3.6 Adjusting the lighthead brake

The lighthead brakes are adjusted during installation. Brakes, like all mechanical parts, are subject to wear. If the lighthead no longer remains steady in all positions the brake must be readjusted.

The light unit becomes hot during operation. Allow it to cool down before adjusting the brake.

bLUE 30 lighthead figure 35

• Turn off the BLUE 30 lighthead and allow it to cool down.

• Press the locking tab (1) to remove the bezel.

• Adjust the lighthead braking by tightening or loosening the brake screw (2). The retaining screw (3) holds the lighthead in place and must not be loosened.

• Close the cover and check that the locking tab clicks audibly into place.

• Check that the cover is correctly in place.

bLUE 80 lighthead figure 36

• Turn off the BLUE 80 lighthead and allow it to cool down.

• Adjust the lighthead braking by tightening or loosening the brake screw (1) above the lighthead fork

(2).

Technical manual

0136201