oxford diffraction Xcalibur Series User Manual
Add to My manuals48 Pages
The oxford diffraction Xcalibur Series is a powerful and versatile X-ray diffractometer designed for single-crystal analysis. It offers a point detector option for collecting diffraction data, which is ideal for crystals of all shapes and sizes. Xcalibur Series allows for accurate and efficient data collection, enabling users to determine crystal structures and analyze molecular properties.
advertisement
▼
Scroll to page 2
of
48
User Manual Xcalibur May 2004 Version 1.2 Oxford Diffraction Limited 68, Milton Park, Abingdon, Oxfordshire. OX14 4RX. UK Tel: +44 (0)1235 443630 Fax: +44 (0)1235 443631 http://www.oxford-diffraction.com Point detector operation Important Information This user manual applies to the Xcalibur systems manufactured in Poland by Oxford Diffraction. It is a supplementary document to accompany the full Xcalibur User Manual regarding the installation and use of a point detector fitted on an Xcalibur 2 diffractometer. Product: Model Type: Electrical Ratings: XCALIBUR PD 1/N AC 230 V 50/60 Hz 4200 Watts Before attempting to operate the system, PLEASE READ THE INSTRUCTIONS. This product should only be used by persons legally permitted to do so. If the equipment is used in a manner not specified in the User Manual, the protection provided by the equipment may be impaired. Important Health and Safety Notice When returning components for service or repair it is essential that the item is shipped together with a signed declaration that the product has not been exposed to any hazardous contamination or that appropriate decontamination procedures have been carried out so that the product is safe to handle. Care has been taken to ensure the information in this manual is accurate and at an appropriate level. Please inform Oxford Diffraction if you have any suggestions for corrections or improvements to this manual. Xcalibur service and support is available for technical and operational issues as indicated below. • E-mail: [email protected] • Phone: +44 (0) 1235 443630 between 8 a.m. and 4.30 p.m. (UK time), Monday to Friday • Fax: +44 (0) 1235 443631 This users' manual has been written according to standard 89/392/EEC and further modifications. Xcalibur is a trademark of Oxford Diffraction Limited in some jurisdictions and a registered trademark of Oxford Diffraction Limited in other jurisdictions. Oxford Diffraction acknowledges all trademarks and registrations. Copyright 2000 Oxford Diffraction Limited. All rights reserved. No part of this document may be reproduced or distributed in any form, or by any means, or stored in a database or retrieval system, without prior written permission of Oxford Diffraction. Version 1.1 XcaliburPD v1.2.doc Page i XcaliburPD USER MANUAL Contents Contents................................................................................. 2 Table of Figures .................................................................... 3 1. Health and Safety Information ......................................... 4 1.1 General ...................................................................................................................................... 4 1.2 Electrical Safety ......................................................................................................................... 5 1.3 Mechanical Handling Safety ...................................................................................................... 6 1.4 Safe Mechanical Practice .......................................................................................................... 6 1.5 Moving Parts .............................................................................................................................. 6 1.6 X-ray Radiation .......................................................................................................................... 7 1.7 Extreme Temperatures .............................................................................................................. 8 1.8 Vacuum...................................................................................................................................... 8 1.9 Hazardous or Toxic Materials .................................................................................................... 8 1.10 Modifications and Service........................................................................................................ 8 2. Normal Operation Using a Point Detector ...................... 1 2.1 Installation of the point detector................................................................................................. 1 2.2 Removal of the point detector.................................................................................................... 6 3 General Commands ........................................................... 7 3.1 Gt - Goto Angles Commands..................................................................................................... 7 3.2 Ty – Type Details Commands ................................................................................................... 7 3.3 Single measurements ................................................................................................................ 7 3.4 Commonly used unit cell/Indexing commands .......................................................................... 9 3.5 Peak table commands ............................................................................................................. 10 3.6 System commands .................................................................................................................. 10 3.7 Writing to disk .......................................................................................................................... 11 3.8 Reading from disk .................................................................................................................... 11 3.9 Exiting the CrysAlis CCD program........................................................................................... 11 4 Standard Point Detector Experiment ............................. 12 4.1 Crystal Mounting and Alignment.............................................................................................. 12 4.2 Setting the data collection parameters .................................................................................... 14 4.3 Peak Hunting ........................................................................................................................... 18 4.4 Unit cell determination ............................................................................................................. 20 4.5 Data Collection......................................................................................................................... 21 4.6 Data Processing and Reduction .............................................................................................. 21 Version 1.1 XcaliburPD v1.2.doc Page ii XcaliburPD USER MANUAL 4.7 Dc Movie - Replay of Data Collection Movie ........................................................................... 23 4.8 Absorption Correction .............................................................................................................. 23 4.9 GRAL - Space Group Determination ....................................................................................... 24 5 Glossary of point detector commands .......................... 25 Table of Figures Figure 2.1 The universal theta arm ..................................................................................................... 2 Figure 4.1 Optical alignment of the crystal ........................................................................................ 13 Figure 4.2 The reflections conditions programme............................................................................ 17 Figure 4.3 The peak hunting process................................................................................................ 19 Figure 4.4 The centring procedure .................................................................................................... 20 Figure 4.5 Dataproc programme opened by using the dc redpd command...................................... 22 Figure 4.6 DC MOVIEPD programme ............................................................................................... 23 Version 1.1 XcaliburPD v1.2.doc Page iii XcaliburPD USER MANUAL HEALTH AND SAFETY INFORMATION 1. Health and Safety Information 1.1 General In normal operation the system is designed to operate safely. All users of Xcalibur should be aware of potential hazards which exist in and around equipment of this type and the ways of avoiding possible injury and equipment damage which may result from inappropriate ways of working. A description of such potential hazards and how to avoid them is given in this section. This manual adopts the following convention: WARNING Indicates a potential hazard which may result in injury or death CAUTION Indicates a potential hazard which may result in damage to equipment Warning symbols on the equipment are: Protective conductor terminal Earth (ground) terminal CAUTION Risk of electric shock CAUTION Refer to accompanying documents WARNING Radiation Hazard See original manufacturers' manuals for further safety data on third party equipment supplied with the system. A list of these is given in this manual. WARNING Do not take risks. You have a responsibility to ensure the safe condition and safe operation of equipment. WARNING Xcalibur should only be operated and maintained by authorised operators of the system. An authorised operator is a person who has undergone specialist radiation training and has been trained in the use of Xcalibur by Oxford Diffraction personnel. Version 1.1 XcaliburPD v1.2.doc Page iv XcaliburPD USER MANUAL HEALTH AND SAFETY INFORMATION 1.2 Electrical Safety In normal use the user is protected from the dangers associated with the voltage, current and power levels used by the equipment. Only personnel qualified to work with the voltages and currents used by this equipment should attempt to disconnect, dismantle or modify the equipment. 1.2.1 Potential Electrical Hazards The following list is not intended as a complete guide to all the electrical hazards on the system, but serves to illustrate the range of potential hazards that exist: • • • • electric shock electric burn fire of electrical origin electric arcing 1.2.2 Recommended Precautions WARNINGS All of the electrical equipment supplied as part of the system should be provided with a protective ground. Do not remove protective grounds as this may give rise to an electrical safety hazard. It is vitally important that the system is properly grounded at all times. Follow local and national electrical regulations and procedures. Do not defeat interlocks, remove connectors, disconnect equipment, open safety covers, dismantle or modify equipment unless you are qualified and authorised to do so and you are fully conversant with its operation and potential hazards, or have total assurance through your local electrical permit to work system that the equipment has been made safe. Ensure that the mains supply is fused at an appropriate rating, or fitted with a circuit breaker, and that it can be isolated locally via a clearly labelled, clearly visible and easily accessible isolating switch. Isolate the supply before carrying out any maintenance work. Do not touch any unshielded wires or connectors while mains power is supplied to the system. Do not allow water or any other foreign objects to come into contact with Xcalibur’s electrical components. Version 1.1 XcaliburPD v1.2.doc Page v XcaliburPD USER MANUAL HEALTH AND SAFETY INFORMATION 1.2.3 First Aid A course in first aid to include methods of artificial respiration is recommended for those whose work involves equipment that may produce a high voltage. WARNING Do not attempt to administer first aid to someone who may have suffered electric shock until the source of the shock has been isolated. Mains voltages are present in the system. High voltages are used by the X-ray tube and power supply. These can cause serious injury or death. Only personnel qualified to work with high voltages and currents should perform service or maintenance work on such equipment. 1.3 Mechanical Handling Safety WARNING Lifting points are provided for safe handling of components and safe handling practice must be observed to comply with local regulations. Check that lifting points are used only for the job intended. The system itself and some components are heavy and require careful handling. Use safe lifting procedures for heavy items to prevent possible strain injury. 1.4 Safe Mechanical Practice In normal use personnel are not required to undertake mechanical work. However, servicing or repair may necessitate access to any part of the system. Only personnel who have been trained by Oxford Diffraction to carry out service work on this equipment should attempt to dismantle, modify or repair the equipment. Water connections should be made and tested in accordance with any local and national safety regulations. 1.5 Moving Parts There are a number of moving parts in the system which are powered by electric motors. WARNING Injury could result if clothing or body parts become caught in moving mechanisms. Keep clothing, hands and body parts away from moving mechanisms. Version 1.1 XcaliburPD v1.2.doc Page vi XcaliburPD USER MANUAL HEALTH AND SAFETY INFORMATION 1.6 X-ray Radiation WARNING This equipment contains an X-ray tube. Ensure that safe working practices relating to radiation are employed. Follow any local, national or international rules and guidelines. Intentional or reckless misuse of the X-ray generator or its safety devices including safety interlocks and cabinet shielding can result in serious injury or even death. During operation, there is an acceptable level of X-ray radiation as based on the recommendations on risk published by the International Commission of Radiological Protection (ICRP) and endorsed by the National Radiological Protection Board (NRPB) in the UK. For use in the UK, the Ionising Radiations’ Regulations 1999 should be adhered to. For countries outside the UK the appropriate laws apply such as registration and inspection. Customers should be aware of their duty of safety to their employees and visitors. WARNINGS To prevent injury to personnel and possible damage to the equipment, please note the following guidelines: 1. Only authorised personnel who have received appropriate instruction and are aware of the laboratory rules that govern the use of this type of system should operate the system. 2. Never dismount the beam stop when the system is operational. 3. Do not operate the system without the collimator, unless performing the beam alignment procedure. 4. Use appropriate X-ray detection equipment to perform regular radiation checks as per any laboratory rules Use only genuine firmware X-ray tubes, X-ray generators, monochromators, goniometer heads and collimators, as recommended by your Xcalibur supplier. Use of other products may compromise the performance of the shielding and safety system, and may invalidate your warranty. Version 1.1 XcaliburPD v1.2.doc Page vii XcaliburPD USER MANUAL HEALTH AND SAFETY INFORMATION 1.7 Extreme Temperatures WARNINGS 1. Systems fitted with the low temperature option use liquid nitrogen and/or liquid helium as a coolant. Liquid nitrogen and liquid helium are cryogenic liquids and can cause cold burns. Wear gloves when handling cryogenic liquids and use eye protection. Refer to the information supplied with the equipment for more information. 2. During operation the X-ray tube becomes hot. In normal use they are located inside a cabinet and hot parts are not accessible. During maintenance periods, however, it may be necessary to override the interlock so that adjustments can be made. Therefore great care must be taken to avoid touching the X-ray tube when it is operating and for a period of 20 minutes after operation. 1.8 Vacuum WARNING When handling and using X-ray tube, particular care should be taken to avoid injury caused by possible implosion of the vacuum tube. Wear eye protection. 1.9 Hazardous or Toxic Materials Beryllium and beryllium oxide are toxic materials. Follow appropriate handling, shipping, use, storage and disposal procedures and regulations. Refer to BrushWellman Material Safety Data Sheet No. M10 for further information. WARNING If Beryllium is exposed to fire, it may oxidise to highly toxic beryllium oxide powder. Do not attempt to clear up the remains of any fire, but contact the relevant local agency stating that there is an incident involving possible beryllium or beryllium oxide contamination. 1.10 Modifications and Service The manufacturer will not be held responsible for the safety, reliability or performance of the equipment unless assembly operations, extensions, re-adjustments, modifications and repairs are carried out only by persons authorised by the manufacturer. It should be stressed that those parts of the equipment which are interchangeable, and which are subject to deterioration during operation, may significantly affect the safety of the equipment. Version 1.1 XcaliburPD v1.2.doc Page viii XcaliburPD USER MANUAL OPERATION 2. Normal Operation Using a Point Detector This section of the manual describes the installation and operation of a point detector on an Xcalibur diffractometer fitted with a universal theta arm. For information regarding the use of a CCD detector, refer to the main Xcalibur operators manual. 2.1 Installation of the point detector If a CCD camera is currently installed, determine the orientation matrix of the standard crystal before removing the CCD camera from the diffractometer (this will save a lot of time): Mount the cubic test crystal (either CaF2, or KAl(SO4)2·12H2O for example) and optically align it. Carry out a short data collection (e.g. unit cell in 5 minutes) and issue the peak hunting command (PH S). Find the orientation matrix and unit cell (UM F). Index the cell (UM I) and save it by typing WD T and giving it a file name. If further information is required about the operation of a CCD camera, refer to the XcaliburCCD manual. 1. Switch off the CCD camera by turning the key anticlockwise on the front panel of the KMW200CCD chiller. In the pull down menu: tools/setup file of the CrysAlis CCD (and CrysAlis RED) programs, swap the setup file from the CCD *.par file to the one relating to the PD (also see next point). 2. Setup file preparation: In the case of a new system, the setup file needs to be checked and altered if necessary. Browse for the setup file using windows explorer. Make a copy of the setup file and save it to disk for backup purposes. Open the original setup file with notepad, scroll down the file and, if necessary, change the last two parameters in the following line of text from: GONIOMETER TYPE KUMA_KM4NEW TOP 15.00000 3.00000 BC2_SAPPHIRE FALSE 1.00000 1.00000 1.00000 TIME FALSE FALSE TRUE TRUE FALSE To: GONIOMETER TYPE KUMA_KM4NEW TOP 15.00000 3.00000 BC2_SAPPHIRE FALSE 1.00000 1.00000 1.00000 TIME FALSE FALSE TRUE FALSE TRUE 3. Restart the CrysAlis CCD (and CrysAlis RED) programmes Version 1.1 XcaliburPD v1.2.doc Page 1 XcaliburPD USER MANUAL OPERATION 4. Drive the goniometer to theta = 90 deg (GT T 90), and drive the camera distance to 60 mm (GT D 60). 5. Unplug all the connectors to the CCD head, note which fibre optics cable is connected to which socket; mark them if necessary to save time when reconnecting. The order of the water pipes is not important. CAUTION Take care when disconnecting water pipes to prevent drips of water landing on the camera. 6. Unscrew the two upper M4 screws connecting the slider to the lead screw which defines the camera distance. Turn the end plate of the slider anticlockwise by 180 degrees. Remove the CCD camera from the slider carefully and place in the storage box. The empty slider will appear as in Figure 2.1 Figure 2.1 The universal theta arm CAUTION Take care not to allow the CCD head to collide with the beam stop or kappa block as the slider has very low friction. WARNING IMPORTANT! Switch the interface off at this point using the power switch on the front of the interface before the high voltage cable is plugged into the point detector. This cable could carry a voltage of up to 1000V. 7. Replace the CCD camera by the point detector. 8. Turn the end fixing plate clockwise by 180 degrees and fix the PD slider into place using the two screws that were removed in step 6. Connect the PD interface cable and screw in the fixing screws to prevent accidental disconnection. Version 1.1 XcaliburPD v1.2.doc Page 2 XcaliburPD USER MANUAL OPERATION WARNING The point detector interface cable carries a high voltage 9. Switch the interface back on and issue the command GON REINIT. Check that no error messages appear. Drive all angles back to zero (GT a 0 0 0 0). 10. In the case of a new machine, set the detector distance to 130mm (GT D 130) and open the cabinet door. Drive theta to 90 degrees (GT T 90) and -90 (GT T -90) and check that there are no potential collisions with the cabinet. Move the goniometer if necessary. 11. Issue the command TY P to display the data collection parameters in a table in the history window. Check and change if necessary the following parameters: SC S 0.15 Omega scan speed SC W 1.3 0.35 1.008 Scan width SC T 0 2 Type of scan MO B 0.5 Background mode MO S 1 61 3 25 0.015 Scan mode DA 1.33 1.33 Detector aperture for the typically used slits DL 2 100 Discrimination level SW CE 1.5 Centering conditions SW SMI 0 Mode of operation for SM I (0.01 for synchrotrons only) TR 2 60 Theta range FI 100000 Filter setting HV 170 Counting chain high voltage level GA 200 Counting chain gain level LL 30 Counting chain low level setting WI 170 Counting chain window setting 12. In the pull down menu tools/options check and set to zero all the correction factors except for: alpha and beta (which should be the same as the CCD setting), detector distance (which should be 130) and the X and Y positions of the detector (which should both be the theoretical values of 512). Don’t recalculate the peak table (if any) on exit. Version 1.1 XcaliburPD v1.2.doc Page 3 XcaliburPD USER MANUAL OPERATION 13. If the orientation matrix and unit cell are unknown find them using peak hunting (for example: PH S 25 10 20 -60 -20 0 359). NOTE to find a cell more rapidly, it may be better to increase the size of the slits from the front of the detector. If you do this, you must type the command DA X Y to update the new detector aperture setting (where X and Y are the slit size). 14. Once the data collection has finished issue the command UM F (or UM C) to find the unit cell then UM I to index it. NOTE at this stage the unit cell will be of poor quality due to the lack of calibration. 15. If the orientation matrix is known, read the peak table from disk if needed using the command RD T. If the data were from a CCD data collection, use the command PT ANGLES to calculate the setting angles. 16. Update the peak table using the command UM U. This will cause the diffractometer to search for all of the theoretical peaks in the peak table and will take about 20 minutes to refine 20 peaks. 17. Save the final model by issuing the command WD CAL. A backup file will be written at the same time. Repeat the UM U procedure to update the peak table. If the standard deviations for the unit cell lengths are less than 0.001, the model is finished. If the model is insufficient, further corrections must be carried out. This may be in the case of a new machine or a machine that has undergone extensive adjustments. The following additional procedure should then be followed: 1. Issue the command PT E to examine the peak table. Click on the radio button in the coordinates section at the bottom of the window labelled angles. Find a strong reflection. 2. After finding the strong reflection, replace the 1.33 slits (if they were previously removed) and type DA 1.33 1.33. Centre the reflection by typing CE. Check the counter line electronics characteristics by issuing the command SM C 170 200 10 200 1 1. Repeat this if necessary changing HV and/or GA and if necessary adjust LL (value just to the left of the peak) and WI (value just to the right of the peak) parameters. To examine the profile of the peak type SM S 30 0.5. Save the new parameters by typing WD CAL. To find the initial zero correction for theta and horizont: 3. If the goniometer is not already set at the reflection position, type GT R h k l (where h k and l are the coordinates of the reflection that was chosen in point 2). Find the initial zero Version 1.1 XcaliburPD v1.2.doc Page 4 XcaliburPD USER MANUAL OPERATION correction for theta and horizontal using the ZC A command. This command will examine the peak from two opposite directions and will output a suggested correction in the history window. Apply the suggested correction by copying the line of text from the history window and pasting it into the command line and pressing return. Re-issue the ZC A command and repeat this procedure approximately 5 times until there is little difference between the current and suggested correction (you may type TY P to see the current setting). Note that if the corrections are very large or the reflections are very narrow, ZC A may fail on the first attempt. In this case, use larger slits and/or wide centring parameters (e.g. SW CE 2). NOTE before reissuing the ZC A command, you must GT R h k l (or GT A o t k p) to recover the starting reflection position. To calculate the zero corrections for omega and kappa: 4. Build a peak table containing 24 reflections of the 3 3 5 family (for Mo radiation) or the 2 2 4 family (for Cu radiation): Type the command GT R 3 3 5 then PT A (this adds the current goniometer setting to the peak table as a peak), UM I (to index the new peak), then PT E (to open the peak table editor). Note that UM I will fail if the peak table contains fewer than 3-4 reflections. Erase all but the last added reflection and click on exit. Issue the command PT L 12 (this will add peaks to the peak table which are related by the Laue symmetry to the 3 3 5 reflection. Issue the command PT E to examine the peak table. Click on the radio button at the bottom of the window in the coordinates section labelled angles. Sort the reflections by descending kappa (by clicking at the top of the kappa column). Remove any reflections with a kappa angle of less than 2 degrees as the centring process will be very slow for these reflections. Click on the exit sorted button. Save the peak table using the command WD T. 5. Calculate the initial zero corrections for omega and kappa by typing ZC T. The suggested correction will be typed into the history window. Issue the command TY P to display the current values. Copy and paste the ZC S line from the table in the history window to the command line and modify the 1.st and 3.rd value by adding approximately 90% of the suggested change. NOTE Check that the suggested change is a sensible value 6. Repeat points 4 and 5 approximately 5 times until the suggested changes are negligible. 7. Issue the command REFINE MODEL. Set the cell type to LAT_AAA and ANG_909090. In each cycle check only ONE radio button at a time to refine in order: alpha, beta, kappa and omega. Refine theta separately otherwise there is a risk that, because the command Version 1.1 XcaliburPD v1.2.doc Page 5 XcaliburPD USER MANUAL OPERATION was originally designed to deal with CCD data, the refinement may become unstable due to the small number of observables. 8. Note the calculated corrections for omega, theta and kappa and add the new values to the parameters by issuing the command ZC S do dt dk NOTE the only way to measure the correction for horizont (4.th parameter) is ZC A. 9. Using tools/options reset the corrections for omega theta kappa and phi to zero - don’t recalculate the peak table on exit from the panel. 10. Save the final model by issuing the command WD CAL. A backup file will be written at the same time. Repeat the UM U procedure to update the peak table. If the standard deviations for the unit cell lengths are less than 0.001, the model is finished. 2.2 Removal of the point detector 1. Drive the goniometer to theta = 90 deg (GT T 90), and drive the camera distance to 60 mm (GT D 60). WARNING IMPORTANT! Switch the interface off at this point using the power switch on the front of the interface before the high voltage cable is plugged into the point detector. This cable could carry a voltage of up to 1000 V. 2. Unscrew the PD interface cable fixing screws and disconnect the PD interface cable. 3. Unscrew the two upper M4 screws connecting the slider to the lead screw which defines the camera distance. Turn the end plate of the slider anticlockwise by 180 degrees. CAUTION Take care not to allow the point detector head to collide with the beam stop or kappa block as the slider has very low friction 4. Place point detector in storage case. The point detector interface cable may be left in place but should be fastened securely to prevent the connector colliding with another detector, the goniometer, beam stop or crystal. Version 1.1 XcaliburPD v1.2.doc Page 6 XcaliburPD USER MANUAL OPERATION 3 General Commands 3.1 Gt - Goto Angles Commands When Xcalibur is not collecting data the goniometer axes can be driven to accessible positions using the following commands: gt a om th ka ph go to angles with values omega (om), theta (th), kappa (ka) and phi (ph) gt o om go to omega angle ‘om’ gt t th go to theta angle ‘th’ gt k ka go to kappa angle ‘ka’ gt p ph go to phi angle ‘ph’ gt d det go to detector distance ‘det’ in mm gt r h k l go to a reflection with specific h k l value 3.2 Ty – Type Details Commands The ty command allows the user to print a variety of current settings to the history window: ty p print current Xcalibur parameter settings ty z print the current zero correction parameters ty u print current UB matrix ty l print current unit cell and lattice settings ty t print current contents of peak table 3.3 Single measurements sm i time repeats sm r h k l psiang psistart psistep Version 1.1 stationary intensity measurement with exposure time (time) and number of repetitions of the measurement (repeats) scan a reflection with the given h k l values with the optional parameters: psiang (psi angle for XcaliburPD v1.2.doc Page 7 XcaliburPD USER MANUAL OPERATION the scan), psistart (starting psi angle for a sequence of scans), psistep (psi step angle for measuring a sequence of scans) sm s steps time step scan with a number steps (steps) and an exposure time (time) sm ao om th ka ph scanwidth time Single measurement omega scan with angles sm eta starteta step steps time om, th, ka and ph are the omega, theta kappa and phi settings in degrees scanwidth = degrees time = seconds step scan moving every reflection vertically in the aperture window starteta = starting eta value. Actual goniometer position is assumed to be theta 0.0. step = scanstep in deg. steps = number of steps for scan time = exposure time in sec per step sm help help overview of sm commands sm i time single static measurement of exposure time in seconds sm rp 30 secs phi rotation photo sm o startangle scanwidth time omega scan sm p startangle scanwidth time phi scan sm q h k l st #s1 sw1 u1 v1 w1 [#s2 sw2 u2 v2 w2] [filename] records a q scan sm r h k l [psistart [psiend psistep]] H K L = indices of selected peak st = measurement time of one scanning point [sec] #s1 = number of scanning points in the [U1 V1 W1] direction sw1 = increment of scanning angle for [U1 V1 W1] direction [Deg] U1 V1 W1 = indices of the first scanning direction #s2 = number of scanning points in the [U2 V2 W2] direction sw2 = increment of scanning angle for [U2 V2 W2] direction [Deg] U2 V2 W2 = indices of the second scanning direction [filename] = optional filename. measure a single reflection h k l = reflection index, may be fractional. psistart = optional psi angle. psiend = optional psi start angle for measuring a sequence. psistep = optional psi step angle for measuring a Version 1.1 XcaliburPD v1.2.doc Page 8 XcaliburPD USER MANUAL OPERATION sm t startangle scanwidth time sequence theta scan 3.4 Commonly used unit cell/Indexing commands um f [lengthdeviation angledeviation fractionindexed] Automatic unit cell determination (indexation, refinement, reduction) um i [indexrejectioncriterion] [lengthdeviation angledeviation fractionindexed] = The defaults are 0.05, 0.1, 0.7. You can loosen this condition in case of an unsuccessful indexing: typically 0.3 0.3 0.5. Index and refine unit cell um c [change #]|[c11 c12 c13 .. c31 c32 c33] [indexrejectioncriterion] = rejection criterion which determines the maximum allowed deviation for which a reflection is considered indexed. Change orientation matrix um reduce um r [symmetrycode] [change #] = transformation number from table obtained by typing um c: [c11 c12 c13 .. c31 c32 c33] – direct space transformation matrix Apply Niggli reduction of unit cell refine UB under symmetry constraint um sarray arraynum ub11 ub12 ub13 .. ub31 ub32 ub33 um u Version 1.1 [symmetrycode] – The necessary code can be obtained by just typing um r Store the defined ub unit cell matrix to a buffer array. 8 available. Used to define a number of ub matrices for data reduction of twinned data. Arraynum (0..7) storage buffers for unit-cell matrices Ub11 … ub33 unit-cell ub matrix activates the procedure to refine the crystal orientation automatically XcaliburPD v1.2.doc Page 9 XcaliburPD USER MANUAL OPERATION 3.5 Peak table commands pt clear pt a pt e pt expand n mmin mmax intmin dmax dmin pt l lauecode clearing of the peak table add a peak to the peak table Peak table edit Peak table expand n = number of reflections required mmin = min order of difference -1 for CCD (-3 Point detector) mmax = max order of difference +1 for CCD (+3 Point detector) intmin = Intensity threshold dmax = max d spacing dmin = min d spacing add reflections to the peak table according to their Laue symmetry lauecode = The following codes can be selected: 1: -1 2: 2/m 3: mmm 4: 4/m 5: 4/mmm 6: -3 7: -3m1 8: -31m 9: 6/m 10 : 6/mmm 11 : m-3 12 : m-3m pt sa settingno add reflections from a different setting position settingno = The settingno refers to the basic measurement settings. 3.6 System commands The following commands may be issued to access the Windows system operations: System dos spawns a MSDOS window with the current directory path being used in the CrysAlis program. System explorer spawns an Explorer window with the current directory path being used in the CrysAlis program Version 1.1 XcaliburPD v1.2.doc Page 10 XcaliburPD USER MANUAL OPERATION 3.7 Writing to disk Current machine parameters, images and the contents of peak hunting tables can be written to disk wd p write disk parameter settings. Saves the current machine parameters to disk wd ph write disk peak hunting. Saves the current contents of the peak hunting table to disk wd t write disk table. Saves the current contents of the peak table to disk 3.8 Reading from disk Machine parameters, images and peak tables can be read from disk using the following commands: rd p read disk parameter settings. Reads stored machine parameters from disk rd ph read disk peak hunting. Reads a stored peak hunting table from disk rd t read disk table. Reads a stored peak table from disk 3.9 Exiting the CrysAlis CCD program To exit the CrysAlis CCD program the command en should be issued. This drives the goniometer axes to their home zero positions and exits the CrysAlis CCD program. Version 1.1 XcaliburPD v1.2.doc Page 11 XcaliburPD USER MANUAL OPERATION 4 Standard Point Detector Experiment In order to use the point detector (PD), the correct parameter file must be loaded. This is due to subtle changes between the layout of the GUI for the CCD experiments and the PD experiments. The location and type of parameter file may be checked by using the tools/setup file pull down menu. If the setup file is changed, the programme must be restarted in order to load this new parameter file. NOTE It is important to make sure the setup file is correctly formatted for point detector operation and not for CCD operation otherwise data collection will not be possible. See section 3.5.1 for details. A standard crystallography experiment using a point detector consists of 7 main steps: 1. Crystal mounting and alignment 5. Peak hunting 6. Unit cell determination 7. Data collection 8. Data processing 9. Space group determination 10. Structure solution and refinement The procedure for a standard experiment follows: 4.1 Crystal Mounting and Alignment Caution Press ‘STOP’ on the remote control or ‘Ctrl’ on the keyboard to stop movement of the equipment in an emergency. Mechanical movement of the goniometer and CCD detector may be performed using the remote control. 1. Start the CrysAlis CCD application 2. Press F12 key to release control from the computer to the remote control unit 3. Press the 0 and HOME buttons on the remote control to drive the goniometer angles to the zero / home position. Version 1.1 XcaliburPD v1.2.doc Page 12 XcaliburPD USER MANUAL OPERATION 4. Mount the xyz goniometer head with crystal attached to, for example a glass fibre or nylon loop with oil or glue. 5. Press Lower and 0 on the remote control. This will drive the goniometer to the correct orientation to allow optical alignment of the crystal. NOTE The settings lower and upper refer to the glass stick position on the video monitor Upper Setting Lower Setting Figure 4.1 Optical alignment of the crystal 6. Use the tool provided with the goniometer head to adjust the vertical height and horizontal position of the crystal, such that the crystal is in the centre of the video monitor screen. 7. Press 180 on the remote control to rotate the crystal through 180 degrees. If the crystal’s horizontal position has moved on rotation adjust the position. Press 0 and repeat this procedure until rotation gives no movement of the crystal. 8. Repeat the above process, rotating between 90 and 270 degrees. 9. Press Upper on the remote control. The goniometer will now move to the upper position such that the goniometer head is located behind the collimator. If the vertical height of the crystal has changed, adjust and return to the lower position. Repeat until the vertical position is unchanged between the upper and lower positions. 10. Press Lower and check alignment of the crystal on rotation between 0 and 180, 90 and 270 degrees. 11. Press 0 and Home to return the goniometer to its zero position. 12. Exit alignment procedure by selecting OK on the computer screen. This will return goniometer control to the computer and prevent use of the remote control. Version 1.1 XcaliburPD v1.2.doc Page 13 XcaliburPD USER MANUAL OPERATION 4.2 Setting the data collection parameters • Two theta range The two-theta range may be set by the command TR tthmin tthmax (where TR = theta range, tthmin is the minimum value of 2 theta and tthmax is the maximum 2 theta value) typical values are: TR 4 60 or TR 4 80 for Mo radiation. • Scan width The scan width depends on the type and quality of the crystal. The sc w command sets the scan width parameters: SC W Oa Ob C where Oa = scan width in omega (0.0≤Oa≤180), Ob = scan widening parameter (0.0≤Ob≤2.0) and C = scan centre shift parameter (1.0≤C≤1.1)The scan width increase parameter and scan centre shift are often left unchanged, typically: SC W 1.2 0.35 1.0008 NOTE The Oa and Ob parameters of the SC W command determine the omega scanwidth. The total scanwidth is calculated according to the formula: Oa + Ob ´ tan (theta). The last parameter allows to adjust the scan centre to any position between Ka1 and Ka2 peaks. The position of a reflection obtained from the orientation matrix corresponds to the Ka1 peak position i.e. Bragg angle Ta1. The scan centre can be changed by calculating new theta angle according to the formula: sin(theta) = C * sin(Ta1). • Scan type The scan type command SC T (Tx) (f) determines the type of scan to be made. Tx = the threshold angle (0.0≤Tx≤90.0) and f = the multiplier for theta motor (0.0≤f≤2.0). Omega scans are made for reflections with the theta Bragg angle less than Tx. Omega-theta scans are made for reflections with the greater Bragg angle. The scan type is usually set to Theta-Twotheta: SC T 0 2 NOTE The theta motor speed is calculated as f * Vo and it must be inside the same range as the omega scan speed, (sc s). • Scan speed The scan speed is dependent upon the crystal quality and the requested quality of data. The sc s command SC S (Vo) sets the omega scan speed in deg/sec. The ranges for Vo are 0.004768 ≤ Vo ≤ 1.5 For routine organometalic data collections use: SC S 0.15 • Mode of scan The mode of scan command MO S m s n1 [n2 Vmin] specifies the type of scan to be made during data collection. The parameters are: m - mode of scan (0≤m≤2; 0=BPB scan, 1=continuous-integrative step scan and 2=stationary step scan); s - number of scan steps (not used for m=0) (10≤s≤1024); n1 and n2 minimum accepted I/sigma(I) and requested I/sigma(I) (-3≤n1≤n2≤50) and Vmin - Minimum rescan speed in deg/sec Version 1.1 XcaliburPD v1.2.doc Page 14 XcaliburPD USER MANUAL OPERATION (0.004768≤Vmin ≤1.5). The most typical mode of scan is an integrating stepscan with prescan. This skips reflections that have I/Sigma(I) values worse than 3.0, whilst trying to obtain I/Sigma(I) ≤25. This should give an R1 factor of the refined structure ≤5% For example: MO S 1 55 3 25 0.015 or MO S 1 60 1 10 0.0050 NOTE m=0 A continuous scan according to the parameters given by SC S , SC W and SC T commands will be made. In this case the parameters must be omitted. For n1 = n2 parameters n2 and Vmin may be omitted also. m=1 An integrative step scan according to parameters given by SC S , SC W and SC T will be made . The total scan width is divided into s steps and for each step a continuous scan is made. m=2 Is equal to mode = 1 with an exception that a stationary measurement of each centre of step is made. The step measuring time is determined by the step width and the scan speed Vo, and is equal to step width/Vo. In all scan modes the attenuation filter is automatically inserted when the counting rate is higher than the level set by FI and scans are repeated. • Background mode The mode of background calculation is set by the command MO B f (0.01 <= f <= 2.0). It is usually set to 0.5, MO B 0.5, which means that 50% of scan steps (25% from each side) is used to calculate background. This must be taken into account when choosing the scan width. NOTE For MO S 0 the background is measured stationary at each side of the scan during time determined by t = f*scantime/2 . The total background time is twice the given value. For MO S 1 and MO S 2 the parameter f determines the division of the measured profile into reflection peak and background regions. The left and right backgrounds include the same number of outer profile points : s1 = 0.5*s*f/(f+1) , Version 1.1 XcaliburPD v1.2.doc Page 15 XcaliburPD USER MANUAL OPERATION where s is the total number of scan points. • Index limit The index limit command IL hmin hmax kmin kmax lmin lmax sets the index limits for the data collection. The range is dependent upon the unit cell lengths. Signs are dependent on the crystal system (unique reflections set) and on the crystal orientation. If possible, the kappa axis should be on the negative side during data collection. Typical values are: IL 0 10 0 10 - 0 0 • Reference reflections Reference reflections are particularly important in point detector data collections where the long data collection time may allow for crystal decay or movement. The RR command: RR rr_num [interval om_tol int_tol [h k l at [h k l at [h k l at]]]] specifies the reference reflections and conditions for initialization of the reorientation procedure. • • • • • rr_num (number of reference reflections: 0≤rr_num≤3) and if rr_num>0: interval (interval between reference reflections measured: 1≤interval≤32000) om_tol (requested omega repeatability: 0.0025≤om_tol≤10) int_tol (allowed intensity fluctuations parameter (number of sigmas by which a fluctuation triggers the re-centring): 1≤int_tol≤100) h k l (Miller indices of ref. reflection:-126≤h,k,l≤126) at (attenuation filter use: 0,1 (0 = filter out, 1 = filter in)). Up to 3 reference reflections may be measured during data collection. When the number of reference reflections n1 is equal to zero the further parameters should not be entered. The interval between two successive measurements of reference reflections is specified by the given number n2 of measured reflections. The reference reflections are measured with a step scan according to parameters given by SC W, SC T commands and a step number specified by n3. The step measuring time and the attenuation filter setting are determined for individual reflection by t and fil respectively. For each reference reflection one has to specify its Miller indices h,k and l which you choose from the peak table which was recorded during the peak hunting procedure. They should be strong, but not need an attenuator. Typically they are measured every 50 - 100 reflections, for example RR 3 100 0.15 12 1 2 3 0 2 3 4 0 3 4 5 0 • Reflection conditions The reflection conditions command RC allows the user to select which reflection conditions are applied during the data collection due to the centred cells, glide planes and screw axes (For further explanations see "International Tables for Crystallography" (1983), edited By Theo Hahn, Vol A, pp. 27-29 and 41-47, Dordrecht (Holland) / Boston (U.S.A): D. Reidel Publishing Company). Typically all reflections should be measured, just in case the a-priori assumption of lattice centring or space group etc. is wrong. It is possible to skip extinct reflections by using the RC command. It will display a menu allowing the extinction rules to be set. These extinction rules are not saved with the parameters so have to be input at the start of each data collection. This is because these rules may easily be overlooked (after typing parameters from the previous data collection on screen) causing the measurement of an incomplete reflections set. There is also a Version 1.1 XcaliburPD v1.2.doc Page 16 XcaliburPD USER MANUAL OPERATION ‘negated’ option, allowing the measurement of, for example, forbidden or mistakenly skipped reflections. Figure 4.2 The reflections conditions programme NOTE One has to be very careful not to create contradicting extinction conditions (the conditions are always logically interpreted by "and"). • Order of index change The order of index increase may be set using MA S command: MA S n1 n2 n3 (where n1, n2, n3 – define the order in which the indexes change (first-second-third) and 1=h,2=k,3=l) Typically one would use MA S 3 2 1 • Setting type Typically, the setting type should be chosen before peak hunting, but it is also possible to do it here, for example SW S 1 • Slits The appropriate slits must be set DA 2.0 0.5 or DA B 2.0 1.0 NOTE To check the result of choice one can use dc t emulator which prints kappa angle for each reflection. Version 1.1 XcaliburPD v1.2.doc Page 17 XcaliburPD USER MANUAL OPERATION 4.3 Peak Hunting The detector aperture is determined by two sets of manually exchangeable slits which are inserted in the front of the detector. The inserted combination of slits is expressed in degrees (DH and DV for the horizontal and vertical apertures, respectively). NOTE The detector aperture parameters are vital in the centring, zero corrections and peak hunting procedures. When the DA settings are wrong, these procedures may fail. The slits are available in a range of sizes 1. Check the aperture size of both the vertical and horizontal slits. If necessary, adjust the detector aperture parameter setting by typing DA DH DV where DH is the aperture of the horizontal slit and DV is the aperture of the vertical slit. NOTE You may use a bigger aperture (e.g. DA 4 2) to speed up peak hunting. Then use smaller slits (e.g. DA 1.33 1.33 depending upon the size of your reflections) and update the peak table before going to the auto-indexing routines to increase the accuracy. This is usually much faster than doing peak hunting immediately using the small slits. In the peak hunting procedure an area of the reciprocal space is investigated by phi scans with a scan speed ‘v’ The peak centring procedure will occur when a signal is observed above the discrimination level (N). Set the phi scanning speed and discrimination level parameters using the DL command: DL v N (typical starting values for v and N are 3 and 1000 respectively). The value of v can range from 0.01 to 3 (mostly a maximum of 2 is used) and the value for N ranges from 0 to 10000. NOTE If you do not know the parameters, the discrimination level parameter should be set to a high value (e.g. 1000) and shortly after peak hunting has started (< 1 min. typically) the procedure should be interrupted and a threshold 20 - 30% higher than the noise level threshold should be set. The speed parameter should be decreased only in case of weak diffractors. 2. Select the setting type for operation using the command SW S setting (where setting is the setting option). The options for setting are 0 = bisecting mode, 1 = traditional mode and 2 = high pressure mode (‘eularian’, phi at zero). Normally one would choose the bisecting setting: SW S 0 During peak hunting, the area is sampled starting at T1, K1 and P1. The omega angle is determined by the standard (traditional) setting requirement. The phi scans oscillate between P1 and P2. At the end of each phi scan the crystal is rotated, increasing the kappa angle. When kappa reaches its maximum value K2, the 2theta angle is changed and kappa is reduced to its minimum value K1 and so on. Version 1.1 XcaliburPD v1.2.doc Page 18 XcaliburPD USER MANUAL OPERATION NOTE All peaks found will be appended to peak table, so in the case of a new data collection, any existing peaks should be erased first by using the PT CLEAR command. 3. Start the peak hunting procedure either by using default values, using the command PH S, or by using custom settings by typing the command PH S n T1 T2 K1 K2 P1 P2 • • n is the number of reflections to find T1 and T2 are the minimum and maximum values for theta (0≤T1≤T2≤90) • K1 and K2 are the minimum and maximum values of Kappa (-130≤K1≤K2≤130) • P1 and P2 are the minimum and maximum values for phi (0≤P1≤P2≤360)). CAUTION It is recommended to routinely use values of Kappa that are between -70 and 70 to reduce the risk of finding reflections to close to the collision limits of the goniometer. 4. When a peak is found the centring procedure is carried out and the result is typed into the history window in the following format: (No) (omega) (2theta) (kappa) (phi) (Int) (filter setting) Where No is the number of the reflection in the peak table, omega, 2theta, kappa and phi are the goniometer settings and Int is the measured peak intensity. Figure 4.3 The peak hunting process Version 1.1 XcaliburPD v1.2.doc Page 19 XcaliburPD USER MANUAL OPERATION Figure 4.4 The centring procedure 5. The top intensity Int is the number of counts recorded in one second. The angles are converted to reciprocal coordinates x(1),x(2) and x(3) and these are compared with the coordinates already stored in the peak table. A new peak is said to be found, when the table does not contain a peak with coordinates y(1) , y(2) and y(3). When the new peak has been found the coordinates and intensity are added to the end of the peak table. If a duplicate peak is observed, the output contains (=No) instead of (No).When the intensity is greater than that already stored in the table the new peak position is stored instead of the old one. When the search is complete or when the peak hunting procedure is interrupted, the unit cell may be determined. 4.4 Unit cell determination After successful peak hunting: 1. Find the orientation matrix - currently two procedures are available, UM F working in reciprocal space and UM T in direct space. 2. Once the orientation matrix and unit cell have been found, there are a number of functions which may be utilised: • • • • • The found orientation matrix may then be refined by issuing the command UM I The orientation matrix may be changed in direct space using the command UM C. Using this command will cause a number of options to be typed to the history window. Typing UM C followed by the number of the desired option will change the orientation matrix. Alternatively, a transformation matrix may be typed after UM C to carry out the desired transformation. The orientation matrix may be changed in reciprocal space using the command UM CREC in the same way as using the command UM C A constrained cell may be displayed by typing UM R followed by a Laue symmetry option. UM REDUCE will facilitate additional unit cell reduction if required. 3. Once the desired cell and setting have been determined, the peak table and orientation data may then be written to disk using the command WD T. Version 1.1 XcaliburPD v1.2.doc Page 20 XcaliburPD USER MANUAL OPERATION 4.5 Data Collection 1. If the orientation matrix of the crystal is known, you may import an existing peak table. Alternatively, you may input the matrix using the command line. Then using sequence: GT R (go to reflection), CE 1 (centre reflection) and PT A (add to peak table) fill the peak table with approximately 10 reflections, refine the existing matrix and skip to point 4. If necessary - extend the existing peak table. 2. Before starting data collection it is advisable to update the peak table using the command UM U (This will cause the diffractometer to search for the reflections in the peak table and re-centre them). 3. To start the data collection procedure, type DC S. If you start the data collection without any parameters the program will use minimal values of HKL taken from IL parameter. After interrupting, a data collection may be restarted by typing the command DC R. A set of reference reflections will be collected followed by data collection from the next HKL from sequence: 4. When data collection finishes, go to the data processing section. 4.6 Data Processing and Reduction Reflection data collected during the data collection commands need to be converted into standard Shelx *.hklf4 format using the DC REDPD command. Additionally some corrections may be applied - typically spherical absorption or empirical absorption using the shape of the crystal. 1. After invoking the DC REDPD command, an option screen appears. It is necessary to select the relevant data collection *.DC1 file by clicking the "File name" button - an open file window will appear. 2. A number of options will be displayed. Typically the information in the data collection file is converted to Shelx format using default parameters. Processing may now be started by clicking on the process now button and the output will be in the form of *.hkl, *.sum and *.lst files. 3. If necessary, by clicking "Enable recalculation", all the reflections may be reintegrated and corrected using the following options: • • • • • Version 1.1 Attenuation factor – a correction is applied to reflections measured with attenuator on (see AF I command). The allowed range is 1 to 1000. Back/peak ratio - allows redefinition of the ratio between the number of scan steps used for background evaluation and the steps integrated as intensity. This means overriding the parameter set by the mo b command. This is the most common reason for needing to recalculate. Polarisation/Monochromator – this is useful only in cases when data were collected using the wrong option. The options are: e1-e3-plane (Xcalibur1), e1-e2-plane (MARHuber) and synchrotron. Typically e1-e3-plane (Xcalibur1). d-value of monochromator/Polarization factor - this is also only useful in cases where data were collected using the wrong option. The allowed ranges are 0.1 to 10000 but typically the value should be 0.98. Dead time corr/Dead time corr(mon) - this is also only useful in cases where data were collected using the wrong option. The ranges are 0 to 1 but typically the value should be 1.7 x 10-6 for both of these parameters. XcaliburPD v1.2.doc Page 21 XcaliburPD USER MANUAL OPERATION • Cut off multiplier – This will clean the output file by removing reflections below a certain I/sigma value. Tick the radio button to apply the cut-off and click on the box to edit the limit. Data may also be corrected for absorption: • • Abs. for empirical shape - EDIT/IMPORT SHAPE – this allows the user to apply a correction according to the shape defined using the abs display function. Click on the radio button to apply the correction and click on the box to open the absorption correction module and import the crystal shape model. Abs. for sphere correction (µr) – this allows the user to apply a spherical absorption correction. Click on the radio button to apply the correction and click on the box to edit the absorption coefficient of the crystal. NOTE The "Screen output" option can only be used in debug mode, and should be switched off. The reference reflection correction is on by default and rescales data collection for decay of crystal comparing intensities of reference reflections. 4. The hkl file format may be chosen. The options are: SHELX hkl F2 s(F2) b, SHELX hkl F2 s(F2) b dircos (which includes the direction cosines) and SHELX hkl F2 s(F2) psi schwaba (which includes Schwarzenbach psi information for use with the SCALE3 ABS command). By default the SHELX hkl F2 s(F2) b output format is selected. 5. Clicking "Process now" will start the data processing operation. Figure 4.5 Dataproc programme opened by using the dc redpd command Version 1.1 XcaliburPD v1.2.doc Page 22 XcaliburPD USER MANUAL OPERATION 4.7 Dc Movie - Replay of Data Collection Movie The CrysAlis software package enables the user to examine the whole of the data collection as a movie. The user can move back and forth through the reflections. 1. Type either dc redpd or dc moviepd and select the relevant *.dc1 file 2. A control window will appear. 3. Click on the Play button and a continuous movie will be played of the data collection. Select forwards or backwards play of the movie by the relevant arrow button. Select the hkl button to see a profile of the reflection with the chosen hkl value. 4. For each reflection, a number of parameters are displayed in the dialogue box: the hkl value, the goniometer angular settings for omega, theta, kappa and phi and the strength of the reflection (ST = strong, WK = weak and LT means not measured). 5. Click on Exit to finish data collection movie playback. Figure 4.6 DC MOVIEPD programme 4.8 Absorption Correction As X-rays pass through the crystal sample, a percentage of the X-rays will be absorbed by the sample. The degree of absorption is related to the distance travelled through the sample and also the composition of the sample. To minimise absorption a spherical crystal is ideal, however, this is often unobtainable. As a result the diffraction data is often corrected for absorption. Version 1.1 XcaliburPD v1.2.doc Page 23 XcaliburPD USER MANUAL OPERATION The absorption correction incorporated into the CrysAlis software package can be summarised into the following five steps: 1. Record a jpeg movie of the crystal sample 2. Build / modify a 3-D model of the crystal sample 3. Refine / optimise the 3-D model against the X-ray diffraction data 4. Examine 3-D model 5. Repeat steps 2-4 until a satisfactory model and absorption correction have been obtained. For details please refer to the main Xcalibur operation manual 4.9 GRAL - Space Group Determination The CrysAlis RED, GRAL plug-in wizard guides the user through space group determination. For full details, please refer to the main Xcalibur operator manual. Version 1.1 XcaliburPD v1.2.doc Page 24 XcaliburPD USER MANUAL GLOSSARY 5 Glossary of point detector commands Command abs display abs grab af i af o ce Meaning View movie recording and build absorption correction routine Initiate movie recording routine for absorption correction Applies an attenuation filter Removes the attenuation filter centre a peak Example abs display abs grab af i af o CE [n|HP] n - setting number: n=0 current angle setting - i.e. do not recover standard setting during refinement n = 1 to 8 standard settings; default – 1 da dc help dc applycorrections dc hkl dc movie dc visualizecorrections n = HP (string) carry out centring procedure in 8 symmetrical positions to calculate the crystal displacements from the sphere of confusion of the goniometer as described by A. Ross in Rev. in Miner. & Geochem. 41, 2000, 559-596 DA dh dv detector aperture setting help overview of dc commands Application of data reductions corrections on current frame dh = horizontal slit in deg dv = vertical slit in deg dc help dc applycorrections typeofcorrection ([L][P][A][W][S]) L = lorentz correction. P = polarization correction. A = air path correction. W = window absorption correction. S = scintillator correction. dc hkl dc movie dc visualizecorrections typeofcorrection ([L][P][A][W][S])scalefactor Import hkl file and apply outlier rejection. Data collection inspection Visualize data reduction corrections typeofcorrection L = lorentz correction. P = polarization correction. A = air path correction. W = window absorption correction. S = scintillator correction. scalefactor = scale factor with respect to the ideal image 1.0. dc s dc r dc red dc rrp dc unwarp dl Data collection start Data collection restart Data collection reduction (initiate guided routine) Data reduction finalization Reciprocal space reconstruction speed parameter setting dc s dc r [[filepath]filename] dc red dc rrp dc unwarp DL v N en ga End program and park goniometer Gain setting – amplification in the counting chain gt a goto angles (omega, theta, kappa, phi) g = gain for the counting chain. 0.0 - 150.0 gt a om th ka ph om = omega angle th = theta angle ka = kappa angle ph = phi angle gt d dd gt d Moves the CCD camera to a requested distance. gt e The gt e command has the same meaning as the gt a command, but its parameters are the Euler geometry setting angles (omega, 2theta, chi and phi) instead of kappa geometry setting angles (omega, 2theta, kappa and phi) Version 1.1 v [Deg]/[Sec] = phi scanning speed: 0.01 <= v <= 3.0 N = discrimination level: 0 < N <= 10000 En GA g dd – detector distance in mm GT E ome the chi phi ome = omega angle in deg. the = detector angle in deg. XcaliburPD v1.2.doc Page 25 XcaliburPD USER MANUAL GLOSSARY gt o goto omega chi = chi angle in deg. phi = phi angle in deg. gt o om gt t goto theta angle om = omega angle gt t th gt k goto kappa angle th = theta angle gt k ka gt p goto phi angle ka = kappa angle gt p ph gt r goto reflection ph = phi angle gt r h k l gt chi goto equivalent omega, theta, kappa, phi values for chi gt s Positions the goniometer at a specified symmetric setting gt help gt orient h = h indice k = k indice l = l indice gt c chi chi = chi angle in degrees. GT S curset goset curset = setting at current pos. goset = required setting for next move gt help gt orient Help overview of gt commands Initialise guided axial photo routine (once unit cell is known) Activation of the CrysAlis online help Activation of the optical alignment menu Stadi4 goniometer zero check help overview of gon commands Xcalibur goniometer initialisation Stadi4 goniometer synchronization Xcalibur goniometer re-initialisation high voltage of photomultiplier F1 F12 gon check gon help gon init gon sync gon reinit HV hv il ll Index limits Low level of the photomultiplier 100.0 - 1000.0 V Il hmin hmax kmin kmax lmin lmax LL ll [llm] ma b matrix boundary (PD) F1 F12 gon check gon help gon init gon sync gon reinit hv hv = high voltage for the counting [beam monitor] chain. KM4: ll [llm] = low level for the counting chain. 0.0 - 2.55 MA B (n(1,1)) (n(1,2)) (n(1,3)) (n(2,1)) (n(2,2)) (n(2,3)) (n(3,1)) (n(3,2)) (n(3,3)) or MA B <predef. set no> n - matrix elements -32 <= n (i,j) <= 32 mo b mode of background (PD) predef. set no: The following codes can be selected: 5: 4/mmm 7: -3m1 8: -31m 10 : 6/mmm 11 : m-3 12 : m-3m MO B (f) mo s mode of scan (PD) f - background parameter: 0.01 <= f <= 2.0 . MO S m s n1 [n2 Vmin] m = mode of scan: 0<=m<=2; 0=BPB scan; 1=continuous-integrative step scan; 2=stationary step scan; ph e Version 1.1 edit peak hunting table s = number of scan steps (not for m=0): 10<=s<=1024 n1 - minimum accepted I/sigma(I) and n2 = requested I/sigma(I): -3 <= n1 <= n2 <= 50 Vmin - Minimal rescan speed: 0.004768 [Deg]/[Sec] <= Vmin <= 1.5 [Deg]/[Sec] ph e XcaliburPD v1.2.doc Page 26 XcaliburPD USER MANUAL GLOSSARY ph extractprofiles ph help ph s ph reconstruct pt clear pt a pt e pt expand pt l extract profiles from data collection help overview of ph commands Peak hunting start reconstruct the peak table with current instrument model clearing of the peak table add a peak to the peak table Peak table edit Peak table expand ph extractprofiles ph help ph s ph reconstruct pt clear pt e pt expand n mmin mmax intmin dmax dmin add reflections to the peak table according to their Laue symmetry n = number of reflections required mmin = min order of difference -1 for CCD (-3 Point detector) mmax = max order of difference +1 for CCD (+3 Point detector) intmin = Intensity threshold dmax = max d spacing dmin = min d spacing PT L lauecode lauecode = The following codes can be selected: 1: -1 2: 2/m 3: mmm 4: 4/m 5: 4/mmm 6: -3 7: -3m1 8: -31m 9: 6/m 10 : 6/mmm 11 : m-3 12 : m-3m pt sa add reflections from a different setting position PT SA settingno settingno = The settingno refers to the basic measurement settings. rd help rd jpg help overview of rd commands JPEG image rd help rd jpg [filepath[filename]] [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. rd p rd ph rd t rd jpgheader refine model refine export rr Read parameter file rd p [filepath[filename]] Read peak hunting table from file [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. rd ph [filepath[filename]] Read peak table from file [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. rd t [filepath[filename]] JPEG image header [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. rd jpgheader [[filepath]filename.jpg] Refine diffractometer geometry model Export details of model refinement to file reference reflections (PD) [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. refine model refine export RR rr_num [interval om_tol int_tol [h k l at [h k l at [h k l at]]]] rr_num = number of reference reflections: 0 <= rr_num <= 3 if rr_num >0 : interval = interval between reference reflections measured: 1 <= interval <= 32000 om_tol = requested omega repeatability: 0.0025 [Deg] <= om_tol <= 10 [Deg] int_tol = allowed intensity fluctuations parameter (number of sigmas by which a fluctuation triggers the recentering): 1 <= int_tol <= 100 h k l = Miller indices of ref. reflection: -126 <= h,k,l <= 126 Version 1.1 XcaliburPD v1.2.doc Page 27 XcaliburPD USER MANUAL GLOSSARY scale3 abs scale3 pack script script help sc s sc t sc w setup detectortype setup help setup options setup setupfile sh o sh c sm ao sm eta Initialise absorption correction shape optimisation routine Initialise scale3pack data scaling plug-in start a script. A script is an ASCII file with the extension *.mac and contains a sequence of CrysAlis commands, which will be, executed one after another. at = attenuation filter use: 0,1 ( 0 = filter out , 1 = filter in ) scale3 scale3abs scale3 scale3pack script [[filepath]filename] help overview of script commands scan speed (PD) [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. script help SC S (Vo) scan type (PD) Vo = scan speed: 0.004768 [Deg]/[Sec] <= Vo <= 1.5 [Deg]/[Sec] SC T (Tx) (f) scan width Tx = threshold angle [deg]: 0.0 [Deg] <= Tx <= 90.0 [Deg] f = multiplier for theta motor: 0.0 <= f <= 2.0 SC W Oa Ob C detector type help overview of setup commands program options start-up file; setup file X-ray shutter open X-ray shutter closed Single measurement omega scan with angles Oa = scan width (omega): 0.0 [Deg] <= Oa <= 180.0 [Deg] Ob = scan widening parameter: 0.0 [Deg] <= Ob <= 2.0 [Deg] C = scan centre shift parameter: 1.0 <= C <= 1.1 setup detectortype setup help setup options setup setupfile sh o sh c sm ao om th ka ph scanwidth time step scan moving every reflection vertically in the aperture window sm help sm i time help overview of sm commands single image photo (static) of exposure time (secs) sm o sm p sm q omega scan phi scan records a q scan om = omega degrees th = theta degrees ka = kappa degrees ph = phi degrees scanwidth = degrees time = seconds SM ETA starteta step steps time starteta = starting eta value. Actual goniometer position is assumed to be theta 0.0. step = scanstep in deg. steps = number of steps for scan time = exposure time in sec per step sm help sm i time time = time in secs sm o startangle scanwidth time sm p startangle scanwidth time SM Q h k l st #s1 sw1 u1 v1 w1 [#s2 sw2 u2 v2 w2] [filename] HKL = indices of selected peak, for which the Omega and Theta angles will be assumed by the program as centre of the scanning area st = measurement time of one scanning point [sec] #s1 = number of scanning points in the [U1 V1 W1] direction sw1 = increment of scanning angle for [U1 V1 W1] direction [Deg] U1 V1 W1 = indices of the first scanning direction #s2 = number of scanning points in the [U2 V2 W2] direction sw2 = increment of scanning angle for [U2 V2 W2] direction [Deg] U2 V2 W2 = indices of the second scanning direction [filename] = optional filename. sm r measure a single reflection SM R h k l [psistart [psiend psistep]] h k l = reflection index, may be fractional. psistart = optional psi angle. psiend = optional psi start angle for measuring a sequence. psistep = optional psi step angle for measuring a sequence. sm s Version 1.1 single step scan SM S steps time XcaliburPD v1.2.doc Page 28 XcaliburPD USER MANUAL GLOSSARY steps = number of steps for scan time = exposure time in sec per step sm t system dos system explorer system help Qvector tr ty help ty l ty p ty t ty imageinfo um crec theta scan Open MSDOS prompt in current directory Open current windows directory Help overview of system commands Refine incommensurate q vector theta range sm t startangle scanwidth time system dos system explorer system help qvector mmax crithkl q1 q2 q3 [q1 q2 q3 [q1 q2 q3]] TR tthmin, tthmax help overview of ty commands Print lattice information to history Print parameter file to history Print peak table to history Print image information to history change orientation matrix in reciprocal space tthmin =minimal two-theta tthmax = maximum two-theta 0.0 [deg] <= tthmin < tthmax <= 180 [deg] ty help ty l ty p ty t ty imageinfo um crec [c11 c12 c13 .. c31 c32 c33] [c11 c12 c13 .. c31 c32 c33] = reciprocal space transformation matrix um clearskipd um help UM HPPOLYNOMIAL [d0 d1 d2 d3 d4 d5 d6 d7 d8] um clearskipd um help um hppolynomial clear skip list help overview of um commands um ip indexing with peak table printing d0-d9 = polynomial coefficients um ip [indexrejectioncriterion] indexing with real indices overlay skip list generate goniometer angles from peak table refine UB under symmetry constraint [indexrejectioncriterion] = rejection criterion which determines the maximum allowed deviation for which a reflection is considered indexed. um ir um overlayskipd um pointdetector um r [symmetrycode] set UB matrix or enter known orientation matrix [symmetrycode] = The necessary code can be obtained by just typing um r um s ub11 ub12 ub13 .. ub31 ub32 ub33 [sub11 sub12 sub13 .. sub31 sub32 sub33] um ir um overlayskipd um pointdetector um r um s um shape um showskipd um skipd um setqvector um f um i um c um reduce um r um sarray Version 1.1 View absorption correction model (wire frame less movie overlay) show skip list add item to skip list set incommensurate q-vector Automatic unit cell determination (indexation, refinement, reduction) ub11 ub12 ub13 .. ub31 ub32 ub33 = orientation matrix [sub11 sub12 sub13 .. sub31 sub32 sub33] = sigma of orientation matrix um shape um showskipd um skipd um setqvector q1 q2 q3 mmax q1 = component of q-vector along a*. q2 = component of q-vector along b*. q3 = component of q-vector along c*. mmax = maximum satellite order. um f [lengthdeviation angledeviation fractionindexed] Index and refine unit cell [lengthdeviation angledeviation fractionindexed] = The defaults are 0.05, 0.1, 0.7. You can loosen this condition in case of an unsuccessful indexing: typically 0.3 0.3 0.5. um i [indexrejectioncriterion] Change orientation matrix [indexrejectioncriterion] = rejection criterion which determines the maximum allowed deviation for which a reflection is considered indexed. um c [change #]|[c11 c12 c13 .. c31 c32 c33] Apply Niggli reduction of unit cell refine UB under symmetry constraint [change #] = transformation number from table obtained by typing um c: [c11 c12 c13 .. c31 c32 c33] – direct space transformation matrix um reduce um r [symmetrycode] Store the defined ub unit cell matrix to a buffer array. 8 available. Used to define a number of [symmetrycode] – The necessary code can be obtained by just typing um r Um sarray arraynum ub11 ub12 ub13 .. ub31 ub32 ub33 XcaliburPD v1.2.doc Page 29 XcaliburPD USER MANUAL GLOSSARY ub matrices for data reduction of twinned data. um u wd i wd flood wd t wd p wd cal wd help wd inc wd ph wd t activates the procedure to refine the crystal orientation automatically Write disc image Arraynum (0..7) storage buffers for unit-cell matrices Ub11 … ub33 unit-cell ub matrix um u wd i [[filepath]filename] Save current image as flood image Write peak table to file [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. wd flood wd t [[filepath]filename] Write parameter file [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. wd p [[filepath]filename] Save current parameters into current setup file help overview of wd commands Save current image non-compressed [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. wd cal wd help wd inc [[filepath]filename] Save peak hunting table (raw profiles) [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. wd ph [[filepath]filename] [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. wd t [[filepath]filename] Save peak table (indexed xyzs without profiles) [[filepath]filename] – Optionally you can put the path and filename on the command line. Note that you have to use quote for filenames with spaces. WI wi [wim] wi window height of the analyser zc a finds the theta zero and the equator/horizon of the machine wi [wim] = window level for the counting chain: 0.0 - 2.55 ZC A [h k l] [h k l]: (optional) Indices of the reflection to use, otherwise the programme will use the current angular setting. Version 1.1 XcaliburPD v1.2.doc Page 30 XcaliburPD USER MANUAL APPENDIX 1 – HIGH PRESSURE WORK Appendix I - High Pressure Crystallography using a DAC on an Xcalibur system This is intended as a short guide to setting up a DAC for a data collection on an Xcalibur diffractometer. The user is assumed to be familiar with the CrysAlis software and commands and their use for data collections from crystals in air. In the following documentation, commands to be typed into the command line of the Crysalis GUI are indicated thus: gt r 4 0 0. Command line entries where numerical values should be substituted are indicated by italics, thus: gt r h k l. The procedure for different system configurations are outlined, Xcalibur1 and Xcalibur2. Both have interchangeable point detector and area detector. The Xcalibur 1 diffractometer is equipped with a dual detector arm (see below). This diffractometer has a dovetail slide for mounting a CCD camera. When the point detector is used for DAC data collections this slide is used to hold a set of additional collimation slits as shown in the picture. The Xcalibur2 system, (and all the other, more recently developed Xcalibur configurations) has a universal theta arm mount which allows easy interchange between point detector and CCD detector. In addition to the normal equipment provided as part of the Xcalibur diffractometer system you will also need: • a small spirit level slightly larger than the size of your DAC to align the DAC. • For the Xcalibur1 system, a dial gauge mounted on a stand so that its axis is horizontal and at the same height as the centre of the goniometer. Used for adjusting the position of the DAC along the beam. The centering of the DAC may also require the use of the WinIntegrStp software available from /www.crystal.vt.edu/crystal/software/ I.1 Preparation Create a filespace for the data collection. Open a file browser and create a new directory for this data (e.g. \P1). Start the CrysAlis program by double-clicking on the desktop icon for CrysAlis CCD Check that the software is set for point detector operation (GUI should have angles and scan display). If the software is in CCD mode, change to PD mode using the following procedure: VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK 1. Select Tools|Setup file/Xcalibur1PD.par 2. Exit from Tools|Setup 3. Exit from program (en) 4. Restart Crysalis CCD from desktop Switch to DAC mode (sw s 2) and set the DAC opening angle to 40deg (or to the value corresponding to your DAC (sw a angle). On the Xcalibur 1 diffractometer, remove the slit assembly from the sled on the dovetail by unscrewing three screws. On the Xcalibur2 diffractometer, ensure that the correct short collimator and long beam stop are installed together with the correct detector limit flag. I.2 Physical alignment of DAC Xcalibur 1 Drive the diffractometer to the alignment position (gt a 0 21.75 0 0). Load the DAC onto the diffractometer and tighten the base screw firmly. Align the DAC by eye, perpendicular to the beam: Loosen the locking screw for the height adjustment on the goniometer head and rotate the cell until it looks perpendicular to the beam direction. Accurately align the DAC perpendicular to the beam. Slide the sled on the dovetail to the back of the dovetail. Mount the aluminium alignment tool on the sled. Carefully slide the tool in to touch the DAC. Rotate the DAC until the face of the DAC is exactly parallel to the end of the alignment tool. Gently tighten the height locking screw on the goniometer head. Remove the alignment tool. Slide the sled back on the dovetail and lift off the alignment tool, taking care not to hit the beam stop. Set the focus of video microscope and view an image of the cell. Loosen the locking screw of video camera and move it to focus (gt e 0 0 90 -90). If image not in focus, adjust the half way to focus with the goniometer head slide, and half with camera adjustment. Repeat until cell is in focus at both of these two positions .Set the height of the DAC (gt e 0 0 90 90). Observe the position of gasket hole centre on video screen. (gt e 0 0 –90 –90). Compare position of gasket hole and adjust height. Repeat until image of gasket hole does not move vertically between these two positions. Tighten the height locking screw. Check that DAC is still perpendicular to beam at zero (gt a 0 21.75 0 0) See the instructions above for using alignment tool on sled. Correct alignment if necessary. Set cell translation across the beam (x direction) (gt e 0 0 90 –90). Observe position of centre of gasket hole (gt e 0 0 90 90). Compare position and adjust with slide on goniometer head. Repeat until image of gasket hole does not move between these two positions. Tighten slide locking screw. The DAC may be additionally centred along the beam by use of the beam rocking method described below. VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK Xcalibur 2/Xcalibur 3 etc. Follow the procedure as above except use the alignment position using the F12 key and the remote pilot. Align the DAC by eye so that it is perpendicular to the beam. Align the aperture in the XY direction using the video microscope and then align along the beam axis as above or use the beam rocking method below. For systems where the video microscope is aligned vertically, use the following positions to centre the DAC instead of the remote pilot: Gt e 0 0 90 90 Gt e 0 0 90 -90 Gt e 0 0 -90 -90 Beam rocking method Accurate alignment of a Diamond anvil cell is essential in order to acquire good quality high pressure data. The Xcalibur system allows very easy alignment of the diamond anvil cell along the beam axis using the ‘beam rocking’ method devised by Ross Angel and Mathias Meyer. • Flip beamstop out of the way • Mount the DAC and align the X-Y direction by centring the aperture using the video microscope. • Record and save images at phi = -30 degrees and +30 degrees (gt p -30, sm I 0.1, wd I, gt p 30, sm I 0.1, wd i) • Subtract one image from the other. (rd I, ip copy i2 i1, rd I, ip subtract i3 i2 i1) to observe a difference map of the two images. • Make small adjustments to the beam axis and repeat. When the alignment is complete, the difference map will be flat. I.3 Determine the initial orientation matrix There are two possibilities, depending on the stage of the high-pressure experiment: 1. The UB from a measurement on the Xcalibur diffractometer at a previous pressure is known 2. The UB from another diffractometer is known 3. The UB is not known. VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK I.3.1 UB matrix known from previous measurement Copy the peak table from the previous measurement. Use the Windows file browser to copy the *.tab file to the current working directory (rd t). Or input values into the command line (um s u11 u12 u13 u21 u22 u23 u31 u32 u33). Check the table is correct by calculating lattice parameters from UB (ty l). I.3.2 UB matrix unknown Switch to the CCD camera and perform a short data collection. Find peaks, index and save a peak table. Switch back to the point detector Go to step I.3.3 I.3.3 Look for reflections Insert the pair of slits labeled 3.3 (horizontal) and 2.0 (vertical) with notches towards the back of the cabinet. • Set slit values in software da 3.3 2.0 • Drive to a strong reflection gt r h k l • Scan the position....set the scan width sc w 2.0 0.0 • Do the omega scan sm s 40 0.1 If the maximum is in the scan, check two more reflections. If all ok, go to step 4. If maximum is not in the scan, drive around in omega and rescan until you find it (gt o omega) (omega should be shifted by 1 deg) (sm s 40) 0.1. When a peak is found, drive to position of maximum (gt o omega), centre the reflection (ce 0) and add reflection to peak table (pt a) Look this reflection up in the peak list (pt e). Select it, edit it, insert the correct hkl values, exit the list and save the table (wd t). Repeat for another reflection and do two-reflection calculation of UB (um f2 a b c α β γ), where abc and α β γ are the estimated cell parameters. Check the result: If the lattice parameters change a lot, then your indexing was incorrect; change the indexing and try again. When you have a valid UB, save it by typing (wd t). Check the UB finds other reflections: Drive to a strong reflection (gt r hkl) then see if it is in the detector window (sm i 1 (or F7) and observe counts) I.4 Refine UB Enter the peak table window by typing pt e. Make a note of the hkl and then delete all reflections (see note at end of this table). Insert the hkl of one of each symmetry-equivalent set VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK Exit from the peak table editor and expand the peak list by Laue symmetry (pt l n) (if you do not know n for your Laue group then type pt l to obtain a list). At this stage you need 20-30 strong reflections (pt e). Edit the peak list to remove reflections with low kappa angle: • Select “angles” at bottom of display • Click on “kappa” column header to order reflections • Delete reflections with –15o < κ < 15o. • Exit the peak list editor • Save the peak list (wd t) • Start reflection centering (um u) Note: If a reflection in the list cannot be centered it will be skipped and deleted from the list. If many reflections are skipped then either the UB matrix and the cell parameters are wrong, or the software has the wrong values loaded (with da) for the detector slits. To recover from this problem, read the original table back into software, rd t. Enter the correct slit values da hslit vslit and repeat centering um u. At the end of centering, the UB is determined. Save it and the peak positions wd t. Switch to smaller slits, h=2.0, v=1.0 and update values in software: da 2.0 1.0. Check the UB finds reflections with smaller slits: Drive to a strong reflection and see if it is in the detector window gt r h k l. Record a still image sm i 1 (or F7) and observe counts. If ok, repeat centering um u. At the end of centering, the UB is determined. Save it and the peak positions wd t Note on the peak table: The Crysalis centering procedure um u works by first driving to the angular positions given in the peak table. This is different from the Single software in which the starting position for centering is calculated from the hkl in the peak table, and the current UB matrix. This means that in Crysalis when the UB is changed significantly, the peak table must be cleared and the indices of reflections be reloaded into the table; this procedure ensures that the peak positions are calculated from the current UB. I.5 Determine crystal offsets At this stage the gasket hole of the DAC has been well-centered optically across the beam, but the positioning along the beam has relied on focusing the video microscope on the sample. The centering along the beam can be improved by “diffracted beam centering”. There are two ways to achieve this: 1. By 8-position centering of a single reflection with a Eulerian-chi value between 80o and 90o. 2. By collecting data scans of 30 or more low-angle reflections and refining the crystal offsets by the method of Dera and Katrusiak (1999, Journal of Applied Crystallography 32:510-515). VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK Method 1 takes less time, but method 2 is often more reliable. Both alternatives are described below: Method 1 - 8-position centering • Find a strong reflection with Eulerian chi > 80o gt r h k l • Take a still image sm i 1 • Do 8-position centering ce HP • Adjust goniometer position according to offsets from ce HP procedure • Repeat until offsets are small or zero Method 2 - Crystal offsets from data collection Set up parameters for a short low-angle data collection, as follows. Set detector slits to h=2.0, v=0.5 and exchange the brass slits on the detector Update values in software: da 2.0 0.5 Set scan parameters to stop rescanning mo s 1 60 10 0 0.005 Set fast scan speed sc s 0.05 Set background calc mo b 0.5 Set scan width sc w 1.200 0.000 1.00300 Set omega scan sc t 0.0 0.0 Set index limits to cover all reciprocal space Note: this sets the maximum values of indices to be tested against 2theta limits etc. Just make them sufficiently large il –10 10 –10 10 –10 10 Clear ma limits ma b 0 0 0 0 0 0 0 0 0 Set 2theta limits tr 0 25 Set absence conditions rc Clear reference reflections rr 0 Check all values are correct ty p Check number of reflections that will be collected dc t Adjust 2theta limits until you have 30-50 reflections to be collected VTX tr tthmin tthmax Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK Save parameters (it is useful to call this something like orient.par) wd p Start data collection (use a distinct filename such as orientn) where n indicates the iteration through this process dc s When data collection is complete, export the data to a dca file Open the Crysalis Reduce software and type dc redpd on its command line. Select your data file and convert to Ascii. Open the WinIntegrStp program by double-clicking on desktop icon. Select the dca file you just created. Select Xcalibur.par as the instrument parameter file. Run preprocessing option to obtain the peak positions • Set I/sigma to 10.0 • Set Intensity, peak width, position to be refined • Set background to not refined, with default value “D” • Set eta and Iratio to not refined • Start preprocess with “Go” If insufficient (<20) reflections are stored after Preprocess, reduce I/sigma or adjust the test limits on the parameters. Once you have >20 reflections stored from Preprocess, calculate the UB (Utilities|Calc UB). Select refine crystal offsets and run. Record the crystal offsets reported (in mm). The X and Z offsets should already be small (<50micron). The Y offset is along the beam. If it is less than 30 micron go to step I.3.6. If Y offset >30 micron proceed as follows: • Drive goniometer to zero gt a 0 0 0 0 • Place the dial gauge in contact with the downstream face of the cell. • Adjust the cell position along the beam One division on the dial gauge is 25 micron. • If the Y offset is positive, move the DAC towards the X-ray tube. • If the Y offset is negative, move the DAC away from the X-ray tube. • Repeat step 5.2 until Y offset is <30 micron. I.6 Data Collection Re-determine the UB matrix: VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK Exchange brass slits on detector. Update values in software: da 1.33 0.5 then um u. Save peak table to disk wd t. On the Xcalibur 1 system, Install the additional slits on detector arm. Screw down the slits onto the carrier on the dovetail. Set the slide to 9.55. Set the detector slits to h=2.0, v=0.5. Exchange the brass slits on detector and update values in software: da 2.0 0.5 Set up parameters for the data collection as follows: Set scan parameters mo s 1 60 1 10 0.005 Set fast scan speed sc s 0.05 Set background calc mo b 0.5 Set scan width sc w 1.200 0.000 1.00300 Set omega scan sc t 0.0 0.0 Set index limits to cover required portion of reciprocal spaceNote: this sets the maximum values of indices to be tested against 2theta limits etc. Just make them sufficiently large. il hmin hmax kmin kmax lmin lmax Set ma limits if required ma b n Set 2theta limits tr thmin thmax Set absence conditions rc Set reference reflections rr 3 200 0.15 15.0 h k l 0 h k l 0 h k l 0 Check all values are correct ty p Check number of reflections to be collected dc t Check that parameters and UB are ok by scanning several reflections sm r h k l Save parameters wd p Start data collection dc s I.6 When Data Collection Is Completed • Open the Crysalis Reduce software • type dc redpd on its command line • Select your data file • Select Convert to Ascii • Open the WinIntegrStp program by double clicking on desktop icon • Select the dca file you just created VTX Xcalibur DAC Data Collection APPENDIX 1 – HIGH PRESSURE WORK • Select Xcalibur.par as the instrument parameter file. • Check the scans are ok and centered; Use Integrate | Manual Profile Fit to review the dataset • Integrate the data and refine the structure VTX Xcalibur DAC Data Collection
advertisement
Key Features
- Point detector operation
- Data collection parameters
- Peak hunting
- Unit cell determination
- Data processing
- Absorption correction
- Space group determination
Related manuals
Frequently Answers and Questions
What is the purpose of the Xcalibur Series?
The Xcalibur Series is designed for single-crystal X-ray diffraction analysis to determine crystal structures and analyze molecular properties.
What kind of detector does the Xcalibur Series support?
It supports both CCD and point detectors, offering flexibility for various crystal types and data collection needs.
How do I align a crystal for data collection?
The manual provides step-by-step instructions for optical alignment of the crystal using the goniometer.
What are the main steps involved in a standard point detector experiment?
The steps include crystal mounting, alignment, parameter setting, peak hunting, unit cell determination, data collection, processing, absorption correction, and space group determination.
advertisement