auto - scanning with digital control lcd color


Add to my manuals
43 Pages

advertisement

auto - scanning with digital control lcd color | Manualzz

AUTO - SCANNING WITH DIGITAL CONTROL

LCD COLOR MEDICAL MONITOR

AMM215MWTD

Operation Manual

For future reference, record the serial number of your display monitor in the space below :

Serial Number

The serial number is located on the back of the monitor

back of the monitor

WARNING

The title «WARNING» is used to inform the users of possible causes that could inflict the injury, death, or property damage to the patients.

AMM215MWTD 1

CAUTION

The title «CAUTION» is used to inform the users of possible causes that could inflict the patients although it might not severe enough to cause deaths.

NOTE

The title «NOTE» is used to inform the users of items that are of importance in terms of installation, operation, or maintenance of the Equipment although the failure does not inflict the bodily harm to the patients.

ADVAN INT’L CORP.2013

All other trademarks are the property of their reference owners.

This document is subject to change without notice.

Advan provides this information as reference only. Reference to other vendor’s products does not imply any recommendation or endorsement.

Revision Control

Document Number

Rev. A

2 User’s Guide

TABLE OF CONTENTS

Product Description and Intended Use ......................................................

Warnings and Cautions .............................................................................

Symbol Explanations .................................................................................

EU Declaration of Conformity for Medical Applications .............................

Safety Precaution ......................................................................................

Cleaning Your Monitor ...............................................................................

Recommendation to Use more than One Unit...........................................

On Burn-in .................................................................................................

Power Management Function....................................................................

Preset Modes..............................................................................................

Video Signals .............................................................................................

PIP / POP / PBP function...........................................................................

DDC ...........................................................................................................

Installation .................................................................................................

Connecting the Power Cord ......................................................................

User Function / Messages.........................................................................

OSD Section ..............................................................................................

Troubleshooting .........................................................................................

Specification of (AMM215MWTD) .............................................................

Classification .............................................................................................

Electromagnetic Compatibility ...................................................................

Dimension Drawing (mm) of AC Adapter ...................................................

Dimension Drawing (mm) of AMM215MWTD ............................................

Connectors ................................................................................................

Description of Warranty .............................................................................

Touchscreen User’s Manual ......................................................................

Accessory Item List ...................................................................................

28

30

31

32

17

25

26

27

33

37

39

14

14

15

16

10

11

12

13

9

9

9

9

3

4

7

8

AMM215MWTD 3

Product Description and Intended Use

Please check the following items are present when you unpack the box, and save the packing materials in case you will need to ship or transport the monitor in future.

• AMM215MWTD LCD Monitor(the equipment that displays information, graphic data, image and video on its LCD screen) and two video cables

(1) HD15 VGA cable (1) DVI-I cable

• AC Power cord*

• AC-Adapter

CAUTION

Model No : JMW1100KB1300F02

• USB Cable**

• User Manual, Quick Reference Guide and 4pcs VESA screws

Cable Cover

*Might vary pending on region standard

**Only available for touch models

4 User’s Guide

Warnings and Cautions

Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed:

Warning The personal safety of the patient or physician may be

involved. Disregarding this information could result in

injury to the patient or physician.

Caution Special service procedures or precautions must be

followed to avoid damaging the instrument.

Note Special information to make maintenance easier or important information more clear.

An exclamation mark within a triangle is intended to alert the user to the presence of important operating and maintenance instructions in the literature accompanying the product.

A lightning bolt within a triangle is intended to warn of the

presence of hazardous voltage. Refer all service to authorized personnel.

Warning TO AVOID POTENTIAL SERIOUS INJURY TO THE USER

AND THE PATIENT AND/OR DAMAGE TO THIS DEVICE,

THE USER MUST :

Warranty is void if any of these warnings are disregarded.

ADVAN Int’l Corp accepts full responsibility for the effects on safety, reliability, and performance of the equipment only if:

• Re-adjustments, modifications, and/or repairs are carried out exclusively

by ADVAN Int’l Corp.

• The electrical installation of the relevant operating room complies with the

applicable IEC and CE requirements.

Warning Federal law (United States of America) restricts this

device to use by, or on order of a physician.

The ADVAN Int’l Corp AMM215MWTD monitor has been tested under UL 60601-1 standard and UL listed for Medical application.

ADVAN Int’l Corp reserves the right to make improvements in the product(s) described herein. Product(s), therefore, may not agree in detail to the published design or specifications. All specifications are subject to change without notice.

Please contact ADVAN Int’l Corp directly or phone your local ADVAN Int’l Corp sales representative or agent for information on changes and new products.

AMM215MWTD 5

Warnings

1. Read the operating manual thoroughly and be familiar with its contents prior to using

this equipment.

2. Carefully unpack the unit and check if any damage occurred during shipment.

3. Should any solid object or liquid fall into the panel, unplug the unit and have it checked

by qualified personnel before operating it any further.

4. Uplug the unit if it is not to be used for an extended period of time. To disconnect the

cord, pull it out by the plug. Never pull the cord itself.

5. Be a qualified physician, having complete knowledge of the use of this equipment.

6. Test this equipment prior to a medical procedure. This monitor was fully tested at the

factory before shipment.

7. Avoid removing covers on control unit to avoid electric shock.

8. Attempt no internal repairs or adjustments not specifically detailed in this operating

manual.

9. Pay close attention to the care, cleaning instructions in this manual. A deviation may

cause damage (refer to the Cleaning section).

10. DO NOT STERILIZE MONITOR.

11. Read the entire instruction manual before assembling or connecting the camera.

12. Do not place the monitor or any other heavy object on the power cord. Damage to the

cable can cause fire or electirc shock.

13. Monitor with power supply is suitable for use in patient environment.

14. DO NOT stack more than 3 boxes high

This equipment has been tested and found to comply with the limits for medical devices in

IEC 60601-1-2:2003. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.

This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

- Reorient or relocate the receiving device.

- Increase the separation between the equipment.

- Connect the equipment into an outlet on a circuit different from that to which the other

device(s) are connected.

- Consult the manufacturer or field service technician for help.

NOTICES TO USER

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

FCC WARNING

This equipement generates or uses radio frequency energy. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. The user could lose the authority to operate this equipment if an unauthorized change or modification is made.

6 User’s Guide

Cautions

1. The AC Adapter must be plugged into a Grounded power outlet.

2. Use only the proprietary AMM215MWTD power supply for the

AMM215MWTD monitor. Make a proper connection by ensuring that the

shrink tubing completely secures the connection between the DC power cord and the

extension cord.

3. Turn power off when unit is not in use.

4. Never operate the unit right after having transported from a cold location directly to a

warm location.

5. Do not expose the monitor to moisture or directly apply liquid cleaners directly to the

screen. Spray the cleaning solution into a soft cloth and clean gently.

6. Handle the monitor with care. Do not strike or scratch the screen.

7. Do not block the monitor cooling vents. The monitor is cooled by natural convection

and has no fan.

8. Do not force the monitor past 28 degrees of vertical when adjusting the screen

position. (For monitors equipped with stands only.)

9. Remove the power module and connection when transporting the unit.

10. Save the original carton and associated packing material. They will be useful should

you have to transport or ship the unit.

11. Allow adequate air circulation to prevent internal heat buildup.

12. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains,

draperies) that may block the ventilation slots.

13. Do not install the unit near sunlight, excessive dust, mechanical vibration or shock.

14. The unit is designed for operation in a horizontal position. Never operate the unit in

a vertical position.

15. Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker.)

16. Do not expose the monitor to moisture or excessive dust.

17. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or

IEC 60601-1-1 harmonized national standard or the combination should be evaluated.

Do not touch the patient with signal input or output connectors.

18. Use only a hospital grade power supply cord.

19. This equipment has been tested and found to comply with the limit for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable generate uses and can radiate radio frequency energy, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation, which can be detemined by turing the equipment off and on, the user is encourage to try to correct the interference by one or more of the following measures:

• Reorient or relocate the receiving device.

• Increase the separation distance between the equipment.

• Connect the equipment to an outlet on a circuit different from that to which the

other device(s) are connected.

• Consult the manufacturer or field service technican for help.

20. Grounding reliability can only be achieved when the equipment is connected to an

equipment receptacle labeled “Hospital Only” or “Hospital Grade.”

Note To connect to an international power supply, use a an attachment plug

appropriate for the power outlet.

Note Refer to the “Electromagnetic Compatibility” (EMC) section of this manual to

ensure EMC. The AMM215MWTD must be installed and operated

according to the EMC information provided in this manual.

AMM215MWTD 7

Symbol Definitions

Dangerous: High Voltage

Direct Current

Serial Number

Indicates protective earth ground.

For indoor use only.

Tested to comply with FCC

Class B standards.

This way up DC power control switch

Fragile, handle with care

Stacking limit by number

Keep dry

Recyclable

See accompanying document

Tested to comply with Japan

VCCI certification

Tested to comply with China

CCC certification

China RoHS symbol. Number indicates EFUP(Environmental

Friendly Use Period) from the date of manufacturing.

51LJ

Medical Equipment

E215822

Medical Equipment is in accordance with UL 60601-1 and

CAN/CSA C22.2 No. 601.1 in regards to electric shock, fire hazards, and mechanical hazards

Tested to comply with IP

(ingression protection) Rating

Indicates proof of conformity to applicable European

EconomicCommunity Council directives and to harmonized standards published in the official journal of the European

Communities.

WEEE symbol for the recyclable product. Please do not throw the product with this symbol in the bin.

Authorized Representative in

Europe

Date of Manufacture

Manufacturer

8 User’s Guide

EU Declaration of Conformity for Medical Applications

A Declaration of Conformity has been filed for this product. A sample of this document may be found in the addendum which accompanied this manual. For a copy of the Declaration of Conformity document, please contact ADVAN Int’l Corp. and request for AMM215MWTD

DOC.

Prepare for Unpack

Before you unpack your monitor, prepare a suitable workspace. You need a stable and level surface near a grounded wall outlet in an area which is relatively free of glare from sunlight or other sources of bright light. The monitor is cooled by natural convection (it has no fan).

For optimum performance, do not block the cooling vents.

While unpacking the monitor, inspect it and other package contents for shipping damage that could cause a fire or shock hazard. Immediately report any shipping damage to the carrier or transportation company and contact customer service for monitor in the future or in case of return.

After you unpack the monitor, make sure the following items are included

• Monitor with video cable

• AC adapter with cable

CAUTION: AC Adapter must be plugged into Grounded a power outlet

CAUTION : AC adapter

Model No: JMW1100KB1300F02

This operations manual

Note: Your system provider may offer alternative cords or cables depending on the installation requirement and local geography issues.

AMM215MWTD 9

SAFETY PRECAUTION

· Avoid placing the monitor, or any other heavy object, on the power cord

to prevent fire or electrical shock from damage to the power cord.

· Do not expose the monitor to rain, excessive moisture, or dust to avoid

fire or shock hazard.

· Do not cover the slots or openings of the monitor for proper heat

dissipation. Always put the monitor in a place where there is adequate

ventilation.

· Avoid placing the monitor against a bright background or where sunlight

or other light sources may reflect on the area of the monitor. Place the

monitor just below eye level.

· Handle with care when transporting the monitor.

· Refrain from giving the shock or scratch to the screen, as screen is

fragile.

CLEANING YOUR MONITOR

No specific liquid or chemical necessary when cleaning this LCD monitor

However, we suggest to clean the monitor with non-abrasive cloths and cleaning solutions used in hospitals to clean similar equipment. We recommend using 70%

Isopropyl alcohol for the screen surface and warm water and a mild detergent for all other surfaces. Other acceptable cleaning agents are listed below:

• 70% isopropyl alcohol

• Cidex (2.4% glutaraldehyde solution)

• 0.5% Chlorhexidine in 70% isopropyl alcohol

To clean the screen, do not spray liquid cleaners directly on to the unit. Stand away

Form the monitor and spray cleaning solution onto a cloth. Without applying excessive pressure, clean the screen with the slightly dampened rag.

Recommendation to Use more than One Unit

As unforeseen situation can occasionally occur for the monitor, when the display is used for safety control of human, equipments or unstable picture, or for emergencies, we strongly recommend you use more than one unit or prepare a spare unit.

On Burn-in

Permanent burn-in may occur from the followings:

• Displaying color bar or static images repeately or for a long period of time.

• Using the unit repeatedly in a high temperature/ high humidity environment

• Displaying an image smaller than the monitor continuously.

To reduce the risk of burn- in, we recommend turning off the power of the unit, and reducing the brightness when the unit is not in use.

10 User’s Guide

POWER MANAGEMENT FUNCTION

The monitor is equipped with the power management function which automatically reduce the power consumption when not in use in three power level modes.

• Stand-by Mode

The monitor goes into stand-by mode when the horizontal sync signal is off for about 10 seconds. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The screen is displayed after the horizontal sync signal is restored.

• Suspend Mode

The monitor goes into suspend mode when the vertical sync signal is off for about 10 seconds. The power consumption during this is 8 W. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off.

The screen is displayed after the vertical sync signal is restored.

• Off Mode

The monitor goes into power-off mode when the vertical and horizontal sync signals are off for about 10 seconds. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The screen is displayed after the vertical and horizontal sync signals are restored.

Power Management System

The AMM215MWTD Medical Monitor power management proposal. Provides four phases of power-saving modes by detecting the horizontal sync signal as shown in the table below.

State

Horizontal Sync

Vertical Sync

Video

Power Indicator

Normal

Operation

Active

Active

Active

Green

DPMS

Standby

Inactive

Active

Blanked

Green Flashing (1 sec. Interval)

DPMS

Suspend

Active

Inactive

Blanked

Green Flashing (1 sec. Interval)

DPMS Off

Inactive

Inactive

Blanked

Green Flashing (1 sec. Interval)

Power

Consumption

43 W 8 W 8 W 8W

When the monitor is power saving mode or detects an incorrect timing, the screen will be blank and power LED indicator will blink.

PRESET MODES

DVI and VGA input signal formats

Resolution

640 x 350 @70Hz

640 x 480 @60Hz

640 x 480 @75Hz

800 x 600 @56Hz

800 x 600 @60Hz

800 x 600 @75Hz

1024 x 768 @60Hz

1024 x 768 @70Hz

1024 x 768 @75Hz

1152 x 864 @60Hz

1152 x 864 @70Hz

1152 x 864 @75Hz

1280 x 1024 @60Hz

1680 x 1050 @60Hz

1680 x 1050 @60Hz

1600 x 1200 @60Hz

1920 x 1080@60Hz

46.875

48.363

56.476

60.023

54.348

63.955

67.500

63.981

64.742

65.160

74.077

67.500

Horizontal frequency

( KHz )

31.469

31.469

37.500

35.156

37.879

75.000

60.004

70.069

75.029

60.053

70.016

75.000

60.020

59.946

59.944

59.981

60.00

Vertical frequency

( Hz )

70.087

59.940

75.000

56.250

60.317

AMM215MWTD 11

Pixel clock

( MHz )

25.175

25.175

31.500

36.000

40.000

49.500

65.000

75.000

78.750

80.000

94.200

108.000

108.000

119.125

147.000

130.375

148.500

12 User’s Guide

Video Signals

DVI 1 Format

480/59.94i

480/59.94p

576/50i

576/50p

720/50p

720/59.94p

720/60p

1080/23.98p

1080/24p

1080/25p

1080/29.97p

1080/30p

1080/50i

1080/59.94i

1080/60i

1080/50p

1080/60P

O

O

O

O

O

O

O

O

O

O

O

O

O

O

O

VGA

O

O

O

O

O

O

O

O

O

O

O

O

O

O

O

O

PIP / POP / PBP function*

The following combination options are available to you:

PIP

Main

DVI 1

DVI 2

VGA

DVI 1 DVI 2 x x

O x x

O

VGA

O

O x

AMM215MWTD 13

*Not available for touch models

14 User’s Guide

DDC

To make your installation easier, the monitor is able to Plug and Play with your system if your system also supports DDC protocol. The DDC (Display Data

Channel) is a communication protocol through which the monitor automatically informs the host system about its capabilities, for example, supported resolutions and corresponding timing. The monitor supports DDC1 and DDC2B standard.

INSTALLATION

To install the monitor to your host system, please follow the steps as given below:

Steps

1. Use the supplied cable (DVI, VGA, USB*) then connect to the

host system accordingly.

2. Connect the DC power to the DC power connector on the monitor.

3. Connect one end of AC power cord into the AC Adapter and the other end to AC

power outlet.

4. Then turn the host system on and then the monitor.

5. If the monitor still does not function properly, please refer to the troubleshooting

section to diagnose the problem.

DC Power DVI VGA USB*

*Only available for touch models

AMM215MWTD 15

CONNECTING the POWER CORD

* Check first to make sure that the power cord you use is the correct type required for your area.

* This monitor has an universal AC adapter that allows operation in either

AC 100 - 240 V ac voltage area. No user-adjustment is required.

* Plug one end of the power cord to the AC adapter, plug another end to a proper

AC outlet.

The cord set should have the appropriate safety approvals for the country in which the equipment will be installed and marked HAR.

For 120 volt Applications, use only UL Listed deachable power cord with NEMA configuration 5-15P type (parallel blades) plug cap. For 240 volt applications use only UL Listed Detachable power supply cord with NEMA configuration 6-15P type

(tandem blades) plug cap.

Hospital-grade power cord

Extension power cord (optional)

Manufacturer: Bridgepower

Part No: CB-47D2004P57MF (15ft) or

CB-47D2022P86MF (75ft)

Power Supply

Model No: JMW1100KB1300F02

AMM215MWTD

16 User’s Guide

User Functions / Messages

Key Name and Function

Power LED : Lights up to indicate the power is turned ON.

Standby Power Button: To power ON or OFF the monitor.

Rotary switch : With the OSD menu activated, increases the value of the selected

(Turn Right) parameter or moves rightward in OSD menu.

Rotary switch : With the OSD menu activated, increases the value of the selected

(Turn Left) parameter or moves leftward in OSD menu.

PIP Switch* : To enable or disable Picture In Picture function.

PBP Switch* : To enable or disable Picture By Picture function.

POP Switch*: To enable or disable Picture On Picture function.

Adjust Switch* : To activate or deactivate image adjustment in sub screen of PIP,

PBP, POP mode.

Input Switch : To change the video source.

Power Switch : To turn ON or OFF the system.

Standby Power Button

Power Switch

How to access the menu.

1. Push the “Knob” button to activate the OSD menu.

2. Able to turn left or right the button to select those icons. The icon will be highlighted when is selected.

3. Push the “knob” button to go into the another OSD menu.

4. Turn left or right “Knob” button to increase or decrease the number of selected function.

5. In order to exit from the OSD menu in a different layer, then choose the “Exit option”. In case of long pushing and holding the button around two or three seconds, then it will completely out from the menu regardless wherever you are.

6. Turn the “Knob” button in a fast way to the left or right, while the menu is not activated, then input signal menu will turns up with “V” mark which is chosen current available input signal. You can switch to another input signal source by turning the knob slowly then select the target source by pushing the knob.

*Not available for touch models

AMM215MWTD 17

OSD Section

Main Menu

Chromatic Adjust Color Temp, Gamma

Visual Adjust Brightness, Contrast, Phase*, Chroma*,

Setting Adjust Scale Mode, Color Space*, Speaker, Freeze Frame,

Zoom / Pan, PIP**, POP**, PBP**

Advanced Adjust OSD Position Control, Screen Control, DPMS, Auto

Source Select, Smart Select, Restore Factory, Key Lock,

Overscan

Information Adjust Custom Entry, SN, Run Time, Input Format

Exit Exit the menu

*Functions not available for AMM215MWTD

**Functions not available for touch models

18 User’s Guide

OSD Section

Chromatic

Color Temp Change the color temperature - D65, D93, S1, S2

Gamma

Exit

Change the gamma value - 1.8, 1.9, 1.95, 2.1, 2.1S, 2.2, 2.3,

2.4, 2.4S, S0, Radio graph

Exit the menu

Visual

Brightness Adjust the brightness of panel (Range 0 - 100)

Contrast Adjust the contrast of video (Range 0 - 100)

Phase* Adjust the phase of video (Range 0 - 100)

Chroma* Adjust the chroma of video (Range 0 - 100)

Sharpness-V Set the sharpness of vertical image (Range 0 - 20)

Sharpness-H Set the sharpness of horizontal image (Range 0 - 20)

Exit Exit the menu

*Functions not available for AMM215MWTD

AMM215MWTD 19

OSD Section

Setting

Scale Mode Change the scale mode - Fill all, One to One,

Vertical - fill, Horizontal - fill, Fill aspect ratio

Color Space* Change color space between RGB and YPbPr

Speaker Adjust volume (Range 0 - 100)

Freeze Frame Off / On freeze frame

Zoom / Pan Enable zoom - in and pan function

PIP**

POP**

Enable PIP (Picture In Picture) function

Enable POP (Picture On Picture) function

PBP**

Exit

Enable PBP (Picture By Picture) function

Exit the menu

*Functions not available for AMM215MWTD

**Functions not available for touch models

20 User’s Guide

OSD Section

Sub menu of PIP*

Mode

Source

PIP Mode ON / Off

PIP sub screen source - VGA, DVI1/2

Position PIP sub screen position - Top L, Top R, Bottom L, Bottom R

Size PIP sub screen size - Small, Medium, Large

Blending

Swap

Exit

PIP sub screen blending (Range 0 - 20)

PIP sub screen swap

Exit the menu

Sub menu of PBP*

Mode

Source

Swap

Exit

PBP Mode ON / Off

PBP sub screen source - VGA, DVI1/2

PBP sub screen swap

Exit the menu

*Functions not available for touch models

AMM215MWTD 21

OSD Section

Sub menu of POP*

Mode

Source

Swap

Exit

POP Mode ON / Off

POP sub screen source - VGA, DVI1/2

Position POP sub screen position - Top L, Top R, Bottom L, Bottom R

Size POP sub screen size - Small, Medium, Large

Blending POP sub screen blending (Range 0 - 20)

POP sub screen swap

Exit the menu

*Functions not available for touch models

22 User’s Guide

OSD Section

Advanced

OSD Position Change the OSD menu - Position, Background,

Control OSD time out, Language

Screen Control Control and adjust H and V position, Freqency, Phase

Noise Reduction*, Motion Offset*

DPMS Change the DPMS

Auto Source

Select

Smart Select

Adjust Auto Source Select between on and off

Enable / Disable smart select sub menu

Restore Factory Changes the all OSD value to factory outgoing status

Key Lock

Overscan

Exit

Set to key lock mode

Adjust Overscan ratio

Exit the menu

*Functions not available for AMM215MWTD

OSD Section

Sub menu of OSD Control

H-position Adjust OSD H - position (Range 0 - 100)

AMM215MWTD 23

V-position Adjust OSD V - position (Range 0 - 100)

Back ground Adjust transparency of OSD back ground (Range 0 - 20)

OSD Time out Adjust OSD time out - 5s, 10s, 15s, 20s, 1m, 2m

Language OSD language - English, Japanese, Chinese, Korean,

Exit Exit the menu

Sub menu of Screen Control

H-position Adjust screen H - position (Range 0 - 100)

*available for VGA

V-position Adjust screen V - position (Range 0 - 100)

*available for VGA

Frequency Adjust frequency (Range 0 - 100)

*available for VGA

Phase Adjust phase (Range 0 - 100)

*available for VGA

Noise

Reduction*

Noise reduction (Range 0 - 31)

Motion Offset* Motion offset (Range 0 - 100)

Exit Exit the menu

*Functions not available for AMM215MWTD

24 User’s Guide

OSD Section

Information

Custom Entry Change the user or monitor’s name

SN Display the serial number

Run Time

Exit

Display the total run time

Input Format Display the current input resolution and vertical frequency

Exit the menu

AMM215MWTD 25

TROUBLESHOOTING

Before sending your LCD monitor for servicing, please check the troubleshooting list below to see if you can self-diagnose the problem.

Problems

No Picture

Abnormal Picture

Current Status

LED ON

LED OFF

LED Blinking

Unstable Picture

Display is missing, center shift, or too small or too large in display size

Remedy

· Using OSD, adjust brightness and contrast to maximum or reset to their default settings.

· Check the Rocker switch and Power switch.

· Check if AC power cord is properly connected to the AC adapter.

· Check if video signal cable is properly connected at the back of monitor.

· Check if the power to computer system is ON.

· Check if the specification of graphics adapter and monitor is in compliance which may be causing the input signal frequency mismatch.

· Using the Screen Control Menu, adjust the Phase, Frequency, Horizontal, and

Vertical settings in order to correct the display image.

26 User’s Guide

SPECIFICATION*

Model

Description

Active Screen Size

Resolution

Pixel Pitch

Display Color

Color Tone

Response Time

Face Finishing

Viewing Angle

Brightness**

Contrast Ratio

Input

VGA

AMM215MWTD a-Si TFT Active Matrix, LED Backlight

21.46 inches (545.22mm) diagonal

1920 (H) x 1080 (V) @60Hz

0.2475 mm

16.7M colors

Up to 256 color tone

<25ms Typ.

Protective Filter with Anti-Reflected Hard Coated

R/L 178°, U/D 178° (CR > 10)

250 cd/m2 (Typ.)

1000:1 (Typ.)

15pin D-Sub x 1

R/G/B : 0.7 ± 0.1 Vp-p

H/V Sync : TTL Level (V high ≥2.3V, V low ≤0.5V)

DVI1

DVI2 / Optical

(Optional)

USB (Only for Touch model)

Remote Input

Audio

Scanning Frequency

DVI-I x 1

DVI-I x 1

USB-B Type

USB 1.1 for Touch screen

9–pin D-Sub (RS-232C) x 1

Audio Input: 500mVrms, 3.5mm Stereo Jack x 1

Speaker Out: 1.0 W x 2 EA

Horizontal : 31.47~80.0Khz Vertical : 50~85Hz

General

Power Adaptor

Power Consumption

Dimension

Weight

VESA Mounting

AC 100 ~ 240V 50-60Hz, 2.0A

DC 13V, 6.92A

43W

523 (W) x 321 (H) x 110 (D) mm

Monitor without touch screen: 5.05Kg (Approximately)

Monitor with touch screen: 5.40Kg (Approximately)

AC adaptor : 720g

100mm x 100mm

Operating/Storage Environment

Operating Temperature

32° ~ 95°F (0° - 35°C)

Operating Humidity

Storage Temperature

20% ~ 80%, non-condensing

-4° ~ 140°F (-20° - 60°C)

Storage Humidity 10 ~ 85%RH (without condensation)

Compliance &Certification

Safety

UL (UL60601-1), cUL (CAN/CSA-C22.2 No.6011-M90), CE (EN60601-1), AS/NZS 3200-1-0.

CCC (GB4943-2001), CB-ITE (IEC60950-1), IP23 Compliance

EMC

FCC (Part Class B), CE (EN60601-1-2), AS/NZS 3200-1-2, VCCI (Class B), CCC (GB9254,

GB17625.1)

Optional Module

DC Extension Cable

Base Stand

5ft, 15ft, 75ft length

Adjustable high, swivel and tilt base stand

* All contents are subject to change without notice.

**Brightness shown is without touch screen or A/R filter installed.

AMM215MWTD 27

: Direct Current

51LJ

Medical Equipment

E215822

MEDICAL EQUIPMENT

WITH RESPECT TO ELECTRIC SHOCK,

FIRE AND MECHANICAL HAZARDS ONLY

IN ACCORDANCE WITH UL 60601-1, AND

CAN/CSA C22.2 NO. 601.1

: UL approval mark according to the safety standard for Medical equipment

This monitor is intended for use in Health Care Facilities model AMM215MWTD

Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide.

No user serviceable parts inside, ask qualified personnel when accessing inside.

For disposal of waste product, follow the requirement of local code.

Electrical input rating: 13V DC 6.92A

Classification

Type of protection against electric shock: Class I Equipment.

Degree of protection against the ingress of water: IP23 compliance.

Mode of operation: Continuous

This monitor has been tested to comply with IEC/EN 60601-1, IEC/EN60601-1-2 and is certified by UL to medical standard UL60601-1(UL/cUL Mark).

Because many medical offices are located in residential areas, this monitor, in addition to the medical requirements, has also been tested and found to comply with the limits for FCC Class B computing devices in a typically configured system. It is the system integrator or configurer’s responsibility to test and ensure that the entire system complies with applicable EMC laws.

-

-

Environmental conditions for transport and storage:

- Temperature range within -4° to 140° F (-20° to 60°C)

Relative humidity range within 10% to 85%

Atmospheric pressure range within 500 to 1060 hPa.

Like other electrical medical equipment, the AMM215MWTD requires special precautions to ensure electromagnetic compatibility with other electrical medical devices. To ensure electromagnetic compatibility (EMC), the AMM215MWTD must be installed and operated according to the EMC information provided in this manual.

Note The AMM215MWTD has been designed and tested to comply with IEC

60601-1-2:2001 requirements for EMC with other devices.

Caution Portable and mobile RF communications equipment may affect the normal function of the AMM215MWTD.

Warning Do not use cables or accessories other than those provided with the AMM215MWTD, as this may result in increased electromagnetic emis-

sions or decreased immunity to such emissions.

Warning If the AMM215MWTD is used adjacent to or stacked with other equipment, observe and verify normal operation of the AMM215MWTD in the configuration in which it will be used prior to using it in a surgical procedure. Consult the tables below for guidance in placing the AMM215MWTD.

Manufacturer’s declaration - electromagnetic emission

The Model AMM215MWTD is intended for use in the electromagnetic environment specified below.

The customer or the user of AMM215MWTD should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment - guidance

RF emissions CISPR 11 Group 1

The Model AMM215MWTD uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment

RF emissions CISPR 11

Harmonic emissions

IEC61000-3-2

Voltage Fluctuations

IEC61000-3-3

Class B

Class D

Complies

AMM215MWTD is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes

Manufacturer’s declaration - electromagnetic immunity

The Model AMM215MWTD is intended for use in the electromagnetic environment specified below.

The customer or the user of AMM215MWTD should assure that it is used in such an environment.

Immunity Test

Electrostatic discharge

(ESD)

IEC61000-4-2

6 kV contact

8 kV air

Electrical fast transient

/burst IEC61000-4-4

2 kV for power supply lines

1 kV for input / output lines

Surge

IEC61000-4-5

IEC 60601 Test Level

1 kV differential mode

2 kV common mode

Compliance Level

6 kV contact

8 kV air

2 kV for power supply lines

1 kV for input / output lines

1 kV differential mode

2 kV common mode

Electromagnetic environment - guidance

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the relative humidity should be at least

30%.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment.

Power frequency

(50/60Hz) magnetic field

IEC 61000-4-8

3.0 A/m 3.0 A/m

Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11

<5% Ut (>95% dip in Ut) for 0.5 cycle

40% Ut (60% dip in Ut) for 5 cycles

70% Ut (30% dip in Ut) for 25 cycles

<5% Ut (>95% dip in Ut) for 5 sec.

<5% Ut (>95% dip in Ut) for 0.5 cycle

40% Ut (60% dip in Ut) for 5 cycles

70% Ut (30% dip in Ut) for 25 cycles

<5% Ut (>95% dip in Ut) for 5 sec.

Note: Ut is the A.C. mains voltage prior to application of the test level.

AMM215MWTD 29

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Mains power quaility should be that of a typical commercial or hospital environment. If the user of the AMM215MWTD image intensifier requires continued operation during power mains interruptions, it is recommended that the AMM215MWTD image intensifier be powered from an uninterruptible power supply or a battery.

Manufacturer’s declaration - electromagnetic immunity

The Model AMM215MWTD is intended for use in the electromagnetic environment specified below.

The customer or the user of AMM215MWTD should ensure that it is used in such an environment.

Immunity Test Compliance Level

Conducted RF

IEC61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 KHz to 80 MHz

3 V/m

80MHz to 2.5 GHz

3 Vrms

150 KHz to 80 MHz

3 V/m

80MHz to 2.5 GHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz where P is the maximum oputput power according to the transmitter manufacturer and d is the recommended separation distance in meters ( m).

Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey (a), should be less than the of equipment marked with the follwing symbol:

NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects, and people.

(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.

30 User’s Guide

Recommended Separation Distances Between Portable and Mobile RF Communications Equipement and the AMM215MWTD System

The AMM215MWTD system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the AMM215MWTD system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications output power of the communications equipment.

Separation distance (m) according to frequency of transmitter

Rated maximum output power (W) of transmitter

150kHz to 80MHz d=1.17, p

80MHz to 800MHz d=1.17, p

800MHz to 2.5GHz

d=2.33, p

0.01

0.1

1

10

100

0.12

0.37

1.17

3.70

11.70

0.12

0.37

1.17

3.70

11.70

0.23

0.74

2.33

7.37

23.30

For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Dimension Drawing (mm) of AC Adapter

Dimension Drawing (mm) of AMM215MWTD

AMM215MWTD 31

Standby Power Button

Power Switch

32 User’s Guide

Connectors

DC Input

Connector Jack Power Input

Pin

1

Description

GND

2

3

GND

13 V Input

Video Input

24 pin DVI-I connector.

DDWG or equivalent connector.

Pin

1

6

7

8

4

5

2

3

Description

T.M.D.S. Data2-

T.M.D.S. Data2+

T.M.D.S. Data2 Shield

NC

NC

DDC Clock

DDC Data

NC

Pin

9

14

15

16

10

11

12

13

Pin

4

5

Description

T.M.D.S. Data1-

T.M.D.S. Data1+

T.M.D.S. Data1 Shield

NC

NC

+5V Power

Ground (for +5V

Hot Plug Detect

Description

NC

13V Input

Pin

17

22

23

24

18

19

20

21

Description

T.M.D.S. Data0-

T.M.D.S. Data0+

T.M.D.S. Data0 Shield

NC

NC

T.M.D.S. Clock Shield

T.M.D.S. Clock+

T.M.D.S. Clock-

15 pin VGA connector.

Pin

1

4

5

2

3

Description

RED

GREEN

BLUE

GND

GND

Pin

6

9

10

7

8

Description

GND- RED

GND- GREEN

GND- BLUE

G5V

GND

Pin

11

12

13

14

15

Description

GND

DDC SDA

HSYNC

VSYNC

DDC SCL

AMM215MWTD 33

Description of Warranty

ADVAN warrants to the first Buyer (Buyer) that the product purchased when shipped in its original container will conform to ADVAN specifications, and to any

ADVAN approved specifications furnished to ADVAN by the Buyer, and will be free of defects in materials and workmanship. ADVAN warrants that the product purchased is manufactured from new components which meet ADVAN standards, quality and specifications.

Subject to the conditions and limitations set forth below, ADVAN will, at its option, either repair any component of its products that prove defective by reason of improper workmanship or materials or ADVAN has the exclusive right to replace with refurbished units or with an equivalent product. ADVAN warrants that the components used for repair, refurbished units or equivalent product will meet

ADVAN standards, quality and specifications.

Commencement and Duration of Warranty

The product purchased will be warranted for a period of eighteen (18) months

(excluding the LCD panel, touch screen, and the protection filter) from the date of shipment.

LCD panels and touch screens are warranted for a period of twelve (12) months from the date of shipment.

Protection filters are not warranted as damage to the protection filter is considered to be normal wear and tear and can be replaced at Buyer’s cost.

Components used for repair, refurbished units or equivalent product will be warranted for a period of twelve (12) months from the date of repair.

Limitation of Warranty

This limited warranty does not cover any damage to this product or other non-

ADVAN products that results from any of the following: improper installation or operation; accident; abuse; misuse; natural disaster; war; insufficient or excessive electrical supply; abnormal mechanical or environmental conditions; any unauthorized disassembly, repair or modification; normal wear and tear; tampering by anyone other than an ADVAN engineer or an ADVAN Authorized Service Center

(ASC); the use of supplies, consumable items and conditions beyond the control of ADVAN, such as common carrier provided equipment and/or facilities; operation of ADVAN product in excess of the specifications. This limited warranty also does not apply to any product that has not been handled or packaged correctly, that has been sold as second-hand or has been resold contrary to the US export regulations, on which the original identification information (i.e. serial number, rating and/or warranty label) has been altered, obliterated or removed.

34 User’s Guide

Return Material Authorization (RMA) Procedure

All claims must be submitted through the ADVAN website, whether in warranty or out, first Buyer, distributor or OEM. End Users who have purchased through a distributor or OEM, please contact the distributor or OEM.

http://www.advancorprma.com

Follow the instructions to receive an RMA number, shipping instructions and a shipping label to be placed on the outside of the shipping container. A serial number and a detailed reason for return are required.

In-transit damage is not covered by warranty. ADVAN will only pay for the return shipment by surface transportation.

Returns without an RMA number will not be accepted.

For product support, please e-mail details of your inquiry including product model: [email protected]

In Warranty

ADVAN or its ASC will repair or replace if defective in material or workmanship without cost, for a period of eighteen (18) months, (LCD panels and touch screens for a period of twelve (12) months) after the date of shipment.

Buyer must notify ADVAN or its ASC of the defect before expiration of the warranty period, and request an RMA number. If the configuration has been modified in any manner, the product must be returned to its original configuration before any warranty service will be performed by ADVAN or its ASC. No goods are to be returned to ADVAN or its ASC without prior authorization. Buyer will be responsible for packaging (preferably original container) and shipping the defective goods to

ADVAN or its ASC, shipping charges prepaid.

ADVAN or its ASC will return the in warranty product, at no cost to the buyer.

AMM215MWTD 35

Out-of-Warranty

ADVAN or its ASC will repair or replace if defective in material or workmanship with fee, any product which the warranty period has expired (out-of-warranty).

Buyer must notify ADVAN or its ASC of the defect and request an RMA number. If the configuration has been modified in any manner, the product must be returned to its original configuration before any service will be performed by ADVAN or its ASC.

No goods are to be returned to ADVAN or its ASC without prior authorization. Buyer will be responsible for packaging (preferably original container) and shipping the defective goods to ADVAN or its ASC, shipping charges prepaid.

ADVAN or its ASC will return the out-of-warranty product, at cost to the buyer.

Product End of Life (EOL)

In the event of an RMA of an EOL product(s), ADVAN will hold or store major components of its products for a period of five (5) years, after the EOL of its products. ADVAN shall continue to perform the service of its products as long as

ADVAN holds or stores said components of the products, with reasonable charge.

The forgoing Warranty and Out-of-Warranty terms apply.

36 User’s Guide

Disclaimer

THE FORGOING IS THE COMPLETE WARRANTY FOR ADVAN PRODUCTS

AND SUPERSEDES ALL OTHER WARRANTIES AND REPRESENTATIONS,

WHETHER ORAL OR WRITTEN. EXCEPT AS EXPRESSLY SET FORTH

ABOVE, NO OTHER WARRANTIES ARE MADE WITH RESPECT TO ADVAN

PRODUCTS AND ADVAN EXPRESSLY DISCLAIMS ALL WARRANTIES NOT

STATED HEREIN, INCLUDING, TO THE EXTENT PERMITTED BY APPLICABLE

LAW, ANY WARRANTY THAT MAY EXIST UNDER NATIONAL, STATE,

PROVINCIAL OR LOCAL LAW INCLUDING BUT NOT LIMITED TO ANY IMPLIED

WARRANTY ON NON-INFRINGEMENT, MERCHANTABILITY OR FITNESS

FOR A PARTICULAR PURPOSE. ALL WARRANTIES, WHETHER EXPRESS

OR IMPLIED, ARE LIMITED TO THE PERIODS OF TIME SET FORTH ABOVE.

SOME STATES OR OTHER JURISDICTIONS DO NOT ALLOW THE EXCLUSION

OF IMPLIED WARRANTIES OR LIMITATIONS ON HOW LONG AN IMPLIED

WARRANTY LASTS, SO THE ABOVE LIMITATIONS MAY NOT APPLY.

ADVAN’S TOTAL LIABILITY UNDER THIS OR ANY OTHER WARRANTY,

EXPRESSED OR IMPLIED, IS LIMITED TO REPAIR OR REPLACEMENT, AND

ARE THE SOLE AND EXCLUSIVE REMEDIES FOR BREACH OF WARRANTY

OR ANY OTHER LEGAL THEORY. TO THE FULLEST EXTENT PERMITTED

BY APPLICABLE LAW, ADVAN SHALL NOT BE LIABLE TO THE PURCHASER

OR END USER CUSTOMER OF AN ADVAN PRODUCT FOR ANY DAMAGES,

EXPENSES, LOST DATA, LOST REVENUES, LOST SAVINGS, LOST PROFITS,

OR ANY OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING

FROM THE PURCHASE, USE OR INABILITY TO USE THE ADVAN PRODUCT,

EVEN IF ADVAN HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH

DAMAGES. SOME STATES OR OTHER JURISDICTIONS DO NOT ALLOW

THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL

DAMAGES, SO THE ABOVE LIMITATIONS MAY NOT APPLY.

THE LIMITED WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS, AND YOU

MAY ALSO HAVE OTHER RIGHTS WHICH VARY FROM STATE TO STATE OR

JURISDICTION TO JURISDICTION.

37 User’s Guide

Touchscreen User’s Manual

(Optional)

Touch Driver Installation for Microsoft Windows 7 and XP(Plug and Play)

Windows touch driver will automatically install when you connect the USB cable between the monitor and a computer. The driver may take a few minutes to install.

After the driver is installed, try using your finger to tap on the screen to see if it works.

Touch Function

The touch function may replace the mouse after you start the Windows operating system and introduce appropriate gesture Gesture and related Windows software applications are listed below:

Gesture

Tap

Windows 7

OK

Press and hold OK

Pinch or stretch to zoom OK

Rotate

Slide to scroll

Slide to rearrange

OK

OK

OK

Windows XP

OK

OK

NA

NA

NA

OK

NOTE Before using the touch function, make sure the USB cable is connected, and

the Window’s operating system is started.

NOTE The touch function may need about 7 seconds to resume if the USB cable is

re-plugged or the computer resumes from sleep mode (suspend mode).

AMM215MWTD 38

Touchscreen User’s Manual

Calibrate touch screen

Windows 7

1. Make sure a USB cable is connected from the computer to the monitor.

2. Select Start > Control Panel and change View by to Large icons or Small icons

3. Select the Tablet PC Settings control panel. The Tablet PC Settings control panel will only appear after Microsoft Windows 7 recognizes a connected touch monitor.

4. In the Tablet PC Settings, select the name of the monitor you want to calibrate from the drop down menu.

5. Click on Calibrate.to start the calibration

39 User’s Guide

Accessory Item List

No. Parts Name Q’ty

1 AC- Adaptor 1

7

8

9

4

5

6

2

3

VGA Cable

DVI cable

USB Cable*

User’s Manual

Quick Reference

Guide

Power Cord*

VESA Screw

Extension Power

Cord**

1

1

1

1

1

1

4

1

10 Cable Cover 1

Description

Model Number: JMW1100KB1300F02

Parts Number: PS-52141209002E

RGB 15P TO 15P, 1.8M, IVORY

DVI-I type CABLE 29P TO 29P

Remark

BridgePower

Rev. A

Rev. A

Medical Grade 115V, GRAY

BH,+,M4x10,Ni

Part No: CB-47D2001P50MF (5ft) or

CB-47D2004P57MF (15ft) or

CB-47D2022P86MF (75ft)

BridgePower

Adaptor VGA Cable DVI Cable

USBCable* User’s Manual Quick Reference

Guide

Power Cord* VESA Screw Cable Cover

Extension

Power Cord **

* Might vary pending on region standard

**Optional item, check with local representative

Authorized Representative in Europe and Exclusive Distributor

EIZO Europe GmbH

Helmut-Grashoff-Strasse 18, 41179 Mönchengladbach, Germany

Phone: +49 2161 8210120

Fax: +49 2161 8210150

Web Page : http://www.eizo.eu

Distributor

ADVAN Korea

#605, Kolon Science Valley 2-cha, 55, Digital-ro, 34-gil, Guro-gu, Seoul, 152-728

Korea

Tel : +82 2 783 5197

Fax : +82 2 868-0880

Web Page : http://www.advancorp.com

Manufacturer

ADVAN INT’L CORP

47817 Fremont Blvd. Fremont, CA 94538, USA

Tel : 1 510 490 1005

Fax : 1 510 490 1151

Web Page : http://www.advancorp.com

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement