advertisement
26
Additel 161, 161Ex
Intelligent Digital Pressure Modules
Pressure ranges to 60,000 psi (4,200 bar)
Precision accuracy to 0.01% RD
Intrinsically Safe (Ex) models available
Fully temperature compensated accuracy
OVERVIEW
Gauge pressure Differential pressure
With advanced microprocessor technology and state-of-the-art silicon pressure sensors, Additel’s 161 and 161Ex series Digital Pressure Modules provide an accurate, reliable, and economic solution for wide range of pressure applications. Our intrinsically safe (Ex) models are up to the task of providing the best possible results, even in hazardous environments. In order to reach the best performance, every silicon pressure sensor in the module has been specially aged, tested and screened before assembly. Designed as external pressure modules for Additel’s 760 automatic handheld pressure calibrator, the ADT761 automated pressure calibrator, ADT226/227 series handheld process calibrator and Additel’s flagship 780 pressure controller, the Additel 161 is unmatched in performance and reliability. If intrinsic safety is a critical requirement for your workload, we have you covered when you combine our model ADT226Ex/227Ex process calibrator with any of our ADT161Ex pressure modules.
FEATURES
Precision sensor measurement accuracy to 0.01% RD
Gauge pressure measurement accuracy of 0.02% FS
Pressure ranges to 60,000 psi (4,200 bar)
Intrinsically Safe (Ex) models available
Advanced temperature compensation
ISO 17025 accredited calibration and data included
PRESSURE RANGE
Differential Pressure
P/N
Pressure Range [1]
Media
(inH
2
0) (mbar)
DP1
DP2
DP5
±1
±2
±5
±2.5
±5.0
±10
G
G
G
Accuracy
(%FS)
DP10
DP20
±10
±20
DP30
DP50
±30
±50
DP100 ±100
DP150 ±150
DP300 ±300
±25
±50
±75
±160
±250
±350
±700
G
G
G
G
G
G
G
0.05
0.05
0.02
0.02
0.02
0.05
[2]
0.05
[2]
0.05
[2]
0.05
[2]
0.05
Burst
Pressure
100×
100×
50×
20×
20×
Static
Pressure
Range
±10 psi
±10 psi
±10 psi
±10 psi
±10 psi
20×
3×
3×
3×
3×
±10 psi
±10 psi
±15 psi
50 psi
50 psi
1 year stability.
[2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year stability.
Gauge Pressure [1]
Pressure Range
P/N
(psi) (bar)
Media
[2]
Accuracy(%FS)
V15
GP2
GP5
GP10
GP15
GP30
GP50
GP100
GP150
GP300
GP500
GP600
GP1K
GP1.5K
1,500
GP2K 2,000
GP3K
GP5K
-15
2
5
10
15
30
50
100
150
300
500
600
1,000
3,000
5,000
GP10K 10,000
GP15K 15,000
GP20K 20,000
GP25K 25,000
-1.0
0.16
0.35
0.7
1.0
2.0
3.5
7.0
10
20
35
40
70
100
140
200
350
700
1,000
1,400
1,600
G
G
G
G
G
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
G , L
0.02
0.05
0.05
0.02
0.02
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.0
1 (0.02)
0.05
0.05
0.05
GP30K 30,000
GP36K 36,000
GP40K 40,000
GP50K 50,000
2,000
2,500
2,800
3,500
G , L
G , L
G , L
G , L
0.05
0.05
0.05
0.1
GP60K 60,000 4,200 G , L 0.1
[1] Sealed gauge pressure for above 1000 psi
[2] G=Gas, L=Liquid
[3] 2x for 0.01% FS
[4] 1.2x for 0.01% FS
Burst
Pressure
2x
1.5x
1.5x
1.5x
1.5x
1.35x
1.2x
1.1x
3×
3×
3×
3×
3× [3]
3× [3]
3×
[3]
3× [3]
3× [3]
3× [3]
2× [4]
3×
3× [3]
3× [3]
3× [3]
3× [3]
3× [3]
3× [3]
Note: and cannot be read by Ex devices.
Visit our website at www.additel.com or call 714-998-6899
SPECIFICATIONS
CP600
CP1K
CP2K
CP3K
CP5K
Note:
CP10K
CP2
CP10
CP15
CP30
CP100
CP150
CP300
CP500
Compound Pressure
P/N
Pressure Range psig bar.g
±2
±10
±15
-15 to 30
-15 to 100
-15 to 150
-15 to 300
-15 to 500
-15 to 600
-15 to 1,000
-15 to 2,000
-15 to 3,000
-15 to 5,000
-15 to 10,000
±0.16
±0.7
±1.0
-1 to 2.0
-1 to 7.0
-1 to 10
-1 to 20
-1 to 35
-1 to 40
-1 to 70
-1 to 140
-1 to 200
-1 to 350
-1 to 700
Media
G
G
G
G
G,L
G,L
G,L
G,L
G,L
G,L
G,L
G,L
G,L
G,L
Accuracy(%FS)
Precision Sensors [1]
P/N
Pressure Range psi bar
AP15R 0 to 15 0 to 1
CP30M -15 to 30 -1 to 2.0
CP50M
CP100M
-15 to 50
-15 to 100
-1 to 3.5
-1 to 7.0
CP300M
CP500M
CP1KM
CP2KM
CP3KM
-15 to 300
-15 to 500
-15 to 1,000
-15 to 2,000
-15 to 3,000
-1 to 20
-1 to 35
-1 to 70
-1 to 140
-1 to 200
Media
G
Accuracy [2]
0.01% FS
G, L
G, L
G, L
G, L
G, L
G, L
G, L
G, L
0.005% rdg + 0.005% FS
0.005% rdg + 0.005% FS
0.005% rdg + 0.005% FS
0.005% rdg + 0.005% FS
0.005% rdg + 0.005% FS
0.01% rdg or 0.003% FS whichever is greater
0.01% rdg or 0.003% FS whichever is greater
0.01% rdg or 0.003% FS whichever is greater
CP5KM -15 to 5,000 -1 to 350 G, L 0.01% rdg or 0.003% FS whichever is greater
CP10KM -15 to 10,000 -1 to 700 G, L 0.01% rdg or 0.003% FS whichever is greater
[1] Contact Additel for other range options.
[2] Accuracy includes calibration uncertainty, linearity and long-term stability.
Note:
0.05% FS
0.02% FS
0.02% FS
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
0.01
(0.02)
Burst
2x
Pressure Rating
Over Pressure
1.2x
2x
2x
2x
2x
2x
2x
2x
2x
2x
1.2x
Pressure Rating
Burst Over Pressure
3x
3x
3x
3x
3x
3x
3x
3x
3x
3x
3x
3x
3x
2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.2x
1.1x
1.1x
1.1x
1.1x
Barometric Pressure
P/N
BP
Pressure Range
Low High
60 kPa 110 kPa
Media
G
Accuracy
55 Pa
Burst Pressure
3×
Visit our website at www.additel.com or call 714-998-6899 27
ORDERING INFORMATION
Model Number
ADT161 01 GP300 PSI N Accessories included
Model:
ADT161
ADT161Ex
Accuracy:
01 - 0.01% FS
(non-Ex models only)
02 - 0.02% of full span
05 - 0.05% of full span
10 - 0.1% of full span
01RD - See precision sensor table
(non-Ex models only)
Range type:
PSI - range by psi
BAR - range by bar
Pressure range P/N:
See pressure range table
Pressure port type:
N - 1/4NPT male
N2 - 1/2NPT male
B - 1/4BSP male
B2 - 1/2BSP male
M - M20 X 1.5 male
AF - Autoclave F-250-C female
AM - Autoclave M-250-C male
Optional Accessories
Model number
Description
9060 Pressure module connection cable
Picture
Additel 226Ex with ADT161Ex Pressure Module
SPECIFICATIONS
Operating temperature
Storage temperature
Relative humidity
Pressure connections
(for external use only)
Enclosure
(for external use only)
Intrinisic Safety
(ADT161Ex models only)
Dimensions (Dia x H)
Weight
Warranty
CPXX
Standard Accuracy
DPXX
-10°C to 50°C (14°F to 122°F)
GPXX
-30°C to 70°C (-22°F to 158°F)
95% RH
1/4NPT, 1/2NPT, 1/4BSP, 1/2BPS, M20 x15
Precision Accuracy
CPXM
10°C to 30°C (50°F to 86°F)
-30°C to 70°C (-22°F to 158°F)
90% RH
SS enclosure
II 1G EX ia IIC T4 Ga
0.4 kg (0.99 lb)
1 Year
28 Visit our website at www.additel.com or call 714-998-6899
advertisement