Heraeus UT 6 P, UT 12 P, UT 20 P Laboratory Air Circulation Oven INSTRUCTION MANUAL
Below you will find brief information for Laboratory Air Circulation Oven UT 6 P, Laboratory Air Circulation Oven UT 12 P, Laboratory Air Circulation Oven UT 20 P. These Heraeus ovens are designed for laboratory applications such as drying, aging, analysis, fusing, burning in, oxidizing, reducing, preheating, and hot-air sterilization. They feature a temperature range of +20°C to 250°C, a programmed controller for precise temperature control and time settings, and a fresh-air flap for controlling the fresh-air supply to the inner compartment. The ovens are also equipped with an overtemperature protection device for added safety.
Advertisement
Advertisement
(GET) INSTRUCTION MANUAL
Heraeus
INSTRUMENTS
Laboratory Air Circulation Ovens
Models UT 6 P, UT 12 P, UT 20 P
Comply with the operating instructions and keep this manual in the vicinity of the unit I
Issue: May, 1994
50 042 750
Heraeus Instruments GmbH
Heraeusstrafie 12 -14
D- 63450 Hanau
DD
National sales
International sales
Technical support
National sales
International sales
Technical support
Spare parts
++ (0 61 81) 3 5 - 3 00
++ (0 61 81) 35- 4 22
++ (0 61 81) 35- 53 05
++(0 61 8 1 ) 3 5 - 5 9 73
+ + (0 61 8 1 ) 3 5 - 7 49
++ (0 61 81) 3 5 - 8 80
++ (0 61 81) 3 5 - 6 67
O Heraeus Instruments GmbH. 63450 Hanau, Germany function line* is a registered trademark of Heraeus Instruments GmbH
The German version of this Instruction manual applies in the event of any deviations in the translation
Nominal charge
This instruction manual applies for the following models:
Order No.
50 042 298
Model
UT6P
Equipment
Air circulation oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
50 042 300
SO 042 306
50 042 312
UT6P
UT12P
UT20P
Air circulation oven, programmed controller, 1/PE AC, 120 V; 50/60 Hz
Air circulation oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
Air circulation oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
These heating cabinets must be operated in accordance with the instruction manual and
may only be used for their intended applications in order to ensure the safety of personnel, the environment and the processed items and materials.
Read the instruction manual carefully before using the equipment and comply with the instructions that it contains to avoid making mistakes and to prevent any personal injury or
damage to property.
2/34 50 042 750
Issue: May, 1994
CONTENTS
1. GENERAL SAFETY INSTRUCTIONS 4 - 6
Explanation of icons 4
General information 5
Operating instructions 5
Field of application 5
Safely instructions 6
2. SETUP AND INSTALLATION 7 - 9
Transport 7
Unpacking the unit 7
Installation 7
Room ventilation 8
Mains connection 8
Noise insulation 8
Connecting to an air extraction system 9
3. UNIT SPECIFICATION 10-17
Overview of the control elements 10
Switchgear unit 10
On / Off switch 10
Overtemperature protection device 10-11
Fresh-air flap control 11
Automatic control unit 12
Signal lamps, status function 12
Basic functions: temperature control, time switch, air speed adjustment 12
Added-feature programming functions, fixed programs 13
Fixed program segments 14
Compilation of a fixed program 14
Example of a fixed program 15
Programmed controller function 16
Termination commands 16
Example of a temperature/time program, definitions 16
What happens If... ? 17
4. OPERATION 18-28
Starting up with
Basic functions - temperature control, time switch, blower speed adjustment 18-19
Added-feature programming functions, example 2 0 - 2 1
Programmed controller functions, example 2 2 - 2 7
Operating guidelines 28
Loading 28
Shutdown 28
5. MAINTENANCE 2 9 - 3 0
Decontamination / disinfection 29
Cleaning 29
Servicing 30
Tests 30
Repairs 30
Approved spare parts / accessories 30
6. TECHNICAL DATA 31-34
Issue: May, 1994 50 042 750 3/34
GENERAL SAFETY INSTRUCTIONS
EXPLANATION OF ICONS
This symbol marks chapters and sections of this instruction manual which are particularly
relevant lo safety.
When attached to the unit, this symbol draws attention to the relevant section of the
instruction manual.
Marks information about optimum utilization of the unit in the instruction manual.
Warns of hot surfaces.
set I
Key for temperature selection, time-switch function and to confirm entered information
prog
Key for the programming function, lime selection and to confirm entered information
Key to set the air speed and to confirm entered information
vent
Value adjustment key, INCREASE VALUE
Value adjustment key, REDUCE VALUE
Information about proper disposal / recycling
Overtemperature protection device
: resh-alr flap control
Continuous adjustment from minimum to maximum values
Quick reference guide to operation o
Select the status of the control unit (enabled / disabled)
Temperature control function, select temperature
(T)
Time-switch function, sel the operating times
Programming function
Blower function
4/34
50 042 750 Issue: May, 1994
GENERAL SAFETY INSTRUCTIONS
General information
The unit fulfils the following safety standards:
DIN EN 61 0 1 0 - 1 / 0 3 . 9 4 . VDE 0411 Part 1 / 0 3 / 9 4 , E D I N V D E 0411 Part 1 1 1 .
DIN 1 2 880 Part 1 / 1 1 . 7 8 and DIN E N 6 0 335 - 1 , V D E 0 7 0 0 Part 1
DIN EN 55 011 Part 3, DIN VDE 0875 Part 11 / 07. '92
"Q ' Always quote the data on the rating plate w h e n requesting information or ordering spare parts
A
Operating instructions
The operator must provide anybody working on or with this equipment with written instructions for any operations to be performed. Such instructions must be easy to understand and must be available in the language of the respective personnel (FRG: UW V B G 1 § 7 (2)).
• Log book
We recommend that a log book is kept, which records any tests and calibration operations performed on the unit and any other work (repairs, modifications etc.).
Field of application
The unit has been designed as an item of laboratory equipment for technical applications.
• Heat treatment of samples or materials with operating temperatures of between room temperature + approx. 20*C a n d 250 ' C , e.g. for drying, ageing, analysis, fusing, burning In, oxidizing, reducing, preheating, etc.
• Hot-air sterilization, sterilization by heating the article up to a temperature of at least 160 'C
(FRG: also refer to DIN 58 947).
Note. Th« ilerillzation temperature usually required in Ihe fleW of medical laboratory work l> 110 *C.
The unit has been designed for installation a n d operation in the following fields:
• Laboratories, e.g. in the commercial or industrial sector, schools, universities, hospitals and biology
The unit is equipped with an independent overtemperature protection device, which means that it does not require constant supervision.
The heating unit has been designed for continuous use.
Make sure that an adequate clearance is maintained between the inside walls of the inner compartment and the loaded articles, and between the various layers of loaded articles, to prevent uneven heating of the materials.
Protective gloves, goggles etc. must be m a d e available to the operator because of the potentially high operating temperatures.
The unit must not be used to dry or heat substances w h i c h may release combustible gases or vapours into the atmosphere, w h i c h could b u m or explode w h e n mixed with air.
The heating unit is equally unsuitable for the heat treatment of combustible dusts or fibrous materials.
Issue: May, 1994 50 042 750 5/34
1 GENERAL SAFETY INSTRUCTIONS
Safety instructions
Comply with the Instructions in this manual and keep it In the vicinity of the unit.
These units must be operated in accordance with the Instruction manual and may only be used for their Intended applications in order to ensure the safety of personnel, the environment and the processed items and materials.
Read the Instruction manual carefully and comply with the instructions that it contains to avoid making mistakes and to prevent any peronal injury or damage to properly.
Unit must be operated by instructed personnel.
Keep unit out of reach of children.
Apart from the Instructions in this manual, the respective national regulations must also be observed for the installation and operation of this unit (FRG: ZH 1/119, DIN 12 880 Part 1)
The applicable national environmental regulations must be observed for the extraction of exhaust gases that are produced during heat treatment. Suitable measures must be Implemented to ensure that such gases are safely led outside (FRG: BlmSchG, U V P G , AbfG, W H G , ChemG, ...).
The unit must not be used to dry or heat substances which may release combustible gases or vapours Into the atmosphere, which could bum or explode when mixed with air. The unit is equally unsuitable for the heat treatment of combustible dusts or fibrous materials.
Check the mains lead and connector for damage before using the unit. If there are any signs of damage, do not connect the unit up to the mains.
The voltage quoted on the rating plate (rated voltage) must agree with the mains supply voltage.
Protective gloves, goggles etc. must be made available to the operator because of the potentially high operating temperatures.
The surfaces around the opening of the Inner compartment become hot at high operating temperatures • RISK OF BURNING - Avoid contact with these surfaces.
Ensure that the overtemperature protection device works properly by conducting a functional check at regular Intervals, every 3 months at least, and check the electrical equipment at least once a year.
Any work to be conducted on the electrical equipment of the unit may only be performed by a qualified electrician. The unit must be disconnected from the mains supply before commencing service or repair work.
Only use approved accessories and approved genuine spare parts. The use of any other parts may result In unforeseen problems and should be avoided under all circumstances.
The serviceability and safety of the unit can only be guaranteed if the necessary tests, maintenance, servicing and repair work Is carried out by the Heraeus service personnel or other agencies who are authorized to act on our behalf.
Heraeus Instruments GmbH cannot accept any liability for any damage that occurs as a result or improper use or repair work, which has not been performed by Heraeus service centers, or if parts other than the approved genuine spare parts / accessories are used.
6/34 50 042 750 Issue: May, 1994
SETUP AND INSTALLATION
Transport
Handle the unit with care. Avoid jarring and protect against damage. Do not lift by the door handle, the door itself, the control unit, the blower box, the spacer or the power lead.
Refer to the TECHNICAL DATA for dimensions and weight.
• Unpacking the unit:
Unpack the unit, remove the transport safety fixtures from the inner compartment and remove any dirt or dust which may have accumulated in transit.
Installation
The unit must be installed in a dry environment.
The relative humidity should be between 60 and ZO % under normal conditions.
Condensation should be avoided. If condensation has formed on the unit following a change of location or during transporalion. for example, wait until the unit has dried out completely before putting into operation.
The ambient air should not contain excessive amounts of dust.
(Comply with the instructions in the MAINTENANCE section).
Do not cover or obstruct the ventilation or exhaust vents in the unit housing, remove any dust or dirt with a vacuum cleaner.
The ambient temperature should not exceed 40 *C.
Place the unit on a firm, fireproof surface (e.g. laboratory table, support frame) in a stable, perfectly upright position.
Keep the unit away from direct sunlight (UV radiation).
If you wish to stack two units on top of one another, use the stacking frame (available as an option). Never stack more than two units on top of one another. When stacking units of different types, make sure that e.g. an Incubator is always at the bottom, and prevent any mutual influence.
Fig. 1/2: Clearances to adjacent surfaces / objects:
: 5cm
30 cm
*m>JI
5 c m >
•
Maintain a distance from the back:
U T 6 P : at least 12 cm
U T 1 2 P : at least 8 cm
U T 2 0 P : at least 8 cm
The wall spacer may only be removed if the unit is connected up to an extraction system.
The exhaust gases must be led through / past adjacent surfaces / objects located above or behind the unit with heat insulation to avoid the danger of fire.
If the unit UT 6 P is to be fitted under a table or into a laboratory fixture, it must be connected to an exhaust air extraction system. The top clearance may only be reduced to min. 5 cm under such circumstances. Ensure adequate ventilation.
Types UT 12 P and UT 20 P are not suitable to be fitted under a table.
Issue: May, 1994
50 042 750
7/34
SETUP AND INSTALLATION
Room ventilation
Adequate ventilation must be provided wherever the unit is installed.
Do not operate the unit In non-ventilated enclosed spaces.
Special ventilation measures (e.g. ventilation of designated work areas, FRG also refer to VDI
1946 Part 7) when several units are Installed in one room.
Mains connection
The laboratory unit is supplied with a permanently connected, flexible mains power lead with connector (grounded).
The connector serves to disconnect the unit from the mains power supply.
Examine the power lead and connector for damage before using the unit. If there are any signs
of damage, do not connect the unit to the mains supply.
The voltage quoted on the rating plate (rated voltage) must agree with the nominal mains
voltage.
The unit may only be connected up to the mains via a properly Installed socket outlet with PE
conductor (protection class I) or a permanent terminal connection.
In the event of a permanent terminal connection, a master switch (lockable In the "Off"
position) must also be installed to disconnect the unit from the mains.
An earth-leakage circuit-breaker should also be fitted in the supply system to provide
additional protection against electric shocks.
Rubber-sheathed flexible cable of at least 2.5 mm
3
diameter (H 07 RN-F ...) should be used for
any extensions that may be required to connect the unit.
Refer to the TECHNICAL DATA section for the rated power consumption, circuit protection with a T 1 6 A fusible link or a 8 16 circuit-breaker.
Noise Insulation
The unit develops a constant noise load.
Refer to the TECHNICAL DATA section for the noise level.
The unit may produce higher noise levels In unfavorable Installation conditions. Supplementary
measures may have to be Implemented to Improve the sound insulation in the room.
8/34 50 042 750 Issue: May, 1994
SETUP AND INSTALLATION
/j\ Connecting to an extraction system
Comply with the applicable national environmental regulations for the extraction of exhaust gases released during the heat treatment. Suitable measures must be implemented to ensure that such gases are safely led outside (FRG: BlmSchG). Such measures may involve thermal or catalytic treatment of flue gases.
Refer to the TECHNICAL DATA section for the volume flow and temperature of the exhaust gases.
Fig. 2/2: Example of an installation
1 Inlet air flow
2 Heating cabinet
3 Adjacent surface / object
4 Exhaust-air blower
5 Exhaust air flow
Draft breaker
Implementation instructions:
• A draft breaker must be used to connect the unit to an extraction system.
• All flue ducts must be made of fireproof materials.
• Apertures must be provided to enable the removal of residue and condensate. The flue ducts must be thermally insulated (recommendation: half-shell sleeves made of laminated mineral Tiber).
• The exhaust air / gases from the unit must never come into contact with combustion exhaust gases.
• Use the draft breaker to adjust the air flow volume in such a way that exhaust gases can be extracted reliably without producing any eddying at the unit's exhaust-gas connector.
• Technical measures should be implemented to monitor the flow of air and flue gases and to prevent operation of the heating unit if the ventilation system Is switched off / defective
(electrical interlock).
Issue: May, 1994
50 042 750
9/34
3. UNIT SPECIFICATION
Fig. 1/3: Overview of the control elements
I 1 c
A Switchgear unit
B Automatic control unit
C Door handle
~
i—.
— B
I 4 -
U LJ
A
Item A: Swltchgear unit
Fig.
2/3: Switchgear unit control panel
1 -
- 2
4 -
1 On / Off switch
2 Overtemperature protection, control element
3 "Fault" signal lamp
4 Fresh-air flap adjustment
• 1 ON / OFF switch:
To switch the unit on and off:
"on" = Unit switched on, the temperature Inside the unit appears on the display panel of the control unit,
"off' = Unit switched off, temperature display off.
• 2 & 3 Overtemperature protection device / signal lamp:
The unit is equipped with an electronic upper-limit cut-out device (TWB) of protection class 2 as defined by DIN 12 880 Part 1. II Is electrically and functionally independent of the temperature control system. In the event of a fault In the unit heating system, the operational TWB shuts down the unit heating across all poles as soon as the temperature exceeds the preset value.
The red signal lamp Indicates activation of the protection function (Fig. 2/3: Item 3).
The unit must be reset manually, by pressing the control element.
* Allow the unit to cool down (by approx. 30 *C), press the control element, the red signal lamp will go out.
The operate value Is adjusted with the help of a tool (coin, screwdriver...).
Adjust (o the required protection level:
• to upper temperature limit» unit protection (protects the unit and its environment).
*• to approx. 10 *C higher than the operating temperature set on the control unit = material protection (protects the unit, Its environment and the loaded material).
Ensure that the overtemperature protection device works properly by conducting a functional check at regular Intervals, every 3 months at least, and check the electrical equipment at least once a year.
10/34 50 042 750
Issue: May, 1994
3. UNIT SPECIFICATION
• 2 & 3 Overtemperature protection device / signal lamp:
Functional test of the TWB
(Prerequisite: required working temperature (control unit) has been reached /remains constant)
Set the TWB to approx. 10 - 20 'C below the temperature displayed on the control unit. The
TWB must respond, the red signal lamp indicates "Fault".
The TWB is working properly. Press the control element to release the TWB.
Now set the TWB to the temperature limit for the required type of protection.
If "Fault" is signalled during operation:
Check the settings of the TWB and the control unit and correct if necessary. If set for "material protection", this may indicate a blower malfunction. If the problem persists, contact the service center.
• 4 Fresh-air flap control
For continuous control of the fresh-air supply to the inner compartment.
•* Set to left-hand stop = minimum fresh-air supply
*• Set to right-hand stop = maximum fresh-air supply
Issue: May, 1994 50 042 750 11/34
3. UNIT SPECIFICATION
Item B: Automatic control unit
The automatic control unit contains the basic functions of a 2-posilion controller with PD/PID control response, time switch to switch the heating and ventilation on / off after the set time, the
added features of permanently programmed heating process programs and the programmed
controller functions of a freely programmable process controller.
Fig. 3/3: Control unit control panel
1
\
\
"C
I
2
*vl
IB
/
/
3 i
/ /
A
5
M
1
2
3
4
5
8
9
6
7
Signal lamp - healing &talus
Signal lamp - control uml status
Display panel
Control key
"Increase displayed value" key
"Reduce displayed value" key
"Programming function" key
"Air speed adjustment" key
Display panel
\ \
The status of the control unit is Indicated by signal lamps in the display panel.
THE YELLOW SIGNAL LAMP (Item 1) indicates the current status of the heating control system:
»• Lamp switched on •
unit heating activated
•• Lamp switched off
: heating off
Rhythmic flashing of this signal lamp indicates that the temperature controller is regulating the operating temperature.
• THE GREEN SIGNAL LAMP (Item 2) indicates the status of the control unit:
> Lamp on » control unit switched on: heating and blower enabled.
The heating and blower are activated when the unit is switched on (On / Off switch).
•• Lamp off » control unit switched off: heating and blower disabled.
The heating and blower are not activated when the unit is switched on
*• Lamp flashes * time-switch function activated.
Basic functions
The control unit is equipped with the "temperature control", "time switch" and "inside air
speed adjustment" basic functions.
The unit can set an operating temperature of up to 250 *C, switching times of between
0:01 and 99:00 hours and and Inside air speed in stages of approx. 40 %, approx. 70 % and
100%.
Refer to STARTING UP WITH BASIC FUNCTIONS for information about the commands for the basic functions.
12/34
50 042 750 Issue: May, 1994
3. UNIT SPECIFICATION
Added-feature programming functions, fixed programs
The control unit is capable of running heating process programs with fixed, preselected parameters from a database.
The program segment and the parameters are stored in the database.
The fixed program contains a maximum of four program segments and a status command.
Fig. 4/3: Example of a temperalure/time process, controlled by the fixed program t
T f C )
P I . 0 - P I . 9 : j P 2 . 0 - P 2 . 9 P 3 . 0 . P 3 . » j.l P 4 . 0 - P 4 . 9 P. OF
P.OFI
P. re
Each program segment, P I - P 4, is assigned to a typical temperature / time response. 10 selectable sets of parameters are permanently stored In the database for each program segment. These sets of parameters can. be assigned within the program.
A status command is provided as the fifth program segment. This determines the subsequent action with the program selected from the fixed program memory:
* Program on standby, do not execute yet = Select" P. o F'
•• Program on standby, wait for program start • Select" P. o n"
•• Delete program from memory completely • Select" P. r E "
The program memory is erased on delivery. If the operator Interrogates the program segments,
"Parameter set off' appears in the display panel = e.g." P. I -".
When compiling a program, selecting "parameter set off" within the selected program segment causes the control unit to bypass this segment for the fixed program function.
The selected program always runs from program segment P I. n to program segment P 4. n.
If program segment P 3. n is combined with program segment P 4. n, the heating and blower are only switched off following termination of segment P 4. n.
If a program is interrupted and started again, the sequence commences at segment P I. n.
The previously selected operating temperature cannot be changed once the program has started.
The heating is switched off in the event of a power failure (> 20 ms) with subsequent restoration of the power supply, the temperature inside the unit Is displayed as a flashing value and the program must be started again.
Issue: May, 1994
50 042 750 13/34
• Fixed program segments:
Program segment" P 1. n " (n = -, 0... 9):
Assigning a parameter (0 ... 9) from the database causes the unit to switch on heating and ventilation on expiry of a preset period.
Selectable parameters In this program segment:
PI-
Off
PL 0
2 h
PI.
I PI.
2 PI.
3 PI.
4 PI.
5
P I . 6 P I . 7 P I . 8 PI. 9
4 h
6 h 8 h
10h 12h 16 h 20 h 24 h 72 h
Program segment" P 2. n " (n «-, 0 ... 9):
Assigning a parameter (0 ... 9) causes the unit to heat up to the preset operating temperature at a specific rate of temperature change per hour.'
1
Selectable parameters In this program segment:
P 2 . off
P 2 . 0 P2.1
P2.
2
P2.3
P2. 4 P2. 5 P 2 . 6
P2. 7
P2. 8 P2. 9
10*C/h 20'C/h 30"C/h 40'C/h 50'C/h 60*C/h 80'C/h 100'C/h 120*C/h 200'C/h
Program segment" P 3. n " (n • -, 0... 9):
Assigning a parameter ( 0 . . . 9) causes the unK to heat up to the preset operating temperature, to maintain this temperature for the preset period and to subsequently switch off heating and ventilation
Selectable parameters In this program segment:
P 3 . off
P3.0
P3.1
P3.2
15 min. 30 mln.
1h
P3.3
1.5 h
P3.4
2 h
P3.5
4 h
P3.8
6 h
P3.7
8 h
P 3 . 8
12 h
P3. 9
24 h
Program segment" P 4. n " (n « - , 0 . * 9 ) : , .
Assigning a parameter ( 0 . . . 9) causes the unit to cool down from the previously reached Inside temperature at a specific rate of temperature Change and to subsequently switch off heating and ventilation.
11
•*: ;
Selectable parameters In this program segment:
P4.-
P 4 . 0 P4.1 P 4 . 2 J»4
!
/3' P 4 . 4 P 4 . 5 P 4.6 P 4.7 P 4 . 8 P 4.9
off 1'C/h 2*C/h 5'C/h 10*C/h 20"C/h 30"C/h 40'C/h 50'C/h 60"C/h 80'C/h
1) KOORDng lo nvxM »nd load
• Compilation of a fixed program
After activating the added-feature function, the sequence is compiled by marking the required program segments.
Example: Fixed program comprising P I. 2; P 2. -; P 3.7; P 4.4 and P. o n :
14/34
50 042 750 Issue: May, 1994
3. UNIT SPECIFICATION
• Example of a fixed program
*• Example E1: Hot-air sterilization process
The heating should switch on in 8 hours, the unit should heat up to an operating temperature of
190 *C (sterilization temperature) at a rate of 200 *C/h, maintain this temperature for 4 hours
(equalization phase and sterilization period) and should subsequently cool down at a rate of 20
•C/h.
The program should start following program selection (status command: P. on).
Fig. 5/3: Program profile - hot-air sterilization process El
190
-
C c
1
S2 S3 S4 ' S1 - S 4«Programmsegment
:
/A
mi P3.5
P.onl
Sequence E 1 :
After sketching the process characteristic, select the segments which correspond to the required sequence from the matrix .
Mark the program segments and the " P. o n" status command. Leave program Input mode.
Select 190 "C as the required operating temperature. Start the fixed program.
While the program is running the display alternates between the temperature inside the cabinet and the current program segment.
Refer to STARTING UP WITH ADDED-FEATURE PROGRAMMING FUNCTIONS for further information.
Issue: May, 1994
50 042 750
15/34
OPERATION
Programmed controller function
The programmed controller function enables the arbitrary definition of temperature / time processes.
You may select up to 9 program segments. Operating temperature, period and air speed level can be selected for each segment.
Termination commands
A termination command can be selected as the last program segment. The termination command determines the function to be performed by the unit after running through the program
There are three optional termination commands:
•• " hold " = Once the program has terminated, the operating temperature set for the last program segment is malntaned until a new operating temperature has been selected.
* •• stOP " = Once the program has terminated, Ihe heating and blower are switched off until the "stop" command Is cancelled.
>• •• CYCL " - Once the program has terminated, It returns to the beginning and is repeated continuously until the "CYCL" command is cancelled.
A termination command must always be entered at the end of every program.
Fig. 6/3: Example of a temperature / time / ventilation program with "CYCL" termination command
40
%
T1
\
7
100%
\
T7 T8
r
1
M
\
.
i
\
S
T3 T4
4
• »
T5
V
/ n 8
\
V
>
\
T9
11 t2 t3 14 t5 re t7 t8 t9 t
16/34 n 10«CYCL
• Definitions n 1 ... n9 n10
T 1 ... T 9
11 ... t 9
Program segments
Termination commands (hold, StOP or CYCL)
Specified operating temperature (SO - 250 *C)
Specified time - program segment (0:01 - 99:0 h)
U 1 ... U 3 Specified ventilation level (approx. 40 %. approx 70 % and 100 %)
• Refer to STARTING UP WITH PROGRAMMED CONTROLLER FUNCTIONS for details.
50 042 750 Issue: May, 1994
OPERATION
What happens if ... ?
Possible display
8 8 8
O F F
One digit in the display flashes
P1.3
P 1.3 alternating with inside temperature
P. o n
P. OF
P. r E
P r o alternating with inside temperature n I hold
stop
CYCL
E - 3
E - 4
E-S
E - 6
E - 7
E - 8
E - 9
Explanation
Control unit test phase
Status of the control unit, heating and ventilation disabled
The flashing value can be changed
Program segment when entering an added-feature program function
Program segment while the added-feature program is in progress
Status command for an added-feature program:
Program on standby, wait for program start
Status command for an added-feature program:
Program on standby, do not execute yet
Status command for an added-feature program:
Delete program from memory completely
Entered program on standby
Power failure (> 20 ms) with subsequent restoration: the heating switches off, "P r o" alternates with the temperature inside the unit on the display, if a program is in progress It is Interrupted and must be started again.
Program segment when entering and activating the programmed controller function.
Termination command when entering the programmed controller function: the temperature set for the last program segment is maintained on termination of the program
Termination command when entering the programmed controller unction: the heating and ventilation is switched off on termination of he program
Termination command when entering the programmed controller function: when the program is terminated, it retums to the beginning and starts again
A termination command is entered for this program segment when entering the programmed controller function. The termination command must be changed to a time value to enable the input of a emperature value.
Stored data defective
Switch the unit off and on again. If the problem persists, contact the service center
Error of measurement in the reference values contact (he service center -
Upper-limit violation of actual temperature tolerance contact the service center -
.ower-limit violation of actual temperature tolerance contact the service center -
Temperature sensor or input defective contact the service center -
Stored data defective contact the service center -
Erroneous reference values contact the service center -
Issue: May, 1994
50 042 750 17/34
OPERATION
Afler letup and Installation, proceed as follows to start up:
Starting up with basic functions, without programming
Instruction
Switch unit on
Key(s)
On / Off switch = "on"
Example ol display Remarks
The green status display indicates the status.
Display panel: inside temperature
The unit runs through a self-test phase after switching on, display: " 888 ".
If the " o f f ' status was selected for the previous operating phase, the display alternates between the lemperature Inside the cabinet and the "oFF" status.
Selecting the control unit status function
Control unit "on":
(heating and ventilation enabled)
1 . 1 l i
Inside temperature, green signal lamp lights up continuously
Select this status function to operate the unit, or to use the time-switch function to switch the heating and ventilation off after a preset period ontrol unit "off":
(heating and ventilation disabled)
I approx. 3 sec.
II I I
: D
Inside temperature, green signal lamp remains dark
Select this status function to switch the heating and ventilation off, or
If the heating and ventilation Is to remain off during programming, or to use the time-switch function to switch the heating and ventilation on after a preset period
Salting the operating temperature
Display the preset operating temperature
Preset operating temperature
Increase the preset operating temperature
Display flashes
Value increases
'I I
Reduce the preset operating temperature
Display flashes
Value decreases
Store new preset iperatlng temperature
II I I
I ""II I
Display flashes
New operating temperature is stored, then Inside temperature appears on display
Unit regulates the operating temperature when status " o n " , yellow signal lamp indicates the status of
Ihe heating control system.
Adjusting the overtemperature protection device: refer to Chapter 3 ettlng the position of the fresh-air flap
18/34
50 042 750
Issue: May, 1994
OPERATION
Setting the switching time
Instruction
Select time-switch function, display time remaining for timing period
Increase preset time
Key(s)
Example of display
I I II I
I.I
II I
Remarks
Display flashes, lime-swilch funcition is displayed. If liming period in progress, displays time remaining
Reduce preset time
l I II
.l I l
Value can be adjusted between 1 minule and 99 hours Adjustable in full hours from 10.0 hour selling
The heating is switched on or off. according to sel status.
once timing period has expired
Store new preset time value set
I I
.II
The lime value is transferred to the memory, the inside temperature then appears on the display
The flashing status display Indicates that the time-switch function has been activated.
If you do not wish to use the time-switch function, enter 0.0 0 as the preset time.
Selecting the air speed level
Display the preset air speed level
•"I
1 1
1 1
' - ' 1 Preset air speed level
Increase preset air speed level
r
i n
1
1 1 '—
-
1
• - • 1
Display flashes, speed symbol
Value increases
(e.g. 7 0 % / 1 0 0 %)
I
IUU|
Reduce preset air speed level
l t ~ ^ —** ~ ^ i
Display flashes, speed symbol
Value decreases
(e.g. 70 % / 40 %)
Store new preset air speed level
Display flashes, speed symbol
Air speed level is stored and the display then goes dark
Issue: May, 1994 50 042 750
19/34
4 OPERATION
Fig. 1/4: Example of a hot-air sterilization process
S 1 - S 4 • Programmsegment
190 *C
P 3 . 5 P.onl
The heating is to switch on in 8 hours, the unit should heat up to 190 'C at 200 'C/h, maintain this temperature for 4 hours, then switch off.
The program should start following program selection (P. o n).
Starting up with added-feature programming functions
Instruction
|Key(s).
| Example of display | Remarks
Switch unit on, select operating temperature: refer to start-up with basic functions
Adjusting the overtemperature protection device: refer to Chapter 3
Control unit "oir, unit should not heat up during this input
- = = — I approx. 3
> / sec.
II
U I
Status display = "off", control unit is disabled, Inside temperature Is displayed
Call fixed program memory
— — I approx. 6 set sec.
Display flashes, program segment S1 appears after 6
Select parameter PI. 3
Mark P I. 3 as the required parameter in program segment S1
Select program segmenl S2
set
I I I
U I _J
I L I
I I I
_
I. I
Programmable digit flashes.
P I. Is displayed if no parameter has been marked
Programmable digit flashes
P I. 3 is transferred lo the memory, programmable digit flashes
Programmable digit flashes
Continue In the same manner to enter / mark program segments P 2. 9, P 3. 5 und P 4.
20/34
50 042 750
Issue: May, 1994
OPERATION
Entering the status command
Instruction Key(s)
Select and mark status command P. o n
Example of display
1
Remarks
P. o n is displayed, mark with the "set" key
Leave the program memory
U
Display alternates between inside temperature and
" P r o "
Input terminated
Starting the added-feature program
Start the program approx. 3 sec.
I . - I
Program is started, the display alternates between the temperature and the program segment
Terminating the added-feature program
Terminate the program f — = = — | approx. 3
| v 1 sec.
a
l_l
1 1 l_l
Display alternates between the inside temperature and
" P r o "
Issue: May, 1994 50 042 750
21/34
UKtKMIION
Fig. 2/4: Example of the programmed controller function with " StOP " termination command
40% 100%
T7 \ T8
/
vV I
T9 t2 13
t5 t6 t7
t8 t9 t n 10 = StOP
The program sequence from a total of 9 segments must be entered, after which the healing and ventilation is to be switched off.
The air speed levels to be set for each program segment are given in percent.
Starting up with programmed controller (unctions instruction
Key(s)
|Example of display Remarks
Switch the unit on, refer to start-up with basic functions
Adjusting the overtemperature protection device: refer to Chapter 3 ontrol unit "off, unit should not heal up during this input
ca. 3 Sek.
n n I
J U.U
Inside temperature, green signal lamp lights up continuously all main program memory set
I I
Program segment
"n I" is displayed, programmable digit flashes
22/34
50 042 750
Issue: May, 1994
OPERATION
Entering the temperature for a program segment
Instruction
Enter operating temperature T 1,
230 *C
Key (s) Example of display Remarks
Preset operating temperature for program segment
1(n I) Mashes in display
If" —" appears In the display when the set key is pressed, then "hold" or "StOP" has been enlered as the termination command for program segment n I. If you wish to enter a temperature value, you must first change the termination command to a time value (refer to Ihe section on "Changing the time for a program segment").
Enter 230 "C
I
II I
"II I
Transfer temperature
T1=230 *C to the memory
I I
Program segment
" n I" is displayed, programmable digit flashes
Entering the time for a program segment
Enter the time for program segment n I,
2:00 h
Transfer the preset time for program segment n I to the memory
The time for n I flashes in the display
I I
Timet 1= 2 . 00
I I
I I I
n I is displayed, programmable digit flashes
Issue: May, 1994 50 042 750
23/34
OPERATION
Entering the air speed level for a program segment
nstruclion
Key(s)
Example of display
Select Ihe preset air speed level lor program segment n I
I I
Remarks
Air speed level flashes in the display, speed symbol
ransferthe preset air speed level lor program segment n I to the memory
I I
I I
Select air speed level,
approx. 40 % in this
case
n I is displayed, programmable digit flashes
Selecting the next program segment
Select program segment n 2
n 2 Is displayed, programmable digit flashes
Enter the preset operating temperatures, time values and air speed levels for program
segments n 2 - n 9 in the same manner as for program segment n 1 and select program
segment n 10.
Selecting the termination command
Select the "StOP" termination command
_ I I I
I I I I I
Termination command flashes in the display
II I
II I
Enter the "StOP"
termination command
Transfer the termination command to the memory prog
II I
n I n 10 is displayed, programmable digit flashes
24/34
50 042 750
Issue: May, 1994
OPERATION
Leaving the program memory
Input terminated
Starting the program
Instruction
Start the program
Key (s)
| approx. 6 sec.
Example of display
II I I I
Remarks
Inside temperature and program segment are displayed
Terminating the program
Instruction Key(s)
Terminate the program I I approx. 6 sec.
1 prog 1
Example of display
D
- r n
1 1
Remarks
Inside temperature and control unit status are displayed
Issue: May. 1994 50 042 750
25/34
OPERATION
Fig. 3/4: Example of the programmed controller function with " hold " termination command
T1
100%
T2
/
•4
/
/
40%
n 1
\
•4
70% n 3
11 12
T3
nA 'hold
Starting up with programmed controller function
The "hold" termination command has been selected here as an alternative to example 2/4.
Enter the preset temperature, time and air speed level values for the program segments according to the same procedure as example 2/4.
Instruction Key(s) Example of display
Select program
segment n 4
I I
Remarks
n 4 is displayed, programmable digit flashes
Select the "hold" termination command prog
I i
"I " I
Termination command
flashes in the display
I I
Enter the "hold" termination command
Transfer Ihe termination command lo the memory
Leave the program memory and slart the program n 4 is displayed, programmable digit
flashes
26/34
50 042 750
Issue: May. 1994
4 OPERATION
Fig. 3/4: Example of the programmed controller function with " CYCL " lerminalion command
4 0 %
T1 T2
100%
70%
\
T3 T4
40 % ^
4
n 1
» < "
2
• t1 12 13 14 n 5 • CYCL
Starting up with the programmed controller function
The "CYCL" termination command has been selected here as an alternative to example 274.
Enter the preset temperature, time and air speed level values for the program segments
according to the same procedure as example 2/4.
Instruction
Select program segment n 5
Key(s) Example of display
i i
Remarks n 5 is displayed, programmable digit flashes
Select the "CYCL" termination command prog
Termination command flashes in the display
Enter the "CYCL" termination command
Transfer the termination command to the memory prog
Leave the program memory and start the program
I I I
n 5 is displayed, programmable digit flashes
Issue: May, 1994
50 042 750 27/34
UfCKAIlUN
Operating guidelines
Remember to put on protective garments, e.g. gloves, goggles, mask, body protection, remove any items of jewellery before starting work.
Do not cover or obstruct the ventilation or exhaust vents in the unit housing. Always keep these vents free of dust and dirt
• Loading:
Circulating air conducts the heat to the samples in the unit. Make sure that the samples are evenly distributed on the trays, and that they are arranged in such a way as to ensure an unimpeded (low of warm air throughout the cabinet, in order to prevent local overheating. Do not position the samples too close to the inner walls.
Fig. 4/4: Loading diagram
:i a
"a" = approx. 1/10 of the internal dimensions
Be careful not to damage the temperature sensors in the inner chamber when loading.
I- Ji Never place samples on the floor of the chamberl
Keep the inner surfaces of the chamber free of dust and dirt. Never use aggressive chemicals which could damage the stainless steel finish.
Continually heating and cooling the unit over wide temperature ranges can detrimentally affect its service life If used more frequently, it is better to keep the unit running in continuous mode at moderate temperatures.
A If not used for an extended period, disconnect the unit from the power supply.
Shutdown
• Remove the samples from the chamber
• Allow the heating unit to cool down to room temperature
• Switch the unit off and disconnect from the mains
• Clean the inside surfaces of the chamber
(also refer to the section on DECONTAMINATION, CLEANING ...)
28/34
50 042 750
Issue: May. 1994
OPERATION
The serviceability and safety of the heating unit can only be guaranteed if the necessary
tests, maintenance, servicing, and repair work is carried out by the Heraeus service
personnel or other agencies who are authorized to act on behalf of Heraeus Instruments
GmbH.
Heraeus Instruments GmbH cannot accept any liability for any damage that occurs as a result of improper use or repair work, which has not been performed by Heraeus service
centers, or if parts other than the approved genuine spare parts / accessories are used.
We recommend that you enter into a service contract with our company Please conlacl us (or a quotation.
Decontamination / disinfection
The operator Is responsible for implementing suitable, effective measures to
decontaminate the unit if it is used In connection with hazardous materials. Such
measures must be conducted on a regular basis under normal operating conditions and
in particular if such materials have been spilt or before commencing service work.
The stainless steel components inside the unit may be decontaminated by wiping out with a
disinfectant solution In cases of biological contamination
Recommended disinfectant: Barrycidal 36, available from: Interchem AG. CH-6340 Baar
(FRG sales: Handelsvertretung Helmut Schroder, GaufistraBe 30/2, 70193 Stuttgart).
The applicable national regulations must be observed with respect to disinfection (FRG:
ZH 1/598 ...)
Once removed, the internal stainless steel components may also be subjected to steam
sterilization or autoclave treatment.
Cleaning
Use a mild soap solution (water and mild detergent) and a soft lint-free cloth lo wipe the outer
surfaces and control elements of the unit.
Remove any dust deposits from the air vents with a vacuum cleaner
Remove any dust or dirt from the inner chamber. Use stainless steel cleaning agents in
moderate quantities. Never use acids, chloric solvents or salt solutions to clean the stainless
steel surfaces inside the unit.
Avoid scouring and scratching, or damage to labels and paintwork will be unavoidable.
Include flue/exhaust ducts In the cleaning operations if the unit is connected up to an extraction
system.
Consult the manufacturer before implementing any other cleaning or disinfection
measures than those mentioned in this manual to ensure that the intended methods will not damage the unit.
Issue: May, 1994 50 042 750 29/34
MAINTENANCE
Servicing
The heating unil is virtually maintenance-free
Lubricate the bearing s of the door and door-handle mechanisms with a little synthetic oil from time to time.
Mineral oil products must not be used where they may come into contact with plastic components, as they could cause damage to such parts.
Tests
The following items should be tested at least once a year to ensure that the unit remains in good working order:
• Mechanical (unctions
• Operation in accordance with the specified technical data
• Electrical system (FRG: UVV VBG 4, OIN VOE 0701 Part 1, E DIN VOE 0702 /11.93)
• Safety devices
Repairs
Only approved genuine spare parts and accessories may be used.
The use of other parts may result in unforeseen problems and should be avoided under all circumstances.
Approved spare parts /accessories
Spare part /accessory
Size / model
Instruction manual
Circuit diagram 1/PE AC, 230 V
Circuit diagram 1/PE AC, 120 V
Wire tray with supporting straps
Seasoned silicone rubber door seal
Stacking frame
Space for back panel
Foot
Clasp for the door latching mechanism
Heraeus order number
U T 6 P
SO 040 640
SO 042 916
U T 1 2 P
SO 043 079
SO 041 987
-
50 042 917
U T 2 0 P
-
50 042 918
SO 042 499
50 042 906
SO 042 484
50 042 907
50 028 137
50 042 541
SO 042 284
SO 042 351
50 042 908
Other spare parts and accessories to order
30/34
50 042 750
Issue: May, 1994
TECHNICAL DATA
DATA 1 Model U T 6 P U T 1 2 P
U T 2 0 P
$Wffl^^&$mSS^S.^ *• - ' • • : ••.;•
Dimensions
(W x H x D)
Housing
Chamber
552 x 700 X 685
408 X 459 x 334
696 x 850 x 685
5 5 2 x 6 1 0 x 3 3 4
7 5 4 x 9 1 0 x 8 6 5
6 1 0 x 6 7 2 x 5 1 4
Useful space '• 326 X 367 x 275 441 x 4 8 8 x 2 6 7
4 8 8 x 5 3 7 x 4 1 1
Minimum clearances to adjacent surfaces / objects
Volumes
Left
Right
Top
Rear
Chamber
(empty)
Useful space ''
Steam space
Unit
33 approx. 73 approx. 40
50
SO
300 (SO when connected to an extraction system)
120
64.4
80
112
80
210
57.6
approx. 131 approx. 55
107.8
approx. 234 approx. 75 Weights
Trays
Max. tray load Lumped 15
20
15
20
10
Number
Dimensions
Surface
Total
Standard / Max.
( W x D ) :
50
2 / 9
362 x 333
50
2 / 1 4
506 x 323
15
70
2 / 1 6
564 x 503
Unit mm mm mm mm mm mm mm liter liter liter kg kg kg kg piece mm
Operating temperatures Room temperature + 20 to 250 |'C
Temperature deviations, circulating mode, air flap closed
Spatial" at 70 *C + 0.6
at 150 "C
± 1 8
+ 0.7
+ 1.9
+ 1.5
+ 2.8
K
K
Temporal '> at 250 *C + 0.7
Temperature deviations, fresh-air mode, air flap open
Operating temperature at 250 *C at 70 *C at150*C
70 *C
150 "C
250 *C
+ 3.4
+ 0.2
+ 0.4
8
19
42
+ 3.5
+ 0.2
±0.3
+ 0.8
+ 6.0
+ O.1
+ 0 7
+ 0.8
K
K
K
K
Spatial"
Temporal '• at 70 *C at150*C at 250 *C at 70 *C at 150 'C at 250 'C
+ 1.3
+ 3.5
+ 5.3
+ 0.2
+ 0.4
+ 1.0
• 0.7
+ 2.1
+ 3.5
+ 0.2
+ 0.4
+ 0.8
± 1 - 7
• 3.1
+ 5 6
+ 0.2
+ 0.5
+ 1.5
|K_
K
K
K
K
K
Warm-up times, circulating mode (unit empty, air flap closed, to 98 % of the operating temperature)
8
12
50
9
20
60 min min min
1) in accordance with DIN 12 880 Part 2 / 1 1 . 7 8
Issue: May, 1994
50 042 750
31/34
TECHNICAL DATA
DATA
Model U T 6 P UT12P U T 2 0 P
THERMAL
Warm-up-timos, (resh-air mode (unit empty, air (lap open, to 98 % of the operating temperature)
Unit
Operating temperature
70 'C
150
#
C
10
25
10
20
250'C 55
50
Cool-down times, circulating mode (unit empty, air flap closed, to 50 *C)
10
25
55 min min min
Operating temperature
Recovery times
Operating temperature
250 *C
150'C
70'C unit empty, door open 60 s, to 98 % of the operating temperature)
70
-
C 2 1
150 'C 5 4
250 "C
120
90
32
9
150
120
40
8
1
4
8
180
140
50 min min min min min min
Heat radiation to surrounding areas (at a room temperature of 25 "C)
Operating temperature
70 "C
150 *C approx. 0.16
approx. 0.43
approx. 0.18
approx. 0.42
250
#
C approx. 0.7
approx. 0.6
VENTILATION •-• ;:.
; ysjMfe-s ; , , ; , - : : • v;;#li||V:!f<j,':,
: f
;
¥;
Air volume flow inside the chamber
2.7
4.7
Circulating mode at 25 'C
Fresh-air mode at 250 *C
Rate of air replacment
Fresh-air mode at 250 'C
Exhaust air volume flow
3.7
50
5.1
40
Fresh-air mode at 250 'C
Speed of exhaust air
Air outlet 0 / air inlet 0
Flue gas temperature at duel at 250 *C
6.5
1,44
40
250
10.1
2,23
40
250 approx. 0.2
approx. 0.5
approx. 0.6
4.8
5.1
20
10.1
2,23
40
250 kW kW kW m h ' m"/h m/s mm
•c s
/min m'/h
NOISE [In accordance with DIN 45 635] ;
;
^ : ;
Sound level [20 uPa) 47
Sound power level (1 pW] 60
54
68
. . • • • - v-^sv*;;*!-;;. •
54
68 dB(A) dB
32/34
50 042 750 Issue: May, 1994
TECHNICAL DATA
DATA
ELECTRICAL (rated values)
Rated voltage
Model
Rated frequency
Power consumption
Connected load
Current input
Protection class
Protective measure
Degree of protection
U T 6 P
1.27
1.4
5.5
U T 1 2 P
1/PEAC. 230
5 0 / 6 0
2.32
2 5
10.1
I
UT20P [Unit
2 32
2 5
10.1
PE terminal
IP 20 |
V
Hz
KW kVA
A
Circuit protection to be provided at the installation location "
T 1$ A fusible link (slow-blow) or B 16 circuit-breaker
Connection via an earth-leakage circuit-breaker (Iripping current
< 30 mA) is recommended.
£llCJ|((CAl<(ra(0<) values) - > • ,
Rated voltage
Rated frequency
Power consumption
Connected load
Current Input
Protection class
Protective measure
Degree of protection
Circuit protection to be provided at the installation location "
<
;
' . \
1/PEAC, 120
5 0 / 6 0
1.27
1.4
10.6
I -
-
-
-
-
-
V
Hz kW kVA
A
PE terminal
IP 20
-
-
-
T16 A fusible link (slow-blow) or B 16 circuit-breaker
Connection via an earth-leakage circuit-breaker (tripping current
< 30 mA) is recommended.
5) The applicable national electrical engineering regulations and technical requirements must be observed when connecting up to mains power supply circuits
Issue: May. 1994
50 042 750 33/34
TECHNICAL DATA
©
Component
Outer casing
Heat sink (rear panel)
Inner chamber
Components litted in the chamber
Trays
Door seal
Door bearing
Door handle
Control unit
Switchgear unit
Operator control membrane
Feet
Thermal insulation
Filling medium - overtemp. protection device
Leads
Electronic / electrical components
Heating element
Materials used
Galvanized sheet steel, painted RAL 9002
Aluminium
Stainless steel, material No. 1.4301
Stainless steel, material No. 1.4301
Chromium-plated steel wire
Seasoned silicone rubber
PA, glass fiber reinforced
PA, glass liber reinforced
ABS
PC, glass fiber reinforced
PVS, printed polyester foil
PPN
Mineral fiber wool
Polydimethyldisiloxane
PVC -sheathed copper wire
Encapsulated components coated with various plastics, some mounted on glass fiber reinforced
PCBs with epoxy resin
Stainless steel tubing filled with magnesium oxide and heating conductors
Fibrous materials used for insulation sre ctassiried is health hazards in somt countrias. AHhough the fibers which may be reteased into the environment during normat operation do not constitute a serious risk, greater, more hazardous quantities may be released as a result of Improper handling during repair work or as a result ol mechanical damage.
34/34
50 042 750
Issue: May. 1994
Download
Advertisement
Key features
Temperature range: +20°C to 250°C
Programmed controller for precise temperature control and time settings
Fresh-air flap for controlling the fresh-air supply to the inner compartment
Overtemperature protection device
Hot-air sterilization capability
Frequently asked questions
The temperature range is +20 'C to 250 'C.
Refer to the "Starting Up with Basic Functions" section of the user manual for detailed instructions on using the programmed controller.
Yes, these ovens are designed for hot-air sterilization. The sterilization temperature should be at least 160 'C.
The fresh-air flap allows you to control the amount of fresh air that enters the inner compartment. This is useful for applications where you need to maintain a specific humidity level or for preventing condensation.
You need to conduct a functional check at regular intervals, at least every 3 months.