Swann AMI Oxytrace Operator's Manual

Add to My manuals
98 Pages

advertisement

Swann AMI Oxytrace Operator's Manual | Manualzz

AMI Oxytrace

AMI Oxytrace QED

Version 6.20 and higher

A-96.250.531 / 300617

Customer Support

SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest

SWAN representative, or the manufacturer:

SWAN ANALYTISCHE INSTRUMENTE AG

Studbachstrasse 13

8340 Hinwil

Switzerland

Internet: www.swan.ch

E-mail: [email protected]

Document Status

Title:

Monitor AMI Oxytrace QED Operator’s Manual

ID:

A-96.250.531

Revision Issue

00 Jan. 2008 First Edition

01

02

Nov. 2008

July 2011

Including AMI Oxytrace QED

Chapter Quality Assurance added

03

04

Oct. 2013

Sept. 2016

New mainboard V2.4, update to FW Release 5.4

New mainboard V2.5, update to FW Release 6.20

© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved subject to change without notice.

AMI Oxytrace QED

Table of Contents

1.

Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.1.

Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.2.

General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.

Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2.1.

Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2.2.

Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

2.3.

Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

2.4.

Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

2.4.1 AMI Oxytrace Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

2.4.2 Sensor OXYTRACE G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

2.4.3 QV-Flow PMMA OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

2.4.4 B-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

4

5

7

3.

Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.1.

Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.2.

Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

3.3.

Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 23

3.3.1 Swagelok Fitting Stainless Steel at Sample Inlet . . . . . . . . . . . . . 23

3.3.2 FEP Tube at Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

3.4

Install the Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . 24

3.5.

Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

3.5.1 Electrical Connection Diagram AMI Oxytrace . . . . . . . . . . . . . . . 27

3.5.2 Electrical Connection Diagram AMI Oxytrace QED . . . . . . . . . . . 28

3.5.3 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

3.6.

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

3.7.

Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

3.7.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

3.7.2 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

3.8.

Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

3.8.1 Signal output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . . 33

3.9.

Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

3.9.1 Signal Output 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

3.9.2 Profibus, Modbus Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

3.9.3 HART Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

3.9.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

4.

Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

4.1.

Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

4.2.

Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

A-96.250.531 / 300617

1

AMI Oxytrace QED

2

5.

Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.1.

Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.2.

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.3.

Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.4.

Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . .

6.

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.1.

Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.2.

Stop of Operation for Maintenance . . . . . . . . . . . . . . . . . . . . . . . .

6.3.

Maintenance of the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . .

6.3.1 Electrolyte exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.3.2 Clean Swansensor Oxytrace G and Flow Cell . . . . . . . . . . . . . . .

6.4.

Maintenance of the Faraday electrode . . . . . . . . . . . . . . . . . . . . .

6.5.

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.6.

Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.7.

Faraday Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.8.

Quality Assurance of the Instrument. . . . . . . . . . . . . . . . . . . . . . .

6.8.1 Activate SWAN Quality assurance procedure . . . . . . . . . . . . . . .

6.8.2 Pre-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.8.3 Connect the sample lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.8.4 Carry out comparison measurement. . . . . . . . . . . . . . . . . . . . . . .

6.8.5 Completion of the measurement. . . . . . . . . . . . . . . . . . . . . . . . . .

6.9.

Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.10. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7.

Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.

Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.1.

Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.2.

Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.3.

Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.4.

Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.5.

Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9.

Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . .

1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10.

Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10.1. Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

41

53

53

55

56

49

50

51

52

57

58

41

41

42

42

44

45

47

37

37

38

39

40

59

63

63

64

65

66

67

69

69

69

71

73

74

89

89

A-96.250.531 / 300617

AMI Oxytrace QED

11.

Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

12.

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

13.

Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

A-96.250.531 / 300617

3

AMI Oxytrace QED

Safety Instructions

AMI Oxytrace QED -

Operator’s Manual

This document describes the main steps for instrument setup, operation and maintenance.

1.

Safety Instructions

General

The instructions included in this section explain the potential risks associated with instrument operation and provide important safety practices designed to minimize these risks.

If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environment.

More safety instructions are given throughout this manual, at the respective locations where observation is most important.

Strictly follow all safety instructions in this publication.

Target audience

Operator: Qualified person who uses the equipment for its intended purpose.

Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.

OM Location

The AMI Operator’s Manual shall be kept in proximity of the instrument.

Qualification,

Training

To be qualified for instrument installation and operation, you must:

 read and understand the instructions in this manual as well as the Material Safety Data Sheets.

 know the relevant safety rules and regulations.

4

A-96.250.531 / 300617

AMI Oxytrace QED

Safety Instructions

1.1.

Warning Notices

The symbols used for safety-related notices have the following significance:

DANGER

Your life or physical wellbeing are in serious danger if such warnings are ignored.

 Follow the prevention instructions carefully.

WARNING

Severe injuries or damage to the equipment can occur if such warnings are ignored.

Follow the prevention instructions carefully.

CAUTION

Damage to the equipment, minor injury, malfunctions or incorrect process can be the consequence if such warnings are ignored.

Follow the prevention instructions carefully.

Mandatory

Signs

The importance of the mandatory signs in this manual.

Safety goggles

Safety gloves

A-96.250.531 / 300617

5

AMI Oxytrace QED

Safety Instructions

Warning Signs

The importance of the warning signs in this manual.

Electrical shock hazard

Corrosive

Harmful to health

Flammable

Warning general

Attention general

6

A-96.250.531 / 300617

AMI Oxytrace QED

Safety Instructions

1.2.

General Safety Regulations

Legal

Requirements

Spare Parts and

Disposables

The user is responsible for proper system operation.

All precautions must be followed to ensure safe operation of the instrument.

Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.

Modifications

Modifications and instrument upgrades shall only be carried out by an authorized Service Technician. SWAN will not accept responsibility for any claim resulting from unauthorized modification or alteration.

WARNING

Risk of Electrical Shock

If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.

To prevent from electrical shock, always make sure that the ground wire is connected.

Service shall be performed by authorized personnel only.

Whenever electronic service is required, disconnect instrument power and power of devices connected to.

– relay 1,

– relay 2,

– alarm relay

WARNING

For safe instrument installation and operation you must read and understand the instructions in this manual.

WARNING

Only SWAN trained and authorized personnel shall perform the tasks described in this document.

A-96.250.531 / 300617

7

AMI Oxytrace QED

Product Description

2.

Product Description

This manual describes the functions of the following monitors

 AMI Oxytrace

 AMI Oxytrace QED

Both monitors are mainly identical except that AMI Oxytrace QED includes a faraday verification.

2.1.

Description of the System

Application

AMI Oxytrace is used to measure low levels of oxygen in high-purity water. Especially in power plant water cycles (e.g. feedwater), a very low level of oxygen is needed to prevent corrosion.

Measuring principle

Clark principle:

The sensor consists of one noble metal electrode (e.g. platinum or gold), a reference electrode (mostly Ag/AgCl) and optionally a metal guard electrode.

The Clark-type electrode is the most widely used oxygen sensor for measuring oxygen dissolved in a liquid. The basic principle is that there is a cathode and an anode submersed in an electrolyte and a voltage is applied between the two parts. Oxygen enters the sensor through a permeable membrane by diffusion, and is reduced at the cathode according to

-

O

2

+ 4e

-

+ 2 H

2

O --> 4 OH

This reaction creates a measurable current. There is a linear correlation between the oxygen concentration and the electrical current.

The guard electrode is on the same voltage level as the cathode but there is no current measurement. Oxygen which diffuses from the electrolyte to the cathode is consumed by the guard electrode.

As a consequence, residual oxygen in the electrolyte will no more disturb the measurement signal and the response time to low oxygen levels will be shorter.

Temperature compensation

The measuring signal depends on temperature, but is automatically compensated to 25 °C. The sample temperature is determined continuously by a temperature sensor inside the oxygen electrode.

8

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

40

35

30

5

0

25

20

15

10

Faraday

Verification

For AMI Oxytrace QED only.

When a direct current is passed through water by means of two electrodes, electrolysis of the liquid takes place, according to the laws of Michael Faraday. Water is converted to molecular oxygen and hydrogen.

Thus, by controlling the current, a fixed, known amount of oxygen, which is independent of temperature and pressure, may be generated. If the sample flow is known, an exact oxygen concentration increment can be generated at low levels. This increment is used to verify the flawless operation of the whole system. Thereby the response characteristics (incremental change and response time) of the sensor are taken into account. Sensor failures (loss of electrolyte etc.) can be detected very easily as shown in the plot below.

operational sensor defective sensor

30 60 90 120 150 180 210 240 270 300 330

Time [Sec]

Unusual sensor performance is recognized immediately and gives also a maintenance indication to the operator/user. The faraday verification is therefore an excellent QA/QC tool.

A-96.250.531 / 300617

9

AMI Oxytrace QED

Product Description

Signal

Outputs

Two signal outputs programmable for measured values (freely scaleable, linear or bilinear) or as continuous control output (control parameters programmable).

Current loop:

Maximal burden:

0/4–20 mA

510

Third signal output available as an option.

Relay

Two potential-free contacts programmable as limit switches for measuring values, controllers or timer for system cleaning with automatic hold function. Both contacts can be used as normally open or normally closed.

Maximum load: 1 A/250 VAC

Alarm Relay

One potential free contact.

Alternatively:

 Open during normal operation, closed on error and loss of power.

 Closed during normal operation, open on error and loss of power.

Summary alarm indication for programmable alarm values and instrument faults.

Input

For potential-free contact to freeze the measuring value or to interrupt control in automated installations (hold function or remote-off)

Safety

Features

No data loss after power failure. All data is saved in non-volatile memory. Over voltage protection of in- and outputs.Galvanic separation of measuring inputs and signal outputs.The analyzer is factory tested and ready for installation and operation.

Communication Interface

(optional)

 USB Interface for logger download

 RS485 with Fieldbus protocol Modbus or Profibus DP.

 HART interface

10

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

Fluidics

AMI Oxytrace

Swansensor oxygen combined with QV-flow PMMA OTG flow cell.

The sample flows via sample inlet [G] through the flow regulating valve [D], where the flow rate can be adjusted. Then the sample flows into the measuring cell [C] were the Oxygen concentration and temperature of the sample is measured.

The sample leaves the measuring cell via flow sensor [B] through the sample outlet [E] and the drain funnel [F].

A

A-96.250.531 / 300617

D

E

F

B

C

A

B

C

D

Oxygen sensor

Flow sensor

Flow cell

Flow regulating valve

G

E

F

G

Sample outlet

Drain funnel

Sample inlet

11

AMI Oxytrace QED

Product Description

Fluidics

AMI Oxytrace QED

Swansensor oxygen combined with QV-flow PMMA OTG flow cell.

The sample flows via sample inlet [H] through the flow regulating valve [E], where the flow rate can be adjusted. Then the sample flows through the faraday electrode [D] into the measuring cell [C] were the Oxygen concentration of the sample is measured.

The sample leaves the measuring cell via flow sensor [B] through the sample outlet [F] and the drain funnel [G].

A

12

A

B

C

D

Oxygen sensor

Flow sensor

Flow cell

Faraday electrode

B

C

D

E

F

G

H

E

F

G

H

Flow regulating valve

Sample outlet

Drain funnel

Sample inlet

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

2.2.

Technical Data

Power Supply

Voltage:

Power consumption:

Electronics

Aluminium with a protection degree of IP 66 / NEMA 4X

housing

Ambient temperature:

Humidity:

Display:

-10 to +50 °C

10–90% rel., non condensing backlit LCD, 75 x 45 mm

Sample requirements

100–240 VAC (± 10%)

50/60 Hz (± 5%) or 24 VDC (± 10%) max. 30 VA

Flow cell and connection

On-Site requirements

Flow rate:

Temperature:

Inlet pressure:

Outlet pressure pH:

Suspended solids:

Sample outlet:

8–25 l/h

15–45 °C

0.2 to 1 bar pressure free not lower than pH 4 less than 10 ppm

Flow cell made of acrylic glass with built-in flow adjustment valve and digital sample flow meter

Sample inlet:

Sample outlet:

1/4” Swagelok tube adapter for flexible tube 15x20 mm

Sample inlet: Swagelok connection for stainless steel tube 1/4”.

15x20 mm (1/2”) hose nozzle which must end in a pressure free waste of sufficient capacity.

Accuracy

Reproducibility

Measuring

Range

Temperature

± 1.5% of measured value or ± 0.2 ppb

±1% of measured value or ± 0.15 ppb up to 60°C

Resolution: 0.1°C

Sample Flow

Measurement

With Swan sample flow meter and alarm in case of insufficient sample flow.

A-96.250.531 / 300617

13

AMI Oxytrace QED

Product Description

Dimensions

(Oxytrace and

Oxytrace QED)

Panel:

Mounting hole distance

Screws:

Weight:

Stainless steel

280x850x150 mm

254x824

8 mm diameter

12.0 kg

280 mm/ 11”

254 mm/ 10”

AMI Oxytrace

/ 16

14

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

2.3.

Instrument Overview

AMI Oxytrace

A

B

I

C

D

E

F

G

H

A

B

C

D

E

Panel

AMI Transmitter

Oxygen sensor

Flow sensor

Flow cell

F

G

I

H

Flow regulating valve

Sample outlet

Drain funnel

Sample inlet

A-96.250.531 / 300617

15

AMI Oxytrace QED

Product Description

AMI Oxytrace QED

C

D

I

E

F

G

H

J

K

A

B

D

E

F

A

B

C

Panel

AMI Transmitter

Faraday control unit

Oxygen sensor

Sample outlet

Flow sensor

G

I

H

J

K

Flow cell

Flow regulating valve

Faraday electrode

Drain funnel

Sample inlet

16

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

2.4.

Single Components

2.4.1

AMI Oxytrace Transmitter

Electronic transmitter and controller for oxygen measurement.

18.5

140

103

4.5

Dimensions

Width:

Height:

Depth:

Weight:

Specifications

Electronics case:

Protection degree:

Display:

Electrical connectors:

140 mm

180 mm

70 mm

1.5 kg

Cast aluminum

IP 66 / NEMA 4X backlit LCD, 75 x 45 mm screw clamps

A-96.250.531 / 300617

17

AMI Oxytrace QED

Product Description

2.4.2

Sensor OXYTRACE G

Sensor for the measurement of dissolved oxygen in ultra pure water. Precise oxygen measuring cell with integrated temperature sensor and guard electrode for faster initial response time after maintenance.

59

100

Technical data :

Clark oxygen electrode

Cathode gold, anode silver, guard silver

Zero current-free electrode system

Robust 25 µm fluoropolymer diaphragm

Measuring range :

0–20 ppm O

Range

2

(25 °C)

Automatic range switching

0.1–9.99 ppb

10–199.9 ppb

200–1999 ppb

2–20 ppm

0 – 200% saturation

Resolution

0.01 ppb

0.1 ppb

1.0 ppb

0.01 ppm

Temp. sensor :

NT5K

Accuracy :

0,3 % if calibration temperature = measuring temp.

1,5 % at ± 10 °C deviation to cal. temperature

Precision

Response time

Minimal flow

Operating temp.

:

:

:

:

± 1 % of reading or ±0.15 ppb t

90

< 30 seconds (rising concentration)

50 cm/s Pressure resistance: 3 bar max. 50 °C

Material

:

Protection

Weight

:

:

shaft: polyacetal copolymer cathode/anode/guard: gold/silver/silver membrane: fluoropolymer

IP 68

150 g

18

A-96.250.531 / 300617

AMI Oxytrace QED

Product Description

2.4.3

QV-Flow PMMA OTG

Flow cell

Flow cell made of acrylic glass with integrated flow sensor.

Sample temp.

max. 45 °C

Inlet pressure

max. 1 bar

Outlet pressure

Pressure free

Sample flow

6–25 l/h

Process connection inlet

Swagelok connection for 1/4” tube

Outlet

SERTO angle for 6 mm flexible tube

Dimensions

see picture below

80

21

38

5.3

8

62

A-96.250.531 / 300617

19

AMI Oxytrace QED

Product Description

2.4.4

B-Flow SS316L OTG

Flow cell

B-Flow SS316L OTG is made of stainless steel without flow sensor and can be used for higher operating pressures and temperatures.

Operating temp.

-10 to + 130 °C

Sensor

max. 50 °C

Operating pressure

max. 5 bar at 130 °C

Sensor

max. 3 bar

Flow cell connection

2x female thread 1/8" ISO

Dimensions

see picture below

52

40

20

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.

Installation

Restrictions on use

The first part of this chapter describes the preparing and placing of the system in position for use. The second part describes the wiring and the properties of the signal outputs and relay.

If exposed to weather, install a cover to protect the instrument from insolation and precipitation. See

Technical Data, p. 13 for tempera-

ture limitations.

Check

3.1.

Installation Check List

On site requirements

Installation

Electrical

Connections

Swansensor

Oxytrace G

Instrument’s specification must conform to the National Electrical

Code, all state and local codes, and all plant codes and standards for electrical equipment.

100–240 VAC (

 10%), 50/60 Hz (  5%) or 24 VDC (±10%), isolated power outlet with ground connection and 30 VA

Sample line with 8–25 l/h and 0.2–3 bar.

Waste line with pressure free drain.

Mount the instrument in vertical position. Display should be at eye level.

Connect the sample and waste line, see

Connecting Sample Inlet and Outlet, p. 23 .

Connect all external devices like limit switches, current loops and pumps, see

Electrical Connections, p. 25

.

Connect the power cord, do not switch on power until all external devices are connected.

The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber. A transport protection cap filled with water keeps the sensor wet during transport and storage. Before installation, remove the transport protection cap and clean the sensor surfaces with water. Afterwards install the oxygen sensor and connect the cable.

See Install the Swansensor Oxytrace G, p. 24

.

A-96.250.531 / 300617

21

AMI Oxytrace QED

Installation

Turn-on power

Run-in Period

Calibration

Open

Sample Flow

Programming

Switch on power.

First, the analyzer performs a self test, displays the firmware version and then starts normal operation

Leave the sensor at the air. The sensor has to run-in for at least

30 min, better 1 h. During this time it has to be at air with a dry and clean membrane and has to be connected to the powered instru-

menty. Calibrate the sensor after this time, see Calibration, p. 47

. and mount it into the flow cell afterwards.

Open the flow regulating valve to allow sample flow into the flow cell and into the waste. Sample must always overflow into waste

Program all parameters for external devices (interface, etc.). Set all parameters for instrument operation (limits, alarms).

3.2.

Mounting of Instrument Panel

Mounting requirements

The first part of this chapter describes the preparing and placing of the system for use.

The instrument must only be installed by trained personnel.

 Mount the instrument in vertical position.

 For ease of operation mount it so that the display is at eye level.

 For the installation a kit containing the following installation material is available:

– 4 Screws 8x60 mm

– 4 Dowels

– 4 Washers 8.4/24 mm

The instrument is only intended for indoor installation.

For dimensions see Dimensions (Oxytrace and Oxytrace QED), p.

14 .

22

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.3.

Connecting Sample Inlet and Outlet

3.3.1

Swagelok Fitting Stainless Steel at Sample Inlet

Preparation

Cut the tube to length and deburr it. The tube must be straight and free from blemishes for approximately 1,5 x tube diameter from the end.

Lubrication with lubricating oil, MoS2, Teflon etc. is recommended for the assembly and reassembly of bigger sized unions (thread, compression cone).

Installation

1 Insert the compression ferrule [C] and the compression cone [D] into the union nut [B].

2 Screw on the union nut onto the body, do not tighten it.

3 Push the stainless steel pipe through the union nut as far as it reaches the stop of the body.

4 Mark the union nut at 6 o’clock position.

5 While holding the fitting body steady, tighten the nut union 1¼ rotation using an open ended spanner.

A B C D E

12

9

3

6

A

B

C

Stainless steel tube

Union nut

Compression ferrule

F

D

E

F

Compression cone

Body

Tightened connection

A-96.250.531 / 300617

23

AMI Oxytrace QED

Installation

3.3.2

FEP Tube at Sample Outlet

1/2” Tube at waste funnel.

A

A

B

C

Waste funnel

Hose nozzle

1/2” tube

B

C

Connect the 1/2” tube [C] to the hose nozzle [B] and place it into a drain with atmospheric pressure.

3.4

Install the Swansensor Oxytrace G

The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber [E]. A transport protection cap [B] filled with water [D] keeps the sensor wet during transport and storage. To install the sensor proceed as follows:

A

B

C

A

B

C

D

E

Fixing sleeve

Transport protection cap

Swansensor Oxytrace G

Water

Sensor cap filled with electrolyte

D

E

24

1 Unscrew the fixing sleeve [A].

2 Remove the transport protection cap [B].

3 Clean the Swansensor Oxytrace G [C] with water.

4 Install the Swansensor Oxytrace G into the flow cell

5 Connect the sensor cable to the transmitter, see

Electrical Connection Diagram AMI Oxytrace, p. 27

.

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.5.

Electrical Connections

Cable thicknesses

WARNING

Risk of electrical shock.

Do not perform any work on electrical components if the transmitter is switched on. Failure to follow safety instructions could result in serious injury or death.

 Always turn off AC power before manipulating electric parts.

 Grounding requirements: Only operate the instrument from an power outlet which has a ground connection.

 Make sure the power specification of the instrument corresponds to the power on site.

In order to comply with IP66, use the following cable thicknesses

Wire

A B C

A

B

C

PG 11 cable gland: cable Ø outer

PG 7 cable gland: cable Ø

PG 9 cable gland: cable Ø

5–10 mm outer

3–6.5 mm outer

4–8 mm

NOTICE: Protect unused cable glands

 For Relays: Use max. 1.5 mm

2 end sleeves.

/ AWG 14 stranded wire with

 For Signal Outputs and Input: Use 0.25 mm

2 stranded wire with end sleeves.

/ AWG 23

A-96.250.531 / 300617

25

AMI Oxytrace QED

Installation

WARNING

External Voltage.

External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks

 Make sure that the devices connected to the following contacts are disconnected from the power before resuming installation.

– relay 1

– relay 2

– alarm relay

WARNING

To prevent from electrical shock, do not connect the instrument to the power unless the ground wire (PE) is connected.

 Do not connect unless specifically instructed to do so.

WARNING

The mains of the AMI Transmitter must be secured by a main switch and appropriate fuse or circuit breaker.

26

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.5.1

Electrical Connection Diagram AMI Oxytrace

A-96.250.531 / 300617

CAUTION

Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.

27

AMI Oxytrace QED

Installation

3.5.2

Electrical Connection Diagram AMI Oxytrace QED

28

CAUTION

Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.5.3

Power Supply

WARNING

Risk of electrical shock

Do not perform any work on electrical components if the transmitter is switched on. Failure to follow safety instructions could result in serious injury or death.

 Always turn off AC power before manipulating electric parts.

 Installation and maintenance of electrical parts must be performed by professionals.

A

B

C

D

Installation requirements

A

B

C

D

Power supply connector

Neutral conductor, Terminal 2

Phase conductor, Terminal 1

Protective earth PE

NOTICE: The protective earth wire (Ground) has to be connected to the grounding terminal.

The installation must meet the following requirements.

 Mains fuse 1.6 AT

 Mains cable to comply with standards IEC 60227 or IEC

60245; flammable rating FV1

 Mains equipped with an external switch or circuit-breaker

– near the instrument

– easily accessible to the operator

– marked as interrupter for AMI Oxytrace QED

A-96.250.531 / 300617

29

AMI Oxytrace QED

Installation

3.6.

Input

NOTICE: Use only potential-free (dry) contacts.

The total resistance (sum of cable resistance and resistance of the relay contact) must be less than 50 Ω.

Terminals 16 and 42

If the signal output is set to hold, the measurement is interrupted if input is active.

For programming see Menu Installation

5.3.4, p. 85

.

3.7.

Relay Contacts

3.7.1

Alarm Relay

NOTICE: Max. load 1 A/250 VAC

Alarm output for system errors.

Error codes see

Error List, p. 59 .

NOTICE: With certain alarms and certain settings of the AMI transmitter the alarm relay does not switch. The error, however, is shown on the display.

NC

1)

Normally

Closed

NO

Normally

Open

Terminals

10/11

12/11

Description

Active (opened) during normal operation.

Inactive (closed) on error and loss of power.

Active (closed) during normal operation.

Inactive (opened) on error and loss of power.

0V

Relay connection

1)

11

0V

10

12

11

10

12

1) usual use

30

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.7.2

Relay 1 and 2

NOTICE: Max. load 1 A/250 VAC

Relay 1 and 2 can be configured as normally open or as normally closed. Standard for both relays is normally open. To configure a

Relay as normally closed, set the jumper in the upper position.

NOTICE: Some error codes and the instrument status may influence the status of the relays described below.

Relay config.

Terminals

Normally

Open

6/7: Relay 1

8/9: Relay 2

Jumper pos.

Normally

Closed

6/7: Relay 1

8/9: Relay 2

Description

Inactive (opened) during normal operation and loss of power.

Active (closed) when a programmed function is executed.

Relay configuration

0V

Inactive (closed) during normal operation and loss of power.

Active (opened) when a programmed function is executed.

0V

6

7

6

7

A

B

A-96.250.531 / 300617

A

B

Jumper set as normally open (standard setting)

Jumper set as normally closed

For programming see Menu Installation 5.3.2 and 5.3.3, p. 81 .

31

AMI Oxytrace QED

Installation

CAUTION

Risk of damage of the relays in the AMI Transmitter due to heavy inductive load.

Heavy inductive or directly controlled loads (solenoid valves, dosing pumps) may destroy the relay contacts.

 To switch inductive loads > 0.1 A use an AMI relay box available as an option or suitable external power relays.

Inductive load

Small inductive loads (max 0.1A) as for example the coil of a power relay can be switched directly. To avoid noise voltage in the

AMI Transmitter it is mandatory to connect a snubber circuit in parallel to the load.

A snubber circuit is not necessary if an AMI relaybox is used.

B

D

A

C

E

A

B

C

D

E

AC or DC power supply

AMI Transmitter

External power relay

Snubber

Power relay coil

Resistive load

Resistive loads (max. 1A) and control signals for PLC, impulse pumps and so on can be connected without further measures

A B

C

A

B

C

AMI Transmitter

PLC or controlled pulse pump

Logic

Actuators

Actuators, like motor valves, are using both relays: One relay contact is used for opening, the other for closing the valve, i.e. with the

2 relay contacts available, only one motor valve can be controlled.

Motors with loads bigger than 0.1A must be controlled via external power relays or an AMI relay box.

B

A

C

A

B

C

AC or DC power supply

AMI Transmitter

Actuator

M

32

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.8.

Signal Outputs

3.8.1

Signal output 1 and 2 (current outputs)

NOTICE: Max. burden 510 Ω.

If signals are sent to two different receivers, use signal isolator

(loop isolator).

Signal output 1:

Signal output 2:

Terminals 14 (+) and 13 (-)

Terminals 15 (+) and 13 (-)

Programming see menu

5.2 Signal Outputs, p. 75

3.9.

Interface Options

A

A

B

C

D

B

AMI Transmitter

Slot for interfaces

Frontend PCB

Screw terminals

C

D

A-96.250.531 / 300617

The slot for interfaces can be used to expand the functionality of the AMI instrument with either:

 an additional signal output

 a Profbus or Modbus connection

 a HART connection

 an USB Interface

33

AMI Oxytrace QED

Installation

3.9.1

Signal Output 3

Terminal 38 (+) and 37 (-).

Requires the additional board for the third signal output 0/4–20 mA

PCB. The third signal output can be operated as a current source or as a current sink (switchable via switch [A]). For detailed information see the corresponding installation instruction.

NOTICE: Max. burden 510

.

A

Third signal output 0/4 - 20 mA PCB

A Operating mode selector switch

3.9.2

Profibus, Modbus Interface

Terminal 37 PB, Terminal 38 PA

To connect several instruments by means of a network or to configure a PROFIBUS DP connection, consult the PROFIBUS manual.

Use appropriate network cable.

NOTICE: The switch must be ON, if only one instrument is installed, or on the last instrument in the bus.

34

A

Profibus, Modbus Interface PCB (RS 485)

A On - OFF switch

A-96.250.531 / 300617

AMI Oxytrace QED

Installation

3.9.3

HART Interface

Terminals 38 (+) and 37 (-).

The HART interface PCB allows for communication via the HART protocol. For detailed information, consult the HART manual.

HART Interface PCB

3.9.4

USB Interface

The USB Interface is used to store Logger data and for Firmware up load. For detailed information see the corresponding installation instruction.

A-96.250.531 / 300617

USB Interface

A

B

35

AMI Oxytrace QED

Instrument Setup

4.

Instrument Setup

4.1.

Establish Sample Flow

A

1 Open the flow regulating valve [A] and wait until the flow cell is completely filled.

2 Switch on power.

3 Adjust the sample flow to 8–25 l/h.

4.2.

Programming

Program all parameters for external devices (interface, recorders, etc.)

Program all parameters for instrument operation (limits, alarms),

see Program List and Explanations, S. 69

36

A-96.250.531 / 300617

AMI Oxytrace QED

Operation

5.

Operation

5.1.

Keys

A B C D

Exit Enter

A

to exit a menu or command (rejecting any changes) to move back to the previous menu level

B

to move DOWN in a menu list and to decrease digits

C

to move UP in a menu list and to increase digits

D

to open a selected sub-menu to accept an entry

Program

Access, Exit

RUN

R1

R2

9 l/h

14:10:45

1.05 ppb

25.4°C

Enter

Exit

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

1

A-96.250.531 / 300617

37

AMI Oxytrace QED

Operation

5.2.

Display

A B C D

RUN

R1

R2

12 l/h

15:20:18

0.15

ppm

23 °C

E

H

G

F

A RUN

HOLD

OFF

B ERROR normal operation input closed or cal delay: Instrument on hold (shows status of signal outputs).

input closed: control/limit is interrupted (shows status of signal outputs).

Error Fatal Error

C Keys locked, transmitter control via Profibus

D Time

E Process values

F

Sample Temperature

G Sample flow in i/h

H Relay status

Relay status, symbols upper/lower limit not yet reached upper/lower limit reached control upw./downw. no action control upw./downw. active, dark bar indicates control intensity motor valve closed motor valve: open, dark bar indicates approx. position timer timer: timing active (hand rotating)

38

A-96.250.531 / 300617

AMI Oxytrace QED

Operation

5.3.

Software Structure

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

Messages

Pending Errors

Maintenance List

Message List

1

1.1

Menu Messages 1

Reveals pending errors as well as an event history

(time and state of events that have occurred at an earlier point of time).

It contains user relevant data.

Menu Diagnostics 2

Provides user relevant instrument and sample data.

Diagnostics

Identification

Sensors

Sample

I/O State

Interface

2.1

Maintenance

Calibration

Service

Simulation

3.1

Set Time 23.09.06 16:30:00

Operation

Sensors

Relay Contacts

Logger

4.1

Installation

Sensors

Signal Outputs

Relay Contacts

Miscellaneous

Interface

5.1

Menu Maintenance 3

For instrument calibration, relay and signal output simulation, and to set the instrument time.

It is used by the service personnel.

Menu Operation 4

User relevant parameters that might need to be modified during daily routine. Normally password protected and used by the process-operator.

Subset of menu 5 - Installation, but process-related.

Menu Installation 5

For initial instrument set up by SWAN authorized person, to set all instrument parameters. Can be protected by means of password.

A-96.250.531 / 300617

39

AMI Oxytrace QED

Operation

5.4.

Changing Parameters and values

Changing parameters

The following example shows how to change the logger interval:

Log interval

Clear logger

30 min no

1 Select the parameter you want to change.

2 Press [Enter]

Temperature

Standards

NT5K

Logger

Log interval

Interval.

5 min

10 min

30 min

1 Hour

4.1.3

30min no

3 Press [ ] or [ ] key to highlight the required parameter.

4 Press [Enter] to confirm the selection or [Exit] to keep the previous parameter).

Logger

Log interval

Clear logger

4.1.3

10 min no

The selected parameter is highlighted (but not saved yet).

5 Press [Exit].

40

Changing values

Logger

Log interval

Clear logger

Yes

No

4.1.3

no

Yes is highlighted.

6 Press [Enter] to save the new parameter.

The system reboots, the new parameter is set.

Alarm oxygen

Alarm High

Alarm Low

Hysteresis

Delay

5.3.1.1.1

10.00 ppm

1.00 ppb

0.10 ppb

5 Sec

Alarm oxygen

Alarm High

Alarm Low

Hysteresis

Delay

5.3.1.1.1

8.00 ppb

1.00 ppb

0.10 pp

5 Sec

1 Select the value you want to change.

2 Press [Enter].

3 Set required value with [

[ ] key.

] or

4 Press [Enter] to confirm the new value.

5 Press [Exit].

Yes is highlighted.

6 Press [Enter] to save the new value.

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.

Maintenance

Maintenance frequency depends strongly on the water quality. The

AMI Oxytrace QED is designed for determination of low level of dissolved oxygen in high purity water.

It is not suitable for the measurement of dissolved oxygen in waste water.

6.1.

Maintenance Table

Weekly

Monthly

Half-yearly

Yearly

Check sample flow.

If necessary, perform an air calibration.

Clean Oxytrace G membrane with a soft tissue, see

Maintenance of the Oxygen Sensor, p. 42 .

Clean faraday electrode, see Maintenance of the

Faraday electrode, p. 45

If necessary, replace filling electrolyte.

If the sensor is exposed to air frequently and during long time intervals, the electrolyte and membrane may have to be changed earlier (see below*).

Clean flow cell and flow meter, if dirty.

Replace Swansensor Oxytrace G membrane by using a new, pre filled sensor cap.

Every 2nd year

*A change of membrane and electrolyte is recommended:

 if indicated in the maintenance list (remaining amount <10%)

 if the response of the sensor is slow

 if the sensor can not be calibrated any more and/or the instrument shows a corresponding error message

 if the sensor signal is very unstable.

6.2.

Stop of Operation for Maintenance

1 Shut off power of the instrument.

2 Stop sample flow by closing the flow regulating valve.

A-96.250.531 / 300617

41

AMI Oxytrace QED

Maintenance

6.3.

Maintenance of the Oxygen Sensor

WARNING

Etching liquid

The electrolyte is alkaline and caustic. It contains less than 1% of potassium hydroxide.

Do not ingest. Wear protective goggles and gloves during handling. Avoid contact with clothes.

 In case of accidental contact with the eyes, wash immediately with clear water and contact a physician. Show him the label of the bottle or this section of the manual.

Short contact with skin is harmless, nevertheless wash with lots of water.

6.3.1

Electrolyte exchange

An electrolyte exchange is indicated in the maintenance list as soon as the remaining amount is below 10%.

A

B

C

A

B

C

Fixing sleeve

Sensor

Flow cell

1 Unscrew the fixing sleeve [A].

2 Remove the sensor from the flow cell.

NOTICE: Do not pull out the sensor with force!

To remove the sensor from the flow cell, open the flow regulating valve during removing the sensor. This prevents the formation of a vacuum, and the sensor can easily be removed from the flow cell.

42

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

A

G

H

E

F

B

C

D

E

F

G

H

A

B

C

D

Swansensor Oxytrace G

Thread

Groove

Guard electrode

Anode

Measuring head

Cathode

Sensor cap with membrane

A-96.250.531 / 300617

3 Unscrew and remove the sensor cap [H] from the Swansensor

Oxytrace G [A].

4 Empty the remaining electrolyte.

5 Refill the sensor cap with fresh electrolyte.

NOTICE: There is a groove [C] in the thread [B] of the sensor, where excessive air and electrolyte can escape while screwing the sensor cap onto the sensor. Hold the sensor in vertical position, measuring head pointing downwards.

6 Slowly screw the sensor cap onto the sensor to allow the excess electrolyte to escape without building up too much pressure inside the electrode. Fasten sensor cap tightly.

7 Clean the sensor thoroughly and dry the sensing membrane with a soft tissue.

8 Switch on power.

9 Let the sensor run-in at air for at least 30 min, better 1 h.

10 Afterwards perform an air calibration.

11 Install the sensor into the flow cell.

12 Select “New Filling” to reset the counter for remaining electro-

lyte, see <Maintenance>/<Service> 3.2.1, p. 71 ).

43

AMI Oxytrace QED

Maintenance

6.3.2

Clean Swansensor Oxytrace G and Flow Cell

Depending on the water quality, the Swansensor Oxytrace G and the flow cell will necessitate a cleaning.

Before cleaning, stop operation as described in Stop of Operation for Maintenance, p. 41 .

1 Dismount the Swansensor Oxytrace G, see Electrolyte exchange, p. 42 .

2 Clean the sensor with a soft tissue and rinse it with water afterwards.

3 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.

4 Flush the flow cell with clean water.

5 Install the Swansensor Oxytrace G and start sample flow.

44

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.4.

Maintenance of the Faraday electrode

A

D

E

F

G

A

B

C

Fixing sleeve

Electrode body

Washer

O-ring

Inner electrode

O-ring

Hollow electrode

B

C

D

E

F

G

A-96.250.531 / 300617

1 Switch off the instrument and close the flow regulating valve.

2 Open the faraday control unit.

3 Disconnect and remove the cable from the faraday control unit.

4 Unscrew and remove the fixing sleeve (A).

5 Remove the faraday electrode from the flow cell, do not pull on the cable.

6 Remove the washer (C) and the o-ring [D] from the electrode body (B).

7 Unscrew the electrode tip containing the hollow electrode (G).

45

AMI Oxytrace QED

Maintenance

8 Clean the inner electrode (E) with a tissue and the hollow electrode with a pipe cleaner.

The electrode surfaces should be shining metallic after cleaning. Use a polishing detergent or a small amount of toothpaste.

9 Rinse all parts well with water.

10 Replace the O-ring and the washer if necessary.

11 Screw the hollow electrode finger-tight onto the electrode body.

12 Insert the faraday electrode into flow cell.

13 Tighten fixing sleeve firmly.

14 Feed the electrode cable through the cable gland of the faraday control unit.

15 Connect the electrode cable to terminal 5 (green) and terminal 6

(white).

16 Switch the instrument on.

17 Open the flow regulating valve and adjust the sample flow between 8 and 25 l/h.

46

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.5.

Calibration

The sensing part of the sensor must not be in direct contact with water!

In the wet flow cell, the atmosphere will be saturated with water vapor. This atmosphere will produce the most accurate calibration results.

The required time for a calibration depends mainly on the difference between temperature and oxygen content in the sample and in the air. It can take 15 – 20 minutes. This is also the case, if the electrolyte was exchanged.

As soon as the reading is stable, the microprocessor will store the calibration data in the memory.The end of the calibration is indicated on the display.

To perform a calibration proceed as follows:

1 Navigate to menu <Maintenance> /<Calibration>.

2 Press [Enter] to start the calibration and follow the dialog on the display.

Calibration

3.1.5

3 Stop the sample flow at with the flow regulating valve.

Close regulating valve to turn off sample flow.

<Enter> to continue

Calibration

Take sensor out of flow cell and dry membrane and sensor

3.1.5

4 Unscrew and remove the thread-

ed sleeve [A], see Electrolyte exchange, p. 42

.

5 Remove the oxygen sensor [B] from the flow cell [C].

6 Dry the sensor membrane and the flow cell with a soft paper tissue.

<Enter> to continue

A-96.250.531 / 300617

47

AMI Oxytrace QED

Maintenance

Calibration

Place the electrode into the wet flow cell at a slightly tilted angle.

3.1.5

<Enter> to continue

7 Place the electrode slightly tilted into the flow cell, so that the sensor cap rests on the rim for the O-ring.

A

B

48

Calibration

Saturation

Sat. Current

Progress

Calibration

Saturation

Sat. Current

3.1.1

98.7 %

32

A

3.1.1

98.7 %

32

A

A Tilted sensor

B Flow cell

8 Press [Enter] to start the calibration measurement.

The saturation should reach

100%, the saturation current should be about 22

A to 33

period, the calibration will be discarded.

9 Press [Enter] to confirm the calibration.

A.

If the measuring values are not stable during the measuring

Calibration Successful

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.6.

Zero-Verification

Swansensor Oxytrace G for the measurement of low oxygen content (< 1 ppb).

1 Calibrate the sensor according to chapter

Calibration, p. 47 .

2 Prepare a 5%-sodium sulfite solution with demineralized water.

3 Put the electrode into the sodium sulfite solution afterwards. Assure that there are no air bubbles in front of the sensor.

4 The measured value should now be < 1 ppb.

NOTICE: Depending on the state of the electrode this process can take several hours. In case of an electrode-refill, this can take days until the measured value is lower than 1 ppb.

A-96.250.531 / 300617

49

AMI Oxytrace QED

Maintenance

6.7.

Faraday Verification

Automatic

Verification

Manual

Verification

The Faraday verification works only for oxygen concentrations below 200 ppb. If automatic Faraday verification is enabled, a periodic check of the system is performed. A manual verification can be started for test purposes.

Per default the instrument performs an automatic faraday verification every 3 hours. To change the settings for automatic verification go to menu Faraday Parameter- Timer mode, see menu

4.1.3, p.

73 for details.

Possible settings are:

 off

 interval

 daily

 weekly

To start a manual verification:

1 Navigate to menu 3.2.2 <Maintenance>/<Service>/<Faraday

Verification>.

Faraday Verification

Current Value

3.3

1.62 ppb

Faraday Conc.

Progress

12.85 ppb

2 Press [Enter] to start the Faraday

Verification.

The verification then starts immediately.

<Enter> to stop

3 Press [Enter] to confirm the Faraday Verification.

Faraday Verification

Efficiency

Faraday Conc.

Progress

3.3

91.5 %

12.85 ppb

Done

Results are saved in the Verification history menu 2.2.1.5

50

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.8.

Quality Assurance of the Instrument

Quality assurance level

Every SWAN on-line instrument is equipped with integrated, autonomous quality assurance functions to survey the plausibility of each measurement.

For the AMI Oxytrace/AMI Oxytrace QED these are:

 continuous monitoring of sample flow

 continuous monitoring of the temperature inside the transmitter case

 periodic accuracy test with ultra high precision resistors

Further a manual, menu driven inspection procedure can be carried out using a certified reference instrument. Running at the same sampling point as an inspection equipment, the AMI Inspector Oxygen verifies the measuring results. After enabling the quality assurance procedure by defining the quality assurance level, the instrument reminds the user periodically to run the procedure and results are stored in a history for review.

Central feature of the quality assurance function is the assignment of the monitored process to a Quality assurance level.

There are three predefined levels plus a user level. Hereby the inspection interval, the deviation limits of temperature and measuring result between the inspection equipment and the monitoring instrument are defined.

 Level 1: Trend; Measurement used as an additional information to follow the process indicating trends.

Level 2: Standard; Monitoring of several parameters of a process (e.g. oxygen, saturation). In case of instrument failure, other parameters can be used for process monitoring.

 Level 3: Crucial; Monitoring of critical processes, value is used for control of another part or subsystem (valve, dosing unit, etc.).

Additional level:

 Quality level 4: User; User defined inspection interval, maximal deviation of temperature and measuring result.

A-96.250.531 / 300617

51

AMI Oxytrace QED

Maintenance

Quality Level

0: Off

1: Trend

2: Standard

3: Crucial

4: User max. deviation temperature [°C] a)

Off

0.5 °C

0.4 °C

0.3 °C

0 - 2°C

a) sample temperature must have 25°C +/- 5°C.

max. deviation result [%]

Off

10 %

5 %

5 %

0 - 20% min. inspection interval

Off annual quarterly monthly annual, quarterly, monthly

Procedure

The standard workflow contains following procedures:

1 Activate SWAN Quality assurance procedure

2 Pre-test

3 Connect instruments

4 Carry out comparison measurement

5 Completion of the measurement

NOTICE: The procedure should only be carried out through qualified personnel.

Materials / Inspection equipment:

Reference instrument: AMI INSPECTOR Oxygen

 Two tubes made of PA

6.8.1

Activate SWAN Quality assurance procedure

Enable quality assurance procedure at each instrument by selecting the quality level in menu 5.1.4.

The corresponding submenus are then activated.

NOTICE: The activation is necessary the first time only.

52

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.8.2

Pre-test

 Reference instrument: AMI INSPECTOR Oxygen

– Check certificate; reference instrument certificate not older than one year.

– Check battery; Battery of the AMI INSPECTOR Oxygen should be completely charged. Remaining operating time on display minimum 20 hours.

– Sensor is in working condition.

 On-line instrument: Monitor AMI Oxytrace QED

– Good order and condition; Flow cell free of particles,

Sensor surface free of deposits.

– Check message list; Review the message list in menu 1.3 and check for frequently alarms (as for example flow alarms). If alarms occur frequently remove cause before starting the procedure.

6.8.3

Connect the sample lines

The choice of sampling depends strongly on local conditions on site. Possible sampling:

 via sample point,

 via T-fitting or

 via piggyback/downstream

NOTICE:

• avoid ingress of air, use screwed fitting,

• sample as near as possible to the process monitor,

• wait approx. 10 minutes, whilst measurement is running, until measurement value and temperature are stabilized.

Example As an example following picture shows the connection of the reference instrument via T-fitting to the process monitor.

A-96.250.531 / 300617

53

AMI Oxytrace QED

Maintenance

54

A

B

C

D

E

F

G

A

B

C

D

Monitor AMI Oxytrace

AMI INSPECTOR Oxygen

Reference flow cell

On-line flow cell

E

F

G

Sample outlet

Sample inlet

T-fitting

1 Stop sample flow to the monitor AMI Oxytrace by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell.

2 Connect sample line of the monitor AMI Oxytrace [A] with the sample inlet of the reference instrument AMI INSPECTOR Oxygen [B]. Use the supplied tube, made of PA.

3 Connect sample outlet of the reference instrument AMI IN-

SPECTOR Oxygen to the sample outlet funnel of the monitor.

4 Switch on the AMI INSPECTOR Oxygen. Open the flow regulating valve and regulate the sample flow to 10 l/h. The actual flow is shown on the transmitter.

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.8.4

Carry out comparison measurement

The comparison measurement is menu driven. Start by selecting

Quality Assurance in menu 3.5 of the monitor AMI Oxytrace.

1 Navigate to menu Maintenance /Quality Assurance.

2 Press [Enter].

3 Follow the dialog on the Display.

Quality Assurance

- Carry out preparations

- Install Inspector

- Sample flow to 10 l/h

3.5.5

4 Carry out pre test preparations

Connect instruments.

Regulate sample flow to 10 l/h using the appropriate valve.

<Enter> to continue

Quality Assurance

Value O2

Value Temp.

Wait 10 Minutes

3.5.5

0.05 ppb

25.00

C

<Enter> to continue

5 Wait 10 minutes whilst measurement is running.

Press [Enter] to continue.

Quality Assurance

Value O2

Value Temp.

Inspector O2

Inspector Temp.

<Enter> to continue

3.5.3

0.05 ppb

25.00

C

0.06 ppb

25.0

C

6 Read the ppb value of the reference instrument and enter under

[

“Inspector” by using the [

] keys.

] or

7 Press [Enter] to confirm.

Quality Assurance

Value H2

Value Temp.

Inspector

Inspector Temp.

<Enter> to continue

3.5.4

0.05 ppb

25.00

C

0.06 ppb

25.0

C

Quality Assurance

Max. Dev. O2

Max. Dev. Temp.

Dev. O2

Dev. Temp.

QA-Check succesful

3.5.5

0.5 %

0.4 °C

0.1 %

0.4 °C

8 Read temperature value of the reference instrument and enter un-

[ der “Inspector Temp.” by using the

] or [ ] keys.

9 Press [Enter] to confirm.

10 Press [Enter] to continue.

The results are saved in QA-

History regardless if successful or not

A-96.250.531 / 300617

55

AMI Oxytrace QED

Maintenance

6.8.5

Completion of the measurement

1 Stop the sample flow to the AMI Oxytrace QED by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell again.

2 Close flow regulating valve of the AMI Inspector.

3 Disconnect the AMI Inspector by removing the tubes and connect the sample outlet of the Monitor AMI Oxytrace QED to the sample outlet funnel again.

4 Start sample flow again and regulate sample flow.

5 Shutdown the AMI INSPECTOR Oxygen.

If the AMI Inspector is put out of operation for a longer period of

time, see Longer Stop of Operation, p. 58

.

56

A-96.250.531 / 300617

AMI Oxytrace QED

Maintenance

6.9.

Replacing Fuses

WARNING

External Voltage.

External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.

 Make sure that the devices connected to the following contacts are disconnected from the power before continuing the installation.

– relay 1

– relay 2

– alarm relay

When a fuse has blown, find out the cause and fix it before replacing it with a new one.

Use tweezers or needle-nosed pliers to remove the defective fuse.

Use original fuses provided by SWAN only.

A B C D E F G

A-96.250.531 / 300617

E

F

G

A

B

C

D

1.6 AT/250V Instrument power supply

1.0 AT/250V Relay 1

1.0 AT/250V Relay 2

1.0 AT/250V Alarm relay

1.0 AF/125V Signal output 2

1.0 AF/125V Signal output 1

1.0 AF/125V Signal output 3

57

AMI Oxytrace QED

Maintenance

6.10. Longer Stop of Operation

1 Shut off power of the instrument.

2 Stop sample flow.

3 Remove the Swansensor Oxytrace G.

4 Clean the sensor with a soft tissue and rinse it with water afterwards.

5 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.

6 Fill the flow cell with water.

7 Install the Swansensor Oxytrace G.

58

A-96.250.531 / 300617

AMI Oxytrace QED

Error List

7.

Error List

Error

Non-fatal Error. Indicates an alarm if a programmed value is exceeded.

Such Errors are marked E0xx (bold and black).

Fatal Error (blinking symbol)

Control of dosing devices is interrupted.

The indicated measured values are possibly incorrect.

Fatal Errors are divided in the following two categories:

 Errors which disappear if correct measuring conditions are recovered (i.e. Sample Flow low).

Such Errors are marked E0xx (bold and orange)

 Errors which indicate a hardware failure of the instrument.

Such Errors are marked

E0xx

( bold and red )

HOLD

R1

R2

8 l/h

14:10:45

7.04 ppm

25.4°C

Error or fatal Error

Error not yet acknowledged.

Check Pending Errors 1.1.5 * and take corrective action.

Press [ENTER].

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

1

Navigate to menu Messages.

Press [ENTER].

Messages

Pending Errors

Maintenance List

Message List

1.1

Navigate to menu Pending Errors.

Press [ENTER].

A-96.250.531 / 300617

Pending Errors

Error Code

Alarm low

<Enter> to Acknowledge

1.1.5

E002

Press [ENTER] to acknowledge the

Pending Errors. The Error is reset and saved in the Message List.

* Menu numbers see

Program Overview, S. 63

59

AMI Oxytrace QED

Error List

60

Error

E001

E002

Description

Oxygen Alarm high

Oxygen Alarm low

E003

Saturation Alarm high

E004

E007

E008

E009

E010

E011

E012

Saturation Alarm low

Sample Temp. high

Sample Temp. low

Sample Flow high

Sample Flow low

Temp. shorted

Temp. disconnected

Corrective action

– check process

– check programmed value, see

5.3.1.1.1, S. 79

– check process

– check programmed value, see

5.3.1.1.25, S. 79

– check process

– check programmed value, see

5.3.1.4, S. 80

– check process

– check programmed value, see

5.3.1.4, S. 80

– check process

– check programmed value, see

5.3.1.3.1, S. 80

– check process

– check programmed value, see

5.3.1.3.25, S. 80

– check sample flow

– check programmed value, see

5.3.1.2.2, S. 80

– establish sample flow

– clean instrument

– check programmed value, see

5.3.1.2.35, S. 80

– Check wiring of sensor

– Check sensor

– Check wiring of sensor

– Check sensor

A-96.250.531 / 300617

AMI Oxytrace QED

Error List

Error

E013

E014

E017

E018

Description

Case Temp. high

Case Temp. low

Control Timeout

Faraday Efficiency

E019

Quality Assurance

E024

Input active

E026

E028

E030

E031

IC LM75

Signal output open

EEProm Frontend

Calibration Recout

Corrective action

– check case/environment temperature

– check programmed value,

see 5.3.1.5.1, S. 80

– check case/environment temperature

– check programmed value,

see 5.3.1.5.2, S. 80

– check control device or programming in

Installation, Relay contact, Relay 1/2

see 5.3.2 and 5.3.3, S. 81

– clean Faraday electrode, see

Maintenance of the Faraday electrode,

S. 45

– perform an air calibration, see

Calibration, S. 47

– perform maintenance of the oxygen

sensor, see Maintenance of the Oxygen

Sensor, S. 42

– Perform QA Procedure using a reference instrument, e.g. AMI Inspector

– See If Fault Yes is programmed in Menu

see 5.3.4, S. 85

– call service

– check wiring on signal outputs 1 and 2

– call service

– call service

A-96.250.531 / 300617

61

AMI Oxytrace QED

Error List

Error

E032

E033

E034

E065

Description

Wrong Frontend

Power-on

Power-down

Electrolyte depleted

Corrective action

– call service

– none, normal status

– none, normal status

– Refill electrolyte, see Electrolyte exchange, S. 42

62

A-96.250.531 / 300617

AMI Oxytrace QED

Program Overview

8.

Program Overview

For explanations about each parameter of the menus see

Program

List and Explanations, S. 69

.

Menu 1 Messages informs about pending errors and maintenance tasks and shows the error history. Password protection possible. No settings can be modified.

 Menu 2 Diagnostics is always accessible for everybody. No password protection. No settings can be modified.

 Menu 3 Maintenance is for service: Calibration, simulation of outputs and set time/date. Please protect with password.

 Menu 4 Operation is for the user, allowing to set limits, alarm values, etc. The presetting is done in the menu Installation

(only for the System engineer). Please protect with password.

 Menu 5 Installation: Defining assignment of all inputs and outputs, measuring parameters, interface, passwords, etc.

Menu for the system engineer. Password strongly recommended.

8.1.

Messages (Main Menu 1)

Pending Errors

1.1.5*

Pending Errors

1.1*

Maintenance List

1.2*

Message List

1.3*

Maintenance List

Number

Date, Time

1.2.5*

1.3.1*

* Menu numbers

A-96.250.531 / 300617

63

AMI Oxytrace QED

Program Overview

Identification

2.1*

8.2.

Diagnostics (Main Menu 2)

Desig.

Version

Factory Test

2.1.3*

AMI Oxytrace

6.20-06/16

Instrument

Motherboard

2.1.3.1*

Front End

Years / Days / Hours / Minutes / Seconds

Sensors

2.2*

Operating Time

2.1.4*

Oxytrace G

2.2.1*

Current Value

(Raw value tc)

Saturation

Cal. History

2.2.1.5*

Number

Date, Time

Sat. Current

Air pressure

2.2.2.1*

Sample

2.3*

I/O State

2.4*

Interface

2.5*

Miscellaneous

2.2.2*

QA History

2.2.3*

Sample ID

Temperature °C

Nt5K Ohm

Alarm Relay

Relay 1/2

Input

Signal Output 1/2

Protocol

USB Stick

Case Temp.

Air pressure

QA History

2.3.1*

2.4.1*

2.4.2*

2.5.1*

2.2.3.1*

* Menu numbers

2.1.4.1*

2.2.1.5.1*

64

A-96.250.531 / 300617

AMI Oxytrace QED

Program Overview

Calibration

3.1*

Service

3.2*

Simulation

3.3*

8.3.

Maintenance (Main Menu 3)

Calibration

3.1.5

Electrolyte

3.2.1*

Last filling

Remaining amount

Remaining time

New Filling

Faraday Verification

Current Value

3.2.2

Faraday Conc.

Alarm Relay

Relay 1

Relay 2

Signal Output 1

Signal Output 2

(Date), (Time)

Progress

3.2.1*

3.2.2*

3.2.3*

3.2.4*

3.2.5*

3.2.1.5*

Set Time

3.4*

Quality Assurance

3.5*

Quality Assurance 3.5.5*

* Menu numbers

A-96.250.531 / 300617

65

AMI Oxytrace QED

Program Overview

Sensors

4.1*

Relay Contacts

4.2*

8.4.

Operation (Main Menu 4)

Filter Time Const.

Hold after Cal.

4.1.1*

4.1.2*

Faraday Parameter

Mode

4.1.3*

Interval

Alarm Relay

4.2.1*

Delay

Signal Outputs

Output/Control

Alarm oxygene

4.2.1.1*

Logger

4.3*

Relay 1/2

4.2.2* - 4.2.3*

Input

4.2.4*

Log Interval

Clear Logger

Eject USB Stick

Alarm Saturation

4.2.1.2*

Setpoint

Hysteresis

Delay

Active

Signal Outputs

Output / Control

Fault

Delay

4.3.1*

4.3.2*

4.3.3*

4.1.3.1*

4.1.3.20*

4.1.3.3*

4.1.3.4*

4.1.3.5*

Alarm High

Alarm Low

Hysteresis

Delay

Alarm High

Alarm Low

Hysteresis

Delay

4.2.x.100*

4.2.x.200*

4.2.x.30*

4.2.4.1*

4.2.4.2*

4.2.4.3*

4.2.4.4*

4.2.4.5*

* Menu numbers

4.2.1.1.1*

4.2.1.1.25*

4.2.1.1.35*

4.2.1.1.45*

4.2.1.2.1*

4.2.1.2.25*

4.2.1.2.35*

4.2.1.2.45*

66

A-96.250.531 / 300617

AMI Oxytrace QED

Program Overview

8.5.

Installation (Main Menu 5)

Sensors

5.1*

Miscellaneous

5.1.1*

Flow

Offset

Quality Assurance

Level

5.1.2*

Signal Outputs Signal Output 1/2

5.2* 5.2.1* - 5.2.2*

Parameter

Current Loop

Function

Scaling

Relay Contacts

5.3*

Alarm Relay

5.3.1*

5.2.x.40

Alarm oxygen

5.3.1.1*

5.1.1.1*

5.1.1.2*

5.1.2.1*

Relay 1 and 2

5.3.2* and 5.3.3*

Input

5.3.4*

Sample Flow

5.3.1.2*

Sample Temp.

5.3.1.3*

Alarm Saturation

5.3.1.4*

Case Temp.

5.3.1.5*

Function

Parameter

Setpoint

Hysteresis

Delay

Active

Signal Outputs

Output/Control

Fault

Delay

* Menu numbers

5.2.1.1 - 5.2.2.1*

5.2.1.2 - 5.2.2.2*

5.2.1.3 - 5.2.2.3*

Range Low

Range High

Alarm High

Alarm Low

Hysteresis

Delay

Flow Alarm

Alarm High

Alarm Low

Alarm High

Alarm Low

Alarm High

Alarm Low

Hysteresis

Delay

Case Temp. high

Case Temp. low

5.3.2.1–5.3.3.1*

5.3.2.20–5.3.3.20*

5.3.2.300–5.3.3.301*

5.3.2.400–5.3.3.401*

5.3.2.50–5.3.3.50*

5.3.4.1*

5.3.4.2*

5.3.4.3*

5.3.4.4*

5.3.4.5*

5.2.x.40.10/11*

5.2.x.40.20/21*

5.3.1.1.1*

5.3.1.1.25

5.3.1.1.35

5.3.1.1.45

5.3.1.2.1

5.3.1.2.2*

5.3.1.2.35*

5.3.1.3.1*

5.3.1.3.25*

5.3.1.4.1*

5.3.1.4.25

5.3.1.4.35

5.3.1.4.45

5.3.1.5.1*

5.3.1.5.2*

A-96.250.531 / 300617

67

AMI Oxytrace QED

Program Overview

Miscellaneous

5.4*

Interface

5.5*

Language

Set defaults

Load Firmware

Password

5.4.4*

5.4.1*

5.4.2*

5.4.3*

Messages

Maintenance

Operation

Sample ID

Installation

5.4.5*

Line Break Detection

5.4.6*

Protocol

Device Address

Baud Rate

Parity

5.5.1*

5.5.21*

5.5.31*

5.5.41*

5.4.4.1*

5.4.4.2*

5.4.4.3*

5.4.4.4*

* Menu numbers

(only with RS485 interface)

68

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

9.

Program List and Explanations

1 Messages

1.1 Pending Errors

1.1.5

Provides the list of active errors with their status (active, acknowledged). If an active error is acknowledged, the alarm relay is active again. Cleared errors are moved to the Message list.

1.2 Maintenance List

1.2.5

Provides the list of necessary maintenance. Cleared maintenance messages are moved to the Message list.

1.3 Message List

1.3.1

Shows the error history: Error code, date / time of issue and status

(active, acknowledged, cleared). 65 errors are memorized. Then the oldest error is cleared to save the newest error (circular buffer).

2 Diagnostics

In diagnostics mode, the values can only be viewed, not modified.

2.1 Identification

Desig.: Designation of the instrument.

Version: Firmware of instrument (e.g. 6.20-06/16)

2.1.3

Factory Test: Test date of the Instrument and Motherboard.

2.1.4

Operating Time: Shows the operating time in Years, Days, Hours,

Minutes and Seconds.

2.2 Sensors

2.2.1

Oxytrace G

Current value: Shows the actual measuring value in ppb.

Raw value tc:

Shows the actual temperature compensated measuring value in mA.

Saturation

Shows the actual saturation in %

A-96.250.531 / 300617

69

AMI Oxytrace QED

Program List and Explanations

2.2.1.4

Cal. History

Review the diagnostic values of the last calibration of the oxygen sensor. Max. 64 data records are memorized.

Number: Calibration counter.

Date, Time: Date and time of the calibration.

Sat. Current: Saturation current at that time of calibration.

Air pressure: Air pressure at that time of calibration.

2.2.2

Miscellaneous:

2.2.2.1

Case Temp: Shows the actual temperature in °C inside the transmitter.

Air pressure: Shows the actual air pressure in hPa

2.2.3

QA History

Review QA values (Number, Date, Time, Deviation oxygen, Deviation Temperature, Status of QA check) of the last quality assurance procedures.

2.3 Sample

2.3.301

Sample ID: Shows the assigned sample identification. This identification is defined by the user to identify the location of the sample

Temperature: Shows temperature in °C.

(Nt5K): Shows raw value of the temperature in 

Sample Flow: Shows the sample flow in l/h

(Raw value) Shows the sample flow in Hz

2.4 I/O State

Shows actual status of all in- and outputs.

2.4.1

Alarm Relay:

Relay 1 and 2:

Input:

Active or inactive.

Active or inactive.

Open or closed.

Signal Output 1 and 2: Actual current in mA

Signal Output 3:

Actual current in mA (if option is installed)

2.5 Interface

2.5.1

Only available if optional interface is installed.

Shows the programmed communication settings.

70

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

3 Maintenance

3.1 Calibration

3.1.1

Start a calibration and follow the instructions on the screen. Displayed values are saturation in % and the saturation current in mA.

The indication bar shows the progress. Detailed explanation see

Calibration, p. 47

.

3.2 Service

3.2.1

Electrolyte

Last Filling: Shows the date of the last filling of electrolyte.

Remaining Amount: Remaining amount of electrolyte in %.

Remaining Time: Remaining time in days until electrolyte exchange recommended.

3.2.1.5

New Filling: Select “Yes” after electrolyte exchange to reset the counter.

3.2.2

Faraday Verification

Start a manual faraday verification. Displayed values are current value in ppb and the faraday concentration in %.

Current value: Measuring value in ppb

Faraday conc.: Oxygen concentration in % after activating the faraday verification.

Progress: The progress bar shows the progress of the faraday verification.

A-96.250.531 / 300617

71

AMI Oxytrace QED

Program List and Explanations

3.3 Simulation

In this menu the following relays and signal outputs can be tested:

 Alarm relay

 Relay 1 and 2

Signal output 1 and 2

 Signal output 3 (if option is installed)

Select a relay or signal output with the [ ] or [ ] keys, press the [Enter]> key to confirm. Then change the value with the

[ ] or [ ] keys. After confirming the setting with the [Enter] key, the value is simulated by the relay/signal output.

Alarm Relay:

Relay 1 and 2:

Input:

Active or inactive.

Active or inactive.

Open or closed.

Signal Output 1 and 2: Current in mA

Signal Output 3:

Current in mA (if option is installed)

At the absence of any key activities, the instrument will switch back to normal mode after 20 min. If you quit the menu, all simulated values will be reset.

3.4 Set Time

Adjust date and time.

3.5 Quality Assurance

Performs a Quality Assurance according to your settings. Follow

the commands on the screen. Detailed explanation see Quality Assurance of the Instrument, p. 51

.

72

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

4 Operation

4.1 Sensors

4.1.1

Filter Time Constant: Used to damp noisy signals. The higher the filter time constant, the slower the system reacts to changes of the measured value.

Range: 5–300 Sec

4.1.2

Hold after Cal: Delay permitting the instrument to stabilize again after calibration. During calibration plus hold-time, the signal outputs are frozen (held on last valid value), alarm values, limits are not active.

Range: 0–6‘000 Sec

4.1.3

Faraday Parameter

4.1.3.1

Mode: Can be set to Interval, daily, weekly or off. If Mode is set to

“Off”, no further settings are available. The Faraday Verification has to be started manually.

4.1.3.20

Interval: The interval can be set between 1h and 12h

4.1.3.21

Start Time: Start time appears if Mode is set to daily, how to set the start time see 5.3.2.341, p. 84 .

4.1.3.22

Calendar: Calendar appears if Mode is set to weekly, how to set the Calendar see

5.3.2.342, p. 85

.

4.1.3.3

Delay: during Faraday Verification plus the delay time the signal and control outputs are held in the operating mode programmed below.

Range: 0–6’000 Sec.

4.1.3.4

Signal Outputs: Select operating mode of the signal output:

Cont.:

Signal outputs continue to issue the measured value.

Hold:

Signal outputs hold the last valid measured value.

Measurement is interrupted. Errors, except fatal errors, are not issued.

Off:

Signal outputs are switched off (set to 0 or 4 mA).

Errors, except fatal errors, are not issued.

4.1.3.5

Output/Control: Select operating mode of the controller output:

Cont.:

Controller continues normally.

Hold:

Controller continues based on the last valid value.

Off:

Controller is switched off.

A-96.250.531 / 300617

73

AMI Oxytrace QED

Program List and Explanations

4.2 Relay Contacts

See Relay Contacts, p. 30

4.3 Logger

The instrument is equipped with an internal logger. The data can be copied to the USB stick installed in the transmitter.

The logger can save approx. 1500 data records. The Records consists of: Date, time, alarms, measuring values, raw values, case temperature, flow.

4.3.1

Log Interval: Select a convenient log interval. Consult the table below to estimate the max logging time. When the logging buffer is full, the oldest data record is erased to make room for the newest one (circular buffer).

Range: 1 Second to 1 hour

Interval

Time

1 s 5 s

25 min 2 h

1 min

25 h

5 min

5 d

10 min

10 d

30 min

31 d

1 h

62 d

4.3.2

Clear Logger: If confirmed with yes, the complete logger data is deleted. A new data series is started.

5 Installation

5.1 Sensors

5.1.1

Miscellaneous

5.1.1.1

Flow: If a flow cell without flow measurement (e.g. B-Flow) is used, choose none. With flow measurement select Q-Flow

5.1.1.2

O2 Offset: Manual, small correction of the offset.

Range -5 to 5 ppb

5.1.2

Quality Assurance: Switch the Quality Assurance on or off.

5.1.2.1

Level: Select quality level:

 Level 0: Off

Quality assurance procedure switched off. Any additional QA menus are hidden.

 Level 1: Trend

Level 2: Standard

 Level 3: Crucial

 Level 4: User

Edit user specific limits in menu 5.1.2.2

74

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.2 Signal Outputs

NOTICE: The navigation in the menu <Signal Output 1> and

<Signal Output 2> is equal. For reason of simplicity only the menu numbers of Signal Output 1 are used in the following.

5.2.1 and 5.2.2

Signal Output 1 and 2: Assign process value, the current loop range and a function to each signal output.

5.2.1.1

Parameter: Assign one of the process values to the signal output.

Available values:

 Oxygen

 Temperature

Sample Flow (if a flow sensor is selected)

 Saturation

5.2.1.2

Current Loop: Select the current range of the signal output.

Make sure the connected device works with the same current range.

Available ranges: 0–20 mA or 4–20 mA

5.2.1.3

Function: Define if the signal output is used to transmit a process value or to drive a control unit. Available functions are:

 Linear, bilinear or logarithmic for process values.

See As process values, p. 75

 Control upwards or control downwards for controllers.

See As control output, p. 77

As process values

The process value can be represented in 3 ways: linear, bilinear or logarithmic. See graphs below.

[mA]

20

B A

10 12

A

B

0 / 4

0.0

linear

0.1

bilinear

0.2

0.3

0.4

X

0.5

X Measured value

A-96.250.531 / 300617

75

AMI Oxytrace QED

Program List and Explanations

[mA]

20

10 12

0 / 4

1

0

4 6

10

1

X

2

100

2

1’000 10’000

3 4

X Measured value (logarithmic)

5.2.1.40

Scaling: Enter beginning and end point (Range low & high) of the linear or logarithmic scale. In addition, the midpoint for the bilinear scale.

Parameter: Oxygen.

Range low: 0.00 ppb –20.00 ppm

Range high: 0.00 ppb –20.00 ppm

Parameter: Temperature

Range low: -30 to + 130 °C

Range high: -30 to + 130 °C

Parameter: Sample flow

Range low: 0–50 l/h

Range high: 0–50 l/h

Parameter: Saturation

Range low: 0–200 %

Range high: 0–200 %

76

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

As control output

Signal outputs can be used for driving control units. We distinguish different kinds of controls:

P-controller: The controller action is proportional to the deviation from the setpoint. The controller is characterized by the

P-Band. In the steady-state, the setpoint will never be reached. The deviation is called steady-state error.

Parameters: setpoint, P-Band

PI-controller: The combination of a P-controller with an

I-controller will minimize the steady-state error. If the reset time is set to zero, the I-controller is switched off.

Parameters: setpoint, P-Band, reset time.

PD-controller: The combination of a P-controller with a

D-controller will minimize the response time to a fast change of the process value. If the derivative time is set to zero, the

D-controller is switched off.

Parameters: setpoint, P-Band, derivative time.

PID-controller: The combination of a P-, an I - and a D-controller allows a proper control of the process.

Parameters: setpoint, P-Band, reset time, derivative time.

Ziegler-Nichols method for the optimization of a PID controller:

Parameters: Setpoint, P-Band, Reset time, Derivative time

Y

B

A

A-96.250.531 / 300617

X

a

A

B

X

L

Response to maximum control output

Tangent on the inflection point

Time

Xp

Tn

Tv

= 1.2/a

= 2L

= L/2

The point of intersection of the tangent with the respective axis will result in the parameters a and L.

Consult the manual of the control unit for connecting and programming details. Choose control upwards or downwards.

77

AMI Oxytrace QED

Program List and Explanations

78

Setpoint: User defined precess value for the selected parameter.

P-Band: Range below (upwards control) or above (downwards control) the set-point, within which the dosing intensity is reduced from

100% to 0% to reach the set-point without overshooting.

5.2.1.43

Control Parameters: if Parameter = Oxygen

5.2.1.43.10

Setopint:

Range: 0.00 ppb –20.00 ppm

5.2.1.43.20

P-Band:

Range: 0.00 ppb –20.00 ppm

5.2.1.43

Control Parameters: if Parameter = Temperature

5.2.1.43.11

Setopint:

Range: -30 to + 130 °C

5.2.1.43.21

P-Band:

Range: 0 to + 100 °C

5.2.1.43

Control Parameters: if Parameter = Sample flow

5.2.1.43.12

Setopint:

Range: 0–50 l/h

5.2.1.43.22

P-Band:

Range: 0–50 l/h

5.2.1.43

Control Parameters: if Parameter = Saturation

5.2.1.43.13

Setopint:

Range: 0–200%

5.2.1.43.23

P-Band:

Range: 0–200%

5.2.1.43.3

Reset time: The reset time is the time till the step response of a single I-controller will reach the same value as it will be suddenly reached by a P-controller.

Range: 0–9’000 sec

5.2.1.43.4

Derivative time: The derivative time is the time till the ramp response of a single P-controller will reach the same value as it will be suddenly reached by a D-controller.

Range: 0–9’000 sec

5.2.1.43.5

Control timeout: If a controller action (dosing intensity) is constantly over 90% during a defined period of time and the process value does not come closer to the setpoint, the dosing process will be stopped for safety reasons.

Range: 0–720 min

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.3 Relay Contacts

5.3.1

Alarm Relay: The alarm relay is used as cumulative error indicator.

Under normal operating conditions the contact is active.

The contact is inactive at:

 Power loss

 Detection of system faults like defective sensors or electronic parts

High case temperature

 Process values out of programmed ranges.

Program alarm levels for the following parameters:

 Oxygen

 Temperature

 Sample Flow

 Case Temperature high

 Case Temperature low

5.3.1.1

Alarm oxygen

5.3.1.1.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E001 is displayed in the message list.

Range: 0.00 ppb –20.00 ppm

5.3.1.1.25

Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E002 is displayed in the message list.

Range: 0.00 ppb –20.00 ppm

5.3.1.1.35

Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.

Range. 0.00 ppb –20.00 ppm

5.3.1.1.45

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range: 0–28‘800 Sec

5.3.1.2

Sample Flow: Define at which sample flow a flow alarm should be issued.

5.3.1.2.1

Flow Alarm: Program if the alarm relay should be activated if there is a flow alarm. Choose between yes or no. The flow alarm will always be indicated in the display, pending error list, saved in the message list and the logger.

Available values: Yes or no

NOTICE: Sufficient flow is essential for a correct measurement.

We recommend to program yes.

A-96.250.531 / 300617

79

AMI Oxytrace QED

Program List and Explanations

80

5.3.1.2.2

Alarm High: If the measuring values rises above the programmed value E009 will be issued.

Range: 12–50 l/h

5.3.1.2.35

Alarm Low: If the measuring values falls below the programmed value E010 will be issued.

Range: 8–11 l/h

5.3.1.3

Sample Temp.: Define at which sample temperature an alarm should be issued.

5.3.1.3.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E007 is issued.

Range: 30–100 °C

5.3.1.3.25

Alarm Low: If the measured value rises above the alarm high value, the alarm relay is activated and E008 is issued.

Range: -10 to + 20 °C

5.3.1.4

Alarm Saturation

5.3.1.4.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E003, is displayed in the message list.

Range: 0.00 –200 %

5.3.1.4.25

Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E004 is displayed in the message list.

Range: 0.00 –200 %

5.3.1.4.35

Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.

Range. 0.00 –200 %

5.3.1.4.45

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range: 0–28‘800 Sec

5.3.1.5

Case Temp.

5.3.1.5.1

Case Temp. high: Set the alarm high value for temperature of electronics housing. If the value rises above the programmed value

E013 is issued.

Range: 30–75 °C

5.3.1.5.2

Case Temp. low: Set the alarm low value for temperature of electronics housing. If the value falls below the programmed value

E014 is issued.

Range: -10 to + 20 °C

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.3.2 and 5.3.3

Relay 1 and 2: The function of relay contacts 1 or 2 are defined by the user

NOTICE: The navigation in the menu <Relay 1> and <Relay 2> is equal. For reason of simplicity only the menu numbers of

Relay 1 are used in the following.

1 First select the functions as:

- Limit upper/lower,

- Control upwards/downwards,

- Timer

- Fieldbus,

2 Then enter the necessary data depending on the selected function. The same values may also be entered in menu 4.2 Relay

Contacts, p. 74

5.3.2.1

Function = Limit upper/lower:

When the relays are used as upper or lower limit switches, program the following:

5.3.2.20

Parameter: choose one of the following process values

 Oxygen

 Temperature

Sample Flow

 Saturation

5.3.2.300

Setpoint: If the measured value rises above respectively falls below the set-point, the relay is activated.

Parameter

Oxygen

Temperature

Sample flow

Saturation

Range

0.00 ppb –20.00 ppm

-30 to + 130 °C

0–50 l/h

0–200%

5.3.2.400

Hysteresis: within the hysteresis range, the relay does not switch.

This prevents damage of relay contacts when the measured value fluctuates around the alarm value.

Parameter

Oxygen

Temperature

Sample flow

Saturation

Range

0.00 ppb –20.00 ppm

0–100 °C

0–50 l/h

0–200%

A-96.250.531 / 300617

81

AMI Oxytrace QED

Program List and Explanations

5.3.2.50

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range. 0–600 Sec

5.3.2.1

Function = Control upwards/downwards

The relays may be used to drive control units such as solenoid valves, membrane dosing pumps or motor valves. When driving a motor valve both relays are needed, relay 1 to open and relay 2 to close the valve.

5.3.2.22

Parameter: choose one of the following process values

 Oxygen

 Temperature

Sample Flow

 Saturation

5.3.2.32

Settings

Choose the respective actuator:

 Time proportional

 Frequency

 Motor valve

Actuator = Time proportional

Examples of metering devices that are driven time proportional are solenoid valves, peristaltic pumps.

Dosing is controlled by the operating time.

5.3.2.32.20

Cycle time: duration of one control cycle (on/off change).

Range: 0–600 sec.

5.3.2.32.30

Response time: Minimal time the metering device needs to react.

Range: 0–240 sec.

5.3.2.32.4

Control Parameters:

Range for each Parameter same as

5.2.1.43, p. 78

Actuator = Frequency

Examples of metering devices that are pulse frequency driven are the classic membrane pumps with a potential free triggering input.

Dosing is controlled by the repetition speed of dosing shots.

5.3.2.32.21

Pulse frequency: Max. pulses per minute the device is able to respond to. Range: 20–300/min.

82

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.3.2.32.31

Control Parameters:

Range for each Parameter same as 5.2.1.43, p. 78

Actuator = Motor valve

Dosing is controlled by the position of a motor driven mixing valve.

5.3.2.32.22

Run time: Time needed to open a completely closed valve

Range: 5–300 Sec.

5.3.2.32.32

Neutral zone: Minimal response time in % of the runtime. If the requested dosing output is smaller than the response time, no change will take place.

Range: 1–20 %

5.3.2.32.4

Control Parameters:

Range for each Parameter same as 5.2.1.43, p. 78

5.3.2.1

Function = Timer

The relay will be activated repetitively depending on the programmed time scheme.

5.3.2.24

Mode: Operating mode (interval, daily, weekly)

5.3.2.24

Interval

5.3.2.340

Interval: The interval can be programmed within a range of 1–1’440 min.

5.3.2.44

Run Time: Enter the time the relay stays active.

Range: 5–32’400 sec.

5.3.2.54

Delay: during run time plus the delay time the signal and control outputs are held in the operating mode programmed below.

Range: 0–6’000 Sec.

A-96.250.531 / 300617

83

AMI Oxytrace QED

Program List and Explanations

5.3.2.6

Signal Outputs: Select operating mode of the signal output:

Cont.:

Signal outputs continue to issue the measured value.

Hold:

Signal outputs hold the last valid measured value.

Measurement is interrupted. Errors, except fatal errors, are not issued.

Off:

Signal outputs are switched off (set to 0 or 4 mA).

Errors, except fatal errors, are not issued.

5.3.2.7

Output/Control: Select operating mode of the controller output:

Cont.:

Controller continues normally.

Hold:

Controller continues based on the last valid value.

Off:

Controller is switched off.

5.3.2.24

daily

The relay contact can be activated daily, at any time of a day.

5.3.2.341

Start time: to set the start time proceed as follows:

1 Press [Enter], to set the hours.

2 Set the hour with the [ ] or [ ] keys.

3 Press [Enter], to set the minutes.

4 Set the minutes with the [ ] or [

5 Press [Enter], to set the seconds.

6 Set the seconds with the [ ] or [

] keys.

] keys.

Range: 00:00:00–23:59:59

5.3.2.44

Run Time: see Interval

5.3.2.54

Delay: see Interval

5.3.2.6

Signal Outputs: see Interval

5.3.2.7

Output/Control: see Interval

5.3.2.24

weekly

The relay contact can be activated at one or several days, of a week. The daily starting time is valid for all days.

84

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.3.2.342

Calendar:

5.3.2.342.1

Start time: The programmed start time is valid for each of the programmed days. To set the start time see 5.3.2.341, p. 84 .

Range: 00:00:00–23:59:59

5.3.2.342.2

Monday: Possible settings, on or off to

5.3.2.342.8

Sunday: Possible settings, on or off

5.3.2.44

Run Time: see Interval

5.3.2.54

Delay: see Interval

5.3.2.6

Signal Outputs: see Interval

5.3.2.7

Output/Control: see Interval

5.3.2.1

Function = Fieldbus

The relay will be switched via the Profibus input. No further parameters are needed.

5.3.4

Input: The functions of the relays and signal outputs can be defined depending on the position of the input contact, i.e. no function, closed or open.

5.3.4.1

Active: Define when the input should be active:

No:

Input is never active.

When closed Input is active if the input relay is closed

When open:

Input is active if the input relay is open

5.3.4.2

Signal Outputs: Select the operation mode of the signal outputs when the relay is active:

Cont.:

Hold:

Off:

Signal outputs continue to issue the measured value.

Signal outputs issue the last valid measured value.

Measurement is interrupted. Errors, except fatal errors, are not issued.

Set to 0 or 4 mA respectively. Errors, except fatal errors, are not issued.

A-96.250.531 / 300617

85

AMI Oxytrace QED

Program List and Explanations

5.3.4.3

Output/Control: (relay or signal output):

Cont.:

Hold:

Off:

Controller continues normally.

Controller continues on the last valid value.

Controller is switched off.

5.3.4.4

Fault:

No:

Yes:

No message is issued in pending error list and the alarm relay does not close when input is active. Message E024 is stored in the message list.

Message E024 is issued and stored in the message list. The Alarm relay closes when input is active.

5.3.4.5

Delay: Time which the instrument waits, after the input is deactivated, before returning to normal operation.

Range: 0–6‘000 Sec

86

A-96.250.531 / 300617

AMI Oxytrace QED

Program List and Explanations

5.4 Miscellaneous

5.4.1

Language: Set the desired language.

Available settings: German /English/French/Spanish

5.4.2

Set defaults: Reset the instrument to factory default values in three different ways:

Calibration: Sets calibration values back to default. All other values are kept in memory.

In parts: Communication parameters are kept in memory. All other values are set back to default values.

Completely: Sets back all values including communication parameters.

5.4.3

Load Firmware: Firmware updates should be done by instructed service personnel only.

5.4.4

Password: Select a password different from 0000 to prevent unauthorized access to the menus “Messages”, “Maintenance”, “Operation” and “Installation”.

Each menu may be protected by a different password.

If you forgot the passwords, contact the closest SWAN representative.

5.4.5

Sample ID: Identify the process value with any meaning full text, such as KKS number.

5.4.6

Line Break Detection: If activated, error message E028 is shown in case of line break on signal outputs 1 and 2.

A-96.250.531 / 300617

87

AMI Oxytrace QED

Program List and Explanations

5.5 Interface

5.5.1

5.5.20

5.5.30

5.5.40

5.5.1

5.5.21

5.5.31

5.5.41

5.5.1

5.5.1

5.5.24

Select one of the following communication protocols. Depending on your selection, different parameters must be defined.

Protocol: Profibus

Device address:

ID-Nr.:

Local operation:

Range: 0–126

Range: Analyzer; Manufacturer; Multivariable

Range: Enabled, Disabled

Protocol: Modbus RTU

Device address:

Range: 0–126

Baud Rate:

Range: 1200–115 200 Baud

Parity:

Range: none, even, odd

Protocol: USB-Stick

Only visible if an USB interface is installed. No further settings are possible.

Protocol: HART

Device address:

Range: 0–63

88

A-96.250.531 / 300617

AMI Oxytrace QED

Material Safety Data sheets

10. Material Safety Data sheets

10.1. Reagents

Catalogue No.:

Product name:

102751.10

Filling solution 1ALK

Part of: A-87.290.060

Download

MSDS

The current Material Safety Data Sheets (MSDS) for the above listed Reagents are available for downloading at www.swan.ch.

A-96.250.531 / 300617

89

AMI Oxytrace QED

Default Values

11.

Default Values

Operation:

Sensors: Filter Time Const.:..................................................................10 Sec

Hold after Cal.:.....................................................................300 Sec

Faraday Parameter:

Mode: ................................................................................ Interval

Interval:..............................................................................3 hours

Delay .................................................................................. 60 sec

Signal Outputs........................................................................ hold

Output/Control ....................................................................... hold

Alarm Relay ...................................................................... same as in Installation

Relay 1/2 ...................................................................... same as in Installation

Input ...................................................................... same as in Installation

Logger: Logger Interval:.............................................................. 30 Minutes

Clear Logger: ................................................................................no

Installation:

Sensors Miscellaneous:

Flow:.................................................................................. Q-Flow

O2 Offset: ..........................................................................0.0 ppb

Quality Assurance:

Level:.................................................................................... 0: Off

Signal Output 1 Parameter: ........................................................................... Oxygen

Current loop: .....................................................................4 –20 mA

Function: .................................................................................. linear

Scaling: Range low: ............................................................0.00 ppb

Scaling: Range high:........................................................10.00 ppm

Signal Output 2 Parameter: ....................................................................Temperature

Current loop: .....................................................................4 –20 mA

Function: .................................................................................. linear

Scaling: Range low: ................................................................0.0 °C

Scaling: Range high:.............................................................50.0 °C

Alarm Relay: Alarm oxygen; Alarm high:...............................................10.00 ppm

Alarm oxygen; Alarm low: ...................................................0.00 ppb

Alarm oxygen; Hysteresis: ...................................................100 ppb

Alarm oxygen; Delay:...............................................................5 Sec

Sample Flow, Flow Alarm: .......................................................... yes

Sample Flow, Alarm high: .................................................... 25.0 l/h

Sample Flow, Alarm low:........................................................ 8.0 l/h

90

A-96.250.531 / 300617

AMI Oxytrace QED

Default Values

Sample Temp., Alarm High: ..................................................... 50 °C

Sample Temp., Alarm Low:........................................................ 0 °C

Alarm Saturation; Alarm high..................................................120 %

Alarm Saturation; Alarm low ....................................................0.0 %

Alarm Saturation; Hysteresis ......................................................2 %

Alarm Saturation; Delay...........................................................5 Sec

Case temp. high: ..................................................................... 65 °C

Case temp. low:......................................................................... 0 °C

Relay 1 Function:.......................................................................... limit upper

Parameter:............................................................................Oxygen

Setpoint: .......................................................................... 10.00 ppm

Hysteresis:........................................................................... 100 ppb

Delay: ....................................................................................30 Sec

Relay 2 Function:.......................................................................... limit upper

Parameter:....................................................................Temperature

Setpoint: .................................................................................. 50 °C

Hysteresis:.............................................................................. 1.0 °C

Delay: ....................................................................................30 Sec

If Function = Control upw. or dnw:

Parameter:............................................................................Oxygen

Settings: Actuator: .......................................................... Frequency

Settings: Pulse Frequency: .............................................120/min.

Settings: Control Parameters: Setpoint:....................... 10.00 ppm

Settings: Control Parameters: P-band: ......................... 100.0 ppb

Settings: Control Parameters: Reset time:...........................0 Sec

Settings: Control Parameters: Derivative Time: ...................0 Sec

Settings: Control Parameters: Control Timeout:.................. 0 Min.

Settings: Act. Time prop.: Cycle time: ................................... 60 s

Settings: Act. Time prop.: Response time: ............................ 10 s

Settings: Act. Motor valve: Run time: .................................... 60 s

Settings: Act. Motor valve: Neutral zone: ................................5%

If Function = Timer:

Mode:.................................................................................... Interval

Interval: ................................................................................ 1 min

Mode: ........................................................................................daily

Start time:........................................................................ 00.00.00

Mode:..................................................................................... weekly

Calendar; Start time: ....................................................... 00.00.00

Calendar; Monday to Sunday:.................................................. Off

Run time: ...............................................................................10 Sec

Delay: ......................................................................................5 Sec

A-96.250.531 / 300617

91

AMI Oxytrace QED

Default Values

Signal output:............................................................................. cont

Output/Control: .......................................................................... cont

Input: Active ............................................................................when closed

Signal Outputs ........................................................................... hold

Output/Control .............................................................................. off

Fault..............................................................................................no

Delay......................................................................................10 Sec

Miscellaneous Language:............................................................................. English

Set default:....................................................................................no

Load firmware: ..............................................................................no

Password: ........................................................... for all modes 0000

Sample ID: ....................................................................... - - - - - - - -

Line break detection .....................................................................no

Interface Protocol:....................................................... depending on interface

92

A-96.250.531 / 300617

AMI Oxytrace QED

Index

12. Index

A

Alarm Relay

. . . . . . .

10

,

30

C

Cable thicknesses

Calendar

. . . . . .

25

. . . . . . . . . . .

85

Changing values

. . . . . . .

40

D

Default Values

. . . . . . . .

90

E

Electrolyte exchange

. . . .

42

Establish Sample Flow

. . .

36

F

Fluidics

. . . . . . . . . .

11

12

I

Input

. . . . . . . . . . . .

10

,

30

Instrument Setup

. . . . . .

55

Interface

HART

USB

. . . . . . . . . .

35

Profibus, Modbus

. . . .

34

. . . . . . . . . . .

35

L

Longer Stop of Operation

. .

58

M

Mounting requirements

. . .

22

P

P-Band

. . . . . . . . . . . .

78

Power Supply

PROFIBUS

. . . . . . . .

13

. . . . . . . . . .

34

R

Relay

. . . . . . . . . . . . .

10

S

Sample Flow

. . . . . . . . .

36

Sample requirements

Signal Outputs

Software

. . . .

13

. . . . . .

10

,

33

. . . . . . . . . . .

39

T

Temperature compensation

Terminals 27

.

8

28

,

30

31

,

33

34

U

USB Interface

. . . . . . . .

35

W

Wire

. . . . . . . . . . . . . .

25

Z

Zero-Verification

. . . . . . .

49

A-96.250.531 / 300617

93

AMI Oxytrace QED

Notes

13. Notes

94

A-96.250.531 / 300617

AMI Oxytrace QED

Notes

A-96.250.531 / 300617

95

AMI Oxytrace QED

SWAN

is represented worldwide by subsidiary companies and distributors.

cooperates with independent representatives all over the world.

SWAN Products

Analytical Instruments for:

High Purity Water

Feedwater, Steam and Condensate

Potable Water

Pool and Sanitary Water

Cooling Water

Waste Water and Effluents

Made in Switzerland

96

A-96.250.531 / 300617

advertisement

Related manuals

Download PDF

advertisement