Configuration and Use Manual
MMI-20019053
NE 53 history
October 2022
F NE 53 history
F.1 NE 53 history
Important
Not all features and capabilities described in this section may apply to your transmitter or configuration.
August 2000, Version 1.x
Modification type
Expansion
Adjustment
Feature
Change
Added writing of the device tag using Modbus ®
Improved communication handling with the HART ® Tri-Loop product
Indication of outputs option board type appears on display at power-up
May 2001, version 2.x
Modification type
Expansion
Adjustment
Feature
Change
• Added alarm A106 to indicate that HART burst mode is enabled
• Indication of outputs option board type appears on display at power-up
• Added alarm A106 to indicate that HART burst mode is enabled
• Control of HART burst mode now available via Modbus
• Added support for the Model 1700 transmitter
• Added support for the intrinsically safe transmitter option
• Added support to configure the process variable units for mass flow, volume flow, density and temperature from the display
• Added support for assigning process variables to the mA and Frequency
Output from the display
Clarified the interaction of the digital fault setting and the fault timeout (last measured value timeout)
• Drive gain can be assigned to mA Output
• Pressure compensation added via HART
• Channel B can be configured as a Discrete Output
Configuration and Use Manual 249
NE 53 history
October 2022
Configuration and Use Manual
MMI-20019053
December 2001, version 3.x
Modification type
Expansion
Adjustment
Feature
Change
• Added support for the configurable I/O option board
• Software version information available via the display or Modbus
• Configurable density cutoff
• Additional HART variables can be assigned to QV
• The display start/stop totalizers function can be enabled or disabled
• Petroleum measurement application improvements
• Live zero available as display variable
• Increased options for fault output settings
• New cryogenic application temperature algorithms
• Improved Frequency Output stability and unit conversions
• Improved the handling of volume flow rate when two-phase flow is detected
• Improved handling of density values and calibrations during fault conditions
• Display configuration, screen flow and optical switch changes
• HART communication and burst mode improvements
• Petroleum measurement application added
• Custody transfer option added to the configurable I/O option board
• HART polling for external pressure/temperature added
June 2003, version 4.x
Modification type
Expansion
Adjustment
Feature
Change
• Added support for the Model 1500 transmitter
• Additional variables displayed by the Model 1700 transmitter
• Improved the handling of certain alarm conditions
• Clarified the behavior of certain Modbus calibration coils
• Clarified the interaction between certain density measurement units and density cutoff values
• Improved the handling of the mA source setting via the display
• Improvements to pressure and temperature polling
• HART Tri-Loop and other communication improvements
• Clarified the value returned by Modbus scaled integer registers during a fault condition
Discrete values now available through Modbus
250 Micro Motion 2700 Transmitters with Configurable Input/Outputs
Configuration and Use Manual
MMI-20019053
NE 53 history
October 2022
September 2006, version 5.x
Modification type
Expansion
Adjustment
Feature
Change
• Discrete Output assignable as a flow switch
• Discrete Output fault indication configurability
• Discrete Input support for multiple action assignments
• Added support for querying the display LED status via Modbus
• Additional HART and Modbus commands
• Process comparator expanded to five configurable events
• Factory configuration restore function
• Factory zero restore function
• Alarm history expanded
• Selectable write protection for configuration data
• Expanded selection of source assignments for mA Output
• Expanded storage of mA range values
• Expanded custody transfer application for independent implementation of
NTEP and OIML compliance
Display improvements for floating-point data
• Configurable alarm severity
• Gas standard volume functionality
• Meter verification availability as an option
• Multiple display language selections
September 2009, version 6.x
Modification type
Expansion
Change
• Frequency Output configurable as Discrete Output on Series 1000 transmitters
• Discrete Output assignable as flow switch on Series 1000 transmitters
• Display Variable 1 optionally fixed to process variable assigned to primary mA Output
• Frequency Output scaling method and related parameters configurable from display
• For enhanced density and petroeum measurement process variables, display cycles among variable name, current value and unit, and reference temperature
Configuration and Use Manual 251
NE 53 history
October 2022
Configuration and Use Manual
MMI-20019053
Modification type
Adjustment
Feature
Change
• mA Output Fault Action = None and Digital Communications Fault Action =
NAN is no longer allowed
• Frequency Output Fault Action = None and Digital Communications Fault
Action = NAN is no longer allowed
• Display variables set to a volume process variable automatically switch between liquid and GSV, according to current setting of Volume Flow Type
• Configurable hysteresis for flow switch
• Field Verification Zero added to support Weights & Measures application
• Transmitter firmware checksum and core processor firmware checksum assignable as display variables and viewable in ProLink
February 2018, version 8.x
Modification type
Expansion
Change
• Polling an external device for base density Gas Standard Volume (GSV)
• NE 53 version added to the display and accessed using Modbus and HART
• Improved data update rate for Advanced Zero Check using ProLink III
• Byte order configuration for double totals
• Modbus Function 43 basic objects for FDI identification
• Fast access to Live Zero in 100 Hz mode
• Display used to disable user access to the Smart Meter Verification menu.
Once disabled, access can only be enabled using ProLink III or a HART device
252 Micro Motion 2700 Transmitters with Configurable Input/Outputs
Configuration and Use Manual
MMI-20019053
Modification type
Adjustment
NE 53 history
October 2022
Change
• Sensors that are not straight tube sensors are now correctly identified
• The mA Output fixed alert is now set
• The Factory Configuration Invalid status bit is now set correctly when connected to a 700 core processor — as the 700 core processor does not support saving and restoring the factory configuration
• Enabling a new feature on a flow-only device does not reset the output configuration to factory defaults
• Polled variables are now converted to the transmitter units before being used
• HART Primary, Secondary, and Tertiary Variables can be changed using
HART when the transmitter is configured for gas standard volume
• Petroleum measurement variables are removed from the display when petroleum measurement is disabled
• The mA Output can be fixed when assigned to a concentration measurement, petroleum measurement, or GSV variable when the transmitter is not connected to a core processor
• The units for concentration curve data are correctly converted when the data is entered using HART
• Meter verification failed and aborted alerts are now reflected in the HART
More Status flag
• The HART Squawk command is functional when the HART write protect is enabled
• Locking the transmitter, unlocking the transmitter, or changing write protection does not set the HART Configuration Changed flag
• If connected to a 700 core processor with software version 3.1 or earlier, there is no longer a core processor write failed alert when setting volume flow units to beer barrels per time unit
• The factory zero value is immediately updated from the core processor after running a zero calibration and saving the factory configuration
• A firmware upgrade done without clearing non-volatile memory no longer causes an EEPROM Failure alert
• The status bit for the undefined A141 alert (Device Data Capture Triggered) no longer appears in the Modbus status register
• When using AMS to configure the transmitter:
— Changing the output configuration for mA 1, mA 2, or the Frequency
Output, does not change the configuration of all three outputs
— mA configuration methods can be cancelled
— Density and temperature units display on the concentration measurement configuration page
— An AMS Field Device did not respond message no longer displays when the concentration offset is configured even though the value was changed
Configuration and Use Manual 253
NE 53 history
October 2022
Configuration and Use Manual
MMI-20019053
Modification type
Feature
Change
— When the volume flow type is changed, the new setting is updated from the transmitter without having to rescan the device
• Basic meter verification
• Production Volume Reconciliation (PVR)
• Transient Mist Remediation (TMR)
• Transient Bubble Remediation (TBR)
• Piecewise Linearization for Gas
• Fuel consumption
• Support for Micro Load
May 2018, version 8.02/v6.82
• Added Basic Smart Meter Verification (SMV).
• Changed the default for the HART version to HART 7.
• Added Production Volume Reconciliation (PVR) that requires an 800 Enhanced Core Processor with version 4.42 or greater software and HART 7.
• Added Transient Mist Remediation (TMR) that requires an 800 Enhanced Core Processor with version 4.42
or greater software and HART 7.
• Added Transient Bubble Remediation (TBR) that requires an 800 Enhanced Core Processor with version
4.42 or greater software and HART 7.
• Added Piecewise Linearization for gas that requires an 800 Enhanced Core Processor with version 4.42 or greater software and HART 7.
• Added fuel consumption to require HART 7.
• Added polling an external device for base density for Gas Standard Volume.
• Added support for Micro Load .
• The NE53 version is added to the display and can be accessed using Modbus and HART.
• Improved the update rate of the data for Advanced Zero Check using ProLink III.
• Added byte order configuration for double totals.
• Added Modbus Function 43 basic objects for FDI identification.
• Added fast access to Live Zero in 100 Hz mode.
• Added the ability to use the display to configure user access to the Smart Meter Verification menu on the display. This feature can be used only to disable access. Once disabled, access can only be enabled using
ProLink III or a HART device.
254 Micro Motion 2700 Transmitters with Configurable Input/Outputs
Configuration and Use Manual
MMI-20019053
October 2022, version 8.02
Updated SMV documentation to add or clarify the following sections:
• SMV test preparation
• Smart Meter Verification capabilities
• Interpreting Smart Meter Verification results
• Resolving a failed Smart Meter Verification test
NE 53 history
October 2022
Configuration and Use Manual 255
For more information: www.emerson.com
© 2022 Micro Motion, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson
Electric Co. Micro Motion, ELITE, ProLink, MVD and MVD Direct
Connect marks are marks of one of the Emerson Automation
Solutions family of companies. All other marks are property of their respective owners.
*MMI-20019053*
MMI-20019053
Rev. AC
2022
Télécharger
Publicité