POSITIONING HOLE
12
≥
1786-Guide HUBER 360-MD-GB-VECT.indd 12
18/02/15 09:26
>>
13
1786-Guide HUBER 360-MD-GB-VECT.indd 13
18/02/15 09:26
>>
14
1786-Guide HUBER 360-MD-GB-VECT.indd 14
18/02/15 09:26
>>
15
1786-Guide HUBER 360-MD-GB-VECT.indd 15
18/02/15 09:26
>>
16
≥
1786-Guide HUBER 360-MD-GB-VECT.indd 16
18/02/15 09:26
>>
17
1786-Guide HUBER 360-MD-GB-VECT.indd 17
18/02/15 09:26
>>
18
1786-Guide HUBER 360-MD-GB-VECT.indd 18
18/02/15 09:26
>>
19
1786-Guide HUBER 360-MD-GB-VECT.indd 19
18/02/15 09:26
>>
20
1786-Guide HUBER 360-MD-GB-VECT.indd 20
18/02/15 09:26
>>
>>
21
1786-Guide HUBER 360-MD-GB-VECT.indd 21
18/02/15 09:26
>>
22
1786-Guide HUBER 360-MD-GB-VECT.indd 22
18/02/15 09:26
>>
≥
1786-Guide HUBER 360-MD-GB-VECT.indd 23
23
18/02/15 09:26
>>
WARNING : KEEP THESE INSTRUCTIONS
24
1786-Guide HUBER 360-MD-GB-VECT.indd 24
18/02/15 09:26
>>
≥
1786-Guide HUBER 360-MD-GB-VECT.indd 25
25
18/02/15 09:26
>>
26
1786-Guide HUBER 360-MD-GB-VECT.indd 26
18/02/15 09:26
>>
>>
≥
1786-Guide HUBER 360-MD-GB-VECT.indd 27
27
18/02/15 09:26
>>
28
1786-Guide HUBER 360-MD-GB-VECT.indd 28
18/02/15 09:26
>>
29
1786-Guide HUBER 360-MD-GB-VECT.indd 29
18/02/15 09:26
>>
30
1786-Guide HUBER 360-MD-GB-VECT.indd 30
18/02/15 09:26
>>
31
1786-Guide HUBER 360-MD-GB-VECT.indd 31
18/02/15 09:26
>>
32
1786-Guide HUBER 360-MD-GB-VECT.indd 32
18/02/15 09:26
>>
33
1786-Guide HUBER 360-MD-GB-VECT.indd 33
18/02/15 09:26
34
1786-Guide HUBER 360-MD-GB-VECT.indd 34
18/02/15 09:26
>>
>>
35
1786-Guide HUBER 360-MD-GB-VECT.indd 35
18/02/15 09:26
>>
36
1786-Guide HUBER 360-MD-GB-VECT.indd 36
18/02/15 09:26
>>
TABLE 1: DIRECTIVES AND MANUFACTURER DECLARATION – ELECTROMAGNETIC EMISSIONS
HUBER 360 MD is intended for use in the electromagnetic environment specified below. The HUBER 360 MD customer or patient should ensure
that it is used in such an environment.
Emissions test
Conformity
Electromagnetic Environment - Directives
RF emissions CISPR 11
Group 1
HUBER 360 MD uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and unlikely to cause
interference in nearby electronic devices.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
HUBER 360 MD may be used in all establishments, including
domestic sites and sites that are directly connected to the low
voltage public power grid, which supplies domestic buildings.
Voltage fluctuations and flicker IEC 61000-3-3
Conforms
TABLE 2: DIRECTIVES AND MANUFACTURER DECLARATION – ELECTROMAGNETIC IMMUNITY – EMISSIONS TESTS
HUBER 360 MD is intended for use in the electromagnetic environment specified below. The HUBER 360 MD customer or patient should ensure
that it is used in such an environment.
Emissions test
Conformity
Conformity
Electromagnetic Environment - Directives
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV on contact
± 8 kV in the air
± 6 kV on contact
± 8 kV in the air
Flooring should be made of wood, concrete, or ceramic tile.
If the floor is covered with synthetic material, relative humidity
should be at least 30%.
Fast transient/burst
IEC 61000-4-4
± 2 kV for electrical
power lines
± 1 kV for input/output
lines
± 2 kV for electrical
power lines
± 1 kV for input/
output lines
The quality of the electrical power network should be equivalent
to that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV between
phases
± 2 kV between phase
and earth
± 1 kV between
phases
± 2 kV between phase
and earth
The quality of the electrical power network should be equivalent to
that of a typical commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations
IEC 61000-4-11
<5% UT
(>95% voltage dip)
during 0.5 cycle
40% UT
(60% UT voltage dip)
during 5 cycles
70% UT
(30 % UT voltage dip)
during 25 cycles
<5% UT
(>95% voltage dip)
during 5s
<5% UT
(>95% voltage dip)
during 0.5 cycle
40 % UT
(60% UT voltage dip)
during 5 cycles
70% UT
(30 % UT voltage dip)
during 25 cycles
<5% UT
(>95% voltage dip)
during 5s
The quality of the electrical power network should be equivalent to
that of a typical commercial or hospital environment. If the HUBER
360 MD patient requires operation to continue during power outages,
the HUBER 360 MD should be powered through an uninterruptible
power supply or a battery.
Power frequency
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should have levels that are
characteristic of a typical hospital or commercial environment.
NOTE: UT is the AC voltage prior to the application of the test level.
37
1786-Guide HUBER 360-MD-GB-VECT.indd 37
18/02/15 09:26
>>
TABLE 3: DIRECTIVES AND MANUFACTURER DECLARATION – ELECTROMAGNETIC IMMUNITY – IMMUNITY TESTS
HUBER 360 MD is intended for use in the electromagnetic environment specified below. The HUBER 360 MD customer or patient should ensure
that it is used in such an environment.
Immunity
Test
Test level
according to IEC
60601
Conformity
level
Electromagnetic environment - directives
Conducted RF
disturbances
IEC 61000-4-6
3 Veff
150 kHz to 80 MHz
3V
Perturbations
RF rayonnées
CEI 61000-4-3
3 V/m
De 80 MHz
to 2,5 GHz
3 V/m
Portable and mobile RF communications devices should not be used closer to any part
of the HUBER 360 MD, including its cables, than the recommended separation distance,
calculated from the applicable equation on transmitter frequency.
Recommended separation distance:
d = 1.2√P
d = 1.2√P 80 MHz at 800MHz
d = 2.3√P 800 MHz at 2.5 GHz
where P is the transmitter’s maximum power output in watts (W), according to the
transmitter’s manufacturer, and d is the recommended separation distance in meters (m).
The field strength for fixed RF transmitters, as determined by an electromagnetic survey
at site a, should be less than the conformity level for each range of frequencies b.
Interference may be produced near the device marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the highest frequency range applies. NOTE 2: These directives may not apply in all situations.
Electromagnetic propagation is affected by absorption and by reflections of structures, objects, and people.
a The field strength of fixed transmitters, such as radio/telephone (cellular/wireless) base stations and land mobile radios, amateur radios,
AM and FM radio broadcasting, and TV broadcasting cannot be theoretically predicted with accuracy. To measure the electromagnetic
environment due to fixed RF transmitters, consider performing an electromagnetic survey of the site. If the field strength, measured where
the HUBER 360 MD is used, exceeds the applicable RF conformity level above, watch the HUBER 360 MD to ensure that it is operating as normal.
If there is any abnormal performance, additional measures may be necessary, such as reorienting or repositioning
the HUBER 360 MD.
b For the frequency range of 150 kHz to 80 MHz, field strength should be less than 3 V/m.
TABLE 4: RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATION DEVICES
AND HUBER 360 MD
HUBER 360 MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The HUBER 360
MD patient or customer can prevent electromagnetic interference by maintaining a minimal distance between the portable or mobile RF
communications device (transmitter) and the HUBER 360 MD, as recommended below, based on the maximum transmission power of the
communication device.
Maximum rated
output of the
transmitter W
d=1,2√P
Separation distance according to the transmitter frequency m
d=1,2√P
d = 2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters whose maximum rated output is not shown above, the recommended separation distance d in meters (m) can be estimated using
the applicable equation for the transmitter frequency, where P is the transmitter’s maximum output in watts (W), according to its manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the highest frequency range applies.
NOTE 2: These directives may not apply in all situations.
Electromagnetic propagation is affected by absorption and by reflections of structures, objects, and people.
38
1786-Guide HUBER 360-MD-GB-VECT.indd 38
18/02/15 09:26
TECHNOPARC DE LA PLAINE
30, RUE DU DR. ABEL - CS 90035 - 26902 VALENCE CEDEX 09 - FRANCE
TEL.: +33 (0)4 75 78 69 00 - FAX: +33 (0)4 75 42 80 85
INTERNATIONAL/MARKETING
ECOLUCIOLES – BAT A
955 RTE DES LUCIOLES - BP 243 - 06905 SOPHIA-ANTIPOLIS - FRANCE
TEL.: +33 (0)4 92 38 39 00 - FAX: +33 (0)4 92 96 09 65
1786-Guide HUBER 360 i-GB-VECT.indd 36
1786 - Ed 01/24 - GB - © LPG Systems 2024 - LPG Systems S.A.S. au capital de 6 404 850 euros - Siren RCS 335.183.836 Romans - Any reproduction – even in part – is strictly prohibited. Non-binding photos.
HEADQUARTERS LPG SYSTEMS S .A .S .
18/02/15 09:08
">