SZM 45B2, SZM 45T2, SZM 45V2 microscope User Manual
Oglas
Oglas
Contents
1.
Before use ………………………………………… 1
2.
Nomenclature ……………………………………… 2
3.
Assemblage ………………………………………… .5
4.
Operation …………………………………………… 7
5.
Configuration ……………………………………… 13
6.
Technical parameter ……………………………… 15
7.
Troubleshooting …………………………………… 17
1
1 Before use
1----1 NOTICE
1) Microscope ought to be placed in a dry and clean place. Do not expose the microscope in the sun directly. Avoid high temperature and violent vibration.
2) As microscope is a precision instrument, handle with care, avoiding impact or abrupt movement during transportation.
3) To keep the image clear, do not leave fingerprints or stains on the surfaces of the lens.
4) Never turn the left and right focusing knob in the adverse direction at the same time, otherwise the microscope will be damaged.
5) Hold the camera with one hand for fearing of falling when you take the films out of the big camera.
1----2 MAINTENA NC E
1) All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue; Fingerprints or oil marked on it should be wiped off with a tissue moistened with a small amount of xylene or a
3:7 mixture of alcohol and ether.
2) Never use the organic solution to clean the other surface (especially the plastic surfaces).
If necessary, please choose the neutral detergent.
3) Do not take the microscope apart for fearing that it is damaged.
4) After using, cover the microscope with the dust-cover provided and store it in a dry and clean place free from moisture to prevent rust.
5) To keep the performance of the microscope, please check it periodically. The detail can be gotten from the agent nearby.
1
2 N o m e n c l a t u r e
2----1 SZM-45B 2
Eyepiece shield
Eyepiece
Diopter adjusting ring
Head
Illumination device
Zoom control knob
Focusing
Lamp lock-screw
Glass stage
Clip
Base
Top variable light control
Fuse
Bottom variable light control
Switch
2
2----2 SZMT2
D i o p te r a d ju s tin g r in g E ye p ie c e sh ie l d
Zo o m c o n tro l k n o b
E y e p ie c e
H e a d lo c k -sc re w
H e a d
Fo c u sin g k n o b
G la ss sta g e
B a se C l i p
3
2----2 SZM-45 V2 ZM
Eyepiece shield
Eyepiece
Diopter adjusting ring
Illumination device
Focusing knob Digtal head
Lamp lock-screw
Glass stage Base
Clip
4
3 A s s e m b l a g e
3----1 SZM-45B 2
Eyepiece shield
Eyepiece
Head
Head lock-screw Illumination device
Clips
Glass stage
Base
5
3----2 SZM-45 T2+S ZM -PH/S ZM -CTV
PK-mount
Photo eyepiece
Eyepiece shield
Eyepiece
Head
Clip
Glass stage
Base
6
C-mount
Illumination device
4 O p e r a t i o n
4----1 Use the glass stage
1) Press the glass stage on the sunken place then the other side of the glass stage will be lifted. (Fig.1)
Fig.1
Fig.2
4----2 Adjust the degree of tightness of the focusing arm.
1) If you want to adjust degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight, otherwise, is loose.
2) The suitable position of tightness can make the adjustment more comfortable and prevent the focusing bracket from slipping down by its weight during the observation. (Fig.2)
4----3 Set the s pecim en slid e
1) Set the specimen on the center of stage plate. If necessary, clamp the slide with the clips.
2) Turn on the light.
4----4 Adjust the s peci m en slid e
1) Turn the zoom control knob to the maximum magnification.
2) Turn the diopter adjusting rings to the zero.
3) Observe the specimen through the right eyepiece and make the jkj image clear by turning the focusing knob.
Fig.3 4) Rotate the zoom control knob to the minimum magnification.
5) Observe the specimen through the right eyepiece and make the image clear by turning the right diopter adjusting ring② .(Fig.3)
6) Redo the step(1),(3),(4)and (5) till the right adjusting ring is more precise.
7) Do the step (4) and make the image clear which is observed through the left eyepiece by turning the left diopter adjusting ring
①. (Fig.3)
7
4----5 Adjust the int er p upillar y distan ce
1) Adjust the prism housing along the direction of arrowhead of the
Fig.4 till the observation is comfortable.
Fig.4
4----6 Use Eyepiece shields
1) For user who does not wear glasses, hold the diopter-adjusting ring to prevent them from rotating and turn the eyepiece till the eyepiece shields fit the observer well.
2) For user who wears glasses, take the eyepiece shields off before
observation
4----7 Mount and Remove the Optional Eyepiece Micrometer
1) Turn and remove the mounting ring② from the eyepiece.(Fig.5)
2) Clean the eyepiece micrometer① and mount it to the mounting
ring with the inscription side downward.
3) Gently twist the mounting ring with the eyepiece micrometer into the eyepiece till tightening② securely.
4) To remove the eyepiece micrometer, take down the mounting ring③
Fig.5 by twisting and take out of the micrometer, and then wrap it.
4----8 Install the illum ination devi ce
1)Insert the illumination device① in the bracket with the protrudent
side toward the lock-screw② and tighten the lock-screw. (Fig.6)
2) Put the plug into the socket of the pillar stand③.
Fig.6
Pole
4----9 Choose th e o pti cal s yst em
1) You can alternate the binocular observation and video capture by
pushing or pulling the pole. You can attain binocular observation by
pushing the pole inside, or attain video capture by pulling it outside.
No matter what optical system is chosen, push or pull the pole
thoroughly.
Fig.7
8
4----10 Mo u n t th e p h o to e y e p ie c e a n d th e P K -m o u n t
a d a p te r
1) Put the photo eyepieces socket of the tri-ocular.
2) Connect the PK-mount adapter with the photo eyepiece, and then tighten the lock-screw. (Fig.8)
Fig.8
C-mount
Fig.9
Fig.10
Fig.11
4----11 Adjust the CTV
1) Adjust the CTV to a suitable position by rotating C-mount.
Note: The range of the adjustment: 1~2mm in general.(Fig.9)
1 2 C o n n e c t th e D ig i ta l h e a d Wi th th e Mo n i to r o r T V
se t
1) Plug one end of the PVA cable into the socket of the digital head. (Fig.10)
2) Plug the C-VIDEO or S-VIDEO of the PVA cable into the correct socket of Monitor (TV set).
3) Connect the 12V DC power with the power socket of the
PVA cable.
4----13 Appea r th e im a ge on th e M onito r o r TV
1) Connect the power supply and then turn on the Monitor or TV.
2) For the monitor, the connect sign model must be chosen
(C-video or S-video) and for TV, the channel must be set to the video channel.
3) Pull the pole out and adjust the focusing knob and then the image will appear on the screen clearly.
4----14 Connect with t he com puter
1) Plug one end of the PVA cable into the socket of the digital head.
2) Plug one of the C-VIDEO or S-VIDEO into the A/D board.
3) Plug the USB of the A/D board into the USB socket of the computer. (Fig.11)
4) If your computer is mounted capture card, you can connect the C-VIDEO or S-VIDEO with the computer directly.
5) Connect the 12V DC power with the power socket of the PVA cable.
9
4----15 Appea r th e im a ge on th e com puter
1) Turn on the power supply and let the computer work.
2) Install the software and the driver of the A/D board. (If they have been installed, this step can be omitted.
3) Double click the icon of the software, and then the video window will appear. You can set the size of the window according to your linking
4) Draw out the pole and adjust the focusing knob, and then the image will appear on the computer screen clearly.
5) If no image or the image without color, it may be because the model of the input signal does not match the output signal of
CCD or the model of C-VIDEO/S-VIDEO is not correct. The detail of operation refers to《Software operation manual》.
1 6 Appear th e im age on th e com put er an d th e
Monito r s yn chron ou sly
1) Do step 4 and step 4----14 to connect the computer and the Monitor.
2) Operate step 4
and step
4----15
, we can make the image appear on the computer and Monitor at the same time.
4----17 Adjust the im age
1) Put the base, stand and digital head correctly, then fix the lock-screw tightly.
2) Put the object on the base stage.
3) Observe the object through the eyepiece and adjust the focusing knob to make the image of the object clearly.
4) Move the digital head or the object gently to adjust the image agreeing with observer.
4----18 Brief instru cti o n fo r th e s oftwa re
1) The program design of the software is up to date, and the
Chinese/English interface can berth powerful delineation bar which be used much conveniently and rapidly. You can finish most of analyze work only to click the mouse.
2) Can afford many powerful area choosing tools which can analyse any area your linking at will, such as adjusting hue and image, dealing with mathematical morphology, image matching, texture analyse, character identify and so on.
10
Fig.12
3) Geometry character measuring function, automatically analyzing function such as slightness body, grain body, line body and so on. The outcome can be kept in data and can be made into chart and so forth.
4 Use th e whi te balan ce
1) The CCD has auto white balance when the white balance switch is on“ON”.
2) Please put the switch on “ON” in general. Let the switch be
“OFF” only in special, for example, observing the red cell, otherwise the color of red cell will be adjusted into white.
3) If you want to observe another single color, please let the switch be“ON”again when you finish the observation, and put the switch on “OFF” again after auto balance, or the color of the image will be distortion.(Fig.12)
4----20 Adjust the brightness of the bottom light
1) Turn the adjustable light knob① according to the sign marked on
the base, along the clockwise the brightness will be added,
otherwise it will be weakened. (Fig.13)
Fig.13
4----21 Replace th e la m ps
1) Press the stage on the sunken place then the other side will be
lifted. (Fig.14)
2) Take the lamp out of the jack.
3) Put a new lamp into the jack thoroughly.
Fig.14 4) Recover the stage plate. (Fig.15)
Note: ① Before replacing the lamps, turn off the power first.
② Avoid violence while the lamp is plugged into the jack.
Fig.15
11
Fig.16
4----22 Replace th e fus e
1) Screw the fuse tube out with a screwdriver and then pull the fuse
out of the tube①.
2) Renew the fuse and mount it in an adverse way. (Fig.16)
12
C o n f i g u r a t i o n
5----1
Parts
Configuration Model
Specification SZM45B1 SZM45B2 SZM45B3 SZM45T1 SZM45T2 SZM45T3 SZM45V2 SZM45P2
Eyepieces
Binocular
Tri-ocular
Digital head
A/D board
PAV cable
Power adaptor W-BW230-12-4.8
Software
SZMEWh10X20
SZMEWh15X15
SZMEHWh20X10
SZM7045
SZM7045TR
SZM7045V
W-A/D
W-RVB-PAV
Pholib3.0
O
O
O
O
Auxiliary objective
SZMAO0.5/165mm
SZMAO1.5/45mm
Focusing arm
Stand
Transformer
Epi-illuminator
Hold for illuminator
Photo device
TV adapter
SZMAO2/30mm
SZMA1
SZMST1
SZMST2
SZMST3
SZSTL1
SZSTL2
SZT1
SZML1
SZFH1
SZPD1
SZMPH
SZMCTV
O
O
O
O
O
O
O
O
Gem clamp
Dark field condenser
Light
Ring
S/ST-GC
ST-30-2L-HJ-01
SZRL O O fluorescence
Box
Inside foam
Outside carton
O O O
Note: The items marked“O”included and others for option
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
13
SZM series configuration
6 T e c h n i c a l p a r a m e t e r
6----1 SZM7045/SZM7045TR
Eyepiece
10X/20
15X/15
20X/10
Standard configuration
Working distance100mm
Magnification
7X
Field of view
28.6
45X
10.5X
67.5X
14X
90X
4.4
21.4
3.3
14.3
2.2
0.5X
Working distance165mm
Magnification
Field of view
57.1 3.5X
22.5X
5.25X
33.75X
7X
45X
8.9
42.8
6.7
28.6
4.4
Auxiliary objectives
1.5X
Working distance
45mm
Magnification
10.5X
Field of view
19
67.5X
15.75X
101.25X
21X
135X
3
14.3
2.2
9.5
1.5
2X
Working distance
30mm
Magnification
Field of view
14.3 14X
90X
21X
135X
28X
180X
2.2
10.7
1.7
7.1
1.1
6----2 The base electronic specification of SZM series
Model
SZMST1 SZMST2
Parts
Power supply
Transformer
Top light
Illuminator Bottom light
No
No
No
SZMST3
220V-50Hz、
110V-50/60Hz
Input: 220/110VAC
Output: 12V DC/45W
220V-50Hz、
110V-50/60Hz
Input: 220/110VAC
Output: 12V DC/45W
12V/15W halogen lamp 12V/15W halogen lamp
12V/15W halogen lamp
220/110V、7W fluorescent lamp
15
6----3 CCD specification
CCD Model
Specification
Pick-up Device
Vertical Resolution
Number of Pixels
Scanning System
White Balance
Back light compensation
Signal/Noise ratio
Gamma Characteristic
Minimum illumination
Input voltage
NTSC PAL
1/3 ″
450 TV line
(H)768*(V)494
1/3 ″
450 TV line
(H)712*(V)582
525 lines, 60Field/Second 625lines,50 Field/Second
Can be switched between auto white balance and hand white balance
Auto Auto
More than 46db More than 46db
0.45
3 Lux
12V DC(9V-14V)
0.45
3 Lux
12V DC(9V-14V)
Power consumption 1.85 or less
★ Working distance is fixed regardless of the magnification factor.
★ Total mag.= Zoom mag. X Eyepiece mag. X Auxiliary objective mag.
1.85W or less
Field number of eyepiece
Diameter of field of view(mm)=
Zoom mag.X Auxiliary objective mag.
★ Photo adaptor mag.= Zoom mag.( ×Auxiliary objective mag.)×Eyepiece mag.
★ TV adaptor mag.=Zoom mag.(Xauxiliary objective mag.)X C-mount TV adaptor middle
★ Field of video view is 83%
★ Total video magnification range is 18∽117
16
7 T r o u b l e s h o o t i n g
The performance of the microscope can’t be made fully because of unfamiliar using, this table will give some advices.
7----1
General troubleshooting
Trouble
1、Double images
Cause
Interpupillary distance is not correct
Remedy
Readjust it
Diopter adjustment is not correct Readjust it
Magnification of each eyepiece is not the same size
Mount the same size eyepiece
2、Dirt appears in the field of view
3、Image is not clear
10、Bulb does not work when the switch is on
Dirt on the specimen Clean the specimen
Dirt on the surfaces of eyepiece Clean the surface
Dirt on the surfaces of the objectives
Clean the objectives
4、Image is not clear while the focus changing
Diopter adjustment is not correct
Focus is not correct
5、The focusing knob is not smooth The focusing knob is too tight
Readjust the diopter
Readjust the focus
Loosen it to a suitable position
6、The image is obscure because of the head slipping down by itself during observation
7、Incision image appears in the field of view or of the video view
8、The image on the monitor is not clear when the focusing knob is turned.
The focusing knob is too loose
The pole is not in correct position
The focus of video is not correct
Tighten it to a suitable position
Pull or push it to the correct position
Readjust the focus of video to a correct position
9、Eyes fell tired easily
Diopter adjustment is not correct Adjust the diopter
Brightness of light is not correct Adjust the brightness
No power supply Check the connection with the power supply
Insert it correctly
11、Bulb is burned out suddenly
12、Brightness is not enough
13 、 The bulb flickers or the brightness is unstable
The bulb was not inserted correctly
Bulb is wrong
Use the wrong bulb
The voltage is too high
Use a wrong bulb
The voltage is too low
The bulb will burn out soon
The bulb was not inserted correctly
Replace with a new one
Replace with a correct one
Control the voltage
Eg: use voltage regulator
Replace with a correct one
Increase the input voltage
Replace with a new one
Insert it correctly
17
7----2 Vid eo troubl esh ooting
Trouble
1、Incision image appears in the video view
2、Dirt appears in the video view
3、Image is not clear while the focus changing
4、No image on the TV screen
5、No image on the Monitor
6、The software run slowly or the window of the view does not come out
7、The image is not correct on the view window
Cause
The pole is not in correct position
Draw it to the correct position
Dirt on the specimen Clean the specimen
Dirt on the surface of objective Clean the surface
The image is not clear in the high magnification
The draw pole is not in correct position
Remedy
Readjust the high magnification
Draw it to the correct position
Objective cover is not open
TV is not on Video channel
Connection is not correct
Objective cover is not open
The input signal does not accord with the signal be chosen on the
Monitor
12V DC power does not be connect
No input signal of A/D board
Open it
Choose the correct one
Reconnect the circuit
Open it
Choose the correct signal model
Connect the 12V DC power
Reconnect the C-Video or
S-Video signal
Choose the correct signal model which match the input signal
The input signal does not accord with the signal which is chosen in the driver of the A/D board
The CCD model chosen in the driver of the A/D board does not accord with the real CCD
Choose the correct CCD model
18