Arjohuntleigh Maxi Sky 1000 Instructions For Use Manual
Advertisement
Advertisement
Maxi Sky
1000
Instructions for Use
001.14160.33.EN rev. 8 • December 2010
ArjoHuntleigh International AB.
© ArjoHuntleigh International 2010
ArjoHuntleigh products are patented or patent pending. Patent information is available by contacting
ArjoHuntleigh International AB.
Our policy is one of continuous development, and we therefore reserve the right to make technical alterations without notice. The content of this publication may not be copied either whole or in part without the consent of
ArjoHuntleigh International AB.
Table of Contents
Authorized European Representative ........................................................... 6
Definitions Used in this Manual: ................................................................. 6
Verifying the Package Contents ................................................................... 7
Human and Environmental Safety Practices ................................................... 11
Battery and Battery Charger Safety Practices ................................................. 11
Maxi Sky
1000
Ceiling Lift and Charger Station .............................................. 12
1000
Ceiling Lift............................................................. 15
Using the FIDO Function (Pre-Programmed Positions) ................................. 20
To Lift a Patient from a Chair ....................................................................... 21
3
Table of Contents
Procedure for Using Bariatric Loop Slings .................................................. 23
Procedure for Attaching Bariatric Loop Slings
to the Four-Point Spreader Bar ..................................................................... 23
Maxi Sky
1000
Programming ................................................................................... 24
Changing the Speed of Movement ............................................................... 24
Adjusting the Spreader Bar Height .............................................................. 25
Enable/Disable Return to Charge (RTC) ..................................................... 26
Inspections by an Authorized Service Technician .......................................... 28
Verification of the Charger’s Power Source ................................................ 31
Electromagnetic Compatibility .............................................................................. 39
4
General Information
Thank you for purchasing the
Maxi Sky 1000
ArjoHuntleigh.
ceiling lift from
Your Maxi Sky 1000 is part of a series of quality products designed specially for home care, nursing homes and other health care uses.
We are dedicated to serving your needs and providing the best products available along with training that will bring your staff maximum benefit from every
ArjoHuntleigh product.
Please read this manual thoroughly, and contact us if you have any questions about the operation or maintenance of your ArjoHuntleigh equipment.
Foreword
Please read this manual in its entirety before using your
Maxi Sky
1000
. The information in this manual is crucial to the proper operation and maintenance of the equipment, and will help protect your product as well as ensure that the equipment performs to your satisfaction. Lifting and transferring a person always presents a potential risk. Some of the information in this manual is important for your safety and must be read and understood to help prevent injuries. ArjoHuntleigh strongly advises and warns that to avoid injuries that can be attributed to the use of inadequate parts, only parts designated by ArjoHuntleigh should be used on equipment and other appliances supplied by
ArjoHuntleigh.
Furthermore, unauthorized modifications on any
ArjoHuntleigh equipment may affect its safety.
ArjoHuntleigh will not be held responsible for any accidents, incidents or deficiencies of performance that occur as a result of any unauthorized modification to its products.
Tested according to standards by TÜVProduct Service and
Service and support
A service routine must be performed on your Maxi Sky 1000 by qualified service personnel. This will ensure the safety and good functioning of your product. See section called
“Care and Maintenance” in this manual .
If you require further information, please contact your local
ArjoHuntleigh representative which can offer comprehensive support and service programs to maximize the long-term reliability, safety and value of the product.
Contact your local representative for replacement parts.
Additional copies of this manual can be purchased from your local ArjoHuntleigh representative. When ordering, include the User Manual product number ( see front page) and equipment identification number.
5
General Information
Manufacturer Information
This product has been manufactured by:
BHM Medical Inc.
2001 Tanguay Street
Magog (Quebec)
Canada J1X 5Y5
Authorized European Representative
Huntleigh Healthcare Ltd.
310-312 Dallow rd.
Luton, UK
LU1 1TD
Definitions Used in this Manual:
WARNING:
Operational Life
The equipment is designed and tested for a useful life of seven (7) years or 10 000 transfers—whichever comes first—subject to preventative maintenance as specified in the “Care and Maintenance” section in this manual. Time equivalence between the number of transfers versus the number of years is made clear
.
Transfers per Day
4
6
8
Fig. 1
Years
(10,000 transfers)
7
4.5
3.5
Means: Failure to understand and follow these instructions may result in injury to yourself and others.
CAUTION:
WARNING: The manufacturer cannot ensure full safety for a ceiling lift or an accessory of which the life span has been exceeded.
Means: Failure to follow these instructions may cause damage to the product.
NOTE:
Means: This is important information regarding the correct use of the equipment.
Intended Use
The Maxi Sky
1000
is designed for lifting patients in a homecare setting, at nursing homes and other assisted living centers. Patient transferring is performed under the supervision of an appropriately trained caregiver staff in accordance with the instructions outlined in this manual. All other uses must be avoided.
The equipment must only be used for the purposes stated above, and must be installed by ArjoHuntleigh authorized personnel and in accordance with local codes.
The red indicator light on the ceiling lift will blink when it is about halfway to its useful life, and again to indicate the end of the useful life period.
The operating life of this equipment corresponds directly to the safe operating time period before a complete overhaul is required. Aging of the cassette, frequency of use (transfers per day), the weight of the patient and maintenance frequency are factors that have an impact on the Maxi Sky
1000
’s life span. A transfer is defined as the displacement of a patient from one point to another. A transfer cycle includes a lifting and a descending action.
The expected operational life for fabric slings and fabric stretchers is approximately two years from date of purchase. This life expectancy only applies if the slings and stretchers have been cleaned, maintained and inspected in accordance with the
ArjoHuntleigh Sling Application Guide , the
Instructions for Use and the Preventive Maintenance
Schedule .
The expected life for other consumable products, such as batteries, fuses, lamps, slings, straps and cords is dependent upon the care and usage of the
6
General Information equipment concerned. Consumables must be maintained in accordance with published
Instructions for Use and the “Preventive
Maintenance Schedule”.
Equipment Identification
The unit's identification number (specification, model, serial number) appears on a silver nameplate attached at one end of the lift.
Verifying the Package Contents
Always ensure that the ceiling lift will be installed by a contractor or installer that has been authorized by
ArjoHuntleigh.
Upon receipt of the equipment, verify it against the packing list to ensure it is complete and inspect the equipment for possible damage due to shipping. If there is any damage, notify the carrier immediately to file a claim. Provide complete information concerning damage claims or shipping errors to your local ArjoHuntleigh representative. Include all equipment identification numbers and group part numbers (if any) as described above along with a full description of damaged parts.
How to Use this Manual
WARNING: Do not attempt to use this equipment without fully understanding the information contained in this manual.
To ensure the safe operation of the Maxi Sky
1000
, read the entire manual carefully, especially the
“Safety Instructions” section, before installing, operating, or servicing this equipment.
If anything is not completely understood, please contact your local ArjoHuntleigh representative for more details. Failure to comply with warnings in this manual may result in injuries.
Keep this manual with the lift and refer to it as required. Make sure that all operators are regularly trained in the use of the equipment according to the information found therein.
7
General Information
Symbols Used
Symbol Key to symbols
This symbol is accompanied by the name and the address of the authorized representative in the European Community.
This symbol is accompanied by a date to indicate the date of manufacture and by the address of the manufacturer.
This symbol indicates the products comply with the medical device directive
93/42/EEC.
This symbol is accompanied by the manufacturer's catalogue number.
This symbol is accompanied by the manufacturer's serial number.
This symbol indicates “separate collection” for all batteries and accumulators as per the WEEE Directive.
These symbols refer to the Instructions for Use .
This symbol indicates a class II electrical equipment: term referring to electrical equipment in which protection against electric shock does not rely on basic insulation only.
This symbol indicates a type B applied part.
This symbol indicates a risk of pinching
This symbol locates the emergency stop system on the lift
This symbol locates the emergency lowering system on the lift.
Fig. 2
8
Safety Instructions
The equipment must be used in accordance with these safety instructions.
Anyone using the equipment must also have read and understood the instructions in this manual.
If there is anything you are not sure about, contact your local
ArjoHuntleigh representative.
General Instructions
Keep these safety instructions with the ceiling lift at all times.
Read the instructions for use in this manual before installing, operating, or servicing this equipment.
WARNING: The Maxi Sky 1000 is for transferring patients only. Do not use the lift for any other purpose.
WARNING: Always place the sling around the patient according to the instructions enclosed.
Failure to do so may result in injuries to you or to others.
CAUTION: Do not drop either the ceiling lift cassette or the batteries, since they may cause internal damage that is not easily seen. If the ceiling lift is suspected to be damaged, contact your local ArjoHuntleigh representative for servicing.
Safe Working Load
The Maxi Sky 1000 has been designed with a lifting capacity of 454 kg (1000 lb).
WARNING: The Maxi Sky 1000 is intended to be used for patients whose weight is within a specified safe working load. Do not attempt to lift more than the lowest weight limit indicated on the following:
• the track system;
• the “maximum load” label on the
Maxi Sky 1000 ;
• on the spreader bar;
• on the accessories;
• on the sling.
9
Safety Instructions
Important Safety Directions
Always ensure that:
• The ceiling lift is installed by an authorized
ArjoHuntleigh contractor or installer.
• The equipment is used by trained staff.
• The track installation will accept a load equal to that of the ceiling lift.
• Before an attempt is made to move the patient, an assessment is performed by a qualified professional.
• You are prepared before attempting to transfer a patient.
• An assessment of the suitability for transfer of a patient who is connected to electrodes, catheters or other medical devices is conducted by qualified personnel before performing the transfer.
• Violent impact during transfers is avoided.
• The spreader bar being used is intended to be used with the Maxi Sky 1000 bearing the patient’s weight.
and is capable of
• The sling is intended to be used for this lift and can take the weight of the patient.
• The sling is not damaged, torn or frayed.
• The lifting procedures outlined in this manual are followed.
• All controls and safety features are used only according to the rules specified in this manual.
Never attempt to force a control or button on the lift.
• The charger is not stored in a shower, bath or other areas with high humidity.
• The sling straps are in good condition and properly fastened.
• The daily maintenance is carried out before using the lift.
• Any precautionary or instruction labels that cannot be easily read are replaced.
• If storing the Maxi Sky
80 % at 20 ºC (68 ºF).
1000 when it is not in use, the humidity in the storage area does not exceed
• The ambient temperature range when storing the
Maxi Sky
1000 to 122 ºF).
is between –10 ºC and 50 ºC (14 ºF
NOTE: ArjoHuntleigh ceiling lifts are specifically designed for KWIKtrak ceiling rail systems, and ArjoHuntleigh slings and accessories.
CAUTION: Keep all components of the lift clean and dry, and have electrical and mechanical safety checks done as instructed in the “Care and Maintenance” section of this manual
CAUTION: Excessive exposure of the handset to water (or other liquid) could cause malfunction of the device.
Shock Prevention
• Do not touch or use a lift with bare conductors or a damaged power cord. Electrically live equipment can result in serious injuries. If the lift or charger has any exposed or damaged wires, contact your local ArjoHuntleigh representative immediately.
• Do not splash or expose electric parts of the device to water or moisture.
• Check nameplate for input voltage and frequency requirements. These requirements differ by country. Do not attempt to use the lift in an area that has a different voltage and frequency requirement.
• Do not attempt to expose, service or repair the lift, battery or charger. If any unit is malfunctioning, contact your local
ArjoHuntleigh representative.
• Read batteries and charger instructions thoroughly before using or storing them.
10
Safety Instructions
Fire and Explosion Prevention
WARNING: Do not place or store the battery under direct sunlight or near a heat source. Do not expose the batteries or battery charger to flames. Do not use the charger in the presence of flammable anaesthetic gases.
• Do not short circuit the battery terminals.
• Do not incinerate the battery.
Human and
Environmental Safety
Practices
• Should the battery casing crack, allowing its contents to come into contact with skin or clothing, rinse immediately with water. If the contents comes in contact with the eyes, rinse immediately with plenty of water and seek medical attention.
• Inhalation of the contents can cause respiratory irritation. Sensitivity to nickel can cause allergic asthma to result. Seek out fresh air and medical attention.
• For recycling and disposal of the batteries, the rules according to the WEEE directive (Waste of
Electronic and Electrical Components) as well as local laws and regulations must be followed. If not they may explode, leak and cause personal injury. When returning batteries, insulate their terminals with adhesive tape. Otherwise, the residual electricity in used batteries may cause
fire or explosion. Fig. 3 below shows the symbols
for disposal and recycling.
Battery and Battery Charger Safety
Practices
WARNING: Do not expose the battery connector or the battery charger to water.
The charger is designed for dry areas only and for normal air humidity conditions.
• Be careful not to drop the batteries.
• Only use the charger that has been supplied with the equipment.
• Do not charge the batteries in an unventilated area.
• The charger must not be covered or exposed to dust.
• Do not crush, puncture, open, dismantle or otherwise mechanically interfere with the batteries.
• Do not store batteries at a temperature higher than
50 ºC (122 ºF).
Equipment Warning Labels
• Carefully read the labels on the battery and follow the instructions.
• Inspect all precautionary labels on the equipment.
Order and replace all labels that cannot be easily read.
Fig. 3
11
Part Designation
Maxi Sky
1000
Ceiling Lift and Charger Station
OR OR
Legend
1) Ceiling lift
2) Track
3) Up button
4) Red maintenance/overload light
5) Down button
6) Emergency lowering mechanism
7) Charging station
8) Green power light
9) Amber charging light
Fig. 4
10) Red emergency pull cord
11) Spreader bar
12) Safety latch
13) Allen key
14) Strap
15) Charger (cord-connected or wall-adapted)
16) Handset
12
Part Designation
The following refers to Fig. 4 on previous page:
• The amber charging light flashes while charging and goes solid when charge is finished.
• The green power light illuminates once the lift is on and ready for use; the green light flashes when the batteries are low.
• The red light illuminates to confirm that the lift is in the programming mode.
• The red light also illuminates in the normal mode when the lift goes into overheat protection caused by overuse.
• The red light flashes when servicing is required
(contact customer service).
Handset
The Maxi Sky 1000 ’s handset unit is used to operate the
ceiling lift. Each funtion is described in Fig. 5. The
UP and DOWN buttons raise or lower the spreader bar.
The “ PROGRAMMING MODE ” button allows you to modify the functions of the lift. Refer to the
“Maxi Sky 1000 information.
Programming” section for more
700.136XX HANDSET
UP button
LEFT button
RIGHT button
DOWN button
RETURN TO
CHARGE button
700-138XX
HANDSET
Chargers
The Maxi Sky 1000 units are equipped with either a cord-connected universal charging system
(700.15500) or a wall-adapted charger (700-15501) that can both be customized to fit the AC voltage
outlets where they are sold (see Fig. 4).
PROGRAMMING
MODE button
UP button
RIGHT button
DOWN button
RETURN TO
CHARGE button
PROGRAMMING
MODE button
Fig. 5
13
Sling
Compatible Sling
Bariatric sling
(MAA8000-L)
Fig. 6
For more information about sling use, please refer to the documentation that comes with your sling.
14
How to Use the Maxi Sky
1000
Ceiling Lift
WARNING: Always read the “Safety
Instructions” before using the
Maxi Sky
1000
.
• The track must be installed and modified only by ArjoHuntleigh authorized personnel and in accordance with local codes.
• All tracks must be closed with end stoppers or connected to other closed track components.
• Before use, make sure all end stoppers are in place and secured.
WARNING: Pay close attention to the safety of the patient as you press the control buttons. Before lifting the patient, make sure that all straps are attached to the spreader bar.
Make sure the sling is not caught on any obstructions (for instance, the wheelchair brakes or armrests).
Also make sure that the spreader bar is correctly attached to the ceiling lift. See
Fig. 7 below.
With split ring No split ring
NOTE: The unit will not lift or lower when it is in charging position.
To begin the transferring procedure:
1) Install the patient sling around the patient. See
“ Use of Slings” section in this manual for further information.
2) Move the ceiling lift directly over the patient. Use the LEFT and RIGHT buttons on the handset (see
3) Use the DOWN button on the handset to lower the spreader bar to a point below the chin of the patient (to avoid the risk of it striking the patient’s face due to a sudden movement).
If the patient is lying down, lower the spreader bar down near the patient’s thorax, then install the straps.
WARNING: Hold the ceiling lift spreader bar with one hand at all times when near a patient.
4) Attach the straps to the desired position. See “Use of Slings ” section.
Fig. 7
5) To lift the patient, press the UP button.
6) Make sure the patient is clear of any obstacles before moving him/her. Guide the patient’s legs if necessary.
7) When the patient is located above the desired point of transfer, press the DOWN button.
8) Use the handles on the back of the sling to position the patient when transferring into a chair.
Hold the handle firmly as the sling will tilt back to position the patient.
9) Once the patient is properly seated and the straps have slackened, remove the sling from around the patient, and off the ceiling lift.
10) Move the ceiling lift away from the patient.
11) When the Maxi Sky
1000
is no longer required, return the ceiling lift back to the charging station.
The charging indicator light on the clip-on charging station must be on for the ceiling lift to detect it.
Verify the green light on the charger station and the amber light (flashing or solid) on the ceiling lift to ensure that the charging function is working properly.
15
How to Use the Maxi Sky
1000
Ceiling Lift
NOTE: The Maxi Sky 1000 is equipped with a safety system to prevent a misuse of the ceiling lift. If the lift is used above the amount specified by the duty cycle, a heat detection system will signal the processor to block the lifting of a load until the temperature of the transmission cools down. In the meantime, it will still be possible to activate the horizontal displacement and down functions. When the overheat protection is engaged, the red light will stay on during the cooling period.
In addition, a buzzer will sound if the UP button is pressed. The cooling period is between 10 to 30 minutes depending on ambient conditions.
Return to Charge (RTC)
The RTC function is disabled by default. To activate this function, please refer to the “Maxi Sky 1000
Programming” section in this manual.
To engage the RTC function, press on the return to charge button on the handset for 3 seconds. The spreader bar will raise all the way up to avoid any obstacles during the run. When the lift is at the charging station, the spreader bar lowers by itself to the preselected height so as to be easily accessible.
NOTE: You can stop the return to charge at any time by pressing any button on the handset or pulling on the red emergency cord. Moreover, a weight detector prevents the RTC from functioning while a person is in the lift.
Emergency Stop (red cord)
The emergency stop can be activated at any time to stop the functioning of the ceiling lift.
1) To stop the ceiling lift in any emergency, gently pull the red emergency cord once, until you hear a “click” (see Fig. 4 in the “Part Designation” section of this manual).
CAUTION: Do not pull the red emergency cord with excessive force. If the cord is jerked too hard, the ceiling lift may become inoperable.
2) To reactivate the ceiling lift, pull the red cord gently once. A green light confirms that the
Maxi Sky 1000 is on and ready for use.
Emergency Lowering
In the unlikely event of an electrical failure, the
Maxi Sky 1000 has an emergency manual lowering feature.
CAUTION: The Emergency lowering feature is to be used only in case of emergency.
If the ceiling lift malfunctions when a patient is being transferred, the emergency lowering device provides a safe way of getting the patient down onto a chair, bed or wheelchair (see
1) Pull the red emergency cord .
2) Remove the little black cap located on each side of the lift.
3) Find the 8 mm Allen key that came with the ceiling lift.
Insert the Allen key deep into the axle .
Turn it counterclockwise in order to lower the patient. Each turn of the key will slowly lower the patient. Repeat on the other side of the lift.
4 Turn the Allen key counter-clockwise to slowly lower the patient. Repeat on the other side of the lift.
Once the patient is lowered and secure within a chair, bed or wheelchair, call a qualified technician to inspect the ceiling lift.
Emergency Brake
The emergency brake is an additional safety feature that automatically prevents the patient from falling in the unlikely event that of a transmission or motor failure.
16
How to Use the Maxi Sky
1000
Ceiling Lift
Access emergency lowering system.
Turn Allen key counterclockwise
Pull red emergency cord.
Fig. 8
17
How to Use the Maxi Sky
1000
Ceiling Lift
Battery Information
The life cycle (number of charging cycles) of the batteries is largely dependent on the depth of discharge within each cycle. The more the batteries are drained, the shorter their overall life span. The life of the batteries is also related to such factors as varying temperatures and rest periods between when they are charged and discharged.
NOTE: To prolong battery life, return the ceiling lift to the charger whenever the ceiling lift is not in use. If the low battery indicator beeps, and a green light flashes, be sure to recharge the batteries as soon as possible. Charge the batteries until the charging indicator light is a solid amber before using the lift again. This will extend the life of the batteries.
NOTE: ArjoHuntleigh uses sealed leadacid batteries. These batteries do not have any memory effect. Therefore, batteries should not be completely drained before recharging.
CAUTION: Never leave a ceiling lift with the power on for an extended period of time without returning it to the charger. The batteries will be drained and damaged.
See Fig. 9 for a graph illustrating the relationship
between the number of lifts versus the load being lifted.
Indicator Lights
The ceiling lift and the charging system have many indicator lights. It is important to understand their significance for the proper use and comprehension of
the ceiling lift (see Fig. 10).
Charging the Batteries
WARNING: Do not operate the charger unit with a damaged cord or if the unit has been dropped or damaged.
Do not bend the power cord by force, or place a heavy object on it. This will damage the cord and may cause fire or electrical shock.
The steps for recharging the batteries are as follows:
1) Press the return to charge button on the handset.
A protection device prevents the ceiling lift from returning to the charger if a patient is in the ceiling lift during this operation.
2) The green indicator light will begin flashing if batteries are low and need to be recharged. If the light does not turn on, verify the
“Troubleshooting” section of this manual.
3) The amber indicator light will flash when the ceiling lift cassette has returned to the charger and the batteries are being charged. Once the charge is complete the light will stop flashing and become a solid amber light.
WARNING: Batteries need to be charged for a minimum of 8 hours prior to the initial use of the lift.
18
How to Use the Maxi Sky
1000
Ceiling Lift
Fig. 9
Green power light
Amber charging light
Red maintenance/overheat light
OPERATING THE LIFT
Flashing Low batteries
Solid The lift is on and ready to use
STATE OF BATTERIES
Flashing In process of charging batteries
Solid
MAINTENANCE
Charging done; batteries charged
Flashing Maintenance required by your local representative
Solid
Lift was powered up in “service mode” or is under overheat protection
CHARGING
Clip-on charger station indicator Solid green
Solid red
Clip-on charging station power is on
Problem with charger; do not use the charger
Fig. 10
19
How to Use the Maxi Sky
1000
Ceiling Lift
If the batteries have been completely drained it could take up to 6-8 hours to completely recharge them. In order to reduce the charging time, refrain from completely draining the batteries and leave the ceiling lift on the charging station between uses.
Use of Slings
Spreader Bar
WARNING: The Maxi Sky 1000 comes with a four-hook spreader bar and must be used only with BHM bariatric slings, as indicated in the list of accessories.
Programming the Ceiling
Lift
The Maxi Sky
1000
can be programmed so that the user can easily adjust the speed of the horizontal movement, the spreader bar height and the return to charge parameters of the ceiling lift. To learn about the programming mode, see the “Maxi Sky
1000
Programming Guide” on pages 24-26 of this manual.
WARNING: Spreader bars must only be installed by a qualified person.
Using the FIDO Function
(Pre-Programmed
Positions)
WARNING: This function can only be activated by a qualified technician.
1) Using the pre-programmed positions allows the unit to detect stations (positions) located along the track path. The unit will go to the indicated position by using the handset.
2) To do so, indicate to the unit the desired position to go to. For an example, to go to the third station push three times on the PROG button followed by the direction ( LEFT or RIGHT buttons).
Then the unit will go by itself to the requested station and stay on hold there.
If the position you asked for is not correct, you may stop the ceiling lift at any time by pressing any button on the handset. From this position, reprogram the ceiling lift to the new desired position. Count the number of stations from where you are and indicate the direction to go.
WARNING: Before using the Maxi Sky 1000 , always ensure the strap attachment pin is installed correctly through the spreader bar socket and lift strap, and that the split ring is correctly inserted through the hole
Before Approaching the Patient
Attendants should always tell the patient what they are about to do. Make sure to have on hand a sling that is of the correct model and of adequate size for transfer with the Maxi Sky 1000 unit.
To Lift a Patient from a Bed
The lifting technique described here can be used for reclining patients regardless of where they may be lying: on the bed or on the floor of the exercise area.
WARNING: It is vital to always check that the sling attachment loops are correctly attached and remain in tension as the weight of the patient is gradually taken up.
1) Roll the patient towards you, fold the sling in half and place it on the bed, along the back of the
patient (see Fig. 11 and Fig. 12).
Ensure that the sling is positioned so that when rolled back, the patient will lie in the center of the sling. Also check that the head support area of the sling lies directly beneath the neck of the patient.
2) Roll the patient back onto the sling and then slightly further in the opposite direction, so that the folded part of the sling can be unfolded from beneath the patient..
20
How to Use the Maxi Sky
1000
Ceiling Lift mean lowering the frame a little more.
When lifting from the bed, some operators prefer to connect the leg strap loops first. This in particular applies to patients with large thighs. To use this method, raise the hip and knee into maximum flexion, and hook up the leg strap loops.
6) Ensure that all four points of the sling are securely attached to the frame, then lift the patient using the handset, positioning him/her comfortably for transfer.
Fig. 11
Fig. 12
NOTE: If the patient can sit easily, then the sling can be positioned in the same manner as if the patient was seated in a chair. See
“ To lift from a Chair ” section.
3) Position the Maxi Sky 1000 patient.
directly over the
4) Using the handset, lower the spreader bar and connect the sling shoulder strap loops, taking care not to lower the frame onto the patient.
5) The straps must be brought under the thighs. This may involve lifting one leg at a time, and may
WARNING: To avoid the patient swinging excessively during lifting, the lift strap should be as vertical as possible, immediately prior to the lift.
Proceed with the patient transfer to a chair as in section “To lift from a Chair”, or place the patient onto a bed, (reversing the above procedure), by moving and then lowering the patient into the desired position.
When the patient’s body weight is fully supported, detach the leg strap loops, then the shoulder connections.
Move the spreader bar away before removing the sling from under the patient.
To lift a Patient from a chair
The lifting techniques described here, and represented in the images in following pages, can be used for seated patients regardless of where they may be seated (e.g. on the edge of a bed, in a chair, wheelchair or similar).
1) Place the sling around the patient so that the base of his/her spine is covered, and the head support area of the sling is behind the head.
Pull each leg strap under the thigh so that it
emerges on the inside of the thigh (Fig. 13).
21
How to Use the Maxi Sky
1000
Ceiling Lift
3) During the transfer, turn the patient to face the direction of travel, and keep at a normal chair height This affords the patient a measure of confidence and dignity. The patient should not be lifted above the caregiver’s eye-level; this can lessen the impression of height that may confuse the patient. Support the patient to stop excessive swinging and to give a feeling of added security.
4) Proceed with the patient transfer to a chair, or place the patient onto a bed, by moving and then lowering the patient into the desired position
Fig. 13
2) Ensure that all four strap loops are securely connected, then raise the patient using the handset.The head support area of the sling should now come into use. This is the most comfortable position for the patient transfer, as it reduces pressure on the patient’s thighs. The angle of recline can be increased for the comfort of more restless patients, when lowering the patient back into a chair
or transferring to a chair from the bed (Fig. 14).
Fig. 15
Fig. 14
22
How to Use the Maxi Sky
1000
Ceiling Lift
To Lift a Patient from the Floor
WARNING: Always check that the sling attachment loops are correctly attached and remain in tension as the weight of the patient is gradually taken up.
The patient can be lifted from a completely reclined position on the floor, but for his/her comfort, a pillow can be placed under his/her head first.
When lifting a patient from the floor it is essential to ensure the patient is situated directly underneath the
Maxi Sky
1000 suspended.
cassette, to avoid excessive swinging or dragging as the weight of the patient becomes
1) Place the sling around the patient as in section
“Lifting from a Bed” (see Fig. 11 and Fig. 12).
2) Attach the shoulder strap loops first, then connect the leg strap loops. Adjustment of the spreader bar height may have to be made before connection is possible.
When lifting from the floor, some operators prefer to connect the leg strap loops first (this in particular applies to patients with large thighs). To use this method, raise the hip and knee into maximum flexion and hook up the leg strap loops.
3) Attach the shoulder strap attachment loops. It will probably be easier to attach the longest loops.
4) When all the connectors are securely attached, raise the patient from the floor in a semi-
recumbent position (see Fig. 16).
Note: To avoid the patient swinging excessively during lifting, the lift strap should be as vertical as possible, immediately prior to the lift.
The patient may then be positioned in a chair, or placed on a bed. patients with extensor spasm may be lifted by the Maxi Sky 1000 , but special attention should be paid to supporting the patient's legs during the early part of the transferring process.
Procedure for Using Bariatric Loop
Slings
The slings to be used with the four-point spreader bar are the ArjoHuntleigh bariatric loop slings. They are available in four sizes (medium, large, extra large and XXL) all colour coded. A range of more specialised slings are available. Please contact
ArjoHuntleigh or their authorised distributors for details. The correct size sling will support the shoulders.
Procedure for Attaching Bariatric
Loop Slings to the Four-Point
Spreader Bar
WARNING: When attaching a loop sling to the four-point spreader bar, always ensure the sling attachment loops are positioned correctly into the
retaining hooks (see Fig. 17).
When the patient has been returned to the bed he/she maybe reclined before the sling is detached from the spreader bar.
Fig. 16
Fig. 17
23
Maxi Sky
1000
Programming
Entering the Programming Mode Speed Adjustment
Moving speed of
10 cm/s
Final Steps
To select another feature
Moving speed of
15 cm/s
Moving speed of
20 cm/s
(factory default)
Moving speed of
25 cm/s
Exit programming mode
Fig. 18
Changing the Speed of Movement
Turn the ceiling lift off by pulling on the red cord. The green LED will turn off.
Press the PROG button on the handset. At the same time, pull the red cord once again. The green light will flash, and you will hear three beeps.
The red LED will then illuminate. You can now release the PROG button.
Press once the UP button (you will hear one beep) for speed adjustment feature.
Now press the PROG button to confirm the selection.
Using the LEFT button, select one of four predetermined speeds.
Now press the PROG button to confirm the selected speed.
Press UP to continue within the programming mode, or to return to regular mode, pull on the red cord to turn the ceiling off. Pull the red cord once again. The lift will return to normal use.
24
Maxi Sky
1000
Programming
Entering the Programming Mode Spreader Bar Height
Adjustment
Final Steps
To select another feature
Exit programming mode
Fig. 19
Adjusting the Spreader Bar Height
Turn the ceiling lift off by pulling on the red cord. The green LED will turn off.
Press the PROG button on the handset. At the same time, pull the red cord once again. The green light will flash, and you will hear three beeps.
The red LED will then illuminate. You can now release the PROG button.
Press twice the UP button (you will hear two beeps)to change the spreader bar height.
Now press the PROG button to confirm the selection.
The strap will begin winding up. Once it is completely wound, press the UP and DOWN buttons to set the height the spreader bar’s rises to once it is sent to the charging station.
Now press the PROG button to confirm the selected height.
Press UP to continue within the programming mode, or to return to regular mode, pull on the red cord to turn the ceiling off. Pull the red cord once again. The lift will return to normal use.
25
Maxi Sky
1000
Programming
Entering the Programming Mode Set Return to Charge
Option
Return to the last charging station travelled.
Disable
(factory default)
Enable left
Final Steps
To select another feature
Exit programming mode
Enable right
Fig. 20
Enable/Disable Return to Charge (RTC)
Turn the ceiling lift off by pulling on the red cord. The green LED will turn off.
Press the PROG button on the handset. At the same time, pull the red cord once again. The green light will flash, and you will hear three beeps.
The red LED will then illuminate. You can now release the PROG button.
Press three times on the UP button (you will hear three beeps) to enable/disable the RTC.
Now press the PROG button to confirm the selection.
Using the left button, select one of the four predetermined selection.
Now press the PROG button to confirm the selected return to charge.
Press UP to continue within the programming mode, or to return to regular mode, pull on the red cord to turn the ceiling off. Pull the red cord once again. The lift will return to normal use.
26
Care and Maintenance
Preventive Maintenance Schedule
The equipment is subjected to wear and tear, and the following maintenance instructions must be acted upon when specified to ensure that the equipment remains within its original manufacturing specifications. Care and maintenance must be carried out in accordance with the preventive maintenance schedule below.
Customer obligations must be carried out by qualified personnel in accordance with the instructions in this manual.
WARNING: The maintenance described in the following checklist is the minimum that the manufacturer recommends. In some cases more frequent inspections should be carried out. Continuing to use this equipment without conducting regular inspections or when a fault is found will seriously compromise the safety of the user and of the patient. Local regulations and standards may be higher than those of the manufacturer. A load test is recommended. Service and preventive maintenance can be arranged with the manufacturer. Preventive maintenance specified in this manual can prevent accidents and reduce repair costs.
WARNING: Safety related maintenance and authorized service must be carried out by qualified personnel, fully trained in servicing procedures by
ArjoHuntleigh, and equipped with correct tools. Failure to meet these requirements could result in personal injuries and/or unsafe equipment.
User Inspections
Inspections for lift cassette and track system
Inspect for evidence of external damage, missing parts or broken panels.
Make sure that end stoppers and rail caps are in place and tightened.
Inspect strap for wear, discoloration or loose threads.
Recharge batteries.
Inspect the spreader bar on the strap of the lift for damage or cracks. Make sure all attachments are properly fixed
(split ring).
Initially
X
X
Before every use
FREQUENCY
Every two months or
500 cycles
Every four months or
1000 cycles
X
Every year or
2500 cycles
Every two years or
5000 cycles
X
X
X
X
X
27
Care and Maintenance
Inspections for lift cassette and track system (continued)
Inspect wheels in rail for damage, rust or cracks.
Replace if damaged.
Clean the rail and the clip-on charging station contacts.
Overall inspection by authorized personnel.
Verify emergency stop cord.
Verify emergency lowering device.
Initially
Before every use
FREQUENCY
Every two months or
500 cycles
Every four months or
1000 cycles
Every year or
2500 cycles
Every two years or
5000 cycles
X
X
X
X
X
Inspections for spreader bar and slings
Initially
FREQUENCY
Before every use
Every two months or
500 cycles
Every four months or
1000 cycles
Every year or
2500 cycles
Every two years or
5000 cycles
Inspect all sling parts
(attachments, fabric, stitch areas and strap) for signs of wear, discoloration, deterioration or loose threads.
Clean sling as indicated on the tag.
Inspect the spreader bar on the strap of the lift for damage or cracks. Make sure all attachments are properly secured (e.g. split ring).
X
X
When necessary
X
Inspections by an Authorized Service Technician
Inspection for lift cassette
Replace strap.
Inspect frame parts interlock and hardware for malfunction and make sure there are no parts missing.
Initially
Before every use
FREQUENCY
Every two months or
500 cycles
Every four months or 1000 cycles
Every year or
2500 cycles
Every two years or
5000 cycles
X
X
28
Care and Maintenance
Inspection for lift cassette
Inspect gears for wear.
Inspect connecting joints for proper attachment (trolley and spreader bar).
Verify that the emergency brake on the drum is turning freely.
Verify the emergency brake.
Verify emergency lowering mechanism.
Load test with the SWL (safe working load) recommended.
Inspect track end stoppers.
Initially
Before every use
FREQUENCY
Every two months or
500 cycles
Every four months or 1000 cycles
Every year or
2500 cycles
X
Every two years or
5000 cycles
X
X
X
X
X
X
Inspections for rails
Torque end stoppers to 20 N. m.
(15 lbf.ft).
Make sure that the bracket locking device is not visible.
Make sure rail joints are closed and that the spring pins are centered.
Make sure the rail is straight when it is not loaded.
Make sure the adjusted load setting of the lift is equal or lower than the safe working load of the installation.
Check that the accessories
(turntable and exchanger) are complete and correctly maintained.
Make sure that the attachments
(ceiling brackets, wall post, wall brackets) have not been displaced.
Inspect track end stoppers.
Inspect and tighten hardware (if necessary).
Initially
X
X
X
X
X
X
X
Before every use
FREQUENCY
Every two months or
500 cycles
Every four months or 1000 cycles
Every year or
2500 cycles
X
Every two years or
5000 cycles
X
X
X
X
X
X
X
NOTE: If the product does not work as intended, immediately contact your local ArjoHuntleigh representative for support.
29
Care and Maintenance
Daily Checklist
The following procedures must be followed before each use:
• Charge the batteries. Park the lift on the charger station whenever the lift is not in use.
• Inspect the lift for any damage. If the lift casing does not look properly aligned, or there are any cracks or other damage on the lift, or there are parts missing, do not use it. Contact your Arjo-
Huntleigh representative to have the lift serviced.
• Inspect the strap for any visible signs of wear, frays, loose threads or other damage. If there is any evidence of damage, do not use it. Contact your local ArjoHuntleigh representative to have the lift serviced.
• Inspect the sling for tears, frayed straps or loose stitching. If the sling has any of the above damage, do not use it. Contact your local ArjoHuntleigh representative to have the sling replaced or repaired.
• Inspect the spreader bar for any signs of cracking or damage.
• Ensure the split ring and cotter pins that attach the spreader bar to the strap are secured.
WARNING: Always reinstall the rail end stoppers (if removed) after servicing.
Strap Inspection
If the strap is damaged or shows signs of wear or discoloration, the acceptable load on the strap before rupture can drop rapidly and present a danger for the patient or caregiver. ArjoHuntleigh recommends a thorough inspection of the straps every 2 months as follows:
1) Completely unwind the strap.
2) Look for any signs of wear or discoloration (see
Loose threads in stitched areas
Noticeable discoloration
(strap color is lighter than color in the stitched area)
Edge wear (fraying)
Middle wear
WARNING: Before each use, make sure all end stoppers are in place and secured.
Inspection and Cleaning
To clean the Maxi Sky
1000
, wipe it down with a damp cloth using warm water and a disinfectant cleaner. Disinfectant wipes, supplied already impregnated with a 70% v/v solution of isopropyl alcohol, can also be used.
Rub the lift vigorously when using the wipes, to promote an effective disinfection of its entire surface.
Do not use phenol, chlorine or any other type of solvent that may damage the finish.
To ensure a better rolling surface for the trolley wheels, clean the inside of the track every 4 months.
To do so, insert a damp cloth in the opening and slide it from one end of the track to the other.
Fig. 21
WARNING: If there is any sign of wear as indicated above or any other visual defects, the strap must be inspected by qualified personnel and changed if required. By continuing to use the lift without changing a damaged strap, the safety of the caregiver or patient is greatly compromised.
NOTE: The manufacturer recommends changing the strap at least every two years.
By continuing to use the lift without changing the strap, the safety of the caregiver or patient is greatly compromised.
30
Care and Maintenance
Handling and Storage
Avoid violent impacts while transporting the lift.
The lift should not remain stored for long periods of time without recharging the batteries.
2) Make sure that there is contact between the contact blades of the lift and the contact plates of the charging station.
3) Check the power of the AC outlet on the wall.
4) If the charge station’s green light does not light up, contact your local ArjoHuntleigh representative for assistance.
NOTE: Even if the lift is not used,
ArjoHuntleigh recommends charging the batteries at least every two weeks. This will prevent premature aging of batteries.
If you store or ship the Maxi Sky 1000 , ensure that the power is turned off (green light) beforehand.
Sling Inspection and
Care
For maximum patient safety and hygiene, read the following instructions:
Battery Replacement
ArjoHuntleigh uses sealed lead-acid batteries in the
Maxi Sky 1000 ceiling lifts. ArjoHuntleigh batteries do not have any memory effect. Therefore, batteries should not be completely discharged before recharging.
Verification of the Charger’s Power
Source
If the light does not illuminate when there are batteries correctly installed in the Maxi Sky
1000
, try the following:
Regular Inspections
It is essential that the slings, their straps, and loops are carefully inspected before each and every use. If the slings, loops or straps are frayed, the sling should be withdrawn from use immediately and replaced.
Replace the battery when there is a noticeable reduction in the number of transfers that can be performed between charges. If you hear the
Maxi Sky 1000 lift beeping and notice a red light flashing, see the instructions in the
“Troubleshooting” section of this manual to determine if it is a problem with the battery.
To replace batteries, be sure to contact your local
ArjoHuntleigh representative.
CAUTION: Do not attempt to use a battery that was not supplied by ArjoHuntleigh.
ArjoHuntleigh batteries are specially designed for ArjoHuntleigh charging systems. Attempting to use an unauthorized battery may seriously damage the lift and/or the charger.
Sling Laundering
WARNING: The slings should be checked before and after use and, if necessary, washed according to instructions on the sling. This is especially important when using the same equipment for another patient. This minimizes the risk of cross infection.
Before washing the slings equipped with head support pockets, always remove the plastic reinforcement inserts. Always refit the inserts before reusing the sling.
Mechanical pressure should be avoided during the washing and drying procedure (e.g. rolling or pressing), as these can damage parts vital to the safe and comfortable operation of the sling.
The stretcher cross straps and suspension straps should be checked and washed if necessary. Washing and drying temperatures must not exceed 80°C
(176°F). Wash using normal detergents and do not iron.
1) Make sure that the charger is correctly plugged into the AC outlet, and that the green light on the clip-on charging station is on.
31
Care and Maintenance
NOTE: With regard to laundering, slings should not be classified as linen, but as an accessory to a patient transfer lifting and therefore classified as a medical device.
Slings should be cleaned and disinfected only in strict accordance with the manufacturer’s instructions.
Annual Inspection
The Maxi Sky 1000 and its accessories must be inspected annually by a qualified technician.
Sky
1000
and accessories must be serviced every 12 months as a minimum requirement. Do not attempt to do the inspection unless you are qualified to do so.
Maintenance Requirements
The Maxi Sky
1000
is equipped with an electronic monitor that causes a red light to flash when a maintenance inspection is necessary. Arranging for scheduled inspections ensures the durability of the unit and the security of the patient and user.
Once this red light begins to flash, please contact your local ArjoHuntleigh representative in order to perform the necessary maintenance inspection.
32
Troubleshooting
WARNING: Do not attempt to open the Maxi Sky 1000 ceiling lift cassette. Only a qualified technician is authorized to open it.
Alterations made to the Maxi Sky 1000 ceiling lift cassette by someone other than a qualified technician may cause serious injury.
PROBLEM
The red “service” light is on and flashing.
TO CHECK
The red light is solid.
• Contact your local ArjoHuntleigh representative to do maintenance.
• The ceiling lift cassette is under its overheat protection. Wait between
10 to 30 minutes until the red light turns off and press on the UP button to use the ceiling lift cassette again.
• If the load is over a safe working load, the unit will not work due to the overload protection on the motor.
The unit starts and stops repetitively.
The ceiling lift cassette emits a beep during utilization. The unit may stop lifting but the lowering function can still be used.
The charger indicator (amber) on the ceiling lift cassette does not light up when the lift is on the charger.
When returning to charge, the ceiling lift cassette passes the clip- on the charging station, or goes in the wrong direction.
Batteries are always dead after only a few transfers (3 to 5).
•
• Batteries are low. Return the ceiling lift cassette to the charging station.
Check that the charger is plugged into a standard outlet, and that the outlet has power. The green light on the clip-on charging station indicates that it is functioning.
• Clean the contact blades of the charging station with mild detergent.
Pass the ceiling lift cassette through the charging station manually once, then retry the return to charger function.
• Verify the function of the ceiling lift charger and the contact plates on the clip-on charging station.
• Replace batteries with new ones. The life of the current batteries could be almost finished. It is important to always change both batteries at the same time. Contact your local ArjoHuntleigh representative to have the batteries replaced.
The amber light on the unit is solid, yet the ceiling lift cassette will only perform one or two transfers.
The amber light on the unit is solid, yet the ceiling lift will only work when there is no one on the lift.
When you try to transfer someone, the ceiling lift stops.
• Contact your local ArjoHuntleigh representative to have batteries replaced.
• Contact your local ArjoHuntleigh representative to have the batteries replaced.
33
Troubleshooting
The charging light on the ceiling lift cassette continues flashing amber and the light does not turn solid even after recharging the unit overnight.
PROBLEM
The ceiling lift does not work when you press the buttons on the handset.
When you press the button to return the ceiling lift to its charger, the ceiling lift goes past the charger.
TO CHECK
• If the charging station light is on, move the ceiling lift away from the charging station in order to operate the lift.
• If the emergency stop is activated, gently pull the red cord once until the green light is on.
• Check if the buttons on the ceiling lift cassette are working. If so, the problem may be coming from the handset. If not, check the charge on the ceiling lift.
• Check if the handset is plugged in properly into the ceiling lift cassette. The handset may be slightly pulled out from its socket and yet appear as though it is plugged in. Pull out the plastic cover to check the connection.
• Slide the ceiling lift over the clip-on charging station. Verify if the amber light turns on.
• If, after testing all of the above, the ceiling lift will not operate, contact your local ArjoHuntleigh representative.
• If available, try another integrated clip-on charging station from another ceiling lift, or a spare one; clip it to the rail and charge the unit for 3 hours. If the amber light is still flashing, contact your local Arjo-
Huntleigh representative.
• If, after testing all of the above, the ceiling lift will not operate, contact your local ArjoHuntleigh representative.
• The charger either plugged inproperly to an electrical outlet or is not working properly. Contact your local ArjoHuntleigh representative.
34
Labels on the Lift
1) Maximum capacity/Service telephone number
2) Date of manufacture, serial number, product code/
Manufacturer identification
3) Product name
4) Emergency lowering system access identification
5) Emergency stop identification
6) ArjoHuntleigh label
Fig. 22
35
Technical Specifications
PRODUCT INFORMATION Maxi Sky
1000
Weight, complete
Lifting capacity
Strap length
Lifting speed
Maximum stroke (from ceiling)
Horizontal displacement speeds
Horizontal axis motor
Vertical axis motor (qty 2)
ELECTRICAL
22 kg (48 lb)
454 kg (1000 lb)
2300 mm (90.6 in)
6.0 cm/s (2.4 in/s) without load)
3.5 cm/s (1.4 in/s) at 230 kg (500 lb)
2.7 cm/s (1.2 in/s) at 454 kg (1000 lb)
2300 mm (90.6 in)
10, 13, 15 and 20 cm/s. Speed is 15 cm/s (6 in/s) by default
24 VDC at 62 watts
24 VDC at 110 watts
Duty cycle
Rating
Noise level for either raising or lowering, with or without load
Max 10%, 1 minute continuously
24 VDC, 15 A max.
61 dBA max.
Medical equipment Type B protection against electrical shock in accordance with IEC 60601-1
The Maxi Sky
1000
is compliant with CAN/CSA-C22.2, CSA-Z323.5.98, IEC 60601-1, UL 60601-1 AND ISO 10535.
WARNING: Radio transmitting devices such as mobile telephones, two-way radios, etc., should never be used near the Maxi Sky 1000 , since they can interfere with the function of the lift. Cables from potentially strong sources of electromagnetic fields should not be placed near the unit. See “Electomagnetic Compatibility” section of this manual for full details.
Battery type
Battery capacity
Degree of protection - handset
Degree of protection - Maxi Sky 1000
Lift - protection class - shock prevention
Battery Charger input
Battery Charger output
Sealed rechargeable valve regulated lead acid battery
Constant voltage charge
Cycle used 14.1 - 14.4 V
Standby use: 13.5 - 13.8 V
Initial current: Less than 2.00 A
Rating: 12 V, 5 Ah
Provides up to 80 transfers with a load of 230 kg (500 lb), up to 25 transfers with a load of 454 kg (1000 lb).
IPX4
IP23
Internally powered equipment
100-240 VAC, 50-60 Hz
28.1 VDC, 1 A max
Battery Charger safety protection Class 2, double insulated
OPERATION AND STORAGE CONDITIONS
Ambient temperature range -10
°
C to 50
°
C (14
°
F to 122
°
F)
Relative humidity range 10 % to 80 % incl. condensation
Atmospheric pressure range Operation: 700 hPa to 1060 hPa Storage: 500 hPa to 1060 hPa
36
Technical Specifications
RECYCLING
Battery
Package
The lift
WARNING: This equipment is not suitable in the presence of flammable anesthetic mixtures with air or oxygen, or with nitrous oxide.
Sealed lead-acid, rechargeable, recyclable
Cardboard recyclable
Separated and recycled, according to the European Directive 2002/96/EG
(WEEE).
37
Technical Specifications
Lift Dimensions
Fig. 23
38
Electromagnetic Compatibility
Electromagnetic Compliance
The Maxi Sky
1000 has been tested for compliance with current regulatory standards regarding its capacity to block
EMI (electromagnetic interference) from external sources.
Nonetheless, some procedures can help reduce electromagnetic interferences:
• Use only ArjoHuntleigh cables and spare parts to avoid increased emissions or decreased immunity which can compromise the correct functioning of the equipment.
• Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted emissions standards.
• Maximize the distance between electro-medical devices. High-powered devices may produce EMI that can affect the ceiling lift. Refer to separation distance table further on in this document.
For more information on how to manage the unit’s RF electromagnetic environment, please consult the
AAMI TIR 18-1997 - Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical
Engineers .
Electromagnetic Emissions
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions - For all Equipment and Systems
The Maxi Sky
1000
is intended for use in the electromagnetic environment indicated below. The customer or the user of the Maxi Sky
1000
should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Maxi Sky
1000
uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Class B
Not applicable
Not applicable
The Maxi Sky
1000
is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
39
Electromagnetic Compatibility
Electromagnetic Immunity
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity - For all Equipment and
Systems
The Maxi Sky
1000
is intended for use in electromagnetic environment specified below. The customer or the user of the Maxi Sky
1000
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-5
Surge
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV for input/output
±6 kV contact
±8 kV air
±2 kV for power supply lines
Not applicable for input/ output lines
Not applicable
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
IEC 61000-4-5
<5% U T
(>95% dip in U T) for 0.5 cycle
Voltage dips, short interruptions and voltage variations on power supply input lines
40% U T
(60% dip in U T) for 5 cycles
70% U T
(30% dip in U T) for 25 cycles
Not applicable
IEC 61000-4-11
<5% U T
(>95% dip in U T) for 5 sec.
Power frequency
(50/60 Hz) magnetic field
3 A/m 3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercials or hospital environment.
NOTE: U T is the AC mains voltage prior to application of the test level.
40
Electromagnetic Compatibility
(continued)
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity - For Equipment and
Systems that are Not Life-Supporting
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
Mhz
3 Vrms
150 kHz to 80
Mhz
Portable and mobile RF communications equipment should be used no closer to any part of the Maxi Sky
1000
, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance: d =
3.5
V 1
P
Radiated RF
IEC 61000-4-3 d =
3.5
E 1
P
80 MHz to 800 MHz
3 V/m
80 MHz to 2.5
GHz
3 V/m
80 MHz to 2.5
GHz d =
E
7
1
P
800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
(a)
should be less than the compliance level in each frequency range.
(b)
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: Theses guidelines may not apply in all situations. Electromagnetic propagation if affected by absorption and reflection from structures, objects and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Maxi Sky
1000 is used exceeds the applicable RF compliance level above, the Maxi Sky
1000
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Maxi Sky
1000
.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
41
Electromagnetic Compatibility
(continued)
Recommended Separation Distance Between Portable and Mobile RF Communications
Equipment and the Maxi Sky
1000
- for Equipment and Systems that are not
Life-Supporting
Recommended separation distances between portable and mobile RF communications equipment and the
Maxi Sky
1000
.
The Maxi Sky
1000 is intended for use in electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Maxi Sky
1000 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the
Maxi Sky
1000
as recommended below, according to the maximum output power of the communications equipment.
Separation distances according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Rated maximum output power of transmitter
W
d =
3.5
V 1
P d =
3.5
E 1
P d =
E
7
1
P
0.01
0.1
1
10
0.12
0.38
1.2
3.8
0.12
0.38
1.2
3.8
0.24
0.74
2.4
7.4
100 12 12 24
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
42
AUSTRALIA
ArjoHuntleigh Pty Ltd
78, Forsyth street
O’Connor
AU-6163 Western Australia
Tel: +61 89337 4111
Free +1800 072 040
Fax: + 61 89337 9077
AUSTRIA
ArjoHuntleigh GmbH
Dörrstrasse 85
AT-6020 INNSBRUCK
Tel: +43(0)512 204 160 -0
Fax: +43(0)512 204 160-75
BELGIUM
ArjoHuntleigh NV/SA
Evenbroekveld 16
B-9420 ERPE-MERE
Tel: +32 (0)53 60 73 80
Fax: +32 (0)53 60 73 81
E-mail: [email protected]
Website: www.arjohuntleigh.be
CANADA
ArjoHuntleigh Canada Inc.
1575 South Gateway Road
Unit "C"
MISSISSAUGA, ON, L4W 5JI
Tel: +1 905 238 7880
Fax: +1 905 238 7881
E-mail: [email protected]
Web page: www.arjohuntleigh.com
CZECH REPUBLIC
ARJO Hospital Equipment s.r.o.
Hlinky 118
CZ- 603 00 BRNO
Tel.: +420 549 254 252
Fax: +420 541 213 550
DENMARK
ArjoHuntleigh A/S
Vassingerödvej 52
DK-3450 LYNGE
Tel: +45 49 13 84 86
Fax: +45 49 13 84 87
FAR EAST
ARJO Far East Limited
Unit 3A, 4/F., block B Hoi Luen
Industrial Centre
55 Hoi Yuen Road,
Kwun Tong, Kowloon
HONG KONG
Tel: +2508 9553
Fax: +2508 1416
FINLAND
ArjoHuntleigh OY
Vanha Porvoontie 229
FI-01380 VANTAA
Tel: +358 9 4730 4320
Fax: +358 9 4730 4999
FRANCE
ArjoHuntleigh SAS
2 Avenue Alcide de Gasperi
BP 133
59436 RONCQ CEDEX
Tel: +33(0)3 20 28 13 13
Fax: +33(0)3 20 28 13 14
E-mail: [email protected]
GERMANY
ArjoHuntleigh GmbH
Peter-Sander-Strasse 10
D-55252 MAINZ-KASTEL
Tel: +49(0)6134-186-0
Fax: +49(0)6134 186 160
E-mail: [email protected]
GREECE
C. Psimitis Co Ltd
Dimitriou Andr. 59
GR-16121 KAISARIANI ATTIKIS
Tel: 21 0724 36 68
Fax: 21 0721 55 53
INTERNATIONAL
ArjoHuntleigh International Ltd.
310-312, Dallow Road
Luton
UK-BEDFORDSHIRE LU1 1TD
Tel: +44 1582 745800
Fax: +44 1582 745866
E-mail: [email protected]
ITALY
ArjoHuntleigh S.p.A.
Via Tor Vergata 432
00133 ROMA - ITALIA
Tel: +39(0)6-87426211
Fax: +39(0)6-87426222
E-mail:
Website: www.arjohuntleigh.it
THE NETHERLANDS
ArjoHuntleigh BV
Biezenwei 21
4004 MB TIEL
Postbus 6116
NL-4000 HC TIEL
Tel: +31(0)344-64 08 00
Fax: +31(0)344-64 08 85
E-mail: [email protected]
NORWAY
ARJO Scandinavia AB
Ryenstubben 2
NO-0679 OSLO
Tel: +47 22 08 00 50
Fax: +47 22 57 06 52
POLAND
ArjoHuntleigh Healthcare Polska
Sp. z o.o.
ul. Ks Wawrzyniaka 2
PL 62-052 KOMORNIKI (Poznan)
Tel. +48 61 662 15 50
Fax +48 61 662 15 90
Email: [email protected]
PORTUGAL
ArjoHuntleigh Iberica, SL -
Sucursal Portugal
Rua Poeta Bocage Nº 2 - 2G
1600-233 Lisboa
Tel: +351 217 163 546
Fax: +351 217 162 534
E-mail: [email protected]
SPAIN
ArjoHuntleigh Ibérica S.L.
Ctra. de Rubí, 88 1ª planta - A1
08173 Sant Cugat del Vallés
ES- BARCELONA 08173
Tel: +34 93 583 11 20
Fax: +34 93 583 11 22
E-mail: [email protected]
SWEDEN
ARJO Scandinavia AB
Verkstadsvägen 5
Box 61
SE-241 21 ESLÖV
Tel: +46(0)413-645 00
Fax: +46(0)413-645 83
E-mail: [email protected]
SWITZERLAND
ArjoHuntleigh AG
Florenzstrasse 1d
Postfach
CH-BASEL4023
Tel: +41(0)61-337 97 77
Fax: +41(0)61-311 97 42
UNITED KINGDOM
ARJO Med AB Limited
St Catherine Street
UK-GLOUCESTER GL1 2SL
Tel: +44 1452 428200
Fax: +44 1452 428344
USA
ArjoHuntleigh Inc.
2349 W Lake Street Suite 250
Addison, IL 60101
Tel: +1 630-307-2756
Free +1 800 323 1245
Fax: +1 630 307 6195
E-mail: [email protected]
www.ArjoHuntleigh.com
BHM Medical Inc.
2001 Tanguay
Magog (Quebec) J1X 5Y5
CANADA
E-mail: [email protected]
GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh , GETINGE and
MAQUET . ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences. MAQUET specializes in solutions, therapies and products for surgical interventions and intensive care.
® and ™ are trademarks belonging to the ArjoHuntleigh group of companies.
© Arjo AB 2010
As our policy is one of continuous development we reserve the right to modify designs and specifications without prior notice.
Download
Advertisement