Standard Operation

5.5 Viewing Calibration Reports (HQ40d only)

A report can be printed for current calibration information or

calibration history as described in section 5.3.1.2 on page 32 .

5.5.1 Current Calibration Reports

A Current Calibration Report contains two lines of information.

Figure 15 details the information available in this type of report.

Figure 15 Current Calibration Report

1

Report Type (IC = Current Calibration)

2

Parameter Type (pH, LDO, CDC, etc.)

3

Calibration Time: prefaced by "CAL" and displayed as hh:mm in 24 h format

4

Calibration Date (DD-MM-YY or user-defined format)

5

Calibration Slope

6

Offset: contents vary depending on type of parameter being read and configuration of specific method. May be blank.

7

r

2: contents vary depending on type of parameter being read, configuration of specific method and number of calibration standards read. May be blank.

8

Calibration Operator ID: user-defined Operator ID of person who performed this calibration. Displays "- - -" if undefined.

9

Method Name: user-defined Method Name used to take this reading

10 Probe Model Name

11 Probe Serial Number: prefaced by "<" or ">" to indicate which channel the probe was connected to during a dual reading.