advertisement
ENMET
680 Fairfield Court
Ann Arbor, MI 48108
734.761.1270 Fax 734.761.3220 www.enmet.com
AM-5150
Ambient Air Monitor
Operation and Maintenance Manual
AM-5150 ENMET
Table of Contents
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 1
AM-5150 ENMET
List of Figures
List of Tables
Reference Information:
N OTE : [important information about use of instrument]
C AUTION : [affects equipment – if not followed may cause damage to instrument, sensor etc.…]
W ARNING : [affects personnel safety – if not followed may cause bodily injury or death.]
! Attention / Warning
Earth Ground
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 2
AM-5150
1.0 Introduction
ENMET
The AM-5150 is an ambient air monitoring instrument that measures and detects gases utilizing a non-specific MOS sensor. The AM-
5150 is NOT in an enclosure rated for use in a Class I, Div. 1, Groups B, C, D classified area and cannot be installed in a hazardous location. However, the remote sensor housing and sensor are rated for Class I, Div. 1, Groups A, B, C & D hazardous locations and may be installed in such locations with appropriate wiring.
Features of the AM-5150:
continuous monitoring of the ambient air and continuous LCD display of gas and vapor concentrations
menu driven operational and maintenance controls
menu driven calibration procedure
audio and visual alarms indicate unsafe conditions
alarm relay contacts available on terminals
a fault relay and visual fault alarm
alarm acknowledgement capability including audio defeat
mA outputs for target gas
N OTE : All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the AM-5150 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N OTE : It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor any claims for damage.
•
This shipment was carefully inspected, verified and properly packaged at ENMET and delivered to the carrier in good condition.
•
When it was picked up by the carrier at ENMET , it legally became your company’s property.
•
If your shipment arrives damaged: o Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and request immediate inspection of the carton and the contents. o After the inspection and after you have received written acknowledgment of the damage from the carrier, contact ENMET
Customer Service for return authorization and further instructions. Please have your Purchase Order and Sales Order numbers available.
•
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually
$100.00. Repair or replacement charges above that value are your company’s responsibility.
•
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to
ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
ENMET
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 Fax 734-761-3220
Toll Free: 800-521-2978
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the AM-5150 is received as ordered. Each AM-5150 is labeled with its target gas. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each AM-5150 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 3
AM-5150
2.0 Components of the AM-5150
2.1 AM-5150 elements
See Figure 1 for location of elements:
Feature Description
ENMET
Enclosure
Front Cover
A polycarbonate box, approximately 7 x 5 x 3, with a detachable front cover.
4 holes for mounting the enclosure to a vertical surface. Located at the corners of the bottom of the enclosure, directly beneath the 4-front cover retaining screws. See Figure 3
Detachable front cover of AM-5150 with Display Panel. See Section 2.2 and Figure 1
There are 4 Screws that hold the front cover in place.
2.2 AM-5150 Operational Features
The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the panel, it is swung upward, exposing the interior of the enclosure. See Figure 1 for location of features.
Feature Description
Display
Audio Alarm(Horn)
Visual:
Indicators and Alarms
Membrane Switches
A single line, 8-character LCD with backlight. Indicates the level of gas detected by sensor.
The numerical value of gas concentration and other information is displayed.
Audio alarm (105 dB at 30cm/12in). The audio alarm is activated when the unit is in alarm.
LED indicators:
Power / Fault Indicator LED, Green / Red
Alarm (3) Indicator LED, Red
2 Pushbutton Switches on front panel control the instrument maintenance functions. The pushbutton switch locations are indicated by:
M ENU
↓
: Advances the instrument display through operation information and maintenance menus
S ELECT
: Disables audio alarm temporally and
Selects the maintenance menu operations such as, Zero, Span, Exit menu or sets proper calibration values for Zero or Span
See Section 4.0 and 5.0
for operational and maintenance flow charts.
Three alarm points are preprogrammed into the AM-5150 . At each alarm point, an LED on the front panel is activated. There are
4, 10 Amp relay contacts at each alarm point, plus a fault relay. See Section 3.2 for wiring information.
Visual Indicator
Power/Fault
Visual Indicators:
Alarm1, Alarm2, Alarm3
Sensor/Remote Sensor
Wiring Access
Menu Select
Pushbutton/Membrane
Switches
Audio Alarm
Front Cover
Retaining Screws
4 places
Power Wiring
Strain Relief
Figure 1: External AM-5150 Features
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 4
AM-5150 ENMET
2.3 Circuit Board Features
The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the panel, it is swung upward, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the enclosure interior. Features are shown in Figure 2 .
Feature Description
Relay Terminals:
J14, J15, J16, J17
Terminal J12
Sensor Terminal J8
Data Terminal J19
Potentiometer 4
Potentiometer 2 & 3
This group of terminals is located on the Circuit Board.
For the contacts for each of three alarm relays, and for the contacts of a fault relay.
See Section 3.3
For V DC back-up power in and the 4-20 mA output. See Section 3.1.2
Sensor connection, See Section 3.2
RS-485 input/output
Sensor heater voltage adjustment, See Section 3.2.1
Not used in AM-5150 Do Not Adjust
RS-485 Input/ Output
Terminal J19
POT 2
Relay Terminals
J14, J15, J16, J17
Relays
K1, K2, K3, K4
Terminal J12
DC Power In
4-20mA Output
Sensor Connection
Terminal J8
POT 3
Sensor Heater Voltage
Adjustment POT 4
Figure 2: AM-5150 Circuit Board Features
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 5
AM-5150
3.0 Installation of the AM-5150
ENMET
The AM-5150 is supplied with a strain relief for a power line cord. Use this fitting or connect a conduit fitting when supplying power to the unit.
N OTE : This control panel is N OT rated for hazardous locations. The control panel must be in a Non-Hazardous area.
3.1 Mounting AM-5150
Mount the AM-5150 instrument on an appropriate vertical surface, leaving room for lid to be opened, using the mounting holes provided. Avoid areas with excessive vibration or temperature extremes. The holes in the bottom of the enclosure are 0.18 inch in diameter and form a 6.44
″
x 4.47
″
rectangle. See Figure 3
It is recommended to use #8 drywall anchors and screws for mounting the AM-5150 to a drywall/sheetrock surface.
Right Side View
Figure 3: Mounting AM-5150
3.1.1 Wiring the AM-5150
The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United
States.
W ARNING : The compliance of the installation to appropriate codes is not ENMET’s responsibility.
The AM-5150 should be powered through circuit breakers provided for this purpose.
3.1.2 Power Supply
The input power can vary from 100 to 240 V AC , 50/60 Hz. Power should be connected to the Power Input Terminal TB1 and the Ground screw . Se e Figure 4 for location.
For DC wiring 24V DC may be wired to J12, (J12-1) position 1 + with ground connected to (J12-2) position 2.
Upon supplying power to the AM-5150 :
The green power on LED is lit.
The display backlight is lit, and instrument will step through a start-up sequence: unit serial number and software revision may be shown on the display.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the S ELECT button. If alarm persists longer than 30 minutes, call ENMET customer service personnel.
W ARNING : Continuous gas detection and alarm systems (110V AC /220V AC / 24V DC /12V DC powered) become inoperative upon loss of
primary power. Contact factory for specifications and pricing of backup battery systems.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 6
AM-5150
AC Power Supply Terminal: TB1
Label on PCB
110V AC TB1 ACN
TB1 ACL
Ground
Screw
220V AC TB1 ACN
Optional TB1 ACL
Ground
Screw
C AUTION : 110/220 V AC when applied
DC Power Supply/4-20mA
Terminal: J12
Function
Neutral
Line
AC GND
Neutral
Line
AC GND
Position
1 +
2
3
Function
24 V DC power
GND
4 - 20 mA out
Sensor Terminal: J8
Position Function
1
2
3
Heater Orange
Color
Signal Blue
Ground 2 wires, Brown &
Yellow
Cover Inside View
Opened Upward
ENMET
Sensor Heater Voltage Label
Sensor Heater Voltage
Adjustment POT 4
Figure 4: Power Terminal Connections AM-5150
3.2 Sensor Location
Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some examples are given below.
Heavier than Air Gas Sensor Location
Bottled LP (liquefied petroleum)
Propane
Butane
Gasoline
Trichloroethylene
Vaporized hydrocarbons
Hydrogen sulfide
Lighter than Air Gas
Natural gas (methane)
Ammonia
Hydrogen
Same Density as Air Gas
Carbon Monoxide
Interior wall; 18-24" from floor.
D O N OT locate directly above or beside gas appliances (ovens, heaters).
Avoid locating anywhere near a vent or window or near an outside doorway.
Sensor Location
Near ceiling.
D O N OT locate directly above appliances where it is subject to direct exposure to heat or steam.
Sensor Location
4-6 feet above the (generally uniform) floor.
D O N OT locate in direct air currents of windows, doors, or vents.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 7
AM-5150 ENMET
3.2.1 Sensor Hook-up
The MOS sensor is connected to the AM-5150 control unit with four-conductor wiring, use the correct oil tight fitting. Two conductors supply heater voltage and heater ground for the sensor. The third and fourth conductors are signal and signal ground wires. Size of the wire depends on the distance between the sensor and the control unit.
Sensor Type
High Voltage:
812, 813, 826, etc.
Low Voltage: 109
Recommended Wire Gauge
Distance from Sensor to Control Unit
≤
500 feet
501 – 800 feet
Longer Distances
≤
50 feet
Longer Distances
Recommended Wire Gauge
16 AWG
14 AWG
Contact ENMET Corp
16 AWG
Contact ENMET Corp
Optional
Splash Shield
Wiring for 4-wire Terminal Sensors
Position
1
2
3
4
Function
Heater
Signal
Heater Ground
Signal Ground
Wire Color
Orange
Blue
Brown
Yellow
Optional
Splash Shield
Wiring for 3-wire Terminal Sensors
Position Function
1
2
Heater
Signal
3 Heater Ground
Wire Nut Signal Ground
Wire Color
Orange
Blue
Brown
Yellow
N OTE : The yellow signal wire is not connected to the terminal block, use supplied wire nut.
Sensor Wiring Terminal*
The 4 th
(yellow) wire is not connected to the terminal block, use supplied wire nut
Figure 5: Internal View of Sensor Wiring
Wire length between the AM-5150 and the sensor greater then 100ft will require that the sensor heater voltage be reset. After you mount and install the AM-5150 and Sensor, you must verify the sensor heater voltage. Use position 2 and 3 to measure heater voltage.
Locate the sensor heater voltage table label inside the instrument, see Figure 4 . Measure the sensor heater voltage at the sensor, see Figure 5 and adjust the heater adjustment POT 4 until required voltage is reached, see Figure 4 .
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 8
AM-5150 ENMET
3.3 Relay Contacts
Relay contacts are available for each alarm; these are SPDT, rated at 10Amp at 110V AC , and may be latching or non-latching as required by the application.
They are accessed on the terminals next to each relay see Figure 6 . The contact positions are noted on the circuit board next to each terminal.
The following table is for the relays in their un-energized state. This is also the alarm condition state. Non-failsafe configured relays in the alarm state, are the reverse of the PC board labeling. Note that the Fault(FLT) relay cannot be set to operate in a Non-Failsafe mode. Please see Table 1 below:
Table 1: Relay Failsafe Settings
Alarm
Alarm 1
Alarm 2
Alarm 3
Fault Alarm
Position
J14 (K1) Relay 1 - NO Normally Open
J14 (K1) Relay 1 - NC Normally Closed
J14 (K1) Relay 1 - COM Common
J15 (K2) Relay 2 - NO Normally Open
J15 (K2) Relay 2 - NC Normally Closed
J15 (K2) Relay 2 - COM Common
J16 (K3) Relay 3 - NO Normally Open
J16 (K3) Relay 3 - NC Normally Closed
J16 (K3) Relay 3 - COM Common
J17 (K4) Relay 4 - NO Normally Open
J17 (K4) Relay 4 - NC Normally Closed
J17 (K4) Relay 4 - COM Common
These relay contacts can be used to operate auxiliary alarms or other functions. The relay contacts are DRY, power must be supplied. It is recommended that power for auxiliary equipment be supplied from an independent power source separate from the
AM-5150 . Use the existing hole in the enclosure for wire to enter and exit and use appropriate cable fittings. Wiring should be grouped together, V AC wires should be separated for V DC wires.
Relay Terminals
J14, J15, J16, J17
Relays
K1, K2, K3, K4
Sensor Terminal J8
Figure 6: Relay Terminal Connections AM-5150
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 9
AM-5150
4.0 Operation
ENMET
When the AM-5150 is installed as described in Section 3 , and in clean air, the POWER green LED is on, the display is lit and the information on the display is measurement of the target detected by the AM-5150 .
The red alarm and fault LEDs are not lit.
4.1 Start Up AM-5150
When the AM-5150 is first powered up, it goes through a series of momentary screens, which identify the instrument model number, serial number and software revision. After all the momentary screens have been displayed, the instrument arrives at the
Main Gas Display showing the gas concentration and unit of measurement, ppm or %LEL. Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter indicating the instrument status.
See the Section 4.1.1 below
4.1.1 Typical Start Up
When power is supplied to the AM-5150, the instrument will display the following sequence of information:
Typical start up sequence of information displayed.
Example of Typical Start Up Display Function
AM-5150
The instrument: Model AM-5150
Example for reference only The instrument: Serial Number
311- 20
Example for reference only
S/W 6.6F
The instrument: Software Revision
IF the right most character is a flashing W
0ppW
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
IF the right most character is a flashing C
0ppm
OR
0ppC
0LEL
The instrument has failed Calibration
The last good calibration values are retained, but the sensor may not be responsive to gas
A new Calibration should be performed As Soon As Possible
The instrument: Normal Display Mode
Measurement of the target gas
0ppm
0ppm
The instrument: in Purge Mode Display
0ppP
Optional feature , not required in all instruments
The instrument: in Recovery Mode Display
0ppR
Optional feature , associated with purge mode
N OTE : Software revision may cause variations of display output.
Prolonged storage of the AM-5150-MOS may result in some contamination of the sensor. Following start up, if the display reads above 000 a sensor purge function should be performed.
Sensor Purge: A sensor purge is performed to clean the MOS sensor (Metal Oxide Sensor) of molecules that may have settled on the sensor.
N OTE : Not all instruments are equipped with or require the Purge feature.
To determine if the purge feature is operational, press and hold the S ELECT switch, 3 to 5 seconds, if purge feature is operational 0ppP appears in the display. If the AM-5150-MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault LED will flash. In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes. When the purge cycle is complete, the display will become stable at 0ppm and the green
Power/Fault LED will stop flashing.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 10
AM-5150 ENMET
4.2 Normal Display Mode
When the AM-5150 is installed as described in section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display in measurement units of ppm or %LEL detected by the AM-5150 .
The red alarm and fault LEDs are not lit.
To advance through displays of operational information press the M ENU button.
N OTE : Software revision may cause variations of display output.
See sequence of operational information below:
Display Measurement of the target gas
Press M ENU button 0ppm S ELECT
M
ENU
Display indicates Alarm 1 Set point
Press M ENU button
Display indicates Alarm 2 Set point
Press M ENU button
Display indicates Alarm 3 Set point
Press M ENU button
A1:
M
ENU
05
A2:
M ENU
10
A3:
M ENU
20
S ELECT
S ELECT
S ELECT
Press S ELECT button
To temporally disable audio alarm, see Section 4.2.1
Display indicates mA Span range
(Full Scale)
Press M ENU button
Display returns to gas measurent mA: 50 S ELECT
M ENU
Operatio nal Display Flo w Chart
4.2.1 Alarm Conditions AM-5150
There are three alarm set points available. These alarm points are normally set at established safety levels, such as the OSHA
Permissible Exposure Limit (PEL) for toxic gases or recognized standards below the Lower Explosive Limit for combustible gases.
The alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.
If the gas concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.
Pressing the S ELECT button can temporally disable the Audio Alarm. The horn will be disabled for about five minutes. If a second alarm condition occurs during this time the horn will re-activate. If the alarm condition(s) have ended during this time the horn will not re-activate.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 11
AM-5150
5.0 Maintenance
ENMET
The AM-5150 maintenance menus that are accessed by pressing the M ENU button and S ELECT button as described in the maintenance menu section.
5.1 Maintenance Menus
Pushbutton switches control the M ENU and S ELECT functions. The M ENU and S ELECT button locations are indicated on the display panel, see Figure 1.
The M ENU button is used to display the various menu options and make incremental changes to numbers such as alarm points, calibrations gas, etc. The S ELECT button is used to select that option, set zero or span digit.
To enter the maintenance menu, press and hold the M ENU button for 2 to 4 seconds
Table 2 indicates the maintenance menu sequence see Figure 7 for a detailed maintenance menu flow chart.
C AUTION : Do Not Attempt a Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the instrument is forced into a calibration fault mode.
N OTE : Software revision may cause variations of display output.
Table 2: AM-5150 Maintenance Menu Sequence
Example of Display Function
Normal Display Mode
5ppm
Measurement of CO
Press and hold the M ENU button for 2 – 4 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the AM-5150 is in Maintenance Mode
Exit
To exit the maintenance Menu and return to the
Normal Display Mode:
If intended function Press S ELECT button
Press the M ENU button to advance to the Zero procedure
Zero
For adjusting Zero:
If intended function Press S ELECT button
Press the M ENU button to advance to the Span procedure
Span
For adjusting the Span:
If intended function Press S ELECT button
Press the M ENU button to advance to each Alarm set point procedures
Alarm1
Alarm2
For adjusting the Alarm 1, 2 and 3 set points:
If Intended function, press S ELECT button.
Alarm3
Press the M ENU button to advance the mA Span set point procedure mA Span
For adjusting the mA Span set point:
If intended function Press S ELECT button
Pressing the M ENU button without pressing the S ELECT button will allow you to cycle through the menu options.
You must Press the S ELECT button to initiate the desired operation.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 12
AM-5150
N OTE : Software revision may cause variations of display output.
Normal Gas Display
5ppm
M ENU
Press and H
OLD
the M
ENU
button for 2 – 4 seconds to enter the Maintenance Menus
Exit
S ELEC T Press the S ELECT button to return to the Normal Gas Display. See Section 5.2.1
Zero
M ENU
Press the M
ENU
button to cycle through Maintenance
S ELECT
PV: 0
S ELECT
Press the S
ELECT
button to initiate Zero
Span
M ENU
M ENU
OR
S ELECT
Cal OK
Bad ZERO
PV: 0
20
If the Zero signal is within Preset Specs the AM-5150 will display
Cal OK, See Section 5.2.2
If the Zero signal is not within Preset Specs the AM-5150 will display Bad ZERO
S ELECT
You can Press and H
OLD
the S
ELECT
button to change the Calibration Gas Level
See Section 5.2.3
S ELECT
Apply Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 8 Cal OK
OR
Same mV
OR
Bad Sens
Press S
ELECT
to enter the cal signal:
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will indicate Bad Sensor
Alarm1
S ELECT
5
M ENU
Alarm2
M ENU
Alarm3
S ELECT
S ELECT
5
10
10
20
To change Alarm set points:
Press Menu button until Alarm to be changed is displayed
Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S ELECT button locks underscored digit and moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to next menu
See Section 5.2.4
M ENU
ENMET
mA Span
M ENU
S ELECT
0050
To change mA Span set point:
Press Menu button until mA Span is displayed
Press Select button to display the set point
The M ENU button changes digit indicated by underscore cursor
The S ELECT button locks underscored digit and moves to next digit
See Section 5.2.5
To return to Normal Gas Display:
Press M
ENU
button until EXIT is displayed
Then press S
ELECT
button
Figure 7: AM-5150 Maintenance Menu Flow Chart
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 13
AM-5150 ENMET
5.2 Calibration of the AM-5150
Sensor Purge: Prior to calibration some MOS (Metal Oxide Sensor) sensor should be purged. A sensor purge is performed to clean the MOS sensor of molecules that may have settled on the sensor.
Press and hold the S ELECT switch, 3 to 5 seconds, if the purge feature is operational 0ppP appears in the display. If the AM-5150-
MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault LED will flash. Example of purge display
0ppm 0ppP
In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes.
Example of purge recovery display
0ppm 0ppR
When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing.
Calibration is the process of setting the instrument up to read accurately when exposed to the target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait at least 3 – 4 hours after initially supplying power to the AM-5150 instrument before initial calibration, overnight is preferred. The AM-5150 has been pre-calibrated at the factory, and initial field calibration should result in only finetuning to circuit, as well to check that installation is successful. It is not necessary to open the enclosure to make adjustment. The calibration functions are operated with pushbuttons from outside the enclosure through the MENU and SELECT switches.
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure.
ENMET recommends at least quarterly calibration of the AM-5150 instrument .
Calibration equipment is available from ENMET to calibrate the AM-5150 instrument.
•
Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a calibration cup to connect to the sensor of the AM-5150 on the other.
•
Gas cylinder, zero gas 20.9% oxygen or Span gas, factory determined, varies by intent.
Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.
N OTE : Software revision may cause variations of display output.
Menu Select
To Cal Cup
To Regulator
Top View of Humidifier
Calibration Cup
Sensor
Humidifier
Sensor
Enclosure
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
Figure 8: Calibration Adapter
Regulator
Gas Cylinder
P a g e | 14
AM-5150 ENMET
5.2.1 Exit Maintenance Menu
Exit maintenance, by pressing the Exit appears on the display. Press the S ELECT button to return to the instrument Normal Gas
Display.
Example of Exit menu:
Exit
S ELECT
Press the S
ELECT
button to return to the Normal Gas Display.
M ENU
Press the M
ENU
button to cycle through Maintenance Menus
5.2.2 Zero Adjust
The ZERO function must be performed by exposing the AM-5150 instrument to clean fresh air. If the air at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference.
Enter the maintenance menu by pressing and holding M ENU button for 2 to 4 seconds. See Figure 7, AM-5150 Maintenance
Menu flow chart.
After entering the maintenance menu, Press the M ENU button until the Zero menu is displayed.
Press the S ELECT button to perform a Zero.
The display will alternate between Zero and PV: To abort Zero function press and hold M ENU button for 3 – 4 seconds,
Abort? will appear, press S ELECT button to return to Zero.
Press the S ELECT button to initiate a Zero adjustment.
An auto detect sequence is initiated. After 15 seconds, the AM-5150 will monitor the zero reading for stability.
If the reading stabilizes , within the pre-programmed perimeters, an automatic zero adjustment will be made.
Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor press the S ELECT button you are now ready to apply gas. Proceed to gas span step 2
If you wish to Exit the maintenance menu, press M ENU button until Exit is displayed, then press S ELECT button to return to the instrument Normal Gas Display
If the reading does not stabilize, within 255 seconds, the procedure will be aborted. Sensor is outside of safe parameters to be zeroed, the display will read Bad Zero. Repeat Section 5.2.2 Zero Adjust making sure to use a Zero gas of 20.9%
Oxygen. ENMET part number 03296-209.
Example of Zero adjustment display:
S ELECT
Zero
PV:
M ENU
Note: You can Press and H OLD the M ENU button to abort Zero
When “Abort?” appears, press S ELECT button to
S ELECT return to Zero menu
0
Abort?
S ELECT
Press the S
ELECT
button to force Zero adjustment
M ENU
Zero
Cal OK
OR
Bad ZERO
If the Zero signal is within Preset Specs the AM-5150 will display Cal OK momentarily then advance to Span menu
If the Zero signal is not within Preset Specs the AM-5150 will display Bad Zero and return to Zero menu
N OTE : Software revision may cause variations of display output.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 15
AM-5150 ENMET
5.2.3 Gas Span
It is recommended that the Zero Function be performed first.
Do not perform a calibration unless span gas is applied to sensor. Calibration can be aborted by pressing and holding the
M ENU button for 3 – 4 seconds.
Enter the maintenance menu. See Figure 7, AM-5150 Maintenance Menu flow chart.
1. Press the M ENU button until Span display.
2. Press the S ELECT button to perform a Span procedure.
The display will alternate between the calibration gas concentration ( Example : Cal 20) and a signal level (PV).
•
To Abort calibration press and Hold M ENU button for 3 – 4 seconds, “Abort?” will appear, press S ELECT button to return to Span.
N
OTE
:
To abort calibration. Press and H OLD the M ENU button to abort Calibration
M ENU
When “Abort?” appears, press S ELECT button to advance to desired menu
Abort?
S ELEC T
•
To change calibration gas level to be used, press and Hold S ELECT button for 3 – 4 seconds, use menu button to change digit and select button to move to next digit.
N
OTE
:
To change calibration gas level. Press and H OLD the S ELECT button to change the Calibration Gas Level
S ELECT
-Use the M ENU button to change digits
-Use the S ELECT
button to move to next digit
20
3. Attach the associated calibration gas cylinder to the calibration adapter. See Figure 8 calibration adapter.
4. Open the valve to apply the calibration gas to the sensor.
An auto detect sequence is initiated after 30 seconds, the AM-5150 will monitor the cal reading for stability.
5. Watch for the signal level to stabilize. This should take about 1 – 4 minutes.
6. Once the signal level has stabilized,
If the Span is successful, Cal OK appears momentarily, then will advance to Alarm1 menu.
To exit cal, press M ENU button until Exit appears and press S ELECT button
If the sensor is outside of acceptable parameters, Bad Span is displayed.
If the sensor did not respond, an incompatible span gas was applied and the sensor did not respond at all,
Same mV is displayed then will return to Span.
If calibration is not successful, it is suggested that calibration be attempted again in 30-60 minutes.
If the sensor will not calibrate See Section 5.3
for sensor replacement.
7. Remove the calibration gas.
8. Press the M ENU button to advance to Exit menu or desired menu.
N OTE : Software revision may cause variations of display output.
Example of Calibration Display:
Span
S ELECT
PV: 0
S ELECT
M ENU
PV:
Cal OK
Bad Sens
OR
Same mV
0
20
Apply Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 8 Calibration
Adapter
When cal signal is stable AM-5150 will automatically update:
If cal is good display will indicate OK or Same and advance to Alarm1
If cal is not within preset “range” display will indicate
Bad Sensor Same mV the AM-5150 will return to the
Span Menu
To exit press M ENU button until Exit appears and press
S ELECT
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 16
AM-5150 ENMET
5.2.4 Alarm Set Points
The AM-5150 has three alarm set points set at the factory. These alarm points are normally set at established safety levels.
Alarm set points can be changed within limits.
To change any of the three alarm points:
Enter the maintenance menu as shown in Figure 7 AM-5150 Maintenance Menu flow chart.
1.
Press the M ENU button until to display Alarm1 is displayed.
2.
Press the S ELECT button to initiate alarm set point change
3.
Press the M ENU button to change the digit indicated by the underscore cursor
4.
Press the S ELECT button to move the cursor to the next digit
When last digit is entered the AM-5150 will advance to the next menu
5.
Press the M ENU button to advance to the next menu
N OTE : Alarms 2 and 3 cannot be set below the Alarm 1 setting.
Example of Alarm Set Point menus:
Alarm1 S ELECT
5
M ENU
Alarm2
M
ENU
Alarm3
S ELECT
S ELECT
5
10
10
20
To change Alarm set points:
Press Menu button until Alarm to be changed is displayed
Press Select button to display the set point
The M ENU button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
If change is not within range display returns to previous digit
If change is within range display moves to next menu
M ENU
N OTE : Software revision may cause variations of display output.
5.2.5 mA Span Set
The AM-5150 4-20mA span range is set at the factory, normally to the full scale of the measurement and can be changed within limits.
To change the span range:
Enter the maintenance menu as shown in Figure 7 AM-5150 Maintenance Menu flow chart.
1.
Press the M ENU button until to display Span is displayed.
2.
Press the S ELECT button to initiate the mA Span menu
3.
Press the M ENU button to change the digit indicated by the underscore cursor
4.
Press the S ELECT button to move the cursor to the next digit
When last digit is entered the AM-5150 will advance to the next menu
5.
Press the M ENU button to advance to the next menu
Example of mA Span menu:
mA Span
S ELECT
M ENU
0200
To change mA Span set points:
Press Menu button until mA Span is displayed
Press Select button to display the set point
The M ENU button changes digit indicated by underscore cursor
The S ELECT button locks underscored digit and moves to next digit
Default mA Span
4mA 20mA
0 ppm 200 ppm
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 17
AM-5150 ENMET
5.3 Sensor Replacement
The MOS sensor is durable, it can be purged of contaminants by operating in PURGE for a sufficient length of time and at regular intervals.
Gross contamination usually occurs during unavoidable misuse. Close exposure to an open gas flame, submersion of the sensor in a liquid, or continuous exposure to heavy concentrations of industrial vapors will grossly contaminate a sensor. A grossly contaminated sensor causes a continuous alarm.
If a sensor fails to calibrate, replace it.
P ROCEDURE :
W ARNING : Power must be removed from the AM-5150 before this or any internal procedure. Failure to do so may cause damage
to equipment, bodily injury or death.
1.
Obtain a new sensor assembly. Make sure the sensor type is identical to your original sensor
Sensor marking is on side of sensor see Figure 9
( Example : 813-4 the first 3 digits are sensor ID the 4 digit indicates a 4-wire sensor).
2.
Disconnect the instrument for power.
3.
Disconnect the orange, brown, blue and yellow sensor wires.
4.
Unscrew the assembly from the sensor enclosure.
5.
Remove the bad sensor.
6.
Replace with the new sensor and reconnect the wires (See Section 3.2.1).
N OTE : The user must perform the four color-coded wiring attachments when replacing the sensor. If the yellow signal wire is not connected to the terminal block, use supplied wire nut.
7.
Reconnect the instrument to power and verify sensor heater voltage. (See Section 3.2.1)
8.
Recalibrate the instrument (See Section 5.3.1).
Spacer Ring
*Replace facing direction as shown
Optional Splash Shield
Standard Sensor Shield
View of 4-wire Terminal
Sensor
View of 3-wire Terminal
Figure 9: AM-5150 Sensor Replacement
After the new sensor assembly has been installed, it is suggested to allow the sensor to stabilize for at least 3 – 4 hours, preferably overnight.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 18
AM-5150 ENMET
5.3.1 Sensor Replacement Calibration
Sensor Purge: Prior to calibration some MOS (Metal Oxide Sensor) sensor should be purged. A sensor purge is preformed to clean the MOS sensor of molecules that may have settled on the sensor.
Press and hold the S ELECT switch, 3 to 5 seconds, if the purge feature is operational 0ppP appears in the display. If the AM-
5150-MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault
LED will flash. Example of purge display
0ppm 0ppP
In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes.
Example of purge recovery display
0ppm 0ppR
When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing.
Calibration is the process of setting the instrument up to read accurately when exposed to the target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait at least 3 – 4 hours after initially supplying power to the AM-5150 instrument before initial calibration, overnight is preferred. The AM-5150 has been pre-calibrated at the factory, and initial field calibration should result in only fine-tuning to circuit, as well to check that installation is successful. It is not necessary to open the enclosure to make adjustment.
The calibration functions are operated with pushbuttons from outside the enclosure through the MENU and SELECT switches.
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure.
ENMET recommends at least quarterly calibration of the AM-5150 instrument .
Calibration equipment is available from ENMET to calibrate the AM-5150 instrument.
•
Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a calibration cup to connect to the sensor of the AM-5150 on the other.
•
Gas cylinder, zero gas 20.9% oxygen or Span gas, factory determined, varies by intent.
Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.
N OTE : Software revision may cause variations of display output.
After entering the Maintenance menu, press and hold the M ENU button for 2-4 seconds while viewing the Zero menu.
After 2-4 seconds, an F will appear on the far-right hand side of the display. The F indicates that the instrument is in Factory mode.
Perform the calibration Zero and Span procedures as outlined in Section 5.2
. Be sure that the F is present when selecting the
Zero and Span functions.
The Factory calibration sets a calibration window for future standard instrument calibrations.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 19
AM-5150
6.0 Accessory and Replacement Parts
ENMET accessory part numbers:
Description of Part Part Number
ENMET
Sensor Marking, 109-4
Sensor Marking, 812-4
Sensor Marking, 813-4
Sensor Marking, 018-4
Sensor Marking, 019-4
Sensor Marking, 821-4
Sensor Marking, 830-4
Sensor Marking, 832-4
Zero Gas, 20.9% O2, 17 liters
Span Gas
Cylinder of Calibration Gas
03034-109
03034-812
03034-813
03034-018
03034-019
03034-821
03034-830
03034-832
03296-209
Contact ENMET for part number of target gas for each instrument.
See note below.
N OTE : The gases should be identical, or equivalent (correlation gas) to those used to initially calibrate the unit, unless you want to recalibrate to a new gas or different concentration. In addition, calibration gases must be in a background of air; do not use gases with an inert gas background (such as nitrogen or argon).
7.0 Technical Data and Specifications
Electrical Power
Storage and Transport
Operation
15 Amp fused branch circuit
100-240 V AC
0.45A, 50/60 Hz
0.6A, 24V DC
Temperature: preferred
Relative Humidity
Atmospheric Pressure
Temperature:
Relative Humidity
Atmospheric Pressure
-20
°
to +60
°
C (-4
°
to +140
°
F)
0
°
to +20
°
C (32
°
to 68
°
F)
10-99% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
-15
°
to +40
°
C (5
°
to +104
°
F)
10-99% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
Mechanical
Outputs
Dimensions:
Weight:
Material:
Strain relief:
Relays:
Analog:
Digital:
Audio:
7.1 x 5.1 x 3 in(180x130x75mm)
2 lbs (0.9 kg)
Polycarbonate
3-6.5mm OD
SPDT
Resistive Load Inductive Load
10A at 110 V AC 7.5A at 110
V AC
10A at 30 V DC 5A at 30 V DC
4-20mA
RS-485-modbus
105 dB at 30cm/12in
N OTE : All specifications stated in this manual may change without notice.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 20
AM-5150
Notes:
ENMET
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 21
AM-5150
8.0 Terms and Conditions
ENMET
8.1 Ordering Information
Address orders to:
ENMET
Attention: Customer Service Department
680 Fairfield Court
Ann Arbor, MI 48108
Email Orders: [email protected]
Phone: 734-761-1270 Fax: 734-761-3220
You may also contact our customer service department by email [email protected]. MINIMUM ORDER IS $50.00.
8.2 Delivery
Unless Seller otherwise specifies, delivery will be made: FOB Ann Arbor, MI and/or FOB Bowling Green, KY. Title and risk of loss shall pass to Buyer at that point. Shipping and handling charges will be Prepaid and Added to Buyer’s invoice. Buyer may request shipping be charged to their own account with a preferred carrier. Seller shall have the right to choose means of transportation and to route shipment when specific instructions are not included with Buyer’s order. Seller agrees to deliver the goods and services, within the time, in accordance with specifications, at the prices specified on the face hereof. Buyer’s orders to this quotation are not subject to cancellation or deferment of delivery without indemnification of loss to the Seller resulting there from. Seller shall not be liable to Buyer for any loss or damage sustained on account of this delay or nonperformance due to causes beyond Seller’s control and without his fault or negligence. Where performance of the terms here is contingent upon timely delivery of goods or services by the Buyer and such delivery is in default, Seller shall be indemnified for any damage or loss resulting there from, and/or by extension of Seller’s delivery commitment, as applicable.
8.3 Payment Terms
Payment Terms are Net 30 Days from the date of shipment from Seller unless otherwise noted. All shipping and handling costs will be charged to Buyer on a Prepaid and Add basis. Buyer has the option of paying for shipping by charging its own account with a carrier
8.4 Warranty Information and Guidelines
The Seller warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment. The warrant covers both parts and labor excluding calibration and expendable parts such as filters, detector tubes, batteries, etc. If the inspection by the Seller confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States. The Seller shall not be liable for any loss or damage caused by the improper use or installation of the product. The Buyer indemnifies and saves harmless the Seller with respect to any loss or damages that may arise through the use by the Buyer or others of this equipment. This warranty is expressly given in lieu of all other warranties, either expressed, implied or statutory, including that of merchantability, and all other obligations, or liabilities of ENMET, LLC for damages arising out of or in connection with the use or repair or performance of the product. In no event shall ENMET, LLC, be liable for any indirect, incidental, special or consequential damages or for any delay in the performance by ENMET, LLC, which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representatives or other persons to assume for it any obligation or liability other than that which is set forth herein.
Buyer agrees to indemnify and save harmless Seller for any damage or loss from lawsuits against Seller by reason of manufacture of sale of materials, parts, or use of processes resulting from Buyer’s design specifications. Any patent, design, pattern, tool, die, jig, fixture, drawing, test equipment, or process furnished by Seller; whether possessed by the Seller before the date of this quotation, or devised or acquired by Seller during performance of the terms of this quotation, shall remain the property of the
Seller except by specific stipulation on the face hereof. Seller reserves the right, without liability, for damage or loss, to destroy
Buyer’s drawings, specifications, patterns and special tools supplied by Buyer for performance of the terms on the face hereof, unless Buyer gives notice of the disposition of such items.
8.5 Return Policy
All returns for credit must be approved in advance by ENMET, LLC. Such returns are subject to a minimum $50.00 or 20% restocking charge, whichever is greater. Approval of equipment for return is totally at the discretion of ENMET, LLC. All requests for return/exchange must be made no later 30 days of the original shipping date from ENMET. The actual amount of any resulting credit will not be determined prior to a complete inspection of the equipment by ENMET. Calibration gas cylinders cannot be returned or restocked.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 22
AM-5150
9.0 Instructions for Returning an Instrument for Service
ENMET
Contact the ENMET Service Department for all service requests.
Phone: 734-761-1270
Email: [email protected]
Fill out the “Service Request Form” found at the end of this manual and return with your instrument for all needs. Please send your instrument for service to the site in which the product was purchased. A new “Service Request Form” may be requested if the one found in the manual is not available. All instruments should be shipped prepaid to ENMET.
Address for Service:
Michigan Location:
ENMET
Attention: Service Department
680 Fairfield Court
Ann Arbor, MI 48108
Kentucky Location:
ENMET
62 Corporate Court
Bowling Green, KY 42103
Providing the “Service Request Form” assists in the expedient service and return of your unit and failure to provide this information can result in processing delays. ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered after a period of 60 days may, be disposed of or returned unrepaired COD and the customer will be expected to pay the evaluation fee. Serviced instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
NOTE : Warranty of customer installed components.
For Warranty Repairs, please reference ENMET’s “Warranty Information and Guidelines” (found earlier in this section).
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-550
P a g e | 23
Mailing/Shipping Address:
ENMET
680 Fairfield Court
Ann Arbor, MI 48108 [email protected]
Phone: 734.761.1270
Fax: 734.761.3220
Product Name or Number:
Product Serial Number:
Describe Problem or Needed Service:
Service Request Form
Warranty Claim?
☐
Yes
☐
No
Billing Address:
Contact Name:
Email:
PO/Reference
#:
CUSTOMER INFORMATION
Shipping Address:
Phone #:
Fax #:
☐
COD
Card Number
Name as it Appears on
Card:
☐
PAYMENT METHOD
VISA/MasterCard
Exp. Date
☐
American Express
Security Code:
☐
UPS Ground
RETURN SHIPPING METHOD
☐
UPS 3 Day
☐
UPS Next Day
☐
UPS ND Air
Select Air Saver
UPS Account #:
☐
FedEx Ground
☐
FedEx Air
Express Saver
FedEx Account #:
Insure Shipment:
☐
Yes
☐
No
☐
FedEx Air
Overnight Std.
☐
FedEx Air 2
Day
Insurance
Amount:
$
☐
UPS 2 Day Air
☐
FedEx Air
Overnight P-1
ENMET
Service Request Form
Rev.2 – 9/15/2016
advertisement
Related manuals
advertisement
Table of contents
- 4 1.1 Unpack
- 4 1.2 Check Order
- 4 1.3 Serial Numbers
- 5 2.1 AM-5150 elements
- 5 2.2 AM-5150 Operational Features
- 6 2.3 Circuit Board Features
- 7 3.1 Mounting AM
- 7 3.1.1 Wiring the AM
- 7 3.1.2 Power Supply
- 8 3.2 Sensor Location
- 10 3.3 Relay Contacts
- 11 4.1 Start Up AM
- 11 4.1.1 Typical Start Up
- 12 4.2 Normal Display Mode
- 12 4.2.1 Alarm Conditions AM
- 13 5.1 Maintenance Menus
- 15 5.2 Calibration of the AM
- 16 5.2.1 Exit Maintenance Menu
- 16 5.2.2 Zero Adjust
- 17 5.2.3 Gas Span
- 18 5.2.4 Alarm Set Points
- 18 5.2.5 mA Span Set
- 19 5.3 Sensor Replacement
- 20 5.3.1 Sensor Replacement Calibration
- 23 8.1 Ordering Information
- 23 8.2 Delivery
- 23 8.3 Payment Terms
- 23 8.4 Warranty Information and Guidelines
- 23 8.5 Return Policy
- 23 Manual Revision Date – June