Cooper Atkins 15400 NotifEye Software Instruction manual
The Cooper Atkins 15400 is a wireless temperature monitoring system that helps you stay informed about the temperature of your environment. With its easy-to-read display and simple setup, you can quickly monitor temperatures and receive alerts when they fluctuate outside of your desired range. The system includes a gateway that connects to your Wi-Fi network and up to four sensors that can be placed in different locations to monitor temperatures remotely. You can access the system's data and configure alerts through a user-friendly online interface.
Reklama
Reklama
EYE
POWERED BY TEMPTRAK
TM
NotifEye ™
Wireless Sensors and Gateway Instructions
1
Contents
Creating an Account
• Creating An Account...................................................2
2
Setting Up the Hardware
• Setting Up and Using the Gateway.............................3
• Inserting Batteries.......................................................4
3
Setting Up the Online System
• Logging into the System ............................................4
• Adding Sensors, Gateways and Networks..............4-5
• Overview of the NotifEye System............................6-7
• Configuring Sensors...................................................7
• Setting Notifications...............................................8-10
• My Account Settings.................................................10
4
Mounting Sensors
• Mounting Sensors......................................................11
5
Advanced
• Adding, Editing and Setting User Permissions.....11-12
• Sensor Maps.............................................................12
• Sensor History and Chart Views...............................13
• Viewing and Exporting Sensor Data.........................14
• System Reports....................................................14-15
User Information
................................................15
Warranty Information
.....................................15
Identify Your Kit
Kit #15906
• (4) Temperature Sensors #15200
• (4) CR123A 3.6V Lithium batteries (Included)
Kit #15907
• (2) Temperature Sensors #15200
• (2) CR123A 3.6V Lithium batteries (Included)
Each Kit includes:
• (1) Gateway #15515
- (1) Buffer
- (1) Receiver
• (1) Buffer Cable (25’)
• (1) Ethernet Cable
• (1) Reference Guide
• 1 Year of Hosting
Creating An Account
1. In a web browser, navigate to: notifeye.digitalcoldchain.com
2. Click the “Get Started Here” button. Please ensure all fields are completed.
3. Account Information.
Key Fields:
• Time Zone - To select a time zone, locate the city closest to you within your time zone. This can be updated later if needed.
4. Primary Contact Information
Key Fields:
• User Name - this is automatically generated during the initial setup. It consists of the first and last name that is entered into the name fields separated by a period.
This username will be set as an administrator.
• Password - must be at least 8 characters
Note: We recommend writing down your username and/or password on the reference guide that shipped with your kit and keeping it in a secure location.
5. Assign Network
Key Fields:
• Network Name - enter a name for your first network (e.g., Main Kitchen).
6. Assign Gateway (referred to as “base station” in the software and is comprised of two components
- a Buffer and a Receiver)
Key Fields:
• Gateway – select Base Station from the dropdown list.
• Serial Number – enter the serial number from the Buffer label.
• Mac Address – enter the Mac address, including colons.
• Assign to Network – select the network to which the Buffer will be assigned.
7. Gateway successfully setup
Once you have successfully created an account and assigned your first gateway, you will get a
“congratulations” screen. Click on the “Manage” button to continue setting up your NotifEye system.
At this point, we recommend you set up your hardware.
2 Contents Page
Setting Up the Hardware
Setting Up and Using the Gateway
• Plug one end of the Ethernet cable into the available Ethernet port on the Buffer ( 1 ); plug the other end into an available ethernet port.
• Plug the power cable ( 2 ) into a power outlet.
• Place gateway in a secure location.
2
1
Understanding the Gateway Lights
Note: When the Gateway is first powered, the lights 1 and 2 will turn on for a moment, turn off, and then turn back on after 30 – 60 seconds. When this cycle is complete, the Buffer is connected. Light 1 will be green and Light 2 will be amber.
Lights
Configuring The Gateway
The Gateway collects data from all sensors within range. If the internet connection is interrupted, the gateway’s internal storage will store readings to prevent data loss. Once the internet connection is re-established, any stored data is automatically transmitted. Since the gateway requires power for this operation, a battery backup is recommended in the event of a power outage.
Contents Page 3
Inserting Batteries
1
2
Remove the sensor cover.
Insert the battery into the battery slot ( 1 ).
Press the reset button ( 2 ) to initialize the transmitter.
Replace the sensor housing.
Manual Sensor Reset Process:
If you want to force a reading to be sent, remove the transmitter cover and press the reset button
Important: Do not mount sensors before entering their IDs and codes into the online system (see page 5).
Setting Up The Online System
Logging into the System
On the home page, enter your username and password you just created
(page 2, #4.) in the login fields and click the “Login” button.
Adding Sensors, Gateways and Networks
These steps will allow you to add sensors, gateways and networks to your account. If you have additional gateways you can either set them up as their own individual networks or add them to your existing network and manage them from one account.
1. Adding Sensors
• Choose the “Manage” tab from the main navigation bar.
• From the “Network” dropdown on the right, select the network to add the sensor to.
4 Contents Page
• In the section “Sensor List”, click the “Add Sensor” button.
• Enter the Sensor ID and Sensor Code from the label on the sensor.
• Press the “Assign Sensor” button. The sensor will now appear in the sensor list.
Repeat this process to add more sensors to this network.
2. Adding Additional Gateways
• Choose the “Manage” tab from the main navigation bar.
• From the “Network”dropdown on the right, select the network to which you would like to add the gateway.
• In the section “Gateway List”, click the “Add Gateway” button.
• From the Gateway or Base Station dropdown, select Base Station.
• Enter the Serial Number and Mac Address (including colons) from label of the Buffer you want to add, and select the network to which it should be added. .
• Press the “Assign Gateway” button. The gateway will now appear in the gateway list.
• To change the default name of this Buffer, (e.g., Buffer - 1232343456) in the “Gateway List” pane, click the “Edit Gateway Details” icon .
• Enter a new name and then click “Save” and the Buffer will appear in the settings list.
3. Creating Additional Networks
• Choose the “Manage” tab from the main navigation bar.
• From the “Network” pane, click the “Add Network” button and a new network will appear in the list.
• To change the default name of this network, (e.g., Network 3355), click the “Edit Network” button.
• Enter a new name and then click “Save” and the network will appear in the settings list.
Contents Page 5
Overview of the NotifEye System
The Online Interface
When you have logged in to the online system, click on the “Overview” tab. This is the default view.
2. Sensor List
1.Main Navigation 5. My Account
3. Manage Gateways
6. Current Sensor Info
1. Main Navigation
Overview - click to return to “Home” view. This shows all the sensors and gateways.
Mouse over and click to select a particular network from the dropdown menu.
Notifications - click to view Notifications.
Manage
Reports
Sensor Maps - click to view a visual map of your active sensors. (Upload an image file and then
Support
- click to add and delete Networks, Gateways and Sensors.
- click to access system reports (sent notifications and access logs) for your account.
visually place your active sensors onto it)
- click to access additional support (i.e. manuals, video tutorials, etc.)
2. Sensor List
Displays all sensors that are currently assigned to your sensor network. Clicking on the sensor names allows you to view information for that specific sensor. Clicking the edit button by a sensor’s name allows you to change the sensor specific settings such as sensor name and heartbeat.
3. Manage Gateways
Displays all Gateways that are currently assigned to your sensor network.
4. Sensor Status Indicators and Icons
Displays the status for each individual sensor.
Sensor is checking in and within user defined safe parameters
Sensor has met or exceeded temperature limits or triggered a notification
Sensor has not checked in
No sensor readings since shipping
No sensor readings will be recorded (Inactive)
5. My Account
Click to display and edit account information.
6. Current Sensor Information
Displays the most current information of the selected sensor, including: last check-in, signal strength, battery power and last sensor reading.
6 Contents Page
Sensor Data Window
Main navigation to display information about the sensor. To display, click on a sensor from the sensor list.
Date Range Selector
Date Range Selector
Allows you to choose the date range for viewable information such as sensor history, notifications sent, charts and sensor data export.
Select a tab to change between:
History
Chart
Notifications - allows you to view, add, edit or delete notifications for the sensor and view/add
Export
Edit
Scale
- displays a history (in list form) of the sensor’s data.
- displays a graphical view of the sensor’s data.
corrective actions
- allows you to archive data by exporting as a .csv file.
- allows you to change configurations such as sensor name and heartbeat.
- allows you to choose the temperature scale (Fahrenheit or Celsius)
Note: The tab highlighted in green is your current selection.
Configuring Sensors
To set a sensor’s configurations, from the Sensor List, click the sensor that you would like to configure, and then click the “Edit” tab. The Configuration window allows you to change the name of the sensor and set the heartbeat (how often the sensor checks-in with the software - 10 minutes is default/recommended setting. When you have finished making changes, press the “Save” button at the bottom of this section.
Note: Be sure to click the “Save” button any time you make a change to any of the sensor parameters.
Sensor Name
The name of the assigned sensor, e.g., Freezer 1.
Heartbeat Interval
The default Heartbeat Interval.
Contents Page 7
Setting Notifications
Automated notifications can be set up to alert you via SMS text or email if a wireless sensor meets a set threshold or condition. To create a new notification or edit/delete an existing notification, click on the “Notifications” tab in the main menu area of the site.
1. The Notification List Window
Notifications
Tab
Create a New
Notification
Enable / Disable a Notification
Edit an Existing
Notification
Delete
Notification
2. Creating a New Notification
Click on the “Add New Notification” button.
There are four notification options:
Class of Notification
There are four notification options available when creating a new notification.
• Sensor Reading Notification: Application notifications are sensor specific (temp sensor = trigger alert when temp is above 70°F, etc.). The notification you create will be based on the selected sensor type.
• Battery Notification: Allows users to define a battery power percentage level that will trigger an alert from the system, warning them to replace batteries.
• Inactivity Notification: Set up “Inactivity” notifications to alert you when your sensors and/or gateways have stopped communicating with the servers. Failure to set up an “Inactivity” notification will result in no email/
SMS text being sent should your sensors stop communicating with the servers. This can also be used for inactivity in the gateway.
• Advanced Notification: Allows the user to set notifications based on more advanced rules.
Advanced Temperature Range: Sends a notification any time a sensor exceeds set measurement limits for a designated period of time.
Advanced Humidity: Sends a notification any time a sensor exceeds set measurement limits for a designated period of time.
Advanced Open/Close: Sends a notification any time a sensor exceeds set measurement limits for a designated period of time.
8 Contents Page
When you are finished, click the “Create” button.
3. Setting and Editing Notifications
You will need to choose what sensor type you are creating the alert for either temperature, humidity or open/close .
Notification Name
Name of Notification.
Notification Text
Message that will be displayed when notification is sent.
Notify After __ Minutes
Defines a delay, in minutes, to wait before sending a notification. If a normal reading (equipment returned to valid state) is recorded during the delay time, no notification will be sent out.
Less than Temperature
Allows you to set the minimum limit. Any readings under this value will trigger a notification.
Greater than Temperature
Allows you to set the maximum limit. Any readings above this value will trigger a notification.
Schedule Notification Times
Schedules notifications for the entire week. Default is set to always notification.
Don’t Alert again for (snooze )
Defines how frequently you want to be notified once the alert is active. Default is set to 60 minutes.
Snooze each trigger
Sensor will alert independent of other sensors. Default is set to independently.
Notification is Active
Allows you to turn off a notification temporarily without deleting it. When you are finished, click the
“Continue” button.
4. Assigning Notifications Delivery Settings
Contents Page 9
Sent From
Each notification needs to be assigned to a sensor. From the Sensor List panel on the left, choose the sensor to attach to this notification by clicking in the checkbox next to the sensor name then clicking the >> button.
People that will receive this notification
For each person that you want to receive notifications, select from the User List panel on the left or type a name into the search box and the system will automatically populate the name of a user within your sensor network. Select the name of the user for the notification.
5. Deleting a Notification
To delete a notification, in the “Notification List” pane, click on the name of the notification you wish to edit and at the bottom of the pane, click on the “trashcan” button to remove the notification.
6. Adding Corrective Actions to Sent Notifications
If a user receives a notification, they can document any corrective actions taken to resolve the issue.
Selecting the “Notifications” tab under the transmitter displays the recently sent notifications. Here you can “Add” a corrective action or “View” any corrective actions that have already been recorded.
My Account Settings
Clicking on the “My Account” link in the top right of the screen will open the Settings pane where you have access to change account information and setup account users.
Account Information
Information that can be edited include: Company Name, Primary Contact, Time Zone and Address.
It will also tell you the date that your account needs reviewing.
10 Contents Page
Mounting Sensors
Sensors should be placed on a clean, dry, flat surface. Place them at least 3 ft away from each other and any wireless gateway for clearest transmissions.
Do not place the sensor body inside any refrigerator/freezer cabinet or storage area since this will interfere with sensor performance.
• Peel off the wax paper from the sticky back of the foam tape that is attached to the sensor and mount the sensor just outside the door of the unit to be monitored, being sure the surface is clean, dry and flat.
• Determine final placement of the probe (back corner/middle of cooler/freezer ceiling), e.g., don’t place near door or fan or boxes where employees may be stocking/removing product.
• Run the sensor probe through the door seal.
• If installing through the cooler wall, start by drilling a hole just larger than the size of the temperature probe in the seal or wall.
• Push the temperature probe and wire through the hole.
• Use clips or twist ties to secure the probe wire to the interior wall or shelving.
• Once the probe is placed in the desired location, use silicone caulking to seal the hole around the wire.
Advanced
Adding, Editing and Setting User Permissions
Adding New Users
To add new users, click on the “My Account” link at the top right hand of the screen.
From the “My Account” page, click the “Add New User” button.
Note: In order to make changes to users, ensure you have administration privileges to the account.
Fill in their corresponding information and assign them a username and password.
If you want this user to receive cell phone or text alerts from the system, you’ll need to enter their cell phone number and cell carrier from the dropdown.
Contents Page 11
If you are creating another administrator, who will have permission to edit information, you’ll need to make sure to check the “Is Administrator” box. If this user is only to have specific permissions, you’ll need to go back and make these changes once the user has been added to the account.
Once you click the “Save New User” button, you will be taken back to the Account settings page where your new user will be displayed.
Editing Existing Users
To make edits to a user, in the “Users” list, locate the user and click on the name of the user you wish to edit. From here you can change their notification settings, reset their password, edit their personal contact information and edit their permissions. If you want to make the user an admin, checking “Is
Administrator” will give them full administrative access to view and make changes to the account.
Setting User Permissions
If this user is to have limited permissions, you’ll need to make these changes once the user is added to the account. To make changes to an existing account users permissions, click on the “Permissions” tab. You can choose which networks each user can see and whether they are allowed to make any changes to network & sensor settings by checking and unchecking the boxes next to each one.
After making changes, select “Same Permissions”.
Sensor Maps
The first time the Sensor Map tab is selected you will be prompted to choose and upload an image of your facility from your own files. Your sensors can be superimposed onto this image to represent their physical location in the facility.
Load Data Snapshot gives an overview of the last readings for all sensors shown on map.
12 Contents Page
Clicking and dragging with the mouse allows you to view different portions of the map. To place a sensor on the map, under the Sensor List, click the “Add to Map” link next to the chosen sensor, or click and drag the sensor image to its desired location. Mousing over any sensors that are on the map will display the last reading while clicking a sensor will display the sensor information window.
To zoom in and out of the map, click on the “+” and “-” buttons in the bottom right of the image map.
Remove a sensor by right-clicking on the sensor and select “Remove From map”. To add additional maps, in the Sensor maps pane, click “New” link.
Note: All your available sensors will be listed in the left, Sensor List pane. To only view sensors in a particular network, select the network from the dropdown box.
Sensor History and Chart Views
The tab highlighted in green is your current selection.
Select the “Overview” tab in the main navigation bar. Click on a sensor name from the sensor list to bring up the sensor navigation tabs. Clicking on the “History” or “Chart” tabs within the sensor data window allows you to view the sensor data history as text or in a graphical chart for the date range specified in the top right corner. To change the date range of the viewable information, click on the date range box at the top right of the sensor data window.
Note: If you view data immediately after setting up your system, the chart will not have enough readings to create a graph; you must wait until your sensor has checked in multiple times.
Contents Page 13
Zoom
Zoom allows you to view a particular portion of the top graph. When you click and drag with your mouse between two points on the graph, a magnified view will appear in the top graph area. Click
“Reset Zoom” to restore to the default view.
Clicking and dragging between two points on the graph will select the zoom area. This area will be shown in the top graph .
The Account Report gives you a list of all sensors on the account with the associated network, corresponding ID# and security code. The Battery Health Report shows the current battery life of all sensors and visual indicators.
Viewing and Exporting Sensor Data
Exporting to .csv file
Clicking on the “Export” tab within the sensor data window allows you to export sensor data to a comma separated value (.csv) file or send the sensor data to an external web source. To export sensor data you must first select the date range for the data you want to export. Once the date range is selected, determine whether you want sensor data from the selected sensor only, from all sensors in the network or all sensors assigned to the account.
When you are finished, click on “Export Data” at the bottom of this window. The data will be exported to a comma separated value (.csv) file for use in spreadsheet software such as Microsoft Excel®.
Depending on your browser settings you may be prompted for a save location. If not, the file will be downloaded to your browser’s default download directory.
Note: Only the first 5,000 records within the selected date range can be exported at a time.
System Reports
The NotifEye system’s “Report” tab allows you to view four types of reports: Sent Notifications,
Access Log, Account Report, and Battery Health. To create these reports, click on the “Reports Tab” in the top navigation bar.
14 Contents Page
To view sent notifications, simply click on Sent Notifications from the left panel and enter the preferred date range (default range is one week). Once you hit submit, it will show all notifications sent regarding your wireless system. If you choose the Access Log for a specified timeframe, you will see an access log showing all users that have logged in to the account, and the date & time of login.
The Account Report gives you a list of all sensors on the account with the associated network, corresponding ID# and security code. The Battery Health Report shows the current battery life of all sensors, complete with percentages and visual indicators.
User Information
For detailed instructional videos on how to use your NotifEye™ Wireless Sensors or the NotifEye™
Online System, please visit us at: www.cooper-atkins.com/notifeye/support
This equipment has been tested and found to comply with the limits for a Class B digital devices, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation.
WARNING
Changes or modifications not expressly approved by Cooper-Atkins could void the users’ authority to operate the equipment.
Limited Warranty
Any instrument which proves to be defective in material or workmanship within one year of original purchase will be repaired or replaced without charge upon receipt of the unit prepaid with proof of purchase. This Limited Warranty does not cover damage in shipment or failure caused by tampering, obvious carelessness, or abuse, and is the purchaser’s sole remedy.
Cooper-Atkins Corporation
33 Reeds Gap Road • Middlefield CT
06455-0450 U.S.A.
Email: www.cooper-atkins.com
For all Technical Support Contact:
I-Care Customer Service Support
Tel: 888.533.6900, Opt. #2
Email: [email protected]
Contents Page
67-1593
V0618
15
Stáhnout
Reklama