DJO TRILLIANT Gridlock Ankle Plating System Instructions
DJO TRILLIANT is a versatile ankle plating system designed for a range of surgical procedures, including trauma and corrective surgeries. It features titanium plates with varying screw options for secure fixation, ensuring minimized post-operative palpability and optimal patient comfort. The triple relief hole configuration enhances plate malleability while maintaining strength, and the skive feature aids in precise plate placement. Pre-bent and pre-contoured plates minimize the need for adjustments, simplifying the surgical process.
Реклама
Реклама
GRIDLOCK ANKLE
PLATING
SYSTEM
The Gridlock Ankle Plating System consists of titanium plates, screws, and instrumentation for applications in the bones of the ankle including trauma and corrective surgical procedures.
PLATING SYSTEMS
• Screw options include 2.7, 3.5, and 4.0mm low
profile solid core locking and non-locking screws,
4.0mm cannulated screws, and syndesmotic screws
with a tapered core designed for greater strength
• Minimized post operative plate palpability with vanishing
edge design and limited surface contact with
Gridlock technology
• Triple relief hole configuration design intended to
increase plate malleability while maintaining
plate strength
• Skive feature on proximal and distal edges to aid in
plate placement
• Low profile, pre-bent, pre-contoured specialty plates
GRIDLOCK ANKLE PLATING SYSTEM
Buttress Plate
VL Distal
Fibula Plate
SURGICAL TECHNIQUE
STEP 1: Identify, expose, and prepare the surgical site.
STEP 2: Reduce the fracture, osteotomy, or fusion site.
STEP 3: Select appropriate plate for fixation of fracture, osteotomy, or fusion. Though plates are precontoured, slight adjustments may be required and made using the plate bending instruments. Warning: Excessive or multiple plate bends may cause weakness in the plate.
STEP 4: Apply the plate to the prepared site. The plate may be temporarily fixated with either olive wires or
K-wires.
STEP 5: Select desired screw diameter and the corresponding pilot drill.
STEP 6: Select the corresponding drill guide and place it into the first plate hole nearest the fracture line or osteotomy site.
STEP 7: Drill the pilot hole with the corresponding drill diameter at the desired angle of approach through the drill guide. Note: It is recommended to irrigate during pilot drilling.
STEP 8: Insert the depth gauge into the pilot hole to determine screw length. Note: Gridlock ankle screw length is measured tip to tip. There are no threads on the distal end of the screw.
STEP 9: Select the desired screw length and type. Insert screw into the pilot hole and drive into position rotating clockwise with the driver.
VL Fibula
Plates
Right/Left
Contoured
VL Fibula
Plates
Universal
Contoured
VL Fibula
Plates
1/3
Tubular
Plates
FDA cleared 510(k) K160177 & K081510.
Trilliant products are made in the U.S.A.
VL Tibia
Plates
STEP 10: Repeat steps 5-9 until all screw holes are filled. When placing additional screws, make sure placement does not interfere with other screws. Note:
The syndesmotic holes are not intended for locking screws.
T 800.495.2919 F 877.778.3864
727 North Shepherd Drive, Suite 100 I Houston, TX 77007 I U.S.A.
djoglobal.com
Copyright © 2021 by DJO, LLC
900-00-054 Rev K
Individual results may vary. DJO, LLC is a manufacturer of orthopedic implants and does not practice medicine. Only an orthopedic, or foot and ankle surgeon can determine what treatment is appropriate. The contents of this document do not constitute medical, legal, or any other type of professional advice. This material is intended for the sole use and benefit of the DJO, LLC sales force and physicians.