advertisement
58
Startup and Basic Operation
Current Sink
MC5-IS has a possibility to sink current, i.e. MC5-IS acts as a device controlling the flow of current while a external device generates the supply voltage for the loop, both in the
E and the
ET
module. The maximum current it 25 mA.
Required settings Options/description
Quantity
Funct/Port
Current
ET: I(control)
or
E: I(control)
Using the ET module’s output terminals
When sinking current using the ET module’s output terminals, select
Funct/
Port
option
ET: I(control)
. The following picture displays the active terminals:
T / C I N T . R J
T / C , L o w V R , R T D
4 w m e a s
, ,
H A R T
T / C
W I R E S
O N L Y m e a s / s i m
S E N S O R M E A S U R E & S I M U L A T E
3 w m e a s
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
O U T P U T
E T
V , ,
M E A S U R E
m e a s / s i n k
C o m
L o w V
Note.
If you are simultaneously using the ET module’s output terminals and some of the ET module’s measuring terminals, there is no galvanic isolation between the measuring circuit and the current sinking circuit.
Using the E module’s output terminals
Select
Funct/Port
option
E: I(control) to sink current using the E modules terminals.
Active terminals:
T / C I N T . R J
4 w m e a s
Note.
V , ,
H A R T
T / C , L o w V R , R T D
T / C
W I R E S
O N L Y m e a s / s i m
S E N S O R M E A S U R E & S I M U L A T E
3 w m e a s
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
O U T P U T
E T
V , ,
M E A S U R E
L o w V
The output current of the E module is a “pure” analog signal without any embedded HART signal.
Generating/Simulating
Notes applying to both the
ET: I(control)
and the
E: I(control) ports:
Check the polarity of your connections. The arrows in the previous pictures describe the correct flow of current.
If the loop resistance of the external circuit is too high or infinite, the message “ O.LOAD” is displayed until the load is reasonable.
Warnings!
To avoid damages to MC5-IS' electronics, first select the correct Quantity and Function/Port. Then connect the circuit including the external supply to MC5-IS.
Make sure that you don’t exceed the maximum current allowed by the instrument under test.
If you open the mA generation loop, MC5-IS tries to maintain the current by decreasing the internal impedance. If you then close the loop again, the current is first too high, but returns quickly to the correct level. If this current peak could damage the components of the loop, make sure that the loop will not open or protect it against overcurrent. For the same reason, always enter 0 mA output before connecting the loop.
Next…
Current Measurement on page 39
Special Generations on page 67
Calibration, see Part D.
59
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 8 About This Manual
- 9 Typographical Conventions
- 9 Unpacking and Inspection
- 10 Connections
- 11 The Upper Panel
- 12 the Left Side of MC5-IS
- 12 The Front Panel
- 15 Memory
- 16 Batteries
- 17 the Charging Procedure
- 18 the Battery Pack
- 19 Support for Table Top Use
- 19 the Neck Support Strap
- 20 General Description
- 20 Startup Procedure
- 21 Basic Mode
- 21 Maintenance
- 21 Calibration Mode
- 22 Help Function
- 23 The User Interface
- 23 The Status Bar
- 24 The Function Key Bar
- 24 Menus
- 25 The Display Area
- 29 Hardware Modules/Options
- 30 Other Connectable Devices
- 30 Firmware Options
- 31 Ex Information and Approvals
- 32 Safety Precautions and Warnings
- 32 General Warnings
- 33 of Electrical Modules (E and ET)
- 33 Pressure Measurement
- 34 High Pressure
- 35 Firmware Update
- 35 Recalibrating MC5-IS
- 35 The Battery Charger
- 36 Cleaning MC5-IS
- 36 Junction Module
- 38 Startup Procedure
- 39 Basic Mode, Defined
- 43 Pressure Measurement
- 43 Using Internal Modules
- 43 Using External Modules
- 44 Zeroing a Pressure Module
- 45 Current Measurement
- 46 Voltage Measurement
- 46 Measuring Low Voltages
- 47 Measuring Voltages up to ±30 V
- 48 Resistance Measurement
- 49 Switch State Sensing
- 50 Limit Switch Test
- 50 Performing the Limit Switch Test
- 52 Frequency Measurement
- 53 Pulse Counting
- 54 RTD Measurement (Temperature)
- 55 Measurement (Temperature)
- 55 Internal Reference Junction
- 55 External Reference Junction
- 58 Minimum value
- 58 Maximum value
- 58 Min/Max value
- 58 Rate of Change
- 59 Special Filtering and Resolution
- 59 Deviation Measurement
- 60 Two Ports Simultaneously
- 60 Difference Measurement
- 61 Redundant Measurement
- 62 General
- 63 the Generated/Simulated Signal
- 64 Current Sink
- 64 output terminals
- 66 Voltage Generation
- 66 between +10 V and -2.5 V
- 66 Low Voltage Generation
- 68 Frequency Generation
- 69 Pulse Generation
- 70 RTD and Resistance Simulation
- 71 Thermocouple Simulation
- 71 Internal Reference Junction
- 71 External Reference Junction
- 73 Configuration Window
- 74 Stepping
- 75 Ramping
- 80 Settings
- 82 Setting Time and Date
- 83 Display Mode Settings
- 84 Scaling
- 85 Displaying Error Values
- 86 Transmitter/Switch Simulation
- 87 Transmitter Simulation
- 88 Switch Simulation
- 89 Data Logging
- 89 General
- 89 Configuring
- 90 Starting the Data Log
- 91 Viewing the Results
- 91 a Personal Computer
- 93 Measuring Pressure
- 93 General
- 93 Pressure Type
- 94 their Naming Conventions
- 94 Square Rooting
- 95 Connections and Troubleshooting
- 95 Internal Reference Junction
- 96 External Reference Junction
- 98 Error situations
- 99 RTD Measurement, Connections
- 99 4-wire System
- 99 3-wire System
- 100 Using a Compensation Loop
- 100 2-wire System
- 101 to a Test Diode, Connections
- 102 Parallel Functions in MC5-IS
- 105 Phases of Instrument Calibration
- 106 As Found Calibration
- 106 Adjustment
- 107 As Left Calibration
- 108 Input/Output Signal Combinations
- 110 to Be Calibrated
- 111 The Instrument Window
- 112 Using MC5-IS
- 114 About Automatic Calibration
- 114 About Manual Calibration
- 115 Examples of Instrument Calibration
- 116 Pressure Transmitters
- 118 Temperature Sensors
- 120 Recorders
- 122 Electrical Limit Switches
- 124 Instrument Adjustment
- 125 Adding New Instruments
- 126 Editing Instrument Data
- 126 General Data Page
- 127 Instrument Input Page
- 127 Instrument Output Page
- 128 Calibration Settings Page
- 129 Calibration Instructions Page
- 129 Deleting Instruments
- 131 Calibration Result Windows
- 131 Calibration Run is Viewed
- 131 Deleting Calibration Results
- 135 MC5-IS and HART
- 136 Calibration Database Software
- 136 External Supply
- 137 Test Diode Connections
- 137 a Hart Communicator
- 139 Connecting to a HART Instrument
- 140 HART Instrument
- 140 Device Information
- 141 Device Settings
- 141 in Basic Mode
- 143 MC5-IS’ Instrument Database
- 144 to be Calibrated
- 145 The Calibration Procedure
- 145 in Calibration Mode
- 150 (INT-IS & EXT-IS)
- 151 Module (ET)
- 151 and Simulation
- 152 Module (RJ)
- 153 Electrical Module (E)
- 155 General
- 155 Installing MC5 Datalog Viewer
- 155 Communication Settings
- 155 File Management
- 155 Transferring data from MC
- 156 Saving Data on Hard Disk
- 156 Opening Existing Data Files
- 156 Importing and Exporting
- 156 Other Utilities
- 156 Chart Properties
- 157 Channel Properties
- 157 Zooming
- 157 Printing the Graph