advertisement
AMI INSPECTOR
Oxygen
Version 6.00 and higher
A-96.250.701 / 140716
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
Title: Monitor AMI INSPECTOR Oxygen Operator’s Manual
ID: A-96.250.701
Revision Issue
00 Sep. 2011 First Edition
01
02
Nov. 2013
Jan. 2016
New main board, integrated USB interface
AMI Inspector Version 2-A (with AMIAKKU mainboard) and Firmware version 6.00
© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved subject to change without notice.
AMI INSPECTOR Oxygen
Table of Contents
1.
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1.
Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2.
General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.
Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2.
Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3
4
6
3.
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.
Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.
Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 14
3.2.1
Connect the Sample Inlet to the Quick-Lock Coupling . . . . . . . 14
3.2.2
Connect the Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3
Install the Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.
Connection Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.1
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.6.
Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6.1
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6.2
Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6.3
Relay Contacts 1 and 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7.
Signal Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.
Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1.
Establish sample flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.
Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.1.
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2.
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.3.
Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.4.
Changing Parameters and values . . . . . . . . . . . . . . . . . . . . . . . . 27
6.
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1.
Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.2.
Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 28
6.3.
Maintenance of the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . 29
6.3.1
Electrolyte exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
A-96.250.701 / 140716
1
AMI INSPECTOR Oxygen
6.3.2
Clean Swansensor Oxytrace and G Flow Cell . . . . . . . . . . . . .
6.4.
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.5
Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.6.
Quality Assurance of the Instrument. . . . . . . . . . . . . . . . . . . . . . .
6.6.1
Activate SWAN Quality assurance procedure. . . . . . . . . . . . . .
6.6.2
Pre-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.6.3
Connect the sample lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.6.4
Carry out comparison measurement . . . . . . . . . . . . . . . . . . . . .
6.6.5
Completion of the measurement . . . . . . . . . . . . . . . . . . . . . . . .
6.7.
Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.8.
Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.9.
Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.
Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.
Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.1.
Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2.
Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.3.
Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.4.
Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.5.
Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.
Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . .
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.
Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11.
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12.
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
50
50
50
52
53
54
45
45
46
47
48
48
36
36
36
38
31
32
34
34
39
40
41
41
42
69
72
73
2
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Safety Instructions
AMI INSPECTOR Oxygen -
Operator’s Manual
This document describes the main steps for instrument setup, operation and maintenance.
1.
Safety Instructions
General The instructions included in this section explain the potential risks associated with instrument operation and provide important safety practices designed to minimize these risks.
If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environment.
More safety instructions are given throughout this manual, at the respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target audience
Operator: Qualified person who uses the equipment for its intended purpose.
Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instrument.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as the Material Safety Data Sheets.
know the relevant safety rules and regulations.
A-96.250.701 / 140716
3
AMI INSPECTOR Oxygen
Safety Instructions
1.1.
Warning Notices
The symbols used for safety-related notices have the following significance:
DANGER
Your life or physical wellbeing are in serious danger if such warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incorrect process can be the consequence if such warnings are ignored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
4
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Safety Instructions
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
A-96.250.701 / 140716
5
AMI INSPECTOR Oxygen
Safety Instructions
1.2.
General Safety Regulations
Legal
Requirements
Spare Parts and
Disposables
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation of the instrument.
Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by an authorized Service Technician. SWAN will not accept responsibility for any claim resulting from unauthorized modification or alteration.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instrument power and power of devices connected to.
– relay 1,
– relay 2,
– alarm relay
WARNING
For safe instrument installation and operation you must read and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the tasks described in this document.
6
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Product Description
2.
Product Description
This chapter contains technical data, requirements and performance data.
2.1.
Description of the System
The portable AMI INSPECTOR instrument is a complete monitoring system mounted on a small panel with supporting stand and a rechargeable battery for stand-alone operation (>24h), designed as an inspection equipment for quality assurance of online process monitors.
Features General Features of AMI INSPECTORs are:
Battery life after full charge:
– >24h at full load (use of 3 relays, USB, signal output, logger)
– >36h at minimum load (use of logger only)
Charging time: approx. 6 hours
Controlled shut-down if battery is empty.
Display of remaining battery life in hours.
For longer battery life the back light of the LC Display is disabled.
Continuous operation using power adapter. The battery should be discharged at least once a month (normal usage until the monitor automatically shuts down).
Battery The Li-Ion battery is located in the housing of the AMI transmitter.
See chapter Power Supply, p. 19 regarding power supply and charging of the battery.
Safety features No data loss after power failure, all data is saved in non-volatile memory. Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
USB interface Built in USB interface for logger download. Use the USB stick supplied by Swan only (other USB sticks can dramatically reduce battery life).
Signal Output
(optional)
One signal output programmable for measured values (freely scaleable, linear or bilinear) or as continuous control output (control parameters programmable).
Current loop:
Maximal burden:
0/4–20 mA
510
A-96.250.701 / 140716
7
AMI INSPECTOR Oxygen
Product Description
Relay Two potential-free contacts programmable as limit switches for measuring values, controllers or timer for system cleaning with automatic hold function.
Maximum load: 100 mA/50 VAC
Alarm Relay One potential free contact.
Alternatively:
Open during normal operation, closed on error and loss of power.
Closed during normal operation, open on error and loss of power.
Summary alarm indication for programmable alarm values and instrument faults.
Input For potential-free contact to freeze the measuring value or to interrupt control in automated installations (hold function or remote-off)
Measuring principle
Clark principle:
The sensor consists of one noble metal electrode (e.g. platinum or gold), a reference electrode (mostly Ag/AgCl) and optionally a metal guard electrode.
The Clark-type electrode is the most widely used oxygen sensor for measuring oxygen dissolved in a liquid. The basic principle is that there is a cathode and an anode submersed in an electrolyte and a voltage is applied between the two parts. Oxygen enters the sensor through a permeable membrane by diffusion, and is reduced at the cathode according to
-
+ 2 H
-
O
2
+ 4e
This reaction creates a measurable current. There is a linear correlation between the oxygen concentration and the electrical current.
2
O --> 4 OH
The guard electrode is on the same voltage level as the cathode but there is no current measurement. Oxygen which diffuses from the electrolyte to the cathode is consumed by the guard electrode.
As a consequence, residual oxygen in the electrolyte will no more disturb the measurement signal and the response time to low oxygen levels will be shorter.
Temperature compensation
The measuring signal depends on temperature, but is automatically compensated to 25 °C. The sample temperature is determined continuously by a temperature sensor inside the oxygen electrode.
8
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Product Description
Fluidics Swansensor oxygen combined with QV-flow PMMA OTG flow cell.
The sample flows via sample inlet [E] through the flow regulating valve [D], where the flow rate can be adjusted. Then the sample flows into the measuring cell [C] were the Oxygen concentration and temperature of the sample is measured.
The sample leaves the measuring cell via flow sensor [B] through the sample outlet [F].
A-96.250.701 / 140716
A
B
C
Oxygen sensor
Flow sensor
Flow cell
A
B
C
D
E
F
D
E
F
Flow regulating valve
Sample inlet
Sample outlet
9
AMI INSPECTOR Oxygen
Product Description
2.2.
Instrument Overview
A
B
C
D
E
F
G
10
A
B
C
D
E
AMI Transmitter
Oxygen sensor
Flow cell
Flow sensor
Flow regulating valve
F
G
Sample inlet
Sample outlet
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Product Description
2.3.
Technical Data
Power Supply Battery
Use original, supplied power adapter only.
Voltage:
Power consumption:
Charging time:
Battery type:
85–265 VAC, 50/60 Hz max. 20 VA
6h
Li-Ion
During charging protect from heat impact and keep splash-proof
(not IP66).
Operating time Stand-alone (Battery):
Connected adapter:
> 24h continuous
Controlled shut-down when battery is empty, remaining time is displayed.
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X housing Ambient temperature:
Humidity:
Display:
-10 to +50 °C
10–90% rel., non condensing backlit LCD, 75 x 45 mm
Sample requirements
Flow cell and connection
Accuracy
Reproducibility
Flow rate:
Temperature:
Inlet pressure: pH:
Suspended solids:
Outlet pressure:
8 to 25 l/h up to 45 °C
0.2 to 1 bar not lower than pH 4 less than 10 ppm pressure free
Flow cell made of acrylic glass with built-in flow adjustment valve and digital sample flow meter
Sample inlet:
Sample outlet:
1/4” Swagelok tube adapter flexible tube 8x6 mm
±1.5 % of measured value or ±0.2 ppb
±1 % of measured value or ±0.15 ppb
A-96.250.701 / 140716
11
AMI INSPECTOR Oxygen
Product Description
Sensor
Oxytrace G
Technical data:
Measuring range:
Accuracy:
Precision:
Response time:
Minimal flow:
Operating temperature:
Material:
Protection:
Weight:
Sensor for the measurement of dissolved oxygen in ultra pure water. Precise oxygen measuring cell with integrated temperature sensor and guard electrode for faster initial response time after maintenance.
Clark oxygen electrode
Cathode gold, anode silver, guard silver
Zero current-free electrode system
Robust 25 µm fluoropolymer diaphragm
0–20 ppm O
2
(25 °C)
Automatic range switching
Range
0.1 to 9.99 ppb
10 to 199.9 ppb
200 to 1999 ppb
2 to 20 ppm
0 – 200% saturation
Resolution
0.01 ppb
0.1 ppb
1.0 ppb
0.01 ppm
0,3 % if calibration temperature = measuring temp. 1,5% at ± 10
°C deviation to cal. temperature
± 1% of reading or ±0.15 ppb t90 < 30 seconds (rising concentration)
50 cm/s Pressure resistance: 3 bar max. 50 °C polyacetal copolymer
IP 68
150 g
12
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Installation
3.
Installation
3.1.
Installation Check List
Check
Installation
Power-up
Instrument
Setup
Run-in period
Instrument’s specification must conform to your AC power ratings. See External power adapter, p. 20 .
Check if the battery is fully charged.
Connect the sample inlet and outlet, see Connecting Sample
Inlet and Outlet, p. 14 .
Install the Swansensor Oxytrace G into the flow cell, see Install the Swansensor Oxytrace G, p. 16 .
Open the flow regulating valve and adjust the sample flow to
8–25 l/h.
Switch on power.
Program all parameters, see chapter 4.
Let the instrument run continuously for 1 h.
A-96.250.701 / 140716
13
AMI INSPECTOR Oxygen
Installation
3.2.
Connecting Sample Inlet and Outlet
3.2.1
Connect the Sample Inlet to the Quick-Lock Coupling
The AMI INSPECTOR Oxygen is delivered with a quick-lock coupling. To connect the sample line to the AMI INSPECTOR Oxygen, simply push the nipple into the quick-lock coupling.
A
B
C
D
Flow cell block
Quick-lock coupling
Nipple
Sample outlet
A
B
C
D
14
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Installation
3.2.2
Connect the Sample Outlet
Installation 1 Loosen the union nut [B] but do not remove it.
2 Push the FEP tube [A] through the union nut [B] as far as it reaches the stop of the threaded tube [E].
3 Tighten the union nut 1¾ rotation using an open ended spanner.
Hold Body from turning with a second wrench.
4 Put the FEP Tube into a pressure free drain with sufficient capacity.
A B C D E
F
A
B
C
FEP tube 8x6
Union nut
Compression ferrule
D
E
F
Compression cone
Threaded tube
Tightened connection
A-96.250.701 / 140716
15
AMI INSPECTOR Oxygen
Installation
3.3
Install the Swansensor Oxytrace G
The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber [E]. A transport protection cap [B] filled with water [D] keeps the sensor wet during transport and storage. To install the sensor proceed as follows:
A
B
C
A
B
C
D
E
Fixing sleeve
Transport protection cap
Swansensor Oxytrace G
Water
Sensor cap filled with electrolyte
D
E
1 Unscrew the fixing sleeve [A].
2 Remove the transport protection cap [B].
3 Clean the Swansensor Oxytrace G [C] with water.
4 Install the Swansensor Oxytrace G into the flow cell
Connect the sensor cable to the transmitter, see Connection Diagram, p. 18 .
16
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Installation
3.4.
Electrical Connections
Cable thicknesses
WARNING
Always turn off DC power before manipulating electric parts.
Make sure the power specification of the instrument corresponds to the power on site.
In order to comply with IP66, use the following cable thicknesses
Wire
A
B
A B
PG 9 cable gland: cable Ø outer
PG 7 cable gland: cable Ø outer
4–8 mm
3–6.5 mm
NOTICE: Protect unused cable glands
For Relays: Use max. 1.5 mm
2 end sleeves.
/ AWG 14 stranded wire with
For Signal Outputs and Input: Use 0.25 mm
2 stranded wire with end sleeves.
/ AWG 23
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.
Make sure that the devices connected to the following contacts are disconnected from the power before continuing the installation.
– relay 1
– relay 2
– alarm relay
A-96.250.701 / 140716
17
AMI INSPECTOR Oxygen
Installation
3.5.
Connection Diagram
18
CAUTION
Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Installation
3.5.1
Power Supply
Contrary to all other Swan online process monitors the
AMI INSPECTOR transmitter is supplied with power by battery only. The rechargeable battery (Li-Ion) enables a stand-alone operation for at least 24 hours.
WARNING
Do not provide power directly to the transmitter as this will destroy the motherboard. All AMI INSPECTOR transmitters are supplied with power by battery only.
Charging Use the original supplied power adapter to charge AMI INSPEC-
TOR only. Charging time: approx 6h.
Fully charged a stand-alone operating time of at least 24h is guaranteed:
>24h at full load (use of 3 relays, USB, signal output, logger)
>36h at minimal load (use of logger only)
In case that the battery is discharged completely the firmware will automatically shut down.
Switch Power
ON - OFF
Continuous operation
Switch the instrument ON or OFF using the toggle switch on the transmitter.
For continuous operation use the power adapter as well.
CAUTION
If the AMI powers ON and then immediately shuts OFF, the battery is empty. Do not hold the toggle switch in ON position, as this can damage the battery.
CAUTION
During charging protect from heat impact and keep splashproof (plug of power adapter is not IP66).
Do not supply external devices, e.g. pumps, magnetic valves or any other current consumers with AMI INSPECTOR
CAUTION
Use the original supplied power adapter to charge AMI IN-
SPECTOR only. Use of any other power adapter can damage the battery or cause malfunction.
A-96.250.701 / 140716
19
AMI INSPECTOR Oxygen
Installation
External power adapter
Universal input range
85 - 265 VAC
Continuous short circuit protection
Over voltage protection
LED indicator for power on
2-pin AC inlet (IEC 320-C8) for detachable country-specific power cord
Power cords Two different power cords are supplied:
Power cord with type C plug (Europlug)
Power cord with type A plug (NEMA-1)
If a different plug type is needed, please purchase a suitable power cord from your local supplier.
NOTICE: Do not use any plug adapters.
Dimensions
Unit: mm
20
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Installation
3.6.
Relay Contacts
Programming of the relay contacts see 5.3 Relay Contacts, p. 59
3.6.1
Input
NOTICE: Use only potential-free (dry) contacts.
Terminals 13/14
If signal output is set to hold, measurement is interrupted if input is active.
For programming see menu 5.3.4, p. 65
3.6.2
Alarm Relay
NOTICE: For resistive loads only; do not use with capacitive or inductive loads. Max. load 1 A / 250 VAC.
Alarm output for system errors.
Error codes see Error List, p. 42
Programming see menu 5.3.1, p. 59
NOTICE: With certain alarms and certain settings of the AMI transmitter the alarm relay does not switch. The error, however, is shown on the display.
NC
1)
Normally
Closed
NO
Normally
Open
Terminals
5/4
5/3
Description
Active (opened) during normal operation.
Inactive (closed) on error and loss of power.
Active (closed) during normal operation.
Inactive (opened) on error and loss of power.
0V
Relay connection
1) 5
0V
4
3
5
4
3
1) usual use
A-96.250.701 / 140716
21
AMI INSPECTOR Oxygen
Installation
3.6.3
Relay Contacts 1 and 2
NOTICE: For resistive loads only; do not use with capacitive or inductive loads.Rated load 100 mA/ 50 VAC.
For programming see Menu Installation 5.3.2 and 5.3.3, p. 61
NO
Normally
Open
Terminals
6/7: Relay 1
8/9: Relay 2
Description
Inactive (opened) during normal operation and loss of power.
Active (closed) when a programmed function is executed.
Relay connection
0V
6/8
7/9
3.7.
Signal Output
The optional signal output 0/4–20 mA PCB can be plugged onto the USB interface PCB.
NOTICE: Max. burden 510
Terminals 16 (+) and 15 (-).
For programming see menu 5.2 Signal Outputs, p. 55 .
A B
22
A
B
Signal output 0/4–20 mA PCB
USB interface
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Instrument Setup
4.
Instrument Setup
4.1.
Establish sample flow
1 Open flow regulating valve, see Fluidics, p. 9 .
2 Wait until the flow cell has been completely filled.
3 Switch on power.
4 Adjust the sample flow to 8–25 l/h.
5 Let the instrument run-in for 1 h.
4.2.
Programming
Programming Program all parameters for external devices (interface, recorders, etc.)
Program all parameters for instrument operation (limits, alarms), see Program List and Explanations, p. 50
A-96.250.701 / 140716
23
AMI INSPECTOR Oxygen
Operation
5.
Operation
5.1.
Keys
A B C D
Exit Enter
A to exit a menu or command (rejecting any changes) to move back to the previous menu level
B to move DOWN in a menu list and to decrease digits
C to move UP in a menu list and to increase digits
D to open a selected sub-menu to accept an entry
Program
Access, Exit
RUN
R1
R2
9 l/h
14:10:45
1.05 ppb
25.4°C
Enter
Exit
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
1
24
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Operation
5.2.
Display
A B C D E
RUN
R1
R2
15 15:20:18
21.5
ppb
25 °C
F
G
A RUN
HOLD
OFF
B ERROR normal operation input closed or cal delay: Instrument on hold (shows status of signal outputs).
input closed: control/limit is interrupted (shows status of signal outputs).
Error Fatal Error
C Relay status
D Battery status (remaining operating time in h)
E Time
F Process values
G Sample Temperature
Relay status, symbols upper/lower limit not yet reached upper/lower limit reached control upw./downw. no action control upw./downw. active, dark bar indicates control intensity motor valve closed motor valve: open, dark bar indicates approx. position timer timer: timing active (hand rotating)
A-96.250.701 / 140716
25
AMI INSPECTOR Oxygen
Operation
5.3.
Software Structure
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
Messages
Pending Errors
Maintenance List
Message List
1
1.1
Menu Messages 1
Reveals pending errors as well as an event history
(time and state of events that have occurred at an earlier point of time).
It contains user relevant data.
Menu Diagnostics 2
Provides user relevant instrument and sample data.
Diagnostics
Identification
Sensors
Sample
I/O State
Interface
2.1
Maintenance 3.1
Calibration
Service
Simulation
Set Time 23.09.06 16:30:00
Operation
Sensors
Relay Contacts
Logger
4.1
Installation
Sensors
Signal Outputs
Relay Contacts
Miscellaneous
Interface
5.1
Menu
For instrument calibration, relay and signal output simulation, and to set the instrument time.
It is used by the service personnel.
Menu
Maintenance 3
Operation 4
User relevant parameters that might need to be modified during daily routine. Normally password protected and used by the process-operator.
Subset of menu 5 - Installation, but process-related.
Menu Installation 5
For initial instrument set up by SWAN authorized person, to set all instrument parameters. Can be protected by means of password.
26
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Operation
5.4.
Changing Parameters and values
Changing parameters
The following example shows how to change the logger interval:
1 Select the parameter you want to change.
2 Press [Enter]
Temperature
Standards
NT5K
Logger
Log interval Interval.
5 min
10 min
30 min
1 Hour
4.1.3
30min no
3 Press [ ] or [ ] key to highlight the required parameter.
4 Press [Enter] to confirm the selection or [Exit] to keep the previous parameter).
Logger
Log interval
Clear logger
4.1.3
10 min no
The selected parameter is highlighted (but not saved yet).
5 Press [Exit].
Logger
Log interval
Clear logger
Yes
No
4.1.3
no
Yes is highlighted.
6 Press [Enter] to save the new parameter.
The system reboots, the new parameter is set.
Changing values
A-96.250.701 / 140716
Alarm oxygen
Alarm High
Alarm Low
Hysteresis
Delay
5.3.1.1.1
10.00 ppm
1.00 ppb
0.10 ppb
5 Sec
Alarm oxygen
Alarm High
Alarm Low
Hysteresis
Delay
5.3.1.1.1
8.00 ppb
1.00 ppb
0.10 pp
5 Sec
1 Select the value you want to change.
2 Press [Enter].
3 Set required value with [
[ ] key.
] or
4 Press [Enter] to confirm the new value.
5 Press [Exit].
Yes is highlighted.
6 Press [Enter] to save the new value.
27
AMI INSPECTOR Oxygen
Maintenance
6.
Maintenance
Maintenance frequency depends strongly on the water quality. The
AMI INSPECTOR Oxygen is designed for determination of low level of dissolved oxygen in high purity water.
It is not suitable for the measurement of dissolved oxygen in waste water.
6.1.
Maintenance Table
Monthly
Half-yearly
Yearly
If necessary, perform an air calibration.
Clean Swansensor Oxytrace G membrane with a soft tissue.
If necessary, replace filling electrolyte.
If the sensor is exposed to air frequently during long time intervals, the electrolyte and membrane may have to be changed earlier (see below*).
Clean flow cell and flow meter, if dirty.
Replace Swansensor Oxytrace G membrane by using a new, prefilled sensor cap.
Every 2nd year
*A change of membrane and electrolyte is recommended:
if indicated in the maintenance list (remaining amount
<10%)
if the response of the sensor is slow
if the sensor can not be calibrated any more and/or the instrument shows a corresponding error message
if the sensor signal is very unstable.
6.2.
Stop of Operation for Maintenance
1 Shut off power of the instrument.
2 Stop sample flow by closing the flow regulating valve.
28
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
6.3.
Maintenance of the Oxygen Sensor
WARNING
Etching liquid
The electrolyte is alkaline and caustic. It contains less than 1% of potassium hydroxide.
Do not ingest. Wear protective goggles and gloves during handling. Avoid contact with clothes.
In case of accidental contact with the eyes, wash immediately with clear water and contact a physician. Show him the label of the bottle or this section of the manual.
Short contact with skin is harmless, nevertheless wash with lots of water.
6.3.1
Electrolyte exchange
An electrolyte exchange is indicated in the maintenance list as soon as the remaining amount is below 10%.
A
A
B
C
Fixing sleeve
Sensor
Flow cell
B
C
1 Unscrew the fixing sleeve [A].
2 Remove the sensor from the flow cell.
A-96.250.701 / 140716
29
AMI INSPECTOR Oxygen
Maintenance
A
G
H
E
F
B
C
D
E
F
G
H
A
B
C
D
Swansensor Oxytrace G
Thread
Groove
Guard electrode
Anode
Measuring head
Cathode
Sensor cap with membrane
30
3 Unscrew and remove the sensor cap [H] from the Swansensor
Oxytrace G [A].
4 Empty the remaining electrolyte.
5 Refill the sensor cap with fresh electrolyte.
NOTICE: There is a groove [C] in the thread [B] of the sensor, where excessive air and electrolyte can escape while screwing the sensor cap onto the sensor. Hold the sensor in vertical position, measuring head pointing downwards.
6 Slowly screw the sensor cap onto the sensor to allow the excess electrolyte to escape without building up too much pressure inside the electrode. Fasten sensor cap tightly.
7 Clean the sensor thoroughly and dry the sensing membrane with a soft tissue.
8 Switch on power.
9 Let the sensor run-in at air for at least 30 min, better 1 h.
10 Afterwards perform an air calibration.
11 Install the sensor into the flow cell.
12 Select “New Filling” to reset the counter for remaining electrolyte, see <Maintnence>/<Service> 3.2.1, p. 52 ).
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
6.3.2
Clean Swansensor Oxytrace and G Flow Cell
Depending on the water quality, the Swansensor Oxytrace G and the flow cell will necessitate a cleaning.
Before cleaning, stop operation as described in Stop of Operation for Maintenance, p. 28 .
1 Dismount the Swansensor Oxytrace G, see Electrolyte exchange, p. 29 .
2 Clean the sensor with a soft tissue and rinse it with water afterwards.
3 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.
4 Flush the flow cell with clean water.
5 Install the Swansensor Oxytrace G and start sample flow.
A-96.250.701 / 140716
31
AMI INSPECTOR Oxygen
Maintenance
6.4.
Calibration
The sensing part of the sensor must not be in direct contact with water!
In the wet flow cell, the atmosphere will be saturated with water vapor. This atmosphere will produce the most accurate calibration results.
The required time for a calibration depends mainly on the difference between temperature and oxygen content in the sample and in the air. It can take 15 – 20 minutes. This is also the case, if the electrolyte was exchanged.
As soon as the reading is stable, the microprocessor will store the calibration data in the memory. The end of the calibration is indicated on the display.
To perform a calibration proceed as follows:
1 Navigate to menu <Maintenance> /<Calibration>.
2 Press [Enter] to start the calibration and follow the dialog on the display.
Calibration 3.1.5
3 Stop the sample flow at with the flow regulating valve.
Close regulating valve to turn off sample flow.
<Enter> to continue
Calibration
Take sensor out of flow cell and dry membrane and sensor
3.1.5
<Enter> to continue
4 Unscrew and remove the threaded sleeve [A], see Electrolyte exchange, p. 29 .
5 Remove the oxygen sensor [B] from the flow cell [C].
6 Dry the sensor membrane and the flow cell with a soft paper tissue.
32
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
Calibration
Place the electrode into the wet flow cell at a slightly tilted angle.
3.1.5
<Enter> to continue
A
B
A-96.250.701 / 140716
Calibration
Saturation
Sat. Current
Progress
Calibration
Saturation
Sat. Current
3.1.1
98.7 %
32 A
3.1.1
98.7 %
32 A
A Tilted sensor
B Flow cell
7 Press [Enter] to start the calibration measurement.
The saturation should reach
100%, the saturation current should be about 24 A to 41 period, the calibration will be discarded.
8 Press [Enter] to confirm the calibration.
A.
If the measuring values are not stable during the measuring
Calibration Successful
33
AMI INSPECTOR Oxygen
Maintenance
6.5
Zero-Verification
Swansensor Oxytrace G for the measurement of low oxygen content (< 1 ppb).
1 Calibrate the sensor according to chapter Calibration, p. 32 .
2 Prepare a 5%-sodium sulfite solution with demineralized water.
3 Put the electrode into the sodium sulfite solution afterwards. Assure that there are no air bubbles in front of the sensor.
4 The measured value should now be < 1 ppb.
NOTICE: Depending on the state of the electrode this process can take several hours. In case of an electrode-refill, this can take days until the measured value is lower than 1 ppb.
6.6.
Quality Assurance of the Instrument
Quality assurance level
Every SWAN on-line instrument is equipped with integrated, autonomous quality assurance functions to survey the plausibility of each measurement.
For the AMI Oxytrace / AMI Oxytrace QED these are:
continuous monitoring of sample flow
continuous monitoring of the temperature inside the transmitter case
periodic accuracy test with ultra high precision resistors
Further a manual, menu driven inspection procedure can be carried out using a certified reference instrument. Running at the same sampling point as an inspection equipment, the AMI Inspector Oxygen verifies the measuring results. After enabling the quality assurance procedure by defining the quality assurance level, the instrument reminds the user periodically to run the procedure and results are stored in a history for review.
Central feature of the quality assurance function is the assignment of the monitored process to a Quality assurance level.
There are three predefined levels plus a user level. Hereby the inspection interval, the deviation limits of temperature and measuring result between the inspection equipment and the monitoring instrument are defined.
34
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
Level 1: Trend; Measurement used as an additional information to follow the process indicating trends.
Level 2: Standard; Monitoring of several parameters of a process (e.g. oxygen, saturation). In case of instrument failure, other parameters can be used for process monitoring.
Level 3: Crucial; Monitoring of critical processes, value is used for control of another part or subsystem (valve, dosing unit, etc.).
Additional level:
Quality level 4: User; User defined inspection interval, maximal deviation of temperature and measuring result.
Quality Level
0: Off
1: Trend
2: Standard
3: Crucial
4: User max. deviation temperature [°C] a)
Off
0.5 °C
0.4 °C
0.3 °C
0 - 2°C max. deviation result [%]
Off
10 %
5 %
5 %
0 - 20% min. inspection interval
Off annual quarterly monthly annual, quarterly, monthly a) sample temperature must be 25 °C +/-5 °C.
Procedure The standard workflow includes the following steps:
1 Activate SWAN Quality assurance procedure
2 Pre-test
3 Connect instruments
4 Carry out comparison measurement
5 Completion of the measurement
NOTICE: The procedure should only be carried out through qualified personnel.
Materials / Inspection equipment:
Reference instrument: AMI Inspector Oxygen
Two tubes made of PA
A-96.250.701 / 140716
35
AMI INSPECTOR Oxygen
Maintenance
6.6.1
Activate SWAN Quality assurance procedure
Enable quality assurance procedure on the process monitor(s) which shall be checked by selecting the quality level in menu
5.1.2.1. The corresponding submenus are then activated.
NOTICE: The activation is necessary the first time only.
6.6.2
Pre-test
Reference instrument: AMI Inspector Oxygen:
– Check certificate; reference instrument certificate not older than one year.
– Check battery; Battery of the AMI Inspector Oxygen should be completely charged. Remaining operating time on display minimum 20 hours.
– Make sure that all instruments are set to the same temperature compensation.
– Sensor is in working condition
On-line instrument: Monitor AMI Oxytrace:
– Good order and condition; Flow cell free of particles,
Sensor surface free of deposits.
– Check message list; Review the message list in menu 1.3 and check for frequently occurring alarms (as for example flow alarms). If alarms occur frequently remove cause before starting the procedure.
6.6.3
Connect the sample lines
See corresponding chapter in the manual of the process monitor which shall be checked with a reference instrument.
The choice of sampling depends strongly on local conditions on site. Possible sampling:
via sample point,
via T-fitting or
via piggyback/downstream
NOTICE:
• avoid ingress of air, use screwed fitting,
• sample as near as possible to the process monitor,
• wait approx. 10 minutes, whilst measurement is running, until measurement value and temperature are stabilized.
36
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
Example As an example following picture shows the connection of the reference instrument via T-fitting to the process monitor.
A
B
C
D
E
F
A-96.250.701 / 140716
G
A
B
C
D
Monitor AMI Oxytrace
AMI Inspector Oxygen
Reference flow cell
On-line flow cell
E
F
G
Sample outlet
Sample inlet
T-fitting
1 Stop sample flow to the monitor AMI Oxytrace by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell.
2 Connect sample line of the monitor AMI Oxytrace [A] with the sample inlet of the reference instrument AMI Inspector Oxygen
[B]. Use the supplied tube made of PA.
3 Connect sample outlet of the reference instrument AMI Inspector Oxygen to the sample outlet funnel of the monitor.
4 Switch on the AMI Inspector Oxygen. Open the flow regulating valve and regulate the sample flow to 10 l/h. The actual flow is shown on the transmitter.
37
AMI INSPECTOR Oxygen
Maintenance
6.6.4
Carry out comparison measurement
Start by selecting Quality Assurance in menu 3.5 of the monitor AMI
Oxytrace.
1 Navigate to menu Maintenance /Quality Assurance.
2 Press [Enter].
3 Follow the dialog on the Display.
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
3.5.5
4 Carry out pre test preparations
Connect instruments.
Regulate sample flow to 10 l/h using the appropriate valve.
<Enter> to continue
Quality Assurance
Value O2
Value Temp.
Wait 10 Minutes
3.5.5
0.05 ppb
25.00 C
5 Wait 10 minutes whilst measurement is running.
Press [Enter] to continue.
<Enter> to continue
Quality Assurance
Value O2
Value Temp.
Inspector O2
Inspector Temp.
<Enter> to continue
3.5.3
0.05 ppb
25.00 C
0.06 ppb
25.0 C
6 Read the ppb value of the reference instrument and enter under
[
“Inspector.” by using the [
] keys.
] or
7 Press [Enter] to confirm.
Quality Assurance
Value O2
Value Temp.
Inspector
Inspector Temp.
3.5.4
0.05 ppb
25.00 C
0.06 ppm
25.0 C
<Enter> to continue
Quality Assurance
Max. Dev. O2
Max. Dev. Temp.
Dev. O2
Dev. Temp.
QA-Check succesful
3.5.5
0.5 %
0.4 °C
0.1 %
0.4 °C
8 Read temperature value of the reference instrument and enter un-
[ der “Inspector Temp.” by using the
] or [ ] keys.
9 Press [Enter] to confirm.
10 Press [Enter] to continue.
The results are saved in QA-
History regardless if successful or not
38
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
6.6.5
Completion of the measurement
1 Stop the sample flow to the AMI Oxytrace by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell again.
2 Close flow regulating valve of the AMI Inspector.
3 Disconnect the AMI Inspector by removing the tubes and connect the sample outlet of the Monitor AMI Oxytrace to the sample outlet funnel again.
4 Start sample flow again and regulate sample flow.
5 Shutdown the AMI Inspector Oxygen.
For longer stop of operation see Longer Stop of Operation, p. 41
A-96.250.701 / 140716
39
AMI INSPECTOR Oxygen
Maintenance
6.7.
Replacing Fuses
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.
Make sure that the devices connected to the following contacts are disconnected from the power before continuing the installation.
– relay 1
– relay 2
– alarm relay
Find and repair the cause for the short circuit before replacing the fuse. Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A
40
A 1.25 AF/250V Instrument power supply
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
6.8.
Replacing the Battery
A B C A
B
C
Battery
Battery plug
Ribbon cable
1 Switch the AMI Inspector off.
2 If connected, disconnect the power adapter from the power jack.
3 Open the transmitter housing.
4 Pull out the ribbon cable [C] from the mainboard.
5 Disconnect battery plug [B] and replace the battery.
6.9.
Longer Stop of Operation
1 Shut off power of the instrument.
2 Stop sample flow.
3 Remove the Swansensor Oxytrace G.
4 Clean the sensor with a soft tissue and rinse it with water afterwards.
5 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.
6 Fill the flow cell with water.
7 Install the Swansensor Oxytrace G.
A-96.250.701 / 140716
41
AMI INSPECTOR Oxygen
Error List
42
7.
Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is exceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are recovered (i.e. Sample Flow low).
Such Errors are marked E0xx (bold and orange)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx ( bold and red )
HOLD
R1
R2
8 l/h
14:10:45
7.04 ppm
25.4°C
Error or fatal Error
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and take corrective action.
Press [ENTER].
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
1
Navigate to menu Messages.
Press [ENTER].
Messages
Pending Errors
Maintenance List
Message List
1.1
Navigate to menu Pending Errors.
Press [ENTER].
Pending Errors
Error Code
Alarm low
<Enter> to Acknowledge
1.1.5
E002
Press [ENTER] to acknowledge the
Pending Errors. The Error is reset and saved in the Message List.
* Menu numbers see
Program Overview, p. 45
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Error List
Error
E001
E002
Description oxygen Alarm high oxygen Alarm low
E003 Saturation Alarm high
E004
E007
E008
E009
E010
E011
E012
Saturation Alarm low
Sample Temp. high
Sample Temp. low
Sample Flow high
Sample Flow low
Temp. shorted
Temp. disconnected
Corrective action
– check process
– check programmed value, see 5.3.1.1.1, p. 59
– check process
– check programmed value, see 5.3.1.1.25, p. 59
– check process
– check programmed value, see 5.3.1.4, p. 60
– check process
– check programmed value, see 5.3.1.4, p. 60
– check process
– check programmed value, see 5.3.1.3.1, p. 60
– check process
– check programmed value, see 5.3.1.3.22, p. 60
– check sample flow
– check programmed value, see 5.3.1.2.2, p. 60
– establish sample flow
– clean instrument
– check programmed value, see 5.3.1.2.32, p. 60
– Check wiring of sensor
– Check sensor
– Check wiring of sensor
– Check sensor
A-96.250.701 / 140716
43
AMI INSPECTOR Oxygen
Error List
Error
E013
Description
Case Temp. high
E014 Case Temp. low
E017 Control Timeout
E019 Quality Assurance
E024 Input active
E026
E030
E031
E032
E033
E034
E065
IC LM75
EEProm Frontend
Calibration Recout
Wrong Frontend
Power-on
Power-down
Electrolyte depleted
Corrective action
– check case/environment temperature
– check programmed value, see 5.3.1.5.1, p. 61
– check case/environment temperature
– check programmed value, see 5.3.1.5.2, p. 61
– check control device or programming in
Installation, Relay contact, Relay 1/2 see 5.3.2 and 5.3.3, p. 61
– Perform QA Procedure using a reference instrument, e.g. AMI Inspector
– See If Fault Yes is programmed in Menu see 5.3.4, p. 65
– call service
– call service
– call service
– call service
– none, normal status
– none, normal status
– Refill electrolyte, see Electrolyte exchange, p. 29
44
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program Overview
8.
Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, p. 50 .
Menu 1 Messages is always accessible for everybody. No password protection. No settings can be modified.
Menu 2 Diagnostics is always accessible for everybody. No password protection. No settings can be modified.
Menu 3 Maintenance is for service: Calibration, simulation of outputs and set time/date. Please protect with password.
Menu 4 Operation is for the user, allowing to set limits, alarm values, etc. The presetting is done in the menu Installation
(only for the System engineer). Please protect with password.
Menu 5 Installation : Defining assignment of all inputs and outputs, measuring parameters, interface, passwords, etc.
Menu for the system engineer. Password strongly recommended.
8.1.
Messages (Main Menu 1)
Pending Errors 1.1.5* Pending Errors
1.1*
Maintenance List
1.2*
Message List
1.3*
Maintenance List
Number
Date, Time
1.2.5*
1.3.1*
* Menu numbers
A-96.250.701 / 140716
45
AMI INSPECTOR Oxygen
Program Overview
Identification
2.1*
8.2.
Diagnostics (Main Menu 2)
Desig.
Version
Factory Test
2.1.3*
AMI Oxytrace
6.00-11/15
Instrument
Motherboard
2.1.3.1*
Front End
Years / Days / Hours / Minutes / Seconds
Sensors
2.2*
Operating Time
2.1.4*
Sensor
2.2.1*
Current Value
(Raw value tc)
(Raw value)
Saturation
Cal. History
2.2.1.5*
Number
Date, Time
Sat. Current
Air pressure
2.2.2.1*
Sample
2.3*
I/O State
2.4*
Interface
2.5*
Miscellaneous
2.2.2*
QA History
2.2.3*
Sample ID
Temperature °C
Nt5K Ohm
Alarm Relay
Relay 1/2
Input
Signal Output 3
Protocol
USB Stick
Case Temp.
Air pressure
QA History
2.3.1*
2.4.1*
2.4.2*
2.5.1*
2.2.3.1*
* Menu numbers
2.1.4.1*
2.2.1.5.1*
46
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program Overview
Calibration
3.1*
Service
3.2*
Simulation
3.3*
8.3.
Maintenance (Main Menu 3)
Calibration 3.1.5
Electrolyte
3.2.1*
Alarm Relay
Relay 1
Relay 2
Signal Output 3
(Date), (Time)
Last filling
Remaining amount
Remaining time
New Filling
3.3.1*
3.3.2*
3.3.3*
3.3.6*
3.2.1.5*
Set Time
3.4*
Quality Assurance
3.5*
Quality Assurance 3.5.5*
* Menu numbers
A-96.250.701 / 140716
47
AMI INSPECTOR Oxygen
Program Overview
8.4.
Operation (Main Menu 4)
Sensors
4.1*
Relay Contacts
4.2*
Filter Time Const.
Hold after Cal.
Alarm Relay
4.2.1*
4.1.1*
4.1.2*
Alarm Oxygen
4.2.1.1*
Logger
4.3*
Relay 1/2
4.2.2* - 4.2.3*
Input
4.2.4*
Log Interval
Clear Logger
Eject USB Stick
Alarm Saturation
4.2.1.2*
Setpoint
Hysteresis
Delay
Active
Signal Outputs
Output / Control
Fault
Delay
4.3.1*
4.3.2*
4.3.3*
Alarm High
Alarm Low
Hysteresis
Delay
Alarm High
Alarm Low
Hysteresis
Delay
4.2.x.100*
4.2.x.200*
4.2.x.30*
4.2.4.1*
4.2.4.2*
4.2.4.3*
4.2.4.4*
4.2.4.5*
4.2.1.1.1*
4.2.1.1.22*
4.2.1.1.32*
4.2.1.1.42*
4.2.1.2.1*
4.2.1.2.22*
4.2.1.2.32*
4.2.1.2.42*
* Menu numbers
8.5.
Installation (Main Menu 5)
Sensors
5.1*
Miscellaneous
5.1.1*
Flow
Offset
Quality Assurance Level
5.1.2*
Signal Outputs Signal Output 1/2
5.2* 5.2.1* - 5.2.2*
Parameter
Current Loop
Function
Scaling
Relay Contacts
5.3*
Alarm Relay
5.3.1*
5.2.x.40
Alarm oxygen
5.3.1.1*
5.1.1.1*
5.1.1.2*
5.1.2.1*
5.2.1.1 - 5.2.2.1*
5.2.1.2 - 5.2.2.2*
5.2.1.3 - 5.2.2.3*
Range Low
Range High
Alarm High
Alarm Low
48
* Menu numbers
5.2.x.40.10/11*
5.2.x.40.20/21*
5.3.1.1.1*
5.3.1.1.22
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program Overview
Miscellaneous
5.4*
Language
Set defaults
Load Firmware
Password
5.4.4*
Interface
5.5*
Relay 1/2
5.3.2* - 5.3.3*
Input
5.3.4*
Sample ID
Protocol
5.5.1*
Case Temp.
5.3.1.5*
Function
Parameter
Setpoint
Hysteresis
Delay
Active
Signal Outputs
Output/Control
Fault
Delay
5.4.1*
5.4.2*
5.4.3*
Messages
Maintenance
Operation
Installation
5.4.5*
USB Stick
Sample Flow
5.3.1.2*
Sample Temp.
5.3.1.3*
Alarm Saturation
5.3.1.4*
Hysteresis
Delay
Flow Alarm
Alarm High
Alarm Low
Alarm High
Alarm Low
Alarm High
Alarm Low
Hysteresis
Delay
Case Temp. high
Case Temp. low
5.3.2.1–5.3.3.1*
5.3.2.20–5.3.3.20*
5.3.2.300–5.3.3.301*
5.3.2.400–5.3.3.401*
5.3.2.50–5.3.3.50*
5.3.4.1*
5.3.4.2*
5.3.4.3*
5.3.4.4*
5.3.4.5*
5.3.1.1.32
5.3.1.1.42
5.3.1.2.1
5.3.1.2.2*
5.3.1.2.32*
5.3.1.3.1*
5.3.1.3.22*
5.3.1.4.1*
5.3.1.4.22
5.3.1.4.32
5.3.1.4.42
5.3.1.5.1*
5.3.1.5.2*
5.4.4.1*
5.4.4.2*
5.4.4.3*
5.4.4.4*
* Menu numbers
A-96.250.701 / 140716
49
AMI INSPECTOR Oxygen
Program List and Explanations
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active, acknowledged). If an active error is acknowledged, the alarm relay is active again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5
Provides the list of necessary maintenance. Cleared maintenance messages are moved to the Message list.
1.3 Message List
1.3.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Desig.
: Designation of the instrument.
Version : Firmware of instrument (e.g. V6.00-11/15)
2.1.3
Factory Test : Test date of the Instrument, Motherboard and
Frontend.
2.1.4
Operating Time: Shows the operating time in Years, Days, Hours,
Minutes and Seconds.
2.2 Sensors
2.2.1
Oxytrace G
Current value: Shows the actual measuring value in ppb.
Raw value tc: Shows the actual temperature compensated measuring value in mA.
Saturation Shows the actual saturation in %
50
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
2.2.1.4
Cal. History
Review the diagnostic values of the last calibration of the oxygen sensor. Max. 64 data records are memorized.
Number : Calibration counter.
Date, Time : Date and time of the calibration.
Sat. Current : Saturation current at that time of calibration.
Air pressure : Air pressure at that time of calibration.
2.2.2
Miscellaneous :
2.2.2.1
Case Temp: Shows the actual temperature in °C inside the transmitter.
Air pressure : Shows the actual air pressure in hPa
2.2.3
QA History
Review QA values (Number, Date, Time, Deviation oxygen,
Deviation Temperature, Status of QA check) of the last quality assurance procedures.
2.3 Sample
2.3.301
Sample ID : Shows the assigned sample identification. This identification is defined by the user to identify the location of the sample
Temperature: Shows temperature in °C.
(Nt5K) : Shows raw value of the temperature in
Sample Flow : Shows the sample flow in l/h
(Raw value) Shows the sample flow in Hz
2.4 I/O State
Shows actual status of all in- and outputs.
2.4.1
Alarm Relay:
Relay 1 and 2:
Input:
Signal Output 3:
Active or inactive
Active or inactive
Open or closed.
Actual current in mA (if option is installed)
2.5 Interface
2.5.1
Protocol USB Stick.
A-96.250.701 / 140716
51
AMI INSPECTOR Oxygen
Program List and Explanations
3 Maintenance
3.1 Calibration
3.1.1
Start a calibration and follow the instructions on the screen.
Displayed values are saturation in % and the saturation current in mA. The indication bar shows the progress. Detailed explanation see Calibration, p. 32 .
3.2 Service
3.2.1
Electrolyte
Last Filling : Shows the date of the last filling of electrolyte.
Remaining Amount : Remaining amount of electrolyte in %.
Remaining Time : Remaining time in days until electrolyte exchange recommended.
3.2.1.5
New Filling : Select “Yes” after electrolyte exchange to reset the counter.
3.3 Simulation
To simulate a value or a relay state, select the
alarm relay
relay 1 and 2
signal output 3 (signal outputs 1 and 2 are deactivated) with the [ ] or [ ] keys.
Press the [Enter]> key.
Change the value with the [ ] or [ ] keys.
Press the [Enter] key.
The value is simulated by the relay/signal output.
Alarm Relay:
Relay 1:
Relay 2:
Signal Output 3:
Active or inactive
Active or inactive
Active or inactive
Current in mA (if option is installed)
At the absence of any key activities, the instrument will switch back to normal mode after 20 min. If you quit the menu, all simulated values will be reset.
3.4 Set Time
Adjust date and time.
52
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
3.5 Quality Assurance
Performs a Quality Assurance according to your settings. Follow the commands on the screen. Detailed explanation see Quality
Assurance of the Instrument, p. 34 .
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant: Used to damp noisy signals. The higher the filter time constant, the slower the system reacts to changes of the measured value.
Range: 5–300 Sec
4.1.2
Hold after Cal: Delay permitting the instrument to stabilize again after calibration. During calibration plus hold-time, the signal outputs are frozen (held on last valid value), alarm values, limits are not active.
Range: 0–6‘000 Sec
4.2 Relay Contacts
See Relay Contacts, p. 21
4.3 Logger
The instrument is equipped with an internal logger. The data can be copied to the USB stick installed in the transmitter.
The logger can save approx. 1500 data records. The Records consists of: Date, time, alarms, measuring values, raw values, case temperature, flow.
4.3.1
Log Interval: Select a convenient log interval. Consult the table below to estimate the max logging time. When the logging buffer is full, the oldest data record is erased to make room for the newest one (circular buffer).
Range: 1 Second to 1 hour
Interval
Time
1 s 5 s
25 min 2 h
1 min
25 h
5 min
5 d
10 min
10 d
30 min
31 d
1 h
62 d
4.3.2
Clear Logger: If confirmed with yes , the complete logger data is deleted. A new data series is started.
4.3.3
Eject USB Stick
With this function all logger data are copied to the USB stick before the USB stick is deactivated.
A-96.250.701 / 140716
53
AMI INSPECTOR Oxygen
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1
Miscellaneous
5.1.1.1
Flow : If a flow cell without flow measurement (e.g. B-Flow) is used, choose none. With flow measurement select Q-Flow
5.1.1.2
O2 Offset : Manual, small correction of the offset.
Range -5 to +5 ppb.
5.1.2
Quality Assurance: Switch the Quality Assurance on or off.
5.1.2.1
Level : Select quality level:
Level 0: Off
Quality assurance procedure switched off. Any additional QA menus are hidden.
Level 1: Trend
Level 2: Standard
Level 3: Crucial
Level 4: User
Edit user specific limits in menu 5.1.2.2
54
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.2 Signal Outputs
5.2.1
Signal Output 3 (signal outputs 1 and 2 are deactivated)
5.2.1.1
Parameter: Assign one of the process values to the signal output.
Available values:
Oxygen
Temperature
Sample Flow (if a flow sensor is selected)
Saturation
5.2.1.2
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3
Function: Define if the signal output is used to transmit a process value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 55
Control upwards or control downwards for controllers.
See As control output, p. 57
As process values
The process value can be represented in 3 ways: linear, bilinear or logarithmic. See graphs below.
[mA]
20
B A
10 12
A
B
0 / 4
0.0
linear
0.1
bilinear
0.2
0.3
0.4
X
0.5
X Measured value
A-96.250.701 / 140716
55
AMI INSPECTOR Oxygen
Program List and Explanations
[mA]
20
10 12
0 / 4
1
0
2 4 6
10
1
100
2
1’000 10’000
3 4
X
X Measured value (logarithmic)
5.2.1.40
Scaling: Enter beginning and end point (Range low & high) of the linear or logarithmic scale. In addition, the midpoint for the bilinear scale.
Parameter: Oxygen.
Range low: 0.00 ppb –20.00 ppm
Range high: 0.00 ppb –20.00 ppm
Parameter: Temperature
Range low: -30 to + 130 °C
Range high: 30 to + 130 °C
Parameter: Sample flow
Range low: 0–50 l/h
Range high: 0–50 l/h
Parameter: Saturation
Range low: 0–200 %
Range high: 0–200 %
56
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
As control output
Signal outputs can be used for driving control units. We distinguish different kinds of controls:
P-controller: The controller action is proportional to the deviation from the setpoint. The controller is characterized by the
P-Band. In the steady-state, the setpoint will never be reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-controller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
Ziegler-Nichols method for the optimization of a PID controller:
Parameters : Setpoint, P-Band, Reset time, Derivative time
Y
B
A
A-96.250.701 / 140716
X a
A
B
X
L
Response to maximum control output
Tangent on the inflection point
Time
Xp
Tn
Tv
= 1.2/a
= 2L
= L/2
The point of intersection of the tangent with the respective axis will result in the parameters a and L.
Consult the manual of the control unit for connecting and programming details. Choose control upwards or downwards.
57
AMI INSPECTOR Oxygen
Program List and Explanations
58
Setpoint : User-defined process value for the selected parameter.
P-Band: Range below (upwards control) or above (downwards control) the set-point, within which the dosing intensity is reduced from 100% to 0% to reach the set-point without overshooting.
5.2.1.43
Control Parameters : if Parameter = Oxygen
5.2.1.43.10
Setpoint:
Range: 0.00 ppb –20.00 ppm
5.2.1.43.20
P-Band:
Range: 0.00 ppb –20.00 ppm
5.2.1.43
Control Parameters : if Parameter = Temperature
5.2.1.43.11
Setpoint:
Range: -30 to + 130 °C
5.2.1.43.21
P-Band:
Range: 0 to + 100 °C
5.2.1.43
Control Parameters : if Parameter = Sample flow
5.2.1.43.12
Setpoint:
Range: 0–50 l/h
5.2.1.43.22
P-Band:
Range: 0–50 l/h
5.2.1.43
Control Parameters : if Parameter = Saturation
5.2.1.43.13
Setpoint:
Range: 0–200%
5.2.1.43.23
P-Band:
Range: 0–200%
5.2.1.43.3
Reset time: The reset time is the time till the step response of a single I-controller will reach the same value as it will be suddenly reached by a P-controller.
Range: 0–9’000 sec
5.2.1.43.4
Derivative time: The derivative time is the time till the ramp response of a single P-controller will reach the same value as it will be suddenly reached by a D-controller.
Range: 0–9’000 sec
5.2.1.43.5
Control timeout: If a controller action (dosing intensity) is constantly over 90% during a defined period of time and the process value does not come closer to the setpoint, the dosing process will be stopped for safety reasons.
Range: 0–720 min
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.3 Relay Contacts
5.3.1
Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
Power loss
Detection of system faults like defective sensors or electronic parts
High case temperature
Process values out of programmed ranges.
Program alarm levels for the following parameters:
Meas. Value
Temperature
Sample Flow (if a flow sensor is selected)
Case Temperature high
Case Temperature low
5.3.1.1
Alarm Oxygen
5.3.1.1.1
Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E001, is displayed in the message list.
Range: 0.00 ppb –20.00 ppm
5.3.1.1.25
Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E002 is displayed in the message list.
Range: 0.00 ppb –20.00 ppm
5.3.1.1.35
Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.
Range. 0.00 ppb –20.00 ppm
5.3.1.1.45
Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.
Range: 0–28‘800 Sec
A-96.250.701 / 140716
59
AMI INSPECTOR Oxygen
Program List and Explanations
60
5.3.1.2
Sample Flow: Define at which sample flow a flow alarm should be issued.
5.3.1.2.1
Flow Alarm: Program if the alarm relay should be activated if there is a flow alarm. Choose between yes or no. The flow alarm will always be indicated in the display, pending error list, saved in the message list and the logger.
Available values: Yes or no
NOTICE: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.2.2
Alarm High: If the measuring values rises above the programmed value E009 will be issued.
Range: 12–50 l/h
5.3.1.2.32
Alarm Low: If the measuring values falls below the programmed value E010 will be issued.
Range: 8–11 l/h
5.3.1.3
Sample Temp.: Define at which sample temperature an alarm should be issued.
5.3.1.3.1
Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E007 is issued.
Range: 30–100 °C
5.3.1.3.22
Alarm Low: If the measured value rises above the alarm high value, the alarm relay is activated and E008 is issued.
Range: -10 to + 20 °C
5.3.1.4
Alarm Saturation
5.3.1.4.1
Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E001, is displayed in the message list.
Range: 0.00 –200 %
5.3.1.4.22
Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E002 is displayed in the message list.
Range: 0.00 –200 %
5.3.1.4.32
Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.
Range. 0.00 –200 %
5.3.1.4.42
Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.
Range: 0–28‘800 Sec
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.1.5
Case Temp.
5.3.1.5.1
Case Temp. high: Set the alarm high value for temperature of electronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.5.2
Case Temp. low: Set the alarm low value for temperature of electronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to + 20 °C
5.3.2 and 5.3.3
Relay 1 and 2: The function of relay contacts 1 or 2 are defined by the user
NOTICE: The navigation in the menu <Relay 1> and <Relay 2> is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected function. The same values may also be entered in menu 4.2 Relay
Contacts, p. 53
5.3.2.1
Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program the following:
5.3.2.20
Parameter: choose one of the following process values
Oxygen
Temperature
Sample Flow
Saturation
5.3.2.300
Setpoint : If the measured value rises above respectively falls below the set-point, the relay is activated.
Parameter
Oxygen
Temperature
Sample flow
Saturation
Range
0.00 ppb –20.00 ppm
-30 to + 130 °C
0–50 l/h
0–200 %
A-96.250.701 / 140716
61
AMI INSPECTOR Oxygen
Program List and Explanations
62
5.3.2.400
Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value fluctuates around the alarm value.
Parameter
Oxygen
Temperature
Sample flow
Saturation
Range
0.00 ppb –20.00 ppm
0–100 °C
0–50 l/h
0–200 %
5.3.2.50
Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.
Range. 0–600 Sec
5.3.2.1
Function = Control upwards/downwards
The relays may be used to drive control units such as solenoid valves, membrane dosing pumps or motor valves. When driving a motor valve both relays are needed, relay 1 to open and relay 2 to close the valve.
5.3.2.22
Parameter: choose one of the following process values
Oxygen
Temperature
Sample Flow
Saturation
5.3.2.32
Settings
Choose the respective actuator:
Time proportional
Frequency
Motor valve
Actuator = Time proportional
Examples of metering devices that are driven time proportional are solenoid valves, peristaltic pumps.
Dosing is controlled by the operating time.
5.3.2.32.20
Cycle time: duration of one control cycle (on/off change).
Range: 0–600 sec.
5.3.2.32.30
Response time: Minimal time the metering device needs to react.
Range: 0–240 sec.
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.2.32.4
Control Parameters:
Range for each Parameter same as 5.2.1.43, p. 58
Actuator = Frequency
Examples of metering devices that are pulse frequency driven are the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
5.3.2.32.21
Pulse frequency: Max. pulses per minute the device is able to respond to. Range: 20–300/min.
5.3.2.32.31
Control Parameters:
Range for each Parameter same as 5.2.1.43, p. 58
Actuator = Motor valve
Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22
Run time: Time needed to open a completely closed valve
Range: 5–300 Sec.
5.3.2.32.32
Neutral zone: Minimal response time in % of the runtime. If the requested dosing output is smaller than the response time, no change will take place.
Range: 1–20 %
5.3.2.32.4
Control Parameters:
Range for each Parameter same as 5.2.1.43, p. 58
5.3.2.1
Function = Timer
The relay will be activated repetitively depending on the programmed time scheme.
5.3.2.24
Mode: Operating mode (interval, daily, weekly)
5.3.2.24
Interval
5.3.2.340
Interval: The interval can be programmed within a range of 1–1440 min.
5.3.2.44
Run Time : Enter the time the relay stays active.
Range: 5–32’400 Sec.
5.3.2.54
Delay : during run time plus the delay time the signal and control outputs are held in the operating mode programmed below.
Range: 0–6’000 Sec.
A-96.250.701 / 140716
63
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.2.6
Signal Outputs : Select operating mode of the signal output:
Cont.: Signal outputs continue to issue the measured value.
Hold: Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors, are not issued.
Off: Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
5.3.2.7
Output/Control : Select operating mode of the controller output:
Cont.: Controller continues normally.
Hold: Controller continues based on the last valid value.
Off: Controller is switched off.
5.3.2.24
daily
The relay contact can be activated daily, at any time of a day.
5.3.2.341
Start time : to set the start time proceed as follows:
1 Press [Enter], to set the hours.
2 Set the hour with the [ ] or [
3 Press [Enter], to set the minutes.
] keys.
4 Set the minutes with the [ ] or [
5 Press [Enter], to set the seconds.
6 Set the seconds with the [ ] or [
] keys.
] keys.
Range: 00:00:00–23:59:59
5.3.2.44
Run Time : see Interval
5.3.2.54
Delay : see Interval
5.3.2.6
Signal Outputs : see Interval
5.3.2.7
Output/Control : see Interval
5.3.2.24
weekly
The relay contact can be activated at one or several days, of a week. The daily starting time is valid for all days.
64
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.2.342
Calendar :
5.3.2.342.1
Start time : The programmed start time is valid for each of the programmed days. To set the start time see 5.3.2.341, p. 64 .
Range: 00:00:00–23:59:59
5.3.2.342.2
Monday : Possible settings, on or off to
5.3.2.342.8
Sunday : Possible settings, on or off
5.3.2.44
Run Time : see Interval
5.3.2.54
Delay : see Interval
5.3.2.6
Signal Outputs : see Interval
5.3.2.7
Output/Control : see Interval
5.3.2.1
Function = Fieldbus
The relay will be switched via the Profibus input. No further parameters are needed.
5.3.4
Input: The functions of the relays and signal outputs can be defined depending on the position of the input contact, i.e. no function, closed or open.
5.3.4.1
Active: Define when the input should be active:
No: Input is never active.
When closed Input is active if the input relay is closed
When open: Input is active if the input relay is open
5.3.4.2
Signal Outputs: Select the operation mode of the signal outputs when the relay is active:
Cont.:
Hold:
Off:
Signal outputs continue to issue the measured value.
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal errors, are not issued.
Set to 0 or 4 mA respectively. Errors, except fatal errors, are not issued.
A-96.250.701 / 140716
65
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.4.3
Output/Control: (relay or signal output):
Cont.:
Hold:
Off:
Controller continues normally.
Controller continues on the last valid value.
Controller is switched off.
5.3.4.4
Fault:
No:
Yes:
No message is issued in pending error list and the alarm relay does not close when input is active.
Message E024 is issued and stored in the message list. The Alarm relay closes when input is active.
5.3.4.5
Delay: Time which the instrument waits, after the input is deactivated, before returning to normal operation.
Range: 0–6‘000 Sec
66
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.4 Miscellaneous
5.4.1
Language: Set the desired language.
Available settings: German /English/French/Spanish
5.4.2
Set defaults: Reset the instrument to factory default values in three different ways:
Calibration: Sets calibration values back to default. All other values are kept in memory.
In parts: Communication parameters are kept in memory. All other values are set back to default values.
Completely: Sets back all values including communication parameters.
5.4.3
Load Firmware: Firmware updates should be done by instructed service personnel only.
5.4.4
Password: Select a password different from 0000 to prevent unauthorized access to the menus “Messages”, “Maintenance”,
“Operation” and “Installation”.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representative.
5.4.5
Sample ID: Identify the process value with any meaningful text, such as KKS number.
A-96.250.701 / 140716
67
AMI INSPECTOR Oxygen
Program List and Explanations
5.5 Interface
Select one of the following communication protocols. Depending on your selection, different parameters must be defined.
5.5.1
5.5.20
5.5.30
5.5.40
5.5.1
5.5.21
5.5.31
5.5.41
5.5.1
Protocol: Profibus
Device address:
ID-Nr.:
Local operation:
Range: 0–126
Range: Analyzer; Manufacturer; Multivariable
Range: Enabled, Disabled
Protocol: Modbus RTU
Device address: Range: 0–126
Baud Rate: Range: 1 200–115 200 Baud
Parity: Range: none, even, odd
Protocol: USB-Stick:
Only visible if an USB interface is installed. No further settings are possible.
68
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Default Values
10. Default Values
Operation:
Sensors: Filter Time Const.: .................................................................10 Sec
Hold after Cal.:.....................................................................300 Sec
Alarm Relay ...................................................................... same as in Installation
Relay 1/2 ...................................................................... same as in Installation
Input ...................................................................... same as in Installation
Logger: Logger Interval:.............................................................. 30 Minutes
Clear Logger:................................................................................ no
Installation:
Sensors Miscellaneous; Flow: ............................................................... None
Offset: ................................................................................... 0.0 ppb
Quality Assurance; Level: ........................................................ 0: Off
Signal Output Parameter:............................................................................ oxygen
Current loop:..................................................................... 4 –20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................ 0.00 ppb
Scaling: Range high: ....................................................... 10.00 ppm
Alarm Relay: Alarm oxygen; Alarm high:............................................... 10.00 ppm
Alarm oxygen; Alarm low: ................................................... 0.00 ppb
Alarm oxygen; Hysteresis:................................................... 100 ppb
Alarm oxygen; Delay: ..............................................................5 Sec
If Flow = Q-Flow
Sample Flow, Flow Alarm: .......................................................... yes
Sample Flow, Alarm high: .....................................................25.0 l/h
Sample Flow, Alarm low: ........................................................8.0 l/h
Sample Temp., Alarm High: ..................................................... 50 °C
Sample Temp., Alarm Low:........................................................ 0 °C
Alarm Saturation; Alarm high..................................................120 %
Alarm Saturation; Alarm low ....................................................0.0 %
Alarm Saturation; Hysteresis ......................................................2 %
Alarm Saturation; Delay...........................................................5 Sec
Case temp. high: ..................................................................... 65 °C
Case temp. low:......................................................................... 0 °C
Relay 1 Function:.......................................................................... limit upper
Parameter:............................................................................ oxygen
Setpoint: .......................................................................... 10.00 ppm
Hysteresis:........................................................................... 100 ppb
Delay: ....................................................................................30 Sec
A-96.250.701 / 140716
69
AMI INSPECTOR Oxygen
Default Values
Relay 2 Function: .......................................................................... limit upper
Parameter: ....................................................................Temperature
Setpoint:...................................................................................50 °C
Hysteresis: ..............................................................................1.0 °C
Delay:.....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter: .................................................................... Meas. Value
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .......................10.00 ppm
Settings: Control Parameters: P-band:.............................100 ppb
Settings: Control Parameters: Reset time: ...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:.................. 0 Min.
Settings: Act. Time prop.: Cycle time: .................................... 60 s
Settings: Act. Time prop.: Response time: ............................ 10 s
Settings: Act. Motor valve: Run time: .................................... 60 s
Settings: Act. Motor valve: Neutral zone: ................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval:................................................................................. 1 min
Mode: ....................................................................................... daily
Start time: ........................................................................00.00.00
Mode:.....................................................................................weekly
Calendar; Start time: .......................................................00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay:.......................................................................................5 Sec
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont
Input: Active ............................................................................when closed
Signal Outputs ........................................................................... hold
Output/Control .............................................................................. off
Fault..............................................................................................no
Delay......................................................................................10 Sec
70
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Default Values
Miscellaneous Language:............................................................................. English
Set default: ................................................................................... no
Load firmware:.............................................................................. no
Password:........................................................... for all modes 0000
Sample ID:....................................................................... - - - - - - - -
Interface Protocol: ...........................................................................USB Stick
A-96.250.701 / 140716
71
AMI INSPECTOR Oxygen
Index
11.
Index
A
Alarm Relay
. . . . . . . . . .
21
C
Cable thicknesses
Calendar
. . . . . .
17
. . . . . . . . . . . .
65
Changing values
Charging
. . . . . . .
27
. . . . . . . . . . . .
19
Clean Flow Cell and Swansensor Oxytrace G
. . . . . . . . . . . . .
31
D
Default Values
. . . . . . . .
69
E
Electrolyte exchange external devices
. . . . .
29
Establish Sample Flow
. . .
23
. . . . . . .
19
F
Fluidics
. . . . . . . . . . . . .
9
I
Input
. . . . . . . . . . . . . .
21
Instrument Setup
. . . .
13
,
38
L
Longer Stop of Operation
. .
41
P
Power ON - OFF
Power Supply
. . . . . . .
19
. . . . . .
11
,
21
R
Relay Contacts 1 and 2
. . .
22
S
Sample requirements shut-down
. . . .
11
. . . . . . . . . . .
19
Signal Outputs
Software
. . . . . . . .
22
. . . . . . . . . . . .
26
T
Temperature compensation
Terminals
. .
8
. . . . . .
18
,
21
–
22
W
Wire
. . . . . . . . . . . . . .
17
Z
Zero-Verification
. . . . . . .
34
72
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Notes
12. Notes
A-96.250.701 / 140716
73
AMI INSPECTOR Oxygen
SWAN
is represented worldwide by subsidiary companies and distributors.
cooperates with independent representatives all over the world.
SWAN Products
Analytical Instruments for:
High Purity Water
Feedwater, Steam and Condensate
Potable Water
Pool and Sanitary Water
Cooling Water
Waste Water and Effluents
Made in Switzerland
74
A-96.250.701 / 140716
advertisement