Platinum Series XL, 5, 10 Operator’s Manual


Add to my manuals
32 Pages

advertisement

Platinum Series XL, 5, 10 Operator’s Manual | Manualzz
Operator’s Manual
Platinum™ Series
XL, 5, 10
Oxygen Concentrators Standard,
with
SensO2®and HomeFill®
Invacare 5 with SensO2
Model IRC5LXO2
Invacare 5
Model IRC5LX
Invacare 10 with SensO2
Model IRC10LXO2
Invacare 10
Model IRC10LX
Dealer: This manual MUST be given to the end
user.
User: Before using this product, read this manual
and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
WARNING
DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment ‐ otherwise, injury or damage may occur.
ACCESSORIES WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.
Platinum™ Series
2
Part No 1118353
TABLE OF CONTENTS
TABLE OF CONTENTS
ACCESSORIES .................................................. 2
REGISTER YOUR PRODUCT .............................. 4
SPECIAL NOTES ............................................... 5
LABEL LOCATION ............................................ 7
SECTION 1—GENERAL GUIDELINES .................. 8
To Reduce the Risk of Burns, Electrocution, Fire or
Injury to Persons ................................................................................. 8
Operating Information ....................................................................... 9
Radio Frequency Interference.......................................................... 9
SECTION 2—FEATURES .................................. 11
SECTION 3—HANDLING ................................. 12
Unpacking............................................................................................ 12
Inspection ............................................................................................ 12
Storage................................................................................................. 12
SECTION 4—TYPICAL PRODUCT
PARAMETERS ............................................... 13
SECTION 5—OPERATING INSTRUCTIONS ....... 16
Introduction........................................................................................ 16
Select a Location ............................................................................... 16
Set-Up .................................................................................................. 17
Plug in Power Cord...................................................................... 17
Connect Humidifier (If So Prescribed) .................................... 17
Power Switch ..................................................................................... 19
Flowrate............................................................................................... 20
Front Panel Indicators - Units with SensO2 ............................... 21
Initial Startup of the Concentrator........................................... 21
Front Panel Indicators - Units without SensO2 ......................... 22
Initial Startup of the Concentrator........................................... 22
Part No 1118353
3
Platinum™ Series
TABLE OF CONTENTS
TABLE OF CONTENTS
SECTION 6—MAINTENANCE .......................... 23
Routine Maintenance........................................................................ 23
Cleaning the Cabinet Filter......................................................... 23
Cleaning the Cabinet ................................................................... 24
Cleaning the Humidifier .............................................................. 24
SECTION 7—TROUBLESHOOTING GUIDE ......... 26
NOTES .......................................................... 28
LIMITED WARRANTY .................................... 30
REGISTER YOUR
PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips
and industry news.
Register ONLINE at
warranty.invacare.com
Please have your model number and purchase date available to
complete your registration.
Any registration information you submit will only be used by
Invacare Corporation and protected as required by applicable laws
and regulations.
Platinum™ Series
4
Part No 1118353
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.
SIGNAL WORD
MEANING
DANGER
Danger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
WARNING
Warning indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
CAUTION
Caution indicates a potentially hazardous situation
which, if not avoided, may result in property
damage or minor injury or both.
NOTICE
The information contained in this document is subject to change without notice.
DANGER
DO NOT SMOKE while using this device. Keep all
matches, lighted cigarettes or other sources of ignition
out of the room in which this product is located and away
from where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
Part No 1118353
5
Platinum™ Series
SPECIAL NOTES
CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or mechanical failure. Consult your physician or equipment provider for the type of reserve system required.
This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.
Platinum™ Series
6
Part No 1118353
LABEL LOCATION
LABEL LOCATION
IRC5LXO2 and IRC10LXO2
IRC5LX and IRC10LX
HomeFill II Compatib
Compatible
le
HomeFill II Compatible
with
Operating Instructions
Alarms
SensO2™
Operating Instructions
(See Operator's Manual)
(See Operator's Manual)
I/O
Alarms
(See Operator's Manual)
(See Operator's Manual)
WARNING - RISK OF FIRE
WARNING - RISK OF FIRE
NO SMOKING - Keep ALL sources of ignition out of the room in
which this product is located and away from areas where oxygen
is being delivered. Textiles, oils and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
NO SMOKING - Keep ALL sources of ignition out of the room in
which this product is located and away from areas where oxygen
is being delivered. Textiles, oils and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
CAUTION - RISK OF ELECTRIC SHOCK
CAUTION - RISK OF ELECTRIC SHOCK
Do not remove cover. Refer servicing to qualified service
personnel.
Do not remove cover. Refer servicing to qualified service
personnel.
Flow rates below 0.5 L/min. will trigger Low Flow Alarm
(rapid audible alarm).
Flow rates below 0.5 L/min. will trigger Low Flow Alarm
(rapid audible alarm).
Set flowmeter at 3 L/min. or less when filling
HomeFill II cylinders.
Set flowmeter at 3 L/min. or less when filling
HomeFill II cylinders
cylinders..
CAUTION: Federal (USA) law restricts this
device to sale by or on the order of a physician.
CAUTION: Federal (USA) law restricts this
device to sale by or on the order of a physician.
P/N 1108755
P/N 1108820
Serial Number Label is located on
the resonator intake assembly
Specification
Label is located
on the back near
the bottom
Part No 1118353
7
Platinum™Series
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed.
WARNING
This section contains important information for the safe operation and use of this product. To Reduce the Risk of Burns,
Electrocution, Fire or Injury to Persons
DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing
to qualified service personnel. No user serviceable parts.
Avoid using while bathing. If continuous usage is required
by the physician’s prescription, the concentrator MUST be
located in another room at least 7 ft from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into
water or other liquid.
DO NOT reach for product that has fallen into water.
Unplug IMMEDIATELY.
If the concentrator has a damaged cord or plug, if it is not
working properly, if it has been dropped or damaged, or
dropped into water, call qualified technician for
examination and repair.
A spontaneous and violent ignition may occur if oil, grease
or greasy substances come in contact with oxygen under
pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all
other oxygen equipment. DO NOT use any lubricants
unless recommended by Invacare.
Avoid creation of any spark near medical oxygen
equipment. This includes sparks from static electricity
created by any type of friction.
Platinum™ Series
8
Part No 1118353
SECTION 1—GENERAL GUIDELINES
Operating Information
For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life. Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances. DO NOT move or relocate concentrator by pulling on the power cord.
NEVER drop or insert any object or liquid into any opening.
Invacare recommends that Crush‐Proof oxygen tubing be used with this product and not exceed 15.2 m (50 ft) in length.
There are no user serviceable parts. This does not include normal maintenance items. See maintenance section for user maintenance items.
A product should NEVER be left unattended when plugged in. Make sure the Platinum is Off when not in use.
Close supervision is necessary when this product is used near children or physically‐challenged individuals.
Additional monitoring or attention may be required for patients using this device who are unable to hear or see alarms or communicate discomfort. DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
Radio Frequency Interference
This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601‐1‐2. These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation.
Part No 1118353
9
Platinum™ Series
SECTION 1—GENERAL GUIDELINES
Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards. To determine if the emissions from the Platinum is causing the interference, turn the Platinum Off. If the interference with the other device(s) stops, then the Platinum is causing the interference. In such rare cases, interference may be reduced or corrected by one of the following measures:
• Reposition, relocate, or increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.
Platinum™ Series
10
Part No 1118353
SECTION 2—FEATURES
SECTION 2—FEATURES
Oxygen Purity Indicator Lights
(SensO2 model only)/ Fault
and Power Indicator Lights
(LX model only)
Oxygen Outlet
Flowmeter
Power Switch
Circuit Breaker
Elapsed Time Meter
Cabinet Filter
(one of two shown)
ACCESSORIES (NOT SHOWN):
Remote Humidifier/Flowmeter Accessory IRCRF5 (Platinum 5 and XL Only)
REAR VIEW
PreciseRX™ Pediatric Humidifier/Flowmeter
Accessory - IRCPF16 (Platinum 5 and XL Only)
HomeFill home oxygen compressor - IOH200
Power
Cord
Part No 1118353
Outlet
Fitting
NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor. The outlet fitting does not affect concentrator performance. Refer to the HomeFill ownerʹs manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill, contact your Invacare dealer.
11
Platinum™ Series
SECTION 3—HANDLING
SECTION 3—HANDLING
The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported.
If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare.
Unpacking
NOTE: For this procedure, refer to FIGURE 3.1.
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
FIGURE 3.1 Unpacking
NOTE: Unless the oxygen concentrator is to be used immediately, retain containers and packing materials for storage until use of the concentrator is required.
Inspection
Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
Platinum™ Series
12
Part No 1118353
SECTION 4—TYPICAL PRODUCT PARAMETERS
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input:
Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level:
Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Altitude:
Platinum 5 and XL - Up to 6,000 ft (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft (1828
meters) to 13,129 ft (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft (1230 meters)
above sea level without degradation of concentration levels. From 4,000 ft (1230
meters) to 13,129 ft (4000 meters) below
90% efficiency
*Oxygen Output
Concentration Levels:
All 5LXO2/5LX models 87% to 95.6% for
flows ranging 0.5 to 5 L/min and all
10LXO2/10LX models 87% to 95.6% for
flows ranging 2 L to 10 L/min
Maximum Outlet Pressure:
Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa
± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45
kPa)
*NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).
Part No 1118353
13
Platinum™ Series
SECTION 4—TYPICAL PRODUCT PARAMETERS
Flow Range:
Platinum 5 and XL - 0.5 to 5 L/min For
flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Platinum 10 - 2 L/min to 10 L/min
Low Flow Alarm:
Platinum 5 - 0 L/min to 0.5 L/min,
Platinum 10 - 0 L/min to 1 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected)
Average Power
Consumption:
Platinum 5 and XL: 400W
Platinum 10: 585W
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recommended flow when back pressure of 7kPa is applied:
0.7 L/min
Filters:
Cabinet (2), Outlet HEPA and Compressor
Inlet.
Safety System:
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO2 Oxygen System (SensO2 Model)
Low Flow Alarm.
Width:
18 3/8 in ± 3/8 in (46.7 cm ± 1 cm)
Height:
26 3/8 in ± 3/8 in (67.0 cm ± 1 cm)
Depth:
14 3/8 in ± 3/8 in (36.5 cm ± 1 cm)
Weight:
Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ±
1 kg)
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg)
Shipping Weight:
Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ±
1 kg)
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg)
Platinum™ Series
14
Part No 1118353
SECTION 4—TYPICAL PRODUCT PARAMETERS
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity
Exhaust:
Platinum 5 and XL: Less than Ambient
+ 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F
(+25°C)
Oxygen Output Temperature:
Platinum 5 and XL: Less than Ambient + 4°F
(+2°C)
Platinum 10: Less than Ambient + 8°F
(+4.4°C)
Cabinet:
Impact Resistant flame-retardant plastic cabinet that conforms to UL 94-V0
Regulatory Listing:
Double Insulated Product ETL certified
complying with UL 1097 ETL certified to
CSA C22.2 No. 68
Electrical:
No extension cords.
Placement:
No closer than three inches from any wall,
furniture, draperies, or similar surfaces.
Tubing:
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft
high flow tubing with high flow cannula at all
flow rates.
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity
Part No 1118353
15
Platinum™ Series
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
Introduction
Your oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.
Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.
Select a Location
WARNING
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.
Keep unit at least 3 inches away from walls, draperies, furniture, and the like.
You may select a room where using your oxygen concentrator would be most convenient. Your concentrator can be easily rolled from room to room on its casters.
The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes. DO NOT place in a closet.
Platinum™ Series
16
Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
Set-Up
Plug in Power Cord
Plug in power cord to an electrical outlet.
POLARIZED PLUG WARNING
As a safety feature, this appliance may have a polarized plug (one blade is wider than the other). This plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.
Connect Humidifier (If So Prescribed)
WARNING
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient.
NOTE: For this procedure, refer to FIGURE 5.1, FIGURE 5.2 and FIGURE 5.3 on page 18 and FIGURE 5.4 on page 19.
NOTE: For Platinum 5, XL ‐ recommended humidifiers are: Invacare 3260, 003‐40, or 006‐40. For Platinum 10, recommended humidifiers are high flow: 7900‐10, 7900‐25.
1. Remove cap and fill humidifier bottle with distilled water to the level indicated by the manufacturer.
2. Replace the humidifier cap and securely tighten. Humidifier Bottle
FIGURE 5.1
Part No 1118353
Filling the Humidifier
17
Platinum™ Series
SECTION 5—OPERATING INSTRUCTIONS
Humidifier Bottle
3.
4.
5.
6.
FIGURE 5.2 Humidifier Compartment
Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3.
The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.
Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. Refer to Detail “C” in FIGURE 5.3.
Place the humidifier bottle in the humidifier compartment.
Plate Groove (Insert
Flat Head Screwdriver
Here)
Filter
Access
Door
DETAIL
“A”
DETAIL
“B”
Humidifier
Bottle
Adapter
DETAIL “C”
Filter
Access
Door
FIGURE 5.3
Platinum™ Series
Humidifier Bottle
Adapter
Attaching the Humidifier Bottle Adapter
18
Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
7. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator (FIGURE 5.4).
8. Attach the cannula/patient supply tubing to the humidifier bottle outlet (FIGURE 5.4).
9. After assembly, ensure that oxygen is flowing through the cannula.
Oxygen Outlet
Connector
Oxygen
Tubing
Humidifier Outlet
FIGURE 5.4
Attaching the Oxygen Tubing
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5 on page 19.
1. Press power switch to on position. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functioning properly.
On/Off (I/O)
Power Switch
FIGURE 5.5
Part No 1118353
Power Switch
19
Platinum™ Series
SECTION 5—OPERATING INSTRUCTIONS
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 20.
1. Turn the flowrate knob to the setting prescribed by your physician or therapist.
CAUTION
Never set the flow greater than 5 L/min on the IRC5. Never set the flow greater than 10L/min on the IRC10.
WARNING
DO NOT change the L/min setting on the flowmeter unless a change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed. 2. If the flowrate on the flowmeter ever falls below 0.5 L/min for Platinum 5 (1 L/min for Platinum 10) for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min for Platinum 5 (1 L/min for Platinum 10), the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball
FIGURE 5.6
Platinum™ Series
20
Flowrate
Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
Front Panel Indicators - Units with SensO2
NOTE: For this procedure, refer to FIGURE 5.7.
The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
OXYGEN PURITY
Auto Shut - Down
RED
Use Backup
Call Supplier
Below Normal
Normal
FIGURE 5.7
YELLOW
O2
GREEN
Front Panel Indicators - Units with SensO2
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.
LABEL
SYMBOL
O2
O2 PURITY
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O2 over 85% (± 2%)
GREEN Indicator Light
O2 Between 73% (± 3%) YELLOW Indicator light
A. YELLOW Solid
to 85% (±2%)
B. YELLOW Flashing Sensor Failure
Call a qualified technician. Have a
back-up supply of oxygen ready.
SYSTEM FAILURE
O2 Below 73% (±3%)
Part No 1118353
21
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shutdown. Call a qualified technician.
Switch to a back-up supply of oxygen immediately.
Platinum™ Series
SECTION 5—OPERATING INSTRUCTIONS
Front Panel Indicators - Units without
SensO2
NOTE: For this procedure, refer to FIGURE 5.8.
Auto Shut - Down
Use Backup
RED
Call Supplier
Normal
FIGURE 5.8
I/O
GREEN
Front Panel Indicators - Units without SensO2
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min). O2 purity will build to a maximum during this period.
LABEL
SYMBOL
I/O
Platinum™ Series
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
GREEN Indicator Light
SYSTEM FAILURE
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician. Switch
to a back-up supply of oxygen
immediately.
22
Part No 1118353
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
WARNING
The Invacare concentrators are specifically designed to minimize routine preventive maintenance. Only qualified personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. In places with high dust or soot levels, maintenance may need to be performed more often. Refer to Preventive Maintenance Record on page 25.
Routine Maintenance
NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record. Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 24. NOTE: There are two cabinet filters one located on each side of the cabinet.
1. Remove each filter and clean at least once a week depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
Part No 1118353
23
Platinum™ Series
SECTION 6—MAINTENANCE
2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.
Cabinet Filter
(located on both
sides of
concentrator)
FIGURE 6.1 Cleaning the Cabinet Filter
Cleaning the Cabinet
WARNING
UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.
Clean the cabinet with a mild household cleaner and non‐abrasive cloth or sponge.
Cleaning the Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these STEPS:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.
Platinum™ Series
24
Part No 1118353
Part No 1118353
25
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
2,160 hours are equivalent to usage 24 hours per day, for 90 days.
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Rebuild Top End of Compressor
Replace Exhaust Muffler
Clean Heat Exchanger
AS REQUIRED
Check Power Loss Alarm
Check Compressor Inlet Filter*
Check Outlet HEPA Filter*
Clean/Replace Cabinet Filter(s)
I RC
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
Check Oxygen Concentration
( UNITS WITH SensO 2 )
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST
Check Oxygen Concentration
(UNITS WITHOUT SensO 2 )
MODEL NO.
EVERY 4,380 HOURS FOR IRC5LX (EVERY 2,160 HOURS FOR IRC10LX)
Check Prescribed L/min. Flow Rate
Clean Cabinet Filter(s)
Record Elapsed Hours On Hour Meter
Record Date of Service
ON EACH INSPECTION
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
SECTION 6—MAINTENANCE
SERIAL NO.
FIGURE 6.2 Preventive Maintenance Record
Platinum™ Series
SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
Alarm:
SHORT BEEPS,
LONG PAUSE
PROBABLE CAUSE:
Main Power Loss:
1. Power cord NOT
plugged in.
2. No power at outlet.
SOLUTION:
1. Insert plug into outlet.
2. Inspect house circuit breakConcentrator
ers or fuses. If problem recurs,
NOT operating,
use a different outlet.
power switch ON. 3. Tripped circuit breaker. 3. Call Service Provider.
Beep......Beep......
Alarm:
CONTINUOUS
System Failure:
1a. Remove and clean cabinet
1. Unit overheating due to filters.
blocked air intake.
1b. Move oxygen concentrator
Concentrator
at least three inches away from
NOT operating,
walls, draperies or furniture.
Power Switch ON. 2. Insufficient power at
2. DO NOT use extension
outlet.
cords. Move to another elecBeeeeeeeep....
trical outlet or circuit.
3. Internal repairs
3. Call Service Provider.
required
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Platinum™ Series
1. Inspect for kinks or
Low Flow Alarm:
1. Kinked or blocked tub- blockages. Correct, clean or
ing, cannula or humidifier. replace item. Once
corrected, turn power OFF
for 60 seconds and then turn
power back ON.
2. Flowmeter set too low: 2. Reset flowmeter to prescribed flowrate. If prescribed
0.5 L/min (or less) for
Platinum 5; 1.0 L/min (or flowrate is below 0.5 L/min,
use the Pediatric Flowmeter
less) for Platinum 10.
IRCPF16 (Only with Platinum
5 and XL).
26
Part No 1118353
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
YELLOW or RED 1. Low oxygen purity.*
LIGHT Illuminated. 2. Kinked or blocked tubing, cannula or humidifier.*
Alarm:
CONTINUOUS
On red light only.
* Only applies to
5LXO2 and
10LXO2 SensO2
models.
GREEN LIGHT
with YELLOW
Light Flashing.
Part No 1118353
1. Clean or Replace filters.
2. Inspect for kinks or blockages. Correct, clean or replace
item. Once corrected, turn
power OFF for 60 seconds and
then turn power back ON.
3. Flowmeter set too low: 3. Check flowmeter is set to
1.0 L/min. or more.
0.5 L/min (or less) for
Platinum 5; 1.0 L/min (or
less) for Platinum 10.
4. Unit overheating due to 4.a. Remove and clean cabinet
blocked air intake.
filters.
4b. Move concentrator at least
three inches from walls, draperies, and furniture.
5. Insufficient power at
5. DO NOT use extension
outlet.
cords. Move to another electrical outlet or circuit.
6. Internal repairs
6. Call Service Provider.
required.
1. Internal repairs
1. Call Service Provider.
required.
27
Platinum™ Series
NOTES
NOTES
Platinum™ Series
28
Part No 1118353
NOTES
NOTES
Part No 1118353
29
Platinum™ Series
LIMITED WARRANTY
LIMITED WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY
WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who purchases
this product when new and unused from Invacare Corporation or a
dealer. This warranty is not extended to any other person or entity and it
is not transferable or assignable to any subsequent purchaser or owner.
Coverage under this warranty will end upon any such subsequent sale or
other transfer of title to any other person. This warranty gives you specific
legal rights and you may also have other legal rights which may vary from
state to state.
Invacare Corporation warrants its Platinum XL Oxygen Concentrators
when purchased new and unused to be free from defects in materials and
workmanship for a period of five years from date of purchase from
Invacare or a dealer, with a copy of the seller’s invoice required for
coverage under this warranty.
Invacare Corporation warrants its Platinum 10 Oxygen Concentrators
when purchased new and unused to be free from defects in materials and
workmanship for a period of three years from date of purchase from
Invacare or a dealer, with a copy of the seller’s invoice required for
coverage under this warranty.
DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE
BODY. This will void any and all warranty applicable to the valve body. If
within such warranty periods any such product shall be proven to Invacare
Corporation's satisfaction to be defective, such product shall be repaired
or replaced, at Invacare Corporation's option. This warranty only applies
to the labor for repairs performed by the Invacare Service Department or
Invacare Authorized Service Centers. It does not apply to the labor
performed by the purchaser or user. This warranty does not include
normal wear and tear or shipping charges. Invacare Corporation's sole
obligation and your exclusive remedy under this warranty shall be limited
to such repair or replacement. Routine maintenance items, such as filters,
are excluded from this warranty.
Platinum™ Series
30
Part No 1118353
LIMITED WARRANTY
For warranty service, please contact Invacare Corporation's Service
Department at the toll free number on the back cover. Upon receiving
notice of an alleged defect in a product, Invacare Corporation will issue a
serialized return authorization. It shall then be the responsibility of the
purchaser to return the entire unit or remove, at purchaser's cost, the
defective component part(s) identified, pack the component part(s) in a
manner to avoid shipping damage and to ship the component part(s) to
either Invacare Corporation's plant or service center as specified by
Invacare Corporation in advance. Defective component part(s) MUST be
returned for warranty inspection using the serial number as identification
within thirty (30) days of return authorization date. DO NOT return
products to our factory without prior consent. C.O.D. shipments will be
refused; please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY
SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE,
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE,
SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN
NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS
MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS
WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO,
MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS
OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF
REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC
CONSENT OF INVACARE CORPORATION OR TO PRODUCTS
DAMAGED BY CIRCUMSTANCES BEYOND INVACARE
CORPORATION'S CONTROL.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN
LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER
EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT
OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS
CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED
WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE
DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN.
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR
INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION
OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION
OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE
EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Part No 1118353
31
Platinum™ Series
This product may be manufactured under one or more of the following patents: US 5,906,672 and US 5,917,135. Other patents pending.
Invacare Corporation
www.invacare.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
800-333-6900
Technical Services
800-832-4707
Trademarks are identified by ™ and ®. All
trademarks are owned by or licensed to
Invacare Corporation or its subsidiaries
unless otherwise noted.
© 2009 Invacare Corporation
Part No 1118353
Rev J - 4/09