Cambridge Sensors MicroDot Xtra Data Sheets

Below you will find brief information for Xtra Blood Glucose Monitoring System. This document serves as an Operations & Quality Assurance Procedure Manual, guiding healthcare professionals through the initial setup, control solution testing, and patient testing procedures. It also covers meter specifications, in-service training outcomes, material safety data sheets, and troubleshooting tips, ensuring proper and accurate use of the device.

PDF
Document
Microdot Xtra Manual | AI Chat & PDF Download | Manualzz

Advertisement

Advertisement

/

Advertisement

Key features

  • Fast, accurate, and easy to use system

  • Portable whole blood testing system

  • Requires only 0.6 microliters (600 nanoliters) sample size

  • Provides results in 10 seconds

  • Features auto-code technology

  • Stores 500 most recent blood glucose and control solution test results

  • Includes a strip ejection push button

  • Operates within 50° - 104°F (10° - 40°C) temperature range

  • One-year manufacturer warranty

Frequently asked questions

You should conduct a control solution test any time you open a new vial of test strips, whenever you suspect the system is not working properly, if blood glucose results differ from a resident's symptoms, if you believe results are inaccurate, if the meter is dropped, or if the test strip vial has been left open for an extended period.

First, repeat the test. If the result is still out of range, check if the vial was at room temperature, if the control solution bottle was shaken well, and if the control solution's expiration or discard date has passed. Do not test blood until control results are within the expected ranges. If issues persist, contact Customer Service.

'Lo' indicates a blood glucose level lower than 20 mg/dL (severe hypoglycemia). 'Hi' indicates a blood glucose level higher than 525 mg/dL (severe hyperglycemia). In both cases, you should recheck the level and follow your facility's protocol for treating severe low or high blood sugar.

Used lancets should be disposed of in an approved sharps container. Used test strips can be disposed of by simply pressing the eject button on the meter.

First, remove the anti-microbial meter cover. Clean the outside with a lint-free cloth dampened with soapy water or isopropyl (70%-80%). To disinfect, use a 1:10 dilution of household bleach (0.5%-0.6% sodium hypochlorite) to dampen a paper towel, then wipe the meter thoroughly, or use commercially available germicidal wipes. Be careful not to get liquid into the test strip dock or key code parts.