Bedfont Smokerlyzer Range User Manual


Add to my manuals
25 Pages

advertisement

Bedfont Smokerlyzer Range User Manual | Manualzz

Sm

o

ker

lyzer ® Range

For use with piCO, piCO

baby

and Micro

+

User manual

Scientific contributions to health.

 

 

 

 

 

 

 

 

Contents  

Intended  Use  ......................................................................................................................................  2  

Introduction  ........................................................................................................................................  2  

Taking  a  breath  test  ............................................................................................................................  2  

Conducting  an  Ambient  Test  –  Micro

+

™  only  .....................................................................................  3  

Reviewing  history  –  Micro

+

™  only  ......................................................................................................  3  

Changing  date  and  time  –  Micro

+

™  only  ............................................................................................  3  

Traffic  Light  Pre-­‐set  Cut-­‐Off  Levels  .....................................................................................................  4  

Technical  Specification  .......................................................................................................................  4  

Safety  Information  and  Device  Symbols  .............................................................................................  4  

Environment  .......................................................................................................................................  5  

Instrument  Layout  ..............................................................................................................................  5  

Display  and  Device  Symbols  ...............................................................................................................  6  

Warnings  and  Maintenance  ...............................................................................................................  8  

Cleaning  ..............................................................................................................................................  8  

Routine  Maintenance  .........................................................................................................................  9  

Troubleshooting  .................................................................................................................................  9  

COdata

+

™  Software  –  connecting  to  the  PC  .....................................................................................  10  

Returns  Procedure  ............................................................................................................................  11  

Spares  ...............................................................................................................................................  11  

Warranty  ..........................................................................................................................................  11  

 

 

  1  

 

Intended  Use  

 

The   piCO™,   piCO baby

™   and   Micro

+

™   Smokerlyzer®   products   are   breath   carbon   monoxide   monitors   intended   for   multi-­‐patient   use   by   healthcare   professionals   in   smoking   cessation   programmes   and   research.  

 

Introduction    

 

Carbon   monoxide   (CO)   is   a   toxic,   odourless,   colourless   and   tasteless   gas.   It   is   formed   by   the   incomplete  combustion  of  organic  material  at  high  temperatures,  with  an  insufficient  oxygen  supply.    

When   inhaled,   CO   displaces   oxygen   in   the   bloodstream   to   form   carboxyhaemoglobin   (COHb).   This   starves  body  tissue  of  oxygen  restricting  the  body’s  performance  and  causing  health  implications.  CO   can   remain   in   the   bloodstream   for   up   to   24   hours,   depending   on   factors   such   as   physical   activity,   sex,  and  inhalation  intensity.  The  half-­‐life  is  about  5  hours.    

CO   (ppm),   carboxyhaemoglobin   (%COHb)   and   foetal   carboxyhaemoglobin   (%fCOHb)   correlation:  

Breath   carbon   monoxide   is   measured   in   parts   per   million   (ppm)   and   blood   carboxyhaemoglobin   is   measured  in  percentages  (%COHb  and  %fCOHb).  The  three  are  compatible  and  convertible:  the  ppm  

CO  reading  relates  to  gas  in  the  lungs  and  on  the  breath  (i.e.  the  amount  of  poisonous  CO  that  has   been  inhaled),  and  the  %COHb  and  %fCOHb  readings  relate  to  the  percentage  of  vital  oxygen  that   has  been  replaced  in  the  bloodstream.    

 

The  piCO both   CO

 

 Smokerlyzer®  displays  both  CO  ppm  and  %COHb,  the  piCO baby ™  

Smokerlyzer®  displays   ppm,%FCOHb   and   %COHb   and   the   Micro

+

™   Smokerlyzer®   displays   CO   ppm,   %COHb   and  

%FCOHb.   However   all   devices   only   directly   measure   CO   ppm,   the   %COHb   and   %fCOHb   is   a   calculation  based  on  clinical  evidence.  Clinical  research  has  demonstrated  that  a  useful  relationship   between   carbon   monoxide   and   carboxyhaemoglobin   is   obtained   after   a   short   period   of   breath-­‐ holding.    

 

The  cut-­‐off  point  between  smoker  and  non-­‐smoker  differs  based  on  local  regulation,  Bedfont  have   created  cut-­‐off  levels  as  a  guide  only,  see  ‘Traffic  Light  Pre-­‐set  Cut  off  Levels’,  these  can  be  altered   in-­‐line  with  your  local  regulations  from  the  settings  screen.  

 

 

Taking  a  breath  test  

1.

Attach  a  breath  sampling  D-­‐piece  and  new  mouthpiece.  

 

2.

Turn  on  the  monitor  by  pressing  the  power  button  once.  

 

 

 

 

 

 

3.

Press  ‘Breath  Test’  symbol  on  screen,  

 

4.

To  cancel  the  breath  test,  press   .   or .  

5.

Inhale  and  hold  breath  for  the  pre-­‐set  15  second  countdown.    

6.

A  beep  will  sound  during  the  last  three  seconds  of  the  countdown.    

7.

Blow  slowly  into  mouthpiece,  aiming  to  empty  lungs  completely.  

8.

The  ppm  and  equivalent  %COHb  and/or  %fCOHb  levels  will  rise  and  hold  on-­‐screen.  

 

  2  

 

 

9.

On  the  piCO

 and  piCO baby ™

,  when  the  test  is  finished      will  appear  at  the  bottom   of  the  screen.  

 

10.

On   the   Micro

+

™,   when   the   test   is   finished           will   appear   at   the   bottom   of   the   screen.  

 

11.

If  a  high  reading  has  been  recorded,  you  can  mute  the  sound  by  pressing   .  

 

12.

To  repeat  breath  test,  press    once  to  return  to  the  home  screen  and  repeat  steps  3-­‐8.  

 

13.

To  save  the  reading  (Micro

+

™  only)  press    and  select  the  relevant  patient  profile.  

 

14.

Remove  the  D-­‐piece  between  tests  to  purge  sensor  with  fresh  air.  

 

15.

To  switch  off,  press  and  hold  the  power  button  for  3  seconds,  unit  will  also  power  off  after  2   minutes  of  inactivity  to  save  power.    

 

 

Conducting  an  Ambient  Test  –  Micro

+

™  only    

 

 

The   environmental   air   can   be   tested   to   check   for   CO   in   the   environment   using   the   Micro

+

™  

Smokerlyzer.  This  is  particularly  useful  in  the  event  of  a  patient  reporting  a  non-­‐smoker  status  but   providing  a  breath  reading  of  that  of  a  smoker  suggesting  they  have  inhaled  environmental  CO.    

This   is   carried   out   by   pressing     on   the   screen;   the   Micro

+

™   will   then   carry   out   real-­‐time  

  monitoring  of  the  environmental  air.  To  stop  the  test,  press   .  

Bedfont   do   not   recommend   that   the   environmental   test   alone   is   used   as   a   measure   of   passive   smoking  as  the  very  high  dispersiveness  of  CO  at  the  levels  involved  with  smoking  can  lead  to  false  

  negative  readings.  

 

Reviewing  history  –  Micro

+

™  only    

The  Micro

+

™  will  record  every  reading  taken  in  its  history  up  to  150  readings,  to  access  this  press    

  and  then    the  history  will  then  be  shown.  

 

Changing  date  and  time  –  Micro

+

™  only    

Press    and  then    you  can  then  select  either  d-­‐m-­‐y  or  m-­‐d-­‐y  for  the  date  format  and  12h  or  

24h  for  the  time  format.  Dates  and  times  are  then  adjusted  by  selecting  the  number  you  wish  to  

  change  and  pressing   or    to  increase  and  decrease.  Press    to  save  your  settings.  

 

 

  3  

 

Traffic  Light  Pre-­‐set  Cut-­‐Off  Levels  

Traffic  light  colour  

Green  

Amber  

1  Red  

2  Red  

Description   Reading  (ppm)   piCO ™  &  Micro

+

™   piCO baby

™  

0-­‐6   0-­‐3   Non-­‐Smoker  

Borderline   7-­‐9  

Smoker  –  Low  addicted   10-­‐15  

Smoker  –  Moderately  addicted   16-­‐25  

4-­‐6  

7-­‐10  

11-­‐15  

3  Red  

3  Red  Flashing  

Smoker  –  Heavily  addicted   26-­‐35  

Smoker  –  Very  heavy  addicted   36+  

16-­‐25  

26+  

 

To  adjust  the  pre-­‐set  cut  off  levels,  press    from  the  settings  menu  and  drag  the  cut  off  levels  up   and  down  to  the  desired  ppm  level.  These  settings  will  remain  until  adjusted  again.  

 

Technical  Specification  

 

Concentration  Range  (CO)   piCO ™  

0-­‐150ppm   piCO baby

™  

0-­‐150ppm  

Micro+ ™  

0-­‐500ppm  

Display  

Detection  Principle  

Repeatability  

H

2

 Cross  Interference  

Power  

Full  colour  touchscreen  

Electrochemical  sensor  

<3%  on  consecutive  readings  

<6%  

3  x  AA  (LR6  or  Equivalent)  

1  x  CR2032  Lithium  coin  cell  

Approx.  3  x  AA  Battery  Life    

Approx.  1  x  CR2032  Battery  Life  

T

90  

Response  Time    

Operating  Temperature  Range  

Operating/Transport/Storage  Pressure  

Operating  Humidity  

Sensor  Operating  Life  

1,000  uses  on  new  alkaline  batteries  (approx.  

3  min  per  use)  

175,000  hours    

≤30  seconds  

0-­‐40°C  (storage  0-­‐50°C)  

800-­‐1200mbar  

10-­‐90%  non-­‐condensing  (storage  0-­‐95%)  

5  years  –  5  year  warranty  

Sensor  Sensitivity  

Sensor  Drift  

Dimensions  

Weight  

Monitor  Construction  

1ppm  

<5%  per  annum  

Approx.  34  x  75  x  140  mm  

Approx.  200g  (including  batteries)  

Case:  Polycarbonate/ABS  blend  

SteriTouch  anti-­‐Microbial  additive  

D-­‐Piece:  Polypropylene  

 

Safety  Information  and  Device  Symbols  

Degree  of  protection  against  electric  shock  

Type  of  protection  against  electric  shock  

Degree  of  protection  against  ingress  of  liquid  

Degree  of  safety  application  in  the  presence  of  a   flammable  anaesthetic  mixture  with  air,  oxygen   or  nitrous  oxide  

Refer  to  Manual    

Type  BF  applied  part  

Internally  powered  equipment  

IPXO  -­‐  not  protected  against  water  ingress  

Equipment  not  suitable  for  use  in  the   presence  of  flammable  mixtures.  

 

Direct  Current  

 

CE  Mark    

 

 

  4  

 

Type  BF  applied  part  

Dispose  of  according  to  WEEE  

 

 

Serial  Number  

 

Manufacture  by  and  Date  

 

Bedfont  Logo  

 

 

Environment  

 

The   piCO™,   piCO baby

™   and   Micro

+

™   Smokerlyzer®   products

  comply   with   the   directive   EN60601-­‐1-­‐2   electromagnetic   compatibility   but   can   be   affected   by   cellular   phones   and   by   electromagnetic   interference   exceeding   the   levels   specified   in   EN50082:1.   This   equipment   should   be   moved   if  

  necessary  to  avoid  interference.  

Instrument  Layout  

 

1.

Display  

2.

Power  button  

3.

Breath  sampling  D-­‐Piece™  

4.

D-­‐Piece™  aperture  

5.

Exhaust  port  for  breath  sample  

6.

USB  connector  (for  use  with  COdata  Software)  

7.

Single-­‐use  SteriBreath™  mouthpiece  

8.

Battery  compartment    

9.

Battery  compartment  clip  

10.

Reset  button  

11.

Programming  switch    

 

 

 

5  

 

Display  and  Device  Symbols  

Description  

Battery  condition:  

Full  

Battery  condition:  

Low  

Battery  condition:  

Empty  

Breath  Test     piCO ™  

 

 

 

  piCO baby

™  

 

 

 

Environmental  Test  

Settings  

Patient  Profiles  

View  patient  results  

Inhale  

N/A  

 

N/A  

 

N/A  

N/A  

 

N/A  

 

N/A  

 

Hold  Breath  

 

   

Countdown  Timer  

Exhale  

    ppm  Reading    

%COHb  Reading  

 

Show  %fCOHb  

 

N/A  

 

 

 

 

 

 

 

 

 

Micro+ ™  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6  

 

 

Description  

%fCOHb  Reading   piCO ™  

N/A  

Ambient  CO  reading  

N/A  

Home    

Change  D-­‐Piece  

 

 

Change  Traffic  Light  

Cut-­‐off  Levels  

Save  

Change  Breath  Hold  

Time  

History  

Mute  

Next  Step  

Countdown  to  

Sensor  Change  

Change  Sensor  

N/A  

 

N/A  

N/A  

 

 

 

  piCO baby

™  

N/A  

 

N/A  

 

N/A  

N/A  

 

 

 

 

 

   

Calibrate  Device  

 

   

 

 

Micro+ ™  

 

 

 

 

 

 

 

 

 

 

 

 

 

7  

 

 

Description  

Attach  Flow  Meter   to  Gas    

Canister   piCO ™   piCO baby

™   Micro+ ™  

     

Attach  Calibration  

Adaptor  to  D-­‐Piece  

     

Attach  D-­‐Piece  to   monitor  and  Turn  on  

Gas  Flow  

     

Unit  Calibrating  

     

Calibration  

Successful  

Calibration  Failed  

     

     

Retry  Calibration  

 

 

 

   

Firmware  Version  

   

 

N/A  =  Not  Available  on  this  device  

 

Warnings  and  Maintenance  

1.

Mouthpieces  should  be  replaced  after  every  use.    

2.

Hands  should  be  washed  regularly  in  accordance  with  infection  control  practice.    

3.

Please  do  not  attempt  to  modify  the  equipment  in  any  way  or  use  accessories  not  specified   by   the   manufacturer.   Any   attempt   to   do   so   will   invalidate   the   warranty   and   may   compromise  the  safety  of  the  device.    

4.

Bedfont  will  make  available  upon  request  service  training  to  appropriately  qualified  persons.  

5.

Holding  the  reset  button  down  for  30  seconds  will  perform  complete  device  reset,   this  will  

 

 

Cleaning   clear  any  saved  data  and  revert  all  settings  to  the  factory  defaults .  After  performing  a  reset   the  device  will  need  to  have  the  date/time  set  and  be  calibrated  before  it  can  be  used.    

1.

The  Smokerlyzer®  products  are  moulded  with  SteriTouch®  technology  for  optimum  infection   control  and  bacterial  efficacy.  However  Bedfont  recommend  wiping  the  instrument  and  D-­‐ piece   external   surfaces   with   a   product   specifically   developed   for   this   purpose.   Bedfont   provides  instrument  cleaning  wipes.  The  D-­‐Piece  cannot  be  sterilised.  

2.

NEVER   use   alcohol   or   cleaning   agents   containing   alcohol   or   other   organic   solvents   as   long   term  exposure  to  these  vapours  will  damage  the  CO  sensor  inside.    

3.

Under   no   circumstances   should   the   instrument   be   immersed   in   liquid   or   splashed   with   liquid.  

 

 

  8  

 

 

 

Routine  Maintenance  

1.

Replace  batteries  when  indicated  by  the  empty  symbol   .    

2.

Bedfont   recommend   removal   of   the   batteries   when   the   device   is   not   used   for   prolonged   periods  of  time  to  prevent  leakage.  

3.

Replace   breath   sampling   D-­‐piece   every   30   days   or   if   visibly   soiled   or   contaminated.   The   piCO™,   piCO baby™

  and

 

Micro

+

™   Smokerlyzer®   products   will   give   a   reminder   during   start-­‐up   when  the  D-­‐piece  should  be  replaced,  see  ‘Change  D-­‐Piece’  symbol.  

4.

The  sensor  should  be  replaced  every  5  years,  60  days  prior  to  the  sensor  change  ‘Countdown   to   sensor   change’   symbol   will   be   shown   with   the   date   on   which   the   sensor   should   be   changed.  This  can  be  ignored  by  pressing    until  the  date  at  which  the  sensor  should  be   changed   arrives,   at   this   point   the   ‘Change   Sensor’   symbol   will   be   shown.   Change   sensor   if   trained  to  by  an  approved  Bedfont  engineer  or  send  to  Bedfont  or  the  local  representative.  

5.

Additional  technical  information  can  be  made  available  on  request;  please  contact  Bedfont   or  its  distributor.  

 

Troubleshooting  

 

The  unit  fails  to  turn  on:  

If  the  unit  fails  to  turn  on,  replace  the  batteries.  Ensure  that  the  batteries  are  inserted  the  correct   way  around,  matching  the  symbols  moulded  into  the  plastic.  

 

High  readings  recorded  for  reported  non-­‐smoker:  

If   a   breath   test   on   a   non-­‐smoker   gives   an   indication   of   CO   being   present,   it   may   indicate   the  

  following:  

1.

A  high  ambient  level  of  CO  (possibly  due  to  a  faulty  combustion  appliance  or  exhaust  fumes).  

2.

The  effects  of  passive  smoking.  

3.

The  device  is  contaminated  by  alcohol  or  another  organic  substance.  

4.

The  patient  may  have  a  very  high  level  of  hydrogen  on  their  breath,  which  can  be  generated   in  the  human  digestive  system.  

5.

The  sensor  may  have  drifted  out  of  specification  (see  below).  

 

The  sensor  has  drifted  out  of  specification:  

Smokerlyzer®   products   are   calibrated   before   leaving   Bedfont,   however   we   recommend   if   you   suspect   the   instrument   is   reading   incorrectly,   try   the   test   again   with   another   device   if   available   to   get   a   comparison.   Alternatively   you   check   the   function   using   Bedfont   check   gas   or   send   back   to  

Bedfont.  The  check  gas  required  is  Bedfont  20ppm  carbon  monoxide  in  air,  procedure  below.  

   

1.

Ensure  the  fine  control  valve  is  in  the  off  position.    

 

2.

Screw  the  fine  control  valve  and  flow  indicator  assembly  to  the  gas  can.  This  is  best  done  by   screwing  the  gas  can  into  the  valve.    

3.

Press  ‘Breath  Test’  symbol  on  screen,   or ,  to  start  a  breath  test.    

4.

When   prompted   by   the   ‘Exhale’   symbol,   or regulator  and  allow  the  gas  to  flow  at  1  litre  per  minute.  

 

,   open   the   fine   control   valve   or  

 

  9  

 

 

5.

Allow  the  gas  to  flow  through  the  instrument  for  the  duration  of  the  test,  again  monitoring   the  rate  of  flow.    

 

6.

If   the   final   displayed   value   is   less   than   15ppm   or   higher   than   25ppm,   stop   the   test   and   perform  a  calibration  or  send  back  to  Bedfont  for  a  free  of  charge  calibration,  following  the   instructions  below.  

 

 

Calibration  procedure:  

1.

Turn  on  the  monitor  by  pressing  the  power  button  once.  

2.

Press  

.

 

3.

Start   the   calibration   process,   pressing   the   proceed.   symbol   5   times   and   then   once   to  

 

4.

Once  this  has  been  successfully  carried  out  the  device  will  perform  a  zero   do  not  apply  gas   at  this  stage  

 

5.

The   first   step   of   the   calibration   process   will   be   shown   onscreen.   Follow   the   onscreen   instructions  to  finish  the  calibration  process.  

 

6.

Gas  flow  should  be  set  to  1.0  litres  per  minute.  

 

7.

A  successful  calibration  will  be  indicated  by  the ,  press   to  return  to  the  home  screen.  

 

8.

A  failed  calibration  will  be  indicated  by ,  press   to  attempt  calibration  again.  If  the   problem  persists  see  ‘Troubleshooting’  or  call  your  local  supplier  of  Bedfont  products.    

 

9.

Return  to  the  home  screen  by  pressing  

 

COdata

+

™  Software  –  connecting  to  the  PC  

.

 

 

Place  one  end  of  the  connection  lead  into  the  USB  socket  on  the  top  of  the  Smokerlyzer®  product,  

  connecting  the  other  end  to  the  USB  port  on  the  PC.  Before  starting  the  software,  ensure  that  the  

Smokerlyzer®   product   is   connected   to   the   PC   and   switched   on.   Double   click   the   COdata

+

™   icon   on   the  PC  to  start  the  programme.  Refer  to  the  supplied  documentation  for  how  to  operate  COdata

+

™.    

 

 

  10  

 

 

 

 

Returns  Procedure  

If   your   equipment   requires   servicing,   please   contact   your   local   Bedfont   Customer   Service  

Department,  distributor  or  supplier  before  returning  any  goods.    

1.

When  you  have  supplied  the  Customer  Repairs  Department  with  the  monitor  serial  number   and   description   of   the   fault,   you   will   be   issued   with   a   returns/   ticket   number.   Please   state   the   returns/ticket   number   on   the   outside   of   the   box   when   returning   the   monitor,   and   ensure  that  your  full  details,  telephone,  fax  numbers  and  return  address  are  clearly  stated.    

2.

The   product   must   also   be   decontaminated   before   it   is   returned   according   to   your   local   regulations.  Bedfont  can  provide  you  with  a  decontamination  certificate  to  complete,  which   also  needs  to  be  attached  to  the  outside  of  the  box.  Failure  to  do  so  will  result  in  the  product   being  subject  to  Bedfont  decontamination  procedure  and  will  delay  your  service/repair.  

3.

Bedfont  advise  that  you  use  a  courier  service  when  returning  monitors.  This  enables  you  to   insure  goods  for  loss  or  damage  in  transit.  When  your  goods  are  received,  you  will  be  sent   an  email  stating  so.    

4.

If  the  device  has  been  returned  for  repair  it  will  then  be  examined  and  you  will  be  sent  an  

‘Engineer’s  Report’  and  a  quotation  for  the  repair,  which  will  include  an  authorisation  form.  

Complete  the  authorisation  form,  and  ensure  that  you  include  the  ‘Official  Purchase  Order  

Number’.   Please   contact   the   Customer   Repairs   Department   if   you   are   unable   to   supply   an  

‘Official  Purchase  Order  Number’.  

5.

If   your   monitor   is   still   in   warranty   and   the   fault   is   covered   by   warranty,   see   ‘Warranty’   section  of  this  manual,  Bedfont  will  repair  it  and  return  it  to  you  with  an  ‘Engineer’s  Report’,   free  of  charge.    

6.

If  you  choose  not  to  proceed  with  the  repair,  a  handling  fee  will  be  charged.  Ensure  that  you   return  the  completed  authorisation  form  with  an  ‘Official  Purchase  Order  Number’.    

7.

The   equipment   will   be   returned   to   you   as   soon   as   Bedfont   has   received   all   relevant  

  paperwork.  A  carriage  fee  will  be  charged  if  the  monitor  is  no  longer  in  warranty.  

Spares  

D-­‐PIECE-­‐2  

STERIBREATH  

WIPE-­‐V  

Box  of  12  D-­‐Pieces  (non-­‐sterile)  

Box  of  250  SteriBreath  Mouthpieces  (non-­‐sterile)  

Pack  of  50  Cleaning  Wipes    

 

FXS397  

 

Warranty  

USB  Cable  

Bedfont   Scientific   Limited   warrants   the   piCO™,   piCO baby

  and   Micro

+

™   Smokerlyzer®   products  

(excluding   batteries)   to   be   free   of   defects   in   materials   and   workmanship   for   a   period   of   five   years   from   the   date   of   shipment.   Bedfont’s   sole   obligation   under   this   warranty   is   limited   to   repairing   or   replacing,  at  its  choice,  any  item  covered  under  this  warranty  when  such  an  item  is  returned,  intact   and   prepaid,   to   Bedfont   Scientific   Limited   or   the   local   representative.   These   warranties   are   automatically   invalidated   if   the   products   are   repaired;   altered,   void   labels   removed   or   otherwise   tampered  with  by  unauthorised  personnel,  or  have  been  subject  to  misuse,  neglect  or  accident.  

Never   dispose   of   any   electronic   instrument   or   batteries   in   domestic   waste.   At   the   end   of   the  

 

 

 

 

  product’s  life,  contact  Bedfont  or  its  distributor  for  disposal  instructions.  

 

  11  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

Contenu    

Domaine  d’utilisation………………………………………………………………………………………………………………..  13  

Introduction………………………………………………………………………………………………………………………………13  

Prendre  un  test  respiratoire………………………………………………………………………………………………………13  

Effectuer  un  test  ambiant  –  Seulement  Micro

+

™………………………………………………………………….……14  

Afficher  l’historique–  Seulement  Micro

+

™……………………………………………………………………….………..  14  

Modifier  le  date  et  l’heure    –  Seulement  Micro

+

™……………………………………………………………………..14  

Classification  du  fumeur  et  code  couleur  prédéfini……………………………………………………………………15  

Spécifications  techniques………………………………………………………………………………………………..………..  15  

Informations  de  sécurité  des  icônes………………………………………………………………………………………....  15  

Environnement………………………………………………………………………………………………………………………….16  

Schéma  d’appareil…………………………………………………………………………………………………………………….17  

Affichage  et  icônes  d’appareil……………………………………………………………………………………………………17  

Avertissements  et  maintenance………………………………………………………………………………………………..  20  

Nettoyage………………………………………………………………………………………………………………………………...  20  

Entretien  courant…………………………………………………………………………………………………………………….21  

Dépannage………………………………………………………………………………………………………………………………21  

Logiciel  COdata

+™  

-­‐  branchement  au  PC……………………………………………………………………………………22  

Procédure  de  retour………………………………………………………………………………………………………………..22  

Pièces  consommables………………………………………………………………………………………………………………23  

Garantie…………………………………………………………………………………………………………………………………..23  

 

  12  

 

Utilisation  prévue  

 

Le  piCO™,  piCO baby

™    et  le  Micro

+

™  Smokerlyzer®  sont  des  analyseurs  de  monoxyde  de  carbone   expiré  destinés  aux  professionnels  de  la  santé  dans  le  cadre  des  programmes  d’aide  au  sevrage   tabagique,  ainsi  que  pour  les  études  cliniques  et  la  recherche.    

 

Introduction    

 

Le  monoxyde  de  carbone  est  un  gaz  toxique,  inodore,  incolore  et  sans  saveur.  Il  résulte  de  la   combustion  incomplète,  à  très  hautes  températures,  de  matières  organiques  combinées  à  une   alimentation  en  oxygène  insuffisante.    

 

Quand  il  est  inhalé,  le  CO  prend  la  place  de  l’Oxygène  sur  l’Hémoglobine  formant  le  complexe  

Carboxyhémoglobine  (HbCO).  Ceci  a  pour  effet  de  limiter  l’oxygénation  des  tissus  qui  est  vital  pour  le   bon  fonctionnement  de  l’organisme.  Le  CO  peut  rester  dans  le  sang  pendant  plus  de  24h  en  fonction   de  l’activité  physique,  du  sexe  de  la  personne  et  de  l’intensité  de  l’inhalation.  La  demi-­‐vie  du  CO  est   de  l’ordre  de  5h.  

Corrélation  entre  le  CO  (ppm),  la  carboxyhémoglobine  (%COHb)  et  la  carboxyhémoglobine  fœtale  

(%fCOHb):    

Le  monoxyde  de  carbone  expiré  se  mesure  en  partie  par  million  (ppm  CO)  et  la  carboxyhémoglobine   dans  le  sang  en  pourcentage  (%COHb  et  %fCOHb).  Les  trois  sont  compatibles  et  convertibles,  le  CO   se  rapportant  à  la  quantité  de  gaz  dans  les  poumons/les  voies  respiratoires  (correspondant  à  la   quantité  de  CO  nocif  qui  a  été  inhalée)  et  la  COHb  et  fCOHb  à  la  quantité  de  gaz  présente  dans  le   sang.  

   

L’analyseur  piCO™  Smokerlyzer®  affiche  la  concentration  en  CO  ainsi  que  %COHb.  Le  piCO baby

Smokerlyzer®  et  le  Micro

™  

+

™  Smokerlyzer®  affichent  en  plus  %fCOHb.  Cependant,  tous  ces  appareils   ne  mesurent  que  directement  ppm  CO  ;  la  %COHb  et  %fCOHb    est  un  calcul  basé  sur  des  données   cliniques.  Des  études  cliniques  ont  montré  qu’une  relation  existe  entre  le  monoxyde  de  carbone  et  la  

  carboxyhémoglobine  lorsque  l’utilisateur  retenait  sa  respiration  pendant  quelques  secondes.  

Le  point  de  coupure  entre  fumeur  et  non-­‐fumeur  diffère  en  fonction  de  la  réglementation  locale.    

Les  niveaux  de  coupures  créés  par  Bedfont  sont  à  prendre  en  tant  que  guide  uniquement;  voir  

«  Système  de  feu  de  circulation  prédéfini    »,  ceux-­‐ci  peuvent  être  modifiés  en  ligne  avec  votre   réglementation  locale  dans  le  menu  des  paramètres.  

 

 

Prendre  un  test  respiratoire  

 

16.

Fixez  le  piégeur  Dpiece  de  la  fin  de  l’air  expiré  et  un  nouvel  embout  buccal.  

 

17.

Allumez  l’appareil  en  appuyant  une  fois  sur  le  bouton  d’alimentation.  

 

 

 

18.

Appuyez  sur  l’icône  «  Test  Respiratoire  »  de  l’écran,  

 

19.

Pour  annuler  le  test  respiratoire,  appuyez  sur  l’icône .   ou .  

20.

 Inspirez  et  retenez  votre  respiration  pendant  le  compte  à  rebours  de  15  secondes.  

21.

Un  son  retentira  pendant  les  trois  dernières  secondes  du  compte  à  rebours.  

 

  13  

 

 

22.

Soufflez  lentement  dans  l’embout  buccal  afin  de  vider  complètement  vos  poumons.  

 

23.

La  concentration  de  CO  en  ppm  et  les  pourcentages  équivalents  à  la  COHb  et/ou  fCOHb   augmenteront  et  se  stabiliseront.  

 

24.

Avec  le  piCO

 et  piCO baby ™

,  lorsque  le  test  est  fini,  les  icônes      s’afficheront  en  bas  

  de  l’écran.  

25.

Avec  le  Micro

+

™,  lorsque  le  test  est  fini,  les  icônes          s’afficheront  en  bas  de  

  l’écran.  

26.

Si  un  résultat  élevé  a  été  obtenu,  vous  pouvez  couper  le  son  en  appuyant  sur .  

 

27.

Pour  répéter  le  test  respiratoire,  appuyez  sur  l’icône   une  fois  pour  revenir  à  l’écran   d’accueil,  et  répétez  les  étapes  3  à  8.  

 

28.

Pour  enregistrer  la  lecture  (seulement  Micro

+

™)  appuyez  sur  l’icône    et  choisissez  le   profil  du  patient  concerné.  

 

29.

Enlevez  le  piégeur  Dpiece  entre  chaque  test  pour  purger  le  capteur  avec  de  l’air  frais.  

 

30.

Pour  éteindre  l’appareil,  maintenez  le  bouton  enfoncé  pendant  3  secondes.  L’analyseur   s’éteindra  automatiquement  après  2  minutes  d’inactivité.    

 

 

Effectuer  un  test  ambiant  –  Seulement  Micro

+

™    

 

Avec  le  Micro+™  Smokerlyzer®,  l’air  ambiant  peut  être  analysé  pour  vérifier  la  concentration  de  CO   dans   l’environnement.   Ceci   est   particulièrement   utile   dans   le   cas   où   un   patient   qui   est   signalé   comme  étant  non-­‐fumeur  et  obtient  un  résultat  correspondant  à  celui  d'un  fumeur,  suggérant  ainsi  

  une  exposition  au  CO  ambiant.  

Effectuez   le   test   ambiant   en   appuyant   sur   l’icône   ;   le   Micro

+

™   effectuera   un   test   ambiant   en  

  temps  réel.  Pour  annuler  le  test,  appuyer  sur .  

Bedfont  ne  recommande  pas  que  seul  le  test  ambiant  soit  utilisé  comme  une  mesure  du  tabagisme   passif,  car  il  est  possible  que  des  niveaux  élevés  de  CO  provenant  d’autres  sources  puissent  fausser   les  données.  

 

Afficher  l’historique–  Seulement  Micro

+

™  

 

Le  Micro+™  enregistrera  chaque  lecture  d’un  test  respiratoire  et  peut  garder  jusqu’à  150  résultats  en  

  mémoire.  Pour  afficher  le  résultat  à  l’écran,  appuyez  sur ,  puis  sur .  

 

Modifier  le  date  et  l’heure    –  Seulement  Micro

+

™  

 

  14  

 

 

Appuyez  sur    et  puis .  Ici  vous  pouvez  choisir  d’afficher  la  date  au  format  j-­‐m-­‐a  ou  m-­‐j-­‐a,  et   l’heure  peut  être  affichée  au  format  12H  ou  24H.  La  date  et  l’heure  peuvent  être  modifiées  en   sélectionnant  les  chiffres  ou  en  appuyant  sur ou pour  augmenter  et  diminuer  la  valeur.  

 

 

Appuyez  sur    pour  enregistrer  les  changements.  

Classification  du  fumeur  et  code  couleur  prédéfini  

Couleur  des  feux   Description  

Vert  

Ambre  

1  Rouge  

2  Rouge  

Non-­‐fumeur  

Fumeur  

Fumeur–  fréquent  

Fumeur  –  dépendant  

Lecture  (ppm)   piCO ™  &  Micro

+

™   piCO baby

™  

0-­‐6   0-­‐3  

7-­‐9  

10-­‐15  

16-­‐25  

3  Rouge  

3  Rouge  clignotant  

Fumeur–  fortement  dépendant   26-­‐35  

Fumeur  –  dangereusement   dépendant  

36+  

4-­‐6  

7-­‐10  

11-­‐15  

16-­‐25  

26+  

Pour   modifier   les   seuils   prédéfinis,   appuyer   sur     dans   le   menu   des   paramètres   et   faites   glisser   votre   doigt   vers   le   haut   ou   le   bas   pour   modifier   le   seuil   en   ppm   jusqu’à   la   valeur   désirée.   Ces  

  modifications  resteront  jusqu’à  ce  qu’elles  soient  de  nouveau  modifiées.  

 

Spécifications  techniques  

  piCO ™   piCO baby

™   Micro+ ™  

0-­‐500ppm   Plage  de  concentration  :  (CO)  

Affichage  

Principe  de  détection  

Répétabilité  

0-­‐150ppm   0-­‐150ppm  

Ecran  tactile  couleur  

Capteur  électrochimique  

<3%  sur  lectures  consécutives  

Interférence  avec  H2  

Alimentation  électrique  

Autonomie  avec  3  piles  AA    

Autonomie  avec  1  pile  CR2032  

Temps  de  réponse  à  T

90

 

Plage  de  températures  de  fonctionnement  

Pression  de  service/transport/stockage  

Humidité  en  fonctionnement  

Durée  de  vie  du  capteur  en  service  

Sensibilité  du  capteur  

<6%  

3  x  AA  (LR6  ou  équivalent)  

1  x  CR2032  pile  au  lithium  

1,000  utilisations  avec    des  piles  alcalines   neuves  (environ  3  minutes  par  utilisation)  

175,000  heures    

≤30  secondes  

0-­‐40°C  (stockage  0-­‐50°C)  

800-­‐1200mbar  

10-­‐90%  sans  condensation  (stockage  0-­‐95%)  

Dérive  du  capteur  

Dimensions  

Poids  

Structure  de  l’analyseur  

5  ans  –  garantie  de  3ans  

1ppm  

<5%  par  an  

Environ  34  x  75  x  140  mm  

Environ  200g  (avec  piles)  

Corps:  Polycarbonate/ABS  fusion  

SteriTouch  additif  antimicrobien  

Piégeur  D:  Polypropylène  

 

Informations  de  sécurité  et  icônes  d’appareil  

Degré  de  protection  contre  l’électrocution   Pièce  appliquée  de  type  BF  

 

  15  

 

Type  de  protection  contre  l’électrocution  

Degré  de  protection  contre  la  pénétration  de   liquide  

Degré  de  sécurité  de  l’application,   en  présence  de  mélange  anesthésique   inflammable  à  l’air,  avec  de  l’oxygène   ou  du  protoxyde  d’azote:  

Référer  aux  instructions  pour  l’utilisation  

Courant  continu  

Marquage  CE  

Pièce  appliquée  de  type  BF  

Eliminer  selon  DEEE  

Equipement  alimenté  de  manière  interne.  

IPXO  -­‐  non  protégé  contre  la  pénétration  de   l’eau  

L’appareil  ne  convient  pas  à  l’utilisation  en   présence  des  mélanges  inflammables.  

 

 

 

 

 

Numéro  de  série  

 

Fabricant  et  date  de  fabrication  

 

Logo  de  Bedfont  

 

 

Environnement  

 

Le   piCO™,   piCO baby

™   et   le   Micro

+

™   Smokerlyzer®   sont   conformes   à   la   norme   EN60601-­‐1-­‐2   sur   la   compatibilité   électromagnétique   mais   peuvent   être   affectés   par   les   téléphones   portables   et   les   interférences  électromagnétiques  dépassant  les  limites  spécifiées  dans  la  norme  EN50082:1.  Le  cas  

échéant  il  faut  déplacer  ces  appareils  pour  éviter  les  interférences.    

 

 

  16  

Schéma  d’appareil  

 

 

1.

Ecran  

2.

Bouton  marche/arrêt    

3.

Le  piégeur  Dpiece  

4.

Emplacement  pour  le  piégeur  Dpiece  

5.

Port  de  sortie  de  l’échantillon  d’air  expiré    

6.

Port  de  connexion  USB  (pour  utilisation  avec  le  logiciel  

COdata)  

7.

Embout  buccal  SteriBreath™  à  usage  unique  

8.

Compartiment  des  piles    

9.

Clip  de  compartiment  des  piles    

10.

Bouton  de  réinitialisation  

11.

Interrupteur  de  programmation  

 

 

Affichage  et  icônes  d’appareil  

Description   piCO ™   piCO baby

™  

Etat  des  pile:  Pleine  

   

Etat  des  piles:  Faible  

   

Etat  des  piles:  Vide  

   

Test  respiratoire  

Test  ambiant  

Paramètres  

N/A  

 

 

 

 

N/A  

 

 

Micro+ ™  

 

 

 

 

 

 

 

17  

Profils  des  patients  

Afficher  les  résultats   mémorisés  

Inspirer  

N/A  

N/A  

 

Retenir  la   respiration  

Compte  à  rebours  

Expirer  doucement   et  complètement  

 

Lecture  en  ppm    

Lecture  en  %COHb  

 

Afficher  la  %fCOHb  

 

 

N/A  

 

 

 

 

 

N/A  

N/A  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  18  

 

 

Description  

Lecture  en  %fCOHb   piCO ™  

N/A  

Lecture  du  CO   ambiant  

Accueil    

Remplacer  le   piégeur  Dpiece  

Modifier  les  seuils   de  classification  du   fumeur  

Enregistrer  

Modifier  le  compte    

à  rebours  pour  le   temps  d’apnée  

Historique  

Volume  

Prochaine  étape  

Date  de   remplacement  du   capteur  

Remplacer  le   capteur  

N/A  

N/A  

 

N/A  

N/A  

 

 

 

 

 

 

Etalonner  l’appareil  

 

 

  piCO baby

™  

N/A  

 

N/A  

 

N/A  

N/A  

 

 

 

 

 

 

 

 

 

Micro+ ™  

 

 

 

 

 

 

 

 

 

 

 

 

 

19  

 

 

Description  

Raccorder  le   débitmètre  a  la   bouteille  de  gaz   piCO ™   piCO baby

™   Micro+ ™  

     

Raccorder   l’adaptateur   d’étalonnage  au   piégeur  Dpiece  

Raccorder  le  piégeur  

à  l’analyseur  et   ouvrir  le  gaz  

     

     

Etalonnage  en  cours  

   

 

Etalonnage  réussi  

     

Etalonnage  échoué  

 

Réessayer   l’étalonnage  

 

Version  du  logiciel  

 

N/A  =  N’est  pas  disponible  sur  ce  dispositif  

 

 

 

 

 

 

 

 

Avertissements  et  maintenance  

6.

L’embout  buccal  doit  être  remplacé  après  chaque  test  respiratoire.  

7.

Il  faut  se  laver  les  mains  pour  éviter  les  contaminations  croisées.  

8.

Ne   tenter   pas   de   modifier   l’appareil   de   quelque   façon   que   ce   soit,   ou   l’utiliser   avec   des   accessoires   non   spécifiés   par   le   fabricant.   Toute   tentative   invalidera   la   garantie   et   peut   compromettre  la  sécurité  de  l’appareil.  

9.

Sur  demande,  Bedfont  peut  former  les  personnes  qualifiées  pour  l’entretien  du  dispositif.  

10.

En  maintenant  le  bouton  de  réinitialisation  pendant  30  secondes  le  dispositif  se  réinitialisera   complétement,   ceci   effacera   toutes   les   données   enregistrées   et   rétablira   tous   les   paramètres   aux   valeurs   par   défaut.

  Après   une   réinitialisation,   il   faut   vérifier   l’heure   et   la   date  et  étalonner  le  dispositif  avant  utilisation.

 

 

 

Nettoyage  

4.

Les  produits  de  la  gamme  Smokerlyzer®  sont  moulés  avec  la  technologie  SteriTouch®  pour  la   filtration  des  bactéries  et  le  contrôle  contre  infections.  Cependant,  Bedfont  recommande   d’essuyer  le  dispositif  avec  un  produit  adapté,  tel  que  les  lingettes  nettoyantes  sans  alcool   fournis  par  Bedfont.  Le  piégeur  Dpiece  ne  doit  pas  être  stérilisé.  Il  se  change  tous  les  mois.  

5.

Lors  du  nettoyage  de  l’analyseur,  n’utiliser  jamais  d’alcool,  des  agents  contenant  de  l’alcool   ou  d’autres  solvants  organiques,  car  ces  vapeurs  peuvent  endommager  le  capteur  de  CO  à   l’intérieur  de  l’appareil.  

6.

En  aucun  cas,  l’appareil  ne  devrait  être  pas  immergé  ou  éclaboussé  par  un  liquide.    

 

  20  

 

 

 

 

Entretien  courant  

6.

Remplacer  les  piles  lorsque  l’icône  indique  que  la  pile  est  vide    

7.

Bedfont   recommande   de   retirer   les   piles   lorsque   le   dispositif   n’est   pas   utilisé   pour   éviter   tout  risque  de  fuite  et  de  corrosion.  

8.

Il  faut  remplacer  le  piégeur  tous  les  30  jours  ou  s’il  est  visiblement  souillé  ou  contaminé.  Le   piCO™,   piCO baby™

  et

 

Micro

+

™   Smokerlyzer®   vous   aviseront   lors   du   démarrage   si   vous   devez   remplacer  le  piégeur,  vous  verrez  l’icône  «  Remplacer  le  piégeur  D  ».  

9.

Le   capteur   doit   être   remplacé   tous   les   cinq   années.   60   jours   avant   le   changement   du   capteur,  le  «Compte  à  rebours  de  remplacement  du  capteur»  affichera  la  date  à  laquelle  le   capteur  doit  être  remplacé.  Ceci  peut  être  ignoré  en  appuyant  sur  l’icône    jusqu’à  ce  que   la   date   de   remplacement   arrive;   à   ce   moment   l’icône   «  Remplacer   capteur  »   s’affichera.   Si   vous  avez  reçu  la  formation  d’entretien  par  un  ingénieur  approprié  de  Bedfont,  vous  pouvez   le  remplacer  votre-­‐même.  Sinon,  vous  pouvez  nous  retourner  le  dispositif.  

10.

Sur   simple   demande,   Bedfont   mettra   à   votre   disposition   les   renseignements   techniques   supplémentaires  ;  veuillez  contacter  Bedfont  ou  votre  revendeur  local.  

 

Dépannage  

 

L’appareil  échoue  le  démarrage:  

Si   l’appareil   échoue   à   démarrer,   remplacer   les   piles.   Assurez-­‐vous   que   les   piles   sont   bien   placées   dans  le  bon  sens  dans  le  compartiment.  

 

Résultats  élevés  pour  un  non-­‐fumeur:  

 

Si   un   non-­‐fumeur   obtient   des   résultats   élevés   lors   du   test   respiratoire,   les   explications   sont   les   suivants  :  

6.

Un  niveau  élevé  de  CO  ambiant  (peut-­‐être  dû  à  un  appareil  à  combustion  défectueux  ou  à   des  gaz  d'échappement)  

7.

Les  effets  du  tabagisme  passif  

8.

L’appareil  est  contaminé  par  l’alcool  ou  une  autre  matière  organique.  

9.

Le   patient   peut   avoir   une   concentration   élevée   d’hydrogène   dans   sa   respiration  

(Intolérances  alimentaires).  

10.

Le  capteur  pourrait  avoir  dérivé  et  se  trouve  hors  spécifications  (voir  en  dessous).  

 

Le  capteur  a  dérivé  hors  spécifications:  

 

Les   produits   Smokerlyzer®   sont   calibrés   par   Bedfont   avant   livraison,   cependant   si   vous   croyez   que  

  l’appareil   est   défectueux,   nous   vous   recommandons   d’essayez   avec   un   autre   appareil   pour   faire   la   comparaison.  Alternativement,  vous  pouvez  vérifier  le  dispositif  en  utilisant  le  gaz  étalon  de  Bedfont,   ou  de  le  renvoyer  chez  Bedfont.  La  vérification  à  l’aide  de  gaz  requiert  du  monoxyde  de  carbone  à  20   ppm  dans  l'air  de  Bedfont  et  suivre  la  procédure  ci-­‐dessous.  

7.

Assurer  que  la  vanne  de  régulation  est  en  position  d'arrêt.  

8.

Visser  la  vanne  de  régulation  avec  le  débitmètre  sur  la  bouteille  de  gaz.  La  meilleure  façon   est  de  visser  la  bouteille  avec  la  vanne  dans  la  main  gauche.  

9.

Appuyez  sur  l’icône   ou  de  l’écran  pour  commencer  un  test  respiratoire.  

 

  21  

 

10.

Lorsque   vous   êtes   invité   par   l’icône   à   «  Expirer  »,   ou ,   ouvrez   la   vanne   de   régulation  et  laissez  le  gaz  s’écouler  à  1  litre  par  minute.  

11.

Faire  circuler  le  gaz  pendant  le  test  respiratoire  en  surveillant  le  débit.  

12.

Si   la   valeur   affichée   est   moins   de   15ppm   ou   plus   élevée   que   25ppm,   arrêtez   le   test   et  

étalonnez   le   dispositif   en   suivant   les   étapes   ci-­‐dessous,   ou   le   renvoyez   chez   Bedfont   pour   l’étalonnage.  

 

 

Procédure  d’étalonnage:  

10.

 Allumez  l’appareil  en  appuyant  une  fois  sur  le  bouton  d’alimentation.  

11.

Appuyez  sur  l’icône

.

 

 

12.

Commencer  l’étalonnage  en  appuyant  5  fois  sur  l’icône   et  puis  une  fois  sur

.

 

 

13.

Après  cela,  l’appareil  fera  sa  mise  à  zéro,   ne  pas  utilisez  le  gaz  pendant  la  purge.  

 

14.

La  première  étape  de  l’étalonnage  s’affichera  sur  l’écran.  Suivez  les  instructions  de  l’écran   pour  compléter  l’étalonnage.  

 

15.

Le  débit  de  gaz  doit  être  réglé  à  1,0  litre  par  minute.  

 

16.

Si  l’étalonnage  a  réussi,  l’écran  affichera  l’icône ,  appuyez  sur  l’icône    pour  revenir  à   l’écran  d’accueil.  

 

17.

Si  l’étalonnage  a  échoué,  l’écran  affichera  cette  icône ,  appuyez  sur  l’icône    pour   réessayer  l’étalonnage.  Si  le  problème  persiste,  voir  «  Dépannage  »  ou  appelez  votre  

  revendeur  local.  

18.

Appuyez  sur  l’icône    pour  revenir  à  l’écran  d’accueil.  

 

Logiciel  COdata

+™  

-­‐  branchement  au  PC  

 

Branchez  le  câble  dans  le  port  USB  de  l’ordinateur  et  le  port  de  connexion  USB  en  haut  de  l’appareil.  

Avant  de  démarrer  le  logiciel,  assurez-­‐vous  que  le  Smokerlyzer®  est  raccordé  à  l’ordinateur  et  qu’il   est   allumé.   Double-­‐cliquez   sur   l’icône   COdata

+

™   de   l’ordinateur   pour   démarrer   le   logiciel.   Referez-­‐ vous  à  la  documentation  fournie  pour  savoir  comment  utiliser  le  logiciel  COdata

+

™.  

 

Procédure  de  retour  

 

Si  le  dispositif  nécessite  un  entretien  ou  une  réparation,  veuillez  contacter  le  service  après-­‐vente  de  

 

Bedfont  ou  votre  revendeur  local  avant  de  retourner  le  dispositif.  

8.

Un   fois   le   numéro   de   série   de   l’analyseur   et   une   description   défaut   fournis,   nous   vous   donnerons  un  numéro  de  retour.  Lors  du  retour  d’un  appareil,  assurez-­‐vous  que  le  numéro   de  retour  et  tous  les  détails,  y  compris  les  numéros  de  téléphone,  l’adresse  de  retour  soient   clairement  indiqués.  

 

  22  

 

 

9.

L’appareil   doit   être   décontaminé   avant   de   son   retour   conformément   à   la   réglementation   locale.  Bedfont  peut  fournir  un  certificat  de  décontamination  que  vous  devez  compléter  et   fixer   à   l'extérieur   du   colis.   Ne   pas   le   faire   imposera   la   procédure   de   décontamination   par  

Bedfont  et  retardera  votre  service/réparation.  

10.

Bedfont  conseille  d’utiliser  votre  transporteur  et  assurer  vos  produits  contre  les  risques  de   perte  ou  de  dommage  lors  du  transit.  Nous  vous  enverrons  la  confirmation  de  réception  une   fois  la  marchandise  reçue.  

11.

Après  vérification,  un  rapport  d’ingénieur  et  un  devis  pour  la  réparation  vous  seront  envoyés   avec  un  formulaire  d’acceptation.  Complétez  le  formulaire  et  assurez-­‐vous  que  le  «  numéro   de  commande  officielle  »  soit  précisé.  Veuillez  contacter  notre  service  après-­‐vente  si  vous  ne   pouvez  fournir  un  «  numéro  de  commande  officielle  ».  

12.

Si  le  défaut  est  couvert  par  la  garantie,  voir  «  Garantie  »  dans  ce  manuel,  Bedfont  réparera   retournera  l’appareil  avec  un  rapport  d’ingénieur,  gratuitement.    

13.

Si  vous  refusez  de  faire  la  réparation,  des  frais  de  traitement  de  dossier  seront  facturés.    

14.

Vous   recevrez   votre   matériel   une   fois   que   Bedfont   a   reçu   les   documents   demandés.   Si   l’opération  est  hors  garantie,  des  frais  de  livraison  seront  facturés.  

 

 

Pièces  consommables  

D-­‐PIECE-­‐2  

STERIBREATH  

WIPE-­‐V  

FXS397  

Boîte  de  12  piégeurs  Dpiece  (non  stérilisable)  

Boîte  de  250  embouts  buccaux  SteriBreath™  (non   stérilisable)  

Paquet  de  50  lingettes  de  nettoyage  

Câble  USB  

 

Garantie  

 

Bedfont  Scientific  Limited  garantit  les  piCO™,  piCO baby

 

&  Micro

+

™  Smokerlyzer®  (à  l'exception  des   piles)  contre  tout  défauts  de  fabrication  pendant  une  période  de  5  années  à  partir  de  la  date  de   livraison.  Les  seules  obligations  de  Bedfont  concernant  cette  garantie  sont  limitées,  à  sa  discrétion,  à   la  réparation  ou  au  remplacement  de  tout  élément  couvert  par  la  garantie  lorsque  ledit  élément  est   renvoyé,  intact  et  prépayé,  à  Bedfont  Scientific  Limited  ou  votre  son  revendeur  local.  Cette  garantie   sera  automatiquement  annulée  en  cas  de  réparation,  modification  et  si  l'étiquette  de  garantie  est   retirée  ou  modifiée,  ou  tout  autre  opération  réalisée  sur  le  produit  par  du  personnel  non  autorisé  ou  

à  la  suite  d’une  mauvaise  utilisation  du  produit,  d’une  négligence  ou  d’un  accident.  

À  la  fin  du  cycle  de  vie  du  produit,  veuillez-­‐vous  abstenir  de  jeter  les  instruments  électroniques  dans  

  les  déchets  domestiques  et  contacter  Bedfont  ou  le  revendeur  local  pour  connaître  la  procédure   d’élimination  à  suivre.  

 

  23  

Contact Bedfont or one of our worldwide

Sm

o

ker

lyzer

®

distributors for a free demonstration www.bedfont.com

Tel:+44 (0)1622 851122

Email: [email protected]

A full list of our worldwide distributors can be found at http://www.bedfont.com/uk/english/distributors

Scientific contributions to health.

Station Road, Harrietsham, Maidstone,

Kent, ME17 1JA England

Tel: +44 (0)1622 851122 Fax: +44 (0)1622 854860

Email: [email protected] www.bedfont.com

© Bedfont Scientific Limited 2012

Issue 1- January 2016, Part No: LAB679_CAN

Bedfont Scientific Limited reserves the right to change or update this literature without prior notice.

Registered in: England and Wales. Registered No: 1289798

ISO 9001:2008

Cert No. FM 31664

ISO 13485:2003

Cert No. MD 502905