advertisement
Recommended separation distances between portable and mobile RF communications equipment and the RadiForce Series
TheRadiForceseriesisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbances arecontrolled.ThecustomerortheuseroftheRadiForceseriescanhelppreventelectromagneticinterferenceby maintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andthe
RadiForceseriesasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.
Ratedmaximumoutput poweroftransmitter
0.01
0.1
1
10
100
W
150kHzto80MHz d=1.2√ P
0.12
0.38
1.2
3.8
12
Separationdistanceaccordingtofrequencyoftransmitter
80MHzto800MHz d=1.2√ P
0.12
0.38
1.2
3.8
12 m
800MHzto2.5GHz
d=2.3√ P
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance “d” in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where “P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE2 Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionand reflectionfromstructures,objectsandpeople.
Cable length
PowerCord:
SignalCable(PP200):
SignalCable(DD200-DL):
SignalCable(FD-C39):
USBCable(MD-C93):
Accessary
Accessary
Accessary(RX340)
Accessary(RX240)
Accessary
2.0m
2.0m
2.0m
2.0m
1.8m
50
Appendix
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Key Features
- Medical 53.9 cm (21.2")
- 1536 x 2048 pixels
- 22 ms 1000 cd/m² 1400:1
- 56 W
Related manuals
advertisement
Table of contents
- 39 Cover
- 41 Notice for this monitor
- 42 Cleaning
- 42 To use the monitor comfortably
- 43 CONTENTS
- 45 Chapter 1 Introduction
- 45 1-1. Features
- 46 1-2. EIZO LCD Utility Disk
- 46 Diskcontentsandsoftwareoverview
- 46 forMedical
- 47 1-3. Basic Operations and Functions
- 47 BasicAdjustmentmenuoperations
- 48 Functions
- 49 Chapter 2 Adjusting Screens
- 49 2-1. Compatible Resolutions
- 49 2-2. Setting the Resolution
- 49 Windows
- 49 WindowsVista
- 49 WindowsXP
- 49 MacOSX
- 50 2-3. Adjusting Color
- 50 mode)
- 50 Toperformadvancedadjustments
- 51 Adjustablesettingsineachmode
- 51 Toadjustbrightness
- 52 Toadjustcontrast
- 52 Toadjustcolortemperature
- 53 Toadjustgamma
- 53 Toadjusthue
- 53 Toadjustsaturation
- 54 Toadjustgain
- 54 displayarea
- 55 Tosetthemodeinthebasedisplayarea
- 55 displayarea
- 56 2-4. Selecting Screen Size
- 57 Chapter 3 Setting Monitor
- 57 3-1. Setting Orientation
- 57 3-2. Enabling/Disabling Mode Selection
- 58 3-3. Setting Language
- 58 3-4. Setting Adjustment Menu Display
- 58 3-5. Setting the EIZO Logo Display
- 59 3-6. Locking Buttons
- 59 3-8. Enabling/Disabling DC5V Output
- 60 3-9. Return To Default Setting
- 60 Toresetthecoloradjustments
- 60 Toresetallsettings
- 61 Chapter 4 Controlling Monitor Quality
- 61 4-1. Executing Calibration
- 61 ExecutingSelfCalibration
- 61 ViewingSelfCalibrationresult
- 62 4-2. Checking Grayscale
- 62 ExecutingGrayscaleCheck
- 62 ViewingGrayscaleCheckresult
- 62 4-3. Acquiring Life Prediction Data
- 63 History