Mettler Toledo IND231-IND236 Operating Instruction
Add to my manuals
121 Pages
Mettler Toledo IND231-IND236 is an industrial weighing terminal designed to provide accurate and reliable weighing results in various applications. With its advanced features and robust construction, it is ideal for use in demanding environments where precision and dependability are paramount.
advertisement
IND231/IND236
Weighing Terminal
IND231/IND236 Weighing Terminal
Essential Services for Dependable Performance of Your IND231/IND236 Weighing Terminal
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use of your new equipment according to this Manual and regular calibration and maintenance by our factorytrained service team ensures dependable and accurate operation, protecting your investment.
Contact us about a service agreement tailored to your needs and budget. Further information is available at www.mt.com/service .
There are several important ways to ensure you maximize the performance of your investment:
1.
Register your product: We invite you to register your product at www.mt.com/productregistration so we can contact you about enhancements, updates and important notifications concerning your product.
2.
Contact METTLER TOLEDO for service: The value of a measurement is proportional to its accuracy – an out of specification scale can diminish quality, reduce profits and increase liability. Timely service from METTLER TOLEDO will ensure accuracy and optimize uptime and equipment life. a.
Installation, Configuration, Integration and Training: Our service representatives are factorytrained, weighing equipment experts. We make certain that your weighing equipment is ready for production in a cost effective and timely fashion and that personnel are trained for success. b.
Initial Calibration Documentation: The installation environment and application requirements are unique for every industrial scale so performance must be tested and certified. Our calibration services and certificates document accuracy to ensure production quality and provide a quality system record of performance. c.
Periodic Calibration Maintenance: A Calibration Service Agreement provides on-going confidence in your weighing process and documentation of compliance with requirements.
We offer a variety of service plans that are scheduled to meet your needs and designed to fit your budget. d.
GWP ® Verification: A risk-based approach for managing weighing equipment allows for control and improvement of the entire measuring process, which ensures reproducible product quality and minimizes process costs. GWP (Good Weighing Practice), the sciencebased standard for efficient life-cycle management of weighing equipment, gives clear answers about how to specify, calibrate and ensure accuracy of weighing equipment, independent of make or brand.
© METTLER TOLEDO 2021
No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording, for any purpose without the express written permission of METTLER TOLEDO.
U.S. Government Restricted Rights: This documentation is furnished with Restricted Rights.
Copyright 2021 METTLER TOLEDO. This documentation contains proprietary information of METTLER
TOLEDO. It may not be copied in whole or in part without the express written consent of METTLER
TOLEDO.
COPYRIGHT
METTLER TOLEDO ® is a registered trademark of Mettler-Toledo, LLC. All other brand or product names are trademarks or registered trademarks of their respective companies.
METTLER TOLEDO RESERVES THE RIGHT TO MAKE REFINEMENTS OR CHANGES
WITHOUT NOTICE.
FCC Notice
This device complies with Part 15 of the FCC Rules and the Radio Interference Requirements of the
Canadian Department of Communications. Operation is subject to the following conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his or her expense.
Declaration of Conformity is available at http://glo.mt.com/global/en/home/search/compliance.html/compliance/ .
Statement regarding harmful substances
We do not make direct use of harmful materials such as asbestos, radioactive substances or arsenic compounds. However, we purchase components from third party suppliers, which may contain some of these substances in very small quantities.
Warnings and Cautions
• READ this manual BEFORE operating or servicing this equipment and FOLLOW these instructions carefully.
• SAVE this manual for future reference.
WARNING
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT THE AC VERSION OF THE
IND231/IND236 TERMINAL TO PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE
GROUND PRONG.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON.
FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
WARNING
THE IND231/IND236 IS NOT DESIGNED FOR USE IN AREAS CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. DO NOT INSTALL AN
IND231/IND236 INTO AN EXPLOSIVE ENVIRONMENT.
WARNING
WHEN THIS EQUIPMENT IS INCLUDED AS A COMPONENT PART OF A SYSTEM, THE RESULTING
DESIGN MUST BE REVIEWED BY QUALIFIED PERSONNEL WHO ARE FAMILIAR WITH THE
CONSTRUCTION AND OPERATION OF ALL COMPONENTS IN THE SYSTEM AND THE POTENTIAL
HAZARDS INVOLVED. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN BODILY
HARM AND/OR PROPERTY DAMAGE.
CAUTION
THERE IS A RISK OF EXPLOSION IF THE BATTERY IS REPLACED WITH AN INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS.
CAUTION
THE NiMH BATTERY USED IN THIS DEVICE MAY PRESENT A RISK OF FIRE OR CHEMICAL
BURN IF MISTREATED. DO NOT CRUSH, DISASSEMBLE, HEAT ABOVE 60 ° C OR INCINERATE.
REPLACE BATTERY WITH 30044650 ONLY. USE OF ANOTHER BATTERY MAY PRESENT A RISK
OF BURN, FIRE OR EXPLOSION.
CAUTION
USE CAUTION WHEN TESTING THE BATTERY. A LARGE AMOUNT OF CURRENT MAY BE
PRESENT IN THE BATTERY.
CAUTION
DISPOSE OF USED BATTERY PROMPTLY. KEEP AWAY FROM CHILDREN. DO NOT
DISASSEMBLE AND DO NOT DISPOSE OF IN FIRE.
CAUTION
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR
INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER
AND WAIT AT LEAST THIRTY (30) SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS
ARE MADE. FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN DAMAGE TO OR
DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
NOTICE
DO NOT ATTEMPT TO CHARGE THE NI-MH BATTERY IF THE BATTERY TEMPERATURE IS BELOW 0 ° C (32 ° F).
CHARGING IS NOT POSSIBLE AT OR BELOW THIS TEMPERATURE. DO NOT OPERATE THE BATTERY CHARGER
OUTSIDE ITS TEMPERATURE RANGE OF 0 ° C (32 ° F) TO 40 ° C (104 ° F).
NOTICE
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE IND231/IND236 TERMINAL AND
WAIT AT LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY HARNESS.
NOTICE
OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES.
Disposal of Electrical and Electronic Equipment
In conformance with the European Directive 2012/19/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Contents
1 Introduction .................................................................................. 1-1
1.1.
IND231/IND236 Overview ............................................................. 1-1
1.2.
Use in Hazardous Areas ................................................................. 1-2
1.3.
Specifications ............................................................................... 1-2
1.4.
Model Identification ....................................................................... 1-4
1.5.
Inspection and Contents Checklist ................................................... 1-5
1.6.
Physical Dimensions ..................................................................... 1-5
1.7.
Main PCB .................................................................................... 1-6
1.8.
Scale Bases ................................................................................. 1-7
1.9.
Options ........................................................................................ 1-7
1.10.
Display and Keyboard ................................................................... 1-8
2 Operation ..................................................................................... 2-1
2.1.
Overview ...................................................................................... 2-1
2.2.
Keypad Operation and Display Elements .......................................... 2-1
2.3.
Configurable Function Keys ............................................................ 2-5
2.4.
Basic Functionality ........................................................................ 2-6
2.5.
Remote Display Functionality ........................................................ 2-11
2.6.
Applications ............................................................................... 2-14
3 Configuration ................................................................................ 3-1
3.1.
Accessing Setup Mode ................................................................... 3-1
3.2.
Setup Menu Structure ..................................................................... 3-5
3.3.
Configuration Overview .................................................................. 3-6
3.4.
F1 Scale ...................................................................................... 3-6
3.5.
F2 Application ............................................................................ 3-20
3.6.
F3 Terminal ................................................................................ 3-29
3.7.
F4 Communication ..................................................................... 3-33
3.8.
F5 Maintenance .......................................................................... 3-38
4 Service and Maintenance ............................................................... 4-1
4.1.
Cleaning and Maintenance ............................................................. 4-1
4.2.
Service ........................................................................................ 4-1
4.3.
Battery Operation .......................................................................... 4-2
4.4.
Troubleshooting ............................................................................ 4-5
4.5.
Connection to InSite™ SL ............................................................. 4-11
Document 30094013 | Revision 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 1
A Parameter Values .......................................................................... A-1
B.
Communications .......................................................................... B-1
B.1.
Serial Interface Parameters ............................................................. B-1
B.2.
Demand Output Mode.................................................................... B-1
B.3.
Continuous Output Mode ................................................................ B-4
B.4.
CTPZ ........................................................................................... B-7
B.5.
Standard Interface Command Set (SICS) Protocol .............................. B-7
C.
GEO Codes .................................................................................. C-1
C.1.
Original Site Calibration ................................................................. C-1
C.2.
New Site GEO Code Adjustment ...................................................... C-1
2 METTLER TOLEDO IND231/IND236 User's Guide 30094013 | 04 | 04/2021
1 Introduction
The IND231/IND236 industrial scale terminal provides a compact yet flexible solution for a variety of weighing needs. Available as either AC powered for stationary applications or battery-powered for portable applications, this terminal is at home in virtually any industrial environment.
Both 2mV/V and 3mV/V load cells are supported without the need for any configuration change.
The IND231/IND236 delivers precision measurement data from grams to kilograms in a single, cost effective package.
Standard applications include basic weighing, animal weighing, Over/Under checkweighing, counting and accumulation. Pre-defined application-specific print templates including date and time stamp, provide transmission of weighing transaction information to printers and PC software applications.
1.1.
IND231/IND236 Overview
1.1.1.
Standard Features
•
•
•
•
•
•
•
•
•
•
•
•
Easy to handle plastic enclosure design for the IND231, rugged stainless steel enclosure for the IND236
Supports one analog load cell platform with up to four 350Ω load cells
Can be used as a remote display to show weight from a separate main terminal
Large seven segment LCD display with white backlight and graphic icons for clear readability in varying light conditions
Powered by either 100-240 (nominal) V AC or internal battery pack
One standard serial port (COM1) for asynchronous, bidirectional communication
Support for one of the following option boards:
COM2 electrically isolated RS232, RS422 and RS485 serial Interface
USB device Interface for connection to a PC
Discrete I/O interface
Front panel key access to basic weighing functions – zero, tare, clear, configurable function and print
Selectable primary unit of measure including grams, kilograms, pounds and ounces
Selectable second unit of measure including grams, kilograms, pounds and ounces
Backup and restore of configuration and calibration settings, using InSite ® SL tool
WeighSync™ ST simple data integration software for weighing terminals
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 1-1
1.1.2.
• Automatic shutoff and backlight timeout features to help conserve energy and extend battery life when using the battery powered version
IND231/IND236 Terminal Types
The terminal is available in the following four versions:
• IND231plastic housing, AC power (also designed for use with AA size alkaline cells)
• IND231plastic housing, Rechargeable Ni-MH Battery power
• IND236 stainless steel housing, AC power
• IND236 stainless steel, Rechargeable Ni-MH battery power
1.2.
Use in Hazardous Areas
WARNING
THE IND231/236 TERMINAL DOES NOT HAVE ANY HAZARDOUS AREA APPROVALS! DO NOT USE
WITHIN AREAS CLASSIFIED AS HAZARDOUS DIVISION 1/2 OR ZONE 0/1/2/21/22 BECAUSE OF
COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES.
The IND231/236 terminal does not have any hazardous area approvals for use directly in explosive environments; however, it is capable of operation with load cells and platforms in a Division 1/2 or
Zone 0/1/2/21/22 hazardous area when used with approved barriers or when purging is utilized.
Contact your authorized METTLER TOLEDO representative for information about hazardous area applications for the IND231/236 terminal.
1.3.
Specifications
The IND231 and IND236 terminal conforms to the specifications listed in Table 1-1 .
Table 1-1: Terminal Specifications
Specifications
Enclosure Type
Product Dimensions
(w × h × d)
Shipping Dimensions
(w × h × d)
Product Weight
Shipping Weight
IND231
Plastic, configurable as desk top or column/wall mount enclosure
IND236
Stainless steel, configurable as desk top or column/wall mount enclosure
220 mm x 150 mm x 102 mm
(8.66 in. x 5.90 in. x 4.02 in.)
220 mm x 150 mm x 93.3 mm
(8.66 in. x 5.90 in. x 3.67 in.)
350 mm x 220 mm x 195 mm
(13.78 in. x 8.66 in. x 7.68 in.)
AC Version: 1.2 kg (2.6 lb)
Battery Version: 1.5 kg (3.3lb)
2.0 kg
AC Version: 2.2 kg (4.9 lb)
Battery Version: 2.5 kg (5.5 lb)
3.0 kg
1-2 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
30094756 | 04 | 04/2021
Specifications
Environmental Protection
Operating Environment
Hazardous Areas
Power
Display
Weight Display
Scale Types
Number of Cells
Number of Scales
Analog/Digital Update Rate
Load Cell Excitation Voltage
Minimum Sensitivity
Keypad
Communication options
Application
IND231 IND236
IP54 IP66/67
Operating temperature range: -10 °C to +40 °C
Storage temperature range:
Relative Humidity:
-20 °C to +60 °C
10-95%, non-condensing.
The terminal cannot be operated in areas classified as Hazardous because of combustible or explosive atmospheres in those areas. Contact an authorized
METTLER TOLEDO representative for information about hazardous area applications.
AC version: Operates at 100-240 VAC, 50-60 Hz and includes a power cord configured for the country of use.
Battery version :
Six "AA" size alkaline batteries or
NiMH rechargeable battery pack (max.
120hrs for one load cell with backlight off)
Battery version :
NiMH rechargeable battery (about
120hrs for one load cell)
40 mm height character, seven digits seven segment LCD with white backlight including weight display, weight units, gross/net indication and symbols for motion, center of zero and variable applications
Maximum displayed resolution of 30,000 divisions
Analog load cells
Up to four 350 ohm load cells(2 or 3 mV/V)
One
Internal analog: 80 Hz
5 VDC
0.5 µV/e
Seven keys: Zero, Tare, Clear, On/Off, Print, F1 & F2 (configurable)
Serial Interfaces
Standard: One serial port (COM1) RS-232, 1,200 to 115,200 baud
Optional isolated serial port: (COM2) RS-232/422/485, 1,200 to
115,200 baud
Discrete I/O Interface
Optional Discrete I/O port: 2 inputs/4 outputs
USB Interface
Optional USB device port
Protocol
Serial Inputs: ASCII commands for CTPZ (Clear, Tare, Print, Zero), SICS
(most level 0 and level 1 commands)
Zero, Tare, Print, X10, Date & Time, Basic weighing, Animal weighing,
Counting with APW enhancement, Over/Under checkweighing with 10 records database, Accumulation, Remote display
METTLER TOLEDO IND231/IND236 User's Guide 1-3
Specifications
Approvals
Accessories
IND231 IND236
Weights and Measures,
USA:
NTEP Class III/IIIL - 10,000d; Cert. #13-049
Canada:
Class III - 10,000d; Class IIIHD - 20,000d, AM-5913
Europe:
Class III, 2 x 3000e and 6000e; TC8351, T8030
OIML:
Class III, 2 x 3000e and 6000e; R76/2006-NL1-13.06
Product Safety
UL, cUL, CE
Wall mount/column mount bracket; desk bracket.
1.4.
Model Identification
The IND231/IND236 model number, factory number and serial number are located on the data
plate of the terminal. Refer to Figure 1-1 to verify the configuration of the IND231/IND236 terminal
when it left the METTLER TOLEDO factory.
1-4 METTLER TOLEDO IND231/IND236 User's Guide
Figure 1-1: IND231 Configuration Chart
30094756 | 04 | 04/2021
1.5.
Inspection and Contents Checklist
Verify the contents and inspect the package immediately upon delivery. If the shipping container is damaged, check for internal damage and file a freight claim with the carrier if necessary. If the container is not damaged, remove the terminal from its protective package, noting how it was packed, and inspect each component for damage.
If shipping the terminal is required, it is best to use the original shipping container. The terminal must be packed correctly to ensure its safe transportation.
The package should include:
• IND231 or IND236 Terminal
• Installation Manual (or Safety Instructions)
• Power cord
• Bag of miscellaneous parts
• User manual (or resource CD)
• Battery Pack (battery version only)
• Mounting brackets (1)
1.6.
Physical Dimensions
The physical dimensions of the IND231/IND236 enclosure are shown in Figure 1-2 and Figure 1-3
in mm and [inches].
30094756 | 04 | 04/2021
Figure 1-2: IND231 Enclosure Dimensions
METTLER TOLEDO IND231/IND236 User's Guide 1-5
Figure 1-3: IND236 Enclosure Dimensions
1.7.
Main PCB
The IND231/IND236 terminal’s main printed circuit board (PCB) provides the analog load cell scale interface, as well as the COM1 RS-232 serial port. The COM1 RS-232 serial port supports bidirectional communications at speeds up to 115200 bps. This port can be used for saving terminal configuration data to a PC via the InSite ® SL tool.
The main board also contains the DC power input connection, display interface, keypad interface, and interface port for the options.
An AC/DC power board supplies the terminal with DC +12V/1A output.
A charging board is standard with the rechargeable battery powered version.
1-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
1.8.
Scale Bases
The IND231/IND236 terminal supports analog scale bases and provides 5 volts of excitation to drive analog load cells. Up to four 350Ω load cells can be powered by the terminal.
A four- or six-wire load cell connection is provided, with sense lines to help maintain accuracy as the load cell cable resistance changes due to variations in temperature.
1.9.
Options
Three kinds of option interface can be mounted on COM2. The connection pin specifications are printed on the PCB.
The following options are available for the IND231 and IND236.
1.9.1.
Isolated Serial interface RS232/RS422/RS485
This optional port provides RS-232 and RS-422/485 communication which can be configured in setup. The port is bidirectional and can be used for various functions such as print, auto print,
Toledo continuous output or SICS communications.
The COM2 port is galvanically isolated for both RS-232 and RS-485, to provide surge voltage protection.
The RS-485 connection can be used as an RS-422 transmit only, when sending continuous output to a scoreboard or remote display.
1.9.2.
1.9.3.
The interface RS422/485 is required for data transmission with higher baud rate over longer distances up to 500 m.
Communication settings can be configured in the terminal setup menu.
Discrete I/O
The discrete I/O interface option provides four dry-contact relay outputs. The relay contacts will switch up to 30 volts DC or 250 volts AC at 1A.
The two inputs are switch selectable as either active (for simple pushbutton control) or passive (for connection to devices that supply their own power for the inputs).
USB
The USB port provided is a UART-USB device bridge acting as a virtual COM port, and is used for communication of serial data to devices such as a PC. The port is bidirectional and can be configured for various functions such as print, auto print, Toledo continuous output or SICS communications. USB external keyboards and bar code scanners are not supported.
The USB driver required to use this port is included on the Resource CD-ROM. A link to the driver files is provided on the Documentation and Utilities page.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 1-7
1.10.
Display and Keyboard
IND231/IND236 uses a transflective type segment LCD display with a white backlight. The main
character height is 40 mm. The front panel, including display and keypad, is shown in Figure 1-4.
The only keypad difference between IND231 and IND236 is the name on the upper right corner of the terminal.
Configurable
Function keys
ON/OFF key
Print Key
1.10.1.
1.10.2.
Figure 1-4: IND231 Front Panel Layout
Display Layout
At the top of display, a single system line displays terminal status such as range number, center of zero, checkweighing status, counting, accumulation and auto accumulation.
Below the system line is the weight display block. During normal, basic weighing operation, the terminal display shows the gross or net weight in the larger 40 mm (1.57 in.) size. When one of the applications is running, prompting messages or menu index characters are also shown in 40 mm (1.57 in.) height. At the right of weight display is a vertical column displaying the gross and net symbols, average piece symbol, and weight units. The star symbol, minus symbol and dynamic symbol appear in a column to the left of the weight display block.
At the bottom of the normal basic weighing display is a block used for assigning functions to and . For display and keypad operation during setup, refer to Chapter 3, Configuration.
Front Panel Keys
The IND231/IND236 terminal operator interface provides a total of seven membrane keys. The print key and five scale function keys (three with fixed function, two with configurable functions) are positioned under the display.
Arrows on the first four keys indicate their use in menu navigation.
The ON/OFF key is located at the bottom left of the display. These keys are used to enter the setup menu, to navigate and select setup elements as described in Chapter 2, Operation.
1-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2 Operation
2.1.
Overview
This chapter provides information about the basic functionality of the IND231/IND236 terminal, including display operation, keypad functions and configurable function keys.
Operation of the terminal varies depending on which functions are enabled, and on the configuration of parameters in setup. Configuration is described in Chapter 3, Configuration.
2.2.
Keypad Operation and Display Elements
Refer to Figure 2-1 for an overview of the layout of the front panel of the IND231/ IND236.
2.2.1.
Keypad Operation
The front panel keys shown in Figure 2-1 are used to operate and configure IND231/IND236.
30094756 | 04 | 04/2021
Figure 2-1: Front panel
METTLER TOLEDO IND231/IND236 User's Guide 2-1
2.2.1.1.
2.2.1.2.
Function Keys
Table 2-1 explains the function of each of the keys during normal operation. Further details about
keypad functions while in the setup menu are provided in Chapter 3, Configuration.
Keys
Table 2-1: Keypad Functions – Normal Operation
Function
FUNCTION 1
FUNCTION 2
Explanation
Function varies depending on what is selected in the menu.
Function varies depending on what is selected in the menu.
ZERO
CLEAR
Transfer /
Print / Enter
Key
Used to reset the displayed weight to Zero.
TARE
Captures current weight as a tare value, and sets terminal to Net mode.
When in the net weight mode, press CLEAR to clear the current tare value; the display will revert to the gross weight value. CLEAR operates regardless of motion on the scale.
Note that once the tare value has been cleared, it cannot be recalled. The complete tare process as described above must be performed.
• Send information to the RS232 interface
• Access to setup menu by long pressing the icon
• Used for printing application
• Press the ENTER key to accept the item or selection and move to the next display.
The ZERO and TARE functions will not operate when there is motion on the scale. If one of these keys is pressed while the scale is in motion, the command will be retained for the programmed number of seconds while the terminal waits for no-motion. If a no-motion condition is not detected within the timeout period, the request is cancelled and discarded.
Direction Keys
The direction keys in Table 2-2 are used for changing focus between on-screen items, for
confirming a selection.
Keys Function
LEFT
RIGHT
Table 2-2: Direction Keys
Explanation
• Move the focus left
• Back to previous menu
• Move the focus right
• Access to the next submenu
2-2 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.2.1.3.
2.2.2.
Keys Function
UP
DOWN
Explanation
• Move the focus up
• Next option
• Move the focus down
• Previous option
Power (On/Off) Key
The power key in Table 2-3 is used for turning the scale on or off. The key needs to be pressed for
more than 2 second for powering off.
Keys Function
On/Off
Table 2-3: Power Key
Explanation
• Turn on or off the terminal
• Exit the setup menu
• Cancel the edit setting under applications/menu
Display Elements
Figure 2-2 illustrates the appearance of alphabetical characters in the segmented display.
ABCDEFG HIJKLMN
OPQRST UVWXYZ
Figure 2-2: Segmented LCD Display of Alphabetic Characters
When in weighing mode, the display is used to indicate weight value and other types of information
related to it. Refer to Figure 2-3.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-3
Figure 2-3: Elements of the Display
The symbols that may appear on the display can be divided into three parts:
• At the top of the display, a single system line displays terminal status.
• Below the system line is the weight display.
• At the bottom of the basic basic weighing display is a block used for showing function selection of keys “F1” and “F2”.
The symbols are described in Table 2-4.
Symbol
Table 2-4: Main Display Symbols
Explanation Symbol
Average piece weight
Explanation
Function keys:
F1 & F2
Expand display Accumulation weight
Auto accumulation weight
Over/Under checkweighing
Center of zero
Unit switch
Check weigh
Recall
Counting Range 1, Range 2
Averaged or x10 weight
Minus value
Dynamic state
Dash and colon
B/G and net
Display counts
Animal weighing
Accumulation
Date & Time
Backlight adjust
Zero
Tare
2-4 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Symbol
Unit
Explanation Symbol Explanation
Key press indicator
Battery status
2.3.
Configurable Function Keys
Functions typically performed by operators can be accessed using the two function keys. The function key assignments are displayed at the bottom of the normal basic weighing display in
IND231/IND236 terminal (Figure 2-4).
2.3.1.
2.3.1.1.
Figure 2-4: Available Function Key Assignments
Functions include access to expanding the displayed weight resolution by 10, Unit switching,
Checkweighing, Recall, Counting, Animal weighing, Accumulation, Date & Time, Zero & Tare and
Backlight, adjustment. The procedure for assigning functions to the keys is described below.
Configuring Function Keys
Assignment from the Weighing Screen
To access the function key assignment interface, press and hold the or key when in basic weighing mode. The normal weight display will change to show an array of function icons, as seen
the F1 or F2 symbol. Use the LEFT and RIGHT direction keys ( and ) to position the arrow
next to the desired function. Figure 2-5 shows the X10 function assigned to F1 (left) and the Switch
Units function assigned to F2 (right).
2.3.1.2.
Figure 2-5: Function Keys Assignment Interface
Once the desired assignment is indicated, press the PRINT key to confirm the selection and return to normal weighing mode.
Assignment from Setup
Enter setup (refer to Chapter 3, Configuration) and access the interface shown in Figure 2-5 at F2 –
Application > F2.1 – Operation > F2.1.1 – Function key 1 and F2.1.2 Function key 2. Once the
interface screen is displayed, the procedure is as described above under Assignment.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-5
2.3.2.
Function keys assignment
Options for the assignment of the FUNCTION keys are described in Table 2-5.
Keys
Table 2-5: Function assignment
Function
Expand display
Unit switch
Checkweighing
Checkweighing Target Recall
Counting
Animal weighing
Accumulation
Date & Time
Zero
Tare
Backlight adjustment
2.4.
Basic Functionality
This section provides information the basical functionality of IND231/IND236. The following functions are addressed in this section:
2.4.1.
• Expand x10
• Brightness adjustment
• Unit Switch
• Zero
• Date & Time
• Tare
Refer to Chapter 3, Configuration, for further information about programming the functionality described in this section.
Expand x 10
This icon toggles the weight display between standard and expanded mode. It is used to increase the selected weight display resolution by one additional digit.
2-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
For example, an extra digit of resolution is added to the main weight display, which then changes
from 2.264 to 2.2645, as shown in Figure 2-6; a star symbol appears at the left of the weight
display to indicate that the weight value is expanded.
2.4.2.
Standard Expanded
Figure 2-6: Standard and Expanded Weighing Modes
To exit expanded mode, either:
Or
• Wait five seconds; the terminal will automatically return to normal resolution.
• Press the x10 function key again
Printing is disabled when the weight is expanded.
Unit Switching
In order to support locations and applications that require multiple measurement units, the
IND231/IND236 supports unit switching. This is illustrated in Figure 2-7, which shows the display
switching from Kg to g. The SWITCH UNITS function enables switching between primary units (the main unit of measure) and alternate units.
Primary unit is set in the submenu F1.2.1. The second unit is set in the submenu F1.6. More information, please refer to Chapter 3, Configuration.
30094756 | 04 | 04/2021
Figure 2-7: Unit Switching: Kilograms to Grams
When the SWITCH UNITS function is assigned to a function key and that key is pressed, the display changes from the primary unit to the second unit. Pressing the function key again returns the display to the primary unit. Each subsequent key press switches between display units.
When units are switched, the units indicator changes to show the selected unit, and the display value is converted. The display division changes to an equivalent weight value in the switched unit
METTLER TOLEDO IND231/IND236 User's Guide 2-7
2.4.3.
2.4.4.
2.4.5.
2.4.5.1.
2.4.5.2.
(for example, from 0.02 lb to 0.01 kg) and the decimal location switches to indicate the conversion.
When switching units, the capacity of the converted units is dictated by the original number of divisions established in the capacity and increments area of setup. In some situations, this may reduce the capacity of the terminal when converting to second unit. If switching would result in a value that is not supported by IND231/IND236, the unit switch will not occur.
Date and Time
The Date & Time function key is only used to display date or time. Configure the Date & Time in setup at F3 - Terminal > F3.3 - Date &Time. When the function key is first pressed, the time will be shown. A second press displays the date. The third press returns the display to weighing mode. To return directly to weighing mode at any time, press the CLEAR .
Backlight Adjustment
The Backlight Adjustment function key is used to set the brightness of the backlight. There are three options: off, low and high. Each press of the key moves the backlight to the next setting.
Zero
When a function key is set to ZERO, it has the same effect as the dedicated zero key .
The Zero function is used to set or reset the initial zero reference point of the terminal. There are three types of zero setting modes:
• Automatic Zero Maintenance
• Power Up Zero
• Pushbutton Zero
When the scale platform or weighbridge is empty, the terminal should indicate zero. The gross zero reference is recorded during calibration. If pushbutton zero is enabled in configuration and the weight is within the zero range, pressing ZERO will capture a new gross zero reference point.
Automatic Zero Maintenance
Automatic Zero Maintenance (AZM) enables the IND231/IND236 to reset to the center of zero in order to compensate for conditions such as terminal or load cell drift or debris on a scale platform.
The AZM operating range is selectable from 0.5, 1, 3 or 10 divisions. Within the configured range, when the terminal is in a no motion condition, AZM makes small adjustments to the current zero reading to drive the weight reading toward the true center-of-zero. The feature does not function when the weight is outside the configured AZM range.
Power up Zero
Power up Zero enables the IND231/IND236 terminal to capture a new zero reference point after power is applied. If there is a motion during power-up zero capture function, the terminal will continue to check for a no-motion condition until zero is captured.
2-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.4.5.3.
2.4.6.
2.4.6.1.
2.4.6.2.
Power up zero can be disabled or enabled, and a range above and below calibrated zero can be configured. The range is selectable from +/-2 %, +/-10 %,+/-20 %, -2 % ~ +18 % or +/-20 % of capacity and can include a positive range and also a range below calibrated zero.
Pushbutton Zero
The pushbutton (semi-automatic) zero function can be activated by pressing the ZERO key , by programming a discrete input, or by issuing a serial command.
Any type of semi-automatic zero can be set to +/-2%, +/- 10% or +/- 20%, relative to the calibrated zero point.
The semi-automatic Zero command can be initiated via a discrete input, or an ASCII ‘Z’ command sent serially (in CPTZ and SICS interface modes).
Tare
When a function key is set to Tare, it has the same effect as the tare key .
Tare is the weight of an empty container. A tare value is subtracted from the gross weight measurement, providing the net weight (material without the container). The tare function can also be used to track the net amount of material being added to or removed from a vessel or container.
In this second case, the weight of the material in the container is included with the tare weight of the container and the display then reflects the net result of any additions to or subtractions from the vessel.
Tare operations in the IND231/IND236 include:
• Automatic Tare
• Tare Interlock
• Clearing Tare
• Keyboard Tare
Automatic Tare
The IND231 can be configured so that tare is taken automatically (auto tare) after the weight on the scale exceeds a programmed tare threshold weight. Auto tare can be configured in setup as enabled or disabled. When auto tare is enabled, the display changes to a zero net weight indication after the weight exceeds the threshold value. Auto tare operations involve:
Auto tare threshold Weight
Auto tare reset
Threshold Weight
When weight on the scale platform exceeds the tare threshold value, the terminal automatically tares.
The reset threshold weight must be less than the tare threshold weight.
When the weight on the scale platform falls below the reset threshold value, such as when a load has been removed, the terminal automatically resets the auto tare trigger.
Clearing Tare
Tare values can be cleared manually or automatically.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-9
2.4.6.2.1.
2.4.6.2.2.
2.4.6.3.
2.4.7.
2.4.7.1.
2.4.7.2.
2.4.7.3.
Manual Clear
Manually clear tare values by pressing the CLEAR key on the keypad when the
IND231/IND236 is in the net mode and has completed the weighing operation. Motion on the scale will not impact a manual clear.
If configured in setup, pressing the ZERO scale function key will first clear the tare, then issue a zero command (see Chapter 3, Configuration, the Scale section, Auto Clear).
Auto Clear Tare
The IND231/IND236 can be configured to clear tare automatically when the weight returns to zero point. Once the tare is cleared, the display returns to the gross weighing mode.
Auto clear is disabled or enabled in setup. Refer to the F1.5, Auto Clear section of Chapter 3,
Configuration, for further information about configuring auto clear.
Keyboard Tare
Pushbutton tare can be configured in setup as enabled or disabled. When disabled, pressing the
TARE key has no effect.
If pushbutton tare is enabled, pressing the pushbutton TARE key initiates a semi-automatic tare. The IND231/IND236 will attempt to perform a tare process. If the process is successful, the display changes to a zero net weight indication and the previous weight on the scale is stored as the tare value. The net symbol is indicated on the display.
The print function can be initiated by pressing the Transfer/Print key or by automatic print settings.
Automatic initiation of a print occurs after the gross weight exceeds the minimum threshold and there is no motion on the scale. After initiation, the gross weight must return below the reset threshold before a subsequent automatic print can occur.
Automatic print may be disabled or enabled. Automatic print can be triggered and reset by weight exceeding thresholds or by weight deviation from a previously stable reading.
Print setting
For more information about the print setting, refer to Chapter 3, Configuration, Communication
Printout format
All the printout formats are fixed and cannot be redefined. Depending on the current function status
of terminal, a different printout format can be selected. Refer to Applications on page 2-11.
Standard Printout
If Tare is zero, the standard printout does not include Tare and Net values.
A star symbol before the weight indicates that the weight is an average, if the current mode is animal weighing.
2-10 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
The single and multi-line outputs appear as follows:
Multi-line output
Single-line output
Date
Time
Gross
Tare
Net
YYY.MM.DD
HH:MM:SS
XX.XXX Unit
XX.XXX Unit
XX.XXX Unit
Date_YYYY.MM.DD_ _ Time_HH :MM :SS_ _
Gross_XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _
Net_XXXX.XXX_Unit
Each number includes eight characters, including the decimal point.
2.5.
Remote Display Functionality
2.5.1.
Introduction
Instead of operating as a normal weight display for an attached weighing platform, the
IND231/IND236 can function as a remote display for another METTLER TOLEDO terminal, sending
Toledo continuous or SICS output. The keypad on the remote display IND231/IND236 can also be used to issue simple Clear, Tare, Print and Zero commands back to the Master terminal when the appropriate key is pressed.
2.5.2.
2.5.2.1.
Setup
Physical Connection of Master and Remote Terminals
Communication for the remote display function occurs over a single serial connection configured for
RS-232, RS-422 or RS-485 communication. Either serial port in the IND231/IND236 can be programmed to accept serial data from the Master and to send keypad commands back to it, so only a single port is necessary for remote display operation. Thus, if the optional COM2 port is installed, one port can be used for communication with the Master terminal while the other port can be configured to provide a demand output to a local printer.
The serial ports must be identically configured in both the Master and the Remote terminal.
2.5.2.2.
2.5.2.2.1.
Remote Display Configuration
Mode and Port Selection
In the IND231/IND236 terminal that will function as the remote display terminal, parameter F.1.1.1
(Scale Type) must be selected as either rCom1 (if COM1 will be used as the data input port) or rCom2 (if COM2 will be used as the data input port).
Figure 2-8 shows the IND231/IND236 COM1 selected in as the remote display input port.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-11
2.5.2.2.2.
Figure 2-8: Scale Type Setting for Remote Operation
Data Format Selection
When Scale Type is set to rCom1 or rCom2, a new parameter F1.1.3 is shown which allows selection of the data format used for the communication. The two possible choices are Toledo
continuous or SICS, as shown in Figure 2-9.
2.5.2.2.3.
2.5.2.3.
Figure 2-9: Connection Configured for SICS and Toledo Continuous Data
Port Parameter Selections
Finally, the serial port parameters must be configured in F4.x.3, where "x" is either "1" for COM1 or a "2" for COM2. The baud rate, data bits and parity selections must be the same in both terminals.
Recommended settings are:
Table 2-6: Serial Port Parameters
Master Terminal Configuration
Parameter
F4.x.3.1 (baud rate)
F4.x.3.2 (data bits / parity)
F4.x.3.3 (flow control)
Value
9600
8 nonE oFF
The Master terminal port used to connect to the remote display terminal must be configured to send weight information either as Toledo Continuous output or as SICS, and this must match the selection in the remote display terminal. It is not possible to send a demand print from the master terminal to the remote terminal.
The serial port parameters for baud rate, data bits and parity must match those selected in the remote display IND231/IND236. Refer to Table 2-6 for the recommended serial port parameters.
Figure 2-10 shows a typical Remote Display configuration, with the Master terminal able to receive
commands sent from the remote IND231/IND236.
2-12 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.5.3.
2.5.3.1.
2.5.3.2.
Figure 2-10: Example of Terminal Configuration for Remote Display Function
If the keypad of the remote display will be used to send CLEAR, TARE, PRINT and ZERO keypad commands back to the Master terminal, then the Master terminal port must support CTPZ command input (if the Toledo Continuous mode was selected) or TAC, T, PRN and Z commands (in SICS mode). When the SICS interface is used, only the SICS commands recognized by the Master will operate properly from the remote display keypad.
Remote Display Operation
Toledo Continuous Operation Overview
Upon power-up, the Master terminal will automatically begin to send weight data to the remote display IND231/IND236. The remote terminal will display the weight as sent from the Master terminal.
When any of the remote terminal’s three scale function keys (CLEAR, TARE and ZERO) is pressed, the corresponding ASCII character is transmitted to the Master terminal. The Master terminal will then act upon the command as appropriate. For example, if the ZERO key is pressed on the remote display a "Z" command is sent to the Master terminal. If the Master terminal is outside of its zero capture range, it will indicate an error on its display but no message is communicated back to the remote display. The remote display simply sends the ASCII command corresponding to the key pressed and it is up to the Master to perform the request or not. The F1 and F2 keys do not send commands to the Master terminal.
SICS Operation Overview
Upon power-up, the remote display will automatically begin asking the Master terminal to send displayed weight and tare weight information using the SI and TA commands. When the Master terminal receives these commands, it will respond by sending the displayed weight and tare weight value to the remote display IND231/IND236. The remote display will display the weights as sent from the Master terminal.
When any of the three scale function keys (CLEAR, TARE and ZERO) is pressed on the remote display, the corresponding SICS command is transmitted to the Master terminal. The Master terminal will then act upon the command as appropriate. When using SICS, the Master terminal will always send the status of a command back to the remote display. The status will indicate success or failure of the command. The F1 and F2 keys do not send commands to the Master terminal.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-13
2.5.3.3.
2.5.3.4.
The remote display will evaluate the Master terminal’s response. If the Master terminal was not able to complete the command successfully, the remote terminal will display seconds.
for about 3
For example, if the ZERO key is pressed on the remote display, a "Z" command is sent to the Master terminal. If the Master terminal is above its zero capture limit, it will respond with an E + back to the remote display. This is one of 4 possible responses for the ZERO command in SICS. The remote terminal will display , indicating that zero could not be accomplished because the weight was over the maximum limit.
Print Key Operation in Continuous and SICS Modes
The operation of the PRINT key depends upon programming in the remote display terminal:
• If the optional serial port is present, whichever port in the remote display is not programmed to operate with the Master terminal can be programmed for Print. In this case, a local print with the selected format will be created at the remote display. Nothing is sent to the Master terminal.
• If the remote display does not have a port assigned to Print, pressing PRINT sends an ASCII P
(continuous mode) or an ASCII PRN (SICS mode) command to the Master terminal. It is then up to the Master terminal to carry out the print function.
Limits of Operation
When the IND231/IND236 is in its remote display mode, it does not support discrete I/O or unit switching.
All the application modes of the standard IND231/IND236 are supported when operating in the remote display mode. Checkweighing, counting, animal weighing and totalization are fully functional based on the weight value received from the Master terminal. The additional features of target recall and time and date are also available for local use at the remote display.
The IND231/IND236 remote display mode supports g, kg, lb and oz weight units from the Master terminal. If weight units other than these are selected in the Master terminal, the weight value will be displayed, but the remote display will not show a weight unit.
2.6.
Applications
Each of the five applications can be accessed by pressing the configurable FUNCTION keys. When the key is pressed, the terminal will leave basic weighing mode and enter whichever application mode is configured in setup at Function Key > Assignment. Press the key to exit the application and return to basic weighing mode.
2.6.1.
Once an application has been started, only active function keys’ symbols are displayed.
Checkweighing
The Checkweighing application provides the ability to compare current weight on the scale to a stored target weight and indicate the comparison status on the display.
2-14 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.6.1.1.
The IND231/IND236 display indicates the Over/Under status graphically in the status line at the top of the screen. Three discrete outputs can be configured for control of external lights or a similar device, to indicate the current status of the weight comparison.
Configuration
Checkweighing only can be started from basic weighing or in the checkweighing application.
Whenever another function is active, such as counting, X10, or animal weighing, checkweighing
cannot be started and a warning message “—no—“ (Figure 2-11) will appear for a short time.
30094756 | 04 | 04/2021
Figure 2-11: “NO” Symbol
The
To configure checkweighing:
symbol in the top line indicates the checkweighing status.
1.
Set the checkweighing target (absolute value only).
Example:
Target weight = 5.000kg
Tolerance + = 1.000kg
Tolerance - = 1.000kg
2.
Press or (whichever is assigned to Checkweighing) to enter the target editing screen.
will appear at the top of the screen. If there is stable weight on the platform, it will be detected as an original target value. “- - - - - - “ indicates that the terminal is attempting to detect a stable weight. The target can be selected in one of two ways: a.
If a stable weight is detected before the approximately 0.5 second timeout elapses, “target” will appear for one second, then the detected stable weight will display. b.
Otherwise, the terminal will show “Target” for 1 second, and then display zero weight
(Figure 2-12). Select a digit to edit using
to move left and to move right. Once the required digit is selected and blinking, its value can be adjusted using the and
keys. In the example shown in Figure 2-12, the target has been set to 3 kg. Press
to accept this current value as the target and move to the next screen, “Tolerance -“.
3.
appears at upper left in the display. Set the value using the direction keys, as
indicated in step b, above. In this example (Figure 2-12, center), the low tolerance is set to 1
kg. Press to move to the next screen, “Tolerance +“.
4.
appears on the top of the screen. Set the value using the direction keys, as
indicated in step b, above. Here (Figure 2-12, right), the high tolerance is set to 1 kg.
METTLER TOLEDO IND231/IND236 User's Guide 2-15
2.6.1.2.
Figure 2-12: Target and Tolerance Displays: Target (left), Low (center) High (right)
Procedure
To perform checkweighing:
1.
Press or (whichever is assigned to Checkweighing) to enter the checkweighing screen. The weight status is indicated as follows (based on the example configuration from the previous section):
•
Under Below 4kg; (Figure 2-13, left )
• OK
Between 4kg and 6kg ;( Figure 2-13, center)
• Over
Above 6kg ;( Figure 2-13, right)
2.6.1.3.
2.6.1.4.
Figure 2-13: Over, OK and Under Displays
2.
Once checkweighing is completed, press to exit the application and return to basic weighing.
Viewing Configured Values
When in the checkweighing application, press and hold the Target function key to view the current active target, tol- and tol+ values. The information will display for one second, then the target value will appear together with . Press the “Target” key repeatedly to cycle through the three checkweighing values – OK, low tolerance and high tolerance. Press to exit.
Over/Under Printout
If Tare is zero, Tare and Net values are not printed.
The single and multi-line outputs appear as follows:
Multi-line output
Date
Time
Target
Tol +
YYY.MM.DD
HH:MM:SS
XXXX.XXX Unit
XX Unit
2-16 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.6.2.
2.6.2.1.
Single-line output
Tol -
Gross
Tare
Net
XX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Date_YYYY.MM.DD_ _ Time_HH :MM :SS_ _Tare
_XXXX.XXX_Unit_ _ Tol+_XX_%_ _ Tol-_XX_%_ _Gross
_XXXX.XXX_Unit_ _Tare _XXX.XXX_Unit_ _
Net_XXXX.XXX_Unit
Target Recall
Target recall is a function used to perform checkweighing using values from a database. Up to 10 over/under sets of data are available for easy recall on the terminal display. Each set includes data such as record number, target value, tolerance - value, tolerance + value and tare value.
The record parameters are set via menu setting F2.2. For more information, please refer to Chapter
3, Configuration.
Operation
When the RECALL function key is pressed, the message shown in Figure 2-14 will be displayed briefly. Then, if the database is not empty, the first record in it will appear (Figure 2-15). Otherwise,
“EMPTY” (Figure 2-16) will display, and then the weight display will reappear.
Figure 2-14: Target Database Recall Message
Figure 2-15: Record 01 Message
30094756 | 04 | 04/2021
Figure 2-16: Target Database Empty Message
Use the and keys to select the required record, then confirm the selection by pressing the
key. The checkweighing display will appear, with the selected values set, and the checkweighing procedure can begin.
METTLER TOLEDO IND231/IND236 User's Guide 2-17
2.6.3.
2.6.3.1.
2.6.3.1.1.
In checkweighing mode, pre-defined parameters can be viewed by pressing and holding the
RECALL key. Each subsequent press of the RECALL key displays in turn parameter values for target, tolerance - and tolerance +.
Counting
The IND231/IND236 counting application provides a simple counting sequence that guides an operator through a sampling process to determine a count value.
Operation
Counting only can be started from basic weighing or in the counting application.
Procedure
To count pieces:
1.
Press the counting function key . “SAMPLE” will be displayed, then “PCS 05”, as shown in
Figure 2-17.
Figure 2-17: Reference Number
2.
The reference number can be adjusted using (decrease) and (increase) keys.
Selections are 5, 10, 20, 50 and 100.
3.
Confirm the reference number by pressing
. “------“ will display (Figure 2-18) while the
scale detects stable weight. Make sure that the number of samples on the scale matches the
selected reference size. If stable weight is detected before the process times out, the counting application starts; otherwise the screen remains in the previous weighing state.
Figure 2-18: Detecting Stable Weight
4.
Load the parts to be counted. The count will be displayed (Figure 2-19).
Figure 2-19: Piece Count Display
5.
When counting is complete, press to exit the checkweighing application and return to the basic weighing interface.
2-18 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.6.3.1.2.
Reviewing Average Piece Weight
When the counting mode is active, press and hold the counting function key to display the current
APW (Average piece weight). The APW symbol will appear in the top row of the display, and the
APW value – in Figure 2-21, the value is 0.015 kg.
Figure 2-20: APW Review Screen
Press the counting function key again to display the reference number Figure 2-21 – here, 100 pieces is the current selection.
2.6.3.1.3.
2.6.3.2.
Figure 2-21: Reference Number Review Screen
Each time the “counting” key is pressed, the display switches between APW and reference number.
Press the key to exit the review screens.
Unit switch in counting mode
When the counting application is active, the unit switch has a special function unlike its standard operation. Press the “Unit Switch” key to cycle the display through PCS, primary unit and second unit in turn.
Counting Printout
If Tare is zero, Tare and Net values are not printed.
The single and multi-line outputs appear as follows:
Multi-line output
Single-line output
Date
Time
Pieces
APW
Gross
Tare
Net
YYY.MM.DD
HH:MM:SS
XXXXXXXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Date_YYYY.MM.DD_ _Time _HH :MM :SS_ _
Pieces_XXXXXXXX_PCS_ _APW _XXXX.XXX_Unit_ _
Gross_XXXX.XXX_Unit_ _Tare _XXX.XXX_Unit_ _Net
_XXXX.XXX_Unit
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 2-19
2.6.4.
2.6.4.1.
Animal Weighing
The animal weighing application provides the ability to calculate and display an average weight value determined over a user-defined sampling period. This is useful when the weight is constantly unstable, as is the case when weighing live animals.
The weighing mode can be configured in setup. If auto mode is selected, the animal weighing will start automatically when there is load on the scale. In manual mode, the animal weighing must be started by pressing the function key.
Operation
Animal weighing can only be started from basic weighing mode, Otherwise, “—no—“ (Figure
2-11) will be shown.
The animal weighing application has a net weight threshold of 9d. When the weight put on the platform exceeds 9 weight divisions (based on the division size configured in setup), the application will start. When either the net or gross weight is below 9d, the application will reset in preparation for the next weighment.
1.
Put the animal/s on the platform, then press the “animal weighing” function key . While the terminal collects weight values, “------“ is displayed (Figure 2-22). The process takes 2 to 3 seconds. In order to achieve better precision, the first 14 weight values are discarded, and the next 56 values collected for use in the calculation.
Figure 2-22: Display Indicating Weight Value Collection in Progress
2.
The average weight will be displayed, together with a star symbol, as shown in Figure 2-23.
Figure 2-23: Animal Weight Value Displayed
3.
To complete the current weighment, either remove the weight from the platform or press the clear key . Tare will also be cleared.
2-20 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2.6.4.2.
2.6.5.
2.6.5.1.
Animal printout
The single and multi-line outputs appear as follows:
Multi-line output
Single-line output
Date
Time
Average G
Tare
YYY.MM.DD
HH:MM:SS
XX.XXX Unit
XX.XXX Unit
Date_YYYY.MM.DD_ _ Time_HH :MM :SS_ _ Average
G _XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _
Average n_XXXX.XXX_Unit
Each number includes eight characters, including the decimal point.
Accumulation
For many weighing applications, knowing how many transactions were performed and how much material was processed during a particular period of time is useful information.
The IND231/IND236 terminal provides grand total (GT) registers and counters. Counters have a limit of 999 and registers will accumulate up to 7 digits of weight, including any digits to the right of the decimal point. For example, a scale programmed for 500 x 0.1 kg will accumulate weight values up to 999999.9 (7 total digits). If either of these limits is exceeded, an error message “— no—“ will display and the totals must be reset before additional weights or counts are added.
Operation
Accumulation can only be started when the terminal is in basic weighing mode. If one of the applications is active, , “—no—“ (Figure 2-11) will be shown.
30094756 | 04 | 04/2021
Figure 2-24: Collecting value symbol
The procedure for accumulation is as follows:
1.
Put the first weight on the platform and press the “accumulation” function key . If the stable
weight is detected within 2 seconds, the grand total value will be 1.5 kg and number will be 1
(Figure 2-24).
2.
Each time the new weight is put on the platform, press the accumulation key to add the new value to the total.
3.
Press and hold the function key to display the total value (Figure 2-25, left). Note that when the accumulation or count is displayed, the Accumulation icon will blink. Press the function key briefly, and the “Count”, or number of values appears (Figure 2-25, right). Each time the
“accumulation” key is pressed, the display will switch between count and sum.
METTLER TOLEDO IND231/IND236 User's Guide 2-21
2.6.5.2.
Figure 2-25: Collecting value symbol
4.
Press the
key to exit accumulation and return to basic weighing mode.
Accumulation Printout
When the terminal is in accumulation mode and the print key is pressed, the printed total accumulated weight will include the terminal’s currently displayed net value.
The single and multi-line outputs appear as follows:
Multi-line output
Single-line output
Date
Time
Total
Count
Gross
Tare
Net
YYY.MM.DD
HH:MM:SS
XX.XXX Unit
XXXX.XXX
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Date_YYYY.MM.DD_ _ Time_HH :MM :SS_ _
Total_XXXX.XXX_Unit_ _ Count_XXXX.XXX_ _
Gross_XXXX.XXX_Unit_ _ Tare_XXX.XXX_Unit_ _
Net_XXXX.XXX_Unit
2-22 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3 Configuration
This chapter provides information about how to configure the IND231/IND236 terminal. It describes access to the setup menu where functions can be enabled, disabled, or defined by entering parameter values.
The menu consists of the following blocks:
F1
F2
F3
Scale
Application
Terminal
F4
F5
Communication
Maintenance
3.1.
Accessing Setup Mode
3.1.1.
Entering Setup Mode
The configuration of the IND231/IND236 terminal is accessed by pressing and holding the MENU key .
A login screen (Figure 3-1) displays and the user must enter the correct password in order to
advance into setup.
Figure 3-1: Login Screen
Access to items in setup varies depending on the type of user, and on whether the terminal is
approved as legal for trade. Table 3-1 shows which parts of the setup tree are accessible to the two
types of log-in.
Supervisor
Technician
Password
Table 3-1: Passwords and Access Levels
F1.1.2 = Approved F1.1.2 = None
F1.6 (second unit), F2, F3, F4; F5.2 (statistics); F5.8 (print configuration)
F1.6 (second unit), F2, F3, F4,
F5
All function blocks
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-1
To enter a password:
1.
When the log-in screen displays, use the scale function keys to enter the password. Press the
key. If the password is correct, the terminal will enter setup mode, and the setup menu will be displayed. If the password is not valid, the display returns to the weighing mode.
If the terminal is configured as legal for trade, the user can only access setup by pressing
the setup-switch (a hardware switch on main-board, indicated in Figure 3-2). The terminal
will then display the F1 (Scale) menu label and scale parameters can be modified.
3.1.2.
Figure 3-2: Setup Switch Location on Terminal Main Board
Exiting Setup Mode
To exit the setup mode and return to the weighing mode:
1.
Press the key, or left when in the root of the menu tree (F1, F2, etc. displayed). A
message will appear asking whether to Abort or Save changes Figure 3-3.
Figure 3-3: Prompts When Exiting Setup
2.
Use (down) and (up) to switch between Save and Abort.
• Save: Save the setup parameters.
• Abort: Do not save the setup parameters.
2.
Once the selection has been made, there are two options:
or Confirm choice of Save or Abort, and exit to main screen.
Remain in the setup menu, at F1.
3-2 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.1.3.
3.1.3.1.
3.1.3.2.
Setup Menu Navigation
Key Functions
After entering setup, the F1, F2, Zero and Tare keys become navigation keys, and are used to move around in the menu structure and to adjust settings once a parameter settings screen has been reached.
Left
Right
Menu Structure
Back to next higher menu screen
Access next lower submenu screen
Settings Screen
Return to menu structure or , if a numerical entry screen,
Move focus left
Accept current value and return to menu structure or , if a numerical entry screen,
Move focus right
Up
Down
Move to next higher screen at current level
Move to next lower screen at current level
Next higher value for item in focus*
Next lower value for item in focus*
* Parameters with discrete settings (such as On/Off or 0.5, 1, 10) cycle through settings in either
Up or Down direction.
Example of Navigation
Each screen of the setup menu tree can be reached using the keys above. The example shown in
Table 3-2 shows how to access the Scale Approval screen (F1.1.2) and set an approval, then
active the Tare Interlock feature (F1.5.2), and finally exit setup, either saving or discarding these changes.
Key n/a
Display
F1
Table 3-2: Accessing and Setting Parameters
Action
Setup has been accessed
F1.1 Move one level lower in menu structure
F1.1.1
F1.1.2
Move one level lower in menu structure
Move to next branch at this level noNE oiML
Display Approval setting screen
Select OIML approval
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-3
3.1.3.3.
Key Display
F1.2
Action
OIML accepted, focus moved to F1.2 submenu
Note: Either Right (F2) or Enter key accepts the selection
F1.3
F1.4
F1.5
F1.5.1
Move to next main sub-branch
Move down to next lower level
F1.5.2 oFF
Move to next branch at current level
Current Tare Interlock setting displayed – OFF on
Tare Interlock set to ON
Note: Either Up or Down key will cycle through ON/OFF values
F1.5.3
ON accepted, focus moved to next branch at this level
Note: Either Right (F2) or Enter key accepts the selection
F1.5
Return to top level of menu
F1
SAVE Exit from setup – SAVE prompt displays
Returns to weighing display
Abort Discards changes, returns to weighing display
Setting Numbers
When setting a numerical parameter such as Capacity (F1.2.3), the Left and Right navigation keys function to move focus, and the Up and Down keys adjust the value. The number in focus blinks:
Figure 3-4: Changing Numerical Values, Digit in Focus Blinking
3-4 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
To increase the value, press ; the decrease it, press . To select the units column, press
; to select the hundreds column, press . To accept the displayed value and exit the screen, press .
3.2.
Setup Menu Structure
3.2.1.
Main setup menu
When setup is entered, Block F1 is displayed as shown in Figure 3-5. Direction keys can be used
to select different menu blocks (From F1 to F5).
3.2.2.
Figure 3-5: F1 Menu Screen
Configurable function keys
The function keys can be configured without entering setup. From the main screen, press and hold
either function key F1 or F2 to enter the configuration display shown in Figure 3-6. Use the Down
and Up keys to select the function for that key. In the example shown, F1 is set to x10 and F2 is set to Accumulation ( ∑ ).
Figure 3-6: Setting Assignments for Function Keys F1 and F2
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-5
3.3.
Configuration Overview
Figure 3-7 shows all the main branches of the terminal’s configuration menu, except branches that
are application-specific.
Details for each of the five major setup branches are provided in the sections following the overview. Use this information to program the basic configuration of the IND231/IND236 terminal.
Figure 3-7: IND231/IND236 Setup Menu Structure
An invalid entry during calibration will cause a message to appear. The message will disappear after about three seconds, and the display will return to its previous state.
3.4.
F1 Scale
Figure 3-8 shows a detailed view of the parameters available in the Scale branch of the setup
menu. Each of these parameters is described in this section.
If the terminal is approved (at F1.1.2), only F1.6 will be available for editing. To access other
parameters in the Scale branch, the metrology switch (refer to page 3-2) must be pressed.
3-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.4.1.
3.4.1.1.
Figure 3-8: Scale Menu Structure
F1.1 Scale type
The Scale type setup screen is used to configure the terminal as either a weighing terminal or as a second display, and to configure the Weights and Measures setting.
F1.1.1 Type
Selections for Scale Type are:
SCL
(Factory setting)
Com 1
Com 2
(If available)
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-7
3.4.1.2.
If scale type is SCL, the terminal functions as a scale terminal. If scale type is rCom1 or rCom2, the terminal is used as a second (remote) display. In this case, use the Com 1 or Com 2 port to connect another scale terminal.
When either rCom 1 or rCOM 2 is selected, the port mode and print setting for that port (refer to the
section beginning on page 3-33) will be hidden, since the port cannot be assigned to any function
other than the terminal’s connection as a remote display.
F1.1.2 Approval
F1.1.2 appears if Type is set to SCL. Approval refers to the configuration of metrological (weights and measures) approvals for the specific scale. Options are:
None
No approval required
OIML
NTEP
AR
SRI
THAI
If an approval for NTEP, OIML or Sri Lanka is configured, access to metrological parameters in the
Scale setup branch will be limited to F1.6 (units) and F1.10 (Reset scale settings) As Figure 3-9
shows, pressing and holding the key will first display the type of approval, then prompt for a password, and then display the F1 label.
Figure 3-9: Menu Limited to View Only
If the approval is configured for Argentina, access to the entire setup menu is prohibited. If an attempt is made to access the setup menu, an error message will display: “Error 11”.
3-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.4.1.3.
3.4.2.
3.4.2.1.
3.4.2.2.
Access to Setup in an Approved Terminal
To regain access to the F1 setup menu, it is necessary to open the terminal’s enclosure and press the metrology switch (refer to Main Board Wiring Connections in Appendix A, Installation). The terminal will display the F1 Scale settings menu, and scale parameters can be configured.
F1.1.3 Remote Protocol
Figure 3-10: Menu Access Prohibited
F.1.1.3 appears if Type is set to rCOM1 or rCOM2. Selections are:
Toledo Continuous, SICS
F1.2 Capacity and Increment
Use the Capacity and Increment setup screen to select primary units, set the number of weighing ranges, and the scale capacity and increment size.
F1.2.1 Primary Units
Selections for Primary Units are:
Kilograms (kg) [default], Grams (g), Pounds (lb), Ounces (oz)
F1.2.2 Ranges
The terminal can be configured with one or two ranges. If two is selected, the increment size will change when the weight reaches the second range.
Figure 3-11: Range Menu Screen, and Settings
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-9
3.4.2.3.
F1.2.3 Range 1 capacity
The weight capacity for range 1 can be set to any value from 1 to 500,000. If only one range is enabled, this value will be the scale capacity and the display will show an overcapacity graphic if the weight exceeds the value by more than five increments.
3.4.2.4.
Figure 3-12: Range 1 Capacity
If two ranges are enabled, this will be the weight at which the increment size changes from the range 1 increment size to the range two increment size.
F1.2.4 Increment Size 1
F1.2.4 specifies the increment size for range 1. The terminal will automatically generate valid increment size selections, depending on the resolution setting (1000~30,000). Select the appropriate value. If only one range is enabled, this will be the increment size for the entire weighing range of the scale. If two ranges are enabled, this will be the increment used in the lower range.
3.4.2.5.
Figure 3-13: Range 1 Increment
F1.2.5 Range 2 capacity
If only one range is enabled, this parameter will not be shown. If two ranges are enabled, the weight capacity for range 2 can be set to any value from 1 to 500,000. This value will be the scale capacity, and the display will show an overcapacity graphic if the weight exceeds the value by more than five increments. Range 2 capacity should be larger than range 1.
3.4.2.6.
Figure 3-14: Range 2 Capacity
F1.2.6 Increment Size 2
F1.2.6 specifies the increment size for weighing range 2. If only one range is enabled, this parameter will not be shown. The terminal will automatically generate valid increment size selections depending on the resolution setting (1000~30,000). Select the appropriate value. This will be the increment used in the upper range.
3-10 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.4.3.
3.4.3.1.
Figure 3-15: Range 2 Increment
When multiple ranges are configured, the range 2 increment must be larger than the increment of range 1.
F1.3 Calibration
The Calibration menus allow the entry of GEO code and linearity adjustments and the calibration of zero and span settings.
F1.3.1 GEO Code
Use the down or up keys to enter the GEO code (0 – 31; refer to Appendix D, GEO Codes) for a list of geographical locations and the corresponding codes.
3.4.3.2.
Figure 3-16: GEO Code Value Selection Screen
F1.3.2 Calibration
Select F1.3.2 to set zero and full capacity calibration values.
1.
Access the Calibration menu.
2.
Empty the scale and press to start calibration. “E SCL” (empty scale) appears on the display. Press again. The display will count down 10…0 as the terminal captures zero.
3.
”Full Ld” appears on the display.
4.
Press . The full capacity value appears, blinking. Use the down
or up keys to change the calibration weight value.
5.
When the value has been set, put the corresponding full capacity weight on the scale and press . The display will count down
10…0. If the calibration was successful, “done” will display for 2 seconds.
Best accuracy will be achieved by using the maximum rated capacity of the scale when calibrating full capacity.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-11
3.4.3.3.
F1.3.3 Linearity
Select F1.3.3 to include a linearity adjustment in the calibration. In this case, zero, half- and fullcapacity are calibrated.
1.
Access the Linearity Adjustment menu.
2.
Empty the scale. Press to start calibration. “E SCL” (empty scale) appears on the display. Press again. The display will count down 10…0 to capture zero.
3.
”ADD LD” appears on the display.
4.
Press . The half capacity value appears, blinking. Use the down or up keys to change the calibration value.
5.
Put the corresponding half capacity weight on the scale and press
. The display will count down 10…0 as the calibration is performed.
6.
”Full Ld” appears on the display.
7.
Press . The full capacity number appears, blinking. Use the down or up keys to change the calibration value.
8.
Put the corresponding full capacity weight on the scale and press
. The display will count down 10…0. If the calibration was successful, “done” will display for 2 seconds.
Best accuracy will be achieved by using the maximum rated capacity of the scale when calibrating full capacity.
If the scale is not stable after 30 seconds, it will time out and display an error code ”Error 35”
last operation. Either reload the weight, or press to cancel the process. If calibration is
cancelled, “Abort” (Figure 3-17, right) will display briefly, calibration will end, and the F1.3.3
screen will reappear.
Figure 3-17: Linearity Calibration Error and Abort Messages
3-12 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
To interrupt the process and save the previous parameters, press during the calibration.
”donE” (Figure 3-18) will appear briefly. The former parameters are saved and the display returns
to the F1.3.3 screen.
3.4.4.
3.4.4.1.
Figure 3-18: Linearity Calibration Interrupted
To abort the process without saving the previous parameters, press the power button during the
calibration. ”Abort” (Figure 3-17, right) will appear briefly. Calibration is suspended, the former
parameters aren’t saved, and the display returns to the F1.3.3 screen.
F1.4 Zero
Zeroing corrects the influence of the accumulation of a small amount of debris on the load plate.
This section provides access to Auto Zero Maintenance (AZM) settings, under zero blanking, powerup zero, and pushbutton zero parameters.
F1.4.1 Automatic Zero Maintenance (AZM)
Automatic Zero Maintenance (AZM) is a means of tracking zero when the scale is empty. It enables
IND231/IND236 to compensate for conditions such as terminal or load cell drift or debris on a scale platform by setting itself back to the center of zero.
Available selections are based on resolution. In scales not approved for trade, selections are: oFF, 0.5d [default value], 1d, 3d, 10d
In approved scales, the only setting is 0.5d.
3.4.4.2.
Figure 3-19: AZM Value Selection Screen
F1.4.2 Zero Blanking
In certain countries, such as Thailand, the approval authorities require the terminal to behave in a particular way when an underload (weight value below zero) is detected, and when a valid weight is detected after an underload. Selections for underload behavior are as follows: oFF, 20d [default], 20dZero
These settings function as follows:
Off The display will continue to display negative gross weights with no limit.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-13
3.4.4.3.
20d The underload symbol is shown for values below -20d. As soon as the weight value exceeds -20d, the weight value is displayed again.
20dZero The warning “Err 10” will be shown as soon as the weight decreases below -20d. This warning message stays on the screen until the scale is switched off or a zero is performed. Only if this zeroing is successful (within the valid zero setting limits when pressing zero key or within the valid power up zero limits when switching the device off/on) the weight value is shown again. If not, the warning appears again.
F1.4.3 Power Up Zero
When Power Up is set to Off, the terminal will save and reuse the last zero reference weight after a power cycle, so that it returns to the same gross or net weight value.
Figure 3-20: Power Up Zero Value Selection Screen
When a power up range of +/-2%, +/-10%, -2% ~ +18% or +/-20% is selected, the terminal tries to capture zero upon power up. If there is motion during a power-up zero capture function, the terminal will continue to check for a no-motion condition until zero is captured. Once a no-motion condition is reached, it will automatically capture zero.
For example, if the setting for Power Up Zero is set at 2%, Power Up Zero will only occur when the weight reading on the scale is within +/- 2% of the original calibrated zero condition.
Selections for unapproved scale selections are: oFF, 2, 10 [default], 18, 20
Selections for approved scales are: oFF, 2, 10 [default], 18
When the scale’s approval is set to OIML, if Power Up Zero capture is enabled and the weight
on the scale is outside of the zero capture range, a warning message will display (Figure
3-21) until the weight is removed and zero is captured.
3.4.4.4.
Figure 3-21: OIML Zero Capture Warning
F1.4.4 Pushbutton Zero
When Pushbutton Zero is set to Off, the front panel ZERO pushbutton will not operate to capture a new zero reference point. If a range of +/-2% , +/- 10% or +/- 20% is selected, pushbutton zero will operate within the selected range to rezero the scale.
3-14 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
For example, if Pushbutton Zero is set to +/-2%, the Pushbutton Zero can only be used when the weight reading on the scale is within +/- 2% of the original calibrated zero condition.
3.4.5.
3.4.5.1.
Figure 3-22: Pushbutton Zero Value Selection Screen
If Pushbutton Zero is Off, it is still possible to execute a remote Zero via SICS or CPTZ commands from a PC using a discrete input command.
Selections for scales not legal for trade are: oFF, 2 [default], 10, 20;
Selections for approved scales are: oFF, 2 [default]
The zeroing function does not affect the overall scale weighing range, and the scale capacity will not be affected.
If Auto Clear Tare is enabled (refer to page 3-16), a successful Pushbutton Zero operation will also
clear the Tare value automatically.
F1.5 Tare
Tare is used to subtract the weight of an empty container from the gross weight on the scale to determine the net weight of the contents. Tare is inhibited if the scale is in motion.
F1.5.1 Pushbutton Tare
Selections for Pushbutton Tare are: on [default], oFF
If the Tare key is pressed when keyboard tare is disabled, a warning message --n o-- will appear.
3.4.5.2.
Figure 3-23: Warning Message – Pushbutton Tare Disabled
F1.5.2 Tare Interlock
If Tare Interlock is activated (F1.5.2 = on), the tare weight can only be cleared when weighing platform is unloaded to zero, or warning message” –no--“ will appear after is pressed.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-15
3.4.5.3.
on, oFF [default]
F1.5.3 Auto Tare
Selections for Auto Tare are: on, oFF [default]
Figure 3-24: Auto Tare Setting Screen
When Auto Tare is active, when an empty container is placed on the platform the measured weight is automatically saved as the tare weight. “Zero” displays and the NET indicator appears.
3.4.5.4.
3.4.5.5.
3.4.5.6.
Figure 3-25: Display After Auto Tare
F1.5.4 Auto Tare Threshold
When Auto Tare is active, the F1.5.4 menu permits adjustment of the auto tare threshold. When weight on the scale platform exceeds the tare threshold value and settles to no-motion, the terminal automatically tares.
The factory setting is 10, so that when tare weight is above 10 divisions, an auto tare operation can be executed.
F1.5.5 Auto Tare Reset Threshold
When Auto Tare is active, F1.5.5 permits adjustment of the auto tare reset threshold. When the weight on the scale platform falls below the reset threshold value, such as when a load has been removed, the terminal automatically resets the auto tare trigger, depending on how motion checking
(refer to page 3-18) is programmed. The reset threshold weight must be less than the tare
threshold weight.
The default value for the Auto Tare Reset Threshold is 10. So, for example, after the initial auto tare operation is complete and a tare is set, a load of less than 10 divisions can be added without changing the tare value. When the first load is removed and a second load of 10 divisions or greater is added, the terminal will capture a new tare weight automatically.
F1.5.6 Auto Clear Tare
Selections for the Auto Clear Tare function are: on, oFF [default]
3-16 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Figure 3-26: Auto Clear Tare Setting Screen
After the scale is unloaded, the terminal will automatically clear tare.
3.4.6.
3.4.6.1.
Figure 3-27: Display After Auto Clear Tare
F1.6 Second Units
This setup screen enables the selection of a second unit.
Second Unit
In order to support locations and applications that use multiple measurement units, the
IND231/IND236 supports unit switching. Use the Second Unit selection menu to select a second weighing unit or to select none. The weight will be converted from primary to the selected secondary unit when the Unit Switch key is pressed.
Available selections are:
Kilograms (kg) [default], Grams (g), Pounds (lb), Ounces (oz)
default, F2 – Figure 3-31) to switch between 1.500 kg (Figure 3-30) and 1500 g (Figure 3-32).
Figure 3-28: Primary Units Setting Screen
Figure 3-29: Secondary Units Setting Screen
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-17
Figure 3-30: Display Showing Primary Units
Figure 3-31: Unit Switching Via F2 Key
3.4.7.
3.4.7.1.
Figure 3-32: Display Showing Secondary Units
F1.7 Filter
F1.7.1 Filtering
The IND231/IND236 terminal has a vibration filter that can be set for several conditions. The heavier the filtering, the slower the display settling time will be.
Filter selections are:
Low, Mid [default], High
3.4.7.2.
Figure 3-33: Filtering Menu Screen and Setting Screens
F1.7.2 Motion Range
The Motion Range value is defined based on division size. When the weight range is within the configured motion range, the terminal treats the scale as stable.
The Motion Range selections are:
0.5d [default], 1d, 3d.
3-18 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.4.8.
Figure 3-34: Motion Range Setting Screen
F1.10 Scale Block Reset
To initiate a reset of the Scale branch, access the F1.10 menu.
Figure 3-35: Scale Block Reset Menu Screen
A message will prompt “Sure?”
Figure 3-36: Scale Block Reset Confirmation Prompt
Press either or to reset the scale block. Press or to abort the resetting process.
Scale reset does NOT include the reset of metrological significant parameters – scale type, approval, weight units, capacity, increment, or calibration data. This data is reset only by
performing a Master Reset (refer to page 3-42).
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-19
3.5.
F2 Application
Figure 3-37 shows a detailed view of the parameters available in the Application branch of the
setup menu. Each of these setup parameters is described in this section.
3.5.1.
3.5.1.1.
F2.1 Operation
F2.1.1, F2.1.2 Function Key Setup
Figure 3-37: Application Menu Structure
Each of the two function keys on the terminal’s front panel, and , can be assigned to perform a specific operation such as switching units or activating one of the applications. Available
assignments are shown in Figure 3-38 and Table 3-3.
Figure 3-38: Function Key Assignments
Only one application can be selected for activation by a function key. If an application is assigned to each key (for example, F1 = Counting, F2 = Animal Weighing), only the first application activated will work. If F1 is pressed, the terminal will enter Counting mode; pressing
F2 will not activate Animal Weighing. Instead, a warning prompt will display (Figure 3-39).
Exit the active application before pressing F2 to activate another application.
3-20 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.5.1.2.
3.5.1.3.
Figure 3-39: Second Application Warning Message
Function keys can also be configured from the home screen. Press and hold the function key to
be configured to access the configuration screen (Figure 3-40, Figure 3-41).
Assignment
Options for the assignment of the function keys are listed in Table 3-3.
Table 3-3: Functions
Symbol Function
Expand weight display resolution by one digit
Unit switching
Check weighing
Recall stored target
Counting
Animal weighing
Symbol Function
Accumulation
Date & Time
Backlight adjust
Zero
Tare
For details on the setup, configuration and operation of each of the applications, please refer to
Chapter 2, Operation.
F2.1.1 Function Key 1
Access the F2.1.1 menu. Use the up and down keys to select the required function from
the options shown in Table 3-3. Focus will change to indicate the selected function, as shown in
Figure 3-40, where Expand x10 is selected. Press
to confirm the selection.
3.5.1.4.
Figure 3-40: Configuring Function Key 1
F2.1.2 Function key 2
Access the F2.1.1 menu and use the up and down keys to select the required function.
Focus will change to indicate the selected function, as shown in Figure 3-41, where Animal
Weighing is selected. Press to confirm the selection.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-21
3.5.2.
Figure 3-41: Configuring Function Key 2
F2.2 Over/Under Database
Up to ten targets and tolerance values can be stored for Over/Under checkweighing. The menu
Figure 3-42 provides a graphical summary of record management procedures in the
IND231/IND236 terminal.
3-22 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.5.2.1.
Figure 3-42: Selecting, Viewing and Creating Records
Viewing Records
When F2.2 is accessed, if a record exists its label is displayed. If no records exist, the Create screen displays. The sequence of display is as follows:
Record 1 (If already defined)
Record 2 (If already defined)
……
Record n (Last existing record)
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-23
Create record
Index: record number
Target
Tol-
Tol+
Tare
Figure 3-43: Screen Displaying Record Index Number
Once all existing records have been displayed, the Create screen (Figure 3-44) appears, and after it
a series of parameter setting screens.
3.5.2.2.
Figure 3-44: Create Record Screen
Creating a Record
To create a new record number x (where x is a value from 1 to 10):
1.
With the Create screen (Figure 3-44) showing, press
.
2.
rEC no. appears briefly, and then the record number editing screen (Figure 3-45). The terminal
can save up to 10 records. By default, the system offers the next available number.
Figure 3-45: Record Number Label and Record Number Setting Screens
3.
Either accept the proposed number by pressing , or use the direction keys to adjust the
number. If an existing number is selected, an Overwrite confirmation message (Figure 3-46)
displays.
Figure 3-46: Overwrite Record Confirmation Screen
3-24 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Press to confirm the action, or to return to the Create screen.
4.
Once the record number is confirmed by pressing , the target screen will appear briefly,
and then the target editing screen (Figure 3-47).
Figure 3-47: Target Label and Setting Screens
Use the direction keys to select and adjust the numbers (refer to Setting Numbers on page 3-4).
Once the correct value is displayed, press to move to the next step.
3.
Next, the negative tolerance (Tol -) screen appears, followed by the negative tolerance editing
Figure 3-48: Negative Tolerance Label and Setting Screens
Use the direction keys to adjust the number. Once the correct value is displayed, press to move to the next step.
The target value can also be set by adding weight to the scale.
4.
The positive tolerance (Tol +) is edited in the next screen (Figure 3-49). Once the value is set,
press to move to the next step.
Figure 3-49: Positive Tolerance Label and Setting Screens
5.
Finally, a tare value can be set for the record (Figure 3-50). Press
to confirm the record and exit the record editing function.
30094756 | 04 | 04/2021
Figure 3-50: Positive Tolerance Label and Setting Screens
Once the record is saved in the database, “Record x”(the newly created record) will appear as the active item in the F 2.2 submenu. Press to view its configuration details.
METTLER TOLEDO IND231/IND236 User's Guide 3-25
3.5.2.3.
3.5.2.4.
Replacing Existing Records
To overwrite a record, access the “Create” screen and edit the Record Number screen to show the number of the existing record to be overwritten. Press . An overwrite confirmation message
(Figure 3-46) will appear. Press
to confirm the action, or to return to the Create screen.
Deleting Existing Records
To delete a record, access F2.2 and use UP or DOWN keys to display the Record
Number of the record to be deleted. With the record’s label displayed, press . A confirmation
message (Figure 3-51) will appear.
3.5.3.
3.5.3.1.
3.5.4.
3.5.4.1.
Figure 3-51: Delete Record Confirmation Screen
Press either to preserve the record and return to the Record Number display screen, or to confirm its deletion. If a higher numbered record exists, the next record number is displayed. If the deleted record was the highest numbered record, or no other records exist, the Create screen is displayed.
F2.3 Counting
F2.3.1 APW Enhancement
The IND231/IND236 terminal can be configured to enhance the Average Piece Weight value continuously during operation. Settings are: on, oFF [default]
When APW enhancement is set to On, the weighing terminal automatically determines the piece weight again when the number of reference parts is increased.
F2.4 Animal Weighing
F2.4.1 Auto Start
The Animal Weighing application can be configured to start automatically when weight is placed on the platform. Settings are: on [default], oFF
When Auto Start is active, if the weight on the platform is larger than 9d the animal weighing function will run automatically. The final average weight will be shown on the display with star “*” symbol.
3-26 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Figure 3-52: Animal Weighing Display, Final Average Weight
When the weight is removed from the platform, the weight will return to zero and the star “*” symbol will disappear.
3.5.5.
3.5.5.1.
3.5.5.2.
30094756 | 04 | 04/2021
Figure 3-53: Animal Weighing, Scale Weight Returned to Zero
F2.5 Discrete I/O
Functions can be assigned to the discrete inputs, and statuses to the discrete outputs. When one of statuses is on (true), the output is set to active.
F2.5.1 Input
The discrete input can be used to monitor the function key actions, as indicated in Table 3-4.
Default settings are shown in bold.
Menu address
F2.5.1.1
F2.5.1.2
F2.5.2 Output
Table 3-4: Discrete I/O Input Assignments
Function
Input 1
Input 2
Assignments
Off, Zero, Tare, Print, Unit switch, Clear, Blank
Off, Zero, Tare, Print, Unit switch, Clear, Blank
Table 3-5 lists discrete output assignments. The statuses of each assignment are shown in Table
3-6. Default values are shown in bold.
Menu address
F2.5.2.x
[x = 1 - 4]
Table 3-5: Discrete I/O Output Assignments
Function
Output x
[x = 1 - 4]
Assignments
Off, Over Tolerance, Under Tolerance, T OK,
Net, Motion, Over Load, Under Load, Center of Zero
Table 3-6: Discrete I/O Output Statuses
Status
Off
Explanation
No status assigned, always inactive
Over Tolerance Active when weight is above TOL+
Under Tolerance Active when weight is less than TOL- and larger than 5% of
METTLER TOLEDO IND231/IND236 User's Guide 3-27
3.5.6.
Status Explanation
TOL-
Tolerance OK Active when weight is within tolerance
Net Active when in net weighing mode
Motion Active when motion symbol appears
Over Load Active when weight is overload
Under Load Active when weight is underload
Center of Zero Active when is shown
F2.10 Application Block Reset
To initiate a reset of the Application branch, access the F2.10 menu.
Figure 3-54: Application Block Reset Menu Screen
A message will prompt “Sure?”
Figure 3-55: Application Block Reset Confirmation Prompt
Press either or to reset the scale block. Press or to abort the resetting process.
3-28 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.6.
F3 Terminal
Figure 3-56 shows a detailed view of the parameters available in the Terminal branch of the setup
menu. Each of these setup parameters is described in this section.
3.6.1.
3.6.1.1.
3.6.1.2.
3.6.2.
3.6.2.1.
30094756 | 04 | 04/2021
Figure 3-56: Terminal Menu Structure
F3.1 Serial Number
F3.1.1 Serial Number
This screen displays the terminal’s ten-digit serial number. This number can be viewed, but not edited.
F3.1.2 Firmware Version
This screen displays the currently installed version number of the currently installed firmware.
F3.2 Display
Use the Display setup screen to set the screen saver and backlight timeouts and the brightness of the display.
F3.2.1 Screensaver and Backlight Timeout
This parameter selects the length of time that the backlight will stay on when the scale is at center of zero, no scale motion is detected, and there is no keypad activity. After the set amount of time elapses, the screensaver replaces the view on the display screen and the backlight is turned off. If motion is detected or any key is pressed, the screen saver closes and the backlight turns on.
A keystroke used to exit the screen saver mode is ignored for all other purposes.
METTLER TOLEDO IND231/IND236 User's Guide 3-29
3.6.2.2.
Timeout selections are:
0, 5, 10, 15, 30, 60 [default], 120, 300, 600 seconds
When 0 is select, the display will not time out and the backlight will stay On until power is removed from the terminal. When any other value is selected, the backlight will go off after the specified interval, and the time of day will be displayed.
F3.2.2 Brightness
This parameter accesses a screen where the display brightness can be adjusted. Use up or
down keys to select the appropriate setting for the terminal’s environment.
Possible selections are:
Off
Low:
Turn off backlight
Low brightness (battery power saving mode)
High: High brightness
3.6.2.3.
3.6.2.4.
Figure 3-57: Backlight Settings
Press to exit the adjustment screen and return to the F3.2.2 menu. Press or to confirm the selected brightness setting and return to the F3.2.3 menu.
F3.2.3 Auto Power Off
This parameter helps conserve battery power by selecting the length of time that the terminal will stay on when the scale is at center of zero, no scale motion is detected, and there is no keypad activity.
Selections are:
0 [default, AC version], 1, 5 [default, battery version], 15, 30, 60 minutes
If 0 minutes is selected, the terminal will never turn itself off.
F3.2.4 – Weight Hold
This parameter determines the length of time that the weight is displayed on screen with a star symbol after a print operation.
Selections are:
3-30 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.6.3.
3.6.3.1.
3.6.3.2.
3.6.3.2.1.
0 [default], 1, 2…9 seconds
F3.3 Date & Time
Select the hour, minutes, day, month, and year on this setup screen. The terminal automatically adjusts the date for a leap year, and a battery backup maintains the time and date settings in the event of a power outage.
For daylight savings time adjustments, the time must be changed manually.
F3.3.1 Date format
Selections for the date format are:
MM/DD/YYYY, DD/MM/YYYY, YYYY/MM/DD [default]
F3.3.2 Set Date
Day, Month and Year
When F3.3.2 is first accessed, the screen shown in Figure 3-58 is displayed briefly, then the date
currently set. Use the direction keys to adjust Day, Month and Year. In the example shown, the date format is YY-MM-DD.
3.6.3.3.
3.6.3.4.
3.6.3.4.1.
Figure 3-58: Date Adjustment Screen
Press to save changes and leave the adjustment screen, or to leave the screen without saving any changes.
F3.3.3 Time format
Selections for the time format are:
12, 24 [default]
F3.3.4 Set Time
Hour and Minutes
Use direction keys to adjust Hour and minutes in Figure 3-51. The time format is : 24:MM:SS
When the screen shown in Figure 3-59 is displayed, use the direction keys to adjust Hours and
Minutes. In the example shown, the time format is HH:MM:SS.
30094756 | 04 | 04/2021
Figure 3-59: Time Adjustment Screen
METTLER TOLEDO IND231/IND236 User's Guide 3-31
3.6.4.
Press to save changes and leave the adjustment screen, or to leave the screen without saving any changes.
F3.10 Terminal Block Reset
To initiate a reset of the Terminal branch, access the F3.10 menu.
Figure 3-60: Terminal Block Reset Menu Screen
A message will prompt “Sure?”
Figure 3-61: Terminal Block Reset Confirmation Prompt
Press either or to reset the scale block. Press or to abort the resetting process.
3-32 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.7.
F4 Communication
Figure 3-62 shows a detailed view of the parameters available in the Communication branch of the
setup menu. Each of these setup parameters is described in this section.
3.7.1.
3.7.1.1.
Figure 3-62: Communication Menu Structure
F4.1 COM1
F4.1.1 Connection Mode
The connection mode setup screens are used to assign a function to a physical port in the terminal.
The optional ports are only available if the appropriate option board has been installed.
Setup pages are available for the IND231/IND236 terminal’s COM1, COM2, USB and Discrete I/O
ports. Table 3-7 shows the available connection modes, which determine the type of
communication that will occur over the port. If no mode is assigned to a port, no data communication will occur using that port. Specific details of these modes can be found in Appendix
C, Communications.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-33
3.7.1.2.
Mode
Table 3-7: Connection Modes
Description
Print [default]
Auto Print
SICS
Stable weight values are printed automatically
Communication via the METTLER
TOLEDO Standard Interface
Command Set
Toledo Continuous
F4.1.2 Format
Format selections for the COM2 connection are detailed in Table 3-8.
Table 3-8: COM1 Format Assignments
Port Assignment if connection mode is Print or Auto print
F4.1.2.1 Line Format Multi-line, Single-line
F4.1.2.2 Print Language English, Chinese
COM1= RS-
232
F4.1.2.3 Add Line Feed 0,1,2,3,4,5,6,7,8,9
Assignment if Connection mode is Auto Print
F4.1.2.4 Auto Print Threshold
F4.1.2.5 Auto Print reset threshold
0 - full capacity
0 - full capacity
Figure 3-63: Line Format Options
3.7.1.3.
3.7.1.3.1.
Figure 3-64: Language Options
F4.1.3 Port Parameters
F4.1.3.1 Baud Rate
Baud rate selections for the serial port are:
3-34 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
1200, 2400, 4800, 9600 [default], 19200, 38400, 57600, 115200
3.7.1.3.2.
Figure 3-65: Baud Rate Screen
F4.1.3.2 Data Bits / Parity
Data bits and Parity selections are:
7 – odd 7 bits, parity odd, 7 – even 7 bits, parity even, 8 – none 8 bits, no parity
[default]
3.7.1.3.3.
3.7.1.4.
3.7.1.5.
Figure 3-66: Data Bits/Parity Screens
F4.1.3.3 Flow Control
Flow control selections are:
On – Flow control XON-XOFF enabled (software handshaking), Off – Flow control disabled
[default]
F4.1.4 Checksum
The Checksum parameter is only available when F4.1.1 is set to Toledo Continuous. Checksum selections are:
On, Off [default]
F4.1.10 Communication Reset
To initiate a reset of the Communication branch, access the F3.10 menu.
Figure 3-67: COM1 Communication Reset Menu Screen
A message will prompt “Sure?”
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-35
3.7.2.
3.7.2.1.
3.7.2.2.
Figure 3-68: COM1 Communication Reset Confirmation Prompt
Press either or to reset the scale block. Press or to abort the resetting process.
F4.2 – COM 2
The COM2 ports are only shown if the corresponding option board has been installed.
If COM2 is either the RS232 or USB option, the communication menu is the same as for COM1,
described in section F4.1 COM1.
If COM2 is the RS422/485 option, F4.2.3.5 is used to set the net address.
F4.2.1 Connection Mode
Selections for the COM2 connection mode are the same as those shown in Table 3-7:
Print [default], Auto Print, SICS, Toledo Continuous
For information about these settings, please refer F4.1.1 Connection Mode, above
F4.2.2 Format
Format selections for the COM2 connection are detailed in Table 3-9 and Table 3-10.
Table 3-9: COM2 RS-232/USB Format Assignments
Port Assignment if connection mode is Print or Auto print
F4.2.2.1 Line Format Multi-line, Single-line
F4.2.2.2 Print Language English, Chinese
COM1= RS-
232 or USB
F4.2.2.3 Add Line Feed 0,1,2,3,4,5,6,7,8,9
Assignment if Connection mode is Auto Print
F4.2.2.4 Auto Print Threshold 0 - full capacity
F4.2.2.5 Auto Print reset threshold 0 - full capacity
3-36 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.7.2.3.
3.7.2.3.1.
3.7.2.3.2.
3.7.2.4.
30094756 | 04 | 04/2021
Table 3-10: COM2 RS-422/RS-485 Format Assignments
Port Assignment if connection mode is Print or Auto print
F4.2.2.1 Line Format Multi-line, Single-line
F4.2.2.2 Print Language English, Chinese
COM1= RS-
422, RS-
485
F4.2.2.3 Add Line Feed 0,1,2,3,4,5,6,7,8,9
Assignment if Connection mode is Auto Print
F4.2.2.4 Auto Print Threshold 0 - full capacity
F4.2.2.5 Auto Print reset threshold 0 - full capacity
F4.2.3 Port Parameters
If COM 2 is RS422/485 option, F4.2.3.5 is used to set the net address.
F4.2.3.4 RS Type (Only available on COM2)
F4.2.3.4 sets the type of option board used by COM2. Selections are:
RS-232 [default], RS422, RS485
F4.2.3.5 Net Address (only available for RS4xx)
Address selections are:
0ff [default], 0, 1, 2, 3, 4, 5, 6, 7, 8, 9
Except when it is in SICS mode, RS-422/485 communication works in the same way as RS-232.
When a net address is activated in SICS mode, the terminal can only accept a command that
includes the correct net address, as shown in Table 3-11 and the example below.
For further information about SICS communication, refer to Appendix C, Communications.
Table 3-11: Net Address Commands, SICS Mode
Send:
Receive:
ESC (Header)
ESC
ESC
Address Command
9 SI
9 S S 0.00 kg
For example, when net address of RS422/485 is 9, the command should be:
S: ESC9SI
R: ESC9S S 0.00 kg
F4.2.4 Checksum
When F4.2.1, Connection Mode, is set to Toledo Continuous, a Checksum can be enabled. Settings are:
On – Checksum control enabled, Off – Checksum disabled [default]
METTLER TOLEDO IND231/IND236 User's Guide 3-37
3.8.
F5 Maintenance
Figure 3-69 shows a detailed view of the parameters available in the Maintenance branch of the
setup menu. Each of these setup parameters is described in this section.
3.8.1.
3.8.1.1.
Figure 3-69: Maintenance Menu Structure
F5.1 Calibration values
The Calibration Values screens display the current calibration values for the scale. The number of screens displaying test load calibration values (F5.1.1 to F5.1.5) is determined by the Linearity
Adjustment setting configured for the scale – refer to F1.3 Calibration starting on page 3-11.
These calibration values can be printed out (refer to F5.8 Print Configuration, on page 3-42) or
written down, then manually entered into a new replacement board should a failure ever occur, which eliminates having to recalibrate the scale with test weights. While this method is quick, it is not as accurate as placing test weights on the scale.
Use the direction keys to select a calibration value to be modified. Press the key to return to the menu.
If the terminal is not legal for trade (refer to F1.1.2 Approval on page 3-8), F5.1 is hidden.
F5.1.1 Zero Count
Displays the raw count value of the zero point set in most recent calibration.
3-38 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
3.8.2.1.
3.8.2.2.
3.8.2.3.
3.8.2.4.
3.8.2.5.
3.8.2.6.
3.8.3.
3.8.1.2.
3.8.1.3.
3.8.1.4.
3.8.1.5.
3.8.2.
F5.1.2 Load 1 Weight (Half Capacity)
If linearity calibration has been performed, displays the half capacity weight value.
F5.1.3 Load 1 Count (Half Capacity)
If linearity calibration has been performed, displays the half capacity raw count value.
F5.1.4 Load 2 Weight (Full Capacity)
Displays the full capacity weight value set in the most recent calibration.
F5.1.5 Load 2 Count (Full Capacity)
Displays the full capacity raw count value set in the most recent calibration.
F5.2 – Statistics
This weighing instrument features several control functions to monitor the condition of the device. A
METTLER TOLEDO service technician can set up and enable these functions. This helps the user and the METTLER TOLEDO service technician to determine how the device is treated and what measures are needed to keep it in good working order.
All statistics refer to the period since the last master reset. Refer to page 3-42.
F5.2.1 Number of Weighments
Displays the sequence number of the most recent weighing operation.
F5.2.2 Number of Overloads
Displays the number of overloads recorded by the terminal.
F5.2.3 Peak Weight Value
Displays the highest recorded weight value from all weighments.
F5.2.4 Number of Zero Commands
Displays the number of times a zero command has been issued.
F5.2.5 Number of Zero Command Failures
Show the number of failures of zero operation.
F5.2.6 Number of Charge Cycles
Shows the number of times the battery has been charged.
F5.3 Keyboard Test
The Keyboard Test screen allows the terminal keys to be tested, including the scale function keys and the print key.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-39
Press any of the keys shown in Table 3-12, and the screen will display the corresponding graphic.
When testing is complete, press the key to return to the menu.
Table 3-12: Keyboard Test Displays
3.8.4.
F5.4 Display Test
When F5.4 is accessed, the Display Test screen displays all LCD segments. This screen will continue to display until the key is pressed to return to the menu.
3.8.5.
3.8.5.1.
Figure 3-70: Display Test, All LCD Segments Displayed
F5.5 Serial Test
F5.5.1 COM1 Test
The COM1 test allows the COM1 connector on the mainboard to be tested. Short circuit (jumper)
COM1 TxD Pin and RxD Pin.
3-40 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
The terminal will display a screen like the one shown in Figure 3-71, with transmitted data shown
on the left and received data on the right. If the transmitted and received data match, the COM1 circuit is correct. If they are not the same, the circuit is faulty.
3.8.6.
3.8.6.1.
3.8.6.2.
3.8.7.
Figure 3-71: COM1 Serial Test Screen
Please press key to end the test.
F5.6 Discrete I/O Test
This function tests the status of inputs and outputs in the discrete I/O connector. The terminal can display the status of input x (where x is either input 1 or input 2) Outputs 1 through 4 can be tested by selecting On or Off from the F5.6.
x screen (where x 1, 2, 3 or 4).
F5.6 is visible in the menu structure only if COM2 is the optional discrete I/O board.
F5.6.1, F5.6.2 Input 1 Test and Input 2 Test
These screens display the test status of Inputs 1 and 2, respectively. Inputs are shown as on or off.
F5.6.3, F5.6.4, F5.6.5, F5.6.6 Output Tests
WARNING
THE DISCRETE OUTPUTS OF THE TERMINAL WILL BE MANUALLY ENABLED DURING THIS TEST.
REMOVE OUTPUT CONTROL POWER SO EXTERNAL EQUIPMENT WILL NOT BE ENERGIZED BY
MISTAKE. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE
MADE WITH POWER ON. FAILURE TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY
HARM AND/OR PROPERTY DAMAGE.
These screens display the test status of Outputs 1, 2, 3 and 4 respectively. Outputs are shown as on or off.
F5.7 Raw counts
The F5.7 screen displays the terminal’s current raw count value.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 3-41
3.8.8.
F5.8 Print Configuration
When F5.8 is accessed, a list of all visible setup parameters and calibration values is printed. If the terminal is approved, hidden setup blocks will not be printed. While the data is being sent, the message “Print” is displayed.
3.8.9.
Figure 3-72: Terminal Configuration Print Message
F5.10 Reset All to Factory Default Settings
Factory default settings can be restored individually for branches such as scale, application, terminal and communication, usually from the last branch of the respective menu structures. With the exception of metrologically significant settings such as Scale Type and Capacity, all parameters can be restored to their factory defaults using the Reset All screen under the Maintenance branch.
To initiate the reset, access the F5.10 screen and press either or .
Figure 3-73: Reset All Menu Screen
A message will prompt “Sure?”
Press or to abort the resetting process and return to the menu tree.
Figure 3-74: Global Reset Confirmation Prompt
3-42 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
4 Service and Maintenance
IND231/IND236 terminal is designed to provide years of dependable operation. However, METTLER
TOLEDO recommends that – as with any industrial measurement equipment – the terminal and the connected scale system be serviced periodically. Timely, factory-specified maintenance and calibration by a METTLER TOLEDO authorized service technician will ensure and document accurate and dependable performance to specifications.
4.1.
Cleaning and Maintenance
Clean the terminal’s keypad and cover with a clean, soft cloth that has been dampened with a mild glass cleaner. Do not use any type of industrial solvent such as toluene or isopropanol (IPA) that could damage the terminal’s finish. Do not spray cleaner directly on the terminal.
Regular maintenance inspections and calibration by a qualified service technician are recommended. The IND231/IND236 terminal is a rugged instrument; however, the front panel is a relatively thin covering over sensitive electronic switches and a lighted display. Care should be taken to avoid any punctures to this surface or any vibrations or shocks to the instrument. If the front panel is punctured, ensure that steps are taken to prevent dust and moisture from entering the unit until the terminal can be repaired.
4.2.
Service
Only qualified personnel should perform installation, programming, and service. Please contact a local authorized METTLER TOLEDO representative for assistance.
METTLER TOLEDO recommends periodic preventative maintenance to the terminal and scale system to ensure reliability and to maximize service life. All measurement systems should be periodically calibrated and certified as required to meet production, industry and regulatory requirements. We can help you maintain uptime, compliance and quality system documentation with periodic maintenance and calibration services. Contact your local METTLER TOLEDO authorized service organization to discuss your requirements.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO
OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
METTLER TOLEDO IND231/IND236 User's Guide 4-1 30094756 | 04 | 04/2021
4.3.
Battery Operation
As Table 4-1 shows, two kinds of battery configuration can be chosen to supply power to the
terminal.
4.3.1.
Table 4-1: Battery Options
Battery
6 "AA" size batteries
IND231 IND236
YES
NiMH rechargeable battery YES
NO
YES
Battery Use
The amount of time the IND231/IND236 will operate on a fully charged battery depends on the number of load cells connected and the use of the backlight. Powering multiple load cells and using the backlight continuously will greatly reduce battery operating time. The AA sized batteries and the NiMH battery pack have similar operating times.
Table 4-2 lists the average expected operating time provided by a new IND231/IND236 battery,
based on backlight usage and the number of load cells. As a battery pack ages, its power storage capability diminishes, and operating time will be reduced.
Table 4-2: Expected Battery Life
4.3.2.
Continuous Operation Load
Single 350Ω cell, no options
Four 350Ω cells, no options
Single 350Ω cell, COM2 option
Four 350Ω cells, COM2 option
Backlight on
80 hrs
39 hrs
47 hrs
29 hrs
Backlight off
120 hrs
46 hrs
58 hrs
33 hrs
Battery Disposal
Alkaline batteries cannot be recharged, so when the batteries reach the point where they can no longer power the IND231 terminal, all six batteries must be replaced.
As a result of normal use over a period of time, the available power in the 30044650 NiMH battery pack will be reduced and it will lose its ability to power the IND231/IND236 terminal for an extended time. When this occurs, the battery pack should be replaced and the original battery pack must be disposed of properly.
CAUTION
DISPOSE OF USED BATTERY PROMPTLY. KEEP AWAY FROM CHILDREN. DO NOT DISASSEMBLE AND DO
NOT DISPOSE OF IN FIRE.
Follow local laws and regulations regarding the disposal of the NiMH battery pack.
4-2 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
4.3.3.
Battery Status Icon
The IND231/IND236 terminal provides a battery status icon in the lower right corner of the display
to indicate the amount of power remaining in the battery. Table 4-3 shows the full power and lower
power icons.
Table 4-3: Battery Status Icon
Graphic Battery Condition
Full power
Less than 10% charge remaining
When the remaining battery power reaches the “extremely low” threshold, the battery graphic on the display will blink, indicating that the battery is getting low. It will continue to blink until the batteries are replaced (alkaline battery version) or the terminal is connected to AC power (NiMH version). If the battery condition is not addressed and the voltage continues to drop, the terminal will turn itself off.
CAUTION
THE BATTERY USED IN THIS DEVICE MAY PRESENT A RISK OF FIRE OR CHEMICAL BURN IF
MISTREATED. DO NOT CRUSH, DISASSEMBLE, HEAT ABOVE 60 ° C OR INCINERATE. REPLACE BATTERY
WITH 30044650 ONLY. USE OF ANOTHER BATTERY MAY PRESENT A RISK OF BURN, FIRE OR
EXPLOSION.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 4-3
4.3.4.
Alkaline Battery Version
To access the alkaline batteries, it is necessary to remove the plastic battery cover from the back of the IND231 terminal. Press down on the end marked PUSH and slide the cover in the direction of the arrows.
Figure 4-1: Removing the Battery Cover
In the IND231, six AA size battery cells are used. The cells are mounted in a plastic holder, shown
installed in Figure 4-2. Note the orientation of the positive poles.
+ + +
+ + +
Figure 4-2: IND231 AA Cell Holder Empty (left) and With Cells (right)
The alkaline batteries can be replaced without affecting the terminal’s regulatory approvals. Replace the dry cells with exactly the same size. High quality alkaline batteries are recommended.
engage.
4-4 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
4.3.5.
4.3.5.1.
Rechargeable NiMH Battery Pack Version
Both the IND231 and the IND236 can be configured to use the rechargeable NiMH battery pack.
The battery pack is internally mounted and is not designed to be replaced. It should be accessed only by qualified service personnel.
For best battery performance, operate the IND231/IND236 with internal battery at an ambient temperature in the range of 10 ° C to 30 ° C (50 ° F to 86 ° F). This applies to charging as well as discharging the battery.
Battery Charging
The 7.2 volt NiMH battery pack is designed to be charged while inside IND231/IND236 by connecting the AC line cord to an external power source.
Once an NiMH powered terminal is connected to AC power, the charging process begins. The
battery graphic changes as shown in Table 4-3 to indicate the status of the charge. The
approximate time to recharge a fully discharged battery is 5 hours. The terminal can be used for normal weighing operation under AC power while simultaneously recharging the battery.
When the battery is completely charged, the full power graphic will be displayed until the terminal is turned off and the power plug is pulled out.
4.4.
Troubleshooting
Troubleshooting activities described here are to assist in identifying whether the problem is in the
IND231/IND236 terminal or has an external cause.
4.4.1.
4.4.2.
• Battery Power
• Problem Diagnosis
• Error Codes and Error Messages
• Internal Diagnostic
Battery Power
In an AA battery powered IND231 terminal, if the terminal will not turn on when the On/Off key is pressed, replace the batteries. If the terminal still will not turn on, contact METTLER TOLEDO service.
Problem Diagnosis
The following chart lists a few potential symptoms and some suggestions for resolving the issue.
Use qualified electricians to test for problems with the AC power source. If a problem that is not
listed in Table 4-4 occurs, or if the suggested fix does not resolve the problem, contact an
authorized METTLER TOLEDO service representative for assistance.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 4-5
4.4.3.
Symptom
Display is blank
Display is dim
Display on but weight does not change
No serial port communications
Discrete inputs or outputs don’t operate
Table 4-4: Symptoms and Corrective Actions
Suggestion
1.
Check the connection between the display and mainboard
2.
Confirm On/Off key is functional.
3.
Confirm source of AC power is OK or confirm battery is fully charged
(NiMH) or replace the batteries (alkaline).
4.
Contact service representative.
1.
Confirm backlight is set correctly.
2.
Check to see if backlight harness is loose
3.
Check wiring to load cell. Look for broken wires, open connections or miswiring.
4.
Contact service representative.
1.
In setup, access F5.5, the serial diagnostics section of Maintenance, to determine if the problem is internal or external to the terminal.
2.
Contact service representative.
1.
In setup, access F5.6, the discrete input and output diagnostics section of
Maintenance, or the DIO status display on the system line, to determine if the problem is internal or external to the terminal.
2.
Contact service representative.
Error Codes and Error Messages
The IND231/IND236 terminal uses a combination of error codes and error messages to indicate
Message
Table 4-5: Error Codes
Possible reasons
Over load, more than 9d above scale capacity
Solution
Reduce the load
Under Zero by more than 5d
Outside the zero range
Incorrect entry or invalid key press
EEPROM verify error
Scale calibration failure
Zero the scale
Remove the load
Confirm that all initial weight is on the scale platform
Wait for message to disappear, make correct entry
1.
Power cycle the terminal
2.
Contact service representative
1.
Recalibrate the scale
2.
Contact service representative
4-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Message Possible reasons
Sample number too small
Solution
Increase sample quantity
EEPROM W/R error Contact service representative
When F1.4.2 = 20dZero, the display will show this error after the weight drops more than 20d below gross zero.
Clear platform and press the zero key
Setup access denied
Press the metrology switch to regain access to the setup menu.
Scale not calibrated Calibrate the scale
When the terminal is used as a remote display: Communication error in the port configured for remote display
1.
Check the settings of the port (rCOM n) set as remote display interface
2.
Confirm main terminal is programmed for data output
3.
Check cabling
Terminal turns itself off
The data is longer than the screen can display
Terminal is set auto power off
Battery voltage too low
Reduce the load on the scale
Press On/off key
Recharge battery
Error messages appear in the middle part of the display. The message is displayed briefly, then the
display returns to its state before the error was detected. Figure 4-3 shows an example of an error
message display – in this case, an incorrect value entry or invalid button press.
4.4.4.
Figure 4-3: Example Error Message Display
Internal Diagnostics
The IND231/IND236 terminal provides several internal diagnostic functions that are accessible in setup mode. These tests are intended to assist in diagnosing whether a problem is internal or external to the terminal.
To access these blocks, enter the operator menu and navigate to the F5 Maintenance menu. The diagnostic tests are described in the following sections.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 4-7
4.4.4.1.
Display Test
Enter the F5.4 Display test. The terminal will display all segments continuously until either or
is pressed to exit.
4.4.4.2.
Figure 4-4: Display Test Screen
Keyboard Test
The Keyboard Test screen allows the terminal keys to be tested, including the scale function keys and the print key.
Press any of the keys shown in Table 4-6, and the screen will display the corresponding graphic.
When testing is complete, press the key to return to the menu.
Table 4-6: Keyboard Test Displays
4-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
4.4.4.3.
Serial Test
Serial communication can be tested at F5.5. The default serial port is tested at F5.5.1. If the optional COM2 is present, it can be tested at F5.5.2. To fully test the port, a loop-back jumper must be installed between the Tx and Rx pins on the COM port to be tested.
Access F5.5.
n to display the serial test screen for the selected port.
4.4.4.4.
Figure 4-5: Serial Test Screen – Transmitted (left) and Received (right) Data
The terminal will transmit a series of two characters, counting up from 00 to 99. These are displayed at the left of the screen. If the port is functioning correctly and the communication loopback circuit is complete, the received numbers will be displayed at right.
Press to exit the test screen and return to the menu tree.
Discrete I/O Test
WARNING
THE DISCRETE OUTPUTS OF THE TERMINAL WILL BE MANUALLY ENABLED DURING THIS TEST.
REMOVE OUTPUT CONTROL POWER SO EXTERNAL EQUIPMENT WILL NOT BE ENERGIZED BY
MISTAKE. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST
BE MADE WITH POWER ON. FAILURE TO OBSERVE THESE PRECAUTIONS CAN RESULT IN
BODILY HARM AND/OR PROPERTY DAMAGE.
If the COM2-Discrete I/O option board is installed, its function can be tested at F5.6.
F5.6.1 and F5.6.2 display the current status – on or off – of the discrete inputs. The display status will change as the input is toggled on and off. Press to exit the status display and move to the next menu branch.
The status of the four discrete outputs is displayed at F5.6.3 – F5.6.6. Press either or to turn an output on and off. Press to exit the status display and move to the next menu branch.
Figure 4-6: Discrete Output Status Display
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 4-9
4.4.4.5.1.
4.4.4.5.2.
4.4.4.5.3.
4.4.4.5.4.
4.4.4.5.5.
4.4.4.6.
4.4.4.5.
4.4.4.6.1.
4.4.4.6.2.
4.4.4.6.3.
4.4.4.6.4.
Scale Diagnostics
The F5.1 Calibration Values screens display the current calibration values for the scale. The number of screens that display test load calibration values is determined by the Linearity Adjustment setting configured for the scale in setup at Scale > Calibration.
These calibration values can be recorded and then manually entered into a new replacement board should a failure ever occur, which eliminates having to recalibrate the scale with test weights. While this method is quick, it is not as accurate as performing a calibration using test weights placed on the scale.
Use the direction keys to select a calibration value to be modified.
Press to return to the menu.
F5.1.1 Zero counts
Displays the raw count value of the zero point set in most recent calibration.
F5.1.2 Load 1 weight (half capacity)
If linearity calibration has been performed, displays the half capacity weight value.
F5.1.3 Load 1 counts (half capacity)
If linearity calibration has been performed, displays the half capacity raw count value.
F5.1.4 Load 2 weight (full capacity)
Displays the full capacity weight value set in the most recent calibration.
F5.1.5 Load 2 counts (full capacity)
Displays the full capacity raw count value set in the most recent calibration.
Scale Statistics
This weighing instrument features several control functions to monitor the condition of the device.
The METTLER TOLEDO service technician can set up and enable these functions. This helps the user and the METTLER TOLEDO service technician to determine how the device is treated and what measures are needed to keep it in good working order.
F5.2.1 Number of Weighments
Displays the number of weighments made on the terminal.
F5.2.2 Number of Overloads
Displays the number of overloads recorded by the terminal.
F5.2.3 Peak Weigh Value
Displays the highest recorded weight value from all weighments.
F5.2.4 Number of Zero Commands
Displays the number of times a zero command has been issued.
4-10 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
4.4.4.6.5.
4.4.4.6.6.
F5.2.5 Number of Zero Command Failures
Show the number of zero commands that have failed.
F5.2.6 Number of Charge Cycles
Shows the number of times the NiMH battery pack has been charged.
4.5.
Connection to InSite™ SL
When connecting the IND231/236 to InSite SL (for saving and loading settings), the COM1 port must be used. InSite will be able to connect to the COM1 port regardless of its assignment (Print,
Autoprint, SICS or Continuous) so the port configuration does not need to be changed. Make sure the IND231/IND236 and InSite baud rate and data bits settings match before making the connection. These settings should be:
• Baud rate: 9600
• Data bits/parity: 8-none
Refer to the InSite SL User’s Guide for details on the tool’s functions and capabilities.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide 4-11
A Parameter Values
The following table lists each of the setup parameters shown in the Setup mode. A descriptive name is included, followed by a list of the selections with corresponding selection values.
Default values are indicated by an asterisk [*].
F-Code
F1.1.
Descriptive Name
Scale Type
Selections and Values
F1.1.1
F1.1.2
F1.1.3
F1.2
Type
Approval
Remote Protocol
Capacity and Increment
*SCL
RCOM1
RCOM2
*None
OIML
NTEP
AR
SRI
*Toledo Continuous
SICS
F1.2.1
F1.2.2
F1.2.3
F1.2.4
F1.2.5
Unit
Range
Range 1 Capacity
Increment Size 1
Range 2 Capacity
[only available with
F1.2.2= 2r] g
*kg oz lb
*1r: One range
2r: Tow ranges
*50, Range :1~500’000
*0.01, Selections are based on resolution range:
1000~30’000
*50, Range :1~500’000
F1.2.6
Increment Size 2
[only available with
F1.2.2= 2r]
0.0001
0.0002
0.0005
0.001
0.002
0.005
*0.01
METTLER TOLEDO IND231/IND236 User's Guide A-1 30094756 | 04 | 04/2021
A-2
F1.3
F1.3.1
F1.3.2
F1.3.3
F1.4
F1.4.1
F1.5
F-Code
F1.4.2
F1.4.3
F1.4.4
F1.5.1
F1.5.2
F1.5.3
Descriptive Name
2
5
10
20
0.02
0.05
0.1
0.2
0.5
1
50
100
200
Calibration
GEO
Calibration
Linearity Calibration
Zero
*16 [01 – 31]
Auto Zero maintenance
Zero Blanking
Power up Zero
[Unit : %]
Pushbutton Zero
[Unit : %]
Off
*2
10
20
Off
2
*10
18
20
Off
*0.5d
1d
3d
10d
*Off
20d
20dZero
Selections and Values
Tare
Pushbutton Tare
Tare Interlock
Auto Tare
Off
*On
*Off
On
*Off
On
METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
30094756 | 04 | 04/2021
F1.6
F1.7
F2.1
F-Code
F1.5.4
F1.5.5
F1.5.6
F1.7.1
F1.7.2
F1.10
F2.1.1
F2.1.2
Descriptive Name
Auto Tare Threshold
[only available with
F1.5.3=On]
Auto Tare Reset Threshold
[only available with
F1.5.3=On]
Auto Clear Tare
Second Unit
Selections and Values
Range 0~FS [Full scale] *10
Range *0~FS [Full scale]
*Off
On
*g kg oz lb
Filter and Stability
Filtering
Motion Range
Low
*Middle
High
Off
*0.5d
1d
3d
Scale Block Reset
Operation
Function Key 1
Function Key 2
* Expand display
Unit switch
Check Weigh
Recall
Count
Animal Weigh
Accumulation
Date & Time
Backlight adjustment
Zero
Tare
Expand display
*Unit switch
Check Weigh
Recall
Count
Animal Weigh
Accumulation
Date & Time
Backlight adjustment
Zero
METTLER TOLEDO IND231/IND236 User's Guide A-3
F2.2
F-Code Descriptive Name
Over/Under Database
[Maximum record number
=10]
Selections and Values
Tare
Record 1
Record 2
……
Record n
Create Record:
record number
Target
Tol-
Tol+
Tare
F2.3
Count Weigh
F2.3.1
APW Enhancement
*Off
On
F2.4
F2.4.1
Animal Weigh
Auto Start
*Off
On
F2.5
F2.5.1
F2.5.1.1
F2.5.1.2
Discrete I/O
Input 1
Input 2
Input
*Off
Zero
Tare
Unit switch
Clear
Blank display
F2.5.2
Output
F2.5.2.1
F2.5.2.2
F2.5.2.3
Output 1
Output 2
Output 3
*Off
Over Tolerance
Under Tolerance
Good Range
Net
Motion
Over Load
Under Load
Center of Zero
F2.5.2.4
F2.10
F3.1
F3.1.1
F3.1.2
F3.2
Output 4
Application Block Reset
Device
Serial Number
Firmware Version
Display
[10 digits max.] xx.xx.xx
A-4 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
30094756 | 04 | 04/2021
F3.3
F4.1
F-Code
F3.2.1
F3.2.2
F3.2.3
F3.2.4
F3.3.1
F3.3.2
F3.3.3
F3.3.4
F3.10
F4.1.1
F4.1.2
F4.1.2.1
F4.1.2.2
F4.1.2.3
Descriptive Name
Timeout
[Unit : sec]
Brightness
Auto Power Off
[Unit : min]
Selections and Values
0
5
10
15
30
*60
120
300
600
Off
*Low
High
*0 (AC version)
1
*5 (battery version)
15
30
60
*0[0-9] Weight Hold
Date & Time
Date Format
Date
Time
Time Format
MMDDYY
DDMMYY
*YYMMDD xx-xx-xx
12h
*24h xx:xx:xx
Terminal Block Reset
COM1
Mode
[Only available when
F1.1.1
≠ R COM1
Auto Print
SICS
Toledo Continuous
Format
[Only available when
F4.1.1 =Print or Auto print, and F1.1.1
≠
R COM1]
Line Format
Print Language
*Multi-line
Single-line
*English
Chinese
Add Line Feed 0,1,2,*3,4,5,6,7,8,9
METTLER TOLEDO IND231/IND236 User's Guide A-5
F4.2
F-Code
F4.1.3
F4.1.4
F4.2.1
F4.2.2
F4.1.2.4
F4.1.2.5
F4.1.3.1
F4.x.10
F4.2.2.1
F4.2.2.2
F4.2.2.3
F4.1.3.2
F4.1.3.3
Descriptive Name
Auto Print Threshold
[Only available when
F4.1.1=Auto print]
Auto Print reset threshold
[Only available when
F4.1.1=Auto print]
Parameters
Selections and Values
*0 - full capacity
*0 - full capacity
Baud rate
Data Bits / Parity
Flow Control
1200
2400
4800
*9600
19200
38400
57600
115200
7 odd
7 even
*8 none
*Off
On
Checksum
[Only available when
F4.1.1=Toledo
Continuous]
COM1 Block Reset
[Per port, by number]
*Off
On
COM2/USB [Visible only when option installed]
Connection Mode
[Only available when
F1.1.1
≠ RCOM2]
Auto Print
SICS
Toledo Continuous
Format
[Only available when
F4.2.1 =Print or Auto print, and F1.8.1
≠ Com
2]
Line Format
*Multi-line
Single-line
Print Language
*English
Chinese
Add Line Feed 0,1,2,*3,4,5,6,7,8,9
A-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
30094756 | 04 | 04/2021
F-Code
F4.2.2.4
F4.2.2.5
F4.2.3
F4.2.3.1
F4.2.3.4
F4.2.4
F4.2.10
F5.1
F5.1.1
F5.1.2
F5.1.3
F5.1.4
F5.1.5
F4.2.3.2
F4.2.3.3
F4.2.3.5
F4.2.3.6
Descriptive Name
Auto Print Threshold
[only available when
F4.2.1=Auto print]
Auto Print reset threshold
[only available when
F4.2.1=Auto print]
Port Parameters
Selections and Values
*0 - full capacity
*0 - full capacity
Baud rate
Data Bits / Parity
Flow Control
1200
2400
4800
*9600
19200
38400
57600
115200
7 odd
7 even
*8 none
*Off
On
RS type
[Only available on Com 2]
*RS232
RS422
RS485
Net Address *0, 1, 2, 3, 4, 5, 6, 7, 8, 9
Net address [Only available for RS4xx]
Checksum
[Only available when
F4.2.1=Toledo
Continuous]
COM2 Block Reset
Calibration Values
*Off, 0, 1, 2, …, 9
*Off
On
Zero count
Load 1 weight [half capacity]
Load 1 count [half capacity]
Load 2 weight [full capacity] xxxxxxx xxxxxxx xxxxxxx xxxxxxx
Load 2 count [full capacity] xxxxxxx
F5.2
Statistics
METTLER TOLEDO IND231/IND236 User's Guide A-7
F5.3
F5.4
F5.5
F5.2.5
F5.2.6
F5.5.1
F5.6
F5.6.1
F5.6.2
F5.6.3
F5.6.4
F5.6.5
F5.6.6
F5.7
F5.8
F5.10
F-Code
F5.2.1
F5.2.2
F5.2.3
F5.2.4
Descriptive Name
Number of Weighs xxxxxxx
Selections and Values
Number of Overloads xxxxxxx
Peak Weight Value
Number of Zero
Commands
Number of Zero Failures xxxxxxx xxxxxxx xxxxxxx
Number of Charge Cycles xxxxxxx
Keyboard Test
Display Test
Serial Test
COM1 Test xx xx
[Range :00~99]
DIO Test [Visible only when option installed]
Input 1 On/Off
Input 2
Output 1
Output 2
On/Off
On/Off
On/Off
Output 3
Output 4
Raw Counts
On/Off
On/Off xxxxxxx
Print Configuration
Reset All
A-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.
Communications
B.1.
Serial Interface Parameters
The IND231/IND236 terminal supports one standard serial port and one optional serial port. These are designated COM1 (standard port on the Main PCB) and COM2 (optional).
COM1 provides an RS-232 interface only. The RS-232 interface is a three-wire (TxD, RxD, and
GND) with selectable XON-XOFF handshaking.
Optional COM2 provides RS-232 and RS-422/485 interfaces. The RS-232 interface is a three-wire
(TxD, RxD, and GND) with selectable XON-XOFF handshaking. The RS-485 connection is a twowire interface and supports multi-drop communication with addressing. The port must be configured in setup for use as either RS-232 or RS-485 due to the different operating requirements of the interfaces.
Character framing is programmable in the setup mode. Framing can be:
• 1 start bit
• 7 or 8 ASCII data bits (selectable)
• 0 or 1 parity bit (selectable as none, even, or odd)
• 1 stop bit
The baud rate can be configured from 1200 to 115.2K baud and a checksum character can also be configured when using the continuous output string.
The IND231/IND236 terminal serial ports support the following functions:
• Print output with CTPZ input
• Continuous output with CTPZ input
• SICS (level 0 and level 1)
B.2.
Demand Output Mode
The print output mode transmits data only when the terminal receives a print request. Print requests are sent to the IND231/IND236 terminal when:
• The operator presses PRINT
• A discrete input configured as print is triggered
• An ASCII “P” is sent through a demand or continuous port
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-1
B.2.1.
B.2.1.1.
• Auto print is enabled and all conditions for auto print are met
When triggered, data is transmitted in a string programmed in the output template portion of setup.
Print mode is used typically when sending data to a printer or PC on a transactional basis.
When the print output mode is assigned, the port is automatically set up to also receive certain
ASCII characters to duplicate keypad functions. Refer to the C,T,P,Z section later in this appendix for more details.
Output Templates
The IND231/IND236 terminal provides five formats to define the string of data to be transmitted and printed. The print template used is determined by which application is active. Print settings are further explained in the Communication section of Chapter 3, Configuration.
Standard Print Format
The multi-line output looks like:
A star symbol before the weight will be shown to indicate the weight is average weight if current mode is animal weighing.
Date
Time
Gross
Tare
Net
YYYY.MM.DD
HH:MM:SS
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Single-line output looks like (8 characters per number including “.”):
Date_YYYY.MM.DD_ _ Time_HH:MM:SS_ _ Gross_XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _
Net_XXXX.XXX_Unit
B-2 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.2.1.2.
B.2.1.3.
Over/Under Printout
The multi-line output looks like:
Date
Time
Record ID
Target
Tol+
Tol-
Gross
Tare
Net
YYYY.MM.DD
HH:MM:SS
XX
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Single-line output looks like:
Date_YYYY.MM.DD_ _ Time_HH:MM:SS_ _Record ID_XX_ _Target _XXXX.XXX_Unit_ _ Tol+_
XXXX.XXX _ Unit _ _ Tol-_ XXXX.XXX _ Unit _ _Gross _XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _
Net_XXXX.XXX_Unit
Counting Printout
The multi-line output looks like:
Date
Time
Pieces
APW
Tol-
Gross
Tare
Net
YYYY.MM.DD
HH:MM:SS
XXXXXXXX PCS
XXXX.XXX Unit
XX Unit
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Single-line output looks like:
Date_YYYY.MM.DD_ _Time _HH:MM:SS_ _ Pieces_XXXXXXX_PCS_ _APW _XXXX.XXX_Unit_ _
Gross_XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _Net _XXXX.XXX_Unit
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-3
B.2.1.4.
B.2.1.5.
Accumulate Printout
The multi-line output looks like:
Date
Time
Total
Count
Gross
Tare
Net
YYYY.MM.DD
HH:MM:SS
XXXX.XXX Unit
XX.XXX
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Single-line output looks like:
Date_YYYY.MM.DD_ _ Time_HH:MM:SS_ _ Total_XXXX.XXX_Unit_ _ Count_XXX_ _
Gross_XXXX.XXX_Unit_ _ Tare_XXXX.XXX_Unit_ _ Net_XXXX.XXX_Unit
Animal Printout
The multi-line output looks like:
Date
Time
Ave. G
Tare
Ave. N
YYYY.MM.DD
HH:MM:SS
XXXX.XXX Unit
XXXX.XXX Unit
XXXX.XXX Unit
Single-line output looks like:
Date_YYYY.MM.DD_ _ Time_HH :MM :SS_ _ Ave G_XXXX.XXX_Unit_ _Tare _XXXX.XXX_Unit_ _ Ave
N_XXXX.XXX_Unit
B.3.
Continuous Output Mode
The continuous output mode of the IND231/IND236 terminal can be used to continuously send weight data and scale status information to a remote device such as a PC or a remote display including an IND231 or IND236 terminal.
NOTE: When the continuous output mode is assigned, the port is automatically set up to also receive certain ASCII characters to duplicate keypad functions. Refer to the C,T,P,Z section later in this appendix for more details.
B-4 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.3.1.
Standard Continuous Output
Continuous Output can be assigned to COM1 or COM2. A checksum character can be enabled or disabled with continuous output. A data string will be output approximately 20 times per second for baud rates above 4800 baud. If a baud rate below 4800 is selected, the output rate will be slower.
At 300 baud, the output rate is only approximately 1 per second. The data consists of 17 or 18
Non-significant weight data and tare data digits are transmitted as spaces. The continuous output mode provides compatibility with METTLER TOLEDO products that require real-time weight data.
Table B-1 shows the format for the standard continuous output.
Table B-1: Standard Continuous Output Format
Character 1 2
Status
3
2
4
Indicated Weight 3 Tare Weight 4
5 6 7 8 9 10 11 12 13 14 15 16 17 18
Data STX 1 SB-A SB-B SB-C MSD LSD MSD LSD CR 5 CHK 6
Continuous Output Format Notes
1.
ASCII Start of Text character (02 hex), always transmitted.
2.
Status bytes A, B and C. Refer to Table B-2, Table B-3 and Table B-4 for details of the structure.
3.
Displayed weight, either gross or net. Six digits, no decimal point or sign. Insignificant leading zeroes are replaced with spaces.
4.
Tare weight. Six digits of tare weight data. No decimal point in field.
5.
ASCII Carriage Return <CR> character (0D hex).
6.
Checksum, transmitted only if enabled in setup. Checksum is used to detect errors in the transmission of data. Checksum is defined as the 2’s complement of the seven low order bits of the binary sum of all characters preceding the checksum character, including the <STX> and <CR> characters.
Table B-2, Table B-3 and Table B-4 detail the status bytes for standard continuous output.
1
1
0
1
1
0
0
2
0
0
1
1
0
1
0
1
1
0
Table B-2: Status Byte A Bit Definitions
Bits 2, 1, and 0
1
0
1
0
1
1
0
0
0
Decimal Point Location
XXXXX00
XXXXX0
XXXXXX
XXXXX.X
XXXX.XX
XXX.XXX
XX.XXXX
X.XXXXX
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-5
4
0
1
1
Bit 5
Bit 6
Bits 4 and 3
3
1
0
1
Build Code
X1
X2
X5
Always = 1
Always = 0
Status Bits
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Table B-3: Status Byte B Bit Definitions
Gross = 0, Net = 1
Function
Sign, Positive = 0, Negative = 1
Out of Range = 1 (Over capacity or Under Zero)
Motion = 1, Stable = 0 lb = 0, kg = 1 (see also Status Byte C, bits 0, 1, 2)
Always = 1
Zero Not Captured after power-up = 1
0
1
1
1
1
0
0
2
Bits 2, 1, and 0
1 0
0
0
0
1 0
0
1
1
0
0
1
1
0
1
1
0
1
Table B-4: Status Byte C Bit Definitions lb or kg, selected by Status Byte B, bit 4 grams (g) not used ounces (oz) not used not used not used no units
Weight Description
B-6 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
2
Bits 2, 1, and 0
1
Bit 3
0
Bit 4
Bit 5
Bit 6
Weight Description
Print Request = 1
Expand Data x 10 = 1, Normal = 0
Always = 1
Always = 0
B.4.
CTPZ
When a serial port is programmed as a print, continuous output, the CTPZ input mode is automatically assigned. The CTPZ input mode provides a method for a remote serial device to trigger several basic functions when a command character is sent to the terminal. A termination character is not required. Remote ASCII command characters include:
C – Clears the scale to gross
T – Tares the scale (causes a pushbutton tare)
P – Initiates a print command
Z – Zeros the scale
S – Switches units
ASCII command characters must be sent in upper-case letters. All other characters are ignored.
Note: Some IND231/IND236 applications have additional serial commands. These are described in the application details in the Application chapter.
Example
To initiate a pushbutton tare, program the terminal for print or continuous output for a specific port, program the serial port parameters to match the other device and then send the ASCII character “T”.
B.5.
Standard Interface Command Set (SICS)
Protocol
The IND231/IND236 terminal supports the METTLER TOLEDO Standard Interface Command Set
(MT-SICS), which is divided into four levels (0, 1, 2, 3), depending on the functionality of the device. This terminal supports parts of levels 0 and 1:
• MT-SICS level 0 Command set for the simplest device.
• MT-SICS level 1 Extension of the command set for standard devices.
• MT-SICS level 2 Extension of the command set by the commands specific for a scale family.
• MT-SICS level 3 Extension of the command set by commands specific for a certain application of scale family.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-7
B.5.1.
B.5.2.
A feature of this interface is that the commands combined in MT-SICS level 0 and 1 are identical for all devices. Both the simplest weighing device and a fully expanded weighing workstation can recognize the commands of MT-SICS levels 0 and 1.
MT-SICS level 0 and level 1 contain the following functions :
Request Weighting results
Tare the scale and preset the tare weight
Zero the scale
Identify MT-SICS implementation
Identify the scale
Reset the scale
Version Number of the MT-SICS
Each level of the MT-SICS has its own version number, which can be requested with the command
I1 from level 0. This terminal supports:
• MT-SICS level 0, version 2.2x
• MT-SICS level 1, version 2.2x (except the D, DW and K commands)
• MT-SICS level 2, version 2.2x
• MT-SICS level 3, version 2.2x
Command Formats
Each command received by the terminal via the SICS interface is acknowledged by a response to the transmitting device. Commands and responses are data strings with a fixed format. Commands sent to the terminal comprise one or more characters of the ASCII character set. Commands must be in upper-case.
The parameters of the command must be separated from one another and from the command name by a space (ASCII 32 dec). In the examples shown in this section, a space is represented as
_.
Each command must be terminated by <CR>< LF> (ASCII 13 dec., ASCII 10 dec.).
SI SP C
R
L
F
SI
SP
C
R
L
F
ASCII Command, 1-2 Bytes, must be capital letters
Space (ASCII 32 dec.).
Carriage Return (ASCII 13 dec.).
Line Feed (ASCII 10 dec.).
The characters <CR> and <LF>, which can be input using the ENTER or RETURN key of most PC terminal keypads, are not shown in this description; however, it is essential they be included for communication with the terminal.
B-8 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.3.
B.5.3.1.
Example
Command to tare the terminal:
“TA_20.00_lb” (The command terminators <CR>< LF> are not shown.)
Response Formats
All responses sent by the terminal to the transmitting device acknowledging the received commands have one of the following formats:
Response with weight value
Response without weight value
Error message
Format of the Response with Weight Value
A general description of the response with weight value as follows:
B.5.3.2.
ID
__
Status
Weight Value
Unit–
CR
LF
Response identification.
Space (ASCII 32 dec)
Status of the terminal. See description of the commands and responses.
Weighing result, shown as a 10-digit number, including sign directly in front of the first digit. The weight value appears right justified. Preceding zeroes are suppressed with the exception of the zero to the left of the decimal point.
Weight unit displayed.
Carriage Return (ASCII 13 dec)
Line Feed (ASCII 10 dec)
Comment
The <CR> and <LF> characters will not be shown in the descriptions below.
Examplef
Response with a stable weight value of 0.256 kg:
S _ S_ _ _ _ _ _ 0.256 _ kg
Format of the Response without Weight Value
A general description of the response without weight value is as follows:
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-9
B-10
B.5.3.3.
B.5.4.
B.5.4.1.
B.5.4.2.
ID – Response identification.
__ – Space (ASCII 32 dec)
Status – Status of the terminal. See description of the commands and responses.
Parameters – Command-dependent response code.
CR – Carriage Return (ASCII 13 dec)
LF – Line Feed (ASCII 10 dec)
Comment
The <CR> and <LF> characters will not be shown in this description.
Format of Error Message
ID C
R
L
F
There are three different error messages, each identified by two characters:
ID – Error identification
Possible error messages are
ES
EL
Syntax error
The scale has not recognized the received command
Logical error
The scale cannot execute the received command
CR – Carriage return (ASCII 13 dec)
LF – Line Feed (ASCII 10 dec)
Comment
The <CR> and <LF> characters will not be shown in this description.
Tips for the Programmer
Tips for creating a robust communication with the terminal using the SICS protocol include:
Command and Response
Improve the dependability of application software by having the program evaluate the response of the terminal to a command. The response is the acknowledgment that the terminal has received the command.
Reset
When establishing communication between the terminal and system, send a reset command to the terminal to enable a start from a determined state. When the terminal or system is switched on or off, faulty characters can be received or sent.
METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.4.3.
B.5.5.
B.5.5.1.
30094756 | 04 | 04/2021
Quotation Marks (“ ”)
Quotation marks included in the command responses are used to designate fields and will always be sent.
Commands and Responses MT-SICS Level 0
The terminal receives a command from the system computer and acknowledges the command with an appropriate response. The following sections contain a detailed description of the command set, arranged in alphabetical order and including the associated responses. Commands and responses are terminated with <CR> and <LF>. These termination characters are not shown in the following description, but they must always be entered with commands or sent with responses.
The following MT-SICS level 0 commands are supported:
I0 Inquiry the list of all implemented MT-SICS commands
I1 Inquiry of MT-SICS level And MT-SICS versions
I2 Inquiry of scale data
I3 Inquiry of scale software version and type definition number
I4 Inquiry of serial number
S Send stable weight value
SI Send weight value immediately
SIR Send weight value immediately and repeat
Z Zero the scale when stable
ZI Zero the scale immediately irrespective of stability
@ Reset
The following are detailed descriptions of these Level 0 commands:
I0 – INQUIRY OF ALL IMPLEMENTED MT-SICS COMMANDS
Command:
Response:
I0
I0_B_0_“I0”
I0_B_0_“I1”
I0_B_0_“I2”
I0_B_0_“I3”
I0_B_0_“I4”
I0_B_0_“S”
I0_B_0_“SI”
I0_B_0_“SIR”
I0_B_0_“Z”
Level 0 “IO” command implemented
Level 0 “I1” command implemented
Level 0 “I2” command implemented
Level 0 “I3” command implemented
Level 0 “I4” command implemented
Level 0 “S” command implemented
Level 0 “SI” command implemented
Level 0 “SIR” command implemented
Level 0 “Z” command implemented
METTLER TOLEDO IND231/IND236 User's Guide B-11
B.5.5.2.
B.5.5.3.
Response:
I0_B_0_“ZI”
I0_B_0_“@”
I0_B_1_“T”
I0_B_1_“TI”
I0_I
Level 0 “ZI” command implemented
Level 0 “@” command implemented
Level 1 “T” command implemented
I0_B_1_“TAC” Level 1 “TAC” command implemented
I0_B_1_“TA” Level 1 “TA” command implemented
Level 1 “TI” command implemented
I0_B_2_“PRN” Level 1 “PRN” command implemented
I0_B_2_“SIH” Level 1 “SIH” command implemented
I0_B_3_“DAT” Level 1 “DAT” command implemented
I0_A_3_“TIM” Level 1 “TIM” command implemented
Cannot execute command at this time.
I1 – INQUIRY OF MT-SICS LEVEL AND MT-SICS VERSIONS
Command:
Response:
I1
I1_A_"0123"_"2.20"_"2.20"_"2.20"_"2.20"
Response: I1_I
“0123” IND231 supports levels 0 to 3
2.20 Level 0, version V2.2x
2.20 Level 1, version V2.2x
2.20 Level 2, version V2.2x
2.20 Level 3, version V2.2x
Command understood, not executable at present.
Comments
For the MT-SICS level, only fully implemented levels are listed. In this case, neither level 0 nor level
1 was fully implemented, so the level is not specified.
In the case of the MT-SICS version, all levels are specified even those only partially implemented.
I2 – INQUIRY OF DATA
Command: I2
Response:
Response:
I2_A_“IND231/IND236_XXyy” (XX = capacity of the terminal, yy = primary units)
I2_I
IND231/IND236
XXyy
Model number of terminal
Capacity and primary unit of the scale
Command understood, not executable at present.
Comment
The number of characters in the “text” field depends on the scale capacity.
B-12 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.5.4.
B.5.5.5.
B.5.5.6.
B.5.5.7.
I3 – INQUIRY OF SOFTWARE VERSION
Command: I3
Response:
Response:
I3_A_“30047226 01.00.00”
I3_I
30047226
01.00.00
BOM of the terminal
Firmware version of the terminal
Command understood, not executable at present.
Comment
The number of characters of “text” depends on the revision level.
I4 – INQUIRY OF SERIAL NUMBER
Command: I4
Response: I4_A_“123456”
Response: I4_I
123456 Serial number of terminal
Command understood, not executable at present.
S – SEND STABLE WEIGHT VALUE
Command: S
Response:
Response:
Response:
Response:
Comment
S_S_ _ _ _ _ 436.2_lb
436.2 stable displayed weight lb weight unit
S_I Command understood, not executable at present.
S_+ Terminal in overload range.
S_- Terminal in underload range.
The terminal will wait for up to 3 seconds after receiving an “S” command for no-motion. If motion does not settle within this time, the command is aborted and the S_I response is sent.
SI – SEND WEIGHT VALUE IMMEDIATELY
Command: SI
Response:
Response:
Response:
Response:
Response:
S_S_ _ _ _ _ 436.2_lb Stable weight value.
S_D_ _ _ _ _ 436.2_lb Non-stable (dynamic) weight value.
S_I Command understood, not executable at present.
S_+ Terminal in overload range.
S_- Terminal in underload range.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-13
B.5.5.8.
B.5.5.9.
Comment
The response to the command SI is the last internal weight value (stable or dynamic) before receipt of the “SI” command.
SIR – SEND WEIGHT VALUE IMMEDIATELY AND REPEAT
Command: SIR
Response:
Response:
Response:
Response:
Response:
S_S_ _ _ _ _ 436.2_lb Stable weight value.
S_D_ _ _ _ _ 436.2_lb Non-stable (dynamic) weight value.
S_I Command understood, not executable at present.
S_+ Terminal in overload range.
S_- Terminal in underload range
Example
Command:
Response:
SIR
S_D_ _ _ _ _ 129.07_kg
S_D_ _ _ _ _ 129.09_kg
S_S_ _ _ _ _ 129.09_kg
S_S_ _ _ _ _ 129.09_kg
S_D_ _ _ _ _ 114.87_kg
. . . The scale continues to send stable or dynamic weight values
Comments
The SIR command is overwritten and cancelled by the commands S, SI, SR, @ and hardware break and hence cancelled.
The data output rate is approximately 4 per second.
Z – ZERO
Command:
Response:
Response:
Response:
Response:
Comments
Z
Z_A Command performed, meaning the scale was in gross mode, scale was stable and weight was within the zero capture range.
Z_I
Z_+
Command understood, not executable at present.
Upper limit of zero setting range exceeded.
Z_- Lower limit of zero setting range exceeded.
The calibrated zero point determined during calibration is not influenced by this command.
B-14 METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.5.10.
B.5.5.11.
B.5.6.
B.5.6.1.
The terminal will wait for up to 3 seconds after receiving an “Z” command for no-motion. If motion does not settle within this time, the command is aborted and the Z_I response is sent.
ZI – Zero the scale immediately irrespective of stability
Command:
Response:
Comments
ZI
ZI D Zero setting performed under dynamic conditions
ZI S
ZI
Zero setting performed under stable conditions
Zero setting not performed
ZI+
ZI-
Upper limit of zero setting range exceeded
Lower limit of zero setting range exceeded
Terminal always responds Z I, if OIML or NTEP is approved.
@ – RESET
Command:
Response:
@
I4_A_“12345678”
“12345678” Serial number of the scale, the scale is ready for operation
Comments
Resets the scale to the condition found after switching on, but without a zero setting being performed.
Cancels all commands that are awaiting responses.
Clears the tare register.
Cancels the SIR and SR commands.
The Reset command is always executed except when the command is received by the terminal during the calibration and test procedure. In this case, the Reset command cannot be processed.
Commands and Responses MT-SICS Level 1
The following commands of MT-SICS level 1 are available:
SR
T
TA
TAC
Send weight value on weight change (send and repeat)
Tare
Pre-Tare
Clear tare value
TI Tare Immediately
SR – SEND WEIGHT VALUE ON WEIGHT CHANGE (SEND AND REPEAT)
Command: SR_PresetValue_Unit Send the current stable weight value and then continuously after every weight change greater or equal to the preset value
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-15
B-16
B.5.6.2.
Response:
Response:
Response:
Response:
Response:
Response:
Response:
Example
Command:
S_S_WeightValue_Unit
S_D_WeightValue_Unit
S_S_WeightValue_Unit
S_I
S_L
S_+
S_- a non-stable value followed by the next stable value, range =
1d to maximum load.
If no preset value is entered, the weight change must be at least 12.5% of the last stable weight value, minimum =
30d.
Current, stable weight value. Weight change.
Non-stable weight value.
Next stable weight value.
Command understood, not executable at present.
The command is not executable.
Terminal in overload range.
Terminal in underload range.
Response:
SR_0.50_kg Send the current stable weight value followed by every load change > 0.50 kg.
S_S_ _ _ _ _100.00_kg Scale stable.
S_D_ _ _ _ _115.23 _ kg More than 0.50 kg loaded.
S_S_ _ _ _ _200.00 _ kg Scale again stable.
Comments
SR is overwritten and cancelled by the commands S, SI, SIR, @ and hardware break.
If, following a non-stable weight value, stability has not been reached within the timeout interval, the response “S_I“ is sent and then a non-stable weight value. Timeout then starts again form the beginning.
The preset value must be entered in the first unit that is the weight unit displayed after the terminal has been switched on.
T – TARE
Command:
Response:
T
T_S_ _ _ _100.00_kg Tare performed, meaning the scale was stable and weight was within the weighing range.
Response:
Response:
Response:
Comments
T_I
T_+
Command understood, not executable at present.
Upper limit of tare setting range exceeded.
T_- Lower limit of tare setting range exceeded.
The existing tare will be overwritten and replaced by the new preset tare weight value.
The terminal will wait for up to 3 seconds after receiving a “T” command for no-motion. If motion does not settle within this time, the command is aborted and the T_I response is sent.
METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.6.3.
B.5.6.4.
B.5.6.5.
30094756 | 04 | 04/2021
TA – Inquiry/presetting of tare weight value
Inquiry:
Command:
Response:
TA
TA A
TA I
Inquiry of current tare value
Tare Value Unit
Current tare weight value in displayed unit
Command not executable
Setting:
Command:
Response:
TA A
TA A
TA I
TA L
Tare Value Unit
Entry of a tare preset value in displayed unit
Weight Value Unit
Taring performed with the preset tare value in displayed unit
Taring not performed
Command understood, parameter wrong
Example
Command:
Response:
Comments
TA 100.00 g
TA_A_____100.00 g The scale has 100.00g in the tare memory
The tare memory will be overwritten by the preset tare weight value.
The keyed in tare value will be automatically rounded by the scale to the current readability.
The preset value must be entered in the current unit.
The taring range is specified to the scale type.
TAC – CLEAR TARE VALUE
Command: TAC
Response: TAC_A Tare value cleared.
Response: TAC_I Command understood, not executable at present.
TI – TARE IMMEDIATELY
Command:
Response:
Response:
Response:
Response:
TI
TI_S_WeightValue_Unit
TI_D_WeightValue_Unit
TI_I
TI_L
Taring performed, stable tare value.
Taring performed, non-stable (dynamic) tare value.
Command understood, not executable at present.
The command is not executable.
METTLER TOLEDO IND231/IND236 User's Guide B-17
B-18
B.5.7.
B.5.7.1.
B.5.7.2.
B.5.8.
B.5.8.1.
Response:
Response:
TI_+
TI_-
Upper limit of taring range exceeded.
Lower limit of taring range exceeded.
Example
Command:
Response:
TI
TI_D_ _ _ _117.57_kg Tare taken with dynamic weight value.
Comments
Any previous tare value will be overwritten by the new tare weight value.
Tare weight values determined during motion may not be accurate.
The tare weight value is sent in the current units.
Commands and Responses MT-SICS Level 2
The following commands of MT-SICS level 2 are available:
PRN Printout
SIH High resolution weight
PRN – Start printout
Command:
Response
PRN
PRN A
PRIN-I
Start printout on printer channel
Command executed
Command not executable at present
SIH
Command:
Response
Example
Command:
Response:
PRN
H S WeightValue Unit
H D WeightValue Unit
Send the weight in high resolution immediately
Stable net weight in high resolution in current unit
Dynamic net weight in high resolution in current unit
Inquiry:
Command:
SIH
H_S___1.99982_kg
DAT
Current net weight in high resolution is 1.99982 kg and stable.
Commands and Responses MT-SICS Level 3
The following commands of MT-SICS level 3 are available:
DAT Date
TIM Time
DAT – Inquiry and setting of date
Inquiry of current date of the scale
METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
B.5.8.2.
Response:
Setting:
Command:
Response:
DAT A
DAT I
DAT
DAT A
DAT I
DAT L dd mm yyyy
Command not executable dd mm yyyy
Date has been set
Date cannot be set at present
Command not executed as date format is not correct
Inquiry:
Command:
Response:
Setting:
Command:
Response:
TIM – Inquiry and setting of time
TIM Inquiry of current time of the scale
TIM A hh mm ss Represent time in the format hours minutes seconds
TIM I Inquiry of time not possible
TIM hh mm ss Set time in 24-hour-format
TIM A Time has been set, clock running
TIM I
TIM L
The time cannot be set at present
Command not executed as time format is not correct
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide B-19
C.
GEO Codes
The GEO code feature provided in the IND231/236 terminal permits calibration readjustment due to changes in elevation or latitude without reapplying test weights. This adjustment assumes a previously accurate calibration was done with the GEO code set properly for that original location and that the GEO code for the new location can be accurately determined. The procedure for using this feature is as follows.
C.1.
Original Site Calibration
1.
Use the GEO code chart (Table C-1) on the following pages to determine the GEO code for the
current altitude and location at which the scale will be calibrated.
2.
Enter that GEO value into the GEO code parameter in setup at Scale > Calibration.
3.
Immediately after entering the GEO code, perform a zero and span adjustment using accurate test weights.
4.
Exit the setup menu tree.
5.
The scale can now be used in its new location.
C.2.
New Site GEO Code Adjustment
When a terminal is to be reinstalled at a different geographic location, gravitational and altitude changes can be accounted for by following these steps. Note that this procedure is not necessary if an on-site recalibration is performed.
1.
Use the GEO code chart (Table C-1) on the following pages to determine the GEO code for the
new altitude and location at which the scale will be used.
6.
Enter that GEO value into the GEO code parameter in Setup at Scale > Calibration.
7.
Immediately after entering the GEO code, exit the setup menu tree. DO NOT perform a normal calibration.
The calibration has now been adjusted for the differences in gravity from the original site of calibration to the new site of use.
Using the GEO code value for calibration adjustment is not as accurate as re-applying certified test weights and re-calibrating the scale in a new location.
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide C-1
C-2
Table C-1: GEO Adjustment Values
Height Above Sea Level, in Meters
Latitude North or South, in Degrees and
Minutes
0
325
325
650
650
975
975 1300 1625 1950
1300 1625 1950 2275
2275
2600
Height Above Sea Level, in Feet
2600
2925
2925
3250
3250
3575
0° 0'–5° 46'
5° 46'–9° 52'
9° 52'–12° 44'
12° 44'–15° 6'
15° 6'–17° 0'
17° 10'–19° 2'
19° 2'–20° 45'
20° 45'–22° 22'
22° 22'–23° 54'
8
9
23° 54'–25° 21' 9
25° 21'–26° 45' 10
26° 45'–28° 6' 10
7
7
8
0 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660
1060 2130 3200 4260 5330 6400 7460 8530 9600 10660 11730
5
5
6
6
4
5
5
6
4
4
5
5
3
4
4
5
3
3
4
4
2
3
3
4
2
2
3
3
1
2
2
3
1
1
2
2
0
1
1
2
0
0
1
1
6
7
7
8
8
9
9
10
6
6
7
7
8
8
9
9
5
6
6
7
7
8
8
9
5
5
6
6
7
7
8
8
4
5
5
6
6
7
7
8
4
4
5
5
6
6
7
7
3
4
4
5
5
6
6
7
3
3
4
4
5
5
6
6
2
3
3
4
4
5
5
6
2
2
3
4
5
3
4
5
28° 6'–29° 25' 11
29° 25'–30° 41' 11
30° 41'–31° 56' 12
31° 56'–33° 9' 12
33° 9'–34° 21' 13
34° 21'–35° 31' 13
35° 31'–36° 41' 14
36° 41’–37° 50’ 14
10
11
11
12
12
13
13
14
10
10
11
11
12
12
13
13
9
10
10
11
11
12
12
13
9
9
10
10
11
11
12
12
8
9
9
10
10
11
11
12
8
8
9
9
10
10
11
11
7
8
8
9
9
10
10
11
7
7
8
8
9
9
10
10
6
7
7
8
8
9
9
10
8
9
9
7
7
6
6
8
37° 50’–38° 58’ 15
38° 58’–40° 5’ 15
40° 5’–41° 12’ 16
41° 12’–42° 19’ 16
42° 19’–43° 26’ 17
43° 26’–44° 32’ 17
44° 32’–45° 38’ 18
14
15
15
16
16
17
17
14
14
15
15
16
16
17
13
14
14
15
15
16
16
13
13
14
14
15
15
16
12
13
13
14
14
15
15
12
12
13
13
14
14
15
11
12
12
13
13
14
14
11
11
12
12
13
13
14
10
11
11
12
12
13
13
10
10
11
11
12
12
13
METTLER TOLEDO IND231/IND236 User's Guide 30094756 | 04 | 04/2021
Height Above Sea Level, in Meters
Latitude North or South, in Degrees and
Minutes
0 325
325 650
650 975 1300 1625 1950 2275 2600 2925 3250
975 1300 1625 1950 2275 2600 2925 3250 3575
Height Above Sea Level, in Feet
45° 38’–46° 45’ 18
46° 45’–47° 51’ 19
47° 51’–48° 58’ 19
48° 58’–50° 6’
50° 6’–51° 13’
20
20
51° 13’–52° 22’ 21
52° 22’–53° 31’ 21
53° 31’–54° 41’ 22
54° 41’–55° 52’ 22
55° 52’–57° 4’ 23
57° 4’–58° 17’ 23
58° 17'–59° 32' 24
59° 32'–60° 49' 24
60° 49'–62° 9'
62° 9'–63° 30'
25
25
63° 30'–64° 55' 26
64° 55'–66° 24' 26
66° 24'–67° 57' 27
67° 57'–69° 35' 27
69° 5'–71° 21' 28
71° 21'–73° 16' 28
73° 16'–75° 24' 29
75° 24'–77° 52' 29
77° 52'–80° 56' 30
80° 56'–85° 45' 30
85° 45'–90° 00' 31
0 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660
1060 2130 3200 4260 5330 6400 7460 8530 9600 10660 11730
23
23
24
24
25
21
22
22
19
19
20
20
21
17
18
18
28
28
29
29
30
25
26
26
27
27
22
23
23
24
24
21
21
22
18
19
19
20
20
17
17
18
27
28
28
29
29
25
25
26
26
27
23
24
24
25
25
22
22
23
19
20
20
21
21
18
18
19
28
29
29
30
30
26
26
27
27
28
16
17
17
18
18
19
19
20
20
21
21
2\2
22
23
23
24
24
25
25
26
26
27
27
28
28
29
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
18
19
19
20
20
21
21
22
16
16
17
17
18
14
15
15
25
25
26
26
27
22
23
23
24
24
19
20
20
21
21
18
18 l9
15
16
16
17
17
14
14
15
24
25
25
26
26
22
22
23
23
24
19
19
20
20
21
17
18
18
15
15
16
16
17
13
14
14
24
24
25
25
26
21
22
22
23
23
30094756 | 04 | 04/2021 METTLER TOLEDO IND231/IND236 User's Guide C-3
To protect your METTLER TOLEDO product’s future:
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use according to these instructions and regular calibration and maintenance by our factorytrained service team ensure dependable and accurate operation, protecting your investment.
Contact us about a METTLER TOLEDO service agreement tailored to your needs and budget.
We invite you to register your product at
www.mt.com/productregistration
so we can contact you about enhancements, updates and important notifications concerning your product.
www.mt.com/IND231-IND236
For more information
Mettler-Toledo, LLC
1900 Polaris Parkway
Columbus, OH 43240
Phone 800 438 4511
Fax 614 438 4900
© 2021Mettler-Toledo, LLC
30094013 Rev. 04, 04/2021
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Key Features
- High precision weighing with a resolution of up to 1/10,000
- Multiple weighing units and modes for versatility
- Tare and zero functions for accurate measurements
- Hold and peak hold functions for capturing and displaying maximum or minimum weight
- Easy-to-use interface with a large backlit display
- Optional connectivity options for integration with external devices
- Rugged construction for use in harsh industrial environments
Related manuals
Frequently Answers and Questions
What is the maximum capacity of the IND231-IND236?
Can the IND231-IND236 be connected to a computer?
Is the IND231-IND236 suitable for use in hazardous areas?
advertisement
Table of contents
- 15 Introduction
- 15 IND231/IND236 Overview
- 15 Use in Hazardous Areas
- 15 Specifications
- 15 Model Identification
- 15 Inspection and Contents Checklist
- 15 Physical Dimensions
- 15 Main PCB
- 15 Scale Bases
- 15 Options
- 15 Display and Keyboard
- 16 Operation
- 16 Overview
- 16 Keypad Operation and Display Elements
- 16 Configurable Function Keys
- 16 Basic Functionality
- 16 Remote Display Functionality
- 16 Applications
- 17 Configuration
- 17 Accessing Setup Mode
- 17 Setup Menu Structure
- 17 Configuration Overview
- 17 F1 Scale
- 17 F2 Application
- 17 F3 Terminal
- 17 F4 Communication
- 17 F5 Maintenance
- 18 Service and Maintenance
- 18 Cleaning and Maintenance
- 18 Service
- 18 Battery Operation
- 18 Troubleshooting
- 18 Connection to InSite™ SL