Operation Method
Other Functions
Use calibration offset function
Calibration offset is a function which corrects the difference between the temperature in furnace and that of controller (sensor temperature) if arises. The function parallel corrects the difference either to the plus or minus side within the whole temperature range of unit. The function can be set or cancelled by the SUBMENU key.
Corrected temp. to plus side
Current temperature
Corrected temp. to minus side
① Start operation with the target setting temperature. Check the temperature in furnace (temperature of sample) with a thermograph after it is stabilized.
②
Check the difference between the setting temperature and that in furnace
(temperature of sample).
③
Press the SUBMENU key. Select the character "cAL", which indicates the calibration offset, using the "▲▼", and then press the ENTER key.
④
Input the difference using the "▲▼" and then press the ENTER key.
This completes the setting.
The setting range of offset correction temperature is +99℃ to plus side and
-99℃ to minus side respectively.
When it is set to the minus side, the temperature on the measurement temperature display screen falls by the setting temperature, while the temperature on furnace rises.
When it is set to the minus side, the temperature on the measurement temperature display screen rises by the setting temperature, while the temperature on furnace falls.
The unit has two-point correction function, which performs offset between low-temperature zone and high-temperature zone. Please consult our local branch office when carrying out validation of temperature controller.
Use lock function
This function locks the operation status previously set. The function can be set or cancelled by the
SUBMENU key.
①
Press the SUBMENU key. Select the character" "Lock", which indicates the lock of setting value, using the "▲▼", and then press the ENTER key.
②
The setting temperature screen displays "oFF". The setting value is locked when it is turned to "on" using the "▲".
③
Press the SUBMENU key again to cancel the lock. Select the character"
"Lock", which indicates the lock of setting value, using the "▲▼", and then press the ENTER key. Select "oFF" with the "▼" and then press the ENTER key to cancel the function.
All keys other than the START/STOP and SUBMENU keys are lock when the lock function is on.
21
Handling Precautions
WARNING!
If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the power key right away, and then turn off the circuit breaker and the main power. Immediately contact a service technician for inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair work yourself, since it is dangerous and not recommended.
Substances that cannot be used
Never use explosive substances, flammable substances and substances that include explosive
or flammable ingredients in this unit. Explosion or fire may occur. (Refer to page31 “List of
CAUTION!
Do not step on this unit
Do not step on this unit. It will cause injury if this unit fall down or break.
Do not put anything on this unit
Do not put anything on this unit. It will cause injury if fall.
During a thunder storm
During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main power. If this procedure is not followed, fire or electrical shock may be caused.
Do not touch high-temperature parts
Some parts of this unit may become hot during and just after operation. It may cause burns.
Sample loading
Overload of samples may cause the malfunction of the temperature control. Do not use this unit with too much samples. Leave a certain space between samples.
Return after power failure
When power is supplied after a power failure, the device automatically starts operation again with the same state as just before the power failure. It is danger that the device starts unattached operation after a power failure. We recommend for you to turn off the switch of this unit if a power failure occurs during operation.
22
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 Explanation
- 5 Table of Illustrated Symbols
- 6 Fundamental Matters of “WARNING!” and “CAUTION!”
- 7 Requirements for Installation
- 8 Requirements for Installation
- 9 Requirements for Installation
- 10 Main Unit
- 11 Control Panel
- 12 Characters of Thermo Controller
- 13 Preparation for Operation
- 13 Read the following without fail before using the vacuum pump
- 14 Operation Mode and Function List
- 16 Operation Mode, Function Setting Key, and Characters
- 17 Setting of Overheating Prevention Device
- 18 Fixed Temperature Operation
- 19 Quick Auto Stop Operation
- 20 Auto Stop Operation
- 21 Auto Stop Operation
- 22 Auto Start Operation
- 23 Auto Start Operation
- 24 Other Functions
- 24 Use calibration offset function
- 24 Use lock function
- 26 Daily Inspection and Maintenance
- 27 When not using this unit for long term / When disposing
- 27 Environmental protection should be considered
- 28 Safety Device and Error Code
- 29 Trouble Shooting
- 34 List of Dangerous Substances