advertisement
ENMET
680 Fairfield Court
Ann Arbor, MI 48108
734.761.1270 Fax 734.761.3220 www.enmet.com
GSM-60
Operation Manual
GSM-60 ENMET
Table of Contents
1.0
I NTRODUCTION .................................................................................................................................................................................... 3
1.1 Unpack ............................................................................................................................................................................................. 3
1.2 Check Order ..................................................................................................................................................................................... 3
1.3 Serial Numbers................................................................................................................................................................................. 3
2.0
I NSTRUMENT F EATURES ..................................................................................................................................................................... 4
2.1 Exterior Features .............................................................................................................................................................................. 4
2.2 Display Panel Features ..................................................................................................................................................................... 4
2.3 Circuit Board Features ..................................................................................................................................................................... 6
3.0
I NSTALLATION ..................................................................................................................................................................................... 7
3.1 Mounting of Instrument ................................................................................................................................................................... 7
3.2 Power Supply ................................................................................................................................................................................... 8
3.3 Inputs / Outputs ................................................................................................................................................................................ 8
3.3.1 Sensor/Transmitter Connection ...................................................................................................................................................................... 9
3.3.2 Relay Contacts ................................................................................................................................................................................................ 9
3.3.3 Optional 4-20mA Outputs ............................................................................................................................................................................. 10
3.4 Installation Verification ................................................................................................................................................................. 10
4.0
O PERATION ........................................................................................................................................................................................ 11
4.1 Normal Operation Condition ......................................................................................................................................................... 11
4.2 Alarm Set Points ............................................................................................................................................................................ 11
4.3 Alarm Latching or Differential Settings ........................................................................................................................................ 12
4.4 Audio Defeat .................................................................................................................................................................................. 12
4.5 Display ........................................................................................................................................................................................... 12
4.6 Operational Menu .......................................................................................................................................................................... 13
4.7 Fault Indications ............................................................................................................................................................................. 14
4.7.1 Low Flow Indication ..................................................................................................................................................................................... 14
4.7.2 Other Fault Indications ................................................................................................................................................................................. 14
4.8 Hydrocarbon Sensor Response ...................................................................................................................................................... 14
5.0
M AINTENANCE .................................................................................................................................................................................. 15
5.1 Cleaning Instructions ..................................................................................................................................................................... 15
5.2 Maintenance Menu ......................................................................................................................................................................... 15
5.2.1 Accessing Maintenance Menu ....................................................................................................................................................................... 15
5.2.2 Maintenance Menu Flow Chart .................................................................................................................................................................... 16
5.3 Calibration for CO, O
2,
HC and CO
2
(Gas Channels) ..................................................................................................................... 17
5.3.1A Low Cal/Zero Cal Adjust ............................................................................................................................................................................ 18
5.3.1B High Cal/Span Gas Adjust .......................................................................................................................................................................... 19
5.3.2 Set 4 –20mA Transmitter Scale ..................................................................................................................................................................... 19
5.3.3 Set Alarm Points ........................................................................................................................................................................................... 20
5.3.4 Set Alarm Delay ............................................................................................................................................................................................ 22
5.3.5 Relay Configuration ...................................................................................................................................................................................... 23
5.3.6 Failsafe Configuration .................................................................................................................................................................................. 24
5.3.7 Set Output Span Range ................................................................................................................................................................................. 24
5.3.8 Set New Password ......................................................................................................................................................................................... 25
5.3.9 Exit Maintenance Menu ................................................................................................................................................................................ 25
5.4 Sensor Replacement ....................................................................................................................................................................... 26
5.4.1 Gas/Oxygen Sensor ....................................................................................................................................................................................... 26
5.4.2 Calibration/Sensor Replacement .................................................................................................................................................................. 26
5.4.3A Low Cal/Zero Cal Adjust ............................................................................................................................................................................ 28
5.4.3B High Cal/Span Gas Adjust .......................................................................................................................................................................... 29
5.5 Flow Control Orifice ...................................................................................................................................................................... 29
6.0
T ECHNICAL D ATA AND S PECIFICATIONS ......................................................................................................................................... 29
7.0
R EPLACEMENT P ART N UMBERS ....................................................................................................................................................... 30
7.1 ENMET part numbers for sensors and replacement parts: ............................................................................................................. 30
7.2 ENMET part numbers for Calibration equipment:......................................................................................................................... 30
A PPENDIX A: CO C HARACTERISTICS .................................................................................................................................................... 31
A PPENDIX B: G AS I ONIZATION P OTENTIALS ......................................................................................................................................... 32
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 1
Rev. 1 – December 15, 2016
GSM-60 ENMET
List of Figures
Figure 1: External Features of the GSM-60 ....................................................................................................................... 5
Figure 2: GSM-60 Interior Features ................................................................................................................................... 6
Figure 3: GSM-60 Mounting Dimensions ......................................................................................................................... 7
Figure 2A: Relay, Input and Output Terminals .................................................................................................................. 8
Figure 4: GSM-60 Operational Display ........................................................................................................................... 11
Figure 5: GSM-60 Operation Menu Flow Chart .............................................................................................................. 13
Figure 6: GSM-60 Maintenance Menu Flow Chart. ........................................................................................................ 16
Figure 7: GSM-60 Calibration Connections .................................................................................................................... 18
Figure 9: Location of Gas Sensor and PiD(HC) Manifolds ............................................................................................. 26
Figure 7: GSM-60 Calibration Connections .................................................................................................................... 28
Figure 10: Carbon Monoxide Concentration.................................................................................................................... 31
List of Tables
Table 1 : Relay Failsafe Settings ...................................................................................................................................... 10
Table 2: Sensor Output .................................................................................................................................................... 10
Table 3: Typical Factory Alarm Set Points ...................................................................................................................... 11
Table 4: Factory Set Gas alarms Delay ............................................................................................................................ 22
Reference Information:
N OTE : [important information about use of instrument]
C AUTION : [affects equipment – if not followed may cause damage to instrument, sensor etc.…]
W ARNING : [affects personnel safety – if not followed may cause bodily injury or death.]
Attention / Warning
!
Earth Ground
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 2
Rev. 1 – December 15, 2016
GSM-60 ENMET
1.0 Introduction
The GSM-60 is a sample draw monitoring instrument that measures and detects certain toxic hazards gases. The GSM-60 was designed primarily for monitoring one area with a single sampling hose, either for an individual gas or multiple gases. The instrument has one sampling pump and one flow sensor. The instrument is available with up to 4 internal sensors. Available sensors include, but are not, limited to carbon monoxide (CO), carbon dioxide (CO
2
), Hydrocarbons (HC) and variations in the oxygen (O
2
) content. The sensors can be used alone or up to four sensors can be used together. Some applications require monitoring of more than one area. If two sampling hoses are connected to one GSM-60 the flow sensor can only detect a total flow fault condition. For example, a pump failure or both sampling hoses being blocked. The system cannot detect a flow fault if only one of the two sampling hoses is blocked.
Flowmeters should be installed in both sampling lines to provide a visual indication of proper flow. In the instrument, a 24 V DC sampling pump pass air over each sensor and the resulting electrical outputs are used to evaluate the air for the target gases.
The GSM-60 is a highly adaptable instrument. Some adaptations require an addendum be added to the manual to facilitate use of the instruments with these adaptations. If addendum is needed, see page(s) between sections replacement part numbers and warranty.
Some features of the instruments are as follows:
continuous monitoring of the sample air
continuous LCD display of gas and vapor concentrations
menu driven operational and maintenance controls
menu driven calibration procedure
audio and visual alarms indicate unsafe conditions
alarm relay contacts available on terminals
a fault relay and visual fault alarm
low air flow fault indication and display
alarm acknowledgement capability including audio defeat mA outputs for each target gas
Hydrocarbons (HC) are limited to gases with an ionization potential of 10.6 eV or less. See Appendix B for a list of gases and IPs.
N OTE : All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the GSM-60 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N OTE : It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor any claims for damage.
•
This shipment was carefully inspected, verified and properly packaged at ENMET and delivered to the carrier in good condition.
•
When it was picked up by the carrier at ENMET , it legally became your company’s property.
•
If your shipment arrives damaged: o Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and request immediate inspection of the carton and the contents. o After the inspection and after you have received written acknowledgment of the damage from the carrier, contact ENMET Customer
Service for return authorization and further instructions. Please have your Purchase Order and Sales Order numbers available.
•
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00.
Repair or replacement charges above that value are your company’s responsibility.
•
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
ENMET
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 Fax 734-761-3220
Toll Free: 800-521-2978
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the GSM-60 is received as ordered. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each GSM-60 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 3
Rev. 1 – December 15, 2016
GSM-60
2.0 Instrument Features
2.1 Exterior Features
The exterior of the instrument is shown in Figure 1 . The exterior features are as follows:
Feature Description
ENMET
Enclosure
Input Port
Visual
Alarms & Indicators
Pushbutton Switches
•
•
•
O
S
PTION
ELECT
Switch
Switch
Audio Defeat / Alarm
Acknowledge
Switch
An engineered thermoplastic box, approximately 10x8x6, with a clear hinged front cover.
The entrance for the air sample and calibration gas. The quick release fitting mates with one on the calibration adapter.
Front Cover Latch
Audio Alarm
Mounting Flanges
Output Port
A quick-release latch that holds the clear front cover in place, and is capable of being padlocked if desired.
A loud horn activated by certain alarm conditions.
Flanges with holes for mounting the enclosure to a vertical surface.
The exit to exhaust the air sample and calibration gas. Luer fitting.
2.2 Display Panel Features
The display panel, shown in Figure 1 , is viewed through the clear front cover of the enclosure, and is accessed by opening the cover. Features are as follows:
Feature Description
Display A 2 line, 16 character per line, LCD with backlight.
The numerical values of gas concentrations, and other information are displayed.
On either side of the display:
A red alarm LED for each sensor installed in the instrument, Low level alarm.
The top center of the panel:
A red alarm LED for all sensors installed in the instrument, High level alarm.
Near the center of the panel:
A green power indicator LED
A red fault alarm indicator LED
There are three of these, located near the center of the panel; they are yellow rectangular membrane switches. They are:
The top left switch.
Directly to the right of the OPTION switch.
Directly below the OPTION switch.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 4
Rev. 1 – December 15, 2016
GSM-60 ENMET
*Visual Alarm CO
(See *note below)
Visual Alarm 2
Display, see *note below
CO reading O
2
reading
H2S reading
*Visual Alarm O2
(If Oxygen option is installed)
See *note below
*Visual Alarm
Dew Point
(If Dew Point option is installed)
See *note below
OPTION
Switch
A LARM
A CKNOWLEDGE /
A LARM D EFEAT
Switch
CO ppm
H2S
ALARM 2
Channel #1 Channel #3
Channel #2 Channel #4
OPTION SELECT
ALARM
ACKNOWLEDGE
POWER
FAULT
O
2
%
Quick Release Fitting
Input Port
Sample/Calibration
SELECT
Switch
Power LED
Front Cover
Fault LED
Latch
Audio Alarm
Mounting Flanges
2 places
Sample Output Port
Figure 1: External Features of the GSM-60
* N OTE : Typical gas reading & alarm locations, depending on instrument configuration, alarms & readings may be in alternate locations
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 5
Rev. 1 – December 15, 2016
GSM-60 ENMET
2.3 Circuit Board Features
The Display Panel is hinged on the left and is released by unscrewing the 2 screws located in the right corners. After releasing the panel, it is swung to the left, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the enclosure interior. Features are shown in Figure 2 .
Feature Description
Relay Terminals
Output Terminals
HC Manifold
Sensor Manifold
Filter, Particulate
Sensor Terminals
J16, J18, J19
This group of terminals is located at the left side of the Circuit Board.
For the contacts for each of four alarm relays, and for the contacts of a fault relay.
One 4-20mA output per active channel. 2 channels/outputs per connector.
The PiD sensor is installed into this housing.
The sample manifold, the carbon monoxide, carbon dioxide and oxygen sensors are located under this housing.
Removes contaminate from air sample line.
Sensor/Transmitter connectors 24V DC 4-20mA Input
Figure 2: GSM-60 Interior Features
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 6
Rev. 1 – December 15, 2016
GSM-60
3.0 Installation
ENMET
3.1 Mounting of Instrument
The GSM-60 should be located near the air to be monitored. Sampling lines should be no more than 50 feet long. It is recommended that Teflon® (PTFE) lined tubing be used. Quick disconnect fittings are supplied for use with 1/8” ID tubing.
Mount the instrument on an appropriate vertical surface using the mounting flanges provided. Avoid areas with excessive vibration or temperature extremes. The holes in the flanges are 0.31 inch in diameter and form a 6 x 10.75-inch rectangle. See Figure 3.
It is recommended to use #8 drywall anchors and screws for mounting the GSM-60 to a drywall/sheetrock surface.
Mounting Holes
0.31” dia. 4 places
Dimensions are in inches.
Figure 3: GSM-60 Mounting Dimensions
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 7
Rev. 1 – December 15, 2016
GSM-60 ENMET
3.2 Power Supply
The input power can vary from 100 to 240V AC , 50/60 Hz. Mains power should be connected to the Power Input Terminal J23 and the ground screw J21 . Se e Figure 2 for location.
W ARNING : Continuous gas detection and alarm systems (110V AC /220V AC / 24V DC /12V DC powered) become inoperative upon loss of primary power. Contact factory for specifications and pricing of backup battery systems.
Upon supplying air and power to the instrument:
The green power on LED is lit.
The display backlight is lit, and instrument will step through a start-up sequence: unit serial number, software revision and gases monitored may be shown on the display.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch. If alarm persists longer than 30 minutes, call ENMET customer service personnel.
3.3 Inputs / Outputs
Two types of alarm outputs are available, relay contacts and 4-20mA outputs.
Connector 2
Channel 3 & 4
4-20mA Output
Relay 1
Channel 1
Alarm 1
Relay 2
Channel 2
Alarm 1
Relay 3
Channel 3
Alarm 1
Relay 4
Channel 4
Alarm 1
Relay 5
Channel 1-4
Alarm 2
Relay 6
Ch 1-4 / System
Fault
Connector 1
Channel 1 & 2
4-20mA Output
Connector RS485
Connector RS232
Connector RS485
Input Connector
24V DC
GND mA
Input Connector
24V DC
GND mA
Input Connector
24V DC
GND mA
Ground Screw J21
Power Input
Terminal J23
Figure 2A: Relay, Input and Output Terminals
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 8
Rev. 1 – December 15, 2016
GSM-60 ENMET
3.3.1 Sensor/Transmitter Connection
Sensor/Transmitters are connected to the GSM-60 control unit with two or three-conductor wiring, use the correct oil tight fitting. Size of wire depends on the distance between the sensor/transmitter and the control unit.
See Recommended Wire Gauge Table below.
2 Wire for Sensors/Transmitter 3 Wire for Sensors/Transmitter
Position
1
2
3
Function
Power +24 V DC
Not Used
Signal/Return to Ground
Position
1
2
3
Function
Power +24 V DC
Power Ground
Signal
Recommended Wire Gauge
Distance from Sensor to Control Unit Recommended Wire Gauge
< 500 feet
501 – 800 feet
Longer Distances
16 AWG
14 AWG
Contact Factory
N OTE : Sensor Location
Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some examples are given below.
Heavier than Air Gas Sensor Location
Bottled LP (liquefied petroleum)
Propane
Butane
Gasoline
Trichloroethylene
Vaporized hydrocarbons
Hydrogen sulfide
Lighter than Air Gas
Natural gas (methane)
Ammonia
Hydrogen
•
Interior wall; 18-24" from floor.
D O N OT locate directly above or beside gas appliances (ovens, heaters).
•
Avoid locating anywhere near a vent or window or near an outside doorway.
Sensor Location
Near ceiling.
•
D O N OT locate directly above appliances where it is subject to direct exposure to heat or steam.
Sensor Location Same Density as Air Gas
Carbon Monoxide 4-6 feet above the (generally uniform) floor.
•
D O N OT locate in direct air currents of windows, doors, or vents.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
3.3.2 Relay Contacts
Relay contacts are available for each alarm; these are SPDT, rated at 10Amp at 110V AC , and may be latching or non-latching as required by the application.
They are accessed on the terminals next to each relay see Figure 2 & 2A . The contact positions are noted on the circuit board next to each terminal.
Relays may also be configured as failsafe or non-failsafe. The default alarm relay configuration is for latching mode, and failsafe. They may be reconfigured in the maintenance menu. See section 5.3.5 & 5.3.6
The PC Board is labeled for the relays in their un-energized state. If the relay is configured for failsafe, then this is also the alarm condition state. Non-failsafe configured relays in the alarm state, are the reverse of the PC board labeling. Note that the
Fault(FLT) relay cannot be set to operate in a Non-Failsafe mode. Please see the Table 1 on page 10:
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 9
Rev. 1 – December 15, 2016
GSM-60 ENMET
Table 1 : Relay Failsafe Settings
Position
J5 Relay 1 - NO
J5 Relay 1 - COM
J5 Relay 1 - NC
J6 Relay 2 - NO
J6 Relay 2 - COM
J6 Relay 2 - NC
J8 Relay 3 - NO
J8 Relay 3 - COM
J8 Relay 3 - NC
J10 Relay 4 - NO
J10 Relay 4 - COM
J10 Relay 4 - NC
J14 Relay 5 - NO
J14 Relay 5 - COM
J14 Relay 5 - NC
J15 Relay 6/FLT - NO
J15 Relay 6/FLT - COM
J15 Relay 6/FLT - NC
Failsafe-Alarm
Normally Open
Common
Normally Closed
Normally Open
Common
Normally Closed
Normally Open
Common
Normally Closed
Normally Open
Common
Normally Closed
Normally Open
Common
Normally Closed
Normally Open
Common
Normally Closed
Non-Failsafe-Alarm
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
Normally Closed
Common
Normally Open
N/A
N/A
N/A
Relays can be linked to specific alarms. The table below shows the default relay links. They may be changed in the maintenance menu if required. See Section 5.0
.
Relay 1
Relay 2
Relay 3
Relay 4
Relay 5
Channel 1
Low Alarm
High Alarm
Channel 2
Low Alarm
High Alarm
Channel 3
Low Alarm
High Alarm
Channel 4
Low Alarm
High Alarm
In addition, there is a fault relay, which changes state whenever the instrument is in a fault condition. The contact positions are noted on the circuit board next to each terminal. See Figure 2A.
The coil of this relay is energized when the instrument is in the non-fault state; the contact conditions given on the circuit board next to the terminal, are for the non-energized state, which is identical to the fault state.
These relay contacts can be used to operate auxiliary alarms or other functions. It is recommended that power for auxiliary equipment be supplied from an independent power source, separate from the GSM-60 . Place a hole in the enclosure for a wire exit, and use appropriate cable fittings. Be sure to note the location and depth of hardware inside the enclosure.
3.3.3 Optional 4-20mA Outputs
Isolated 4-20 mA outputs are available for data logging or other purposes. An output is supplied for each sensor supplied in an instrument, and can be added when a sensor is added in the field. These outputs are available on the Connector 1 for channels 1
& 2 and Connector 2 for channels 3 & 4.
4mA corresponds to a sensor reading at the bottom of the instrument range and 20mA corresponds to a full-scale reading.
Standard ranges are shown in Table 2 .
Table 2: Sensor Output
Sensor 4mA 20mA
CO
O2
CO2
HC
0
0
0
0
50
30
5000
100
Wiring requirements are the same as for the relays.
3.4 Installation Verification
All instruments are calibrated at the factory. You may, if a calibration kit is available, calibrate the all gas channels of the instrument 24 hours after installation to verify proper installation and instrument operation. See Section 5.0
, Maintenance, for calibration instructions. Calibration is also recommended after the first month of operation. Subsequent calibrations should be performed every 3 months. The dew point sensor cannot be calibrated in the field.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 10
Rev. 1 – December 15, 2016
GSM-60
4.0 Operation
ENMET
4.1 Normal Operation Condition
With the GSM-60 installed as described in Section 3 , and in clean air, the POWER green LED is on, the display is lit and the information on the display is as shown in Figure 4 Display , for the sensor(s) installed in the GSM-60 .
The red alarm and fault
LEDs are not lit.
CO
PPM
O
2
%
H2S
PPM
CO 0
H2S 0
O2 20.9
CO2 300 CO2
PPM
Example of display with CO (ch 1), H
2
S (ch 2), Oxygen (ch 3) and CO
2
(ch 4) options installed
Figure 4: GSM-60 Operational Display
4.2 Alarm Set Points
There are two alarm set points for each installed channel of the GSM-60 . The factory settings of these alarm set points are shown in Table 3 .
Typical Channel # Gas
Table 3: Typical Factory Alarm Set Points
Alarm 1, Flashing LED Alarm 2, Steady LED
1
2
3
4
4
Carbon Monoxide
Hydrogen Sulfide
Oxygen Deficiency
Carbon Dioxide
Hydrocarbon
10 ppm
10 ppm
19.5 % by volume
500 ppm
5 ppm
20 ppm
20 ppm
23.5 % by volume
1000 ppm
10 ppm
The alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.
If the CO concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.
If the dew point increases above that of the alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.
If the oxygen content of the sample air decreases below the deficiency alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.
If the oxygen content of the sample air exceeds that of the abundance alarm set point, the associated red LED is lit, the audio alarm is activated, and both the oxygen alarm relay and the oxygen high alarm relay change state.
The HC sensor can only detect and alarm to hydrocarbons with an Ionization Potential of less than 10.6 eV. See Appendix
B .
The HC sensor is broad range in nature and is unable to differentiate between different hydrocarbons.
The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value. Example : If the alarm set point is
Λ
10 and the differential is 2, the
GSM-60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 11
Rev. 1 – December 15, 2016
GSM-60 ENMET
4.3 Alarm Latching or Differential Settings
An instrument is shipped with the alarms in the latching mode. The alarms may be independently configured in the non-latching mode or differential setting by use of the maintenance menu.
See Section 5.3.3
, for setting alarm 1 and alarm 2.
Standard Setting
I N THE LATCHING MODE : at the cessation of the condition which causes an alarm, the alarm indications do not cease, and the alarm relay contacts do not revert to the non-alarm state, until the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch is pressed.
An alarm can also be acknowledged by pressing the switch during the alarm condition; then at the cessation of the alarm condition, alarm indications cease and alarm relays revert to the non-alarm state. After an alarm is acknowledged, alarms in the latching configuration are re-armed to latch at the next alarm condition.
I N THE NON LATCHING MODE : at the cessation of the condition that causes an alarm, the alarm indications automatically cease, and the alarm relay contacts revert to the non-alarm state.
Differential Setting
The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value. Example : If the alarm point is
Λ
10 and the differential is 2, the GSM-
60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.
4.4 Audio Defeat
Pressing the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch during an alarm temporarily silences the audio alarm. Relays and alarm LEDs continue to function, in the alarm state, during an alarm condition. If the alarm condition persists, the audio alarm will
“chirp” every 20 seconds.
If after 15 minutes the alarm condition continues the audio alarm will reactivate at full intensity.
If any other alarm condition occurs while the audio alarm has been silenced it will force the audio alarm to reactivate immediately.
4.5 Display
In clean air a display is shown in Figure 4 .
This position of the display is termed the " operational display ". As explained below, the display can be used to view other information by using the OPTION and SELECT switches.
Concentrations of CO and CO2 are given in PPM (parts per million parts of air). Dew point is given in degrees Fahrenheit at 55
PSIG ; this can be changed to degrees Centigrade by pressing the SELECT switch . Oxygen concentration is given in percent by volume.
When sample flow is reduced below a limit, the bottom line of the display flashes “Low Flow Alarm”.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 12
Rev. 1 – December 15, 2016
GSM-60 ENMET
4.6 Operational Menu
The operational menu allows the user to:
View alarm set point concentration values
View alarm ascending/descending trigger, latching and delay configurations
Enter the maintenance menu with the proper Password.
The operational menu is accessed with the OPTION and SELECT switches. The operational menu flow chart is shown in Figure 5 ,
Pressing the OPTION switch is indicated with a " O "
Pressing the SELECT switch is indicated with a " S ".
If the instrument is left at any location in the operational or maintenance menus, other than the operational display, with no action taken for a period of 45 seconds, it returns to the operational display.
CH-1 CH-3
CH-2 CH-4
O
S
No
Function
O = Press Option switch
S = Press Select switch
O
Relays 1 2 3 4 5 6
[ =ON [[[[[[
O
CH1 SCALE (CO)
0 – 50
PPM
O
Λ L10 A1 vL19.5
vL-40 Λ D 500
O
Λ L20 A2 Λ 23.5
Λ L39 Λ 1000
O
ALARM1 Delays
(Seconds)
O
50 mASPAN 30.0
68 5000
O
Enter Maint Menu
S
S
O
O
S
S
O
S
S
S
CH3 SCALE
(O2)
S
CH4 SCALE (CO2)
0 – 5000
PPM
No
Function for each active channel
CH2 SCALE (H2S)
0 – 50 PPM
S
Alternating
No
Function
No
Function
5
5
No
Function
_
Enter Password
5
5
S
Displays are examples of gases:
Channel 1 = Carbon Monoxide
Channel 2 = Hydrogen Sulfide
Channel 3 = Oxygen
Channel 4 = Carbon Dioxide
Displays are examples of Alarms
Λ
- Indicates alarm triggered on v - Indicates alarm triggered on decreasing value of reading
Displays are examples of Alarms
L – Indicates alarm is in latching
(no L present) – Indicates alarm is in non-latching mode.
Displays are examples of Alarms
D – Indicates alarm is in
(no D present) – Indicates alarm is in Standard Setting.
No increasing value of reading mode.
Differential Setting.
Function
See Maintenance Menu Diagram
Figure 5: GSM-60 Operation Menu Flow Chart
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 13
Rev. 1 – December 15, 2016
GSM-60 ENMET
4.7 Fault Indications
4.7.1 Low Flow Indication
A flow sensor is used to furnish a low flow indication. When the sample air pressure drops below preset levels, the fault light and audio alarm are activated, and the display flashes “Low Flow Alarm”. If two sampling hoses are connected to one GSM-
60 , the flow sensor can only detect a total flow fault condition. For example: a pump failure or both sampling hoses being blocked. The system cannot detect a flow fault if only one of the two sampling hoses is blocked.
4.7.2 Other Fault Indications
Other fault indications are associated with sensor zero and calibration activities, and are described in the maintenance Section
5.0
of this manual.
4.8 Hydrocarbon Sensor Response
If a Hydrocarbon (HC) sensor is supplied with the GSM-60 instrument, it designed to detect hydrocarbon gases and vapors with an ionization potential (IP) of 10.6 eV or less. Hydrocarbons with an IP of greater than 10.6 eV will NOT be detected. Please see
Appendix B for a list of common gases and vapors and their respective IP rating.
Unless otherwise noted Isobutylene is used as a calibration and reference gas.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 14
Rev. 1 – December 15, 2016
!
GSM-60
5.0 Maintenance
ENMET
The GSM-60 requires periodic sensor calibration and replacement. Calibration of toxic gas and oxygen sensor should be performed immediately following installation, one month after installation and every 3 months thereafter. HC sensor should be calibrated monthly. Oxygen and CO sensor have an estimated lifetime of 1 – 2 years. The CO2 sensor has an estimated lifetime of 3 years.
Other sensors vary. Sensors should be replaced when they will not calibrate or shortly before the end of the estimated lifetime.
5.1 Cleaning Instructions
C AUTION : Never spray a cleaning solution on the surfaces of the GSM-60 devices.
Clean the exterior of the GSM-60 enclosures with a mild soap solution on a clean, damp cloth. Do not soak the cloth with solution so that moisture drips onto, or lingers on, external surfaces.
Under no circumstances should organic solvents such as paint thinner be used to clean instrument surfaces.
5.2 Maintenance Menu
5.2.1 Accessing Maintenance Menu
The GSM-60 maintenance menu is accessed by entering the proper password with the OPTION and SELECT switches. See
Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.
Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password is 1270.
When a valid numerical password is inserted, the user can enter the maintenance menu.
To enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.
In the "Enter Maint Menu" position
Press the SELECT switch "Enter Password
ζ
0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.
In the
ζ
000 position, the underline cursor is under the left digit.
Press the OPTION switch to change the left digit; select the correct digit.
Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint
Menu display.
CO
PPM
H2S
PPM
Enter Password
1270
O
2
%
Example : Password Display ( with factory installed password entered ) and Flow Chart below.
O
O (6)
Enter Maint Menu
S Enter Password Changes digit indicated by underscore cursor
ζ
0000
S
Locks underscored digit and moves cursor to next digit
Invalid
O
= Option Switch
Valid
S
= Select Switch
To Calibration
See Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 15
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.2.2 Maintenance Menu Flow Chart
The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart . From the operational display, press the
OPTION switch 6 times; "Enter MAINTENANCE Menu" is displayed.
To/From
Operational
Display
From Operational Menu
Press
O PTION
(6 times)
O
= Press Option
O (6)
Enter Maint Menu S Enter Password
O
S
= Press Select
Changes digit indicated by underscore cursor
0000
S
Locks underscored digit and moves cursor to next digit
Invalid
Valid
MAINTENANCE MENU
Calibration
S
O
If installed
MAINTENANCE MENU
Scale mA Xmtrs
O
MAINTENANCE MENU
Set Alarm1
S
S
Calibration
Select (Gas)
Scale mA Xmtrs
Select (Gas)
Alarm 1
Select: XX
Press O PTION until the channel to be Calibrated is displayed
See Section 5.3 for calibration instructions.
Press O PTION until gas to be Set-Up is displayed
See Section 5.3.2 for transmitter set-up instructions.
Press O PTION until the gas alarm to be Set is displayed
See Section 5.3.3 for setting alarms instructions.
O
MAINTENANCE MENU
Set Alarm2
S
O
MAINTENANCE MENU
Set Alarm Delays
S
O
MAINTENANCE MENU
Configure Alarms
S
O
MAINTENANCE MENU
Relay Failsafes
S
O
MAINTENANCE MENU mA Output Span
S
Alarm 2
Select: XX
Alarm Delay
Select: XX
Ch 1, 2, 3, 4
R1 L
Relay Failsafes
R:1 Failsafe ON mA Output Span
Select: (Gas)
Press O PTION
until the gas alarm to be Set is displayed
See Section 5.3.3 for setting alarms instructions.
Press O PTION until the gas alarm delay to be Set is displayed
See Section 5.3.4 for setting alarms instructions.
Pressing O PTION changes letter indicated by underscore cursor
See Section 5.3.5 for configuring relay instructions.
Pressing O PTION changes Failsafe setting from ON to OFF
See Section 5.3.6 for configuring relay failsafe instructions.
Press O PTION until the gas span to be Set is displayed
See Section 5.3.7 for output span instructions.
O
MAINTENANCE MENU
Set New Password
S
S
O
O
MAINTENANCE MENU
Exit maint menu
Password
XXXX
O
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
See Section 5.3.8
for changing password instructions.
Press O PTION to return to top of maintenance menu.
Press S ELECT to return to operational menu.
See Section 5.3.9 for instructions on how to exit Maintenance Menus.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
Figure 6: GSM-60 Maintenance Menu Flow Chart.
P a g e | 16
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3 Calibration for CO, O
2,
HC and CO
2
(Gas Channels)
Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. This is a two-step process.
A Low Calibration sets clean air reference point and the High Calibration function sets the sensitivity of the instrument.
Calibration equipment is available from ENMET to calibrate the GSM-60 . A list of needed material is in Section 7.0. A calibration adapter will have a fitting for the gas cylinder on one side, and a quick-disconnect to attach to the instrument on the other.
You may exit the calibration section, at any time, by pressing and holding the OPTION switch for 3 seconds, if entering calibration section by mistake or calibration gas is not available.
Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the OPTION and SELECT switches on the front panel.
After entering a valid password to maintenance menu, see Section 5.2.1
, the calibration section is the first menu section; enter by pressing the SELECT switch.
Supply sensor with clean air for Low Cal/Zero Cal setting and apply calibration gas for Hi Cal/Span Gas setting.
Press the SELECT switch "Calibration Select XX " is displayed. XX = the gas to be calibrated
Press the OPTION switch, if needed, to change to the gas to be calibrated.
Press the SELECT switch, the gas & current reading is displayed in upper portion of display. The mV reading & "Low Cal 0" is displayed in the lower portion of display. This is the Low Cal setting, usually zero , clean air must be supplied to the sensor.
This reading needs to be at or near zero. If it is not, then a cylinder of clean 20.9 air should be used. See Figure 7 if this is required.
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to
"Hi Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.
Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.
Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration Menu
Repeat above steps for each channel to be calibrated.
N OTE : Instruments equipped with an oxygen sensor programmed with a range of 0 to 15% by volume require 99.999 % Nitrogen
( ENMET part number 03295-100) be used for the Low Cal/Zero Cal setting and 5% by volume Oxygen ( ENMET part number 03296-050) be used for Hi Cal/Span setting
To continue to next section, Press the OPTION switch.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Full Calibration Flow Chart, for CO
From Valid Password Entry
Default Calibration Points
Gas Low Cal Hi Cal
MAINTENANCE MENU S
Calibration
Calibration
Select (Gas)
O Press O PTION until the gas to be
Calibrated is displayed
CO
O2
0
N/A*
20
20.9*
CO2 0 1000
S
CO: XX
LowCal: 0000
11
S each digit
CO: XX
HiCal: 0000
14
HC 0 10
O
S each digit
O
= Press Option
S
= Press Select
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 17
Rev. 1 – December 15, 2016
GSM-60
Calibration/Sample Port
Fitting
N OTE : With twin tube input, the noncalibration line must be plugged.
Quick Disconnect
ENMET
Cylinder Valve and Regulator
Cylinder of Gas
To Cylinder of Gas
GSM-60
with
O
PTIONAL
T
WIN
T
UBE
Example of Regulator
Humidifier
Calibration Adapter with Humidifier - MOS
Figure 7: GSM-60 Calibration Connections
Calibration Process for MOS Sensors
The following calibration procedure must be followed whenever a 60-series part number appear in the part number of the instrument, examples 046526200 -0000, 04652-70836000
Humidification is required when calibration is performed. In addition to the standard flow demand regulator, cylinder of calibration gas you need ENMET humidifier assembly part number 037000-000
Fill the humidifier with clean water to about ¾ full, connect the flow demand regulator to one side of the humidifier and connect the other side of the humidifier to the GSM-60 calibration/sampling port.
Warning: Be sure that the flow direction is correct, note the arrow on the humidifier , failure to do so will cause damage to the
GSM-60 instrument
With the cylinder regulator and humidifier assembled follow the above steps to complete the calibration process
5.3.1A Low Cal/Zero Cal Adjust
A Low-Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.
Enter the maintenance menu by repeatedly pressing OPTION switch, until the maintenance menu is displayed. See Figure 6,
GSM-60 Maintenance Menu flow chart.
The first menu available is the Low Cal/Zero Cal.
Press the SELECT switch 4 times to perform a Low Cal.
If the Low Cal/Zero Cal is successful , the display will change to Hi Cal/Span Gas.
If you wish to Hi Cal/Span Gas the sensor apply calibration gas. Proceed to gas calibration Section 5.3.1B
If you wish to Exit the maintenance menu, Press and hold OPTION switch until the Maintenance Menu is displayed then release. Then press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the Operational Display
If the Low Cal/Zero Cal is Not successful, sensor is outside of safe parameters to Low Cal, a “SLP/Off Set err” will be indicated. Repeat Section 5.3.1 Low Cal/Zero Cal Adjust making sure to use a cylinder of 20.9% Oxygen.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 18
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.1B High Cal/Span Gas Adjust
A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed.
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to
"HI Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.
Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.
Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration
Menu
Repeat above steps for each channel to be calibrated.
To continue to next section, Press the OPTION switch.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
5.3.2 Set 4 –20mA Transmitter Scale
This section of the maintenance menu is installed when there are 4-20mA style sensors for dew point or other gases. This function is normally performed at the factory and is not usually required to be performed in the field unless a new transmitter is installed.
After entering a valid password into maintenance menu, the Scale mA Xmtrs section is the second menu section, if it is installed, enter by pressing the S ELECT switch
Press the SELECT switch "mA Xmter Scale: Select XX " is displayed. XX = the gas to be set up.
Press the OPTION switch, if needed, to change to the gas to be set up.
Press the SELECT switch, “Ch#: mAXmter: 4mA: 0000” is displayed
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display move to the full-Scale mA Xmtrs menu
Press the SELECT switch, “Ch#: mAXmter: 20mA: 0000” is displayed
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will return to the Scale mA Xmtrs menu
Repeat these steps for each 4 –20mA transmitter.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Sensor/Transmitter Set Up Flow Chart
MAINTENANCE MENU S mA Xmtr Scale
Scale mA Xmtrs
Select (Gas)
O Press O PTION until the gas to be
Set Up is displayed
O
S
= Press Option
= Press Select
O S
CH#: XX mA Xmter
4mA: 0000
S each digit
CH#: XX mA Xmter
20mA: 0000
S each digit
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 19
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.3 Set Alarm Points
Factory alarm set points are discussed in Section 4.2, See Table 1. To change the alarm points, you must enter the maintenance menu.
Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password is 1270.
When a valid numerical password is inserted, the user can enter the maintenance menu.
In the "Enter Maint Menu" position
Press the SELECT switch "Enter Password
ζ
0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.
In the
ζ
000 position, the underline cursor is under the left digit.
Press the OPTION switch to change the left digit; select the correct digit.
Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint
Menu display.
After entering a valid password:
Press the OPTION switch until; “Maintenance Menu Set Alarm1” appears on display.
Press the SELECT switch, "ALARM1 Select: XX " is displayed. XX = the gas of alarm point to be changed.
Press the OPTION switch until, desired gas is displayed.
Press the SELECT switch; "ALARM 1 V " is displayed, with the flashing placeholder underscore cursor, under the left most character,
Λ
for ascending trigger point or V for descending trigger point indicator.
Press the OPTION switch to toggle between
Λ
and V ; select the correct indicator.
Press the SELECT switch to lock in the correct indicator. "ALARM 1 STD " is displayed
Press the OPTION switch to toggle between STD and DIFF ; select the correct indicator.
Press the SELECT switch to lock in the correct indicator.
If STD is selected, "ALARM 1 V L " is displayed.
The next character is the latching indicator L or NO L press the OPTION switch to toggle the latching mode.
The next character is the negative sign – press the OPTION switch to toggle the negative sign.
The next characters are the alarm 1 value, press the OPTION switch to select each digit of the value
When the last digit is accepted display returns to the "Set Alarm1" position.
If DIFF is selected, "ALARM 1 DI
FF Λ
000 " is displayed.
The next characters are the alarm 1 value, press the OPTION switch to select each digit of the value
Press the SELECT switch to lock in the correct character and move the cursor to the right.
"ALARM 1 DIFF BAND 000 " is displayed, press the OPTION switch to select each digit of the value.
The next characters are the alarm 1 differential value, press the OPTION switch to select each digit of the value
Press the SELECT switch to lock in the correct character and move the cursor to the right.
When the last digit is accepted, display returns to the "Set Alarm1" position.
Note : The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value.
Example : If the alarm set point is
Λ
10 and the differential is 2, the GSM-60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.
Repeat for each sensor alarm 1 to be changed.
Press the OPTION switch to move to alarm 2, "Set ALARM2" is displayed.
Repeat as for alarm 1 using the STD section.
Press OPTION switch until “Exit Maint Menu” appears, then press SELECT switch to return the instrument to the Operational
Display
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 20
Rev. 1 – December 15, 2016
GSM-60 ENMET
Example: Set Alarms Flow Chart
Displays are examples of Alarms
Λ
- Indicates alarm triggered on increasing value of reading v - Indicates alarm triggered on decreasing value of reading
O
= Press Option
S
= Press Select
L - Indicates alarm is set for latching
NO L - Indicates alarm is set for non-latching
STD – Indicates alarm in standard setting, can be set in latched or non-latched mode
DIFF – Indicates alarm in differential setting, instrument will stay in alarm beyond the alarm set point by the differential value
MAINTENANCE MENU
Set Alarm1
S Alarm 1 -
Select: XX
O Press O PTION until the channel to be Set is displayed
O
S
Alarm 1
Λ
O
S
Toggles status between
Locks selection
Λ
increasing trigger and v decreasing trigger
S
Alarm 1
STD
DIFF S
O
S
Toggles status between Standard and D IFF
Locks selection
Alarm 1
Λ DIFF 000
O
Changes character indicated by underscore cursor
S
Locks underscored character and moves cursor to next digit
S TD
S
Alarm 1
Λ ζ
L
Alarm 1
Λ
L–010
O
Alarm 1 DIFF BAND
O
000 S
Changes character indicated by underscore cursor
Locks underscored character and moves cursor to next digit
Toggles status between
Latching and Non Latching
S
Locks selection
O
S
Changes character indicated by underscore cursor
Locks underscored character and moves cursor to next digit
MAINTENANCE MENU
Set Alarm2
S
O
Alarm 2 -
Select: XX
S
Alarm 2
V L 0000
O Press O PTION until the channel to be Set is displayed
O Changes character indicated by underscore cursor
S
Locks underscored character and moves cursor to next
See Section 4.2 Table 3 for factory alarm set points.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 21
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.4 Set Alarm Delay
The alarms may be set to delay by 1 second increments, up to 255 seconds. Alarm delays are factory set to 5 seconds.
To change an alarm delay, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1
.
See Table 4 below for factory set delays. A space is provided to record changes.
After entering a valid password:
Press the OPTION switch until; “Maintenance Menu Set Alarm Delay” appears on display.
Press the SELECT switch, "ALARM Delay Select: XX " is displayed. XX = the gas alarm to be changed.
Press the OPTION switch until, desired gas is displayed.
Press the SELECT switch; "ALARM Delay = 0000" is displayed, with the underscore cursor under the left digit.
Press the OPTION switch to change the left digit; select the correct digit.
Press the SELECT switch to lock in the correct digit and move the cursor one digit to the right. When the last digit is accepted, display returns to the "Set Alarm Delay" position.
Repeat for each sensor alarm delay to be changed.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Set Alarm Delay Flow Chart
MAINTENANCE MENU
Set Alarm Delays
S
Alarm Delay -
Select: XX
S
O
Alarm Delay -
0005
O
O
S
Press Option until the channel to be Set is displayed
O
S
= Press Option
= Press Select
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
Table 4: Factory Set Gas alarms Delay
Gas Delay
CO 5 sec
H
2
S 5 sec
O
2
5 sec
CO
2
5 sec
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 22
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.5 Relay Configuration
To change a relay configuration, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.
In the "Enter Maint Menu" position
Press the SELECT switch "Enter Password
ζ
0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.
In the
ζ
000 position, the underline cursor is under the left digit.
Press the OPTION switch to change the left digit; select the correct digit.
Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint
Menu display.
After entering a valid password:
Press the OPTION switch until “Configure Alarms” is displayed
Press the SELECT switch to enter the Configure Alarms menu
Press the OPTION switch to set relay configuration as needed, see below for indications
L = Low Alarm, H = High Alarm, B = Both Alarms,
ζ
= No Relay linked to channel
Press the SELECT switch to lock setting and move to next, channel and relay
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Set Relay Configuration Flow Chart
MAINTENANCE MENU S Ch 1,2,3,4
R1 L
Configure Alarms
Ch 1 2 3 4
R2 L
O
Ch 1 2 3 4
R3 L
Pressing O PTION changes letter indicated by underscore cursor
Pressing S ELECT Locks underscored digit and moves cursor to next digit
Ch 1 2 3 4
R4 L
Ch 1 2 3 4
R5
H H H H
O
S
= Press Option
= Press Select
The table below shows the default relay links.
Relay 1
Relay 2
Relay 3
Relay 4
Relay 5
Channel 1
Low Alarm
High Alarm
Channel 2
Low Alarm
High Alarm
Relays can be linked to specific alarms.
N OTE : Each operating channel must be linked to at least 1 relay.
Channel 3
Low Alarm
High Alarm
Channel 4
Low Alarm
High Alarm
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 23
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.6 Failsafe Configuration
To change a relay failsafe configuration, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint
Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.
In the "Enter Maint Menu" position
Press the SELECT switch "Enter Password
ζ
0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.
In the
ζ
000 position, the underline cursor is under the left digit.
Press the OPTION switch to change the left digit; select the correct digit.
Press the switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the
Enter Maint Menu display.
After entering a valid password:
Press the OPTION switch until “Relay Failsafes” is displayed
Press the SELECT switch to indicate relay to be set.
Press the OPTION switch to set relay indicated, On or Off as appropriate.
Press the SELECT switch to cycle through each of the 5 relays, return to “Maintenance Menu Relay Failsafes”
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Set Relay Failsafe Configuration Flow Chart
MAINTENANCE MENU
Relay Failsafes
O
S
Relay Failsafes
R: 1 Failsafe ON
Relay Failsafes
Pressing O PTION changes Failsafe setting from ON to OFF
Continue pressing S ELECT to cycle through all 5 relays
R: 2 Failsafe ON
Relay Failsafes
R: 3 Failsafe ON
Relay Failsafes
Changing the setting of each relay as needed, using the Option switch
S
R: 4 Failsafe ON
Relay Failsafes
R: 5 Failsafe ON
O
S
= Press Option
= Press Select
5.3.7 Set Output Span Range
To change 4-20 mA output range. This range is set at the factory and should not be changed, contact ENMET for information.
Press the OPTION switch to continue to next section of maintenance menu.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Set Output Span Flow Chart
MAINTENANCE MENU mA Output Span
S mA OUTPUT SPAN
Select: CO
S
O mA OUTPUT SPAN
_ 50 CO
O Press O PTION until the channel to be Set is displayed.
O
S
Changes digit indicated by underscore cursor
O
S
= Press Option
= Press Select
Locks underscored digit and moves cursor to next digit
S
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 24
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.3.8 Set New Password
To change the password, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1.
To set a new password, after inserting a valid password,
Press the OPTION switch until; "Set New Password” is displayed.
Press the SELECT switch; "Password
ζ
1270" is displayed, with the underscore cursor under the left digit.
Use the OPTION switch to change the left digit, when the desired digit is displayed.
Press the SELECT switch to lock the digit in place and move the cursor one digit to the right.
When all four digits of the new password have been selected, "Set New Password" is displayed.
Record the new password; without it, the maintenance menu cannot be reentered once you exit the Maintenance Menu. If the password is lost, call ENMET customer service personnel.
From the "Password XXXX " position,
Press the SELECT switch to return to Set New Password section.
Press the OPTION switch; to continue to "exit MAINTENANCE Menu"
Example: Set Password Flow Chart
MAINTENANCE MENU S
O
Set New Password
O
Password
0000
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
= Press Option
S
= Press Select
5.3.9 Exit Maintenance Menu
From the "exit MAINTENANCE Menu" position
Press the SELECT switch to resume the operational display.
Press the OPTION switch to reenter the maintenance menu at the "Calibration" position.
Example: Exit Maintenance Menu Flow Chart
MAINTENANCE MENU
Exit Maint Menu
CH-1 CH-3
CH-2 CH-4
MAINTENANCE MENU
Calibration
O S
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 25
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.4 Sensor Replacement
5.4.1 Gas/Oxygen Sensor
A Gas sensor must be replaced when it can no longer be calibrated. To replace a sensor, perform the following steps.
Turn off the electrical power. The sample air can continue to flow.
Open the display panel and remove the four manifold retention screws and remove the manifold. See Figure 9 .
Remove the old sensor, and replace it with a new sensor. It is recommended that the sensor gasket/manifold also be checked and potentially replaced at the time the sensor is replaced.
C AUTION : Some new sensors come with a shorting clip that must be removed before installation, for proper operation. See
Figure 8 .
Replace the manifold.
Turn on the electrical power.
Remove Shorting Clip from Cell if Present
Orifice
To Pump
PiD (HC) Sensor Manifold
Bottom View of Sensor
Figure 8: Shorting Clip
Gas Sensor Manifold
Retaining Screw, 4 places
Figure 9: Location of Gas Sensor and PiD (HC) Manifolds
5.4.2 Calibration/Sensor Replacement
Sensor replacement requires that a Factory Calibration be performed. Factory Calibration allows the instrument to properly set operational parameters for each sensor.
Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. This is a two-step process. A Low Calibration sets clean air reference point and the High Calibration function sets the sensitivity of the instrument.
Calibration equipment is available from ENMET to calibrate the GSM-60 . A list of needed material is in Section 7.0. A calibration adapter will have a fitting for the gas cylinder on one side, and a quick-disconnect to attach to the instrument on the other.
You may exit the calibration section, at any time, by pressing and holding the OPTION switch for 3 seconds, if entering calibration section by mistake or calibration gas is not available.
Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the OPTION and SELECT switches on the front panel.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 26
Rev. 1 – December 15, 2016
GSM-60 ENMET
After entering a valid password to maintenance menu, see Section 5.2.1
, the calibration section is the first menu section; enter by pressing the SELECT switch.
Supply sensor with clean air for Low Cal/Zero Cal setting and apply calibration gas for Hi Cal/Span Gas setting.
Press the SELECT switch "Calibration Select XX " is displayed. XX = the gas to be calibrated
Press and Hold the OPTION switch, until the letter F appears in the upper right hand corner of the display. The F indicates that the instrument is in the Factory Calibration Mode.
Press the OPTION switch, if needed, to change to the gas to be calibrated.
Press the SELECT switch, the gas & current reading is displayed in upper portion of display. The mV reading & "Low Cal
0" is displayed in the lower portion of display. This is the Low Cal setting, usually zero , clean air must be supplied to the sensor. This reading needs to be at or near zero. If it is not, then a cylinder of clean 20.9 air should be used. See Figure 7 if this is required.
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to
"Hi Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.
Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.
Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration
Menu
Repeat above steps for each channel to be calibrated.
N OTE : Instruments equipped with an oxygen sensor programmed with a range of 0 to 15% by volume require 99.999 %
Nitrogen ( ENMET part number 03295-100) be used for the Low Cal/Zero Cal setting and 5% by volume Oxygen
( ENMET part number 03296-050) be used for HI Cal/Span setting
To continue to next section, Press the OPTION switch.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
Example: Full Calibration Flow Chart, for CO
From Valid Password Entry
Default Calibration Points
Gas Low Cal Hi Cal
MAINTENANCE MENU
Calibration
S Calibration
Select (Gas)
O Press O PTION until the gas to be
Calibrated is displayed
CO
O2
0
N/A*
20
20.9*
O
S
CO: XX
LowCal: 0000
11
S each digit
CO2
HC
0
0
1000
10
CO: XX
HiCal: 0000
14
S each digit
O
= Press Option
S
= Press Select
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 27
Rev. 1 – December 15, 2016
GSM-60
Calibration/Sample Port
Fitting
N OTE : With twin tube input, the noncalibration line must be plugged.
Quick Disconnect
2 places
Plug
ENMET
Cylinder Valve and Regulator
Cylinder of Gas
To Cylinder of Gas
GSM-60
with
O
PTIONAL
T
WIN
T
UBE
Example of Regulator
Humidifier
Calibration Adapter with Humidifier - MOS
Figure 7: GSM-60 Calibration Connections
Calibration Process for MOS Sensors
The following calibration procedure must be followed whenever a 60-series part number appear in the part number of the instrument, examples 046526200 -0000, 04652-70836000
Humidification is required when calibration is performed. In addition to the standard flow demand regulator, cylinder of calibration gas you need ENMET humidifier assembly part number 037000-000
Fill the humidifier with clean water to about ¾ full, connect the flow demand regulator to one side of the humidifier and connect the other side of the humidifier to the GSM-60 calibration/sampling port.
Warning: Be sure that the flow direction is correct, note the arrow on the humidifier , failure to do so will cause damage to the
GSM-60 instrument
With the cylinder regulator and humidifier assembled follow the above steps to complete the calibration process
5.4.3A Low Cal/Zero Cal Adjust
A Low Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.
Enter the maintenance menu by repeatedly pressing OPTION switch, until the maintenance menu is displayed. See Figure 6,
GSM-60 Maintenance Menu flow chart.
The first menu available is the Low Cal/Zero Cal.
Press the SELECT switch 4 times to perform a Low Cal.
If the Low Cal/Zero Cal is successful , the display will change to Hi Cal/Span Gas.
If you wish to Hi Cal/Span Gas the sensor apply calibration gas. Proceed to gas calibration Section 5.3.1B
If you wish to Exit the maintenance menu, Press and hold OPTION switch until the Maintenance Menu is displayed then release. Then press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the Operational Display
If the Low Cal/Zero Cal is Not successful, sensor is outside of safe parameters to Low Cal, a “SLP/Off Set err” will be indicated. Repeat Section 5.3.1 Low Cal/Zero Cal Adjust making sure to use a cylinder of 20.9% Oxygen.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 28
Rev. 1 – December 15, 2016
GSM-60 ENMET
5.4.3B High Cal/Span Gas Adjust
A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed.
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to
"Hi Cal xx " gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.
Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.
Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration
Menu
Repeat above steps for each channel to be calibrated.
To continue to next section, Press the OPTION switch.
Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the
Operational Display
5.5 Flow Control Orifice
A 0.0225-inch diameter orifice is used to set the flow rate. In well-maintained medical air systems, this orifice should not clog.
However, if difficulty is experienced in maintaining flow rate examine this orifice; replace it if necessary.
Orifice location will depend on instrument sensor configuration.
6.0 Technical Data and Specifications
N OTE : All specifications stated in this manual may change without notice.
Electrical Power
Storage and
Transport
Operation
Mechanical
Outputs
Examples of Sensors
15 Amp fused branch circuit
100-240 V AC
0.9 A
50/60 Hz
Board Mounted Fuse FH2, 0.630A, 5 x
20mm
Temperature: preferred
Relative Humidity
Atmospheric Pressure
Temperature:
Relative Humidity
Atmospheric Pressure
Dimensions:
Weight:
Material:
Strain relief:
Relays:
Analog:
Digital:
Audio:
Type
CO
O
2
CO
2
HC
Others
N OTE : Response time is dependent on sampling length.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
Range
0 – 50 ppm
0 – 30%
0 – 5000 ppm
0 – 100 ppm
-20
°
to +60
°
C (-4
°
to +140
°
F)
0
°
to +20
°
C (32
°
to 68
°
F)
0 - 99% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
0
°
to +40
°
C (32
°
to +104
°
F)
0 - 99% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
11 x 9 x 8 inches (27.9 x 22.9 x 20.3 cm)
8 lbs (3.6 kg)
Engineered thermoplastic with hinged front cover
5 – 12 mm OD
SPDT
Resistive Load Inductive Load
10A at 110 VAC 7.5A at 110 VAC
10A at 30 VDC 5A at 30 VDC
4-20 mA x 3
RS-232 – Modbus
RS-485 – Modbus
95 db at 2 ft
*Typical Response Time
T
90
= 30 seconds
T
90
= 15 seconds
T
90
= 30 seconds
T
90
= 30 seconds
Contact ENMET
Life
1 – 3 years
1 – 2 years
3 – 5 years
1 – 2 years
P a g e | 29
Rev. 1 – December 15, 2016
GSM-60
7.0 Replacement Part Numbers
7.1 ENMET part numbers for sensors and replacement parts:
Part number Description
03053-000
67025-1114
03015-014
03016-014
03028-XXX
67025-XXXX
03028-005
73540-002
64002-1000
06008-004
65057-011
65057-012
65057-010
73089-004
04018-128
04018-118
Sensor, CO2
Sensor, Oxygen
Sensor, MOS 812
Sensor, MOS 813
Sensor, PID, 10.6 eV
Contact ENMET
Sensor, Toxic
Contact ENMET
Sensor, Replacement lamp, PID, 10.6 eV
Orifice
Fuse, 1.0 Amp 5x20mm
Sensor Gasket
Terminal plug, 3 position
Terminal plug, 4 position
Terminal plug, 2 position
Filter Assembly, Particulate
Pump, Replacement for instrument s/n 118 and above
Pump, Replacement for instrument s/n 117 and below
Consult ENMET or your distributer for more information
7.2 ENMET part numbers for Calibration equipment:
Part number Description
03219-020
03296-209
03510-001
03223-1000
03510-002
03290-010
Gas Cylinder, 20 ppm CO in air
Gas Cylinder, 20.9% oxygen in nitrogen
Regulator Assembly, Flow Demand CO, O
2
(steel cylinders)
Gas Cylinder, 1000 ppm CO
2
in air,
Regulator Assembly, Flow Demand CO
2
(aluminum cylinders)
Gas Cylinder, 10 ppm Isobutylene in air
Consult ENMET or your distributer for more information
ENMET
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 30
Rev. 1 – December 15, 2016
GSM-60
Appendix A: CO Characteristics
ENMET
The Characteristics and Effects of Carbon Monoxide
Carbon monoxide is a colorless odorless toxic gas generated by incomplete combustion of a hydrocarbon fuel in air. It may be present where internal combustion engines, furnaces, boilers, and other combustion devices are present. It is toxic when inhaled because of its great affinity to hemoglobin, the oxygen carriers in the red cells of the blood. CO replaces the oxygen normally carried by the hemoglobin, and thus inhibits the delivery of oxygen throughout the body; the victim suffers from oxygen deficiency, and may die from asphyxiation. The symptoms and degree of danger resulting from exposure to CO depend upon the concentration of the gas and the length of exposure; this is shown in Figure 10 . The GSM-60 instrument is employed to warn the user of the presence of CO, and to facilitate the assessment of the degree of danger that he or she is exposed to.
Based upon knowledge of the effects of CO, the Occupational Safety and Health Authority (OSHA) has set limits on exposure to CO in the workplace. These are 35 ppm (parts CO per million parts air) as a time weighted average for an eight-hour day, and a maximum exposure of 200 ppm. For compressed air line applications, OSHA requires Grade D breathing air supplied, using a Compressed Gas
Association (CGA) definition (G-7.1). Depending on interpretation of the OSHA respiratory standard, 10 ppm and 20 ppm CO have been used as maximum limits and standard instrument alarm points.
If a CO sensor is installed, the GSM-60 has two preset alarm set points, at 10 ppm and 20 ppm CO, which are adjustable, but cannot be set below 5 ppm or above 50 ppm.
The curves below are for percent Carboxyhemoglobin with 50% being the top curve, 5% the bottom. %COHb is a measure of the amount of hemoglobin occupied by CO rather than oxygen. CO effects upon children, adults engaging in physical activity, and smokers, are more pronounced.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
Figure 10: Carbon Monoxide Concentration
P a g e | 31
Rev. 1 – December 15, 2016
GSM-60
Appendix B: Gas Ionization Potentials
Chemical Name
A
2-Amino pyridine
Acetaldehyde
Acetamide
Acetic acid
Acetic anhydride
Acetone
Acetonitrile
Acetophenone
Acetyl bromide
Acetyl chloride
Acetylene
Acrolein
Acrylamide
Acrylonitrile
Allyl alcohol
Allyl chloride
Ammonia
Aniline
Anisidine
Anisole
Arsine
B
1,3-Butadiene (butadiene)
1-Bromo-2-chloroethane
1-Bromo-2-methylpropane
1-Bromo-4-fluorobenzene
1-Bromobutane
1-Bromopentane
1-Bromopropane
1-Bromopropene
1-Butanethiol
1-Butene
1-Butyne
2,3-Butadione
2-Bromo-2-methylpropane
2-Bromobutane
2-Bromopropane
2-Bromothiophene
2-Butanone (MEK)
3-Bromopropene
3-Butene nitrile
Benzaldehyde
Benzene
Benzenethiol
Benzonitrile
Benzotrifluoride
Biphenyl
Boron oxide
Boron trifluoride
Bromine
Bromobenzene
Bromochloromethane
Bromoform
Butane
Butyl mercaptan cis-2-Butene m-Bromotoluene n-Butyl acetate n-Butyl alcohol n-Butyl amine n-Butyl benzene n-Butyl formate n-Butyraldehyde n-Butyric acid n-Butyronitrile o-Bromotoluene
IP (eV)
10.55
11.02
11.41
10.10
9.50
10.91
9.67
9.90
10.20
7.70
7.44
8.22
9.89
8.00
10.21
9.77
10.69
10.00
9.69
12.20
9.27
10.39
9.53
9.25
8.33
9.71
9.68
8.27
13.50
15.56
10.54
8.98
9.30
9.14
9.58
10.18
9.23
9.89
9.98
10.08
8.63
9.54
9.70
9.07
10.63
10.09
8.99
10.13
10.10
10.18
8.71
8.69
10.50
9.86
10.16
11.67
8.79
10.77
10.48
10.63
9.15
9.13
8.81
10.01
10.04
IP (eV)
8.67
10.66
9.21
10.67
10.82
10.61
10.65
10.78
8.68
10.04
8.76
13.79
10.07
14.01
11.47
8.28
8.70
8.68
9.91
8.64
8.68
9.13
8.95
7.99
8.56
10.53
9.26
9.01
10.06
8.83
8.83
8.70
12.45
11.37
12.91
7.59
8.14
9.73
8.75
13.80
9.80
9.75
9.14
11.48
10.36
12.65
10.61
9.44
9.07
10.77
10.07
10.85
10.06
10.35
9.82
10.02
9.17
9.65
10.19
11.12
9.65
7.28
9.45
12.20
11.12
10.87
Chemical Name p-Bromotoluene p-tert-Butyltoluene s-Butyl amine s-Butyl benzene sec-Butyl acetate t-Butyl amine t-Butyl benzene trans-2-Butene
C
1-Chloro-2-methylpropane
1-Chloro-3-fluorobenzene
1-Chlorobutane
1-Chloropropane
2-Chloro-2-methylpropane
2-Chlorobutane
2-Chloropropane
2-Chlorothiophene
3-Chloropropene
Camphor
Carbon dioxide
Carbon disulfide
Carbon monoxide
Carbon tetrachloride
Chlorine
Chlorine dioxide
Chlorine trifluoride
Chloroacetaldehyde
α -Chloroacetophenone
Chlorobenzene
Chlorobromomethane
Chlorofluoromethane (Freon 22)
Chloroform
Chlorotrifluoromethane (Freon 13)
Chrysene
Cresol
Crotonaldehyde
Cumene (isopropyl benzene)
Cyanogen
Cyclohexane
Cyclohexanol
Cyclohexanone
Cyclohexene
Cyclo-octatetraene
Cyclopentadiene
Cyclopentane
Cyclopentanone
Cyclopentene
Cyclopropane m-Chlorotoluene o-Chlorotoluene p-Chlorotoluene
D
1,1-Dibromoethane
1,1-Dichloroethane
1,1-Dimethoxyethane
1,1-Dimethylhydrazine
1,2-Dibromoethene
1,2-Dichloro-1,1,2,2-tetrafluoroethane (Freon 114)
1,2-Dichloroethane
1,2-Dichloropropane
1,3-Dibromopropane
1,3-Dichloropropane
2,2-Dimethyl butane
2,2-Dimethyl propane
2,3-Dichloropropene
2,3-Dimethyl butane
3,3-Dimethyl butanone cis-Dichloroethene
ENMET
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 32
Rev. 1 – December 15, 2016
GSM-60 o-Dichlorobenzene p-Dichlorobenzene p-Dioxane trans-Dichloroethene
E
Epichlorohydrin
Ethane
Ethanethiol (ethyl mercaptan)
Ethanolamine
Ethene
Ethyl acetate
Ethyl alcohol
Ethyl amine
Ethyl benzene
Ethyl bromide
Ethyl chloride (chloroethane)
Ethyl disulfide
Ethyl ether
Ethyl formate
Ethyl iodide
Ethyl isothiocyanate
Ethyl mercaptan
Ethyl methyl sulfide
Ethyl nitrate
Ethyl propionate
Ethyl thiocyanate
Ethylene chlorohydrin
Ethylene diamine
Ethylene dibromide
Ethylene dichloride
Ethylene oxide
Ethylenelmine
Ethynylbenzene
Chemical Name
(D continued)
Decaborane
Diazomethane
Diborane
Dibromochloromethane
Dibromodifluoromethane
Dibromomethane
Dibutylamine
Dichlorodifluoromethane (Freon 12)
Dichlorofluoromethane
Dichloromethane
Diethoxymethane
Diethyl amine
Diethyl ether
Diethyl ketone
Diethyl sulfide
Diethyl sulfite
Difluorodibromomethane
Dihydropyran
Diiodomethane
Diisopropylamine
Dimethoxymethane (methylal)
Dimethyl amine
Dimethyl ether
Dimethyl sulfide
Dimethylaniline
Dimethylformamide
Dimethylphthalate
Dinitrobenzene
Dioxane
Diphenyl
Dipropyl amine
Dipropyl sulfide
Durene m-Dichlorobenzene
N,N-Diethyl acetamide
N,N-Diethyl formamide
N,N-Dimethyl acetamide
N,N-Dimethyl formamide
IP (eV)
9.06
8.95
9.13
9.66
10.20
11.65
9.29
8.96
10.52
10.11
10.48
8.86
8.76
10.29
10.98
8.27
9.51
10.61
9.33
9.14
9.29
8.55
11.22
10.00
9.89
10.52
8.60
10.37
11.05
10.57
9.20
8.82
10.00
8.24
10.00
8.69
7.13
9.18
9.64
10.71
9.19
7.95
7.84
8.30
8.03
9.12
8.60
8.89
8.81
9.12
9.70
8.01
9.53
9.32
8.43
9.68
11.07
9.88
9.00
12.00
10.59
11.07
10.49
7.69
12.31
12.39
11.35
8.34
9.34
7.73
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
IP (eV)
9.18
9.21
9.19
9.26
9.09
9.17
9.28
8.73
10.57
9.97
10.12
8.70
10.46
9.74
10.02
10.32
9.07
8.85
9.99
10.16
8.72
8.69
9.20
9.71
8.61
8.62
8.50
9.61
8.85
8.85
8.85
9.46
9.33
9.35
10.08
11.10
10.18
8.10
15.43
11.62
12.74
13.91
15.77
10.38
9.88
10.46
9.14
7.95
12.31
12.91
12.45
8.89
9.21
8.92
8.66
9.21
15.70
9.20
10.87
10.25
11.05
11.77
11.30
11.78
12.20
8.92
8.79
2-Iodopropane
Iodine
Iodobenzene
Isobutane
Isobutyl acetate
Isobutyl alcohol
Isobutyl amine
Isobutyl formate
Isobutyraldehyde
Isobutyric acid
Isopentane
Isophorone
Isoprene
Isopropyl acetate
Isopropyl alcohol
Isopropyl amine
Isopropyl benzene
Isopropyl ether
Isovaleraldehyde m-Iodotoluene o-Iodotoluene p-Iodotoluene
K
Ketene
L
2,3-Lutidine
2,4-Lutidine
2,6-Lutidine
Chemical Name
F
2-Furaldehyde
Fluorine
Fluorobenzene
Formaldehyde
Formamide
Formic acid
Freon 11 (trichlorofluoromethane)
Freon 112 (1,1,2,2-tetrachloro-1,2-difluoroethane)
Freon 113 (1,1,2-trichloro-1,2,2-trifluororethane)
Freon 114 (1,2-dichloro-1,1,2,2-tetrafluoroethane)
Freon 12 (dichlorodifluoromethane)
Freon 13 (chlorotrifluoromethane)
Freon 22 (chlorofluoromethane)
Furan
Furfural m-Fluorotoluene o-Fluorophenol o-Fluorotoluene p-Fluorotoluene
H
1-Hexene
2-Heptanone
2-Hexanone
Heptane
Hexachloroethane
Hexane
Hydrazine
Hydrogen
Hydrogen bromide
Hydrogen chloride
Hydrogen cyanide
Hydrogen fluoride
Hydrogen iodide
Hydrogen selenide
Hydrogen sulfide
Hydrogen telluride
Hydroquinone
I
1-Iodo-2-methylpropane
1-Iodobutane
1-Iodopentane
1-Iodopropane
2-Iodobutane
ENMET
P a g e | 33
Rev. 1 – December 15, 2016
GSM-60
Chemical Name
M
2-Methyl furan
2-Methyl napthalene
1-Methyl napthalene
2-Methyl propene
2-Methyl-1-butene
2-Methylpentane
3-Methyl-1-butene
3-Methyl-2-butene
3-Methylpentane
4-Methylcyclohexene
Maleic anhydride
Mesityl oxide
Mesitylene
Methane
Methanethiol (methyl mercaptan)
Methyl acetate
Methyl acetylene
Methyl acrylate
Methyl alcohol
Methyl amine
Methyl bromide
Methyl butyl ketone
Methyl butyrate
Methyl cellosolve
Methyl chloride
Methyl chloroform (1,1,1-trichloroethane)
Methyl disulfide
Methyl ethyl ketone
Methyl formate
Methyl iodide
Methyl isobutyl ketone
Methyl isobutyrate
Methyl isocyanate
Methyl isopropyl ketone
Methyl isothiocyanate
Methyl mercaptan
Methyl methacrylate
Methyl propionate
Methyl propyl ketone
-Methyl styrene
Methyl thiocyanate
Methylal (dimethoxymethane)
Methylcyclohexane
Methylene chloride
Methyl-n-amyl ketone
Monomethyl aniline
Monomethyl hydrazine
Morpholine n-Methyl acetamide
N
1-Nitropropane
2-Nitropropane
Naphthalene
Nickel carbonyl
Nitric oxide, (NO)
Nitrobenzene
Nitroethane
Nitrogen
Nitrogen dioxide
Nitrogen trifluoride
Nitromethane
Nitrotoluene p-Nitrochloro benzene
O
Octane
Oxygen
Ozone
P
1-Pentene
1-Propanethiol
2,4-Pentanedione
IP (eV)
10.00
9.85
11.32
9.30
7.32
7.67
8.20
8.90
9.90
10.85
8.97
10.54
9.34
10.07
9.60
11.28
11.00
8.46
9.53
10.82
9.54
9.30
9.98
10.67
9.32
9.25
9.44
9.70
10.15
9.39
10.80
9.08
8.40
12.98
9.44
10.27
10.37
8.39
7.96
7.96
9.23
9.12
10.12
9.51
8.67
10.08
8.91
8.35
10.07
10.88
10.71
8.12
8.27
9.25
9.92
10.88
15.58
9.78
12.97
11.08
9.45
9.96
9.82
12.08
12.08
9.50
9.20
8.87
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
IP (eV)
9.51
8.47
12.30
15.33
9.66
13.00
7.78
11.30
9.91
10.00
11.07
10.51
9.70
9.98
10.24
11.84
10.04
10.20
8.78
8.72
9.27
10.54
9.73
10.87
9.00
10.22
10.36
9.32
8.20
8.82
7.64
8.77
8.52
6.89
11.77
9.87
9.38
9.02
9.02
9.04
11.07
10.40
10.35
9.32
8.18
8.50
10.04
11.00
11.78
9.86
7.44
11.62
9.32
11.47
9.54
9.25
10.00
8.86
8.82
9.27
10.67
10.51
9.45
9.47
11.77
11.42
7.50
11.40
Chemical Name
(P continued)
2-Pentanone
2-Picoline
3-Picoline
4-Picoline n-Propyl nitrate
Pentaborane
Pentane
Perchloroethylene
Pheneloic
Phenol
Phenyl ether (diphenyl oxide)
Phenyl hydrazine
Phenyl isocyanate
Phenyl isothiocyanate
Phenylene diamine
Phosgene
Phosphine
Phosphorus trichloride
Phthalic anhydride
Propane
Propargyl alcohol
Propiolactone
Propionaldehyde
Propionic acid
Propionitrile
Propyl acetate
Propyl alcohol
Propyl amine
Propyl benzene
Propyl ether
Propyl formate
Propylene
Propylene dichloride
Propylene imine
Propylene oxide
Propyne
Pyridine
Pyrrole
Q
Quinone
S
Stibine
Styrene
Sulfur dioxide
Sulfur hexafluoride
Sulfur monochloride
Sulfuryl fluoride
T o-Terphenyls
1,1,2,2-Tetrachloro-1,2-difluoroethane (Freon 112)
1,1,1-Trichloroethane
1,1,2-Trichloro-1,2,2-trifluoroethane (Freon 113)
2,2,4-Trimethyl pentane o-Toluidine
Tetrachloroethane
Tetrachloroethene
Tetrachloromethane
Tetrahydrofuran
Tetrahydropyran
Thiolacetic acid
Thiophene
Toluene
Tribromoethene
Tribromofluoromethane
Tribromomethane
Trichloroethene
Trichloroethylene
Trichlorofluoromethane (Freon 11)
Trichloromethane
Triethylamine
Trifluoromonobromo-methane
ENMET
P a g e | 34
Rev. 1 – December 15, 2016
GSM-60
Chemical Name
(T continued)
Trimethyl amine
Tripropyl amine
Notes:
V o-Vinyl toluene
Valeraldehyde
Valeric acid
Vinyl acetate
Vinyl bromide
Vinyl chloride
Vinyl methyl ether
IP (eV)
7.82
7.23
8.20
9.82
10.12
9.19
9.80
10.00
8.93
Chemical Name
W
Water
X
2,4-Xylidine m-Xylene o-Xylene p-Xylene
IP (eV)
12.59
7.65
8.56
8.56
8.45
ENMET
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 35
Rev. 1 – December 15, 2016
GSM-60
8.0 Terms and Conditions
ENMET
8.1 Ordering Information
Address orders to:
ENMET
Attention: Customer Service Department
680 Fairfield Court
Ann Arbor, MI 48108
Email Orders: [email protected]
Phone: 734-761-1270 Fax: 734-761-3220
You may also contact our customer service department by email [email protected]. MINIMUM ORDER IS $50.00.
8.2 Delivery
Unless Seller otherwise specifies, delivery will be made: FOB Ann Arbor, MI and/or FOB Bowling Green, KY. Title and risk of loss shall pass to Buyer at that point. Shipping and handling charges will be Prepaid and Added to Buyer’s invoice. Buyer may request shipping be charged to their own account with a preferred carrier. Seller shall have the right to choose means of transportation and to route shipment when specific instructions are not included with Buyer’s order. Seller agrees to deliver the goods and services, within the time, in accordance with specifications, at the prices specified on the face hereof. Buyer’s orders to this quotation are not subject to cancellation or deferment of delivery without indemnification of loss to the Seller resulting there from. Seller shall not be liable to Buyer for any loss or damage sustained on account of this delay or nonperformance due to causes beyond Seller’s control and without his fault or negligence. Where performance of the terms here is contingent upon timely delivery of goods or services by the Buyer and such delivery is in default, Seller shall be indemnified for any damage or loss resulting there from, and/or by extension of Seller’s delivery commitment, as applicable.
8.3 Payment Terms
Payment Terms are Net 30 Days from the date of shipment from Seller unless otherwise noted. All shipping and handling costs will be charged to Buyer on a Prepaid and Add basis. Buyer has the option of paying for shipping by charging its own account with a carrier
8.4 Warranty Information and Guidelines
The Seller warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment. The warrant covers both parts and labor excluding calibration and expendable parts such as filters, detector tubes, batteries, etc. If the inspection by the Seller confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States. The Seller shall not be liable for any loss or damage caused by the improper use or installation of the product. The Buyer indemnifies and saves harmless the Seller with respect to any loss or damages that may arise through the use by the Buyer or others of this equipment. This warranty is expressly given in lieu of all other warranties, either expressed, implied or statutory, including that of merchantability, and all other obligations, or liabilities of ENMET, LLC for damages arising out of or in connection with the use or repair or performance of the product. In no event shall ENMET, LLC, be liable for any indirect, incidental, special or consequential damages or for any delay in the performance by ENMET, LLC, which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representatives or other persons to assume for it any obligation or liability other than that which is set forth herein.
Buyer agrees to indemnify and save harmless Seller for any damage or loss from lawsuits against Seller by reason of manufacture of sale of materials, parts, or use of processes resulting from Buyer’s design specifications. Any patent, design, pattern, tool, die, jig, fixture, drawing, test equipment, or process furnished by Seller; whether possessed by the Seller before the date of this quotation, or devised or acquired by Seller during performance of the terms of this quotation, shall remain the property of the
Seller except by specific stipulation on the face hereof. Seller reserves the right, without liability, for damage or loss, to destroy
Buyer’s drawings, specifications, patterns and special tools supplied by Buyer for performance of the terms on the face hereof, unless Buyer gives notice of the disposition of such items.
8.5 Return Policy
All returns for credit must be approved in advance by ENMET, LLC. Such returns are subject to a minimum $50.00 or 20% restocking charge, whichever is greater. Approval of equipment for return is totally at the discretion of ENMET, LLC. All requests for return/exchange must be made no later 30 days of the original shipping date from ENMET. The actual amount of any resulting credit will not be determined prior to a complete inspection of the equipment by ENMET. Calibration gas cylinders cannot be returned or restocked.
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 36
Rev. 1 – December 15, 2016
GSM-60
9.0 Instructions for Returning an Instrument for Service
ENMET
Contact the ENMET Service Department for all service requests.
Phone: 734-761-1270
Email: [email protected]
Fill out the “Service Request Form” found at the end of this manual and return with your instrument for all needs. Please send your instrument for service to the site in which the product was purchased. A new “Service Request Form” may be requested if the one found in the manual is not available. All instruments should be shipped prepaid to ENMET.
Address for Service:
Michigan Location:
ENMET
Attention: Service Department
680 Fairfield Court
Ann Arbor, MI 48108
Kentucky Location:
ENMET
62 Corporate Court
Bowling Green, KY 42103
Providing the “Service Request Form” assists in the expedient service and return of your unit and failure to provide this information can result in processing delays. ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered after a period of 60 days may, be disposed of or returned unrepaired COD and the customer will be expected to pay the evaluation fee. Serviced instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
NOTE : Warranty of customer installed components.
For Warranty Repairs, please reference ENMET’s “Warranty Information and Guidelines” (found earlier in this section).
Manual Revision Date – June 15, 2017
Manual Part No. – 80003-600
P a g e | 37
Rev. 1 – December 15, 2016
Mailing/Shipping Address:
ENMET
680 Fairfield Court
Ann Arbor, MI 48108
[email protected]
Phone: 734.761.1270
Fax: 734.761.3220
Product Name or Number:
Product Serial Number:
Describe Problem or Needed Service:
Service Request Form
Warranty Claim?
☐
Yes
☐
No
Billing Address:
Contact Name:
Email:
PO/Reference
#:
CUSTOMER INFORMATION
Shipping Address:
Phone #:
Fax #:
☐
COD
Card Number
Name as it Appears on
Card:
☐
PAYMENT METHOD
VISA/MasterCard
Exp. Date
☐
American Express
Security Code:
☐
UPS Ground
RETURN SHIPPING METHOD
☐
UPS 3 Day
☐
UPS Next Day
☐
UPS ND Air
Select Air Saver
UPS Account #:
☐
FedEx Ground
☐
FedEx Air
Express Saver
FedEx Account #:
Insure Shipment:
☐
Yes
☐
No
☐
FedEx Air
Overnight Std.
☐
FedEx Air 2
Day
Insurance
Amount:
$
☐
UPS 2 Day Air
☐
FedEx Air
Overnight P-1
ENMET
Service Request Form
Rev.2 – 9/15/2016
advertisement