1 The installation and components of equipment—————————————1

1.1 Instruction——————————————————————————1

1.2 Components—————————————————————————1

1.3 The main technical specifications—————————————————1

1.4 Components instruction—————————————————————2

2. Installation and adjustment————————————————————2

3. Operation methods and function instruction—————————————3

4. Sterilization——————————————————————————4

5. Contraindication————————————————————————4

6. Storage and maintenance—————————————————————4

7. Troubleshooting and notes————————————————————5

7.1 Troubleshooting————————————————————————5

7.2 Notice———————————————————————————7

8. Precaution——————————————————————————9

9. Transportation————————————————————————10

10. Working condition——————————————————————10

11. After service—————————————————————————10

12. Symbol instruction——————————————————————11

13. Environmental protection———————————————————12

14. Manufacturer’s right—————————————————————12

15. For technical data, please contact————————————————12

16. Declaration of conformity———————————————————13

16.1 Product conformity the following standards————————————13

16.2 EMC - Declaration of conformity————————————————13

17. Statement——————————————————————————16

1 The installation and components of equipment

1.1 Instruction

Guilin Woodpecker Medical Instrument Co., Ltd. is a professional manufacturer in researching, developing and producing ultrasonic scalers. The product is mainly used for teeth cleaning and also an indisensable equipment for tooth disease prevention and treatment.

The ultrasonic scaler UDS-J has scaling, perio functions with the following features:

1.1.1 Automatic frequency tracking ensures that the machine always works on the best frequency and more steadily.

1.1.2 Digitally controlled, easy operation and more efficient for scaling.

1.2 Components

1.2.1 The components of machine are listed in the packing list.

The scaling tips and other accessories are not listed in this instruction manual completely. The detail can be found in the instruction for tips and packing list.

1.2.2 Product performance and structure

Ultrasonic scaler is composed of electro circuit, water way and ultrasonic transducer.

1.2.3 Scope of application

Ultrasonic scaler UDS-J is used for the dental calculus elimination.

1.3 The main technical specifications

1.3.1 Power source input: 220-240V~ 50Hz/60Hz 150mA

1.3.2 Main unit input: 24V~ 50Hz/60Hz l.3A

13.3 Output power: 3W to 20W

1.3.4 Output tip vibration frequency: 30kHz±3kHz

1.3.5 Output half-excursion force: <2N

1.3.6 Output primary tip vibration excursion: ≤100μm

1.3.7 Main unit fuse: T1.6AL 250V

1.3.8 Power source fuse: T0.5AL 250V

1

1.3.9 Water pressure: 0.01MPa to 0.5MPa

1.3.10 Main unit weight: 0.8 kg

1.3.11 Power supply weight: 1.2kg

1.3.12 Operating mode: Continuous operation

1.3.13 Type of protection against electric shock: class II equipment

1.3.14 Degree of protection against electric shock: Type BF applied part

1.3.15 Degree of protection against harmful ingress of water: Ordinary equipment

(IPX0), the foot switch is drip-proof equipment (IPX1)

1.3.16 Applied part of the equipment: handpiece and tip

1.3.17 Degree of safety of application in the presence of a Flammable Anesthetic

Mixture with air, Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.

1.4 Components instruction

The components of the equipment are as showed in picture 1.

Picture 1

2. Installation and adjustment

2.1 Open the packing box, make sure that all the parts and accessories are

2

complete according to the packing list.

2.2 Take the main unit out of the box and put it on a stable plane, keep the main unit straight to the operator.

2.3 Turn the water control knob towards clockwise direction to the max and turn the potentiometer towards clockwise direction to a suitable position. [note2]

2.4 Insert the plug of the foot switch to its socket.

2.5 Connect one end of the water pipe to the water entrance, and the other end to the clean water source.

2.6 Connect the output end of power supply with main unit and get through to the power

2.7 Press the power switch of the main unit, then the power indicator shines.

3. Operation methods and function instruction

3.1 Direct the pit of the potentiometer at the "1" dial on the corer before turning on the scaler, make the main unit straight to the operator and turn the water control knob towards clockwise direction about three circles to the maximum.

3.2 The normal frequency is 30kHz±3kHz. With the high frequency , a light touch and a certain to-and-fro motion will eliminate the tartar without obvious heating. Overexertion and overstay are forbidden.

3.3 The way of assembly and disassembly of the scaling tips is as showed in picture 2.

3.4 The choice and operation methods of tips is shown in detail in attached materials of the equipment.

3.5 Vibrating intensity: Adjust the vibrating intensity as you need, generally turn the knob to the middle grade. According to patient's different sensitivity and the rigidity of the gingival tartar, to adjust the vibrating intensity during the clinical treatment.

3.6 Water volume adjust: Step on the foot switch, and

Picture 2:

Fasten the scaling tip by wrench with the handbetween the tip begins to vibrate, then turn the water control thumb and index finger.

knob to form spray, so as to cool down the handpiece and clean the teeth.

3

3.7 The handpiece can be handled in the same gestureas a pen in hand.

3.8 During the clinical treatment, be sure not to make the end of tip touch the teeth vertically and not to make the tip overexert on the surface of the teeth, in case of hurting the teeth and damaging the tip.

3.9 After finishing operation, keep the machine working for 30 seconds with the water supply in order to clean the handpiece and the scaling tip.

3.10 Unscrew the scaling tip and sterilize it.

4. Sterilization

4.1 All the scaling tips can be autoclaved.

4.2 Handpiece can be sterilized by any neutral sterilized liquid for cleaning and sterilizing. Do not sterilize under the high temperature and pressure.

4.3 The scaling tip and wrench can be cleaned by ultrasonic cleaner.

5. Contraindication

5.1 The patient who has hemophilia is not allowed to use this equipment.

5.2 The patient or doctor who with heart pacemaker is forbidden to use this equipment.

5.3 The heart disease patient, pregnant woman and children should be cautious to use the equipment.

6. Storage and maintenance

6.1 The equipment should be handled carefully and lightly. Be sure that it is far from the vibration, and is installed or kept in a cool, dry and ventilated place.

6.2 Don't store the machine together with the articles that are combustible poisonous, caustic, or explosive.

6.3 This equipment should be stored in a room where the relative humidity is ≤

80%, atmospheric pressure is 50kPa to106kPa, and the temperature is -10℃ to

+50℃.

6.4 Please turn off the power switch and pull out the power plug when the

4

equipment is not used. If the machine is not used for a long time, please make it get through to the power and water once per month for five minutes.

7. Troubleshooting and notes

7.1 Troubleshooting

Fault

The scaling tip doesn’t vibrate and there is no water flowing out when The foot switch is in loose Insert the foot switch to its stepping on the foot loose contact.

contact.

socket well.

socket tightly.

switch.

Possible

The power line plug is in

Solutions

Make the plug insert to the

The fuse of transformer is Open the power box, change broken.

The fuse in the main unit is broken.

a new T0.5AL250V fuse.

Take off the cover, change a new T1.6AL250V fuse.

The scaling tip doesn’t vibrate but there is

The tip hasn’t been screwed on the handpiece

Screw the tip on the handpiece tightly (picture 2).

water flowing out when tightly.

stepping on the foot The connect plug between Contact our dealers or us.

switch.

The handpiece generates heat.

the handpiece and the circuit board is in loose contact.

Something wrong with the handpiece

The water control knob is in a low grade.

The potentiometer is damaged [note 2].

Contact our dealers or us

[note 1].

Turn on the water control knob to a higher grade [note

2].

Change a new one.

5

Fault Possible Solutions

The scaling tip vibrates but there is no fine

The water control knob is turn off.

Turn on the water control knob [note 2].

spray when stepping on the foot switch.

There is impurity in the solenoid valve.

Clean inside of solenoid valve ( picture 5).

The water pipe is blocked. Clean the water pipe by

There is still water flowing out after the power is off.

The amount of spouting water is too little.

There is impurity in the solenoid valve.

The water control knob is in a low grade.

multi-function syringe [note

3].

Clean inside of solenoid valve (picture 5).

Turn on the water control knob to a higher grade [note

2].

The water pressure is not high enough.

Make the water pressure higher.

The water pipe is blocked Clean the water pipe by multi-function syringe

The vibration of the tip becomes weak.

The tip hasn't been screwed on to the handpiece tightly.

[note2].

Screw the tip on the handpiece tightly (picture 2).

The vibration is too strong and the potentiometer is failure.

The tip is vibrated loose. Screw the tip on the handpiece tightly (picture 2).

The tip is damaged

[ note3 ].

Change a new one.

The potentiometer is damaged [note 2].

Change a new one.

If the troubles still can't be solved, please contact with the local distributors or our company.

6

7.2 Notice

[ Note 1 ] The disassembly of the handpiece (As showed in picture 3): a) Remove the screw from the cover, pull out the potentiometer vertically, then take off the cover from the end of the machine lightly (there is a line connecting the cover and the machine, don't use too much strength). The inner of the machine is as showed in picture 3.

Picture 3 The disassemblt of handpiece b) Pull out the water pipe in the handpiece cable from the coupling between the water control knob and the water pipe.

c) Pull out the lead plug from circuit board and untie the string.

d) Hold the joint of handpiece cable and the main unit and push it into the main unit about lcm, then pull out the cutting sleeve from the cable.

e) Take off the handpiece from the main unit, and the disassembly is finished.

The assembly of handpiece is on the contrary. Be sure not to assemble the lead

7

plug in a wrong direction, otherwise the tip will be electriferous.

Check-up method: get through to the power, step on the foot switch, and checkup the scaling tip with electric pen. If the electric pen shines, the scaling tip is electriferous. Turn off the machine and then insert the lead plug correctly.

[ Note 2] Turn the water control knob towards anticlockwise direction, when the knob can't be turned any more, it comes to the min. On the contrary direction, the water volume increases step by step till the knob is back-out. The grade of the potentiometer is from gear l to 9. The ninth grade is the max. Be sure not to overdo.

[ Note 3] To clean the water pipe with the multi-function syringe of the dental unit

(as showed in the picture 4): a) Snip the water pipe at a distance of 10cm to 15cm from the water entrance.

b) Turn on the power switch, get through to the power.

c) Connect the multi-function syringe of the dental unit to the water pipe.

d) Step on the foot switch.

e) Turn on the switch of the multi-function syringe, press the air or water into the water pipe in the machine, then eliminate the impurity in the water pipe.

Picture 4 Multi-fuction syringe

[ Note 4] If the scaling tip has been screwed on tightly and there is fine spray too, the following phenomena shows that the scaling tip was damaged: a) The vibrating intensity and the pulverization degree become weak obviously.

b) When operating, there is some buzz when the scaling tip is working.

8

8. Precaution

8.1 Notice when using equipment

8.2 Keep the scaler clean before and after operation.

8.3 The scaling tip, wrench and handpiece must be sterilized before each treatment.

8.4 Don't screw the scaling tip when stepping on the foot switch.

8.5 The scaling tip must be fastened. There must be fine spray coming out from the tip when operating.

8.6 Change a new one when the tip is damaged or worn excessively.

8.7 While scaler working ,the heat of scaling tip will become higher if there is no water flowing out.Please keep the water flow smoothly.

8.8 Don't twist or rub the tip.

8.9 Don't use impure water source, and be sure not to use normal brine instead of pure water source.

8.10 If use the water source without hydraulic pressure, the water surface should be one meter higher than the head of the patient.

8.11 After operating, turn off electrical source, and then pull out the plug.

8.12 As a professional manufacturer of medical instruments, we are only responsible for the safety on the following conditions:

8.12.1 The maintenance, repair and modification are made by the manufacturer or the authorized dealer.

8.12.2 The changed components are original of "WOODPECKER" and operated correctly according to instruction manual.

8.1.13 Please put the power plug into the socket easy to pull out, to make sure it can be pull out in emergency.

8.1.14 The power supply is considered as a part of ME equipment.This device can only be equipped with the special power supply of Guilin Woodpecker Medical

Instrument Co., Ltd.

8.1.15 The power supply is NOT waterproof. Please keep it dry and away from the water.

8.16 The screw thread of the scaling tips produced by other manufacturers maybe coarse, rusty and collapsed, which will damage the screw thread of the handpiece irretrievably. Please use "WOODPECKER" brand scaling tip.

9

9. Transportation

9.1 Excessive impact and shake should be prevented in transportation. Lay it carefully and lightly and don't invert it.

9.2 Don't put it together with dangerous goods during transportation.

9.3 Avoid solarization and getting wet in rain or snow during transportation.

10. Working condition

Environment temperature: +5℃ to +40℃

Relative humidity: ≤80%

Atmosphere pressure: 70kPa to106kPa

11. After service

We offer one year free repair to the equipment according to the warranty card.

The repair of the equipment should be carried out by professional technician. We are not responsible for any irretrievable damage caused by the not professional person.

Note: "P" was put on the valve seat to designate the water entrance.

Picture 5 The assembly and disassembly of the solenoid valve

10

12. Symbol instruction

Trademark

Consult the accompanying documents

Date of manufacture

Manufacturer

Class II equipment

Type BF applied part

Ordinary equipment

Drip-proof

Used indoor only

Appliance compliance WEEE directive

Alternating current

~24V power supply input interface

Foot switch interface

Water control knob

Atmospheric pressure for storage

Temperature limitation

11

Humidity limitation

Water entrance pressure 0.01MPa-0.5MPa

CE marked product

FDA marked product

Authorised Representative in the EUROPEAN COMMUNITY

Got the quality management system certification and CE certification issued by TüV Rheinland

13. Environmental protection

There are no harmful factors in our product. You can deal with it based on the local law.

14. Manufacturer’s right

We reserve the rights to change the design of the equipment, the technique, fittings, the instruction manual and the content of the original packing list at any time without notice. If there are some differences between blueprint and real equipment, take the real equipment as the norm.

15. For technical data, please contact

12

16. Declaration of conformity

16.1 Product conformity the following standards

EN 60601-1:2006 EN ISO 9687:1995

EN 60601-1-2:2007

EN 61000-3-2:2006

EN 61000-3-3:2008

EN 60601-1-4:1996

EN 1041:2008

EN ISO 14971:2009

EN ISO 7405:2008

EN 60601-1-6:2007

EN 61205:1994

EN ISO 22374:2005

EN ISO 17664:2004

EN ISO 17665-1:2006

EN ISO 10993-1:2009

EN ISO 10993-5:2009

EN 62304:2006

EN 980:2008

EN ISO 10993-10:2010

16.2 EMC - Declaration of conformity

13

14

15

The device has been tested and homologated in accordance with EN 60601-1-2 for EMC. This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment.

17. Statement

All rights of modifying the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD.

The industrial design, inner structure, etc, have claimed for several patents by