advertisement
Catalog Number HQ40d18
HQ Series Portable Meters
USER MANUAL
March 2006, Edition 3
© Hach Company, 2006. All rights reserved.
Visit us at www.hach.com
Table of Contents
3
Table of Contents
4
Table of Contents
5
Visit us at www.hach.com
Section 1 General Information
1.1 Safety Information
Please read this entire manual before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
1.1.1 Use of Hazard Information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Important Note: Information that requires special emphasis.
Note: Information that supplements points in the main text.
1.1.2 Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment and manufacturer supplied accessories marked with this symbol may not be disposed of in
European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This symbol, if noted on the product, indicates the need for protective eye wear.
7
General Information
1.2 Product Overview
1.3 Meter Description
The HQ Series Portable Meters measure various parameters when used with IntelliCAL™ probes such as pH, conductivity, salinity, total dissolved solids (TDS), or dissolved oxygen (using Hach's patented luminescent dissolved oxygen probes, LDO
®
). The meter automatically recognizes the type of probe that is connected to the meter. The probes store the unique serial number, current calibration, and calibration history. When the default settings are used, an operator can take measurements right out of the box.
Data is easily managed by using the settings for operator ID, sample ID, and data storage. Supervisory control can be set by using the access function. Settings for measurement and calibration are stored as methods. The default method for each parameter follows suggested USEPA measurement techniques.
The HQ series meters are available in four models:
•
HQ11d—pH/mV
•
HQ14d—conductivity, salinity, total dissolved solids (TDS)
•
HQ30d—pH, conductivity, salinity, total dissolved solids
(TDS) or dissolved oxygen (LDO), 1 probe connector
•
HQ40d—pH, conductivity, salinity, total dissolved solids
(TDS), dissolved oxygen (LDO), 2 probe connectors,
PC/printer/flash memory stick/keyboard connectivity
Other features:
•
Auto probe recognition including serial number
•
Methods containing parameter settings for regulatory control
•
Supervisory access control
•
Long sensor life, LDO
•
No polarization time, LDO
•
Internal data storage of 500 results
•
Sample ID and Operator ID for data traceability
•
Adjustable automatic shut-off for extended battery life
•
Automatic correction for barometric pressure and temperature, LDO
•
IP67 (waterproof to 1 meter for 30 minutes, excluding battery housing. Battery compartment submersible to 2 feet for 15 seconds)
•
Connectivity to PC/printer/flash memory stick/keyboard
(HQ40d only)
•
Power from four alkaline or Nickel Metal Hydride (NiMH)
AA batteries, or AC adapter
8
Section 2 Specifications
Specifications are subject to change without notice.
Controller Enclosure
Enclosure
Power Requirements (internal)
Power Requirements (external USB/DC
Power Adaptor)
Storage Temperature
Operating Temperature
Operating Humidity
Weight
Inputs
5-pin Custom M-12 for probes
8-Pin Connector for USB and external DC power
PH Probes (standard and rugged)
pH Range
Sodium (Alkalinity) Error
Temperature Range
Temperature Accuracy
Warranty
LDO Probes (standard and rugged)
Dissolved Oxygen Range
Dissolved Oxygen Accuracy
% Saturation
Temperature Range
Temperature Resolution
Temperature Accuracy
Warranty
Conductivity Probe
Conductivity Range
Conductivity Resolution
Conductivity Accuracy
TDS Range
TDS Accuracy
Meter: IP67, waterproof to 1 meter for 30 minutes
Battery Compartment: water resistant to 2 feet for 15 seconds
AA Alkaline or Nickel Metal Hydride (NiMH) Batteries (4)
100–240 V, 50/60 Hz input;
4.5 to 7.5 V (7 VA) output (center contact +, outer shield -)
–20 to +60 °C (–4 to +140 °F)
0 to +60 °C (32 to 140 °F)
90% (non-condensing)
0.75 lb/11.6 oz/330 g
0.95 lb/15.2 oz/430 g (with four AA alkaline batteries installed)
Meters accept IntelliCAL probes
(HQ11d pH/ORP only; HQ14d conductivity only)
The 8-pin connector enables USB and external DC power connectivity
PHC301 (refillable): 0.0–14.0 pH
PHC101 (gel filled): 2.0–14.0 pH
–0.6 pH at pH 12.6 in 1 M NaOH
0.0–80.0 ºC
±0.3 ºC
PHC301 probe is covered by a one-year warranty
PHC101 probe is covered by a six-month warranty
0.1–20.0 mg/L (ppm)
1–200% saturation
±0.1 mg/L for 0–8 mg/L
±0.2 mg/L for greater than 8 mg/L
1.0%
0–50 ºC
0.1 ºC
± 0.3 ºC
Probe is covered by a three-year warranty.
Sensor cap is covered by a one-year warranty.
0.01 µS/cm to 200.0 mS/cm
0.0–19.99 µS/cm: 0.01 µS/cm
20.0–199.9 µS/cm: 0.1 µS/cm
200.0–1999.0 µS/cm: 1.0 µS/cm
2.0–19.99 mS/cm: 0.01 mS/cm
20.0–200.0 mS/cm: 0.1 mS/cm
±0.5% of Reading
0 to 50,000 mg/L as NaCl
±0.5% of Reading
9
Specifications
TDS Resolution
Conductivity Probe (continued)
Salinity Range
Salinity Accuracy
Salinity Resolution
Temperature Range
Temperature Accuracy
Warranty
Outputs
USB
0.0–199.9 mg/L: 0.1 mg/L
200.0
–
1999.0 mg/L: 1.0 mg/L
2.0
–19.99 g/L: 0.01 g/L
20.0–50.0 g/L: 0.1 g/L
0 to 42 ppt (‰)
±0.1 ppt
0.01 ppt
–10.0 to 110.0 ºC
±0.3 ºC
Probe is covered by a one-year warranty.
Peripheral and Host
10
Section 3 Installation
3.1 Unpacking the Instrument
Remove the instrument and accessories from the shipping container and inspect each item for damage. Verify that all items listed on the packing slip are included. If any items are missing or damaged, contact the manufacturer or distributor (outside US).
1
HQd Meter
2
User Manual
3
AA Batteries (4)
Figure 1 HQ11d, HQ14d, HQ30d Instrument Components
4
AC-DC Power Supply (optional)
5
Power Cord (optional)
6
DC Power Adapter (optional)
1
HQ40d Meter
2
AC Power Cord
3
AA Batteries (4)
4
User Manual
Figure 2 HQ40d Instrument Components
5
HQ40d PC Application Software
6
AC-DC Power Supply
7
USB/DC Power Adapter
11
Installation
3.2
AC Power and Batteries
DANGER
Use only alkaline or nickel metal hydride type batteries in the meter. Other battery types might cause a fire or explosion.
DANGER
Make sure that the batteries are installed according to the polarity markings in the meter battery compartment. Failure to correctly install the batteries can result in damage to the meter, fire, or explosion.
DANGER
AC mains outlets in wet or potentially wet locations MUST
ALWAYS be provided with a Ground Fault Circuit Interrupting
(GFCI/GFI) circuit breaker. The AC-DC power adapter provided with this product is not sealed and must not be used on wet benches or in wet locations without GFCI protection.
CAUTION
Never mix battery types in the meter. Use four AA alkaline, or four AA nickel metal hydride batteries.
Important Note: The battery compartment of the meter and the
USB/DC power adapter are not waterproof. Use care when operating these devices on a bench in wet environments. Water may infiltrate these devices and eventually cause performance or quality problems. Periodic inspection of the batteries and battery compartment is recommended, if the meter is used in wet environments: remove, clean, and dry the batteries, the interior of the battery compartment, and the battery contacts; then reinsert the batteries and close the compartment cover.
The meter can be battery powered using four AA batteries (alkaline or nickel metal hydride) or by AC power. Connection to AC power requires additional components (
3.2.1 Battery Power
Important Note: Rechargeable alkaline or nickel metal hydride batteries may also be used in the meter (do not mix battery types).
Batteries are not charged in the meter.
1. Pull the release tab on the battery cover and remove the cover as shown in Figure 3 .
2. Insert four AA batteries (alkaline or nickel metal hydride) following polarity markings inside the battery housing.
3. Replace the battery cover.
12
Installation
Figure 3 Battery Installation
1
AA Alkaline or Nickel Metal Hydride Battery (4)
(Do not mix battery types.)
2
Battery Cover
3
Release Tab
A battery icon appears in the top right corner of the display to indicate current battery status.
When batteries are installed, the meter will automatically shut off after five minutes of sitting idle (this is the default setting). This auto shut-off feature can be changed in the Display Options menu
(see
).
Note: When using nickel metal hydride (NiMH) batteries, the battery icon will not indicate a full charge after freshly charged batteries have been inserted (NiMH batteries are 1.2 V versus 1.5 V for alkaline batteries). Even though the icon does not indicate complete battery charge, if you use 2500 mAH NiMH batteries you will achieve 90% of instrument operation lifetime
(before you need to recharge) versus new alkaline batteries.
Note: NiMH batteries self-discharge during storage. If you do not insert freshly charged NiMH batteries, operational lifetime will be reduced from this 90%.
Note: As batteries age, their output voltage decreases. Whenever battery voltage drops below 4-volts, the meter will shut itself down, to assure no loss of data. Insert fresh batteries and meter functionality will be restored.
13
Installation
3.2.2 AC Power
All meters can be powered by AC power using a power supply, adapter, and cord. The HQ40d meter ships with an AC-DC Power
Supply, a USB/DC Power Adapter, and a Power Cord (see Figure 6 on page 17 ). The USB/DC Power Adaptor allows the meter to transfer data to a computer or flash memory stick (
The HQ30d, HQ11d, and HQ14d meters can be powered by AC power using optional power adapter kits (Cat. No. 58263-00 for 115
V or 58311-00 for 230 V). Both kits include an AC-DC Power
Supply, a DC Power Adapter, and an appropriate power cord
3.3 Turning the Meter On and Off
Note: The meter can be operated in several different languages. When the meter is turned on for the first time, the user must select a language before any other meter functions can be accessed. Additionally, the operator is prompted to enter the correct time and date during initial use, and to verify
correct time and date whenever the batteries are changed. See section 4.3 on page 22 .
Press the power
ON/OFF
key to turn the meter on. If the meter does not turn on, be sure the batteries are installed properly or that the
AC-DC Power Supply is connected properly to an electrical outlet.
Press the power
ON/OFF
key to turn the meter off. When batteries are used, the display backlight will turn off after 1 minute, and the meter will automatically turn off after 5 minutes (default settings).
These features can be changed in the Meter Options>Display
Options>Auto Shut Off/Backlight menus.
3.4 Probe Connection
CAUTION
BEFORE ATTACHING THE PROBE FOR THE FIRST TIME: Set the date and time in the meter before attaching the probe for its first use. If the meter date and time are incorrect when the probe is installed, the probe will retain this incorrect time stamp for the remainder of its service life, even if the meter time and date have subsequently been corrected.
14
Installation
The HQ11d, HQ14d, and the HQ30d support single connection and display of IntelliCAL™ probes (see Figure 4 ).
Figure 4 Connectors on HQ11d, HQ14d, HQ30d Meters
1
USB/DC Power Adapter Port (8-pin)
2
Probe Port (5-pin)
The HQ40d supports dual connection and display of IntelliCAL™ probes (see Figure 5 ).
1
Probe Port (5-pin)
Figure 5 Connectors on HQ40d Meter
2
USB/DC Power Adapter Port (8-pin)
15
Installation
3.5 Data Transfer (HQ40d only)
Important Note: The battery compartment of the meter and the
USB/DC power adapter are not waterproof. Use care when operating these devices on a bench in wet environments. Water may infiltrate these devices and eventually cause performance or quality problems. Periodic inspection of the batteries and battery compartment is recommended, if the meter is used in wet environments: remove, clean, and dry the batteries, the interior of the battery compartment, and the battery contacts; then reinsert the batteries and close the compartment cover.
USB - Peripheral
USB - Host
Data can be transferred to a PC, printer, or flash memory stick by using the USB/DC power adapter. Refer to Figure 6 for USB/DC
Adapter connections.
The USB peripheral connector on the USB/DC power adapter is used for data transfer to a flash memory stick or printer. The flash stick can be connected to a PC for data transfer.
The USB host connector on the USB/DC power adapter is used for direct connection to a PC using a standard USB cable. The HQ40d
Application Software must be installed onto a PC for direct communication with the meter. Start the Application Software to transfer data.
To conserve battery life, USB functionality is enabled only when the meter is connected to AC power. To enable USB:
1. Turn the instrument off.
2. Connect the USB/DC adapter to the instrument.
3. Plug the AC power cord into the AC power adapter. Connect the power output jack from the AC power adapter to the
USB/DC power adapter.
4. Plug the AC power cord into an AC receptacle.
5. Turn on the instrument and plug in the desired USB device
(refer to Figure 6 ).
See section 5.3 on page 31 for more information on data transfer
using the HQ40d meter.
16
Installation
1
AC-DC Power Supply
2
Power Cord
3
USB/DC Power Adapter
Figure 6 USB/DC Adapter on HQ40d Meter
4
Personal Computer Connection (USB Host)
5
Flash Memory Stick/Printer Connection (USB Peripheral)
17
Visit us at www.hach.com
Section 4 System Start Up
4.1 Basic Start Up Overview
1. Attach the probe and power the meter.
When a probe is connected to a HQ30d or HQ40d portable meter, the meter automatically recognizes the parameter and is ready for use. The HQ11d only measures pH. The HQ14d only measures conductivity, salinity, and total dissolved solids
(TDS).
2. Select the language to display on the screen.
Refer to
3. Set the date and time. Refer to
4. Set the Sample and Operator IDs. Refer to
5. Calibrate the probe.
•
pH Probe,
•
Conductivity Probe,
•
LDO Probe (use factory-default setting)
6. Take a measurement reading.
•
pH Probe,
•
Conductivity Probe,
•
LDO Probe,
7. Run Check Standards (pH and Conductivity only).
•
pH Probe,
•
Conductivity Probe,
8. Set the method.
•
pH Probe,
•
Conductivity Probe,
•
LDO Probe,
9. Modify Meter Options. Refer to
.
19
System Start Up
4.2 Meter User Interface and Navigation
4.2.1 Keypad Description
Figure 7 shows the meter keypad and key descriptions common to all models.
1 POWER ON/OFF
Figure 7 HQ Series Keypad Description
6
UP
and
DOWN
Softkeys: function changes with software menus
2 OPERATOR ID 7
GREEN/RIGHT
Softkey: function changes with software menu
3 BACKLIGHT
4 SAMPLE ID
5
BLUE/LEFT
Softkey: function changes with software menu
8 DATA LOG
9 METER OPTIONS AND PARAMETER METHODS
20
System Start Up
4.2.2 Display Description (Single and Dual)
4.2.2.1 Using Single Screen Mode
The meter displays the concentration, units, temperature, calibration status, operator ID, sample ID, date, and time as shown in Figure 8 .
The HQ40d meter can display two parameters simultaneously in the dual screen mode. See
.
1
Calibration Indicator
2
IntelliCAL™ Probe Type
3
Main Measurement Unit
4
Battery Status
5
Sample Temperature
6
Additional Units
Figure 8 Single Screen Mode
7
Time
8
Date
9
Menu Driven Function Bar (Operated by
GREEN/RIGHT
key,
UP
and
DOWN
keys, and
BLUE/LEFT
key)
10 Sample and Operator Identification
11 Stability or Display Lock Indicator
4.2.2.2 Using Dual Screen Mode (HQ40d only)
When two probes are connected to the HQ40d meter, the screen can show the reading from both probes simultaneously, or show just one probe.
With two probes connected to the meter, use the
UP
and
DOWN
keys to change the screen mode to single or dual screen. In dual screen mode, the
UP
key will select the left probe for single view and the
DOWN
key will select the right probe.
21
System Start Up
4.3 Selecting the Language
The meter can be operated in several different languages. When the meter is turned on for the first time, the user must select a language before any other meter functions can be accessed.
To select or change the language:
1. Press the
METER OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Language. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select a language. Press the
GREEN/RIGHT
key under OK.
4.4 Setting the Date and Time
CAUTION
BEFORE ATTACHING THE PROBE FOR THE FIRST TIME: Set the date and time in the meter before attaching the probe for its first use. If the meter date and time are incorrect when the probe is installed, the probe will retain this incorrect time stamp for the remainder of its service life.
To change the time and date that is displayed on the meter:
1. Press the
METER OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Time. Press the
GREEN/RIGHT
key under Select.
22
System Start Up
3. By default, time entries use the 24-hour clock format. Use the
UP
and
DOWN
keys to highlight Set Time, Time Format, Set
Date, or Date Format. Press the
GREEN/RIGHT
key under
Select.
SET TIME
Use the
UP
and
DOWN
keys to change the time. Press the
GREEN/RIGHT
key to advance to the next space.
Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry. All time entry is in
24-hour format.
TIME FORMAT
Use the
UP
and
DOWN
keys to select an AM/PM (12-hour) or
24-hour time clock. Press the
GREEN/RIGHT
key under OK.
SET DATE
By default, date entries use the day-month-year format. Use the
UP
and
DOWN
keys to change the date. Press the
GREEN/RIGHT
key to advance to the next space. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
DATE FORMAT
Use the
UP
and
DOWN
keys to select a date format. Press the
GREEN/RIGHT
key under OK.
23
Visit us at www.hach.com
Section 5 Standard Operation
Important Note: Screen shot examples in this manual are included for illustrative purposes, and may not reflect actual results.
5.1 Setting the Sample and Operator Identification
Use the
SAMPLE ID
key to associate sample readings with a particular sample location. The Sample ID will be shown in the lower left corner of the display, and all stored data will include this
ID. If no Sample ID is entered, the meter will display a generic
“Sample ID”.
Use the
OPERATOR ID
key to associate sample readings with an individual. The Operator ID will be shown in the lower left corner of the display, and all stored data will include this ID. If no Operator ID is entered, the meter will display three dashes in the display.
5.1.1 Sample ID
Note: Sample and Operator ID text can be entered using the optional keyboard (HQ40d only).
5.1.1.1 Creating a New Sample ID
To enter a new Sample ID:
1. Press the
SAMPLE ID
key.
2. Use the
UP
and
DOWN
keys to highlight Create New Sample
ID. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to scroll through the letters and numbers. To accept a letter or number, press the
GREEN/RIGHT
key. The cursor will advance to the next space.
4. Repeat the previous step to add additional letters or numbers until the name is complete. To add a space, scroll to the blank space between A and 9 using the
UP
and
DOWN
keys and press the
GREEN/RIGHT
key. To replace a letter or number, press the
BLUE/LEFT
key and re-enter the letter or number.
5. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
25
Standard Operation
5.1.1.2 Selecting a Sample ID
To select a different Sample ID:
1. Press the
SAMPLE ID
key.
2. Use the
UP
and
DOWN
keys to highlight Current ID. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select the correct Sample ID.
Press the
GREEN/RIGHT
key under OK.
5.1.1.3 Deleting a Sample ID
To delete an existing Sample ID:
1. Press the
SAMPLE ID
key.
2. Use the
UP
and
DOWN
keys to highlight Delete Sample ID.
Press the
GREEN/RIGHT
key under Select.
Note: The default Sample ID cannot be deleted.
3. Use the
UP
and
DOWN
keys to select the Sample ID to be deleted. Press the
GREEN/RIGHT
key under Delete.
26
Standard Operation
5.1.2 Operator ID
5.1.2.1 Creating a New Operator ID
To create a new Operator ID:
1. Press the
OPERATOR ID
key.
2. Use the
UP
and
DOWN
keys to highlight Create New Operator
ID. Press the
GREEN/RIGHT
key under Select.
5.1.2.2 Selecting an Operator ID
3. Use the
UP
and
DOWN
keys to scroll through the letters and numbers. To accept a letter or number, press the
GREEN/RIGHT
key. The cursor will advance to the next space.
4. Repeat the previous step to add additional letters or numbers until the name is complete. To add a space, scroll to the blank space between A and 9 using the
UP
and
DOWN
keys and press the
GREEN/RIGHT
key. To replace a letter or number, press the
BLUE/LEFT
key and re-enter the letter or number.
5. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Press the
GREEN/RIGHT
key under OK.
To select an existing Operator ID:
1. Press the
OPERATOR ID
key.
2. Use the
UP
and
DOWN
keys to highlight Current ID. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select the Operator ID. Press the
GREEN/RIGHT
key under OK.
27
Standard Operation
5.1.2.3 Deleting an Operator ID
To delete an existing Operator ID:
1. Press the
OPERATOR ID
key.
Note: The Default Operator ID cannot be deleted
2. Use the
UP
and
DOWN
keys to highlight Delete Operator ID.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select the Operator ID to be deleted. Press the
GREEN/RIGHT
key under Delete.
Note: If the Current Operator ID is deleted, Default becomes the
Current Operator ID.
5.2 Using the Data Log
5.2.1 Storing Data
5.2.2 Viewing Stored Data
The HQ series portable meters can store up to 500 sample measurement, calibration, or check standard measurement results.
Data is stored automatically when Press to Read or Interval is selected for Measurement Mode in the Meter Options menu. When
Continuous is selected, data will only be stored when the
GREEN/RIGHT
key under Store is pressed.
The data log records events chronologically, and displays the current number of data records (for example Data Record 250 of
500). Perform the following to recall data:
1. Press the
DATA LOG
key.
2. Use the
UP
and
DOWN
keys to highlight View Data Log. Press the
GREEN/RIGHT
key under Select.
3. The display shows the most recent measurement, calibration, or check standard. Use the
UP
and
DOWN
keys to scroll through the stored data.
28
5.2.3 Viewing Probe Data
Standard Operation
READING LOG
The measurement or Reading Log shows the most recent measurement value with associated time, date, operator and sample ID. A warning message appears if an error is associated with the measurement such as over limit or expired calibration.
The first screen displays information associated with the reading.
Press the
GREEN/RIGHT
key to access the calibration details for the reading. Press the
UP
and
DOWN
keys under Details. Scroll (if necessary) to the desired information associated with the reading.
CALIBRATION LOG
The Calibration Log shows the most recent calibration data. Any error messages appear before other calibration data. If there are multiple error messages, they are listed in the Details screen.
The first screen displays information associated with the calibration.
Press the
GREEN/RIGHT
key to access the calibration details for the calibration. Press the
UP
and
DOWN
keys under Details. Scroll (if necessary) to the desired information associated with the reading.
CHECK STANDARD LOG
The Check Standard Log shows the most recent check standard data with results.
The first screen displays the results of the check standard.
Press the
GREEN/RIGHT
key to access the calibration details for the reading. Press the
UP
and
DOWN
keys under Details. Scroll (if necessary) to the desired information associated with the reading.
Note: A probe must be connected to the meter to use the View Probe Data function.
1. Press the
DATA LOG
key on the meter.
2. Use the
UP
and
DOWN
keys to highlight View Probe Data.
Press the
GREEN/RIGHT
key under Select.
29
Standard Operation
3. If only one probe is connected, proceed to Step 4. If two probes are connected (HQ40d only), the connected probes will be displayed. Use the
UP
and
DOWN
keys to highlight the probe.
Press the
GREEN/RIGHT
key under Select.
4. Use the
UP
and
DOWN
keys to highlight one of the selections for Probe Data. Press the
GREEN/RIGHT
key under Select.
5.2.4 Deleting Data
Data will be automatically deleted when the data log is full on a first in first out basis (oldest data deleted first).
Data can be deleted manually when Access Control is off, or when a valid password is entered.
1. Press the
DATA LOG
key.
2. Use the
UP
and
DOWN
keys to highlight Delete Data Log.
Press the
GREEN/RIGHT
key under Select.
3. The display will show “Delete All Data?”. Press the
GREEN/RIGHT
key under Select to delete all stored data.
30
Standard Operation
5.3 Transferring Data (HQ40d meter only)
Note: If the response time is slow when transferring data, reformat the flash memory stick or computer to use the file allocation table
(FAT 16 or 32) format.
Data can be transferred to a printer, flash memory stick, or printer using the USB connector on the HQ40d USB/DC adapter. The meter must be powered up after being connected to AC power for data transfer to occur.
5.3.1 Sending Data to a Printer or Flash Memory Stick
USB - Peripheral
1. Connect the HQ40d meter to an electrical outlet using the
USB/DC power adapter, and then turn on the meter.
2. Connect the printer cable or flash memory stick to the peripheral USB connector on the USB/DC Power Adapter (for more information, see
).
3. Press the
DATA LOG
key on the HQ40d meter.
4. Use the
UP
and
DOWN
keys to highlight Send Data Log. Press the
GREEN/RIGHT
key under Select.
5.3.1.1 Selecting Report Options
5. The display will show “Sending Data” to indicate that the data is being stored to the flash memory stick or sent to the printer. All measurement data, calibration data, and check standard results will be printed at the level of detail selected under
Report Options.
Important Note: The HQ40d meter can connect directly to a USB compatible 72-column page printer.
Note: In addition to data, all user-created methods present on a meter are copied to the flash memory stick. When the memory stick is used by another meter, these methods are copied to the new meter (refer to section
5.6 Archiving and Exchanging Methods (HQ40d only) on page 40
).
Note: Do not unplug the flash memory stick or the printer cable while
"Sending Data" screen is displayed, or the
meter may lock up
.
the
The data log can be printed at three levels of detail: Basic Report,
Advanced Report or Total Report (
). The report option displayed on the data log screen indicates the level of detail which is currently selected.
To select a different report option:
31
Standard Operation
1. Press the
DATA LOG
key.
2. Use the
UP
and
DOWN
keys to highlight Report Options. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select the desired report type.
Press the
GREEN/RIGHT
key under OK.
5.3.1.2 Printing Calibration Data
Calibration data is printed as part of the data log, but can also be printed from the stored information in the probe. Printed calibration
reports are shown in section 5.5 on page 38 .
To print calibration data that is stored in the probe:
1. Press the
DATA LOG
key.
2. Use the
UP
and
DOWN
keys to highlight View Probe Data.
Press the
GREEN/RIGHT
key under Select.
Note: A probe must be connected to the meter to use the View Probe
Data function.
3. If only one probe is connected, proceed to Step 4.
If two probes are connected, the connected probes will be displayed. Use the
UP
and
DOWN
keys to highlight the desired probe. Press the
GREEN/RIGHT
key under Select.
32
Standard Operation
4. Use the
UP
and
DOWN
keys to highlight Send Current
Calibration or Send Calibration History:
•
Send Current Calibration: sends the most recent calibration information.
•
Send Calibration History: sends all calibration information that is stored in the probe.
Press the
GREEN/RIGHT
key under Select.
5. The display will show “Sending Data” to indicate the data is being sent to the printer.
Important Note: Do not disconnect the printer cable while the
"Sending Data" screen is displayed, or the
meter may lock up
.
5.3.2 Sending Data to a Computer
USB - Host
1. Load the HQ40d PC Application Software supplied with the
HQ40d meter onto a PC.
2. Turn off the meter, connect the HQ40d meter to an electrical outlet using the USB/DC power adapter, and then turn on the meter.
3. Connect a USB cable (Cat. No. 59240-00) to the host USB connector on the USB/DC adapter and to the computer.
4. Open the HQ40d PC Application Software on the PC. Click on the green triangle shown in the menu bar to initiate a connection.
5. Press the
DATA LOG
key on the HQ40d meter.
6. Use the
UP
and
DOWN
keys to highlight Send Data Log. Press the
GREEN/RIGHT
key under Select.
7. The meter display will show “Sending Data”. The data will appear in the PC Application Software window. The file is saved in Comma Separated Values (.csv) file format.
Note: For more information on working with CSV files, please reference the
"Working with Data" section of the HQ40d CD.
33
Standard Operation
5.4 Viewing Data Log Reports (HQ40d only)
When the data log is sent to a printer ( section 5.3.1 on page 31 ),
the printed report contains all stored sample data, check standard data, and calibration information.
Note: All error messages will print at the end of each report option selected
(Basic Report, Advanced Report, or Total Report).
5.4.1 Report Names
The first line of each report displays the report name, which is associated with the data log file. Figure 9 shows an example of the report name on a printed report.
1
Serial Number
2
Report Label
5.4.2 Sample Results
5.4.2.1 Basic Reports
Figure 9 Report Name
3
Date and Time (24 h) (YYMMDDhhmm)
4
File Type Extension
The amount of information that is printed for sample readings varies with the selected Report Option (Basic, Advanced, or Total).
A Basic Report contains a single line of information per sample reading. Figure 10 details the information available in this type of report.
Figure 10 Sample Data on a Basic Report
1
Measurement Type (RD = Reading)
2
Parameter Type (pH, LDO, CDC, etc.)
3
Time (hh:mm:ss in 24 h format)
4
Date (DD-MM-YY or user-defined format)
5
Reading Value
6
Reading Units
7
Sample ID: user-defined, displays "SAMPLE ID" if undefined
8
Sample ID Counter
9
Operator ID: user-defined, displays "- - -" if undefined
34
5.4.2.2 Advanced Reports
Standard Operation
An Advanced Report contains two lines of information per sample reading. The first line of information is the same as the information in a Basic Report. The second line includes additional information as shown in Figure 11 .
Figure 11 Sample Data on an Advanced Report
1
Probe Model Name
2
Error Message (if applicable)
3
Probe Serial Number: prefaced by "<" or ">" to indicate which channel the probe was connected to during a dual reading.
4
Additional Units: displays all additional units associated with the reading. Contents vary depending on type of parameter being read.
5
Method Settings: displays highest-priority method setting associated with the reading. Contents vary depending on type of parameter being read and configuration of specific method.
5.4.2.3 Total Reports
A Total Report contains three lines of information per sample reading. The first two lines of information are the same as the information in an advanced report. The third line includes additional information as shown in
.
Figure 12 Sample Data on a Total Report
1
Method Name: user-defined Method Name used to take this reading
2
Calibration Time: time of last calibration, prefaced by "CAL" and displayed as hh:mm in 24 h format
3
Calibration Date: date of last calibration (DD-MM-YY or user-defined format)
4
Calibration Slope
5
Offset: contents vary depending on type of parameter being read and configuration of specific method. May be blank.
6
Calibration Operator ID: user-defined Operator ID of person who performed the last calibration.
Displays "- - -" if undefined.
35
Standard Operation
5.4.3 Calibration Results
Calibration data is included in the data log printout as two lines of information (
Figure 13 ).The amount of information that is displayed
for calibration data does not vary with the selected Report Option.
Figure 13 Calibration Data Report
1
Measurement Type (CL = Calibration)
2
Parameter Type (pH, LDO, CDC, etc.)
3
Calibration Time: prefaced by "CAL" and displayed as hh:mm in 24 h format
4
Calibration Date (DD-MM-YY or user-defined format)
5
Calibration Slope
6
Offset: contents vary depending on type of parameter being read and configuration of specific method. May be blank.
7
r
2: contents vary depending on type of parameter being read, configuration of specific method and number of calibration standards read. May be blank.
8
Calibration Operator ID: user-defined Operator ID of person who performed this calibration. Displays "- - -" if undefined.
9
Method Name: user-defined Method Name used to take this reading
10 Probe Model Name
11 Probe Serial Number: prefaced by "<" or ">" to indicate which channel the probe was connected to during a dual reading
36
5.4.4 Check Standard Results
Standard Operation
Check standard results are included in the data log printout as a single line of information ( Figure 14 ). The amount of information that is displayed for check standard results does not vary with the selected Report Option.
Figure 14 Check Standard Report
1
Measurement Type (ST = Check Standard)
2
Parameter Type (pH, LDO, CDC, etc.)
3
Check Standard Time (hh:mm:ss in 24 h format)
4
Check Standard Date (DD-MM-YY or user-defined format)
5
Reading Value
6
Reading Units
7
Check Standard Pass/Fail: identifies whether Check Standard acceptance criteria has been met
8
Check Standard Operator ID: user-defined Operator ID of person who performed the check standard.
Displays "- - -" if undefined.
37
Standard Operation
5.5 Viewing Calibration Reports (HQ40d only)
A report can be printed for current calibration information or
calibration history as described in section 5.3.1.2 on page 32 .
5.5.1 Current Calibration Reports
A Current Calibration Report contains two lines of information.
Figure 15 details the information available in this type of report.
Figure 15 Current Calibration Report
1
Report Type (IC = Current Calibration)
2
Parameter Type (pH, LDO, CDC, etc.)
3
Calibration Time: prefaced by "CAL" and displayed as hh:mm in 24 h format
4
Calibration Date (DD-MM-YY or user-defined format)
5
Calibration Slope
6
Offset: contents vary depending on type of parameter being read and configuration of specific method. May be blank.
7
r
2: contents vary depending on type of parameter being read, configuration of specific method and number of calibration standards read. May be blank.
8
Calibration Operator ID: user-defined Operator ID of person who performed this calibration. Displays "- - -" if undefined.
9
Method Name: user-defined Method Name used to take this reading
10 Probe Model Name
11 Probe Serial Number: prefaced by "<" or ">" to indicate which channel the probe was connected to during a dual reading.
38
Standard Operation
5.5.2 Calibration History Reports
A Calibration History Report contains two lines of information per calibration. Figure 16 details the information available in this type of report.
Figure 16 Calibration History Report
1
Report Type (CH = Calibration History)
2
Parameter Type (pH, LDO, CDC, etc.)
3
Calibration Time: prefaced by "CAL" and displayed as hh:mm in 24 h format
4
Calibration Date (DD-MM-YY or user-defined format)
5
Calibration Slope
6
Offset: contents vary depending on type of parameter being read and configuration of specific method. May be blank.
7
Probe Model Name
8
Probe Serial Number: prefaced by "<" or ">" to indicate which channel the probe was connected to during a dual reading.
39
Standard Operation
5.6 Archiving and Exchanging Methods (HQ40d only)
In addition to capturing real-time data and transferring data from the data log, user-created methods can be saved to a USB flash memory stick. These methods can then be archived to a PC or transferred to another HQd meter.
Whenever a flash memory stick is connected to a meter, a 'User' folder is created on the memory stick, and all user-created methods that are stored in the meter are archived to this folder.
When the memory stick is connected to another meter, all methods in the 'User' folder will be uploaded to the new meter. In addition, any user-created methods in the new meter that are not on the memory stick will be downloaded to the memory stick.
The method names will appear in the 'User' folder with a prefix by the parameter type and an underscore (e.g., LDO_, Cond_, pH_), and a file extension of .NGM. Only the METHOD NAME appears in the method library in the meter.
If a method on a meter has the same name as a method on a memory stick, the method in the meter will be overwritten. For example, if a file on the USB flash memory stick has the name
"LDO_SOUR TEST.NGM", this file will overwrite an LDO method on another meter with the same displayed method name (SOUR
TEST).
When a memory stick is connected to a PC, a user can delete or change the name of any method, or delete the entire 'User' folder on the memory stick. Method names that are changed must preserve the prefix (e.g. LDO_) and the file extension (.NGM). The
METHOD NAME is limited to 12 characters. Allowable characters in the METHOD NAME are A through Z, 0 through 9, and one or more spaces.
When a memory stick is connected to a meter, any user-created methods on the meter will be downloaded to the memory stick. The
'User' folder will be recreated each time the memory stick is connected. If the 'User' folder does not exist, a new 'User' folder will be created. The meter will not upload any methods from the memory stick if the 'User' folder does not exist or is empty.
40
Section 6 pH Operation and Methods
6.1 Calibrating the pH Probe
Note: Use pH buffer solutions to calibrate the pH probe. The minimum number required and values of the pH buffer solutions are specified in the
Calibration Options menu. A maximum of three buffers can be entered by using the “Color Coded” and DIN buffer sets. A maximum of four buffers can be entered by using the IUPAC buffer set.
Note: If using the HQ40d meter with two probes, the display must be in single display mode.
1. Press the
BLUE/LEFT
key under Calibrate.
2. The display will show the buffer values to be measured. These values are set in the Calibrations Options menu. Rinse the probe and place it in the first buffer solution. The probe automatically recognizes buffer values from a selected set, therefore entering the buffer values in a specific order is not required. However, it is recommended to begin with the lowest pH buffer for greatest accuracy.
3. Press the
GREEN/RIGHT
key under Read. The meter will automatically detect which buffer is being measured.
When the reading is stable, the display will highlight the buffer that has been read and display the temperature corrected pH value.
Temperature correction is automatic: the displayed pH is the true pH value of the buffer at the measured temperature.
4. Rinse the probe and place it in the next buffer solution.
5. Press the
GREEN/RIGHT
key under Read.
6. Repeat this procedure until the minimum number of calibration points specified in the pH Method have been acquired.
7. When the minimum number of buffer points are collected, Done appears above the
UP
key. Continue calibrating with the
Method-specified calibration buffers (up to the maximum number specified in the pH Method), until all have been used, or press the
UP
key to review the Calibration Summary.
41
pH Operation and Methods
8. The Calibration Summary will appear. Press the
GREEN/RIGHT
key under Store to accept the calibration and return to measurement mode. The calibration is recorded in the
500-result data log. For an HQ40d model, the calibration information is also sent to a PC/printer/flash memory stick if connected.
9. When the calibration is successful, the display will show OK in the upper left corner. The icon will appear as a question mark if the calibration information has expired or if a check standard has failed or been delayed.
6.1.1 Calibration Errors
6.1.1.1 Standard Not Recognized
If the mV reading of the buffer does not fall within the limits set for auto detection, the display will show “Standard not recognized.” If this happens, perform the following steps:
1. Rinse the probe and place it in a fresh buffer solution.
2. Press the
GREEN/RIGHT
key under Read. If the meter still does not recognize the buffer, be sure that the buffers used are the ones specified for the method. If so, refer to the probe instruction sheet for cleaning and troubleshooting procedures.
6.1.1.2 Slope Error
If the calibration slope does not meet the acceptance criteria, the display will show “Slope out of range”. If this happens, perform the following steps:
1. Obtain fresh buffer solutions. Rinse the probe and repeat the calibration.
2. If the meter still gives a Slope error, be sure the buffers that are used are the ones specified for the method. If the error continues, there may be a problem with the probe.
42
pH Operation and Methods
6.2 Taking a pH Measurement
If complete traceability is required, set the Sample ID and
Operator ID before taking a measurement.
Note: The default setting for Measurement Mode is “Press to Read”. If a different mode is required, change the Measurement Mode (see
1. Place the pH probe into the sample.
2. Press the
GREEN/RIGHT
key under Read.
3. The display will show “Stabilizing…” and a progress bar will fill from 0 to 100% as the probe stabilizes in the sample. Stability is determined by using a fixed change in the signal/time equation.
The lock icon will appear and the result will be automatically stored in the 500-result data log.
4. To take another measurement, repeat this procedure.
6.3 Running Check Standards Manually or Automatically
The Run Check Standards option for pH verifies reading accuracy by measuring a buffer solution of known pH and comparing actual values to theoretical values. When the probe is placed in the solution, the meter will indicate if the check standard passed or failed based on user-selected criteria.
A check standard can be measured at any time by using the
Meter
Options
menu or at specific intervals. Set the criteria for Check
Standards from the
Meter Options
menu.
Note: Access Control must be off or a valid password entered to change the acceptance criteria.
When the Check Standard reminder is ON, the meter will automatically display the Check Standard screen. The check standard can either be measured immediately, or be delayed and measured at a later time (this is a user-specified option, see
Parameter Methods
in
).
To measure the Check Standard:
1. Obtain the pH buffer solution specified for the check standard.
The buffer solution to be used is shown in the display.
2. Place the probe in the buffer solution.
3. Press the
GREEN/RIGHT
key under Read.
43
pH Operation and Methods
4. The display shows the value of the check standard and either
“Check Standard Passed” or “Check Standard Failed” appears.
If “Check Standard Passed” appears, the success criteria has been met, and the measurement has been verified to be accurate. Press the
GREEN/RIGHT
key under Done to proceed with sample measurements.
5. If “Check Standard Failed” appears, the measurement is outside of the accepted limits.
If the acceptance criteria is set to “Cal Expires on Failure: Yes”, the instrument will display the
CALIBRATION ?
icon until it is re-calibrated. Press the
BLUE/LEFT
key under Calibrate and follow the steps for calibration.
6.3.1 Deferring a Check Standard
A Check Standard Reminder can be deferred to a later time. This option is set in the Parameter Methods menu so that a supervisor can control this function. A password is required to change this setting. To defer the Check Standard measurement to a later time, use the Reading Check Standard screen.
1. Press the
UP
key under Defer.
2. Use the
UP
and
DOWN
keys to select when the next reminder will be displayed.
3. Press the
GREEN/RIGHT
key under OK. The Check Standard reminder will appear after the selected time has passed.
44
pH Operation and Methods
6.4 Setting the pH Method
pH MAIN MENU
Current Method
Save Current Method As
Modify Current Method
Delete a Method
The pH Method menu is available via the
METER OPTIONS/PARAMETER METHODS
key when Access Control is off, or when a valid password is entered. A pH probe must be connected to the meter to edit pH methods. Method selection is not restricted.
outlines the menu options for a pH Method. These options do not need to be changed if the default method is used. Modify
Current Method submenus and default settings are described in
.
The HQ series meters contain a default method for pH with settings for measurement, calibration, check standards, and units. The default settings cannot be changed.
Use the Save Current Method As function to save the selected method with a new name. The meter settings for this new method can then be modified.
Note: To enter options that are different from the default settings, a new method must be created and then modified, as described in
Table 1 pH Methods Menu Summary pH SUBMENU
Select Method
Enter New Method Name
Measurement Options
Set Calibration Options
Set Check Standards Options
Units
Select Method
6.5 Modify Current Method Menu for pH Summary
pH METHOD OPTION
Measurement Options
AVAILABLE SELECTIONS
Resolution
Measurement Limits
0.1 Fast
0.01 Fast
0.01 Medium
0.01 Slow
0.001 Slow
Lower Limit: 0.00–14.00 pH
Upper Limit 0.00–14.00 pH
Calibration Options
Buffer set
Calibration Reminder
Color Coded 4, 7, 10
IUPAC 1.679, 4.005, 7.000, 10.012
DIN 1.09, 4.65, 9.23
Reminder: On or Off
Repeat: 30 min, 2 h, 4 h, 8 h, 2 d, 5 d, 7 d
Expires: Immediately, Reminder +30 min, + 1 h, + 2 h, continue reading
DEFAULT SETTING
0.01 Medium
Lower: 0.00 pH
Upper: 14.00 pH
Color Coded 4, 7, 10
Reminder: Off
Repeat: 8h
Expires: +30 min
45
pH Operation and Methods
6.5 Modify Current Method Menu for pH Summary (continued)
pH METHOD OPTION
Minimum Cal Points
Acceptance Criteria
Check Standards Options
Check Standard Values
AVAILABLE SELECTIONS
1, 2, or 3 Calibration Points
Slope Limit: 0 to ±10%
Check Standard
Reminder
Acceptance Criteria
Units
0–14 pH
Reminder: On or Off
Repeat: Off, 0.5 h, 2 h, 4h, 8h, 12 h, 24 h
Allow Defer: Yes or No
Limits: ±1 to ±10%
Calibration Expires On Failure: Yes or No pH or mV
DEFAULT SETTING
1 Calibration Point
±5%
7 pH buffer
Reminder: Off
Repeat: 4h
Defer: Yes
±5%
No pH
6.5.1 Modifying pH Methods
A new pH method can be entered when Access Control is off, or when a valid password is entered.
1. Press the
OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight pHC101 Method.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to highlight Save Current Method
As. Press the
GREEN/RIGHT
key under Select.
Note: The default method cannot be modified or deleted, but can be saved with a new name (Save Current Method As) and then modified.
4. Use the
UP
and
DOWN
keys to scroll through the letters and numbers. To select a letter or number, press the
GREEN/RIGHT
key. The cursor will advance to the next space.
5. Repeat the previous step to add additional letters or numbers until the name is complete. To add a space, scroll to the blank space (between A and 9) using the
UP
and
DOWN
keys and press the
GREEN/RIGHT
key. To delete a letter or number, press the
BLUE/LEFT
key and re-enter the letter or number.
6. Press the
GREEN/RIGHT
key until OK replaces the Right arrow in the function bar. Press the
GREEN/RIGHT
key under OK.
Alternately, use the accessory USB keyboard option
(HQ40d only).
46
pH Operation and Methods
6.5.2 Modifying the pH Measurement Options
Edit measurement options to change the displayed resolution or upper and lower pH limit.
1. With Measurement Options highlighted in the Modify Current
Method menu, press the
GREEN/RIGHT
key under Select.
To Edit the Resolution
1. With Measurement Options highlighted in the Modify Current
Method menu, press the
GREEN/RIGHT
key under Select.
2. With Resolution highlighted, press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select the desired resolution and speed of response. The most accurate readings are obtained at the “slow” response settings. Press the
GREEN/RIGHT
key under OK.
47
pH Operation and Methods
To Edit the Upper and Lower pH Limits
1. With Measurement Options highlighted in the Modify Current
Method menu, press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select Measurement Limits.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select Lower Limit or Upper
Limit. Press the
GREEN/RIGHT
key under Select.
4. Set Lower Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
5. Set Upper Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
48
pH Operation and Methods
6.5.3 Modifying the pH Calibration Options
Edit calibration options to change the specified buffer sets for calibration, calibration reminders, minimum required number of calibration points, and slope acceptance criteria.
To modify the current calibration method options, use the
UP
and
DOWN
keys to highlight Calibration Options in the Modify Current
Method menu. Press the
GREEN/RIGHT
key under Select.
The Calibrations Options menu will appear. Edit the buffer sets, calibration reminder, calibration points, and slope limits using the following steps.
Editing the Buffer Sets for Automatic Recognition.
1. Use the
UP
and
DOWN
keys to highlight Buffer Set. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the desired buffer set for calibration. Press the
GREEN/RIGHT
key under OK. The meter will use these buffers for auto recognition.
Editing the Calibration Reminder
1. Use the
UP
and
DOWN
keys to highlight Set Calibration
Reminder. Press the
GREEN/RIGHT
key under Select.
49
pH Operation and Methods
To turn Reminder On or Off:
2. Use the
UP
and
DOWN
keys to highlight Reminder. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select On or Off. Press the
GREEN/RIGHT
key under OK.
To set the Reminder frequency:
4. Use the
UP
and
DOWN
keys to highlight Repeat. Press the
GREEN/RIGHT
key under Select.
5. Use the
UP
and
DOWN
keys to select the desired reminder frequency. Press the
GREEN/RIGHT
key under OK.
To edit the calibration expiration:
6. Use the
UP
and
DOWN
keys to highlight Expires. Press the
GREEN/RIGHT
key under Select.
50
pH Operation and Methods
7. Use the
UP
and
DOWN
keys to select how long after the reminder the calibration will expire. Press the
GREEN/RIGHT
key under OK.
Note: The meter cannot be used for measuring samples after the calibration expires unless the Continue Reading setting is enabled.
Editing the Calibration Points
The meter can be set to require a minimum number of calibration points be entered before calibration can be completed. To set the minimum number of buffers that are required to complete calibration:
1. Use the
UP
and
DOWN
keys to highlight Minimum Cal Points.
Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the desired minimum number of calibration points. Press the
GREEN/RIGHT
key under OK.
Setting the Calibration Slope Limit
The meter can detect a calibration point slope that is not within tolerance. The tighter the limits the more accurate a successful calibration.
To change the acceptable slope tolerance for calibration:
1. Use the
UP
and
DOWN
keys to highlight Slope Limit. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to enter a value for acceptable slope limit. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
The meter will reject a calibration if the slope falls outside of the specified slope tolerance limit.
51
pH Operation and Methods
6.5.4 Modifying the pH Check Standard Options
Edit check standard options to change the buffer solution used for the check standard, reminder, and acceptance criteria.
Editing the Check Standard Buffer
From the Modify Current Method screen:
1. Use the
UP
and
DOWN
keys to highlight Check Standard
Options (not pictured).
2. To edit the buffer of a known value used for the Check
Standard, use the
UP
and
DOWN
keys to highlight Check
Standard. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to enter a value for the buffer solution to use for the Check Standard. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
Editing the Check Standard Reminder Options
To turn the Check Standard reminder on or off, set the frequency of the check standard reminder, and edit the defer option refer to the following steps.
To edit Check Standard Reminder Options:
1. Use the
UP
and
DOWN
keys to highlight Check Standard
Reminder. Press the
GREEN/RIGHT
key under Select.
2. To turn the Check Standard on or off, use the
UP
and
DOWN
keys to highlight Reminder. Press the
GREEN/RIGHT
key under Select.
52
pH Operation and Methods
3. Use the
UP
and
DOWN
keys to select On or Off. Press the
GREEN/RIGHT
key under OK.
4. To set the Check Standard Reminder frequency, use the
UP
and
DOWN
keys to select Repeat. Press the
GREEN/RIGHT
key under Select.
5. Use the
UP
and
DOWN
keys to select the reminder frequency.
Press the
GREEN/RIGHT
key under OK.
6. To edit the defer options, use the
UP
and
DOWN
keys to select
Allow Defer. Press the
GREEN/RIGHT
key under Select.
7. Use the
UP
and
DOWN
keys to select Yes or No. Press the
GREEN/RIGHT
key under OK.
Editing the Acceptance Criteria for Check Standards
1. Use the
UP
and
DOWN
keys to highlight Acceptance Criteria.
Press the
GREEN/RIGHT
key under Select.
Refer to the following steps to edit the Check Standard acceptance criteria or edit whether a failed Check Standard requires re-calibration before continuing measurements.
1. Use the
UP
and
DOWN
keys to highlight Acceptance Limits.
Press the
GREEN/RIGHT
key under Select.
53
pH Operation and Methods
2. Use the
UP
and
DOWN
keys to enter a percent range of accuracy that the Check Standard must fall within. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
3. Use the
UP
and
DOWN
keys to highlight Cal Expires on
Failure. Press the
GREEN/RIGHT
key under Select.
4. Use the
UP
and
DOWN
keys to select Yes or No. Press the
GREEN/RIGHT
key under OK.
When set to Yes, the meter must be calibrated when a Check
Standard falls outside of the specified acceptance limits. No measurements can be taken until the meter is successfully re-calibrated.
When set to No, the meter will operate normally in the measurement mode.
6.5.5 Modifying the pH Measurement Units
The meter will display both pH and mV values in the measurement mode.
To change which unit is prominent:
1. Use the
UP
and
DOWN
keys to highlight Units. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select pH or mV. Press the
GREEN/RIGHT
key under OK.
54
Section 7 Conductivity Operation and Methods
7.1 Calibrating the Conductivity Probe
Calibrating a conductivity probe establishes the linear cell constant of the probe. Use a conductivity standard solution to calibrate the conductivity probe. The conductivity standard can be specified in
the Conductivity Options menu (see section 7.4 on page 58 ).
1. Press the
BLUE/LEFT
key under Calibrate.
Note: If using the HQ40d meter with two probes, the display must be in single screen mode.
2. The display will show the required conductivity standard solution. Rinse the probe and place it in the standard solution.
Note: All conductivity calibrations are performed using conductivity units (µS/cm, mS/cm) regardless of whether conductivity, resistivity, salinity, or total dissolved solids (TDS) is measured.
3. Press the
GREEN/RIGHT
key under Read. The meter will calculate the cell constant based on the selected calibration standard.
4. When the reading is stable, the display will show the temperature corrected value of the conductivity reading of the standard solution.
5. Press the
UP
key under Done.
6. The Calibration Summary will appear. Press the
GREEN/RIGHT
key under Store to accept the calibration and return to the measurement mode. The calibration is stored in the meter data log. If using the HQ40d meter, the calibration information is also sent to a PC/printer/flash memory stick if connected.
55
Conductivity Operation and Methods
7. When the calibration is successful, the display will show OK in the upper left corner. A question mark will be displayed if a calibration has expired or if a check standard has failed or been delayed.
7.2 Taking a Conductivity, Salinity, Resistivity, or TDS Measurement
If complete traceability is required, enter an Operator ID and
Sample ID before measuring samples: be sure the Operator ID and
Sample ID shown in the display are current. Press the
OPERATOR
ID
and
SAMPLE ID
keys to update.
Note: The default setting for Measurement Mode is “Press to Read”. If a different mode is required, change the Measurement Mode (see
1. Place the conductivity probe into the sample.
2. Press the
GREEN/RIGHT
key under Read.
3. The display will show “Stabilizing…” and a progress bar will fill from 0 to 100% as the probe stabilizes in the sample. The lock icon will appear and the result will be stored automatically in the
500-result data log.
4. This screen is an example of a conductivity measurement displayed in conductivity units. Repeat this procedure to take additional measurements. See
,
Figure 19 on page 57 .
Note: When a conductivity probe is attached to the meter, the measurement results will be displayed in units of conductivity, salinity, resistivity, or total dissolved solids (TDS). To change the measurement units, save the default method under a new name, and select the unit of choice in the new method.
Figure 17 Example of Conductivity Measurement Displayed in Salinity Units
56
Conductivity Operation and Methods
Figure 18 Example of Conductivity Measurement Displayed in TDS Units
Figure 19 Example of Conductivity Measurement Displayed in Resistivity Units
7.3 Running Check Standards Manually or Automatically
When the Check Standard reminder is on, the meter will automatically display the Check Standard screen. The check standard can either be measured immediately, or the measurement can be delayed until a later time.
To measure the Check Standard:
1. Obtain the conductivity standard solution specified for the check standard. The conductivity solution to be used is shown on the display.
2. Place the probe in the conductivity standard.
3. Press the
GREEN/RIGHT
key under Read.
4. The display will show the value of the check standard and either “Check Standard Passed” or “Check Standard Failed”.
5. If “Check Standard Passed” is displayed, the reading is verified to be accurate. Press the
GREEN/RIGHT
key under Done to proceed with sample measurements.
57
Conductivity Operation and Methods
6. If “Check Standard Failed” is displayed, the measurement is outside of the accepted limits.
7. If the acceptance criteria does not allow failed Check
Standards, all results will be displayed with the
CALIBRATION ?
icon, and will be stored with a flag indicating a suspect calibration. Press the
BLUE/LEFT
key under Calibrate and follow the steps for calibration.
7.3.0.1 Deferring a Check Standard
A Check Standard Reminder can be deferred to a later time. This option is set within the Method to allow for supervisor control of this function. A password may be required to change this setting.
To defer the Check Standard measurement to a later time:
1. Press the
UP
key under Defer.
2. Use the
UP
and
DOWN
keys to select when the next reminder will be displayed.
3. Press the
GREEN/RIGHT
key under OK. The Check Standard reminder will re-appear after the selected time has passed.
7.4 Setting the Conductivity Method
The Conductivity Method menu is available via the
METER OPTIONS/PARAMETER METHODS
key when Access Control is off, or when a valid password is entered. A conductivity probe must be connected to the meter to change these options.
Table 2 outlines the menu options for Conductivity Method. These options do not need to be changed if the default method is used.
The Modify Current Method submenus and default settings are described in detail in
CONDUCTIVITY MAIN MENU
Current Method
Save Current Method As
Modify Current Method
Delete a Method
The HQ series meters contain a default method for conductivity with settings for measurement, calibration, check standards, and units. The default settings cannot be changed.
To enter options that are different from the default settings, a new method must be created and then modified.
Table 2 Conductivity Parameter Method Menu Summary
CONDUCTIVITY SUBMENU
Select Method
Enter New Method Name
Parameter
Measurement Options
Set Calibration Options
Set Check Standards Options
Select Method
58
Conductivity Operation and Methods
7.5 Modify Current Method Menu for Conductivity Summary
CONDUCTIVITY OPTION
Parameter
AVAILABLE SELECTIONS
Conductivity
Salinity
TDS
Resistivity
Measurement Options
Units
Conductivity: Auto range between µS/cm and mS/cm, fixed µS/cm, or fixed mS/cm
Salinity: ppt, g/kg, <unitless>
TDS: no options other than mg/L
Resistivity: no options other than Ohm·cm
Measurement Limits
(conductivity, salinity,
TDS, and resistivity respectively)
Lower Limit: 0.01 µS/cm, 0 ppt, 0 mg/L, 5 Ohm·cm
Upper Limit: 200,000 µS/cm, 40 ppt, 50,000 mg/L, 5 x 10
7
Ohm·cm
Temperature Correction
1
Correction Factor
,2
Reference
Temperature
Calibration Options
DEFAULT SETTING
Conductivity
Auto range
(conductivity) ppt (salinity)
Lower: 0.01 µS/cm
Upper: 200,000 µS/cm
None
Linear (conductivity 1.9%C)
Non-Linear/NaCl
Natural Water
Enter Factor (Available only for conductivity and resistivity with linear temperature correction)
20 °C
25 °C
Parameter Based
Defaults
1.90%/°C
25 °C
Calibration Standard
Calibration Reminder
Standard Value
Reference Temperature
Temperature Correction
Check Standards Options
Check Standard Value
1 D, 111.3 mS/cm, 25 °C
0.1 D 12.85 mS/cm, 25 °C
0.01 D, 1.41 mS/cm, 25 °C
0.1 M, 12,880 µS/cm, 25 °C
0.01 M, 1413 µS/cm, 25 °C
0.001 M, 146.93 µS/cm, 25 °C
NaCl, 18 mS/cm, 25 °C
NaCl, 1000 µS/cm, 25 °C
NaCl, 25 µS/cm, 25 °C
NaCl, 0.05%, 1015 µS/cm, 25 °C
Seawater
Custom
Reminder: On or Off
Repeat: 30 min, 2 h, 4 h, 8 h, 2 d, 5 d, 7 d
Expires: Immediately, Reminder +30 min, + 1 h, + 2 h, continue reading
(For custom calibration standard only)
(For custom calibration standard only)
(For custom calibration standard only)
NaCl, 1000 µS/cm,
25 °C
Reminder: Off
Repeat: 8h
Expires: +30 min
User Selectable
User Selectable
User Selectable
Check Standard
Reminder
Enter value
Reminder: On or Off
Repeat: Off, 0.5 h, 2 h, 4h, 8h, 12 h, 24 h
Allow Defer: Yes or No
1413 µS/cm
Reminder: Off
Repeat: 4h
Defer: No
59
Conductivity Operation and Methods
7.5 Modify Current Method Menu for Conductivity Summary (continued)
CONDUCTIVITY OPTION
Acceptance Criteria
AVAILABLE SELECTIONS
Limits: ±1 to ±10%
Check Standard Reference Temperature: Enter temperature
Calibration Expires On Failure: Yes or No
1
Automatic temperature correction for salinity or TDS.
2
Available for conductivity and resistivity with linear temperature correction only.
3
Available for conductivity and resistivity with linear on non-linear/NaCl temperature correction only.
DEFAULT SETTING
±5%
25 °C
Off
7.5.1 Modifying a Conductivity Method
A method for conductivity can be edited when Access Control is off, or when a valid password is entered.
1. From the Conductivity Parameter Methods menu, use the
UP
and
DOWN
keys to highlight Modify Current Method. Press the
GREEN/RIGHT
key under Select.
Note: The default method cannot be modified or deleted, but can be saved with a new name (Save Current Method As) and then modified.
7.5.2 Modifying the Conductivity Parameter
Change the parameter to measure Conductivity, TDS, Salinity, or Resistivity.
1. With Parameter highlighted in the Modify Current Method menu, press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the desired parameter.
Press the
GREEN/RIGHT
key under OK.
60
Conductivity Operation and Methods
7.5.3 Modifying the Conductivity Measurement Options
Use Measurement Options to change units for conductivity upper and lower measurement limits, or temperature correction for conductivity or resistivity.
1. Use the
UP
and
DOWN
keys to highlight Measurement
Options in the Modify Current Method menu. Press the
GREEN/RIGHT
key under Select.
The available Measurement Options will vary depending on which parameter is selected.
Changing Measurement Units
The units for conductivity can be fixed or “auto-scaling”. The units for Salinity can also be changed. To change the units for either parameter:
1. Use the
UP
and
DOWN
keys to highlight Measurement Units.
Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the desired units. Press the
GREEN/RIGHT
key under OK.
Changing Measurement Limits
Upper and lower limits can be set for conductivity, salinity, TDS, and resistivity. To change limits:
1. Use the
UP
and
DOWN
keys to select Measurement Limits.
Press the
GREEN/RIGHT
key under Select.
61
Conductivity Operation and Methods
2. Use the
UP
and
DOWN
keys to select Lower Limit or Upper
Limit. Press the
GREEN/RIGHT
key under Select.
3. Set Lower Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
4. Set Upper Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
Note: Upper and lower limits only use conductivity units.
Changing Temperature Correction
Temperature correction options are available for conductivity or resistivity.
To change the Temperature Correction Options:
1. Use the
UP
and
DOWN
keys to highlight Temperature
Correction. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the temperature correction to be used. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
62
Conductivity Operation and Methods
Changing the Correction Factor
When the temperature correction is set to linear, the correction factor can be changed.
To change the Correction Factor:
1. Use the
UP
and
DOWN
keys to highlight Correction Factor.
Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to change the correction factor.
Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
Changing the Reference Temperature
The reference temperature can be changed when the temperature correction is set to linear or NaCl/non-linear.
To change the Reference Temperature:
1. Use the
UP
and
DOWN
keys to highlight Reference
Temperature. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the reference temperature to be used. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
63
Conductivity Operation and Methods
7.5.4 Modifying the Conductivity Calibration Options
Calibration Options changes the specified conductivity standard for calibration and calibration reminders.
1. Use the
UP
and
DOWN
keys to highlight Calibration Options.
Press the
GREEN/RIGHT
key under Select.
Changing the Conductivity Standard
1. Use the
UP
and
DOWN
keys to highlight Std. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the calibration standard solution to be used. Press the
GREEN/RIGHT
key under OK.
Editing the Calibration Reminder
1. Use the
UP
and
DOWN
keys to highlight Set Calibration
Reminder. Press the
GREEN/RIGHT
key under Select.
2. To turn Reminder On or Off, use the
UP
and
DOWN
keys to highlight Reminder. Press the
GREEN/RIGHT
key under Select.
64
Conductivity Operation and Methods
3. Use the
UP
and
DOWN
keys to select On or Off. Press the
GREEN/RIGHT
key under OK.
4. To set the Calibration Reminder frequency, use the
UP
and
DOWN
keys to highlight Repeat. Press the
GREEN/RIGHT
key under Select.
Note: The frequency setting is only available when the Calibration
Reminder selection is on.
5. Use the
UP
and
DOWN
keys to select the desired reminder frequency. Press the
GREEN/RIGHT
key under OK.
6. To edit when the Calibration Reminder expires, use the
UP
and
DOWN
keys to highlight Expires. Press the
GREEN/RIGHT
key under Select.
7. Use the
UP
and
DOWN
keys to select how long after the reminder the calibration will expire. Press the
GREEN/RIGHT
key under OK.
65
Conductivity Operation and Methods
Editing the Custom Calibration Standard
When Custom is selected for the conductivity calibration standard, the concentration, reference temperature, and temperature correction can be set for the calibration standard.
STANDARD VALUE
To enter the conductivity Standard Value of the custom calibration solution:
1. Use the
UP
and
DOWN
keys to highlight Standard Value. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to change the value of conductivity standard for calibration. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
REFERENCE TEMPERATURE
To enter the Reference Temperature for the custom calibration standard:
1. Use the
UP
and
DOWN
keys to highlight Reference
Temperature. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to change the Reference
Temperature for calibration. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
TEMPERATURE CORRECTION
To enter the Temperature Correction to be used with the custom calibration standard:
1. Use the
UP
and
DOWN
keys to highlight Temp Correction
(%/C). Press the
GREEN/RIGHT
key under Select.
66
Conductivity Operation and Methods
2. Use the
UP
and
DOWN
keys to change the Correction Factor.
Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
7.5.5 Modifying Conductivity Check Standard Options
Check Standard Options changes the standard solution used for the check standard, the reminder, and the acceptance criteria.
To edit the Check Standard Options, use the
UP
and
DOWN
keys to highlight Check Standards Options. Press the
GREEN/RIGHT
key under Select. The Check Standard menu appears with the following sections.
Editing the Check Standard Value
1. To edit the Check Standard value, use the
UP
and
DOWN
keys to highlight Standard Value. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the Check Standard Value to be used. Press the
GREEN/RIGHT
key under OK.
67
Conductivity Operation and Methods
Editing the Check Standard Reminder
Refer to the following steps to turn the Check Standard Reminder on or off, to set the Check Standard repeat, and to defer the
Check Standard.
1. Use the
UP
and
DOWN
keys to highlight Check Standard
Reminder. Press the
GREEN/RIGHT
key under Select.
2. To turn Check Standard Reminder on or off, use the
UP
and
DOWN
keys to highlight Reminder. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select On or Off. Press the
GREEN/RIGHT
key under OK.
4. To set the Check Standard Reminder repeat, use the
UP
and
DOWN
keys to highlight Repeat. Press the
GREEN/RIGHT
key under Select.
5. Use the
UP
and
DOWN
keys to select the desired reminder frequency. Press the
GREEN/RIGHT
key under OK.
6. To edit the Allow Defer option, use the
UP
and
DOWN
keys to highlight Allow Defer. Press the
GREEN/RIGHT
key under Select. The defer option allows the operator to continue with sample measurements instead of measuring a check standard when the reminder appears.
7. Use the
UP
and
DOWN
keys to select Yes or No. Press the
GREEN/RIGHT
key under OK.
68
Conductivity Operation and Methods
Editing the Acceptance Criteria for Check Standards
1. Use the
UP
and
DOWN
keys to highlight Acceptance Criteria.
Press the
GREEN/RIGHT
key under Select.
2. To edit Acceptance Limits, use the
UP
and
DOWN
keys to highlight Acceptance Limits. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to enter a percent range that the
Check Standard must fall within. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
4. To edit the result of a failed Check Standard, use the
UP
and
DOWN
keys to highlight Cal Expires on Failure. Press the
GREEN/RIGHT
key under Select.
5. Use the
UP
and
DOWN
keys to select Yes or No. Press the
GREEN/RIGHT
key under OK.
When set to Yes, the meter must be re-calibrated when a
Check Standard falls outside of the specified acceptance limits.
If the meter is not re-calibrated, each result will be displayed with the
CALIBRATION ?
icon and stored with a flag indicating a questionable calibration.
When set to No, the meter will operate normally in the measurement mode.
69
Visit us at www.hach.com
Section 8 LDO Operation and Methods
8.1 Taking a Dissolved Oxygen Measurement
If complete traceability is required, enter a Sample ID and
Operator ID before measuring.
Note: The default setting for Measurement Mode is “Press to Read”. To
change the mode, see Measurement Mode in section 9.5 on page 84 .
Important Note: A count down message appears on the screen thirty days before the sensor-cap expiration date. This message will continue to display until there are zero days remaining and the sensor cap must be replaced. All measurements taken after the sensor cap expiration date appear with the calibration ? icon at the top left corner of the screen.
1. Place the LDO probe into the sample.
2. Press the
GREEN/RIGHT
key under Read.
3. The display will show “Stabilizing…” and a progress bar will fill from 0 to 100% as the probe stabilizes in the sample. When the result has stabilized, the lock icon will appear and the result will be stored automatically in the data log.
4. To make another measurement, repeat this procedure.
The display will also show the temperature and pressure. If a salinity correction was entered, the correction will appear on the display.
8.2 Calibrating the LDO Probe
Important Note: Factory calibration is the default setting. Factory calibration coefficients are stored in the iButton
®
*. Each lot of LDO sensor caps is factory calibrated. Performance will vary slightly as a function of usage history. For best performance, a one-time calibration initialization can be performed when a new sensor is installed. Additional calibrations can be performed at the operator's discretion but are not required. Manually calibrating the LDO probe will require creating and modifying a new method (see
).
LDO calibration can be performed manually using one of two standards:
•
Water-saturated air (recommended). For example, use a bottle with a narrow neck such as a BOD bottle (Cat. No. 621-00).
Add a small amount (1-cm) of water to the bottle, stopper and shake vigorously for several minutes, then insert the probe.
•
A water sample with a known dissolved oxygen concentration.
The concentration must be determined by Winkler titration, or by calculation of a saturated-air water sample using existing pressure, temperature, and salinity conditions.
Note: Modify the LDO Method to use a water sample as a standard.
* iButton is a registered trademark of Maxim Integrated Products, Inc.
71
LDO Operation and Methods
To calibrate using water-saturated air:
1. Press the
BLUE/LEFT
key under Calibrate.
Note: If using the HQ40d meter with two probes, the display must be in single screen mode.
2. Dry the probe and place it in the calibration chamber.
3. Press the
GREEN/RIGHT
key under Read.
4. When the reading is stable the standard value will be highlighted on the screen and the calibrated reading value will appear on the screen. Press the
UP
key under Done.
5. The Calibration Summary will appear. Press the
GREEN/RIGHT
key under Store to accept the calibration and return to the measurement mode. The calibration is recorded in the data log.
If using the HQ40d meter, the calibration information is also sent to a PC/printer/flash memory stick if connected.
6. When the calibration is successful, the display will show OK in the upper left corner. A question mark will be displayed if the calibration has expired or if a check standard has failed or been delayed.
72
LDO Operation and Methods
8.2.1 Calibration Error- Slope Out of Range
If the calibration slope does not meet the acceptance criteria, the display will show “Slope out of range”. If this happens, allow the probe to stand in the water-saturated air for several minutes to reach equilibrium and re-press the
GREEN/RIGHT
key under Read.
8.3 Setting LDO Methods
The LDO Method menu is available via the
METER OPTIONS/PARAMETER METHODS
key when Access Control is off, or when a valid password is entered. An LDO probe must be connected to the meter to change these options.
outlines the software menu for LDO options. These options do not need to be changed if the default method is used. The
Modify Current Method submenus and default settings are described in detail in
The HQ series meters contain a default method for LDO with settings for measurement, calibration, and units. The default settings cannot be changed. The default method must be saved and then modified. Use the Save Current Method As function to save the method as a new method that can be modified.
To enter options that are different from the default settings, a new method must be entered and then modified..
Table 3 LDO Options Main Menu Summary
LDO MAIN MENU
Current Method
Save Current Method As
Modify Current Method
Delete a Method
LDO SUBMENU
Select Method
Enter New Method Name
Measurement Options
Select Units
Calibration Standard
Select Method
73
LDO Operation and Methods
8.4 Modify Current Method Menu Summary
LDO METHOD OPTION
Measurement Options
AVAILABLE SELECTIONS
Resolution
Measurement Limits
Salinity Correction
Pressure Units
Averaging Interval
Units
Calibration Standard
hPa mBar inHg mmHg
Off
30 s
60 s
90 s
3 min
5 min
0.1 Fast
0.01 Fast
0.01 Medium
0.01 Slow
Lower Limit: 0–20 mg/L
Upper Limit: 0–20 mg/L
0–70 mg/L
%
100% mg/L
DEFAULT SETTING
0.01 Medium
Lower: 0 mg/L
Upper: 20 mg/L
0 hPa
Off mg/L
100%
8.4.1 Entering a New LDO Method
A new method for LDO can be entered when Access Control is off, or when a valid password is entered.
1. Press the
OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight LDO101 Method.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to highlight Save Current Method
As. Press the
GREEN/RIGHT
key under Select.
Note: Modify Current Method and Delete a Method are not available until a new method is created.
74
LDO Operation and Methods
4. Use the
UP
and
DOWN
keys to scroll through the letters and numbers. To select a letter or number, press the
GREEN/RIGHT
key. The cursor will advance to the next space.
5. Repeat the previous step to add additional letters or numbers until the name is complete. To add a space, scroll to the blank space (between A and 9) using the
UP
and
DOWN
keys and press the
GREEN/RIGHT
key. To delete a letter or number, press the
BLUE/LEFT
key.
6. Press the
GREEN/RIGHT
key until OK replaces the Right arrow in the function bar. Select OK to complete the entry.
8.4.2 Modifying an LDO Method
A method for LDO can be edited when Access Control is off, or when a valid password is entered.
1. From the LDO Method menu, use the
UP
and
DOWN
keys to highlight Modify Current Method. Press the
GREEN/RIGHT
key under Select.
Note: The default method cannot be modified or deleted, but can be saved with a new name (Save Current Method As) and then modified.
8.4.3 Modifying LDO Measurement Options
Edit Measurement Options to change the displayed resolution, upper and lower limits, salinity correction, pressure units, or averaging interval.
1. With Measurement Options highlighted in the Modify Current
Method menu, press the
GREEN/RIGHT
key under Select.
Changing Resolution
1. With Resolution highlighted, press the
GREEN/RIGHT
key under Select.
75
LDO Operation and Methods
2. Use the
UP
and
DOWN
keys to select the desired resolution and speed of response. Press the
GREEN/RIGHT
key under OK.
Changing Measurement Limits
1. Use the
UP
and
DOWN
keys to select Measurement Limits.
Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select Lower Limit or Upper
Limit. Press the
GREEN/RIGHT
key under Select.
3. Lower Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
4. Upper Limit: Use the
UP
and
DOWN
keys to change the limit value. Use the
BLUE/LEFT
key to move to the left. Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
76
LDO Operation and Methods
Changing Salinity Correction
Correct dissolved oxygen values for high concentrations of dissolved salts by entering the sample salinity. Use a conductivity probe to measure the salinity.
1. Use the
UP
and
DOWN
keys to enter the sample salinity/salinity correction factor. Use the
BLUE/LEFT
key to move to the left.
Use the
GREEN/RIGHT
key to move to the right. When the cursor is at the far right, press the
GREEN/RIGHT
key under OK.
Changing Pressure Units
1. Use the
UP
and
DOWN
keys to select the desired pressure units. Press the
GREEN/RIGHT
key under OK.
Changing the Averaging Interval
In samples containing a high amount of air bubbles such as aeration basins, results will appear unstable or noisy. Use the averaging function to improve stability.
To select the interval for averaging results:
1. Use the
UP
and
DOWN
keys to select the desired Averaging
Interval. Press the
GREEN/RIGHT
key under OK.
Displayed results will be averaged over the selected interval.
8.4.4 Modifying the LDO Measurement Units
The meter will display both mg/L DO and % saturation in the measurement mode. To change which unit is prominent:
1. Use the
UP
and
DOWN
keys to highlight Select Units. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the units. Press the
GREEN/RIGHT
key under OK.
77
LDO Operation and Methods
8.4.5 Selecting a LDO Calibration Setting
Important Note: Factory calibration is the default setting. Factory calibration coefficients are stored in the iButton.
Water-saturated air is recommended for calibration of the HQ series meters, however a water sample with a known dissolved oxygen concentration can also be used. The dissolved oxygen concentration must be determined by Winkler titration, or by calculation of a saturated-air water sample using existing pressure, temperature, and salinity conditions.
Note: Factory calibration is the default setting. For information about manual calibration, see
.
Each lot of LDO sensor caps is factory calibrated. Performance will vary slightly as a function of usage history. For best performance, a one-time calibration initialization can be performed when a new sensor is installed. Additional calibrations can be performed at the operator's discretion but are not required. Use the Calibration
Standard option to use a water sample as the calibration standard:
1. Use the
UP
and
DOWN
keys to highlight Calibration Standard in the Modify Current Method menu. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select mg/L. Press the
GREEN/RIGHT
key under OK.
To use water-saturated air as the calibration standard:
1. Use the
UP
and
DOWN
keys to select 100%. Press the
GREEN/RIGHT
key under OK.
8.4.6 Selecting a LDO Method
Select which method to use when additional methods have been entered in the LDO Options menu.
1. Use the
UP
and
DOWN
keys to highlight Current Method.
Press the
GREEN/RIGHT
key under Select.
78
LDO Operation and Methods
2. Use the
UP
and
DOWN
keys to select the desired method.
Press the
GREEN/RIGHT
key under OK.
8.4.7 Deleting a Method
To delete an existing LDO method:
1. Use the
UP
and
DOWN
keys to highlight Delete a Method.
Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to select the desired method.
Press the
GREEN/RIGHT
key under Delete. Once a method has been deleted, it cannot be recovered.
79
Visit us at www.hach.com
Section 9 Advanced Operations
The various meter features that can be changed via the
OPTIONS
key is displayed in section 9.1
Method menu selection is a dynamic selection screen that can change depending on how many probes are attached to the meter.
The Full Access Meter Options menu is displayed when Access
Control is off, or when a valid password is entered. These options do not need to be changed if the default factory settings are used.
9.1 Meter Options Menu- Full Access
FULL ACCESS METER OPTIONS (Access Control Off or valid password entered)
RUN CHECK STANDARD
SETUP MEASUREMENT MODE
INSTRUMENT INFORMATION
ACCESS CONTROL
DISPLAY OPTIONS
SOUND
TIME
TEMPERATURE UNITS
LANGUAGE
Measure standard solution (available for pH and conductivity)
Press To Read
Interval: Duration and Interval
Continuous
Probe Information
Instrument Information
On or Off
Set Access Password
Contrast
Auto-Shutoff
Backlight
Key Press
Stability Alert
Calibration Reminder
Set Time
Time Format
Set Date
Date Format
Select Temperature Units
Select Language
9.2 Operator Meter Options Menu
OPERATOR METER OPTIONS (Access Control On)
RUN CHECK STANDARD
INSTRUMENT INFORMATION
ACCESS PASSWORD
DISPLAY OPTIONS
SOUND
Measure standard solution or pH buffer
Probe Information
Instrument Information
Enter Password
Contrast
Auto-Shutoff
Backlight
Key Press
Stability Alert
Calibration Reminder
81
Advanced Operations
9.3 Using Access Control
Access Control is used to protect parameter methods and meter setup. When Access Control is on, options for Setup Measurement
Mode, Access Control, Time, Temperature Units, and Language will not appear in the Setup menu.
The Access Control option is available in the Meter Options>Full
Access menu, which is available upon initial startup when Access
Control is set to OFF, or when Access Control is set to ON and a valid password is entered.
Important Note: If a password is specified and Access Control is set to ON, make sure the password is stored in a safe place. If the password is forgotten, the operator will be locked out of the meter.
9.3.1 Turning Access Control On
To protect parameter methods and meter setup:
1. Press the
METER OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Access Control.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to highlight Set Access
Password. Press the
GREEN/RIGHT
key under Select.
4. Use the
UP
and
DOWN
keys to scroll through the letters and numbers. To accept a letter or number, press the
GREEN/RIGHT
key. The cursor will advance to the next space.
5. Repeat the previous step to add additional letters or numbers until the password is complete. To add a space, scroll to the blank space between A and 9 using the
UP
and
DOWN
keys and press the
GREEN/RIGHT
key. To replace a letter or number, press the
BLUE/LEFT
key and re-enter the letter or number.
6. Press the
GREEN/RIGHT
key until OK replaces the right arrow in the function bar. Select OK to complete the entry.
82
Advanced Operations
7. Use the
UP
and
DOWN
keys to highlight Access Control.
Press the
GREEN/RIGHT
key under Select.
8. Use the
UP
and
DOWN
keys to select ON. Press the
GREEN/RIGHT
key under Select.
Access is now restricted. The Operator Meter Options menu will be displayed when the
OPTIONS
key is pressed.
9.3.2 Turning Access Control Off
To turn Access Control off:
1. Press the
OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Access Control.
Press the
GREEN/RIGHT
key under Select. Select Off to turn
Access Control off.
9.4 Running Check Standards
Run Check Standards verifies equipment accuracy by measuring a solution of known concentration or pH. When the probe is placed in the solution, the meter will indicate if the Check Standard passed or failed.
The meter can automatically display a reminder to measure a check standard at a specified interval with a specified acceptance criteria. These options are changed in the Parameter Method menu for each parameter.
83
Advanced Operations
9.5 Setting the Measurement Mode
Three measurement modes affect the way measurements are taken and data is stored:
PRESS TO READ:
The GREEN/RIGHT key must be pressed for each sample measurement. Each result is stored in the Data Log automatically when the Set Stability Criteria are met. The result is also sent simultaneously to any device (PC/printer/flash memory stick) that is connected to the USB/DC power adaptor (HQ40d only).
INTERVAL:
The meter measures the sample at a user defined interval for a user defined duration and stores the data in the Data Log automatically. The result is also sent simultaneously to any device
(PC/printer/flash memory stick) that is connected to the USB/DC power adaptor (HQ40d only).
CONTINUOUS:
The meter continuously measures the sample, and data can be stored manually in the Data Log. When stored, the data point is also sent simultaneously to any device (PC/printer/flash stick) that is connected to the USB/DC power adaptor (HQ40d only).
To select a measurement mode:
1. Press the
OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Setup Measurement
Mode. Press the
GREEN/RIGHT
key under Select.
3. With Mode highlighted, press the
GREEN/RIGHT
key under Select.
84
Advanced Operations
4. Use the
UP
and
DOWN
keys to select Press to Read, Interval, or Continuous. Press the
GREEN/RIGHT
key under OK.
9.5.1 Setting Auto Measurement Intervals
When using the Interval Mode, it must be specified how often and for how long the measurements will be taken. Calibration
Reminders, Auto Shut Off, and Check Standard Reminders do not interrupt interval measurements. However, a missed calibration will store readings as Cal? rather than as Cal OK. After interval measurements are completed, any missed reminders appear and
Auto Shut Off is enabled.
To specify the measurement interval and duration:
1. Select Interval as the Measurement Mode.
2. Use the
UP
and
DOWN
keys to select Duration.
Press the
GREEN/RIGHT
key under Select to display the Set
Duration screen.
3. Use the
UP
and
DOWN
keys to select the duration or total time that measurements will be taken. Press the
GREEN/RIGHT
key under OK.
4. Use the
UP
and
DOWN
keys to select Interval.
Press the
GREEN/RIGHT
key under Select to display the Set
Interval screen.
5. Use the
UP
and
DOWN
keys to select how often measurements will be taken. Press the
GREEN/RIGHT
key under OK.
85
Advanced Operations
9.5.2 Starting Interval Measurements
From the Main Measurement screen, press the
GREEN/RIGHT
key under Start to begin interval measurements. The remaining duration for the measurement is displayed in the lower right corner of the screen. The Sample ID automatically advances by number for each reading.
The auto-shutoff feature is disabled during interval measurements.
The meter goes into a standby state between readings to conserve power. Measurements are suspended when performing calibration, running check standards, or using the Meter Options menu.
Measurements resume when returning to the reading mode.
Measurements stop when the selected interval duration has passed. The auto-shutoff feature becomes active. To repeat the interval measurement, press the
GREEN/RIGHT
key under Start.
9.5.3 Preventing Data Log Overflow in Interval Reading Mode
When measurements are taken at specified intervals, each result will be stored automatically in the Data Log. The meter can store up to 500 data records. If the number of accumulated results exceeds
500, data will be replaced on a first in first out basis (FIFO). The
HQ40d meter can be connected to a PC/printer/flash memory stick to prevent loss of data.
Table 4 Suggested Combinations of Interval/Duration to
Prevent Data Log Overflow
Interval
10 seconds
30 seconds
1 minute
5 minutes
Duration
1 hour
4 hours
8 hours
24 hours
86
Advanced Operations
9.6 Viewing Instrument Information
Use the Instrument Information menu to find the serial number, software version, and model number of the meter or probes connected to the meter.
1. Press the
OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Instrument
Information. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to choose probe information or instrument information. Press the
GREEN/RIGHT
key under Select.
9.7 Setting the Display Options
Use Display Options to change the display contrast, battery saving auto-shutoff options, and the backlight option.
1. Press the
METER
OPTIONS
key. Use the
UP
and
DOWN
keys to highlight Display Options. Press the
GREEN/RIGHT
key under Select.
2. Use the
UP
and
DOWN
keys to choose Contrast,
Auto-Shutoff, or Backlight. Press the
GREEN/RIGHT
key under Select.
CONTRAST
Use the
UP
and
DOWN
keys to adjust the contrast of the display.
Zero is the lightest setting and 9 is the darkest setting. Press the
GREEN/RIGHT
key under OK to accept the setting.
87
Advanced Operations
AUTO-SHUTOFF
Auto-shutoff maximizes battery life and is not active when the meter is connected to AC power or in Interval Reading Mode.
Use the
UP
and
DOWN
keys to select a time period after which the meter will shut off if no keys are pressed. Press the
GREEN/RIGHT
key under OK.
BACKLIGHT
The display is illuminated when the
BACKLIGHT
key is pressed. To maximize battery life, set a time period after which the backlight will automatically turn off if no key is pressed.
Use the
UP
and
DOWN
keys to select a time period after which the backlight will shut off if no keys are pressed. Press the
GREEN/RIGHT
key under OK.
9.8 Setting the Sound Options
The meter can make an audible sound when a key is pressed, when stability is reached, or when the calibration reminder is due.
Note: For the HQ40d only, the meter makes an audible sound when it begins transferring data to a flash memory stick or printer and again when the data transfer is complete. To use this feature, select Stability Alert in the Sound Options menu.
To turn the sound on or off:
1. Press the
METER OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Sound. Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to highlight Key Press, Stability
Alert, or Cal Reminder. Press the
BLUE/LEFT
key under the check mark. Multiple items can be selected.
88
Advanced Operations
9.9 Setting the Date and Time
Refer to
section 4.4 on page 22 for more information.
9.10 Changing the Temperature Units
To select degrees Celsius or Fahrenheit:
1. Press the
METER OPTIONS
key.
2. Use the
UP
and
DOWN
keys to highlight Temperature Units.
Press the
GREEN/RIGHT
key under Select.
3. Use the
UP
and
DOWN
keys to select Celsius or Fahrenheit.
Press the
GREEN/RIGHT
key under OK.
9.11 Language
Refer to
section 4.3 on page 22 for more information.
89
Visit us at www.hach.com
Section 10 Maintenance
DANGER
Only qualified personnel should conduct the tasks described in this section of the manual.
10.1 General Meter Cleaning
The meter is designed to be maintenance-free. If the meter is dirty, wipe the surface with a damp cloth. Use a cotton-tipped applicator to clean or dry the connectors if they get wet.
10.2 General Probe Cleaning
For information about cleaning the probes, see the instruction sheet that comes with the probe.
10.3 Battery Replacement
10.4 Updating Instrument Software
For the HQ40d, a flash stick connected to the USB/DC power adapter can be used to update the instrument software or transfer data from the instrument. The HQ11d, HQ14d, and the HQ30d can only use the USB update function to update the instrument software
(data transfer is not possible) if necessary.
Note: The manufacturer will determine if an update to the software is necessary to make sure that the HQd Series correctly functions.
91
Visit us at www.hach.com
Section 11 Parts and Accessories
11.1 Parts
Description
LDO Probe, standard, with 1 m cable
LDO Probe, standard, with 3 m cable
LDO Probe, rugged, with 5 m cable
LDO Probe, rugged, with 10 m cable
LDO Probe, rugged, with 15 m cable
LDO Probe, rugged, with 30 m cable pH Gel Probe, standard, with 1m cable pH Gel Probe, standard, with 3m cable pH Liquid Probe, standard, with 1m cable pH Liquid Probe, standard, with 3m cable pH Gel Probe, rugged, with 5m cable pH Gel Probe, rugged, with 10m cable pH Gel Probe, rugged, with 15m cable pH Gel Probe, rugged, with 30m cable
Conductivity Probe, standard with 1m cable
Conductivity Probe, standard with 3m cable
Conductivity Probe, rugged with 5m cable
Conductivity Probe, rugged with 10m cable
Conductivity Probe, rugged with 15m cable
Conductivity Probe, rugged with 30m cable
11.2 Accessories
Description
Probe Depth Marker (Rugged Cable Markers)
LDO Sensor Cap, Replacement (includes iButton, cap seal, and probe-tip o-ring)
Glove Kit
Field Kit
Standard Probe Holder
Rugged Shroud Kit
Color Coded Probe Clips (five colors, two clips in each color)
AC-DC Power Supply 115 V
AC-DC Power Supply 230 V
USB/DC Adapter
USB Cable, 6 ft (1.8 m), Type A Male, Type B Male
Keyboard (QWERTY)
BOD Bottle (300 mL)
BOD Stirrer/Funnel Accessory Kit (US)
BOD Stirrer/Funnel Accessory Kit (EU)
Quantity
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Catalog Number
LDO101-01
LDO101-03
LDO101-05
LDO101-10
LDO101-15
LDO101-30
PHC101-01
PHC101-03
PHC301-01
PHC301-03
PHC101-05
PHC101-10
PHC101-15
PHC101-30
CDC401-01
CDC401-03
CDC401-05
CDC401-10
CDC401-15
CDC401-30
Quantity
5
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Catalog Number
58286-10
58112-00
58287-00
52258-00
58294-00
58259-00
58184-00
58263-00
58311-00
58134-00
59240-00
LZV582
621-00
58266-00
58267-00
93
Parts and Accessories
11.3 Consumables
Description
IUPAC Series Certified pH Standards (Buffers)
1
:
pH 1.679 ± 0.010 @ 25 °C pH 4.005 ± 0.010 @ 25 °C pH 7.000 (Radiometer Analytical) ± 0.010 @ 25 °C pH 10.012 ± 0.010 @ 25 °C
Color-coded pH Standards (Buffers):
pH 4.01 ±0.02 @ 25 °C pH 7.00 ±0.02 @ 25 °C pH 10.01 ±0.02 @ 25 °C pH 4.01 ±0.02 @ 25 °C pH 7.00 ±0.02 @ 25 °C pH 10.01 ±0.02 @ 25 °C pH 4.01 ±0.02 @ 25 °C pH 7.00 ±0.02 @ 25 °C pH 10.01 ±0.02 @ 25 °C
Certified Conductivity Standards
1
:
KCl, 1 Demal, 111.3 mS/cm ± 0.5% @ 25 °C
KCl, 0.1 Demal, 12.85 mS/cm ± 0.35% @ 25 °C
KCl, 0.01 Demal, 1408 µS/cm ± 0.5% @ 25 °C
NaCl, 0.05%, 1015 µS/cm ± 0.5% @ 25 °C
KCl Conductivity Standards:
0.1 Molar KCl, 12.88 mS/cm @ 25 °C
0.01 Molar KCl, 1413 µS/cm @ 25 °C
0.001 Molar KCl, 148 µS/cm @ 25 °C
NaCl Conductivity Standards:
180 µS/cm @ 25 °C
1000 µS/cm @ 25 °C
18.00 mS/cm @ 25 °C
BOD Consumables:
Nitrification Inhibitor (TCMP) - (200 tests)
Nitrification inhibitor (TCMP) - (2500 tests)
Nitrification Inhibitor (ATU)
BOD Seed (50 tests)
BOD Standard Solution, 300 mg/L GGA, 10-mL Voluette
®
Ampules
Nutrient Buffer Pillow, 0.5 mL (for preparing 300 mL of dilution water)
Nutrient Buffer Pillow, 3 mL (for preparing 3 L of dilution water)
Nutrient Buffer Pillow, 4 mL (for preparing 4 L of dilution water)
Nutrient Buffer Pillow, 6 mL (for preparing 6 L of dilution water)
Nutrient Buffer Pillow, 19 mL (for preparing 19 L of dilution water)
Buffer Solution, APHA, for BOD, pH 7.2, phosphate type
Calcium Chloride Solution, APHA
Magnesium Sulfate Solution, APHA
94
Quantity
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
4 L
4 L
4 L
20 L
20 L
20 L
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
500 mL
100 mL
100 mL
100 mL
35 g
500 g
50 g
50 capsules
25/pkg
50/pkg
50/pkg
50/pkg
50/pkg
25/pkg
1 L
1 L
1 L
Catalog Number
S11M001
S11M002
S11M004
S11M007
22834-49
22835-49
22836-49
22834-56
22835-56
22836-56
22834-61
22835-61
22836-61
S51M001
S51M002
S51M003
S51M004
C20C250
C20C270
C20C280
23075-42
14400-42
23074-42
2533-35
2533-34
28454-25
24712-00
14865-10
14160-66
14861-66
24364-66
14862-66
14863-98
431-53
428-53
430-53
Parts and Accessories
11.3 Consumables (continued)
Description
Ferric Chloride Solution, APHA
Miscellaneous:
pH Filling Solution (for PHC301), 3M KCl, sat'd with AgCl pH Electrode Storage Solution
1
Certified standards ship with certificates for traceability to Standard Reference Materials
Quantity
1 L
30 mL
500 mL
Catalog Number
429-53
28417-00
27565-49
95
Visit us at www.hach.com
Section 12 How to Order
U.S.A. Customers
By Telephone:
6:30 a.m. to 5:00 p.m. MST
Monday through Friday
(800) 227-HACH (800-227-4224)
By Fax:
(970) 669-2932
By Mail:
Hach Company
P.O. Box 389
Loveland, Colorado 80539-0389 U.S.A.
Ordering information by e-mail: [email protected]
Information Required
•
Hach account number (if available) • Billing address
•
Your name and phone number
•
Shipping address
•
Purchase order number
•
Catalog number
•
Brief description or model number
•
Quantity
International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send an e-mail to: [email protected] or contact:
Hach Company World Headquarters; Loveland, Colorado, U.S.A.
Telephone: (970) 669-3050; Fax: (970) 669-2932
Technical and Customer Service (U.S.A. only)
Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use.
Specialists in analytical methods, they are happy to put their talents to work for you.
Call 1-800-227-4224 or e-mail [email protected]
97
Visit us at www.hach.com
Section 13 Repair Service
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the Hach Service
Center serving your location.
In the United States:
Hach Company
Ames Service
100 Dayton Avenue
Ames, Iowa 50010
(800) 227-4224 (U.S.A. only)
FAX: (515) 232-3835
In Canada:
Hach Sales & Service Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only)
Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: [email protected]
In Latin America, the Caribbean, the Far East,
Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters,
P.O. Box 389
Loveland, Colorado, 80539-0389 U.S.A.
Telephone: (970) 669-3050
FAX: (970) 669-2932
E-mail: [email protected]
99
Visit us at www.hach.com
Section 14 Certification
Product Safety (power supply only)
115/230 VAC External Power Supply
Certified to CSA and Listed to UL safety standards (cULus mark),
TUV-GS & CE marked per 73/23/EEC
EMI Immunity
Hach Company certifies this instrument was tested thoroughly, inspected and found to meet its published specifications when it was shipped from the factory.
The HQ Series Portable Meters have been tested and are certified as indicated to the following instrumentation standards:
Instrument tested with the external Power Supply:
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement, control and laboratory use— EMC requirements)
Supporting test records by Hach Company, certified compliance by
Hach Company.
Standards include:
IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge
Immunity (Criteria B)
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF
Electro-Magnetic Field Immunity (Criteria B)
IEC 1000-4-4:1995 (EN 61000-4-5:1995) Electrical Fast
Transients/Burst (Criteria B)
IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances
Induced by RF Fields (Criteria A)
IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short
Interruptions (Criteria B)
Additional immunity Standard/s include:
ENV 50204:1996 Radiated Electro-Magnetic Field from Digital
Telephones (Criteria B)
Emissions
Instrument tested with the external Power Supply:
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement, control and laboratory use—EMC requirements)
Class “B” emission limits all models except HQ40D Multi-Portable
Meter with “2” LDO probes. The HQ40D when used with “2” LDO probes meets only Class “A” limits. Supporting test records by
Hewlett Packard, Fort Collins, Colorado Hardware Test Center
(A2LA # 0905-01) and certified compliance by Hach Company.
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by Electrical
Equipment
EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical Equipment
101
Certification
Additional Emissions Standard/s include
EN 55011 (CISPR 11), Class “B” emission limits all models except
HQ40D Multi-Portable Meter with “2” LDO probes. The HQ40D when used with “2” LDO probes meets only Class “A” limits.
CANADIAN INTERFERENCE-CAUSING EQUIPMENT REGULATION
IECS-003: Class “A” emission limits. Supporting test records by
Hewlett Packard, Fort Collins, Colorado Hardware Test Center
(A2LA # 0905-01) and certified compliance by Hach Company.
This Class “A” digital apparatus meets all requirements of the
Canadian Interference- Causing Equipment Regulations.
Cet appareil numérique de la classe “A” respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
FCC PART 15: Class emission A limits
Supporting test records by Hewlett Packard, Fort Collins, Colorado
Hardware Test Center (A2LA # 0905-01) and certified compliance by Hach Company.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class “A” digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their own expense. The following techniques of reducing the interference problems are applied easily.
1. Disconnect the external power supply from the meter and/or remove one of the meter’s batteries to verify that meter is or is not the source of the interference.
2. Move the meter and it’s power supply away from the device receiving the interference.
3. Reposition the receiving antenna for the device receiving the interference.
4. Try combinations of the above.
102
Section 15 Limited Warranty
Hach Company warrants its products to the original purchaser against any defects that are due to faulty material or workmanship for a period of one year from date of shipment unless otherwise noted in the product manual.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Limitations
This warranty does not cover:
• Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
• Damage caused by misuse, neglect, accident or improper application or installation
• Damage caused by any repair or attempted repair not authorized by Hach Company
• Any product not used in accordance with the instructions furnished by Hach Company
• Freight charges to return merchandise to Hach Company
• Freight charges on expedited or express shipment of warranted parts or product
• Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
103
Visit us at www.hach.com
A
AC power supply
Access Control
passwords
turning off
turning on
accessories
auto-shutoff, batteries
,
averaging interval, LDO methods
B
Backlight options
batteries
auto-shutoff
,
features
installing
beep options
box contents
C cables, part numbers
calibrating probes
conductivity
LDO
pH
Slope out of range error
,
Standard not recognized error
Calibration
Calibration Log
displaying
calibration options
conductivity
,
LDO
pH
calibration reports
printing
viewing
Canadian safety regulations
Celsius, settings
certification
Check Standards
conductivity method options
,
deferring (conductivity)
deferring (pH)
log, displaying
pH method options
running manually or automatically (conductivi-
ty)
running manually or automatically (pH)
settings
cleaning
components, unpacking
computers, transferring data to
,
conductivity
calibrating probe
deferring Check Standards
measuring
probe specifications
running Check Standards
conductivity methods
calibration options
,
Check Standards options
,
limits
measurement options
,
menu options
modifying
parameters
parameters, changing
reference temperature
temperature correction
units
connectors, probe
contact information
ordering parts
repair service
support
Continuous measurement mode
contrast, display
Current
customer support
D data deleting
overflow, preventing
Report Options
viewing probe
viewing stored
data transfer
archiving and exchanging methods
features
to computer
to flash stick or printer
date, setting
deferring Check Standards conductivity
pH
deleting methods
Index
105
Index
Operator ID
Sample IDs
stored data
Display Options
display screen
dissolved oxygen, measuring
See also LDO methods; LDO probe
dual display screen
E
e-mail customer support
EMI Immunity
Emissions Standards
end-of-life waste disposal
errors, probe calibration
Slope out of range
,
Standard not recognized
European Union regulations
F
Fahrenheit, settings
FCC statement
features
flash sticks, transferring data to
,
I installing batteries
Instrument Information, viewing
Interval measurement mode
auto measurement intervals, setting
data log overflow
selecting
starting measurements
K
key press options
keypad
L
labels, precautionary
language, selecting
LDO methods averaging interval
calibration options
limits
measurement options
modifying
new, creating
pressure units
resolution
salinity correction
units
LDO probe
calibrating
106
measuring dissolved oxygen
specifications
limits conductivity
LDO
pH
logs
Calibration
Check Standards
data overflow
Reading
Report Options
luminescent dissolved oxygen. See LDO methods;
LDO probe
M maintenance
measurement modes, selecting
measurement options conductivity
,
LDO
pH
measuring
conductivity, salinity, resistivity, or TDS
dissolved oxygen
pH
meter cleaning
methods archiving and exchanging
deleting
pH
selecting
model number
models
O
on/off key
Operator ID creating
deleting
displaying
selecting
ordering parts
overflow, data
oxygen, measuring dissolved
See also LDO methods; LDO probe
P
packing slip
part number, AC power supply
parts numbers
Index
ordering
passwords, Access Control
PCs, transferring data to
,
pH calibrating probe
deferring Check Standards
measuring
probe specifications
running Check Standards
pH methods calibration options
Check Standard options
creating new
limits
measurement options
modifying
resolution
units
phone numbers ordering parts
repair service
support
power on/off key
power requirements
power supply, AC
,
Press to Read measurement mode
pressure units, LDO methods
printers, transferring data to
,
probes calibrating conductivity
calibrating LDO
calibrating pH
cleaning
connectors
display screen
Instrument Information
measuring conductivity, salinity, resistivity, or
TDS
measuring dissolved oxygen
measuring pH
part numbers
Slope out of range error
,
specifications
Standard not recognized error
viewing data
R
Reading Log
displaying
rechargeable batteries
recycling
reference temperature, conductivity methods
reminders, calibration conductivity
pH
repair
Report Options
resistivity, measuring
resolution
LDO methods
pH method
Run Check Standards conductivity
pH
settings
S
safety information
,
salinity
correction, LDO methods
measuring
Sample ID creating
deleting
Reading Log
selecting
screen, display
sending data features
to computer
to flash stick or printer
serial number
,
service
single display screen
Slope out of range error
,
softkeys
software updating
sound options
specifications
Standard not recognized calibration error
standards. See Check Standards
stored data deleting
viewing
support
symbols, safety
T
TDS, measuring
technical support
107
Index
telephone numbers ordering parts
repair service
support
temperature correction, conductivity methods
specifications
units, setting
time, setting
transferring data
features
to computer
to flash stick or printer
troubleshooting probe calibration
Slope out of range error
,
Standard not recognized error
U units
conductivity methods
LDO methods
pH methods
temperature, setting
unpacking instrument components
updating software
USB connector
USB update function
user interface
V
version number
Viewing
W warranty
waste disposal
waterproof features
108
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 7 Section 1 General Information
- 7 1.1 Safety Information
- 7 1.1.1 Use of Hazard Information
- 7 1.1.2 Precautionary Labels
- 8 1.2 Product Overview
- 8 1.3 Meter Description
- 9 Section 2 Specifications
- 11 Section 3 Installation
- 11 3.1 Unpacking the Instrument
- 12 3.2 AC Power and Batteries
- 12 3.2.1 Battery Power
- 14 3.2.2 AC Power
- 14 3.3 Turning the Meter On and Off
- 14 3.4 Probe Connection
- 16 3.5 Data Transfer (HQ40d only)
- 19 Section 4 System Start Up
- 19 4.1 Basic Start Up Overview
- 20 4.2 Meter User Interface and Navigation
- 20 4.2.1 Keypad Description
- 21 4.2.2 Display Description (Single and Dual)
- 21 4.2.2.1 Using Single Screen Mode
- 21 4.2.2.2 Using Dual Screen Mode (HQ40d only)
- 22 4.3 Selecting the Language
- 22 4.4 Setting the Date and Time
- 25 Section 5 Standard Operation
- 25 5.1 Setting the Sample and Operator Identification
- 25 5.1.1 Sample ID
- 25 5.1.1.1 Creating a New Sample ID
- 26 5.1.1.2 Selecting a Sample ID
- 26 5.1.1.3 Deleting a Sample ID
- 27 5.1.2 Operator ID
- 27 5.1.2.1 Creating a New Operator ID
- 27 5.1.2.2 Selecting an Operator ID
- 28 5.1.2.3 Deleting an Operator ID
- 28 5.2 Using the Data Log
- 28 5.2.1 Storing Data
- 28 5.2.2 Viewing Stored Data
- 29 5.2.3 Viewing Probe Data
- 30 5.2.4 Deleting Data
- 31 5.3 Transferring Data (HQ40d meter only)
- 31 5.3.1 Sending Data to a Printer or Flash Memory Stick
- 31 5.3.1.1 Selecting Report Options
- 32 5.3.1.2 Printing Calibration Data
- 33 5.3.2 Sending Data to a Computer
- 34 5.4 Viewing Data Log Reports (HQ40d only)
- 34 5.4.1 Report Names
- 34 5.4.2 Sample Results
- 34 5.4.2.1 Basic Reports
- 35 5.4.2.2 Advanced Reports
- 35 5.4.2.3 Total Reports
- 36 5.4.3 Calibration Results
- 37 5.4.4 Check Standard Results
- 38 5.5 Viewing Calibration Reports (HQ40d only)
- 38 5.5.1 Current Calibration Reports
- 39 5.5.2 Calibration History Reports
- 40 5.6 Archiving and Exchanging Methods (HQ40d only)
- 41 Section 6 pH Operation and Methods
- 41 6.1 Calibrating the pH Probe
- 42 6.1.1 Calibration Errors
- 42 6.1.1.1 Standard Not Recognized
- 42 6.1.1.2 Slope Error
- 43 6.2 Taking a pH Measurement
- 43 6.3 Running Check Standards Manually or Automatically
- 44 6.3.1 Deferring a Check Standard
- 45 6.4 Setting the pH Method
- 45 6.5 Modify Current Method Menu for pH Summary
- 46 6.5.1 Modifying pH Methods
- 47 6.5.2 Modifying the pH Measurement Options
- 49 6.5.3 Modifying the pH Calibration Options
- 52 6.5.4 Modifying the pH Check Standard Options
- 54 6.5.5 Modifying the pH Measurement Units
- 55 Section 7 Conductivity Operation and Methods
- 55 7.1 Calibrating the Conductivity Probe
- 56 7.2 Taking a Conductivity, Salinity, Resistivity, or TDS Measurement
- 57 7.3 Running Check Standards Manually or Automatically
- 58 7.3.0.1 Deferring a Check Standard
- 58 7.4 Setting the Conductivity Method
- 59 7.5 Modify Current Method Menu for Conductivity Summary
- 60 7.5.1 Modifying a Conductivity Method
- 60 7.5.2 Modifying the Conductivity Parameter
- 61 7.5.3 Modifying the Conductivity Measurement Options
- 64 7.5.4 Modifying the Conductivity Calibration Options
- 67 7.5.5 Modifying Conductivity Check Standard Options
- 71 Section 8 LDO Operation and Methods
- 71 8.1 Taking a Dissolved Oxygen Measurement
- 71 8.2 Calibrating the LDO Probe
- 73 8.2.1 Calibration Error- Slope Out of Range
- 73 8.3 Setting LDO Methods
- 74 8.4.1 Entering a New LDO Method
- 74 8.4 Modify Current Method Menu Summary
- 75 8.4.2 Modifying an LDO Method
- 75 8.4.3 Modifying LDO Measurement Options
- 77 8.4.4 Modifying the LDO Measurement Units
- 78 8.4.5 Selecting a LDO Calibration Setting
- 78 8.4.6 Selecting a LDO Method
- 79 8.4.7 Deleting a Method
- 81 Section 9 Advanced Operations
- 81 9.1 Meter Options Menu- Full Access
- 81 9.2 Operator Meter Options Menu
- 82 9.3 Using Access Control
- 82 9.3.1 Turning Access Control On
- 83 9.3.2 Turning Access Control Off
- 83 9.4 Running Check Standards
- 84 9.5 Setting the Measurement Mode
- 85 9.5.1 Setting Auto Measurement Intervals
- 86 9.5.2 Starting Interval Measurements
- 86 9.5.3 Preventing Data Log Overflow in Interval Reading Mode
- 87 9.6 Viewing Instrument Information
- 87 9.7 Setting the Display Options
- 88 9.8 Setting the Sound Options
- 89 9.9 Setting the Date and Time
- 89 9.10 Changing the Temperature Units
- 89 9.11 Language
- 91 Section 10 Maintenance
- 91 10.1 General Meter Cleaning
- 91 10.2 General Probe Cleaning
- 91 10.3 Battery Replacement
- 91 10.4 Updating Instrument Software
- 93 Section 11 Parts and Accessories
- 93 11.1 Parts
- 93 11.2 Accessories
- 94 11.3 Consumables
- 97 Section 12 How to Order
- 99 Section 13 Repair Service
- 101 Section 14 Certification
- 103 Section 15 Limited Warranty
- 105 Index