Motif Duo Operation Manual

Motif Duo Double Electric Breast Pump

Operation Manual

1

Motif Duo Operation Manual

Table of Contents

INTENDED USE

PRODUCT DESCRIPTION

CONTRAINDICATIONS

1. SAFETY INFORMATION

2. PRODUCT CONFIGURATION

3. PREPARING THE PUMP & PARTS

3.1 Assembling the Tubing and Connector

3.2 Assembling Pump Parts

3.3 LCD Display

3.4 Operating Functions

3.5 Performance Functions

4. PUMPING

4.1 Preparing to Pump

4.2 Pumping

5. CLEANING THE PUMP

5.1 Washing the Pump and Parts

5.2 Sanitation

6. STORING BREAST MILK

7. TROUBLESHOOTING

8. MEANING OF SYMBOLS

9. MANUFACTURER’S DECLARATION ON EMC

10. TECHNICAL SPECIFICATIONS

PRODUCT WARRANTY

14

20

24

26

15

16

18

19

3

5

7

3

3

3

2

Motif Duo Operation Manual

Intended Use

The Motif Duo Double Electric Breast Pump is intended for use by lactating women to express and collect breast milk. The Duo can be used as a single or double pump and is intended for personal use only — sharing a breast pump can present a health hazard.

Product Description

The Motif Duo is a single user, double electric breast pump that includes customizable massage and expression modes and is capable of single and double pumping.

Contraindications

There are no known contraindications for this breast pump.

1. Safety Information

Take a moment to read through this manual and the included safety precautions before using your breast pump.

Warning Symbols

The following symbols are important for user safety. Failure to follow these instructions can lead to personal injury or damage to the breast pump. When used in conjunction with the following words, the symbols indicate:

WARNING A hazard of serious injury or death.

CAUTION A hazard of minor injury or damage to the system.

WARNING

Avoid electric shock — do not touch or operate the Motif Duo with wet hands.

Do not handle device if it falls into water. Unplug the power adapter immediately.

Only use the Motif Duo breast pump for its intended use as described in this manual.

Never use a breast pump while pregnant.

The breast pump is for personal use only. Using secondhand or rental pumps can cause cross contamination and can result in health problems.

Do not use the device while sleeping or if you are overly drowsy.

Keep the breast pump away from direct sunlight.

The Motif Duo and its accessories are not heat-resistant. Keep away from radiators, open flames, and heated surfaces.

Supervision is necessary when breast pump is in the vicinity of children.

Do not use a damaged power adapter.

Do not use this product in a high humidity environment — damage may occur.

The Motif Duo electric breast pump should not be used adjacent to or with other equipment.

Use only AC/DC Adaptors supplied by Motif Medical.

3

Motif Duo Operation Manual

Caution:

If you have been diagnosed with Hepatitis B, Hepatitis C or Human

Immunodeficiency Virus (HIV), pumping breast milk will not reduce or remove the

• risk of transmitting the virus to your baby through your breast milk.

Before each use, visually inspect the individual components of your breast pump for cracks, chips, tears, discoloration or deterioration. In the event that damage to the device is observed, please discontinue use until the parts have been replaced.

If you feel pain in your breast or nipple, turn the pump off or slide a finger between the breastshield and your breast to break the suction.

You might experience some discomfort in the early stages of pumping. If discomfort is persistent or injury or bleeding around the nipple occurs, contact a healthcare professional or Lactation Consultant immediately.

While expressing breast milk, make yourself comfortable and do not incline the bottle.

While operating the breast pump, firmly connect the diaphragm to prevent milk flow into the main body of the pump or the tubing.

Clean and sanitize all components that come into contact with breast milk before first use and after every use.

Air-dry all pump components after cleaning.

Be sure that expressed milk does not flow into the main body by checking the tubing during and after each use. If milk flows into the tubing, turn the power off, remove the milk from air tube or diaphragm, and replace or clean the tubing as instructed in this manual.

Do not fill the bottle past the recommended capacity.

Make sure that the tubing does not get tangled while double pumping.

When single pumping, cover the auxiliary tubing port with the port cover to prevent airflow into the breast pump.

Unplug the power adapter from the breast pump immediately after you are finished charging your device.

The battery charge level displayed on the LCD screen can differ from the actual battery capacity. Use your breast pump with a fully charged battery whenever possible.

Charge the battery immediately when [ Lo Bt ] is displayed on the LCD screen.

If the breast pump remains unused for a long period of time, make sure that the battery is fully charged before use.

Only authorized service personnel should repair or replace this device.

Never use a damaged device. Inspect all appropriate components before each use.

Do not use this pump outdoors, anywhere that aerosol products are being used, or where oxygen is being administered.

If pump tubing becomes moldy, discontinue use and replace the tubing before resuming pumping sessions.

Mobile and portable radio frequency communication equipment can affect the operation of this breast pump.

The breast pump, plastic bottles, and component parts may break when dropped. Use caution when transporting your device.

Do not pump while bathing or showering.

Do not thaw frozen breast milk in a microwave or in a pan of boiling water (this can

4

Motif Duo Operation Manual

• burn the breast milk or change its composition)

Thaw breast milk overnight in the refrigerator. Thawed breast milk is safe in the refrigerator for 24 hours. Do not refreeze thawed breast milk.

Quickly thaw breast milk by holding the bottle under warm running water, or place the sealed bottle in a bowl of warm water for 20 minutes to bring it to a warmer temperature.

Always use accessories supplied by Motif Medical.

Shape

Qty.

2. Product Configuration

The Motif Duo double electric breast pump utilizes a vacuum to express breast milk. The motor is classified as a medium vacuum with an intermittent suction device.

When unpacking the box, ensure that all of the following components are enclosed.

Name User manual Power adapter Main Body Nipple

Shape

Qty.

Name

4

1

Diaphragm

Top

1

Diaphragm

1

Diaphragm

Bottom

2

Tubing

Connector

Shape

Qty.

2

Name Breastshield

2

Breastshield

Connector

2

Valve

2

Tubing

Shape

Qty.

2

Name Bottle/180 ml

2

Bottle cap

2

Bottle cover

2

Bottle disc

2 2 2 2

5

Motif Duo Operation Manual

1.

LCD Display 2. Buttons

1.

Basic tubing port: Tubing port for single pumping

2.

Auxiliary tubing port: tubing port for double pumping

3.

Port cover

4.

Power adapter port

5.

Port cover stabilizer

6

Motif Duo Operation Manual

3. Preparing the Pump & Parts

3.1 Assembling the Tubing and Connector

Insert the metal part of the tubing connector into the end of tubing as shown.

3.2 Assembling Pump Parts

1.

Place the diaphragm so the cross-lined surface faces the bottom. Cover the diaphragm top and turn it to lock.

2.

Place the breastshield into the breastshield connector.

3.

Place the valve on the bottom of the breastshield connector. Screw the bottle into the breastshield connector.

4.

Connect the diaphragm bottom to the breastshield connector.

5.

Connect the tubing to the port on main body of the pump.

6.

Connect the tubing to the bottle block as shown in the picture.

7.

To clean or store after use, disassemble the bottle in the reverse order.

7

Motif Duo Operation Manual

5

CAUTION

6 l l

When operating the pump, connect diaphragm tightly to prevent the expressed milk flow from entering the main body.

When single pumping, cover the auxiliary tubing port with port cover to prevent airflow into the breast pump.

3.3 LCD Display

No.

Name

Memory

Expression mode

Adapter connection

Status Description

Off No stored operating sequence.

Flashing Storing operating sequence.

On

On

There is a stored operating sequence.

Operating in expression mode.

Flashing Operating in stored expression mode sequence.

Off Power adapter is not connected.

On Power adapter is connected.

8

Motif Duo Operation Manual

Battery

Speed/Vacuum level

Massage mode

Operating time

Percent

Charge

On

Fully charged

75 - 100%

50 - 75%

25 - 50%

0 - 25%

0%

Charge battery when symbol is flashing. The device will shut down automatically when battery dies.

Indicates operating time since device has been powered on. After 30 minutes, the main body will turn off automatically.

When the battery needs to be charged, [ Lo bt ] will be displayed.

Speed/vacuum level display On

On Operating in massage mode

Flashing Operating in stored massage mode sequence.

Speed/

Vacuum level

On Speed/vacuum level displayed in gradation.

9

Motif Duo Operation Manual

Button

Application method

Basic function

Press for less than 0.5 seconds.

Power On/Off button

Turns power on and off.

Frequency function

Press from 0.5 to 1.5 seconds.

-

-

Frequency up

Increase the frequency one step.

Frequency down

Memory function

Press longer than 2 seconds until the function begins to operate.

Memory start/stop button

Press and hold this button for more than two seconds after turning the pump on to start the memory function operating sequence.

Operate the pump as you want to save and press and hold again to save the whole sequence operated.

When the memory function is activated, the memory icon will flash in the LCD display. When the operating sequence is saved, the memory icon will be turned on.

The saved operating sequence will not be erased upon power off.

Stop button

Stop to operate on saved sequence.

No response when not operating according to saved sequence.

Start button

Speed/vacuum increase button

Increase the speed/vacuum in the operation mode.

Speed/vacuum can be changed from 1 to 10.

Speed/vacuum decrease button

10

Motif Duo Operation Manual

Decrease the speed/vacuum in the operating mode.

Mode change button

Toggle massage between and expression modes.

Decrease the frequency one step.

-

-

Start to operate on saved sequence.

No response when no sequence saved.

Delete button

Delete the saved sequence.

Memory icon will be off.

After deleting the saved sequence, the pump will operate at your last set-up.

No response when no sequence saved.

3.4 Operating Functions

The Motif Duo has three operating functions: basic massage/express function, frequency change function within the massage level and memory function to save the operating sequence. The memory function, massage/expression mode, level of speed/vacuum strength, and massage/expression time can be saved and restored at any time.

Buttons

Each button has three functions. Pressing the button for less than 0.5 seconds will perform basic functions, pressing for a moderate duration of time (from 0.5 to 1.5 seconds) will perform the frequency function in expression mode, and pressing the button for longer than 2 seconds will perform memory functions.

Using the Memory Function

You can save and restore your personal settings for massage/expression modes, speed/vacuum levels as well as your optimal pumping session time.

1.

Determine the operating sequence that you want to save (massage/expression sequence, speed/vacuum level, operating time)

2.

Turn on the pump by pressing the power on/off button then press and hold for memory session.

3.

Operate the pump at the sequence you want to save.

11

Motif Duo Operation Manual

4.

After selecting the sequence that you want to save, press and hold the power on/off button for more than two seconds to save the sequence. When saved, the memory icon will turn on.

5.

To start the saved operation sequence, press and hold the play button for more than 2 seconds while the device is on. The massage mode icon or the expression mode icon will be flashing.

6.

To stop the saved operation sequence, press and hold the stop button for more than 2 seconds. The pump will operate on your last session setting and the massage mode icon or the expression mode icon will be on.

Example of Memory Function

If you want to store the operation sequence of the following:

Operating

Mode

Massage

Level 4

Massage

Level 5

Express

Level 3

Express

Level 4

Operating time

1 minute 30 seconds 1 minute Until pumping session completed

1.

Check the LCD display to see if there is an existing stored operating sequence.

Icon is highlighted when there is a stored operating sequence.

2.

If there is a stored sequence, delete it by pressing button more than 2 seconds.

The icon will disappear.

3.

Set the massage mode on level 4 and expression mode on level 3.

4.

Change the operating mode to massage mode level 4.

5.

To start the memory function, press the button for more than 2 seconds. The

icon will begin to flash to indicate that the memory function has stored the operation sequence.

6.

Operate the pump for 1 minute at massage mode level 4. Switch to level 5 by pressing button shortly and operating for another 30 seconds.

7.

Change to expression mode level 3 by pressing the button shortly.

8.

Operate the pump for 1 minute at level 3 and press button shortly to change to level 4.

12

Motif Duo Operation Manual

9.

Press the button to save the operating sequence or turn off the breast pump.

The pump will store the operating sequence. If you turn on the breast pump, the

will appear, which indicates that the operating sequence is stored. The operating mode stored last will operate until terminated by user or the operating time reaches 30 minutes.

10.

To check the stored sequence, press the button until the operation starts. The pump should start with the stored massage level.

Frequency Function

This function operates in the expression mode.

• On this setting, the frequency settings can be changed without altering the pressure level.

• With the frequency function you can change the mode (expression or massage) by pressing the mode button. Then press the + or – buttons to adjust frequency. The display will show [CY: --] to indicate that you are in frequency mode.

3 frequency levels can be changed for each expression level. Select the level that you want by pressing or for 0.5 to 1.5 seconds in the expression mode. The selected frequency value will be displayed by 1, 2 or 3 for 4 seconds and will then return to the operating time display.

• Note that increasing the frequency might be accompanied by increased noise and vibration.

Using the Frequency Function

1. Turn on the pump by pressing the power on/off button

2. Press the mode button, then will be displayed to indicate that you are in expression mode. Set your preferred level.

3. Press the mode button once again to get to frequency. The time indicator will be changed to [CY: --]. This indicates that you are in frequency mode.

4. There are 3 different frequencies for each level of expression. You can change the frequency level by pressing or .

3.5 Essential Performance Functions

The essential performance of the Motif Duo limits its vacuum pressure lower than 250 mmHg. l

WARNING

Only use the Motif Duo for its intended use as described in this manual.

13

Motif Duo Operation Manual

4. Pumping

4.1 Preparing to Pump

1.

2.

3.

4.

5.

6.

Wash hands thoroughly with soap and water before touching the breast and pump parts.

Check pump components for wear or damage before use and replace parts when/if necessary.

Make sure that all components that come into contact with breast milk are cleaned and sanitized appropriately. Assemble the breastshield, diaphragm and bottle according to this manual and connect to the pump body via the tubing.

Ensure tubing is connected tightly to prevent air leakage.

Check battery capacity before use to ensure the device is fully charged.

Massage the breast for about 5 minutes with a warm towel to promote expression.

4.2 Pumping

To prevent backflow in tubing, always use the breast pump while seated.

1.

Place the breastshield on your breast so that your nipple is centered within breastshield.

2.

Turn the breast pump on to begin pumping.

3.

Adjust the speed/vacuum settings to a comfortable level.

4.

Normal pumping time ranges from 15 minutes to 30 minutes; this can vary based on the individual.

5.

When your pumping session is over, turn off the power and slide a finger between the breastshield and your breast to separate the breastshield from your breast.

CAUTION

Never use while pregnant.

A breast pump is a single user device. Use by more than one person may present a health risk.

If you have been diagnosed with Hepatitis B, Hepatitis C, or Human

Immunodeficiency Virus (HIV), pumping breast milk will not reduce or remove the risk of transmitting the virus to your baby through your breast milk.

Before each use visually inspect the individual components for cracks, chips, tears, discoloration, or deterioration. In the event that damage to the device is

14

Motif Duo Operation Manual

• observed, please discontinue use until the parts have been replaced.

If you feel pain in your breast or nipple, turn the pump off or slide a finger between the breastshield and your breast to break the suction.

You can experience some inconvenience in the early stage of usage. If the inconvenience is persistent or injury/ bleeding around the nipple appears, contact a healthcare professional immediately.

5. Cleaning the Pump

Before using the pump, wash the breast pump and all parts thoroughly. Clean the main body of the pump with a clean, dry towel after each use.

5.1 Washing Pump & Parts

1.

2.

3.

4.

5.

6.

Disassemble the bottle and parts.

Rinse all parts that came in contact with the breast and breast milk in cool water to remove breast milk residue.

Wash parts separately in sink or soak all parts in warm, soapy water for 5 minutes.

Wash each part with a clean dishcloth or a soft brush.

Dishwasher — w ash all separated parts on top rack of dishwasher only.

Place parts on a clean surface and/or towel and allow all parts to air dry.

Store dry parts in a clean, cool place when not in use. Do not store wet or damp parts.

CAUTION

Disassemble and wash all parts that come in contact with the breast and breast milk immediately after use to avoid milk residue and to prevent the growth of bacteria.

Only use drinking-quality water for washing pump parts.

Main body

Wipe with a clean, dry towel after each pumping session.

15

Motif Duo Operation Manual

Tubing

Do not clean tubing: replace damaged or dirty tubing with a new set.

5.2 Sanitation

1.

Clean all parts before using the pump for the first time.

2.

Clean the device (main body) after each use with a dry towel.

Sanitizing Pump Parts

1.

2.

3.

4.

5.

6.

7.

Disassemble all parts that come in contact with breast milk.

Fill a pot with enough water to cover all parts.

Boil the water for 5 minutes with parts in the water.

Allow water to cool and gently remove parts from water with tongs.

Place parts on a clean surface and/or towel.

Allow all parts to air dry.

Store dry parts in a clean, cool place when not in use. Do not store wet or damp parts.

6. Storing Breast Milk

The information below is referenced from the Office of Women’s Health.

Store your breast milk in clean glass or hard BPA-free plastic bottles with tight-fitting lids. You can also use milk storage bags, which are made for freezing human milk. Do not use disposable bottle liners or other plastic bags to store breast milk.

After Each Pumping Session

• Label the date on the milk storage container. Include your child’s name if you are giving the milk to a childcare provider.

• Gently swirl the containers to mix the cream part of the breast milk that may rise to the top back into the rest of the milk. Do not shake the milk. This can cause some of the milk nutrients to break down.

• Refrigerate or chill breast milk immediately after expression. You can refrigerate breast milk in a cooler or insulated cooler pack, or freeze the milk in small (2 to 4 ounce/60 to 110 g) batches for later feeding.

16

Motif Duo Operation Manual

Freezing Breast Milk

Wait to tighten bottle caps or lids until milk is completely frozen.

Try to leave an inch or so from the milk to the top of the container, as it will expand when freezing.

Store milk in the back of the freezer, away from freezer door.

Thawing and Warming Milk

• Clearly label milk containers with the date the milk was expressed. Use oldest stored milk first.

• Thaw the bottle or bag of frozen milk by putting it in the refrigerator overnight, holding it under warm running water, or by placing it in a container of warm water for several minutes.

• Never put a bottle or bag of breast milk in the microwave. Microwaving the milk creates hot spots that could burn or damage the milk.

• Swirl the milk, and test the temperature by dropping some on your wrist. Milk should be comfortably warm — not hot.

Use thawed breast milk within 24 hours. Do not refreeze thawed breast milk.

Storing Fresh Breast Milk

Storage

Method

Temperature Longevity Tips

Countertop or table

Room temperature

Up to 77°F

(25° ℃ )

Up to 3 to 4 hours is best.

Up to 6 to 8 hours is okay for very clean expressed milk.

Containers should be covered and kept as cool as possible. Discard leftover milk within 1 to 2 hours after baby is finished feeding.

Refrigerator

39°F (4°C) or colder

Up to 3 days is best.

Up to 5 days is okay for very clean expressed milk.

Store milk in the back of the main body of the refrigerator. While at work, you can place your expressed milk in the refrigerator. Use an insulated bag and place the milk at the back of refrigerator.

17

Motif Duo Operation Manual

Freezer

0° ℉ (-18° ℃ ) or colder

Up to 3-6 months is best.

Up to 9 months is okay for very clean expressed milk.

Store milk toward the back of the freezer where the temperature is most constant.

Deep freezer

-4° ℉ (-20° ℃ ) or colder

Up to 6 months.

Up to 12 months is okay for very clean expressed milk.

Store milk toward the back of the freezer where the temperature is most constant. Milk stored at 0 degrees or colder is safe for longer durations but the milk quality might not be as high as when first expressed.

Storing Thawed Breast Milk

Thawed breast milk

Room temperature

(60°F to 85°F/16°Cto 29°C)

1 to 2 hours is best. Up to 4 hours is okay.

Refrigerator

(39°F/4°C or colder)

24 hours

Freeze

Options

Do not refreeze.

7. Troubleshooting

No or low expression

Check the power button.

Ensure the breastshield, diaphragm assembly, and bottles are assembled correctly.

Inspect all components for tears, cracks, or bending.

Battery not charged

Does not power on

Ensure that the power adapter is connected correctly.

If your pump will not hold a charge, replace the adapter with new one.

Inspect the power on/off button and exterior of pump for any foreign materials.

Main body becomes wet

Unplug the power adapter from the socket.

Turn off the main body.

Wipe main body with dry cloth and store the main body in a warm, dry place for more than 12 hours.

18

Motif Duo Operation Manual

Breast milk in tubing

Turn off the main body.

Disconnect the tubing from diaphragm top.

Remove the milk from tubing and diaphragm.

If you are having trouble with your breast pump or have further questions, please contact the

Motif Medical customer center.

8. Meaning of Symbols

The following symbols are placed on the product, label, packaging and this manual for ease of use.

Symbols on product label and packaging:

Manufacturer

Date of manufacture

Serial number of the device

Refer to user manual

Type BF applied part

Keep the device dry

Device is fragile

Temperature limitation for operation, transport and storage

Humidity limitation for operation, transport and storage

Atmospheric pressure limitation for operation, transport and storage

IP21

Keep the device away from sunlight

Degree of protection against ingress of water or solid particles

Do not dispose the device together with unsorted municipal waste (for EU only). The solid bar symbol indicates that the mains adapter was put on the market after 13 August 2005.

This symbol indicates unique GS1 Global Trade Item Number (GTIN).

(01) Barcode (11) Date of manufacture (21) Serial No.

This symbol indicates the compliance with the essential requirements and provisions of the Medical Device Directive 93/42/EEC as amended by

19

Motif Duo Operation Manual

2007/47/EEC.

Symbols on the adaptor:

Power adaptor is a class II device.

Power adaptor is for indoor use only.

This symbol indicates direct current.

This symbol indicates alternating current.

This symbol indicates compliance with Canadian and U.S. component requirements.

(Recognized Component Mark for Canada and the United States)

9. Manufacturer’s Declaration on EMC

The Motif Duo requires special precautions regarding EMC (Electromagnetic compatibility) and should be used in accordance with the EMC information provided in this user manual. Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, and walkie-talkies can affect the functionality of the device. Wireless equipment of this type should be kept at least 1 m away from the equipment.

The set pressure value and operating mode should be unchanged during EMC testing. It is the essential performance of the pump and is checked during EMC testing.

Electromagnetic Emissions

The Motif Duo is intended for use in the electromagnetic environment specified below.

Emissions test Compliance Electromagnetic environment-guidance

RF emissions

CISPR 11

Group 1

This electric breast pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause interference in nearby electronic equipment.

RF emissions

CISPR 11

Harmonic emission

IEC61000-3-2

Class B

Class A

The electric breast pump is suitable for use in all establishments, including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

20

Motif Duo Operation Manual

Voltage fluctuations

/flicker emissions

IEC61000-3-3

Complies

WARNING

The Motif Duo breast pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the breast pump should be observed to verify normal operation in the configuration where it will be used.

Recommended separation distances between portable and mobile RF communications equipment and the electric breast pump

The Motif Duo is intended for use in the electromagnetic environment in which radiated

RF disturbances are controlled. The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitte r [W]

Separation distance according to frequency of transmitter [M]

150 ㎑ ~ 80 ㎒ 80 ㎒ ~ 800 ㎒ 800 ㎒ ~ 2.5 ㎓

0.01

0.1

1

10

100 d = 1.2

0.12

0.38

1.2

3.8

12 d = 1.2

0.12

0.38

1.2

3.8

12 d = 2.3

0.23

0.73

2.3

7.3

23

For transmitters at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1) At 80 ㎒ and 800 ㎒ , the separation distance for the higher frequency range applies.

NOTE2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Electromagnetic Immunity

21

Motif Duo Operation Manual

The electric breast pump is intended for use in the electromagnetic environment specified below. The user of the electric breast pump should ensure that it is used in such an environment.

Immunity test Test level

Compliance level

Electromagnetic

Environment-guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

±6 ㎸

Contact

±8 ㎸ air

±6 ㎸

Contact

±8 ㎸ air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrostatic discharge (ESD)

FDA home use guidance

±8 ㎸

Contact

±15 ㎸ air

Electrical fast transient/burst

IEC61000-4-4

Surge

IEC61000-4-5

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC61000-4-11

±2 ㎸ for power supply lines

±1 ㎸ for input/output lines

±1 ㎸ line(s) to line(s)

±2 ㎸ line(s) to earth

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycle, 6 cycle

70% U

T

(30% dip in

U

T

) for 25 cycle,

30 cycle

<5% U

T

(<95% dip in

U

T

)

±8 ㎸

Contact

±15 ㎸ air

±2 ㎸ for power supply lines

±1 ㎸ for input/output lines

±1 ㎸ line(s) to line(s)

±2 ㎸ line(s) to earth

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycle, 6 cycle

70% U

T

(30% dip in

U

T

) for 25 cycle,

30 cycle

<5% U

T

(<95% dip in

U

T

)

Mains power quality should be that of a typical commercial or domestic environment.

Mains power quality should be that of a typical commercial or domestic environment.

Mains power quality should be that of a typical commercial or domestic environment.

If the user of the electric breast pump requires continued operation during power mains interruptions, it is recommended that the electric breast pump be powered from an uninterruptible power supply or a battery.

22

Motif Duo Operation Manual for 5s for 5s

Power frequency

(50/60 ㎐ ) magnetic field

IEC61000-4-8

Power frequency

(50/60 ㎐ ) magnetic field

FDA home use guidance

3 A/m

30 A/m

3 A/m

30 A/m

Power frequency magnetic fields should be at levels characteristic of a typical commercial or domestic environment.

U

T

is the a.c. mains voltage prior to application of the test level.

Immunity Test

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

Test Level

150

80 rms

3 V/m

~ 80

~ 2.5

Compliance

Level

3 V rms

3 V/m

Electromagnetic environmentguidance

Portable and mobile RF communications equipment, including cables, should be used no closer to any part of the electric breast pump than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance: d = 1.2 d = 1.2 80 ㎒ ~ 800 ㎒

23

Motif Duo Operation Manual

Conducted RF

FDA home use guidance

Radiated RF

FDA home use guidance

6 V rms

150

80

10 V/m

~ 80

~ 2.6

6 V rms

10 V/m d = 2.3 800 ㎒ ~ 2.5 ㎓

Where p is the maximum output power rating of the transmitter (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m)

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a , should be less than the compliance level in each frequency range b .

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1) At 80 ㎒ and 800 ㎒ , the higher frequency range applies.

NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the electric breast pump is used exceeds the applicable RF compliance level above, the electric breast pump should be observed to verify normal operation.

If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the pump. b Over the frequency range 150 ㎑ to 80 ㎒ , field strengths should be less than 3 v/m.

10. Technical Specifications

24

Motif Duo Operation Manual

Functional Characteristics

Type

Electric Breast Pump

Display

Type

Pressure Level

Operating Time

Battery Level indication

Pumping Pressure

Vacuum

Function

Mode 3" Mono LCD

Numeric & 10 levels

Numeric Display (min:sec) Auto Shut Off

3 levels Memory

Backflow Prevention

32 - 250 mmHg

Expression 10 levels

Massage 10 levels

30 min

- 60 operations

Power

Adaptor Input: 100-240Vac,

50-60Hz, 0.5A

Output: DC +12V,

1.5A (MAX)

Rechargeable

Battery

7.4V Li-ion Polymer,

900mAh

Operating Time: 150min

Standard Configuration

Charging Time: 90min

Operation Manual 1ea

Breastshield &

Bottle

1 set

Power Adaptor 1ea

(Standard breastshield

24mm)

Options

Carrying bag

Breastshield 21mm / 27mm

Breastshield & Bottle

Set

(for double pumping)

Warranty

Main Unit

1 year for motor, 90 days for parts

Physical Characteristics

Dimension Weight

Main Unit

Packing

Carton Box (10ea)

95(W)x49(D)x132(H)mm Main Unit

250(W)x 55(D)x180(H)mm Packing

795(W)x270(D)x400(H)mm

Carton Box

(10ea)

Environmental Conditions

Operating

Temperature

5 to 40°C (41-104°F)

Operating Humidity 15-95% non-condensing

Storage

Temperature

260g

0.9Kg

10Kg

–25 to 70°C (–13-158°F)

Storage Humidity 15-95% non-condensing

Certificates

25

Motif Duo Operation Manual

Certificates

Standard

CE0434 Classifications

EN60601-1, EN60601-1-2 Waterproof

Disposing of the Pump

Class II, Type BF

IP21

In order to comply with EU Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE): This product may contain material that could be hazardous to human health and the environment. DO

NOT DISPOSE of this product as unsorted municipal waste. This product needs to be RECYCLED in accordance with local regulations, contact your local authorities for more information. This product may be returnable to your distributor for recycling - contact the distributor for details.

Product Warranty

Product Name Motif Duo Double Electric Breast Pump

Brand name Motif Medical

Model Name Duo

Serial No.

Warranty Period 1 year for motor, 90 days for parts

Date of Purchase

Manufacturer Motif Medical

Thank you for purchasing the Motif Duo Double Electric Breast Pump. This product has been passed through strict quality control and inspection.

Motif Duo Double Electric Breast Pump

Distributed by Motif Medical LLC.

Motif Medical

8 The Green

Suite #5601

Dover, DE 19901 www.motifmedical.com

[email protected]

(844)-272-8390

26

advertisement

Related manuals

Frequently Answers and Questions

What is the intended use of the Motif Medical Duo?
The Motif Medical Duo is a double electric breast pump used by lactating women to express and collect breast milk.
Can I use the Motif Medical Duo if I am pregnant?
No, the Motif Medical Duo should not be used during pregnancy.
How do I assemble the tubing and connector for the Motif Medical Duo?
Insert the metal part of the tubing connector into the end of the tubing.
How do I assemble the pump parts for the Motif Medical Duo?
Place the diaphragm so the cross-lined surface faces the bottom. Cover the diaphragm top and turn it to lock.
How do I clean and sanitize the Motif Medical Duo?
Clean and sanitize all components that come into contact with breast milk before first use and after every use.
How do I store breast milk expressed with the Motif Medical Duo?
Store expressed milk in the refrigerator for up to 24 hours or in the freezer for up to 6 months.
What should I do if the pump tubing becomes moldy?
Discontinue use of the pump and replace the tubing before resuming pumping sessions.
What are the safety precautions I should take when using the Motif Medical Duo?
Avoid electric shock by not touching or operating the pump with wet hands. Do not use the pump while sleeping or overly drowsy.