Understanding the GLUCOCARD 01 Blood Glucose Meter. Arkray Glucocard 01


Add to my manuals
76 Pages

advertisement

Understanding the GLUCOCARD 01 Blood Glucose Meter. Arkray Glucocard 01 | Manualzz

UNDERSTANDING THE GLUCOCARD 01 BLOOD GLUCOSE METER

Top Bottom

Battery Compartment

Cover

Tab

Display

Test Strip

Port

Data Connection

Terminal

Gray/Down

Arrow

Orange/Up

Arrow

7

Explanation of the full display screen:

Memory Mode Symbol

Battery Symbol Audible Beep Off

Apply Blood Symbol

Date

Temperature Warning

Unit of Measurement

Time Format

Delete Symbol Control Solution

8

NOTE: Meters in the USA are programmed to show results in mg/dL. Meters used outside the USA are programmed to show results in either mg/dL or mmol/L. Results displayed in mmol/L will always have a decimal point. Results displayed in mg/dL will never have a decimal point.

GLUCOCARD

®

01 SENSOR

These test strips need only 0.3 µL (microliter) of blood. Apply blood to the sample application tip. The strip draws in the blood by capillary action. The blood glucose reacts with the enzyme on the test strip. This produces a current proportional to the blood glucose level. The meter detects the current and converts it into a blood glucose reading.

Test Strip Test Strip Bottle

L : XXX - XXX mg/dL

N : XXX - XXX mg/dL

H : XXX - XXX mg/dL

06E1A01

2011-11

Item No., N. de artículo : 78497

F-4

Code No.

(F-2, 3 or 4)

F- 4

Expiration date

Front Back

Insert this way into meter

Contact Bars

Sample

Application Tip

(Apply blood to end of test strip.

Capillary draw will pull blood to reaction site).

9

Changing the Battery

The GLUCOCARD 01 blood glucose meter comes with one lithium battery CR2032 (or

DL2032).

CAUTION : Replacing the battery will not delete meter settings or test results stored in the meter unless you:

• Touch the metal portion on main unit with your fingers or any metal.

• Take over 5 minutes to replace battery.

• Load the battery incorrectly.

• Load a used or weak battery.

NOTE:

• To save battery power, the meter turns itself off 3 – 5 minutes after last use.

• Replace batteries when low battery symbol appears. Otherwise

inaccurate results could be obtained.

• Discard used batteries as per local guidelines.

10

WARNING: Keep all batteries out of children’s reach. If a battery is swallowed, consult a doctor immediately.

To replace battery:

1. Open the battery cover on the back

side of the meter by pushing the tab

in the direction of the arrow and

pulling the cover up.

2. The battery is held in place on the

underside of the cover. Slide the

battery out of the cover.

CAUTION: Do not touch the metal contacts inside the battery compartment with your finger or any metal.

3. Load a new battery into the battery

cover. Place the “+” side down when

loading the battery into the cover.

4. Make sure the battery is loaded in the

proper position. Otherwise, the

preprogrammed times will be reset.

11

5. Put the battery cover back into correct position and snap to close.

If the clock setting has been deleted, either “12:00” or “0:00” will blink when you put a test strip into the meter. The low battery symbol means you need to replace the battery as soon as possible. Discard used batteries as per local guidelines.

12

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals