Swann AMI INSPECTOR Oxygen Operator's Manual

Add to My manuals
76 Pages

advertisement

Swann AMI INSPECTOR Oxygen Operator's Manual | Manualzz

AMI INSPECTOR

Oxygen

Version 6.00 and higher

A-96.250.701 / 140716

Customer Support

SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest

SWAN representative, or the manufacturer:

SWAN ANALYTISCHE INSTRUMENTE AG

Studbachstrasse 13

8340 Hinwil

Switzerland

Internet: www.swan.ch

E-mail: [email protected]

Document Status

Title: Monitor AMI INSPECTOR Oxygen Operator’s Manual

ID: A-96.250.701

Revision Issue

00 Sep. 2011 First Edition

01

02

Nov. 2013

Jan. 2016

New main board, integrated USB interface

AMI Inspector Version 2-A (with AMIAKKU mainboard) and Firmware version 6.00

© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved subject to change without notice.

AMI INSPECTOR Oxygen

Table of Contents

1.

Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.1.

Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.2.

General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.

Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.1.

Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.2.

Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2.3.

Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3

4

6

3.

Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

3.1.

Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

3.2.

Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 14

3.2.1

Connect the Sample Inlet to the Quick-Lock Coupling . . . . . . . 14

3.2.2

Connect the Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

3.3

Install the Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . 16

3.4.

Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

3.5.

Connection Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

3.5.1

Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3.6.

Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.6.1

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.6.2

Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.6.3

Relay Contacts 1 and 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

3.7.

Signal Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

4.

Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

4.1.

Establish sample flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

4.2.

Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

5.

Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

5.1.

Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

5.2.

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

5.3.

Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

5.4.

Changing Parameters and values . . . . . . . . . . . . . . . . . . . . . . . . 27

6.

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

6.1.

Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

6.2.

Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 28

6.3.

Maintenance of the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . 29

6.3.1

Electrolyte exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

A-96.250.701 / 140716

1

AMI INSPECTOR Oxygen

6.3.2

Clean Swansensor Oxytrace and G Flow Cell . . . . . . . . . . . . .

6.4.

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.5

Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.6.

Quality Assurance of the Instrument. . . . . . . . . . . . . . . . . . . . . . .

6.6.1

Activate SWAN Quality assurance procedure. . . . . . . . . . . . . .

6.6.2

Pre-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.6.3

Connect the sample lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.6.4

Carry out comparison measurement . . . . . . . . . . . . . . . . . . . . .

6.6.5

Completion of the measurement . . . . . . . . . . . . . . . . . . . . . . . .

6.7.

Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.8.

Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6.9.

Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7.

Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.

Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.1.

Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.2.

Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.3.

Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.4.

Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8.5.

Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9.

Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . .

1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10.

Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11.

Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

12.

Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

50

50

50

52

53

54

45

45

46

47

48

48

36

36

36

38

31

32

34

34

39

40

41

41

42

69

72

73

2

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Safety Instructions

AMI INSPECTOR Oxygen -

Operator’s Manual

This document describes the main steps for instrument setup, operation and maintenance.

1.

Safety Instructions

General The instructions included in this section explain the potential risks associated with instrument operation and provide important safety practices designed to minimize these risks.

If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environment.

More safety instructions are given throughout this manual, at the respective locations where observation is most important.

Strictly follow all safety instructions in this publication.

Target audience

Operator: Qualified person who uses the equipment for its intended purpose.

Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.

OM Location The AMI Operator’s Manual shall be kept in proximity of the instrument.

Qualification,

Training

To be qualified for instrument installation and operation, you must:

 read and understand the instructions in this manual as well as the Material Safety Data Sheets.

 know the relevant safety rules and regulations.

A-96.250.701 / 140716

3

AMI INSPECTOR Oxygen

Safety Instructions

1.1.

Warning Notices

The symbols used for safety-related notices have the following significance:

DANGER

Your life or physical wellbeing are in serious danger if such warnings are ignored.

 Follow the prevention instructions carefully.

WARNING

Severe injuries or damage to the equipment can occur if such warnings are ignored.

 Follow the prevention instructions carefully.

CAUTION

Damage to the equipment, minor injury, malfunctions or incorrect process can be the consequence if such warnings are ignored.

 Follow the prevention instructions carefully.

Mandatory

Signs

The importance of the mandatory signs in this manual.

Safety goggles

Safety gloves

4

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Safety Instructions

Warning Signs The importance of the warning signs in this manual.

Electrical shock hazard

Corrosive

Harmful to health

Flammable

Warning general

Attention general

A-96.250.701 / 140716

5

AMI INSPECTOR Oxygen

Safety Instructions

1.2.

General Safety Regulations

Legal

Requirements

Spare Parts and

Disposables

The user is responsible for proper system operation.

All precautions must be followed to ensure safe operation of the instrument.

Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.

Modifications Modifications and instrument upgrades shall only be carried out by an authorized Service Technician. SWAN will not accept responsibility for any claim resulting from unauthorized modification or alteration.

WARNING

Risk of Electrical Shock

If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.

 To prevent from electrical shock, always make sure that the ground wire is connected.

 Service shall be performed by authorized personnel only.

 Whenever electronic service is required, disconnect instrument power and power of devices connected to.

– relay 1,

– relay 2,

– alarm relay

WARNING

For safe instrument installation and operation you must read and understand the instructions in this manual.

WARNING

Only SWAN trained and authorized personnel shall perform the tasks described in this document.

6

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Product Description

2.

Product Description

This chapter contains technical data, requirements and performance data.

2.1.

Description of the System

The portable AMI INSPECTOR instrument is a complete monitoring system mounted on a small panel with supporting stand and a rechargeable battery for stand-alone operation (>24h), designed as an inspection equipment for quality assurance of online process monitors.

Features General Features of AMI INSPECTORs are:

 Battery life after full charge:

– >24h at full load (use of 3 relays, USB, signal output, logger)

– >36h at minimum load (use of logger only)

 Charging time: approx. 6 hours

 Controlled shut-down if battery is empty.

 Display of remaining battery life in hours.

 For longer battery life the back light of the LC Display is disabled.

 Continuous operation using power adapter. The battery should be discharged at least once a month (normal usage until the monitor automatically shuts down).

Battery The Li-Ion battery is located in the housing of the AMI transmitter.

See chapter Power Supply, p. 19 regarding power supply and charging of the battery.

Safety features No data loss after power failure, all data is saved in non-volatile memory. Over voltage protection of in- and outputs.

Galvanic separation of measuring inputs and signal outputs.

USB interface Built in USB interface for logger download. Use the USB stick supplied by Swan only (other USB sticks can dramatically reduce battery life).

Signal Output

(optional)

One signal output programmable for measured values (freely scaleable, linear or bilinear) or as continuous control output (control parameters programmable).

Current loop:

Maximal burden:

0/4–20 mA

510 

A-96.250.701 / 140716

7

AMI INSPECTOR Oxygen

Product Description

Relay Two potential-free contacts programmable as limit switches for measuring values, controllers or timer for system cleaning with automatic hold function.

Maximum load: 100 mA/50 VAC

Alarm Relay One potential free contact.

Alternatively:

 Open during normal operation, closed on error and loss of power.

 Closed during normal operation, open on error and loss of power.

Summary alarm indication for programmable alarm values and instrument faults.

Input For potential-free contact to freeze the measuring value or to interrupt control in automated installations (hold function or remote-off)

Measuring principle

Clark principle:

The sensor consists of one noble metal electrode (e.g. platinum or gold), a reference electrode (mostly Ag/AgCl) and optionally a metal guard electrode.

The Clark-type electrode is the most widely used oxygen sensor for measuring oxygen dissolved in a liquid. The basic principle is that there is a cathode and an anode submersed in an electrolyte and a voltage is applied between the two parts. Oxygen enters the sensor through a permeable membrane by diffusion, and is reduced at the cathode according to

-

+ 2 H

-

O

2

+ 4e

This reaction creates a measurable current. There is a linear correlation between the oxygen concentration and the electrical current.

2

O --> 4 OH

The guard electrode is on the same voltage level as the cathode but there is no current measurement. Oxygen which diffuses from the electrolyte to the cathode is consumed by the guard electrode.

As a consequence, residual oxygen in the electrolyte will no more disturb the measurement signal and the response time to low oxygen levels will be shorter.

Temperature compensation

The measuring signal depends on temperature, but is automatically compensated to 25 °C. The sample temperature is determined continuously by a temperature sensor inside the oxygen electrode.

8

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Product Description

Fluidics Swansensor oxygen combined with QV-flow PMMA OTG flow cell.

The sample flows via sample inlet [E] through the flow regulating valve [D], where the flow rate can be adjusted. Then the sample flows into the measuring cell [C] were the Oxygen concentration and temperature of the sample is measured.

The sample leaves the measuring cell via flow sensor [B] through the sample outlet [F].

A-96.250.701 / 140716

A

B

C

Oxygen sensor

Flow sensor

Flow cell

A

B

C

D

E

F

D

E

F

Flow regulating valve

Sample inlet

Sample outlet

9

AMI INSPECTOR Oxygen

Product Description

2.2.

Instrument Overview

A

B

C

D

E

F

G

10

A

B

C

D

E

AMI Transmitter

Oxygen sensor

Flow cell

Flow sensor

Flow regulating valve

F

G

Sample inlet

Sample outlet

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Product Description

2.3.

Technical Data

Power Supply Battery

Use original, supplied power adapter only.

Voltage:

Power consumption:

Charging time:

Battery type:

85–265 VAC, 50/60 Hz max. 20 VA

6h

Li-Ion

During charging protect from heat impact and keep splash-proof

(not IP66).

Operating time Stand-alone (Battery):

Connected adapter:

> 24h continuous

Controlled shut-down when battery is empty, remaining time is displayed.

Electronics Aluminium with a protection degree of IP 66 / NEMA 4X housing Ambient temperature:

Humidity:

Display:

-10 to +50 °C

10–90% rel., non condensing backlit LCD, 75 x 45 mm

Sample requirements

Flow cell and connection

Accuracy

Reproducibility

Flow rate:

Temperature:

Inlet pressure: pH:

Suspended solids:

Outlet pressure:

8 to 25 l/h up to 45 °C

0.2 to 1 bar not lower than pH 4 less than 10 ppm pressure free

Flow cell made of acrylic glass with built-in flow adjustment valve and digital sample flow meter

Sample inlet:

Sample outlet:

1/4” Swagelok tube adapter flexible tube 8x6 mm

±1.5 % of measured value or ±0.2 ppb

±1 % of measured value or ±0.15 ppb

A-96.250.701 / 140716

11

AMI INSPECTOR Oxygen

Product Description

Sensor

Oxytrace G

Technical data:

Measuring range:

Accuracy:

Precision:

Response time:

Minimal flow:

Operating temperature:

Material:

Protection:

Weight:

Sensor for the measurement of dissolved oxygen in ultra pure water. Precise oxygen measuring cell with integrated temperature sensor and guard electrode for faster initial response time after maintenance.

Clark oxygen electrode

Cathode gold, anode silver, guard silver

Zero current-free electrode system

Robust 25 µm fluoropolymer diaphragm

0–20 ppm O

2

(25 °C)

Automatic range switching

Range

0.1 to 9.99 ppb

10 to 199.9 ppb

200 to 1999 ppb

2 to 20 ppm

0 – 200% saturation

Resolution

0.01 ppb

0.1 ppb

1.0 ppb

0.01 ppm

0,3 % if calibration temperature = measuring temp. 1,5% at ± 10

°C deviation to cal. temperature

± 1% of reading or ±0.15 ppb t90 < 30 seconds (rising concentration)

50 cm/s Pressure resistance: 3 bar max. 50 °C polyacetal copolymer

IP 68

150 g

12

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Installation

3.

Installation

3.1.

Installation Check List

Check

Installation

Power-up

Instrument

Setup

Run-in period

 Instrument’s specification must conform to your AC power ratings. See External power adapter, p. 20 .

 Check if the battery is fully charged.

 Connect the sample inlet and outlet, see Connecting Sample

Inlet and Outlet, p. 14 .

 Install the Swansensor Oxytrace G into the flow cell, see Install the Swansensor Oxytrace G, p. 16 .

 Open the flow regulating valve and adjust the sample flow to

8–25 l/h.

 Switch on power.

 Program all parameters, see chapter 4.

 Let the instrument run continuously for 1 h.

A-96.250.701 / 140716

13

AMI INSPECTOR Oxygen

Installation

3.2.

Connecting Sample Inlet and Outlet

3.2.1

Connect the Sample Inlet to the Quick-Lock Coupling

The AMI INSPECTOR Oxygen is delivered with a quick-lock coupling. To connect the sample line to the AMI INSPECTOR Oxygen, simply push the nipple into the quick-lock coupling.

A

B

C

D

Flow cell block

Quick-lock coupling

Nipple

Sample outlet

A

B

C

D

14

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Installation

3.2.2

Connect the Sample Outlet

Installation 1 Loosen the union nut [B] but do not remove it.

2 Push the FEP tube [A] through the union nut [B] as far as it reaches the stop of the threaded tube [E].

3 Tighten the union nut 1¾ rotation using an open ended spanner.

Hold Body from turning with a second wrench.

4 Put the FEP Tube into a pressure free drain with sufficient capacity.

A B C D E

F

A

B

C

FEP tube 8x6

Union nut

Compression ferrule

D

E

F

Compression cone

Threaded tube

Tightened connection

A-96.250.701 / 140716

15

AMI INSPECTOR Oxygen

Installation

3.3

Install the Swansensor Oxytrace G

The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber [E]. A transport protection cap [B] filled with water [D] keeps the sensor wet during transport and storage. To install the sensor proceed as follows:

A

B

C

A

B

C

D

E

Fixing sleeve

Transport protection cap

Swansensor Oxytrace G

Water

Sensor cap filled with electrolyte

D

E

1 Unscrew the fixing sleeve [A].

2 Remove the transport protection cap [B].

3 Clean the Swansensor Oxytrace G [C] with water.

4 Install the Swansensor Oxytrace G into the flow cell

Connect the sensor cable to the transmitter, see Connection Diagram, p. 18 .

16

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Installation

3.4.

Electrical Connections

Cable thicknesses

WARNING

Always turn off DC power before manipulating electric parts.

Make sure the power specification of the instrument corresponds to the power on site.

In order to comply with IP66, use the following cable thicknesses

Wire

A

B

A B

PG 9 cable gland: cable Ø outer

PG 7 cable gland: cable Ø outer

4–8 mm

3–6.5 mm

NOTICE: Protect unused cable glands

 For Relays: Use max. 1.5 mm

2 end sleeves.

/ AWG 14 stranded wire with

 For Signal Outputs and Input: Use 0.25 mm

2 stranded wire with end sleeves.

/ AWG 23

WARNING

External Voltage.

External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.

 Make sure that the devices connected to the following contacts are disconnected from the power before continuing the installation.

– relay 1

– relay 2

– alarm relay

A-96.250.701 / 140716

17

AMI INSPECTOR Oxygen

Installation

3.5.

Connection Diagram

18

CAUTION

Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Installation

3.5.1

Power Supply

Contrary to all other Swan online process monitors the

AMI INSPECTOR transmitter is supplied with power by battery only. The rechargeable battery (Li-Ion) enables a stand-alone operation for at least 24 hours.

WARNING

Do not provide power directly to the transmitter as this will destroy the motherboard. All AMI INSPECTOR transmitters are supplied with power by battery only.

Charging Use the original supplied power adapter to charge AMI INSPEC-

TOR only. Charging time: approx 6h.

Fully charged a stand-alone operating time of at least 24h is guaranteed:

 >24h at full load (use of 3 relays, USB, signal output, logger)

 >36h at minimal load (use of logger only)

In case that the battery is discharged completely the firmware will automatically shut down.

Switch Power

ON - OFF

Continuous operation

Switch the instrument ON or OFF using the toggle switch on the transmitter.

For continuous operation use the power adapter as well.

CAUTION

 If the AMI powers ON and then immediately shuts OFF, the battery is empty. Do not hold the toggle switch in ON position, as this can damage the battery.

CAUTION

 During charging protect from heat impact and keep splashproof (plug of power adapter is not IP66).

 Do not supply external devices, e.g. pumps, magnetic valves or any other current consumers with AMI INSPECTOR

CAUTION

 Use the original supplied power adapter to charge AMI IN-

SPECTOR only. Use of any other power adapter can damage the battery or cause malfunction.

A-96.250.701 / 140716

19

AMI INSPECTOR Oxygen

Installation

External power adapter

 Universal input range

85 - 265 VAC

 Continuous short circuit protection

 Over voltage protection

 LED indicator for power on

 2-pin AC inlet (IEC 320-C8) for detachable country-specific power cord

Power cords Two different power cords are supplied:

 Power cord with type C plug (Europlug)

 Power cord with type A plug (NEMA-1)

If a different plug type is needed, please purchase a suitable power cord from your local supplier.

NOTICE: Do not use any plug adapters.

Dimensions

Unit: mm

20

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Installation

3.6.

Relay Contacts

Programming of the relay contacts see 5.3 Relay Contacts, p. 59

3.6.1

Input

NOTICE: Use only potential-free (dry) contacts.

Terminals 13/14

If signal output is set to hold, measurement is interrupted if input is active.

For programming see menu 5.3.4, p. 65

3.6.2

Alarm Relay

NOTICE: For resistive loads only; do not use with capacitive or inductive loads. Max. load 1 A / 250 VAC.

Alarm output for system errors.

Error codes see Error List, p. 42

Programming see menu 5.3.1, p. 59

NOTICE: With certain alarms and certain settings of the AMI transmitter the alarm relay does not switch. The error, however, is shown on the display.

NC

1)

Normally

Closed

NO

Normally

Open

Terminals

5/4

5/3

Description

Active (opened) during normal operation.

Inactive (closed) on error and loss of power.

Active (closed) during normal operation.

Inactive (opened) on error and loss of power.

0V

Relay connection

1) 5

0V

4

3

5

4

3

1) usual use

A-96.250.701 / 140716

21

AMI INSPECTOR Oxygen

Installation

3.6.3

Relay Contacts 1 and 2

NOTICE: For resistive loads only; do not use with capacitive or inductive loads.Rated load 100 mA/ 50 VAC.

For programming see Menu Installation 5.3.2 and 5.3.3, p. 61

NO

Normally

Open

Terminals

6/7: Relay 1

8/9: Relay 2

Description

Inactive (opened) during normal operation and loss of power.

Active (closed) when a programmed function is executed.

Relay connection

0V

6/8

7/9

3.7.

Signal Output

The optional signal output 0/4–20 mA PCB can be plugged onto the USB interface PCB.

NOTICE: Max. burden 510

Terminals 16 (+) and 15 (-).

For programming see menu 5.2 Signal Outputs, p. 55 .

A B

22

A

B

Signal output 0/4–20 mA PCB

USB interface

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Instrument Setup

4.

Instrument Setup

4.1.

Establish sample flow

1 Open flow regulating valve, see Fluidics, p. 9 .

2 Wait until the flow cell has been completely filled.

3 Switch on power.

4 Adjust the sample flow to 8–25 l/h.

5 Let the instrument run-in for 1 h.

4.2.

Programming

Programming Program all parameters for external devices (interface, recorders, etc.)

Program all parameters for instrument operation (limits, alarms), see Program List and Explanations, p. 50

A-96.250.701 / 140716

23

AMI INSPECTOR Oxygen

Operation

5.

Operation

5.1.

Keys

A B C D

Exit Enter

A to exit a menu or command (rejecting any changes) to move back to the previous menu level

B to move DOWN in a menu list and to decrease digits

C to move UP in a menu list and to increase digits

D to open a selected sub-menu to accept an entry

Program

Access, Exit

RUN

R1

R2

9 l/h

14:10:45

1.05 ppb

25.4°C

Enter

Exit

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

1

24

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Operation

5.2.

Display

A B C D E

RUN

R1

R2

15 15:20:18

21.5

ppb

25 °C

F

G

A RUN

HOLD

OFF

B ERROR normal operation input closed or cal delay: Instrument on hold (shows status of signal outputs).

input closed: control/limit is interrupted (shows status of signal outputs).

Error Fatal Error

C Relay status

D Battery status (remaining operating time in h)

E Time

F Process values

G Sample Temperature

Relay status, symbols upper/lower limit not yet reached upper/lower limit reached control upw./downw. no action control upw./downw. active, dark bar indicates control intensity motor valve closed motor valve: open, dark bar indicates approx. position timer timer: timing active (hand rotating)

A-96.250.701 / 140716

25

AMI INSPECTOR Oxygen

Operation

5.3.

Software Structure

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

Messages

Pending Errors

Maintenance List

Message List

1

1.1

Menu Messages 1

Reveals pending errors as well as an event history

(time and state of events that have occurred at an earlier point of time).

It contains user relevant data.

Menu Diagnostics 2

Provides user relevant instrument and sample data.

Diagnostics

Identification

Sensors

Sample

I/O State

Interface

2.1

Maintenance 3.1

Calibration

Service

Simulation

Set Time 23.09.06 16:30:00

Operation

Sensors

Relay Contacts

Logger

4.1

Installation

Sensors

Signal Outputs

Relay Contacts

Miscellaneous

Interface

5.1

Menu

For instrument calibration, relay and signal output simulation, and to set the instrument time.

It is used by the service personnel.

Menu

Maintenance 3

Operation 4

User relevant parameters that might need to be modified during daily routine. Normally password protected and used by the process-operator.

Subset of menu 5 - Installation, but process-related.

Menu Installation 5

For initial instrument set up by SWAN authorized person, to set all instrument parameters. Can be protected by means of password.

26

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Operation

5.4.

Changing Parameters and values

Changing parameters

The following example shows how to change the logger interval:

1 Select the parameter you want to change.

2 Press [Enter]

Temperature

Standards

NT5K

Logger

Log interval Interval.

5 min

10 min

30 min

1 Hour

4.1.3

30min no

3 Press [ ] or [ ] key to highlight the required parameter.

4 Press [Enter] to confirm the selection or [Exit] to keep the previous parameter).

Logger

Log interval

Clear logger

4.1.3

10 min no

 The selected parameter is highlighted (but not saved yet).

5 Press [Exit].

Logger

Log interval

Clear logger

Yes

No

4.1.3

no

 Yes is highlighted.

6 Press [Enter] to save the new parameter.

 The system reboots, the new parameter is set.

Changing values

A-96.250.701 / 140716

Alarm oxygen

Alarm High

Alarm Low

Hysteresis

Delay

5.3.1.1.1

10.00 ppm

1.00 ppb

0.10 ppb

5 Sec

Alarm oxygen

Alarm High

Alarm Low

Hysteresis

Delay

5.3.1.1.1

8.00 ppb

1.00 ppb

0.10 pp

5 Sec

1 Select the value you want to change.

2 Press [Enter].

3 Set required value with [

[ ] key.

] or

4 Press [Enter] to confirm the new value.

5 Press [Exit].

 Yes is highlighted.

6 Press [Enter] to save the new value.

27

AMI INSPECTOR Oxygen

Maintenance

6.

Maintenance

Maintenance frequency depends strongly on the water quality. The

AMI INSPECTOR Oxygen is designed for determination of low level of dissolved oxygen in high purity water.

It is not suitable for the measurement of dissolved oxygen in waste water.

6.1.

Maintenance Table

Monthly

Half-yearly

Yearly

If necessary, perform an air calibration.

Clean Swansensor Oxytrace G membrane with a soft tissue.

If necessary, replace filling electrolyte.

If the sensor is exposed to air frequently during long time intervals, the electrolyte and membrane may have to be changed earlier (see below*).

Clean flow cell and flow meter, if dirty.

Replace Swansensor Oxytrace G membrane by using a new, prefilled sensor cap.

Every 2nd year

*A change of membrane and electrolyte is recommended:

 if indicated in the maintenance list (remaining amount

<10%)

 if the response of the sensor is slow

 if the sensor can not be calibrated any more and/or the instrument shows a corresponding error message

 if the sensor signal is very unstable.

6.2.

Stop of Operation for Maintenance

1 Shut off power of the instrument.

2 Stop sample flow by closing the flow regulating valve.

28

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

6.3.

Maintenance of the Oxygen Sensor

WARNING

Etching liquid

The electrolyte is alkaline and caustic. It contains less than 1% of potassium hydroxide.

 Do not ingest. Wear protective goggles and gloves during handling. Avoid contact with clothes.

 In case of accidental contact with the eyes, wash immediately with clear water and contact a physician. Show him the label of the bottle or this section of the manual.

 Short contact with skin is harmless, nevertheless wash with lots of water.

6.3.1

Electrolyte exchange

An electrolyte exchange is indicated in the maintenance list as soon as the remaining amount is below 10%.

A

A

B

C

Fixing sleeve

Sensor

Flow cell

B

C

1 Unscrew the fixing sleeve [A].

2 Remove the sensor from the flow cell.

A-96.250.701 / 140716

29

AMI INSPECTOR Oxygen

Maintenance

A

G

H

E

F

B

C

D

E

F

G

H

A

B

C

D

Swansensor Oxytrace G

Thread

Groove

Guard electrode

Anode

Measuring head

Cathode

Sensor cap with membrane

30

3 Unscrew and remove the sensor cap [H] from the Swansensor

Oxytrace G [A].

4 Empty the remaining electrolyte.

5 Refill the sensor cap with fresh electrolyte.

NOTICE: There is a groove [C] in the thread [B] of the sensor, where excessive air and electrolyte can escape while screwing the sensor cap onto the sensor. Hold the sensor in vertical position, measuring head pointing downwards.

6 Slowly screw the sensor cap onto the sensor to allow the excess electrolyte to escape without building up too much pressure inside the electrode. Fasten sensor cap tightly.

7 Clean the sensor thoroughly and dry the sensing membrane with a soft tissue.

8 Switch on power.

9 Let the sensor run-in at air for at least 30 min, better 1 h.

10 Afterwards perform an air calibration.

11 Install the sensor into the flow cell.

12 Select “New Filling” to reset the counter for remaining electrolyte, see <Maintnence>/<Service> 3.2.1, p. 52 ).

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

6.3.2

Clean Swansensor Oxytrace and G Flow Cell

Depending on the water quality, the Swansensor Oxytrace G and the flow cell will necessitate a cleaning.

Before cleaning, stop operation as described in Stop of Operation for Maintenance, p. 28 .

1 Dismount the Swansensor Oxytrace G, see Electrolyte exchange, p. 29 .

2 Clean the sensor with a soft tissue and rinse it with water afterwards.

3 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.

4 Flush the flow cell with clean water.

5 Install the Swansensor Oxytrace G and start sample flow.

A-96.250.701 / 140716

31

AMI INSPECTOR Oxygen

Maintenance

6.4.

Calibration

The sensing part of the sensor must not be in direct contact with water!

In the wet flow cell, the atmosphere will be saturated with water vapor. This atmosphere will produce the most accurate calibration results.

The required time for a calibration depends mainly on the difference between temperature and oxygen content in the sample and in the air. It can take 15 – 20 minutes. This is also the case, if the electrolyte was exchanged.

As soon as the reading is stable, the microprocessor will store the calibration data in the memory. The end of the calibration is indicated on the display.

To perform a calibration proceed as follows:

1 Navigate to menu <Maintenance> /<Calibration>.

2 Press [Enter] to start the calibration and follow the dialog on the display.

Calibration 3.1.5

3 Stop the sample flow at with the flow regulating valve.

Close regulating valve to turn off sample flow.

<Enter> to continue

Calibration

Take sensor out of flow cell and dry membrane and sensor

3.1.5

<Enter> to continue

4 Unscrew and remove the threaded sleeve [A], see Electrolyte exchange, p. 29 .

5 Remove the oxygen sensor [B] from the flow cell [C].

6 Dry the sensor membrane and the flow cell with a soft paper tissue.

32

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

Calibration

Place the electrode into the wet flow cell at a slightly tilted angle.

3.1.5

<Enter> to continue

A

B

A-96.250.701 / 140716

Calibration

Saturation

Sat. Current

Progress

Calibration

Saturation

Sat. Current

3.1.1

98.7 %

32  A

3.1.1

98.7 %

32  A

A Tilted sensor

B Flow cell

7 Press [Enter] to start the calibration measurement.

 The saturation should reach

100%, the saturation current should be about 24  A to 41  period, the calibration will be discarded.

8 Press [Enter] to confirm the calibration.

A.

If the measuring values are not stable during the measuring

Calibration Successful

33

AMI INSPECTOR Oxygen

Maintenance

6.5

Zero-Verification

Swansensor Oxytrace G for the measurement of low oxygen content (< 1 ppb).

1 Calibrate the sensor according to chapter Calibration, p. 32 .

2 Prepare a 5%-sodium sulfite solution with demineralized water.

3 Put the electrode into the sodium sulfite solution afterwards. Assure that there are no air bubbles in front of the sensor.

4 The measured value should now be < 1 ppb.

NOTICE: Depending on the state of the electrode this process can take several hours. In case of an electrode-refill, this can take days until the measured value is lower than 1 ppb.

6.6.

Quality Assurance of the Instrument

Quality assurance level

Every SWAN on-line instrument is equipped with integrated, autonomous quality assurance functions to survey the plausibility of each measurement.

For the AMI Oxytrace / AMI Oxytrace QED these are:

 continuous monitoring of sample flow

 continuous monitoring of the temperature inside the transmitter case

 periodic accuracy test with ultra high precision resistors

Further a manual, menu driven inspection procedure can be carried out using a certified reference instrument. Running at the same sampling point as an inspection equipment, the AMI Inspector Oxygen verifies the measuring results. After enabling the quality assurance procedure by defining the quality assurance level, the instrument reminds the user periodically to run the procedure and results are stored in a history for review.

Central feature of the quality assurance function is the assignment of the monitored process to a Quality assurance level.

There are three predefined levels plus a user level. Hereby the inspection interval, the deviation limits of temperature and measuring result between the inspection equipment and the monitoring instrument are defined.

34

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

 Level 1: Trend; Measurement used as an additional information to follow the process indicating trends.

 Level 2: Standard; Monitoring of several parameters of a process (e.g. oxygen, saturation). In case of instrument failure, other parameters can be used for process monitoring.

 Level 3: Crucial; Monitoring of critical processes, value is used for control of another part or subsystem (valve, dosing unit, etc.).

Additional level:

 Quality level 4: User; User defined inspection interval, maximal deviation of temperature and measuring result.

Quality Level

0: Off

1: Trend

2: Standard

3: Crucial

4: User max. deviation temperature [°C] a)

Off

0.5 °C

0.4 °C

0.3 °C

0 - 2°C max. deviation result [%]

Off

10 %

5 %

5 %

0 - 20% min. inspection interval

Off annual quarterly monthly annual, quarterly, monthly a) sample temperature must be 25 °C +/-5 °C.

Procedure The standard workflow includes the following steps:

1 Activate SWAN Quality assurance procedure

2 Pre-test

3 Connect instruments

4 Carry out comparison measurement

5 Completion of the measurement

NOTICE: The procedure should only be carried out through qualified personnel.

Materials / Inspection equipment:

 Reference instrument: AMI Inspector Oxygen

 Two tubes made of PA

A-96.250.701 / 140716

35

AMI INSPECTOR Oxygen

Maintenance

6.6.1

Activate SWAN Quality assurance procedure

Enable quality assurance procedure on the process monitor(s) which shall be checked by selecting the quality level in menu

5.1.2.1. The corresponding submenus are then activated.

NOTICE: The activation is necessary the first time only.

6.6.2

Pre-test

 Reference instrument: AMI Inspector Oxygen:

– Check certificate; reference instrument certificate not older than one year.

– Check battery; Battery of the AMI Inspector Oxygen should be completely charged. Remaining operating time on display minimum 20 hours.

– Make sure that all instruments are set to the same temperature compensation.

– Sensor is in working condition

 On-line instrument: Monitor AMI Oxytrace:

– Good order and condition; Flow cell free of particles,

Sensor surface free of deposits.

– Check message list; Review the message list in menu 1.3 and check for frequently occurring alarms (as for example flow alarms). If alarms occur frequently remove cause before starting the procedure.

6.6.3

Connect the sample lines

See corresponding chapter in the manual of the process monitor which shall be checked with a reference instrument.

The choice of sampling depends strongly on local conditions on site. Possible sampling:

 via sample point,

 via T-fitting or

 via piggyback/downstream

NOTICE:

• avoid ingress of air, use screwed fitting,

• sample as near as possible to the process monitor,

• wait approx. 10 minutes, whilst measurement is running, until measurement value and temperature are stabilized.

36

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

Example As an example following picture shows the connection of the reference instrument via T-fitting to the process monitor.

A

B

C

D

E

F

A-96.250.701 / 140716

G

A

B

C

D

Monitor AMI Oxytrace

AMI Inspector Oxygen

Reference flow cell

On-line flow cell

E

F

G

Sample outlet

Sample inlet

T-fitting

1 Stop sample flow to the monitor AMI Oxytrace by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell.

2 Connect sample line of the monitor AMI Oxytrace [A] with the sample inlet of the reference instrument AMI Inspector Oxygen

[B]. Use the supplied tube made of PA.

3 Connect sample outlet of the reference instrument AMI Inspector Oxygen to the sample outlet funnel of the monitor.

4 Switch on the AMI Inspector Oxygen. Open the flow regulating valve and regulate the sample flow to 10 l/h. The actual flow is shown on the transmitter.

37

AMI INSPECTOR Oxygen

Maintenance

6.6.4

Carry out comparison measurement

Start by selecting Quality Assurance in menu 3.5 of the monitor AMI

Oxytrace.

1 Navigate to menu Maintenance /Quality Assurance.

2 Press [Enter].

3 Follow the dialog on the Display.

Quality Assurance

- Carry out preparations

- Install Inspector

- Sample flow to 10 l/h

3.5.5

4 Carry out pre test preparations

Connect instruments.

Regulate sample flow to 10 l/h using the appropriate valve.

<Enter> to continue

Quality Assurance

Value O2

Value Temp.

Wait 10 Minutes

3.5.5

0.05 ppb

25.00  C

5 Wait 10 minutes whilst measurement is running.

Press [Enter] to continue.

<Enter> to continue

Quality Assurance

Value O2

Value Temp.

Inspector O2

Inspector Temp.

<Enter> to continue

3.5.3

0.05 ppb

25.00  C

0.06 ppb

25.0  C

6 Read the ppb value of the reference instrument and enter under

[

“Inspector.” by using the [

] keys.

] or

7 Press [Enter] to confirm.

Quality Assurance

Value O2

Value Temp.

Inspector

Inspector Temp.

3.5.4

0.05 ppb

25.00  C

0.06 ppm

25.0  C

<Enter> to continue

Quality Assurance

Max. Dev. O2

Max. Dev. Temp.

Dev. O2

Dev. Temp.

QA-Check succesful

3.5.5

0.5 %

0.4 °C

0.1 %

0.4 °C

8 Read temperature value of the reference instrument and enter un-

[ der “Inspector Temp.” by using the

] or [ ] keys.

9 Press [Enter] to confirm.

10 Press [Enter] to continue.

 The results are saved in QA-

History regardless if successful or not

38

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

6.6.5

Completion of the measurement

1 Stop the sample flow to the AMI Oxytrace by closing the appropriate valve, e.g. back pressure regulator, sample preparation or flow regulating valve at flow cell again.

2 Close flow regulating valve of the AMI Inspector.

3 Disconnect the AMI Inspector by removing the tubes and connect the sample outlet of the Monitor AMI Oxytrace to the sample outlet funnel again.

4 Start sample flow again and regulate sample flow.

5 Shutdown the AMI Inspector Oxygen.

For longer stop of operation see Longer Stop of Operation, p. 41

A-96.250.701 / 140716

39

AMI INSPECTOR Oxygen

Maintenance

6.7.

Replacing Fuses

WARNING

External Voltage.

External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.

 Make sure that the devices connected to the following contacts are disconnected from the power before continuing the installation.

– relay 1

– relay 2

– alarm relay

Find and repair the cause for the short circuit before replacing the fuse. Use tweezers or needle-nosed pliers to remove the defective fuse.

Use original fuses provided by SWAN only.

A

40

A 1.25 AF/250V Instrument power supply

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Maintenance

6.8.

Replacing the Battery

A B C A

B

C

Battery

Battery plug

Ribbon cable

1 Switch the AMI Inspector off.

2 If connected, disconnect the power adapter from the power jack.

3 Open the transmitter housing.

4 Pull out the ribbon cable [C] from the mainboard.

5 Disconnect battery plug [B] and replace the battery.

6.9.

Longer Stop of Operation

1 Shut off power of the instrument.

2 Stop sample flow.

3 Remove the Swansensor Oxytrace G.

4 Clean the sensor with a soft tissue and rinse it with water afterwards.

5 Use a soft brush to remove the dirt which sticks on the walls of the flow cell.

6 Fill the flow cell with water.

7 Install the Swansensor Oxytrace G.

A-96.250.701 / 140716

41

AMI INSPECTOR Oxygen

Error List

42

7.

Error List

Error

Non-fatal Error. Indicates an alarm if a programmed value is exceeded.

Such Errors are marked E0xx (bold and black).

Fatal Error symbol)

Control of dosing devices is interrupted.

The indicated measured values are possibly incorrect.

Fatal Errors are divided in the following two categories:

 Errors which disappear if correct measuring conditions are recovered (i.e. Sample Flow low).

Such Errors are marked E0xx (bold and orange)

 Errors which indicate a hardware failure of the instrument.

Such Errors are marked E0xx ( bold and red )

HOLD

R1

R2

8 l/h

14:10:45

7.04 ppm

25.4°C

Error or fatal Error

Error not yet acknowledged.

Check Pending Errors 1.1.5 * and take corrective action.

Press [ENTER].

Main Menu

Messages

Diagnostics

Maintenance

Operation

Installation

1

Navigate to menu Messages.

Press [ENTER].

Messages

Pending Errors

Maintenance List

Message List

1.1

Navigate to menu Pending Errors.

Press [ENTER].

Pending Errors

Error Code

Alarm low

<Enter> to Acknowledge

1.1.5

E002

Press [ENTER] to acknowledge the

Pending Errors. The Error is reset and saved in the Message List.

* Menu numbers see

Program Overview, p. 45

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Error List

Error

E001

E002

Description oxygen Alarm high oxygen Alarm low

E003 Saturation Alarm high

E004

E007

E008

E009

E010

E011

E012

Saturation Alarm low

Sample Temp. high

Sample Temp. low

Sample Flow high

Sample Flow low

Temp. shorted

Temp. disconnected

Corrective action

– check process

– check programmed value, see 5.3.1.1.1, p. 59

– check process

– check programmed value, see 5.3.1.1.25, p. 59

– check process

– check programmed value, see 5.3.1.4, p. 60

– check process

– check programmed value, see 5.3.1.4, p. 60

– check process

– check programmed value, see 5.3.1.3.1, p. 60

– check process

– check programmed value, see 5.3.1.3.22, p. 60

– check sample flow

– check programmed value, see 5.3.1.2.2, p. 60

– establish sample flow

– clean instrument

– check programmed value, see 5.3.1.2.32, p. 60

– Check wiring of sensor

– Check sensor

– Check wiring of sensor

– Check sensor

A-96.250.701 / 140716

43

AMI INSPECTOR Oxygen

Error List

Error

E013

Description

Case Temp. high

E014 Case Temp. low

E017 Control Timeout

E019 Quality Assurance

E024 Input active

E026

E030

E031

E032

E033

E034

E065

IC LM75

EEProm Frontend

Calibration Recout

Wrong Frontend

Power-on

Power-down

Electrolyte depleted

Corrective action

– check case/environment temperature

– check programmed value, see 5.3.1.5.1, p. 61

– check case/environment temperature

– check programmed value, see 5.3.1.5.2, p. 61

– check control device or programming in

Installation, Relay contact, Relay 1/2 see 5.3.2 and 5.3.3, p. 61

– Perform QA Procedure using a reference instrument, e.g. AMI Inspector

– See If Fault Yes is programmed in Menu see 5.3.4, p. 65

– call service

– call service

– call service

– call service

– none, normal status

– none, normal status

– Refill electrolyte, see Electrolyte exchange, p. 29

44

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program Overview

8.

Program Overview

For explanations about each parameter of the menus see Program

List and Explanations, p. 50 .

 Menu 1 Messages is always accessible for everybody. No password protection. No settings can be modified.

 Menu 2 Diagnostics is always accessible for everybody. No password protection. No settings can be modified.

 Menu 3 Maintenance is for service: Calibration, simulation of outputs and set time/date. Please protect with password.

 Menu 4 Operation is for the user, allowing to set limits, alarm values, etc. The presetting is done in the menu Installation

(only for the System engineer). Please protect with password.

 Menu 5 Installation : Defining assignment of all inputs and outputs, measuring parameters, interface, passwords, etc.

Menu for the system engineer. Password strongly recommended.

8.1.

Messages (Main Menu 1)

Pending Errors 1.1.5* Pending Errors

1.1*

Maintenance List

1.2*

Message List

1.3*

Maintenance List

Number

Date, Time

1.2.5*

1.3.1*

* Menu numbers

A-96.250.701 / 140716

45

AMI INSPECTOR Oxygen

Program Overview

Identification

2.1*

8.2.

Diagnostics (Main Menu 2)

Desig.

Version

Factory Test

2.1.3*

AMI Oxytrace

6.00-11/15

Instrument

Motherboard

2.1.3.1*

Front End

Years / Days / Hours / Minutes / Seconds

Sensors

2.2*

Operating Time

2.1.4*

Sensor

2.2.1*

Current Value

(Raw value tc)

(Raw value)

Saturation

Cal. History

2.2.1.5*

Number

Date, Time

Sat. Current

Air pressure

2.2.2.1*

Sample

2.3*

I/O State

2.4*

Interface

2.5*

Miscellaneous

2.2.2*

QA History

2.2.3*

Sample ID

Temperature °C

Nt5K Ohm

Alarm Relay

Relay 1/2

Input

Signal Output 3

Protocol

USB Stick

Case Temp.

Air pressure

QA History

2.3.1*

2.4.1*

2.4.2*

2.5.1*

2.2.3.1*

* Menu numbers

2.1.4.1*

2.2.1.5.1*

46

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program Overview

Calibration

3.1*

Service

3.2*

Simulation

3.3*

8.3.

Maintenance (Main Menu 3)

Calibration 3.1.5

Electrolyte

3.2.1*

Alarm Relay

Relay 1

Relay 2

Signal Output 3

(Date), (Time)

Last filling

Remaining amount

Remaining time

New Filling

3.3.1*

3.3.2*

3.3.3*

3.3.6*

3.2.1.5*

Set Time

3.4*

Quality Assurance

3.5*

Quality Assurance 3.5.5*

* Menu numbers

A-96.250.701 / 140716

47

AMI INSPECTOR Oxygen

Program Overview

8.4.

Operation (Main Menu 4)

Sensors

4.1*

Relay Contacts

4.2*

Filter Time Const.

Hold after Cal.

Alarm Relay

4.2.1*

4.1.1*

4.1.2*

Alarm Oxygen

4.2.1.1*

Logger

4.3*

Relay 1/2

4.2.2* - 4.2.3*

Input

4.2.4*

Log Interval

Clear Logger

Eject USB Stick

Alarm Saturation

4.2.1.2*

Setpoint

Hysteresis

Delay

Active

Signal Outputs

Output / Control

Fault

Delay

4.3.1*

4.3.2*

4.3.3*

Alarm High

Alarm Low

Hysteresis

Delay

Alarm High

Alarm Low

Hysteresis

Delay

4.2.x.100*

4.2.x.200*

4.2.x.30*

4.2.4.1*

4.2.4.2*

4.2.4.3*

4.2.4.4*

4.2.4.5*

4.2.1.1.1*

4.2.1.1.22*

4.2.1.1.32*

4.2.1.1.42*

4.2.1.2.1*

4.2.1.2.22*

4.2.1.2.32*

4.2.1.2.42*

* Menu numbers

8.5.

Installation (Main Menu 5)

Sensors

5.1*

Miscellaneous

5.1.1*

Flow

Offset

Quality Assurance Level

5.1.2*

Signal Outputs Signal Output 1/2

5.2* 5.2.1* - 5.2.2*

Parameter

Current Loop

Function

Scaling

Relay Contacts

5.3*

Alarm Relay

5.3.1*

5.2.x.40

Alarm oxygen

5.3.1.1*

5.1.1.1*

5.1.1.2*

5.1.2.1*

5.2.1.1 - 5.2.2.1*

5.2.1.2 - 5.2.2.2*

5.2.1.3 - 5.2.2.3*

Range Low

Range High

Alarm High

Alarm Low

48

* Menu numbers

5.2.x.40.10/11*

5.2.x.40.20/21*

5.3.1.1.1*

5.3.1.1.22

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program Overview

Miscellaneous

5.4*

Language

Set defaults

Load Firmware

Password

5.4.4*

Interface

5.5*

Relay 1/2

5.3.2* - 5.3.3*

Input

5.3.4*

Sample ID

Protocol

5.5.1*

Case Temp.

5.3.1.5*

Function

Parameter

Setpoint

Hysteresis

Delay

Active

Signal Outputs

Output/Control

Fault

Delay

5.4.1*

5.4.2*

5.4.3*

Messages

Maintenance

Operation

Installation

5.4.5*

USB Stick

Sample Flow

5.3.1.2*

Sample Temp.

5.3.1.3*

Alarm Saturation

5.3.1.4*

Hysteresis

Delay

Flow Alarm

Alarm High

Alarm Low

Alarm High

Alarm Low

Alarm High

Alarm Low

Hysteresis

Delay

Case Temp. high

Case Temp. low

5.3.2.1–5.3.3.1*

5.3.2.20–5.3.3.20*

5.3.2.300–5.3.3.301*

5.3.2.400–5.3.3.401*

5.3.2.50–5.3.3.50*

5.3.4.1*

5.3.4.2*

5.3.4.3*

5.3.4.4*

5.3.4.5*

5.3.1.1.32

5.3.1.1.42

5.3.1.2.1

5.3.1.2.2*

5.3.1.2.32*

5.3.1.3.1*

5.3.1.3.22*

5.3.1.4.1*

5.3.1.4.22

5.3.1.4.32

5.3.1.4.42

5.3.1.5.1*

5.3.1.5.2*

5.4.4.1*

5.4.4.2*

5.4.4.3*

5.4.4.4*

* Menu numbers

A-96.250.701 / 140716

49

AMI INSPECTOR Oxygen

Program List and Explanations

9.

Program List and Explanations

1 Messages

1.1 Pending Errors

1.1.5

Provides the list of active errors with their status (active, acknowledged). If an active error is acknowledged, the alarm relay is active again. Cleared errors are moved to the Message list.

1.2 Maintenance List

1.2.5

Provides the list of necessary maintenance. Cleared maintenance messages are moved to the Message list.

1.3 Message List

1.3.1

Shows the error history: Error code, date / time of issue and status

(active, acknowledged, cleared). 65 errors are memorized. Then the oldest error is cleared to save the newest error (circular buffer).

2 Diagnostics

In diagnostics mode, the values can only be viewed, not modified.

2.1 Identification

Desig.

: Designation of the instrument.

Version : Firmware of instrument (e.g. V6.00-11/15)

2.1.3

Factory Test : Test date of the Instrument, Motherboard and

Frontend.

2.1.4

Operating Time: Shows the operating time in Years, Days, Hours,

Minutes and Seconds.

2.2 Sensors

2.2.1

Oxytrace G

Current value: Shows the actual measuring value in ppb.

Raw value tc: Shows the actual temperature compensated measuring value in mA.

Saturation Shows the actual saturation in %

50

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

2.2.1.4

Cal. History

Review the diagnostic values of the last calibration of the oxygen sensor. Max. 64 data records are memorized.

Number : Calibration counter.

Date, Time : Date and time of the calibration.

Sat. Current : Saturation current at that time of calibration.

Air pressure : Air pressure at that time of calibration.

2.2.2

Miscellaneous :

2.2.2.1

Case Temp: Shows the actual temperature in °C inside the transmitter.

Air pressure : Shows the actual air pressure in hPa

2.2.3

QA History

Review QA values (Number, Date, Time, Deviation oxygen,

Deviation Temperature, Status of QA check) of the last quality assurance procedures.

2.3 Sample

2.3.301

Sample ID : Shows the assigned sample identification. This identification is defined by the user to identify the location of the sample

Temperature: Shows temperature in °C.

(Nt5K) : Shows raw value of the temperature in 

Sample Flow : Shows the sample flow in l/h

(Raw value) Shows the sample flow in Hz

2.4 I/O State

Shows actual status of all in- and outputs.

2.4.1

Alarm Relay:

Relay 1 and 2:

Input:

Signal Output 3:

Active or inactive

Active or inactive

Open or closed.

Actual current in mA (if option is installed)

2.5 Interface

2.5.1

Protocol USB Stick.

A-96.250.701 / 140716

51

AMI INSPECTOR Oxygen

Program List and Explanations

3 Maintenance

3.1 Calibration

3.1.1

Start a calibration and follow the instructions on the screen.

Displayed values are saturation in % and the saturation current in mA. The indication bar shows the progress. Detailed explanation see Calibration, p. 32 .

3.2 Service

3.2.1

Electrolyte

Last Filling : Shows the date of the last filling of electrolyte.

Remaining Amount : Remaining amount of electrolyte in %.

Remaining Time : Remaining time in days until electrolyte exchange recommended.

3.2.1.5

New Filling : Select “Yes” after electrolyte exchange to reset the counter.

3.3 Simulation

To simulate a value or a relay state, select the

 alarm relay

 relay 1 and 2

 signal output 3 (signal outputs 1 and 2 are deactivated) with the [ ] or [ ] keys.

Press the [Enter]> key.

Change the value with the [ ] or [ ] keys.

Press the [Enter] key.

 The value is simulated by the relay/signal output.

Alarm Relay:

Relay 1:

Relay 2:

Signal Output 3:

Active or inactive

Active or inactive

Active or inactive

Current in mA (if option is installed)

At the absence of any key activities, the instrument will switch back to normal mode after 20 min. If you quit the menu, all simulated values will be reset.

3.4 Set Time

Adjust date and time.

52

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

3.5 Quality Assurance

Performs a Quality Assurance according to your settings. Follow the commands on the screen. Detailed explanation see Quality

Assurance of the Instrument, p. 34 .

4 Operation

4.1 Sensors

4.1.1

Filter Time Constant: Used to damp noisy signals. The higher the filter time constant, the slower the system reacts to changes of the measured value.

Range: 5–300 Sec

4.1.2

Hold after Cal: Delay permitting the instrument to stabilize again after calibration. During calibration plus hold-time, the signal outputs are frozen (held on last valid value), alarm values, limits are not active.

Range: 0–6‘000 Sec

4.2 Relay Contacts

See Relay Contacts, p. 21

4.3 Logger

The instrument is equipped with an internal logger. The data can be copied to the USB stick installed in the transmitter.

The logger can save approx. 1500 data records. The Records consists of: Date, time, alarms, measuring values, raw values, case temperature, flow.

4.3.1

Log Interval: Select a convenient log interval. Consult the table below to estimate the max logging time. When the logging buffer is full, the oldest data record is erased to make room for the newest one (circular buffer).

Range: 1 Second to 1 hour

Interval

Time

1 s 5 s

25 min 2 h

1 min

25 h

5 min

5 d

10 min

10 d

30 min

31 d

1 h

62 d

4.3.2

Clear Logger: If confirmed with yes , the complete logger data is deleted. A new data series is started.

4.3.3

Eject USB Stick

With this function all logger data are copied to the USB stick before the USB stick is deactivated.

A-96.250.701 / 140716

53

AMI INSPECTOR Oxygen

Program List and Explanations

5 Installation

5.1 Sensors

5.1.1

Miscellaneous

5.1.1.1

Flow : If a flow cell without flow measurement (e.g. B-Flow) is used, choose none. With flow measurement select Q-Flow

5.1.1.2

O2 Offset : Manual, small correction of the offset.

Range -5 to +5 ppb.

5.1.2

Quality Assurance: Switch the Quality Assurance on or off.

5.1.2.1

Level : Select quality level:

 Level 0: Off

Quality assurance procedure switched off. Any additional QA menus are hidden.

 Level 1: Trend

 Level 2: Standard

 Level 3: Crucial

 Level 4: User

Edit user specific limits in menu 5.1.2.2

54

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.2 Signal Outputs

5.2.1

Signal Output 3 (signal outputs 1 and 2 are deactivated)

5.2.1.1

Parameter: Assign one of the process values to the signal output.

Available values:

 Oxygen

 Temperature

 Sample Flow (if a flow sensor is selected)

 Saturation

5.2.1.2

Current Loop: Select the current range of the signal output.

Make sure the connected device works with the same current range.

Available ranges: 0–20 mA or 4–20 mA

5.2.1.3

Function: Define if the signal output is used to transmit a process value or to drive a control unit. Available functions are:

 Linear, bilinear or logarithmic for process values.

See As process values, p. 55

 Control upwards or control downwards for controllers.

See As control output, p. 57

As process values

The process value can be represented in 3 ways: linear, bilinear or logarithmic. See graphs below.

[mA]

20

B A

10 12

A

B

0 / 4

0.0

linear

0.1

bilinear

0.2

0.3

0.4

X

0.5

X Measured value

A-96.250.701 / 140716

55

AMI INSPECTOR Oxygen

Program List and Explanations

[mA]

20

10 12

0 / 4

1

0

2 4 6

10

1

100

2

1’000 10’000

3 4

X

X Measured value (logarithmic)

5.2.1.40

Scaling: Enter beginning and end point (Range low & high) of the linear or logarithmic scale. In addition, the midpoint for the bilinear scale.

Parameter: Oxygen.

Range low: 0.00 ppb –20.00 ppm

Range high: 0.00 ppb –20.00 ppm

Parameter: Temperature

Range low: -30 to + 130 °C

Range high: 30 to + 130 °C

Parameter: Sample flow

Range low: 0–50 l/h

Range high: 0–50 l/h

Parameter: Saturation

Range low: 0–200 %

Range high: 0–200 %

56

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

As control output

Signal outputs can be used for driving control units. We distinguish different kinds of controls:

 P-controller: The controller action is proportional to the deviation from the setpoint. The controller is characterized by the

P-Band. In the steady-state, the setpoint will never be reached. The deviation is called steady-state error.

Parameters: setpoint, P-Band

 PI-controller: The combination of a P-controller with an

I-controller will minimize the steady-state error. If the reset time is set to zero, the I-controller is switched off.

Parameters: setpoint, P-Band, reset time.

 PD-controller: The combination of a P-controller with a

D-controller will minimize the response time to a fast change of the process value. If the derivative time is set to zero, the

D-controller is switched off.

Parameters: setpoint, P-Band, derivative time.

 PID-controller: The combination of a P-, an I - and a D-controller allows a proper control of the process.

Parameters: setpoint, P-Band, reset time, derivative time.

Ziegler-Nichols method for the optimization of a PID controller:

Parameters : Setpoint, P-Band, Reset time, Derivative time

Y

B

A

A-96.250.701 / 140716

X a

A

B

X

L

Response to maximum control output

Tangent on the inflection point

Time

Xp

Tn

Tv

= 1.2/a

= 2L

= L/2

The point of intersection of the tangent with the respective axis will result in the parameters a and L.

Consult the manual of the control unit for connecting and programming details. Choose control upwards or downwards.

57

AMI INSPECTOR Oxygen

Program List and Explanations

58

Setpoint : User-defined process value for the selected parameter.

P-Band: Range below (upwards control) or above (downwards control) the set-point, within which the dosing intensity is reduced from 100% to 0% to reach the set-point without overshooting.

5.2.1.43

Control Parameters : if Parameter = Oxygen

5.2.1.43.10

Setpoint:

Range: 0.00 ppb –20.00 ppm

5.2.1.43.20

P-Band:

Range: 0.00 ppb –20.00 ppm

5.2.1.43

Control Parameters : if Parameter = Temperature

5.2.1.43.11

Setpoint:

Range: -30 to + 130 °C

5.2.1.43.21

P-Band:

Range: 0 to + 100 °C

5.2.1.43

Control Parameters : if Parameter = Sample flow

5.2.1.43.12

Setpoint:

Range: 0–50 l/h

5.2.1.43.22

P-Band:

Range: 0–50 l/h

5.2.1.43

Control Parameters : if Parameter = Saturation

5.2.1.43.13

Setpoint:

Range: 0–200%

5.2.1.43.23

P-Band:

Range: 0–200%

5.2.1.43.3

Reset time: The reset time is the time till the step response of a single I-controller will reach the same value as it will be suddenly reached by a P-controller.

Range: 0–9’000 sec

5.2.1.43.4

Derivative time: The derivative time is the time till the ramp response of a single P-controller will reach the same value as it will be suddenly reached by a D-controller.

Range: 0–9’000 sec

5.2.1.43.5

Control timeout: If a controller action (dosing intensity) is constantly over 90% during a defined period of time and the process value does not come closer to the setpoint, the dosing process will be stopped for safety reasons.

Range: 0–720 min

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.3 Relay Contacts

5.3.1

Alarm Relay: The alarm relay is used as cumulative error indicator.

Under normal operating conditions the contact is active.

The contact is inactive at:

 Power loss

 Detection of system faults like defective sensors or electronic parts

 High case temperature

 Process values out of programmed ranges.

Program alarm levels for the following parameters:

 Meas. Value

 Temperature

 Sample Flow (if a flow sensor is selected)

 Case Temperature high

 Case Temperature low

5.3.1.1

Alarm Oxygen

5.3.1.1.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E001, is displayed in the message list.

Range: 0.00 ppb –20.00 ppm

5.3.1.1.25

Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E002 is displayed in the message list.

Range: 0.00 ppb –20.00 ppm

5.3.1.1.35

Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.

Range. 0.00 ppb –20.00 ppm

5.3.1.1.45

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range: 0–28‘800 Sec

A-96.250.701 / 140716

59

AMI INSPECTOR Oxygen

Program List and Explanations

60

5.3.1.2

Sample Flow: Define at which sample flow a flow alarm should be issued.

5.3.1.2.1

Flow Alarm: Program if the alarm relay should be activated if there is a flow alarm. Choose between yes or no. The flow alarm will always be indicated in the display, pending error list, saved in the message list and the logger.

Available values: Yes or no

NOTICE: Sufficient flow is essential for a correct measurement.

We recommend to program yes.

5.3.1.2.2

Alarm High: If the measuring values rises above the programmed value E009 will be issued.

Range: 12–50 l/h

5.3.1.2.32

Alarm Low: If the measuring values falls below the programmed value E010 will be issued.

Range: 8–11 l/h

5.3.1.3

Sample Temp.: Define at which sample temperature an alarm should be issued.

5.3.1.3.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E007 is issued.

Range: 30–100 °C

5.3.1.3.22

Alarm Low: If the measured value rises above the alarm high value, the alarm relay is activated and E008 is issued.

Range: -10 to + 20 °C

5.3.1.4

Alarm Saturation

5.3.1.4.1

Alarm High: If the measured value rises above the alarm high value, the alarm relay is activated and E001, is displayed in the message list.

Range: 0.00 –200 %

5.3.1.4.22

Alarm Low: If the measured value falls below the alarm low value, the alarm relay is activated and E002 is displayed in the message list.

Range: 0.00 –200 %

5.3.1.4.32

Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relays contacts when the measured value fluctuates around the alarm value.

Range. 0.00 –200 %

5.3.1.4.42

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range: 0–28‘800 Sec

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.3.1.5

Case Temp.

5.3.1.5.1

Case Temp. high: Set the alarm high value for temperature of electronics housing. If the value rises above the programmed value

E013 is issued.

Range: 30–75 °C

5.3.1.5.2

Case Temp. low: Set the alarm low value for temperature of electronics housing. If the value falls below the programmed value

E014 is issued.

Range: -10 to + 20 °C

5.3.2 and 5.3.3

Relay 1 and 2: The function of relay contacts 1 or 2 are defined by the user

NOTICE: The navigation in the menu <Relay 1> and <Relay 2> is equal. For reason of simplicity only the menu numbers of

Relay 1 are used in the following.

1 First select the functions as:

- Limit upper/lower,

- Control upwards/downwards,

- Timer

- Fieldbus

2 Then enter the necessary data depending on the selected function. The same values may also be entered in menu 4.2 Relay

Contacts, p. 53

5.3.2.1

Function = Limit upper/lower:

When the relays are used as upper or lower limit switches, program the following:

5.3.2.20

Parameter: choose one of the following process values

 Oxygen

 Temperature

 Sample Flow

 Saturation

5.3.2.300

Setpoint : If the measured value rises above respectively falls below the set-point, the relay is activated.

Parameter

Oxygen

Temperature

Sample flow

Saturation

Range

0.00 ppb –20.00 ppm

-30 to + 130 °C

0–50 l/h

0–200 %

A-96.250.701 / 140716

61

AMI INSPECTOR Oxygen

Program List and Explanations

62

5.3.2.400

Hysteresis: within the hysteresis range, the relay does not switch.

This prevents damage of relay contacts when the measured value fluctuates around the alarm value.

Parameter

Oxygen

Temperature

Sample flow

Saturation

Range

0.00 ppb –20.00 ppm

0–100 °C

0–50 l/h

0–200 %

5.3.2.50

Delay: Duration, the activation of the alarm relay is retarded after the measuring value has risen above/fallen below the programmed alarm.

Range. 0–600 Sec

5.3.2.1

Function = Control upwards/downwards

The relays may be used to drive control units such as solenoid valves, membrane dosing pumps or motor valves. When driving a motor valve both relays are needed, relay 1 to open and relay 2 to close the valve.

5.3.2.22

Parameter: choose one of the following process values

 Oxygen

 Temperature

 Sample Flow

 Saturation

5.3.2.32

Settings

Choose the respective actuator:

 Time proportional

 Frequency

 Motor valve

Actuator = Time proportional

Examples of metering devices that are driven time proportional are solenoid valves, peristaltic pumps.

Dosing is controlled by the operating time.

5.3.2.32.20

Cycle time: duration of one control cycle (on/off change).

Range: 0–600 sec.

5.3.2.32.30

Response time: Minimal time the metering device needs to react.

Range: 0–240 sec.

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.3.2.32.4

Control Parameters:

Range for each Parameter same as 5.2.1.43, p. 58

Actuator = Frequency

Examples of metering devices that are pulse frequency driven are the classic membrane pumps with a potential free triggering input.

Dosing is controlled by the repetition speed of dosing shots.

5.3.2.32.21

Pulse frequency: Max. pulses per minute the device is able to respond to. Range: 20–300/min.

5.3.2.32.31

Control Parameters:

Range for each Parameter same as 5.2.1.43, p. 58

Actuator = Motor valve

Dosing is controlled by the position of a motor driven mixing valve.

5.3.2.32.22

Run time: Time needed to open a completely closed valve

Range: 5–300 Sec.

5.3.2.32.32

Neutral zone: Minimal response time in % of the runtime. If the requested dosing output is smaller than the response time, no change will take place.

Range: 1–20 %

5.3.2.32.4

Control Parameters:

Range for each Parameter same as 5.2.1.43, p. 58

5.3.2.1

Function = Timer

The relay will be activated repetitively depending on the programmed time scheme.

5.3.2.24

Mode: Operating mode (interval, daily, weekly)

5.3.2.24

Interval

5.3.2.340

Interval: The interval can be programmed within a range of 1–1440 min.

5.3.2.44

Run Time : Enter the time the relay stays active.

Range: 5–32’400 Sec.

5.3.2.54

Delay : during run time plus the delay time the signal and control outputs are held in the operating mode programmed below.

Range: 0–6’000 Sec.

A-96.250.701 / 140716

63

AMI INSPECTOR Oxygen

Program List and Explanations

5.3.2.6

Signal Outputs : Select operating mode of the signal output:

Cont.: Signal outputs continue to issue the measured value.

Hold: Signal outputs hold the last valid measured value.

Measurement is interrupted. Errors, except fatal errors, are not issued.

Off: Signal outputs are switched off (set to 0 or 4 mA).

Errors, except fatal errors, are not issued.

5.3.2.7

Output/Control : Select operating mode of the controller output:

Cont.: Controller continues normally.

Hold: Controller continues based on the last valid value.

Off: Controller is switched off.

5.3.2.24

daily

The relay contact can be activated daily, at any time of a day.

5.3.2.341

Start time : to set the start time proceed as follows:

1 Press [Enter], to set the hours.

2 Set the hour with the [ ] or [

3 Press [Enter], to set the minutes.

] keys.

4 Set the minutes with the [ ] or [

5 Press [Enter], to set the seconds.

6 Set the seconds with the [ ] or [

] keys.

] keys.

Range: 00:00:00–23:59:59

5.3.2.44

Run Time : see Interval

5.3.2.54

Delay : see Interval

5.3.2.6

Signal Outputs : see Interval

5.3.2.7

Output/Control : see Interval

5.3.2.24

weekly

The relay contact can be activated at one or several days, of a week. The daily starting time is valid for all days.

64

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.3.2.342

Calendar :

5.3.2.342.1

Start time : The programmed start time is valid for each of the programmed days. To set the start time see 5.3.2.341, p. 64 .

Range: 00:00:00–23:59:59

5.3.2.342.2

Monday : Possible settings, on or off to

5.3.2.342.8

Sunday : Possible settings, on or off

5.3.2.44

Run Time : see Interval

5.3.2.54

Delay : see Interval

5.3.2.6

Signal Outputs : see Interval

5.3.2.7

Output/Control : see Interval

5.3.2.1

Function = Fieldbus

The relay will be switched via the Profibus input. No further parameters are needed.

5.3.4

Input: The functions of the relays and signal outputs can be defined depending on the position of the input contact, i.e. no function, closed or open.

5.3.4.1

Active: Define when the input should be active:

No: Input is never active.

When closed Input is active if the input relay is closed

When open: Input is active if the input relay is open

5.3.4.2

Signal Outputs: Select the operation mode of the signal outputs when the relay is active:

Cont.:

Hold:

Off:

Signal outputs continue to issue the measured value.

Signal outputs issue the last valid measured value.

Measurement is interrupted. Errors, except fatal errors, are not issued.

Set to 0 or 4 mA respectively. Errors, except fatal errors, are not issued.

A-96.250.701 / 140716

65

AMI INSPECTOR Oxygen

Program List and Explanations

5.3.4.3

Output/Control: (relay or signal output):

Cont.:

Hold:

Off:

Controller continues normally.

Controller continues on the last valid value.

Controller is switched off.

5.3.4.4

Fault:

No:

Yes:

No message is issued in pending error list and the alarm relay does not close when input is active.

Message E024 is issued and stored in the message list. The Alarm relay closes when input is active.

5.3.4.5

Delay: Time which the instrument waits, after the input is deactivated, before returning to normal operation.

Range: 0–6‘000 Sec

66

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Program List and Explanations

5.4 Miscellaneous

5.4.1

Language: Set the desired language.

Available settings: German /English/French/Spanish

5.4.2

Set defaults: Reset the instrument to factory default values in three different ways:

 Calibration: Sets calibration values back to default. All other values are kept in memory.

 In parts: Communication parameters are kept in memory. All other values are set back to default values.

 Completely: Sets back all values including communication parameters.

5.4.3

Load Firmware: Firmware updates should be done by instructed service personnel only.

5.4.4

Password: Select a password different from 0000 to prevent unauthorized access to the menus “Messages”, “Maintenance”,

“Operation” and “Installation”.

Each menu may be protected by a different password.

If you forgot the passwords, contact the closest SWAN representative.

5.4.5

Sample ID: Identify the process value with any meaningful text, such as KKS number.

A-96.250.701 / 140716

67

AMI INSPECTOR Oxygen

Program List and Explanations

5.5 Interface

Select one of the following communication protocols. Depending on your selection, different parameters must be defined.

5.5.1

5.5.20

5.5.30

5.5.40

5.5.1

5.5.21

5.5.31

5.5.41

5.5.1

Protocol: Profibus

Device address:

ID-Nr.:

Local operation:

Range: 0–126

Range: Analyzer; Manufacturer; Multivariable

Range: Enabled, Disabled

Protocol: Modbus RTU

Device address: Range: 0–126

Baud Rate: Range: 1 200–115 200 Baud

Parity: Range: none, even, odd

Protocol: USB-Stick:

Only visible if an USB interface is installed. No further settings are possible.

68

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Default Values

10. Default Values

Operation:

Sensors: Filter Time Const.: .................................................................10 Sec

Hold after Cal.:.....................................................................300 Sec

Alarm Relay ...................................................................... same as in Installation

Relay 1/2 ...................................................................... same as in Installation

Input ...................................................................... same as in Installation

Logger: Logger Interval:.............................................................. 30 Minutes

Clear Logger:................................................................................ no

Installation:

Sensors Miscellaneous; Flow: ............................................................... None

Offset: ................................................................................... 0.0 ppb

Quality Assurance; Level: ........................................................ 0: Off

Signal Output Parameter:............................................................................ oxygen

Current loop:..................................................................... 4 –20 mA

Function:.................................................................................. linear

Scaling: Range low:............................................................ 0.00 ppb

Scaling: Range high: ....................................................... 10.00 ppm

Alarm Relay: Alarm oxygen; Alarm high:............................................... 10.00 ppm

Alarm oxygen; Alarm low: ................................................... 0.00 ppb

Alarm oxygen; Hysteresis:................................................... 100 ppb

Alarm oxygen; Delay: ..............................................................5 Sec

If Flow = Q-Flow

Sample Flow, Flow Alarm: .......................................................... yes

Sample Flow, Alarm high: .....................................................25.0 l/h

Sample Flow, Alarm low: ........................................................8.0 l/h

Sample Temp., Alarm High: ..................................................... 50 °C

Sample Temp., Alarm Low:........................................................ 0 °C

Alarm Saturation; Alarm high..................................................120 %

Alarm Saturation; Alarm low ....................................................0.0 %

Alarm Saturation; Hysteresis ......................................................2 %

Alarm Saturation; Delay...........................................................5 Sec

Case temp. high: ..................................................................... 65 °C

Case temp. low:......................................................................... 0 °C

Relay 1 Function:.......................................................................... limit upper

Parameter:............................................................................ oxygen

Setpoint: .......................................................................... 10.00 ppm

Hysteresis:........................................................................... 100 ppb

Delay: ....................................................................................30 Sec

A-96.250.701 / 140716

69

AMI INSPECTOR Oxygen

Default Values

Relay 2 Function: .......................................................................... limit upper

Parameter: ....................................................................Temperature

Setpoint:...................................................................................50 °C

Hysteresis: ..............................................................................1.0 °C

Delay:.....................................................................................30 Sec

If Function = Control upw. or dnw:

Parameter: .................................................................... Meas. Value

Settings: Actuator: ...........................................................Frequency

Settings: Pulse Frequency: ............................................ 120/min.

Settings: Control Parameters: Setpoint: .......................10.00 ppm

Settings: Control Parameters: P-band:.............................100 ppb

Settings: Control Parameters: Reset time: ...........................0 Sec

Settings: Control Parameters: Derivative Time: ...................0 Sec

Settings: Control Parameters: Control Timeout:.................. 0 Min.

Settings: Act. Time prop.: Cycle time: .................................... 60 s

Settings: Act. Time prop.: Response time: ............................ 10 s

Settings: Act. Motor valve: Run time: .................................... 60 s

Settings: Act. Motor valve: Neutral zone: ................................5%

If Function = Timer:

Mode:.................................................................................... Interval

Interval:................................................................................. 1 min

Mode: ....................................................................................... daily

Start time: ........................................................................00.00.00

Mode:.....................................................................................weekly

Calendar; Start time: .......................................................00.00.00

Calendar; Monday to Sunday:.................................................. Off

Run time: ...............................................................................10 Sec

Delay:.......................................................................................5 Sec

Signal output:............................................................................. cont

Output/Control: .......................................................................... cont

Input: Active ............................................................................when closed

Signal Outputs ........................................................................... hold

Output/Control .............................................................................. off

Fault..............................................................................................no

Delay......................................................................................10 Sec

70

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Default Values

Miscellaneous Language:............................................................................. English

Set default: ................................................................................... no

Load firmware:.............................................................................. no

Password:........................................................... for all modes 0000

Sample ID:....................................................................... - - - - - - - -

Interface Protocol: ...........................................................................USB Stick

A-96.250.701 / 140716

71

AMI INSPECTOR Oxygen

Index

11.

Index

A

Alarm Relay

. . . . . . . . . .

21

C

Cable thicknesses

Calendar

. . . . . .

17

. . . . . . . . . . . .

65

Changing values

Charging

. . . . . . .

27

. . . . . . . . . . . .

19

Clean Flow Cell and Swansensor Oxytrace G

. . . . . . . . . . . . .

31

D

Default Values

. . . . . . . .

69

E

Electrolyte exchange external devices

. . . . .

29

Establish Sample Flow

. . .

23

. . . . . . .

19

F

Fluidics

. . . . . . . . . . . . .

9

I

Input

. . . . . . . . . . . . . .

21

Instrument Setup

. . . .

13

,

38

L

Longer Stop of Operation

. .

41

P

Power ON - OFF

Power Supply

. . . . . . .

19

. . . . . .

11

,

21

R

Relay Contacts 1 and 2

. . .

22

S

Sample requirements shut-down

. . . .

11

. . . . . . . . . . .

19

Signal Outputs

Software

. . . . . . . .

22

. . . . . . . . . . . .

26

T

Temperature compensation

Terminals

. .

8

. . . . . .

18

,

21

22

W

Wire

. . . . . . . . . . . . . .

17

Z

Zero-Verification

. . . . . . .

34

72

A-96.250.701 / 140716

AMI INSPECTOR Oxygen

Notes

12. Notes

A-96.250.701 / 140716

73

AMI INSPECTOR Oxygen

SWAN

is represented worldwide by subsidiary companies and distributors.

cooperates with independent representatives all over the world.

SWAN Products

Analytical Instruments for:

High Purity Water

Feedwater, Steam and Condensate

Potable Water

Pool and Sanitary Water

Cooling Water

Waste Water and Effluents

Made in Switzerland

74

A-96.250.701 / 140716

advertisement

Related manuals

Download PDF

advertisement