Helmer PC100i, PC2200h, PC2200i, PC3200i, PC4200i, PC900h Service Manual
Add to My manuals104 Pages
advertisement
Countertop
i.Series
PC100i, PC900i, PC1200i
PC100h, PC900h, PC1200h
Floor
i.Series
PC2200i, PC3200i, PC4200i
PC2200h, PC3200h, PC4200h
Platelet Incubator Service Manual
i.Series
®
and Horizon Series™
360096-1/P
Document History
J
K
L
I
Revision
M
N
O
P
Date
28 JUN 2013
23 DEC 2013
31 JAN 2014
22 APR 2014*
26 MAY 2014*
21 SEP 2015*
12 JUL 2016
6 DEC 2017
CO
8414
8965
9113
9386
9497
9754
12016
13171
Supersession
Supersedes A, B, C,
D, E, F, G, H
Revision Description
Revised layout for ease of navigation and locating information.
J supersedes I
K supersedes J
L supersedes K
M supersedes L
N supersedes M
O supersedes N
P supersedes O
Added 100 V information.
Added caution note for remote alarm interface.
Revised section I for consistency with existing manuals.
Changed voltage for remote alarm contacts.
Revised voltage specification for remote alarm contacts.
Added 0086 to CE certification.
Revised caption under photograph (chapter 16.1).
Revised power failure alarm test (chapter 19.2).
Changed part number for terminal board (chapters 12 and 21)
Revised schematic to correct labeling of J1 and J3 on the display board (page
78).
Changed test procedure for No Battery alarm test (chapter 8.8.7).
Updated Power Failure test for Horizon units
Reformatted content. Updated Emergo address. Updated condensing unit and refrigerant charge information for PC2200, PC3200 and PC4200, 115 V and 230 V models.
* Date submitted for Change Order review. Actual release date may vary.
Document Updates
The document is furnished for information use only, is subject to change without notice and should not be construed as a commitment by Helmer Scientific. Helmer Scientific assumes no responsibility or liability for any errors or inaccuracies that may appear in the informational content contained in this material. For the purpose of clarity, Helmer Scientific considers only the most recent revision of this document to be valid.
Notices and Disclaimers
Confidential / Proprietary Notices
Use of any portion(s) of this document to copy, translate, disassemble or decompile, or create or attempt to create by reverse engineering or otherwise the information from Helmer Scientific products is expressly prohibited.
Copyright and Trademark
Helmer ® , i.Series
® , i.Center
® , Horizon Series™, AgiTrak™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in the United States of America. Copyright © 2017 Helmer, Inc. All other trademarks and registered trademarks are the property of their respective owners.
Helmer, Inc., doing business as (DBA) Helmer Scientific and Helmer.
Disclaimer
This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain
Helmer Scientific products.
Any failure to follow the instructions as described could result in impaired product function, injury to the operator or others, or void applicable product warranties. Helmer Scientific accepts no responsibility for liability resulting from improper use or maintenance of its products.
The screenshots and component images appearing in this guide are provided for illustrative purposes only, and may vary slightly from the actual software screens and/or product components.
Helmer Scientific
14400 Bergen Boulevard
Noblesville, IN 46060 USA www.helmerinc.com
ISO 13485:2003 CERTIFIED Part No. 360096-1/ Rev P
Helmer Scientific Platelet Incubator Service Manual
Contents
1 About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Safety Precautions and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Product Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Control System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Temperature Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5 Chart Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 i.Series
®
Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 Installation and Configuration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Install Batteries for Backup Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Platelet Agitator and Platelet Incubator Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Install a Platelet Agitator in the Platelet Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4 External Monitoring Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Configure Drawers (PC4200i)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.6 Drawer Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3 Temperature Monitor Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.1 Home Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3 Change Configuration Password
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4 Calibrate Temperature Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5 Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.6 Alarm Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7 Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.8 Alarm Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.9 AgiTrak™ System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.10 Additional System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.11 Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.12 System Firmware Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.13 Reset the i.Center Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.14 View Manufacturer and Product Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4 Temperature Controller Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1 Level 1 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2 Level 2 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.3 Level 3 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.4 Level 4 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.5 Platelet Incubator Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.6 Calibrate the Temperature Controller Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.1 Recharge Refrigerant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2 i.Center Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3 Motion Alarm Battery (PC4200i) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.4 Clean the Platelet Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.5 Lubricate the Drawer Slides (PC4200i) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.6 Lubricate and Service Internal Parts (PC4200i) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
360096-1/P ii
Helmer Scientific Platelet Incubator Service Manual
6 Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.1 General Operation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2 Chamber Temperature Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3 Alarm Activation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.4 Testing Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5 Condensation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7 i.Series Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8 Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.1 PC100i, PC900i, PC1200i (100/115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.2 PC100i, PC900i, PC1200i (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.3 PC2200i (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.4 PC2200i (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.5 PC3200i (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.6 PC3200i (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.7 PC4200i (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8.8 PC4200i (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
9 i.Center Screen Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Horizon Series™ Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10 Installation and Configuration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.1 Install Battery for Backup Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.2 Platelet Agitator and Platelet Incubator Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.3 Install a Platelet Agitator in the Platelet Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
10.4 External Monitoring Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
10.5 Configuring Drawers (PC4200h)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
10.6 Drawer Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
11 Alarm Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.1 Temperature and Power Failure Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.2 Motion Alarm (PC4200h) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.3 Alarm Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
12 Temperature Controller Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
12.1 Level 1 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
12.2 Level 2 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
12.3 Level 3 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
12.4 Level 4 Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12.5 Change the Platelet Incubator Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12.6 Calibrate the Temperature Controller Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
13 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
13.1 Recharge Refrigerant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
13.2 Power Failure Alarm Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
13.3 Motion Alarm Battery (PC4200h) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
13.4 Clean the Platelet Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.5 Lubricate the Drawer Slides (PC4200h) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.6 Lubricate and Service Internal Parts (PC4200h) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
360096-1/P iii
Helmer Scientific Platelet Incubator Service Manual
14 Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
14.1 General Operation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
14.2 Chamber Temperature Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
14.3 Alarm Activation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
14.4 Condensation Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
15 Horizon Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
16 Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
16.1 PC100h, PC900h, PC1200h (100/115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
16.2 PC100h, PC900h, PC1200h (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
16.3 PC2200h (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
16.4 PC2200h (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
16.5 PC3200h (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
16.6 PC3200h (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
16.7 PC4200h (115 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
16.8 PC4200h (230 V) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Appendix A: Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Appendix B: Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
360096-1/P iv
Helmer Scientific Platelet Incubator Service Manual
1 About this Manual
This manual covers all platelet incubators, which may be identified singly, by their size, or by their respective “Series” and provides information on how to use the device. It is intended for use by end users of the platelet incubator and authorized service technicians.
Generic references are used throughout this manual to group models that contain similar features. For example, “PC100 models” refers to all models of that size (PC100i, PC100h).
1.1 Safety Precautions and Symbols
Symbols found in this document
The following symbols are used in this manual to emphasize certain details for the user:
Task Indicates procedures which need to be followed.
Note Provides useful information regarding a procedure or operating technique when using Helmer
Scientific products.
NOTICE Advises the user against initiating an action or creating a situation which could result in damage to equipment; personal injury is unlikely.
EC REP
CAUTION Advises the user against initiating an action or creating a situation which could result in damage to equipment or impair the quality of the products or cause minor injury.
WARNING Advises the user against initiating an action or creating a situation which could result in damage to equipment and serious personal injury to a patient or the user.
Authorized representative in the European Community
Symbols found on the units
The following symbols may be found on the refrigerator or refrigerator packaging:
CE Mark (European units only) Caution: Unlock all casters
Caution: Risk of damage to equipment or danger to operator
Caution: Hot surface
Earth / ground terminal
Protective earth / ground terminal
Caution: Shock / electrical hazard Product falls under the scope of the WEEE
(Waste Electrical and Electronic Equipment) directive.
360096-1/P
5
Helmer Scientific Platelet Incubator Service Manual
Avoiding Injury
Review safety instructions before installing, using, or maintaining the equipment.
♦
Before moving unit, remove contents from the drawers (if applicable).
♦
Do not open multiple drawers at the same time (if applicable).
♦
Before moving unit, ensure door(s) are closed and casters (if applicable) are unlocked and free of debris.
♦
Before moving unit, disconnect the AC power cord and secure the cord.
♦
When moving unit, use assistance from a second person.
♦
Never physically restrict any moving component.
♦
Avoid removing electrical service panels and access panels unless so instructed.
♦
Keep hands away from pinch points when closing the door or when agitation motion is enabled (if applicable).
♦
Avoid sharp edges when working inside the electrical compartment.
♦
Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good laboratory practices.
♦
Proceed with caution when adding and removing samples from the platelet incubator.
♦
Use supplied power cord only.
♦ Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment.
♦
Ensure biological materials are stored safely, in accordance with all applicable organizational, regulatory, and legal requirements.
♦ The platelet incubator is not considered to be a storage cabinet for flammable or hazardous materials.
NOTICE
Decontaminate parts prior to sending for service or repair. Contact Helmer Scientific or your distributor for decontamination instructions and a Return Authorization Number.
1.2 Product Labels
Note
Service information varies depending on the model and power requirements.
This information appears on the product specification label, on the back of the platelet incubator.
A
B
REF
SN
PC2200i i.Series®
1000000
Platelet Storage System
Weight 363 lb / 165 kg
Ref Type
R134A
0086
2017-12-05
Made in USA
Helmer, Inc.
14400 Bergen Boulevard
Noblesville, IN USA www.helmerinc.com
Voltage 115 V
HZ 50/60
Amps
Power
Fuse
14.5 A
1.61kW
10.0A
Amount Design Pressures GWP CO2 Equ
8oz/.23kg 88 psig/186 psig 1430 324kg
Sample Product Specification label.
C
Label
A
B
C
Description
Model
Serial number (SN)
Power requirements
360096-1/P
6
Helmer Scientific Platelet Incubator
1.3 Control System
Note
Service information varies depending on the control system.
Service Manual i.Center Control System i.Series platelet incubators are equipped with the i.Center monitoring system. A separate temperature controller is located behind the control panel on the platelet incubator.
i.Center monitoring system. Temperature controller.
Horizon Series Control System
Horizon Series platelet incubators feature the combined temperature monitor and controller. The Horizon Series system controls chamber temperature and monitors and displays operational information.
Horizon Series temperature controller and display.
1.4 Temperature Probes
Number and location of probes varies by model:
Model
PC100
PC900
PC1200
PC2200
PC3200
PC4200
Location
Front, top right corner
Back, right bottom corner
Back, right bottom corner
Front, right, between upper shelf and top of cabinet
Front, right, between upper shelf and top of cabinet
Front, right, next to middle shelf
360096-1/P
7
Helmer Scientific Platelet Incubator Service Manual
1.5 Chart Recorder
If installed, refer to the Temperature Chart Recorder Operation and Service Manual.
The chart recorder has a battery system, enabling a period of continuous operation if power is lost. Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available, backup power for the temperature chart recorder is available for up to 14 hours.
Prior to use:
♦
Install battery.
♦
Add paper.
♦
Calibrate chart recorder to match upper chamber temperature.
Open door by pressing and releasing.
Install Chart Paper
Note
For accurate temperature reading, ensure that current time is aligned with time line groove when chart knob is tightened.
1. Press and hold C
button. When stylus begins to move left, release button. The LED flashes to indicate current temperature range.
2. When stylus stops moving, remove chart knob then move knob up and away.
3. Place chart paper on chart recorder.
4. Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5. Hold chart paper and reinstall chart knob.
6.
Confirm temperature range is set to the correct value.
7. Press and hold C button. When stylus begins to move right, release button.
8.
Confirm stylus is marking temperature correctly.
360096-1/P
8
Helmer Scientific Platelet Incubator
i.Series
®
Models
Service Manual
2 Installation and Configuration
2.1 Install Batteries for Backup Power
The monitoring system and chart recorder each have a battery system, enabling a period of continuous operation if power is lost.
Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available and no battery-related alarms are active, backup power for the monitoring system is available for up to two hours.
The batteries are located on the top of the platelet incubator, under a removable access panel.
Batteries are disconnected for shipping. Reconnect batteries to provide monitoring system with backup power in the event of AC power failure.
Notes
• The monitoring system will start on battery power alone. If the platelet incubator was previously not connected to AC
power and the batteries are connected, the monitoring system will begin running on battery power.
• If AC power is lost, the monitoring system will continue to collect data until battery power is depleted.
CAUTION
Before installing or replacing batteries, switch the power OFF and disable backup power to the i.Center monitoring system by switching the ON/OFF key switch OFF (located behind the control door). Disconnect the platelet incubator from AC power.
NOTICE
Use only manufacturer supplied battery when installing a replacement battery.
2.2 Platelet Agitator and Platelet Incubator Compatibility i.Series platelet incubators may be fitted with i.Series platelet agitators. When installed in an i.Series platelet incubator, the i.Series platelet agitator can be monitored by the platelet incubator’s AgiTrak™ system.
The AgiTrak system receives alarm information through a direct connection from the platelet agitator’s alarm system to the platelet incubator’s alarm system.
Note
Only i.Series platelet incubators can monitor i.Series platelet agitators.
Platelet Incubator
PC100i
PC900i
PC1200i, PC2200i, PC3200i
Compatible Helmer Platelet Agitator
PF15i
PF48i
PF96i
360096-1/P
9
Helmer Scientific Platelet Incubator
2.3 Install a Platelet Agitator in the Platelet Incubator
Service Manual
Notes
• This section is not applicable to the PC4200i platelet incubator, which has a built-in platelet agitator.
•
For additional instruction on configuring a platelet agitator in a platelet incubator, refer to the platelet agitator service manual.
PC900i platelet incubator with a PF48i platelet agitator installed.
Helmer platelet agitators can be installed inside Helmer platelet incubators. i.Series platelet agitators can be connected to i.Series platelet incubators via a data cable. The connection allows the platelet incubator to monitor the platelet agitator motion, and to sound an alarm when agitation motion stops.
PC2200i and PC3200i platelet incubators hold 1 agitator per shelf. Each shelf features a corresponding power connector. In
PC2200i and PC3200i platelet incubators, each shelf includes a data port. Through the data port and data cable, agitation motion is communicated from i.Series platelet agitators to the i.Series platelet incubator.
When installing a platelet agitator on a shelf, use the power connector for that shelf to prevent the power cord from obstructing the agitation motion. When installing an i.Series platelet agitator on a shelf in an i.Series platelet incubator, use the data port for that shelf so agitation motion is communicated appropriately.
2.4 External Monitoring Devices
The remote alarm interface is a relay switch with three terminals:
♦
Common (COM)
♦ Normally Open (NO)
♦
Normally Closed (NC)
Terminals are dry contacts and do not supply voltage. Interface circuit is either normally open or normally closed, depending on terminals used.
Requirements for your alarm system determine which alarm wires must connect to terminals.
CAUTION
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
NOTICE
• For any platelet incubator (except PC4200i) without a platelet agitator installed, a bypass plug is needed for a normally-
closed interface circuit.
• The bypass plug is needed if multiple platelet agitators are installed, and not all are Helmer i.Series platelet agitators.
For each position in which a non-Helmer i.Series platelet agitator is installed, the bypass plug must be installed in the
data port.
• To obtain a bypass plug, contact Helmer Technical Service.
The terminals on the remote alarm interface have the following maximum load capacity:
♦
0.5 A at 30 V (RMS)
♦
1.0 A at 24 V (DC)
360096-1/P
10
Helmer Scientific Platelet Incubator
Connect to Remote Alarm Interface
1. Switch the Battery ON/OFF key switch OFF . Switch the AC ON/OFF switch OFF .
2. On the alarm panel, locate the remote alarm terminals.
3. Connect remote alarm wires to appropriate terminals, according to requirements for your alarm system.
4. Use a cable tie to relieve strain on alarm wires (as necessary).
5. Switch the Battery ON/OFF switch ON . Switch the AC ON/OFF switch ON .
6. Touch MUTE to disable the high temperature alarm while platelet incubator reaches operating temperature.
2.5 Configure Drawers (PC4200i)
CAUTION
• Before moving drawers, ensure they are completely empty for safe lifting.
• Open the drawer using the drawer handle; not the label holder.
Remove / Install Drawer
Service Manual
Remote alarm interface.
Drawer in the full open position. Drawer with the tab in front of the drawer stop.
1. Pull drawer out until it stops (when the tabs on the drawer contact the spring-loaded stops).
2. Depress the spring-loaded drawer stops.
3. Pull the drawer outward until the tabs on the drawer have passed the drawer stops.
4. Pull drawer free of the drawer slides.
5. Align drawer with the drawer slides.
6. Push drawer in until it stops (when the tabs on the drawer contact the spring-loaded stops).
7. Depress the spring-loaded drawer stops.
8. Push the drawer into the storage frame until it stops.
2.6 Drawer Labels
Drawer label holders may be installed on platelet agitator drawers.
Installed label holder.
Install Label Holders:
1. Insert the tabs on the label holder into the slots on the drawer.
2. Pivot the holder around the drawer handle. Align the hole on the label holder with the corresponding hole on the drawer.
3. Insert a screw through the hole in label holder and through the hole in the drawer.
4. Install the nut on the screw, on the back side of the label holder.
360096-1/P
11
Helmer Scientific Platelet Incubator Service Manual
3 Temperature Monitor Settings
Through the i.Center monitoring system, current settings may be viewed and changed. The settings that can be viewed or changed are as follows:
Setting
Description
Clock Mode
Date Format
Door Ajar Timeout
Pwr Failure Timeout
High Alarm Setpoint
Low Alarm Setpoint
Cond. Alarm Setpoint
Alarm Volume
Alarm Pulse
Chart Paper Timer; or Chart
Paper Days Left
Temperature Graph
Agitator Alarms
Time format used on the HOME screen and the Event Log
Date format used on the HOME screen and on the Event Log
Number of minutes the door can be open before an alarm activates
Number of minutes the platelet incubator can be without AC power before an alarm activates
Temperature at which the High Temperature Alarm activates
Temperature at which the Low Temperature Alarm activates
Temperature at which the Condenser Temperature Alarm activates
Volume for audible alarms
Pattern for audible alarms
Setting for the chart paper timer (disabled); or the number of days remaining before chart paper must be changed (if temperature chart recorder is installed)
The setting for the temperature graph feature (Enabled or Disabled)
Settings for the motion alarm for each Helmer i.Series agitator installed in the platelet incubator, or built-in agitator (PC4200i)
• All platelet incubators are programmed to monitor three positions, even though some models accommodate fewer platelet agitators.
• Positions that are not in use are set to OFF .
• Positions that are in use are set to ON and display the number of minutes the agitator can be stopped before an alarm activates.
View settings
Note
The Configuration screen is password protected. The factory default password is “1234”.
1. On the Home screen, press the MAIN button.
2. Press the DOWN button to highlight View Configuration. Press the SELECT button.
3. Press the UP or DOWN button to highlight the desired option. Press the SELECT button.
4. Press the BACK
button to return to the View Configuration screen, or press the
HOME button to exit.
NOTICE
Changing temperature monitor settings affects operation of the platelet incubator. Do not change settings unless instructed in product documentation or by Helmer Technical Service.
360096-1/P
12
Helmer Scientific Platelet Incubator Service Manual
3.1 Home Screen
The Home screen appears when the HOME button is touched from any other screen, or there is no interaction for two minutes on any screen other than the password entry screen.
HOME screen on the monitoring system.
Home Screen Functions
Note
Refer to section 9 (i.Center Screen Reference) for a complete list of screens in the i.Center monitoring system.
♦
View current temperature readings
♦
View the current time and date
♦
View detailed information about current or previous alarm events
♦
View the remaining backup battery charge
♦
View active alarms
♦
Mute audible alarms
♦
Adjust contrast
♦
View 24-hour chamber temperature graph
♦
Access Main screen to view and change settings
360096-1/P
13
Helmer Scientific Platelet Incubator
3.2 Main Screen
The Main screen displays functional options that allow access to all other screens in the system.
Service Manual
MAIN screen functional options.
Functions available from the Main screen:
Option
Event Log
System Alarm Test & Status
Edit Configuration (password required)
View Configuration
Product/Company Information i.Help
Function
View historical information about alarms and operational events
Start or stop an automatic test for temperature alarms
View the number of days remaining before the paper for the temperature chart recorder needs to be changed
View the current status of the door (OPEN or CLOSED)
View the current condenser temperature
Change the language used for text
Change date and time information
Change temperature units
Change the volume and pattern for audible alarms
Enable or disable the chart paper timer
Enable or disable the temperature graph display
Change alarm-related setpoints and timers
Calibrate the temperature probe reading
Change some settings to the factory default values
Change the password, preventing unauthorized changes
Configure the AgiTrak system to monitor agitation
View the date and time formats
View alarm-related setpoints and timers
View the volume and pattern for audible alarms
View the setting for the chart paper timer
View the setting for the temperature graph display
View the settings for agitator alarms
View the software versions for control and display components of the monitoring system
View information to contact Helmer
Access the on-board help system
View real-time agitation speed and cycle counts
View historical agitation event logs and details
Reset cycle counters for agitators
360096-1/P
14
Helmer Scientific Platelet Incubator Service Manual
3.3 Change Configuration Password
The default password is 1234. A new password must use four digits, ranging from 1 to 5.
Change the password:
1. On the Home screen, press the MAIN button.
2. Press the DOWN button to select Edit Configuration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to select Change Password. Press the SELECT button.
5.
Enter the new password, then re-enter the new password when prompted. If password entries match, the “update”
message is displayed. If password entries do not match, the “incorrect match” message is displayed.
6. Repeat the procedure to change the password.
3.4 Calibrate Temperature Probes
Verify the temperature probe(s) are reading chamber temperature correctly by comparing the chamber probe reading to temperature read by an independent thermometer. If the chamber temperature probe(s) are not reading correctly, change the value displayed on the temperature monitoring system.
Notes
• If the variance is within acceptable limits for your organization, changing probe settings is optional.
• Default setting for chamber temperature is 22.0 ºC. Value is factory-preset
• Value can be changed from -40.0 ºC to +40.0 ºC
Calibrate Probe(s)
Model
PC100i
Location
Front, top right corner
PC900i, PC1200i Back, right bottom corner
PC2200i Next to upper temperature probe
Model
PC3200i
PC4200i
Location
Upper reading: next to upper temperature probe
Lower reading: next to lower temperature probe
Upper or lower reading: in the middle shelf on the right side
1. Using the table above, place a calibrated independent reference thermometer traceable per national standards in the
specified location.
2. Close the door and allow the chamber temperature to stabilize for 30 minutes.
3. Observe and note the thermometer temperature.
4. On the Home screen, press the MAIN button.
5. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
6. Enter the password when prompted.
7. Press the DOWN button to highlight Temperature Calibration. Press the SELECT button.
8. Follow the system prompts for Model Type.
♦ For models with two probes , press the UP or DOWN buttons to select the Upper Probe or Lower Probe options. Press
SELECT button, then press the DOWN button to highlight the Temperature option.
♦
For models with one probe , the Upper Probe option is displayed and the Temperature option is highlighted.
9. Press the INC or DEC buttons to change the displayed value to match the value measured by the independent thermometer.
10. Press the DOWN button to highlight Store Calibration.
11. To save the new value, press the ENTER button. The “Calibration Memorized” message appears. New settings are saved.
12. To discard the new value, press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. New settings are not saved.
13. Remove the thermometer from the chamber and close the door.
Notes
• The current temperature displayed by the monitoring system may change so that it no longer matches the new probe
calibration value. This is normal.
• If a new probe value is entered but not saved, the new value will appear when the calibration setting for the probe is
viewed. This is normal.
360096-1/P
15
Helmer Scientific Platelet Incubator
3.5 Factory Default Settings
Settings listed below may be simultaneously returned to factory default values.
Service Manual
Setting
High Alarm Setpoint
Low Alarm Setpoint
Cond. Alarm Setpoint
Door Ajar Timeout
Power Failure Timeout
Agitator 1 Setpoint
Agitator 2 Setpoint
Agitator 3 Setpoint
Restored Value
24.0 °C
20.0 °C
50.0 °C
3 minutes
3 minutes
ON, 3 minutes
ON, 3 minutes
ON, 3 minutes
Restore Factory Default Settings
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Factory Default Settings. Press the SELECT button.
5. Press the ENTER button to restore the Factory default settings.
3.6 Alarm Reference
If an alarm condition is met, an alarm activates. Some alarms are visual only; others are visual and audible. Some alarms are sent through the remote alarm interface.
The table indicates if an alarm is audible (A), visual (V), or sent through the remote alarm interface (R).
Alarm
High Temperature
Low Temperature
Condenser Temperature
Door Open
Low Battery
No Battery
AC Power Failure
Change Chart Paper (1)
Agitator 1 Motion Alarm (2)
Agitator 2 Motion Alarm (3, 5)
Agitator 3 Motion Alarm (4, 5)
Alarm Type
A, V, R
A, V, R
A, V, R (i.Series)
A, V, R (i.Series)
V (i.Series)
A, V, R (i.Series)
A, V, R
V (i.Series)
A, V, R (i.Series); A, R (Horizon Series)
A, V, R (i.Series); R (Horizon Series)
A, V, R (i.Series); R (Horizon Series)
(1) Applicable if a temperature chart recorder is installed. Additional alarms may be available on the stand-alone chart recorder. Refer to the
Temperature Chart Recorder operation and service manual.
(2) Applicable to PC4200 platelet incubators. For other models, this alarm applies if a Helmer i.Series platelet agitator is installed as Agitator 1.
(3) Applicable to PC2200 and PC3200 platelet incubators with a Helmer i.Series platelet agitator installed as Agitator 2.
(4) Applicable to PC3200 platelet incubators with a Helmer i.Series platelet agitator installed as Agitator 3.
(5) Not applicable to PC4200h platelet incubators.
360096-1/P
16
Helmer Scientific Platelet Incubator
3.7 Alarm Settings
Service Manual
The following alarm settings may be changed by the operator. The setpoint for temperature alarms may be changed (where applicable), as well as the time delay between when the alarm condition commences and when the visual and audible alarms are initiated.
Setting
Alarm Volume
Default Setting
10
Default Time Delay n/a
Alarm Pulse
High Alarm Setpoint
Low Alarm Setpoint
Condenser Alarm
Power Failure
Door Ajar Timeout
Motion
Single
24.0 °C
20.0 °C
50.0 °C n/a n/a
On n/a n/a n/a n/a
3 minutes
3 minutes
3 minutes
Alarm Volume
The Alarm Volume controls volume for all audible alarms. The default setting is 10 and can be changed from 1 to 10 with 10 being the maximum setting.
Change the Alarm Volume
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Alarm Volume.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
Alarm Pulse
The alarm pattern can be changed to provide an easily distinguished sound. The default setting is Single, but may be changed to
Double, Triple or Constant.
Change the Alarm Pulse
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Alarm Pulse.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK button to return to the Edit Configuration screen, or press the HOME button to exit. The new settings are saved.
High Chamber Temperature Alarm
The High Alarm setpoint specifies the temperature at which the High Temperature Alarm activates. If the temperature detected by the upper chamber probe is greater than or equal to this value, the alarm activates. The default setpoint is 24 °C and can be changed from -40 °C to +40 °C.
Change High Temperature Setpoint:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight High Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
360096-1/P
17
Helmer Scientific Platelet Incubator Service Manual
Low Chamber Temperature Alarm
The Low Alarm setpoint specifies the temperature at which the Low Temperature Alarm activates. If the temperature detected by the upper chamber probe is less than or equal to this value, the alarm activates. The default setpoint is 20 °C and can be changed from -40 °C to +40 °C.
Change Low Temperature Setpoint:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Low Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
Condenser Temperature Alarm
The Condenser Alarm setpoint specifies the temperature at which the Condenser Temperature Alarm activates. If the temperature of the condenser discharge line is greater than or equal to this value, the alarm activates. The default setpoint is 50 °C and can be changed from -40 °C to +80 °C.
NOTICE
Condenser Temperature Alarm should not be changed unless directed by Helmer Technical Service.
Change Condenser Temperature Setpoint:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Cond. Alarm Setpoint.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
Power Failure Alarm
The Power Failure Timeout specifies longest time the platelet incubator can be without AC power before the alarm activates. If the time elapsed since the last power failure is greater than or equal to this value, the alarm activates. The default setting is three (3) minutes and can be changed from 0 to 60 minutes.
Change the Power Failure Alarm Delay:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Power Failure Timeout.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
360096-1/P
18
Helmer Scientific Platelet Incubator Service Manual
Door Ajar Timeout
The Door Ajar Timeout specifies longest time the platelet incubator door(s) can be open before the alarm activates. If the time elapsed since the last door opening is greater than or equal to this value, the alarm activates. The default delay setting is three (3) minutes and can be changed from 0 to 60 minutes.
Change the Door Ajar Alarm Delay:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Door Ajar Timeout.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
Motion Alarm (PC4200i)
Unlike other i.Series models, the PC4200i platelet incubator has a motion alarm delay control. When the platelet agitator motion stops, this control delays when alarm information is sent to the remote alarm interface on the platelet incubator. The alarm delay control does not delay motion information sent to the AgiTrak system. When the Agitator Alarm Setpoint is reached, the AgiTrak system activates an alarm, which is also sent to the remote alarm interface on the platelet incubator.
Notes
• To prevent a motion alarm being communicated by an external monitoring device but not by the platelet incubator, set
the alarm delay control to a value greater than the Agitator 1 Setpoint.
• The alarm delay control for the motion alarm is located on the rear of the platelet incubator.
Top: Motion alarm volume control. Bottom: Alarm delay control for the motion alarm.
The default motion alarm delay is set at the halfway point (approximately four (4) minutes). To change the motion alarm delay, use a small flat-head screwdriver. The maximum alarm delay is eight (8) minutes.
The default alarm volume is set at the middle position (medium).
Note
Do not set the alarm delay to 0 minutes.
Change Motion Alarm Delay:
Using a small, flat-head screwdriver, rotate the control to the left (counterclockwise) to shorten the motion alarm delay, or to the right (clockwise) to extend the motion alarm delay.
Change Motion Alarm Volume:
1. On back of the platelet incubator, locate the alarm volume control.
2. Position the switch to correspond with the desired motion alarm volume.
360096-1/P
19
Helmer Scientific Platelet Incubator Service Manual
3.8 Alarm Tests
Test alarms to ensure they are working correctly. The platelet incubator has alarms for chamber temperature, compressor temperature, door open (time), agitation motion (for installed platelet agitators and built-in PC4200i platelet agitator), low battery, and power failure.
Note
Before testing alarms, protect items in the platelet incubator from extended exposure to adverse temperature.
Automatic Chamber Temperature Alarm Test
Note
• Calibrate the chamber temperature probe prior to performing the Automatic Chamber alarm test.
• The test can be aborted by selecting the Cancel High or Low Test option.
•
The test takes less than five minutes.
When performing an automatic temperature alarm test, the Peltier device heats or cools the temperature probe until the high or low alarm setpoint is reached. An event is added to the Event Log to indicate a temperature alarm was activated.
Test the low alarm:
1. Identify the current setting for the low alarm setpoint.
2. Press the MAIN button.
3. Press the DOWN button to select System Alarm Test & Status. Press the SELECT button. The System Alarm Test &
Status screen appears.
4. Press the DOWN button to select Start Low Alarm Auto Test. Press the SELECT
button. The “Low Alarm Test in
Progress” message appears. When the test is complete, the message clears.
5. View the Event Log. Note the temperature at which the low alarm occurred. Compare the temperature to the low alarm setpoint. If the values do not match, refer to Troubleshooting.
Test the high alarm:
1. Identify the current setting for the high alarm setpoint.
2. Press the MAIN button.
3. Press the DOWN button to select System Alarm Test & Status. Press the SELECT button. The System Alarm Test &
Status screen appears.
4. Press the DOWN button to select Start High Alarm Auto Test. Press the SELECT
button. The “High Alarm Test in
Progress” message appears. When the test is complete, the message clears.
5. View the Event Log. Note the temperature at which the low alarm occurred. Compare the temperature to the high alarm setpoint. If the values do not match, refer to Troubleshooting.
Cancel the test:
1. Press the MAIN button.
2. Press the DOWN button to select System Alarm Test & Status. Press the SELECT button. The System Alarm Test &
Status screen appears.
3. Press the DOWN button to select Cancel High or Low Test. Press the ENTER button. The test is cancelled.
Note
When cancelling an automatic test, the message indicating the test is in progress clears immediately. If a setpoint was reached before the test was cancelled, the alarm activates and clears, as described earlier.
360096-1/P
20
Helmer Scientific Platelet Incubator
Manual Chamber Temperature Alarm Test (PC100i)
Service Manual
Notes
• Perform the low alarm test before the high alarm test to control the temperature more closely and complete the testing
more quickly.
• Calibrate the chamber temperature probe prior to performing the manual chamber alarm test.
• To perform the manual chamber alarm test on PC100i platelet incubators, obtain the alarm check tube.
•
Fill the tube with 2” (51 mm) of hot or cold water as noted in the instructions, and place the filled alarm check tube over
the chamber probe to perform the test.
• Do not remove the chamber probe from PC100i platelet incubators.
• Temperature probes are fragile; handle with care.
Test the Low Alarm:
1. Identify the current setting for the low alarm setpoint.
2. Remove the installed platelet agitator(s) to allow access to the to the chamber probe.
3. Fill the alarm check tube with cold water.
4. Place the alarm check tube in the holding bracket. Slide the tube upward to immerse the probe tip.
5. When the low temperature alarm sounds, note the temperature on the monitoring and control system display.
6. Compare the temperature at which the alarm sounds to the low alarm setpoint. If the values do not match, refer to Troubleshooting.
Test the High Alarm:
1. Identify the current setting for the high alarm setpoint.
2. Fill the alarm check tube with hot water.
3. Place the alarm check tube in the holding bracket. Slide the tube upward to immerse the probe tip.
4. When the high temperature alarm sounds, note the temperature on the monitoring and control system display.
5. Compare the temperature at which the alarm sounds to the high alarm setpoint. If the values do not match, refer to Troubleshooting.
6. Reinstall the platelet agitator in the chamber.
7. Slide the alarm check tube down so the probe is no longer immersed.
Manual Chamber Temperature Alarm Test (PC900i, PC1200i, PC2200i, PC3200i, PC4200i)
Notes
• Perform the low alarm test before the high alarm test to control the temperature more closely and complete the testing
more quickly.
• Calibrate the chamber temperature probe prior to performing the manual chamber alarm test.
• To perform the manual chamber alarm test on PC100i platelet incubators, obtain the alarm check tube.
•
Fill the tube with 2” (51 mm) of hot or cold water as noted in the instructions, and place the filled alarm check tube over
the chamber probe to perform the test.
• Do not remove the chamber probe from PC100i platelet incubators.
• Temperature probes are fragile; handle with care.
Test the Low Alarm:
1. Identify the current setting for the low alarm setpoint.
2. Remove the installed platelet agitator(s) to allow access to the to the chamber probe.
3. Loosen the screws securing the chamber probe bracket to the chamber wall. Slide the bracket up and away from the wall.
4.
Remove the probe from the bracket by unscrewing and removing the fitting securing the probe to the bracket.
360096-1/P
21
Helmer Scientific Platelet Incubator
5. Slide the probe tip upward and free the probe from the bracket.
Service Manual
6. Immerse the probe in cold water. Ensure the probe is at the bottom of the glass.
7. When the low temperature alarm sounds, note the temperature on the monitoring system display.
8. Compare the temperature at which the alarm sounds to the low alarm setpoint. If the values do not match, refer to Troubleshooting.
Test the High Alarm:
1. Identify the current setting for the high alarm setpoint.
2. Immerse the probe in hot water. Ensure the probe is at the bottom of the glass.
3. When the high temperature alarm sounds, note the temperature on the monitoring system display.
4. Compare the temperature at which the alarm sounds to the high alarm setpoint. If the values do not match, refer to Troubleshooting.
5. Remove the probe from the warm water.
6. Install the probe in the chamber probe bracket.
7.
Install the fitting to secure the probe to the probe bracket, leaving approximately 1” (25 mm) of the probe exposed beyond
the fitting.
8. Install the probe bracket in the chamber and tighten the screws.
9. Reinstall the platelet agitator(s) in the chamber.
Power Failure Alarm Test
Notes
• During a power failure, the power failure alarm sounds and the batteries provide power to the monitoring system.
• If AC power fails, the backup batteries will allow for continued data collection and temperature display.
• If the backup batteries fail, data is not collected and the temperature is not displayed.
• When power is restored, the 24 hours of data prior to the power loss are retained in the system memory. Stored
temperature data is displayed on the graph and the monitoring system resumes data collection and display.
Test Power Failure Alarm
1.
Confirm the platelet incubator is connected to AC power.
2. Ensure the monitoring system backup battery is switched ON .
3. Press the MAIN button.
4. Press the DOWN button to highlight Edit Configuration. Press the SELECT button.
5. Enter the password when prompted.
6. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
7. Press the DOWN button to highlight Power Failure Timeout.
8. Press the DEC button to change the setting to 0.
9. Switch the AC ON/OFF switch OFF . Power failure alarm will activate immediately.
10. Switch the AC ON/OFF switch ON . Power failure alarm will clear and audible alarm will cease.
11. Change the Power Failure Timeout setting to the original setting.
360096-1/P
22
Helmer Scientific Platelet Incubator Service Manual
Door Ajar Alarm Test
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Agitator Setpoints. Press the SELECT button.
5. Press the DOWN button to highlight Door Ajar Timeout.
6. Press the DEC button to change the setting to 0.
7. Press the BACK button to return to the Edit Configuration screen, or press the HOME button to exit. New settings are saved.
8. Open the door. Door ajar alarm will activate immediately.
9. Close the door. Door ajar alarm will clear and audible alarm will cease.
10. Change the Door Ajar Timeout setting to the original setting.
Motion Alarm Test
1. Ensure the Motion Alarm ON/OFF switch is switched ON .
2. Press the MAIN button.
3. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
4. Enter the password when prompted.
5. Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
6. Press the DOWN button to highlight Agitator Alarm Setpoint. Press the SELECT button.
7. Press the DOWN button to highlight Agitator 1 Setpoint.
8. Press the DEC button to change the setting to 0.
9. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. New settings are saved.
10. Open the door. Agitation motion will stop and the motion alarm will activate immediately.
11. Close the door. Agitation motion will resume. The motion alarm will clear and the audible alarm will cease.
12. Change the Motion Alarm Delay setting to the original setting.
No Battery Alarm Test
Test the no battery alarm to ensure that it activates in the event the backup batteries are depleted.
Notes
• During a power failure, the power failure alarm sounds and the batteries provide power to the monitoring system.
• If AC power fails, the backup batteries will allow for continued data collection and temperature display.
• If the backup batteries fail, data is not collected and the temperature is not displayed.
Test No Battery Alarm:
1. Switch the Battery ON/OFF key switch OFF . The No Battery alarm will activate.
2. Switch the battery ON/OFF key switch ON . The No Battery alarm will clear.
NOTICE
Use a battery meeting manufacturer’s specifications.
360096-1/P
23
Helmer Scientific Platelet Incubator Service Manual
3.9 AgiTrak™ System
The AgiTrak system is integrated into the i.Center monitoring system. The AgiTrak system simultaneously receives motion information from as many as three i.Series platelet agitators. Within the AgiTrak interface, the topmost platelet agitator installed in the platelet incubator is numbered 1, and subsequent platelet agitators are numbered in ascending order.
Note
In the PC4200i platelet incubator, the built-in agitator is numbered 1.
Platelet Incubator Model
PC100i
PC900i
PC1200i
PC2200i
PC3200i
PC4200i
3
1
1
2
Agitator Positions
1
1
Note
The PC4200i’s built-in agitator transmits motion and speed information directly to the AgiTrak system without a separate data cable.
The platelet incubator has a data port for each platelet agitator position. The installed platelet agitator transmits motion and speed information to the AgiTrak system through a data cable and the data port in the platelet incubator. The AgiTrak system monitors platelet agitator motion on/off status and current agitation speed.
If motion is not detected for the pre-configured amount of time, the AgiTrak system activates an alarm and records the alarm event(s) in the event log. For agitation information to be monitored correctly, each platelet agitator position must be enabled in the
AgiTrak interface.
Disable a platelet agitator position through the AgiTrak interface under the following conditions:
♦
The position is not available
♦
The position is not in use
♦
The position has a Horizon Series agitator installed
Note
If a Horizon Series platelet agitator is installed in the platelet incubator, its position (in the AgiTrak interface) should be disabled to prevent false alarms.
The Agitator Alarm Setpoint specifies the longest the agitator can be stopped prior to alarm. If the time elapsed is greater than or equal to this value, an alarm activates. The default setting is three minutes. The Setting can be changed from 0 minutes to 60 minutes
Note
The AgiTrak system is intended to replace each i.Series platelet agitator alarm. The motion alarm on each installed i.Series platelet agitator should be disabled when using the alarms in the AgiTrak system.
View Real-Time Agitation Speed
For each i.Series platelet agitator installed in the platelet incubator (or for the built-in platelet agitator, in the case of a PC4200i platelet incubator). agitation speed may be viewed in real-time.
View Agitation Speed:
1. Press the MAIN button.
2. Press the DOWN button to highlight i.Help. Press the SELECT button.
3. Press the DOWN button to highlight Agitator Information. Press the SELECT button. The Agitator RPM screen is displayed. The current speed for each platelet agitator is shown.
4. Press the BACK button to return to the i.Help screen, or press the HOME button to exit.
360096-1/P
24
Helmer Scientific Platelet Incubator Service Manual
View Real-Time Agitator Cycle Counts
For each i.Series platelet agitator installed in the platelet incubator (or for the built-in platelet agitator, in the case of a PC4200i platelet incubator), cycle counts may be viewed in real-time. The cycle counter information is useful for troubleshooting problems with the agitation motion sensor, or for scheduling additional preventive maintenance functions.
View Agitation Cycle Count:
1. Press the MAIN button.
2. Press the DOWN button to highlight i.Help. Press the SELECT button.
3. Press the DOWN button to highlight Agitator Information. Press the SELECT button. The Agitator RPM screen is displayed.
4. Display the Agitator Cycle Count by pressing the DOWN button to highlight Agitator RPM.
5. Press the SELECT button. The Agitator RPM screen is displayed.
6. Press the CYCLES button.
7. Press the DOWN button to highlight Agitator Cycles. The platelet agitator cycle count is displayed, in units of thousands.
8. Press the BACK button to return to the i.Help screen, or press the HOME button to exit.
3.10 Additional System Settings
Display Language
The i.Center monitoring system stores two languages. English is the default language. If a different language is desired, it must be loaded from the flash memory card. If a flash memory card is not included with the platelet incubator, the languages may have been loaded to the i.Center prior to shipment.
To obtain a flash memory card to load an alternate language, contact Helmer Technical Service.
Note
Each time the platelet incubator is powered on, the i.Center display language must be selected.
Set Display Language on Power-on:
1. Switch the AC ON/OFF switch ON .
2. Switch the Battery ON/OFF key switch ON . The platelet incubator powers on and the i.Center will display the System
Options screen.
3. Press the INC or DEC buttons to select the desired language. Press the SELECT button
4. Press the HOME button to return to the Home screen.
5. If a temperature alarm sounds, press the MUTE button.
Change Display Language
1. Press the MAIN button.
2. Press the DOWN button to highlight Edit Configuration. Press the SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the UP or DOWN buttons to select Language. Press the SELECT button.
6. Press the INC or DEC buttons to select the desired language.
7. Press the BACK button to return to the System Options screen, or press the HOME button to exit. The new settings are saved.
360096-1/P
25
Helmer Scientific Platelet Incubator Service Manual
Date and Time
The Date Format setting controls the order in which the month (mm) and day (dd) are displayed. The default date format is mm/dd/yyyy.
The Clock Mode setting controls whether the time is displayed in a 12-hour or 24-hour format. When using the 12-hour format, AM or PM must be specified. Default setting is 12-hour.
Change Date and Time Settings:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Set Date & Time. Press the SELECT button.
5. Press the UP or DOWN buttons to select the date and time settings to change.
6. Press the INC or DEC buttons to change the setting.
7. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new settings are saved.
Temperature Units
Temperature units can be displayed in Celsius (°C) or Fahrenheit (°F). The default temperature unit is Celsius.
Change temperature units:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight Temperature Units.
5. Press the INC or DEC buttons to select the desired temperature units.
6. Press the BACK
button to return to the Edit Configuration screen, or press the
HOME button to exit. The new setting is saved.
Temperature Graph
The home screen includes a user-enabled 24-hour temperature graph. Current readings from the temperature probe(s) are displayed. While the i.Center monitoring system is powered on, temperature data from the upper temperature probe is displayed in real-time. The default setting is Enabled.
Notes
• Inactivity of one minute on any i.Center screen results in an automatic return to the Temperature Graph screen (if the
temperature graph is enabled).
• The graph is cleared if a button is pressed or if an alarm activates.
Home screen with temperature graph.
360096-1/P
26
Helmer Scientific Platelet Incubator Service Manual
Enable or disable the temperature graph:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Temperature Graph.
6. Press the INC or DEC buttons to enable or disable the temperature graph.
7. Press the BACK button to return to the System Options screen, or press the HOME button to exit. The new setting is saved.
View Chart Paper Status
The i.Center monitoring system includes a resettable count-down timer for the chart paper. This serves as a reminder to change the chart paper when the 7-day paper is nearly depleted.
View Chart Paper Status:
1. Press the MAIN button.
2. Press the DOWN button to highlight System Alarm Test & Status.
3. Press the SELECT button. The number of days remaining (if the chart paper timer is enabled) is displayed. If the chart paper timer is not enabled, the status of the chart paper timer is displayed.
4. Press the BACK button to return to the Main screen, or press the HOME button to exit.
Enable, Disable, or Reset the Chart Paper Timer
The default setting for the chart paper timer is Enabled. One sheet of chart paper records temperatures continuously for seven days. The timer activates an alarm 6.5 days from when the timer is reset. The timer period cannot be changed.
Note
Enabling the timer also resets the timer.
Change Chart Paper Timer Setting:
1. Press the MAIN button.
2. Press the DOWN
button to highlight Edit Configuration. Press the
SELECT button.
3. Enter the password when prompted.
4. Press the DOWN button to highlight System Options. Press the SELECT button.
5. Press the DOWN button to highlight Chart Paper Timer.
6. Press the INC or DEC buttons to select Enabled, Disabled, or Reset.
7. If Enabled or Disabled is selected, press the BACK button to return to the System Options screen, or press the HOME button to exit. The new setting is saved.
8. If Reset is selected, press the DOWN button then press the PAPER-CHANGED button. The System Options screen appears with the Chart Paper Timer set to Enabled.
9. Press the BACK button to return to the System Options screen, or press the HOME button to exit. The new setting is saved.
Door Status
The door open/closed status may be viewed through the i.Center monitoring system. This may be helpful in troubleshooting functional problems.
View Door Status:
1. Press the MAIN button.
2. Press the DOWN button to highlight System Alarm Test & Status. Press the SELECT button. The door status (OPEN or
CLOSED) is displayed.
3. Press the BACK button to return to the Main screen, or press the HOME button to exit.
360096-1/P
27
Helmer Scientific Platelet Incubator
Condenser Temperature
The temperature of the condenser discharge line may be referenced for troubleshooting operational problems.
Service Manual
View Condenser Temperature:
1. Press the MAIN button.
2. Press the DOWN button to highlight System Alarm Test & Status. Press the SELECT button. The condenser temperature is displayed.
3. Press the BACK button to return to the i.Help screen, or press the HOME button to exit.
i.Help
The i.Center monitoring system includes an on-board help system.
Access i.Help System:
1. Press the MAIN button.
2. Press the DOWN button to highlight i.Help. Press the SELECT button. The i.Help screen is displayed.
3. Follow on-screen instructions to view additional help topics.
4. Press the BACK button to return to the Main screen, or press the HOME button to exit.
Note
The level of detail in i.Help may differ slightly from the contents of this manual.
360096-1/P
28
Helmer Scientific Platelet Incubator Service Manual
3.11 Event Log
The Event Log allows 50 (most recent) alarm events to be viewed and shows information regarding the number of door opening from the current and previous day. Information regarding agitator cycle counter reset can also be viewed.
View the Event Log:
1. Press the MAIN button.
2. Press the DOWN button to highlight Event Log. Press the SELECT button. The Event Log screen is displayed.
Event log format:
EVENT LOG Door AA/AA BB/BB
Openings: C D
EE F GG HH/HH/HH II:IIJ KK.KK°L
B
C
D
E
F
G
Item
AA/AA
BB/BB
C
D
EE
F
GG
HH/HH/HH
II:II
J
KK.KK
L
Event Details
A
Description
Current date (month and day)
Previous date (month and day)
Number of door openings on current date
Number of door openings on previous date
Event number. The most recent event is numbered as 1. Values can be 1-50.
Event. “S” indicates the start of an alarm condition. “R” indicates the alarm was reset and the system returned to normal.
Alarm Type:
DR Door Open
HI
LO
High temperature
Low temperature
CO
NB
AC
A(n)
Condenser temperature
No Battery
Power failure
Agitator (n) stopped (where n is 1, 2, or 3, depending on model)
Date of event (month and day, and the last two digits of the year)
Time (hours and minutes) of event
Time (appears for 12-hour format). “A” indicates AM. “P” indicates PM.
Upper chamber temperature at time of event
Temperature units. C = Celsius. F = Fahrenheit.
Label Description
C
D
E
F
G
A
B
Event number
Event
Alarm Type
Date and time of event
Upper chamber temperature at time of event
Lower chamber temperature at time of event (PC3200i and PC4200i)
Condenser discharge line temperature at time of event
Event Log Detail screen.
360096-1/P
29
Helmer Scientific Platelet Incubator Service Manual
View an Event:
1. Press the MAIN button.
2. Press the DOWN button to highlight Event Log. Press the SELECT button.
3. From the Event Log screen, press the UP or DOWN buttons to highlight the desired event number. Press the SELECT button. The Event Log Detail screen for the selected event is displayed.
4. Press the BACK button to return to the Event Log screen, or press the HOME button to exit.
Note
If the event is highlighted the alarm for that event was caused by a system self-test, initiated by an operator.
Agitator Count Log
The i.Center monitoring system has three cycle counter logs, one for each agitator. Logs display date, time, and cycle count for the previous 50 reset events. Each time a counter is reset for an agitator, an event is recorded in the log for that agitator.
View agitator count logs:
1. Press the MAIN button.
2. Press the DOWN button to highlight i.Help. Press the SELECT button.
3. Press the DOWN button to highlight Agitator Information. Press the SELECT button.
4. Press the UP or DOWN buttons to highlight Agitator Count Logs. Press the SELECT button.
5. Press the UP or DOWN buttons to highlight the desired Agitator Count Log. Press the ENTER button. The Agitator Count
Log screen is displayed.
6. Press the UP or DOWN buttons to view previous events in the Agitator Count Log.
7. Press the BACK button to return to the Agitator Information screen, or press the HOME button to exit.
Agitator Count Log Format
Total Counts AAAAAAAAA
BB CC/CC/CC DD:DDE FFFFFFFFF
Item
AAAAAAAAA
BB
CC/CC/CC
DD:DD
E
FFFFFFFFF
Description
Total cycles counted
Reset event number. The most recent event is numbered as 1. Values can be 1-50.
Date of event (month and day, and the last two digits of the year)
Time (hours and minutes) of event.
Time (appears for 12-hour format). A indicates AM, and P indicates PM.
Cycles counted since the last reset event
3.12 System Firmware Upgrade
Helmer may occasionally issue updates for the i.Center firmware. Follow upgrade instructions included with the firmware update.
3.13 Reset the i.Center Monitoring System
1. Switch the Battery ON/OFF key switch OFF .
2. Switch the AC ON/OFF switch OFF .
3. Switch the AC ON/OFF switch ON .
4. Switch the Battery ON/OFF key switch ON .
3.14 View Manufacturer and Product Information
1. Press the MAIN button.
2. Press the INC or DEC buttons to select the Product/Company Information option. Press the SELECT button.
Manufacturer contact information appears. Software version appears.
360096-1/P
30
Helmer Scientific Platelet Incubator Service Manual
4 Temperature Controller Setpoints
Temperature controller setpoints are programmed at the factory. Setpoints can be viewed and changed through the temperature controller. Parameter values reside in four program levels.
Temperature controller display.
Changeable Parameters
Parameter
Program
Level
ZEro
AL.hi
AL.Lo
3
1
1
Function
Calibrate the temperature controller readout
Set high temperature alarm value: Horizon Series
Set low temperature alarm value: Horizon Series
View / Change Parameter Values:
NOTICE
Changing parameter values affects platelet incubator operation. Do not change parameter values unless instructed in product documentation or by Helmer Technical Service.
Note
When there is no interaction for 60 seconds, the temperature controller exits program mode.
1. Press and hold the UP and DOWN
buttons simultaneously for approximately three seconds. “tunE” and “oFF” flash on the display. The temperature controller is now in Level 1 program mode.
2. Press and release the UP or DOWN
buttons until the desired parameter flashes on the display.
3. Change a parameter value by pressing and holding the * button.
4. Press the UP or DOWN buttons to change the parameter value.
♦ To access Level 2 parameters, change the value for the “LEVL” parameter to
2 .
♦ To access Level 3 parameters, change the value for the “LEVL” parameter to
3 .
5.
Release all buttons to exit the parameter. New settings are saved. If the “LEVL” parameter value is changed, the temperature controller returns to the selected program level.
6. Repeat steps 2 through 4 to access another program level, or to view or change parameter values in the selected level.
7. Access Level 4 parameters by navigating to Level 3 program mode.
8.
Select the “UEr” parameter.
9. Press and hold the UP and DOWN
buttons simultaneously for approximately 10 seconds. The “LOCK” parameter flashes on the display.
10. Exit program mode by pressing and holding the UP and DOWN arrow buttons simultaneously for approximately three seconds, or until the current chamber temperature is displayed.
360096-1/P
31
Helmer Scientific Platelet Incubator
4.1 Level 1 Parameters and Values
Service Manual
Notes
• Parameters are listed in order of appearance.
• SP1 refers to the chamber heater.
• SP2 refers to the temperature alarm for Horizon Series models.
• Default values for AL.Lo and AL.hi may vary: the values listed are for temperature controllers programmed at the factory
with a platelet incubator setpoint of 22 °C.
Parameter Description
Autotune selection
SP1 proportional band (gain) or hysteresis
(in °C)
SP1 integral time/reset (in minutes)
SP1 derivative time/rate
SP1 derivative approach control
SP1 proportional cycle-time
SP1 offset (manual reset)
Lock main setpoint
Ramp rate (if ramp is on)
Ramp selection
Soak time selection
SP2 low alarm setpoint (in °C)
SP2 high alarm setpoint (in °C)
SP2 hysteresis or proportional band (in °C)
SP2 mode selection (ON/OFF or proportional cycle-time)
Parameter level currently selected
Default Value (Countertop
Models)
Default Value
(Floor Models) oFF oFF
7.5
3.2
0.5
oFF
0 oFF
7.5
oFF
0.5
0.5
--
20.0 (see note above)
24.0 (see note above)
0.1
on.oF
1
0.5
oFF
0 oFF
15 oFF
0.5
0.5
--
20.0 (see note above)
24.0 (see note above)
0.1
on.oF
1
360096-1/P
32
Helmer Scientific Platelet Incubator
4.2 Level 2 Parameters and Values
Parameter Description
Display of SP1 output power percentage (read-only)
Selection for manual control of power for SP1
SP1 power limit percentage
SP2 power limit percentage (cooling)
Main SP2 operating mode (alarm strategy)
Subsidiary SP2 mode
Display resolution
Full scale
Minimum scale
Input sensor type
Units (°C / °F)
Parameter level currently selected
35.0
15.0
rtd
°C
2
0 bAnd nonE
0.1°
Default Value
(Countertop Models)
(read only)
Default Value
(Floor Models)
(read only) oFF
100 oFF
100
0 bAnd nonE
0.1°
35.0
15.0
rtd
°C
2
4.3 Level 3 Parameters and Values
Parameter Description
SP1 output device type
SP2 output device type (read only)
Sensor burn-out protection type (upscale or downscale)
Output mode (reverse or direct)
Default Value (Countertop
Models)
Default Value
(Floor Models)
SSd1 SSd1
SSd2 uP.SC
1r.2r
SSd2 uP.SC
1r.2r
LED indicator modes for SP1 and SP2 (normal or inverted) 1n.2n
Sensor span adjust 0.0
Zero sensor error (calibration across full scale)
Selection for the control accuracy monitor
(varies) oFF
1n.2n
0.0
(varies) oFF
Read control accuracy monitor results (variance) UAr°
Read autotune tuning cycle data
Software version (select this parameter to access Level 4 parameters)
Function reset
Ct A
392b nonE
Parameter level currently selected 3
UAr°
Ct A
392b nonE
3
Service Manual
360096-1/P
33
Helmer Scientific Platelet Incubator
4.4 Level 4 Parameters and Values
Parameter Description
Program security lock
Derivative sensitivity
Display sensitivity (or direct display of input)
Disable SP2 alarm annunciator
Program mode auto-exit switch (returns display to normal mode if no activity for one minute) oFF
Auto
Default Value (Countertop
Models)
Default Value
(Floor Models) nonE nonE
0.5
dir
0.5
dir oFF
Auto
Service Manual
4.5 Platelet Incubator Setpoint
The default setpoint is 22.0 °C. The temperature controller adjusts chamber temperature around the setpoint. When chamber temperature is below setpoint, the controller activates the heater to warm the chamber. When temperature rises above setpoint, the controller deactivates the heater to allow the chamber to cool. The compressor runs continuously.
NOTICE
Do not change the setpoint to a value outside the temperature control range.
Open the control door to access the temperature controller. The current chamber temperature is displayed. The temperature may be different from the setpoint, especially if the chamber temperature has not yet stabilized.
Temperature controller display.
Change Platelet Incubator Setpoint
1. Press and hold the
2. Press and hold the *
button to access the setpoint adjustment function. The controller displays current setpoint value.
button. to change the setpoint.
3. Press the UP or DOWN buttons to increase or decrease setpoint in increments of 0.1 °C.
4. Release all buttons to exit the setpoint parameter. New settings are saved.
360096-1/P
34
Helmer Scientific Platelet Incubator Service Manual
4.6 Calibrate the Temperature Controller Display
The temperature controller senses chamber temperature through a probe in the chamber. The chamber setpoint typically varies from the measured temperature, so an offset value is used by the control system to compensate for the difference.
NOTICE
Do not change the control sensor offset unless instructed in product documentation or by Helmer Technical Service.
Notes
• Value is factory-preset and varies for each unit
• Offset value can be changed from -10.0 °C to +10.0 °C
• If the variance is within acceptable limits for your organization, changing the offset value is optional.
Determine New Offset Value:
1. Use the zip tie to attach a calibrated independent thermometer traceable per national standards to the control probe.
2. Close door and allow chamber temperature to stabilize for 30 minutes.
3. Observe independent thermometer temperature for 10 minutes and determine temperature range.
4. From the range, calculate the average temperature.
5. Remove the thermometer from the control probe.
6. Determine the change in value to reach desired setpoint.
Example ♦ Current setpoint is 22.0 ºC
♦
Average temperature is 23 ºC
♦ Offset adjustment value is +1.0 ºC
Enter New Offset Value:
Note
When there is no interaction for 60 seconds, the temperature controller exits program mode.
1. Press and hold the UP and DOWN
buttons simultaneously for approximately three seconds. “tunE” and “oFF” flash on the display. The temperature controller is now in Level 1 program mode.
2. Press and release the UP or DOWN
buttons until “LEVL” flashes on the display.
3. Press and hold the * button.
4. Press the UP or DOWN buttons to change the parameter value to 3 .
5. Release the * button.
6. Press and release the UP or DOWN
buttons until “ZEro” flashes on the display.
7. Press and hold the * button.
8. Press the UP or DOWN buttons to change the value to the calculated offset adjustment value.
9. Release the * button.
10.
Release all buttons to exit the parameter. New settings are saved. If the “LEVL” parameter value is changed, the temperature controller returns to the selected program level.
11. Press and hold the UP and DOWN arrow buttons simultaneously for approximately three seconds, or until the current chamber temperature is displayed to exit program mode.
360096-1/P
35
Helmer Scientific Platelet Incubator
5 Maintenance
Service Manual
NOTICE
• Before performing maintenance, protect items in platelet incubator from extended exposure to adverse temperature.
• Allow platelet incubator temperature to stabilize at setpoint after performing service or after extended door opening.
Note
Refer to the operation manual for the preventive maintenance schedule.
5.1 Recharge Refrigerant
CAUTION
• Review all safety instructions prior to recharging refrigerant.
• Maintenance should only be performed by trained refrigeration technicians.
NOTICE
Use only non-CFC R-134A refrigerant.
Full initial refrigerant charge by model.
Model
PC100i, PC900i, PC1200i
Volts
Initial Charge
115/230 5.0 oz (142 g)
PC2200i, PC3200i, PC4200i
(serial numbers prior to 2042392)
115
PC2200i, PC3200i, PC4200i
(serial number 2042392 and after)
115
8.0 oz (227 g)
4.0 oz (113 g)
PC2200i, PC3200i, PC4200i
(serial numbers prior to 2046319)
230
PC2200i, PC3200i, PC4200i
(serial number 2046319 and after)
230
8.0 oz (227 g)
12.0 oz (340 g)
Add Refrigerant:
Note
Pressure varies depending on ambient air temperature.
1.
Attach calibrated pressure gauge (0 lbs/in² to 60 lbs/in² (0 kPa to 400 kPa)) to the fittings on the refrigeration lines.
2. Measure the low side (suction) pressure.
3. Add refrigerant so the peak low side pressure corresponds to pressure (at the stated ambient temperature) in the table below.
24
25
26
27
28
29
30
18
19
20
21
22
23
Ambient
Temperature
°C °F
75.2
77.0
78.8
80.6
82.4
84.2
86.0
64.4
66.2
68.0
69.8
71.6
73.4
Low Side Pressure lbs/in²
32.0
32.6
33.3
33.5
34.3
34.8
35.4
29.0
29.5
30.0
30.5
31.0
31.5
kPa
221
225
230
231
236
240
244
200
203
207
210
214
217 bar
2.21
2.25
2.30
2.31
2.36
2.40
2.44
2.00
2.03
2.07
2.10
2.14
2.17
4. Remove pressure gauge.
360096-1/P
36
Helmer Scientific Platelet Incubator Service Manual
5.2 i.Center Batteries
The i.Center monitoring system has visual indicators for battery charge level. If the batteries deplete to a particular voltage output, a flashing Low Battery alarm is initiated. If the batteries are missing or nearly depleted, the flashing No Battery alarm initiates.
Test i.Center Backup Batteries:
1. Switch AC ON/OFF switch OFF . Screen should continue to display information with reduced brightness. If the display is blank, replace battery.
2. Switch AC ON/OFF switch ON .
Replace the i.Center Batteries
NOTICE
Use batteries which meet manufacturer’s specifications. (1.5 V D-cell alkaline batteries or equivalent)
1. Switch AC ON/OFF switch OFF . Disconnect the platelet incubator from AC power.
2. Remove the access panel from the top of the platelet incubator.
3. Remove six batteries and replace with six new batteries.
4. Reinstall access panel.
5. Reconnect the platelet incubator to AC power. Switch AC ON/OFF switch ON .
5.3 Motion Alarm Battery (PC4200i)
The motion alarm battery is applicable to the PC4200i platelet incubator only.
Test Motion Alarm Battery
1. Ensure the Motion Alarm ON/OFF switch is switched ON .
2. Ensure the monitoring system backup battery is switched ON .
3.
Using a small flat-head screwdriver, rotate the Motion Alarm Delay control to the left (counterclockwise) to shorten the
motion alarm delay to five seconds.
4.
Open the door to disable agitation. Motion alarm will activate after five seconds.
Note
• If the agitation motion alarm sounds, the battery does not need to be replaced.
• If the agitation motion alarm does not sound, replace the motion alarm backup battery.
5. Close the door to enable agitation. Agitation motion alarm will clear and audible alarm will cease.
6.
Using a small flat-head screwdriver, rotate the Motion Alarm Delay control to the right (clockwise) to change the motion alarm delay to the original setting.
Replace the Motion Alarm Battery (PC4200i)
NOTICE
Use a battery which meets manufacturer’s specifications. (9 V nickel-metal hydride rechargeable battery or equivalent)
Note
The motion alarm battery is applicable to the PC4200i platelet incubator only.
1. Switch the Battery ON/OFF key switch OFF .
2. Switch the AC ON/OFF switch OFF .
3. Disconnect the platelet incubator from AC power.
4. Remove the access panel on top of the platelet incubator.
5. Remove the battery. Replace with a new battery.
6. Reinstall the access panel.
7. Reconnect the platelet incubator to AC power.
8. Switch the AC ON/OFF switch ON .
9. Switch the Battery ON/OFF switch ON .
360096-1/P
37
Helmer Scientific Platelet Incubator
5.4 Clean the Platelet Incubator
Service Manual
WARNING
Disconnect platelet incubator from AC power when cleaning.
Condenser Grill
In environments where the platelet incubator is exposed to excessive lint or dust, the condenser grill may require cleaning more frequently than stated in preventive maintenance schedule.
Clean the condenser grill using a soft brush and a vacuum cleaner.
Exterior
Clean glass surfaces with soft cotton cloth and glass cleaner. Clean exterior surfaces with soft cotton cloth and non-abrasive liquid cleaner.
CAUTION
The condensate evaporator and water evaporation tray are hot.
NOTICE
For PC4200i platelet incubators, clean the vents in the base as needed.
Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose laboratory cleaner suitable for stainless steel. For PC4200i models, clean textured drawer surfaces with a soft brush.
Door Gaskets
Clean with soft cloth and mild soap and water solution.
5.5 Lubricate the Drawer Slides (PC4200i)
CAUTION
• Before moving drawers, ensure they are completely empty.
• Open the drawer using the drawer handle; not the label holder.
• Do not open more than one drawer at a time.
Clean the slots in the drawer guides with a clean, dry cloth. Apply a thin film of petroleum jelly to the bottom of each drawer where it contacts the drawer guide.
360096-1/P
38
Helmer Scientific Platelet Incubator
5.6 Lubricate and Service Internal Parts (PC4200i)
Service Manual
WARNING
Disconnect platelet incubator from AC power when lubricating internal parts.
Drive Rod Bearings
Clean drive rod end bearings, wheels, and components according to the preventive maintenance schedule in the operation manual.
Lubricate Drive Rod Bearings:
1. Obtain a grease gun with bearing grease.
2.
Inject bearing grease into the grease fitting on the bearing.
Label Description
A
B
C
Drive rod
Grease fitting
Rod end bearing
A B C
Drive rod and rod end bearing detail.
Wheels
NOTICE
Do not use oil or aerosol lubricant.
Lubricate wheel axles with a light machine oil. Remove the cover to access the front wheels for inspection, cleaning, and lubrication. Replace wheels as necessary. The hole in each wheel should be circular. If the hole is out-of-round, replace the wheel.
If the wheels need to be replaced, contact Helmer Technical Service.
360096-1/P
39
Helmer Scientific Platelet Incubator
6 Troubleshooting
CAUTION
Review all safety instructions prior to troubleshooting.
Service Manual
6.1 General Operation Problems
Problem
Display on the monitoring system is hard to read.
Monitoring system is not responding.
Possible Cause
Screen contrast is set too low.
Digital electronics are locked because of an interruption in power.
Debris in the shelf slides.
Roll-out shelf does not slide easily (PC2200i and
PC3200i).
Door does not open easily
(PC100i and floor models).
Shelf slides are not lubricated.
Drawer does not slide easily
(PC4200i).
Shelf slide is faulty.
Debris in slots of drawer guides.
Drawers are not lubricated.
Debris in hinges.
Door hinges are not lubricated.
Hinge cam is faulty.
Hood does not open easily
(PC900i and PC1200i).
Platelet incubator does not roll easily (floor models).
Platelet incubator is connected to AC power, and doors are closed, but built-in platelet agitator is not moving (PC4200i).
Agitation is noisy or rough
(PC4200i).
Damping cylinder or hood hinge is faulty.
One or more casters are locked.
Platelet agitator is obstructed.
Component is faulty or internal connections are loose.
Main power switch is on, but nothing is working.
Platelet incubator is not level.
Moving parts are grinding against each other.
Component is faulty or internal connections are loose.
Outlet connection is faulty.
Power cord is faulty.
Action
Change the screen contrast.
Reset the monitoring system.
Pull the shelf out and confirm the slides are free of debris. Clean if necessary.
Using a lightweight oil, lubricate the bearings in the slides.
Contact a qualified service technician.
Pull the drawer out and confirm the slots are free of debris. Clean if necessary.
Clean and lubricate the drawer slots.
Confirm the hinges are free of debris. Clean the hinges if necessary.
Using a general-purpose grease, lubricate the pivots in the hinges.
Contact a qualified service technician.
Contact a qualified service technician.
Unlock the casters. Push the platelet incubator from the sides, not from the front or back.
Verify the platelet agitator is free of obstructions.
Contact a qualified service technician.
Verify the platelet incubator has been set up appropriately.
Verify that moving parts are lubricated and in good condition. Lubricate or replace if necessary.
Contact a qualified service technician.
Power connector is faulty (all models except 115 V floor models).
Verify power at the outlet. Repair original outlet or connect to a different outlet if necessary.
• Check condition of the power cord. Replace if in poor condition.
• Check that voltage through the cord is appropriate. If not, replace cord.
• Check that connections between power cord and power connector are secure.
• Check that voltage through power connector is appropriate. If not, replace power connector.
Confirm that all circuit breakers are seated. Push circuit breaker to reset if necessary.
Circuit breaker was tripped (230 V countertop models, all 100/115 V models).
One or both fuses have blown (230
V floor models).
Power line filter is faulty (230 V floor models).
Power switch is faulty.
Terminal board is faulty.
Verify fuse has not blown. Replace fuse if necessary.
Verify power output from the filter. Replace power line filter if necessary.
Verify power output from the switch. Replace switch if necessary.
Confirm terminal board is supplying power to other components. Replace terminal board if necessary.
360096-1/P
40
Helmer Scientific Platelet Incubator Service Manual
Problem Possible Cause Action
Main power switch is on, but temperature controller is not working.
Main power switch is on, but compressor is not running.
Connection between terminal board and temperature controller is loose or a part is faulty.
Connection between terminal board and the compressor is loose or a part is faulty.
• Check connections between terminal board and temperature controller.
Secure connections if necessary.
•
Confirm terminal board is supplying power to temperature controller.
Replace temperature controller or terminal board if necessary.
• Check connections between the terminal board and compressor.
Secure connections if necessary.
•
Confirm terminal board is supplying power to compressor by checking whether condensing unit fan is working.
•
Replace terminal strip if necessary (floor models only).
• Replace starting relay coil, compressor/condensing unit, or terminal board if necessary.
Main power switch is on, but chart recorder is not working.
Main power switch is on, and the doors are closed, but built-in platelet agitator is not agitating (PC4200i).
Connection between terminal board and chart recorder is loose or a part is faulty.
Doors are not closing completely.
• Check connections between terminal board and chart recorder. Secure connections if necessary.
•
Confirm terminal board is supplying power to chart recorder.
• Replace chart recorder or terminal board if necessary. Refer to the temperature chart recorder operation and service manual.
•
Confirm hinge cams are not damaged. Replace if necessary.
•
Confirm door gaskets are sealing completely. Replace door gasket if necessary.
Main power switch is on, and door/hood/doors is closed, but platelet agitators installed inside platelet incubator are not agitating (all models except
PC4200i).
One or both door switches are faulty.
Secondary circuit breaker has tripped (100/115 V models).
Agitation motor is faulty.
Terminal board is faulty.
Door(s) are not closing completely
(PC100i and floor models only).
Hood is not closing completely
(PC900i and PC1200i).
Check connections for door switches. Replace one or both door switches if necessary.
Confirm that all circuit breakers are seated. Push circuit breaker to reset it if necessary.
Check connections for agitation motor. Replace motor if necessary. On
230 V models, replace motor capacitor if necessary.
•
Confirm terminal board is supplying power to door switch(es). Replace terminal strip if necessary (floor models only).
• Replace terminal board if necessary.
Confirm hinge cams (floor models) or hinge bearings (PC100i) are not damaged.
•
Confirm damping cylinder is not damaged. Replace it if necessary.
• On chamber wall in the hinge area, check nutsert for looseness.
Replace nutsert if necessary.
Main power switch is on, and door/hood/doors is closed, but platelet agitators installed inside platelet incubator are not agitating (all models except
PC4200i).
Main power switch is on, but one or both chamber fans are not running.
One or both door switches are faulty.
Terminal board is faulty.
Secondary circuit breaker has tripped (100/115 V models).
Chamber fan is faulty (PC100i).
One or both chamber fan motors are faulty (all models except
PC100i).
Check connections for door switch(es). Replace door switch(es) if necessary.
Confirm terminal board is supplying power to door switch(es). Replace terminal strip if necessary (floor models only). Replace terminal board if necessary.
Confirm that all circuit breakers are seated. Push circuit breaker to reset it if necessary.
One or more interior outlets are faulty.
Platelet agitator component is faulty or internal connections are loose.
Check connections for interior outlets. Replace outlets as necessary.
Refer to the platelet agitator service manual.
Check connections for the fan. Replace fan if necessary.
Check connections for the fan. Replace fan motor if necessary.
Slide is faulty.
Confirm slide is operating correctly. Replace slide if necessary.
Roll-out shelf does not slide easily (PC2200i and
PC3200i).
Door does not open easily
(PC100i and floor models).
Hinge cam is faulty (floor models only).
Confirm hinge cam is not damaged. Replace if necessary.
Hood does not open or close easily or slams shut instead of gradually closing
(PC900i and PC1200i).
Hinge bearing is faulty (PC100i).
Damping cylinder is faulty.
Hinge hardware is loose or faulty.
Confirm bearings are not damaged. Replace if necessary.
Replace damping cylinder.
On chamber wall, check nutsert for looseness. Replace nutsert if necessary.
360096-1/P
41
Helmer Scientific Platelet Incubator Service Manual
Problem
Agitation is noisy or rough
(PC4200i).
Agitation is noisy or rough
(PC4200i).
Agitation of installed platelet agitators is noisy or rough (all models except
PC4200i).
Possible Cause Action
One or more wheels are worn.
Bearing on drive rod assembly is grinding or worn.
Check hole through each wheel for wear. If a wheel is out-of-round, replace wheel or wheel assembly.
Drive rod is loose or disconnected.
Check connections from drive rod to drive plate on the storage frame, and to the crank block on the motor. Secure connections if necessary.
On drive rod assembly, check bearing on end connected to the drive plate.
Lubricate bearing by injecting grease through grease fitting. If bearing is worn, replace drive rod assembly.
Crank block on the motor is worn.
Motor is faulty.
One or more feet on platelet incubator are worn (countertop models).
Check crank block on the motor for wear. Replace crank block if worn.
Replace motor.
Check feet for wear. Replace feet if necessary.
Platelet agitator component is faulty or internal connections are loose.
Refer to the platelet agitator service manual.
6.2 Chamber Temperature Problems
Problem
Chamber temperature displayed is higher or lower than actual temperature.
Possible Cause
Monitor is not calibrated.
Action
Confirm upper temperature probe is reading correctly. Calibrate probe if necessary.
Reset monitoring system.
Digital electronics are locked because of an interruption in power.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Chamber temperature does not stabilize at platelet incubator setpoint.
Condenser grill is dirty.
Air circulation at top of chamber is not adequate.
Ambient air temperature around platelet incubator is too high.
Refrigerant level is too low.
Component is faulty or internal connections are loose.
Check condenser grill. Clean if necessary.
Check for items that may obstruct air flow and remove necessary.
Confirm platelet incubator is placed appropriately.
Contact a qualified service technician.
Contact a qualified service technician.
6.3 Alarm Activation Problems
Problem
Platelet incubator is in an alarm condition, but alarms are not audible.
All alarms are activating sporadically.
No Battery alarm is activating sporadically.
Possible Cause Action
Audible alarms have been muted.
Verify audible alarms are not muted. If time remaining on Mute timer is greater than five minutes, change Mute timer value to five minutes and wait until timer resets.
Check volume and change if necessary.
Volume is set too low.
Alarm is not communicated audibly, visually, or through remote alarm interface.
Audible alarm time delay has not elapsed.
Verify whether active alarm should be communicated audibly.
Allow the alarm time delay to elapse or change the alarm delay.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Battery voltage level on backup batteries for monitoring system is low.
Replace backup batteries for monitoring system.
360096-1/P
42
Helmer Scientific Platelet Incubator Service Manual
Problem Possible Cause Action
Platelet incubator meets a high or low temperature alarm condition, but appropriate alarm is not active.
Active temperature alarm, but platelet incubator does not meet high or low temperature condition.
Alarm setpoint was changed.
Component is faulty or internal connections are loose.
Monitor is not calibrated.
Check current setpoints for alarms. Change setpoints if necessary.
Contact a qualified service technician.
Confirm upper temperature probe is reading within appropriate tolerance.
Calibrate probe if necessary.
High or low temperature alarm activated, but temperature recorded at activation does not match alarm setpoint.
Platelet incubator is not receiving AC power, but power failure alarm is not active.
Monitor is not calibrated.
Temperature changed slightly around time of activation.
Alarm setpoint was changed.
Component is faulty or internal connections are loose.
Confirm upper temperature probe is reading within appropriate tolerance.
Calibrate probe if necessary.
No action needed.
Check current setpoints for alarms. Change setpoints if necessary.
Contact a qualified service technician.
Condenser alarm is active.
Platelet incubator is connected to power, but
AC Power Failure alarm is active.
Compressor is overheating due to lack of airflow.
• Check condenser grill. Clean if necessary.
•
Confirm platelet incubator is placed appropriately.
Confirm alarm setpoint is set at appropriate value.
Condenser alarm setpoint is too low.
Refrigerant level is too low.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Contact a qualified service technician.
Condenser probe is not calibrated.
Contact a qualified service technician.
Circuit breakers have tripped or are not seated correctly (115 V floor models, all countertop models).
Confirm circuit breakers are seated. Push circuit breaker(s) to reset if necessary. If breaker(s) cannot be reset or trip again, contact a qualified service technician.
Fuses have blown (230 V floor models).
Contact a qualified service technician.
Door Open alarm is activating sporadically.
Motion alarm is active, but platelet agitator is moving.
No AC power to platelet incubator.
Verify power cord is connected securely to platelet incubator and outlet.
Tighten connections if necessary.
Component is faulty or internal connections are loose.
Verify outlet is operational and meets power requirements for platelet incubator. Repair original outlet or connect to a different outlet if necessary.
Contact a qualified service technician.
Check current setpoint for Door Ajar alarm. Change setpoint if necessary.
Door Ajar Timeout is set to
0, causing alarm to activate immediately when door is opened.
Doors are not closing completely.
Doors are closing but not sealing completely.
Component is faulty or internal connections are loose.
Confirm doors are aligned. Contact a qualified service technician if necessary.
Confirm door gasket seals completely. Contact a qualified service technician if necessary.
Contact a qualified service technician.
A Horizon Series platelet agitator is installed, but alarm for the corresponding position was not disabled.
i.Center monitoring system is not receiving motion information from the platelet agitator (models without built-in platelet agitator).
Component is faulty or internal connections are loose.
Disable motion alarm for that position.
On platelet incubator’s monitoring system, verify agitation information is being received.
Verify data cable is connected securely to platelet agitator and corresponding port in the platelet incubator. Tighten connections if necessary.
Contact a qualified service technician.
360096-1/P
43
Helmer Scientific Platelet Incubator Service Manual
Problem
Agitation has stopped, but motion alarm is not active.
Motion alarm is activating sporadically while platelet agitator is moving.
Motion alarm is active, but there is no platelet agitator installed at that position
(models without a built-in platelet agitator).
Motion alarm is activating immediately when agitation is started (PC4200i).
Possible Cause Action
Motion alarm for position where platelet agitator is installed has been disabled.
Platelet agitator alarm setpoint value has not been reached.
Enable motion alarm.
Wait until setpoint has been reached. Alarm should activate.
Delay period for motion alarm is set too low (PC4200i).
Verify delay period control is set to a value greater than platelet agitator alarm setpoint.
Motion alarm system is not receiving power.
• Check that there is AC power to platelet incubator.
• Test backup battery. If battery is not functioning, replace battery.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Alarm system is not receiving power consistently (PC4200i).
• Check that there is AC power to platelet agitator.
• Test backup battery. If battery is not functioning or has been in use for one year, replace battery.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Alarm is enabled for that position when it should be disabled.
Disable motion alarm for that position.
Delay period is set to 0 minutes, which activates alarm before motion can be sensed.
Component is faulty or internal connections are loose.
Change delay period to a value greater than 0 minutes.
Contact a qualified service technician.
6.4 Testing Problems
Problem Possible Cause
Automatic temperature tests do not work.
High Alarm setpoint is set significantly higher than default value, or Low Alarm setpoint is set significantly lower than default value.
Component is faulty or internal connections are loose.
Connections between upper chamber temperature probe and control board are loose or a part is faulty.
Action
•
Confirm alarm setpoints are set at appropriate values.
• Test temperature alarms manually.
Contact a qualified service technician.
Test probe connections. Secure connections if necessary. Replace probe if necessary. Replace parts with control and display board kit, if necessary.
360096-1/P
44
Helmer Scientific Platelet Incubator Service Manual
6.5 Condensation Problems
Problem
Excessive water in water evaporation tray.
Excessive water in chamber.
Excessive humidity on doors.
Excessive water in water evaporation tray (countertop models).
Possible Cause Action
Humid air is entering chamber.
Confirm platelet incubator is level and doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Heater in evaporation tray is faulty.
Confirm heater is hot. Contact a qualified service technician to confirm heater is drawing the appropriate current.
Humid air is entering chamber.
Humid air is entering chamber.
Relative humidity around the platelet incubator is too high.
Confirm platelet incubator is level and doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Confirm platelet incubator is level, and the doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Confirm platelet incubator is placed appropriately.
Connections between terminal board and condensate evaporator are loose or a part is faulty.
• Check connections between terminal board and condensate evaporator
(heater). Secure connections if necessary.
•
Confirm condensate evaporator is working correctly. It should be hot and drawing appropriate current:
• For 100/115 V models, current should be approximately 0.43 A to 0.55 A.
• For 230 V models, current should be approximately 0.21 A to 0.35 A.
• Replace condensate evaporator or terminal board if necessary.
360096-1/P
45
Helmer Scientific Platelet Incubator
7 i.Series Parts
Service Manual
Note
• Before replacing parts, protect items in platelet incubator from extended exposure to adverse temperature.
• Allow platelet incubator temperature to stabilize at setpoint after replacing parts or after extended door opening.
A
D
B
H
J
I
C
Front parts (multiple models shown).
Label Description
C
D
E
F
G
A
B
Chart recorder
Chart recorder backup battery
Chart paper
Display assembly
LCD Board
Touchpad
Interface cable
* PC100i only
E
K
F
G
L
Part Number Label Description
400409
120218
220273
400509-1
12045
320722-1
400502-3*
400502-2
I
H
J
K
L
Main power switch
Temperature controller
(countertop)
Temperature controller (floor model)
Battery backup key switch, 2-way
Caster (swivel with brake; floor models)
Foot
(countertop models)
Part Number
120478
400600-1
400601-1
120227
220261
220432
360096-1/P
46
A
B
Helmer Scientific Platelet Incubator
Door and Hinge
C
D
Door and Hinge (PC100i)
Label Description
A
B
C
Hinge bearing
Upper hinge bracket
Lower hinge bracket
D
E
Not shown
Door seal kit
Door handle
Door lock
F Door brush seal
Door brush seal
Hood
Service Manual
F
G
E
H
K
L
Model
100
100
100
2200
3200/4200
Door and Hinge (PC2200i, PC3200i, and PC4200i)
Part Number Label Description
220295
320846-1
310846-2 I
G
H
Door handle
Door lock
Upper hinge bracket
(left)
400785-1
320865-1
220374
320920-2
320920-1
J
K
L
Upper hinge bracket
Door seal kit
Hinge cam
(right)
Lower hinge bracket
(left)
Lower hinge bracket
(right)
Model Part Number
320865-1
220374
2200/3200/4200 400647-2
2200/3200/4200 400647-1
2200 400787-1
3200/4200 400788-1
320742-1
2200/3200/4200 400647-1
2200/3200/4200 400647-2
I
J
A
B
C
(PC900i and PC1200i)
Label Description
A Hood assembly
(Excludes damping cylinder and door lock)
Model
900
Part Number
800098-1
Label
C
Description
Hood lock
Not shown
Hood seal kit
B Damping cylinder
1200 800099-1
220210
Model Part Number
220272
900 400786-1
1200 400786-2
360096-1/P
47
Helmer Scientific Platelet Incubator
Top
Service Manual
E
F
G
H
I
PC100i
A
B
C
D
PC2200i, PC3200i and PC4200i
PC900i and PC1200i
B
C
Label
A
D
Description
Compressor/condensing unit
Monitor backup battery
(six required)
Terminal board
Control board
(included in Display assembly)
Model
100/900/
1200
Volts
115
230
115/230
100
Part
Number
400668-1 E
400668-2 F
715031
800285-1
800292-1 I
Label Description
G
H
Power supply board
Motion alarm battery
Power line filter
Agitator board
Condenser probe
400650-1 Not
Shown
Compressor relay
Model
4200
Volts
230
Part
Number
400633-1
120232
120400
400516-1
400674-1
100/115 120528
230 120523
360096-1/P
48
Rear
Helmer Scientific Platelet Incubator Service Manual
I
A
D
E
F
G H
J
K
C
B
PC100i, PC900i, and PC1200i PC2200i, PC3200i, and PC4200i
C
D
E
Label
A
B
Description
Power cord
Model all models
2200/3200
Volts
100/900/1200 100/115 120630
230
115
Part
Number
120156
120130
4200 115 110040
Condensate evaporator kit
(includes tray)
100/115 400791-1
230 400791-2 I
H
Label Description
F Power connector (includes power line filter)
Model
G Circuit breaker
(secondary, 2A)
Circuit breaker
(secondary, 5A)
2200/3200
4200
Volts
115
115
Circuit breaker
(primary, 15A)
Circuit breaker
(primary, 20A)
Motion alarm control board
2200/3200 115
4200
4200
115
Circuit breaker
RS-232 port
Alarm buzzer
(6 A)
(optional)
230 120288
400705-2
120457
J
K
Fuse
(10 A)
Power connector with fuse holder
230
230
Part
Number
120299
120252
120514
120259
120191
400523-1
120174
120379
360096-1/P
49
Helmer Scientific Platelet Incubator
12.4 Interior
A B C D E
F
I
J
K
Service Manual
M N
PC100i
G
H
D
L
PC900i and PC1200i
O
PC2200i and PC3200i
F
G
H
C
D
E
Label
A
B
Description
Chamber fan guard
Chamber fan
Alarm check tube
Control and chart recorder probe
Door switch
Agitator port board
Interior outlet
Interior outlet
Volts
230
Part Number Label Description
220122
100/115 120009
120010
220299
120315
120490
400522-1
100/115 120499
230 120097
I
J
K
L
M
N
O
Chamber fan blade
Chamber fan motor
Chamber heater
Temperature monitoring probe
Shelf slide
Roll-out shelf
Lower chamber probe
(PC3200i)
Volts Part Number
220177
100/115 120170
230 120171
100/115 120313
230 120314
400510-1
400717-1
320971-1
400513-1
360096-1/P
50
Helmer Scientific Platelet Incubator
Agitation Parts (PC4200i)
A
B
G
E
F
C
D
Service Manual
Label Description
A
B
C
Drawer stop
Drawer guide
(Top left)
Drawer guide
(Top right)
Drawer guide
(Bottom left)
Drawer guide
(Bottom right)
Wheel (included in wheel assembly)
Base (PC4200)
Part Number Label Description
400174-1
320540-1
320540-2
320541-1
320541-2
320566-1
D
E
F
E
Wheel track
Drive plate
Part Number
320215-1
321005-1
Wheel assembly
(includes two wheels, axle, mounting bracket)
400159-1
Drawer assembly
(includes label holder kit.)
* = (silver drawer - standard)
** = (red drawer)
*** = (black drawer)
400616-1*
400617-1**
400615-1***
D
E
F
H
A B C
G
B
C
D
Label Description
A Agitation motor
(includes capacitor)
* for serial numbers 968995 and lower
** for serial numbers 968996 and greater
Crank block
Motor mount
Proximity sensor
Volts Part Number Label Description
115 400305-1 E Drive rod assembly
230 400306-1*
230 400638-1**
115 320222-1
F
G
Proximity actuator
Compressor/condensing unit
* = units with serial # 2046319 above
** = units with serial # 2046318 and prior
230 321032-1
220079
400694-1
H
Not
Shown
Condenser probe
Starting relay coil
Model Volts Part Number
400111-1
2200/3200/
4200
120125
115 400669-1
230 800977-1*
230 800978-1**
400674-1
115 120524
230 120525
360096-1/P
51
Helmer Scientific Platelet Incubator
Accessories
A
B
Service Manual
Label
A
B
Not shown
Description
Data Cable
Fan airflow deflector
Power cord, short
(For all models except PC4200)
Model
100
100/900/1200/
2200/3200
Volts Part Number Label Description
320842-1 Not shown
Shelf for platelet rotator
320967-1
100/115 120316
230 120317
Label holder kit
Bypass plug
Model Part Number
900 400618-1
1200 400619-1
4200 400620-1*
400657-1
* Label holder kit if for PC4200h and includes label, label holder, hardware and installation tool. Also included in drawer assembly
360096-1/P
52
Helmer Scientific Platelet Incubator
8 Schematics
8.1 PC100i, PC900i, PC1200i (100/115 V)
Service Manual
360096-1/P
53
Helmer Scientific Platelet Incubator
8.2 PC100i, PC900i, PC1200i (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(AK) POWER
CONNECTOR
WITH POWER
LINE FILTER
N L
(AD) CHART
RECORDER
BACKUP
BATTERY
(P) CIRCUIT
BREAKERS optional feature
12 V dc to J2 on
CONTROL
BOARD
1
1 J9
(F)
TERMINAL
BOARD
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
(AO)
INTERIOR
OUTLET
1 J1 from i.Series
Agitator
J3 1
(AJ) AGITATOR
PORT BOARD
(AL) CHAMBER FAN
(I) DOOR SWITCH
REMOTE ALARM
INTERFACE
COM
NC
NO
(K) DISPLAY
ASSEMBLY
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
(AA)
INTERFACE
CABLE
J9
1
J1
(E) CONTROL
BOARD
J12
1
J15
(D) ALARM
BUZZER
J11
1
1
J8
J2
1
J6
1
1
J4
-
+
(AE) MONITOR
BACKUP
BATTERY
(Z) POWER
SUPPLY BOARD
J10
-
+
12 V dc
J5
1
P1
(W) KEY SWITCH optional feature
(AQ) RS232 PORT
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
(AC) CONDENSER
PROBE
(AB) UPPER
CHAMBER PROBE
THERMOSTAT (C) CHAMBER
HEATER
(AP)
STARTING
RELAY COIL
(A) COMPRESSOR/
CONDENSING UNIT
(J) CONDENSATE
EVAPORATOR
360096-1/P
54
Helmer Scientific Platelet Incubator
8.3 PC2200i (115 V) optional feature
MAIN POWER
115 (±10%) V ac 60 Hz
L N
POWER
CORD
(P) CIRCUIT
BREAKER
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J2 on
CONTROL
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
(AN)
INTERIOR
OUTLETS
THERMOSTAT CHAMBER
HEATER
1 from i.Series
Agitator J3
J1
1
(H) CHAMBER
FAN MOTOR
(AM) CIRCUIT
BREAKER
1 J1 from i.Series
Agitator J3 1
(AJ) AGITATOR
PORT BOARDS
LEFT RIGHT
(I) DOOR SWITCHES
(H) CHAMBER
FAN MOTOR
REMOTE ALARM
INTERFACE
COM
NC
NO
J9
1
-
+
(D) ALARM
BUZZER
J8
J11
1
1
(AE) MONITOR
BACKUP
BATTERY
J10
-
+
12 V dc
(W) KEY SWITCH
1
J1
(E) CONTROL
BOARD
(K) DISPLAY
ASSEMBLY
(AA) INTERFACE
CABLE
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
J12 optional feature
(AQ) RS232 PORT
1
J15
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
J2
J6
1
1
J4
(AC) CONDENSER
PROBE
(Z) POWER
SUPPLY BOARD
J5
1
(AB) UPPER
CHAMBER PROBE
P1
TERMINAL
STRIP
L
N
(AP)
STARTING
RELAY COIL
(A) COMPRESSOR/
CONDENSING UNIT
360096-1/P
55
Helmer Scientific Platelet Incubator
8.4 PC2200i (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N
(Y) POWER
LINE FILTER
IN
OUT
L
Service Manual optional feature
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J2 on
CONTROL
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
(AO)
INTERIOR
OUTLETS
THERMOSTAT CHAMBER
HEATER
TERMINAL
STRIP
J1
1 from i.Series
Agitator
J3
1
(H) CHAMBER
FAN MOTOR
L
1 from i.Series
Agitator
J3
J1
1
(AJ) AGITATOR
PORT BOARDS
REMOTE ALARM
INTERFACE
COM
NC
NO
J9
1
(D) ALARM
BUZZER
-
+
(AE) MONITOR
BACKUP
BATTERY
J10
-
+
12 V dc
J8
J11
1
1
(W) KEY SWITCH
LEFT RIGHT
(I) DOOR SWITCHES
(H) CHAMBER
FAN MOTOR
1
J1
(E) CONTROL
BOARD
(K) DISPLAY
ASSEMBLY
(AA) INTERFACE
CABLE
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
J12 optional feature
(AQ) RS232 PORT
1
J15
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
J2
J6
1
1
J4
(AC) CONDENSER
PROBE
(Z) POWER
SUPPLY BOARD
J5
1 (AB) UPPER
CHAMBER PROBE
P1
N
(AP)
STARTING
RELAY COIL
(A) COMPRESSOR/
CONDENSING UNIT
360096-1/P
56
Helmer Scientific Platelet Incubator
8.5 PC3200i (115 V)
Service Manual optional feature
MAIN POWER
115 (±10%) V ac 60 Hz
L N
POWER
CORD
(P) CIRCUIT
BREAKER
1
J1
(AD) CHART
RECORDER
BACKUP
BATTERY
(X)
CHART
RECORDER
12 V dc to J2 on
CONTROL
BOARD
1
1 J9
(F)
TERMINAL
BOARD
(AN) INTERIOR
OUTLETS
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
THERMOSTAT from i.Series
Agitator
J3
1
1
J1
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
(H) CHAMBER
FAN MOTOR
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
L from i.Series
Agitator
J3
1
(AM) CIRCUIT
BREAKER
(H) CHAMBER
FAN MOTOR
N
1 J1
LEFT RIGHT
(I) DOOR SWITCHES from i.Series
Agitator
J3 1
(AJ) AGITATOR
PORT BOARDS
(AP)
STARTING
RELAY COIL
REMOTE ALARM
INTERFACE
COM
NC
NO
1
J1
(K) DISPLAY
ASSEMBLY
(AA) INTERFACE
CABLE
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
J12 optional feature
(AQ) RS232 PORT
(A) COMPRESSOR/
CONDENSING UNIT
J9
(E) CONTROL
BOARD
FLASH CARD
PORT
1
J15
(D) ALARM
BUZZER
J8
J11
1
1
J2
1
J6
1
1
J4 to J7 on
TERMINAL
BOARD
-
+
(AE) MONITOR
BACKUP
BATTERY
J10
-
+
12 V dc
(Z) POWER
SUPPLY BOARD
J5
1
(AC) CONDENSER
PROBE
(N) LOWER
CHAMBER PROBE
(AB) UPPER
CHAMBER PROBE
P1
(W) KEY SWITCH
360096-1/P
57
Helmer Scientific Platelet Incubator
8.6 PC3200i (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N
(Y) POWER
LINE FILTER
1 from i.Series
Agitator J3 1
IN
OUT
J1
L
Service Manual optional feature
(AD) CHART
RECORDER
BACKUP
BATTERY
(X)
CHART
RECORDER
12 V dc to J2 on
CONTROL
BOARD
(F)
TERMINAL
BOARD
(AO) INTERIOR
OUTLETS
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
THERMOSTAT
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
1 J1
(H) CHAMBER
FAN MOTOR
L from i.Series
Agitator J3 1
(H) CHAMBER
FAN MOTOR
1
J1
N
LEFT RIGHT
(I) DOOR SWITCHES from i.Series
Agitator
J3
1
(AJ) AGITATOR
PORT BOARDS
(AP)
STARTING
RELAY COIL
1
J1
(K) DISPLAY
ASSEMBLY
(AA) INTERFACE
CABLE
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
J12 optional feature
(AQ) RS232 PORT
(A) COMPRESSOR/
CONDENSING UNIT
REMOTE ALARM
INTERFACE
COM
NC
NO
J9
1
(D) ALARM
BUZZER
-
+
(AE) MONITOR
BACKUP
BATTERY
J10
-
+
12 V dc
J8
J11
1
1
(W) KEY SWITCH
(E) CONTROL
BOARD
J2
(Z) POWER
SUPPLY BOARD
1
P1
J15
J6
1
1
J4
J5
1
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
(AC) CONDENSER
PROBE
(N) LOWER
CHAMBER PROBE
(AB) UPPER
CHAMBER PROBE
360096-1/P
58
Helmer Scientific Platelet Incubator
8.7 PC4200i (115 V) optional feature
MAIN POWER
115 (±10%) V ac 60 Hz
L N
POWER
CORD
(P) CIRCUIT
BREAKER
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J2 on
CONTROL
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
J3
(S)
AGITATOR
BOARD
(T)
BATTERY
J7
(O)
PROXIMITY
SENSOR
THERMOSTAT CHAMBER
HEATER
(H) CHAMBER
FAN MOTOR
TERMINAL
STRIP
L
(AM) CIRCUIT
BREAKER
(H) CHAMBER
FAN MOTOR
N
Voltage selection switch
L
N
L
N
J2
1
J1
ALARM
VOLUME
ALARM
DELAY
P1
(V) MOTION
ALARM CONTROL
BOARD
REMOTE ALARM
INTERFACE
COM
NC
NO
(K) DISPLAY
ASSEMBLY
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
(AA)
INTERFACE
CABLE
J9
1
J1
(E) CONTROL
BOARD
J12
1
J15
J8 1
(D) ALARM
BUZZER J2
J11
1
1
J6
1
1
J4
-
+
(AE) MONITOR
BACKUP
BATTERY
(Z) POWER
SUPPLY BOARD
J10
-
+
12 V dc
J5
1
P1
(W) KEY SWITCH
LEFT RIGHT
(I) DOOR SWITCHES optional feature
(AQ) RS232 PORT
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
(AC) CONDENSER
PROBE
(N) LOWER
CHAMBER PROBE
(AB) UPPER
CHAMBER PROBE
(AP)
STARTING
RELAY COIL
(A) COMPRESSOR/
CONDENSING UNIT
M
(M) AGITATION
MOTOR
360096-1/P
59
Helmer Scientific Platelet Incubator
8.8 PC4200i (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N
(Y) POWER
LINE FILTER
IN
OUT
L
Service Manual optional feature
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J2 on
CONTROL
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
J3
(S)
AGITATOR
BOARD
(T)
BATTERY
J7
(O)
PROXIMITY
SENSOR
THERMOSTAT CHAMBER
HEATER
(H) CHAMBER
FAN MOTOR
TERMINAL
STRIP
L
Voltage selection switch
L
N
L
N
J2
1 J1
ALARM
VOLUME
ALARM
DELAY
P1
(V) MOTION
ALARM CONTROL
BOARD
REMOTE ALARM
INTERFACE
COM
NC
NO
(K) DISPLAY
ASSEMBLY
(AA)
INTERFACE
1
CABLE
J9
DISPLAY
BOARD
J4
J1 J2 J3
LCD
BOARD
TOUCHPAD
1
J1
(E) CONTROL
BOARD
J12
J15
J8 1
(D) ALARM
BUZZER J2
J11
1
1
J6
1
1
J4
-
+
(AE) MONITOR
BACKUP
BATTERY
(Z) POWER
SUPPLY BOARD
J10
-
+
12 V dc
J5
1
P1
(W) KEY SWITCH
(H) CHAMBER
FAN MOTOR
LEFT RIGHT
(I) DOOR SWITCHES optional feature
(AQ) RS232 PORT
FLASH CARD
PORT to J7 on
TERMINAL
BOARD
(AC) CONDENSER
PROBE
(N) LOWER
CHAMBER PROBE
(AB) UPPER
CHAMBER PROBE
(AP)
STARTING
RELAY COIL
(A) COMPRESSOR/
CONDENSING UNIT
CAPACITOR
M
(M) AGITATION
MOTOR
N
360096-1/P
60
Helmer Scientific Platelet Incubator
9 i.Center Screen Reference
HOME screen
MAIN button
MAIN screen
MUTE button (changes mute timer)
MAIN screen
Event Log
option
(Press button)
EVENT LOG screen
EVENT LOG DETAIL screen
System Alarm Test & Status option
SYSTEM ALARM TEST & STATUS screen
Start High Alarm Auto Test
option
Start Low Alarm Auto Test option
Cancel High or Low Test
option
Chart Paper Days Left or Chart Paper Timer display
Door Status display
Condenser Temp display
Edit Configuration
option
(Enter the password)
CONFIGURATION screen
Set Date & Time option
SET DATE & TIME screen
System Options option
SYSTEM OPTIONS screen
Language
option
Alarm Volume option
Alarm Pulse option
Temperature Units option
Temperature Graph option
Chart Paper Timer option
Alarm Setpoints option
SET ALARM SETPOINT screen
High Alarm Setpoint
option
Low Alarm Setpoint option
Cond. Alarm Setpoint option
Door Ajar Timeout option
Power Failure Timeout option
Agitator Alarm Setpoints
option
AGITATOR ALARM SETPOINT screen
Agitator 1 Setpoint
option
Agitator 1
option
Agitator 2 Setpoint
option
Agitator 2
option
Agitator 3 Setpoint
option
Agitator 3
option
Temperature Calibration option
TEMPERATURE CALIBRATION screen
Select Temp Probe option or Upper Temperature Probe display
Temperature option
Factory Default Settings
option
FACTORY DEFAULT SETTINGS screen
Change Password
option
Service Manual
360096-1/P
61
Helmer Scientific Platelet Incubator
View Configuration
option
VIEW CONFIGURATION screen
Clock Mode display
Date Format display
Door Ajar Timeout display
Pwr Failure Timeout display
High Alarm Setpoint
display
Low Alarm Setpoint display
Cond. Alarm Setpoint display
Alarm Volume display
Alarm Pulse display
Chart Paper Days Left or Chart Paper Timer display
Agitator Alarms
display
Temperature Graph display
Product/Company Information option
INFORMATION screen i.Help
option i.Help
screen
Agitator Information
option
AGITATOR INFORMATION screen
Agitator RPM
option
AGITATOR RPM screen
Agitator Cycles
option
AGITATOR TOTAL CYCLE COUNT screen
Reset option
RESET AGITATOR screen
Agitator Count Logs
option
AGITATOR COUNT LOGS screen
Service Manual
360096-1/P
62
Helmer Scientific Platelet Incubator Service Manual
Horizon Series™ Models
10 Installation and Configuration
10.1 Install Battery for Backup Power
The monitoring and control system and the chart recorder each have a battery system, enabling a period of continuous operation if power is lost.
Battery life varies by manufacturer as well as voltage level remaining. The power failure backup battery provides power to the power failure alarm in the event of an AC power failure. Backup battery power for the chart recorder is available for up to 14 hours.
CAUTION
Before installing or replacing the battery, switch the power OFF . Disconnect the platelet incubator from AC power.
NOTICE
Use battery meeting manufacturer’s specifications.
The battery is located behind the control door.
Battery is disconnected for shipping. Reconnect the battery to provide power failure alarm with backup power in the event of AC power failure.
10.2 Platelet Agitator and Platelet Incubator Compatibility i.Series countertop platelet agitators (PF15i, PF48i, and PF96i) have a built-in motion alarm system and may be placed in either i.Series or Horizon Series platelet incubators. When agitation motion is not detected, the alarm sounds.
An i.Series platelet agitator placed in a Horizon Series platelet incubator cannot be monitored by the platelet incubator. If the agitation motion stops, the alarm on the i.Series platelet agitator sounds inside the platelet incubator, so the sound may not be apparent. An extension to the alarm system is desirable for this configuration.
Note
Horizon Series platelet incubators cannot monitor i.Series platelet agitators.
Platelet Incubator
PC100h
PC900h
PC1200h, PC2200h,
PC3200h
Compatible Helmer
Platelet Agitator
PF15h
PF48h
PF96h
360096-1/P
63
Helmer Scientific Platelet Incubator
10.3 Install a Platelet Agitator in the Platelet Incubator
Note
This chapter is not applicable to the PC4200h platelet incubator, which has a built-in platelet agitator.
Service Manual
PC900h platelet incubator with a PF48h platelet agitator installed.
Helmer platelet agitators can be installed inside Helmer platelet incubators. PC2200h and PC3200h platelet incubators hold one agitator per shelf. Each shelf features a corresponding power connector.
When installing a platelet agitator on a shelf, use the power connector for that shelf to prevent the power cord from obstructing the agitation motion.
Horizon Series platelet agitators (PF15h, PF48h, and PF96h) are not equipped with motion alarms, but may be placed in any
Series platelet incubator. Without motion alarms, Horizon Series platelet agitator motion cannot be detected by the AgiTrak system in i.Series platelet incubators.
Note
For additional instruction in configuring a platelet agitator in a platelet incubator, refer to the platelet agitator service manual.
10.4 External Monitoring Devices
The remote alarm interface is a relay switch with three terminals:
♦
Common (COM)
♦
Normally Open (NO)
♦
Normally Closed (NC)
Terminals are dry contacts and do not supply voltage. Interface circuit is either normally open or normally closed, depending on terminals used.
Requirements for your alarm system determine which alarm wires must connect to terminals.
CAUTION
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
The terminals on the remote alarm interface have the following maximum load capacity:
♦
0.25 A at 30 V (RMS)
♦
0.25 A at 60 V (DC)
360096-1/P
64
Helmer Scientific Platelet Incubator
Connect to Remote Alarm Interface
1. Switch the Alarm ON/OFF key switch OFF . Switch the AC ON/OFF switch OFF .
2. On the alarm panel, locate the remote alarm terminals.
Service Manual
Remote alarm interface.
3. Connect remote alarm wires to appropriate terminals, according to requirements for your alarm system.
4. Use a cable tie to relieve strain on alarm wires (as necessary).
5. Switch the AC ON/OFF switch ON . Switch the Alarm ON/OFF switch ON .
6. Press the MUTE button to disable the high temperature alarm while platelet incubator reaches operating temperature.
10.5 Configuring Drawers (PC4200h)
CAUTION
• Before moving drawers, ensure they are completely empty for safe lifting.
• Open the drawer using the drawer handle; not the label holder.
Remove / Install a Drawer
Drawer in the full open position. Drawer with the tab in front of the drawer stop.
1. Pull drawer out until it stops (when the tabs on the drawer contact the spring-loaded stops).
2. Depress the spring-loaded drawer stops.
3. Pull the drawer outward until the tabs on the drawer have passed the drawer stops.
4. Pull drawer free of the drawer slides.
5. Align drawer with the drawer slides.
6. Push drawer in until it stops (when the tabs on the drawer contact the spring-loaded stops).
7. Depress the spring-loaded drawer stops.
8. Push the drawer into the storage frame until it stops.
360096-1/P
65
Helmer Scientific Platelet Incubator
10.6 Drawer Labels
Drawer label holders may be installed on platelet agitator drawers.
Service Manual
Installed label holder.
Install label holders:
1. Insert the tabs on the label holder into the slots on the drawer.
2. Pivot the holder around the drawer handle. Align the hole on the label holder with the corresponding hole on the drawer.
3. Insert a screw through the hole in label holder and through the hole in the drawer.
4. Install the nut on the screw, on the back side of the label holder.
360096-1/P
66
Helmer Scientific Platelet Incubator Service Manual
11 Alarm Settings
11.1 Temperature and Power Failure Alarm Volume
The alarm volume control and alarm delay control are located on top of the platelet incubator under the access panel.
NOTICE
Protect items in the platelet incubator from extended exposure to adverse temperatures.
Top: Alarm volume control for temperature and power failure alarms.
Bottom: Alarm delay control for power failure alarms.
Alarm Volume
The alarm volume for high chamber temperature, low chamber temperature, and power failure is set with the alarm volume control.
The default alarm volume is set at the halfway point between low and high setting (medium). The volume can be adjusted by rotating the control from left (quietest) to right (loudest).
Change Alarm Volume:
1. Switch the AC ON/OFF switch OFF . Disconnect the platelet incubator from AC power. The power failure alarm will sound.
2. On top of the platelet incubator, remove the access panel. Locate the alarm volume control on the alarm board.
3.
Using a small, flat-head screwdriver, rotate the control to the left (counterclockwise) to decrease the alarm volume, or right
(clockwise) to increase the alarm volume
4. Reinstall the access panel on the top of the platelet incubator.
5. Reconnect the platelet incubator to AC power. Switch the AC ON/OFF switch ON .
Alarm Delay
The delay period controls the high temperature alarm and low temperature alarm. The default alarm delay is set at the halfway point between low and high setting (approximately four to five minutes). The delay period can be changed from five seconds to eight minutes. The delay period also sets the mute timer for audible high and low temperature alarms.
Note
If the delay period is set to 0 minutes, the mute button will have no effect.
Change Alarm Delay:
1. Disconnect the power failure alarm backup battery.
2. Switch the AC ON/OFF switch OFF . Disconnect the platelet incubator from AC power.
3. On top of the platelet incubator, remove the access panel. Locate the alarm delay control on the alarm board.
4.
Using a small, flat-head screwdriver, rotate the control to the left (counterclockwise) to decrease the alarm delay to
five seconds.
5. Using the screwdriver, rotate the control to the right (clockwise) to increase the alarm delay to eight minutes.
6. Reinstall the access panel on the top of the platelet incubator.
7. Reconnect the platelet incubator to AC power. Switch the AC ON/OFF switch ON .
8. Reconnect the power failure alarm backup battery.
360096-1/P
67
Helmer Scientific Platelet Incubator Service Manual
11.2 Motion Alarm (PC4200h)
Unlike other Horizon Series models, the PC4200h platelet incubator has a motion alarm delay control. When the platelet agitator motion stops, this control delays when alarm information is sent to the remote alarm interface on the platelet incubator.
Note
The delay control for the motion alarm is located on the rear of the platelet incubator.
Top: Motion alarm volume control. Bottom: Alarm delay control for the motion alarm.
The default motion alarm delay is set at the halfway point (approximately four (4) minutes). To change the motion alarm delay, use a small flat-head screwdriver. The maximum alarm delay is eight (8) minutes.
The default alarm volume is set at the middle position (medium).
Note
Do not set the alarm delay to 0 minutes.
Change Motion Alarm Delay:
Using a small, flat-head screwdriver, rotate the control to the left (counterclockwise) to shorten the motion alarm delay, or to the right (clockwise) to extend the motion alarm delay.
Change Motion Alarm Volume:
1. On back of the platelet incubator, locate the alarm volume control.
2. Position the switch to correspond with the desired motion alarm volume.
360096-1/P
68
Helmer Scientific Platelet Incubator Service Manual
11.3 Alarm Tests
Test alarms to ensure they are working correctly. The platelet incubator has alarms for chamber temperature, power failure, and agitation motion (for built-in PC4200h platelet agitator).
NOTICE
Before testing alarms, protect items in the platelet incubator from extended exposure to adverse temperature.
Chamber Temperature Alarm Test (PC100h)
NOTICE
• Perform the low alarm test before the high alarm test to control the temperature more closely and complete the testing
more quickly.
• Calibrate the chamber temperature probe prior to performing the manual chamber alarm test.
• Temperature probes are fragile; handle with care.
Notes
• To perform the manual chamber alarm test on PC100h platelet incubators, obtain the alarm check tube.
•
Do not remove the chamber probe from PC100h platelet incubators; fill the tube with 2” (51 mm) of hot or cold water as
noted in the instructions below.
•
Place the filled alarm check tube over the chamber probe to perform the test.
Test Low Alarm
1. Identify the current setting for the low alarm setpoint.
2. Remove the installed platelet agitator(s) to allow access to the to the chamber probe.
3. Fill the alarm check tube with cold water.
4. Place the alarm check tube in the holding bracket. Slide the tube upward to immerse the probe tip.
5. When the low temperature alarm sounds, note the temperature on the monitoring and control system display.
6. Compare the temperature at which the alarm sounds to the low alarm setpoint. If the values do not match, refer to Troubleshooting.
Test High Alarm
1. Identify the current setting for the high alarm setpoint.
2. Fill the alarm check tube with hot water.
3. Place the alarm check tube in the holding bracket. Slide the tube upward to immerse the probe tip.
4. When the high temperature alarm sounds, note the temperature on the monitoring and control system display.
5. Compare the temperature at which the alarm sounds to the high alarm setpoint. If the values do not match, refer to Troubleshooting.
6. Reinstall the platelet agitator in the chamber.
7. Slide the alarm check tube down so the probe is no longer immersed.
360096-1/P
69
Helmer Scientific Platelet Incubator
Chamber Temperature Alarm Test (PC900h, PC1200h, PC2200h, PC3200h, PC4200h)
Service Manual
NOTICE
• Perform the low alarm test before the high alarm test to control the temperature more closely and complete the testing
more quickly.
• Calibrate the chamber temperature probe prior to performing the manual chamber alarm test.
• Temperature probes are fragile; handle with care.
Note
Place the filled alarm check tube over the chamber probe to perform the test.
Test Low Alarm
1. Identify the current setting for the low alarm setpoint.
2. Remove the installed platelet agitator(s) to allow access to the to the chamber probe.
3. Loosen the screws securing the chamber probe bracket to the chamber wall. Slide the bracket up and away from the wall.
4.
Unscrew and remove the fitting securing the probe to the bracket.
5. Slide the probe tip upward and free the probe from the bracket.
6. Immerse the probe in cold water. Ensure the probe is at the bottom of the glass.
7. When the low temperature alarm sounds, note the temperature on the monitoring system display.
8. Compare the temperature at which the alarm sounds to the low alarm setpoint. If the values do not match, refer to Troubleshooting.
Test High Alarm
1. Identify the current setting for the high alarm setpoint.
2. Immerse the probe in hot water. Ensure the probe is at the bottom of the glass.
3. When the high temperature alarm sounds, note the temperature on the monitoring system display.
4. Compare the temperature at which the alarm sounds to the high alarm setpoint. If the values do not match, refer to Troubleshooting.
5. Remove the probe from the warm water.
6. Install the probe in the chamber probe bracket.
7.
Install the fitting to secure the probe to the probe bracket, leaving approximately 1” (25 mm) of the probe exposed beyond
the fitting.
8. Install the probe bracket in the chamber and tighten the screws.
9. Reinstall the platelet agitator(s) in the chamber.
360096-1/P
70
Helmer Scientific Platelet Incubator
Power Failure Alarm Test
During a power failure, the power failure alarm sounds and the battery provides power to the power failure alarm.
Service Manual
Test Power Failure Alarm
1. Ensure the power failure alarm backup battery is connected.
2. Behind the control door on the front of the platelet incubator, enable the temperature and power failure alarms.
♦
PC4200h: Switch the alarm key switch to position 2 to enable the audible temperature and power failure alarms but disable the motion alarm.
♦ PC100h, PC900h, PC1200h, PC2200h, PC3200h: Switch the alarm key switch ON .
3. Disconnect unit from AC power. The power failure alarm will activate immediately. If the power failure alarm does not activate, replace the power failure alarm battery.
4. Connect unit to AC power. The power failure alarm should clear.
5. Switch the alarm key switch to the original setting.
Motion Alarm Test (PC4200h)
1.
Confirm the platelet incubator is connected to AC power.
2. Ensure the Motion Alarm key switch is switched to position 1 .
3. Switch the AC ON/OFF switch ON .
4. Open the door. Agitation motion will stop and the motion alarm will activate after the delay period elapses.
5. Close the door. Agitation motion will resume. The motion alarm will clear and the audible alarm will cease.
360096-1/P
71
Helmer Scientific Platelet Incubator Service Manual
12 Temperature Controller Setpoints
Temperature controller setpoints are programmed at the factory. Setpoints can be viewed and changed through the temperature controller. Parameter values reside in four program levels.
Temperature controller display.
Changeable Parameters
Parameter
Program
Level
ZEro
AL.hi
AL.Lo
3
1
1
Function
Calibrate the temperature controller readout
Set high temperature alarm value: Horizon Series
Set low temperature alarm value: Horizon Series
View / Change Parameter Values:
NOTICE
Changing parameter values affects platelet incubator operation. Do not change parameter values unless instructed in product documentation or by Helmer Technical Service.
Note
When there is no interaction for 60 seconds, the temperature controller exits program mode.
1. Press and hold the UP and DOWN
buttons simultaneously for approximately three seconds. “tunE” and “oFF” flash on the display. The temperature controller is now in Level 1 program mode.
2. Press and release the UP or DOWN
buttons until the desired parameter flashes on the display.
3. Change a parameter value by pressing and holding the * button.
4. Press the UP or DOWN buttons to change the parameter value.
♦ To access Level 2 parameters, change the value for the “LEVL” parameter to
2 .
♦ To access Level 3 parameters, change the value for the “LEVL” parameter to
3 .
5.
Release all buttons to exit the parameter. New settings are saved. If the “LEVL” parameter value is changed, the temperature controller returns to the selected program level.
6. Repeat steps 2 through 4 to access another program level, or to view or change parameter values in the selected level.
7. Access Level 4 parameters by navigating to Level 3 program mode.
8.
Select the “UEr” parameter.
9. Press and hold the UP and DOWN
buttons simultaneously for approximately 10 seconds. The “LOCK” parameter flashes on the display.
10. Exit program mode by pressing and holding the UP and DOWN arrow buttons simultaneously for approximately three seconds, or until the current chamber temperature is displayed.
360096-1/P
72
Helmer Scientific Platelet Incubator
12.1 Level 1 Parameters and Values
Service Manual
Notes
• Parameters are listed in order of appearance.
• SP1 refers to the chamber heater.
• SP2 refers to the temperature alarm for Horizon Series models.
• Default values for AL.Lo and AL.hi may vary: the values listed are for temperature controllers programmed at the factory
with a platelet incubator setpoint of 22 °C.
Parameter Description
Autotune selection
SP1 proportional band (gain) or hysteresis
(in °C)
SP1 integral time/reset (in minutes)
SP1 derivative time/rate
SP1 derivative approach control
SP1 proportional cycle-time
SP1 offset (manual reset)
Lock main setpoint
Ramp rate (if ramp is on)
Ramp selection
Soak time selection
SP2 low alarm setpoint (in °C)
SP2 high alarm setpoint (in °C)
SP2 hysteresis or proportional band (in °C)
SP2 mode selection (ON/OFF or proportional cycle-time)
Parameter level currently selected
7.5
oFF
0.5
0.5
0.5
oFF
Default Value (Countertop
Models)
Default Value
(Floor Models) oFF oFF
7.5
3.2
15 oFF
0.5
0.5
0.5
oFF
0 oFF
--
20.0 (see note above)
24.0 (see note above)
0.1
on.oF
1
0 oFF
--
20.0 (see note above)
24.0 (see note above)
0.1
on.oF
1
360096-1/P
73
Helmer Scientific Platelet Incubator
12.2 Level 2 Parameters and Values
Parameter Description
Display of SP1 output power percentage (read-only)
Selection for manual control of power for SP1
SP1 power limit percentage
SP2 power limit percentage (cooling)
Main SP2 operating mode (alarm strategy)
Subsidiary SP2 mode
Display resolution
Full scale
Minimum scale
Input sensor type
Units (°C / °F)
Parameter level currently selected
35.0
15.0
rtd
°C
2
0 bAnd nonE
0.1°
Default Value (Countertop
Models)
Default Value
(Floor Models)
(read only) (read only) oFF
100 oFF
100
0 bAnd nonE
0.1°
35.0
15.0
rtd
°C
2
12.3 Level 3 Parameters and Values
Parameter Description
SP1 output device type
SP2 output device type (read only)
Sensor burn-out protection type (upscale or downscale)
Default Value (Countertop
Models)
Default Value
(Floor Models)
SSd1 SSd1
SSd2 uP.SC
SSd2 uP.SC
Output mode (reverse or direct) 1r.2r
LED indicator modes for SP1 and SP2 (normal or inverted) 1n.2n
Sensor span adjust
Zero sensor error (calibration across full scale)
0.0
(varies)
1r.2r
1n.2n
0.0
(varies)
Selection for the control accuracy monitor
Read control accuracy monitor results (variance) oFF
UAr°
Read autotune tuning cycle data
Software version (select this parameter to access Level 4 parameters)
Function reset
Ct A
392b nonE
Parameter level currently selected 3 oFF
UAr°
Ct A
392b nonE
3
Service Manual
360096-1/P
74
Helmer Scientific Platelet Incubator
12.4 Level 4 Parameters and Values
Service Manual
Parameter Description
Program security lock
Derivative sensitivity
Display sensitivity (or direct display of input)
Default Value (Countertop
Models)
Default Value
(Floor Models) nonE nonE
0.5
dir
0.5
dir
Disable SP2 alarm annunciator
Program mode auto-exit switch (returns display to normal mode if no activity for one minute) oFF
Auto oFF
Auto
12.5 Change the Platelet Incubator Setpoint
The default setpoint is 22.0 °C. The temperature controller adjusts chamber temperature around the setpoint. When chamber temperature is below setpoint, the controller activates the heater to warm the chamber. When temperature rises above setpoint, the controller activates the compressor to cool the chamber.
NOTICE
Do not change the setpoint to a value outside the temperature control range.
Open the control door to access the temperature controller. The current chamber temperature is displayed. The temperature may be different from the setpoint, especially if the chamber temperature has not yet stabilized.
Temperature controller display.
1. Press and hold the * button to access the setpoint adjustment function. The controller displays current setpoint value.
button to change the setpoint.
2. Press and hold the *
3. Press the UP or DOWN buttons to increase or decrease setpoint in increments of 0.1 °C.
4. Release all buttons to exit the setpoint parameter. New settings are saved.
12.6 Calibrate the Temperature Controller Display
The temperature controller senses chamber temperature through a probe in the chamber. The chamber setpoint typically varies from the measured temperature, so an offset value is used by the control system to compensate for the difference.
NOTICE
Do not change the control sensor offset unless instructed in product documentation or by Helmer Technical Service.
Notes
• Value is factory-preset and varies for each unit
• Offset value can be changed from -10.0 °C to +10.0 °C
• If the variance is within acceptable limits for your organization, changing the offset value is optional.
360096-1/P
75
Helmer Scientific Platelet Incubator Service Manual
Determine New Offset Value
1. Use the zip tie to attach a calibrated independent thermometer traceable per national standards to the control probe.
2. Close door and allow chamber temperature to stabilize for 30 minutes.
3. Observe independent thermometer temperature for 10 minutes and determine temperature range.
4. From the range, calculate the average temperature.
5. Remove the thermometer from the control probe.
6. Determine the change in value to reach desired setpoint.
Example
♦
Current setpoint is 22.0 ºC
♦
Average temperature is 23 ºC
♦
Offset adjustment value is +1.0 ºC
Enter New Offset Value
Note
When there is no interaction for 60 seconds, the temperature controller exits program mode.
1. Press and hold the UP and DOWN
buttons simultaneously for approximately three seconds. “tunE” and “oFF” flash on the display. The temperature controller is now in Level 1 program mode.
2. Press and release the UP or DOWN buttons until “LEVL” flashes on the display.
3. Press and hold the * button.
4. Press the UP or DOWN buttons to change the parameter value to 3 .
5. Release the * button.
6. Press and release the UP or DOWN
buttons until “ZEro” flashes on the display.
7. Press and hold the * button.
8. Press the UP or DOWN buttons to change the value to the calculated offset adjustment value.
9. Release the * button.
10.
Release all buttons to exit the parameter. New settings are saved. If the “LEVL” parameter value is changed, the temperature controller returns to the selected program level.
11. Press and hold the UP and DOWN arrow buttons simultaneously for approximately three seconds, or until the current chamber temperature is displayed.
360096-1/P
76
Helmer Scientific Platelet Incubator
13 Maintenance
Service Manual
NOTICE
• Before performing maintenance, protect items in platelet incubator from extended exposure to adverse temperature.
• Allow platelet incubator temperature to stabilize at setpoint after performing service or after extended door opening.
Note
Refer to the operation manual for the preventive maintenance schedule.
13.1 Recharge Refrigerant
CAUTION
Review all safety instructions prior to recharging refrigerant. Refer to chapter 2 (Safety).
Maintenance should only be performed by trained refrigeration technicians.
NOTICE
Use only non-CFC R-134A refrigerant.
Full initial refrigerant charge varies by model.
Model
PC100h, PC900h, PC1200h
Volts
115/230
PC2200h, PC3200h, PC4200h
(serial numbers prior to 2042392)
115
Initial Charge
5.0 oz (142 g)
8.0 oz (227 g)
PC2200h, PC3200h, PC4200h
(serial number 2042392 and after)
115
PC2200h, PC3200h, PC4200h
(serial numbers prior to 2046319)
230
PC2200h, PC3200h, PC4200h
(serial number 2046319 and after)
230
4.0 oz (113 g)
8.0 oz (227 g)
12.0 oz (340 g)
Add refrigerant:
Note
Pressure varies depending on ambient air temperature.
1.
Attach calibrated pressure gauge (0 lbs/in² to 60 lbs/in² (0 kPa to 400 kPa)) to the fittings on the refrigeration lines.
2. Measure the low side (suction) pressure.
3. Add refrigerant so the peak low side pressure corresponds to pressure (at the stated ambient temperature) in the table below.
25
26
27
28
29
30
21
22
23
24
18
19
20
Ambient
Temperature
°C °F
77.0
78.8
80.6
82.4
84.2
86.0
64.4
66.2
68.0
69.8
71.6
73.4
75.2
Low Side Pressure lbs/in²
32.6
33.3
33.5
34.3
34.8
35.4
29.0
29.5
30.0
30.5
31.0
31.5
32.0
kPa
225
230
231
236
240
244
200
203
207
210
214
217
221 bar
2.25
2.30
2.31
2.36
2.40
2.44
2.00
2.03
2.07
2.10
2.14
2.17
2.21
4. Remove pressure guage.
360096-1/P
77
Helmer Scientific Platelet Incubator
13.2 Power Failure Alarm Battery
If the backup battery for the Power Failure alarm is depleted, the Power Failure alarm will not activate.
Service Manual
Test Power Failure Alarm Backup Battery
1. Disconnect the platelet incubator from AC power. The Power Failure alarm will sound immediately. If the alarm does not sound, replace the battery.
2. Reconnect the platelet incubator to AC power.
Replace the Power Failure Alarm Battery
NOTICE
Use battery which meets manufacturer’s specifications (9 V non-rechargeable alkaline battery or equivalent).
1. Open the control door on the front of the platelet incubator. Switch the AC ON/OFF switch OFF .
2. Disconnect the platelet incubator from AC power.
3. Disable the Power Failure alarm:
♦
PC100h, PC900h, PC1200h, PC2200h, PC3200h: disconnect the power failure alarm battery.
♦
PC4200h: Turn alarm key to position 3 .
4. Remove the battery. Replace with a new battery.
5. Enable the Power Failure alarm:
♦ PC100h, PC900h, PC1200h, PC2200h, PC3200h: reconnect the power failure alarm battery.
♦
PC4200h: Turn alarm key to the original position.
6. Reconnect the platelet incubator to AC power.
7. Switch the AC ON/OFF switch ON and close the control door.
13.3 Motion Alarm Battery (PC4200h)
Test the motion alarm battery to ensure it can provide power to the motion alarm in the event agitation motion stops.
Notes
• The motion alarm battery is applicable to the PC4200h platelet incubator only.
• If the agitation motion alarm sounds, the battery does not need to be replaced.
• If the agitation motion alarm does not sound, replace the motion alarm backup battery.
Test Motion Alarm Battery
1. Ensure the Motion Alarm ON/OFF switch is switched ON .
2. Ensure the monitoring system backup battery is switched ON .
3.
Using a small flat-head screwdriver, rotate the Motion Alarm Delay control to the left (counterclockwise) to shorten the
motion alarm delay to five seconds.
4.
Open the door to disable agitation. Motion alarm will activate after five seconds.
5. Close the door to enable agitation. Agitation motion alarm will clear and audible alarm will cease.
6.
Using a small flat-head screwdriver, rotate the Motion Alarm Delay control to the right (clockwise) to change the motion alarm delay to the original setting.
Replace the Motion Alarm Battery (PC4200h)
NOTICE
Use battery which meets manufacturer’s specifications (9 V nickel-metal hydride rechargeable battery (or equivalent).
1. Open the control door on the front of the platelet incubator and disconnect the motion alarm battery.
2. Switch the AC ON/OFF switch OFF .
3. Disconnect the platelet incubator from AC power.
4. Remove the battery. Replace with a new battery.
5. Reconnect the platelet incubator to AC power.
6. Switch the AC ON/OFF switch ON .
7. Reconnect the motion alarm battery and close the control door.
360096-1/P
78
Helmer Scientific Platelet Incubator
13.4 Clean the Platelet Incubator
Service Manual
WARNING
Disconnect platelet incubator from AC power when cleaning.
Condenser Grill
In environments where platelet incubator is exposed to excessive lint or dust, condenser grill may require cleaning more frequently than stated in preventive maintenance schedule.
Clean the condenser grill using a soft brush and a vacuum cleaner.
Exterior
Clean glass surfaces with soft cotton cloth and glass cleaner. Clean exterior surfaces with soft cotton cloth and non-abrasive liquid cleaner.
CAUTION
The condensate evaporator and water evaporation tray are hot.
NOTICE
For PC4200h platelet incubators, clean the vents in the base as needed.
Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose laboratory cleaner suitable for stainless steel. For PC4200h models, clean textured drawer surfaces with a soft brush.
Door Gaskets
Clean with soft cloth and mild soap and water solution.
13.5 Lubricate the Drawer Slides (PC4200h)
CAUTION
• Before moving drawers, ensure they are completely empty.
• Open the drawer using the drawer handle; not the label holder.
• Do not open more than one drawer at a time.
Clean the slots in the drawer guides with a clean, dry cloth. Apply a thin film of petroleum jelly to the bottom of each drawer where it contacts the drawer guide.
360096-1/P
79
Helmer Scientific Platelet Incubator
13.6 Lubricate and Service Internal Parts (PC4200h)
Service Manual
WARNING
Disconnect platelet incubator from AC power when lubricating internal parts.
Drive Rod Bearings
Clean drive rod end bearings, wheels, and components according to the preventive maintenance schedule in the operation manual.
Lubricate the drive rod bearings:
1.
Using grease gun, inject bearing grease into the grease fitting on the bearing.
B
C
Label
A
Description
Drive rod
Grease fitting
Rod end bearing
A B C
Drive rod and rod end bearing detail.
Wheels
NOTICE
Do not use oil or aerosol lubricant.
Lubricate wheel axles with a light machine oil. Remove the cover to access the front wheels for inspection, cleaning, and lubrication.
Replace wheels as necessary. The hole in each wheel should be circular. If the hole is out-of-round, replace the wheel. If the wheels need to be replaced, contact Helmer Technical Service.
360096-1/P
80
Helmer Scientific Platelet Incubator
14 Troubleshooting
CAUTION
Review all safety instructions prior to troubleshooting.
Service Manual
14.1 General Operation Problems
Problem
Roll-out shelf does not slide easily (PC2200h and
PC3200h).
Drawer does not slide easily
(PC4200h).
Door does not open easily
(PC100h and floor models).
Possible Cause
Debris in the shelf slides.
Shelf slides are not lubricated.
Shelf slide is faulty.
Debris in slots of drawer guides.
Drawers are not lubricated.
Debris in hinges.
Action
Pull the shelf out and confirm the slides are free of debris. Clean if necessary.
Using a lightweight oil, lubricate the bearings in the slides.
Contact a qualified service technician.
Pull the drawer out and confirm the slots are free of debris. Clean if necessary.
Clean and lubricate the drawer slots.
Confirm the hinges are free of debris. Clean the hinges if necessary.
Using a general-purpose grease, lubricate the pivots in the hinges.
Contact a qualified service technician.
Contact a qualified service technician.
Hood does not open easily
(PC900h and PC1200h).
Platelet incubator does not roll easily (floor models).
Platelet incubator is connected to AC power, and doors are closed, but built-in platelet agitator is not moving (PC 4200h).
Agitation is noisy or rough
(PC4200h).
Door hinges are not lubricated.
Hinge cam is faulty.
Damping cylinder or hood hinge is faulty.
One or more casters are locked.
Platelet agitator is obstructed.
Component is faulty or internal connections are loose.
Platelet incubator is not level.
Moving parts are grinding against each other.
Component is faulty or internal connections are loose.
Outlet connection is faulty.
Unlock the casters. Push the platelet incubator from the sides, not from the front or back.
Verify the platelet agitator is free of obstructions.
Contact a qualified service technician.
Verify the platelet incubator has been set up appropriately.
Verify that moving parts are lubricated and in good condition. Lubricate or replace if necessary.
Contact a qualified service technician.
Main power switch is on, but nothing is working.
Power cord is faulty.
Power connector is faulty (all models except 115 V floor models).
Verify power at the outlet. Repair original outlet or connect to a different outlet if necessary.
Check condition of the power cord. Replace if in poor condition.
Check that voltage through the cord is appropriate. If not, replace cord.
Check that connections between power cord and power connector are secure.
Check that voltage through power connector is appropriate. If not, replace power connector.
Confirm that all circuit breakers are seated. Push circuit breaker to reset if necessary.
Circuit breaker was tripped (230 V countertop models, all 100/115 V models).
One or both fuses have blown (230
V floor models).
Power line filter is faulty (230 V floor models).
Power switch is faulty.
Verify fuse has not blown. Replace fuse if necessary.
Verify power output from the filter. Replace power line filter if necessary.
Terminal board is faulty.
Main power switch is on, but temperature controller is not working.
Connection between terminal board and temperature controller is loose or a part is faulty.
Verify power output from the switch. Replace switch if necessary.
Confirm terminal board is supplying power to other components. Replace terminal board if necessary.
• Check connections between terminal board and temperature controller.
Secure connections if necessary.
•
Confirm terminal board is supplying power to temperature controller.
Replace temperature controller or terminal board if necessary.
360096-1/P
81
Helmer Scientific Platelet Incubator Service Manual
Problem
Main power switch is on, but compressor is not running.
Main power switch is on, but chart recorder is not working.
Main power switch is on, and the doors are closed, but built-in platelet agitator is not agitating (PC4200h).
Main power switch is on, and door/hood/doors is closed, but platelet agitators installed inside platelet incubator are not agitating (all models except
PC4200h).
Main power switch is on, and door/hood/doors is closed, but platelet agitators installed inside platelet incubator are not agitating (all models except
PC4200h).
Main power switch is on, but one or both chamber fans are not running.
Roll-out shelf does not slide easily (PC2200h and
PC3200h).
Door does not open easily
(PC100h and floor models).
Hood does not open or close easily or slams shut instead of gradually closing
(PC900h and PC1200h).
Possible Cause Action
Connection between terminal board and the compressor is loose or a part is faulty.
• Check connections between the terminal board and compressor.
Secure connections if necessary.
•
Confirm terminal board is supplying power to compressor by checking whether condensing unit fan is working.
•
Replace terminal strip if necessary (floor models only).
• Replace starting relay coil, compressor/condensing unit, or terminal board if necessary.
Connection between terminal board and chart recorder is loose or a part is faulty.
Doors are not closing completely.
• Check connections between terminal board and chart recorder. Secure connections if necessary.
•
Confirm terminal board is supplying power to chart recorder.
• Replace chart recorder or terminal board if necessary. Refer to the temperature chart recorder operation and service manual.
•
Confirm hinge cams are not damaged. Replace if necessary.
•
Confirm door gaskets are sealing completely. Replace door gasket if necessary.
One or both door switches are faulty.
Secondary circuit breaker has tripped (100/115 V models).
Agitation motor is faulty.
Terminal board is faulty.
Door(s) are not closing completely
(PC100h and floor models only).
Hood is not closing completely
(PC900h and PC1200h).
One or both door switches are faulty.
Terminal board is faulty.
Check connections for door switches. Replace one or both door switches if necessary.
Confirm that all circuit breakers are seated. Push circuit breaker to reset it if necessary.
Check connections for agitation motor. Replace motor if necessary. On
230 V models, replace motor capacitor if necessary.
•
Confirm terminal board is supplying power to door switch(es). Replace terminal strip if necessary (floor models only).
• Replace terminal board if necessary.
Confirm hinge cams (floor models) or hinge bearings (PC100h) are not damaged.
Confirm damping cylinder is not damaged. Replace it if necessary.
On chamber wall in the hinge area, check nutsert for looseness. Replace nutsert if necessary.
Check connections for door switch(es). Replace door switch(es) if necessary.
Confirm terminal board is supplying power to door switch(es). Replace terminal strip if necessary (floor models only). Replace terminal board if necessary.
Confirm that all circuit breakers are seated. Push circuit breaker to reset it if necessary.
Check connections for interior outlets. Replace outlets as necessary.
Secondary circuit breaker has tripped (100/115 V models).
One or more interior outlets are faulty.
Platelet agitator component is faulty or internal connections are loose.
Refer to the platelet agitator service manual.
Chamber fan is faulty (PC100h).
Check connections for the fan. Replace fan if necessary.
Check connections for the fan. Replace fan motor if necessary.
One or both chamber fan motors are faulty (all models except
PC100h).
Slide is faulty.
Confirm slide is operating correctly. Replace slide if necessary.
Hinge cam is faulty (floor models only).
Hinge bearing is faulty (PC100h).
Damping cylinder is faulty.
Hinge hardware is loose or faulty.
Confirm hinge cam is not damaged. Replace if necessary.
Confirm bearings are not damaged. Replace if necessary.
Replace damping cylinder.
On chamber wall, check nutsert for looseness. Replace nutsert if necessary.
360096-1/P
82
Helmer Scientific Platelet Incubator Service Manual
Problem
Agitation is noisy or rough
(PC4200h).
Agitation of installed platelet agitators is noisy or rough (all models except
PC4200h).
Possible Cause Action
One or more wheels are worn.
Bearing on drive rod assembly is grinding or worn.
Check hole through each wheel for wear. If a wheel is out-of-round, replace wheel or wheel assembly.
Drive rod is loose or disconnected.
Check connections from drive rod to drive plate on the storage frame, and to the crank block on the motor. Secure connections if necessary.
On drive rod assembly, check bearing on end connected to the drive plate.
Lubricate bearing by injecting grease through grease fitting. If bearing is worn, replace drive rod assembly.
Crank block on the motor is worn.
Motor is faulty.
One or more feet on platelet incubator are worn (countertop models).
Check crank block on the motor for wear. Replace crank block if worn.
Replace motor.
Check feet for wear. Replace feet if necessary.
Platelet agitator component is faulty or internal connections are loose.
Refer to the platelet agitator service manual.
14.2 Chamber Temperature Problems
Problem Possible Cause
Chamber temperature does not stabilize at platelet incubator setpoint.
Condenser grill is dirty.
Air circulation at top of chamber is not adequate.
Ambient air temperature around platelet incubator is too high.
Refrigerant level is too low.
Component is faulty or internal connections are loose.
Action
Check condenser grill. Clean if necessary.
Check for items that may obstruct air flow and remove necessary.
Confirm platelet incubator is placed appropriately.
Contact a qualified service technician.
Contact a qualified service technician.
14.3 Alarm Activation Problems
Problem
Platelet incubator is in an alarm condition, but alarms are not audible.
All alarms are activating sporadically.
Platelet incubator meets a high or low temperature alarm condition, but appropriate alarm is not active.
High or low temperature alarm activated, but temperature recorded at activation does not match alarm setpoint.
Possible Cause Action
Audible alarms have been muted.
Verify audible alarms are not muted.
Audible alarms have been disabled.
• Verify audible alarms are not disabled.
• Check the alarm key switch position.
Audible alarm time delay has not elapsed.
Volume is set too low.
Allow the alarm time delay to elapse or change the alarm delay.
Check volume and change if necessary.
Verify whether active alarm should be communicated audibly.
Alarm is not communicated audibly, visually, or through remote alarm interface.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Alarm setpoint was changed.
End of delay period for alarm has not been reached.
Check current setpoints for alarms. Change setpoints if necessary.
Wait until delay period has ended. Alarm should activate.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Temperature changed slightly around time of activation.
No action needed.
360096-1/P
83
Helmer Scientific Platelet Incubator Service Manual
Problem
Platelet incubator is not receiving AC power, but power failure alarm is not active.
Platelet incubator is connected to power, but
AC Power Failure alarm is active.
Possible Cause
Component is faulty or internal connections are loose.
Alarm backup battery is low.
Action
Contact a qualified service technician.
Replace the alarm backup battery.
Circuit breakers have tripped or are not seated correctly (115 V floor models, all countertop models).
Fuses have blown (230 V floor models).
Confirm circuit breakers are seated. Push circuit breaker(s) to reset if necessary. If breaker(s) cannot be reset or trip again, contact a qualified service technician.
Contact a qualified service technician.
No AC power to platelet incubator.
• Verify power cord is connected securely to platelet incubator and outlet.
Tighten connections if necessary.
• Verify outlet is operational and meets power requirements for platelet incubator. Repair original outlet or connect to a different outlet if necessary.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
Motion alarm is active, but platelet agitator is moving
(PC4200h).
Agitation has stopped, but motion alarm is not active
(PC4200h).
Motion alarm is activating sporadically while platelet agitator is moving
(PC4200h).
Motion alarm is activating immediately when agitation is started (PC4200h).
Delay period for motion alarm is set too low.
Verify delay period control is set to a value greater than platelet agitator alarm setpoint.
Wait until delay period has ended. Alarm should activate.
End of delay period for the motion alarm has not been reached.
Motion alarm system is not receiving power.
• Check that there is AC power to platelet incubator.
• Test backup battery. If battery is not functioning, replace battery.
Contact a qualified service technician.
Component is faulty or internal connections are loose.
Alarm system is not receiving power consistently.
Component is faulty or internal connections are loose.
• Check that there is AC power to platelet agitator.
• Test backup battery. If battery is not functioning or has been in use for one year, replace battery.
Contact a qualified service technician.
Change delay period to a value greater than 0 minutes.
Delay period is set to 0 minutes, which activates alarm before motion can be sensed.
Component is faulty or internal connections are loose.
Contact a qualified service technician.
360096-1/P
84
Helmer Scientific Platelet Incubator
14.4 Condensation Problems
Service Manual
Problem
Excessive water in water evaporation tray (countertop models).
Possible Cause Action
Humid air is entering chamber.
Confirm platelet incubator is level and doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Heater in evaporation tray is faulty.
Confirm heater is hot. Contact a qualified service technician to confirm heater is drawing the appropriate current.
Excessive water in chamber.
Excessive humidity on doors.
Connections between terminal board and condensate evaporator are loose or a part is faulty.
Humid air is entering chamber.
Humid air is entering chamber.
Relative humidity around the platelet incubator is too high.
• Check connections between terminal board and condensate evaporator
(heater). Secure connections if necessary.
•
Confirm condensate evaporator is working correctly. It should be hot and drawing appropriate current:
• For 100/115 V models, current should be approximately 0.43 A to 0.55 A.
• For 230 V models, current should be approximately 0.21 A to 0.35 A.
• Replace condensate evaporator or terminal board if necessary.
Confirm platelet incubator is level and doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Confirm platelet incubator is level, and the doors are aligned, closing tightly, and sealing correctly. Contact a qualified service technician if necessary.
Confirm platelet incubator is placed appropriately.
360096-1/P
85
Helmer Scientific Platelet Incubator
15 Horizon Parts
Service Manual
NOTICE
• Before replacing parts, protect items in platelet incubator from extended exposure to adverse temperature.
• Allow platelet incubator temperature to stabilize at setpoint after replacing parts or after extended door opening.
Front
A
D
B
C
J
Front parts (multiple models shown)
B
C
D
E
F
Label Description
A Chart recorder
Chart recorder backup battery
Chart paper
Control Door
Main power switch
Temperature controller
(countertop)
Temperature controller (floor model)
E
F
G
H
I
K
Part Number Label Description
400409 G Key switch, 2-way
(all models except PC4200h)
120218
220273
I
H
Key switch, 3-way
Mute button
(PC4200h)
Power failure battery
120478
400600-1
400601-1
J
K
Caster (swivel with brake; floor models)
Foot
(countertop models)
Part Number
120227
120489
120197
120218
220261
220432
360096-1/P
86
Helmer Scientific Platelet Incubator
Door and Hinges
Service Manual
F
A
B
C
D
Door and Hinge (PC100h)
B
C
D
Label Description
A Hinge bearing
E
Not shown
Upper hinge bracket
Lower hinge bracket
Door seal kit
Door handle
Door lock
F Door brush seal
Door brush seal
Hood
G
E
K
L
H
Door and Hinge (PC2200h, PC3200h, and PC4200h)
Model
100
100
100
2200
Part Number Label Description
220295 G Door handle
320846-1
310846-2
400785-1
I
H Door lock
Upper hinge bracket
(left)
320865-1
220374
J
Model
2200/3200/4200
Part Number
320865-1
220374
400647-2
Upper hinge bracket
(right)
2200/3200/4200 400647-1
Door seal kit 2200
3200/4200
400787-1
400788-1
320920-2
3200/4200 320920-1
K
L
Hinge cam
Lower hinge bracket
(left)
2200/3200/4200
320742-1
400647-1
Lower hinge bracket
(right)
2200/3200/4200 400647-2
I
J
A
B
C
(PC900h and PC1200h)
Label Description
A Hood assembly
B Damping cylinder
Model Part Number Label Description
900 800098-1 C Hood lock
1200 800099-1
220210
Not
Shown
Hood seal kit
Model Part Number
220272
900 400786-1
1200 400786-2
360096-1/P
87
Helmer Scientific Platelet Incubator
Top / Side
A B
Service Manual
E F
D
C
PC100h PC2200h, PC3200h and PC4200h
Label Description
A Compressor/condensing unit
Model
100/900/
1200
B
C
Terminal board
Alarm Board
Volts
115
230
Part
Number
400668-1
Label Description
D
Power line filter
400668-2 E
115/230 800285-1 F
Agitator board
Motion alarm battery
100 800292-1 Not
Shown
400521-1
Compressor relay
Model Volts
230
4200
Part
Number
120400
400516-1
120232
100/115 120528
230 120523
360096-1/P
88
Helmer Scientific Platelet Incubator
Rear - All Models
F G H
Service Manual
I J
A
D
E
C
B
K
.
PC100h, PC900h, and PC1200h PC2200h, PC3200h, and PC4200h
C
D
E
Label
A
B
Description
Power cord
Model Volts
Part
Number
100, 900, 1200 100/115 120630 all models
2200, 3200
4200
Condensate evaporator kit
(includes tray)
230
115
115
100/115
230
120156
120130
110040
400791-1
400791-2
120160
I
H
J Alarm buzzer
Circuit breaker
(6 A)
Power connector (includes power line filter)
230 120288
120299
Label Description
F Motion Alarm buzzer
G Circuit breaker
(secondary, 2 A)
Model
4200
2200, 3200
Circuit breaker
(secondary, 5 A)
4200
Volts
115
115
K
Circuit breaker
Circuit breaker
(primary, 15 A)
(primary, 20 A)
2200, 3200
4200
Power connector with fuse holder
Fuse
(10 A)
Motion alarm control board 4200
115
115
230
230
Part
Number
120160
120252
120514
120259
120191
120379
120174
400523-1
360096-1/P
89
Helmer Scientific Platelet Incubator
Interior - All Models
A
B
PC100h
C
E
D
PC900h and PC1200h
F
G
H
Service Manual
K
L
PC2200h and PC3200h
B
C
D
E
F
Label Description
A Chamber fan
Chamber fan guard
Alarm check tube
Door switch
Control probe / chart recorder probe
Chamber fan blade
I J
Volts
Part
Number
100/115 120009
Label
G
Description
Chamber fan motor
230 120010
220122 H Chamber heater
220299
120490
120315
220177
I
J
K
L
Interior outlet
Interior outlet
Shelf slide
Roll-out shelf
(excludes bumpers and slides)
Volts
Part
Number
100/115 120170
230 120171
100/115 120313
230 120314
100/115 120499
230 120097
400717-1
320971-1
360096-1/P
90
Helmer Scientific Platelet Incubator
Agitation Parts
A
B
C
D
E
F
G
Service Manual
PC4200h
Label Description
A
B
C
Drawer stop
Drawer guide
(Top left)
Drawer guide
(Top right)
Drawer guide
(Bottom left)
Drawer guide
(Bottom right)
Wheel (included in wheel assembly)
Base
(PC4200h)
Part Number Label Description
400174-1
320540-1
320540-2
320541-1
320541-2
320566-1
D
E
F
G
Wheel track
Drive plate
Part Number
320215-1
321005-1
Wheel assembly
(includes two wheels, axle, mounting bracket)
400159-1
Drawer assembly
(includes label holder kit.)
* = (silver drawer - standard)
** = (red drawer)
*** = (black drawer)
400616-1*
400617-1**
400615-1***
D
E
F
A B C
G
B
C
D
Label Description
A Agitation motor
(includes capacitor)
*for serial numbers 968995 and lower
** for serial numbers 968996 and greater
Crank block
Motor mount
Proximity sensor
Volts
115
230
Part
Number
400305-1
400306-1* F
230 400638-1** G
Label Description
E Drive rod assembly
Proximity actuator
115 320222-1
230 321032-1
Compressor/condensing unit
* = units with serial # 2046319 above
** = units with serial # 2046318 and prior
220079
400694-1
Not
Shown
Starting relay coil
Model
2200, 3200 and 4200
Volts
Part
Number
400111-1
120125
115 400669-1
230 800977-1*
230 800978-1**
115 120524
230 120525
360096-1/P
91
Helmer Scientific Platelet Incubator
Accessories
Service Manual
A
Label Description
A
Fan airflow deflector
Not shown
Bypass plug
Power cord, short
(For all models except PC4200h. )
Model Volts
100
Part Number
320967-1
400657-1
110/115 120316
Label
Not shown
Description
Shelf for platelet rotator
Label holder kit
230 120317
Model Part Number
900 400618-1
1200 400619-1
4200 400620-1*
* Label holder kit if for PC4200h and includes label, label holder, hardware and installation tool. Also included in drawer assembly
360096-1/P
92
Helmer Scientific Platelet Incubator Service Manual
16 Schematics
16.1 PC100h, PC900h, PC1200h (100/115 V)
MAIN POWER
100 (±10%) V 50/60 Hz
115 (±10%) V 60 Hz
(AK) POWER
CONNECTOR
WITH POWER
LINE FILTER
N L
(AD) CHART
RECORDER
BACKUP
BATTERY optional feature
(P) CIRCUIT
BREAKER to J1 on
ALARM
BOARD
12 V dc
(F)
TERMINAL
BOARD
1
J6
115 V
J7
1
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J9
J2
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
(B)
TEMPERATURE
CONTROLLER
100-240
N
V ac
L -
SP2
+ -
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
1
J1
(AN)
INTERIOR
OUTLET
THERMOSTAT (C) CHAMBER
HEATER
(AP)
STARTING
RELAY COIL
J3
1
(AJ) AGITATOR
PORT BOARD
(AL) CHAMBER FAN
(A) COMPRESSOR/
CONDENSING UNIT
(I) DOOR SWITCH
(W) KEY
SWITCH
(J) CONDENSATE
EVAPORATOR to J10 on
TERMINAL
BOARD
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
(D) ALARM
BUZZER
360096-1/P
93
Helmer Scientific Platelet Incubator
16.2 PC100h, PC900h, PC1200h (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(AK) POWER
CONNECTOR
WITH POWER
LINE FILTER
N L
(AD) CHART
RECORDER
BACKUP
BATTERY
(P) CIRCUIT
BREAKERS optional feature
12 V dc to J1 on
ALARM
BOARD
1
1 J9
(F)
TERMINAL
BOARD
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
1 J1
(AO)
INTERIOR
OUTLET
THERMOSTAT (C) CHAMBER
HEATER
(AP)
STARTING
RELAY COIL
J3 1
(AJ) AGITATOR
PORT BOARD
(AL) CHAMBER FAN
(A) COMPRESSOR/
CONDENSING UNIT
(I) DOOR SWITCH
(W) KEY
SWITCH
(J) CONDENSATE
EVAPORATOR to J10 on
TERMINAL
BOARD
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
(D) ALARM
BUZZER
360096-1/P
94
Helmer Scientific Platelet Incubator
16.3 PC2200h (115 V) optional feature
MAIN POWER
115 (±10%) V ac 60 Hz
L N
POWER
CORD
(P) CIRCUIT
BREAKERS
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
(AN)
INTERIOR
OUTLETS
THERMOSTAT CHAMBER
HEATER
TERMINAL
STRIP
1 J1
(H) CHAMBER
FAN MOTOR
L
J3 1
(AM) CIRCUIT
BREAKER
1 J1
(H) CHAMBER
FAN MOTOR
N
J3 1
(AJ) AGITATOR
PORT BOARDS
LEFT RIGHT
(I) DOOR SWITCHES
(W) KEY
SWITCH
(AP)
STARTING
RELAY COIL to J10 on
TERMINAL
BOARD
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(D) ALARM
BUZZER
(A) COMPRESSOR/
CONDENSING UNIT
360096-1/P
95
Helmer Scientific Platelet Incubator
16.4 PC2200h (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N
(Y) POWER
LINE FILTER
IN
OUT
L
Service Manual optional feature
(AD) CHART
RECORDER
BACKUP
BATTERY
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(X)
CHART
RECORDER
(Q) MAIN
POWER
SWITCH
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
(AO)
INTERIOR
OUTLETS
THERMOSTAT CHAMBER
HEATER
TERMINAL
STRIP
1
J1
(H) CHAMBER
FAN MOTOR
L
J3
1
1
J1
LEFT RIGHT
(I) DOOR SWITCHES
(H) CHAMBER
FAN MOTOR
N
J3
1
(AJ) AGITATOR
PORT BOARDS
(W) KEY
SWITCH
(AP)
STARTING
RELAY COIL to J10 on
TERMINAL
BOARD
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(A) COMPRESSOR/
CONDENSING UNIT
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
(D) ALARM
BUZZER
360096-1/P
96
Helmer Scientific Platelet Incubator
16.5 PC3200h (115 V) optional feature
MAIN POWER
115 (±10%) V ac 60 Hz
L N
(P) CIRCUIT
BREAKERS
1
J1
(AD) CHART
RECORDER
BACKUP
BATTERY
(X)
CHART
RECORDER
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
(AN) INTERIOR
OUTLETS
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
THERMOSTAT
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
Service Manual
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
J3
1
1
J1
(H) CHAMBER
FAN MOTOR
L
J3
1
(AM) CIRCUIT
BREAKER
(H) CHAMBER
FAN MOTOR
N
1 J1
LEFT RIGHT
(I) DOOR SWITCHES
J3 1
(AJ) AGITATOR
PORT BOARDS
(W) KEY
SWITCH
(AP)
STARTING
RELAY COIL to J10 on
TERMINAL
BOARD
-
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(D) ALARM
BUZZER
(A) COMPRESSOR/
CONDENSING UNIT
360096-1/P
97
Helmer Scientific Platelet Incubator
16.6 PC3200h (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N
(Y) POWER
LINE FILTER
1
IN
OUT
J1
L
Service Manual optional feature
(AD) CHART
RECORDER
BACKUP
BATTERY
(X)
CHART
RECORDER
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
(AO) INTERIOR
OUTLETS
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
THERMOSTAT
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
J3 1
1 J1
(H) CHAMBER
FAN MOTOR
L
J3 1
1
J1
J3
1
(AJ) AGITATOR
PORT BOARDS
LEFT RIGHT
(I) DOOR SWITCHES
(W) KEY
SWITCH
(H) CHAMBER
FAN MOTOR
N
(AP)
STARTING
RELAY COIL to J10 on
TERMINAL
BOARD
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(A) COMPRESSOR/
CONDENSING UNIT
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
(D) ALARM
BUZZER
360096-1/P
98
Helmer Scientific Platelet Incubator
16.7 PC4200h (115 V)
Service Manual
MAIN POWER
115 (±10%) V ac 60 Hz
POWER
CORD
L N
(AD) CHART
RECORDER
BACKUP
BATTERY optional feature
(X)
CHART
RECORDER
(P) CIRCUIT
BREAKERS
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(AG) MOTION
ALARM BUZZER
J4
J3
(S)
AGITATOR
BOARD
(T)
BATTERY
J7
(O)
PROXIMITY
SENSOR
THERMOSTAT
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
(H) CHAMBER
FAN MOTOR
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
SP1 = chamber heater
SP2 = temperature alarm
(G)
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
L
Voltage selection switch
1
J1
ALARM
VOLUME
ALARM
DELAY
P1
(V) MOTION
ALARM CONTROL
BOARD to J10 on
TERMINAL
BOARD
L
N
L
N
J2
(AI) KEY
SWITCH
6
7
5 4
8 1
3
2
(H) CHAMBER
FAN MOTOR
(AM) CIRCUIT
BREAKER
LEFT RIGHT
(I) DOOR SWITCHES
N
(AP)
STARTING
RELAY COIL
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(D) ALARM
BUZZER
(A) COMPRESSOR/
CONDENSING UNIT
M
(M) AGITATION
MOTOR
360096-1/P
99
Helmer Scientific Platelet Incubator Service Manual
16.8 PC4200h (230 V)
MAIN POWER
230 (±10%) V ac 50 Hz
230 (±10%) V ac 60 Hz
(U) POWER
CONNECTOR
WITH
(R) FUSES
N L
(AD) CHART
RECORDER
BACKUP
BATTERY optional feature
(X)
CHART
RECORDER
(Y) POWER
LINE FILTER
IN
OUT
12 V dc to J1 on
ALARM
BOARD
(F)
TERMINAL
BOARD
1
1 J9
J7 J2
J6
115 V
Voltage selection switch
L1
230 V
LINE (L)
J4
NEUTRAL (N)
J5
L2
J1
1 2 3 4 5 6 1 2 3 4 5 6
J10
12 V dc
SP2
(alarm)
J3
J8 N
L
(AG) MOTION
ALARM BUZZER
J4
J3
(S)
AGITATOR
BOARD
(T)
BATTERY
J7
(O)
PROXIMITY
SENSOR
THERMOSTAT
(Q) MAIN
POWER
SWITCH
CHAMBER
HEATER
(H) CHAMBER
FAN MOTOR
(B)
TEMPERATURE
CONTROLLER
100-240
V ac
N L
SP2
+
SP1
+
8 7 6 5 4 3 2 1
(G)
SP1 = chamber heater
SP2 = temperature alarm
TEMPERATURE
CONTROL AND
CHART
RECORDER
PROBE
TERMINAL
STRIP
L
(H) CHAMBER
FAN MOTOR
N
Voltage selection switch
1
J1
ALARM
VOLUME
ALARM
DELAY
P1
(V) MOTION
ALARM CONTROL
BOARD to J10 on
TERMINAL
BOARD
L
N
L
N
J2
(AI) KEY
SWITCH
6
7
5 4
8 1
3
2
LEFT RIGHT
(I) DOOR SWITCHES
(AP)
STARTING
RELAY COIL
(AF)
MUTE
BUTTON
REMOTE ALARM
INTERFACE
COM
NC
NO
J1
J2
MUTE
(L) ALARM
BOARD
J5
P1
ALARM
DELAY
P2
ALARM
VOLUME
J3
J4
-
+
(AH) POWER FAILURE
ALARM BACKUP
BATTERY
(D) ALARM
BUZZER
(A) COMPRESSOR/
CONDENSING UNIT
CAPACITOR
M
(M) AGITATION
MOTOR
END OF MANUAL
360096-1/P
100
Helmer Scientific Platelet Incubator
Appendix A: Compliance
Regulatory Compliance
This device complies with the requirements of directive 93/42/EEC concerning
Medical Devices, as amended by 2007/47/EC.
Sound level is less than 70 dB(A).
Service Manual
EC REP
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
WEEE Compliance
The WEEE (waste electrical and electronic equipment) symbol (right) indicates compliance with European Union
Directive WEEE 2002/96/EC and applicable provisions. The directive sets requirements for labeling and disposal of certain products in affected countries.
When disposing of this product in countries affected by this directive:
♦
Do not dispose of this product as unsorted municipal waste.
♦
Collect this product separately.
♦
Use collection and return systems available locally.
For more information on the return, recovery, or recycling of this product, contact your local distributor.
0086
360096-1/P
101
Helmer Scientific Platelet Incubator Service Manual
Appendix B: Warranty
For technical service needs, please contact Helmer at 800-743-5637 or www.helmerinc.com. Have the model and serial number available when calling.
Rapid Resolution
When a warranty issue arises it is our desire to respond quickly and appropriately. The service department at Helmer is there for you.
Helmer will oversee the handling of your warranty service from start to finish. Therefore, Helmer must give advance authorization for all service calls and/or parts needs relating to a warranty issue. Any repeat service calls must also be authorized as well. This allows for proper diagnosis and action. Helmer will not be responsible for charges incurred for service calls made by third parties prior to authorization from Helmer. Helmer retains the right to replace any product in lieu of servicing it in the field.
Compressor
For the warranty period listed below, Helmer will supply the refrigeration compressor, if it is determined to be defective, at no charge, including freight. Helmer will not be liable for installation, refrigerant, or miscellaneous charges required to install the compressor beyond the first year of the warranty period.
♦ i.Series model compressor warranty period is five (5) years.
♦
Horizon Series model compressor warranty period is three (3) years.
Parts
For a period of two (2) years, Helmer will supply at no charge, including freight, any part that fails due to defects in material or workmanship under normal use, with the exception of expendable items. Expendable items include glass, batteries, and door gaskets.
Inspection of defective parts by Helmer will be final in determining warranty status. Warranty procedures must be followed in all events.
Labor
Floor Models
For a period of one (1) year, Helmer will cover repair labor costs (including travel) and the cost of refrigerant and supplies necessary to perform authorized repairs. Repair service must be performed by an authorized Helmer service agency following the authorization process detailed above. Alternatively, your facility’s staff may work with a Helmer technician to make repairs. Labor costs for repairs made by unauthorized service personnel, or without the assistance of a Helmer technician, will be the responsibility of the end user.
Countertop Models
For a period of one (1) year, Helmer will cover repair labor costs, provided the product is returned to Helmer for warranty service.
Alternatively, your facility’s staff may work with a Helmer technician to make repairs on site. Labor costs for repairs performed at a location other than Helmer, or for repairs made without the assistance of a Helmer technician, will be the responsibility of the end user.
Additional Warranty Information
The time periods set forth above begin two weeks after the original date of shipment from Helmer. Warranty procedures set forth above must be followed in all events.
THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF. THIS WARRANTY
IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR PARTICULAR PURPOSE SHALL APPLY.
THE LIABILITY, IF ANY, OF HELMER FOR DIRECT DAMAGES WHETHER ARISING FROM A BREACH OF ANY SALES
AGREEMENT, BREACH OF WARRANTY, NEGLIGENCE, OR INDEMNITY, STRICT LIABILITY OR OTHER TORT, OR OTHERWISE
WITH RESPECT TO THE GOODS OR ANY SERVICES IS LIMITED TO AN AMOUNT NOT TO EXCEED THE PRICE OF THE
PARTICULAR GOODS OR SERVICES GIVING RISE TO THE LIABILITY. IN NO EVENT SHALL HELMER BE LIABLE FOR ANY
INDIRECT, INCIDENTAL, CONSEQUENTIAL, OR SPECIAL DAMAGES, INCLUDING WITHOUT LIMITATION DAMAGES RELATED
TO LOST REVENUES OR PROFITS, OR LOSS OF PRODUCTS.
This warranty does not cover damages caused in transit, during installation by accident, misuse, fire, flood, or acts of God. Further, this warranty will not be valid if Helmer determines the failure was caused by a lack of performing recommended equipment maintenance
(per Helmer manual) or by using the product in a manner other than for its intended use. Installation and calibration are not covered under this warranty agreement.
Outside of USA and Canada
Consult your local distributor for warranty information.
360096-1/P
102
Helmer Scientific
14400 Bergen Boulevard, Noblesville, IN 46060 USA
Copyright © 2017 Helmer, Inc.
360096-1/P
advertisement
Related manuals
advertisement
Table of contents
- 6 About this Manual
- 6 Safety Precautions and Symbols
- 7 Product Labels
- 8 Control System
- 8 Temperature Probes
- 9 Chart Recorder
- 10 Models
- 10 Install Batteries for Backup Power
- 10 Platelet Agitator and Platelet Incubator Compatibility
- 11 Install a Platelet Agitator in the Platelet Incubator
- 11 External Monitoring Devices
- 12 Drawer Labels
- 14 Home Screen
- 15 Main Screen
- 16 Calibrate Temperature Probes
- 17 Factory Default Settings
- 17 Alarm Reference
- 18 Alarm Settings
- 21 Alarm Tests
- 25 AgiTrakā¢ System
- 26 Additional System Settings
- 30 Event Log
- 31 System Firmware Upgrade
- 31 Reset the i.Center Monitoring System
- 31 View Manufacturer and Product Information
- 32 Temperature Controller Setpoints
- 33 Level 1 Parameters and Values
- 34 Level 2 Parameters and Values
- 34 Level 3 Parameters and Values
- 35 Level 4 Parameters and Values
- 35 Platelet Incubator Setpoint
- 36 Calibrate the Temperature Controller Display
- 37 5 Maintenance
- 37 Recharge Refrigerant
- 38 i.Center Batteries
- 38 Motion Alarm Battery (PC4200i)
- 39 Clean the Platelet Incubator
- 39 Lubricate the Drawer Slides (PC4200i)
- 40 Lubricate and Service Internal Parts (PC4200i)
- 41 General Operation Problems
- 43 Chamber Temperature Problems
- 43 Alarm Activation Problems
- 45 Testing Problems
- 46 Condensation Problems
- 47 i.Series Parts
- 54 8 Schematics
- 54 PC100i, PC900i, PC1200i (100/115 V)
- 55 PC100i, PC900i, PC1200i (230 V)
- 56 PC2200i (115 V)
- 57 PC2200i (230 V)
- 58 PC3200i (115 V)
- 59 PC3200i (230 V)
- 60 PC4200i (115 V)
- 61 PC4200i (230 V)
- 62 i.Center Screen Reference
- 64 Horizon Seriesā¢ Models
- 64 Install Battery for Backup Power
- 64 Platelet Agitator and Platelet Incubator Compatibility
- 65 Install a Platelet Agitator in the Platelet Incubator
- 65 External Monitoring Devices
- 67 Drawer Labels
- 68 Temperature and Power Failure Alarm Volume
- 69 Motion Alarm (PC4200h)
- 70 Alarm Tests
- 73 12 Temperature Controller Setpoints
- 74 Level 1 Parameters and Values
- 75 Level 2 Parameters and Values
- 75 Level 3 Parameters and Values
- 76 Level 4 Parameters and Values
- 76 Change the Platelet Incubator Setpoint
- 76 Calibrate the Temperature Controller Display
- 78 13 Maintenance
- 78 Recharge Refrigerant
- 79 Power Failure Alarm Battery
- 79 Motion Alarm Battery (PC4200h)
- 80 Clean the Platelet Incubator
- 80 Lubricate the Drawer Slides (PC4200h)
- 81 Lubricate and Service Internal Parts (PC4200h)
- 82 General Operation Problems
- 84 Chamber Temperature Problems
- 84 Alarm Activation Problems
- 86 Condensation Problems
- 87 15 Horizon Parts
- 94 16 Schematics
- 94 PC100h, PC900h, PC1200h (100/115 V)
- 95 PC100h, PC900h, PC1200h (230 V)
- 96 PC2200h (115 V)
- 97 PC2200h (230 V)
- 98 PC3200h (115 V)
- 99 PC3200h (230 V)
- 100 PC4200h (115 V)
- 101 PC4200h (230 V)
- 102 Appendix A: Compliance
- 103 Appendix B: Warranty