- No category
advertisement
6. Resolution
Bandwidths
Specification
Bandwidth
Amplitude
Selectivity
10 Hz to 3 MHz; *20%
3
kHz to 1 MHz; flO%
(Referenced to 1 MHz amplitude)
3
MHz; fl.O dB
1 MHz to 100 Hz; f0.5 dB
30
Hz; AO.8 dB
10 Hz; f2.0 dB
(60 dB/3 dB Ratio)
3
MHz to 100 kHz; <15: 1
30
kHz to 10 kHz; <13: 1
3 kHz to 30 Hz; <ll:l
10 Hz; ~100 Hz separation of 60 dB points
Related Performance
Tests
Description
Note
Resolution Bandwidth Accuracy Test
Resolution Bandwidth Selectivity Test
Resolution Bandwidth Switching Uncertainty Test
A signal source is connected to the spectrum analyzer input. The analyzer steps through the bandwidths from 3 MHz to 10 Hz, centers the signal, sets signal peak near the reference level, and measures the frequency of the 3 dB points for each bandwidth. The 3 dB bandwidth is
then
calculated by determining the difference in frequency between the 3 dB points.
Next, the spectrum analyzer steps through the bandwidths and measures the frequency of the 60 dB points of each bandwidth. The
60 dB bandwidth is then calculated by determining the frequency difference between the 60 dB points.
The shape factors are calculated by dividing the 60 dB bandwidths by the 3 dB bandwidths.
This procedure is for instruments other than those having Option 462.
For Option 462 instruments perform the test on the following page.
5-16 Operation Verification
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Certification
- 3 Warranty
- 4 Assistance
- 4 Safety Notes
- 5 General Safety Considerations
- 6 How to Use This Guide
- 6 HP 8568B Documentation Description
- 6 HP 8568B Installation and Verification Manual
- 6 HP 8568B Operating and Programming Manual
- 6 HP 8568B Performance Tests and Adjustments Manual
- 6 HP 85680B RF Section Troubleshooting and Repair Manual
- 6 HP 85662A IF-Display Section Troubleshooting and Repair Manual
- 13 What You'll Find in This Chapter
- 14 Introducing the HP 8568B
- 14 Safety
- 14 Accessories Supplied
- 17 Accessories Available
- 18 Options
- 19 Instruments Covered by This Manual
- 21 Calibration Cycle
- 21 HP-IB Address Selection
- 21 Bench Operation
- 22 Electrostatic Discharge Information
- 23 Static-Safe Accessories
- 24 Routine Maintenance
- 24 Cleaning the Display
- 25 Cleaning the RF Section's Fan Filter
- 25 Replacing the RF Section's Battery
- 30 Ordering Information
- 30 Direct Mail Order System
- 30 Direct Phone-Order System
- 31 Returning the Instrument for Service
- 31 Service Tags
- 31 Original Packaging
- 33 Other Packaging
- 35 Sales and Service Offices
- 36 What You'll Find in This Chapter
- 36 Safety
- 38 Preparation for Use
- 38 Initial Inspection
- 38 Operating Environment
- 38 Power Requirements
- 39 To Install Standard Instruments
- 42 To Install Option 908 and 913 Instruments
- 48 To Install Option 010 Instruments
- 55 To Set the HP-IB Address
- 55 From the Front Panel
- 55 From the HP-IB Bus
- 55 From the HP-IB Address Switch
- 61 Introduction
- 61 Frequency
- 61 Measurement Range
- 61 Displayed Values
- 61 Center Frequency
- 62 Frequency Span
- 63 Frequency Count
- 63 Resolution
- 65 Stability
- 65 Spectral Purity
- 66 Amplitude
- 66 Measurement Range
- 66 Displayed Values
- 67 Reference Lines Accuracy
- 67 Dynamic Range
- 69 Amplitude Uncertainty
- 72 Marker
- 72 Sweep
- 73 Inputs
- 73 INPUT #1
- 73 INPUT #2
- 74 Outputs
- 74 CAL OUTPUT
- 74 1ST LO OUTPUT
- 74 Options
- 74 75 Ohm Input Impedance, Option 001
- 75 400 Hz Power Line Frequency Operation, Option 400
- 76 General
- 76 HP-IB Interface Functions
- 76 Environmental
- 76 Power Requirements
- 76 Humidity
- 76 EMI
- 76 X-Rays
- 78 Warm-Up Time
- 78 Weight
- 79 Dimensions
- 80 Frequency
- 80 Resolution
- 81 Stability
- 81 Spectral Purity
- 82 Amplitude
- 82 Average Noise Level
- 82 Log to Linear Switching
- 82 Frequency Response (Flatness) Uncertainty
- 83 Residual Responses
- 83 Third Order Intermodulation Distortion
- 84 Inputs
- 84 RF INPUTS
- 84 VIDEO INP
- 84 IF INP
- 84 EXT TRIGGER
- 85 FREQ REFERENCE (IN)
- 85 Outputs
- 85 Display Outputs
- 85 Recorder Outputs
- 85 SWEEP
- 86 VIDEO
- 86 PENLIFT
- 86 2 1.4 MHz IF OUTPUT
- 86 FREQ REFERENCE (OUT)
- 86 VIDEO OUT
- 87 IF OUT
- 87 PROBE POWER (front panel)
- 87 General
- 87 Acoustic Noise Emission/Geraeuschemession
- 88 Function Descriptions
- 88 Sweep
- 90 What You'll Find in This Chapter
- 94 Test System Configuration
- 94 Equipment Connections
- 96 Program Loading
- 96 Program Operation
- 98 HP-IB Addresses
- 99 Error Messages
- 99 Test Descriptions
- 100 1. Input Attenuator Switching Uncertainty
- 101 2. IF Gain Uncertainty
- 102 3. Scale Fidelity (Log)
- 103 4. Scale Fidelity (Linear)
- 104 5. Log Scale Switching Uncertainty
- 105 6. Resolution Bandwidths
- 106 7. Impulse and Resolution Bandwidths (Opt. 462 Impulse Bandwidth)
- 107 8. Line Related Sidebands
- 108 9. Residual FM
- 109 10. RF Gain Uncertainty
- 110 11. Sweep Time Accuracy
- 111 12. Average Noise Level
- 112 13. Residual Responses
- 113 14. Frequency Span Accuracy
- 114 15. Frequency Response
- 115 16. Second Harmonic Distortion
- 116 17. Third Order Intermodulation Distortion
- 117 18. Calibrator Output Amplitude Accuracy