ENMET GSM-60 Manual

Add to My manuals
39 Pages

advertisement

ENMET GSM-60 Manual | Manualzz

ENMET

680 Fairfield Court

Ann Arbor, MI 48108

734.761.1270 Fax 734.761.3220 www.enmet.com

GSM-60

Operation Manual

GSM-60 ENMET

Table of Contents

1.0

I NTRODUCTION .................................................................................................................................................................................... 3

1.1 Unpack ............................................................................................................................................................................................. 3

1.2 Check Order ..................................................................................................................................................................................... 3

1.3 Serial Numbers................................................................................................................................................................................. 3

2.0

I NSTRUMENT F EATURES ..................................................................................................................................................................... 4

2.1 Exterior Features .............................................................................................................................................................................. 4

2.2 Display Panel Features ..................................................................................................................................................................... 4

2.3 Circuit Board Features ..................................................................................................................................................................... 6

3.0

I NSTALLATION ..................................................................................................................................................................................... 7

3.1 Mounting of Instrument ................................................................................................................................................................... 7

3.2 Power Supply ................................................................................................................................................................................... 8

3.3 Inputs / Outputs ................................................................................................................................................................................ 8

3.3.1 Sensor/Transmitter Connection ...................................................................................................................................................................... 9

3.3.2 Relay Contacts ................................................................................................................................................................................................ 9

3.3.3 Optional 4-20mA Outputs ............................................................................................................................................................................. 10

3.4 Installation Verification ................................................................................................................................................................. 10

4.0

O PERATION ........................................................................................................................................................................................ 11

4.1 Normal Operation Condition ......................................................................................................................................................... 11

4.2 Alarm Set Points ............................................................................................................................................................................ 11

4.3 Alarm Latching or Differential Settings ........................................................................................................................................ 12

4.4 Audio Defeat .................................................................................................................................................................................. 12

4.5 Display ........................................................................................................................................................................................... 12

4.6 Operational Menu .......................................................................................................................................................................... 13

4.7 Fault Indications ............................................................................................................................................................................. 14

4.7.1 Low Flow Indication ..................................................................................................................................................................................... 14

4.7.2 Other Fault Indications ................................................................................................................................................................................. 14

4.8 Hydrocarbon Sensor Response ...................................................................................................................................................... 14

5.0

M AINTENANCE .................................................................................................................................................................................. 15

5.1 Cleaning Instructions ..................................................................................................................................................................... 15

5.2 Maintenance Menu ......................................................................................................................................................................... 15

5.2.1 Accessing Maintenance Menu ....................................................................................................................................................................... 15

5.2.2 Maintenance Menu Flow Chart .................................................................................................................................................................... 16

5.3 Calibration for CO, O

2,

HC and CO

2

(Gas Channels) ..................................................................................................................... 17

5.3.1A Low Cal/Zero Cal Adjust ............................................................................................................................................................................ 18

5.3.1B High Cal/Span Gas Adjust .......................................................................................................................................................................... 19

5.3.2 Set 4 –20mA Transmitter Scale ..................................................................................................................................................................... 19

5.3.3 Set Alarm Points ........................................................................................................................................................................................... 20

5.3.4 Set Alarm Delay ............................................................................................................................................................................................ 22

5.3.5 Relay Configuration ...................................................................................................................................................................................... 23

5.3.6 Failsafe Configuration .................................................................................................................................................................................. 24

5.3.7 Set Output Span Range ................................................................................................................................................................................. 24

5.3.8 Set New Password ......................................................................................................................................................................................... 25

5.3.9 Exit Maintenance Menu ................................................................................................................................................................................ 25

5.4 Sensor Replacement ....................................................................................................................................................................... 26

5.4.1 Gas/Oxygen Sensor ....................................................................................................................................................................................... 26

5.4.2 Calibration/Sensor Replacement .................................................................................................................................................................. 26

5.4.3A Low Cal/Zero Cal Adjust ............................................................................................................................................................................ 28

5.4.3B High Cal/Span Gas Adjust .......................................................................................................................................................................... 29

5.5 Flow Control Orifice ...................................................................................................................................................................... 29

6.0

T ECHNICAL D ATA AND S PECIFICATIONS ......................................................................................................................................... 29

7.0

R EPLACEMENT P ART N UMBERS ....................................................................................................................................................... 30

7.1 ENMET part numbers for sensors and replacement parts: ............................................................................................................. 30

7.2 ENMET part numbers for Calibration equipment:......................................................................................................................... 30

A PPENDIX A: CO C HARACTERISTICS .................................................................................................................................................... 31

A PPENDIX B: G AS I ONIZATION P OTENTIALS ......................................................................................................................................... 32

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 1

Rev. 1 – December 15, 2016

GSM-60 ENMET

List of Figures

Figure 1: External Features of the GSM-60 ....................................................................................................................... 5

Figure 2: GSM-60 Interior Features ................................................................................................................................... 6

Figure 3: GSM-60 Mounting Dimensions ......................................................................................................................... 7

Figure 2A: Relay, Input and Output Terminals .................................................................................................................. 8

Figure 4: GSM-60 Operational Display ........................................................................................................................... 11

Figure 5: GSM-60 Operation Menu Flow Chart .............................................................................................................. 13

Figure 6: GSM-60 Maintenance Menu Flow Chart. ........................................................................................................ 16

Figure 7: GSM-60 Calibration Connections .................................................................................................................... 18

Figure 9: Location of Gas Sensor and PiD(HC) Manifolds ............................................................................................. 26

Figure 7: GSM-60 Calibration Connections .................................................................................................................... 28

Figure 10: Carbon Monoxide Concentration.................................................................................................................... 31

List of Tables

Table 1 : Relay Failsafe Settings ...................................................................................................................................... 10

Table 2: Sensor Output .................................................................................................................................................... 10

Table 3: Typical Factory Alarm Set Points ...................................................................................................................... 11

Table 4: Factory Set Gas alarms Delay ............................................................................................................................ 22

Reference Information:

N OTE : [important information about use of instrument]

C AUTION : [affects equipment – if not followed may cause damage to instrument, sensor etc.…]

W ARNING : [affects personnel safety – if not followed may cause bodily injury or death.]

Attention / Warning

!

Earth Ground

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 2

Rev. 1 – December 15, 2016

GSM-60 ENMET

1.0 Introduction

The GSM-60 is a sample draw monitoring instrument that measures and detects certain toxic hazards gases. The GSM-60 was designed primarily for monitoring one area with a single sampling hose, either for an individual gas or multiple gases. The instrument has one sampling pump and one flow sensor. The instrument is available with up to 4 internal sensors. Available sensors include, but are not, limited to carbon monoxide (CO), carbon dioxide (CO

2

), Hydrocarbons (HC) and variations in the oxygen (O

2

) content. The sensors can be used alone or up to four sensors can be used together. Some applications require monitoring of more than one area. If two sampling hoses are connected to one GSM-60 the flow sensor can only detect a total flow fault condition. For example, a pump failure or both sampling hoses being blocked. The system cannot detect a flow fault if only one of the two sampling hoses is blocked.

Flowmeters should be installed in both sampling lines to provide a visual indication of proper flow. In the instrument, a 24 V DC sampling pump pass air over each sensor and the resulting electrical outputs are used to evaluate the air for the target gases.

The GSM-60 is a highly adaptable instrument. Some adaptations require an addendum be added to the manual to facilitate use of the instruments with these adaptations. If addendum is needed, see page(s) between sections replacement part numbers and warranty.

Some features of the instruments are as follows:

 continuous monitoring of the sample air

 continuous LCD display of gas and vapor concentrations

 menu driven operational and maintenance controls

 menu driven calibration procedure

 audio and visual alarms indicate unsafe conditions

 alarm relay contacts available on terminals

 a fault relay and visual fault alarm

 low air flow fault indication and display

 alarm acknowledgement capability including audio defeat mA outputs for each target gas

Hydrocarbons (HC) are limited to gases with an ionization potential of 10.6 eV or less. See Appendix B for a list of gases and IPs.

N OTE : All specifications stated in this manual may change without notice.

1.1 Unpack

Unpack the GSM-60 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.

Regarding Damaged Shipments

N OTE : It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor any claims for damage.

This shipment was carefully inspected, verified and properly packaged at ENMET and delivered to the carrier in good condition.

When it was picked up by the carrier at ENMET , it legally became your company’s property.

If your shipment arrives damaged: o Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and request immediate inspection of the carton and the contents. o After the inspection and after you have received written acknowledgment of the damage from the carrier, contact ENMET Customer

Service for return authorization and further instructions. Please have your Purchase Order and Sales Order numbers available.

ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00.

Repair or replacement charges above that value are your company’s responsibility.

The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.

Regarding Shortages

If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:

ENMET

680 Fairfield Court

Ann Arbor, MI 48108

734-761-1270 Fax 734-761-3220

Toll Free: 800-521-2978

1.2 Check Order

Check, the contents of the shipment against the purchase order. Verify that the GSM-60 is received as ordered. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.

1.3 Serial Numbers

Each GSM-60 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 3

Rev. 1 – December 15, 2016

GSM-60

2.0 Instrument Features

2.1 Exterior Features

The exterior of the instrument is shown in Figure 1 . The exterior features are as follows:

Feature Description

ENMET

Enclosure

Input Port

Visual

Alarms & Indicators

Pushbutton Switches

O

S

PTION

ELECT

Switch

Switch

Audio Defeat / Alarm

Acknowledge

Switch

An engineered thermoplastic box, approximately 10x8x6, with a clear hinged front cover.

The entrance for the air sample and calibration gas. The quick release fitting mates with one on the calibration adapter.

Front Cover Latch

Audio Alarm

Mounting Flanges

Output Port

A quick-release latch that holds the clear front cover in place, and is capable of being padlocked if desired.

A loud horn activated by certain alarm conditions.

Flanges with holes for mounting the enclosure to a vertical surface.

The exit to exhaust the air sample and calibration gas. Luer fitting.

2.2 Display Panel Features

The display panel, shown in Figure 1 , is viewed through the clear front cover of the enclosure, and is accessed by opening the cover. Features are as follows:

Feature Description

Display A 2 line, 16 character per line, LCD with backlight.

The numerical values of gas concentrations, and other information are displayed.

On either side of the display:

A red alarm LED for each sensor installed in the instrument, Low level alarm.

The top center of the panel:

A red alarm LED for all sensors installed in the instrument, High level alarm.

Near the center of the panel:

A green power indicator LED

A red fault alarm indicator LED

There are three of these, located near the center of the panel; they are yellow rectangular membrane switches. They are:

The top left switch.

Directly to the right of the OPTION switch.

Directly below the OPTION switch.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 4

Rev. 1 – December 15, 2016

GSM-60 ENMET

*Visual Alarm CO

(See *note below)

Visual Alarm 2

Display, see *note below

CO reading O

2

reading

H2S reading

*Visual Alarm O2

(If Oxygen option is installed)

See *note below

*Visual Alarm

Dew Point

(If Dew Point option is installed)

See *note below

OPTION

Switch

A LARM

A CKNOWLEDGE /

A LARM D EFEAT

Switch

CO ppm

H2S

ALARM 2

Channel #1 Channel #3

Channel #2 Channel #4

OPTION SELECT

ALARM

ACKNOWLEDGE

POWER

FAULT

O

2

%

Quick Release Fitting

Input Port

Sample/Calibration

SELECT

Switch

Power LED

Front Cover

Fault LED

Latch

Audio Alarm

Mounting Flanges

2 places

Sample Output Port

Figure 1: External Features of the GSM-60

* N OTE : Typical gas reading & alarm locations, depending on instrument configuration, alarms & readings may be in alternate locations

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 5

Rev. 1 – December 15, 2016

GSM-60 ENMET

2.3 Circuit Board Features

The Display Panel is hinged on the left and is released by unscrewing the 2 screws located in the right corners. After releasing the panel, it is swung to the left, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the enclosure interior. Features are shown in Figure 2 .

Feature Description

Relay Terminals

Output Terminals

HC Manifold

Sensor Manifold

Filter, Particulate

Sensor Terminals

J16, J18, J19

This group of terminals is located at the left side of the Circuit Board.

For the contacts for each of four alarm relays, and for the contacts of a fault relay.

One 4-20mA output per active channel. 2 channels/outputs per connector.

The PiD sensor is installed into this housing.

The sample manifold, the carbon monoxide, carbon dioxide and oxygen sensors are located under this housing.

Removes contaminate from air sample line.

Sensor/Transmitter connectors 24V DC 4-20mA Input

Figure 2: GSM-60 Interior Features

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 6

Rev. 1 – December 15, 2016

GSM-60

3.0 Installation

ENMET

3.1 Mounting of Instrument

The GSM-60 should be located near the air to be monitored. Sampling lines should be no more than 50 feet long. It is recommended that Teflon® (PTFE) lined tubing be used. Quick disconnect fittings are supplied for use with 1/8” ID tubing.

Mount the instrument on an appropriate vertical surface using the mounting flanges provided. Avoid areas with excessive vibration or temperature extremes. The holes in the flanges are 0.31 inch in diameter and form a 6 x 10.75-inch rectangle. See Figure 3.

It is recommended to use #8 drywall anchors and screws for mounting the GSM-60 to a drywall/sheetrock surface.

Mounting Holes

0.31” dia. 4 places

Dimensions are in inches.

Figure 3: GSM-60 Mounting Dimensions

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 7

Rev. 1 – December 15, 2016

GSM-60 ENMET

3.2 Power Supply

The input power can vary from 100 to 240V AC , 50/60 Hz. Mains power should be connected to the Power Input Terminal J23 and the ground screw J21 . Se e Figure 2 for location.

W ARNING : Continuous gas detection and alarm systems (110V AC /220V AC / 24V DC /12V DC powered) become inoperative upon loss of primary power. Contact factory for specifications and pricing of backup battery systems.

Upon supplying air and power to the instrument:

 The green power on LED is lit.

The display backlight is lit, and instrument will step through a start-up sequence: unit serial number, software revision and gases monitored may be shown on the display.

The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch. If alarm persists longer than 30 minutes, call ENMET customer service personnel.

3.3 Inputs / Outputs

Two types of alarm outputs are available, relay contacts and 4-20mA outputs.

Connector 2

Channel 3 & 4

4-20mA Output

Relay 1

Channel 1

Alarm 1

Relay 2

Channel 2

Alarm 1

Relay 3

Channel 3

Alarm 1

Relay 4

Channel 4

Alarm 1

Relay 5

Channel 1-4

Alarm 2

Relay 6

Ch 1-4 / System

Fault

Connector 1

Channel 1 & 2

4-20mA Output

Connector RS485

Connector RS232

Connector RS485

Input Connector

24V DC

GND mA

Input Connector

24V DC

GND mA

Input Connector

24V DC

GND mA

Ground Screw J21

Power Input

Terminal J23

Figure 2A: Relay, Input and Output Terminals

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 8

Rev. 1 – December 15, 2016

GSM-60 ENMET

3.3.1 Sensor/Transmitter Connection

Sensor/Transmitters are connected to the GSM-60 control unit with two or three-conductor wiring, use the correct oil tight fitting. Size of wire depends on the distance between the sensor/transmitter and the control unit.

See Recommended Wire Gauge Table below.

2 Wire for Sensors/Transmitter 3 Wire for Sensors/Transmitter

Position

1

2

3

Function

Power +24 V DC

Not Used

Signal/Return to Ground

Position

1

2

3

Function

Power +24 V DC

Power Ground

Signal

Recommended Wire Gauge

Distance from Sensor to Control Unit Recommended Wire Gauge

< 500 feet

501 – 800 feet

Longer Distances

16 AWG

14 AWG

Contact Factory

N OTE : Sensor Location

Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some examples are given below.

Heavier than Air Gas Sensor Location

Bottled LP (liquefied petroleum)

Propane

Butane

Gasoline

Trichloroethylene

Vaporized hydrocarbons

Hydrogen sulfide

Lighter than Air Gas

Natural gas (methane)

Ammonia

Hydrogen

Interior wall; 18-24" from floor.

D O N OT locate directly above or beside gas appliances (ovens, heaters).

Avoid locating anywhere near a vent or window or near an outside doorway.

Sensor Location

Near ceiling.

D O N OT locate directly above appliances where it is subject to direct exposure to heat or steam.

Sensor Location Same Density as Air Gas

Carbon Monoxide 4-6 feet above the (generally uniform) floor.

D O N OT locate in direct air currents of windows, doors, or vents.

If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.

3.3.2 Relay Contacts

Relay contacts are available for each alarm; these are SPDT, rated at 10Amp at 110V AC , and may be latching or non-latching as required by the application.

They are accessed on the terminals next to each relay see Figure 2 & 2A . The contact positions are noted on the circuit board next to each terminal.

Relays may also be configured as failsafe or non-failsafe. The default alarm relay configuration is for latching mode, and failsafe. They may be reconfigured in the maintenance menu. See section 5.3.5 & 5.3.6

The PC Board is labeled for the relays in their un-energized state. If the relay is configured for failsafe, then this is also the alarm condition state. Non-failsafe configured relays in the alarm state, are the reverse of the PC board labeling. Note that the

Fault(FLT) relay cannot be set to operate in a Non-Failsafe mode. Please see the Table 1 on page 10:

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 9

Rev. 1 – December 15, 2016

GSM-60 ENMET

Table 1 : Relay Failsafe Settings

Position

J5 Relay 1 - NO

J5 Relay 1 - COM

J5 Relay 1 - NC

J6 Relay 2 - NO

J6 Relay 2 - COM

J6 Relay 2 - NC

J8 Relay 3 - NO

J8 Relay 3 - COM

J8 Relay 3 - NC

J10 Relay 4 - NO

J10 Relay 4 - COM

J10 Relay 4 - NC

J14 Relay 5 - NO

J14 Relay 5 - COM

J14 Relay 5 - NC

J15 Relay 6/FLT - NO

J15 Relay 6/FLT - COM

J15 Relay 6/FLT - NC

Failsafe-Alarm

Normally Open

Common

Normally Closed

Normally Open

Common

Normally Closed

Normally Open

Common

Normally Closed

Normally Open

Common

Normally Closed

Normally Open

Common

Normally Closed

Normally Open

Common

Normally Closed

Non-Failsafe-Alarm

Normally Closed

Common

Normally Open

Normally Closed

Common

Normally Open

Normally Closed

Common

Normally Open

Normally Closed

Common

Normally Open

Normally Closed

Common

Normally Open

N/A

N/A

N/A

Relays can be linked to specific alarms. The table below shows the default relay links. They may be changed in the maintenance menu if required. See Section 5.0

.

Relay 1

Relay 2

Relay 3

Relay 4

Relay 5

Channel 1

Low Alarm

High Alarm

Channel 2

Low Alarm

High Alarm

Channel 3

Low Alarm

High Alarm

Channel 4

Low Alarm

High Alarm

In addition, there is a fault relay, which changes state whenever the instrument is in a fault condition. The contact positions are noted on the circuit board next to each terminal. See Figure 2A.

The coil of this relay is energized when the instrument is in the non-fault state; the contact conditions given on the circuit board next to the terminal, are for the non-energized state, which is identical to the fault state.

These relay contacts can be used to operate auxiliary alarms or other functions. It is recommended that power for auxiliary equipment be supplied from an independent power source, separate from the GSM-60 . Place a hole in the enclosure for a wire exit, and use appropriate cable fittings. Be sure to note the location and depth of hardware inside the enclosure.

3.3.3 Optional 4-20mA Outputs

Isolated 4-20 mA outputs are available for data logging or other purposes. An output is supplied for each sensor supplied in an instrument, and can be added when a sensor is added in the field. These outputs are available on the Connector 1 for channels 1

& 2 and Connector 2 for channels 3 & 4.

4mA corresponds to a sensor reading at the bottom of the instrument range and 20mA corresponds to a full-scale reading.

Standard ranges are shown in Table 2 .

Table 2: Sensor Output

Sensor 4mA 20mA

CO

O2

CO2

HC

0

0

0

0

50

30

5000

100

Wiring requirements are the same as for the relays.

3.4 Installation Verification

All instruments are calibrated at the factory. You may, if a calibration kit is available, calibrate the all gas channels of the instrument 24 hours after installation to verify proper installation and instrument operation. See Section 5.0

, Maintenance, for calibration instructions. Calibration is also recommended after the first month of operation. Subsequent calibrations should be performed every 3 months. The dew point sensor cannot be calibrated in the field.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 10

Rev. 1 – December 15, 2016

GSM-60

4.0 Operation

ENMET

4.1 Normal Operation Condition

With the GSM-60 installed as described in Section 3 , and in clean air, the POWER green LED is on, the display is lit and the information on the display is as shown in Figure 4 Display , for the sensor(s) installed in the GSM-60 .

The red alarm and fault

LEDs are not lit.

CO

PPM

O

2

%

H2S

PPM

CO 0

H2S 0

O2 20.9

CO2 300 CO2

PPM

Example of display with CO (ch 1), H

2

S (ch 2), Oxygen (ch 3) and CO

2

(ch 4) options installed

Figure 4: GSM-60 Operational Display

4.2 Alarm Set Points

There are two alarm set points for each installed channel of the GSM-60 . The factory settings of these alarm set points are shown in Table 3 .

Typical Channel # Gas

Table 3: Typical Factory Alarm Set Points

Alarm 1, Flashing LED Alarm 2, Steady LED

1

2

3

4

4

Carbon Monoxide

Hydrogen Sulfide

Oxygen Deficiency

Carbon Dioxide

Hydrocarbon

10 ppm

10 ppm

19.5 % by volume

500 ppm

5 ppm

20 ppm

20 ppm

23.5 % by volume

1000 ppm

10 ppm

The alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.

 If the CO concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.

 If the dew point increases above that of the alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.

 If the oxygen content of the sample air decreases below the deficiency alarm set point, the associated red LED is lit, the associated relay changes state, and the audio alarm is activated.

 If the oxygen content of the sample air exceeds that of the abundance alarm set point, the associated red LED is lit, the audio alarm is activated, and both the oxygen alarm relay and the oxygen high alarm relay change state.

 The HC sensor can only detect and alarm to hydrocarbons with an Ionization Potential of less than 10.6 eV. See Appendix

B .

The HC sensor is broad range in nature and is unable to differentiate between different hydrocarbons.

 The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value. Example : If the alarm set point is

Λ

10 and the differential is 2, the

GSM-60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 11

Rev. 1 – December 15, 2016

GSM-60 ENMET

4.3 Alarm Latching or Differential Settings

An instrument is shipped with the alarms in the latching mode. The alarms may be independently configured in the non-latching mode or differential setting by use of the maintenance menu.

See Section 5.3.3

, for setting alarm 1 and alarm 2.

Standard Setting

 I N THE LATCHING MODE : at the cessation of the condition which causes an alarm, the alarm indications do not cease, and the alarm relay contacts do not revert to the non-alarm state, until the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch is pressed.

An alarm can also be acknowledged by pressing the switch during the alarm condition; then at the cessation of the alarm condition, alarm indications cease and alarm relays revert to the non-alarm state. After an alarm is acknowledged, alarms in the latching configuration are re-armed to latch at the next alarm condition.

 I N THE NON LATCHING MODE : at the cessation of the condition that causes an alarm, the alarm indications automatically cease, and the alarm relay contacts revert to the non-alarm state.

Differential Setting

 The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value. Example : If the alarm point is

Λ

10 and the differential is 2, the GSM-

60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.

4.4 Audio Defeat

Pressing the AUDIO DEFEAT / ALARM ACKNOWLEDGE switch during an alarm temporarily silences the audio alarm. Relays and alarm LEDs continue to function, in the alarm state, during an alarm condition. If the alarm condition persists, the audio alarm will

“chirp” every 20 seconds.

If after 15 minutes the alarm condition continues the audio alarm will reactivate at full intensity.

 If any other alarm condition occurs while the audio alarm has been silenced it will force the audio alarm to reactivate immediately.

4.5 Display

In clean air a display is shown in Figure 4 .

This position of the display is termed the " operational display ". As explained below, the display can be used to view other information by using the OPTION and SELECT switches.

Concentrations of CO and CO2 are given in PPM (parts per million parts of air). Dew point is given in degrees Fahrenheit at 55

PSIG ; this can be changed to degrees Centigrade by pressing the SELECT switch . Oxygen concentration is given in percent by volume.

When sample flow is reduced below a limit, the bottom line of the display flashes “Low Flow Alarm”.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 12

Rev. 1 – December 15, 2016

GSM-60 ENMET

4.6 Operational Menu

The operational menu allows the user to:

 View alarm set point concentration values

 View alarm ascending/descending trigger, latching and delay configurations

 Enter the maintenance menu with the proper Password.

The operational menu is accessed with the OPTION and SELECT switches. The operational menu flow chart is shown in Figure 5 ,

 Pressing the OPTION switch is indicated with a " O "

 Pressing the SELECT switch is indicated with a " S ".

If the instrument is left at any location in the operational or maintenance menus, other than the operational display, with no action taken for a period of 45 seconds, it returns to the operational display.

CH-1 CH-3

CH-2 CH-4

O

S

No

Function

O = Press Option switch

S = Press Select switch

O

Relays 1 2 3 4 5 6

[ =ON [[[[[[

O

CH1 SCALE (CO)

0 – 50

PPM

O

Λ L10 A1 vL19.5

vL-40 Λ D 500

O

Λ L20 A2 Λ 23.5

Λ L39 Λ 1000

O

ALARM1 Delays

(Seconds)

O

50 mASPAN 30.0

68 5000

O

Enter Maint Menu

S

S

O

O

S

S

O

S

S

S

CH3 SCALE

(O2)

S

CH4 SCALE (CO2)

0 – 5000

PPM

No

Function for each active channel

CH2 SCALE (H2S)

0 – 50 PPM

S

Alternating

No

Function

No

Function

5

5

No

Function

_

Enter Password

5

5

S

Displays are examples of gases:

 Channel 1 = Carbon Monoxide

 Channel 2 = Hydrogen Sulfide

Channel 3 = Oxygen

 Channel 4 = Carbon Dioxide

Displays are examples of Alarms

Λ

- Indicates alarm triggered on v - Indicates alarm triggered on decreasing value of reading

Displays are examples of Alarms

L – Indicates alarm is in latching

(no L present) – Indicates alarm is in non-latching mode.

Displays are examples of Alarms

D – Indicates alarm is in

(no D present) – Indicates alarm is in Standard Setting.

No increasing value of reading mode.

Differential Setting.

Function

See Maintenance Menu Diagram

Figure 5: GSM-60 Operation Menu Flow Chart

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 13

Rev. 1 – December 15, 2016

GSM-60 ENMET

4.7 Fault Indications

4.7.1 Low Flow Indication

A flow sensor is used to furnish a low flow indication. When the sample air pressure drops below preset levels, the fault light and audio alarm are activated, and the display flashes “Low Flow Alarm”. If two sampling hoses are connected to one GSM-

60 , the flow sensor can only detect a total flow fault condition. For example: a pump failure or both sampling hoses being blocked. The system cannot detect a flow fault if only one of the two sampling hoses is blocked.

4.7.2 Other Fault Indications

Other fault indications are associated with sensor zero and calibration activities, and are described in the maintenance Section

5.0

of this manual.

4.8 Hydrocarbon Sensor Response

If a Hydrocarbon (HC) sensor is supplied with the GSM-60 instrument, it designed to detect hydrocarbon gases and vapors with an ionization potential (IP) of 10.6 eV or less. Hydrocarbons with an IP of greater than 10.6 eV will NOT be detected. Please see

Appendix B for a list of common gases and vapors and their respective IP rating.

Unless otherwise noted Isobutylene is used as a calibration and reference gas.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 14

Rev. 1 – December 15, 2016

!

GSM-60

5.0 Maintenance

ENMET

The GSM-60 requires periodic sensor calibration and replacement. Calibration of toxic gas and oxygen sensor should be performed immediately following installation, one month after installation and every 3 months thereafter. HC sensor should be calibrated monthly. Oxygen and CO sensor have an estimated lifetime of 1 – 2 years. The CO2 sensor has an estimated lifetime of 3 years.

Other sensors vary. Sensors should be replaced when they will not calibrate or shortly before the end of the estimated lifetime.

5.1 Cleaning Instructions

C AUTION : Never spray a cleaning solution on the surfaces of the GSM-60 devices.

Clean the exterior of the GSM-60 enclosures with a mild soap solution on a clean, damp cloth. Do not soak the cloth with solution so that moisture drips onto, or lingers on, external surfaces.

Under no circumstances should organic solvents such as paint thinner be used to clean instrument surfaces.

5.2 Maintenance Menu

5.2.1 Accessing Maintenance Menu

The GSM-60 maintenance menu is accessed by entering the proper password with the OPTION and SELECT switches. See

Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.

Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password is 1270.

When a valid numerical password is inserted, the user can enter the maintenance menu.

To enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.

In the "Enter Maint Menu" position

 Press the SELECT switch "Enter Password

ζ

0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.

In the

ζ

000 position, the underline cursor is under the left digit.

 Press the OPTION switch to change the left digit; select the correct digit.

 Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint

Menu display.

CO

PPM

H2S

PPM

Enter Password

1270

O

2

%

Example : Password Display ( with factory installed password entered ) and Flow Chart below.

O

O (6)

Enter Maint Menu

S Enter Password Changes digit indicated by underscore cursor

ζ

0000

S

Locks underscored digit and moves cursor to next digit

Invalid

O

= Option Switch

Valid

S

= Select Switch

To Calibration

See Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 15

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.2.2 Maintenance Menu Flow Chart

The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart . From the operational display, press the

OPTION switch 6 times; "Enter MAINTENANCE Menu" is displayed.

To/From

Operational

Display

From Operational Menu

Press

O PTION

(6 times)

O

= Press Option

O (6)

Enter Maint Menu S Enter Password

O

S

= Press Select

Changes digit indicated by underscore cursor

0000

S

Locks underscored digit and moves cursor to next digit

Invalid

Valid

MAINTENANCE MENU

Calibration

S

O

If installed

MAINTENANCE MENU

Scale mA Xmtrs

O

MAINTENANCE MENU

Set Alarm1

S

S

Calibration

Select (Gas)

Scale mA Xmtrs

Select (Gas)

Alarm 1

Select: XX

Press O PTION until the channel to be Calibrated is displayed

See Section 5.3 for calibration instructions.

Press O PTION until gas to be Set-Up is displayed

See Section 5.3.2 for transmitter set-up instructions.

Press O PTION until the gas alarm to be Set is displayed

See Section 5.3.3 for setting alarms instructions.

O

MAINTENANCE MENU

Set Alarm2

S

O

MAINTENANCE MENU

Set Alarm Delays

S

O

MAINTENANCE MENU

Configure Alarms

S

O

MAINTENANCE MENU

Relay Failsafes

S

O

MAINTENANCE MENU mA Output Span

S

Alarm 2

Select: XX

Alarm Delay

Select: XX

Ch 1, 2, 3, 4

R1 L

Relay Failsafes

R:1 Failsafe ON mA Output Span

Select: (Gas)

Press O PTION

until the gas alarm to be Set is displayed

See Section 5.3.3 for setting alarms instructions.

Press O PTION until the gas alarm delay to be Set is displayed

See Section 5.3.4 for setting alarms instructions.

Pressing O PTION changes letter indicated by underscore cursor

See Section 5.3.5 for configuring relay instructions.

Pressing O PTION changes Failsafe setting from ON to OFF

See Section 5.3.6 for configuring relay failsafe instructions.

Press O PTION until the gas span to be Set is displayed

See Section 5.3.7 for output span instructions.

O

MAINTENANCE MENU

Set New Password

S

S

O

O

MAINTENANCE MENU

Exit maint menu

Password

XXXX

O

S

Changes digit indicated by underscore cursor

Locks underscored digit and moves cursor to next digit

See Section 5.3.8

for changing password instructions.

Press O PTION to return to top of maintenance menu.

Press S ELECT to return to operational menu.

See Section 5.3.9 for instructions on how to exit Maintenance Menus.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

Figure 6: GSM-60 Maintenance Menu Flow Chart.

P a g e | 16

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3 Calibration for CO, O

2,

HC and CO

2

(Gas Channels)

Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. This is a two-step process.

A Low Calibration sets clean air reference point and the High Calibration function sets the sensitivity of the instrument.

Calibration equipment is available from ENMET to calibrate the GSM-60 . A list of needed material is in Section 7.0. A calibration adapter will have a fitting for the gas cylinder on one side, and a quick-disconnect to attach to the instrument on the other.

You may exit the calibration section, at any time, by pressing and holding the OPTION switch for 3 seconds, if entering calibration section by mistake or calibration gas is not available.

Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the OPTION and SELECT switches on the front panel.

After entering a valid password to maintenance menu, see Section 5.2.1

, the calibration section is the first menu section; enter by pressing the SELECT switch.

Supply sensor with clean air for Low Cal/Zero Cal setting and apply calibration gas for Hi Cal/Span Gas setting.

 Press the SELECT switch "Calibration Select XX " is displayed. XX = the gas to be calibrated

 Press the OPTION switch, if needed, to change to the gas to be calibrated.

 Press the SELECT switch, the gas & current reading is displayed in upper portion of display. The mV reading & "Low Cal 0" is displayed in the lower portion of display. This is the Low Cal setting, usually zero , clean air must be supplied to the sensor.

This reading needs to be at or near zero. If it is not, then a cylinder of clean 20.9 air should be used. See Figure 7 if this is required.

 Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to

"Hi Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.

 Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.

 Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration Menu

Repeat above steps for each channel to be calibrated.

N OTE : Instruments equipped with an oxygen sensor programmed with a range of 0 to 15% by volume require 99.999 % Nitrogen

( ENMET part number 03295-100) be used for the Low Cal/Zero Cal setting and 5% by volume Oxygen ( ENMET part number 03296-050) be used for Hi Cal/Span setting

To continue to next section, Press the OPTION switch.

Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Full Calibration Flow Chart, for CO

From Valid Password Entry

Default Calibration Points

Gas Low Cal Hi Cal

MAINTENANCE MENU S

Calibration

Calibration

Select (Gas)

O Press O PTION until the gas to be

Calibrated is displayed

CO

O2

0

N/A*

20

20.9*

CO2 0 1000

S

CO: XX

LowCal: 0000

11

S each digit

CO: XX

HiCal: 0000

14

HC 0 10

O

S each digit

O

= Press Option

S

= Press Select

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 17

Rev. 1 – December 15, 2016

GSM-60

Calibration/Sample Port

Fitting

N OTE : With twin tube input, the noncalibration line must be plugged.

Quick Disconnect

ENMET

Cylinder Valve and Regulator

Cylinder of Gas

To Cylinder of Gas

GSM-60

with

O

PTIONAL

T

WIN

T

UBE

Example of Regulator

Humidifier

Calibration Adapter with Humidifier - MOS

Figure 7: GSM-60 Calibration Connections

Calibration Process for MOS Sensors

The following calibration procedure must be followed whenever a 60-series part number appear in the part number of the instrument, examples 046526200 -0000, 04652-70836000

Humidification is required when calibration is performed. In addition to the standard flow demand regulator, cylinder of calibration gas you need ENMET humidifier assembly part number 037000-000

 Fill the humidifier with clean water to about ¾ full, connect the flow demand regulator to one side of the humidifier and connect the other side of the humidifier to the GSM-60 calibration/sampling port.

Warning: Be sure that the flow direction is correct, note the arrow on the humidifier , failure to do so will cause damage to the

GSM-60 instrument

With the cylinder regulator and humidifier assembled follow the above steps to complete the calibration process

5.3.1A Low Cal/Zero Cal Adjust

A Low-Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.

Enter the maintenance menu by repeatedly pressing OPTION switch, until the maintenance menu is displayed. See Figure 6,

GSM-60 Maintenance Menu flow chart.

The first menu available is the Low Cal/Zero Cal.

Press the SELECT switch 4 times to perform a Low Cal.

 If the Low Cal/Zero Cal is successful , the display will change to Hi Cal/Span Gas.

If you wish to Hi Cal/Span Gas the sensor apply calibration gas. Proceed to gas calibration Section 5.3.1B

If you wish to Exit the maintenance menu, Press and hold OPTION switch until the Maintenance Menu is displayed then release. Then press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the Operational Display

If the Low Cal/Zero Cal is Not successful, sensor is outside of safe parameters to Low Cal, a “SLP/Off Set err” will be indicated. Repeat Section 5.3.1 Low Cal/Zero Cal Adjust making sure to use a cylinder of 20.9% Oxygen.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 18

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.1B High Cal/Span Gas Adjust

A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed.

 Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to

"HI Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.

 Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.

 Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration

Menu

Repeat above steps for each channel to be calibrated.

To continue to next section, Press the OPTION switch.

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

5.3.2 Set 4 –20mA Transmitter Scale

This section of the maintenance menu is installed when there are 4-20mA style sensors for dew point or other gases. This function is normally performed at the factory and is not usually required to be performed in the field unless a new transmitter is installed.

After entering a valid password into maintenance menu, the Scale mA Xmtrs section is the second menu section, if it is installed, enter by pressing the S ELECT switch

 Press the SELECT switch "mA Xmter Scale: Select XX " is displayed. XX = the gas to be set up.

Press the OPTION switch, if needed, to change to the gas to be set up.

 Press the SELECT switch, “Ch#: mAXmter: 4mA: 0000” is displayed

 Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display move to the full-Scale mA Xmtrs menu

 Press the SELECT switch, “Ch#: mAXmter: 20mA: 0000” is displayed

Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will return to the Scale mA Xmtrs menu

 Repeat these steps for each 4 –20mA transmitter.

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Sensor/Transmitter Set Up Flow Chart

MAINTENANCE MENU S mA Xmtr Scale

Scale mA Xmtrs

Select (Gas)

O Press O PTION until the gas to be

Set Up is displayed

O

S

= Press Option

= Press Select

O S

CH#: XX mA Xmter

4mA: 0000

S each digit

CH#: XX mA Xmter

20mA: 0000

S each digit

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 19

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.3 Set Alarm Points

Factory alarm set points are discussed in Section 4.2, See Table 1. To change the alarm points, you must enter the maintenance menu.

Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password is 1270.

When a valid numerical password is inserted, the user can enter the maintenance menu.

In the "Enter Maint Menu" position

 Press the SELECT switch "Enter Password

ζ

0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.

 In the

ζ

000 position, the underline cursor is under the left digit.

 Press the OPTION switch to change the left digit; select the correct digit.

 Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint

Menu display.

After entering a valid password:

 Press the OPTION switch until; “Maintenance Menu Set Alarm1” appears on display.

 Press the SELECT switch, "ALARM1 Select: XX " is displayed. XX = the gas of alarm point to be changed.

 Press the OPTION switch until, desired gas is displayed.

 Press the SELECT switch; "ALARM 1 V " is displayed, with the flashing placeholder underscore cursor, under the left most character,

Λ

for ascending trigger point or V for descending trigger point indicator.

 Press the OPTION switch to toggle between

Λ

and V ; select the correct indicator.

 Press the SELECT switch to lock in the correct indicator. "ALARM 1 STD " is displayed

 Press the OPTION switch to toggle between STD and DIFF ; select the correct indicator.

Press the SELECT switch to lock in the correct indicator.

If STD is selected, "ALARM 1 V L " is displayed.

The next character is the latching indicator L or NO L press the OPTION switch to toggle the latching mode.

The next character is the negative sign – press the OPTION switch to toggle the negative sign.

 The next characters are the alarm 1 value, press the OPTION switch to select each digit of the value

When the last digit is accepted display returns to the "Set Alarm1" position.

If DIFF is selected, "ALARM 1 DI

FF Λ

000 " is displayed.

 The next characters are the alarm 1 value, press the OPTION switch to select each digit of the value

 Press the SELECT switch to lock in the correct character and move the cursor to the right.

 "ALARM 1 DIFF BAND 000 " is displayed, press the OPTION switch to select each digit of the value.

The next characters are the alarm 1 differential value, press the OPTION switch to select each digit of the value

 Press the SELECT switch to lock in the correct character and move the cursor to the right.

When the last digit is accepted, display returns to the "Set Alarm1" position.

Note : The Alarm 1 differential value is the delay of the GSM-60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value.

Example : If the alarm set point is

Λ

10 and the differential is 2, the GSM-60 will go into alarm at 10 and stay in alarm until the reading has dropped below 8.

 Repeat for each sensor alarm 1 to be changed.

 Press the OPTION switch to move to alarm 2, "Set ALARM2" is displayed.

 Repeat as for alarm 1 using the STD section.

Press OPTION switch until “Exit Maint Menu” appears, then press SELECT switch to return the instrument to the Operational

Display

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 20

Rev. 1 – December 15, 2016

GSM-60 ENMET

Example: Set Alarms Flow Chart

Displays are examples of Alarms

Λ

- Indicates alarm triggered on increasing value of reading v - Indicates alarm triggered on decreasing value of reading

O

= Press Option

S

= Press Select

L - Indicates alarm is set for latching

NO L - Indicates alarm is set for non-latching

STD – Indicates alarm in standard setting, can be set in latched or non-latched mode

DIFF – Indicates alarm in differential setting, instrument will stay in alarm beyond the alarm set point by the differential value

MAINTENANCE MENU

Set Alarm1

S Alarm 1 -

Select: XX

O Press O PTION until the channel to be Set is displayed

O

S

Alarm 1

Λ

O

S

Toggles status between

Locks selection

Λ

increasing trigger and v decreasing trigger

S

Alarm 1

STD

DIFF S

O

S

Toggles status between Standard and D IFF

Locks selection

Alarm 1

Λ DIFF 000

O

Changes character indicated by underscore cursor

S

Locks underscored character and moves cursor to next digit

S TD

S

Alarm 1

Λ ζ

L

Alarm 1

Λ

L–010

O

Alarm 1 DIFF BAND

O

000 S

Changes character indicated by underscore cursor

Locks underscored character and moves cursor to next digit

Toggles status between

Latching and Non Latching

S

Locks selection

O

S

Changes character indicated by underscore cursor

Locks underscored character and moves cursor to next digit

MAINTENANCE MENU

Set Alarm2

S

O

Alarm 2 -

Select: XX

S

Alarm 2

V L 0000

O Press O PTION until the channel to be Set is displayed

O Changes character indicated by underscore cursor

S

Locks underscored character and moves cursor to next

See Section 4.2 Table 3 for factory alarm set points.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 21

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.4 Set Alarm Delay

The alarms may be set to delay by 1 second increments, up to 255 seconds. Alarm delays are factory set to 5 seconds.

To change an alarm delay, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1

.

See Table 4 below for factory set delays. A space is provided to record changes.

After entering a valid password:

 Press the OPTION switch until; “Maintenance Menu Set Alarm Delay” appears on display.

Press the SELECT switch, "ALARM Delay Select: XX " is displayed. XX = the gas alarm to be changed.

 Press the OPTION switch until, desired gas is displayed.

 Press the SELECT switch; "ALARM Delay = 0000" is displayed, with the underscore cursor under the left digit.

Press the OPTION switch to change the left digit; select the correct digit.

 Press the SELECT switch to lock in the correct digit and move the cursor one digit to the right. When the last digit is accepted, display returns to the "Set Alarm Delay" position.

 Repeat for each sensor alarm delay to be changed.

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Set Alarm Delay Flow Chart

MAINTENANCE MENU

Set Alarm Delays

S

Alarm Delay -

Select: XX

S

O

Alarm Delay -

0005

O

O

S

Press Option until the channel to be Set is displayed

O

S

= Press Option

= Press Select

Changes digit indicated by underscore cursor

Locks underscored digit and moves cursor to next digit

Table 4: Factory Set Gas alarms Delay

Gas Delay

CO 5 sec

H

2

S 5 sec

O

2

5 sec

CO

2

5 sec

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 22

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.5 Relay Configuration

To change a relay configuration, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.

In the "Enter Maint Menu" position

Press the SELECT switch "Enter Password

ζ

0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.

 In the

ζ

000 position, the underline cursor is under the left digit.

 Press the OPTION switch to change the left digit; select the correct digit.

 Press the SELECT switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the Enter Maint

Menu display.

After entering a valid password:

Press the OPTION switch until “Configure Alarms” is displayed

 Press the SELECT switch to enter the Configure Alarms menu

 Press the OPTION switch to set relay configuration as needed, see below for indications

L = Low Alarm, H = High Alarm, B = Both Alarms,

ζ

= No Relay linked to channel

Press the SELECT switch to lock setting and move to next, channel and relay

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Set Relay Configuration Flow Chart

MAINTENANCE MENU S Ch 1,2,3,4

R1 L

Configure Alarms

Ch 1 2 3 4

R2 L

O

Ch 1 2 3 4

R3 L

Pressing O PTION changes letter indicated by underscore cursor

Pressing S ELECT Locks underscored digit and moves cursor to next digit

Ch 1 2 3 4

R4 L

Ch 1 2 3 4

R5

H H H H

O

S

= Press Option

= Press Select

The table below shows the default relay links.

Relay 1

Relay 2

Relay 3

Relay 4

Relay 5

Channel 1

Low Alarm

High Alarm

Channel 2

Low Alarm

High Alarm

Relays can be linked to specific alarms.

N OTE : Each operating channel must be linked to at least 1 relay.

Channel 3

Low Alarm

High Alarm

Channel 4

Low Alarm

High Alarm

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 23

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.6 Failsafe Configuration

To change a relay failsafe configuration, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint

Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described below.

In the "Enter Maint Menu" position

 Press the SELECT switch "Enter Password

ζ

0" is displayed. Press SELECT switch once, to move cursor to next digit, this will be the first digit of the password.

 In the

ζ

000 position, the underline cursor is under the left digit.

 Press the OPTION switch to change the left digit; select the correct digit.

 Press the switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted, you are returned to the

Enter Maint Menu display.

After entering a valid password:

Press the OPTION switch until “Relay Failsafes” is displayed

 Press the SELECT switch to indicate relay to be set.

 Press the OPTION switch to set relay indicated, On or Off as appropriate.

 Press the SELECT switch to cycle through each of the 5 relays, return to “Maintenance Menu Relay Failsafes”

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Set Relay Failsafe Configuration Flow Chart

MAINTENANCE MENU

Relay Failsafes

O

S

Relay Failsafes

R: 1 Failsafe ON

Relay Failsafes

Pressing O PTION changes Failsafe setting from ON to OFF

Continue pressing S ELECT to cycle through all 5 relays

R: 2 Failsafe ON

Relay Failsafes

R: 3 Failsafe ON

Relay Failsafes

Changing the setting of each relay as needed, using the Option switch

S

R: 4 Failsafe ON

Relay Failsafes

R: 5 Failsafe ON

O

S

= Press Option

= Press Select

5.3.7 Set Output Span Range

To change 4-20 mA output range. This range is set at the factory and should not be changed, contact ENMET for information.

 Press the OPTION switch to continue to next section of maintenance menu.

Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Set Output Span Flow Chart

MAINTENANCE MENU mA Output Span

S mA OUTPUT SPAN

Select: CO

S

O mA OUTPUT SPAN

_ 50 CO

O Press O PTION until the channel to be Set is displayed.

O

S

Changes digit indicated by underscore cursor

O

S

= Press Option

= Press Select

Locks underscored digit and moves cursor to next digit

S

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 24

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.3.8 Set New Password

To change the password, you must enter the maintenance menu. Press the OPTION switch until “Enter Maint Menu” is displayed then press SELECT switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1.

To set a new password, after inserting a valid password,

 Press the OPTION switch until; "Set New Password” is displayed.

 Press the SELECT switch; "Password

ζ

1270" is displayed, with the underscore cursor under the left digit.

 Use the OPTION switch to change the left digit, when the desired digit is displayed.

 Press the SELECT switch to lock the digit in place and move the cursor one digit to the right.

When all four digits of the new password have been selected, "Set New Password" is displayed.

Record the new password; without it, the maintenance menu cannot be reentered once you exit the Maintenance Menu. If the password is lost, call ENMET customer service personnel.

From the "Password XXXX " position,

 Press the SELECT switch to return to Set New Password section.

 Press the OPTION switch; to continue to "exit MAINTENANCE Menu"

Example: Set Password Flow Chart

MAINTENANCE MENU S

O

Set New Password

O

Password

0000

S

Changes digit indicated by underscore cursor

Locks underscored digit and moves cursor to next digit

O

= Press Option

S

= Press Select

5.3.9 Exit Maintenance Menu

From the "exit MAINTENANCE Menu" position

Press the SELECT switch to resume the operational display.

 Press the OPTION switch to reenter the maintenance menu at the "Calibration" position.

Example: Exit Maintenance Menu Flow Chart

MAINTENANCE MENU

Exit Maint Menu

CH-1 CH-3

CH-2 CH-4

MAINTENANCE MENU

Calibration

O S

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 25

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.4 Sensor Replacement

5.4.1 Gas/Oxygen Sensor

A Gas sensor must be replaced when it can no longer be calibrated. To replace a sensor, perform the following steps.

 Turn off the electrical power. The sample air can continue to flow.

 Open the display panel and remove the four manifold retention screws and remove the manifold. See Figure 9 .

Remove the old sensor, and replace it with a new sensor. It is recommended that the sensor gasket/manifold also be checked and potentially replaced at the time the sensor is replaced.

C AUTION : Some new sensors come with a shorting clip that must be removed before installation, for proper operation. See

Figure 8 .

 Replace the manifold.

 Turn on the electrical power.

Remove Shorting Clip from Cell if Present

Orifice

To Pump

PiD (HC) Sensor Manifold

Bottom View of Sensor

Figure 8: Shorting Clip

Gas Sensor Manifold

Retaining Screw, 4 places

Figure 9: Location of Gas Sensor and PiD (HC) Manifolds

5.4.2 Calibration/Sensor Replacement

Sensor replacement requires that a Factory Calibration be performed. Factory Calibration allows the instrument to properly set operational parameters for each sensor.

Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. This is a two-step process. A Low Calibration sets clean air reference point and the High Calibration function sets the sensitivity of the instrument.

Calibration equipment is available from ENMET to calibrate the GSM-60 . A list of needed material is in Section 7.0. A calibration adapter will have a fitting for the gas cylinder on one side, and a quick-disconnect to attach to the instrument on the other.

You may exit the calibration section, at any time, by pressing and holding the OPTION switch for 3 seconds, if entering calibration section by mistake or calibration gas is not available.

Wait 24 hours after initially supplying air and power to the GSM-60 sensor before initial calibration. It is not necessary to open the Front Panel to make adjustment. The calibration functions are operated through the OPTION and SELECT switches on the front panel.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 26

Rev. 1 – December 15, 2016

GSM-60 ENMET

After entering a valid password to maintenance menu, see Section 5.2.1

, the calibration section is the first menu section; enter by pressing the SELECT switch.

Supply sensor with clean air for Low Cal/Zero Cal setting and apply calibration gas for Hi Cal/Span Gas setting.

Press the SELECT switch "Calibration Select XX " is displayed. XX = the gas to be calibrated

 Press and Hold the OPTION switch, until the letter F appears in the upper right hand corner of the display. The F indicates that the instrument is in the Factory Calibration Mode.

 Press the OPTION switch, if needed, to change to the gas to be calibrated.

Press the SELECT switch, the gas & current reading is displayed in upper portion of display. The mV reading & "Low Cal

0" is displayed in the lower portion of display. This is the Low Cal setting, usually zero , clean air must be supplied to the sensor. This reading needs to be at or near zero. If it is not, then a cylinder of clean 20.9 air should be used. See Figure 7 if this is required.

 Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to

"Hi Cal xx" gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.

 Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.

Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration

Menu

Repeat above steps for each channel to be calibrated.

N OTE : Instruments equipped with an oxygen sensor programmed with a range of 0 to 15% by volume require 99.999 %

Nitrogen ( ENMET part number 03295-100) be used for the Low Cal/Zero Cal setting and 5% by volume Oxygen

( ENMET part number 03296-050) be used for HI Cal/Span setting

To continue to next section, Press the OPTION switch.

 Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

Example: Full Calibration Flow Chart, for CO

From Valid Password Entry

Default Calibration Points

Gas Low Cal Hi Cal

MAINTENANCE MENU

Calibration

S Calibration

Select (Gas)

O Press O PTION until the gas to be

Calibrated is displayed

CO

O2

0

N/A*

20

20.9*

O

S

CO: XX

LowCal: 0000

11

S each digit

CO2

HC

0

0

1000

10

CO: XX

HiCal: 0000

14

S each digit

O

= Press Option

S

= Press Select

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 27

Rev. 1 – December 15, 2016

GSM-60

Calibration/Sample Port

Fitting

N OTE : With twin tube input, the noncalibration line must be plugged.

Quick Disconnect

2 places

Plug

ENMET

Cylinder Valve and Regulator

Cylinder of Gas

To Cylinder of Gas

GSM-60

with

O

PTIONAL

T

WIN

T

UBE

Example of Regulator

Humidifier

Calibration Adapter with Humidifier - MOS

Figure 7: GSM-60 Calibration Connections

Calibration Process for MOS Sensors

The following calibration procedure must be followed whenever a 60-series part number appear in the part number of the instrument, examples 046526200 -0000, 04652-70836000

Humidification is required when calibration is performed. In addition to the standard flow demand regulator, cylinder of calibration gas you need ENMET humidifier assembly part number 037000-000

 Fill the humidifier with clean water to about ¾ full, connect the flow demand regulator to one side of the humidifier and connect the other side of the humidifier to the GSM-60 calibration/sampling port.

Warning: Be sure that the flow direction is correct, note the arrow on the humidifier , failure to do so will cause damage to the

GSM-60 instrument

 With the cylinder regulator and humidifier assembled follow the above steps to complete the calibration process

5.4.3A Low Cal/Zero Cal Adjust

A Low Cal function should be performed only when the GSM-60 sensors are exposed to clean uncontaminated air. Use a cylinder of 20.9% oxygen to provide a clean air reference if necessary. Attach the cylinder to the calibration adapter, attach the adapter to the instrument and allow gas to flow over the sensor for up to 4 minutes.

Enter the maintenance menu by repeatedly pressing OPTION switch, until the maintenance menu is displayed. See Figure 6,

GSM-60 Maintenance Menu flow chart.

The first menu available is the Low Cal/Zero Cal.

Press the SELECT switch 4 times to perform a Low Cal.

If the Low Cal/Zero Cal is successful , the display will change to Hi Cal/Span Gas.

If you wish to Hi Cal/Span Gas the sensor apply calibration gas. Proceed to gas calibration Section 5.3.1B

If you wish to Exit the maintenance menu, Press and hold OPTION switch until the Maintenance Menu is displayed then release. Then press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the Operational Display

 If the Low Cal/Zero Cal is Not successful, sensor is outside of safe parameters to Low Cal, a “SLP/Off Set err” will be indicated. Repeat Section 5.3.1 Low Cal/Zero Cal Adjust making sure to use a cylinder of 20.9% Oxygen.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 28

Rev. 1 – December 15, 2016

GSM-60 ENMET

5.4.3B High Cal/Span Gas Adjust

A High Cal/Span Gas should only be performed after a successful Low Cal/Zero Cal has been completed.

 Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will move to

"Hi Cal xx " gas calibration. xx = the level of gas to be used for calibration. The mV reading is shown in the upper right hand corner of the display.

 Apply calibration gas to sensor. See Figure 7 . After about 1 minute and mV reading has stabilized.

 Press the SELECT switch, that moves the cursor one digit to the right, when the last digit is accepted and the calibration is successful the display will momentarily show Cal OK then slope and off set readings, before returning to the Calibration

Menu

Repeat above steps for each channel to be calibrated.

To continue to next section, Press the OPTION switch.

Press OPTION switch until “Exit Maint Menu” appears and then press SELECT switch to return the instrument to the

Operational Display

5.5 Flow Control Orifice

A 0.0225-inch diameter orifice is used to set the flow rate. In well-maintained medical air systems, this orifice should not clog.

However, if difficulty is experienced in maintaining flow rate examine this orifice; replace it if necessary.

Orifice location will depend on instrument sensor configuration.

6.0 Technical Data and Specifications

N OTE : All specifications stated in this manual may change without notice.

Electrical Power

Storage and

Transport

Operation

Mechanical

Outputs

Examples of Sensors

15 Amp fused branch circuit

100-240 V AC

0.9 A

50/60 Hz

Board Mounted Fuse FH2, 0.630A, 5 x

20mm

Temperature: preferred

Relative Humidity

Atmospheric Pressure

Temperature:

Relative Humidity

Atmospheric Pressure

Dimensions:

Weight:

Material:

Strain relief:

Relays:

Analog:

Digital:

Audio:

Type

CO

O

2

CO

2

HC

Others

N OTE : Response time is dependent on sampling length.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

Range

0 – 50 ppm

0 – 30%

0 – 5000 ppm

0 – 100 ppm

-20

°

to +60

°

C (-4

°

to +140

°

F)

0

°

to +20

°

C (32

°

to 68

°

F)

0 - 99% RH, non-condensing

20 to 36 inHg (68 to 133 kPa)

0

°

to +40

°

C (32

°

to +104

°

F)

0 - 99% RH, non-condensing

20 to 36 inHg (68 to 133 kPa)

11 x 9 x 8 inches (27.9 x 22.9 x 20.3 cm)

8 lbs (3.6 kg)

Engineered thermoplastic with hinged front cover

5 – 12 mm OD

SPDT

Resistive Load Inductive Load

10A at 110 VAC 7.5A at 110 VAC

10A at 30 VDC 5A at 30 VDC

4-20 mA x 3

RS-232 – Modbus

RS-485 – Modbus

95 db at 2 ft

*Typical Response Time

T

90

= 30 seconds

T

90

= 15 seconds

T

90

= 30 seconds

T

90

= 30 seconds

Contact ENMET

Life

1 – 3 years

1 – 2 years

3 – 5 years

1 – 2 years

P a g e | 29

Rev. 1 – December 15, 2016

GSM-60

7.0 Replacement Part Numbers

7.1 ENMET part numbers for sensors and replacement parts:

Part number Description

03053-000

67025-1114

03015-014

03016-014

03028-XXX

67025-XXXX

03028-005

73540-002

64002-1000

06008-004

65057-011

65057-012

65057-010

73089-004

04018-128

04018-118

Sensor, CO2

Sensor, Oxygen

Sensor, MOS 812

Sensor, MOS 813

Sensor, PID, 10.6 eV

Contact ENMET

Sensor, Toxic

Contact ENMET

Sensor, Replacement lamp, PID, 10.6 eV

Orifice

Fuse, 1.0 Amp 5x20mm

Sensor Gasket

Terminal plug, 3 position

Terminal plug, 4 position

Terminal plug, 2 position

Filter Assembly, Particulate

Pump, Replacement for instrument s/n 118 and above

Pump, Replacement for instrument s/n 117 and below

Consult ENMET or your distributer for more information

7.2 ENMET part numbers for Calibration equipment:

Part number Description

03219-020

03296-209

03510-001

03223-1000

03510-002

03290-010

Gas Cylinder, 20 ppm CO in air

Gas Cylinder, 20.9% oxygen in nitrogen

Regulator Assembly, Flow Demand CO, O

2

(steel cylinders)

Gas Cylinder, 1000 ppm CO

2

in air,

Regulator Assembly, Flow Demand CO

2

(aluminum cylinders)

Gas Cylinder, 10 ppm Isobutylene in air

Consult ENMET or your distributer for more information

ENMET

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 30

Rev. 1 – December 15, 2016

GSM-60

Appendix A: CO Characteristics

ENMET

The Characteristics and Effects of Carbon Monoxide

Carbon monoxide is a colorless odorless toxic gas generated by incomplete combustion of a hydrocarbon fuel in air. It may be present where internal combustion engines, furnaces, boilers, and other combustion devices are present. It is toxic when inhaled because of its great affinity to hemoglobin, the oxygen carriers in the red cells of the blood. CO replaces the oxygen normally carried by the hemoglobin, and thus inhibits the delivery of oxygen throughout the body; the victim suffers from oxygen deficiency, and may die from asphyxiation. The symptoms and degree of danger resulting from exposure to CO depend upon the concentration of the gas and the length of exposure; this is shown in Figure 10 . The GSM-60 instrument is employed to warn the user of the presence of CO, and to facilitate the assessment of the degree of danger that he or she is exposed to.

Based upon knowledge of the effects of CO, the Occupational Safety and Health Authority (OSHA) has set limits on exposure to CO in the workplace. These are 35 ppm (parts CO per million parts air) as a time weighted average for an eight-hour day, and a maximum exposure of 200 ppm. For compressed air line applications, OSHA requires Grade D breathing air supplied, using a Compressed Gas

Association (CGA) definition (G-7.1). Depending on interpretation of the OSHA respiratory standard, 10 ppm and 20 ppm CO have been used as maximum limits and standard instrument alarm points.

If a CO sensor is installed, the GSM-60 has two preset alarm set points, at 10 ppm and 20 ppm CO, which are adjustable, but cannot be set below 5 ppm or above 50 ppm.

The curves below are for percent Carboxyhemoglobin with 50% being the top curve, 5% the bottom. %COHb is a measure of the amount of hemoglobin occupied by CO rather than oxygen. CO effects upon children, adults engaging in physical activity, and smokers, are more pronounced.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

Figure 10: Carbon Monoxide Concentration

P a g e | 31

Rev. 1 – December 15, 2016

GSM-60

Appendix B: Gas Ionization Potentials

Chemical Name

A

2-Amino pyridine

Acetaldehyde

Acetamide

Acetic acid

Acetic anhydride

Acetone

Acetonitrile

Acetophenone

Acetyl bromide

Acetyl chloride

Acetylene

Acrolein

Acrylamide

Acrylonitrile

Allyl alcohol

Allyl chloride

Ammonia

Aniline

Anisidine

Anisole

Arsine

B

1,3-Butadiene (butadiene)

1-Bromo-2-chloroethane

1-Bromo-2-methylpropane

1-Bromo-4-fluorobenzene

1-Bromobutane

1-Bromopentane

1-Bromopropane

1-Bromopropene

1-Butanethiol

1-Butene

1-Butyne

2,3-Butadione

2-Bromo-2-methylpropane

2-Bromobutane

2-Bromopropane

2-Bromothiophene

2-Butanone (MEK)

3-Bromopropene

3-Butene nitrile

Benzaldehyde

Benzene

Benzenethiol

Benzonitrile

Benzotrifluoride

Biphenyl

Boron oxide

Boron trifluoride

Bromine

Bromobenzene

Bromochloromethane

Bromoform

Butane

Butyl mercaptan cis-2-Butene m-Bromotoluene n-Butyl acetate n-Butyl alcohol n-Butyl amine n-Butyl benzene n-Butyl formate n-Butyraldehyde n-Butyric acid n-Butyronitrile o-Bromotoluene

IP (eV)

10.55

11.02

11.41

10.10

9.50

10.91

9.67

9.90

10.20

7.70

7.44

8.22

9.89

8.00

10.21

9.77

10.69

10.00

9.69

12.20

9.27

10.39

9.53

9.25

8.33

9.71

9.68

8.27

13.50

15.56

10.54

8.98

9.30

9.14

9.58

10.18

9.23

9.89

9.98

10.08

8.63

9.54

9.70

9.07

10.63

10.09

8.99

10.13

10.10

10.18

8.71

8.69

10.50

9.86

10.16

11.67

8.79

10.77

10.48

10.63

9.15

9.13

8.81

10.01

10.04

IP (eV)

8.67

10.66

9.21

10.67

10.82

10.61

10.65

10.78

8.68

10.04

8.76

13.79

10.07

14.01

11.47

8.28

8.70

8.68

9.91

8.64

8.68

9.13

8.95

7.99

8.56

10.53

9.26

9.01

10.06

8.83

8.83

8.70

12.45

11.37

12.91

7.59

8.14

9.73

8.75

13.80

9.80

9.75

9.14

11.48

10.36

12.65

10.61

9.44

9.07

10.77

10.07

10.85

10.06

10.35

9.82

10.02

9.17

9.65

10.19

11.12

9.65

7.28

9.45

12.20

11.12

10.87

Chemical Name p-Bromotoluene p-tert-Butyltoluene s-Butyl amine s-Butyl benzene sec-Butyl acetate t-Butyl amine t-Butyl benzene trans-2-Butene

C

1-Chloro-2-methylpropane

1-Chloro-3-fluorobenzene

1-Chlorobutane

1-Chloropropane

2-Chloro-2-methylpropane

2-Chlorobutane

2-Chloropropane

2-Chlorothiophene

3-Chloropropene

Camphor

Carbon dioxide

Carbon disulfide

Carbon monoxide

Carbon tetrachloride

Chlorine

Chlorine dioxide

Chlorine trifluoride

Chloroacetaldehyde

α  -Chloroacetophenone

Chlorobenzene

Chlorobromomethane

Chlorofluoromethane (Freon 22)

Chloroform

Chlorotrifluoromethane (Freon 13)

Chrysene

Cresol

Crotonaldehyde

Cumene (isopropyl benzene)

Cyanogen

Cyclohexane

Cyclohexanol

Cyclohexanone

Cyclohexene

Cyclo-octatetraene

Cyclopentadiene

Cyclopentane

Cyclopentanone

Cyclopentene

Cyclopropane m-Chlorotoluene o-Chlorotoluene p-Chlorotoluene

D

1,1-Dibromoethane

1,1-Dichloroethane

1,1-Dimethoxyethane

1,1-Dimethylhydrazine

1,2-Dibromoethene

1,2-Dichloro-1,1,2,2-tetrafluoroethane (Freon 114)

1,2-Dichloroethane

1,2-Dichloropropane

1,3-Dibromopropane

1,3-Dichloropropane

2,2-Dimethyl butane

2,2-Dimethyl propane

2,3-Dichloropropene

2,3-Dimethyl butane

3,3-Dimethyl butanone cis-Dichloroethene

ENMET

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 32

Rev. 1 – December 15, 2016

GSM-60 o-Dichlorobenzene p-Dichlorobenzene p-Dioxane trans-Dichloroethene

E

Epichlorohydrin

Ethane

Ethanethiol (ethyl mercaptan)

Ethanolamine

Ethene

Ethyl acetate

Ethyl alcohol

Ethyl amine

Ethyl benzene

Ethyl bromide

Ethyl chloride (chloroethane)

Ethyl disulfide

Ethyl ether

Ethyl formate

Ethyl iodide

Ethyl isothiocyanate

Ethyl mercaptan

Ethyl methyl sulfide

Ethyl nitrate

Ethyl propionate

Ethyl thiocyanate

Ethylene chlorohydrin

Ethylene diamine

Ethylene dibromide

Ethylene dichloride

Ethylene oxide

Ethylenelmine

Ethynylbenzene

Chemical Name

(D continued)

Decaborane

Diazomethane

Diborane

Dibromochloromethane

Dibromodifluoromethane

Dibromomethane

Dibutylamine

Dichlorodifluoromethane (Freon 12)

Dichlorofluoromethane

Dichloromethane

Diethoxymethane

Diethyl amine

Diethyl ether

Diethyl ketone

Diethyl sulfide

Diethyl sulfite

Difluorodibromomethane

Dihydropyran

Diiodomethane

Diisopropylamine

Dimethoxymethane (methylal)

Dimethyl amine

Dimethyl ether

Dimethyl sulfide

Dimethylaniline

Dimethylformamide

Dimethylphthalate

Dinitrobenzene

Dioxane

Diphenyl

Dipropyl amine

Dipropyl sulfide

Durene m-Dichlorobenzene

N,N-Diethyl acetamide

N,N-Diethyl formamide

N,N-Dimethyl acetamide

N,N-Dimethyl formamide

IP (eV)

9.06

8.95

9.13

9.66

10.20

11.65

9.29

8.96

10.52

10.11

10.48

8.86

8.76

10.29

10.98

8.27

9.51

10.61

9.33

9.14

9.29

8.55

11.22

10.00

9.89

10.52

8.60

10.37

11.05

10.57

9.20

8.82

10.00

8.24

10.00

8.69

7.13

9.18

9.64

10.71

9.19

7.95

7.84

8.30

8.03

9.12

8.60

8.89

8.81

9.12

9.70

8.01

9.53

9.32

8.43

9.68

11.07

9.88

9.00

12.00

10.59

11.07

10.49

7.69

12.31

12.39

11.35

8.34

9.34

7.73

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

IP (eV)

9.18

9.21

9.19

9.26

9.09

9.17

9.28

8.73

10.57

9.97

10.12

8.70

10.46

9.74

10.02

10.32

9.07

8.85

9.99

10.16

8.72

8.69

9.20

9.71

8.61

8.62

8.50

9.61

8.85

8.85

8.85

9.46

9.33

9.35

10.08

11.10

10.18

8.10

15.43

11.62

12.74

13.91

15.77

10.38

9.88

10.46

9.14

7.95

12.31

12.91

12.45

8.89

9.21

8.92

8.66

9.21

15.70

9.20

10.87

10.25

11.05

11.77

11.30

11.78

12.20

8.92

8.79

2-Iodopropane

Iodine

Iodobenzene

Isobutane

Isobutyl acetate

Isobutyl alcohol

Isobutyl amine

Isobutyl formate

Isobutyraldehyde

Isobutyric acid

Isopentane

Isophorone

Isoprene

Isopropyl acetate

Isopropyl alcohol

Isopropyl amine

Isopropyl benzene

Isopropyl ether

Isovaleraldehyde m-Iodotoluene o-Iodotoluene p-Iodotoluene

K

Ketene

L

2,3-Lutidine

2,4-Lutidine

2,6-Lutidine

Chemical Name

F

2-Furaldehyde

Fluorine

Fluorobenzene

Formaldehyde

Formamide

Formic acid

Freon 11 (trichlorofluoromethane)

Freon 112 (1,1,2,2-tetrachloro-1,2-difluoroethane)

Freon 113 (1,1,2-trichloro-1,2,2-trifluororethane)

Freon 114 (1,2-dichloro-1,1,2,2-tetrafluoroethane)

Freon 12 (dichlorodifluoromethane)

Freon 13 (chlorotrifluoromethane)

Freon 22 (chlorofluoromethane)

Furan

Furfural m-Fluorotoluene o-Fluorophenol o-Fluorotoluene p-Fluorotoluene

H

1-Hexene

2-Heptanone

2-Hexanone

Heptane

Hexachloroethane

Hexane

Hydrazine

Hydrogen

Hydrogen bromide

Hydrogen chloride

Hydrogen cyanide

Hydrogen fluoride

Hydrogen iodide

Hydrogen selenide

Hydrogen sulfide

Hydrogen telluride

Hydroquinone

I

1-Iodo-2-methylpropane

1-Iodobutane

1-Iodopentane

1-Iodopropane

2-Iodobutane

ENMET

P a g e | 33

Rev. 1 – December 15, 2016

GSM-60

Chemical Name

M

2-Methyl furan

2-Methyl napthalene

1-Methyl napthalene

2-Methyl propene

2-Methyl-1-butene

2-Methylpentane

3-Methyl-1-butene

3-Methyl-2-butene

3-Methylpentane

4-Methylcyclohexene

Maleic anhydride

Mesityl oxide

Mesitylene

Methane

Methanethiol (methyl mercaptan)

Methyl acetate

Methyl acetylene

Methyl acrylate

Methyl alcohol

Methyl amine

Methyl bromide

Methyl butyl ketone

Methyl butyrate

Methyl cellosolve

Methyl chloride

Methyl chloroform (1,1,1-trichloroethane)

Methyl disulfide

Methyl ethyl ketone

Methyl formate

Methyl iodide

Methyl isobutyl ketone

Methyl isobutyrate

Methyl isocyanate

Methyl isopropyl ketone

Methyl isothiocyanate

Methyl mercaptan

Methyl methacrylate

Methyl propionate

Methyl propyl ketone

 -Methyl styrene

Methyl thiocyanate

Methylal (dimethoxymethane)

Methylcyclohexane

Methylene chloride

Methyl-n-amyl ketone

Monomethyl aniline

Monomethyl hydrazine

Morpholine n-Methyl acetamide

N

1-Nitropropane

2-Nitropropane

Naphthalene

Nickel carbonyl

Nitric oxide, (NO)

Nitrobenzene

Nitroethane

Nitrogen

Nitrogen dioxide

Nitrogen trifluoride

Nitromethane

Nitrotoluene p-Nitrochloro benzene

O

Octane

Oxygen

Ozone

P

1-Pentene

1-Propanethiol

2,4-Pentanedione

IP (eV)

10.00

9.85

11.32

9.30

7.32

7.67

8.20

8.90

9.90

10.85

8.97

10.54

9.34

10.07

9.60

11.28

11.00

8.46

9.53

10.82

9.54

9.30

9.98

10.67

9.32

9.25

9.44

9.70

10.15

9.39

10.80

9.08

8.40

12.98

9.44

10.27

10.37

8.39

7.96

7.96

9.23

9.12

10.12

9.51

8.67

10.08

8.91

8.35

10.07

10.88

10.71

8.12

8.27

9.25

9.92

10.88

15.58

9.78

12.97

11.08

9.45

9.96

9.82

12.08

12.08

9.50

9.20

8.87

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

IP (eV)

9.51

8.47

12.30

15.33

9.66

13.00

7.78

11.30

9.91

10.00

11.07

10.51

9.70

9.98

10.24

11.84

10.04

10.20

8.78

8.72

9.27

10.54

9.73

10.87

9.00

10.22

10.36

9.32

8.20

8.82

7.64

8.77

8.52

6.89

11.77

9.87

9.38

9.02

9.02

9.04

11.07

10.40

10.35

9.32

8.18

8.50

10.04

11.00

11.78

9.86

7.44

11.62

9.32

11.47

9.54

9.25

10.00

8.86

8.82

9.27

10.67

10.51

9.45

9.47

11.77

11.42

7.50

11.40

Chemical Name

(P continued)

2-Pentanone

2-Picoline

3-Picoline

4-Picoline n-Propyl nitrate

Pentaborane

Pentane

Perchloroethylene

Pheneloic

Phenol

Phenyl ether (diphenyl oxide)

Phenyl hydrazine

Phenyl isocyanate

Phenyl isothiocyanate

Phenylene diamine

Phosgene

Phosphine

Phosphorus trichloride

Phthalic anhydride

Propane

Propargyl alcohol

Propiolactone

Propionaldehyde

Propionic acid

Propionitrile

Propyl acetate

Propyl alcohol

Propyl amine

Propyl benzene

Propyl ether

Propyl formate

Propylene

Propylene dichloride

Propylene imine

Propylene oxide

Propyne

Pyridine

Pyrrole

Q

Quinone

S

Stibine

Styrene

Sulfur dioxide

Sulfur hexafluoride

Sulfur monochloride

Sulfuryl fluoride

T o-Terphenyls

1,1,2,2-Tetrachloro-1,2-difluoroethane (Freon 112)

1,1,1-Trichloroethane

1,1,2-Trichloro-1,2,2-trifluoroethane (Freon 113)

2,2,4-Trimethyl pentane o-Toluidine

Tetrachloroethane

Tetrachloroethene

Tetrachloromethane

Tetrahydrofuran

Tetrahydropyran

Thiolacetic acid

Thiophene

Toluene

Tribromoethene

Tribromofluoromethane

Tribromomethane

Trichloroethene

Trichloroethylene

Trichlorofluoromethane (Freon 11)

Trichloromethane

Triethylamine

Trifluoromonobromo-methane

ENMET

P a g e | 34

Rev. 1 – December 15, 2016

GSM-60

Chemical Name

(T continued)

Trimethyl amine

Tripropyl amine

Notes:

V o-Vinyl toluene

Valeraldehyde

Valeric acid

Vinyl acetate

Vinyl bromide

Vinyl chloride

Vinyl methyl ether

IP (eV)

7.82

7.23

8.20

9.82

10.12

9.19

9.80

10.00

8.93

Chemical Name

W

Water

X

2,4-Xylidine m-Xylene o-Xylene p-Xylene

IP (eV)

12.59

7.65

8.56

8.56

8.45

ENMET

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 35

Rev. 1 – December 15, 2016

GSM-60

8.0 Terms and Conditions

ENMET

8.1 Ordering Information

Address orders to:

ENMET

Attention: Customer Service Department

680 Fairfield Court

Ann Arbor, MI 48108

Email Orders: [email protected]

Phone: 734-761-1270 Fax: 734-761-3220

You may also contact our customer service department by email [email protected]. MINIMUM ORDER IS $50.00.

8.2 Delivery

Unless Seller otherwise specifies, delivery will be made: FOB Ann Arbor, MI and/or FOB Bowling Green, KY. Title and risk of loss shall pass to Buyer at that point. Shipping and handling charges will be Prepaid and Added to Buyer’s invoice. Buyer may request shipping be charged to their own account with a preferred carrier. Seller shall have the right to choose means of transportation and to route shipment when specific instructions are not included with Buyer’s order. Seller agrees to deliver the goods and services, within the time, in accordance with specifications, at the prices specified on the face hereof. Buyer’s orders to this quotation are not subject to cancellation or deferment of delivery without indemnification of loss to the Seller resulting there from. Seller shall not be liable to Buyer for any loss or damage sustained on account of this delay or nonperformance due to causes beyond Seller’s control and without his fault or negligence. Where performance of the terms here is contingent upon timely delivery of goods or services by the Buyer and such delivery is in default, Seller shall be indemnified for any damage or loss resulting there from, and/or by extension of Seller’s delivery commitment, as applicable.

8.3 Payment Terms

Payment Terms are Net 30 Days from the date of shipment from Seller unless otherwise noted. All shipping and handling costs will be charged to Buyer on a Prepaid and Add basis. Buyer has the option of paying for shipping by charging its own account with a carrier

8.4 Warranty Information and Guidelines

The Seller warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment. The warrant covers both parts and labor excluding calibration and expendable parts such as filters, detector tubes, batteries, etc. If the inspection by the Seller confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States. The Seller shall not be liable for any loss or damage caused by the improper use or installation of the product. The Buyer indemnifies and saves harmless the Seller with respect to any loss or damages that may arise through the use by the Buyer or others of this equipment. This warranty is expressly given in lieu of all other warranties, either expressed, implied or statutory, including that of merchantability, and all other obligations, or liabilities of ENMET, LLC for damages arising out of or in connection with the use or repair or performance of the product. In no event shall ENMET, LLC, be liable for any indirect, incidental, special or consequential damages or for any delay in the performance by ENMET, LLC, which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representatives or other persons to assume for it any obligation or liability other than that which is set forth herein.

Buyer agrees to indemnify and save harmless Seller for any damage or loss from lawsuits against Seller by reason of manufacture of sale of materials, parts, or use of processes resulting from Buyer’s design specifications. Any patent, design, pattern, tool, die, jig, fixture, drawing, test equipment, or process furnished by Seller; whether possessed by the Seller before the date of this quotation, or devised or acquired by Seller during performance of the terms of this quotation, shall remain the property of the

Seller except by specific stipulation on the face hereof. Seller reserves the right, without liability, for damage or loss, to destroy

Buyer’s drawings, specifications, patterns and special tools supplied by Buyer for performance of the terms on the face hereof, unless Buyer gives notice of the disposition of such items.

8.5 Return Policy

All returns for credit must be approved in advance by ENMET, LLC. Such returns are subject to a minimum $50.00 or 20% restocking charge, whichever is greater. Approval of equipment for return is totally at the discretion of ENMET, LLC. All requests for return/exchange must be made no later 30 days of the original shipping date from ENMET. The actual amount of any resulting credit will not be determined prior to a complete inspection of the equipment by ENMET. Calibration gas cylinders cannot be returned or restocked.

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 36

Rev. 1 – December 15, 2016

GSM-60

9.0 Instructions for Returning an Instrument for Service

ENMET

Contact the ENMET Service Department for all service requests.

Phone: 734-761-1270

Email: [email protected]

Fill out the “Service Request Form” found at the end of this manual and return with your instrument for all needs. Please send your instrument for service to the site in which the product was purchased. A new “Service Request Form” may be requested if the one found in the manual is not available. All instruments should be shipped prepaid to ENMET.

Address for Service:

Michigan Location:

ENMET

Attention: Service Department

680 Fairfield Court

Ann Arbor, MI 48108

Kentucky Location:

ENMET

62 Corporate Court

Bowling Green, KY 42103

Providing the “Service Request Form” assists in the expedient service and return of your unit and failure to provide this information can result in processing delays. ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered after a period of 60 days may, be disposed of or returned unrepaired COD and the customer will be expected to pay the evaluation fee. Serviced instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.

NOTE : Warranty of customer installed components.

For Warranty Repairs, please reference ENMET’s “Warranty Information and Guidelines” (found earlier in this section).

Manual Revision Date – June 15, 2017

Manual Part No. – 80003-600

P a g e | 37

Rev. 1 – December 15, 2016

Mailing/Shipping Address:

ENMET

680 Fairfield Court

Ann Arbor, MI 48108

[email protected]

Phone: 734.761.1270

Fax: 734.761.3220

Product Name or Number:

Product Serial Number:

Describe Problem or Needed Service:

Service Request Form

Warranty Claim?

Yes

No

Billing Address:

Contact Name:

Email:

PO/Reference

#:

CUSTOMER INFORMATION

Shipping Address:

Phone #:

Fax #:

COD

Card Number

Name as it Appears on

Card:

PAYMENT METHOD

VISA/MasterCard

Exp. Date

American Express

Security Code:

UPS Ground

RETURN SHIPPING METHOD

UPS 3 Day

UPS Next Day

UPS ND Air

Select Air Saver

UPS Account #:

FedEx Ground

FedEx Air

Express Saver

FedEx Account #:

Insure Shipment:

Yes

No

FedEx Air

Overnight Std.

FedEx Air 2

Day

Insurance

Amount:

$

UPS 2 Day Air

FedEx Air

Overnight P-1

ENMET

Service Request Form

Rev.2 – 9/15/2016

advertisement

Related manuals

advertisement

Table of contents