Operation Method
Auto Start Operation
3. Stop/terminate timer operation
• The operation starts automatically at setting time.
• Press the START/STOP key for one second to stop or terminate operation. The screen returns to the initial setting screen.
To correct or check setting…
Changing the setting temperature or time during operation is possible by pressing the TIMER key. Use the "▼▲" to change the setting value. Press the ENTER key respectively after changing the setting.
They are not changeable after the unit starts operation. In this case, stop the operation by pressing the
START/STOP key, then set the value again.
Press the "▼" to display the setting temperature, operation mode and residual time on the setting temperature screen.
20
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 Explanation
- 5 Table of Illustrated Symbols
- 6 Fundamental Matters of “WARNING!” and “CAUTION!”
- 7 Requirements for Installation
- 8 Requirements for Installation
- 9 Requirements for Installation
- 10 Main Unit
- 11 Control Panel
- 12 Characters of Thermo Controller
- 13 Preparation for Operation
- 13 Read the following without fail before using the vacuum pump
- 14 Operation Mode and Function List
- 16 Operation Mode, Function Setting Key, and Characters
- 17 Setting of Overheating Prevention Device
- 18 Fixed Temperature Operation
- 19 Quick Auto Stop Operation
- 20 Auto Stop Operation
- 21 Auto Stop Operation
- 22 Auto Start Operation
- 23 Auto Start Operation
- 24 Other Functions
- 24 Use calibration offset function
- 24 Use lock function
- 26 Daily Inspection and Maintenance
- 27 When not using this unit for long term / When disposing
- 27 Environmental protection should be considered
- 28 Safety Device and Error Code
- 29 Trouble Shooting
- 34 List of Dangerous Substances