Section 1 Specifications
Specifications are subject to change without notice.
General
Dimensions
Weight
Temperature
Modes
Chemical Tests
Certification
520 x 450 x 215 mm (20.5 x 17.5 x 8.5 in.)
9 kg (20 lb), fully loaded
Tested from -20 to 55 °C (-4 to 131 °F)
Chemiluminescence Toxicity Testing
Luminescent Bacteria Toxicity Testing
Arsenic, Chemiluminescence Toxicity, Chlorine, Color,
Nerve Agents, Pesticides, pH, Total Dissolved Solids
(Conductivity)
CE marked
Luminometer—Water and chemical warfare agent resistant
Weight
Dimensions
Temperature
Battery Type
Display
Data Logging
Download Capability
Power
Warranty
Light detection for the
Luminescent Bacteria
Toxicity Test
1.4 kg (3.09 lb), including batteries
230 x 77 x 125 mm (9.1 x 3 x 4.92 in.)
Tested from -20 to 55 °C
4 AA alkaline, at least 250 test per set of batteries
Graphical LCD display with backlight for low light conditions
Up to 60 test results recorded in full detail
Up to 100 luminescent measurements for Luminescent
Bacteria Toxicity Test
Up to 100 screening results for Luminescent Bacteria
Toxicity Test
RS232
Battery powered alkaline cell, lithium cell, AA
CAUTION: For quality and safety reasons, only alkaline batteries should be used with this instrument. Use of other batteries may reduce the functioning of and/or damage the instrument electronics by overloading the electronics, or, depending on the battery type, can cause fire or an explosion.
One year
Two decades in two different ranges:
20 - 1000 relative units (default mode)
20 - 2000 relative units
Precision: 2% coefficient of variation
Arsenic
Range
Limit of detection
0 to 4 mg/L
0.01 mg/L
7
8
Chlorine (free)
Range
Chlorine (total)
Range
Color
Range
Pocket Pal™ pH Tester
Range
Accuracy
Operating temperature
Battery life
Enclosure
Warranty
Pocket Pal™ TDS Tester
Range
Accuracy
Operating temperature
Temperature compensation
Battery life
Enclosure
Warranty
0–3.5 mg/L
0–3.5 mg/L
0 to 100, 0 to 500 APHA Platinum-Cobalt Color units
0.0 to 14.0 pH
± 0.1 pH at 20 °C
0 to 50 °C
1000 hours continuous use (approximately)
IP67 Rated, waterproof (immersible); dust proof
Six months from the date of shipment
10 to 1990 µS/cm
± 2% of reading at 25 °C calibration and 25 °C sample
± 10% of temperature compensated µS/cm readings over
0 to 50 °C range
0 to 50 °C
2% per °C
1000 hours (approximately)
IP67 Rated, waterproof (immersible), dust proof
Six months from the date of shipment
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Table of Contents
- 7 Section 1 Specifications
- 9 Section 2 General information
- 9 2.1 Safety information
- 9 2.1.1 Use of hazard information
- 10 2.2 Product overview
- 10 2.2.1 Test descriptions
- 12 2.2.2 Test setup
- 12 2.3 Unpack the instrument
- 13 Section 3 Luminometer
- 13 3.1 Overview
- 14 3.2 Prepare the luminometer for use
- 14 3.2.1 Test the operation
- 15 3.2.2 Erase the results saved on the luminometer
- 16 3.2.3 Set the measurement range
- 18 3.3 Change the default settings
- 18 3.3.1 Set the LCD contrast
- 18 3.3.2 Set the waiting time and measuring time
- 19 3.4 Connect the luminometer to a printer
- 21 3.5 Connect the luminometer to a computer
- 21 3.6 Install LUMISsoft on the computer
- 23 Section 4 Chemiluminescence Toxicity Test
- 23 4.1 Overview
- 23 4.2 Prepare the reagents for luminometer calibration and sample testing
- 24 4.2.1 Prepare CT Reagent 2
- 24 4.2.2 Prepare CT Reagent 3
- 26 4.3 Calibrate the luminometer
- 28 4.4 Measure pollutants in the water sample
- 31 4.5 Show the previous results
- 31 4.6 Send previous results to a computer
- 33 Section 5 EZ Arsenic Test Kit
- 33 5.1 EZ Arsenic, 0–500 ppb (0, 10, 25, 50, 100, 250, 500)
- 34 5.2 EZ Arsenic, 0–4000 ppb (0, 35, 75, 175, 1500, 4000)
- 34 5.3 Interferences
- 35 5.3.1 Removing sulfide (optional)
- 37 Section 6 Pesticide/Nerve Agent Test
- 37 6.1 Pesticide/Nerve Agent procedure
- 39 Section 7 Free and Total Chlorine Test
- 39 7.1 Measuring hints and general information
- 39 7.2 Free chlorine procedure, 0–3.5 mg/L
- 41 7.3 Total chlorine procedure, 0–3.5 mg/L
- 43 Section 8 Color Test
- 43 8.1 Color Test, low range
- 44 8.2 Color Test, high range
- 45 Section 9 Pocket Pal™ pH Tester
- 45 9.1 Use and care of the tester
- 46 9.2 2-point calibration
- 47 Section 10 Pocket Pal™ TDS Tester
- 47 10.1 Use and care of the tester
- 48 10.2 Calibration
- 49 Section 11 Luminescent Bacteria Toxicity Test: Screening and LIMIT measure
- 49 11.1 Overview
- 50 11.2 Accuracy
- 50 11.3 Reagent description
- 50 11.3.1 Quality assurance test
- 51 11.4 Reagent storage and preservation
- 51 11.5 Prepare the reagent
- 51 11.6 Prepare the stock suspension using the LCK491 reagent
- 54 11.7 Prepare the test suspension
- 55 11.8 Sample collection, storage and preservation
- 55 11.9 Interferences
- 56 11.10 Prepare the sample
- 57 11.11 Prepare the test tubes
- 60 11.12 Measure the toxicity of the sample dilutions
- 65 11.13 Show or send previous results
- 65 11.13.1 Description of screening luminescence results
- 66 11.13.2 Description of LIMIT measure results
- 66 11.13.3 Show or send screening luminescence results
- 67 11.13.4 Show or send LIMIT measure results
- 69 Section 12 Luminescent Bacteria Toxicity Test according to ISO 11348 part 3
- 70 12.1 Overview
- 71 12.2 Accuracy
- 72 12.3 Thermostat and PC software requirements
- 73 12.4 Reagent description
- 74 12.5 Reagent storage and preservation
- 75 12.6 Prepare the reagent
- 75 12.6.1 Prepare the stock suspension using the LCK491 reagent
- 76 12.6.2 Prepare the test suspension
- 82 12.7 Sample collection, storage and preservation
- 83 12.8 Interferences
- 84 12.9 Prepare the sample
- 85 12.10 Prepare the dilutions series
- 85 12.10.1 Prepare a 9 dilution series (D 2 values and higher)
- 89 12.10.2 Prepare a 3 dilution series (D 2 values and higher)
- 93 12.10.3 Prepare a 9 dilution series (D 1 values and higher)
- 98 12.10.4 Prepare a 3 dilutions series (D 1 values and higher)
- 101 12.11 Measure the light intensity of the test suspension
- 106 12.12 Measure the light intensity of the test suspension after the sample dilutions are added
- 108 12.13 Show or send previous results
- 111 Section 13 Maintenance
- 111 13.1 General maintenance
- 111 13.1.1 Cleaning the kit
- 111 13.1.2 Cleaning the luminometer
- 111 13.2 Decontamination
- 112 13.3 Battery replacement
- 112 13.3.1 Luminometer battery replacement
- 112 13.3.2 Pocket Pal™ battery replacement (pH and TDS)
- 115 Section 14 Troubleshooting
- 117 Section 15 Replacement parts and accessories
- 125 Appendix A Luminescent bacteria risks
- 125 A.1 Risk specifications
- 126 A.2 Biosafety Level 1 information
- 126 A.3 Disposal