Installation Stamolys CA71SI

The sample can be obtained as follows:

• Directly or via a reversible flow filter or a cross-flow filter by means of a small pump (rating approx. 300 ml/min), suitable for clear media

For questions regarding sample conditioning and its automation, please contact Service or the sales center responsible for your region.

9.

Connect the tubes from canisters containing reagents, standards and cleaner to the following nozzles:

Canister

Sample

Reagent 1

Reagent 2

Reagent 3

Standard solution

Cleaner

Tube designation (mark)

S

R

P

SI-1

SI-2

SI-3

3.4

Installation examples

• Permeate can contain air bubbles (CAT430) or is free of bubbles (customer-supplied ultrafiltration)

• Distance between the analysers as short as possible: sampling line between T-piece and the second analyser (

→ å 11, item 2)

shorter than 1.5 m

• Cross-section of sampling line ID 3.2 - 4 mm

• Only one sample receiver required

NOTICE

Measuring errors due to missing sample

► Ensure that there is always sufficient sample available for both analysers.

► Observe this when selecting maintenance intervals for CAT430 and when setting the buffer volume on the collecting vessel.

3

4

1

2

5

1

CA 71

2

Sample from CAT430

Sampling line

Sample receiver overflow

T-piece

Collecting vessel

3

5

4

3.5

Post-installation check

• After installation, check that all connections are fitted tightly and are leakage resistant.

• Ensure that the hoses cannot be removed without effort.

• Check all hoses for damage.

CA 71 a0001728

14 Endress+Hauser

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • - Measures silicate concentration in ultrapure water
  • - Photometric measurement principle
  • - Measuring range: 0.01 - 1000 µg/l SiO2
  • - Compact and wall-mountable design
  • - Easy-to-use interface with clear display
  • - Automatic calibration and cleaning functions
  • - Low maintenance and operating costs

Related manuals

Frequently Answers and Questions

What is the measuring range of the analyzer?
The measuring range is 0.01 - 1000 µg/l SiO2.
How often does the analyzer need to be calibrated?
The analyzer has automatic calibration functions, but the frequency of calibration may vary depending on the application.
What is the power supply requirement of the analyzer?
The power supply requirement is specified in the technical data section of the manual.
Can the analyzer be used in hazardous areas?
The manual does not provide information on the suitability of the analyzer for hazardous areas.

advertisement