BHM Medical Inc. Voyager, VOYAGER-AU, VOYAGER-CE, VOYAGER-JP, VOYAGER-UK portable ceiling lift User manual
Below you will find brief information for portable ceiling lift Voyager VOYAGER, portable ceiling lift Voyager VOYAGER-AU, portable ceiling lift Voyager VOYAGER-CE, portable ceiling lift Voyager VOYAGER-JP, portable ceiling lift Voyager VOYAGER-UK. Voyager portable ceiling lift is an easy-to-use, safe and efficient system for lifting patients in a variety of situations. It includes a unique quick-release feature that allows for very fast winding and unwinding of the strap before and after transfer operation. The Voyager portable ceiling lift is designed to be used with a variety of slings, and it is easy to use and maneuver. It is also very portable and can be easily moved from room to room.
Advertisement
Advertisement
S50 Century bt
A
. Jeffrey CooK, O T Re
9
(MB) i ?rodStConsuUant
-r , fjfM 949-2300
Fax: (204) ^ ^ ^ , |
I T
E D I C A L I N C .
RMANUAL
Revision 7, March 2004
IMPORTANT: Read these instructions before installing, operating or servicing this system.
P O R T A B L E
VOYAGER®
Product # V O Y A G E R (N. America)
V O Y A G E R - A U (Australia)
V O Y A G E R - C E (Europe)
V O Y A G E R - J P (Japan)
V O Y A G E R - U K (United Kingdom)
CEILING
LIFT S Y S T E M
@ ©
DO NOT D E S T R O Y
KEEP KEY ENCLOSED AND MANUAL WITH
THE LIFT AT ALL TIMES.
BHM MEDICAL INC., 2001 TANGUAY STREET, MAGOG, QUEBEC, CANADA, JIX 5Y5
B H M Medical Inc. designs and manufactures quality engineered patient lifting equipment for the care of the disabled in hospitals, nursing homes and home care.
Please take the time to read the entire manual, including the section on Safety Instructions and Warnings.
It contains important information that will allow you to take advantage of all the characteristics of your
B H M patient lift.
BHM Medical Inc. Credentials
• FDA Registration # 9681684
• CE mark (European Union Listing)
• Global Operations for in-warranty and out-of-warranty service support for Voyager Portable lift.
• Each finished Voyager Portable is individually tested in our laboratory.
• Lifting Capacity 200 kg/440 lbs. Engineering safety factor exceeds European Standards.
• Complete range of patient lift slings and supports designed and manufactured by BHM Medical Inc. using medical quality materials and fabrics.
• ISO 9001
• Quick release US PATENT #6085368 / Patent Pending in other countries.
• Voyager is a registered trademark of BHM Medical inc.
A P P R O V A L S :
CSA 601.1
UL 2601.1
EN 60601.1
EN ISO 10535 (Lifting device) SGS
UK Test Report: DUR22107/AC/OO
Note: BHM Medical is constantly improving its products. For this reason, it may be possible to encounter a modification of product without revision of this guide.
Thank you for trusting your BHM Medical Partners.
How to contact us:
Head Office
BHM Medical Inc.
2001 Tanguay Street
Magog (Quebec) J1X 5Y5
Canada
Phone: (819) 868-0441
& Service calls during business hours
(8:00-12:00 and 1:00-5:00 p.m. E.S.T.)
Sales Headquarters
BHM Medical Inc.
6200A Tomken Road
Mississauga (Ontario) LT5 1X7
Canada
0 Service calls outside regular business hours - extension #3400
Fax: (819) 868-2249
Web site: www.bhm-medical.com
E-mail: bhm(p)bhm-medical.com
User Manual - Portable Voyager 38
TABLE OF CONTENTS
INTRODUCTION 26
H o w T O USE THIS M A N U A L 2 6
E Q U I P M E N T IDENTIFICATION 2 6
R E C E I P T OF E Q U I P M E N T 2 6
K E Y TO S Y M B O L S 2 6
S A F E T Y INSTRUCTIONS AND WARNINGS 26
D E S C R I P T I O N O F EQUIPMENT 26
L I F T U N I T 2 6
W H A T ' S INCLUDED 2 6
P A R T S DESCRIPTION 2 6
D I M E N S I O N 2 6
BHM E R G O F I T S L I N G 26
H A M M O C K S L I N G 2 6
Hammock Sling (THA-S, M & L) 26
Hammock 6 Strap Sling (THA6-S, M & L) 26
Sling Attachment Options 2 6
Q U I C K - F I T S L I N G 2 6
H Y G I E N I C S L I N G 2 6
W A L K I N G S L I N G 2 6
O P E R A T I O N 26
How T O USE T H E V O Y A G E R P O R T A B L E 2 6
B A T T E R Y INFORMATION . 26
C H A R G E R SPECIFICATIONS # A 8 2 0 0 2 6
C H A R G I N G T H E BATTERY WITH C H A R G E R # A 8 2 0 0 2 6
C H A R G E R SPECIFICATIONS # 7 0 0 . 1 3 5 0 0 2 6
C H A R G I N G T H E BATTERY WITH C H A R G E R # 7 0 0 . 1 3 5 0 0 2 6
MAINTENANCE 26
D A I L Y C H E C K LIST 2 6
INSPECTION A N D C L E A N I N G 2 6
S T R A P INSPECTION 2 6
H A N D L I N G A N D S T O R A G E 2 6
BATTERY R E P L A C E M E N T 2 6
C H A R G E R F U S E REPLACEMENT ( C H A R G E R # A 8 2 0 0 ONLY) 2 6
S U N G INSPECTION A N D C A R E 2 6
A N N U A L INSPECTION 2 6
MAINTENANCE I N S P E C T I O N C H E C K L I S T 26
T R O U B L E S H O O T I N G 26
A C C E S S O R I E S 26
L O G B O O K 26
W A R R A N T Y 26
User Manual - Portable Voyager 3 9
INTRODUCTION
How to use this Manual
DO NOT A T T E M P T T O U S E THIS
E Q U I P M E N T WITHOUT UNDER-
STANDING THIS M A N U A L .
To ensure safe operation, read carefully the entire manual, especially the section on
"Safety Instructions and Warnings", before installing, operating, or servicing this equipment.
If anything is not completely understood, please contact your supplier for more details. Failure to comply with warnings in this manual may result in injury.
K e e p this manual with the lift and refer to it a s required. Contents of this manual are subject to change without prior notice to u s e r s .
Throughout this manual, the following symbols may appear. Pay particular attention to the information provided under these headings. These special annotations are easily recognized as follows:
A WARNING: this symbol is intended to alert the user to hazards or unsafe practices, which could result in serious bodily harm.
INTRODUCTION
Equipment Identification
The unit's identification number (specification, model, serial number) appears on a silver nameplate attached to the back of base.
Receipt of Equipment
Upon receipt of the equipment, verify it against the packing list to ensure it is complete and inspect the equipment for possible damage due to shipping. If there is any damage, DO NOT use the equipment and notify the carrier immediately to file a claim. Provide complete information concerning damage claims or shipping errors to your supplier. Include all equipment identification numbers and group part numbers (if any) as described above along with a full description of damaged parts.
Key to Symbols
T H E F O L L O W I N G S Y M B O L S A R E U S E D
ON L I F T E R ' S ATTACHMENT L A B E L S :
This symbol is required to be displayed on regulated products for sale in the European
Market. It indicates that the product complies with applicable European
Directives related to health, safety, environment and consumer protection. o y m Safe Working Load represents the maximum load the lifter is rated for safe operation.
CAUTION: this symbol is intended to alert the user of the presence of important operating and maintenance instructions, which could prevent product damage or possible personal injury.
The following symbols are used on sling labels and related to washing instructions: v
60°c/ Maximum washing temperature 60°C
(140°F) permanent cycle.
Do not use bleach. j^><^ Do not dry clean.
N O T E : this symbol offers helpful information concerning certain operating procedures.
Q| Tumble dry low temperature.
Additional copies of this manual can be purchased by contacting your supplier. Include the User
Manual product number and equipment identification numbers.
Do not iron.
Please refer to individual sling labels for complete instructions regarding washing and drying.
User Manual - Portable Voyager 4 0
SAFETY INSTRUCTIONS AND WARNINGS
S A F E T Y INSTRUCTIONS AND WARNINGS
IMPORTANT - BEFORE USING THE PATIENT LIFTER FOR THE FIRST TIME,
PLEASE PUT THE UNIT UNDER CHARGING AT LEAST 8 HOURS TO
COMPLETELY RECHARGE BATTERIES.
A. G E N E R A L
> IMPORTANT - R E A D T H E S E INSTRUCTIONS C A R E F U L L Y OR S E R I O U S INJURY MAY R E S U L T .
> K E E P T H E S E INSTRUCTIONS AND T H E K E Y P R O V I D E D WITH T H E LIFT A T A L L TIMES.
> R E A D O P E R A T I O N AND MAINTENANCE INSTRUCTIONS IN THIS MANUAL B E F O R E
INSTALLING, O P E R A T I N G , O R S E R V I C I N G THIS EQUIPMENT.
> BHM Medical ceiling lifts are installed by an authorized contractor or installer
> Your lift is for transferring patients only. Do not u s e the lift for any other purpose.
> Always carry out the daily checklist before using the lift.
> BHM Medical ceiling lifts are specifically designed for BHM Medical ceiling rail s y s t e m s , slings and a c c e s s o r i e s . Slings and a c c e s s o r i e s designed by any other manufacturer are prohibited and will void BHM Medical's warranty. U s e only BHM Medical slings and a c c e s s o r i e s to maintain patient safety and product utility.
> BHM Medical ceiling lifts are intended to be u s e d for patients within the specified weight limit indicated for the lift. Do not attempt to lift more than the weight limit indicated.
> Before attempting to transfer, the patient must be a s s e s s e d by a qualified professional.
> BHM Medical ceiling lifts must be used by a caregiver with proper training to work with the patient to be transferred.
> O N L Y trained and qualified caregivers should transfer a patient. DO NOT attempt to u s e the lift if you have not been properly trained to do s o .
> A L W A Y S be prepared before attempting to transfer a patient.
> DO NOT use a sling that is not recommended for the lift.
> N E V E R u s e a damaged, torn or frayed sling.
> Always place the sling around the patient according to the instructions e n c l o s e d .
> F O L L O W lifting procedures outlined in this manual.
> U S E all controls and safety features only according to the rules specified in this manual.
Never attempt to force a control or button on the lift.
> DO NOT store the charger in a shower, bath or other areas with high humidity.
> DO NOT drop the patient lift or battery. Dropping the battery or lift may c a u s e internal damage that is not easily s e e n . If the lift is s u s p e c t e d to be damaged, take to an authorized technician for servicing.
> IMPORTANT: Keep all components of the lift clean and dry, and have electrical and mechanical safety checkpoints done a s instructed in the Maintenance section of this manual.
> Replace any precautionary or instruction labels that cannot be easily read.
> Avoid violent s h o c k during transportation.
User Manual - Portable Voyager 4 1
SAFETY INSTRUCTIONS AND WARNINGS
B. SHOCK PREVENTION
>
DO NOT t o u c h or use a lift w i t h bare c o n d u c t o r s o r a d a m a g e d p o w e r c o r d . Electrically live e q u i p m e n t can e l e c t r o c u t e a patient. If the lift o r c h a r g e r has a n y e x p o s e d o r d a m a g e d w i r e s c o n t a c t y o u r local dealer immediately.
DO NOT cut or remove t h e r o u n d g r o u n d i n g p r o n g f r o m a n y p l u g . A l l BHM lifts are e q u i p p e d w i t h t h r e e - p r o n g p l u g s t o protect f r o m s h o c k hazard o r e l e c t r o c u t i o n . A n y t w o - p r o n g electrical o u t l e t m u s t be replaced w i t h a p r o p e r l y g r o u n d e d t h r e e - p r o n g o u t l e t a c c o r d i n g t o the National
Electrical Code a n d local c o d e s . It is t h e r e s p o n s i b i l i t y o f t h e c u s t o m e r t o have t h e w o r k d o n e by a qualified e l e c t r i c i a n .
Do not s p l a s h o r e x p o s e electric parts o f t h e d e v i c e t o water o r m o i s t u r e .
Check nameplate f o r voltage a n d cycle r e q u i r e m e n t s . These r e q u i r e m e n t s d i f f e r b y c o u n t r y .
Do n o t a t t e m p t t o use t h e lift in an area t h a t has a d i f f e r e n t v o l t a g e a n d c y c l e r e q u i r e m e n t .
>
DO NOT a t t e m p t t o e x p o s e , service o r repair t h e lift, battery o r c h a r g e r . If any u n i t is m a l f u n c t i o n i n g , c o n t a c t y o u r local dealer.
>
READ t h e battery a n d c h a r g e r i n s t r u c t i o n s t h o r o u g h l y before u s i n g o r s t o r i n g t h e m .
C. FIRE AND EXPLOSION PREVENTION
>
Batteries m a y e x p l o d e , leak a n d c a u s e p e r s o n a l i n j u r y if n o t d i s p o s e d o f p r o p e r l y
• D o not place o r s t o r e t h e battery u n d e r d i r e c t s u n l i g h t o r near a heat s o u r c e
• Do not d i s p o s e o f in fire
• Do not s h o r t t h e battery t e r m i n a l s
• Do not incinerate
>
• F l u s h w i t h w a t e r if electrolyte (Acid) c o m e s in c o n t a c t w i t h s k i n o r eyes.
Batteries m u s t be r e c y c l e d , d i s p o s e d o f a c c o r d i n g t o local law r e g u l a t i o n s . W h e n r e t u r n i n g batteries, insulate their t e r m i n a l s w i t h a d h e s i v e tape, etc. O t h e r w i s e , t h e residual electricity in u s e d batteries may cause fire or e x p l o s i o n .
D.
>
Inspect all p r e c a u t i o n a r y labels o n the e q u i p m e n t . O r d e r a n d replace all labels t h a t c a n n o t be easily read.
User Manual - Portable Voyager 4 2
DESCRIPTION OF EQUIPMENT
DESCRIPTION OF EQUIPMENT
Parts Description
Lift Unit
The lift unit is an aluminium frame based system driven by a gear reduced high torque motor.
OPERATING FEATURES
• Ultra-light; 5 . 4 5 kg (12 pounds.) battery included.
• Electronic soft-start and soft-stop motor control.
• Easily accessible controls on both sides of the unit.
• Manual emergency lowering device (in case of electrical failure).
• Emergency brake (in case of mechanical failure). Operates on the same principle as a car safety belt mechanism.
• Quick-Release feature allows very fast winding and unwinding of the strap before and after transfer operation.
• Anti-crush feature stops the lift when there is no tension on the strap, avoiding any injury to the patient. This feature also prevents the strap from getting twisted inside the machine.
Current limiter for circuit protection in case of overload.
Low battery disconnect system to protect batteries from been drained.
Lifting speed: 4 cm/sec. (1.6 in./sec).
Strap length up to 2.2 m (88 inches). Tested for 1363.6 kg (3000 lbs.)
CSA No. 601.1, UL No. 2601-1 and CE certifications.
Respects EMI standards.
4 mm Allen key
Optional hand control
Battery
Battery eject button
Control buttons
Fuse magazine
What's included
ITEM
Voyager lift
Battery
Charger
Power cord
4 mm Allen key
User Manual
QTY
1
1
1
1
1
1
Dimension
Power cord
1 * 0 m m ( 5 . 5 " ) »-
4 0 4 m m ( 1 5 . 9 " )
4 4 m m ( 1 7 . 5 " ) -
1 3 2 m m ( 5 . 2 " ) ~
User Manual - Portable Voyager 4 3
SLINGS
BHM E R G O F I T SLING
Every patient has different needs. It is very important to choose the right sling based on the needs of the patient to be transferred as well as their physical ability and size.
All BHM slings are made from a Polyester/ Nylon material, which is soft and pliable, easy to care for, and install. All slings are machine wash and dry in medium temperature. BHM slings are covered by a one-year warranty.
> U S E A S L I N G R E C O M M E N D E D B Y
T H E P A T I E N T ' S PHYSICIAN, N U R S E
OR M E D I C A L S P E C I A L I S T A S A P -
P R O P R I A T E F O R T H E P A T I E N T ' S
ABILITY AND N E E D S .
DO NOT use any kind of plastic backed incontinence pad or seating c u s h i o n between the patient and sling material.
That may c a u s e the patient to slide out of the sling during a transfer.
DO NOT exceed the maximum weight limitation of the sling. A P L E A S E R E A D T H E F O L L O W I N G
C A R E F U L L Y , B E F O R E USING A
BHM M E D I C A L E R G O F I T S L I N G .
> A L W A Y S C A R R Y O U T T H E DAILY
C H E C K L I S T B E F O R E E A C H L I F T U S E .
> WHENEVER POSSIBLE, HAVE
SOMEONE ASSIST YOU TO ENSURE
THE P A T I E N T S SAFETY AT ALL
TIMES.
> BHM M E D I C A L C E I L I N G L I F T S A R E
S P E C I F I C A L L Y D E S I G N E D F O R BHM
M E D I C A L S L I N G S AND A C C E S -
S O R I E S . S L I N G S AND A C C E S S O R I E S
D E S I G N E D B Y ANY O T H E R
M A N U F A C T U R E R A R E P R O H I B I T E D
AND WILL VOID BHM M E D I C A L ' S
W A R R A N T Y . U S E O N L Y BHM
M E D I C A L S L I N G S AND
A C C E S S O R I E S T O MAINTAIN
PATIENT S A F E T Y AND P R O D U C T
UTILITY.
Laundry Instructions for all slings:
The following symbols are used on sling labels and related to washing instructions:
\60°c/ Maximum washing temperature 60°C
(140°F) permanent cycle.
Do not use bleach.
Do not dry clean.
> PATIENT T O B E T R A N S F E R R E D
MUST B E A S S E S S E D B Y A
Q U A L I F I E D P R O F E S S I O N A L B E F O R E
ATTEMPTING T O T R A N S F E R .
OJ Tumble dry low temperature.
? > x j Do not iron.
Observe a trained expert to perform the transfer procedures and then practice them several times with a capable individual acting a s patient before attempting a transfer with a patient with limited capabilities.
User instructions for:
Hammock Sling
Inspect e a c h sling for wear, tears or loose stitching after e a c h w a s h .
> R E A D C A R E F U L L Y A L L
INSTRUCTIONS B E L O W B E F O R E
ATTEMPTING T O I N S T A L L A S L I N G .
H a m m o c k S l i n g ( T H A - S , M & L )
> IF Y O U H A V E ANY D O U B T S , P L E A S E
C O N T A C T Y O U R L O C A L
R E P R E S E N T A T I V E .
Before transferring a patient from a stationary object (wheelchair, bath or bed), slightly raise the patient off the stationary object and check that all sling attachments are s e c u r e . If any attachment is not correct, lower the patient and correct the problem, then raise the patient and check again.
Our most universal sling, the Hammock sling, is a mesh type material that dries quickly but is soft and pliable. The sling has an incorporated head support and a unique strap system that keeps the patient secure in the sling. A special polyester fabric in the seat area helps to reduce friction and makes the sling easy to place around a patient in a chair without lifting or remove the patient. The patient may be transferred in a seated, semi-reclined or fully reclined position.
Best choice for most situations - including bathing. Fits all BHM ceiling lift models as well
User Manual - Portable Voyager 44
SLINGS
as the Ergolift Mobile Lift. Available in sizes S, M
& L. Weight capacity is 250 kg (550 lbs.)
Hammock 6 Strap Sling (THA6-S, M & L)
Almost identical to the Hammock sling above, this sling provides extra security and a tighter fit for patients who are low-tone or who would prefer a more secure feeling. The head support is tighter and the extra straps on the side provide more lateral support. The patient may be transferred in a seated, semi-reclined or fully reclined position. Fits all BHM ceiling lift models as well as the Ergolift mobile lift. Available in sizes S, M & L. Weight capacity is 250 kg (550 lbs.).
(U-shape and white fabric)
5. Lean the patient forward in the chair and make sure that the patient is supported.
Whenever possible have someone a s s i s t you to e n s u r e the patient's safety at all times.
^
With the white fabric of the sling facing the patient, grab the sling at the top of the Ushape and slide the sling down to the base of the spine or coccyx.
^ ^ ^ ^
* ]
•
SLING SIZES AND PRODUCT NUMBERS:
SMALL MEDIUM LARGE
Approximate weight:
Approximate height:
Color Code
(45-100 lbs.)
120-150 cm (4-
4'11"')
Red
THA-S
(100-210 lbs.)
(210-550 lbs.)
151-180 cm 181 cm +
(5'-5'11") (6' +)
Yellow
THA-M
Green
THA-L Hammock
Product #
Hammock 6
Product #
THA6-S THA6-M THA6-L
Sling Placement
Wheelchair
2.
from Chair or
Sling has two sides, the white fabric indicates the side that is intended to be next to the patient and the label and colored handles are on the outside. The top of the sling is slightly rounded and the bottom of the sling has a U-shaped design.
Tuck the sling down on each side to make sure that it is just slightly under the patient's buttock. Check for the stitching of the sling handle at the center to make sure that it is in line with the patient's spine and bring the patient to the back of the chair or wheelchair.
User Manual - Portable Voyager 4 5
SLINGS
Sling Attachment Options
©pOfleoofl © p t a g i p t a i S
Option 1 - Bridge
(Most recommended)
d) Take the two straps between the legs these are the short loops - and thread one loop through the other to cross them.
7. Kneel in front of the patient so that your back is straight and so you are at eye level with the patient. Grab the end of the sling with the white fabric and make sure there are no twists in the sling. Make sure the white fabric is facing the patient's leg. Place one hand on the patient's knee to hold it in place and gently pull the sling forward with the other hand until the edge of the sling is approximately even with the patient's knee.
8. Place the patient's foot on your thigh and slide the white section of the sling under the leg. e) Now take the long strap on the outside of each leg and string it through the inside strap. By crossing the straps as you see here, form a bridge above the patient's legs and protect them from any pinching or pressure, a s well as keep the knees and hips in line.
9. Repeat steps 5 & 6 f o r the other leg. f) Finally, place the patient's hands inside the sling.
User Manual - Portable Voyager 46
SLINGS
OPTION 2 - Leg Separation
A Using this method of sling placement will pull the patient's legs apart. C h e c k with the patient's physician, nurse or medical specialist before attempting this method. c) Thread the long strap through the short strap on each leg. d) Each short strap should be looped through the long strap on the same leg.
OPTION 3 - Amputee
Do not u s e this method of sling placement for a patient who is agitated or spastic. d) Slide the sling end with the white fabric so that it is all the way under both legs. One sling strap will be on top of the other. d) Place the hands inside the sling. See sling placement for leg separation as installed below.
User Manual - Portable Voyager e) Thread the long strap through the short strap that is next to it. The photo below shows clearly an example without a patient in it.
47
SLINGS
f) When installed, there should not be any sling material between the patient's legs. Place the hands inside the sling.
Sling Placement- from lying position
5. Fold the sling in half lengthwise with the white part of the sling on the inside of the fold and place next to the patient to be transferred.
7. Lower the patient down on top of the sling.
Roll the patient the opposite direction to pull the sling out the other side.
When transferring a patient from
AjyK a bed, have people on each side
/ \ J r \ of it, or raise the bed rails to ensure the patient does not roll off the bed.
6. With someone there to assist you, carefully roll the patient onto one side. Now place the fold of the sling at the center of the bed or where the spine of the patient will be. The top of the U-shape in the sling should be at the base of the spine or coccyx .The top of the sling should be next to the head. Lift the top fold and tuck next to the patient.
User Manual - Portable Voyager
4. Place the straps in the same way as shown from a seated position for the Hammock sling.
4 8
SLINGS
Sling attachment to Carry Bar
9. Ensure the sling is properly placed around the patient according to the instructions above.
3. Choose the color of the loop you will need for the position you will be placing the patient in. If transferring to a seated position, use the loops indicating it.
Sling position
T T
BLACK
B L U E
G R E Y
B l U C
B L U E B L U E
B L U E G R E Y
User Instructions for:
Quick-Fit Sling
Specially designed for patients who are stocky or obese, especially in the midsection. Padding around the buttock and leg area helps to provide extra support. Upper portion of the sling is a mesh type material that dries quickly but is soft and pliable. This sling does not provide head support. The patient may be lifted in a seated, semi-reclined or fully reclined position. Fits all
BHM ceiling lift models as well as the Ergolift mobile lift. Available in sizes S, M & L. Weight capacity is 250 kg (550 lbs.)
Do not use the Quick-Fit sling on a patient who cannot support his or her own head.
A
Always u s e the same color loop on the rear shoulder straps for both sides of the sling.
Use the same color loop on the leg straps for both sides of the sling.
Use the same color loop on the hip straps for both sides of the
Hammock 6-strap sling.
5. Ensure the patient's arms are inside the sling and that it is not caught on any obstructions (e.g. wheelchair brake or handle).
6. See Lift Instructions for next step.
SLING SIZES AND PRODUCT NUMBERS:
SMALL MEDIUM LARGE
Approximate weight:
Approximate height:
Color Code
Quick-fit
(45-100 lbs.)
120-150 cm (4-
4'11-)
Red
TIR-S
(100-210 lbs.)
151-180 cm (5-
5*11")
Yellow
TIR-M
(210-550 lbs.)
181 cm +
(6' +)
Green
TIR-L
Sling Placement
Wheelchair from Chair or
Whenever possible have someone a s s i s t you to ensure the patient's safety at all times.
Sling has two sides, the white fabric indicates the side that is intended to be next to the patient, and the label and colored handles are on the outside. The top of the sling is straight and the bottom of the sling has a U-shaped design.
User Manual - Portable Voyager 49
SLINGS
1 0 . Lean patient forward in the chair and make sure the patient is supported.
11. With the white fabric of the sling facing the patient, grab the sling at the top of the Ushape and slide it down to the base of the spine or coccyx. Tuck the sling down on each side to make sure that it is just slightly under the patient's buttock. Check the seam for the sling handle at the center to make sure that it is in line with the patient's spine and bring the patient to the back of the chair.
1 2 . Kneel in front of the patient so that your back is straight and so you are at eye level with the patient. Grab the end of the sling with the white fabric and make sure there are no twists in the sling. The white fabric should be facing the patient's leg. Place one hand on the patient's knee to hold it in place and gently pull the sling forward with the other hand until the edge of the sling is approximately even with the patient's knee.
15. By crossing the straps as you see here, we keep the knees and hips in line.
16. Place the patient's hands inside the sling before attaching it to the lift.
Sling attachment to Carry Bar
5. Ensure the sling is properly placed around the patient according to the instructions above.
6. Choose the color of the loop you will need for the position you will be placing the patient in. If transferring to a seated position, use the loops indicated for a seated position.
Sling position
1 4 . Take the two straps between the legs.
There is a small loop on each strap right next to the sling. Thread one strap through the other to cross them.
User Manual - Portable Voyager
Bl ACK B L U E
G R C Y • L U C
B L U L
Always use the same color loop on the rear shoulder straps for both sides of the sling.
Use the same color loop on the
5 0
SLINGS
the sling has two long straps and the inside of the sling has a padded section with a belt. hip straps for both sides of the sling.
7. Ensure the patient's arms are inside the sling and that it is not caught on any obstructions (e.g. wheelchair brake or handle)
8. See Lift Instructions for next step.
User Instructions for:
Hygienic Sling
This sling is excellent for changing incontinence pads or to transfer onto a toilet as it provides an open area from the middle of the back to the middle of the thigh. The padding wraps around the patient's midsection and under each leg. jn order to use this sling, the patient must have good muscle-tone in their shoulders and upper body. Fits all BHM ceiling lift models as well as the Ergolift mobile lift and Ergostand standassist. Available in sizes S, M & L. Weight capacity is 250 kg (550 lbs.)
A Do not u s e Hygienic sling on a patient who does not have good muscle-tone in their shoulders and neck.
C h e c k with the patient's physician, nurse or medical specialist before using this sling.
/ \ Whenever possible have
/ I \ someone a s s i s t you to ensure the
^ — - — ^ patient's safety at all times.
5. Lean patient forward in the chair and make sure the patient is supported.
6. With the white fabric of the sling facing the patient, place it approximately halfway down the back, or so the sling is below the shoulder blades. Lean the patient against the back of the chair.
SLING SIZES AND PRODUCT NUMBERS
SMALL MEDIUM LARGE
CHEST SIZE 68-89 cm 89-107 cm 107-125 cm
(27-35") (35-42") (42-49")
PRODUCT # THY-S
COLOR
CODE
Red
THY-M
Yellow
THY-L
Green
4. Sling has two sides, the white fabric indicates the side that is intended to be next to the patient, and the label and colored handles are on the outside. The bottom of
User Manual - Portable Voyager
7. Fold the padded fabric waistband with the belt attached across the mid-section of the patient and clip the belt closed. Make sure the belt is secure.
5 1
8. Tighten the belt until snug by pulling gently on the end of the belt.
9. Kneel in front of the patient so that your back is straight and so you are at eye level with the patient. Grab the end of the sling with the white fabric and make sure there are no twists in the sling. Make sure the white fabric is facing the patient's leg. Place the patient's foot on your thigh and slide the white section of the sling under the leg.
8. The hygienic straps need not cross to allow better access to the toileting area. With the hygienic sling, the arms of the patient are always over top of the sling.
Sling attachment to Carry Bar
5. Ensure the sling is property placed around the patient according to the instructions above.
6. Choose the color of the loop for the position that is best suited for the patient. For a larger patient, choose blue for both the leg and shoulder straps. A smaller patient may be better suited for the black loops on both the shoulder and leg straps.
SLINGS
Sling position tr=7f 1 1
B L U K L 3 L U L
C R F Y
G R F " v
Ik
BLACK BLACK
Always use the s a m e color loop on the shoulder straps for both s i d e s of the sling.
U s e the s a m e color loop on the leg straps for both s i d e s of the sling.
7. Ensure the patient's arms are over top of the sling and that it is not caught on any obstructions (e.g. wheelchair brake or handle).
8. See Lift Instructions for next step.
User Instructions for.
Walking S l i n g
Designed to assist people with limited ability to walk, this sling helps support and assist in rehabilitation. Fully padded with a security buckle, the sling is comfortable and simple to use. The user must have some weight bearing ability. Fits all BHM ceiling lift models as well as the Ergolift mobile lift. Available in sizes S, M &
L. Weight capacity is 250 kg (550 lbs.)
A Do not u s e the Walking sling on a patient who d o e s not have weight bearing ability and good muscletone in their shoulders and neck.
C h e c k with the patient's physician, nurse or medical specialist before using this sling.
User Manual - Portable Voyager 5 2
SLINGS
SLING SIZES AND PRODUCT NUMBERS
SMALL MEDIUM LARGE
CHEST SIZE 68-89 cm 89-107 cm 107-125 cm
(27-35") (35-42") (42-49")
TEM-M TEM-L PRODUCT # TEM-S
COLOR
CODE
Red Yellow Green
10. Sling has two sides, the white fabric indicates the side that is intended to be next to the patient, the label and colored handles are on the outside. The bottom of the sling has two long straps and the inside of the sling has a padded waist section with a belt.
14. Tighten the belt until snug by pulling gently on the end of it.
Whenever possible, have someone a s s i s t you to ensure the patient's safety at all times.
11. Lean patient forward in the chair and make sure the patient is supported.
12. With the white fabric of the sling facing the patient, place the sling approximately halfway down the back, or so it is below the shoulder blades. Lean the patient against the back of the chair.
15. Kneel in front of the patient so that your back is straight and so you are at eye level with the patient. Grab the end of the sling with the white fabric and make sure there are no twists in the sling. The white fabric should be facing the patient's leg. Place the patient's foot on your thigh and slide the white section of the sling under the leg.
13. Fold the padded fabric waistband with the belt attached across the mid-section of the patient and clip the belt closed. Make sure the belt is secure.
16. The straps that come from between the legs do not attach to the lift. Thread the strap at the chest through the longest loop on the same side.
User Manual - Portable Voyager 53
18. Note that with the walking sling, the arms of the patient are always over top of the sling.
Sling attachment to Carry Bar
1. Ensure the sling is properly placed around the patient according to the instructions above.
2. Adjusting the sling loops higher or lower in either the front or the back allows you to tilt the body either direction to correct posture.
Sl
;
ng position
if
F R O N T B A C K
0 S A V f COLOR TOR
A L L 1 S I R A :
J
S
1
- J L U HI. A C K
BLAC-< B L U L
SLINGS
Always u s e the s a m e color loop on the rear shoulder straps for both sides of the sling.
Use the s a m e color loop on the chest straps for both sides of the sling.
3. Ensure the patient's arms are over top of the sling and that it is not caught on any obstructions (e.g. wheelchair brake or handle).
4. See Lift Instructions for next step.
5. When raising the patient in the lift, check to make sure the sling sections between the legs are providing snug support. If there is not enough support, reinstall the sling threading the chest straps through a lower loop (grey or black).
Repeat this step until the lower portion of the sling provides proper support.
OPERATION
How to use the Voyager
Portable
A R E A D " S A F E T Y INSTRUCTIONS
/ | \ AND WARNINGS" B E F O R E
/ • \ ATTEMPTING T O U S E T H E
V O Y A G E R LIFT.
Always carry out the daily checklist before each lift u s e .
Damage to the lift may occur if not properly handled. Carry the
V O Y A G E R P O R T A B L E by its handle only. Do not carry the lift with the reacher a c c e s s o r y or other a c c e s s o r y not intended to be used to carry the lift.
15. Install the patient sling (see sling installation)
16. Take the battery out of the charger and insert it into the lift. Make sure the silver contact points go inside the lift.
17. Once the patient sling is installed, unwind the strap.
User Manual - Portable Voyager 54
There are two ways to unwind the strap:
First, you can press the button while holding tension on the strap with the other hand (there must be tension on the strap for the lift to function).
The other way to do this is to use the "QUICK-
RELEASE"; Push the knob
(1) and pull the strap out of the lift (2).
A P u s h knob completely over to ensure the Q U I C K - R E L E A S E will work properly.
Q U I C K - R E L E A S E will not work if there is tension on the strap.
Make sure the strap h a s some slack in it before attempting to slide the knob over.
DO NOT force the knob. It should slide easily.
18. Attach the round hook to the trolley in the track.
19. Move the lift over to the patient.
Hold the lift with one hand at all times when near a patient.
20. Lower the lift below the chin before beginning to attach the sling straps (this is to avoid the risk of facial contact caused by movement of the lift).
For a patient in a lying position, lower the lift near the thorax, and then install the straps.
2 1 . Attach the straps of the sling as described in the sling attachment instructions for the sling to be used. (See sling installation)
22. Lift the patient slightly by pressing the control button ("#"]
User Manual - Portable Voyager
OPERATION
B E F O R E LIFTING T H E PATIENT:
4. Make sure that all straps are attached to the carry bar.
5. Make sure the patient's arms are safely out of the way.
6. Make sure the sling is not caught on an obstruction
(wheelchair brake or arm of chair).
If any of the above occur - lower the patient immediately and correct the problem.
23. Lift until the patient's buttock clears the arm supports or the top of the bath or bed before moving the patient. Guide the legs past any obstacle.
24. When the patient is located above the desired point of transfer and ready to be lowered, press thef 4 button.
25. Use the handles on the back of the sling to position the patient when transferring into a chair. Hold the handles firmly as the sling lowers and the sling will tilt back to position the patient.
26. Once the patient is properly seated and the straps are loose, remove the sling from the lift.
A
Hold the lift with one hand at all times when near a patient.
27. Slide the lift away from the patient.
28. Unhook the lift from the trolley in the rail and use the "QUICK RELEASE" to shorten the strap. To do so, push the knob (1) and hold the strap while it winds itself into the lift.
P u s h knob completely over to ensure the " Q U I C K - R E L E A S E ' will work properly.
Q U I C K - R E L E A S E " will not work there is tension on the strap.
Make sure the strap h a s some slack in it before attempting to slide the knob over.
5 5
OPERATION
DO NOT force the knob, easily.
It should slide
A When using the "QUICK-
R E L E A S E " hold the strap in your hand and release the knob (1) only when the strap is not going in motion. This will prevent wear of the internal components.
15. Sling may be removed from around the patient.
Emergency Lowering Feature
The Voyager Portable has an emergency manual-lowering feature in the event of an electrical or functional failure.
If the lift malfunctions when a patient is being transferred, the emergency manual lowering device provides a safe way of lowering the patient onto a chair or bed.
To operate the emergency lowering feature:
6. Find the rubber stopper and label on the side of the lift.
7. Pull the rubber stopper out of the Voyager
Portable casing - DO NOT DISCARD.
8. Move the patient over a bed or chair.
9. Using the Allen key provided with this manual, turn the key clockwise to lower the patient into the chair. As the key turns the motor directly, each key turn will move the patient slightly.
User Manual - Portable Voyager
. . „ I - . , .
.. . , ... j
To lower the patient more quickly, use a power drill with a 4 mm Allen key bit.
10. Once the patient is lowered safely into a chair or bed, take the lift to a certified technician to be serviced.
Battery Information
A DO NOT D R O P B A T T E R Y . A
D R O P P E D B A T T E R Y MAY H A V E
INTERNAL D A M A G E THAT IS
NOT V I S I B L E . IT MAY L O S E ITS
ABILITY T O R E C H A R G E .
Due to internal damage c a u s e d when the battery is dropped, the battery may indicate that it is fully charged, when in fact the battery cannot hold a full charge.
A green light indicated on the charger may not accurately reflect a full charge on a damaged battery.
N E V E R attempt to u s e a battery that appears to be damaged or dented.
DO NOT store battery near a heat source.
Heat will damage the battery.
Always keep metal battery contacts clean.
24 Vdc, 2.0 Ah interchangeable and rechargeable battery.
Provides 10 to 30 transfers per battery charge (may vary depending on patient weight and battery condition. See graph below).
56
Life cycle (number of charging cycles) of the battery is largely dependent on the depth of discharge in each cycle. The more the battery is drained, the shorter its life span. The life of the battery is also related to such factors as varying temperatures and rest periods between charge and discharge.
Graph 1: Number of recharge v s . Depth of discharge
GRAPH #1 illustrates the relationship between discharging depth and expected battery life.
If you use only 3 0 % of the battery power between each charge, you can expect the battery to fully charge 1050 times.
If you drain the battery until it beeps every time, you can expect the battery to fully charge only
300 times.
To prolong battery life - Put the battery into the charger
W H E N E V E R T H E LIFT IS NOT IN
U S E .
The battery will not overcharge.
BHM u s e s sealed lead-acid batteries.
Contrary to nickel-cadmium, BHM batteries do not have any memory effect.
Therefore, batteries should not be completely drained before recharge.
Graph 2: Number of lifts v s . Lift load
O P E R A T I O N number of lifts that can be done with one recharged battery.
Do not drain the battery
p\ excessively. This will
**" ^ dramatically reduce the battery life span.
If the low battery buzzer s o u n d s , be sure to recharge the battery a s soon a s possible.
Charger Specifications #A8200
• Hospital grade charger cord and plug.
• Charger input: 120 Vac, 60 Hz, 0.5 A. (North
America)/ 240 Vac, 50 Hz, 0.25 A European.
Charging the Battery with Charger
#A8200
I
DO NOT O P E R A T E T H E
C H A R G E R UNIT WITH A
DAMAGED C O R D O R IF T H E
UNIT HAS B E E N D R O P P E D O R
DAMAGED.
DO NOT forcibly bend the power cord or place a heavy object on it. This will damage the cord and may c a u s e fire or electrical shock.
DO NOT pour liquid on or near the charger.
A DO NOT place the unit in locations that are:
Extremely hot
• Dusty or dirty
• Very humid
• Moving or vibrating
Do not attempt to put a damaged battery into a charger. Inspect the battery first for dents or damage.
The recharging steps should be as follows:
(See step by step instructions)
The graph above (GRAPH #2) illustrates the relationship between the load lifted and the
User Manual - Portable Voyager 5 7
O P E R A T I O N on the charger will turn off, or will not go on.
This is only the case when the battery is
99.9% charged.
• To test the charger, use the battery in the lift a few times then put the battery in the charger. If the light sill does not go on, see the "Troubleshooting" section.
Charger Specifications #700.13500
Eject the battery from the lift unit by using the "Battery Eject Button" (1). Keep your hand near the opening to catch the battery
(2).
Keep one hand near the opening for the battery (2) before pushing eject button. Do not allow the battery to eject and fall.
Insert the battery into the charger (3) with the silver contact points of the battery going into the charger.
The charging indicator light will illuminate red to white to green depending on the battery drain. If the light does not go on, check the
"Troubleshooting" section in this manual for assistance.
If the battery is low, the light will show red or orange when you place the battery in the charger. The light will gradually turn to a white or yellow color, then to green when fully charged.
Whenever possible, leave the battery in the charger when the lift is not in use.
At minimum, charge the battery until the light is green before using again.
This will extend the life of the battery.
When the lift is needed again, pull the battery out of the charger, and gently slide it back into the Voyager Portable lift. The silver contacts should be inside the lift.
The power cord can remain plugged into the wall indefinitely, convenient for overnight recharging.
If the battery has been in the charger for an extended period of time or if the battery has not been drained: it is possible that the light
User Manual - Portable Voyager
Class 2 power cord rated at 2.5 A (100-240
Vac 50-60 Hz).
Charger input: 100-240 Vac, 50-60 Hz, 50VA
Charger output 27 Vdc 1 A.
Input fuse - 1A 250 V located inside the charger.
Green light showing power to the charger
Solid yellow light showing battery 8 5 % charge
Yellow light blinking showing battery under charge
FULL CHARGE
INPUT = 1 0 0 - 2 4 0 V~, 50/60 Hz, 50VA
O U T P U T = 27 V r— , 1.0 A l.R 101445 ( | | »
ITEM : »700.13500 daw ? « )
V M » W M C n > M 0 t v a H M M f O I C M I N C . MAC* m CMiHOA
Charging the Battery with Charger #
700.13500
(Available Oct. 2001)
Check to make sure the green light on the charger is illuminated. This indicates that the charger is on.
5 8
• Insert the battery into the charger with the silver contact points of the battery going into the charger.
• The charging indicator light will flash yellow while the battery is charging. If the battery is approximately within 85% of being fully charged, the charging indicator light will illuminate solid yellow.
• If the battery is low, it will take approximately
2 hours for a full charge.
' Whenever possible, leave the A battery in the charger when the lift is not in u s e .
A At minimum, charge the battery until the light is solid yellow (not flashing) before using again.
This will extend the life of the battery.
• When the lift is needed again, pull the battery out of the charger and, gently slide it back into the Voyager portable lift. The silver contacts should be inside the lift.
• The power cord can remain plugged into the wall indefinitely, convenient for overnight recharging.
• The battery can be left in the charger for an extended period of time without damaging either the charger or battery.
• If the green light does not go on when the charger is plugged into an electrical outlet, see "Troubleshooting"
MAINTENANCE
A L W A Y S C A R R Y OUT T H E DAILY
C H E C K L I S T B E F O R E E A C H LIFT
USE.
DO NOT attempt to open the V O Y A G E R
P O R T A B L E . Only a certified technician is authorized to open the V O Y A G E R
P O R T A B L E .
Alterations made to the V O Y A G E R
P O R T A B L E by someone other than a certified technician may c a u s e serious injury.
V O Y A G E R P O R T A B L E and a c c e s s o r i e s must be inspected A N N U A L L Y by a certified technician in addition to the daily and other periodic visual c h e c k s done by the user specified in this section.
MAINTENANCE
Preventive m a i n t e n a n c e s p e c i f i e d in t h i s m a n u a l can prevent a c c i d e n t s a n d reduce repair c o s t s .
Note all service or repairs t o the VOYAGER or its a c c e s s o r i e s in t h e l o g b o o k at t h e e n d of t h i s m a n u a l . Have t h e d o c u m e n t s i g n e d by a certified t e c h n i c i a n
Daily Check List
T H E F O L L O W I N G P R O C E D U R E M U S T B E
F O L L O W E D B E F O R E E A C H U S E .
• Has the battery been charged? Charge the battery whenever the lift is not in use.
• Inspect the battery before placing it in the
Voyager portable. If the battery appears to have any damage - DO NOT USE IT.
Replace the battery immediately (Product #
A8500)
• Inspect the lift for any damage. If the lift casing does not look properly aligned, or there are any cracks or other damage on the lift, or any parts are missing - DO NOT USE
IT. Contact your local representative to have the lift serviced.
• Attach the lift to the trolley in the rail. Inspect the visible strap for any signs of wear, frays, loose threads or other damage. If there is any evidence of damage- DO NOT USE IT.
Contact your local representative to have the lift serviced.
• Inspect the sling for tears, frayed straps or loose stitching. If the sling has any of the above damage - D O NOT USE IT. Contact your local representative to have the sling replaced or repaired.
• Inspect the reacher and adjustable extension strap (if you have these options) for wear.
Check the adjustable extension strap for frays or loose stitching. Check the reacher for damage in the metal hook or handle. DO
NOT USE if there is any damage.
• Inspect the hook at the top of the strap to ensure it is properly attached and there is no damage to the hook.
Inspection and Cleaning
Clean the voyager body with a dry soft cloth, or a soft cloth lightly moistened with a mild detergent
User Manual - Portable Voyager 59
solution. Do not use any type of solvent that may damage the finish.
MAINTENANCE
and patient safety is greatly compromised.
| \ DO NOT i m m e r s e lift in water.
To ensure a better rolling surface for the trolley wheels, clean the inside of the track every 4 months. To do so, insert a damp cloth in the opening and slide it from one end of the track to the other.
In any c a s e , manufacturer recommends to change strap at least every two years. By continuing to u s e the lift without changing the strap, caregiver and patient safety is greatly compromised.
Take note of results of inspection in the log book at the end of this manual.
Always replace rail end stopper
(if they have been removed) after servicing.
Handling and Storage
Strap Inspection
This equipment is built with a strap that can withstand a load of 1 3 6 4 kg ( 3 0 0 0 lbs.)
Nonetheless, if this strap is damaged or shows signs of wear, the acceptable load on the strap before rupture can drop rapidly and present a danger for the patient or caregiver.
BHM Medical recommends thoroughly inspecting the strap every 2 months as follows:
Completely unwind the strap.
Look for any signs of wear.
Avoid violent shock during transportation.
The lift should not remain stored for long periods of time without recharging the batteries.
BHM Medical recommends charging of batteries at least every two weeks even if the lift is not used. T h i s will prevent premature aging of batteries.
Loose threads in stitched area.
—Noticeable discoloration by having a lighter color strap than the double thickness stitched area.
Side wear. iddle wear.
Battery Replacement
BHM uses sealed lead-acid batteries. Contrary to nickel-cadmium, BHM batteries do not have any memory effect. Therefore, batteries should not be completely drained before recharge.
Replace the battery when it has been dropped or if there is a noticeable reduction in the number of transfers that can be performed between charges. If you notice the Voyager Portable
"beeping' - see the instructions in
"Troubleshooting" section to ensure it is a problem with the battery.
To replace a battery, contact your local dealer or representative and order product # A 8 5 0 0 .
DO NOT A T T E M P T T O U S E A
B A T T E R Y NOT A U T H O R I Z E D B Y
BHM MEDICAL.
If there are any signs of wear a s indicated here or other visual defects, strap should be changed immediately. By continuing to u s e the lift without changing the strap, caregiver
BHM batteries are specially designed for BHM charging s y s t e m s . Attempting to u s e an unauthorized battery may seriously damage the lift and/or the charger.
User Manual - Portable Voyager 6 0
Charger Fuse Replacement
(Charger # A 8 2 0 0 only)
If the charger light does not illuminate when there is a battery in it, do the following steps:
6. Make sure the battery is properly placed in the charger. The silver contact points should be inside the charger.
7. Check to make sure the power cord is correctly plugged in to the charger and the wall.
8 . If the light still doesn't work, plug the charger into a different outlet in a different area to ensure that the outlet is working properly.
9. Pull the power cord out of the charger.
Under the socket on the charger, there is a drawer or magazine where there is a fuse.
Using a pen or screwdriver, pull the drawer out gently. Pull the burnt fuse out of the drawer that is next to the charger (see diagram below) and discard. Take the replacement fuse and place it in the same slot where the burnt fuse was. Slide the drawer back into the charger. Plug the power cord into the charger. Place the battery into the charger and check for the light.
1 0 . If the light does not illuminate, contact your local dealer or representative for assistance.
R e p l a c e m e n t f u s e
MAINTENANCE
Sling Inspection and Care
For maximum patient safety and hygiene, read the following instructions:
Chemical Attack
Oil, grease, creosote or paint stains are harmless but other forms of chemical attack of a sufficient degree may lead to deterioration or extreme softening of the webbing, which would lead to the fibres being rubbed off (almost as powder in extreme cases). It is best to avoid fumes, spray or mists of acid and alkalis or organic solvents. If contamination is suspected, wash out well in warm water. Avoid contact with excessive heat, which is likely to affect the product.
Cleaning
Certain chemical substances mentioned above may be removed by diluted heavy duty cleaner and warm water. Ensure that any cleaning agents are removed by rinsing thoroughly in warm water, maximum 60°C (140°F), and allowed to dry naturally away from any source of direct heat.
For normal washing and cleaning, follow the care instructions on the label of the product.
Storage
Slings should be stored away from direct sunlight, stress, pressure, excessive heat or humidity. The sling should be kept away from contact with sharp objects, corrosives or other possible damaging material.
B u r n e d f u s e
M a g a z i n e
Loose threads in stitched area
Noticeable discoloration by having lighter color strap than the double thickness stitched area ide wear.
Replace the fuse with the following:
Fuse: GMC 0.5 A 250 V North America (When input is 120 Vac) Product # C8FGMC0.5
GMC 0.25 A 250 V European (When input is 240 Vac) Product # C8FGMC0.250
The fuse may also be ordered through your local dealer or representative.
Check for any sign of wear on sling straps as described above and at the strap-sling junction. Buckles should be checked to ensure that the locking/unlocking mechanism opens and closes easily.
User Manual - Portable Voyager
6 1
8. Check for any sign of deterioration of the sling material. Check for general external wear, local abrasion, cuts, holes or burns in the webbing and for chemical attack.
9. Look at "stitched areas" for broken, worn, pulled or unravelled stitches.
10. Sling should be cleaned as indicated on the tag.
If there is any sign of wear or deterioration as indicated above or other visual defect, sling should be changed immediately. By continuing to use the lift without changing the sling, caregiver and patient safety is greatly compromised.
Annual Inspection
The Voyager Portable and its accessories must be inspected annually by a certified technician.
V O Y A G E R P O R T A B L E AND
A C C E S S O R I E S MUST B E
S E R V I C E D E V E R Y 12 MONTHS A S
X MINIMUM R E Q U I R E M E N T .
Do not attempt to do the inspection unless you are certified to do s o .
A s part of the annual inspection, an annual load test with the safe working load must be performed on the V O Y A G E R P O R T A B L E a s required for C E MARK MAINTENANCE.
After annual inspection, have the log book signed by the certified technician.
MAINTENANCE
User Manual - Portable Voyager 62
MAINTENANCE INSPECTION CHECKLIST
Maintenance Inspection Checklist
T O B E I N S P E C T E D B Y U S E R
ITEM INITIALLY B E F O R E
E V E R Y
U S E
L I F T D E V I C E
Inspect for missing hardware or broken panel. X X
X
X
Inspect strap for wear.
Battery recharge.
Inspect round hook on the strap of the lift for damage or cracks.
Inspect trolley in rail for damage, rust or cracks.
Replace if damaged.
Overall inspection by authorized personnel
S L I N G AND H A R D W A R E
Check all sling attachments for sign of wear.
Inspect sling material for wear or deterioration.
Inspect sling straps for wear.
Inspect for any defect or loose threads in the
"stitched areas".
Cleaning sling as indicated on the tag.
A C C E S S O R I E S
Verify strap extension for wear
Verify arm extension for wear
X
X
X
X
When Necessary
E V E R Y
2
MONTHS
E V E R Y
6
MONTHS
X
X
X
E V E R Y
Y E A R
X
X
X
X
E V E R Y
2 Y E A R S
MAINTENANCE B Y A C E R T I F I E D T E C H N I C I A N
Replace strap.
Inspect the welding aspect on the frame to detect cracks.
Inspect connecting joints for proper attachment.
Verify that the emergency brake on the drum is turning freeiy.
Verify emergency devices for good functioning.
X
X
X
X
X
A Maintenance inspection points are suggested by the manufacturer, however in certain cases,
/ 1 \ some inspections should be done on a more frequent basis. By continuing to use the lift
/ • \ without replacing the listed parts in this maintenance inspection checklist, caregiver and patient safety is greatly compromised.
User Manual - Portable Voyager 6 3
TROUBLESHOOTING
TROUBLESHOOTING
DO NOT a t t e m p t t o o p e n t h e VOYAGER P O R T A B L E , Only a c e r t i f i e d t e c h n i c i a n is a u t h o r i z e d t o o p e n the VOYAGER P O R T A B L E .
A l t e r a t i o n s m a d e t o t h e VOYAGER P O R T A B L E by s o m e o n e o t h e r t h a n a certified t e c h n i c i a n may c a u s e serious injury.
PROBLEMS TO CHECK
The unit does not work when you press • Make sure the hand control is plugged correctly into the the ( j T j or ( # ) buttons on the hand control. unit.
• Use the control switches directly on the unit. If they are working, it indicates a problem with the hand control; you would then need to change the hand control.
The unit does not work when vou press • Is the battery charged and correctly installed? the # or # buttons on both sides of the lift.
Ensure the battery is inserted into the lift with the silver contact points inside the lift.
Place the battery into the charger and check the light on the charger. If the light remains red, leave the battery in the charger for approximately 2 hours.
• Is the "QUICK RELEASE" knob completely to the left
(locked) position? If not, the lift will not operate. Tighten the tension on the strap to help move the "Quick Release" knob.
• Is there tension on the strap? The "Anti Crush" feature will prevent the lift from working if there is no tension.
• If all of the previous points have been checked without success, you probably have a motor or electric problem, you then need to call your BHM Medical representative.
The unit starts and stops repetitively. • If the load is more than safe working load, the unit will not work due to an overload protection on the motor.
• Has the lift been making a "beeping" sound? The battery is low and needs to be charged.
The lift emits a "beep" during utilization.
The unit may stop lifting but you can
• Batteries are low, recharge them still lower it.
#A8200 (green-red) does not light up when the battery is in the charger points should be inside the charger.
• Is the charger plugged into a standard outlet?
• Check the condition of the fuse. If needed, change it with a spare one. (see "Maintenance" section of this manual)
• Has the battery been in the charger for many hours? See
"Battery Information" and "Charging the Battery " in
"Maintenance" section.
The light on the charger #A8200 shows • Battery is damaged - have it replaced. green when the battery is inserted; yet the lift will only work when there is no one in the lift. When you attempt to transfer someone, the lift stops.
The light on the charger #A8200 always • Try a different battery in the charger if available. lights up red - does not turn to green • Take the battery and charger to a certified technician.
User Manual - Portable Voyager 6 4
TROUBLESHOOTING
PROBLEMS after an overnight charge.
TO CHECK
PROBLEMS TO CHECK
Battery is always dead after a few • Replace battery with a new one, the life of the current one transfers (3 to 5). is probably nearly over.
(See the "Battery Information")
The light on the charger shows green • Battery has either been dropped or damaged in some other
(yellow for charger #700.13500) when way. Replace the battery. the battery is inserted, yet the lift will only do one or two transfers.
Charger #700.13500: Green light is not illuminated
• Is the charger plugged into a standard outlet?
• Try plugging the charger into another outlet to see if it will work.
• Contact your local dealer or representative.
Charger #700.13500: When battery is • Check the orientation of the battery. The silver contact inserted into the charger, the yellow light does not flash or illuminate solid. points should be inside the charger.
• Contact your local dealer or representative.
The battery does not slide into the lift. • Check if there is any damage to the battery that would interfere in the lift battery's slot. IF THERE IS DAMAGE -
DO NOT USE THE BATTERY.
• Try to slide in another battery to see if there is any damage in the lift battery's slot. If so, call your BHM Medical representative, if not discard the damaged battery.
Cannot remove battery from the lift. • See "Operation" instructions "Charging the Battery" on how to remove the battery.
• If the battery will not come out of the lift, tilt the lift to the side so that the battery opening is under the lift. With your hand guarding the opening for the battery, press the eject button on the front of the Voyager Portable while gently tapping the lift to help jolt the battery out of the lift. ONCE
THE BATTERY IS REMOVED - DISCARD BATTERY AND
REPLACE.
• If you cannot remove the battery after trying the above, contact your local dealer or representative for assistance.
User Manual - Portable Voyager 6 5
A C C E S S O R I E S
B A T T E R Y
P r o d u c t No. A 8 5 0 0
H O O K
P r o d u c t N o . 0 0 0 . 0 8 0 0 0
ACCESSORIES
C H A R G E R S
P r o d u c t N o . A 8 2 0 0
S T A N D A R D T R O L L E Y
P r o d u c t N o . A 5 9 0 0
P r o d u c t N o 7 0 0 . 1 3 5 0 0
S W I V E L A D A P T O R
P r o d u c t N o . 700.10950
H O S P I T A L G R A D E
P O W E R C O R D
P r o d u c t N o . E 7 5 2 5 N o r t h A m e r i c a
E 7 5 2 1 E u r o p e
E 7 5 2 7 U K
E 7 5 2 8 A u s t r a l i a
F U S E
P r o d u c t N o .
C 8 F G M C 0 . 5 ( 1 2 0 / 2 4 volt) N o r t h A m e r i c a
C 8 F G M C 0 . 2 5 ( 2 4 0 / 2 4 volt) E u r o p e a n
A R M E X T E N S I O N
P r o d u c t N o . A 8 3 0 0
H A N D C O N T R O L
P r o d u c t N o . 7 0 0 . 1 3 6 0 0
A D J U S T A B L E S T R A P E X T E N S I O N
P r o d u c t N o . A 3 7 1 0
User Manual - Portable Voyager
6 6
LOG BOOK
LOG BOOK
Product Name :
Product Serial Number
Date of purchase :
Warranty Period :
Distributed By
VOYAGER PORTABLE
Owner Checklist:
Ensure the lift is serviced regularly as the Maintenance Inspection Checklist on page 63.
Contact an authorized BHM Medical Distributor / Service Agent immediately if there are any problems with the operation of the device.
Ensure the log book is completed and signed.
Record any repairs required.
Withdraw the patient lifter from service if inspection reveals that user safety is jeopardized in any way from use of the patient lifter.
TO BE COMPLETED AFTER EACH SERVICE OR INSPECTION
Service Type Pre-delivery Periodic inspection Minor
Condition report
Major
Action taken
Date : Inspected by:
Printed Signature
Service Type Pre-delivery Periodic inspection Minor Major
Condition report
Action taken
Date Inspected by:
Printed Signature
User Manual - Portable Voyager 6 7
LOG BOOK
Service Type i Pre-delivery Periodic inspection Minor Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
Service Type Pre-delivery Periodic inspection ; Minor M Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature i
Service Type Pre-delivery \ Periodic inspection Minor \ i Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
Service Type Pre-delivery Periodic inspection ' Minor ; Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
User Manual - Portable Voyager
6 8
LOG BOOK
Service Type n Pre-delivery u Periodic inspection n Minor • Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
Service Type r i Pre-delivery Periodic inspection i Minor u Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
Service Type Pre-delivery Periodic inspection i Minor ; Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
Service Type r: Pre-delivery Periodic inspection Minor • Major
Condition report:
Action taken :
Date : Inspected by:
Printed Signature
User Manual - Portable Voyager
6 9
LOG BOOK
WARRANTY
This warranty is extended only to the original purchaser/user of BHM products.
B H M Medical Inc. warrants its products to be free from defects in material under normal use and service, within the periods stated below from the date of purchase. If within such warranty period any such product shall be proven to be defective, such product shall be repaired or replaced at BHM Medical's option. This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product. BHM Medical's sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement.
Patient Lifter 1 year
Tracks and installation Lifetime warranty*
Weighing Devices 1 year
Accessories on Lifter 1 year
Slings 1 year
Batteries - Voyager Portable 3 months
Batteries - All other lifts 1 year
Easytrack System 1 year
For warranty service, please contact the dealer from whom you purchased the BHM Medical product. [In the event that you do not receive satisfactory warranty service, please contact BHM Medical (see contact information in page 2 of this guide)].
Do not return products to our factory without prior authorization. BHM Medical will issue a Return
Merchandise Authorization (RMA) Number. C.O.D. shipments will be refused; all shipments to BHM
Medical must be prepaid. For this warranty to be valid, the purchaser must present its original proof of purchase at the moment of the claim. The defective unit, assembly or part must be returned to BHM
Medical for inspection. The part or components repaired or replaced are guaranteed for the remaining period of the initial warranty.
Limitations and E x c l u s i o n s :
The warranty above does not apply to serial numbered products if the serial number has been removed or defaced.
No warranty claim shall apply where the product or any other part thereof has been altered, varied, modified, or damaged; either accidentally or through improper or negligent use and storage. Warranty does not apply to products modified without BHM Medical's express written consent, (including but not limited to products modified with unauthorized parts or attachments); products damaged by reason of repairs made to any component without the specific consent of BHM Medical, or to products damaged by circumstances beyond BHM Medical's control. BHM Medical will solely determine evaluation of warranty claim. The warranty does not apply to problems arising from normal wear or failure to adhere to the instructions in this manual. BHM Medical Inc. slings are void of warranty if not laundered as per instructions on the Sling Label.
B H M Medical Inc. shall not be liable for damages losses or inconveniences caused by a carrier.
This warranty replaces any other warranty expressed or implicit and constitutes BHM Medical Inc. only obligation towards the purchaser. BHM Medical shall not be liable for any consequential or incidental damages whatsoever.
Valid only if BHM Medical did the original installation. Warranty voids if tracks/installation have been modified.
User Manual - Portable Voyager
7 0
Download
Advertisement
Key features
Lightweight and portable
Rechargeable battery
Quick-release strap system
Emergency manual-lowering feature
Available in a variety of models
Lifting capacity of 200 kg/ 440 lbs
Electronic soft-start and soft-stop motor control
Anti-crush feature
Complete range of patient lift slings
Frequently asked questions
The lifting capacity of the Voyager Portable Ceiling Lift is 200 kg/440 lbs.
The battery can be charged using the included charger. Plug the charger into a standard electrical outlet and connect the battery to the charger. The battery will take approximately 8 hours to fully charge.
The emergency manual-lowering feature allows you to safely lower the patient to the ground or a chair in the event of an electrical or functional failure. To use the emergency manual-lowering feature, pull the rubber stopper out of the Voyager Portable casing and turn the Allen key clockwise.
The quick-release strap system allows you to quickly and easily unwind and rewind the strap before and after transferring the patient. To use the quick release, push the knob and pull the strap out of the lift.