2

– hose in front: inlet from canister with standard solution

– hose at the back: inlet from canister with cleaner

a0001352

Endress+Hauser

Stamolys CA71SI

Table of contents

1 Safety instructions . . . . . . . . . . . . . . . . 4

1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.2 Installation, commissioning and operation . . . . . . . . 4

1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.5 Safety messages and their meaning . . . . . . . . . . . . . 5

2 Identification . . . . . . . . . . . . . . . . . . . . 6

2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6

2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8

3 Installation . . . . . . . . . . . . . . . . . . . . . . 9

3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 9

3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . 12

3.4 Installation examples . . . . . . . . . . . . . . . . . . . . . . . 14

3.5 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 14

4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 15

4.1 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . 15

4.2 Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . 18

4.3 Switching contacts . . . . . . . . . . . . . . . . . . . . . . . . 19

4.4 Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

4.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 21

5 Operation . . . . . . . . . . . . . . . . . . . . . . 22

5.1 Display and operating elements . . . . . . . . . . . . . . . 22

5.2 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

5.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

6 Commissioning. . . . . . . . . . . . . . . . . . 31

6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

6.2 Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

7 Maintenance. . . . . . . . . . . . . . . . . . . . 35

7.1 Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . 35

7.2 Replacing reagents . . . . . . . . . . . . . . . . . . . . . . . . . 36

7.3 Replacing pump hoses . . . . . . . . . . . . . . . . . . . . . . 37

7.4 Replacing valve hoses . . . . . . . . . . . . . . . . . . . . . . 38

7.5 Replacing the static mixer . . . . . . . . . . . . . . . . . . . 39

7.6 Replacing the photometer optical cell . . . . . . . . . . . 39

7.7 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

7.8 Placing out of service . . . . . . . . . . . . . . . . . . . . . . . 41

8 Accessories. . . . . . . . . . . . . . . . . . . . . 42

8.1 Collecting vessel . . . . . . . . . . . . . . . . . . . . . . . . . . 42

8.2 Reagents, cleaner, standard solutions . . . . . . . . . . . 42

8.3 Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . 42

8.4 Hose cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

8.5 Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . 43

8.6 Additional accessories . . . . . . . . . . . . . . . . . . . . . . 43

9 Trouble-shooting . . . . . . . . . . . . . . . . . 44

9.1 Trouble-shooting instructions . . . . . . . . . . . . . . . . . 44

9.2 System error messages . . . . . . . . . . . . . . . . . . . . . . 44

9.3 Process errors without messages . . . . . . . . . . . . . . 45

9.4 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

9.5 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

9.6 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

9.7 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

10 Technical data . . . . . . . . . . . . . . . . . . . 50

10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

10.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

10.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

10.4 Performance characteristics . . . . . . . . . . . . . . . . . . 51

10.5 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 52

10.6 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

10.7 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

10.8 Mechanical construction . . . . . . . . . . . . . . . . . . . . 52

11 Appendix . . . . . . . . . . . . . . . . . . . . . . . 54

11.1 Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . . 54

11.2 Analyzer settings . . . . . . . . . . . . . . . . . . . . . . . . . . 58

11.3 Maintenance schedule . . . . . . . . . . . . . . . . . . . . . . 59

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Endress+Hauser 3

4

Safety instructions Stamolys CA71SI

1 Safety instructions

1.1

Designated use

The analyzer is a compact photometric analysis system.

It is designed for the monitoring of the silicate content in ultrapure water and boiler feed water.

In particular, CA71 is designated for:

• boiler feed water cycles

• ultrapure water

• steam and condensate analysis

• reversed osmosis

• demineralizers.

Any other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted.

The manufacturer is not liable for damage caused by improper or non-designated use.

1.2

Installation, commissioning and operation

Please note the following items:

• Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel.

Trained personnel must be authorized for the specified activities by the system operator.

• Electrical connection must only be carried out by a certified electrician.

• Technical personnel must have read and understood these Operating Instructions and must adhere to them.

• Before commissioning the entire measuring point, check all the connections. Ensure that electrical cables and hose connections are not damaged.

• Do not operate damaged products and secure them against unintentional commissioning.

Mark the damaged product as being defective.

• Measuring point faults may only be rectified by authorized and specially trained personnel.

• If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning.

• Repairs not described in these Operating Instructions may only be carried out at the manufacturer’s or by the service organization.

1.3

Operational safety

The analyzer has been designed and tested to the highest standards and left the factory in perfect functioning order.

Relevant regulations and European standards have been met.

As the user, you are responsible for complying with the following safety conditions:

• Installation instructions

• Local prevailing standards and regulations.

1.4

Return

Before returning observe the following items:

• Contact your regional Endress+Hauser sales center to receive detailed shipping conditions.

• Enclose the duly completed "Declaration of de-contamination" with the packaging.

You can find the form as master at the end of these instructions.

Endress+Hauser

Stamolys CA71SI Safety instructions

1.5

Safety messages and their meaning

The structure, signal words and safety colors of the signs comply with the specifications of

ANSI Z535.6 ("Product safety information in product manuals, instructions and other collateral materials").

Safety message structure Meaning

!

DANGER

Cause (/consequences)

Consequences if safety message is not heeded

► Corrective action

This symbol alerts you to a dangerous situation.

Failure to avoid the situation will result in a fatal or serious injury.

!

WARNING

Cause (/consequences)

Consequences if safety message is not heeded

► Corrective action

This symbol alerts you to a dangerous situation.

Failure to avoid the situation can result in a fatal or serious injury.

!

CAUTION

Cause (/consequences)

Consequences if safety message is not heeded

► Corrective action

This symbol alerts you to a dangerous situation.

Failure to avoid this situation can result in minor or medium injury.

NOTICE

Cause/situation

Consequences if safety message is not heeded

► Action/note

This symbol alerts you to situations that can result in damage to property and equipment.

Endress+Hauser 5

6

Identification Stamolys CA71SI

2 Identification

2.1

Device designation

2.1.1

Nameplate

Compare the (extended) order code on the nameplate (on the analyzer) to the product structure and your order.

You can read the following information from the nameplate:

• Order code

• Extended order code (=order code from the product structure)

• Serial number

• Measuring range

• Outputs and communication

• Power supply

• Ingress protection

• Environment conditions

2.1.2

Product structure

CA71SI-

A

B

Y

Measuring range

1.0 to 200 μg/l SiO

2

50 to 5000 μg/l SiO

2

Special version acc. to customer’s specification

Sample transfer

1

2

From one measuring point (one channel version)

From two measuring points (two channel version)

Power supply

0

1

2

3

230 V AC / 50 Hz

115 V AC / 60 Hz

115 V AC / 50 Hz

230 V AC / 60 Hz

Collecting vessel (for up to 3 analyzers)

A

B

C

D

Not selected (without collecting vessel)

Collecting vessel without level measurement

Collecting vessel with level measurement (one-channel version only)

Two collecting vessels without level measurement (two-channel version)

Housing

1

2

3

Open design

GFR housing

Stainless steel (1.4301 / AISI 304) + purge

Output

A 0/4 to 20 mA, RS 232

Reagents

1

2

3

To order separately

One set, inactive

Three sets, inactive

complete order code

Endress+Hauser

Stamolys CA71SI Identification

2.2

Scope of delivery

Order reagents separately with analyzer version CA71XX-XXXXXX1.

With all other versions, inactive reagents are included in the scope of delivery. You have to mix the reagents before using them. Please, read the instructions attached to the reagents.

2.2.1

Version CA71SI-B

The scope of delivery comprises:

• an analyzer with mains plug

• a cleaning injector

• a tube of silicone grease

• a Norprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)

• a C-flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)

• a C-flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)

• two hose fittings of each size:

– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)

– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)

– 6.4 mm x 3.2 mm (0.25 inch x 0.13 inch)

• two T-hose fittings of each size:

– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)

– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)

– 6.4 mm x 6.4 mm x 6.4 mm (0.25 inch x 0.25 inch x 0.25 inch)

• an interference suppressor for the current output

• 4 edge covers (version with GFR housing only)

• PTFE strip

• a quality certificate

• Operating Instructions (English).

2.2.2

Version CA71SI-A

The scope of delivery comprises:

• an analyzer with mains plug

• a cleaning injector

• a tube of silicone grease

• a Norprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)

• a Grifflex hose, length 2.0 m (6.6 ft), ID 19 mm (0.75 inch)

• a C-flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)

• a C-flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)

• two hose fittings of each size:

– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)

– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)

– 6.4 mm x 6.4 mm (0.25 inch x 0.25 inch, version without housing only)

• two T-hose fittings of each size:

– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)

– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)

• an interference suppressor for the current output

• a hose clamp

• 2 pipe clamps (version without housing only)

• a screwed socket for the outlet pipe

• 4 edge covers

• PTFE strip

• a quality certificate

• Operating Instructions (English).

Endress+Hauser 7

Identification Stamolys CA71SI

2.3

Certificates and approvals

2.3.1

4 approval

Declaration of conformity

The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives.

The manufacturer confirms successful testing of the product by affixing the

4 symbol.

2.3.2

Manufacturer certificate

Quality certificate

With the certificate the manufacturer confirms compliance with all technical regulations and the successful testing individually for your product.

8 Endress+Hauser

Stamolys CA71SI Installation

3 Installation

3.1

Incoming acceptance, transport, storage

• Make sure the packaging is undamaged!

Inform the supplier about any damage to the packaging.

Keep the damaged packaging until the matter has been settled.

• Make sure the contents are undamaged!

Inform the supplier about damage to the contents. Keep the damaged products until the matter has been settled.

• Check that the order is complete and agrees with your shipping documents.

• The packaging material used to store or to transport the product must provide shock protection and humidity protection. The original packaging offers the best protection. Also, keep to the approved ambient conditions (see "Technical data").

• If you have any questions, please contact your supplier or your local sales center.

3.2

Installation conditions

3.2.1

Design, dimensions

Version with stainless steel housing

Endress+Hauser

Fig. 2: Stainless steel version

a0001361

9

Installation

10

Stamolys CA71SI

Version with GFR housing, CA71SI-B

163 (6.42)

330 (13.0)

7-11/

0.28 - 0.43

(M 6)

200 (7.87)

326 (12.8)

360 (14.2)

436 (17.2)

465 (18.3)

500 (19.7)

Version with GFR housing, CA71SI-A

38 (1.5)

60 (2.36)

250 (9.84)

160 (6.30)

140 (5.51)

100(3.94)

Ø 26 (1.02)

Ø 16 (0.63)

120(4.72)

Ø 27 (1.06) mm(inch) a0001354-en

60 (2.36)

250 (9.84)

160 (6.30)

140 (5.51)

100(3.94)

330 (13.0)

7-11/

0.28 - 0.43

(M 6)

200 (7.87)

326 (12.8)

360 (14.2)

436 (17.2)

465 (18.3)

500 (19.7)

Ø 26(1.02)

Ø 16 (0.63)

120(4.72)

Ø 27 (1.06) mm (inch) a0001355

Endress+Hauser

Stamolys CA71SI

Open version

381 (15.0)

351 (13.8)

19 (0.75)

197 (7.76)

Installation

325 (12.8)

9 (0.35)

Fig. 5: Open version (without housing)

8 (0.31) mm (inch) a0001356

With the open version, you need an additional platform for the reagents. Mount this platform max.

35 cm (13.8 inch) below the pumps. The reagent bottels have the following dimensions: 90 x 90 x

215 mm (3.54 x 5.54 x 8.46 inch). The number of bottles varies from 2 to 5 depending on the analyzer version.

For these versions, the outlet pipe must be installed right of the analyzer.

The outlet pipe must be mounted to a wall so that the sample outlet hoses from the photometer have a gradient of 5 to 10 %. If neccessary, extend the hoses.

Endress+Hauser 11

Installation

12

Stamolys CA71SI

3.2.2

Connecting the sample line

27 (1.06)

145 (5.71)

60 (2.36)

4 x Ø 6.6 (0.26)

10 (0.39)

115 (4.53)

3 x Ø 28/20 (1.1/0.79)

Ø 34/25 (1.34/0.98)

4 x Ø 5.5/10.4 (0.22/0.41)

Abb. 6:

3

4

1

2

5

Collecting vessel at analyzer (optional)

Ventilation

Sample inlet from sampling

Collecting vessel

Electrical connections

Analyzer sample inlet

Ø 56 (2.2)

Ø 50H7 (1.97H0.28)

*

* a0001363

M20x1.5

Ø 24 (0.94) mm (inch)

*

Abb. 7:

7

8

*

6

Collecting vessel dimensions variable, freely adjustable dimensions

Sampling for analyzer

Sample overflow

Analyzer outlet

70 (2.76) a0001346

One-channel version

Collecting vessel (at analyzer, with or without level measurement)

Connection hose ID 3.2 mm (1/8 inch)

Customer collecting vessel

Connection

Max. distance from collecting vessel to analyzer

Max. height difference from collecting vessel to analyzer hose ID 1.6 mm (1/16 inch)

1 m (3.3 ft)

0.5 m (1.6 ft)

Two-channel version

• Depending on the ordered version, one or two collecting vessels (with or without level measurement) are included in the scope of delivery.

• Only one collecting vessel can be mounted at the housing. The second is to be placed nearby the analyzer.

3.3

Installation instructions

To install the analyzer at the intended location, proceed as follows:

1.

Mount the analyzer and secure it to a vibration-free wall using screws (Ø6 mm / 0.24"). The fastening clips on the housing make sure that the necessary distance to the wall is observed for ventilation purposes.

For the installation dimensions, please see the previous chapter.

2.

Using a spirit level, check that the cabinet is hanging straight. Only in this way can any air bubbles present escape from the cell.

3.

Fit the edge covers (with GFR housing only).

Endress+Hauser

Stamolys CA71SI Installation

4.

Lay the drain pipe for the reaction products. Where possible, use solid pipes (PE, internal diameter 1" with 3% incline).

5.

Screw the screw-in connector ID 16 into the outlet pipe from below. Secure the Grifflex hose

ID 19 to the nozzle with a hose clip (only CA71SI-A).

6.

Insert the valve hoses accordingly. These were partly removed from the valves for transportation. This prevents the hoses from sticking or pressure being applied to a point in the hose over an extended period.

CA71SI-B

→ å 8

CA71SI-A

→ å 9

1 2 3

V3 V1 V2

Fig. 8:

1

7

Valves and valve hoses CA71SI-B

V3 V1

6

2

V2

5

3

V4

a0006362

7 6 5 4

a0001730

1

2

Fig. 9:

V1-4

3

Valves and valve hoses CA71SI-A

Valves

Channel switchover

To sample pump

Connecting hose to valve 1, at rear

6

7

4

5

Outlet hose

Hose, valve 2 at front, standard

Hose, valve 2 at rear, cleaner

Hose, valve 1 at front, sample

7.

Secure the hose boxes in the appropriate pump holders:

Sample pump, left, reagent pump, right. The direction of flow of the sample and reagent must be counterclockwise.

R3

R2

R1

Fig. 10:

R1

R2

R3

Reagent pump, top view

Reagent 1

Reagent 2 (if present)

Reagent 3 (if present)

8.

Connect the sample supply.

a0002206

Endress+Hauser 13

Installation Stamolys CA71SI

The sample can be obtained as follows:

• Directly or via a reversible flow filter or a cross-flow filter by means of a small pump (rating approx. 300 ml/min), suitable for clear media

For questions regarding sample conditioning and its automation, please contact Service or the sales center responsible for your region.

9.

Connect the tubes from canisters containing reagents, standards and cleaner to the following nozzles:

Canister

Sample

Reagent 1

Reagent 2

Reagent 3

Standard solution

Cleaner

Tube designation (mark)

S

R

P

SI-1

SI-2

SI-3

3.4

Installation examples

• Permeate can contain air bubbles (CAT430) or is free of bubbles (customer-supplied ultrafiltration)

• Distance between the analysers as short as possible: sampling line between T-piece and the second analyser (

→ å 11, item 2)

shorter than 1.5 m

• Cross-section of sampling line ID 3.2 - 4 mm

• Only one sample receiver required

NOTICE

Measuring errors due to missing sample

► Ensure that there is always sufficient sample available for both analysers.

► Observe this when selecting maintenance intervals for CAT430 and when setting the buffer volume on the collecting vessel.

3

4

1

2

5

1

CA 71

2

Sample from CAT430

Sampling line

Sample receiver overflow

T-piece

Collecting vessel

3

5

4

3.5

Post-installation check

• After installation, check that all connections are fitted tightly and are leakage resistant.

• Ensure that the hoses cannot be removed without effort.

• Check all hoses for damage.

CA 71 a0001728

14 Endress+Hauser

Stamolys CA71SI Wiring

4 Wiring

!

WARNING

Device is energized

Improper connection can cause injury or death.

► The electrical connection must only be carried out by a certified electrician.

► Technical personnel must have read and understood the instructions in this manual and must adhere to them.

Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.

4.1

Electrical connection

4.1.1

Quick wiring guide

NOTICE

Reaching the terminal block and avoiding damage

► You have to fold out the frame of the analyzer to reach the terminal block.

► Beforehand, release the hoses from the outlet pipe. Otherwise there is a danger of flooding.

► Connect the hoses again after folding the frame back in.

Fold out the frame as follows:

1.

Release the two bottom Allen screws (AF 6) by 3-4 rotations (

→ å 12, item 1).

2.

Unscrew the two upper Allen screws completely so that the frame folds out. In this way, you reach the terminal block (item 2).

2

M E K RS 232

1 a0001370

Fig. 12:

1

2

Folding out the frame

Allen screws AF 6

Terminal strip

NOTICE

The device does not have a mains switch

► Install the device near to a fused socket which is easy to reach.

Endress+Hauser 15

Wiring Stamolys CA71SI

4.1.2

Terminal assignment

!

CAUTION

Shown diagram (

→ å 13) is an example

The terminal assignment and cable colors can deviate from the actual assignment and colors!

► Only use the terminal assignment of the sticker in the device (

→ å 14) to connect your

analyzer!

16

2

1

Fig. 14:

1

2

3

Analyzer from top (open version resp. folded out)

Connection department sticker

Printed circuit board with terminal strip

Backside of the analyzer

Schirm

Screen

Schirm

Shield n.b.

n.c.

n.b.

n.c.

Schirm

Shield n.b.

n.c.

WH keine Probe

No sample

Mess 2

Meas 2

Analog Out

0/4-20mA

Kanal 1

Channel 1

COM

NC

Kanal 2

Channel 2 n.b.

n.c.

n.b.

n.c.

NO

COM

NC

NO

7

6

5

N max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC

GN/YE 4

AC Motor

T 0.1A

Elektronik

Electronics

T 0.5A

3

2

1

L

N

3 a0001369 a0001372

Endress+Hauser

Stamolys CA71SI

Endress+Hauser

Wiring

Function Designation Terminal one channel

Mains

Alarm value 1, channel 1

Alarm value 2, channel 1

Alarm value 1, channel 2

Alarm value 2, channel 2

Fault

Reserve (unassigned terminals)

Analog output 1

0/4 ... 20 mA

Analog output 2

0/4 ... 20 mA

Sample conditioning remote control

Channel switch-over

1)

+

Screen

Input

0 V

Input

0 V

NC

NO

COM

NC

NO

+

Screen

NO

COM

NC

NO

COM

NC

NO

COM

L

N

PE

COM

NC

NO

COM

NC

Brass screw with bolt top right in the connection compartment (marked with

*

)

53

57

20

36

35

PE

1

18

17

22

21

19

26

24

23

28

27

1

25

3

2

38

PE

1)

57

53

55

53

22

21

20

36

35

PE

1

39

14

19

18

17

12

11

16

15

28

27

26

13

Terminal two channels

3

2

1

25

24

23

NOTICE

Terms for special terminals

► Limit switches 1 and 2 do not need to be connected if the PLC sets its own alarm values at the analog output.

► When using a sample conditioning system: Connect terminals 57 and 53 on the analyzer to the corresponding terminals on the sample conditioning system. For the allocation of these terminals see the sample conditioning system operating instructions.

► If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement (sample not ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.

► All external relay contacts have to be connected to 230 V.

17

Wiring Stamolys CA71SI

4.2

Signal connection

4.2.1

Screening of the analog outputs

The interference suppressor attenuates electromagnetic effects on control, power and signal lines.

After the connection of the data transfer cables clip the interference suppressor (in scope of delivery) on the cable cores (not on the outer insulation of the cable!). Place the cable screen out of the interference suppressor and connect it to PE (brass screw with bolt, top right in the connection

compartment) (

→ å 15).

4 3 a0001746

Fig. 15:

3

4

1

2

Interference protection of the signal cable

Cable screen (to PE )

Signal cable

Interference suppressor

Cable cores of the signal cable

With the two-channel version, place the cable cores of all cables (data cables to analog output 1 and to analog output 2) through the interference suppressor.

4.2.2

One-channel version

Connection

Signal inputs

Signal outputs

Designation

Leak

No sample

AV 1

AV 2

Fault

Measurement end l-1 channel 1

Analog output

Function

Liquid has collected in the drip pan

No sample available, measurement is not started, display flashes

Alarm value 1 exceeded or undershot

Alarm value 2 exceeded or undershot

Retrieves error message using operation menu

Displays "Measurement finished" (5 s)

0 or 4 mA = measuring range start

20 mA = measuring range end

18 Endress+Hauser

Stamolys CA71SI

4.2.3

Two-channel version

Connection

Signal inputs

Signal outputs

Analog output

Channel selection

Designation

Leak

No sample

AV 1-1

AV 2-1

AV 1-2

AV 2-2

Function

Liquid has collected in the drip pan

No sample available, measurement is not started, display flashes

Alarm value 1, channel 1 exceeded or undershot

Alarm value 2, channel 1 exceeded or undershot

Alarm value 1, channel 2 exceeded or undershot

Alarm value 2, channel 2 exceeded or undershot

Fault Retrieves error message using operation menu

Channel 1/2 or measurement end

1)

Displays active channel

Displays "Measurement finished" (5 s) l-1 channel 1 l-2 channel 2

0 or 4 mA = measuring range start

20 mA = measuring range end

0 or 4 mA = measuring range start

20 mA = measuring range end

Meas. 2 0 V = channel 1

24 V = channel 2

Wiring

4.3

Switching contacts

One-channel version

Connection

AV 1

AV 2

Fault

Unassigned

A:

R:

A:

R:

A:

R:

Terminal connection for condition fulfilled

25

25

-

-

23

24

28

28

19

19

-

-

-

-

26

27

17

18

22

16

13

-

-

-

20

14

11

A = NO current configured

R = NC current configured

A:

R:

A:

R:

A:

R:

Terminal connection for condition not fulfilled

25

25

-

-

24

23

28

28

19

19

-

-

-

-

27

26

18

17

22

16

13

-

-

-

21

15

12

Terminal connection for power off

25 24

28

19

22

16

13

-

-

-

-

-

27

18

21

15

12

Endress+Hauser 19

Wiring

20

Stamolys CA71SI

Two-channel version

Connection

AV 1 - 1

AV 1 - 2

AV 2 - 1

AV 2 - 2

Fault

Channel 1/2 measurement end

A:

R:

A = NO current configured

R = NC current configured

A:

R:

A:

R:

A:

R:

A:

R:

A:

R:

Terminal connection for condition fulfilled

25

25

-

-

23

24

13

13

28

28

-

-

-

-

11

12

26

27

16

16

19

19

22

22

-

-

-

-

-

-

14

15

17

18

20

21

A:

R:

A:

R:

A:

R:

A:

R:

A:

R:

A:

R:

Terminal connection for condition not fulfilled

25

25

-

-

24

23

13

13

28

28

-

-

-

-

12

11

27

26

16

16

19

19

22

22

-

-

-

-

-

-

15

14

18

17

21

20

Terminal connection for power off

25 24

13

28

16

19

22

-

-

-

-

-

12

27

15

18

21

Condition fulfilled means:

• AV 1: concentration > Alarm value 1

• AV 2: concentration > Alarm value 2

• Fault: error occurred

Contacts AV 1, AV 2 and fault are only affected during automatic operation.

4.4

Serial interface

RS 232 of CA71

SUB-D, nine-pin Function

3

2

8

TxD

RxD

CTS

5

Software protocol:

Output format:

Function

RxD

TxD

COM 1/2 at PC

RTS

CTS

GND GND

9600 baud, 8 data bits, 1 stop bit, no parity (9600, N, 8, 1)

ASCII

SUB-D, nine-pin

2

3

7

8

5

The results (measured value+unit of measure) are output in the "Data memory Measured values" menu.

The calibration results (measured value+unit of measure) are output in the "Data memory-

Calibration factors" menu.

A null modem cable is required (not a crossed one).

The analyzer does not have to be configured for the interface.

The following commands can be sent from the PC, in order to read out data:

• "D" = Data memory-Measured values

• "C" = Data memory-Calibration factors

• "S" = Setup (parameter entry, configuration...)

• "F" = Frequency (current)

Endress+Hauser

Stamolys CA71SI Wiring

4.5

Post-connection check

Carry out the following checks after electrical connection:

Device status and specifications

Is the analyser or cable externally undamaged?

Note

Visual inspection

Electrical connection

Does the supply voltage correspond to the data on the nameplate?

Note

230 V AC / 50 Hz

115 V AC / 60 Hz

Are current outputs screened and connected?

Are the mounted cables relieved of tension?

Cable type properly disconnected?

Guide power supply and signal lines separately over the entire travel distance.

Separate cable channels are ideal.

Cable routing without scuffing or cross-overs?

Are power supply and signal lines connected correctly according to wiring diagram?

Are all screw terminals tightened?

Are all cable entries mounted, tightened and leak-resistant?

Interference suppressor at the analogue output?

Current output simulation See procedure below

Current output simulation:

1.

Hold both arrow keys down (see "Display and operating elements" chapter) and connect the analyser to the mains or switch the mains switch on (if available). Wait until the display "0 mA" appears.

2.

Check on your PLC, PCS or you data logger whether the current value is the same.

3.

Press the

F

key. Browse to the next current values (4, 12, 20 mA, depending on the setting).

4.

Check that the respective current values are also on your PLC, PCS or your data logger.

5.

If the values are not there: a. Check the terminal assignment for analogue output 1 or 2.

b. Disconnect the analogue outputs from your PLC, PLS or your data logger and repeat steps

14 with the restriction that you measure the current values at the terminals of the analyser and not at the PLC, SPL or data logger.

If these current values are correct, please check your PLC, SPL, the data logger or the electrical cables.

Endress+Hauser 21

Operation Stamolys CA71SI

5

5.1

Operation

Display and operating elements

1 2

RS 232

M CE E K

4

Fig. 16: Display operating elements

3

4

1

2

LED (measured value)

LC display (measured value and status)

Serial interface RS 232

Operating keys and control LEDs

3

5.2

Local operation

The operating keys and the integrated indicator LEDs have the following functions:

Key Key function

H

– “Auto measuring” option

– back to the main menu from all sub-menus

K

– backwards in the sub-menu

(horizontal, see Appendix)

V

– backwards in the main menu (vertical)

– increase value

W

– forwards in the main menu (vertical)

– reduce value

F

– select option

– adopt value, forwards in the sub-menu (horizontal)

1)

J

– selection in the sub-menu

Indicator LED function

Alarm value 1 exceeded

Alarm value 2 exceeded

Measuring range exceeded

Measuring range undershot

Retrieve error message

Unassigned

V

or

W

and the

F

key simultaneously, you can set the digit after the decimal point.

a0001679

22 Endress+Hauser

Stamolys CA71SI Operation

5.2.1

Main menu

Access the main menu by holding down the

H

key until “AUTO MEASURING” is displayed.

For the main menu options and information about them, please see the following table.

Selection Display

AUTO MEASURING

Info

Calibration, measurement, flushing time-controlled actions

PARAMETER ENTRY

CONFIGURATION

LANGUAGE

Default settings for measuring ranges, alarm values, calibration, rinsing

Basic settings such as parameters, measuring units, arrangement of analog outputs and alarm values, date, time, offset values

Selecting menu language

ERROR DISPLAY

SERVICE

DATA MEMORY 1

DATA MEMORY 2

(Two-channel version only)

Displaying error messages

Manually switching valves and pumps

Last 1024 measured values channel 1

Last 1024 measured values channel 2

5.2.2

AUTO MEASURING

The actions "calibration", "measuring" and "flushing" are triggered by time-control.

The respective action is displayed in the LC display. The most recently registered concentration value is displayed until the end of the next measurement.

Otherwise, "wait" is displayed when

• the time of the first measurement has not yet been reached or

• the measuring interval has not yet expired.

"Measuring" flashes when the analyzer is ready for the next measurement but has not yet received the enable signal from the sample collector or the sample conditioning unit.

Endress+Hauser 23

Operation

Menu item

Code number

Photometer

Default settings

Measuring unit

Calibration factor

Concentration offset

Dilution

Delay to sample

Stamolys CA71SI

Depending on specification

SI-A

SI-B

Yes / no

μg/l / ppb

20 to 300 s

SI-A: 120 s

SI-B: 160 s

5.2.3

CONFIGURATION

NOTICE

Some settings that can be made in this menu affect the defaults in the PARAMETER

ENTRY menu.

► Complete the CONFIGURATION menu first during initial start-up.

Display Range of adjustment

(factory settings in bold)

03

0.10 to 10.00

0.00 to 50.0 mg/l

0.10 to 10.00

1.00

Info

Input 03.

If an incorrect code is entered the program exits the submenu.

The setting displays the parameter that is being measured.

This is defined by the product specification and set in this menu item at the factory. Do not change the value.

Otherwise, you will receive an "Incorrect photometer" error message.

If "yes" is selected, all the settings are reset to the factory settings.

You must set the current date and time (scroll through the menu, third-last point).

In the event of a reset, the date for the 1st calibration and the 1st flush is set to the following day.

The measuring unit selected depends on the type of photometer. This setting also affects the scope of the measuring range.

The calibration factor is the ratio of the measured concentration of the calibration standard to the predefined concentration of the standard (see

"PARAMETER ENTRY", calibration solution). The deviation results from factors such as reagent aging, aging of constructive components, etc.

The calibration factor compensates for these effects. CA

71 checks the registered calibration factor logically. If the factor is outside of the error tolerance, the calibration is automatically repeated. If the value is still outside the tolerance range after repeating calibration, an error message appears and the analyzer continues to work with the most recently registered, logically correct factor.

The last 100 calibration factors are stored in the memory with the date and time and can be retrieved by pressing the

J

key.

The calibration factor can be changed manually.

Optional dilution modules are taken into consideration in the calibration factor.

The offset gives users the option of adapting to an external comparison measurement.

(Change the sign with the

J

key.)

If the sample is diluted externally between taking the sample and the analyzer, the dilution factor has to be entered here.

If an optional dilution module is used, the factor remains

1. The dilution is taken into consideration by means of the calibration factor.

Dosing time for sample or standard. During this time, the entire system is flushed with sample or standard, so when the reagent is added there is definitely only fresh sample in the mixer. If sufficient sample is available, or in the event of an external collecting vessel with a long supply line, select the highest possible value.

24 Endress+Hauser

Stamolys CA71SI Operation

Menu item

Analog output 1

Analog output 2

Alarm value AV 1-1

Alarm value AV 2-1

Alarm value AV 1-2

Alarm value AV 2-2

Error contact

Current date/time

Calibrate offset

Range of adjustment

(factory settings in bold)

0 to 20 mA / 4 to 20 mA

NO

NC

Changes only take effect after a reset (power off/on)!

01.01.96 00:00 to 31.12.95 23:59

Yes / no

Display Info

Selection for the measuring range scope for channel 1. If the concentration measuring range is 0 to x mg/l, either

4 mA or 0 mA corresponds to 0 mg/l. The end of the measuring range is the same in both cases at 20 mA.

Only two-channel version!Selection for the measuring range scope for channel 2. The scopes of measuring range are independent of each other for channel 1 and channel

2 and are determined by the start of measuring range setting (channel 1 / channel 2) or end of measuring range setting (channel 1 / channel 2) in the PARAMETER

ENTRY menu.

Setting as to whether the contact for alarm value 1, channel 1 works as an NO or NC contact.

Setting as to whether the contact for alarm value 2, channel 1 works as an NO or NC contact.

Two-channel version only!

Setting as to whether the contact for alarm value 1, channel 2 works as an NO or NC contact.

Two-channel version only!

Setting as to whether the contact for alarm value 2, channel 2 works as an NO or NC contact.

Setting as to whether the contact for the fault message works as an NO or NC contact.

Setting the system clock.

Format DD.MM.YY hh:mm.

Frequency offset

1)

Pressing the

J

key starts blank value measurement for compensating the reagent's inherent color.

Manually changing the frequency offset.

1)

Frequency offset

1)

-5000 to +5000

0

1) Redetermine the frequency offset and the calibration factor every time the reagents are changed or the photometer is replaced. To do so, instead of sample connect deionized water to the sample inlet to determine the frequency offset (blank value). The value is generally between 0 and 10 Hz.

Endress+Hauser 25

Operation

Menu item

Measuring range start 1

Measuring range start 2

Measuring range end 1

Measuring range end 2

Alarm value

AV 1 - 1

Alarm value

AV 2 - 1

Alarm value

AV 1 - 2

Alarm value

AV 2 - 2

Time

1. measurement

5.2.4

PARAMETER ENTRY

Range of adjustment

(factory settings in bold)

SI-A: 1 to 200 μg/l / 0 μg/l

SI-B: 50 to 5000 μg/l / 0 μg/l

SI-A: 1 to 200 μg/l / 200 μg/l

SI-B: 50 to 5000 μg/l / 5000 μg/l

SI-A: 1 to 200 μg/l / 100 μg/l

SI-B: 50 to 5000 μg/l / 2500 μg/l

SI-A: 1 to 200 μg/l / 200 μg/l

SI-B: 50 to 5000 μg/l / 5000 μg/l

SI-A: 1 to 200 μg/l / 100 μg/l

SI-B: 50 to 5000 μg/l / 2500 μg/l

SI-A: 1 to 200 μg/l / 200 μg/l

SI-B: 50 to 5000 μg/l / 5000 μg/l

01.01.96 00:00 to 31.12.95 23:59

Measuring interval

Frequency of measurement

Channel 1

Frequency of measurement

Channel 2

2 to 120 minutes

10 min

0 to 9

1

2)

0 to 9

1

Display

Time

1. calibration

01.01.96 00:00 to 31.12.95 23:59

Calibration interval

0 to 720 h

48 h

Calibration solution

SI-A: 1 to 200 μg/l / 500 μg/l

SI-B: 50 to 5000 μg/l / 500 μg/l

26

Stamolys CA71SI

Info

The specified concentration is allocated a value of 0 or 4 mA at analog output 1

1)

.

Only two-channel version!The specified concentration is allocated a value of 0 or 4 mA at analog output 2.

The specified concentration is allocated a value of 20 mA at analog output 1.

Only two-channel version!The specified concentration is allocated a value of 20 mA at analog output 2.

Concentration threshold value for limit relay 1, channel 1

(switching hysteresis 2% of alarm value).

Concentration threshold value for limit relay 2, channel 1

(switching hysteresis 2% of alarm value).

Only two-channel version!Concentration threshold value limit relay 1, channel 2 (switching hysteresis 2% of alarm value).

Only two-channel version!Concentration threshold value limit relay 2, channel 2 (switching hysteresis 2% of alarm value).

Date format DD.MM.YY, time hh.mm. After each change the instrument does not wait for the measuring interval. If the measurement is to start immediately, set the time in the past.

Time between two measurements.

If the setting is 2 minutes, the measurements take place without any pauses.

Two-channel version only!

Number of measurements at channel 1 before switching to channel 2.

Two-channel version only!

Number of measurements at channel 2 before switching to channel 1.

Time of 1st calibration (DD.MM.YY, time hh.mm). After each change the instrument does not wait for the calibration interval. If the calibration is to start immediately, set the time in the past.

Analyzers are delivered pre-calibrated.

– Start the 1st calibration 2 hours after the initial startup at the earliest (warm-up phase)

– Set the time to 8:00 to track the effects of calibration in the curve.

– If you have started a calibration manually, you should redefine the time of the 1st calibration because the interval is calculated starting from the last calibration.

Time between two calibrations.

The "0 h" setting means that calibration does not take place.

Recommended: calibration interval of 48 to 72 h.

Concentration of the calibration standard.

Select a standard whose concentration is in the top third of the measuring range.

Endress+Hauser

Stamolys CA71SI Operation

Menu item Range of adjustment

(factory settings in bold)

Display Info

1. flushing

Flushing interval

Flushing hold on

01.01.96 00:00 to 31.12.95 23:59

0 to 720 h

0 h (SI-A)

48 h (SI-B)

0 to 60 s

0 s (SI-A)

60 s (SI-B)

Time of 1st flushing (DD.MM.YY, time hh.mm). After each change the instrument does not wait for the flushing interval. If the flushing is to start immediately, set the time in the past.

– Set the time to 4:00 to track the effects of flushing in the curve.

– If you have started flushing manually, you should redefine the time of the 1st flushing because the interval is calculated starting from the last flushing.

Time between two flushings.

The "0 h" setting means that cleaning no longer takes place.

Dwell time of flushing solution in the pump-mixerphotometer line.

Recommended: 30 to 60 s.

1)

2)

Option 0 or 4 mA: see "CONFIGURATION".

All channels set to 0 means that the channel is selected externally. All channels set to 1 means alternating, beginning with channel 1.

NOTICE

Measuring errors due to contamination by cleaning agents

► Always coordinate calibration and flushing.

► Carry out flushing with standard cleaners approx. 3-4 hours before the next calibration.

► Flushing with special cleaners (e.g. hydrochloric acid or ammonia) has a lasting effect on calibration. Therefore, carry out this cleaning 3-4 hours before calibration. Then rinse with deionized water afterwards.

5.2.5

LANGUAGE

The following languages are available:

• Deutsch

• English

• Français

• Suomi

• Polski

• Italiano.

5.2.6

ERROR DISPLAY

• This menu is a "read-only menu".

• You can find the individual error messages, their meaning and solutions to problems in chapter

"Trouble-shooting instructions".

• If there is at least one error message, the signal output is set to "fault".

• Causes of faults are requested for every measurement. If an error which occurred previously no longer exists, it is automatically cancelled.

Endress+Hauser 27

Operation

28

Stamolys CA71SI

5.2.7

SERVICE

NOTICE

In this menu, you can trigger direct function controls. If these controls are incorrectly combined, however, this could damage the analyzer.

► Using the Service menu requires detailed knowledge of how the analyzer works and operates.

► Exit the Service menu with the "M" key. All functions will be reset to the delivery state then.

Option Display

Pumps and valves

Signal frequency

Info

"Virtual switching board"

Various valve and pump combinations can be selected.

The setting options are:

– Valve 1:

P (sample) or S (standard)

– Valve 2:

S (standard) or C (cleaning agent)

– Valve 3 (Two-channel version only):

1 (channel 1) or 2 (channel 2)

– Valve 4 (for optical cell outlet, improves cleaning and avoids memory effects): s (stop) or g (go)

– Pump 1 (sample or standard/cleaner) and pump 2

(reagents): s (stop) or g (go)

– Mixture

The reagent and sample pumps can be switched on together, so that they run in the same ratio as for filling the sample-reagent mixture in measuring mode.

s (stop) and g (go)

P1 and P2 are inactiv, when G is at go. If P1 or

P2 is at go, G is not available.

The following valve combinations are possible:(applies to one-channel and two-channel version, whereby where the latter is concerned selection is made by positioning valve 3 between channel 1 and 2)

– V1: P, V2: S

Passage for the sample. This combination is automatically reset on leaving the service menu.

– V1: S, V2: S

Passage for standard solution

– V1: S, V2: R

Passage for cleaning agent

Signal frequency of the photometer

5.2.8

DATA STORAGE-Measured values

Two menus, "DATA MEMORY 1" and "DATA MEMORY 2", are only for the two-channel

version. In the one-channel version there is only one menu "DATA MEMORY".

Selection Display

Measured values

Serial output

Info

The data memory contains the last 1024 concentration measured values with date and time. If there are no values available, "Empty set" appears.

Browse through the data sets by pressing the

V

and

W keys.

You can output all data sets (in ASCII-format) via the serial interface. For this, the receiving end (PC) must be configured like this: 9600, N, 8, 1.

To send data, the receiving end (PC) must send the ASCII character 81 ("Shift", "D").

Endress+Hauser

Stamolys CA71SI

Endress+Hauser

Operation

Selection

Clear data

Display Info

This deletes all data sets.

5.2.9

DATA STORAGE-Calibration data

To enter this menu select the CONFIGURATION menu, browse to the "Calibration factor" option and press the

J

key.

Selection Display

Calibration factor

Serial output

only available via PC!

no display

Info

This data memory contains the last 100 calibration factors with date and time. If there are no values available,

"Empty set" appears.

Browse through the data sets by pressing the

V

and

W keys.

You can output all data sets (in ASCII-format) via the serial interface. For this, the receiving end (PC) must be configured like this: 9600, N, 8, 1.

To send data, the receiving end (PC) must send the ASCII character 81 ("Shift", "C").

Clear data This deletes all data sets.

5.3

Calibration

5.3.1

Standard calibration data

The signal strength is processed device-internally as a frequency.

The following table provides an overview of the standard calibration data.

Compare these values to your own data.

After changes in the CONFIGURATION menu and in case of software updates, you can check and, if necessary, change the calibration data in the sub-menu.

Silicate, lower measuring range

SI-A

Silicate, upper measuring range

SI-B

Measuring range

1 to 200 μg/l

Concentration [mg/l]

0.000

0.020

0.040

0.060

0.080

0.100

0.120

0.140

0.160

0.200

50 to 5000 μg/l 0.00

0.10

0.50

1.00

1.50

2.00

2.50

3.00

4.00

5.00

Frequency [Hz]

0

68

142

217

289

369

443

515

591

742

0

107

211

311

396

515

555

636

671

858

29

Operation Stamolys CA71SI

5.3.2

Calibration example

Complete calibration, consisting of two offset measurements and one standard measurement, is triggered via the automatic calibration mode. In addition, you can start offset and standard calibration manually via the CONFIGURATION menu and view the current calibration data.

Offset calibration with the CAY642-V10C00AAE standard

1.

Hold

H

down until AUTO MEASURING appears.

2.

Use

W

to move through the CONFIGURE menu and press

F

.

3.

Use

V

to enter a 3 in "Code No.?".

4.

Use

F

to go to the "Calibrate offset" option.

5.

Now use

J

to start the offset calibration. Wait until the offset value is displayed.

6.

Press

F again. This takes you back to measuring mode.

Calibration

Ensure that you have changed the reagents, filled the hoses again (no air bubbles) and the analyzer is in measuring mode.

1.

Hold

H

down until AUTO MEASURING appears.

2.

Use

W

to move through the PARAMETER ENTRY menu and press

F

.

3.

Use

F

to go to the "1st calibration" option.

4.

Now use the

W

or

V

and

F

keys to set a time which lies in the past.

5.

Press

F

to accept the value and then press

H

twice to return to the main menu (AUTO

MEASURING).

6.

Press

F again. This takes you back to measuring mode.

The calibration is now carried out automatically.

NOTICE

After the calibration has finished, the analyzer automatically switches to the measuring mode.

► You now have to set the time of the 1st calibration back into the future in order to align the calibration and rinsing times to each other. The rinse must be performed 3-4 hours before the next calibration.

► Proceed as described above to change the setting for the time of the 1st calibration. After changing to measuring mode, the analyzer automatically begins measuring, rinsing and calibrating at the defined times.

30 Endress+Hauser

Stamolys CA71SI Commissioning

6 Commissioning

6.1

Function check

!

CAUTION

Avoid damage due to incorrect power supply and inaccurate or insecure hose connections

► Check that all connections are secure. In particular, ensure that all hose connections are secure, so that no leaks occur.

► Ensure that the power supply voltage corresponds to the voltage specified on the nameplate.

6.2

Switch-on

6.2.1

Dry commissioning

When the analyzer has been configured and calibrated, the measuring cycle starts automatically.

Configuration is no longer necessary.

To perform initial commissioning or to readjust the device parameters, proceed as follows:

1.

Plug in the analyzer's plug with grounding contact into the socket.

2.

Press the

H

key until AUTO MEASURING is displayed.

• If possible, let the analyzer warm up in standby mode before commissioning ("Auto measuring" display). The time can be defined via the "1. measurement" option in the PARAMETER ENTRY menu.

• At the start of measurement with a cold analyzer, the first measuring results will be incorrect. The reaction is temperature-dependent and if the temperature is too low the predefined reaction time is insufficient for a complete reaction. For this reason, never carry out calibration with a cold analyzer. Wait at least two hours before carrying out calibration.

3.

Select the CONFIGURATION menu and configure up to and including "Current date/time".

With

H

you can return to the main menu.

4.

Now work through the PARAMETER ENTRY and SERVICE menus.

With

H

you can return to the main menu.

5.

Select CONFIGURATION again and use

F

to go to the "Calibrate offset" menu item

1)

.

6.

Connect a vessel containing the 0 μg/l standard CAY642-V10C00AAE or a plastic vessel containing ultrapure water to the "Sample" connection and start the frequency offset (

J

key).

The value determined is displayed and saved. Repeat the frequency offset measurement several times.

7.

Then reconnect the sample line.

With

H

you can return to the main menu.

The analyzer starts automatically (triggered by a control signal or integrated timer) with the

"Calibration", "Measurement" and "Flushing" procedures in accordance with the device parameters that you have set (1st calibration, 1st measurement, 1st flushing times and the respective intervals control the procedure).

1) Steps 5 and 6 are not required for version CA71SI-A.

Endress+Hauser 31

Commissioning

32

Stamolys CA71SI

Settings for CA71SI-B

Function

Measurement Flushing (sample)

Delay to sample

Stabilization

1. measurement (basis)

Fill mixture

Reaction

2nd measurement (plateau)

Flushing (sample)

Calibration Flushing (standard)

Delay to standard

Stabilization

1. measurement (basis)

Fill mixture

Reaction

2nd measurement (plateau)

Flushing (sample)

Flushing Pump cleaning solution

Allow to react

Pump cleaning solution

Settings for CA71SI-A

Function

Measurement Flushing (sample)

Delay to sample

Stabilization

1. measurement (basis)

Flushing (reag. line)

Fill mixture

Reaction

2nd measurement (plateau)

Empty cell

Flushing (sample)

Calibration Flushing (standard)

Delay to standard

Stabilization

1. measurement (basis)

Flushing (reag. line)

Fill mixture

Reaction

2nd measurement (plateau)

Empty cell

Flushing (sample)

Duration [s] Range of adjustment

120

20 to 999

8

CONFIGURATION / "Delay to sample"

SERVICE / "Pumps and valves"

17

120

30

120

20 to 999

8

17

120

CONFIGURATION / "Delay to sample"

SERVICE / "Pumps and valves"

30

½ Flush hold on

5

½ Flush hold on

PARAMETER ENTRY / "Flushing hold on"

Duration [s] Range of adjustment

160

20 to 999

8

CONFIGURATION / "Delay to sample"

2

24

90

SERVICE / "Pumps and valves"

30

160

20 to 999

8

2

24

90

CONFIGURATION / "Delay to sample"

SERVICE / "Pumps and valves"

30

6.2.2

Wet commissioning

Wet commissioning is different to dry commissioning in that in wet commissioning you fill the reagent lines before the automatic cycles are started.

Proceed as follows:

1.

Plug in the analyzer's plug with grounding contact into the socket.

2.

Press the

H

key until AUTO MEASURING is displayed.

3.

Select the SERVICE menu.

4.

Switch the P2 reagent pump "on" (select P2 with

F

and set to "g" with

V

) and leave it running until you can tell that there are reagents at the T-hose connector. After this, switch P2 "off" (s) again with

W

.

Endress+Hauser

Stamolys CA71SI Commissioning

5.

Now switch the valves to passage for standard (select V1: S, V2: S; with

F

or

K

switch to

"S" with

V

) and then switch the P1 sample pump "on". Leave the pump running until you can tell that there is standard at the T-hose connector. Switch P1 back "off".

6.

CA71SI-B only:

Now switch the valves to passage for cleaning agent (select V1: S, V2: R; with to "R" with

F

or

K

, switch

V

) and then switch the P1 sample pump "on". Leave the pump running until you can tell that there is cleaning agent at the T-hose connector. Switch P1 back "off".

7.

Now switch the valves to passage for sample (select V1: P, V2: S; with and "S" with

F

or

K

, switch to "P"

V

) and then switch the P1 sample pump "on". If you notice that sample is present at the T-hose connector, let the pump run for another 2 minutes. This removes any standard or cleaner residues.

Then switch P1 "off".

8.

Now proceed as with dry commissioning (from step 2).

6.2.3

Using a dilution module

You can extend the measuring range of your analyzer with the sample dilution module.

How it works:

• The analyzer determines the absorption of the sample at the specified wavelength.

It does not matter whether the sample was diluted previously or not.

• The characteristic of the photometer (--> standard calibration data) refers to the concentrations of the diluted sample and not to those of the (undiluted) original sample.

• The dilution factor is not taken into account until the measured value is calculated.

• The concentration information you specify for the measuring range and the alarm values in

PARAMETER ENTRY refer to the concentrations of the original sample.

You achieve maximum measuring accuracy if you also dilute the standard with the dilution module.

In the event of a 1:10 dilution, you have to use a 10 mg/l standard solution instead of a 1.0 mg/ l standard solution (without a dilution module) for example.

Balance changes to the dilution ratio (e.g. through hose aging) with the calibration. The calibration factor determined takes into account all deviations from the ideal state through reagents, the photometer and dilution.

Installing the dilution module

The dilution module is supplied ready to use.

Installation:

1.

Release the sample pump hose from the hose connector to valve 1 (

→ å 17, item 4).

2.

Connect the sample hose of the dilution module (black/black or yellow/blue, depending on the version) to the hose connector that is now free.

3.

Connect the T-section (item 1) of the dilution module to the input of the sample pump (hose box K1). Use the original sample pump hose for this purpose.

4.

Fit the dilution module onto the free hose box positions of the sample pump.

5.

Connect an outflow hose (O) to the free outlet of the T-piece (item 1).

Route this hose in the device in such a way that is guided at least 20 cm upwards and then into the open outlet. This prevents the hose from running dry.

NOTICE

Measuring errors due to malfunctioning outflow from photometer

Backwater, siphon effects or similar influence the sample outflow and cause measuring errors

► Do not connect the outflow hoses of the dilution module and the photometer.

Endress+Hauser 33

Commissioning Stamolys CA71SI

34 a0006622

2

3

4

O

P

W

1

K1

K2

K3

A

Hose box of the original sample pump

Hose box for additional sample (standard)

Hose box for dilution water

Sample supply to photometer

Overflow (in outlet of analyzer)

Sample hose (sample or standard)

Dilution water inlet (to be provided by the client)

T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)

Static mixer 122-012

T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)

Hose connector 3.2 x 1.6 mm (0.13 x 0.06 inch)

Changing parameters

The following section only lists the parameters where you have to make changes to the standard settings.

Start with CONFIGURATION.

CONFIGURATION

Delay to sample Increase the value of the standard setting by 40 seconds since additional sample volume is drawn in.

PARAMETER ENTRY

Measuring range start 1 / measuring range start 2

Measuring range end 1 / measuring range end 2

Alarm values AV

1-1 / 1-2 / 2-1 / 2-2

Calibration solution

Enter values that refer to the concentration of the undiluted sample.

If you are using a dilution module with the factor 10, for example, multiply the settings to date (without the dilution module) by 10.

Enter the original concentration (undiluted) of the standard here.

Endress+Hauser

Stamolys CA71SI Maintenance

7 Maintenance

!

CAUTION

You must not carry out any procedures not listed in the following chapters, yourself.

► Not listed works must only be carried out by the service.

If you use components for maintenance or modification, which are not approved by

Endress+Hauser, the warranty and the certified conformity of the device will become void.

7.1

Maintenance schedule

All maintenance duties that have to be carried out during normal operation of the analyzer are explained below.

If you are using a sample conditioning unit, e.g. CAT430, coordinate the maintenance work required for it with that of the analyzer. For this, read the maintenance chapter in the respective operating instructions.

Period of time

Weekly

Monthly

Every 3 months

Every 6 months

Duty Note

– Check and note calibration factor (for service purposes)

– Move valve hoses into their position and spray with silicone (extends the service life).

CONFIGURATION

– Flush sample line hose system with pressurized water (disposable syringe), check and replace reagents if necessary

– Remove the cassette of the sample pump.

– Grease pump hoses with silicone grease

– Check sample collector for fouling and clean it if necessary

– Connect the one-way syringe instead of the sample inlet.

– Cleaning the drain lines:

Flush all hoses and then rinse with sample for at least 30 minutes

– Rotate pump hoses

– Cleaning of the filter mat of the housing fan

– Replace pump hoses

– Replace valve hoses

– s. chapter "Cleaning the filter mat of the housing fan"

– see chapter "Replacing pump hoses"

– see chapter "Replacing valve hoses"

Whenever working on the reagent hoses, the hoses must be disconnected from the canisters, in order to prevent contamination of the reagents. Empty the reagent hoses via the service menu.

Endress+Hauser 35

Maintenance Stamolys CA71SI

7.2

Replacing reagents

!

CAUTION

Corrosive chemicals and other risks

Eye or skin injuries or crushes possible

► There is a danger of crushing limbs at doors, inserts and pump heads.

► Refer to the warning instructions in the safety data sheets when handling reagents or cleaning solutions. Wear protective clothing, gloves and goggles.

► If reagents come into contact with the skin or eyes, carefully rinse with copious amounts of water and consult a physician immediately.

► Never add water to reagents. Reagents containing acids may splash and heat may build up.

If you store the reagents correctly (in the dark, not over 20 ˚C) they will be stable for minimum

12 weeks from the date of manufacture (batch number). When this period of time has expired, the reagents must be replaced. Shelf life can be prolonged by keeping the reagents in a dark, cool storage place. The reagents absolutely must be replaced when:

• The reagents have been contaminated by sample (see "Trouble-shooting instructions")

• The reagents are too old

• The reagents have been spoilt by incorrect storage conditions or environmental influences.

7.2.1

Checking reagents

1.

Check the concentration of the standard in the laboratory. Adapt the values (PARAMETER

ENTRY, calibration solution") or replace the standard solution.

2.

Mix 10 ml of standard (c=500 μg/l or higher) and 5 ml of reagent SI 1 in a beaker.

3.

After approx. 5 minutes, add 5 ml each of reagent SI 2 and SI 3.

The finished mixture must be free of particles.

The reagents must be replaced if no visual coloration (blue) occurs after max. 2 minutes or if the mixture is not free of particles.

7.2.2

Replacing reagents

1.

Carefully remove the hoses from the canisters and wipe them with a clean and dry (paper) towel. Wear protective gloves when doing this.

2.

Switch on the reagent pump for about 5 seconds for the hoses to be drained.

3.

Replace the reagent canisters and feed the hoses into the new canisters.

4.

Fill the reagent hose with the new reagents (SERVICE). Switch all pumps to "g" for this purpose. Do not switch the pumps off ("s") until no more air bubbles are to be seen in the hoses.

5.

Connect standard CAY642-V10C00AAE to the sample flow and rinse the system (V1: P, P1: g and after 5 minutes P1: s).

6.

Exit the Service menu and carry out an offset calibration with CAY642-V10C00AAE and a calibration.

36 Endress+Hauser

Stamolys CA71SI Maintenance

7.3

Replacing pump hoses

!

CAUTION

Risk of splashing reagents when removing hoses from the hose nozzles

► Wear protective clothing, gloves and goggles.

The peristaltic pumps used for the analyzer convey the medium in a combination of vacuum and displacement pump. The pump rate is dependent on the elasticity of the pump hoses. Elasticity decreases and the pump rate drops as mechanical stress increases. Wear depends on mechanical stress (measuring interval, pump pressure). The wear effect can be compensated by periodical calibration. If the hose elasticity becomes too low and the pump rate is no longer reproducible, it is necessary to replace the hoses.

Removing the old hoses

1.

Remove the reagent hoses from the reagent canisters in order to prevent contamination of the reagents.

2.

Flush the old hoses with ultrapure water first and then with air to empty them (see SERVICE).

3.

Remove the hoses from the nipples of the hose boxes (item 5).

4.

Release the hose box(es) (1 to 3 per pump):

– Press against the lower retainer (item 3).

– You can now remove the hose box along with the pump hose.

– Remove the old hose from the box any dispose of it.

– Clean the hose box and the roller head (item 1) with a tissue.

1 2

3

4

1

2

5

Roller head

Upper retainer of hose box

Lower retainer of hose box

Guide at pump hose

Nipple with guide

5

3

4

a0001652

Installing the new hoses

1.

Grease the new pump hoses and the pump heads with silicone grease.

NOTICE

Danger of mix-up and resulting malfunction

► Make sure you connect the new pump hoses to the correct connections at the T-hose connector.

► For the order numbers of the pump hoses, see the chapter "Troubleshooting"/"Spare parts".

► The orange/orange Fluran hose must be installed with reagent R3.

2.

Fit the new hose on the hose box.

3.

First pull the hose downwards at both ends and then push the guide on the hose into that of the nipple on the hose box. Make sure it is seated correctly.

4.

First place the hose box into the upper retainer (item 2) of the pump and then press the box into the lower retainer (item 3).

Make sure the hose boxes are in the correct order in the pump (

→ å 19).

Endress+Hauser 37

Maintenance Stamolys CA71SI

R3

R2

R1 a0002206

Fig. 19:

R1

R2

R3

Reagent pump, top view

Reagent 1

Reagent 2 (if present)

Reagent 3 (if present)

5.

Reconnect the reagent hoses to the canisters.

6.

After installation, refill the hoses with sample, standard or cleaner (SERVICE)

2)

.

7.

Carry out an offset calibration and a calibration.

7.4

Replacing valve hoses

To replace the hoses, proceed as follows:

1.

Rinse the old hoses first with water and then with air to empty them (see SERVICE).

2.

Remove the hoses from the valves: a. You can disconnect the front hoses directly because the valves are open when deenergised.

b. To remove the back hoses, press the black button on the valve and disconnect the hoses.

3.

Grease the new hoses with silicone grease before inserting them.

4.

Install the new valve hoses in the reverse sequence of operations.

Ensure that the hoses are connected correctly.

5.

After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).

6.

Carry out an offset calibration and a calibration.

2) Make sure that the hoses are bubble-free when filled. If not, check the sample flow rate (see below).

38 Endress+Hauser

Stamolys CA71SI

Endress+Hauser

1

Maintenance

V1

2

V2

3

V4

V3

7

Fig. 20: Valves and valve hoses

5

6

3

4

7

1

2

V1-4 Valves 1, 2 and 4 (if present)

V3 Valve 3, two channel switch (optional)

Sample

To the pump

Y-piece, connecting hose to valve 1, behind it

Outlet hose

Hose valve 2, front, standard

Hose valve 2 back, cleaning agent

Hose valve 1 front, sample

6 5

7.5

Replacing the static mixer

To replace the mixer, proceed as follows:

1.

Rinse first with water and then with air (see SERVICE).

2.

Unscrew the four screws on the photometer housing and remove it.

3.

Disconnect the mixer from the photometer and from the T-piece below the photometer housing or release the mixer from the holder.

4.

Remove the old mixer and insert the new one.

5.

Connect the new mixer to the photometer and the T-piece again.

6.

Attach the photometer housing and screw it down.

7.

After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).

8.

Carry out an offset calibration and a calibration.

4 a0001730

7.6

Replacing the photometer optical cell

NOTICE

Electrostatic discharge (ESD) can cause damage to electronic assemblies

► Discharge yourself at a protective earth before handling the assemblies.

Removing the old cell

1.

Rinse first with water and then with air (see SERVICE).

2.

Switch the analyzer off.

3.

Unscrew the four screws on the photometer housing and remove it.

4.

Release the nuts from the guide screws and remove the photometer completely.

5.

Unscrew the four screws on the side of the photometer on which there is no ribbon cable.

6.

Take the photometer's electronics apart.

7.

Take out the cell and remove the hoses.

NOTICE

Traces of grease on the optical surfaces can cause corrupted measured values

► Do not under any circumstances touch the optical window of the cell with your fingers.

39

Maintenance Stamolys CA71SI

Installing the new cell

1.

Insert the new cell.

2.

Connect the cell to the hoses such that the sample is fed in from below.

3.

Secure the hoses with the supplied cable connectors to stop the cell from slipping.

4.

Reassemble the photometer and tighten the screws and nuts.

5.

Attach the photometer housing and screw it down.

6.

After installation, refill the hoses with sample, standard or cleaning agent (SERVICE).

7.

Carry out an calibration ("Calibration").

7.7

Cleaning

7.7.1

Housing

NOTICE

When cleaning, ensure that you do not damage the nameplate on the analyzer.

► Do not use any solvent-based cleaning agents.

To clean the analyzer housing, proceed as follows:

• Stainless steel housing (stainless steel SS 1.4301 (AISI 304)): with a lint-free cloth and Glittol RG 10.51

• GFR housing: with a damp cloth or with tenside-based (alkaline) cleaning agent.

7.7.2

Cleaning of the filter mat of the housing fan

The cover of the filter mat is at the backside of the analyzer (center, top).

1.

Remove the cover to replace the filter mat.

2.

To clean the filter mat you have various options: a. Wash the filter mat in water (approx. 40 ˚C) with a mild detergent if necessary.

b. Alternatively you can beat or vacuum the mat or blow it with compressed air.

c. In case of greasy substances: Wash the filter mat in warm water with grease solvent added.

NOTICE

Damage due to residues of organic solvents and incorrectly installed filter mat

Lost of the guaranteed protection degree due to missing, damaged or incorrectly installed filter mat

► If organic solvents are used, the mat has to be completely dried on air before re-assembled.

► Always take care to install the filter mat correctly.

40 Endress+Hauser

Stamolys CA71SI Maintenance

7.8

Placing out of service

You must place the analyzer out of service before shipping or before longer operation breaks (more than 5 days).

!

CAUTION

Residues of chemicals can cause injuries

► Thoroughly rinse all of the lines of the measuring system with ultrapure water.

To place the analyzer out of service, proceed as follows:

1.

Remove the reagent and standard hoses from the canisters and immerse them in a tank containing ultrapure water.

2.

Switch valve 1 to "Standard" and switch pumps 1 and 2 on for one minute (see SERVICE).

3.

Remove the hoses from the water and allow the pumps to run until the hoses are completely dry.

4.

If you are using a continuous sample supply, disconnect sampling line.

5.

Flush sampling hoses with ultrapure water and then with air, in order to completely empty the hoses.

6.

Remove the valve hoses from the valves.

7.

Remove the load from the pump hoses by removing the hose cassette from the bracket below.

Keep opened reagents and standards in a refrigerator. Observe the shelf-life.

Endress+Hauser 41

Accessories

42

Stamolys CA71SI

8 Accessories

In the following sections, you find the accessories available at the time of issue of this documentation.

For information on accessories that are not listed here, please contact your local service or sales representation.

8.1

Collecting vessel

– for sampling from pressurized systems

– results in an unpressurised continuous sample stream

• Collecting vessel without level measurement; order no. 51512088

• Collecting vessel with level monitoring for pure and ultrapure water, level monitoring with polypropylene float lever; Order no. C-A061019-50

8.2

Reagents, cleaner, standard solutions

• Reagent set active, 1 l SI1+SI2+SI3 each; order no. CAY643-V10AAE

• Reagent set, inactive, 1 l SI1+SI2+SI3 each; order no. CAY643-V10AAH

• Cleaning agent, 1 l; order no. CAY641-V10AAE

• Standard solution 0.0 μg/l SiO

2

• Standard solution 50 μg/l SiO

2

; order no. CAY642-V10C00AAE

; order no. CAY642-V10C50AAE

• Standard solution 100 μg/l SiO; order no. CAY642-V10C01AAE

• Standard solution 500 μg/l SiO

2

• Standard solution 1000 μg/l SiO

; order no. CAY642-V10C05AAE

2

; order no. CAY642-V10C10AAE

NOTICE

Reagents can cause environmental damage

► Note the disposal instructions in the safety data sheets of the reagents.

8.3

Maintenance kit

CAV740, maintenance kit for CA71

• Pump hoses

• Valve hoses

• Hose connectors

• Ordering acc. to product structure

CAV740-

For CA71 parameter

4

6

SI-B

SI-A

B

C

Inlet and outlet hoses

A not selected selected, for CA71SI-B selected, for CA71SI-A

complete order code

8.4

Hose cleaner

• Cleaning agent, alkaline, 250 ml (8.5 fl.oz.); order no. CAY746-V02AAE

• Cleaning agent, acidic, 250 ml (8.5 fl.oz.); order no. CAY747-V02AAE

Endress+Hauser

Stamolys CA71SI

8.5

Dilution module

Dilution module

1 set of pump hoses, 2 hose boxes, 1 set of hose connectors, static mixer

• Dilution 1:3

Order no. C-A030228-10

• Dilution 1:10

Order no. C-A030228-11

8.6

Additional accessories

• Interference suppressor for control, power and signal lines order no. 51512800

• Silicon grease, tube, 35 g order no. 71017654

• Valve set, 2 pieces, for two-channel version order no. 51512234

• Upgrade kit for upgrading from one-channel to two-channel version order no. 51512640

Accessories

Endress+Hauser 43

Trouble-shooting Stamolys CA71SI

9 Trouble-shooting

9.1

Trouble-shooting instructions

Although the analyser is not very prone to faults due to its simple assembly, problems can, of course, not be completely ruled out.

Possible errors, their causes and their possible remedies are listed below.

9.2

System error messages

Error message

Calibration failed

Cell dirty

Incorrect photometer

No sample

Spillage error

No measuring signal

Possible cause

Air bubbles in system Start calibration manually (PARAMETER ENTRY, "1. calibration", change the date accordingly, start measurement) or enter a new calibration factor.

Incorrect concentration of standard Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER

ENTRY", "Calibration solution") or replace the standard.

Reagents contaminated or aged Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.

Standard dosing defective

Tests and / or corrective measures

If a calibration fails, you can enter a new calibration factor manually (CONFIGURATION menu,

"Calibration factor"). Cancel the error message by switching the analyzer off and on briefly.

Check the valves for contamination, obstructions (visual inspection). Replace valve hoses if necessary.

Check the setting in the CONFIGURATION menu.

Incorrect photometer

Not enough light intensity at receiver

Incorrect photometer

No sample

Level measurement defective

Leak at canisters or hoses

– Flush with 12.5% bleaching lye.

– When using a CAT430:

Check the filter.

Check the setting in the CONFIGURATION menu, "Photometer".

Establish sample transfer.

Check level measurement at the sample collector.

Replace defective components and clean and dry the analyzer or those components affected by the leak.

Cell filled with air

Photometer defective

Electrical connection

Fuse defective

Pump sample for 1 minute (SERVICE).

Inform Service

Check all electrical connections and make sure the fuses are firmly seated.

Replace fuse F4 or F5 (semi-delay 0.2 A)

44 Endress+Hauser

Stamolys CA71SI Trouble-shooting

9.3

Process errors without messages

Error Possible cause Tests and / or corrective measures

Reagents contaminated or aged

Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.

Measured values always the same

No sample, no reagents Make sure sample and reagents are supplied, check level monitor and clean if necessary

System blocked Flush with 12.5% bleaching lye.

Check the hose from valve 4.

Incorrect concentration of standard

Reagents contaminated or aged

Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER ENTRY",

"Calibration solution") or replace the standard.

Simple check: Mix 5-10 ml of standard solution (of a high concentration) and 5 ml of reagent in a beaker. If it does not change color after max. 10 min, replace the reagents.

Measured values inaccurate

Reagent blank value too high

After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,

"Offset calibration")

Wrong dimension

Wrong cell

Check the setting in the CONFIGURATION menu, "Measuring unit" (e.g. ion instead of element).

Check the setting in the CONFIGURATION menu, "Photometer"

Sample suction time too short

Increase the suction time (CONFIGURATION, "Delay to sample")

Matrix effects

(substances that interfere with the photometric method)

Detect interfering substances (see Technical Information, "Measuring principle"), possibly use sample conditioning

Filter life too long

System blocked or contaminated

Take a counter sample at the analyzer inlet and check the concentration in the laboratory. If there is no deviation to the analyzer measured values, replace the ultrafiltration modules or reversible flow filters more frequently.

Flush with 12.5% bleaching lye.

Dosing

Cell dirty

Replace pump hoses.

Flush with 12.5% bleaching lye first and then with 5% hydrochloric acid

Measured values inaccurate

Counter sample in the laboratory delivers deviating measured values

Standard is dosed to the sample

Sample aging

Wrong size for analog output

Measured value transfer defective

Incorrect measuring range

Check the valves and valve settings. Replace valve hoses if necessary.

Shorten time between sampling and analysis.

Check the setting (CONFIGURATION, "Analog output 1" or " 2").

Adjust measuring range (PARAMETER ENTRY, "Measuring range")

Analyzer will not switch on

Analyzer is running but display is incomplete or off

Pumps fail to run

Measurement does not start

Background noise

No power

Fuse

Initialization failed

Leak

Leak sensor bypassed

Fuse

Pump defective

Leak at photometer

Check electrical cables for interference from strong sources of induction.

Check the electrical connection and ensure power supply.

Replace fuse F1 (slow-blow 0.5 A)

Switch the analyzer off and after approx. 30 seconds back on.

See "Spillage error" error message

Interrupt the contact between the two leak sensors (terminals 67-66)

Check all fuses and replace if necessary.

Service

Service

Endress+Hauser 45

Trouble-shooting

Error

"Measurement" display flashes

Possible cause

Time of 1st measurement not reached

Interval not expired

Time of 1st calibration not reached

Tests and / or corrective measures

The date must be between 01.01.1996 and the current date.

Change parameter setting.

The date must be between 01.01.1996 and the current date.

Calibration does not start Interval not expired or

0 h

Flushing does not start

Change parameter setting.

Leak at photometer

Time of 1st flushing not reached

Service

The date must be between 01.01.1996 and the current date.

Change parameter setting.

Leak at photometer

Blockage, deposits in device

Interval not expired or

0 h

Device or discharge blocked

Not enough sample conditioning

Remove blockage.

Service

Shorten the cleaning intervals of sample conditioning.

9.4

Spare parts

9.4.1

Overview for CA71SI-B

Stamolys CA71SI

46 a0006414

A

B

C

P

H

F

Sample input two-channel version

Standards and reagent canisters

Collecting vessel

Sample

Holder for dosage loop

Fluran hose, orange/orange (Kit CAV740)

R

R1

R2

R3

S

G

Cleaner canister

Reagent 1 canister

Reagent 2 canister

Reagent 3 canister

Standard solution canister

Coded pump hose (see Kit CAV740)

Endress+Hauser

Stamolys CA71SI

9.4.2

Overview for CA71SI-A

Trouble-shooting

Endress+Hauser a0006420

Fig. 22:

A

B

C

P

H

Overview of spare parts

Sample input two-channel version

Standards and reagent canisters

Collecting vessel

Sample

Holder for dosage loop

F

R1

R3

S

G

Fluran hose orange/orange (in Kit CAV740)

Reagent 1 canister

Reagent 3 canister

Standard solution canister

Coded pump hose (see Kit CAV740)

9.4.3

Spare parts for sample and reagent transportation

142

143

154

155

135

136

140

141

160

161

Item

120

121

122

123

130

131

134

Spare part

Norprene hose, 1.6 mm

C-Flex hose, 3.2 mm (permeate inlet and overflow with sample collector.)

C-Flex hose ID 6.4 mm

C-Flex hose, 1.5 mm

Hose connector 1.6 mm x 1.6 mm (10 pieces)

T-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)

Y-hose connector 1.6 mm x 1.6 mm x 1.6 mm (10 pieces)

Connection nipples for sample collector (10 piece)

T-hose connector 3.2 mm x 3.2 mm x 3.2 mm (10 pieces)

Mixer set, complete (2 pieces)

Collecting tank

Outlet pipe with hose nipple (2 pieces)

Kit CA71 Leak sensor for collecting tank (2 pieces)

Kit CA71 Valve, complete (1 piece for one-channel version)

Valve set for two-channel version

Roller head with holder for hose pump

Kit CA71 Hose box for pump (1 piece)

Order number

51504116

51504114

51504115

51512535

51506495

51506490

51512096

51512099

51516166

51512101

51512102

51512104

51512103

51512100

51512234

51512085

51512086

47

Trouble-shooting Stamolys CA71SI

Item

171

182

183

184

185

300

301

302

Spare part

Collecting vessel without level monitoring

Outlet pipe with hose nipple

Leak sensor for collecting tank

Mixer assembly

Drain valve

Kit CA T-hose connectors 1.6 mm x 0.8 mm x 1.6 mm

Kit CA hose connectors 3.2 mm x 1.6 mm

Kit CA hose connectors 1.6 mm x 0.8 mm

9.4.4

Analyzer spare parts

Item

H

144

200

Spare part

Dosing loop (H), complete

Photometer cell, 2 pieces

– SI-B

– SI-A

Photometer

– SI-A

– SI-B

Order number

51512088

51515578

51515581

51515579

51515580

71039848

51506491

71039845

Order number

51512463

51505778

51512409

51512078

51512079

9.5

Software history

Date Version Changes in the software

08/2006 6.0

Extension

• New parameter FE-D

06/2006 5.9

09/2005

06/2005

5.8

5.7

Extension

• New parameter NO-D

• Stepping motor position is also reset in a reset

• New stepping motor setting for CU-A/B (50 rpm)

• Spanish replaces Dutch and Hungarian

Extension

• New parameter PH-E

• Altered factory settings for SI-A and AM-B

• Characteristics for FE-B, CU-A and CR-A extended

• Measuring range limitation increased to 50000 for all parameters and engineering units

• Calibration value can be adjusted to end of measuring range

Improvement

• Stack overflow eliminated

• FE-C valve control corrected

• RTC properly dealt with when setting/reading the time

07/2004 5.6

Extension

• Range of adjustment for delay to sample to 999s

• Range of adjustment for cleaning duration to 300s

• Range of adjustment for frequencies to 5800 Hz

• Factory settings accepted as specified

• Altered measuring cycles for AM-D, AL-A (like hardness)

Documentation

BA360C/07/xx/07.06

BA355C/07/xx/07.06

BA356C/07/xx/07.06

-

BA353C/07/xx/09.04

BA357C/07/xx/10.04

48 Endress+Hauser

Stamolys CA71SI Trouble-shooting

Date Version

05/2004 5.5

Changes in the software

Extension/improvement

• Longer initialization

• Altered measuring cycle for SI

• Shortened reaction time, AM-C (90 s)

• SI-A: 0 Hz signal error only after 90 s

• Reason for crashing during frequency display in SERVICE eliminated

• Restart after power failure

03/2004 5.4

Extension/improvement

• New process: HA

• Menu extension: Cell drain time

• Calibration valve off during reaction

• Hungarian and Spanish no longer available

• New hardness engineering units

• New standard parameters/factory settings

• Stepping motor activation corrected

08/2003 5.3

08/2003

05/2003

12/2002

5.2

5.1

5.0

Improvement

• Fault elimination for measuring range switchover, HA

Extension/improvement

• Factory settings for SI changed

• CL characteristic changed

• New startup message

• Fault elimination for output for current output, two-channel

Extension/improvement

• New parameter CL

• Memory correction

• Time for "Delay to sample" is used for zero measurement

• Factory settings for MN, SI, HA changed

Original software

Documentation

BA364C/07/xx/06.04

BA361C/07/xx/02.04

-

-

BA354C/07/07.03

BA353C/07/xx/01.03

BA355C/07/xx/01.03

BA356C/07/xx/01.03

BA357C/07/xx/01.03

BA358C/07/xx/01.03

BA359C/07/xx/01.03

BA360C/07/xx/01.03

BA361C/07/xx/01.03

BA362C/07/xx/01.03

BA363C/07/xx/01.03

BA364C/07/xx/07.03

9.6

Return

Before returning observe the following items:

• Contact your regional Endress+Hauser sales center to receive detailed shipping conditions.

• Enclose the duly completed "Declaration of de-contamination" with the packaging.

You can find the form as master at the end of these instructions.

9.7

Disposal

The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.

Please observe local regulations.

Endress+Hauser 49

Technical data

Measured variable

Measuring range

Wavelength

Reference wavelength

Output signal

Signal on alarm

Load

Serial interface

Data logger

Load capacity

Supply voltage

Power consumption

Current consumption

Fuses

Stamolys CA71SI

10 Technical data

10.1

Input

SIO

2

[μg/l, mg/l]

SI-A

1 to 200 μg/l

SI-B

50 to 5000 μg/l

810 nm

565 nm (SI-B only)

10.2

Output

0/4 to 20 mA

Contacts: 2 limit contacts (per channel), 1 system alarm contact optional: end of measurement (with two channel version display of channel no. available) max. 500

Ω

RS 232 C

1024 data pairs per channel with date, time and measured value

100 data pairs with date, time and measured value for calibration factor determination (diagnostic tool)

230 V / 115 V AC max. 2 A, 30 V DC max. 1 A

10.3

Power supply

115 V AC / 230 V AC ±10%, 50/60 Hz approx. 50 VA approx. 0.2 A at 230 V approx. 0.5 A at 115 V

1 x slow-blow 0.5 A for electronics

2 x semi-delay 0.2 A for photometer

1 x slow-blow 0.1 A for motors

1 x slow-blow 1 A for fan

50 Endress+Hauser

Stamolys CA71SI Technical data

10.4

Performance characteristics

Time between two measurements

t mes

= reaction time + rinse time + waiting time + rinse again time + filling time + sampling time + reagent refusal time (min. waiting time = 0 min)

Maximum measured error 1 to 200 μg/l:

±2 % of measuring range end there of 5 to 20 μg/l:

±2 μg/l

50 to 5000 μg/l:

±2 % of measuring range end

Measuring interval

t mes

to 120 min

Reaction time

Sample requirement

6 minutes

• SI-A:

90 s

• SI-B:

120 s

20 ml (0.68 fl.oz.) per measurement

Reagent requirement

Calibration interval

SI-B: 3 x 0.18 ml (0.006 fl.oz.)

0.86 l (0.23 US.gal) per reagent per month with 15 minute measuring interval

SI-B: 3 x 0.18 ml (0.006 fl.oz.)

0.52 l (0.14 US.gal) per reagent per month with 15 minute measuring interval

0 to 720 h

Rinse interval (SI-B only)

0 to 720 h

Rinse again time

Filling time

30 s

SI-A

24 s

SI-B

17 s

Empy optical cell

Maintenance interval

Servicing requirement

t refusal

= 30 s (SI-A) t refusal

= 0 s (SI-B)

6 months (typical)

15 minutes per week (typical)

Endress+Hauser 51

Technical data

Mounting location

10.5

Installation conditions

Installation on a vibration-free wall

Ambient temperature

Humidity

Ingress protection

Design, dimensions

Weight

Material

10.6

Environment

5 to 35 ˚C (40 to 95 ˚F), avoid strong fluctuations below the condensation limit, installation in usual, clean rooms outdoor installation only possible with protective devices (customer supplied)

IP 54

10.7

Process

5 to 50 ˚C (40 to 120 ˚F)

Sample temperature

Sample flow rate

min. 5 ml (0.17 fl.oz.) per min

Consistency of the sample

low solid content (< 50 ppm)

Sample inlet

Unpressurized

10.8

Mechanical construction

see chapter "Installation"

GFR housing

Stainless steel housing

Without housing

Housing:

Front windows:

Endless hose:

Pump hose:

Valves: approx. 28 kg (62 lbs) approx. 33 kg (73 lbs) approx. 25 kg (55 lbs)

Stainless steel 1.4301 (AISI 304) or glass-fibre reinforced carbon(GFR)

Polycarbonate

®

C-Flex

®

, Norprene

®

Tygon

®

, Viton

®

Tygon

®

, silicone

Stamolys CA71SI

52 Endress+Hauser

Stamolys CA71SI Technical data

Endress+Hauser 53

Appendix

11 Appendix

11.1

Operating matrix

One-channel version

Stamolys CA71SI

54 a0001907-en

Endress+Hauser

Stamolys CA71SI Appendix

Endress+Hauser a0001908-en

55

Appendix

Two-channel version

Stamolys CA71SI

56 a0001909-en

Endress+Hauser

Stamolys CA71SI Appendix

Endress+Hauser a0001910-en

57

Appendix

11.2

Analyzer settings

Place:

Type:

Serial no. analyzer:

Serial no. Photometer:

Software version:

Date:

Photometer Type:

Measuring unit:

Calibration factor: c-offset:

Dilution:

Delay to sample:

Analog output:

AV 1:

AV 2:

Fault signal:

Frequency offset:

Ground line: (demineralised water without reagent)

• 0-20 mA

• normally closed

• normally closed

• normally closed

Measuring range start:

Measuring range end:

AV 1:

AV 2:

1st measuring:

Measuring interval:

1st calibration:

Calibration interval:

Calibration solution:

1st rinsing:

Rinse interval:

Rinse time:

• 4-20 mA

• normally open

• normally open

• normally open

• mg/l s h s

Hz

Hz

• mg/l

• mg/l

• mg/l

• mg/l min h

• mg/l

• μg/l

• μg/l

• μg/l

• μg/l

• μg/l

• μg/l

Stamolys CA71SI

58 Endress+Hauser

Stamolys CA71SI

11.3

Maintenance schedule

Appendix

Endress+Hauser a0001911-en

59

Stamolys CA71SI

Index

A

Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Alarm values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–26

Analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

AUTO MEASURING . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

C

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Calibration factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Calibration interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Certificates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Checking

Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Cleaner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Collecting vessel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 31

Dry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Wet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Concentration offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Connection

Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

D

DATA STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Dilution module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33, 43

Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

E

Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

ERROR DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Process errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

System errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

F

Filter mat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Flushing interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Frequency offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

G

GFR housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

60

H

Housing

GFR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Stainless steel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Without . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Housing fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

I

Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 9, 12

Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Interference suppressor . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

L

LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

M

Main menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Maintenance kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Menu

Auto Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Error display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Parameter entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

N

Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

O

Offset

Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Open version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Operating matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 22

Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

P

PARAMETER ENTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Performance characteristics. . . . . . . . . . . . . . . . . . . . . . . . 51

Photometer optical cell . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Placing out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Endress+Hauser

Stamolys CA71SI

Pump hoses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Q

Quality certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

R

Reaching the terminal block . . . . . . . . . . . . . . . . . . . . . . . 15

Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36, 42

Replacing

Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Photometer optical cell . . . . . . . . . . . . . . . . . . . . . . . . 39

Pump hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 49

S

Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Screening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

SERVICE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Signal connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Stainless steel housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Standard solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Switching contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

T

Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50–52

Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

U

Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

V

Valve hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Endress+Hauser 61

Stamolys CA71SI

62 Endress+Hauser

Declaration of Hazardous Material and De-Contamination

Erklärung zur Kontamination und Reinigung

RA No.

Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.

Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung.

Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the packaging.

Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene

"Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der

Verpackung an.

Type of instrument / sensor

Geräte-/Sensortyp ____________________________________________

Serial number

Seriennummer ________________________

Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen

Process data/ Prozessdaten

Temperature /

Conductivity /

Temperatur

Leitfähigkeit

_____ [°F]

_____

[°C]

________ [µS/cm]

Pressure

Viscosity

/

/

Druck

Viskosität

_____ [psi]

_____ [cp]

_____

_____

[ Pa ]

Medium and warnings

Warnhinweise zum Medium

Medium /concentration

Medium /Konzentration

Identification

CAS No.

flammable

entzündlich

toxic

giftig

corrosive

ätzend

harmful/ irritant

gesundheitsschädlich/ reizend

other *

sonstiges*

harmless

unbedenklich

Process medium

Medium im

Prozess

Medium for process cleaning

Medium zur

Prozessreinigung

Returned part cleaned with

Medium zur

Endreinigung

* explosive; oxidising; dangerous for the environment; biological risk; radioactive

* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv

Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.

Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.

Description of failure / Fehlerbeschreibung __________________________________________________________________________

______________________________________________________________________________________________________________

______________________________________________________________________________________________________________

Company data / Angaben zum Absender

Company /

Firma ___

________________________________

_________________________________________________

Address /

Adresse

_________________________________________________

_________________________________________________

Phone number of contact person /

Telefon-Nr. Ansprechpartner:

____________________________________________

Fax / E-Mail ____________________________________________

Your order No. /

Ihre Auftragsnr.

____________________________

“We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returned parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.”

“Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen w eiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringender Menge sind.”

( place, date / Ort, Datum) Name, dept./Abt.

(please print /

bitte Druckschrift

)

Signature / Unterschrift

BA364C/07/EN/14.11

FM9

www.endress.com/worldwide

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • - Measures silicate concentration in ultrapure water
  • - Photometric measurement principle
  • - Measuring range: 0.01 - 1000 µg/l SiO2
  • - Compact and wall-mountable design
  • - Easy-to-use interface with clear display
  • - Automatic calibration and cleaning functions
  • - Low maintenance and operating costs

Related manuals

Frequently Answers and Questions

What is the measuring range of the analyzer?
The measuring range is 0.01 - 1000 µg/l SiO2.
How often does the analyzer need to be calibrated?
The analyzer has automatic calibration functions, but the frequency of calibration may vary depending on the application.
What is the power supply requirement of the analyzer?
The power supply requirement is specified in the technical data section of the manual.
Can the analyzer be used in hazardous areas?
The manual does not provide information on the suitability of the analyzer for hazardous areas.

advertisement