Troubleshooting (only for electrical shell chairs). Vermeiren Coraille XXL, Charme Deluxe, Charme, Alesia, Charme Electrical, Altitude XXL, Regina, Altitude, Coraille

Add to My manuals
16 Pages

advertisement

Troubleshooting (only for electrical shell chairs). Vermeiren Coraille XXL, Charme Deluxe, Charme, Alesia, Charme Electrical, Altitude XXL, Regina, Altitude, Coraille | Manualzz

Shell chairs

2017-01

Check the functioning of the leg support (locking, loading, deformation, wear and tear due to loading, gas pressure springs, adjustment levers).

Check the operation of other detachable parts (example: multi-point safety belt, treatment table)

If available: Visually and functionally inspect the gas pressure springs (including the adjusting lever and Bowden cable)

Completeness of the delivery condition, instruction manual available?

Only for electrical shell chair:

Visual inspection of the power cable (EPR or similar).

Condition of the power cable (especially: squashing, abrasion, cuts, visible insulation or inner conductors, visible metallic veins, kinks, bulges, color changes of the outer covering, brittle places).

The power cable and other electric leads must be laid down securely in such a way that shearing, squashing, and other mechanical stresses are improbable.

The service must only be signed off in the maintenance plan if a minimum of all the abovementioned aspects have been checked.

1.5 Troubleshooting (only for electrical shell chairs)

The following list will assist you with the troubleshooting of your shell chair.

Problem

High stresses on the motor, blocking from mechanical system.

The shell chair does not work (no power).

Cause of problem

Not enough space to incline the shell chair to maximum position.

The power cable is not connected to the power socket.

The cables are damaged (power cable, ...) or kinks in the cables.

Not all electrical components are connected or working (motor, handset, power supply).

The thermal fuse of the power supply trigger unit overheats.

Unplug the power cable for 20 à 30 minutes. Try again, if it still not work contact your specialist dealer.