Power On and Check. Agilent Technologies N9320B

Add to my manuals
42 Pages

advertisement

Power On and Check. Agilent Technologies N9320B | Manualzz

Getting Started 1

Power On and Check

1

Connect the AC power cord into the instrument. Insert the plug into a power socket provided with a protective ground.

Set the tilt adjustors for your preference.

Figure 1 Plug in and angle adjustment

NOTE

2

Press the AC line switch on the rear panel. The standby

LED on the front panel will light and the spectrum analyzer is in standby mode (AC power applied).

3

Press the standby switch on the front panel. The On LED will light, and the spectrum analyzer boots up.

Self- initialization takes about 25 seconds; the spectrum analyzer then defaults to the menu mode. After power on, let the spectrum analyzer warm up for 45 minutes for stabilization.

The front panel switch is a standby switch only; it is not a power switch. To completely disconnect the spectrum analyzer from the AC line power, shut off the power switch on the rear panel.

N9320B Spectrum Analyzer 7

8

1 Getting Started

Running Internal Alignments

1

2

To meet the instrument performance specifications, the analyzer must periodically be manually aligned.

Connect a BNC cable with the N- BNC adapter between the

CAL OUT and RF IN front panel connectors.

After 15- minute warm up, press

Preset/

System

> Alignment > Align > All

The self alignment takes about 5 minutes.

Check for Instrument Messages

1

2

3

4

The spectrum analyzer has two categories of instrument messages: error and warning messages. A error message is triggered by operation errors, for example, parameter setting conflicts or data input that is out of the range of a parameter. An warning message may be triggered by hardware defects which could result in damage to instrument.

Here are some tips to check the instrument messages.

Check the display to see if any messages display in the status bar. Press

Preset/

System

> More > Show Errors to review each message. Refer to User’s Guide for detailed system message descriptions.

When you have reviewed and resolved all of the error messages, press

> More > Show Errors > Clear error queue to delete the messages.

Cycle the power to the analyzer and re- check to see if the

Instrument messages are still there.

If the error messages cannot be resolved, contact the Agilent

Customer Contact Center for assistance or service.

N9320B Spectrum Analyzer

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals