Operation Method
Fixed Temperature Operation
Fixed temperature operation procedure
1. Turn on the power (turn on the breaker in front)
The default value is displayed for about four seconds after turning on the power. The screen then displays the initial setting. The current temperature in furnace, operation mode character and setting temperature of overheating prevention device are displayed on respective screens.
Measurement temperature screen:
Displays the current temperature in furnace.
Setting temperature screen:
Displays the operation mode character. (Refer to Page 9)
Overheating prevention screen:
Displays the setting temperature of overheating prevention device
2. Select the operation mode
• Press the FIXED TEMP key to display "FIX", which indicates the fixed temperature operation, on the center display screen.
3. Set the temperature
• Press the FIXED TEMP key again.
• The setting temperature screen displays the character "SV" which indicates the temperature setting. Also it displays the current setting temperature with blinking. The FIXED TEMP lamp blinks, too.
• Set the temperature by pressing the "▼▲".
4. Start operation
• Press the orange START/STOP key for about one second.
The unit starts operation and the blinking FIXED TEMP lamp lights on.
5. Stop operation
• Press the orange START/STOP key for about one second.
The unit stops operation and the FIXED TEMP lamp lights off.
The screen returns to the initial setting screen.
To correct or check setting…
Press the FIXED TEMP key again to correct or check the setting.
Changing the setting temperature during operation is also possible by pressing the FIXED TEMP key.
15
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 Explanation
- 5 Table of Illustrated Symbols
- 6 Fundamental Matters of “WARNING!” and “CAUTION!”
- 7 Requirements for Installation
- 8 Requirements for Installation
- 9 Requirements for Installation
- 10 Main Unit
- 11 Control Panel
- 12 Characters of Thermo Controller
- 13 Preparation for Operation
- 13 Read the following without fail before using the vacuum pump
- 14 Operation Mode and Function List
- 16 Operation Mode, Function Setting Key, and Characters
- 17 Setting of Overheating Prevention Device
- 18 Fixed Temperature Operation
- 19 Quick Auto Stop Operation
- 20 Auto Stop Operation
- 21 Auto Stop Operation
- 22 Auto Start Operation
- 23 Auto Start Operation
- 24 Other Functions
- 24 Use calibration offset function
- 24 Use lock function
- 26 Daily Inspection and Maintenance
- 27 When not using this unit for long term / When disposing
- 27 Environmental protection should be considered
- 28 Safety Device and Error Code
- 29 Trouble Shooting
- 34 List of Dangerous Substances