Cambridge Sensors MicroDot Xtra Data Sheets
Below you will find brief information for Xtra Blood Glucose Monitoring System. This document serves as an Operations & Quality Assurance Procedure Manual, guiding healthcare professionals through the initial setup, control solution testing, and patient testing procedures. It also covers meter specifications, in-service training outcomes, material safety data sheets, and troubleshooting tips, ensuring proper and accurate use of the device.
Advertisement
Advertisement
CAMBRIDGE SENSORS USA, LLC
Customer Service: toll free: 877.374.4062 Monday–Friday 9am–5pm CST www.microdotcs.com
FAST. ACCURATE. EASY TO USE.
Xtra
OPERATIONS & QUALITY ASSURANCE
PROCEDURE MANUAL
For
Healthcare
Professionals
Xtra Xtra
Chapter 10
®
Xtra
Cleaning Your Meter
First, remove the anti-microbial meter cover. Second, use a lint free cloth dampened with soapy water or isopropyl (70% - 80%) to clean the outside of the blood glucose meter.
Microdot Meter Disinfecting Guideline: To disinfect the meter, dilute 1 mL of household bleach (5% - 6% sodium hypochlorite solution) in 9 mL of water to achieve a 1:10 dilution (final concentration of 0.5% - 0.6% sodium hypochlorite). The solution can then be used to dampen a paper towel (do not saturate the towel). Then use the damped towel to thoroughly wipe down the meter.
Please note that there are commercially available 1:10 bleach wipes from a variety of manufactures.
Germicidal wipes are readily available through major medical distributors. To use these products, remove a wipe from container and follow product label instructions to disinfect meter. Take extreme care not to get liquid in the test strip dock or key code parts of the meter.
Cambridge Sensors USA, LLC
Customer Service available from from 8am - 5pm CST Monday - Friday
Toll free 1.877.374.4062 | www.microdotcs.com
26
Cambridge Sensors USA, LLC
Contents
TABLE OF CONTENTS
Chapter
1. Microdot® Xtra Meter Overview & System Components
2. Initial Setup of the Microdot Xtra Meter
3. Control Solution Testing
4. Testing with Patients
5. Microdot Xtra Meter Specifications
6. In-Service Training Outcomes
7. Microdot Xtra Blood Glucose Meter Log Sheets
7A. Microdot Xtra BGMS In-Service Form
7B. Microdot Xtra BGMS Qualified Trainers Form
7C. Microdot Xtra BGMS Qualified Trainer/Operator
Certification Form
7D. Microdot Xtra BGMS Quality Control Record
8. Material Safety Data Sheets
8A. Microdot Xtra Test Strips
8B. Microdot Control Solutions
9. Troubleshooting the Microdot Xtra Meter
10. Manufacturer Cleaning Instructions
Notes
Page
15
16
17
18
19
2 - 4
5 - 7
13
14
8 - 10
11 - 12
20
21 - 24
25
26
27 - 28
Cambridge Sensors USA, LLC
1
Chapter 1
® Xtra BLOOD GLUCOSE METER OVERVIEW
This section provides general information on the Microdot Xtra Blood
Glucose Monitoring System
The Microdot Xtra Blood Glucose Monitoring
System is a portable whole blood testing system which performs tests with quick and accurate results.
It consists of three main parts:
1. Microdot Xtra Blood Glucose Meter
2. Microdot Xtra Test Strips
3. Microdot High and Low Control
Solution
These products have been designed, tested and proven to work together as a system to produce safe and accurate results.
Use only Microdot Xtra Test Strips and Control Solutions with the
Microdot Xtra Glucose Meter.
Your system includes:
Microdot Xtra Meter
Anti-microbial Meter Cover
Quality Assurance Manual
Warranty Registration Card
3V Lithium Battery (installed)
Additional components include:
Microdot Xtra Test Strips
Microdot Control Solutions
2
Cambridge Sensors USA, LLC
Chapter 9
The blood glucose level is higher than
525 mg/dL.
The blood glucose level is lower than
20 mg/dL.
Error message that indicates that there is a problem with the meter, e.g., measurement error (time out, overflow, offset) or temperature out of range.
Error message that indicates that there may be a problem with the test strip, e.g, the test strip may be damaged, moved, or removed during testing, or inserted improperly.
This message indicates very high blood sugar. You should recheck the blood glucose level. If “Hi” again, follow facility protocol.
This message indicates very low blood sugar. You should recheck the blood glucose level. If “Lo” again, follow facility protocol.
Review the instructions and try again with a new test strip. If the problem persists, contact Customer Service.
Check the test strip for damage and retest as necessary. Repeat the test.
If the error message appears again, contact Customer Service.
Error message could be caused by a used or damaged test strip.
Repeat the test with a new strip.
If the error message appears again, contact Customer Service.
Error message indicates serial communications error.
If error persists, contact Customer
Service.
The battery sign appears on the display with the unit of measurement. The power of the battery is getting low. You can complete about 50 more tests from the time this symbol first appears.
Test results will be accurate but replace the battery as soon as possible. Battery type is
CR2032 3 volt.
Cambridge Sensors USA, LLC
25
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section XII - Transport Information (See also Section IX)
ADR / RID
CEFIC Tremcard
Hazchem Code
Kemmler Code
IMDG Classification
IATA Classification
Marine Pollutant
UN Number
UN Class
UN Packing Group
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Section XIV - Regulatory Information
EEC Hazard Classification
Risk Phrases
Safety Phrases
Section XV - Other Information
Not known
Not known
Not known
Directives 88/379/EEC and 91/155/EEC have been considered when compiling this MSDS; the information is provided for health and safety assessment by an industrial user. Reference should be made to any relevant local or national health, safety or environmental legislation.
This information does not constitute indication of suitability for specific uses. The information, data, and recommendations contained herein are based upon information believed by Cambridge
Sensors USA, after reasonable investigation and research, to be accurate. However, Cambridge
Sensors USA does not warrant the accuracy of this information. All materials and mixtures may present unknown hazards and should be used with caution. When necessary or appropriate, independent opinions regarding the risk of handling or exposure should be obtained from trained professionals.
Cambridge Sensors USA disclaims any warranty against patent infringement and the implied warranties of merchantability and fitness for a particular purpose. Customer’s sole and exclusive remedy shall be replacement of the product or return of the product and refund of the purchase price, at Cambridge
Sensors USA option. In no case can Cambridge Sensors USA be liable for incidental or consequential damages, including lost profits.
Chapter 1
SYSTEM COMPONENTS
LCD Screen: Shows blood glucose result and symbols that guide you through the test.
Up/Down Toggle Buttons: Scrolls the memory, sets time and date.
Eject Button: Releases test strip.
Test Strip Port: Insertion site for test strip.
Serial Number: Located in the upper left hand corner of the meter. Required when calling Customer
Service for troubleshooting or meter replacement.
Battery Compartment: Holds one 3V Lithium battery (CR2032).
SN 00023463
Customer Service:
888-425-1149
24
Cambridge Sensors USA, LLC
Customer Service Number: Located on the back of the meter.
Call this number if you have any questions or
problems with the Microdot Xtra Glucose Meter.
(877) 374-4062
Cambridge Sensors USA, LLC
3
Chapter 1
SYSTEM COMPONENTS
The Microdot Xtra Blood Glucose Test Strip offers the latest advances in biosensor, auto-code technology. Blood is applied to the top edge of the Microdot Xtra Test
Strip and is automatically drawn into the white channel where the reaction takes place. Only 600 nanoliters (0.6 microliters) is needed.
The Microdot Xtra Test Strip consists of the following parts:
Top Edge: Apply a drop of blood here, where the white channel meets the top of the edge strip.
White Channel: This is where you check if enough blood has been applied to the top edge.
Contact Bars: Insert this end of the test strip into the meter.
Push firmly until the strip can go no further.
4
Cambridge Sensors USA, LLC
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section IX - Physical and Chemical Properties
Pure Substance or preparation
Appearance/Odor
Odor Threshold
Melting/Freezing
Partition
Evaporation Rate
Vapor Pressure (mmHg)
Vapor Density (air =1)
Viscosity
Volatiles
Volatile Organic Compounds
Auto-flammability
Flash Point
Oxidizing Properties
Stability
Materials to Avoid
Hazardous Decomposition
Products
Preparation Physical Form Liquid
Blue, odorless pH AS is neutral
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Stable
Strong bases, strong acids and water reactive materials
Thermal decomposition may emit carbon monoxide and carbon dioxide.
Section X - Toxicological Information
Route of Entry
Effects of Chronic Exposure
Effects of Acute Exposure
Special Health Effects
Target Organs
Section XI - Ecological Information
Potential Effect on Environment
Potential to Bioaccumulate
Mobility
Ecotoxicity
Persistence and Degradeability
Aqua Toxicity
Section XII - Disposal Considerations
Ingestion, skin and/or eye contact
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Water Disposal
Empty Container Warnings
Please consult local, state and federal regulations for additional guidance on disposal.
Not known
Cambridge Sensors USA, LLC 23
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section V - Fire Fighting Measures
Auto-flammability
Flash Point (test method)
Extinguishing Media
Special Fire Fighting
Procedures
Fire and Explosion Hazards
Hazardous Combustion
Products
Upper Explosion Limit (%)
Lower Explosion Limit (%)
Not determined
Not determined
Use fire extinguishing media appropriate for site collections.
Structural fire fighting gear and self-contained breathing apparatus will provide adequate protection if this product is in a fire area.
Not determined
Thermal decomposition may emit carbon monoxide and carbon dioxide.
Not determined
Not determined
Section VI - Accidental Release Measures
Spill and Leak Procedures Use an absorbent material to contain/pick up the spilled solution. Place all contaminated disposals into a suitable container, seal, label and hold for disposal.
Section VII - Handling and Storage
22
Storage Temperature
Handling/Storage
Ventilation Requirements
Sensitivity to Static Electricity
Sensitivity to Mechanical Impact
Store vials as directed in the package insert.
Handle and store vials as directed by the package insert.
No special requirements.
Not known
Not known
Section VIII - Exposure Controls / Personal Protection
Respiratory Protection
Ventilation
Protective Gloves
Other Protective Equipment
Other Engineering Controls
Work Practices
Hygienic Practices
Cambridge Sensors USA, LLC
Not required under normal use of this product.
Not required under normal use of this product.
Wear appropriate gloves to prevent skin contact. Replace torn or punctured gloves promptly.
Wear appropriate eye protection to prevent eye contact.
Wear appropriate body protection to prevent skin contact.
Eye wash stations and deluge showers.
Good laboratory technique should be used when handling this product. Observe appropriate chemical hygiene.
Do not place in mouth.
Do not eat, drink or smoke while working with this product.
Upon completion of work activities involving this product, wash hands thoroughly with soap and water.
Chapter 2
Before using the Microdot Xtra Meter for the first time, you should set the actual Time,
Date and Year. The unit of measurement is preset and can not be changed.
Setting the Time, Date and Year
Enter the Set Mode.
To Enter the set mode, turn the meter on by pressing the C button. After the segment test, the time and date will start to flash.
Step 1
Set the Time and Date Format.
Pressing the C button again will now toggle between US and International Time and Date formats.
For US: 12h Time format, mm-dd (begins with
AM setting). To accept the desired setting, press the M button.
Step 2
Set the Hour.
The Hour will start to flash. It can now be changed by pressing the C button. The US setting will begin at 12:01 AM and will move to PM. To accept the correct setting, press the M button.
Step 3
Set the Minutes.
The Minute will start to flash, it can now be changed by pressing the C button. To accept the correct setting, press the M button.
Repeat Steps 1 & 2 to set Month and Year.
1.
2.
3.
Cambridge Sensors USA, LLC
5
Chapter 2
Setting the Beeper
After setting the Time and Date, the Sound symbol will appear and can now be changed by pressing the C button. To accept the setting, press the M button.
This option is used to switch on or off the beeper. When turned ON, a sound will be heard when blood or control solution is applied to the strip and when the test is finished. The sound will be heard when an error has occurred or if an alarm is triggered.
When the M button is pressed to accept the beeper option, the meter will display END and switch off.
6
Cambridge Sensors USA, LLC
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section I - Identification
Trade Name:
Common Name:
Supplier:
Emergency Phone Number:
Microdot Control Solutions
Control Solutions for use with Microdot Xtra
Blood Glucose Test System
Cambridge Sensors USA
10051 Bode Rd.
Palinfield, IL 60585
(877) 374-4062
Section II - Composition / Information on Ingredients
Chemical Family:
Chemical Name:
Salt solution
N/A
Section III - Hazards Identification
Eye Contact:
Skin Contact:
Ingestion:
May cause irritation if in contact with eye.
Skin contact may cause irritation.
Though not a likely route of occupational exposure, ingestion of the product may cause choking, vomiting or nausea.
N/A
N/A
Inhalation:
Chronic Exposure
Section IV - First Aid Measures
Eye Contact:
Skin Contact:
Ingestion:
Inhalation:
Chronic Exposure:
Although no adverse health reactions are expected from the normal use of this product, it is recommended to flush eyes with water and seek medical advise whenever there is a potential injury to the eye.
If contact with this product leads to reddening, inflammation or irritation, flush the exposed area with running water. If irritation persists, get medical attention.
As a precaution, get medical attention if there has been ingestion of this product.
If breathing becomes difficult, remove victim to fresh air and get medical attention.
N/A
Cambridge Sensors USA, LLC
21
Chapter 8
MATERIAL SAFETY DATA SHEET
Xtra Blood Glucose Test Strips
Section I - Identification
Trade Name:
Common Name:
Supplier:
Emergency Phone Number:
Microdot Xtra Blood Glucose Test Strips
Blood glucose test strips for use with Microdot
Xtra Blood Glucose Test System
Cambridge Sensors USA
10051 Bode Rd.
Palinfield, IL 60585
(877) 374-4062
Section II - Composition
This product does not present a physical or health hazard under reasonable use or under emergency situations involving a release of only this product. This material is therefore not considered to be a “Hazardous Chemical” as defined by the Federal Occupational Safety and
Health Administration in the Hazard Communication Standard (29 CFR 1910.1200) or the equivalent standards generated by state agencies.
ACCORDINGLY NO MATERIAL SAFETY DATA SHEET IS REQUIRED FOR THIS PRODUCT.
MSDSs that represent non-hazardous chemicals are not covered by the HCS. Paragraph 29
CFR 1910.1200 (g) (8) of the standard requires that “the employer shall maintain in the workplace copies of the required MSDSs for each hazardous chemical, and shall ensure that they are readily accessible during each work shift to employees when that are in their work area(s)”.
OSHA does not require or encourage employers to maintain MSDSs for non-hazardous chemicals. Consequently, an employer is free to discard MSDSs for non-hazardous chemicals.
20
Cambridge Sensors USA, LLC
Chapter 2
Using Meter Memory
Your Microdot Xtra Meter stores the 500 most recent blood glucose and control solution test results and insulin data with date and time in the memory. It also provides you with 14-day averages of your blood glucose test results.
You can review the test results in memory with these easy steps.
1.
Step 1
Enter the Memory Mode.
To Enter Memory Mode, turn on the meter by pressing the M button.
The meter will display the last result with
Mem. symbol, time and date.
Step 2
Recall Test Results.
Previous results can be displayed by pressing the C button. As long as the C button is held, the meter will scroll through the memory displaying the result and the memory location. Once the button is released, the meter will display the selected result with its Time and Date. When the memory is full (500 results stored), the oldest result is dropped as the newest is added.
NOTE: When using the meter for the first time
“Mem.---” will appear, showing that there are no test results stored in memory.
2.
Cambridge Sensors USA, LLC
7
Chapter 3
CONTROL SOLUTION TESTING
This chapter describes the necessary steps to test with the Microdot® Control
Solutions in order to validate the performance of the Microdot Xtra Meter and
Microdot Xtra Strips.
Operating Guidelines: When Should You Conduct a
Control Solution Test?
Any time you open a new vial of test
strips.
Whenever you think the system
is not working properly.
If the blood glucose test results
differ from the resident’s symptoms
or non-symptoms.
If you believe the results are not
accurate.
If you drop the meter.
If the vial of test strips has been
left open for an extended period
of time.
It is critical to follow the Operating
Guidelines, on the right, to obtain
accurate results while using the
Microdot Xtra Glucose System.
Use only Microdot Control Solutions
(High and Low).
Check the expiration date on the
control solution vial. Do not use
if expired or if the discard date
has passed.
Control solution, meter and test
strips should come to room
temperature before testing
(66-77˚F / 20-25˚C).
Use solution for three months
after first opening. Record the
discard date (opening date
plus three months) on the
control solution vial. Discard
after three months.
Close tightly and store the control
solution at temperatures between
50-86˚F (10-30˚C).
Do not refrigerate. Do not freeze.
8
Cambridge Sensors USA, LLC
Chapter 7
® Xtra BLOOD GLUCOSE MONITORING SYSTEM
QUALITY CONTROL RECORD
Microdot Xtra Meter Serial Number
Month/Year
Date
Station/
Initials
Test Strip
Lot #
Low
Control
Range
(mg./dl.)
Low
Control
Result
(mg./dl.)
High
Control
Range
(mg./dl.)
High
Control
Result
(mg./dl.)
Notes:
Cambridge Sensors USA, LLC
19
Chapter 7
QUALIFIED TRAINER/OPERATOR CERTIFICATION FORM
I. Microdot Xtra Meter - Locate the following: a. Battery b. Serial Number c. Toll Free Customer Service Number d. Eject Button
II. Identify and explain the following: a. Blood Test Procedure b. Control Solution Procedure c. Troubleshooting
III. Explain proper procedure for: a. Dating of Microdot Xtra Test Strips and Control Solution b. When to perform a Control Test c. Handling Control Solution results that are out of range
IV. Perform and explain the following procedures: a. Control Solution testing b. Blood testing c. Recalling test results d. Changing the battery
If all tasks were not properly completed, have the operator repeat the procedure correctly. When all tasks have been properly completed, sign bottom of this form and fill out the proper Qualified Trainer or In-service form.
Cambridge Sensors USA Sales Rep.
Qualified Trainer
Operator Date
Date
Date
18
Cambridge Sensors USA, LLC
Notes:
The control solution test is similar to a blood test except that you use
Microdot® Control Solution instead of a drop of blood. The control solution ranges printed on the test strip vial are for Microdot Control
Solutions only. It is used to check the meter and test strip performance.
It is NOT a recommended range for your blood glucose level. When performing a control solution test it does not matter which solution you use first, the Low or High.
Chapter 3
CONTROL SOLUTION TESTING
Control Solution Test Procedure:
1. Shake Low Control Solution bottle
well before using.
2. Remove cap and discard the first drop
of Control Solution and wipe off the
dispenser tip to ensure a good sample
and an accurate result.
3. Insert a test strip into the Microdot Xtra
meter. Be sure the black contact bars
go into the meter. Push the strip in
firmly. Be sure it can go no further.
3.
4. Invert bottle and squeeze out one
drop of control solution. Apply
the drop to the strip by bringing the
meter and the strip to the drop.
Touch the drop with the top edge
of the test strip and wait until the
test pad fills with the solution. Results
appear in 10 seconds.
4.
Cambridge Sensors USA, LLC
9
Chapter 3
CONTROL SOLUTION TESTING
5. Compare the results with the ranges of
expected results shown on the test strip
vial. (Low = Blue Cap, High = Red Cap)
6. You should obtain results within the
expected range printed on the test
strip vial. If this is not so, repeat the test.
If the results are out of the range printed on the test strip vial, check the following:
Was the vial at room temperature?
Did you shake the bottle of control
solution before using?
Has the control solution expiration or discard date expired?
Is the meter malfunctioning?
7. Repeat steps 1-6 for the High
Control Solution procedure.
If the test result is still out of range, call your Microdot® Customer
Service Representative at:
Toll Free (877) 374-4062
DO NOT test blood until you obtain control results within the expected ranges.
Date
Chapter 7
QUALIFIED TRAINERS FORM
Health Care Professional’s Name Title
10
Cambridge Sensors USA, LLC
Cambridge Sensors Representative
Cambridge Sensors Representative Signature
Date
Cambridge Sensors USA, LLC
17
Chapter 7
IN-SERVICE FORM
The Health Care professionals listed below are trained and have demonstrated proficiency using the Microdot Xtra Blood Glucose Monitoring System.
Cambridge Sensors Representative/Qualified Trainer
Date Health Care Professional’s Name
Chapter 4
TESTING WITH PATIENTS
This section describes the procedure to test patient blood samples using the
Microdot® Xtra Blood Glucose System.
Operating Guidelines
Before attempting to test with real blood, make sure you have performed control solution
tests correctly to ensure the meter and test strips are performing properly and to verify technique.
Use the test strips before their expiration date and within three months after opening.
Do not use test strips that are wet, bent, scratched or damaged. Use each test strip immediately
after removing it from the vial.
Patient Test Procedure:
What You Do
1. Press the C button
to turn the meter on
and/or insert test strip
by pushing the two
contact bars in firmly.
2. Obtain a drop
of blood.
Notes
To verify all display symbols are working, all symbols should appear at the same time. If any of the symbols are missing or do not display completely, contact
Cambridge Sensors USA at
(877) 374-4062
Using a safety lancet, lance the side of the finger to obtain a rounded blood sample.
Avoid squeezing the puncture site excessively.
The chart is designed for use during the initial training on the Microdot Xtra Blood Glucose System by a
Cambridge Sensors Representative, In-Service Video or Qualified Trainer.
16
Cambridge Sensors USA, LLC Cambridge Sensors USA, LLC
11
Chapter 4
TESTING WITH PATIENTS
What You Do
3. Apply sample.
4. Results in
10 seconds.
Notes
The blood drop symbol will flash alternately to indicate the meter has detected that a test strip has been inserted and is ready for the blood sample to be added.
Apply the drop of blood directly to the edge of the test strip, and allow the sample to wick automatically into the test zone.
If you do not apply a blood sample within one minute, the meter will turn itself off.
Either reinsert the test strip or press the
M button to turn the meter back on.
After an adequate blood sample has been applied to the test strip, the Microdot®
Xtra Meter will display three running dashes, which indicates the meter is performing the test. After 10 seconds, the test result will display.
5. Dispose of
lancet.
Dispose of used lancet in an approved sharps container.
12
6. Dispose of
test strip.
Simply press the eject button on the meter to dispose of used test strip.
Note: If the blood test result is lower than 20 mg/dL,
“Lo” will appear on the meter display.
This indicates severe hypoglycemia
(low blood glucose). You should immediately treat your hypoglycemia as recommended by your physician’s protocol.
If the blood test result is higher than 525 mg/dL,
“Hi” will appear on the meter display.
This indicates severe hyperglycemia
(high blood glucose). You should immediately treat your hyperglycemia as recommended by your physicians protocol.
Cambridge Sensors USA, LLC
The forms in this chapter are Master Copies .
Please make photocopies for distribution.
If you need replacement documents, please contact Customer Service at (877) 374-4062
Chapter 7
LOG SHEETS
Cambridge Sensors USA, LLC
15
Chapter 6
IN-SERVICE TRAINING OUTCOMES
Once your Microdot® Xtra Meter in-service is complete, Health Care Professionals should be able to:
1. Locate and explain the following
components of the Microdot Xtra Meter:
Eject Button
Test Strip Port
Battery Compartment
Serial Number
Customer Service Number
2. Locate and explain the label
information on the Microdot Xtra Test
Strip Vial:
Discard Date
Lot Number
Expiration Date
Control Ranges
Storage Temperature Range
3. Set Time and Date of meter
4. Identify three parts of the
Microdot Xtra Test Strips
Top Edge
Test Pad
Contact Bars
5. Identify the following parts of the
Microdot Control Solution Bottles:
Low vs. High
Discard Date
Lot Number
Expiration Date
Storage Temperature Range
6. Properly insert the Microdot Xtra Test Strip
7. Perform control solution tests
8. Document and maintain Microdot Xtra
Quality Control Records
9. Obtain a blood sample
10. Perform a blood test
11. Identify meter result range
12. Access meter memory
13. Change the battery
14. Identify and resolve error readings
14
Cambridge Sensors USA, LLC
Sample Size
Total Test Time
Sample Application
Hematocrit Range
Measurement Range
Interference
Strip Removal
Coding
Accuracy of Strip/Meter System
Precision of Strip
Shelf Life of Meter
Operating Humidity range
Operating Temperature
Type of Glucose Result
Strip Vial Packing Size
Typical Control Solution Ranges
when used with Strips/Meter
Strip Shelf Life after Opening
Strip Vial
Altitude
Module Size
Weight
Battery Life
Glucose Units
Memory
Power Source
Automatic Shut-off
Warranty
Chapter 5
0.6 Microliters (600 Nanoliters)
10 Seconds
Strip is placed in the meter and sample wicks into strip
30 - 50%
20 to 525 mg/dL
No interference from over 20 interfering substances
Strip ejection by push button
Auto-code
At least +/- 20% relative to YSI in clinical trials
Correction Coefficient of Regression > 0.969
The strip variation is not greater than 6.4%
Approx. 5 years
10 - 90%; storage 90% max (non-condensing)
50˚ - 104˚F (10˚ - 40˚C)
Plasma equivalent
50 strips per vial
Printed on vial label
3 months
Plastic with desiccant sleeve vial
To 10,000 ft (Target)
65 mm x 85 mm x 16 mm
Approx. 60 grams / 2 ounces
1,500 tests or about 1 year at three tests per day mg/dL or mmol/L
Total of 500 glucose results and insulin-input data
One replaceable 3V lithium battery
(CR2032 or equivalent)
One minute after last user action
1 Year Manufacturer Warranty
Cambridge Sensors USA, LLC
13
Chapter 6
IN-SERVICE TRAINING OUTCOMES
Once your Microdot® Xtra Meter in-service is complete, Health Care Professionals should be able to:
1. Locate and explain the following
components of the Microdot Xtra Meter:
Eject Button
Test Strip Port
Battery Compartment
Serial Number
Customer Service Number
2. Locate and explain the label
information on the Microdot Xtra Test
Strip Vial:
Discard Date
Lot Number
Expiration Date
Control Ranges
Storage Temperature Range
3. Set Time and Date of meter
4. Identify three parts of the
Microdot Xtra Test Strips
Top Edge
Test Pad
Contact Bars
5. Identify the following parts of the
Microdot Control Solution Bottles:
Low vs. High
Discard Date
Lot Number
Expiration Date
Storage Temperature Range
6. Properly insert the Microdot Xtra Test Strip
7. Perform control solution tests
8. Document and maintain Microdot Xtra
Quality Control Records
9. Obtain a blood sample
10. Perform a blood test
11. Identify meter result range
12. Access meter memory
13. Change the battery
14. Identify and resolve error readings
14
Cambridge Sensors USA, LLC
Sample Size
Total Test Time
Sample Application
Hematocrit Range
Measurement Range
Interference
Strip Removal
Coding
Accuracy of Strip/Meter System
Precision of Strip
Shelf Life of Meter
Operating Humidity range
Operating Temperature
Type of Glucose Result
Strip Vial Packing Size
Typical Control Solution Ranges
when used with Strips/Meter
Strip Shelf Life after Opening
Strip Vial
Altitude
Module Size
Weight
Battery Life
Glucose Units
Memory
Power Source
Automatic Shut-off
Warranty
Chapter 5
0.6 Microliters (600 Nanoliters)
10 Seconds
Strip is placed in the meter and sample wicks into strip
30 - 50%
20 to 525 mg/dL
No interference from over 20 interfering substances
Strip ejection by push button
Auto-code
At least +/- 20% relative to YSI in clinical trials
Correction Coefficient of Regression > 0.969
The strip variation is not greater than 6.4%
Approx. 5 years
10 - 90%; storage 90% max (non-condensing)
50˚ - 104˚F (10˚ - 40˚C)
Plasma equivalent
50 strips per vial
Printed on vial label
3 months
Plastic with desiccant sleeve vial
To 10,000 ft (Target)
65 mm x 85 mm x 16 mm
Approx. 60 grams / 2 ounces
1,500 tests or about 1 year at three tests per day mg/dL or mmol/L
Total of 500 glucose results and insulin-input data
One replaceable 3V lithium battery
(CR2032 or equivalent)
One minute after last user action
1 Year Manufacturer Warranty
Cambridge Sensors USA, LLC
13
Chapter 4
TESTING WITH PATIENTS
What You Do
3. Apply sample.
4. Results in
10 seconds.
Notes
The blood drop symbol will flash alternately to indicate the meter has detected that a test strip has been inserted and is ready for the blood sample to be added.
Apply the drop of blood directly to the edge of the test strip, and allow the sample to wick automatically into the test zone.
If you do not apply a blood sample within one minute, the meter will turn itself off.
Either reinsert the test strip or press the
M button to turn the meter back on.
After an adequate blood sample has been applied to the test strip, the Microdot®
Xtra Meter will display three running dashes, which indicates the meter is performing the test. After 10 seconds, the test result will display.
5. Dispose of
lancet.
Dispose of used lancet in an approved sharps container.
12
6. Dispose of
test strip.
Simply press the eject button on the meter to dispose of used test strip.
Note: If the blood test result is lower than 20 mg/dL,
“Lo” will appear on the meter display.
This indicates severe hypoglycemia
(low blood glucose). You should immediately treat your hypoglycemia as recommended by your physician’s protocol.
If the blood test result is higher than 525 mg/dL,
“Hi” will appear on the meter display.
This indicates severe hyperglycemia
(high blood glucose). You should immediately treat your hyperglycemia as recommended by your physicians protocol.
Cambridge Sensors USA, LLC
The forms in this chapter are Master Copies .
Please make photocopies for distribution.
If you need replacement documents, please contact Customer Service at (877) 374-4062
Chapter 7
LOG SHEETS
Cambridge Sensors USA, LLC
15
Chapter 7
IN-SERVICE FORM
The Health Care professionals listed below are trained and have demonstrated proficiency using the Microdot Xtra Blood Glucose Monitoring System.
Cambridge Sensors Representative/Qualified Trainer
Date Health Care Professional’s Name
Chapter 4
TESTING WITH PATIENTS
This section describes the procedure to test patient blood samples using the
Microdot® Xtra Blood Glucose System.
Operating Guidelines
Before attempting to test with real blood, make sure you have performed control solution
tests correctly to ensure the meter and test strips are performing properly and to verify technique.
Use the test strips before their expiration date and within three months after opening.
Do not use test strips that are wet, bent, scratched or damaged. Use each test strip immediately
after removing it from the vial.
Patient Test Procedure:
What You Do
1. Press the C button
to turn the meter on
and/or insert test strip
by pushing the two
contact bars in firmly.
2. Obtain a drop
of blood.
Notes
To verify all display symbols are working, all symbols should appear at the same time. If any of the symbols are missing or do not display completely, contact
Cambridge Sensors USA at
(877) 374-4062
Using a safety lancet, lance the side of the finger to obtain a rounded blood sample.
Avoid squeezing the puncture site excessively.
The chart is designed for use during the initial training on the Microdot Xtra Blood Glucose System by a
Cambridge Sensors Representative, In-Service Video or Qualified Trainer.
16
Cambridge Sensors USA, LLC Cambridge Sensors USA, LLC
11
Chapter 3
CONTROL SOLUTION TESTING
5. Compare the results with the ranges of
expected results shown on the test strip
vial. (Low = Blue Cap, High = Red Cap)
6. You should obtain results within the
expected range printed on the test
strip vial. If this is not so, repeat the test.
If the results are out of the range printed on the test strip vial, check the following:
Was the vial at room temperature?
Did you shake the bottle of control
solution before using?
Has the control solution expiration or discard date expired?
Is the meter malfunctioning?
7. Repeat steps 1-6 for the High
Control Solution procedure.
If the test result is still out of range, call your Microdot® Customer
Service Representative at:
Toll Free (877) 374-4062
DO NOT test blood until you obtain control results within the expected ranges.
Date
Chapter 7
QUALIFIED TRAINERS FORM
Health Care Professional’s Name Title
10
Cambridge Sensors USA, LLC
Cambridge Sensors Representative
Cambridge Sensors Representative Signature
Date
Cambridge Sensors USA, LLC
17
Chapter 7
QUALIFIED TRAINER/OPERATOR CERTIFICATION FORM
I. Microdot Xtra Meter - Locate the following: a. Battery b. Serial Number c. Toll Free Customer Service Number d. Eject Button
II. Identify and explain the following: a. Blood Test Procedure b. Control Solution Procedure c. Troubleshooting
III. Explain proper procedure for: a. Dating of Microdot Xtra Test Strips and Control Solution b. When to perform a Control Test c. Handling Control Solution results that are out of range
IV. Perform and explain the following procedures: a. Control Solution testing b. Blood testing c. Recalling test results d. Changing the battery
If all tasks were not properly completed, have the operator repeat the procedure correctly. When all tasks have been properly completed, sign bottom of this form and fill out the proper Qualified Trainer or In-service form.
Cambridge Sensors USA Sales Rep.
Qualified Trainer
Operator Date
Date
Date
18
Cambridge Sensors USA, LLC
Notes:
The control solution test is similar to a blood test except that you use
Microdot® Control Solution instead of a drop of blood. The control solution ranges printed on the test strip vial are for Microdot Control
Solutions only. It is used to check the meter and test strip performance.
It is NOT a recommended range for your blood glucose level. When performing a control solution test it does not matter which solution you use first, the Low or High.
Chapter 3
CONTROL SOLUTION TESTING
Control Solution Test Procedure:
1. Shake Low Control Solution bottle
well before using.
2. Remove cap and discard the first drop
of Control Solution and wipe off the
dispenser tip to ensure a good sample
and an accurate result.
3. Insert a test strip into the Microdot Xtra
meter. Be sure the black contact bars
go into the meter. Push the strip in
firmly. Be sure it can go no further.
3.
4. Invert bottle and squeeze out one
drop of control solution. Apply
the drop to the strip by bringing the
meter and the strip to the drop.
Touch the drop with the top edge
of the test strip and wait until the
test pad fills with the solution. Results
appear in 10 seconds.
4.
Cambridge Sensors USA, LLC
9
Chapter 3
CONTROL SOLUTION TESTING
This chapter describes the necessary steps to test with the Microdot® Control
Solutions in order to validate the performance of the Microdot Xtra Meter and
Microdot Xtra Strips.
Operating Guidelines: When Should You Conduct a
Control Solution Test?
Any time you open a new vial of test
strips.
Whenever you think the system
is not working properly.
If the blood glucose test results
differ from the resident’s symptoms
or non-symptoms.
If you believe the results are not
accurate.
If you drop the meter.
If the vial of test strips has been
left open for an extended period
of time.
It is critical to follow the Operating
Guidelines, on the right, to obtain
accurate results while using the
Microdot Xtra Glucose System.
Use only Microdot Control Solutions
(High and Low).
Check the expiration date on the
control solution vial. Do not use
if expired or if the discard date
has passed.
Control solution, meter and test
strips should come to room
temperature before testing
(66-77˚F / 20-25˚C).
Use solution for three months
after first opening. Record the
discard date (opening date
plus three months) on the
control solution vial. Discard
after three months.
Close tightly and store the control
solution at temperatures between
50-86˚F (10-30˚C).
Do not refrigerate. Do not freeze.
8
Cambridge Sensors USA, LLC
Chapter 7
® Xtra BLOOD GLUCOSE MONITORING SYSTEM
QUALITY CONTROL RECORD
Microdot Xtra Meter Serial Number
Month/Year
Date
Station/
Initials
Test Strip
Lot #
Low
Control
Range
(mg./dl.)
Low
Control
Result
(mg./dl.)
High
Control
Range
(mg./dl.)
High
Control
Result
(mg./dl.)
Notes:
Cambridge Sensors USA, LLC
19
Chapter 8
MATERIAL SAFETY DATA SHEET
Xtra Blood Glucose Test Strips
Section I - Identification
Trade Name:
Common Name:
Supplier:
Emergency Phone Number:
Microdot Xtra Blood Glucose Test Strips
Blood glucose test strips for use with Microdot
Xtra Blood Glucose Test System
Cambridge Sensors USA
10051 Bode Rd.
Palinfield, IL 60585
(877) 374-4062
Section II - Composition
This product does not present a physical or health hazard under reasonable use or under emergency situations involving a release of only this product. This material is therefore not considered to be a “Hazardous Chemical” as defined by the Federal Occupational Safety and
Health Administration in the Hazard Communication Standard (29 CFR 1910.1200) or the equivalent standards generated by state agencies.
ACCORDINGLY NO MATERIAL SAFETY DATA SHEET IS REQUIRED FOR THIS PRODUCT.
MSDSs that represent non-hazardous chemicals are not covered by the HCS. Paragraph 29
CFR 1910.1200 (g) (8) of the standard requires that “the employer shall maintain in the workplace copies of the required MSDSs for each hazardous chemical, and shall ensure that they are readily accessible during each work shift to employees when that are in their work area(s)”.
OSHA does not require or encourage employers to maintain MSDSs for non-hazardous chemicals. Consequently, an employer is free to discard MSDSs for non-hazardous chemicals.
20
Cambridge Sensors USA, LLC
Chapter 2
Using Meter Memory
Your Microdot Xtra Meter stores the 500 most recent blood glucose and control solution test results and insulin data with date and time in the memory. It also provides you with 14-day averages of your blood glucose test results.
You can review the test results in memory with these easy steps.
1.
Step 1
Enter the Memory Mode.
To Enter Memory Mode, turn on the meter by pressing the M button.
The meter will display the last result with
Mem. symbol, time and date.
Step 2
Recall Test Results.
Previous results can be displayed by pressing the C button. As long as the C button is held, the meter will scroll through the memory displaying the result and the memory location. Once the button is released, the meter will display the selected result with its Time and Date. When the memory is full (500 results stored), the oldest result is dropped as the newest is added.
NOTE: When using the meter for the first time
“Mem.---” will appear, showing that there are no test results stored in memory.
2.
Cambridge Sensors USA, LLC
7
Chapter 2
Setting the Beeper
After setting the Time and Date, the Sound symbol will appear and can now be changed by pressing the C button. To accept the setting, press the M button.
This option is used to switch on or off the beeper. When turned ON, a sound will be heard when blood or control solution is applied to the strip and when the test is finished. The sound will be heard when an error has occurred or if an alarm is triggered.
When the M button is pressed to accept the beeper option, the meter will display END and switch off.
6
Cambridge Sensors USA, LLC
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section I - Identification
Trade Name:
Common Name:
Supplier:
Emergency Phone Number:
Microdot Control Solutions
Control Solutions for use with Microdot Xtra
Blood Glucose Test System
Cambridge Sensors USA
10051 Bode Rd.
Palinfield, IL 60585
(877) 374-4062
Section II - Composition / Information on Ingredients
Chemical Family:
Chemical Name:
Salt solution
N/A
Section III - Hazards Identification
Eye Contact:
Skin Contact:
Ingestion:
May cause irritation if in contact with eye.
Skin contact may cause irritation.
Though not a likely route of occupational exposure, ingestion of the product may cause choking, vomiting or nausea.
N/A
N/A
Inhalation:
Chronic Exposure
Section IV - First Aid Measures
Eye Contact:
Skin Contact:
Ingestion:
Inhalation:
Chronic Exposure:
Although no adverse health reactions are expected from the normal use of this product, it is recommended to flush eyes with water and seek medical advise whenever there is a potential injury to the eye.
If contact with this product leads to reddening, inflammation or irritation, flush the exposed area with running water. If irritation persists, get medical attention.
As a precaution, get medical attention if there has been ingestion of this product.
If breathing becomes difficult, remove victim to fresh air and get medical attention.
N/A
Cambridge Sensors USA, LLC
21
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section V - Fire Fighting Measures
Auto-flammability
Flash Point (test method)
Extinguishing Media
Special Fire Fighting
Procedures
Fire and Explosion Hazards
Hazardous Combustion
Products
Upper Explosion Limit (%)
Lower Explosion Limit (%)
Not determined
Not determined
Use fire extinguishing media appropriate for site collections.
Structural fire fighting gear and self-contained breathing apparatus will provide adequate protection if this product is in a fire area.
Not determined
Thermal decomposition may emit carbon monoxide and carbon dioxide.
Not determined
Not determined
Section VI - Accidental Release Measures
Spill and Leak Procedures Use an absorbent material to contain/pick up the spilled solution. Place all contaminated disposals into a suitable container, seal, label and hold for disposal.
Section VII - Handling and Storage
22
Storage Temperature
Handling/Storage
Ventilation Requirements
Sensitivity to Static Electricity
Sensitivity to Mechanical Impact
Store vials as directed in the package insert.
Handle and store vials as directed by the package insert.
No special requirements.
Not known
Not known
Section VIII - Exposure Controls / Personal Protection
Respiratory Protection
Ventilation
Protective Gloves
Other Protective Equipment
Other Engineering Controls
Work Practices
Hygienic Practices
Cambridge Sensors USA, LLC
Not required under normal use of this product.
Not required under normal use of this product.
Wear appropriate gloves to prevent skin contact. Replace torn or punctured gloves promptly.
Wear appropriate eye protection to prevent eye contact.
Wear appropriate body protection to prevent skin contact.
Eye wash stations and deluge showers.
Good laboratory technique should be used when handling this product. Observe appropriate chemical hygiene.
Do not place in mouth.
Do not eat, drink or smoke while working with this product.
Upon completion of work activities involving this product, wash hands thoroughly with soap and water.
Chapter 2
Before using the Microdot Xtra Meter for the first time, you should set the actual Time,
Date and Year. The unit of measurement is preset and can not be changed.
Setting the Time, Date and Year
Enter the Set Mode.
To Enter the set mode, turn the meter on by pressing the C button. After the segment test, the time and date will start to flash.
Step 1
Set the Time and Date Format.
Pressing the C button again will now toggle between US and International Time and Date formats.
For US: 12h Time format, mm-dd (begins with
AM setting). To accept the desired setting, press the M button.
Step 2
Set the Hour.
The Hour will start to flash. It can now be changed by pressing the C button. The US setting will begin at 12:01 AM and will move to PM. To accept the correct setting, press the M button.
Step 3
Set the Minutes.
The Minute will start to flash, it can now be changed by pressing the C button. To accept the correct setting, press the M button.
Repeat Steps 1 & 2 to set Month and Year.
1.
2.
3.
Cambridge Sensors USA, LLC
5
Chapter 1
SYSTEM COMPONENTS
The Microdot Xtra Blood Glucose Test Strip offers the latest advances in biosensor, auto-code technology. Blood is applied to the top edge of the Microdot Xtra Test
Strip and is automatically drawn into the white channel where the reaction takes place. Only 600 nanoliters (0.6 microliters) is needed.
The Microdot Xtra Test Strip consists of the following parts:
Top Edge: Apply a drop of blood here, where the white channel meets the top of the edge strip.
White Channel: This is where you check if enough blood has been applied to the top edge.
Contact Bars: Insert this end of the test strip into the meter.
Push firmly until the strip can go no further.
4
Cambridge Sensors USA, LLC
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section IX - Physical and Chemical Properties
Pure Substance or preparation
Appearance/Odor
Odor Threshold
Melting/Freezing
Partition
Evaporation Rate
Vapor Pressure (mmHg)
Vapor Density (air =1)
Viscosity
Volatiles
Volatile Organic Compounds
Auto-flammability
Flash Point
Oxidizing Properties
Stability
Materials to Avoid
Hazardous Decomposition
Products
Preparation Physical Form Liquid
Blue, odorless pH AS is neutral
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Not determined
Stable
Strong bases, strong acids and water reactive materials
Thermal decomposition may emit carbon monoxide and carbon dioxide.
Section X - Toxicological Information
Route of Entry
Effects of Chronic Exposure
Effects of Acute Exposure
Special Health Effects
Target Organs
Section XI - Ecological Information
Potential Effect on Environment
Potential to Bioaccumulate
Mobility
Ecotoxicity
Persistence and Degradeability
Aqua Toxicity
Section XII - Disposal Considerations
Ingestion, skin and/or eye contact
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Water Disposal
Empty Container Warnings
Please consult local, state and federal regulations for additional guidance on disposal.
Not known
Cambridge Sensors USA, LLC 23
Chapter 8
MATERIAL SAFETY DATA SHEET
Microdot® Control Solutions
Section XII - Transport Information (See also Section IX)
ADR / RID
CEFIC Tremcard
Hazchem Code
Kemmler Code
IMDG Classification
IATA Classification
Marine Pollutant
UN Number
UN Class
UN Packing Group
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Not known
Section XIV - Regulatory Information
EEC Hazard Classification
Risk Phrases
Safety Phrases
Section XV - Other Information
Not known
Not known
Not known
Directives 88/379/EEC and 91/155/EEC have been considered when compiling this MSDS; the information is provided for health and safety assessment by an industrial user. Reference should be made to any relevant local or national health, safety or environmental legislation.
This information does not constitute indication of suitability for specific uses. The information, data, and recommendations contained herein are based upon information believed by Cambridge
Sensors USA, after reasonable investigation and research, to be accurate. However, Cambridge
Sensors USA does not warrant the accuracy of this information. All materials and mixtures may present unknown hazards and should be used with caution. When necessary or appropriate, independent opinions regarding the risk of handling or exposure should be obtained from trained professionals.
Cambridge Sensors USA disclaims any warranty against patent infringement and the implied warranties of merchantability and fitness for a particular purpose. Customer’s sole and exclusive remedy shall be replacement of the product or return of the product and refund of the purchase price, at Cambridge
Sensors USA option. In no case can Cambridge Sensors USA be liable for incidental or consequential damages, including lost profits.
Chapter 1
SYSTEM COMPONENTS
LCD Screen: Shows blood glucose result and symbols that guide you through the test.
Up/Down Toggle Buttons: Scrolls the memory, sets time and date.
Eject Button: Releases test strip.
Test Strip Port: Insertion site for test strip.
Serial Number: Located in the upper left hand corner of the meter. Required when calling Customer
Service for troubleshooting or meter replacement.
Battery Compartment: Holds one 3V Lithium battery (CR2032).
SN 00023463
Customer Service:
888-425-1149
24
Cambridge Sensors USA, LLC
Customer Service Number: Located on the back of the meter.
Call this number if you have any questions or
problems with the Microdot Xtra Glucose Meter.
(877) 374-4062
Cambridge Sensors USA, LLC
3
Chapter 1
® Xtra BLOOD GLUCOSE METER OVERVIEW
This section provides general information on the Microdot Xtra Blood
Glucose Monitoring System
The Microdot Xtra Blood Glucose Monitoring
System is a portable whole blood testing system which performs tests with quick and accurate results.
It consists of three main parts:
1. Microdot Xtra Blood Glucose Meter
2. Microdot Xtra Test Strips
3. Microdot High and Low Control
Solution
These products have been designed, tested and proven to work together as a system to produce safe and accurate results.
Use only Microdot Xtra Test Strips and Control Solutions with the
Microdot Xtra Glucose Meter.
Your system includes:
Microdot Xtra Meter
Anti-microbial Meter Cover
Quality Assurance Manual
Warranty Registration Card
3V Lithium Battery (installed)
Additional components include:
Microdot Xtra Test Strips
Microdot Control Solutions
2
Cambridge Sensors USA, LLC
Chapter 9
The blood glucose level is higher than
525 mg/dL.
The blood glucose level is lower than
20 mg/dL.
Error message that indicates that there is a problem with the meter, e.g., measurement error (time out, overflow, offset) or temperature out of range.
Error message that indicates that there may be a problem with the test strip, e.g, the test strip may be damaged, moved, or removed during testing, or inserted improperly.
This message indicates very high blood sugar. You should recheck the blood glucose level. If “Hi” again, follow facility protocol.
This message indicates very low blood sugar. You should recheck the blood glucose level. If “Lo” again, follow facility protocol.
Review the instructions and try again with a new test strip. If the problem persists, contact Customer Service.
Check the test strip for damage and retest as necessary. Repeat the test.
If the error message appears again, contact Customer Service.
Error message could be caused by a used or damaged test strip.
Repeat the test with a new strip.
If the error message appears again, contact Customer Service.
Error message indicates serial communications error.
If error persists, contact Customer
Service.
The battery sign appears on the display with the unit of measurement. The power of the battery is getting low. You can complete about 50 more tests from the time this symbol first appears.
Test results will be accurate but replace the battery as soon as possible. Battery type is
CR2032 3 volt.
Cambridge Sensors USA, LLC
25
Chapter 10
®
Xtra
Cleaning Your Meter
First, remove the anti-microbial meter cover. Second, use a lint free cloth dampened with soapy water or isopropyl (70% - 80%) to clean the outside of the blood glucose meter.
Microdot Meter Disinfecting Guideline: To disinfect the meter, dilute 1 mL of household bleach (5% - 6% sodium hypochlorite solution) in 9 mL of water to achieve a 1:10 dilution (final concentration of 0.5% - 0.6% sodium hypochlorite). The solution can then be used to dampen a paper towel (do not saturate the towel). Then use the damped towel to thoroughly wipe down the meter.
Please note that there are commercially available 1:10 bleach wipes from a variety of manufactures.
Germicidal wipes are readily available through major medical distributors. To use these products, remove a wipe from container and follow product label instructions to disinfect meter. Take extreme care not to get liquid in the test strip dock or key code parts of the meter.
Cambridge Sensors USA, LLC
Customer Service available from from 8am - 5pm CST Monday - Friday
Toll free 1.877.374.4062 | www.microdotcs.com
26
Cambridge Sensors USA, LLC
Contents
TABLE OF CONTENTS
Chapter
1. Microdot® Xtra Meter Overview & System Components
2. Initial Setup of the Microdot Xtra Meter
3. Control Solution Testing
4. Testing with Patients
5. Microdot Xtra Meter Specifications
6. In-Service Training Outcomes
7. Microdot Xtra Blood Glucose Meter Log Sheets
7A. Microdot Xtra BGMS In-Service Form
7B. Microdot Xtra BGMS Qualified Trainers Form
7C. Microdot Xtra BGMS Qualified Trainer/Operator
Certification Form
7D. Microdot Xtra BGMS Quality Control Record
8. Material Safety Data Sheets
8A. Microdot Xtra Test Strips
8B. Microdot Control Solutions
9. Troubleshooting the Microdot Xtra Meter
10. Manufacturer Cleaning Instructions
Notes
Page
15
16
17
18
19
2 - 4
5 - 7
13
14
8 - 10
11 - 12
20
21 - 24
25
26
27 - 28
Cambridge Sensors USA, LLC
1
Notes
27
Cambridge Sensors USA, LLC
Notes
Cambridge Sensors USA, LLC
28
Notes
27
Cambridge Sensors USA, LLC
Notes
Cambridge Sensors USA, LLC
28
CAMBRIDGE SENSORS USA, LLC
Customer Service: toll free: 877.374.4062 Monday–Friday 9am–5pm CST www.microdotcs.com
FAST. ACCURATE. EASY TO USE.
Xtra
OPERATIONS & QUALITY ASSURANCE
PROCEDURE MANUAL
For
Healthcare
Professionals
Download
Advertisement
Key features
Fast, accurate, and easy to use system
Portable whole blood testing system
Requires only 0.6 microliters (600 nanoliters) sample size
Provides results in 10 seconds
Features auto-code technology
Stores 500 most recent blood glucose and control solution test results
Includes a strip ejection push button
Operates within 50° - 104°F (10° - 40°C) temperature range
One-year manufacturer warranty
Frequently asked questions
You should conduct a control solution test any time you open a new vial of test strips, whenever you suspect the system is not working properly, if blood glucose results differ from a resident's symptoms, if you believe results are inaccurate, if the meter is dropped, or if the test strip vial has been left open for an extended period.
First, repeat the test. If the result is still out of range, check if the vial was at room temperature, if the control solution bottle was shaken well, and if the control solution's expiration or discard date has passed. Do not test blood until control results are within the expected ranges. If issues persist, contact Customer Service.
'Lo' indicates a blood glucose level lower than 20 mg/dL (severe hypoglycemia). 'Hi' indicates a blood glucose level higher than 525 mg/dL (severe hyperglycemia). In both cases, you should recheck the level and follow your facility's protocol for treating severe low or high blood sugar.
Used lancets should be disposed of in an approved sharps container. Used test strips can be disposed of by simply pressing the eject button on the meter.
First, remove the anti-microbial meter cover. Clean the outside with a lint-free cloth dampened with soapy water or isopropyl (70%-80%). To disinfect, use a 1:10 dilution of household bleach (0.5%-0.6% sodium hypochlorite) to dampen a paper towel, then wipe the meter thoroughly, or use commercially available germicidal wipes. Be careful not to get liquid into the test strip dock or key code parts.