<b>2. SC202 SPECIFICATIONS</b>. YOKOGAWA EXA SC202S, EXA SC202G, SC202G (S)

Add to My manuals
72 Pages

advertisement

<b>2.  SC202 SPECIFICATIONS</b>. YOKOGAWA EXA SC202S, EXA SC202G, SC202G (S) | Manualzz

3-3 Installation and wiring

mA

3-2. Preparation

Refer to figure 3-4. The power/output connections and the sensor connections should be made in accordance with the diagram on page 3-6. The terminals are of a plug in style for ease of mounting.

To open the EXA 202 for wiring:

1. Loosen the four frontplate screws and remove the cover.

2. The terminal strip is now visible.

3. Connect the power supply. Use the gland on the left for this cable.

4. Connect the sensor input, using the gland on the right (see fig. 3-5). Switch on the power. Commission the instrument as required or use the default settings.

5. Replace the cover and secure frontplate with the four screws.

6. Connect the grounding terminals to protective earth.

7. The optional hose connection is used to guide the cables coming from an immersion fitting through a

protective plastic tubing to the transmitter.

3-2-1. Cables, terminals and glands

The SC202 is equipped with terminals suitable for the connection of finished cables in the size range: 0.13 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in the range of 7 to 12 mm (9/32 to 15/32 inches).

SENSOR

CABLE GLAND

POWER/OUTPUT

CABLE GLAND

GROUNDING TERMINAL

Figure 3-5. Glands to be used for cabling

IM 12D7B3-E-E

advertisement

Key Features

  • Wide measurement range: 0.001 μS/cm to 1000 mS/cm
  • High accuracy: ±0.5% of reading
  • Temperature compensation: Automatic or manual
  • Hold function: Freezes the output signal
  • Passcode protection: Prevents unauthorized access
  • Intrinsically safe: Suitable for hazardous areas

Related manuals

Frequently Answers and Questions

What is the installation site requirement for SC202?
The installation site should be free from excessive vibration, shock, dust, corrosive gases, and direct sunlight.
How do I wire the sensor?
Refer to section 3.3 of the manual for detailed wiring instructions.
How do I set the range?
Enter the Commissioning mode and navigate to the 'Range' parameter. Use the arrow keys to set the desired range.