Index. Agilent Technologies G1978B
Add to my manuals
44 Pages
Agilent Technologies G1978B is a Multimode Source for the 6100 Series Single Quad LC/MS. It allows for the analysis of samples using multiple ionization techniques, including electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and atmospheric pressure photoionization (APPI). This makes it a versatile tool for a wide range of applications, including the analysis of small molecules, proteins, and lipids.
advertisement
Index
Index
A autotune,
C converting from ESI, APCI or APPI,
converting to ESI or APCI,
D diverter valve inlet filter,
E
ESI convert from,
convert to,
I installation,
L
LC/MSD sample tubing,
M method alternating ESI and APCI,
basic setup,
positive/negative mixed mode,
multimode nebulizer,
source image,
P parts multimode spray shield,
S spray shield for multimode source,
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide 41
Index
42 Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
www.agilent.com
In This Book
This book contains installation, operation, maintenance and troubleshooting instruction for the Multimode Source for 6100 Series Single Quad
LC/MS.
©
Agilent Technologies, Inc. 2008
Printed in USA
First Edition, December 2008
*G1978-90070*
G1978-90070
Agilent Technologies
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Content
- 7 Installation
- 8 Installation
- 8 Step 1. Prepare to install
- 9 Step 2. Install the HV control PCA and cables
- 13 Changing Sources
- 13 To remove the multimode source
- 14 To convert from multimode to ESI, APCI or APPI
- 15 To convert from ESI, APCI or APPI to the multimode source
- 21 Verification
- 22 To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes
- 24 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes
- 27 To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes
- 30 To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation
- 32 To do an autotune
- 33 Example of multimode verification report
- 35 Methods
- 36 To setup a method to use the multimode source
- 37 To create a method for positive/negative mixed mode operation
- 39 To create a method for alternating ESI and APCI operation
- 41 Index