Below you will find brief information for Eagle 2, GX 2012, GX 2009, Sirius, Altair 5X, Altair 4X, Altair Pro, 450, G460. This manual is intended to provide information on the use of hazardous materials field screening equipment. This document is not intended as a training tool, but as a quick reference for those who have received full and appropriate training. The instruments described in this manual represent the current models that are marketed by a variety of equipment manufacturers. This manual does not include those outdated and relatively unused equipment.
Advertisement
Advertisement
HAZARDOUS MATERIALS FIELD SCREENING
EQUIPMENT MANUAL
Instrumentation for Emergency, Investigatory, and Remedial Response
1.30.15
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 2 of 168
California CUPA Forum
TABLE OF CONTENTS
Preface ......................................................................................................................... 7
Introduction .................................................................................................................. 8
Instrument Applications ................................................................................................ 9
EPA Action Guides ....................................................................................................... 10
Instrument Mailing List ................................................................................................. 11
MULTIPLE GAS METERS
Glossary ....................................................................................................................... 13
RKI Multiple Gas Meters
RKI Eagle 2 .................................................................................................................. 17
RKI GX-2012 ................................................................................................................ 20
RKI GX-2009 ................................................................................................................ 23
MSA Multiple Gas Meters
MSA Sirius Multi-gas Detector ..................................................................................... 27
MSA Altair 5x Multi-gas Detector ................................................................................. 30
MSA Altair 4x Multi-gas Detector ................................................................................. 32
MSA Altair Pro Single Gas Detector ............................................................................. 34
GfG Instrumentation Multi Gas Detector
GfG Instrumentation 450 Multi-gas Detector ................................................................ 37
G460 Multi-gas Detector .............................................................................................. 39
RAE Systems Multi Gas Meters
QRAE II Gas Detector .................................................................................................. 41
QRAE 3 Wireless 4-Gas Monitor .................................................................................. 43
MultiRae Multi-gas Detection Monitors ........................................................................ 45 ppbRAE 3000/MiniRae 3000 ........................................................................................ 47
ToxiRAE Single Gas Monitors ...................................................................................... 49
AreaRAE Steel Multi-Gas Monitoring System .............................................................. 51
Industrial Scientific Multi Gas Detectors
IBRID MX6 Multi-gas Detector ..................................................................................... 53
Ventis MX4 Multi-gas Detector ..................................................................................... 56
Sperian Biosystems Multi Gas Detectors
Sperian PHD6 Multi-gas Detector ................................................................................ 59
Sperian MultiPro Multi-gas Detector ............................................................................. 61
HazMat Field Screening Equipment Manual
1/30/2015
Page 3 of 168
California CUPA Forum
Honeywell Multi Gas Monitors
Honeywell Lumidor MicroMax Pro ................................................................................ 63
Honeywell Lumidor Impact Pro .................................................................................... 65
BW Gas Alert5 by Honeywell ....................................................................................... 67
Chemical Warfare Agents (CWAs)
ChemPro100i ............................................................................................................... 69
GMI Gas Detectors
GMI Visa 5 Multi-gas Detector ..................................................................................... 73
GMI PS200 Multi-gas Detector ..................................................................................... 75
GMI GT Series ............................................................................................................. 77
GMI PSI Single Gas Detector ....................................................................................... 79
Drager Gas Monitors
Drager x-am 7000 ........................................................................................................ 81
Drager x-am 5600 ........................................................................................................ 83
Drager x-am 2500 ........................................................................................................ 85
Drager Pac 3500/5500 ................................................................................................. 87
Scott Bacharach Multi Gas Detectors
Scott Protégé ............................................................................................................... 89
Scott Scout ................................................................................................................... 91
FIELD GAS ANALYZERS
Hazardous Materials Field Gas Analyzer/Glossary ...................................................... 93
Miran SapphlRe Gas Analyzer ..................................................................................... 95
GasMet Field Gas Analyzer
Gasmet DX 4040 Portable FTIR Gas Analyzer……………………………………………97
Smiths Detection HGVI Gas and Vapor Identifier ......................................................... 99
Smiths Detection APD 2000 ......................................................................................... 100
HAZARDOUS MATERIALS SOLIDS AND LIQUID ANALYZERS
Thermo Scientific
First Defender RMX ...................................................................................................... 103
Tru Defender ................................................................................................................ 103
Smiths Detection
Hazmat ID Ranger........................................................................................................ 105
HazMat Field Screening Equipment Manual
1/30/2015
Page 4 of 168
California CUPA Forum
MERCURY VAPOR ANALYZERS
Jerome Mercury Gas Analyzers (Arizona Instruments)
Jerome 411/431 Mercury Vapor Analyzers .................................................................. 107
Jerome J405 Mercury Vapor Analyzer ......................................................................... 110
Mercury Instruments (MI) Mercury Vapor Analyzer
MI Mercury Tracker 3000 IP ......................................................................................... 113
Lumex Mercury Vapor Analyzer
Lumex RA915 Mercury Vapor Analyzer ....................................................................... 115
Genesis Laboratory Systems
Hg253 Portable Mercury Vapor Analyzer ..................................................................... 117
COLORIMETRIC GAS INDICATORS
Colorimetric Gas Indicators .......................................................................................... 119
Draeger Colorimetric Gas Indicators ............................................................................ 120
Draeger Accuro (bellows pump) ................................................................................... 120
Ion Science – Uniphos ASP-40 .................................................................................... 121
MSA RAE LP – 1200 .................................................................................................... 123
MSA Kwik Draw Deluxe (bellows pump) ...................................................................... 123
Sensidine/Kitagawa Detector Tubes ............................................................................ 124
Draeger Chip Measurement System (CMS) ................................................................. 125
RADIATION DETECTION METERS
Radiation Detection Meters/Glossary ........................................................................... 127
Ludlum Radiation Detectors
Ludlum Radiation Meter Model 2241-3RK (Survey Meter) ........................................... 130
Ludlum GM Pancake Model 44-9 (Survey Meter) ........................................................ 132
Ludlum Model AT- 138 Pocket (Dosimeter) ................................................................. 133
Ludlum Model 25 Personal Radiation Detection Meter ................................................ 134
Fluke Radiation Detection Meters
Fluke ASM 990 Advanced (Survey Meter) ................................................................... 135
Fluke Direct Reading Pocket Dosimeter ....................................................................... 136
HazMat Field Screening Equipment Manual
1/30/2015
Page 5 of 168
California CUPA Forum
Civil Defense Victoreen (CDV) Radiation Meters
CDV 700 Radiation Detection Meter ............................................................................ 137
CDV 715 Radiation Detection Meter ............................................................................ 138
Victoreen 450B Ion Chamber Survey Meter (Fluke Inovision 451B) ............................ 139
Mirion Radiation Detection Meters
Mirion DX-1 Radiation Monitor (Survey Meter) ............................................................. 140
Mirion RDS 31 (Survey Meter) ..................................................................................... 141
Mirion dMC 3000 Personal Electronic Dosimeter ......................................................... 142
W.B. Johnson Instruments DSM-500 Digital Radiation (Survey Meter)........................ 143
Polimaster – Model PM 1405 Survey Radiation Monitoring Meter ............................... 145
RADEX RD 1706 Radiation Detector ........................................................................... 147
Mazur Instruments PRM-9000 Nuclear Radiation Detector and Monitor ...................... 149
Radiation Alert Radiation Monitors ............................................................................... 151
Smiths Detection RadSeeker Handheld Radioisotope Identifier .................................. 153
Gamma RAE II R Radiation Detector ........................................................................... 155
DoseRAE 2 – Electronic Dosimeter.............................................................................. 157
HAZCAT KITS
Haztech Systems, Inc ................................................................................................... 159
Microcat/WMD Kit......................................................................................................... 159
Hazcat Kit – Quick Tests .............................................................................................. 162
TELEMETRY
LINC ............................................................................................................................. 167
HazMat Field Screening Equipment Manual
1/30/2015
Page 6 of 168
California CUPA Forum
PREFACE
The California CUPA Forum (CCF) is an organization consisting of representatives from
Certified Unified Program Agencies (CUPAs) and Participating Agencies (PAs) for promoting consistent implementation of hazardous materials/waste regulatory programs. The CCF works cooperatively with Federal, State and local agencies, industry and members of the public to promote a single, united voice statewide for the handling, storage, and disposal of hazardous materials and waste. These local unified program agencies are mostly fire and environmental health organizations with a primary responsibility to inspect regulated facilities and enforce applicable laws and regulations. Hazardous materials emergency response is a component of this responsibility, especially in the more urban areas of the State. The need to have portable and reliable monitoring equipment is necessary for site characterization, initial identification of hazards present, and to ensure public safety during a hazardous materials incident.
Appropriate use of field screening equipment and environmental protection can aid personnel involved with a hazardous materials situation in determining potential or actual effects on public safety, the environment, immediate and long term risks to public health, actions to mitigate the hazards, and appropriate personnel protection equipment to be used.
This manual was compiled to provide a single source of hazardous materials field detection equipment useful for hazardous waste sites and incident response. This document is intended as a quick reference of equipment and specifications for use. The instructions contained within should not replace the manufacturers operating instructions for the device or equipment. Use of any device for hazard assessment and or identification requires a full understanding of the equipment’s responses, limitations, and mode of operation. This document is also not intended to be a training tool for response personnel, but a quick reminder for someone who has received full and appropriate training to include operations, limitations, and interpretations of results.
The instruments described in this manual represent the current models that are marketed by a variety of equipment manufacturers; and models that have been updated or discontinued but are still used by many Fire and Health HazMat agencies. The manual does not include those outdated and relatively unused equipment and also recognizes that new technology and instrumentation is a constantly evolving trend.
Funding for this Instrument Manual was provided through the CUPA Forum Environmental
Protection Trust Fund, established to manage and disburse funds from enforcement case settlements. The document was compiled by Jeff Benedict, former City of Long Beach
Hazardous Waste Operations Officer through document research; and consultation with local
CUPA and fire department hazmat agencies, and equipment vendors. The project manager for the CCF was Division Chief Bill Jones, Los Angeles County Fire Department. Comments and suggestions may be directed to Chief Jones at [email protected] or the CCF
Manager Sheryl Baldwin at [email protected].
HazMat Field Screening Equipment Manual
1/30/2015
Page 7 of 168
California CUPA Forum
INTRODUCTION
Need and technology has produced an array of instruments to comply with OSHAs standards to identify and evaluate unseen hazards at locations requiring an evaluation due to the presence or potential presence of hazardous substances. The instruments and tools in use are designed to indicate specific types of hazards or characteristics of the materials being evaluated within a designated operating or indication range. Trained professionals, such as Environmental or
Hazardous Material Specialists, Industrial Hygienists, Safety Officers, or emergency responders can use hazardous materials field screening equipment for a variety of purposes, not the least of which includes protection of the public and response personnel during an emergency situation or disaster involving unknown or known hazardous materials.
With the large array of field screening tools available, the CCF set out to address two concerns in compiling this manual. First, there are agencies that have many of the instruments and tools described in this manual and need a “quick and dirty” resource to refresh staff on the screening tool and its use. This manual, with the links to manufacturers’ manuals gives the information to the user quickly and conveniently. Second, in some jurisdictions, multiple agencies often have varying tools and this manual will allow users to quickly learn what the tool can do and any concerns regarding its use.
These hazardous materials field screening tools analyze various media (gas, liquids or solids) for a wide variety of parameters and assists responders and other personnel responsible to evaluate site conditions to gauge risk, develop mitigation strategies and options for reducing or eliminating hazardous conditions. This assessment also assists in identifying needed personal protection equipment and actions necessary to keep workers safe during mitigation activities.
This manual has been compiled using instructions and information provided by the manufacturer of the specific instrument or equipment. Users must read the manufacturer’s instructions completely for a full understanding of the equipment. Summary information and specifications are provided with references to website links for more detailed instrument use, calibrations, and maintenance instructions.
Most importantly, this guide does not describe or intend to assist in the interpretation of the results and findings using the field screening tools described herein. Only properly trained professionals, knowledgeable in the many variables that could affect result interpretations should be making decisions or giving directions based on the field results.
HazMat Field Screening Equipment Manual
1/30/2015
Page 8 of 168
California CUPA Forum
______________________________________________________________________
INSTRUMENT APPLICATIONS
Combustible Gas Indicator
(CGI)
Oxygen Meter (O
2
Multi Gas Detector
Color Metric Tube
Fire or Explosion Hazard
Inorganic and Organic Gases and Vapors
Measures the concentration of combustible gas or vapor in air
Determines % O
2 in air and fire hazard
Identifies Hazardous conditions
Specific Gases and Vapors
Measure concentration of specific gases and vapor
Chemical Specific Instruments
Radiation Meters
Photoionization Detectors
(PID)
Flame Ionization Detector
(FID)
Infrared
Toxicity Establish exposure levels
Radiation
Identify presence of a radiation source
Organic and Inorganic Gases and Vapors
Measure total organic vapors
Organic Gases and Vapors Measure total organic vapors
Hazardous Substance Compound identification
HazMat Field Screening Equipment Manual
1/30/2015
Page 9 of 168
California CUPA Forum
______________________________________________________________________
EPA ACTION GUIDES
(Atmosphere Hazards)
Combustible Gases
< 10%
< 5%*
10 – 25%
LEL
LEL
LEL
Continue monitoring
Continue with caution
Continue monitoring with
> 25%
>5%*
LEL
LEL
Explosion Hazard withdraw from area Immediately
Oxygen Concentrations
Must wear SCBA < 19.5%
19.9 – 25%
> 25%
Continue monitoring with caution SCBA not needed based on O
2
Discontinue monitoring Fire, Consult Fire
Toxic Gases and Vapors
> IDLH SCBA required
HazMat Field Screening Equipment Manual
1/30/2015
Page 10 of 168
California CUPA Forum
Instrument Manual Mailing List
RAE Systems, Inc.
3775 North First Street
San Jose, CA 95134 USA
Phone: 408.952.8200
Fax: 408.952.8480
GfG Instrumentation Inc.
1194 Oak Valley Drive, Suite 20
Ann Arbor, Michigan 48108
Western Region: GfG Instrumentation, Inc. (800) 959-0329 Mike Calvo (Irvine)
MSA Manufacturer, (800) 672-2222 /800-672-2222 Lauren – customer service
Pine Environmental Services,
2350 Reynolds, Suite #117
Irvine, CA
(949) 955-3930
RKI Instruments, Inc.
33248 Central Ave.
Union City, CA 94587
RKI Tech Support 800-754-5165 (Mike Jeter)/Shirley Trejis) photos
Rhett Squires Retail Representative (Ventura) 805-218-8853
Smiths Detection Inc.
2202 Lakeside Boulevard
Edgewood, MD 21040
410-844-5186 fax 1 410-510-9496 [email protected]
Don Wolf & Associates, Inc.
21955 Plummer Street
Chatsworth, CA 91311
818-885-5800
800-266-2046
HazMat Field Screening Equipment Manual
1/30/2015
Page 11 of 168
California CUPA Forum
Honeywell Analytics
651 South Main Street
Middletown, CT 06457
800 711-6776 860 344-1079
Fax 860 344 – 1068
HazMat Field Screening Equipment Manual
1/30/2015
Page 12 of 168
California CUPA Forum
MULTIPLE-GAS METERS
Multiple-Gas handheld meters are used to detect multiple gases simultaneously. Gases are detected using sensors to include: catalytic, electrochemical, photoionization (PID) infrared (IR) and thermo-conductivity. The ‘Four (4)-gas’ meter is the most common instrument used and is designed to detect the confined space gases: CO, H2S, O2 and combustible gases (LEL). Five
(5)’ and ‘Six (6)’ gas meters are also available to increase the monitoring capabilities. Sensors can be interchangeable and replaced by sending to the manufacturer or by trained staff.
Glossary of Specifications
Sensors
Photoionization Detection (PID) Use of a high energy ultraviolet (UV) light source to ionize chemicals in an airstream, The charged molecules are collected on a charged surface, which generates a current that is directly proportional to the concentration of the chemical in the air being sampled for the measurement of volatile organic compounds and other gases. PIDs are broadband detectors with a sensitivity that differs for each
Infrared (IR)
VOC.
Useful for measuring a broad range of inorganic and organic chemicals in the air IR analyzers operate by passing Infrared radiation generated from a heated metal source through a gas sample. The IR radiation is absorbed by the chemical in the sample. The IR analyzer is preloaded with library of known chemicals for identification and concentration of
Electrochemical Sensors chemicals in the air. It measures a variety of gases to include CO2, CO, CH4, NO and SO2.
Measures the concentration of a target gas by oxidizing or reducing the gas at an electrode and measuring the resultant current commonly used for CO and H2S.
Catalytic Combustible Sensors Catalytic bead sensors are used primarily for the detection of combustible gases Based on the premises that all
Oxygen Sensors electrically conducted materials change their conductivity as temperature changes. There is a coefficient of temperature
resistance
(ct) for each combustible gas mixture.
Based on electrochemical (galvanic) cells. The generated current in the sensor, which is produced from an oxidation reaction, is directly proportional to the rate of oxygen diffusion into the cell. Meters are generally calibrated oxygen concentrations between 0 -25%. Normal oxygen is 20.9%.
Meter alarms are set at concentrations lower than 19.5% or above 23.5%.
HazMat Field Screening Equipment Manual
1/30/2015
Page 13 of 168
California CUPA Forum
Explosivity Sensors For combustible gases use of an oxidizing agent such as platinum or palladium combustible gas meters measure flammable gas concentration as a percentage of the LEL of the calibrated gas.
Use of electrochemical technology (NH3, HCN, PH3, SO2.) Toxic Gas Sensors
Detection Method
Diffusion
Pump
Passive monitoring installed in a port on the meter.
Internal or external pump to actively collect gases/vapor.
Can be used with attached tubing for more specific or remote gathering points.
External handheld pump (bulb) with tubing Aspirator
Meter Display
LED
LCD
Liquid Emitting Diode
Liquid Crystal Display
Note: Both LED and LCD are considered ‘state of the art’ displays and the differences for use in portable meters are arguably negligible
Power Sources
Alkaline Batteries
Lithium Ion
Nickel Metal Hydride
AAA, AA, C, D size and 9 volt snap on. These batteries are generally single use/disposable. Rechargeable alkaline batteries are available but seldom used. Interchangeable with lithium and nickel rechargeable batteries.
(Li-Ion) most commonly used rechargeable battery.
Nickel Cadmium
Direct Current
(NiMH) Rechargeable battery, common use for AA penlight size
(NiCD) Small rechargeable battery
(DC Power) can be used to recharge batteries and may be plugged into vehicle ‘cigarette lighters’ for long-term monitoring.
Power generated for plug-in to utility supplied systems Alternating Current
Miscellaneous
STEL/TWA Short Term: X Exposure Limit value above which exposure to a chemical substance should not occur. Usually relates to a 15 minute reference period.
Time Weighted Average: The exposure to a chemical can be used when both the chemical concentration and time for
HazMat Field Screening Equipment Manual
1/30/2015
Page 14 of 168
California CUPA Forum
Data Logging
Alarms
Calibration
Modes of Operation exposure varies over time. It is used as the average exposure to a contaminant to which workers may be exposed without adverse effect over a period such as an 8 hour day or 40 hour week.
Turn on/off, gas readings, battery condition, and calibration data. Software is available for transfer of information to a PC for data storage and remote monitoring.
Instruments are equipped with audio, visual and or vibrating alarms that may be set at threshold limits for gas or conditions of concern.
Test/Auto-calibration checks Fresh air/zero calibration. Use of manufacturers supplied known gas exposure for appropriate readings.
Stand Alone Docking Stations. Available to cradle calibration gas and expedite the process/provide additional data storage capabilities.
Full Calibration Outsourced to manufacture/representative or performed in-house by trained technicians to certify calibration of meter accuracy.
Normal Mode: Confined space or area monitoring
Bar Hole Mode: Underground gas leaks/only combustible or
O2.
Leak Check Mode: Location of leaks in valves and piping.
HazMat Field Screening Equipment Manual
1/30/2015
Page 15 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 16 of 168
California CUPA Forum
RKI Multiple Gas Meters
Models: RKI Eagle 2
RKI GX-2012
RKI GX-2009
RKI Eagle 2
Use of Instrument: Handheld meter capable of monitoring up to six different gases simultaneously. For use in confined spaces, ‘Bar Hole’ (underground leaks) and hazardous environments. The meter has interchangeable sensors for detection of different gases.
Specifications
Measurements of Gases and Detectable Ranges
Combustibles (CH4)
Oxygen (O2)
0—100% LEL
0—40% VOL
Carbon Monoxide (CO) 0—500 PPM
Hydrogen Sulfide (H2S) 0—100 PPM
Ammonia (NH3)
Arsine (ASH3)
Phosphine (PH3)
Sulfur Dioxide (SO2)
0—75 PPM
0—1 PPM
0—1 PPM
0—10,000 PPM
Carbon Dioxide (CO2) 0—10,000 PPM
Methane (CH4) 0—100%LEL
Methane (CH4) 0—100% VOL
VOC 0—2,000PPM
VOC 0—50PPM
HazMat Field Screening Equipment Manual
1/30/2015
Page 17 of 168
California CUPA Forum
Sensors
Detection Method
Power Source
Electrochemical, catalytic combustion, galvanic cell, toxic, thermal
Conductivity, PID, IR
Internal pump (draw up to 125 feet)
MIMH (rechargeable) or Alkaline size C batteries
Display LCD
Data Logging Equipped with Eagle 2 Data Logger Management Program. Data is logged to internal memory and can be downloaded to computer infrared capabilities. Data includes gas readings, alarm and calibration information.
Calibration Bump test and auto-calibration using manufacturer issued kit.
Stand Alone: RKI Gas Monitor Calibration Station SDM E-2 for enhanced calibration and record keeping.
RKI Eagle 2
Start Up (Normal Operation):
1. Determine if a sample hose will be needed and configure the instrument based upon the intended tasked and objectives. Connect the sample hose and or probe.
2. Press and hold briefly the POWER ENTER RESET button until you hear a beep.
The machine will cycle through the start- up sequence and enter the monitoring mode. If needed, perform a calibration. To skip calibration, press and release the DISPLAY
ADJUST NO
button.
3. In an environment free of toxic or combustible gases and normal oxygen, perform a demand zero (fresh air set) by pressing and holding the AIR ▲YES button until the unit prompts you to release.
The unit is now ready to monitor. Proceed to the monitoring area then allow 10 – 15 seconds for response (response increases with length of hose).
A rapid increase in combustible gas readings followed by a declining or erratic reading may indicate a gas concentration above the LEL which may be hazardous.
To monitor for combustible gases in the ppm range allow the instrument to run for 3 – 5 minutes after the start-up sequence is complete. This stabilization period is not needed for monitoring in the % LEL or % volume range.
HazMat Field Screening Equipment Manual
1/30/2015
Page 18 of 168
California CUPA Forum
Resetting and Silencing Alarms
If the instrument goes into alarm press and release any button to silence. The concentration must fall below the alarm setting before you can reset the alarm.
If the instrument will not silence, the machine is set to alarm silence off and selfresetting.
Once the condition passes, press and release the POWER ENTER RESET button.
Powering Down the Unit:
Press and hold the POWER ENTER RESET button. Release the button when “Good Bye” and the RKI logo appear.
Calibration Station SDM E-2
Eagle 2 Operator’s Manual: www.rkiinstruments.com/pdf/71-0154RK.pdf
RKI Tech Support (800) 754-5165
Rhett Squires Retail Representative (Ventura) (805) 218-8853
HazMat Field Screening Equipment Manual
1/30/2015
Page 19 of 168
California CUPA Forum
RKI GX-2012
Use of Instrument:
Small handheld meter. Monitors up to five different gases simultaneously; for use in confined space entry, ‘Bar Hole’ monitoring (underground gas leaks and other hazardous environments. (Replaces RKI GX 2003)
Measurements and Detectable Ranges
Combustible Gases (CH4) 0 – 100% LEL
Oxygen (O2) 0 – 40% VOL
Hydrogen Sulfide (H2S) 0 – 100 PPM
Carbon Dioxide (C02) 0 – 500 PPM
Combustible Gases (CH4) 0 – 100% VOL
Sensors Catalytic combustion, thermal conductivity, galvanic and electrochemical cells
Detection Method
Power Source
Internal pump (draw up to 50 feet)
Lithium Ion (rechargeable) or Alkaline size AA batteries
Display LCD
Data Logging Equipped with GX 2012 Logger Management Program software. Data is logged to internal memory and can be downloaded to a computer through infrared capabilities to log gas readings, alarm and calibrated information.
HazMat Field Screening Equipment Manual
1/30/2015
Page 20 of 168
California CUPA Forum
Calibration Bump testing/auto or single calibration uses manufacturer’s calibration kit. Stand Alone Docking Station: RKI SDM 2012 Calibration Station for enhanced calibration and data retrieval.
RKI SDM 2012 Calibration Station
RKI GX – 2012
Start Up (Normal Operation) Only:
1. Determine if a sample hose will be needed and configure the instrument based upon the intended tasked and objectives. Connect the sample hose and or nozzle.
2. Press and hold briefly the POWER ENTER button until you hear a beep. The machine will cycle through the start- up sequence until it reaches the calibration mode.
3. Calibrate using Auto Cal or One Cal.
4. In an environment free of toxic or combustible gases and normal oxygen, perform fresh air adjustments by pressing and holding the AIR ▲ button until the unit prompts you to release.
The unit is now ready to monitor. Proceed to the monitoring area then allow 10 – 15 seconds for response (response increases with length of hose). If beep set is turned on, the unit will beep once every 5 minutes to confirm that it’s operating.
Note: A heart symbol flashes in the display while the instrument is functioning properly
HazMat Field Screening Equipment Manual
1/30/2015
Page 21 of 168
California CUPA Forum
and, a fan symbol spins while the pump is operating.
In a low light environment press and release any button to turn on display backlight.
Resetting and Silencing Alarms
If the instrument goes into alarm press and release the RESET SILENCE button to silence, LEDS will continue to flash. The concentration must fall below the alarm setting before you can reset the alarm.
If the instrument will not silence, the machine is set to alarm silence off. Once the condition passes, press and release the RESET SILENCE button.
Powering Down the Unit:
Press and hold the POWER ENTER button. The buzzer will pulse and “Turn off” will appear in the display. Release the POWER ENTER button when “Turn off” disappears.
Operator’s Manual – RKI Instruments http://www.rkiinstruments.com/pdf/71-0239RK.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 22 of 168
California CUPA Forum
RKI GX-2009
Use of Instrument: Small portable gas meter. Monitors up to four different gases simultaneously.
For use in confined spaces and hazardous environments. (Replaces RKI GX 2001)
Specifications:
Measurements – Gases and Ranges
Combustible Gases (CH4) 0 – 100% LEL
Oxygen (O2) 0 – 40 PPM
Hydrogen Sulfide (H2S) 0 – 100 PPM
Carbon Dioxide (C02) 0 – 500 PPM
Sensors
Detection Method
Catalytic combustion, electrochemical and galvanic cells
Diffusion/attached powered pump/hand aspirated sample draw with hose and probe
Power Source NiMH batteries
Display LCD
Data logging
Calibration
Equipped with GX 2009 Data Logger Management Program software.
Data is logged to internal memory and can be downloaded to a computer through infrared capabilities. Data includes gas readings, alarm and calibration information.
Bump test and auto-calibration using manufacturers issued kit.
Stand Alone Docking: RKI Cal 2000 Docking Station for enhanced calibration and data retrieval.
HazMat Field Screening Equipment Manual
1/30/2015
Page 23 of 168
California CUPA Forum
RKI Cal 2000 Docking Station
RKI GX – 2009
Start Up:
1. In a known fresh air environment press and briefly hold down the POWER MODE button. Release when you hear a beep. The start-up sequence will begin (alarm LEDs and buzzer will pulse several times).
The machine will cycle through the start- up sequence until it reaches the calibration and bump test mode.
2. Press the POWER MODE button to confirm calibration limits and bump test.
3. The unit will perform an automatic fresh air adjustment and proceed into the monitoring mode and is ready for use.
4. Prior to entering the impacted area, with the unit operating in the monitoring mode perform a fresh air adjustment by pressing and holding the AIR button until the unit prompts you to release.
5. Proceed to the monitoring area.
HazMat Field Screening Equipment Manual
1/30/2015
Page 24 of 168
California CUPA Forum
Resetting Alarms
To reset the alarm, the gas reading that caused the alarm must decrease below the alarm set point. Press and release the POWER MODE button.
Powering Down the Unit:
Press and hold the POWER MODE button for about 3 seconds to turn off the unit. Release button when the LCD is blank.
Operator’s Manual – RKI Instruments http://www.rkiinstruments.com/pdf/71-0158RK.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 25 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 26 of 168
California CUPA Forum
MSA Multiple Gas Meters
Models: MSA Sirius Multi-gas Detector
MSA Altair 5X
MSA Altair 4X
MSA Altair Pro Single Gas Detector
MSA Sirius Multi-Gas Detector
Use of Instrument: Handheld multi-gas meter will simultaneously monitor up to four gases with a PID sensor for monitoring of low vapor VOCs. PID capabilities have a library of the detection of up to 100 gases. For confined space entry and other hazardous environments.
Specifications:
Measurements and Detectable Ranges
Combustible Gases 0 - 50% LEL
Oxygen (O2 0 - 25% VOL
Carbon Monoxide (CO) 0 - 500 PPM
Hydrogen Sulfide (H2S) 0 - 200 PPM
VOCs 0 - 2,000 PPM
Sensors Catalytic, electrochemical, PID sensors
PID Sensor measures VOCs with isobutylene as the reference gas.
Detection Method
Power Source
Internal pump
Lithium Ion (rechargeable) or Alkaline batteries
HazMat Field Screening Equipment Manual
1/30/2015
Page 27 of 168
California CUPA Forum
Display LED
Data Logging
Calibration
MSA Link software works with IR reader USB port plugged into a computer for retrieval and storage of events.
Bump test and auto-calibration using manufactures kit.
Stand Alone: MSA Galaxy Automated Calibration System and for enhanced calibration and data retrieval.
MSA Galaxy Automated Calibration System
Operating Manual (turn on Ch. 3.3 page 19)
Reference Link: http://www.livoniafirefighters.com/docs/MSA Sirius Manual.pdf
MSA Manufacturer (800) 672-2222
Sirus MultiGas Detector
Start Up:
1. Determine if a Fresh Air Set-up (FAS) should be used. In a fresh air environment press the POWER ON button to begin the start-up cycle and self-check.
2. Perform the FAS or skip the FAS a. To perform the FAS press ON/OFF button while zero is flashing
HazMat Field Screening Equipment Manual
1/30/2015
Page 28 of 168
California CUPA Forum
b. To skip FAS, press the RESET/▼button.
3. The unit is in the measuring mode and ready for monitoring when the measure gas page appears.
4. Verify pump operation by plugging the free end of the pump inlet or probe. Alarm must activate (pump should be checked before each use).
Press the RESET/▼ button to reset the alarm and restart the pump.
5. Ensure PID bulb corresponds to PID bulb setting on instrument display.
Note: Default bulb is 10.6
Organic silicones, silicates, lead containing compounds, hydrogen sulfide exposures over 200 ppm or exposures over 55 ppm for one minuet may desensitize the combustible gas sensor and/or voc sensor reducing its readings.
(Check each sensors calibration prior to field use.)
Silencing Alarms
Alarms can be silenced by pressing the RESET/▼button.
Combustible gas alarm cannot be reset with the reset button if the indication reaches
100% LEL or 5% CH
4.
. Proceed to a fresh air area and power down then restart the meter to reset.
Powering Down the Unit:
Press and hold the ON/OFF/ACCEPT button for 3 seconds. Four beeps will be heard during the turn off sequence.
HazMat Field Screening Equipment Manual
1/30/2015
Page 29 of 168
California CUPA Forum
MSA ALTAIR 5X Multi-gas Detector
Use of Instrument: Handheld multi-gas meter will simultaneously monitor up to five different gases. For confined space and other hazardous environments.
Specifications:
Measurements and Detectable Ranges
Combustible
Oxygen (O2)
0 - 100% LEL
0 - 30% VOL
Carbon monoxide (CO) 0 - 2,000 PPM
Hydrogen Sulfide (H2S) 0 - 200 PPM
Sulfur Dioxide (SO2)
Chlorine (CL)
0 - 20 PPM
0 - 10 PPM
Ammonia (NH3) 0 - 100 PPM
Nitrogen Dioxide (NO2) 0 - 20 PPM
Hydrogen Cyanide (HCN) 0 - 30 PPM
Phosphine (PH3) 0 - 5 PPM
Carbon Dioxide (CO2) 0 - 10% VOL
Butane (C4H10) 0 - 25% VOL
Methane (CH4)
Propane (C3H8)
0 - 100% VOL
0 - 100% VOL
Sensors IR (Infrared) sensor and interchangeable sensors (X Cell) sensors for the detection of a variety of gas types
Detection Method
Power Source
Internal Pump
Lithium Ion (rechargeable) or Alkaline AA size batteries
HazMat Field Screening Equipment Manual
1/30/2015
Page 30 of 168
California CUPA Forum
Display LED (monochrome or color)
Data Logging MSA Link Software The data link communication port on the meter is aligned to PC IR interface
Calibration Bump test and auto-calibration using manufacturer’s kit.
Stand Alone: MSA Galaxy Calibration System for enhanced calibration and data retrieval.
Altair 5x Multi Gas Detector
Start Up:
1. Determine if a Fresh Air Set-up (FAS) should be used.
2. In a fresh air environment press the ON/OFF button to begin the start-up cycle, selfcheck and sensor warm up period.
3. Verify pump operation by blocking the end of the pump inlet or probe when the alarm sounds within 30 seconds. Watch for a pass or error message.
If the pump passes:
Press the RESET/▲ button to reset the alarm and restart the pump.
If an error occurs check for an leak in the sampling system and retest.
4. Perform the FAS or skip the FAS a. To perform the FAS press ON/OFF button while zero is flashing b. To skip FAS, press the RESET/ ▼ button.
5. The unit is in the measuring mode and ready for monitoring when the measure gas page appears.
Silencing Alarms
Alarms can be silenced by pressing the RESET/ ▼button.
Combustible gas alarm cannot be reset with the reset button if the indication reaches
100% LEL or 5% CH
4
. Proceed to a fresh air area and power down then restart the meter to reset.
Powering Down the Unit: Press and hold the ON/OFF button for 3 seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 31 of 168
California CUPA Forum
MSA ALTAIR 4X Multi-gas Detector
Use of Instrument: Handheld multi-gas meter monitors up to four different gases simultaneously. For use in confined space entry and other hazardous environments.
Fluorescent Model
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 - 100% LEL
0 - 30% VOL
Carbon Monoxide (CO) 0 - 1,999 PPM
Hydrogen Sulfide (H2S) 0 - 200 PPM
Nitrogen Oxide (NO2) 0 - 50 PPM
Sulfur Dioxide (SO2) 0 - 20 PPM
Sensors Dual Tox Sensor (CO/H2S), X Cell (trademark) Sensors
Display LCD
Power Source
Detection
Calibration
Rechargeable Lithium-polymer
Diffusion (Accessories – Universal Pump Probe)
Alarms STEL/WTA;
Data Logging MSA Link Software Data link communication to IR interface with PC
Bump test and auto-calibration using manufacturer supplied kit.
Stand Alone: MSA Galaxy Calibration System for enhanced calibration and data retrieval.
HazMat Field Screening Equipment Manual
1/30/2015
Page 32 of 168
California CUPA Forum
Operating Manual: MSA ALTAIR 4X
MSA Manufacturer (800) 672-2222
Pine Environmental Services, 2350 Reynolds, Irvine, Suite #117 (949) 955-3930
Altair 4x Multi Gas Detector
Start Up:
1. Determine if a Fresh Air Set-up (FAS) should be used.
2. In a fresh air environment press the ON/OFF button to begin the start-up cycle, selfcheck.
3. Verify that LCD and audible alarm sounds briefly.
4. Perform the FAS or skip the FAS a. To perform the FAS press ON/OFF button within 10 seconds b. To skip FAS, press the ▲ button.
5. The unit goes into the measuring mode and ready for use
A ♥ in the display indicates the sensor has reached or near the end of life.
Activate backlight by pressing any button
Silencing Alarms
Alarms can be silenced by pressing the ▲ button. If alarm condition continues to exist, the ▲ will only silence for 5 seconds.
False O2 alarms can occur due to changes in barometric pressure or extreme changes in ambient temperature.
Lock alarm state (“xxx” appears in display) can only be reset by turning the unit off and on in a fresh air environment.
Powering Down the Unit:
Press and hold the ON/OFF button
HazMat Field Screening Equipment Manual
1/30/2015
Page 33 of 168
California CUPA Forum
MSA Altair Pro Single Gas Detector
Use of Instrument: Handheld one gas meter for Oxygen, Hydrogen, Sulfide and Carbon
Monoxide using interchangeable electrochemical sensors for specific gas to be monitored. For measurement of known chemical in potentially hazardous environments.
Specifications:
Measurements and Detectable Ranges
Oxygen (O2) 0 - 25% VOL
Carbon Monoxide (CO) 0 -500 PPM
Hydrogen Sulfide (H2S) 0 - 100 PPM
Sensors Electrochemical Sensors (O2, CO, H2S)
Detection Method Diffusion
Power Source Lithium (rechargeable) batteries
Display LCD
Data Logging Capacity to record events and transfer data to a PC receiving device with
MSA Five Star Link Software
Calibration Bump test and auto-calibration using manufacturers supplied kit.
Stand Alone: MSA Galaxy Calibration System for enhanced calibration and data retrieval.
Operating Manual
MSA ALTAIR SINGLE GAS DETECTOR
HazMat Field Screening Equipment Manual
1/30/2015
Page 34 of 168
California CUPA Forum
Altair Single Gas Detector
Start Up:
1. Press and hold the TEST button until “ON” and “?” appear. Release the button and press again to activate the unit. A 99 second activations cycle begins.
2. After activating, press the TEST button to perform an alarm test. If the alarm LED and sounds do not activate do not proceed with the unit.
Once activated the unit remains activate until low battery error occurs.
Vibrating alarm may cease to operate if temperature is below 0
◦
C.
A confidence indicator flashes every 60 seconds to indicate that the unit is on and operating normally.
Do not block the sensor.
HazMat Field Screening Equipment Manual
1/30/2015
Page 35 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 36 of 168
California CUPA Forum
GfG Instrumentation
Multi-gas
GfG
GfG Instrumentation 450 Multi-Gas Detector
Use of Instrument: Handheld 1 - 4 sensor meter for the monitoring of up to four gases simultaneously Used for confined space and hazardous environments.
Specifications:
Measurements and Detection Ranges
Combustible Gases 0 – 100% LEL
Carbon Monoxide (CO) 0 – 500 PPM
Hydrogen Sulfide (H2S) 0 – 100 PPM
Oxygen (O2) 0 – 25% VOL
Sensors Four “G45” sensors
Detection Method Diffusion mode or attachable motorized pump
Power Source NiMH (rechargeable) or alkaline AA size batteries
Display LCD
HazMat Field Screening Equipment Manual
1/30/2015
Page 37 of 168
California CUPA Forum
Data logging
Calibration
GfG 450 Simulator Software Program Records gas concentrations and alarms. Events can be downloaded to a PC for viewing, storage and evaluation.
Bump test and auto-calibration using manufacturers calibration kit.
Docking Station: OEM GfG DS404
Docking Station: OEM GfG DS404
Operators Manual
G450 MULTI-GAS DETECTOR
GfG Instrumentation, Inc. (800) 959-0329
G450
Start Up (Normal Operation):
1. Press button 3 to activate the unit. Allow the instrument to stabilize then completely warm up (approximately 5 min.).
2. If fresh air adjustment is needed attach the calibration adapter. Push the “AIR” button to initiate fresh air adjustment. Remove adapter once completed, the unit is ready for normal use.
Reset alarms by pushing the RESET key (button 2). Press yes when prompted to resume detection.
Green back light, No alarms
Orange backlight, 1 alarm
Red backlight, 2 or 3 alarms
Alarms in fresh air, perform an auto zero
↑↑↑ Indicates detection range of LEL sensor has been exceeded.
Shut down unit by pressing and holding button 3 for 5 seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 38 of 168
California CUPA Forum
G460 Multi-gas Detector
Instrument Use: Handheld 1 – 6 channels for simultaneous monitoring of a variety of gases.
Features include PID, IR, capability and a wide range of sensors. Use in hazmat response and confined spaces.
Specifications:
Measurements of Gases and Detectable Ranges
Combustibles
Oxygen (O2)
0 – 100% LEL
0– 25% VOL
Carbon monoxide (CO) 0 – 300 PPM
Hydrogen sulfide (H2S) 0 – 100 PPM
Ammonia (NH4)
Carbon dioxide (CO2)
0 – 200 PPM
0 – 10 PPM
Chlorine (CL2) 0 – 2 PPM
Ethylene oxide (C2H4O) 0 - 20 PPM
Hydrogen (H2) 0 – 2,000 PPM
Hydrogen chloride (HCL) 0 - 30 PPM
Hydrogen cyanide (HCN) 0 - 50 PPM
Hydrogen fluoride (HF) 0 - 10 PPM
Nitric oxide (NO2) 0 - 100 PPM
Nitrogen dioxide (NO2) 0 – 50 PPM
Ozone (O3) 0 - 1 PPM
Phosgene (CL2)
Phosphine (PH2)
0 - 10 PPM
0 – 10 PPM
PID (VOC)
Silane (SiH4)
Sulfur dioxide (SO2)
0 - 2,000 PPM
0 - 40 PPM
0 - 10 PPM
HazMat Field Screening Equipment Manual
1/30/2015
Page 39 of 168
California CUPA Forum
Sensors
Detection Method
Power Source
PID sensors for direct reading of VOCs; IR; and smart (electrochemical and catalytic sensors) for detection of a variety of gases
Diffusion mode or attachable motorized pump
NIMH (rechargeable) or alkaline AA size batteries
Display LED
Data Logging GfG Simulator Software Program records gas concentration and alarms. Events can be downloaded to a PC for viewing, storage and evaluation
Calibration Bump test and auto-calibration using manufacturer supplied calibration kit. Docking Station: OEM GfG DS404
Field Operations Manual
G460 MULTI-GAS DETECTOR
G460
Start Up (Normal Operation):
1. Press button 3 to activate the unit.
2. Allow the instrument to stabilize then complexly warm up (approximately 5 min.).
3. If fresh air adjustment is needed attach the calibration adapter.
4. Push the “AIR” button to initiate fresh air adjustment.
5. Remove adapter once completed, the unit is ready for normal use.
Reset alarms by pushing the RESET key (button 2). Press yes when prompted to resume detection.
Green back light, No alarms
Orange backlight, 1 alarm
Red backlight, 2 or 3 alarms
Alarms in fresh air, perform an auto zero
Shut down unit by pressing and holding button 3 for 5 seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 40 of 168
California CUPA Forum
RAE Systems Multi Gas Meters
Models:
QRAE II Gas Detector
QRAE 3 Wireless 4-Gas Monitor
MultiRAE Multigas Detection Monitors ppbRAE3000/MiniRAE3000
ToxiRAE Single Gas Monitors
AreaRAE Steel Multi-Gas Monitoring System
QRAE II GAS Detector
Use of Instrument: Handheld 1 - 4 gas monitor for the simultaneous detection of combustibles,
O2, H2S, or CO. For confined space entry and other hazardous environments.
Specifications:
Measurement and Detection Ranges
Combustible
Oxygen (O2)
0 – 100% LEL
0 – 30% VOL
Carbon Monoxide(CO) 0 – 1,000 PPM
Hydrogen Sulfide (H2S) 0 – 100 PPM
Sensors 2 electrochemical (CO/H2S); 1 solid electrolyte (O2); 1 catalytic
(combustible)
Detection Method Diffusion or pump (depending on the QRAE II model) option – hand pump
(squeeze bulb)
Power Source Lithium Ion (rechargeable) and alkaline size AA batteries
HazMat Field Screening Equipment Manual
1/30/2015
Page 41 of 168
California CUPA Forum
Display LCD
Data logging ProRAE Studio Software Download data to PC capabilities
Calibration Bump test and auto-calibration using manufacturer’s test kit.
Stand Alone: Auto RAE Lite Docking Station for enhanced calibration and data retrieval.
User’s Guide
QRAE II GAS DETECTOR
Auto RAE Lite Docking Station for QRAE II
RAE Systems (San Jose) (408) 952-8200 / (877) 723-2878
HazMat Field Screening Equipment Manual
1/30/2015
Page 42 of 168
California CUPA Forum
QRAE 3 Wireless 4-Gas Monitor
Instrument Use: Handheld meter with 1-4 gas capabilities for simultaneously monitoring with wireless real-time readings and alarm status. For confined space and other hazardous environments.
Specifications:
Measurements and Detection Ranges
Oxygen (O2)
Combustible
0 - 30% VOL
0 - 100% LEL
Carbon Monoxide (CO) 0 - 500 PPM
Hydrogen Sulfide (H2S) 0 – 100 PPM
Sulfur Dioxide (SO2) 0 – 20 PPM
Hydrogen Cyanide (HS) 0 – 50 PPM
Sensors Up to 4 replaceable sensors: Electrochemical (H2S, CO,SO2,HCN), electrolyte Oxygen (O2), catalytic (combustible)
Detection Method Built-in pump or diffusion
Data Logging Data download and instrument setup on PCC via Travel Charger
Wireless data and status transmission via built in RF modem. ProRAE
Guardian Software
Calibration Bump test and auto-calibration using manufacturer supplied test kit.
Stand Alone: AutoRAE Lite Docking Station; AutoRAE 2 Station
User’s Guide
QRAE 3 WIRELESS 4-GAS MONITOR
HazMat Field Screening Equipment Manual
1/30/2015
Page 43 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 44 of 168
California CUPA Forum
MultiRAE Multigas Detection Monitors
Models: MultiRAE Pro, MultiRAE Standard, MultiRAE Lite
Instrument Use: Portable handheld 1 – 5 gas meter that will simultaneously monitor a variety of gases using interchangeable electrochemical sensors and PID capabilities. Used in confined space entry and other hazardous environments. Three MultiRAE models: MultiRAE Pro is the top of the line of the models with 10.6eV PID capability and sensor capability to measure gamma radiation; MultiRAE Standard monitors up to 6 contaminants; MultiRAE Lite best suited for confined space gases. (Replaces MultiRAE Plus).
PRO
Specifications:
Measurements and Detection Ranges
Standard
STANDARD
LITE
Oxygen (O2)
Combustible Gas
VOCs
0 – 30 ppm
0 – 100% LEL
0 – 200 ppm
Carbon Monoxide (CO) 0 – 500 ppm
Hydrogen Sulfide H2S) 0 – 100 ppm
Sulfur Dioxide (SO2) 0 - 20 ppm
VOCs 0 – 5000 ppm
Sensors Interchangeable sensors for radiation (gamma), PID and electrochemical for toxics, oxygen and combustible LEL
Detection Method Diffusion or internal pump (15 foot sampling wand)
Power Source NiCd (rechargeable) and alkaline AA size batteries
Alarms STWL/TWA; Audio/Visual/Vibrating
HazMat Field Screening Equipment Manual
1/30/2015
Page 45 of 168
California CUPA Forum
Data Logging
Calibration
ProRAE Suite Software This meter can store gas concentrations and other data for data downloading into a PC. Wireless data and status can be transmitted via built-in RF modem.
Bump and auto-calibration manually by use of manufacturer supplied gases. Docking Station: AutoRAE2 Test and Calibration Station.
AutoRAE2 Test and Calibration Station
User’s Guide
MULTIRAE MULTIGAS DETECTION GAS MONITORS
HazMat Field Screening Equipment Manual
1/30/2015
Page 46 of 168
California CUPA Forum
ppbRAE3000/MiniRAE 3000
Use of Instrument: Portable handheld meter – single use (display) for VOCs. This instrument has broadband VOC gas monitoring using PID detection capabilities. The ppbRae 3000 detects at 1 ppb – 10,000 ppm; the mini RAE detects at 0 – 15,000 ppm.
This meter features PID capabilities for measuring VOC’s of over 200 different gases. For use in HazMat and environmental hazard areas.
_____________________________________________
______________________________________________
Specifications:
Measurements and Detection Ranges ppbRAE 3000 – VOCs 1 ppb – 10000 ppm miniRAE 3000 – VOCs 0 – 15000 ppm
Gases are organized into four lists:
1. My List – Customized list that can be created. First gas is always isobutylene
2. Last Ten – List of last ten gases used by the instruments
3. Gas Library – List of all gases found in RAE Systems library
4. Custom Gases – Gases with user modified parameters
Sensors
Power Source
Detection Method
PID sensor
Lithium Ion (rechargeable) or Alkaline battery pack
Internal pump (Up to 100 feet of tubing)
HazMat Field Screening Equipment Manual
1/30/2015
Page 47 of 168
California CUPA Forum
Data Logging ProRAE Guardian Software An RF modem allows for real time transmissions. A PC can be used as a base station for remote viewing.
Calibration Calibration regulator and flow controller/isobutylene calibration gas used.
(ppb manually only) Stand Alone: AutoRAE Calibration Station (MiniRAE only).
User’s Guide
ppbRAE3000/MiniRAE 3000
HazMat Field Screening Equipment Manual
1/30/2015
Page 48 of 168
California CUPA Forum
ToxiRAE Pro Single Gas Monitor
Instrument use: Small, portable single gas monitor for detection of a variety of different gases by use of field interchangeable sensors. Product range has three models all using the single interchangeable sensors. For measurement of a known chemical in potentially hazardous environments.
Models: ToxiRAE Pro, ToxiRAE II, ToxiRAE III
__________________________________________
___________________________________________
Specifications:
Measurements and Detection Ranges
Model ToxiRAE Pro
Combustible Gases 0 – 100 LEL
Oxygen (O2)
Carbon Monoxide (CO)
Hydrogen Sulfide (H2S)
Ammonia (NH3)
0 – 30% Vol
0 – 2,000 ppm
0 – 100 ppm
0 – 100 ppm
Chlorine (CL2)
Chlorine Dioxide (CLO2)
Ethylene Oxide (EtO-B)
Hydrogen (H2)
Hydrogen Chloride (HCL)
0 – 50 ppm
0 – 1 ppm
0 – 10 ppm
0 – 1,000 ppm
0 – 15 ppm
Hydrogen Cyanide (HCN)
Hydrogen Flouride (HF)
0 – 50 ppm
0 – 10 ppm
Methyl Mercaptan (CH3-SH) 0 – 10 ppm
Nitric Oxide (NO) 0 – 250 ppm
Nitrogen Dioxide (NO2)
Phosgene (COCL2)
0 – 20 ppm
0 – 1 ppm
Phosphine (PH3)
Sulfur Dioxide (SO2)
0 – 20 ppm
0 – 20 ppm
Note: Model ToxiRAE II NH3, CO, H2S and O2
Model ToxiRAE 3 CO and H2S
HazMat Field Screening Equipment Manual
1/30/2015
Page 49 of 168
California CUPA Forum
Sensors
Power Source
20 field interchangeable replaceable electrochemical sensors
Lithium Batteries (rechargeable)
Display LCD
Detection Method Diffusion (wireless range 300 feet)
Data Logging Wireless downloads to PC via built-in RF modem for Area RAE. Short range radio. Pro Guardian Software.
Calibration Manual bump and auto-calibration using manufacturer’s supplied gases.
Alone: Docking Station: AutoRAE2 Test and Calibration. Station for enhanced calibration and data retrieval.
User’s Guide http://www.enviroequipment.com/rentals/PDF/toxirae-pro-manual.pdf
ToxiRAE Pro
Start Up (Normal Operation):
1. Press and hold [mode] for 3 seconds to activate the unit. A self-test begins.
2. Sensors warm up. Visual count down will appear.
3. Unit is ready for use.
Reset alarm by pressing [Y/+]
Alarms in fresh air, perform an auto zero
↑↑↑ Indicates detection range of LEL sensor has been exceeded.
Shut down unit by pressing and holding [mode]. Continue holding until you see “Unit Off”.
HazMat Field Screening Equipment Manual
1/30/2015
Page 50 of 168
California CUPA Forum
AreaRAE Steel Multi-Gas Monitoring System
Instrument Use: Portable 1 – 5 gas multi-gas detector with a wireless radio frequency (RF) modem which allows communication and data retrieval (real-time) at remote location(s).
AreaRAE detectors can be placed at outlying locations for the establishment and monitoring of an emergency perimeter.
_________________________________________________
___________________________________________________
Specifications:
Measurements and Ranges
Oxygen (O2) 0 – 30%
Combustible 1 – 100% LEL
VOCs 0 – 200 ppm
Carbon Monoxide (CO) 0 – 500 ppm
Hydrogen Sulfide (H2S) 0 – 100 ppm
Sulfur Dioxide (SO2) 0 – 20 ppm
Nitric Oxide (NO) 0 – 250 ppm
Nitrogen Dioxide (NO2) 0 – 20 ppm
Chlorine (Cl2) 0 – 10 ppm
Ammonia (NH3)
Phosphine
0 – 50 ppm
0 – 5 ppm
Sensors
Detection Method
Up to 5 sensors with catalytic, electrochemical and PID capabilities
Internal pump with attached probe
Power Source Lithium-ion or Alkaline size C batteries
Display LCD
HazMat Field Screening Equipment Manual
1/30/2015
Page 51 of 168
California CUPA Forum
Data Logging Download to PC software: Remote Basic Host or ProRAE Suite
Package. Real time wireless transmission with built in RF modem (2 miles)
Field calibration with Manufacturer’s supplied calibration gas kit. Calibration
Website:
AreaRAE Steel Multi-Gas Monitoring System
HazMat Field Screening Equipment Manual
1/30/2015
Page 52 of 168
California CUPA Forum
INDUSTRIAL SCIENTIFIC Multi Gas Detectors
Models: IBRID MX6 Multigas Detector, Ventis MX4 Multigas Detector
(Note: Scientific iTX Gas Monitor – discontinued)
IBRID MX6 Multigas Detector
Use of Instrument: Portable handheld meter capable of simultaneously monitoring from 1 – 6 gases. Detects a variety of gases with electrochemical, catalytic PID and IR sensors. For confined space entry and other environmental environments.
_____________________________________________
________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases 0 - 1-100% LEL
Carbon Monoxide (CO) 0 – 1,500 ppm
Hydrogen Sulfide (H2S) 0 – 500 ppm
Oxygen (O2)
Hydrogen (H2)
0 – 30% Vol
0 – 2,000 ppm
Nitric Oxide (HO)
Chlorine (CL2)
0 – 1,000 ppm
0 – 100 ppm
Nitrogen Dioxide (NO2) 0 – 150 ppm
Sulfur Dioxide (SO2) 0 – 150 ppm
Hydrogen Cyanide (HN) 0 – 30 ppm
Hydrogen Chloride (HCL) 0 – 30 ppm
Ammonia (NH3) 0 – 500 ppm
HazMat Field Screening Equipment Manual
1/30/2015
Page 53 of 168
California CUPA Forum
Chlorine Dioxide (CLO2) 0 – 1 ppm
Phosphine (PH3) 0 – 10 ppm
Carbon Dioxide (CO2)
VOCs (general) PID
0 – 5 ppm
0 – 2,000 ppm
Sensors 25 interchangeable field replaceable PID, IR, catalytic bead and electrochemical sensors.
Display LCD
Power Source
Detection Method
Lithium Ion (replaceable) or Alkaline size AA batteries.
Diffusion or integrated sampling pump
Data logging
Calibration
Internal memory provides real time and historical memory for incident review.
Auto-calibration or bump test using manufacturers supplied test kit.
Stand Alone: DS2 Docking Station for enhanced automatic calibration and record keeping.
Instruction Manual
IBRID MX6 MULTIGAS DETECTOR
DS2 Docking Station
Industrial Scientific Corp (800) 338-3287
HazMat Field Screening Equipment Manual
1/30/2015
Page 54 of 168
California CUPA Forum
MX6 iBrid
Start Up (Normal mode):
1. Press and hold the center navigation button for 3 seconds to activate the unit. A self-test process begins.
2. Unit is ready for use.
A blinking red “OR” is displayed when conditions exceed the range.
Blinking black text indicates an alarm condition.
Solid red numbers indicate an alarm condition.
Bump test should be performed before each use.
Full instrument calibration should be performed monthly to ensure accuracy
Shut down unit by pressing and holding the center navigation button for more than 2 seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 55 of 168
California CUPA Forum
Ventis MX4 Multi-gas Detector
Use of Instrument: Handheld meter for the simultaneous detection of 1 – 4 gases. Used for confined space entry and in other hazardous environments and for personal monitoring.
_______________________________________________
________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases 0 – 100% LEL
Oxygen (O2) 0 – 30% Vol
Carbon Monoxide (CO) 0 – 1,000 ppm
Hydrogen Sulfide (H2S) 0 - 500 ppm
Nitrogen Dioxide (NO2) 0 – 150 ppm
Sulfur Dioxide (SO2) 0 – 150 ppm
Sensors
Detection Method
Changeable sensors (4) electrochemical/catalytic
Diffusion/ modified pump (draw up to 100 feet)
Power Source Lithium Ion (rechargeable) or Alkaline size AAA batteries
Display LCD
Data logging Meter performs continuous data logging. Date and time stamped events records and stores data.
HazMat Field Screening Equipment Manual
1/30/2015
Page 56 of 168
California CUPA Forum
Calibration Manufacturer supplied calibration kit for bump testing and calibration.
Stand Alone: DS2 Docking Station provides automatic calibration and record keeping.
Product Manual http://www.indsci.com/WorkArea/DownloadAsset.aspx?id=9399
(412) 788-4353 located in PA
Ventis MX 4
Start Up (Normal Operation):
1. Determine sampling mode. Ensure that the appropriate battery pack is installed.
2. Press and hold ON/OFF/MODE for 3 to 5 seconds to activate the unit. An internal test and initialization phase begins.
3. Select gas monitoring mode. If aspirated: determine the sampling time.
Minimum time required to draw air and read concentrations is 2 minutes plus 2 seconds for every 12 inches of tubing
4. Ensure portals are exposed to the air. No part of the monitor should be covered by any garment or part of a garment.
5. Unit is ready for use.
High frequency tone indicates pump or sensor failure
Beeping patterns indicate service needs
Shut down unit by pressing and holding ON/OFF/MODE for a full 5 seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 57 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 58 of 168
California CUPA Forum
SPERIAN BIOSYSTEMS Multi Gas Detectors
Models: Sperian PHD6 Multi-gas Detector, Sperian MultiPro Multi-gas Detector
Sperian PHD6 Multi-gas Detector
Use of Equipment: Handheld meter for the measurement of 1 – 6 gases simultaneously. For use in confined space entry and other hazmat response areas.
_________________________________________
_________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gas
Oxygen (O2)
0 – 100 % LEL
0 – 30% Vol.
Carbon Monoxide (CO) 0 – 1,000 ppm
Hydrogen Sulfide (H2S) 0 - 200 ppm
Sulfur Dioxide (SO2)
Nitric Oxide (NO)
0 – 25 ppm
0 – 350 ppm
Nitrogen Dioxide (NO2) 0 – 50 ppm
Hydrogen Cyanide (HCN) 0 – 100 ppm
Phosphine (PH3)
Chlorine (CL)
0 – 20 ppm
0 – 50 ppm
Chlorine Dioxide (CLO2) 0 – 5 ppm
Ammonia (NH3)
VOCs
0 – 100 ppm
0 – 2,000 ppm
HazMat Field Screening Equipment Manual
1/30/2015
Page 59 of 168
California CUPA Forum
Sensors 19 sensor options interchangeable into 5 sensor ports Include PID, IR, electrochemical sensors.
Power Source
Detection Method
Lithium ion (rechargeable) or Alkaline size AA batteries.
Internal motorized draw pump (option – manual sample draw kit with squeeze bulb with 10 foot tubing)
Display LCD
Alarms
Data logging
STEL/TWA; Visual/Audible
Built in IRDA port for configuration with a PC. Allows data to be downloaded in the field to retrieve information on calibration and recorded events.
Calibration Bump test/auto-calibration with manufacturer’s test kit at turn-on. Stand
Alone: Biosystems IQ6 Express Docking Station for enhanced calibration and record keeping.
Biosystems IQ6 Express Dock
Reference Manual
Sperian PHD6 Multi-gas Detector
Sperian Instruments Conn (800) 711-6776
HazMat Field Screening Equipment Manual
1/30/2015
Page 60 of 168
California CUPA Forum
SPERIAN MULTIPRO MULTI-GAS DETECTOR
Use of Instrument: Handheld meter capable of monitoring 4 gases simultaneously. Used for confined space entry and other hazmat response areas.
_____________________________________________________
_______________________________________________________
Specifications:
Measurements and Detections Ranges
Combustible Gases 0 – 100% LEL
Sensors
Detection Method
Oxygen (O2) 0 – 30% Vol.
Carbon Monoxide (CO) 0 – 1,000 ppm
Hydrogen Sulfide (H2S) 0 – 200 ppm
3 sensor capability electrochemical and catalytic bead
Diffusion or sample draw mode (hand aspirated sample draw kit or with motorized sample draw pump)
Power Source
Display
Alarms
Data logging
Lithium ion (rechargeable) or Alkaline size AA batteries
LCD
STEL/TWA; Audio/Visual
Automatic event logging stores information to include gas readings, turn on times, battery conditions and calibration data. Built-in IRDA transceiver with capabilities for transfer to a PC.
HazMat Field Screening Equipment Manual
1/30/2015
Page 61 of 168
California CUPA Forum
Calibration Manufacturer’s MultiPro Kits can be used at start-up. Stand Alone: IQ
Express Docking Station for enhanced automated bump test, calibration and record keeping.
Reference Manual
Sperian MultiPro Detector
Sperian Bio systems (800) 711-6776
HazMat Field Screening Equipment Manual
1/30/2015
Page 62 of 168
California CUPA Forum
HONEYWELL MULTI GAS MONITORS
Models: Lumidor MicroMax Pro Multi-gas Monitor
Honeywell Lumidor Impact Pro Multi-gas Monitor
BW Gas Alert5 by Honeywell
HONEYWELL LUMIDOR MICROMAX PRO
Use of Instrument: Handheld meter for simultaneously monitoring 1 – 5 gases. Used for confined space entry and other hazardous environments.
________________________________________________
________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
Carbon Monoxide (CO)
0 – 100% LEL
0 – 40% Vol.
0 – 999 ppm
Hydrogen Sulfide (H2S) 0 – 500 ppm
Ammonia (NH3)
Chlorine (CL2)
0 – 200 ppm
0 – 20 ppm
Chlorine Dioxide (CLO2) 0 – 10 ppm
Sulfur Dioxide (SO2) 0 – 20 ppm
Phosphine (PH3)
Nitrogen Dioxide (NO2)
0 – 20 ppm
0 – 50 ppm
Hydrogen Cyanide (HCN) 0 – 200 ppm
Combustible Gases (CH4) 0 – 100% Vol.
Sensors
Detection Method
25 interchangeable electrochemical sensors
Diffusion and built-in internal sampling pump (10 foot sampling hose)
HazMat Field Screening Equipment Manual
1/30/2015
Page 63 of 168
California CUPA Forum
Power Source NiMH (rechargeable) or Alkaline size AAA batteries
Data logging Built in data logging memory for recording events. Biosystems IQ
Management System software is available for data retrieval and record keeping.
Calibration Bump test and auto-calibration using manufacturer’s supplied kits. Stand
Alone: Honeywell Biosystems IQ Dock Station has fully automated bump test, calibration, meter diagnostics and data download.
User and Maintenance Manual https://www.instrumart.com/assets/MicroMAXpro_manual.pdf
(847) 955-8200 Honeywell Analytics (Illinois) Or Honeywell Analytics (800) 538-0363
HazMat Field Screening Equipment Manual
1/30/2015
Page 64 of 168
California CUPA Forum
HONEYWELL LUMIDOR IMPACT 4X
Use of Instrument: Handheld meter capable of monitoring 4 gases simultaneously. Two configurations are available - Impact (diffusion) and Impact Pro (sample pump). For use in confined space entry and other hazardous environments.
Specifications:
Measurements and Detection Ranges
Combustible gases
Oxygen (O2)
0 – 100% LEL
0 – 25% Vol.
Carbon Monoxide (CO) 0 – 500 ppm
Hydrogen Sulfide (H2S) 0 – 50 ppm
Sulfur Dioxide (S02) 0 – 20 ppm
Chlorine (CL2) 0 – 10 ppm
Chlorine Dioxide (CL02) 0 – 5 ppm
Nitrogen Dioxide (NO2) 0 – 20 ppm
Ammonia (NH3) 0 – 100 ppm
Carbon Dioxide (C02) 0.2 – 2.2% v/v
Sensors interchangeable
Detection Method Diffusion mode or built-in pump (standard on Impact Pro). Hand aspirator capabilities (option)
Power Source Alkaline dry cell batteries
Display LED
Data Logging Event information is accessed by down loading to a PC using software supplied in manufacturer’s Datalogging Kit
HazMat Field Screening Equipment Manual
1/30/2015
Page 65 of 168
California CUPA Forum
Calibration Bump test or auto-calibration using manufacturer’s target test gases.
Honeywell Enforcer Automatic Calibrator can be used to enhance calibration data retrieval.
Honeywell Enforcer Automatic Calibrator
Impact Pro Manual http://www.buygasmonitors.com/content/10722_Imp-ImpPro_DS0280_V5_09-07_EMEA.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 66 of 168
California CUPA Forum
GAS ALERT MICRO5
Use of Instrument: Handheld meter for the simultaneous measurement of 1 to 5 gases. This instrument has two models for IR and PID sensor capabilities. For use in confined spaces or other hazardous environments.
_______________________________________________
________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 – 100% LEL
0 – 30% Vol.
Carbon Monoxide (CO) 0 – 999 ppm
Hydrogen Sulfide (H2S) 0– 500 ppm
Phosphine (PH3)
Sulfur Dioxide (SO2)
0 – 5 ppm
0 – 150 ppm
Chlorine Dioxide (CL2) 0 – 50 ppm
Ammonia (NH3) 0 – 100 ppm
Nitrogen Dioxide (NO2) 0 – 99 ppm
Chlorine Dioxide (CLO2) 0 – 1ppm
Sensors
Detection Method
Power Source
12 interchangeable electrochemical, catalytic bead and PID sensors
Diffusion or attached external pump
NiMH (rechargeable) or alkaline size AA batteries
HazMat Field Screening Equipment Manual
1/30/2015
Page 67 of 168
California CUPA Forum
Display
Alarms
Data logging
Calibration
LCD
STEL/TWA; Audio/Visual/Vibrating
Continuous data logging while instrument is operating. Data is saved on a multi-media card (mmc) or secure digital (sd) card and can be downloaded onto a PC for review and record keeping.
Bump test and auto-calibration using manufacturer’s calibration kit.
Stand Alone: MicroDock II Automatic Test and Calibration Station is available for enhanced calibration and record keeping.
MicroDock II Automatic Test and Calibration Station
User Manual
Gas Alert Micro 5/PID/IR
HazMat Field Screening Equipment Manual
1/30/2015
Page 68 of 168
California CUPA Forum
ChemPro100i
Use of Instrument:
The ChemPro100i weights 2 lbs and is small enough to be used as a monitor to survey the scene after an event and as a personal detector. It uses an open loop Ion Mobility
Spectrometer technology (IMS) and orthogonal sensors. Its Ion Mobility Cell is designed to detect chemical warfare agents (CWAs) and toxic industrial compounds/Materials
(TICs/TIMs).
Specifications
Sensors Orthogonal detection using and open loop IMS cell, temperature, humidity, pressure and mass flow sensors, 6 semiconductor sensors.
Detection Libraries Chemical Warfare Agent/CWA std, First Responders
(optional), Toxic Industrial Chemical (optional), CWA precursor (optional), Volatile Organic Compounds (optional)
Battery
Operating Hours
Li-ion rechargeable and alkaline packs included
8 hours continuous use of Li-Ion, 5 hours continuous on alkaline batteries, 12 hour continuous on Li-Ion AA batteries
Battery Recharge Time 5 hours using the standard power supplies, unit will run and charge simultaneously
Power
User Interface
110-250 VAC, 50-60Hz Mains Power Supply
Graphical LCD display with blue backlight and 3 operation keys
Sampling
Calibration
1.3 LPM internal pump with continuous airflow
No calibration is required, but a simple confidence check is recommended before use
Alarm Modes Audible 85 db and visual alarm, optional earphone alarm available
Temperature -22 o
to 131 o
F
HazMat Field Screening Equipment Manual
1/30/2015
Page 69 of 168
California CUPA Forum
Dust & Water IP67: dust tight (resists blowing sand MIL-STD-810F) and protected against water immersion when inlet is closed
Shock & Vibration
Resistance Highly resistant to electromagnetic & radio interference and does not generate electromagnetic interference
Use in Potentially
Flammable
Environments
Size
Weight
Will not ignite in a n-Hexane atmosphere (Mil-STD-810
9” L x 4” W x 2” H
31 ounces with battery pack
__________________________________________________ o
F)
__________________________________________________
Website: ChemPro100i Operator’s Manual – Environics ChemPro 100 Quick Reference Link: http://www.environics.fi/us/product/chempro100i/ quick reference.pdf
Contact: Tech support Christopher Wrenn, Environics USA (610)659-4507
Don wolf & Associates, Inc., Retail representative (800)266-2046
HazMat Field Screening Equipment Manual
1/30/2015
Page 70 of 168
California CUPA Forum
Normal Operation of ChemPro100i
Start Up:
1. Startup procedures last 1-3 minutes depending on ambient temperature
2. Make sure that the ChemPro is turned on in a clean environment.
3. Open the storage cap at the top of the instrument by turning it counter-clockwise.
4. In a fresh air environment press the ON/OFF button to right, just below the display screen, to begin the start-up cycle. Builtin test checks RAM and Flash memory, IMCell channels, bias voltage and flow measurements.
Selecting menus
1. Press Menu button in main display
2. Select target menu by pressing the right button several times until the right menu appear.
3. Press the main menu button to select the appropriate menu.
Sensor Test
Sensor test should be performed and passed before every usage to assure the operation of all sensors.
1. Select the sensor test menu
2. Warm the test tube if necessary before the test. (Test tube contains a mixture of Propanol and Di-isopropyl Methyl
Phosphonate (DIMP))
3. Open the Test Tube holder cap until the hole is fully visible.
4. With the field monitoring cap attached, insert to probe into the opening.
HazMat Field Screening Equipment Manual
1/30/2015
Page 71 of 168
California CUPA Forum
5. Keep the test tube in place until the display shows the condition of the sensors. (Test passed, or Test failed)
6. Close the Test Tube
7. Exit the sensor test mode (after 30 seconds)
Trend Function
1. Press the left button in the main menu display. Enable user to switch between gas libraries)
2. Press the right button to select “Gas library”
3. Select the required “Gas library” by using the down arrow button
4. Press the Main-menu button to select or confirm the library selection.
5. Exit back to the Trend screen by pressing the left-button
Shutting Down ChemPro100i
The shutdown process uses a two button process to avoid unwanted shutdown.
1. Press the right button (shut’ down) for about three seconds until the shutdown screen appear
2. Press Menu-button (ok) to accept the shutdown.
3. Change the field monitoring cap to storage cap after shutdown.
Note:
Run the ChemPro100i for a least 30 minutes after a significant chemical exposure to keep chemical contamination from permanently damaging sensors.
HazMat Field Screening Equipment Manual
1/30/2015
Page 72 of 168
California CUPA Forum
Gas Measurement Instruments (GMI)
Models: GMI VISA 5 Gas Detector
GMI P200 Series Multi-Gas Detector
GMI PSI Single Gas Detector
VISA 5 GAS DETECTOR
Use of Instrument: Handheld meter for detection of up to 5 gases with catalytic sensors, PID and IR capabilities. For use in confined spaces and other hazardous environments. (This model is upgrade from Visa 4 Gas Detector).
_____________________________________________
_________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 – 100% LEL
0 – 25% Vol.
Carbon Monoxide (CO) 0 – 1,000 ppm
Hydrogen Sulfide (H2S) 0 – 100 ppm
Sulfur Dioxide (SO2)
Chlorine (CL2)
Ammonia (NH3)
0 – 30 ppm
0 – 10 ppm
0 – 100 ppm
Nitric Oxide (NO) 0 – 300 ppm
Nitrogen Dioxide (NO2) 0 – 20 ppm
Carbon Dioxide (CO2) 0 – 5% Vol.
Phosphine (PH3) 0 – 100 ppm
HazMat Field Screening Equipment Manual
1/30/2015
Page 73 of 168
California CUPA Forum
Sensors 14 interchangeable ‘plug-in’ catalytic, electrochemical, IR and PID sensors
Power Source
Detection Method
NIMH (rechargeable) or alkaline size AA batteries
Diffusion or electric pump (10 foot sample line)
Display LCD
Calibration
Data Logging: Built-in data retrieval which allows the storage of data; to include events and turn-off/on times.
Manufacturer’s Visa Calibration Kit can be used for bump test and autocalibration.
Visa Calibration Kit
Handbook: http://www.gmiuk.com/product/visa/
Detcon (713) 559-9290 or GMI Instruments (716) 609-3053 or GMI main office 877-421-5315
HazMat Field Screening Equipment Manual
1/30/2015
Page 74 of 168
California CUPA Forum
GMI PS200 SERIES MULTI-GAS DETECTORS
Use of Instrument: Handheld meter simultaneously measures up to 4 gases (combustible, O2,
CO and H2S). For use in confined spaces and other hazardous environments.
__________________________________________________
__________________________________________________
Specifications:
Measurements and Detections Ranges
Combustible Gases 0 – 100% LEL
Oxygen (O2) 0 – 25% Vol.
Carbon Monoxide (CO) 0 – 1,000 ppm
Hydrogen Sulfide (H2S) 0 - 100 ppm
Sensors 5 Interchangeable sensors (4) LEL, O2, H2S, CO electrochemical and
(1) dual toxic H2S/CO sensor
Methods of Detection Diffusion or sampling pump (10 foot line). Optional hand aspirator (ten feet of tubing)
Power Source Lithium Ion (rechargeable)
Alarm STEL/TWA;Audio/Visual/Vibrating
Display LED
Data Logging Capabilities for event recording and calibration tests for automatic storage and later downloading into a PC via a USB connection
HazMat Field Screening Equipment Manual
1/30/2015
Page 75 of 168
California CUPA Forum
Calibration Bump/auto-calibration testing using manufacturers supplied test gases.
Stand Alone: PS200 Auto Bump/Calibration Station enhances calibration and data storage.
PS200 Auto Bump/Calibration Station
User Handbook http://www.sedasl.es/resources/64171a_01+PS200+User+Handbook.pdf
PS200
Start Up (Normal Operation):
1. Press and hold the right hand button to warm up begins.
2. Check battery status.
3. Attach accessories if needed. activate the unit. A 30 second
4. Perform a leak test.
5. Unit is ready for use.
Display changes from green to red when an alarm condition exists.
Clear or acknowledge an alarm press the right hand button
Turn the unit off by pressing the left hand button.
HazMat Field Screening Equipment Manual
1/30/2015
Page 76 of 168
California CUPA Forum
GMI GT SERIES
Use of Instrument: Handheld meter with seven (7) application modes. Designed for use by gas technician in the gas industry. Can be used in confined space entry and hazardous environments for monitoring methane gas. Seven application modes:
1. Confined Space Mode measures up to 4 gases simultaneously (O2,LeL, CO, H2S)
2. Leak Test Mode measures methane in 3 ranges (ppm, LEL, % Vol.)
3. Barhole Mode for testing underground leaks
4. Carbon Monoxide Mode measures CO directly in ppm
5. Project Mode measuring methane 0 – 100 ppm
6. Sniffer Mode enables rapid leak detection by measuring methane 0 – 1,000 ppm
7. Pressure Mode measures from 0 – 60 inches of water column via a pressure sensor on the back of the instrument. For tightness tests on gas systems up to 2 psi.
________________________________________
____________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 – 100% LEL
0 – 25% Vol.
Carbon Monoxide (CO) 0 – 2,000 ppm
Hydrogen Sulfide 0 – 100 ppm
Sensors 7 interchangeable electrochemical, semiconductor, catalytic bead and thermal conductivity sensors
Detection Method Integral pump
HazMat Field Screening Equipment Manual
1/30/2015
Page 77 of 168
California CUPA Forum
Display LED
Power Source Alkaline size C batteries
Data Logging
Calibration
Automatic data logging
Alone GT Series Bump Test & Calibration Station Stores calibration test results
User Handbook
GMI GT SERIES
GT series
Start Up (Normal Operation):
1. Tightly screw in ‘clear bulb’.
2. Switch on unit by pressing and holding the right hand button for 1 second to initiate the warm up sequence. When warm up is complete, screen light switches off.
3. Attach accessories if needed. Unit is ready to operate in the configured option.
Power on LEDs will be illuminated (LEDS on the left bottom)
After use allow the unit to run 1 – 2 minutes in fresh air prior to switching off. Switch the unit off by pressing and holding both the left hand and right hand buttons.
Hold the buttons throughout the countdown.
HazMat Field Screening Equipment Manual
1/30/2015
Page 78 of 168
California CUPA Forum
GMI PS1 SINGLE GAS DETECTOR
Use of Instrument: Handheld single gas detection meter with interchangeable sensors to monitor O2, H2S, CO, and SO2. Used for monitoring of single known gases.
__________________________________________________
__________________________________________________
Specifications
Measurements and Detection Ranges
Oxygen (O2)
Hydrogen Sulfide
0 – 25% Vol.
0 – 100 ppm
Carbon Monoxide (CO) 0 – 500 ppm
Sulfur Dioxide (SO2) 0 – 20 ppm
Sensors Interchangeable electrochemical sensors (O2, H2S, CO, SO2)
Display LCD
Power Source Lithium heavy duty 3.6V battery
Data Logging Internal data logging for bump/calibration results
Calibration Manual testing can be performed by using manufacturer’s supplied
calibration gas kit. Stand Alone: PSI Auto Bump/Calibration Station with data record keeping capabilities.
User Handbook http://www.gmiuk.com/images/library/manuals/63183_07.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 79 of 168
California CUPA Forum
PSI Auto Bump/Calibration Station
HazMat Field Screening Equipment Manual
1/30/2015
Page 80 of 168
California CUPA Forum
DRAGER GAS MONITORS
Models: Drager x-am 7000
Drager x-am 5600
Drager x-am 2500
Drager Pac 3500/5500
DRAGER X-AM 7000
Use Instrument: Handheld meter with simultaneous and continuous detection of up to five gases simultaneously. For use in confined space and other hazardous environments.
____________________________________________________
____________________________________________________
Specifications
Measurements and Detection Ranges
Combustible Gas 0 – 100% LEL
Oxygen (O2) 0 – 25% Vol.
Carbon Monoxide (CO) 0 – 2,000 ppm
Hydrogen Sulfide (H2S) 0 – 100 ppm
Carbon Dioxide (CO2) 0 – 5% Vol.
Sulfur Dioxide (SO2) 0 – 100 ppm
Nitrogen Dioxide (NO2) 0 – 50 ppm
COCL2
Ammonia (NH3)
0 – 3 ppm
0 – 50 ppm
Hydrogen Cyanide (HCN) 0 – 50 ppm
Chlorine (CL2) 0 – 20 ppm
Hydrides
Hydrogen (H2)
0 – 20 ppm
0 – 2,000
PID Sensor 0 – 2,000
Sensors
Detection Method
Over 25 interchangeable catalytic bead, electrochemical, PID and IR sensors enables the detection of over 100 gases and vapors.
Diffusion or pump mode; built in pump (up to 150 foot draw)
HazMat Field Screening Equipment Manual
1/30/2015
Page 81 of 168
California CUPA Forum
Power Source NIMH or alkaline size C batteries
Display LCD
Data Logging Use of a data logger to record monitoring information. The data can be transferred to a PC using an infrared interface together with the
Drager GasVision software package.
Calibration Stand Alone: Drager X-am 7000 Bump Test Station with calibration adapter for test gas cylinder auto-calibration.
Drager X-am 7000 Bump Test Station
Handbook:: http://draeger-mo.com/files/x-am_7000_instructions.pdf
Contact – Drager (800) 752-8472
X-am 7000
Start Up (diffusion mode):
1. Press and hold OK key for 3 seconds. A green operating LED will illuminate and countdown sequence will begin, release OK when countdown reaches 0.
The self-test and warm up period automatically initiates. When the warm up is complete the unit is in the measuring mode and ready for use.
Acknowledge an alarm condition by pressing the OK key
↑↑↑ indicates the measuring range has been exceeded.
- - - - ambient concentration lower than minimum value.
Power down the unit by pressing and holding both the ▲ and ▼ keys for more than 3 seconds to initiate the shutdown sequence. Continue holding the keys throughout the count down.
Releasing the buttons before reaching zero will return the unit to the measuring mode
HazMat Field Screening Equipment Manual
1/30/2015
Page 82 of 168
California CUPA Forum
DRAGER X-AM 5600
Use of Instrument: Handheld meter for the monitoring of up to 6 gases simultaneously. Used for confined spaces and other hazardous environments.
_______________________________________________
_________________________________________________
Specifications:
Measurements and Detections Ranges
Combustible Gases
Oxygen (O2)
Carbon Monoxide(CO)
0 – 100 LEL
0 – 100% Vol.
0 – 2,000 ppm
Hydrogen Sulfide (H2S) 0 – 200 ppm
Carbon Dioxide (CO2)
Nitric Oxide (NO)
Nitrogen Dioxide (NO
Sulfur Dioxide (SO2)
0 – 5% Vol.
0 – 200 ppm
0 – 50 ppm
0 – 100 ppm
Phosphine (PH3) 0 – 20 ppm
Hydrogen Cyanide (HCN) 0 – 50 ppm
Ammonia (NH3)
Chlorine (CL2)
0 – 300 ppm
0 – 20 ppm
Hydrogen (H2) 0 – 2,000 ppm
Sensors
Power Source
Interchangeable IR(3) and electrochemical (24) sensors
NiMH (rechargeable) or Alkaline batteries
Detection Methods Diffusion or external pump (hoses draw up to 100 feet). Option: hand pump adapter
Display LCD
HazMat Field Screening Equipment Manual
1/30/2015
Page 83 of 168
California CUPA Forum
Data logging
Calibration
PC Software Drager CC-Vision Software record and storage of events
Manual bump testing/auto-calibration using manufacturer’s gas supply
Stand Alone: Drager Bump Test and Calibration Station
Drager Bump Test and Calibration Station
Manual: http://www.draeger.net/media/10/08/02/10080210/9046571_DraegerSensor_Handbook_en.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 84 of 168
California CUPA Forum
DRAGER X-AM 2500
(Replaces Drager am-x 2000)
Use of Instrument: Handheld meter for the simultaneous monitoring of 1 – 4 gases. For use in confined space and other hazardous environments.
_____________________________________________________
Specifications:
_____________________________________________________
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 – 100% LEL
0 – 25% Vol.
Hydrogen Sulfide (H2S) 0 – 100ppm
Sulfur Dioxide (SO2) 0 – 100 ppm
Sensors
Power Source
Nitrogen Dioxide (NO2) 0 – 50 ppm
5 interchangeable catalytic and electrochemical sensors
NiMH (rechargeable) or Alkaline batteries
Detection Methods Diffusion or external pump (100 foot draw)
Display LCD
Data Logging Data retrievable to PC with Drager CC-Gas Vision software with IR adaptor to USB cable
Calibration Stand Alone: Drager Bump Test and Calibration Station. Bump test using manufacturer’s gas cylinder mix.
HazMat Field Screening Equipment Manual
1/30/2015
Page 85 of 168
California CUPA Forum
Technical Manual: DRAGER X-AM 2500
X-am 2500
Start Up:
1. Press and hold OK key for 3 seconds.
The self-test and warm up period automatically initiates. When the warm up is complete the unit is in the measuring mode and ready for use.
2. Check that the inlet opening is not covered.
Acknowledge an alarm condition by pressing the OK key
Power down the unit by pressing and holding both the OK and + keys for more than 3 seconds to initiate the shutdown sequence. Continue holding the keys throughout the count down.
Releasing the buttons before reaching zero will return the unit to the measuring mode.
HazMat Field Screening Equipment Manual
1/30/2015
Page 86 of 168
California CUPA Forum
DRAGER PAC 3500/5500
Instrument Use: Handheld single detection meter for carbon monoxide, hydrogen sulfide or oxygen. ‘For measurement of a known chemical in potentially hazardous environments.’
_________________________________________
__________________________________________
Specifications:
Measurements and Detection Ranges
Oxygen (O2) 0 – 25% Vol.
Hydrogen Sulfide (H2S) 0 – 100 ppm
Carbon Monoxide (CO) 0 – 500 ppm
Sensors
Power Source
Detection Method
Electrochemical sensors for each gas (CO, H2S, and o2)
Lithium battery (rechargeable)
Diffusion
Display LCD
Data Logging Stores monitoring data/can be linked to a PC via a connecting cable.
Calibration Bump tests and auto-calibration Stand Alone: Drager Bump Test Station
Drager Pac 3500
HazMat Field Screening Equipment Manual
1/30/2015
Page 87 of 168
California CUPA Forum
Drager Bump Test Station Drager Pac 3500
Reference Link: http://www.draeger.net/media/10/08/02/10080210/9046571_DraegerSensor_Handbook_en.pdf
HazMat Field Screening Equipment Manual
1/30/2015
Page 88 of 168
California CUPA Forum
SCOTT BACHARACH Multi Gas Detectors
Models:
Scott Protégé Multi-gas Monitor
Scott Scout Multi-Gas Detector
SCOTT PROTÉGÉ MULTI-GAS MONITOR
Use of Instrument: Handheld 4 gas meter with the capability for simultaneous readings. Used in confined spaces and other hazardous environments. (Replaces Scott Scout Multi-Gas Meter)
_________________________________________________
_________________________________________________
Specifications:
Measurements and Detection Ranges
Combustible Gas 0 – 80% LEL
Carbon Monoxide (CO) 0 – 999 ppm
Hydrogen Sulfide (H2S) 0 – 500 ppm
Oxygen (O2) 0 – 25% Vol.
Sensors 5 electrochemical interchangeable sensor combinations for use in this 4 gas continuous display meter
Detection Methods
Power Source
Diffusion mode with optional pump and 12 inch tubing
Lithium Ion battery
Display LCD
HazMat Field Screening Equipment Manual
1/30/2015
Page 89 of 168
California CUPA Forum
Data Logging
Calibration
Data download capabilities with logging to PC
Automatic zero calibration and bump test using manufacturer’s supply kit. Stand Alone: Protégé Masterdock Automated Bump Test and
Calibration Station for enhanced calibration and data retrieval.
Protégé Masterdock Automated Bump Test and Calibration Station
Instruction Manual https://www.scottsafety.com/en/anzp/DocumentandMedia1/Protege Manual_REVL.pdf
Protege
Start Up:
1. Press and hold down either the right or left operation button until the LCD displays release. The self-test sequence initiates. When the sequence is complete the unit is in the monitoring mode and ready for use.
Blinking heart ♥ on the display indicates the device is in the monitoring mode
If the gas type label blinks with no audible, sensor is detecting a negative drift. Perform a zero calibration.
+++ indicates concentration exceeds sensor range
If an over range condition is encountered, bump test the unit. If it fails the bump test, perform a full calibration.
Power down the unit by pressing both the left and right operation buttons until the countdown and LCD shuts off.
HazMat Field Screening Equipment Manual
1/30/2015
Page 90 of 168
California CUPA Forum
SCOTT SCOUT MULTI-GAS DETECTOR
(This monitor was discontinued by Scott Safety; Effective 12/31/2011)
Use of Instrument: Multi-gas monitor can detect up to 5 gases simultaneously, with up to 9 sensors capable of measuring a variety of gases. For use in confined spaces, site remediation and other hazardous environments.
Specifications:
Measurements and Detection Ranges
Combustible Gases
Oxygen (O2)
0 – 100% LEL
0 – 30%
Carbon Monoxide (CO) 0 – 999 ppm
Hydrogen Sulfide (H2S) 0 – 500%
Sensors 9 electrochemical sensors for CO, O2, H2S and combustible LEL and a variety of chemicals PID sensors can detect HCN, NO2, and SO2.
Detection Methods Diffusion mode and internal pump/option – hand drawn aspirator.
HazMat Field Screening Equipment Manual
1/30/2015
Page 91 of 168
California CUPA Forum
Power Source Lithium Ion or Alkaline size C batteries
Data Logging
Calibration
Event log menu with events and dates transfer to PC
Automatic zero calibration/bump test to known gas concentration
Users Guide: http://www.enviroequipment.com/rentals/PDF/Scott-Scout-Manual.pdf
(800) 247-7259 (Monroe NC)
SCOUT
Start Up:
1. Press and hold any of the four function buttons.
After the initialization process, the unit goes right into monitoring mode and is ready for use
Flashing +++ indicates over range conditions
Interfering gases can potentially cause a negative reading or zero drift
Power down the unit by pressing both the lower left and right operation buttons simultaneously for 3 seconds. Press the upper left button to complete the shutdown sequence.
HazMat Field Screening Equipment Manual
1/30/2015
Page 92 of 168
California CUPA Forum
HAZARDOUS MATERIALS FIELD GAS ANALYZERS
Field Gas Analyzers
Using spectrometry for the measurement of hundreds of different compounds, the field gas analyzer can be used as a survey tool, specific gas monitor or for Chemical Warfare Agents
(CWCs). This is not a monitor to be used for confined space entry. Software allows the use of a library of compounds that can be identified in the field. Gases are drawn into a portable monitor for spectral analysis; with information then transferred to a PC for data retrieval, further analysis and record keeping. Both identification of a specific gas and concentration can be obtained.
Materials Solids and liquid Field Analyzers
Using spectrometry for the identification of unknown materials (solids/liquids). Software allows the use of a library of compounds that can be identified in the field. Solid and liquid chemical samples are placed on a portable monitor for spectral analysis; with information then transferred to a PC for data retrieval, further analysis and record keeping.
Mercury Vapor Analyzers
Using atomic spectroscopy, mercury vapors can be identified by use of portable mercury vapor analyzers. Used for environmental spills, contaminated sites and cleanup analysis. Monitors have capabilities for data retrieval, logging and storage.
Glossary of Specifications
Infrared Spectroscopy Chemicals will absorb infrared light. The absorption pattern is unique to specific chemicals which is called a spectrum. The infrared spectrum is recorded by passing an infrared beam through a sample and light energy absorbed reveals specificities at each frequency or wavelength. The spectrum of the sample can be matched to a library sample for identification and quantification of a chemical.
Ramon Spectroscopy
Ion Mobility Spectrometry
An analytical technique using laser technology.
Use analytical technique to identify ionized molecules in gas phase. Use for mercury vapor detection.
Atomic Spectroscopy Use of electromagnetic radiation absorbed and emitted by atoms
Used in mercury vapor detection
Teguchi Gas Sensors Gas detectors that measure the concentration by oxidizing the target gas at an electrode and measuring the resulting current.
Also known as an electrochemical gas sensor.
Fournier Transform Infrared A measurement technique for recording infrared spectra. (FTIR)
Spectroscopy
Toxic Industrial Chemicals NATO’s International Task Force established a list of 25 Toxic
Industrial Chemicals (TIC’s) that could cause harm in military
HazMat Field Screening Equipment Manual
1/30/2015
Page 93 of 168
California CUPA Forum
Chemical Warfare Agents
Library operations. The TICs are considered easier to use than conventional chemical and biological weapons and are more readily available.
Manufactured as weapons to be used in warfare and must be toxic enough to cause instant harm when inhaled or in contact with skin.
Single gas calibrations are established in a standard library and through the use of a PC matched for selection of a compound from the infrared spectrum.
HazMat Field Screening Equipment Manual
1/30/2015
Page 94 of 168
California CUPA Forum
THERMO MIRAN SAPPHLRE PORTABLE INFRARED AMBIENT AIR (GAS) ANALYZER
Use of Equipment:
Portable infrared ambient air analyzer different from other gas monitors with the ability to measure over 100 different compounds. Can be used as a survey tool or gas monitor by emergency responders in hazardous environments. Not a tool for confined space entry.
Specifications:
Measurement and Detection
Analysis of up to 120 compounds with results displayed in ppm, ppb, percent (%) or Absorbance
Units (AU) with detection range charts for each compound provided by the manufacturer
Infrared Spectroscopy ThermoMatch software/match with manufacturers’ library of over
120 compounds
Detection Method Pump with 4 foot tubing and attached wand.
Power Source NiCD (rechargeable) battery. Connector available for auto cigarette lighter socket, with AC and DC capabilities
Display LCD
Data Logging ThermoConnect software for downloading, analyzing and storing data
Calibration “Zero Score” automatic auto-calibration using a known substance injected into the monitor. Information saved on PC.
HazMat Field Screening Equipment Manual
1/30/2015
Page 95 of 168
California CUPA Forum
Reference Link: https://www.thermo.com/eThermo/CMA/PDFs/Various/File_27169.pdf
Contact (866) 282-0430 (MA)
SapphIRe
Start Up:
1. Connect the wand to the analyzer Press the ON/OFF key. Allow 20- 30 minutes for warm up
2. Select gas. Zero the analyzer (install chemical filter). The unit is ready for use.
3. Replace chemical filter with particulate filter, and press enter to draw sample.
4. Press and hold any of the four function buttons.
After the initialization process, the unit goes right into monitoring mode and is ready for use
Flashing +++ indicates over range conditions
Interfering gases can potentially cause a negative reading or zero drift
Power down the unit by pressing the control key. Then use the arrow keys to select the desired function and press Select or Enter.
HazMat Field Screening Equipment Manual
1/30/2015
Page 96 of 168
California CUPA Forum
GASMET DX 4040 PORTABLE FTIR GAS ANALYZER
Instrument Use: Portable FTIR gas analyzer for measuring multiple gases (both organic and inorganic) at low concentrations in real-time. For use at chemical spills and other hazardous environments. Not used for confined space entry.
__________________________________________________
______________________________________________________
Specifications:
Measurement and Detection Ranges
In standard configuration mode can measure 25 gases simultaneously. A large manufacturer supplied chemical library provided in instrument software provides capability for the measurement of numerous compounds.
Infrared Spectroscopy Fourier Transform Infrared (FTIR) Spectroscopy
Detection Method Internal pump with attached tubing (wand)
Power Source
Data Logging
Battery or AC power supply
Using Calcmet Lite Software: the analyzer is connected to a PDA with Bluetooth for analysis results.
Display LCD
Calibration Zero calibration with clean air or nitrogen
Reference Link: GASMET DX 4040 FTIR Gas Analyzer
Gasmet Technologies (866) 685-0050
HazMat Field Screening Equipment Manual
1/30/2015
Page 97 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 98 of 168
California CUPA Forum
SMITHS DETECTION HGVI HAZARDOUS GAS AND VAPOR IDENTIFIER
Instrument Use: Handheld multi-sensor for the detection of Toxic Industrial Chemicals (TIC) and Chemical Warfare Agents (CWA) used for first responders for unknowns at hazard incidents. Instrument can be operated while wearing Level ‘A’ gear.
____________________________________________
_____________________________________________
Specifications:
Measurements and Detection Ranges
Chemical Warfare Agents- nerve, blister, blood and choking agents
Toxic Industrial Chemicals – high and medium hazard materials from the International
Terror Force 25 List (ITF)
Sensors Ion Mobility Spectrometry (IMS); Photoionization; Taguchi Gas
Sensors (TGS)
Power Source Lithium Ion (rechargeable battery; ‘cigarette’ lighter power adapter
Display LCD
Alarms Audio/Visual/Vibrating
Data Logging Data retrieved using H-Fusion Enhancement Software; HGVI LINK
Wireless Communcations Kit with RF Modem on linx enables transfer of reading to remote locations.
Website: HGVI Hazardous Gas and Vapor Identifier
Federal Resources (800) 892-1099
HazMat Field Screening Equipment Manual
1/30/2015
Page 99 of 168
California CUPA Forum
SMITHS DETECTION APD 2000
Instrument Use: Hand-held monitor and detector designed for emergency first responders to hazardous chemical releases and Chemical Warfare Agents (CWA). Simultaneously detects nerve and blister agents, and recognizes pepper spray and mace. Sensors options can monitor radiation (gamma, x-ray) and hazardous compounds.
____________________________________________
______________________________________________
Specifications:
Measurements and Detection Ranges
Chemical Warfare Agents (CWA) and hazardous chemicals can be identified to include relative concentrations
ID System Ion Mobility Spectrometry (IMS)
Power Source
Alarms
Data Logging
Alkaline size C batteries
Audio/Visual
Communication cable links monitor with PC. Used with APD Datalogger
Software
User’s Manual http://www.livoniafirefighters.com/docs/APD2000 TM 2.0.pdf
Contact Smiths Detection (410) 510-9141
HazMat Field Screening Equipment Manual
1/30/2015
Page 100 of 168
California CUPA Forum
APD 2000
Start Up (Detect Mode):
1. Install the batteries.
2. Remove the protective cap and attach the nozzle
3. Press and hold power button until APD appears
4. Wait 3 minutes for self-test to complete
When display shows Ready CW or CWVX it is ready for use.
Do not touch the nozzle or filter nozzle standoff
Interfering gases can potentially cause a negative reading or zero drift
To shut down, in a clean environment press and hold the power button for 1 second. Remove the nozzle standoff and treat as hazardous waste if CW has been detected. Cover nozzle with protective cap and remove batteries from handle.
HazMat Field Screening Equipment Manual
1/30/2015
Page 101 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 102 of 168
California CUPA Forum
HAZARDOUS MATERIALS SOLIDS AND LIQUID ANALYZERS
Thermo Scientific Field Analyzers
THERMO SCIENTIFIC FIRST RESPONDER/TRUDEFENDER FIELD ANALYZERS
Instrument Use: Portable analyzers for the field identification of unknown chemicals
(solid/liquid/headspace gas), to include toxic industrial chemicals, explosives, narcotics, chemical warfare agents, powders (anthrax) and other common chemicals. Based on Raman
Spectroscopy. Not used for confined space entry.
Models: First Defender RMX (solid/liquid identification)
TruDefender FTG (headspace gas detection)
TruDefender FT (solid/liquid identification)
First Defender RMX and TruDefender FTi
____________________________________________
Specifications:
Measurements
ID System
Detection Method
Power Source
Data Logging
Measures explosives, narcotics, toxic industrial chemicals (TICs), chemical warfare agents (CWAs), powders (anthrax) and other common chemicals.
Fournier Transform Infrared Raman Spectroscopy (FTIR) with manufacturers supplied library for detection of hundreds of chemicals.
TruDefender FTG – internal pump First Responder RMX/TruDefender
FT – manual solid material or liquid sampling
Lithium Ion (rechargeable) battery/DC wall adaptor
Can be transmitted to PC for remote monitoring, data retrieval and storage
HazMat Field Screening Equipment Manual
1/30/2015
Page 103 of 168
California CUPA Forum
Calibration Poly-styrene (known sample) system readout
Notes: Thermo Scientific – Thermo Innova Model has been discontinued
Thermo Scientific – formally AHURA Scientific
Reference Link: http://www.ahurascientific.com/chemical-explosives-id
(978) 657-5555
FirstDefender
Start Up:
1. Press the power key. Enter the password
2. Position the instrument or insert vial. Press the scan key
Hold the instrument still during the scan
To power down, press and hold the power button for several seconds.
TruDefender
Start Up:
1. Press the power key.
2. Clean the tip
3. Press Go to perform a background check
4. Position the unit
5. Press the scan key
Hold the instrument still during the scan
To power down, press and hold the power button for several seconds.
HazMat Field Screening Equipment Manual
1/30/2015
Page 104 of 168
California CUPA Forum
SMITHS DETECTION FIELD ANALYZERS
SMITHS DETECTION HAZMATID RANGER
Use of Instrument: Portable analyzer used for the identification of unknown materials
(solids/liquids) including chemical warfare agents (CWAs), common chemicals, explosives toxic industrial chemicals (TICs) and narcotics. Over 30,000 liquid/solids can be identified using a manufacturer’s supplied chemical library. Not used for confined space entry.
_______________________________________________
Specifications:
Measurements
Detection of over 30,000 chemicals (solid/liquid). Explosives, common chemicals, narcotics, organic (petroleum products, pesticides, fertilizers; in-organics (mineral acids, inorganic oxides, nitrates, chlorates and phosphateS); and unknown powders (anthrax).
ID System Fournier Transform Infrared – FT-IR/Raman Spectroscopy with manufacturer’s chemical library for detection of over 30,000 chemicals
Display LCD
Power Source
Data Logging
Lithium Ion (rechargeable) or AC Power
Spectra can be transmitted to wireless to a PC for remote monitoring and retrieval/storage of information
Calibration Effective cleaning of Diamond Internal Reflection Element is autodisplayed on spectrum readout.
Reference Link: http://www.smithsdetection.com/index.php/products-solutions/chemical-identification.html
HazMat Field Screening Equipment Manual
1/30/2015
Page 105 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 106 of 168
California CUPA Forum
MERCURY VAPOR ANALYZERS
Jerome Mercury Vapor Analyzers (Arizona Instruments)
JEROME 411/431 MERCURY VAPOR ANALYZERS
Use of Instrument: Portable meters (Models 411 and 413) designed to detect mercury vapors in hazardous areas for the location and concentrations of mercury from spills and other environmental contamination.
Model 411
Specifications:
Measurements and Detection Ranges
Model 431
Mercury (Hg) vapors Ranges:
Model 411 0 – 1.999 mg/m3Hg
Model 431 0.001 – 0.999 mg/m3Hg
Sensor A thin gold film in the meter will undergo an increase in electrical resistance proportional to the mass of mercury in a sample.
Power Source NiCd rechargeable batteries or AC power generator
Display LCD
Detection Method
Data Logging
Internal intake pump with probe attachment
Calibration
Model 431 Jerome Communications Interface Software for retrieval and storage of data on a PC
The gold film sensor does not require frequent calibration and manufacturer’s recommendation is annual factory servicing. The manufacturer’s Functional Test Kit AZIP/N-0902 is used in between
HazMat Field Screening Equipment Manual
1/30/2015
Page 107 of 168
California CUPA Forum
annual servicing to determine if the instrument is within calibration tolerances.
Manual Model 411: JEROME MERCURY VAPOR ANALYZER
411
Start Up (Detect Mode):
1. Ensure the zero air filter is installed in the instruments intake. Press the power button.
2. Attach line cord and plug into an outlet
3. Perform a film heat (15 minuet cycle)
4. Wait 30 minutes before continuing to use the instrument
5. Press sensor status if within the range, the unit is ready for use.
411 is intended for vapor use only
Sensor Status must be between 01 and 99 to operate, occasionally check status.
Sampling period is a 10 second cycle
Once initiated, do not interrupt the film heat cycle.
A probe may be used to locate mercury in hard to reach places. Plug probe directly into intake.
To shut down, press the power off button. Plug the zero air filter into the intake during storage.
Manual Model 431: JEROME MERCURY VAPOR ANALYZER
AZI Instruments (800) 528-7411
HazMat Field Screening Equipment Manual
1/30/2015
Page 108 of 168
California CUPA Forum
431-X
Start Up (Sample Mode):
1. Ensure the zero air filter is installed in the instruments intake.
2. Attach line cord and plug into an outlet.
3. Press the power button.
4. Sample continuously (12 sec cycle) until readings stabilize.
5. Perform sensor regeneration (10 minuet cycle). 0.0.0 is displayed when complete.
6. Remove power cord, unit is ready for use after zeroing.
7. Wait 30 minutes before zeroing the instrument for maximum accuracy. Unit can be rezeroed immediately after sensor regeneration if immediate use is needed.
Press and hold sample button to enter survey mode. Take several samples for approximately 1 minuet to zero. (Instrument will sample for 12 seconds then go into 3 second sampling cycles). Release sample button to exit survey mode.
431-x is intended for vapor use only. Do not allow probe or instruments intake to be exposed to any liquid, dust, or other foreign matter
8.8.8. indicates sensor is saturated
Approximately 65 samples at .1mg/m
3
may be taken before regeneration is required.
Use survey mode to locate mercury spills or assess areas of potentially high mercury concentrations
Once initiated, do not interrupt the sensor regeneration cycle. AC power is required to thermally regenerate the sensor. When in storage, at a minimum, regenerate every 30 days.
A probe may be used to locate mercury in hard to reach places. Plug probe directly into intake.
To shut down, press the power off button. Perform the sensor regeneration prior to storing the instrument and ensure the zero air filter remains plugged in during storage
HazMat Field Screening Equipment Manual
1/30/2015
Page 109 of 168
California CUPA Forum
JEROME J405 MERCURY VAPOR ANALYZER
Use of Instrument: Portable mercury vapor analyzer for low level mercury detection for spills and other environmental contamination and cleanup analysis. The J405 in the Jerome 400 series models (preceded by the 411/413) and has lower detection capabilities.
__________________________________________________
_________________________________________________
J405 Mercury Vapor Analyzer
Specifications:
Measurement and Detection Ranges
Sensor
Mercury (Hg) Vapors Range: 0.5 – 999 ug/m3Hg (0.0001 – 0.999 mg/m3Hg)
A thin gold film in the meter will undergo an increase in electrical resistance proportional to the mass of mercury in a sample.
Power Source NiMH rechargeable batteries or 24V DC power adapter/AC power supply/charger
Detection Method Internal intake pump with probe attachment
Display LCD
Data logging
Calibration
USB communications to a PC for data retrieval, storage and reporting
The gold film sensor does not require frequent calibration and manufacturer’s recommendation is annual factory servicing. The manufacturer’s Functional Test Kit AZIP/N-0902 is used between annual servicing to determine if the instrument is within calibration tolerances.
Manual: JEROME J405 MERCURY VAPOR ANALYZER
HazMat Field Screening Equipment Manual
1/30/2015
Page 110 of 168
California CUPA Forum
J405
Start Up:
1. Ensure the zero air filter is installed in the instruments intake. Attach power cord and plug into an outlet. Press the I/O button.
2. Initiate automatic 5 min warm-up by pressing warm-up from Regen Menu
3. Perform sensor regeneration (45minuet cycle).
4. Remove power cord, and power on for 5 minutes prior to sampling
5. Sample every 15 seconds with the zero air filter attached until readings are zero.
6. Remove zero air filter, unit is ready for use. Wait 15 seconds between samples to allow sensor to stabilize. Sampling cycle is 16 seconds.
J405 is intended for vapor use only. Do not allow probe or instruments intake to be exposed to any liquid, dust, or other foreign matter
Approximately 65 samples at .1mg/m
3
may be taken before regeneration is required.
Use survey mode to locate mercury spills or assess areas of potentially high mercury concentrations
Once initiated, do not interrupt the sensor regeneration cycle. AC power is required to thermally regenerate the sensor. When in storage, at a minimum, regenerate every 30 days.
To shut down, press the power off button. Perform the sensor regeneration prior to storing the instrument and ensure the zero air filter remains plugged in during storage.
HazMat Field Screening Equipment Manual
1/30/2015
Page 111 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 112 of 168
California CUPA Forum
Mercury Instruments
Mercury Tracker 3000 IP Portable Mercury Vapor Analyzer
Use of Instrument: Detection of mercury vapors for environmental monitoring of spills, investigation of contaminated sites and cleanup analysis.
_______________________________________________
Mercury Tracker 3000 IP Portable Mercury Vapor Analyzer
_________________________________________________
Specifications:
Measurement and Detection Ranges
Mercury (Hg) vapors
Ranges: Three (3) ranges
0 – 100 ug/m3Hg
0 – 1000 ug/m3Hg
0 – 2000 ug/m3Hg
Sensor
Power Source
Cold vapor atomic spectrometry (UV absorption)
NiMH rechargeable batteries/external DC ‘cigarette lighter’ adapter or AC power 240/110V (50/60 HZ) with power supply connection capabilities
Display LCD
Alarms
Data Logging
Calibration
3 levels programmable (cell soiled, battery state, UV status)
Data logger system integrated with data transfer to PC.
Factory calibration – long lasting. For quality control a test screen is used to check calibration
Website: Mercury Tracker 3000 IP Portable Mercury Vapor Analyzer for Air and Other Gases.
Reference Link: http://www.mercuryinstrumentsusa.com/Brochures/MI Tracker 3000-IP
Brochure.pdf
Contact: (303) 972-3740
HazMat Field Screening Equipment Manual
1/30/2015
Page 113 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 114 of 168
California CUPA Forum
LUMEX MERCURY VAPOR ANALYZERS
LUMEX RA 915 MERCURY VAPOR ANALYZER
Instrument Use: Handheld meter for the detection of mercury vapor in the environment from spills, contaminated sites or cleanup analysis.
__________________________________________________
Specifications:
Measurements and Detections
Sensor
Detection Method
Power Source
Data Logging
Calibration
Mercury (Hg) vapors
Range:
0.1 – 200 ug/m3Hg
Zeeman atomic absorption spectrometry using high frequency modulation of light polarization
Internal pump with attached probe
Rechargeable batteries/12 V DC and 240/110V AC (50/60 HZ)
Data logging integrated into meter with USB transfer to PC
Serviceability test done prior to and after each sampling period to verify proper functions using manufacturer’s quality assurance procedures.
HazMat Field Screening Equipment Manual
1/30/2015
Page 115 of 168
California CUPA Forum
Manual: http://www.cihequipment.com/sites/default/files/manual_RA_915.pdf
RA 915+
Operating:
1. Select the operating mode. Select the power mode.
2. Set the handle of test cell into off position and optical bridge handle to position III.
3. Switch the analyzer on.
4. Press the Ent button.
5. Press and hold the lamp ignition button. The * sign will go out when the lamp turns on.
6. Let the analyzer warm up at least 20 minutes.
7. Device is ready for measurements.
HazMat Field Screening Equipment Manual
1/30/2015
Page 116 of 168
California CUPA Forum
GENESIS LABORATORY SYSTEMS MERCURY VAPOR ANALYZERS
HG 253 PORTABLE MERCURY VAPOR ANALYZER
Instrument Use: Portable monitor used for the detection of mercury vapor in industry and the environment.
_______________________________________________
________________________________________________
Specifications:
Measurement and Detections Ranges
Mercury (Hg) vapors
Ranges:
0.001 mg/m3Hg – 10.000 mg/m3Hg
Sensor
Power Source
Detection Method
Alarm
Calibration
Atomic absorption spectrometry
Lithium-Ion rechargeable batteries
Internal pump with attached probe
Alarm capabilities (programmable)
Annual factory calibration Monitoring for status of batteries, detectors and calibration.
HazMat Field Screening Equipment Manual
1/30/2015
Page 117 of 168
California CUPA Forum
User Manual: http://www.genlabsystems.com/mercury/hg253man.pdf
Genesis Laboratory Systems (970) 241-0889 or (888) 270-0465
Hg 253
Operating:
1. Power on the unit.
2. To operate simply press the touch screen where a visual interface button appears.
3. Press the sample button to enter the sample mode.
A sample will take about 3 minutes.
Press cancel to abort a sample and return to the main screen.
HazMat Field Screening Equipment Manual
1/30/2015
Page 118 of 168
California CUPA Forum
COLORIMETRIC GAS INDICATORS
Detector tubes can be used for spot measurements of over 200 organic and inorganic gases and vapors in the air. These tubes are most useful for screening purposes to determine, if warranted a need for further analysis. Sealed glass tubes are filled with a granular reagent that is coated with an appropriate indicator chemical that will react with a specific gas or vapor to provide a color change. A pump is used to draw a known volume of gas inside the tube to measure the concentration of the substance. Pumps can be hand or battery operated. A number of commercially manufactured hand pump colorimetric gas indicators are available; in addition, a battery pumped Dragger Chip Measurement System (CMS) uses a colorimetric 10 capillary chip for multiple gas detection.
COLORIMETRIC GAS INDICATOR TUBES
Models: Drager
Ion Science – Uniphos ASP-40
Drager Accuro (bellows pump)
MSA RAE LP – 1200
MSA Kwik Draw Deluxe (bellows pump)
Sensidyne
Instrument Use:
Basic use and detection are similar for each model. Colorimetric indicating tubes are used to monitor a variety of chemicals. The indicating tube contains a reagent that responds to specific chemicals in the environment resulting in a color change in the tube. The length of the stain is proportional to the concentration of the chemical in the atmosphere.
Sensors (tubes) Glass tubes with specific reagents for colorimetric evaluation.
Colorimetric tubes are available for over 200 inorganic or organic chemicals. The accuracy of a detector tube, established by the National
Institute of Occupational Safety & Health (NIOSH) and the American
Industrial Hygiene Association (AIHA) is +/- 25% of a reading.
Detection Method The glass ends of a detection tube are broken and inserted into a pump.
Pump strokes are performed per the manufacturer’s information type use of instrument. A color change in the reagent is analyzed by either the length of stain generated inside the tube or the degree of color. The color change is compared to a scale printed on the tube or to a manufacturer’s reference chart to determine concentration.
Calibration Detection tubes are calibrated prior to leaving the factory. Field calibration is not necessary.
HazMat Field Screening Equipment Manual
1/30/2015
Page 119 of 168
California CUPA Forum
__________________________________________
Drager Gas Detector Tube Pump
_________________________________________________
Drager Accuro (bellows pump)
__________________________________________
Draeger Safety
(800) 482-5536
Reference Link: http://www.draegersafetyonline.com/index.php
HazMat Field Screening Equipment Manual
1/30/2015
Page 120 of 168
California CUPA Forum
____________________________________________________
Ion Science – Uniphos ASP-40
_________________________________________
Ion Science Uniphos precision Air Sampling Pump Model ASP-40 http://www.ionscienceusa.com/wp-content/uploads/2013/11/ASP-40-V1.pdf
Contact: (877) 864-7710
ASP 40
Operating:
1. Insert a new detector tube without breaking its ends into the rubber tube connector.
2. Align red dots and pull the pump handle a full stroke.
3. Wait 3 minutes, and release the pump handle by turning 90 degrees left or right.
4. Ensure piston returns to its original position.
5. Cut both ends of the detector tube using the tip cutter.
6. Insert into the rubber tube connector. Ensure the arrow on tube points toward the pump.
7. Align the red dots.
8. Pull pump handle to 50 or 100 cc. Complete the desired number of strokes keeping the tube in the sample source.
Keep the detector tube in the gas source for 1 – 3 minutes.
Empty broken tips by pulling off rubber cap opposite the tip cutter.
HazMat Field Screening Equipment Manual
1/30/2015
Page 121 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 122 of 168
California CUPA Forum
________________________________________________
MSA RAE LP - 1200
________________________________________________
RAE Colorimetric Gas Detection Gas Detection Tubes and Pump http://www.raesystems.com/products
Contact: (408) 952-8200
__________________________________________
MSA Kwik-Draw Pump
____________________________________________
Manual: MSA Kwik-Draw Pump
Contact:
HazMat Field Screening Equipment Manual
1/30/2015
Page 123 of 168
California CUPA Forum
________________________________________________
Sensidine/Kitagawa Detector Tubes
________________________________________________
Website: www.sensidyne.com http://www.sensidyne.com/colorimetric-gas-detector-tubes/detector-tubes/
Contact: (800) 451-9444
HazMat Field Screening Equipment Manual
1/30/2015
Page 124 of 168
California CUPA Forum
DRAEGER CHIP MEASUREMENT SYSTEMS (CMS)
Instrument Use: Handheld meter for spot gas measuring of types and concentrations of chemicals by use of a replaceable chip containing colorimetric tubes. A chip containing up to 10 glass tubes is inserted into the meter. Once the chemical type and concentration is read on the meter LCD display screen, the chip is automatically ejected from the instrument and the analyzer shuts down.
____________________________________________
Specifications:
Measurements
Sensors (tubes)
___________________________________________________
Glass tubes with reagents for over fifty gases and vapors.
Draeger CMS Chip with 10 capillary glass tubes containing reagents with substance-specific granules.
Power Source
Data Logging
Alkaline size AA batteries
Integrated data recorder will store events which may be recalled and retrieved upon demand.
Calibration Chips are calibrated before leaving the factory. No field calibration checks are necessary. After an automatic system self-test, the meter is ready for operation.
Manual: Draeger CMS Operator Guide
Website: Draeger Chip Measurement System
Contact: (412) 787-8383
HazMat Field Screening Equipment Manual
1/30/2015
Page 125 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 126 of 168
California CUPA Forum
RADIATION DETECTION METERS
Radiation detection meters, both survey and dosimeters are used to locate and quantify sources of ionizing radiation. Survey meters are used for radiation detection to check personnel, equipment and facilities for radioactive contamination; or to evaluate radiation fields
(contaminated environments) for direct exposure hazards. Radiation dosimeters are used to measure the dose of individual exposure to radiation energy. Meters are designed to measure
Alpha/Beta/Gamma and X-Ray radiation.
Glossary
Ionizing Radiation Ionizing radiation alters the structure and, by association the electric charge of individual atoms by a process known as ionization. The substances that produce ionizing radiation are said to be radioactive. There are four types of ionizing radiation –
Alpha, Beta, Gamma and Neutron. X-Rays are also electromagnetic radiation as light and radio waves, but are also ionizing.
Non-Ionizing Radiation Electromagnetic radiation that does not carry enough energy per quantum to ionize atoms or molecules Non-ionizing radiation is located at the low end of the electromagnetic spectrum and sources include power lines, microwaves, radio waves, visible light and lasers. Radiation survey meters are not used for the detection of non-ionizing radiation.
Survey Meter Portable radiation detectors and measurement instruments Used to check personnel, equipment and facilities for radioactive contamination. Radiation survey meters are used to locate and quantify sources of ionizing radiation or to quantify the exposure rare from sources of ionizing radiation. To assess the quantity of radioactive materials present, survey meters for Alpha and Beta are typically calibrated to measure counts per minute (cpm). To measure the exposure rate from gamma or X-Ray sources, survey meters are calibrated to measure roentgens per hour (R/h).
Commonly used hand-held survey meters:
Scintillation Counter – Used in measurement of Alpha, Beta and
Neutron particles.
Geiger Counter - Commonly used for measurement of Alpha, Beta and Gamma levels.
Ion Chamber – Used for Beta, Gamma and X-Ray measurements.
Dosimeter Used to measure and record individual and exposure to radiation energy.
HazMat Field Screening Equipment Manual
1/30/2015
Page 127 of 168
California CUPA Forum
Alpha Particles With 2 protons and no electrons positively charged. Paper, clothing or a few inches of air space can effectively shield Alpha particles. Particles can be hazardous if ingested or inhaled.
Generally read in cpm.
Beta Particles Electrons have a negative charge and may travel 6 – 10 feet through the air. Heavy clothing, thick cardboard or one-inch thick wood will provide protection from Beta particles. Generally read in cpm.
Gamma Particles These particles have no charge, with a short wavelength and high energy. Gamma is the most penetrating form of the
Alpha/Beta/Gamma radiation particles and can travel in the air up to 1,500 feet. Lead or concrete barriers are necessary to shield from Gamma particles. Generally read in R/h.
X-Ray X-Rays are electromagnetic radiation of the same nature as light and radio waves, but they are also ionizing. Similar characteristics to Gamma particles. Generally read in R/h.
Neutron Radiation Neutron radiation is a form of ionizing radiation most often found in nuclear reactors and nuclear bombs. Neutrons are found in all elemental nuclei except for hydrogen. Neutron radiation is considered the fourth major type radiation after Alpha/Beta particles and Gamma rays.
Ionizing radiation consists of free neutrons through splitting or nuclear fusion. It consists of the use of free neutrons from atoms and these free neutrons react with nuclei of other atoms to form new isotopes creating radiation (example - radioactive isotope uranium/plutonium).
Neutron radiation is not directly measured by use of survey meters, but a Gamma component in association with neutron particles can be detected and measured.
Radioactive Material Measurements:
Curie (Ci) Measures amount of ionizing radiation released by a material
(point of origin). Expressed in terms of “radioactivity” which represents how many atoms in the material decay in a given time period
Counts Per Minute (cpm) Measures quantity of radioactive materials present (monitoring location). Generally associated with Alpha and Beta particles.
Roetgen Measures exposure or the amount of radiation. Radiation meters often measure exposure in roentgens per hour (R/h).
RAD Measures the amount of radiation absorbed by an object or person.
HazMat Field Screening Equipment Manual
1/30/2015
Page 128 of 168
California CUPA Forum
REM/Sievert (Sr) Dose equivalent (effective dose) combining the amount of radiation absorbed and the medical effects of the radiation type.
Exposure to different types of radiation expressed as RAD do not produce equal biological effects. For example, one RAD of
Alpha will have a greater effect than one RAD of Beta.
Thus radiation effects are expressed in radiation as an effective dose or Roetgen Equivalent Man (REM). REM is also measured in Sieverts (Sv).
100 REM = 1 Sv
HazMat Field Screening Equipment Manual
1/30/2015
Page 129 of 168
California CUPA Forum
LUDLUM RADIATION DETECTORS
Survey Meters
LUDLUM RADIATION METER MODEL 2241-3RK
Equipment Use: This meter is a pancake for detection of beta and gamma radiation. Used for personnel testing, objects and environmental detection.
____________________________________________
______________________________________________
Detection – Beta and gamma radiation
Website: Ludlum Model 2241 3RK Response Kit July 3013
Website: Product Manuals – Ludlum Measurements, Inc http://www.ludlums.com/multisites/medphys/images/stories/product_manuals/M2241-3RK.pdf
Contact: 888-62401
HazMat Field Screening Equipment Manual
1/30/2015
Page 130 of 168
California CUPA Forum
Ludlum 2241
Operating:
1. Connect a detector by firmly pushing the connectors together while twisting clockwise until the connector latches.
2. Place SCA/RATE switch in the Rate position.
3. Check for background readings.
4. Switch the AUD ON/OFF switch to the on position.
When surveying for alpha particles ensure the F/S switch is in the F position.
Instrument response time varies from 2 to 50 seconds.
To survey for beta particle use no shielding.
The tip is the most sensitive. Point the detector towards the radiation being measured.
HazMat Field Screening Equipment Manual
1/30/2015
Page 131 of 168
California CUPA Forum
LUDLUM GM PANCAKE MODEL 44-9
Equipment Use: pancake probe for radiation detection of personnel, objects and hazardous environments.
____________________________________________________
Ludlum GM Pancake Model 44-9
_____________________________________________________
Detection – Alpha, beta, gamma radiation
HazMat Field Screening Equipment Manual
1/30/2015
Page 132 of 168
California CUPA Forum
Dosimeters
LUDLUM MODEL AT-138 POCKET DOSIMETER
Equipment Use: Direct read pocket dosimeter for documentation of exposure to workers and visitors in restricted areas. Has an eye-piece lens.
__________________________________________________
Ludlum Model AT-138 Pocket Dosimeter
________________________________________________
Detection: Gamma and x-ray radiation
HazMat Field Screening Equipment Manual
1/30/2015
Page 133 of 168
California CUPA Forum
LUDLUM MODEL 25 PERSONAL RADIATION DETECTION METER
Equipment Use: Pocket sized handheld meter for radiation detection in hazardous environments.
Ludlum Model 25 Personal Radiation Detection Meter
________________________________________________
Detection: Gamma and X-Ray radiation
HazMat Field Screening Equipment Manual
1/30/2015
Page 134 of 168
California CUPA Forum
FLUKE RADIATION DETECTION METERS
Survey Meters
FLUKE ASM 990 ADVANCED SURVEY METER
Equipment Use: Portable handheld meter with probe which can be used by emergency responders at HazMat incidents. (Model 993 has built-in pancake probe)
______________________________________________
Fluke ASM 990 Advanced Survey Meter
_______________________________________________
Detection: Alpha, beta, gamma (above 60 KeV) and x-ray radiation
Website: Fluke Biomedical ASM 990 Series Operators Manual http://fieldenvironmental.com/assets/files/Manuals/Victoreen 990 ASM Manual.pdf
Contact: (800) 850-4608
ASM 990
Actuating:
1. Press the ON/OFF key to power on the unit.
2. Select the probe to be used.
3. Unit is ready for normal operation.
To turn off, press the ON/OFF button.
HazMat Field Screening Equipment Manual
1/30/2015
Page 135 of 168
California CUPA Forum
Dosimeters
FLUKE DIRECT READING POCKET DOSIMETER
Equipment Use: Direct read pocket dosimeter for measuring accumulated gamma and x-ray radiation. The dosimeter is read through an eye-piece.
________________________________________________
_________________________________________________
Detection – Gamma and x-ray radiation
Website: Fluke Biomedical Nuclear Direct Reading Pocket Dosimeters
Operating Manual: FLUKE DIRECT READING POCKET DOSIMETER
HazMat Field Screening Equipment Manual
1/30/2015
Page 136 of 168
California CUPA Forum
CIVIL DEFENSE VICTOREEN (CDV) RADIATION METERS
CDV 700 RADIATION DETECTION METER
Equipment Use: Radiation survey meter for low-range measurement of gamma and detection of beta radiation. Used for the monitoring of personnel, food and water in shielded areas with low radiation backgrounds.
_______________________________________________
CDV 700 Radiation Detection Meter
Detection: 0 – 50 mr/hr/hr for Gamma dose and detects presence of beta radiation. (When the probe on the instrument is closed, beta is stopped and only gamma is detected and measured.
When the shield is opened, both beta and gamma can be detected.
Website: CDV -700 Gamma Survey Meter Page http://www.civildefensemuseum.com/cdmuseum2/radkits/cdv700.html
Contact: (619) 708-8114
HazMat Field Screening Equipment Manual
1/30/2015
Page 137 of 168
California CUPA Forum
CDV 715 RADIATION DETECTION METER
Equipment Use: This meter detects gamma radiation by design for use only for the period immediately following a nuclear weapon attack.
________________________________________
CDV 715 Radiation Detection Meter
_________________________________________
Detection – Gamma only
Instruction and Maintenance Manual http://www.orau.org/ptp/Library/cdv/victor715-1a.pdf
CD V 715
Start Up:
1. Turn the instrument on by switching the range switch from “Off” to “Zero” .
2. Allow 1 minute to warm up.
3. Adjust the needle to zero.
4. Perform a circuit check.
5. Select the appropriate range.
Meter reading observed must be multiplied by the factor indicated by the position of the range switch. To power down, switch the range switch to the off position.
HazMat Field Screening Equipment Manual
1/30/2015
Page 138 of 168
California CUPA Forum
Victoreen 450-B Ion Chamber Survey Meter
Equipment Use:
This handheld meter monitors for low level alpha, beta and gamma radiation.
Victoreen 450-B Ion Chamber Survey Meter
________________________________________
Detection: Alpha above r MeV, beta above 100keV, gamma, x-ray radiation. The model responds to but is not calibrated for alpha, beta radiation with the slide open.
Reference link: Victoreen 450-B Ion Chamber Survey Meter
Victoreen 450
Actuating:
1. Press the ON/OFF button. A self-test will initiate.
2. If the unit passes the self-test it will go into the normal operation mode.
To turn off, press the ON/OFF button.
HazMat Field Screening Equipment Manual
1/30/2015
Page 139 of 168
California CUPA Forum
MIRION RADIATION DETECTION METERS
Survey Meters
Mirion DX-1 Radiation Monitor
Equipment Use: Handheld pocket radiation detector.
___________________________________________
Mirion DX-1 Radiation Monitor
____________________________________________
Detection: Beta, Gamma and X-Ray radiation.
Website: Mirion DX-1 Instruction Manual
Reference link: http://www.filterwater.com/docs/sensafe/dx1-dx2_instruction_manual.pdf
Contact: [email protected] (770) 432-2744 X156
DX-1
Actuating:
1. Hold the unit in your right hand with your thumb over the push button.
2. Press and hold the button for operation.
LED above the push button indicates the unit is on.
To power down, release the push button.
HazMat Field Screening Equipment Manual
1/30/2015
Page 140 of 168
California CUPA Forum
Mirion RDS 31 Survey Meter
Equipment Use: Handheld dose rate radiation detector for survey in field applications and monitoring of field personnel.
______________________________________________
Mirion RDS 31 Survey Meter
___________________________________________
Detection: Alpha, Beta, Gamma, X-Ray
Website: RDS-31 | Modular Radiation Survey Meter
Reference Link: Mirion RDS 31 Survey Meter
HazMat Field Screening Equipment Manual
1/30/2015
Page 141 of 168
California CUPA Forum
Dosimeters
Mirion DMC 3000 Personal Electronic Dosimeter
Equipment Use: Personal dosimeter for gamma and x-ray radiation.
_______________________________________________
Mirion DMC 3000 Personal Electronic Dosimeter
____________________________________________
Website: Mirion DMC 3000 | Electronic Radiation Dosimeter
DMC 3000 Quick User’s Guide: DMC 3000 Electronic Dosimeter
Reference Link: https://www.mirion.com/products/dmc-3000-electronic-radiation-dosimeter/
Detection: Gamma and x-ray radiation
HazMat Field Screening Equipment Manual
1/30/2015
Page 142 of 168
California CUPA Forum
Survey Meter
W.B. JOHNSON INSTRUMENTS RADIATION DETECTION METER
W.B. Johnson Instruments DSM-500 Digital Radiation Survey Meter
Equipment Use: Handheld meter for the measurement and monitoring of radiation in contaminated environments.
___________________________________________________
_________________________________________________________
Detection: Alpha, Beta, Gamma
Website: B Johnson Instruments DSM Digital Radiation Survey Meter
Reference Link: http:www.jradmeters.com/device.php?device=2
Contact: W.B. Johnson Instruments 208-557-6945 or 208-557-6945 Idaho
HazMat Field Screening Equipment Manual
1/30/2015
Page 143 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 144 of 168
California CUPA Forum
POLIMASTER RADIATION DETECTION METERS
Survey Meter
Model PM 1405 Survey Radiation Monitoring Meter
Equipment Use: This is a survey radiation meter for detection of gamma and x-ray radiation sources. A special screen-filter allows choosing beta radiation flux density measurement mode.
_____________________________________________________
Model PM 1405 Survey Radiation Monitoring Meter
________________________________________________________
Detection: Beta, Gamma and X-Ray radiation (opening special screen filter and choosing beta radiation flux density measurement mode – can detect beta nuclear radiation sources)
Website: Polimaster Survey Meter 1405
Reference Link: Model PM 1405 Survey Radiation Monitoring Meter
Contact: (703) 525-5075
PM 1405
Actuating:
1. Hold the lower button till the backlight comes on
2. After an internal test and calibration the instrument is ready for use.
HazMat Field Screening Equipment Manual
1/30/2015
Page 145 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 146 of 168
California CUPA Forum
RADEX RADIATION DETECTION METER
RADEX RD 1706 Radiation Detector
Equipment Use: Handheld Geiger counter for detection of beta, gamma and x-ray radiation.
________________________________________________
_____________________________________________
Detection: Beta, Gamma and X-Ray radiation
Website: Radiation Monitor Radex RDI 1706 – Quarta
Reference Link: http://www.quarta-rad.ru/en/manuals/RD1706.pdf
Contact: (201) 877-2002
RD 1706
Actuating:
1. Switch on device by pressing the off button. Examination of radioactive environments starts.
2. After 13 seconds, the results of a short cycle are displayed. First reliable results are displayed after 26 seconds.
To find a source of radiation, turn off the threshold signal. Switch on audio signal and pay attention to sound frequency.
Do not use in areas of high humidity.
Keep device away from which generate strong magnetic fields.
Extract batteries if the device is not being used for a long time, keep unit in cool dry conditions.
Check working capacity of unit before using after being in storage for long periods of time.
To turn off, press off button and hold till disappearance of messages.
HazMat Field Screening Equipment Manual
1/30/2015
Page 147 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 148 of 168
California CUPA Forum
MAZUR INSTRUMENTS RADIATION DETECTOR
Survey Meter
PRM-9000 Geiger Counter – Nuclear Radiation Detector and Monitor
Equipment Use: Handheld pancake Geiger detector tube for surveying radiation levels.
_________________________________________________________
PRM-9000 Geiger Counter – Nuclear Radiation Detector and Monitor
_____________________________________________________
Detection: Alpha, Beta, Gamma and X-Ray radiation
Website: Mazur Instruments PRM 8000/9000 Nuclear Radiation Monitor User’s Guide
Reference Link: http://www.mazurinstruments.com/PRM-8000-9000_UG_R2_EN.pdf
Contact:
PRM 8000/9000
Actuating:
1. Press and hold the power button. After 3 seconds the unit will default to the home screen and begin displaying measurements.
Changes in the rate of beeps will be the first indication of a radiation object or area.
If display is off, press any key to wake unit up (the 8000/9000 is designed to be always on. Internal battery life in sleep mode can be up to 60,000 hours).
The unit can be powered down for long time storage by pressing and holding the power key.
HazMat Field Screening Equipment Manual
1/30/2015
Page 149 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 150 of 168
California CUPA Forum
RADIATION ALERT RADIATION DETECTION METERS
Survey Meters
Radiation Alert Radiation Monitors
Models: Digital - Inspector USB; Inspector EXP (pancake external probe); Analog – Monitor
4EC
Equipment Use: Handheld radiation meters
Detection: Alpha, Beta, Gamma
Radiation Alert Reference Link: http://seintl.com/support/manuals/
Inspector USB
Inspector
Actuating:
1. Set the mode switch to the desired mode.
2. Set the bottom switch to On or Audio. After a short system check, the radiation level is displayed. After 30 seconds a beep indicates enough information has been collected to ensure statistical validity.
Be sure there is no obstruction between the detection window and source being monitored.
Data is updated every 3 seconds unless you’re using the timer mode.
HazMat Field Screening Equipment Manual
1/30/2015
Page 151 of 168
California CUPA Forum
Remove batteries when storing for long periods of time
Power down the unit by switching the bottom switch to OFF.
Inspector EXP (pancake external probe)
Analog – Monitor 4EC
Inspector EXP
Actuating:
1. Connect external probe.
2. Install standard 9 volt alkaline battery
3. Set the mode switch to the desired mode.
4. Set the bottom switch to On or Audio. After a short system check, the radiation level is displayed. After 30 seconds a beep indicates enough information has been collected to ensure statistical validity.
Be sure there is no obstruction between the detection window and source being monitored.
Data is updated every 3 seconds unless your using the timer mode.
Remove batteries when storing for long periods of time
Power down the unit by switching the bottom switch to OFF.
HazMat Field Screening Equipment Manual
1/30/2015
Page 152 of 168
California CUPA Forum
SMITHS DETECTION RADIOISOTOPE IDENTIFIER
Smiths Detection RadSeeker Handheld Radioisotope Identifier
Equipment Use: Handheld meter with two types of sensors for both gamma and neutron detection provides for assessment of radiation threat for HazMat emergency or WMD response.
Spectrum processing and identification algorithms provide state of the art identification accuracy.
_______________________________________________
____________________________________________
Detection: Gamma and Neutron
Website: Smiths Detection Radseeker Handheld Radioisotope Identifier
Reference Link: RadSeeker Handheld Radioisotope Identifier
HazMat Field Screening Equipment Manual
1/30/2015
Page 153 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 154 of 168
California CUPA Forum
RAE SYSTEMS
Gamma RAE II R Radiation Detector
Equipment Use: This a handheld radiation meter with dual capabilities of a general environmental response detector and dose measurement (dosimeter).
___________________________________________________
_______________________________________________
Detection: Gamma Radiation
RAE Systems Gamma RAE II R Radiation Detector and Dosimeter
User’s Guide: Gamma RAE II R Radiation Detector
Contact (877) 723-2828
HazMat Field Screening Equipment Manual
1/30/2015
Page 155 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 156 of 168
California CUPA Forum
DoseRAE 2 – Electronic Dosimeter
Equipment Use: Electronic direct reading personal detector to detect gamma and x-ray radiation. Provides real-time monitoring of personal dose and dose rate.
Specifications: Detection – Gamma and x-ray radiation
__________________________________________________
____________________________________________
Detection: Gamma and X-Ray radiation
RAE Systems DoseRae Pro Electronic Dosimeter
User’s Guide: DoseRAE 2 Electronic Dosimeter
Contact (877) 723-2878
HazMat Field Screening Equipment Manual
1/30/2015
Page 157 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 158 of 168
California CUPA Forum
HAZCAT KITS
HAZTECH SYSTEMS, INC
HazCat kits can be used to categorize unknown solids and liquid chemicals in the field. These kits are used for screening purposes and not intended to replace analysis using calibrated field monitors or licensed laboratories. HazCat 2.0 Identification System is designed for substance identification in the field. Extensive training is provided for emergency responders and a kit is issued with appropriate supplies for use in the detection of unknown chemicals.
MICROCAT/WMD KIT
Equipment Use:
This kit is for screening white powder unknowns for potential hazards of nuclear, biological and chemical WMD warfare agents. This kit can enable First Responders to quickly assess the nature of suspicious powder threats in the field. It is designed to augment WMD field response protocol and features an assessment algorithm to categorize a suspect powder as either harmless or a potential WMD threat. Used for screening purposes and not intended to replace sophisticated analysis of licensed laboratories or federal authorities. The kit can reduce the number of samples that must be sent to a laboratory and save time to reduce severe economic impacts downtime at critical facilities.
Kit Supplies
MICROCAT/WMD KIT
________________________________________________
-Clinical microscope, phase optics (10x, 40x, 100x)
-Digital camera microscope attachment
-HazCat kit with an assessment algorithm to sort between WMD threats and non-threats
HazMat Field Screening Equipment Manual
1/30/2015
Page 159 of 168
California CUPA Forum
-Training materials
-Travel cases
Operation A minimum of 40 hours of hands-on training is required to operate this kit. A variety of flow charts are used for screening radiological, biological and chemical WMDs. Chemical weapon screening involves substantial skills in basic chemical hazcatting. Radiological screening requires operating a radiation detection meter. Biological screening requires skills in operating a compound microscope and conducting a variety of organic chemical and immune-assay tests. Other required skills is wearing a
Level A chemical encapsulation suit while using this kit.
Sensitivity Range
Deficiencies
Varies from test to test
Expensive, fragile and bulky components
Vehicles lack space for routine transport
Difficult to use while wearing personal protection equipment
Some test reactions are difficult to interpret, even under ideal conditions
False positives/negatives are likely due to operator’s level of experience and familiarity of equipment.
Regular use and practice is needed to maintain necessary skill sets
Calibration Basic calibration is needed before every microscope use and is part of the 40 hour training exercise. Radiation meter may need annual test for simple qualitative radiation detection.
Website: Haztech Systems, Inc
Reference: http://www.hazcat.com/Products.html
Contact: HazTech Systems, Inc (800) 543-5487
HazMat Field Screening Equipment Manual
1/30/2015
Page 160 of 168
California CUPA Forum
HAZCAT KIT
HazMat Field Screening Equipment Manual
1/30/2015
Page 161 of 168
California CUPA Forum
QUICK TEST KIT
1) BORAX BEAD TEST
Place ½ inch of Sodium Borate in a test tube
Unknown chemical. Add 2-3 granules of solid or ¼ inch liquid
Heat until the borax forms a glassy bead
Allow to cool and observe color
OBSERVATIONS
OXIDIZED ELEMENT
Deep Blue
Aquamarine
COBALT
COPPER
REDUCED
Deep Blue
Red Opaque Grey (usually with specks of free copper metal)
Green to Lime Green
Yellow
Deep Purple/Brownish
Red
CHROMIUM
IRON
Green to Lime Green
Green to Lime Green
MANGANESE Colorless/Clear
Clear MOLYBDENUM
Colorless (Yellow when hot)
TITANIUM
Colorless TUNGSTEN
Yellow URANIUM
Colorless VANDIUM
Powder Blue (Yellow when hot)
Yellow
Green
Green
2) HAIRPIN TEST
The Hairpin Test is a test for explosives.
Unknown chemical. Place a grain size solid or drop of liquid on a watch dish.
Take a hairpin and heat one end in the flame of a torch until cherry red.
Take the heated end of the hairpin and touch it to the unknown solid or liquid.
OBSERVATIONS
The heated end of the hairpin when in contact with the unknown material will create a volatile explosive “puff”
HazMat Field Screening Equipment Manual
1/30/2015
Page 162 of 168
California CUPA Forum
3) CHAR TEST
The Char Test is to determine if unknown is organic or inorganic.
Always do a Hairpin Test before conducting a Char Test
Unknown chemical. Add two peas-size amounts or ½ inch of the liquid to a test tube.
Heat until no further reaction takes place or the test tube is melting.
OBSERVATIONS
ORGANIC: Ignition of head space vapors or smoke.
Exception: halogenated hydrocarbons, which sink in water, will not ignite.
INORGANIC:
4) CHLORINE HOT WIRE TEST
No ignition of head space vapors or smoke
The Chlorine Hot Wire Test is used to test an unknown chemical for determination of chlorine, an amine, a nitrate or ammonium salt.
Unknown chemical. Place liquid into a test tube.
Heat the Chlorine Hot Wire (copper wire) to with a torch flame, until there is not a green flame and then allow the wire to cool. (If the green flame persists, or if the wire is black or discolored, cleaning of the wire will be necessary).
Note: To clean the Chlorine Hot Wire
Heat the Chlorine Hot Wire in a torch flame until heated through
Place the heated wire into a small jar labeled “Chlorine Hot Wire Cleaner Methanol”, stir for a few seconds and remove.
The wire is clean if a bright copper color is observed when the wire is removed from the methanol. BE CAREFUL the methanol can ignite if the
Repeat as necessary to obtain a bright copper color.
Place the cleaned wire into the test tube containing the unknown chemical; then reheat the wire using a torch flame.
OBSERVATIONS
A green flame indicates chlorine, an amine, a nitrate, or an ammonium salt.
5) CHROMIUM TEST
Unknown chemical. Place ½ inch of substance into a test tube.
Adjust the pH to less than 3 by adding a drop of Sulfuric Acid.
Add 10 drops of Diphenylcarbazide.
HazMat Field Screening Equipment Manual
1/30/2015
Page 163 of 168
California CUPA Forum
OBSERVATIONS
A deep color purple indicates a chromate
6) COMBUSTIBILITY TEST
Unknown chemical. Add the chemical to a watch dish to form a pool the size of a quarter.
Try to ignite with a lit match.
OBSERVATIONS
EXTREMELY FLAMMABLE: Flame jumps from match to liquid
FLAMMABLE: Stays ignited after removing the match
COMBUSTIBLE:
NON-FLAMMABLE:
Requires match as a wick
Will not ignite
7) CYANIDE TEST
Unknown chemical. Dilute unknown in water.
Add a drop of Sodium Hydroxide to obtain a pH of 11.
Add 3 – 4 drops of p-Dimethylaminobenzal rhodanine.
Add a drop of Silver Nitrate.
OBSERVATIONS
Positive for Cyanide if NO COLOR CHANGE is observed
8) ORGANOPHOSPHATE TEST
Unknown chemical. Dilute unknown in water (Observe an emulsion)
Add 2 – 3 drops of Silver Nitrate to the emulsified solution
OBSERVATIONS
A lemon yellow color indicates the presence of an organophosphate
Note: Pesticides have a strong odor.
9) OXIDIZER TEST
10) pH TEST
Acidify a Potassium Iodide Starch Paper with a drop of Hydrochloric Acid.
Touch the unknown with the paper.
OBSERVATIONS
Blue or Black or Purple color indicates an oxidizer
HazMat Field Screening Equipment Manual
1/30/2015
Page 164 of 168
California CUPA Forum
Wet pH Test Paper with water
Dip the pH Test Paper into the unknown solution
OBSERVATIONS
Compare the colors with those on the container of the PH Test Paper pH greater than 2 or less than 12. Not hazardous (CCR Title 22) pH greater that 12.5 is corrosive pH less than or equal to 2 is acidic
11) SULFIDE TEST
Wet Lead Acetate Test Paper with a drop of water.
Touch the paper to the unknown on a watch dish.
OBSERVATIONS
A Color change to Brown indicates a Sulfide
12) WATER REACTIVE TEST
Place a pea-size amount of the solid unknown or form a dime size pool of the liquid unknown on a watch glass.
Place a drop of Hydrochloric Acid on Potassium Iodide Starch Paper
OBSERVATIONS
Positive if occurrence of a Reaction (Ignore color of Potassium Iodide Starch
Test Paper)
13) WATER SOLUBILITY TEST
Add a pea-size amount of the unknown to a ½ inch of water in a test tube.
If the unknown is not effervescing, stopper the test tube and shale it vigorously.
OBSERVATIONS
Allow time for reactions to occur and observe the results of unknown solubility
14) WATER TEST
Add a pea-size amount of the unknown to a ½ inch of water in a test tube.
OBSERVATIONS
Effervescence indicates the presence of greater than 1% water in the solution.
Further detailed instruction regarding tests can be found in the Ascot Chemical Identification
Manual by Torrington.
HazMat Field Screening Equipment Manual
1/30/2015
Page 165 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 166 of 168
California CUPA Forum
TELEMETRY
Used to record readings from detection meters at multiple locations to establish perimeter and area data from a contamination source. Specific analyzers (gas/radiation) are placed in strategic areas to provide immediate (real-time) detection of contaminants to communicate the readings to a remote location.
Safe Environment Engineering – Lifeline Wireless Monitoring System
Equipment Use:
This system relays live readings by the use of a Lifeline Interoperable Network Communicator
(LINC) which connects to a variety of instruments to create a wireless connection through a
Lifeline Gateway WiFi to a remote computer system. Each detection meter is connected to a
LINC for transmission to the remote PC which can be located in a safe location away from the potential hazard. Event data can be shared with real-time information between departments and across jurisdictions.
LINC Lifeline Gateway WiFi
______________________________________________
Power Source: Lithium Ion batteries (rechargeable)
Website: LINC – Safe Environment Engineering Contact: (661) 295-5500
Reference Link: http://www.safeenv.com/lifeline-training-video-lifeline-operation/
HazMat Field Screening Equipment Manual
1/30/2015
Page 167 of 168
California CUPA Forum
INTENTIONALLY LEFT BLANK
HazMat Field Screening Equipment Manual
1/30/2015
Page 168 of 168
California CUPA Forum
Download
Advertisement
Key features
Field screening equipment
Multi gas meters
Radiation meters
Mercury vapor analyzers
Colorimetric gas indicators
Hazcat kits
Telemetry
Quick reference guide for emergency responders and environmental specialists
Frequently asked questions
This manual is intended to provide information on the use of hazardous materials field screening equipment.
This document is not intended as a training tool, but as a quick reference for those who have received full and appropriate training.
The instruments described in this manual represent the current models that are marketed by a variety of equipment manufacturers.