Keysight 8753E and 8753ES Option 011 Network Analyzers Option 611 6 GHz Upgrade Kit Installation Guide
Keysight 8753E and 8753ES Option 011 Network Analyzers Option 611 6 GHz Upgrade Kit provides optional 6 GHz operation for the network analyzer. The upgrade kit is compatible with specific models of Keysight network analyzers and is designed to enhance their capabilities. With this upgrade, users can extend the frequency range of their network analyzers and perform measurements up to 6 GHz, enabling them to analyze higher frequency signals and components. This can be particularly useful in applications such as designing and testing microwave circuits, antennas, and other high-frequency devices.
Publicitate
Publicitate
About this Manual
We’ve added this manual to the Agilent website in an effort to help you support your product. This manual is the best copy we could find; it may be incomplete or contain dated information. If we find a more recent copy in the future, we will add it to the Agilent website.
Support for Your Product
Agilent no longer sells or supports this product. Our service centers may be able to perform calibration if no repair parts are needed, but no other support from
Agilent is available. You will find any other available product information on the
Agilent Test & Measurement website, www.tm.agilent.com
.
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that
Hewlett-Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. We have made no changes to this manual copy. In other documentation, to reduce potential confusion, the only change to product numbers and names has been in the company name prefix: where a product number/name was HP XXXX the current name/number is now Agilent XXXX. For example, model number
HP8648A is now model number Agilent 8648A.
Installation Note
Option 611 6 GHz Operation Upgrade Kit
For HP 8753E Option 011 and HP 8753ES Option 011 Network
Analyzers
Network Analyzer Model
Number and Options
HP 8753E Option 011
HP 8753ES Option 011
Applicable Upgrade Kit
Model Number
HP 8753EU Option 611
HP 8753ESU Option 611
HP Part Number 08753-90418 Supersedes March 1998
Printed in USA December 1999
Notice.
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
© Copyright Hewlett-Packard Company 1998, 1999
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
6 GHz Operation Upgrade Kit
Products Affected: . . . . . . . . . . . . . . . . . . . . . . . . .
HP 8753E Option 011 Network Analyzer
HP 8753ES Option 011 Network Analzyer
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . .
Personnel Qualified by HP
Estimated Installation Time: . . . . . . . . . . . . . . . .
45 minutes
Estimated Verification Time: . . . . . . . . . . . . . . . . .
2.0 hours (includes instrument warmup time)
Purpose
This Option 611 upgrade kit provides the Option 011 network analyzer with optional 6 GHz operation. The upgrade kit is keyed to the serial number of an individual network analyzer and cannot be used to upgrade other instruments.
Items Included in the Kit
Table 1 describes the parts included in this upgrade kit. Check the contents of this kit against
Table 1.
Table 1
Quantity
1
1
1
1
1
1
1
1
Option 611 Upgrade Kit Contents
Description
6 GHz source assembly
Coax cable assembly
Keyword label (affixed to
Keyword label (to affix to instrument)
Installation note
6 GHz nameplate
6 GHz nameplate
Envelope
HP Part Number
08753-60230
8120-5026
9320-6215
9320-5805
08753-90418
08753-80157
(HP 8753E Option 011 only)
08753-80205
(HP 8753ES Option 011 only)
9230-0023
Installation Note 08753-90418 3
Equipment and Tools Required
Table 2 Required Equipment and Tools
Item
Power meter
Power sensor
Power sensor
Power splitter, 2 way 50
Ω (2)
Frequency counter
Cable set, RF 24-inch (2)
Cable set, RF 50
Ω⋅
Type-N (2)
Cable, BNC-to-BNC
Adapter, APC 3.5 mm (f) to Type-N (f)
Adapter, Type-N (f) to BNC (m)
Adapter, Type-N (m) to Type-N (m)
Adapter, Type-N (f) to Type-N (f)
Adapter, Type-N (m) to SMA (f)
Adapter, Type-N (f) to SMA (m)
Adapter, BNC Alligator clip
HP-IB cable assembly
Low-pass filter
Attenuator, 10 dB (Option 002 only)
Attenuator, 20 dB (2)
Attenuator, 30 dB (Option 002 only)
3.5-inch floppy disk
Small needle-nose pliers
T-10 TORX screwdriver
T-15 TORX screwdriver
5/16 inch open-end wrench
5/16 inch torque wrench (10 in-lbs)
Electrostatic discharge (ESD) grounding wrist strap and mat
HP 8753E Option 011 Network Analyzer Service Guide
(required for upgrades to 8753E Option 011 models only)
HP 8753ES Option 011 Network Analyzer Service Guide
(required for upgrades to 8753ES Option 011 models only)
08753-90404
08753-90485
HP Part or Model Number
HP 436A/437B/438A or
HP E4418/E4419B
HP 8482A
HP 8481A
HP 11667A Option 001
HP 5350B
HP 11500B
HP 11851B
8120-1840
1250-1745
1250-0077
1250-1475
1250-1472
1250-1250
1250-1562
8120-1292
HP 10833A
9135-0198
HP 8491A Option 010
HP 8491A Option 020
HP 8491A Option 030
HP 92192A (box of 10)
4 Installation Note 08753-90418
Safety Considerations
WARNING Before you disassemble the instrument, turn the power switch OFF and unplug the instrument. Failure to unplug the instrument can result in personal injury.
CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components.
Perform these procedures only at a static-safe workstation and wear a grounding strap. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
Conventions
This installation note uses the following conventions for front-panel keys and softkeys.
Front-Panel Key represents a key physically located on the instrument.
SOFTKEY represents a “softkey,” a key whose label is determined by the instrument’s firmware.
Installation Note 08753-90418 5
Installation Procedure for the Option 611 Upgrade Kit
The Option 011 Network Analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation.
NOTE The HP 8753E Option 011 network analyzer is shown in the illustrations of this installation procedure. If you have an HP 8753ES Option 011 network anlayzer, the details of these illustrations may vary slightly. However, these slight differences will not affect the procedures of this installation.
Verify the Serial Number
Refer to the keyword label in the box below. First, verify that the analyzer’s serial number matches the serial number on the keyword label. Second, verify that the option number on the keyword label matches the option number for this installation note. If in either step, the serial number or the option number does not match the expected results, the keyword will not enable the option. If this is the case, refer to
Table 3, “Hewlett-Packard Sales and Service
Offices,” on page 14 and contact the nearest sales or service office for assistance.
Keyword Label
6 Installation Note 08753-90418
Remove the Top Cover
Refer to
1. Disconnect the power cord.
2. Remove both of the upper rear feet (item 1) by loosening the TORX T-10 screws (item 2).
3. Loosen the top cover screw (item 3).
4. Slide the cover off towards the rear of the analyzer.
Figure 1 Cover and Feet Removal
Installation Note 08753-90418 7
Remove the 3 GHz Source Assembly
Refer to
5. Remove the source bracket (item 1) by removing four screws.
NOTE It might be necessary to disconnect the flexible cable from A6 (the sampler next to the source assembly). To disconnect the cable, use a small needle-nose pliers to grasp the metal connector of the cable and carefully loosen it from its mating connector on A6. See item 7 of
6. Disconnect the semirigid cable W1.
7. Lift the two retention clips (item 2) at the sides of the source assembly to an upright position.
8. Lift the A3 source assembly using the source bracket handle. See item 3 of
9. Remove the source assembly from the instrument.
Figure 2 A3 Source Assembly
8 Installation Note 08753-90418
Installing the 6 GHz Source Assembly
Refer to
10.Check the connector pins to the motherboard before reinstallation.
11.Slide the edges of the sheet metal partition (item 4) of the 6 GHz source assembly into the guides at the front and back of the source compartment.
12.Press down on the assembly to ensure that it is well seated in the motherboard connector.
13.Push down the retention clips. Reconnect the semirigid cable, W1, to the source assembly.
NOTE When connecting semirigid cables, it is recommended that the connections be torqued to 10 in-lb.
Figure 3 A3 Source Assembly Removal
Installation Note 08753-90418 9
Cable Installation
Refer to
14.Remove the two screws from the air flow cover and remove the cover.
15.Remove the screw from the PC board stabilizer and remove the stabilizer.
16.Lift the two extractors located at each end of the A12 board. Lift the board from the card cage slot just enough to connect the flexible cable, provided with the kit (HP part number
8120-5026), to the second connector from the front A12J4 (item 5).
17.Slide the board back into position and route the cable along the board towards the rear of the analyzer and through the notch on the end of the card cage.
18.Route the flexible cable as shown in
and connect it to its connector P2 (item 6) on the board of the source assembly.
19.Lower the two extractors, replace the PC board stabilizer, and secure it with the screw.
Figure 4 Cable Connections
10 Installation Note 08753-90418
20.Replace the air flow cover and secure with the two screws.
21.Reinstall the source bracket, item 1 of
22.Reconnect the flexible cable to the A6 sampler.
Configure the Network Analyzer for 6 GHz Operation
Refer to
23.Remove the instrument bottom cover.
a. Remove both lower rear feet (item 4) by loosening the TORX T-10 screws (item 5).
b. Loosen the bottom cover screw (item 6).
c. Slide the cover off towards the rear of the instrument.
Refer to
24.Locate the A9 switch on the A9 CPU assembly at location S400.
25.Set the A9 switch to Alter Mode.
Figure 5 A9 Switch Location
Refer to
26.Install the instrument top and bottom covers.
27.Reconnect the analyzer power cord and switch on the power.
28.Press
Preset Display MORE TITLE ERASE TITLE to erase the HP logo.
Installation Note 08753-90418 11
29.Enter the keyword from the label located on
page 6 . Use a keyboard attached to the
analyzer or rotate the front panel knob to position the arrow below each character of the keyword. Press SELECT LETTER to enter each character. When all characters are entered, press DONE .
NOTE Be sure to use upper-case letters when entering via a keyboard.
30.Press
System SERVICE MENU TESTS 56 x1
EXECUTE TEST YES .
. At the prompt, press
The analyzer will display Option Cor DONE when the test is complete.
31.Press
RETURN TESTS 44 x1 . At the prompt, press EXECUTE TEST YES .
The analyzer will display *Source Def DONE when the test is complete.
32.Press
RETURN TESTS 45 x1 . At the prompt, press EXECUTE TEST YES .
The analyzer will display *Pretune Def DONE when the test is complete.
33.Press
RETURN TESTS 46 x1 . At the prompt, press EXECUTE TEST YES .
The analyzer will display ABUS Cor DONE when the test is complete.
34.Press
RETURN TESTS 48 x1 . At the prompt, press EXECUTE TEST YES .
The analyzer will display Pretune Cor DONE when the test is complete.
35.Refer to "Adjustments and Correction Constants" in the network analyzer’s service guide.
After the required warmup times, perform the following adjustments:
• RF Output Power Correction Constants
• Sampler Magnitude and Phase Correction Constants
• Cavity Oscillator Frequency Correction Constants
• Source Spur Avoidance Tracking Adjustment
• EEPROM Backup Disk Procedure
36.Refer to "System Verification and Performance Tests" in the network analyzer’s service guide and perform the following system tests:
• Source Frequency Range and Acuracy
• Source Power Range, Linearity and Accuracy
• Source and Receiver Harmonics (required only for analyzers with Option 002)
37.Switch off the analyzer and disconnect the power cord.
Refer to
38.Remove the instrument top and bottom covers.
Refer to
39.Set the A9 switch to Normal Mode.
12 Installation Note 08753-90418
Affix the Keyword Label
40.Locate the keyword label in the small envelope and remove it from its backing. Affix the label onto the location shown in
Figure 6 Location of Keyword Label, Top View of Analyzer
Reassemble the Instrument
Refer to
41.Replace the instrument top and bottom covers by reversing the order of removal.
Replace the Analyzer's Nameplate
42.Use a sharp knife to carefully peel away the old nameplate located on the analyzer’s front panel.
43.Adhere the new nameplate, which is included in the upgrade kit.
Installation Note 08753-90418 13
Table 3 Hewlett-Packard Sales and Service Offices
UNITED STATES
Instrument Support Center
Hewlett-Packard Company
(800) 403-0801
Headquarters
Hewlett-Packard S.A.
150, Route du Nant-d’Avril
1217 Meyrin 2/ Geneva
Switzerland
(41 22) 780.8111
EUROPEAN FIELD OPERATIONS
France
Hewlett-Packard France
1 Avenue Du Canada
Zone D’Activite De Courtaboeuf
F-91947 Les Ulis Cedex
France
(33 1) 69 82 60 60
Germany
Hewlett-Packard GmbH
Hewlett-Packard Strasse
61352 Bad Homburg v.d.H
Germany
(49 6172) 16-0
Great Britain
Hewlett-Packard Ltd.
Eskdale Road, Winnersh
Triangle Wokingham,
Berkshire RG41 5DZ England
(44 118) 9696622
Headquarters
Hewlett-Packard Company
3495 Deer Creek Rd.
Palo Alto, CA 94304-1316
USA
(415) 857-5027
Japan
Hewlett-Packard Japan, Ltd.
Measurement Assistance
Center
9-1, Takakura-Cho, Hachioji-
Shi,
Tokyo 192-8510, Japan
TEL (81) -426-56-7832
FAX (81) -426-56-7840
China
China Hewlett-Packard Co.
38 Bei San Huan X1 Road
Shuang Yu Shu
Hai Dian District
Beijing, China
(86 1) 256-6888
INTERCON FIELD OPERATIONS
Australia
Hewlett-Packard Australia Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
(61 3) 895-2895
Canada
Hewlett-Packard (Canada) Ltd.
17500 South Service Road Trans-
Canada Highway
Kirkland, Quebec H9J 2X8
Canada
(514) 697-4232
Singapore
Hewlett-Packard Singapore (Pte.)
Ltd.
150 Beach Road
#29-00 Gateway West
Singapore 0718
(65) 291-9088
Taiwan
Hewlett-Packard Taiwan
8th Floor, H-P Building
337 Fu Hsing North Road
Taipei, Taiwan
(886 2) 712-0404
14 Installation Note 08753-90418
Descarcă
Publicitate