baxter Auto Syringe AS50 Operator's Manual
Advertisement
Advertisement
Operator’s Manual
Auto Syringe AS50
Model AS50 Infusion Pump
Warning
There are risks associated with using anything other than the recommended IV extension sets with this device. Sets designated for use with this device are identified in the section “Simplified Instructions”. Baxter’s warranty on this device will be null and void and Baxter will assume no responsibility for incidents which may occur if the product is not used in accordance with product labeling. See the section “Precautions” for a complete list of warnings and cautions.
Computer Software Copyrights
©Copyright 1999, Baxter Healthcare Corporation. All rights reserved.
For use only by Baxter Healthcare Corporation. The software contains proprietary information belonging to Baxter
Healthcare Corporation. The software must not be reproduced or disclosed to others without prior written approval.
Any unauthorized use of this information may subject the user to substantial liability.
Documentation Copyrights
Duplication or distribution of this manual and any information contained within is strictly prohibited without the express written permission of Baxter. This manual and any information contained within, may not be reproduced, distributed, or transmitted in any form, or by any means, for any purpose, without the express written permission of
Baxter. To order additional copies of this manual, or other related manuals, contact your local Baxter Service Center.
Disclaimer
The information in this document has been carefully examined and is believed to be entirely reliable. However, no responsibility is assumed for inaccuracies. Furthermore, Baxter reserves the right to make changes to any products herein to improve readability, function, or design. Baxter does not assume any liability arising out of the application or use of any product or circuit described herein; neither does it cover any license under its patent rights nor the rights of others.
Trademark Information
Baxter , the Baxter wordmark, Micro-Volume , and Auto Syringe are trademarks of Baxter International Inc. in the
U.S. Other trademark applications pending. Terumo is a trademark of Terumo Medical Corporation. ASCEND, TOR, and Hi-Tor Plus are trademarks of Huntington Professional Products. LpH SE and Septisol are trademarks of
Steris Corporation. Cidex 7 is a trademark of Johnson & Johnson. Other product names appearing in this manual are the property of their respective owners.
Table of Contents
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Notations Used in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Device Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Special Precautions for Epidural Administration. . . . . . . . . . . . . . . . . . . . . . . . . . 7
Simplified Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Programming the Continuous Infusion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Programming the Timed Infusion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
The Status Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
The Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
The Status Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Operating States. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Run State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Data Display and Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Protocol Library Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Bypassing the Protocol Library Feature. . . . . . . . . . . . . . . . . . . . . . . . . . 22
Serial Communication Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Infusion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Continuous Infusion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Timed Infusion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Syringe Data Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Infusion Rate Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Deliverable Volumes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Detailed Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Preliminary Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Manual Purge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Mount the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Turn the Pump On and Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Check the Lamp Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Using Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Programming mL/hr and mL/min Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Programming Units/hr, Units/min, mUnits/hr, and mUnits/min Modes. . . 35
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page i
Programming mg/hr, mg/min, mcg/hr, and mcg/min Modes . . . . . . . . . . 36
Programming mg/kg/hr, mg/kg/min, mcg/kg/hr, and mcg/kg/min Modes 37
Programming CUSTOM DILUTION Mode. . . . . . . . . . . . . . . . . . . . . . . . 39
Programming SINGLE DOSE Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Programming MANUAL SCHEDULE Mode . . . . . . . . . . . . . . . . . . . . . . 42
Programming AUTO SCHEDULE Mode . . . . . . . . . . . . . . . . . . . . . . . . . 45
General Programming Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Data Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Field Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Editing in Standby State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Programming Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Editing in Run State. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Edit Cancellation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Number Out of Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Data Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Additional Functions and Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Purge Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Purge Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Purge Delivery Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Bolus Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Bolus Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Bolus Size Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Bolus Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Bolus Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Bolus Programming Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Bolus Delivery Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Bolus Review. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Bolus Editing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Repeating a Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Changing the Syringe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Lock and Auto Lock Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
The Total Delivered Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Expanded Data Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Backlight Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Configuration Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Date and Time Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Software and Configuration Version Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Reviewing Configuration Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Configuration Review Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Configurable Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Protocol Library Group ( VIEW LIBRARY ) . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Modes Group ( VIEW MODES ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Page ii Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Syringes Group ( VIEW SYRINGES ). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Defaults Group ( VIEW DFAULTS ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Miscellaneous Group ( VIEW MISC ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Alerts Group ( VIEW ALERTS ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Identifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Terminating Configuration Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Custom Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Factory Standard Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Protocol Libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Pump Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Charging the Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
IV Pole Clamp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
IV Pole Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Routine Cleaning and Disinfection Techniques . . . . . . . . . . . . . . . . . . . . 71
Conditions Requiring Service Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Daily Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Display, Battery, and Syringe Sensor Check . . . . . . . . . . . . . . . . . . . . . . 75
Keypad Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Battery Voltage Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
In Case of Difficulty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Alerts and Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Alerts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Failsafe Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Warranty and Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Service Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page iii
This page is intentionally blank.
Page iv Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
About This Manual
Chapter 1
The Auto Syringe AS50 Infusion Pump Operator’s Manual is intended for use by trained healthcare professionals familiar with infusions and infusion pump procedures.
Notations Used in this Manual
Note: Important information is preceded by the word “Note” in boldface type.
!
BOLD ITALIC TEXT
BOLD TEXT
Denotes text prompts as they appear on the pump’s status display. For example, ML/HR is the text that appears on the bottom line of the status display when selecting the mL/hr infusion mode.
Denotes the keys on the pump’s keypad. For example, the text “press 123 , CONFIRM ” means: press the keys labeled
“1,” “2,” and “3,” and then press the key labeled “ CONFIRM .”
(hh:mm) Denotes hours and minutes. The left two digits represent hours, and the right two digits represent minutes.
Important precautions are denoted by an exclamation point enclosed in a triangle, as shown below:
IMPORTANT : Read and understand all operating instructions and the “Precautions” on page 5 before using the Auto Syringe AS50 Infusion Pump.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 1
About This Manual
This page is intentionally blank.
Page 2 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Device Description
Chapter 2
07-19-B3-430
The Auto Syringe AS50 Infusion Pump (hereinafter referred to as “the pump”) is designed to meet the fluid and drug delivery requirements of today’s changing clinical environment.
It provides accurate, continuous or intermittent infusion of intravenous solutions, drug solutions, whole blood, and packed red blood cells.
The pump is indicated for infusion via intravenous (IV), intra-arterial (IA), epidural, or subcutaneous routes of administration. It can be piggybacked into an ongoing infusion line to deliver secondary solutions automatically, precisely, and economically. Infusion rates are programmable from 0.01 to 438 mL/hr.
The pump accepts standard disposable syringes from 1 mL to 60 mL in size. A numeric keypad simplifies programming and makes the pump easy to use. During programming, the status display prompts the user to enter data about the syringe and the infusion.
While the pump is running, the status display shows information about the infusion.
Safety and effectiveness are reinforced by pre-programmable bolus operation, titration of a dose without interruption of fluid flow, and easily understood alarm and alert messages.
The pump includes an optional serial communication feature, which provides remote computer control and monitoring capabilities. A standard 9-pin connector allows an
RS-232 communication link between the pump and the host computer. The “Auto Syringe
AS50 Infusion Pump Programmer’s Manual” describes the use of this feature.
The pump can be custom-configured to select key features that meet specific requirements. Configurable options include:
● Audio range
● Auto default to previous mode
● Backlight
● Infusion mode selections
● Keypad auto lock
● Maximum infusion rates
● Occlusion pressure sensitivity
● Protocol library
● Remote computer control
● Syringe manufacturer, type and size selections
Auto Syringe AS50 Infusion Pump Operator’s Manual Page 3
Device Description
The selected options can easily be reviewed and the chosen configuration can be changed to meet new or different requirements.
The pump operates on an internal rechargeable battery pack and can also be operated while attached to a “900-Series” battery charger (model number CAS40-01-900) or to a
Multi-Charger (product code 1M8563).
The pump is supplied with a pole clamp and a built-in IV pole loop. The pump can also be used on a tabletop.
Information on writing custom software for the pump is available in the Auto Syringe AS50
Infusion Pump Programmer’s Manual. Information on troubleshooting and repair of the pump is available in the Auto Syringe AS50 Infusion Pump Service Manual.
Page 4 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
07-19-B3-430
Precautions
Chapter 3
!
!
!
!
Federal (USA) law restricts this device to sale by or on the order of a physician.
The charger port is to be used only with “900 Series” chargers or other accessories that are labeled specifically for use with Auto Syringe AS40A or AS50 Series Infusion Pumps, including the 1M8563 Multi-Charger.
CAUTION: Refer all service, repair, and calibration to qualified technical personnel.
CAUTION: Do not operate the Auto Syringe AS50 Infusion Pump in the presence of flammable anesthetics, oxygen-enriched, or explosive atmospheres.
● Though the factory-supplied configuration settings are suitable for most therapies, the operator and hospital professionals should verify that the pump’s settings are appropriate for the clinical application.
● The Protocol Library feature is a configuration option developed to facilitate infusion programming. Before using a protocol for any specific drug, refer to the full prescribing information supplied by the drug manufacturer.
● As with all medical electronic equipment, care must be exercised to avoid exposing this device to powerful sources of electromagnetic interference.
● Do not expose the pump to X-rays, gamma rays, or other ionizing radiation, or to strong electric or magnetic fields.
●
●
Do not autoclave, steam sterilize, EtO sterilize, or subject the pump or charger to temperatures in excess of 104 ° F (40 ° C).
Do not use the following chemicals on the device, as they will damage the front panel: acetone, acetoaldehyde ammonia, benzene, hydroxytoluene, methylene chloride, and ozone, Do not use cleaners containing n-alkyl dimethyl ethylbenzyl ammonium chloride unless they appear in the list of recommended cleaners in the Pump
Maintenance section of this manual.
● Do not use hard or sharp objects on the keypad.
Auto Syringe AS50 Infusion Pump Operator’s Manual Page 5
Precautions
● If a LINE OCCLUDED alarm occurs, RELIEVE THE RESIDUAL PRESSURE IN THE
SYRINGE BY RELEASING THE PLUNGER DRIVER. If the pressure is not relieved prior to clearing the occlusion, an unintentional bolus delivery may occur when the blockage is cleared. Check for kinked tubing, clogged catheter, closed stopcocks, etc.
Check for foreign material preventing movement of the pump mechanism.
● Be sure to PURGE THE SYSTEM OF ALL AIR BEFORE ADMINISTERING ANY
MEDICATION. Failure to follow this normal infusion procedure could precipitate serious consequences.
● Remember that the volume of fluid contained in the connecting tubing is a residual amount and cannot be infused. Allow for this extra volume of fluid when initially filling the syringe.
● In MANUAL SCHEDULE mode, START must be pressed to deliver each dose and continue the infusion regimen. Pressing any other key silences the audio portion of the DOSE DUE alarm and leaves the pump in Standby state, without initiating a dose delivery.
● Although extreme care has been taken in assembly, component selection, and quality control during manufacture of the Auto Syringe AS50 Infusion Pump, routine checks and maintenance procedures must be performed so that patients receiving infusions can be assured of trouble-free operation.
● Caution must be exercised in the selection of drugs intended to be delivered via any infusion pump. If the drug contained in the syringe will be exposed to extreme environmental conditions for prolonged time periods, IT IS IMPORTANT TO SELECT
DRUGS THAT WILL NOT CHANGE PHARMACOLOGICALLY UPON SUCH
EXPOSURE.
● As with all automatic infusion devices, whenever a TOXIC OR DANGEROUS LEVEL of drug is stored in the reservoir, CONSTANT MONITORING of the infusion is required.
● CAUTION must be employed to assure that the Auto Syringe AS50 Infusion Pump is in good working order before putting it into use. If the pump is being operated on battery power alone, ensure that the battery has been charged as described in this manual.
● When the pump is first turned on, verify that the Lamp Test is accurate (see “Display,
Battery, and Syringe Sensor Check” on page 75).
● Verify all program data before pressing START .
● Wipe off spills immediately. Do not allow fluid or residues to remain on the pump.
● Repeated incorrect identification of syringe size by Syringe Recognition may signify a pump fault condition, or that a syringe manufacturer has made a dimensional change.
The pump should be removed from service as soon as possible, so that the problem can be investigated.
Page 6 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Precautions
● If the Size Override configuration option is enabled, an operator can manually override the Syringe Recognition feature. Incorrect syringe information may cause delivery errors.
Special Precautions for Epidural Administration
!
● Epidural administration of anesthetics is limited to short-term infusion (not to exceed
96 hours) with indwelling catheters specifically indicated for short-term anesthetic epidural drug delivery.
● Epidural administration of analgesics is limited to use with indwelling catheters specifically indicated for either short- or long-term analgesic epidural drug delivery.
● To prevent infusion of drugs that are not indicated for epidural use, do not use IV extension sets that incorporate injection sites during epidural delivery.
● It is strongly recommended that pumps used for epidural drug delivery be clearly differentiated from pumps used for other routes of administration.
WARNING: Epidural administration of drugs other than those indicated for epidural use could result in serious injury to the patient.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 7
Precautions
This page is intentionally blank.
Page 8 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Simplified Instructions
Chapter 4
Operation
Refer to the figure below for the location and description of the pump’s operating controls.
The item numbers shown in the figure are referenced throughout the operating instructions also.
1.
Status Display
2.
Plunger Assembly
(Finger Grip, Plunger
Clamp, Plunger Driver)
3.
Status Panel
4.
Keypad
5.
Flange Slot
6.
Barrel Clamp
7.
Cradle
8.
Charger Port
9.
IV Pole Loop
10.
Pole Clamp Mounting
Pins
11.
Battery Cover
12.
Pole Clamp Attach
Recess
13.
Barrel Clamp Release
Levers
14.
ON/OFF Switch
15.
DB-9 RS-232 Serial
Communication Port
14
8
Front View
Bottom View
1
3
4
7
2
5
6 restr by
CA
Top View
Back View
9
10
11
12
15
13
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 9
Simplified Instructions
Note: Steps 1 and 2 can be performed after programming the pump if necessary (for example, if using CUSTOM DILUTION mode).
1.
Attach an IV extension set to the syringe and manually purge the air from the filled syringe and tubing. Baxter recommends the following IV extension sets.
Catalog
Number
2N3345
2N3346
2N3347
2N3348
2N3349
2N3350
36” Micro-Volume IV extension set
Description
36” High Flow Rate IV extension set
36” Micro-Volume IV extension set with 0.22 micron filter
60” Micro-Volume IV extension set
60” High Flow Rate IV extension set
60” Micro-Volume IV extension set with 0.22 micron filter
PLUNGER
2.
Mount the syringe on the pump.
• Pull out the plunger driver (2), slide it to top of the pump, and release it.
FLANGE
• Using the lever (13) on the side or rear of the pump, release the barrel clamp (6).
• Place the syringe in the cradle (7), making sure the syringe flange fits into the slot (5).
BARREL
• Close the barrel clamp (6) firmly against the syringe barrel.
• Pull out the plunger driver (2), slide it down to capture the top of the syringe plunger, and push in firmly to capture syringe plunger.
3.
Slide the ON/OFF switch (14) to the ON position.
4.
Program the pump (see “Programming” on page 11).
5.
Purge the syringe and tubing again to remove slack from the plunger driver.
• Press PURGE , START to begin purging. Repeat if necessary.
• Press STOP if necessary to stop purging.
6.
Connect the IV extension set to the patient.
7.
Press START to begin the infusion.
Page 10 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Simplified Instructions
Programming
Note: If the Protocol Library feature is enabled, a
▲▼
SELECT LIBRARY prompt is displayed first. Use
▲
,
▼
and CONFIRM as required to select a protocol library. Next, select a protocol from the library.
If the selected protocol specifies a default infusion mode, the pump skips step 1.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼ as necessary to display the available infusion mode selections. Press CONFIRM to select an infusion mode.
2.
Select the syringe manufacturer and size using
▲
or
▼
as necessary and complete the data entry by pressing CONFIRM . (For use of the Size Override feature, see “Size
Override” on page 64.)
3.
Enter the remaining program data as required. The current field is the one that is flashing.
• Press number keys to enter the desired value.
• Press CONFIRM to complete the data entry.
4.
To change programmed data:
• Press
▲
or
▼
to move to the desired field.
• Press EDIT . The current value flashes.
• Press
▲
,
▼
, or number keys to change the value.
• Press CONFIRM to complete the data entry.
5.
To program or edit the bolus size:
• Press BOLUS , EDIT .
• Set the bolus size using the number keys.
• Press START to deliver a bolus, or CONFIRM to store the bolus size.
6.
To clear the Total Delivered display:
• Use
▲
or
▼
to move to the TOTAL mL field.
• Press EDIT , CLR to reset the total to 0.00.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 11
Simplified Instructions
Programming the Continuous Infusion Modes
Programming for each infusion mode proceeds in the sequence shown in the following table, in top-to-bottom order.
Step
1
2
3
4
5
6
7
8 mL/hr mL/min
Select mode, syringe manufacturer, syringe size.
Enter infusion rate in mL/hr or mL/ min.
Continuous Infusion Modes
Units/hr
Units/min mUnits/hr mUnits/min
Select mode, syringe manufacturer, syringe size. mcg/min mcg/hr mg/min mg/hr
Select mode, syringe manufacturer, syringe size.
Enter drug concentration in
Units/mL.
Enter dose in
Units/hr, Units/min, mUnits/hr, or mUnits/min.
Enter drug concentration in mg/mL.
Enter dose in mcg/hr, mcg/min, mg/hr, or mg/min.
mcg/kg/min mcg/kg/hr mg/kg/min mg/kg/hr
Select mode, syringe manufacturer, syringe size.
CUSTOM
DILUTION
Select mode, syringe manufacturer, syringe size.
Enter target rate in mL/hr.
Enter body weight in kg.
Enter body weight in kg.
Enter drug concentration in mg/ mL.
Enter drug concentration in mg/ mL.
Enter dose in mcg/kg/hr, mcg/kg/min, mg/kg/hr, or mg/kg/min.
Enter target dose in mcg/kg/min.
Enter volume limit in mL.
Enter bolus size in mL.
Enter bolus size in
Units.
Enter bolus size in mg.
Enter bolus size in mcg/kg or mg/kg.
Enter final volume in mL.
See last Note below.
Enter bolus size in mcg/kg.
Notes:
• mL = milliliters, kg = kilograms, mg = milligrams, mcg = micrograms, mUnits = milliunits.
• If the Protocol Library feature is active, these modes can be accessed directly by selecting NONE (if available) at the SELECT LIBRARY or SELECT PROTOCOL programming step.
• The mode, rate, concentration, dose, bolus size, and maximum bolus rate may be initialized automatically if a protocol is selected.
• Bolus programming is optional. Bolus programming may be disallowed if a protocol is selected.
Note: Custom Dilution mode:
• Press CONFIRM to acknowledge the displayed Drug Volume. Mix the indicated drug volume with sufficient IV fluid to produce the Final Volume and fill the syringe. Follow manual purging and mounting procedures (see “Operation” on page 9).
Page 12 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Simplified Instructions
Programming the Timed Infusion Modes
Programming for each infusion mode proceeds in the sequence shown in the following table, in top-to-bottom order.
Steps
1
2
3
4
5
6
SINGLE DOSE
Timed Infusion Modes
MANUAL SCHEDULE AUTO SCHEDULE
Select mode, syringe manufacturer, syringe size.
Select mode, syringe manufacturer, syringe size.
Select mode, syringe manufacturer, syringe size.
INFUSE field: Enter dose in mL. INFUSE field: Enter dose in mL. INFUSE field: Enter dose in mL.
OVER field: Enter dose duration in hours and minutes (hh:mm).
OVER field: Enter dose duration in hours and minutes (hh:mm).
OVER field: Enter dose duration in hours and minutes (hh:mm).
EVERY field: Enter dose Interval in hours and minutes (hh:mm).
EVERY field: Enter dose Interval in hours and minutes (hh:mm).
NEXT DOSE IN field: Enter delay to start of first dose in hours and minutes (hh:mm).
To start the first dose without delay, program NEXT DOSE IN field to
0:00 .
When the DOSE DUE alarm occurs, press START to deliver one dose.
NEXT DOSE IN field: Enter delay to start of first dose in hours and minutes (hh:mm).
To start the first dose without delay, program NEXT DOSE IN field to 0:00 .
Note:
• If the Protocol Library feature is active, these modes can be accessed directly by selecting NONE (if available) at the
SELECT LIBRARY or SELECT PROTOCOL programming step.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 13
Simplified Instructions
This page is intentionally blank.
Page 14 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
General Information
Chapter 1
Technical Specifications
Model:
Catalog Code:
Size:
Weight:
Flow Rate Accuracy:
Volume Accuracy:
Syringes:
Auto Syringe AS50 Infusion Pump
1M8550 approx. 3.4" x 2.6" x 10" (8.6 x 6.7 x 25 cm) approx. 3.2 lb. (1.45 kg)
±
3% of full scale plunger travel (not including syringe tolerance)
± 3%, or ± 0.007" of travel, whichever is greater (not including syringe tolerance)
B-D 1, 3, 5, 10, 20, 30, 60 mL plastic
Sherwood Monoject 1, 3, 6, 12, 20, 35, 60 mL
Terumo 1, 3, 5, 10, 20, 30, 60 mL
Flow Rate Range:
Deliverable Volume:
Data Display:
Status Display:
0.01 mL/hr to 438 mL/hr, depending on the syringe
Full syringe volume for 1-60 mL syringes
Self-prompting, multi-field LCD (Liquid Crystal Display)
Nine-LED (Light Emitting Diode) array
Power Requirement:
Temperature Range:
AC: 105-125V 60 Hz (battery charger)
DC: internal rechargeable battery pack
Battery Operating Time: 5 hours of operation at 100 mL/hr, or 12 hours of operation at 2 mL/hr using a 60 mL syringe, following a charge of no less than 16 hours.
50 o
F to 104 o
F (10 o
C to 40 o
C) Delivery of high viscosity fluids at low temperatures is not recommended.
Keypad:
Construction:
Elastomeric type, with tactile feedback
Water-resistant, high-impact plastic case with removable elastomeric protective bumpers
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 15
General Information
The Status Display
SYRINGE m f r s i z e
INFUSE
RATE
OVER
BODY
WT
EVERY
CONC.
NEXT
DOSE
IN
T FINAL
VOL LIMITVOL
T DR UG
VOL m L m L min hr kg hr :min
Units mg m L hr :min mg U hr kg hr :min mgrams mLmcg m L
The status display has eight information areas, called fields . Except for the text field at the bottom, each field displays three types of information:
● The markings on the left and right side of each field are called annunciators . The left annunciators identify the type of data contained in the field, while the right annunciators identify the units of measure.
● The central portion of each field displays variable numeric data, either decimal numbers or time. This data generally represents either infusion program information or an active display such as a running total of the amount of drug infused.
● The bottom field of the status display is a two-line text display, used to prompt the programming sequence or to display a message about the infusion.
After the infusion mode has been selected, the pump automatically displays the necessary fields to be programmed. When the pump needs information, it provides a text prompt and flashes the current field.
Note: The pump’s backlight is on whenever the charger is connected and the pump is switched on. The pump can also be configured to turn the backlight on for a specified time (0 to 60 minutes) after each keypress when the charger is not connected. See the Auto Syringe AS50 Infusion Pump Technical Manual for details.
SYRINGE m f r s i z e
Field #1 displays the syringe manufacturer code and syringe size in mL.
[B] = B-D
[M] = Monoject
[T] = Terumo
INFUSE m L m L min hr
Field #2 displays the volume to be infused
(mL), or the infusion rate (mL/hr or mL/min).
Page 16 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Field #3 displays dose duration ( HH:MM ), or patient body weight (kg).
OVER
BODY
WT
Field #4 displays dose interval ( HH:MM ), drug concentration (mg/mL, or U/mL), or bolus size
(Units or mL).
EVERY
CONC.
Field #5 displays time remaining until the next dose is due (hh:mm), dose size (mcg/min, mcg/hr, mg/min, mg/hr, mcg/kg/min, mcg/kg/hr, mg/kg/min, mg/kg/hr, mU/min, mU/hr, U/min, or
U/hr) or bolus size (mg, mcg/kg, mg/kg, or
Units).
NEXT
DOSE
IN
Field #6 displays total delivered (mL, mcg, mg, or grams), volume limit (mL), bolus size (mL), or final volume (mL).
T FINAL
VOL LIMITVOL
Field #7 displays total delivered (mL), or drug volume to be mixed with infusion base solution
(mL).
DRUG
VOL
Field #8 displays text prompts and messages.
General Information kg hr:min
Units mg m L hr:min mg U hr mcg kg hr:min mgrams mLmcg m L
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 17
General Information
The Keypad
7
4
1
0
8
5
2
PURG CLR
9
6
BOLUS
EDIT
3 LOCK
CONFIRM
The keypad includes a standard set of number keys and decimal point (referred to in this document as digit keys), arrow keys, and various function keys.
Arrow Keys The up arrow key (
▲
) increases a number, selects a different field, or displays the next item in a list.
The down arrow key (
▼
) decreases a number, selects a different field, or displays the previous item in a list.
Digit Keys The digit keys are used to enter numeric values directly.
Function Keys These keys activate specific functions.
Page 18
Action Keys
Confirm Key
START STOP
These keys start and stop the pump motor.
This key is used to complete a programming step, to respond to a prompt, or to display additional infusion program information.
Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
The Status Panel
RUN lights
ALERT
BATTERY
ON CHARGE
STANDBY
LOCK
SYSTEM
General Information
When the pump is in the Run state, one or more of the green RUN lights are on.
When the pump is infusing, the three RUN lights flash in sequence, in a “falling drop” pattern.
When the pump is counting time until an infusion is to begin, only the top RUN light flashes.
When a bolus or purge is in progress, all three RUN lights flash simultaneously.
The red ALERT light flashes during an alert or alarm. The status display and remaining status lights may display additional information.
When flashing, the red BATTERY light indicates that the battery charge is low. If the battery discharges significantly, the BATTERY light stays on constantly and the pump goes into a fail-safe alarm state.
The green ON CHARGE light indicates that the battery charger is plugged in.
The yellow STANDBY light flashes to indicate that the pump is in Standby state.
The yellow LOCK light indicates that the keypad is locked.
The red SYSTEM light indicates a malfunction. The pump cannot operate until the problem is corrected. The infusion must then be reprogrammed.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 19
General Information
Operating States
Standby State
When the pump is in Standby state, the infusion is stopped and the STANDBY light flashes. If the pump is left in Standby state for more than two minutes, a PUMP IS IDLE alert or alarm occurs.
Run State
There are two Run states: Run Delivering state and Run Counting state.
In Run Delivering state, the infusion program is running and the pump is delivering an infusion. The three RUN lights flash in a “falling drop” pattern.
In Run Counting state, the infusion program is counting time until the next dose is due, and the pump is not delivering an infusion. The top RUN light flashes.
Note: Bolus and Purge operations are special infusion states. The RUN lights flash in unison when bolusing or purging in both Standby and Run states.
Data Display and Entry
Numeric Data
Time Data Fields
Automatic Decimal
Flashing Data
Flashing Annunciators
Data Test
Numbers are displayed in a right-justified format. A leading zero appears before decimal values when the field size permits. For example, the value “.12” is displayed as 0.12
.
In time fields, values are displayed right justified. Leading zero(es) are automatically entered if needed. For example,
1 minute is entered by pressing 1 , CONFIRM and is displayed as 0:01 .
If a number is entered without a decimal point, the pump automatically enters a decimal point to the right of the number when CONFIRM is pressed.
The data within a field flashes to show that the field has been opened and is ready to be programmed or edited.
The annunciators for the current field flash. The current field will be opened for editing when EDIT is pressed.
All entered data is tested when CONFIRM is pressed. If the entered value is too large or too small, the pump issues a
PUMP LIMIT alert and substitutes the nearest acceptable value. Press CONFIRM to accept the value, or enter a different value.
Page 20 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
General Information
Arrow Keys
Key Sounds An audible click provides confirmation of each keypress.
The pump indicates an invalid keypress by sounding a short beep. Simultaneous keypresses are ignored.
Dashes Dashes are displayed when a field is clear and open for programming. In general, digit keys must be used to enter a value in a dashed field. If an arrow key is pressed while dashes are displayed, the pump beeps and displays the message USE DIGITS .
Protocol Library Feature
The Protocol Library feature is a configurable option that allows authorized personnel to store named sets or protocols of frequently used programming data. Protocols may also be grouped into libraries.
Using a protocol can simplify the programming process by providing default selections at some of the data entry steps. For instructions on selecting a protocol, see “Using
Protocols” on page 32.
The Protocol Library feature is enabled or disabled through the pump configuration mode.
This procedure can be done by authorized personnel on site, and is described in the Auto
Syringe AS50 Infusion Pump Service Manual.
A protocol may specify any of the following:
Default Infusion Mode
Arrow keys can be used instead of individual digits for selecting predefined responses, incrementing or decrementing fields. In general, arrow keys cannot be used to initialize fields that have been cleared to dashes, except for syringe size and manufacturer fields.
Default Concentration
Default Dose or Rate
Bolus Prevention
If a protocol specifies a default infusion mode, the infusion mode programming step is skipped. The pump briefly displays the infusion mode name before proceeding directly to the syringe manufacturer programming step.
If a protocol specifies a default concentration, the default value appears automatically at the start of the concentration programming step. This value may be changed, or it may be accepted by pressing CONFIRM .
If a protocol specifies a default dose or rate, the default value appears automatically at the start of the ENTER DOSE or ENTER RATE programming step. This value may be changed, or it may be accepted by pressing CONFIRM .
If a protocol specifies bolus prevention, the pump will not allow a bolus to be programmed or delivered. If BOLUS is pressed, the pump beeps and displays the message NOT
ALLOWED .
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 21
General Information
Default Bolus Size
Bolus Maximum Rate
If a protocol specifies a default bolus size, the default value appears automatically when the Bolus Size field is first opened for editing. This value may be changed, or it may be accepted by pressing CONFIRM .
The bolus delivery rate is normally limited to the smaller of: the limit specified by the RATE RANGE configuration option, or the maximum rate for the programmed syringe. If a protocol specifies a bolus maximum rate, the pump will further limit the bolus delivery rate so it does not exceed the specified bolus maximum rate, which is defined in terms of dosage (such as mcg/kg/min).
Bypassing the Protocol Library Feature
Note: The NONE selection may not be available, depending upon configuration.
Selecting NONE at the SELECT LIBRARY or SELECT PROTOCOL step bypasses the
Protocol Library feature and allows infusion mode selection and programming to proceed in the normal way.
Serial Communication Feature
The pump provides remote control and monitoring capabilities. A standard 9-pin connector (DB9) allows the connection of an RS-232 cable between the pump and a computer. The cable allows an operator at a remote computer to control pump operations and to query the pump for data such as infusion totals, program values, and operational status. The remote control feature is a configurable option. See the Auto Syringe AS50
Infusion Pump Programmer’s Manual for details.
Page 22 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
General Information
Infusion Modes
The pump provides 18 infusion modes, which fall into two major categories: continuous and timed infusions. Continuous infusions deliver at a steady, programmed rate. Timed infusions deliver one or more equal doses according to a programmed schedule. The infusion modes are briefly described below. For more information see “Detailed
Instructions” on page 27.
Continuous Infusion Modes
mL/hr and mL/min: The rate of infusion is programmed in mL/hr or mL/min, depending on the mode selected. A volume limit can be programmed for these modes only.
Units/hr, Units/min, mUnits/hr and mUnits/min: The drug concentration is entered in units/mL, and the dose is entered in Units/hr, Units/min, mUnits/hr or mUnits/min, depending on the mode selected.
The pump automatically displays the effective rate in mL/hr.
mg/hr, Mg/min, mcg/hr and mcg/Min: The drug concentration is entered in mg/mL, and the dose is entered in mg/hr, mg/min, mcg/hr or mcg/min, depending on the mode selected. The pump automatically displays the effective rate in mL/hr.
mg/kg/hr, mg/kg/min, mcg/kg/hr and mcg/kg/min: The patient's body weight is entered in kg, the drug concentration is entered in mg/mL, and the dose is entered in mg/kg/hr, mg/kg/min, mcg/kg/hr or mcg/kg/min, depending on the mode selected. The pump automatically displays the effective rate in mL/hr.
CUSTOM DILUTION: In this mode the pump is programmed for a continuous infusion and the drug concentration is adjusted for a target infusion rate and dose. The target rate is entered in mL/hr, the patient’s body weight is entered in kg, the initial drug concentration is entered in mg/mL, the target dose is entered in mcg/kg/min, and the final volume is entered in mL. The pump calculates and displays the volume of drug
(in mL) to be added to the infusion base solution to produce a final concentration that achieves the target rate and target dose. The final concentration is displayed when the infusion is initiated.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 23
General Information
Timed Infusion Modes
SINGLE DOSE:
MANUAL SCHEDULE:
AUTO SCHEDULE:
In this mode, the pump is programmed to infuse a single dose over a specified period of time. The dose is entered in mL, and the time period is entered in hours and minutes.
In this mode, the pump is programmed to infuse a dose over a specified period of time, and to repeat the dose at specified intervals. The dose is entered in mL, and the time period and interval are entered in hours and minutes.
The pump alerts the operator when a dose is due to start.
The operator must then press START to initiate each dose.
In this mode, the pump is programmed to infuse a dose over a specified period of time, and to repeat the infusion at specified intervals. The dose is entered in mL, and the time period and interval are entered in hours and minutes.
Whereas MANUAL SCHEDULE mode requires the operator to press START to initiate each dose, AUTO SCHEDULE mode automatically starts each dose when it is due.
Page 24 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
General Information
Syringe Data Tables
This table shows the performance ranges for infusion delivery parameters, which are a function of the syringe dimensions.
Infusion Rate Ranges
Syringe
Manufacturer
B-D Plastic
Monoject
Syringe
Size
(mL)
1
3
6
12
20
35
60
1
3
5
10
20
30
60
Minimum
Flow
Rate
(mL/hr)
0.01
0.02
0.03
0.1
0.1
0.1
0.1
0.01
0.02
0.03
0.1
0.1
0.1
0.1
Maximum
Flow
Rate
(mL/hr)
11
41
83
126
208
285
366
11
38
74
107
188
241
366
Minimum
Flow Rate
(mL/min)
Maximum
Flow Rate
(mL/min)
0.001
0.001
0.001
0.002
0.002
0.002
0.002
0.001
0.001
0.001
0.002
0.002
0.002
0.002
Terumo 1
3
5
10
20
30
60
0.01
0.02
0.03
0.1
0.1
0.1
0.2
11
41
87
129
210
276
438
0.001
0.001
0.001
0.002
0.002
0.002
0.004
Maximum Flow Rates shown apply to pumps configured for:
Rate Range = H (High).
Rate Range = High limits maximum rate to maximum rate for syringe used.
Rate Range = Medium limits maximum rate to 120 mL/hr (2 mL/min).
Rate Range = Low limits maximum rate to 15 mL/hr (0.25 mL/min).
0.183
0.683
1.45
2.15
3.5
4.6
7.3
0.183
0.633
1.233
1.783
3.13
4.01
6.1
0.183
0.683
1.383
2.1
3.46
4.75
6.1
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 25
General Information
Deliverable Volumes
This table shows the minimum and maximum programmable settings for all syringes that can be used with the pump. The limits shown apply to bolus, volume limit, single dose, and scheduled dose deliveries.
Syringe
Manufacturer
B-D Plastic
Monoject
Terumo
Syringe Size (mL)
1
3
6
12
20
35
60
1
3
5
10
20
30
60
1
3
5
10
20
30
60
Minimum Volume
(mL)
Maximum Volume
(mL)
0.03
0.08
0.1
0.1
0.1
0.1
0.1
0.03
0.08
0.1
0.1
0.1
0.1
0.1
0.03
0.08
0.1
0.1
0.1
0.1
0.1
1
3
6
12
20
35
60
1
3
5
10
20
30
60
1
3
5
10
20
30
60
Page 26 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Chapter 1
This section provides detailed descriptions of the seven basic steps required to operate the pump:
Note: Steps 1 and 2 can be performed after programming the pump if necessary (for example, if using CUSTOM DILUTION mode).
1.
Attach a Baxter 2N series (or equivalent) extension set to the syringe and manually purge the air from the filled syringe and tubing.
2.
Mount the syringe on the pump.
3.
Slide the ON/OFF switch (7) to the “on” position (refer to figure on the following page).
4.
Program the pump.
5.
Purge the syringe and tubing again to remove slack from the plunger driver.
6.
Connect the extension set to the patient.
7.
Press START to begin the infusion.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 27
Detailed Instructions
Preliminary Operations
Manual Purge
1.
Attach a Baxter 2N Series or equivalent IV extension set to the filled syringe.
2.
Remove the sterile cap from the end of the tubing.
3.
Manually purge until a drop of fluid is visible at the end of the tubing.
4.
Replace the sterile cap on the end of the tubing.
Mount the Syringe
1.
Grasp the FINGER GRIP (1) on the PLUNGER DRIVER (2) and pull it all the way out.
2.
Slide the PLUNGER DRIVER to the top of the pump and release the FINGER GRIP.
2
1
7
Page 28 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
3.
Open the BARREL CLAMP (3) by moving either of the BARREL CLAMP RELEASE
LEVERS (4). Allow the barrel clamp to slide all the way open.
4.
Place the syringe in the CRADLE (5). Be sure the syringe is centered in the V of the cradle, and that the flange of the syringe fits into the FLANGE SLOT (6).
5.
Push the BARREL CLAMP (3) in firmly so that it securely grips the syringe barrel. The barrel clamp is shaped to fit the heel of the hand, making it easy to firmly clamp the syringe.
6
3
5
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 29
Detailed Instructions
6.
Grasp the FINGER GRIP (1) on the DRIVER (2) and pull it all the way out. Slide the
DRIVER down until it contacts the syringe plunger, then release.
7.
Push the PLUNGER CLAMP in firmly to capture the syringe plunger.
2
1
ON/OFF
SWITCH
Turn the Pump On and Off
The ON/OFF switch is on the left side of the pump, near the bottom. Turn the pump on by sliding the switch up to the ON position.
Note: If the syringe is improperly mounted, the pump displays one of the following alert messages: CHECK BARREL , CHECK FLANGE , or CHECK PLUNGER . Examine the syringe and correct the indicated problem. If a combination of conditions is sensed, the pump displays a CHECK SYRINGE alert or alarm message instead.
The pump is turned off by sliding the switch down to the OFF position.
Page 30 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Check the Lamp Test
As soon as the pump is turned on, an automatic self-test begins. This includes the Lamp Test, which tests the status display and the status panel lights. Verify that all the lights illuminate and that all parts of the status display are legible.
Carefully check the Lamp Test daily, or once per shift. See
“Display, Battery, and Syringe Sensor Check” on page 75.
To examine the status display more closely during the
Lamp Test, press and hold STOP while the status display characters are all visible. The Lamp Test will complete when STOP is released.
If CONFIRM is pressed during the Lamp Test, the battery voltage will be displayed in the text field immediately following completion of the Lamp Test.
The Lamp Test ends with a short beep. If the pump is configured to show an identifier message, it will appear in the text field at the bottom of the status display.
This completes the preliminary operations.
SYRINGE m f r s i z e
INFUSE
RATE
OVER
BODY
WT
EVERY
CONC.
NEXT
DOSE
IN
T FINAL
VOL LIMITVOL
T DR UG
VOL m L m L m L min hr kg hr :min
Units mg m L hr :min mg U hr kg hr :min mgrams mLmcg
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 31
Detailed Instructions
Programming
The pump is ready for programming when the preliminary operations have been completed. The initial programming prompt is determined as follows:
● If the Protocol Library feature is enabled, a protocol selection prompt is displayed.
Continue with the “Using Protocols” section below.
● If the Protocol Library feature is not enabled, a mode selection prompt is displayed.
Skip to the appropriate mode programming section below.
For additional information about programming the pump, see “General Programming
Information” on page 48.
Using Protocols
If the Protocol Library feature is enabled, the pump display alternates the prompts
▲▼
SELECT LIBRARY and
▲▼
SELECT PROTOCOL .
Selecting a Protocol Library
If
▲▼
SELECT LIBRARY is displayed, use the
▲
and
▼
keys as necessary to display the name of the desired protocol library and press CONFIRM to complete the selection.
Note: If the NONE selection is available, the Protocol Library feature may be bypassed by selecting NONE . Continue with the appropriate mode programming section below.
Selecting a Protocol
If
▼▲
SELECT PROTOCOL is displayed, use the
▲
and
▼
keys as necessary to display the name of the desired protocol, and press CONFIRM to complete the selection.
Note: The protocol name is continuously displayed in the text field except when a message or prompt must be displayed.
If the NONE selection is available, the Protocol Library feature may be bypassed by selecting NONE . Continue with the appropriate mode programming section below.
Page 32 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Selecting an Infusion Mode
If a default mode has been assigned to the selected protocol, the pump sets the infusion mode automatically. The name of the mode is displayed briefly in the text field, and programming continues with syringe manufacturer selection.
If no default mode has been assigned to the selected protocol, the pump displays a
SELECT MODE
▼▲
prompt.
Continue with the appropriate mode programming section below.
For additional information about the Protocol Library, see “Protocol Library Feature” on page 21.
Programming mL/hr and mL/min Mode
In these modes, the pump runs at a programmed, constant rate. A volume limit can also be programmed in these modes.
Note: If a protocol has been selected, the infusion mode may appear automatically.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use
▲
or
▼
and CONFIRM as required to select the syringe manufacturer and size.
Note: If a protocol has been selected, a default infusion rate may appear automatically.
Use ▲ and ▼ or the digit keys to modify this value as necessary.
3.
Enter the infusion rate in mL/hr or mL/min, depending on the infusion mode selected.
4.
Press CONFIRM to complete the entry.
Volume Limit
The Volume Limit feature is a configurable option available only in the mL/hr and mL/min modes. The pump stops when the programmed volume limit has been delivered.
Enter the desired volume limit. When the delivery begins, the VOL LIMIT field displays the amount of drug remaining until the programmed volume limit is reached. The pump stops when the VOL LIMIT field reaches zero.
The Volume Limit feature can be temporarily disabled without reconfiguring the pump by programming zero ( 0 ) in the VOL LIMIT field.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 33
Detailed Instructions
Bolus in mL/hr and mL/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default bolus size may appear automatically. If a default bolus size is displayed, use
▲ and ▼ or the digit keys to modify the value as necessary.
If the infusion program is to include a bolus delivery:
1.
Press BOLUS to obtain the bolus programming display.
2.
Press EDIT , then use the digit keys to set the bolus size (in mL).
3.
Press CONFIRM to store the bolus size and return to the normal display. The bolus infusion rate is automatically set.
See “Bolus Function” on page 55 for more information on bolus delivery.
Note: For mL/min mode, the bolus rate is displayed in mL/min. For all other modes, it is displayed in mL/hr.
mL/hr and mL/min Programming Notes
1.
See “Syringe Data Tables” on page 25 for minimum and maximum infusion rates for each syringe.
2.
Pressing CONFIRM during delivery briefly displays additional program data.
3.
A bolus can only be programmed, edited, or delivered after the syringe size has been programmed.
Page 34 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Programming Units/hr, Units/min, mUnits/hr, and mUnits/min Modes
In these modes, the pump runs at a programmed, constant rate. The infusion is programmed in terms of drug concentration and dose.
Note: If a protocol has been selected, the infusion mode may have been selected automatically.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use ▲ or ▼ and CONFIRM as required to select the syringe manufacturer and size.
Note: If a protocol has been selected, a default drug concentration may appear automatically. If a default drug concentration is displayed, use
▲ and
▼ or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
3.
At the ENTER CONCENTR prompt, enter the drug concentration ( CONC.
) in Units/mL.
Press CONFIRM to complete the data entry.
Note: If a protocol has been selected, a default dose may appear automatically. Use
▲ and ▼ or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
4.
At the ENTER DOSE prompt, enter the dose in Units/hr, Units/min, mUnits/hr, or mUnits/min (mUnits = milliUnits), depending on the infusion mode selected. Press
CONFIRM to complete the data entry. The pump automatically calculates and displays the equivalent rate in mL/hr.
Bolus in Units/hr, Units/min, mUnits/hr, and mUnits/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default bolus size may appear automatically. If a default bolus size is displayed, use ▲ and
▼
or the digit keys to modify the value as necessary.
In these modes, bolus size is entered in Units or mUnits. The pump automatically calculates the equivalent bolus size in mL and displays both values. Bolus rate is displayed in mL/hr. See “Bolus Function” on page 55 for more information on bolus delivery.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 35
Detailed Instructions
Programming Notes
1.
The drug concentration range is from 0.01 Units/mL to 9999 Units/mL. The dose limits are determined by the drug concentration and syringe selection. See “Syringe Data
Tables” on page 25 for minimum and maximum infusion rates for each syringe.
2.
The CONC . field cannot be reprogrammed once the infusion has been started.
3.
Pressing CONFIRM during delivery briefly displays additional program data.
4.
A bolus can only be programmed, edited, or delivered after the drug concentration
( CONC .) has been programmed.
Programming mg/hr, mg/min, mcg/hr, and mcg/min Modes
In these modes, the pump runs at a programmed, constant rate. The infusion is programmed in terms of drug concentration and dose.
Note: If a protocol has been selected, the infusion mode may have been selected automatically.
1.
At the SELECT MODE
▲▼ prompt, use ▲ or ▼ to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use
▲
or
▼
and CONFIRM as required to select the syringe manufacturer and size.
Note: If a protocol has been selected, a default drug concentration may appear automatically. If a default drug concentration is displayed, use
▲ and
▼ or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
3.
Enter the drug concentration ( CONC .) in mg/mL. Press CONFIRM to complete the data entry.
Note: If a protocol has been selected, a default dose may appear automatically. Use ▲ and
▼ or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
4.
Enter the dose in mg/hr, mg/min, mcg/hr or mcg/min, depending on the infusion mode selected. Press CONFIRM to complete the data entry. The pump automatically calculates and displays the equivalent rate in mL/hr.
Page 36 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Bolus in mg/hr, mg/min, mcg/hr, and mcg/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default bolus size may appear automatically. If a default bolus size is displayed, use ▲ and ▼ or the digit keys to modify the value as necessary.
In these modes, bolus size is entered in mg or mcg. The pump automatically calculates the equivalent bolus size in mL and displays both values. Bolus rate is displayed in mL/hr.
See “Bolus Function” on page 55 for more information on bolus delivery.
Programming Notes
1.
The drug concentration range is from 0.0001 mg/mL to 100.0 mg/mL. The dose limits are determined by the drug concentration and syringe selection. See “Syringe Data
Tables” on page 25 for minimum and maximum infusion rates for each syringe.
2.
The CONC . field cannot be reprogrammed once the infusion has been started.
3.
Pressing CONFIRM during delivery briefly displays additional program data.
4.
A bolus can only be programmed, edited, or delivered after the drug concentration
( CONC .) has been programmed.
Programming mg/kg/hr, mg/kg/min, mcg/kg/hr, and mcg/kg/min Modes
In these modes, the pump runs at a programmed, constant rate. The infusion is programmed in terms of patient body weight, drug concentration and dose.
Note: If a protocol has been selected, the infusion mode may have been selected automatically.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use ▲ or ▼ and CONFIRM as required to select the syringe manufacturer and size.
3.
Enter the patient body weight ( BODY WT ) in kg. Press CONFIRM to complete the data entry.
Note: If a protocol has been selected, a default drug concentration may appear automatically. If a default drug concentration is displayed, use
▲ and
▼ or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
4.
Enter the drug concentration ( CONC .) in mg/mL. Press CONFIRM to complete the data entry.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 37
Detailed Instructions
Note: If a protocol has been selected, a default dose may appear automatically. Use ▲ and
▼
or the digit keys to modify this value as necessary, and press CONFIRM to complete the data entry.
5.
Enter the dose in mg/kg/hr, mg/kg/min, mcg/kg/hr or mcg/kg/min, depending on the infusion mode selected. Press CONFIRM to complete the data entry. The pump automatically calculates and displays the equivalent rate in mL/hr.
Bolus in mg/kg/hr, mg/kg/min, mcg/kg/hr, and mcg/kg/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default bolus size may appear automatically. If a default bolus size is displayed, use ▲ and
▼
or the digit keys to modify the value as necessary.
In mg/kg/hr and mg/kg/min modes, bolus size is entered in mg/kg. In mcg/kg/hr and mcg/ kg/min, the bolus size is entered in mcg/kg. The pump automatically calculates the equivalent bolus size in mL and displays both values. The bolus rate is displayed in mL/hr.
See “Bolus Function” on page 55 for more information on bolus delivery.
Programming Notes
1.
The patient body weight range is from 0.01 to 200.0 kilograms (kg). The drug concentration range is from 0.0001 mg/mL to 100.0 mg/mL. The concentration range may be further limited for very large or small body weights. The dose limits are determined by the drug concentration and syringe selection. See “Syringe Data
Tables” on page 25 for minimum and maximum infusion rates for each syringe.
2.
The CONC . and BODY WT fields cannot be reprogrammed once the infusion has been started.
3.
Pressing CONFIRM during delivery briefly displays additional program data.
4.
A bolus can only be programmed, edited, or delivered after the drug concentration
( CONC .) has been programmed.
Page 38 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Programming CUSTOM DILUTION Mode
In this mode, the pump runs at a programmed, constant rate. The infusion is programmed in terms of target rate, patient body weight, initial drug concentration, target dose, and final volume. The pump calculates and displays the drug volume to be mixed with sufficient IV base fluid to yield the programmed final volume. The final concentration is displayed when a bolus or infusion is initiated.
Note: The rate, body weight, concentration, and dose fields will not be editable after the confirm key is pressed. Power the pump off and then back on to make these fields editable again.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the CUSTOM DILUTION mode.
2.
Use
▲
or
▼
and CONFIRM as required to select the syringe manufacturer and size.
3.
Enter the target rate ( RATE ) in mL/hr. Press CONFIRM to complete the data entry.
4.
Enter the patient body weight ( BODY WT ) in kg. Press CONFIRM to complete the data entry.
5.
Enter the initial drug concentration ( CONC .) in mg/mL. Press CONFIRM to complete the data entry.
6.
Enter the target dose ( DOSE ) in mcg/kg/min. Press CONFIRM to complete the data entry.
7.
Enter the final volume ( FINAL VOL ) in mL. Press CONFIRM to complete the data entry.
8.
In the DRUG VOL field, the pump displays the volume of drug that must be mixed with
IV base fluid to yield the programmed final volume. The text field displays DRUG VOL .
Press CONFIRM to acknowledge the drug volume.
9.
Mix the indicated drug volume ( DRUG VOL ) with sufficient IV base fluid to produce the final volume that was programmed ( FINAL VOL ).
Note: The solution may be mixed directly in the syringe, or may be mixed in an IV bag and then drawn up into the syringe.
10.
Attach a 2N Series (or equivalent) IV extension set to the syringe and manually purge the air from the filled syringe and tubing.
11.
Mount the syringe.
12.
Perform a purge operation to remove slack from the plunger driver.
13.
After completing the purge procedures, press START . The new final concentration will be displayed in the concentration field.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 39
Detailed Instructions
!
Bolus in CUSTOM DILUTION Mode
In this mode, bolus size is programmed in mcg/kg. The pump automatically calculates the equivalent bolus size in mL and displays both values. Bolus rate is displayed in mL/hr.
See “Bolus Function” on page 55 for more information on bolus delivery.
Programming Notes
IMPORTANT: If the FINAL VOL field is altered after the solution has been mixed, the solution must be re-mixed using the new DRUG VOL quantity.
For maximum accuracy in measuring the drug volume, select the smallest syringe that will contain the entire DRUG VOL quantity.
When CONFIRM is pressed to confirm the DRUG VOL value, the target rate field ( RATE ) is updated to show the actual infusion rate that will be used. (Usually the actual rate is very close to or the same as the programmed target rate.)
The program data fields all remain displayed and available for editing until an infusion or bolus is started. At that point, the pump replaces the original drug concentration ( CONC .) with the final drug concentration, and clears the fields containing the final volume ( FINAL
VOL ) and the drug volume ( DRUG VOL ). The rate of infusion, patient body weight and dose remain displayed. Pressing CONFIRM temporarily displays the hidden program data
(original drug concentration, final volume, and drug volume).
Target rate ( RATE ) limits are determined by the syringe selection and Rate Range configuration (see “Syringe Data Tables” on page 25 for minimum and maximum infusion rates for each syringe). The patient body weight range is from 0.01 to 200.0 kilograms
(kg). The drug concentration range is from 0.0001 mg/mL to 100.0 mg/mL. The concentration range may be further limited for very large or small body weights. Target dose limits are determined by the drug concentration and syringe selection. Final Volume range is from 30 mL to 250 mL.
The CONC . and BODY WT fields cannot be reprogrammed once the infusion has been started.
A bolus can only be programmed, edited, or delivered after the pump has been fully programmed.
Page 40 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Programming SINGLE DOSE Mode
SINGLE DOSE mode delivers a programmed volume of drug over a specified period of time. The infusion is programmed in terms of dose volume and dose duration.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use ▲ or ▼ and CONFIRM as required to select the syringe manufacturer and size.
3.
Enter the dose volume ( INFUSE ) in mL. Press CONFIRM to complete the data entry.
4.
Enter the dose duration ( OVER ) in hours and minutes. Press CONFIRM to complete the data entry.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 41
Detailed Instructions
SINGLE DOSE Mode Programming Notes
1.
All time displays use standard time notation: the two digits to the left of the colon show hours, the two digits to the right show minutes.
2.
If either the INFUSE field or the OVER field is edited, the pump treats the result as a new infusion. As a reminder, DOSE CANCELED appears for a few seconds.
Example : Program the pump to infuse 5 mL over 5 minutes, using a B-D 30 mL syringe.
Stop the infusion when the TOTAL ML field shows that about 2 mL have been delivered.
Now edit the OVER field to 6 minutes (note that DOSE CANCELED appears for a few seconds), and start the infusion.
When the pump stops after 6 minutes, about 7 mL will have been delivered; 2 mL from the original infusion, and 5 mL from the edited infusion.
3.
The pump issues a DOSE COMPLETE alarm at the end of the infusion if the Dose
Complete Alarm option is enabled.
4.
Bolus operation is not available in this delivery mode.
5.
Pressing CONFIRM during delivery briefly displays additional programming data.
Programming MANUAL SCHEDULE Mode
MANUAL SCHEDULE mode is a timed infusion mode for use when a programmed dose is to be repeated at regular intervals. The infusion is programmed in terms of dose volume and dose duration, and the time from the start of one dose to the start of the next.
Each dose delivery must be manually initiated.
MANUAL SCHEDULE mode also allows a delay to the start of the first dose, allowing a time delay before the infusion regimen begins.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use ▲ or ▼ and CONFIRM as required to select the syringe manufacturer and size.
3.
Enter the dose volume ( INFUSE ) in mL. Press CONFIRM to complete the data entry.
Page 42 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
4.
Enter the dose duration ( OVER ) in hours and minutes. Press CONFIRM to complete the data entry.
5.
Enter the interval ( EVERY ) in hours and minutes. This is the time from the start of one dose to the start of the next dose. Press CONFIRM to complete the data entry.
6.
Enter the delay ( NEXT DOSE IN ), then press CONFIRM . This is the time interval from when START is pressed until the first dose becomes due. Enter 0:00 to begin infusing the first dose immediately.
Dose Due
Note: START must be pressed to begin infusing each dose.
The NEXT DOSE IN field shows the time remaining until the next dose is due to be delivered. When this time counts down to 0:00, the pump signals that the dose is due by flashing the ALERT and STANDBY lights, sounding continuous beeps, and showing DOSE
DUE in the text field. START must be pressed to begin infusing each dose.
When a dose is due, pressing any key other than START silences the audio, turns off the
DOSE DUE prompt and ALERT light, and leaves the pump in Standby state, but does not begin dose delivery.
“Counting Time” Display
The three RUN lights flash in a “falling drop” pattern while the pump is infusing. When the pump is counting time between scheduled doses, only the top RUN light flashes.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 43
Detailed Instructions
!
Programming Notes
Precaution: In MANUAL SCHEDULE mode, START must be pressed to deliver each dose and continue the infusion regimen. Pressing any other key silences the audio portion of the DOSE DUE alarm and leaves the pump in Standby state, without initiating a dose delivery.
1.
For MANUAL SCHEDULE mode it is recommended that the Idle Alarm feature be enabled (configured), and that the Auto Lock feature be disabled (not configured).
2.
All time displays use standard time notation: the two digits to the left of the colon show hours, the two digits to the right show minutes.
3.
The programmed delivery can be stopped at any time by pressing STOP . The RUN light(s) then turn off, the STANDBY light begins flashing, and the TOTAL ML field stops counting up. Press START to continue the infusion program.
Note: Pressing STOP affects the NEXT DOSE IN time field in two different ways, depending on the pump’s infusion status at the time:
• If STOP is pressed while the pump is infusing a dose (the three RUN lights are flashing in sequence), the NEXT DOSE IN field will stop counting down, until START is pressed again. This feature allows the schedule to be delayed while ensuring delivery of the full dose.
• If STOP is pressed while the pump is counting down the time to the next dose (only the top RUN light is flashing), the NEXT DOSE IN field will continue to count down while the pump is stopped. This feature allows editing during idle time, without altering the overall schedule. If a dose becomes due while the pump is stopped, the DOSE DUE alarm will be suppressed and delivery of the dose will begin immediately when START is pressed.
Page 44 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
4.
If either the INFUSE field or the OVER field is edited, the pump treats the result as a new infusion. As a reminder, DOSE CANCELED appears for a few seconds.
Example : Program the pump to infuse 5 mL over 5 minutes, using a B-D 30 mL syringe.
Stop the infusion when the TOTAL ML field shows that about 2 mL have been delivered.
Now edit the OVER field to 6 minutes (note that DOSE CANCELED appears for a few seconds), and start the infusion.
When the pump stops after 6 minutes, about 7 mL will have been delivered; 2 mL from the original infusion, and 5 mL from the edited infusion.
5.
Pressing CONFIRM during delivery briefly displays additional program data.
6.
Bolus operation is not available in this delivery mode.
Programming AUTO SCHEDULE Mode
AUTO SCHEDULE mode is a timed infusion mode for use when a programmed dose is to be repeated at regular intervals. The infusion is programmed in terms of dose volume and dose duration, and the time from the start of one dose to the start of the next. Each dose delivery is automatically initiated.
AUTO SCHEDULE mode also adds a delay to the start of the first dose, allowing a time delay before the infusion regimen begins.
1.
At the SELECT MODE
▲▼
prompt, use
▲
or
▼
to display the available infusion mode selections. Press CONFIRM to select the desired infusion mode.
2.
Use
▲
or
▼
and CONFIRM as required to select the syringe manufacturer and size.
3.
Enter the dose volume ( INFUSE ) in mL. Press CONFIRM to complete the data entry.
4.
Enter the dose duration ( OVER ) in hours and minutes. Press CONFIRM to complete the data entry.
5.
Enter the interval ( EVERY ) in hours and minutes. This is the time from the start of one dose to the start of the next dose. Press CONFIRM to complete the data entry.
6.
Enter the delay ( NEXT DOSE IN ), then press CONFIRM . This is the time interval from when START is pressed until the first dose becomes due.
Enter 0:00 to begin infusing the first dose immediately.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 45
Detailed Instructions
Dose Initiation
The NEXT DOSE IN field on the display shows the time remaining until the next dose is due to be delivered. When this time counts down to 0:00, the pump automatically begins infusing the dose.
“Counting Time” Display
The three RUN lights flash in a “falling drop” pattern while the pump is infusing.
When the pump is counting time between scheduled doses, only the top RUN light flashes.
Programming Notes
1.
All time displays use standard time notation: the two digits to the left of the colon show hours, the two digits to the right show minutes.
2.
The programmed delivery can be stopped at any time by pressing STOP .
The RUN light(s) then turn off, the STANDBY light begins flashing, and the
TOTAL ML field stops counting up. Press START to continue the infusion program.
3.
Pressing STOP affects the NEXT DOSE IN time field in two different ways, depending on the pump's infusion status at the time:
• If STOP is pressed while the pump is infusing a dose (the three RUN lights are flashing in sequence), the NEXT DOSE IN field will stop counting down, until START is pressed again. This feature allows the schedule to be delayed while ensuring delivery of the full dose.
• If STOP is pressed while the pump is counting down the time to the next dose (only the top RUN light is flashing), the NEXT DOSE IN field will continue to count down while the pump is stopped. This feature allows editing during idle time, without altering the overall schedule. If a dose becomes due while the pump is stopped, delivery of the dose will begin immediately when START is pressed.
4.
If either the INFUSE field or the OVER field is edited, the pump treats the result as a new infusion. As a reminder, DOSE CANCELED appears for a few seconds.
Page 46 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Example : Program the pump to infuse 5 mL over 5 minutes, using a B-D 30 mL syringe.
Stop the infusion when the TOTAL ML field shows that about 2 mL have been delivered.
Now edit the OVER field to 6 minutes (note that DOSE CANCELED appears for a few seconds), and start the infusion.
When the pump stops after 6 minutes, about 7 mL will have been delivered; 2 mL from the original infusion, and 5 mL from the edited infusion.
Note: Pressing CONFIRM during delivery briefly displays additional program data.
Bolus operation is not available in this delivery mode.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 47
Detailed Instructions
General Programming Information
Data Entry
The digit keys are used to enter program data.
Before pressing CONFIRM to complete a program data entry, the value can be changed by pressing ▲ or ▼ or can be replaced by pressing CLR and entering a new value. A value that appears automatically (as a result of selecting a protocol) can be changed by pressing ▲ or ▼ or can be replaced by entering a new value without pressing CLR .
Field Selection
The currently selected field on the pump’s status display flashes. Use
▲
and
▼ to select the next field. Press EDIT to open the currently selected field for editing. When a field is opened for editing, the annunciators stop flashing and the value itself begins flashing.
Editing in Standby State
All modes except CUSTOM DILUTION modes:
Almost all program data can be changed (edited). The exceptions are:
● Once a protocol library, protocol or infusion mode has been selected, the only way to select a different protocol library, protocol or infusion mode is to turn the pump off, then turn it back on and reprogram.
● The drug concentration ( CONC .) and patient body weight ( BODY WT ) fields cannot be changed (edited) once an infusion or bolus has been started.
To edit a programmed field:
1.
2.
Use
▲
and
▼
to select the field to be changed.
Press EDIT . Use
▲
,
▼
, and the digit keys as necessary to enter the new value. Press CONFIRM to complete the change.
Page 48 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
CUSTOM DILUTION mode:
To make a change to syringe manufacturer, syringe size, target rate, body weight or initial concentration before a value is entered for final volume:
1.
Press CONFIRM while dashes are displayed in the FINAL VOL field.
2.
3.
Use
▲
and
▼
to select the field to be changed.
Press EDIT . Use
▲
,
▼
, and the digit keys as necessary to enter the new value. Press CONFIRM to complete the change.
4.
5.
Use ▲ and ▼ to return to the FINAL VOL field.
Press EDIT . Use ▲ , ▼ , and the digit keys as necessary to enter the final volume. Press CONFIRM to complete the entry.
6.
Finish programming.
Only the final volume field will be editable after the confirm key has been pressed and the pump starts running or delivers a bolus. To make a change to the final volume field:
7.
Press EDIT . Use
▲
,
▼
, and the digit keys as necessary to enter the new value. Press CONFIRM to complete the change.
The final concentration is displayed when a bolus or infusion is initiated.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 49
Detailed Instructions
Programming Notes
!
If the FINAL VOL field is altered after the solution has been mixed, the solution must be re-mixed using the new DRUG VOL quantity.
1.
For maximum accuracy in measuring the drug volume, select the smallest syringe that will contain the entire DRUG VOL quantity.
2.
The DRUG VOL field may not be edited at any time, once it has been programmed.
3.
Refer to “Programming Notes” on page 40.
Editing in Run State
The fields that can be edited while the pump is in Run state are listed below:
Mode mL/hr, mL/min
SINGLE DOSE, MANUAL/AUTO SCHEDULE
CUSTOM DILUTION
All other modes
Fields
RATE, TOTAL ML
TOTAL ML
DOSE, TOTAL ML
RATE, DOSE,
TOTAL ML
The TOTAL ML field can be cleared while the infusion is running. (The TOTAL
ML field is a data display, and cannot be edited.)
Edit Cancellation
Page 50
If a field has been opened for editing, pressing EDIT again restores the previous value and cancels the edit operation. The pump displays EDIT
CANCELED for a few seconds.
Example : Select mL/hr mode, and program a B-D 60 mL syringe.
Enter 99 in the RATE field and a volume limit of 60 mL.
The RATE field annunciators flash to show it is the current field.
Pressing CLR at this point only results in an “invalid key” error beep.
Press EDIT . Enter another valid number.
Press EDIT again. This will undo the change, restore the previous value ( 99 ), and display EDIT
CANCELED .
Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Detailed Instructions
Number Out of Range
Every programmable field is limited to a particular range of allowed program values, depending on the syringe, drug concentration, and other factors.
Attempting to program a value outside the acceptable range results in a PUMP
LIMIT alert. The pump beeps and substitutes the nearest acceptable value.
Press CONFIRM to accept the value, or program a new value.
Data Test
Whenever a data field is edited, the pump tests the remaining program data fields for range limits. If there is a conflict, the affected fields are cleared to dashes, and the first cleared field is automatically opened for editing. As each cleared field is reprogrammed, the next cleared field (if any) is automatically opened. All cleared fields must be reprogrammed before the infusion can be started.
Example : Reprogramming the SYRINGE MFR field automatically clears the
SYRINGE SIZE field.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 51
Detailed Instructions
This page is intentionally blank.
Page 52 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Additional Functions and
Operations
Chapter 1
Purge Function
!
The Purge function advances the syringe plunger driver a preset distance. Manually purge the syringe and IV extension set before mounting the syringe in the pump. Perform the Purge function after the syringe is mounted on the pump, (before connecting to the patient) to eliminate slack in the drive mechanism and thus ensure prompt fluid delivery when the infusion starts.
Precaution: To prevent unintended delivery to the patient, the IV extension set must be disconnected from the patient while purging.
The “Purge Delivery Table” on page 54 lists the nominal amount of fluid delivered in one complete Purge cycle, if there were no slack in the drive mechanism. Use this information for calculating the excess fluid allowance when filling the syringe.
Purge operation is allowed only when the pump is in Standby state.
Purge Procedure
!
Precaution: To prevent unintended delivery to the patient, the IV extension set must be disconnected from the patient while purging.
1.
After programming the pump, press STOP (if necessary), to put the pump in Standby.
2.
Press PURGE . The normal display is temporarily replaced by a display showing the purge rate in mL/hr, and the message PURGE READY .
Note: If START is not pressed within 10 seconds, the purge function is canceled and the pump reverts to Standby state.
If a steady drip of fluid is observed before the full purge volume is delivered, purging can be stopped immediately by pressing STOP .
3.
Press START to begin delivering the fixed purge volume.
The accumulating purge volume is displayed while purging is in progress, then the normal Standby state display resumes, and the text field reads PURGE COMPLETE for a few seconds. The purge may be repeated until the required volume is achieved. The
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 53
Additional Functions and Operations purge accumulated volume will disappear when the START key is pressed and will not be included in the total volume.
4.
Repeat steps 2 and 3 as often as necessary to obtain a steady drip at the end of the tubing.
Purge Delivery Table
Syringe
Manufacturer
B-D Plastic
Monoject
Terumo
Syringe Size
(mL)
1
3
5
10
20
30
60
1
3
6
12
20
35
60
1
3
5
10
20
30
60
Approximate
Purge
Volume (mL)
0.02
0.08
0.15
0.22
0.39
0.50
0.75
0.02
0.09
0.17
0.26
0.43
0.59
0.75
0.02
0.09
0.18
0.27
0.44
0.58
0.75
Purge Rate
(mL/hr)
11
38
74
107
188
241
366
11
41
83
126
208
285
366
11
41
87
129
210
276
438
Purge Rate
(mL/min)
0.19
0.63
1.2
1.7
3.1
4.0
6.1
0.19
0.69
1.4
2.1
3.5
4.6
7.3
0.19
0.68
1.3
2.1
3.4
4.7
6.1
NOTE: Purge volume and rate information shown applies to pumps configured for Rate Range = H (High).
NOTE: Purge volumes are approximately the amount delivered in a single complete purge delivery, without slack in the drive system.
NOTE: The volume of purged fluid is not added to the amount in the TOTAL mL field.
Page 54 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Additional Functions and Operations
Bolus Function
Note: If a protocol is selected, the bolus function may not be allowed. If the protocol disallows the bolus function, the pump beeps and displays the message
NOT ALLOWED when BOLUS is pressed.
A bolus is a fixed dose which may be programmed and delivered, either during an infusion or while the pump is in Standby state. The pump allows the operator to program the rate of infusion for a bolus, and to program the bolus size either in terms of dose or volume where applicable.
Bolus Units
Infusion Mode Bolus Size Units mL/hr, mL/min
Units/hr, Units/min mUnits/hr, mUnits/min mg/hr, mg/min mcg/hr, mcg/min mg/kg/hr, mg/kg/min,
CUSTOM DILUTION mcg/kg/hr, mcg/kg/min mL (milliliters)
Units mg (milligrams) mg/kg (milligrams of drug per kilogram of patient body weight) mcg/kg (micrograms of drug per kilogram of patient body weight)
SINGLE DOSE, MANUAL SCHEDULE,
AUTO SCHEDULE
Bolus not allowed.
Bolus Size Limits
The minimum programmable bolus size is approximately the same as the minimum deliverable volume (see “Syringe Data Tables” on page 25).
The maximum programmable bolus size is the lesser of:
● The syringe size (the entire syringe may be programmed as a single bolus).
● 9999 Units, 9999 mg, 9999 mg/kg or 9999 mcg/kg (depending on the selected mode).
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 55
Additional Functions and Operations
Bolus Rate
The bolus infusion rate is the smallest of:
● The maximum infusion rate for the syringe (see “Syringe Data Tables” on page 25).
● The maximum rate established by the RATE RANGE configuration option.
● The bolus maximum rate specified by the selected protocol (if applicable).
Example 1: If a pump is configured for Rate Range = Medium, the overall maximum infusion rate is 120 mL/hr.
However, the maximum infusion rate for a B-D 10 mL syringe is 107 mL/hr. Therefore, if a B-D
10 mL syringe is used, the bolus rate will be the lesser of the two limits, or 107 mL/hr.
Example 2: Assume that the same pump and syringe from Example 1 are used.
A protocol is selected that specifies a bolus maximum rate of 120 mcg/kg/hr.
If a patient body weight of 50 kg is programmed and the concentration is 0.1 mg/mL, the bolus maximum rate of 100 mcg/kg/hr will translate to a rate of 60 mL/hr, and the bolus rate will be limited to 60 mL/hr.
Bolus Programming
A bolus may be programmed during an infusion without interrupting the normal delivery.
Note: If a protocol is selected, the bolus function may not be allowed. If the protocol disallows the bolus function, the pump beeps and displays the message
NOT ALLOWED when BOLUS is pressed.
Note: If a protocol has been selected, a default bolus value may appear automatically.
Use ▲ and ▼ or the digit keys to modify this value as necessary, and press CON-
FIRM to complete the data entry.
1.
Press BOLUS . The status display changes to the Bolus display.
• If there is no programmed bolus size, the pump displays PRESS EDIT TO SET
BOLUS SIZE . Proceed to step 2.
• If a bolus size has been programmed, the pump displays BOLUS READY . Press
START within 10 seconds to begin bolus delivery, or proceed to step 2 to change the programmed bolus size, or press CONFIRM to save the bolus data and return to the normal display.
Page 56 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Additional Functions and Operations
2.
Press EDIT , then use the ▲ , ▼ or digit keys as necessary to program the bolus size.
When a bolus is fully programmed, the pump displays BOLUS READY .
3.
Press START to begin bolus delivery immediately or press CONFIRM to save the bolus data and return to the normal display.
Bolus Programming Notes
For mL/min mode, the bolus rate is displayed in mL/min. For all other modes it is displayed in mL/hr.
Pressing BOLUS during the BOLUS READY or PRESS EDIT TO SET BOLUS SIZE display cancels bolus programming and restores the normal display immediately.
A bolus can be programmed and delivered from Standby state, even when the infusion has not been fully programmed. The minimum program data required for bolus operation are: Syringe manufacturer and size, body weight (if applicable), and drug concentration (if applicable). In addition, for CUSTOM DILUTION mode, all remaining fields must be programmed.
Bolus Delivery
During bolus delivery, the text display reads BOLUS RUNNING and all three RUN lights flash simultaneously.
1.
To halt bolus delivery before the full bolus amount has been delivered, press STOP .
The normal display and operating state resumes. The text field reads BOLUS
STOPPED for a few seconds.
2.
If the pump is allowed to deliver the full bolus amount, the normal display and operating state are automatically restored. The text field reads BOLUS COMPLETE for a few seconds. In addition, if the BOLUS COMPLETE alert is enabled, the pump sounds two beeps and flashes the ALERT light.
Bolus Delivery Notes
During bolus delivery, the display shows the amount of the bolus that has been infused.
If bolus delivery is started during an infusion, the infusion resumes automatically once the bolus delivery ends. If the bolus delivery is stopped prematurely by pressing STOP , the pump displays DOSE RESUMES for a few seconds.
The bolus volume is added to the TOTAL ML display after the bolus operation stops.
Bolus Review
Press BOLUS to review the Bolus display. The bolus information will be displayed for about 10 seconds, or the normal display can be restored immediately by pressing BOLUS or CONFIRM .
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 57
Additional Functions and Operations
Bolus Editing
The bolus size can be edited during bolus programming and after a bolus has been fully programmed. A bolus may be edited during an infusion without interrupting normal delivery.
1.
Press BOLUS to display the PRESS EDIT TO EDIT BOLUS SIZE prompt.
2.
Press EDIT to open the bolus size field.
3.
Use ▲ and ▼ or the digit keys to modify the value as necessary, and press CONFIRM to complete the edit.
The pump displays BOLUS READY .
4.
Press START to begin bolus delivery immediately or press CONFIRM to save the bolus data and return to the normal display.
Repeating a Bolus Delivery
To repeat a bolus delivery:
1.
Press BOLUS to display the bolus size.
2.
Review the bolus program data.
3.
Press START .
Changing the Syringe
To continue an infusion when a syringe becomes empty or to flush a line, the syringe may be replaced when the pump is in the Standby state.
If the new syringe does not have the same barrel size as the currently programmed syringe:
1.
The pump displays VERIFY (NAME) , where “name” refers to the currently programmed syringe manufacturer.
Use
▲
and
▼
as needed to select the correct syringe manufacturer and press
CONFIRM to complete the entry.
2.
The pump displays VERIFY (SIZE) where “size” is the new syringe size recognized by the pump.
Press CONFIRM after verifying that the pump is displaying the correct syringe size.
Note: Changing syringe size may invalidate the contents of other fields (e.g. VOL LIMIT ).
In such cases, the pump automatically clears the necessary field(s). The infusion cannot be continued until all cleared fields have been reprogrammed.
Page 58 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Additional Functions and Operations
● If the syringe is improperly mounted, the pump displays one of the following alert or alarm messages: CHECK BARREL , CHECK FLANGE , or CHECK PLUNGER . Examine the syringe and correct the indicated problem.
If a combination of such conditions is sensed, the pump displays a CHECK SYRINGE alert or alarm message instead.
Lock and Auto Lock Functions
The Lock function serves two purposes. First, it reduces risk of accidental key activations by disabling all keys except LOCK . Second, the Lock function makes it convenient to program the pump in advance of delivery by disabling the audio portion of the PUMP IS
IDLE alert or alarm.
The Lock function is activated by pressing LOCK . Pressing LOCK again restores normal operation. The LOCK light remains on while the Lock function is active.
Note: The pump will not allow the Lock function to be active during a LINE OCCLUDED alarm condition.
The Auto Lock feature can be enabled through pump configuration. During an infusion, the Auto Lock feature automatically sets the Lock function two minutes after the last key is pressed. This feature is available only while the pump is in Run state.
If the pump enters Standby State because of an alarm condition, the keypad is automatically unlocked.
The Total Delivered Display
mgrams mLmcg mL
The Total Delivered display indicates the accumulated total volume (in mL) that has been delivered. Some infusion modes also show the total delivery in other units, such as mcg, mg, or grams.
The Total Delivered display cannot be programmed, but can be cleared (reset to zero). To clear the total delivered information, press
▲
or
▼
to move to the TOTAL m L field. The annunciators will flash, then press EDIT and CLR to clear the display. After resetting the
Total Delivered display to zero, the previous value cannot be restored.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 59
Additional Functions and Operations
Note: Purge volumes are not included in the total, because purged fluid is not intended to be infused into the patient.
Display totals are rounded up at the digit displayed at the far right.
Example : if a total delivered value is 0.005, the pump displays 0.01
. If a total delivered value is
30.15, the pump displays 30.2
.
If an accumulated total becomes too large to display (that is, over 999), the field changes to EEE .
Expanded Data Display
The status display generally displays only the most important infusion information.
Pressing CONFIRM when the pump is fully programmed causes the pump to temporarily display additional program, syringe, and battery voltage information. Press CONFIRM again to restore the normal display immediately. If no key is pressed, the normal display automatically resumes in a few seconds.
Backlight Feature
The status display backlight is lit whenever the charger is connected and the pump power is on.
Control of the backlight when the charger is not connected is determined by the SET
BACKLITE configuration option group. When the charger is not connected and this option is enabled for use, the pump turns on the backlight for the preconfigured time period following a keypress.
Page 60 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Configuration
Chapter 1
The pump can be custom-configured to meet particular needs, or to simplify infusion programming. Instructions for setting configuration options are given in the Auto Syringe
AS50 Infusion Pump Technical Manual.
Configuration Review
The Configuration Review feature allows the pump’s configuration to be examined without risk of accidental alteration. This feature is activated by entering the code number 123 at the initial prompt ( SELECT MODE
▲▼
or SELECT PROTOCOL
▲▼
). The error beeps that occur when 123 is entered are provided to discourage patient tampering and may be ignored.
Date and Time Display
After 123 is entered, the date and time are briefly displayed. The time is displayed in hours and minutes, separated by a decimal point ( HH.MM
).
Software and Configuration Version Display
Following the time and date display, the software and configuration version numbers are briefly displayed.
The first line of the text display indicates the pump software version. The second line is divided into two fields. The left field displays the factory configuration version, and the right field displays a four character checksum that uniquely identifies the current configuration (including protocols), as shown below.
Example : The display for software version 6.00, factory configuration version D1, and a current configuration checksum of 3AC5, would be shown on the display as follows: software version
→ factory configuration → D 1
V 7.0 0
3 A C 5 ← configuration checksum
The checksum field allows a high-confidence rapid comparison of configurations between two pumps.
Reviewing Configuration Options
The configurable options are organized into several groups, each of which begins with a prompt that requires a “yes” or “no” response. Each option group is identified by name in the text display field (for example: VIEW MODES ). The syringe manufacturer field displays
Y (for “yes”), which represents the pending response.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 61
Configuration
To begin reviewing the options within the current group, press CONFIRM to accept the pending Y response. The text field will display only the options that are configured. Press
CONFIRM to view each additional option. When the last option in a group has been displayed, the next option group prompt is displayed (for example: VIEW SYRINGES ).
To skip an option group, press either
▲
or
▼
to change the Y to N (for “no”). Press
CONFIRM to advance to the next option group.
To review the Modes group, the initial prompt is VIEW MODES . Answering Y (the default response) opens up that group for review. Answering N skips that group and prompts for the manufacturers group ( VIEW MFRS ).
The following table shows the configuration groups, and the options within each group.
Note: Inactive options are not shown in Configuration Review mode.
Configuration Review Table
GROUP PROMPT
VIEW LIBRARY
VIEW MODES
VIEW SYRINGES
VIEW DEFAULTS
VIEW MISC
VIEW ALERTS
GROUP SELECTIONS
List of enabled protocol libraries.
ML/HR
ML/MIN
UNITS/HR
UNITS/MIN
MUNITS/HR
MUNITS/MIN
B-D
MG/HR
MG/MIN
MCG/HR
MCG/MIN
MG/KG/HR
MG/KG/MIN
MONOJECT
MCG/KG/HR
MCG/KG/MIN
CUSTOM DILUTION
SINGLE DOSE
MANUAL SCHEDULE
AUTO SCHEDULE
TERUMO
DFAULT LIBRARY
DFAULT PROTOCOL
REMOTE CONTROL
LIBRARY
SIZE OVERRIDE
IDLE ALARM
NEAR END
DFAULT MODE
DFAULT MFR
PSI RANGE
AUDIO RANGE
RATE RANGE
AUTO LOCK
SET BACKLITE
ENABLE DB9 PORT
BOLUS COMPLETE VOLUME LIMIT
DOSE COMPLETE
Page 62 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Configuration
Configurable Options
The option groups and configurable options are listed here and in the table above. Each option is described in the following paragraphs. The option is displayed exactly as listed in the paragraph title unless otherwise noted.
Protocol Library Group (
VIEW LIBRARY
)
This option group displays the drug protocols that are enabled. The pump displays a
VIEW (library name) prompt for each enabled protocol library. A pending Y response is also displayed. To select a library, press CONFIRM to accept the pending response. To skip a library, press
▲
or
▼
to change the Y to N and press CONFIRM .
When a library has been selected, the pump displays a VIEW (protocol name) prompt for each enabled protocol. A pending Y response is also displayed. To begin reviewing a protocol, press CONFIRM to accept the pending response. To skip a protocol, press
▲
or
▼
to change the Y to N and press CONFIRM .
For each selected protocol, the default mode (if any) is displayed in the text field (for example, MG/KG/HR ), and concentration, dose and rate default values (if any) are shown in their usual fields.
Press CONFIRM to view the default bolus size. BOLUS SIZE is displayed in the text field.
Press CONFIRM again to view the bolus maximum rate. BOLUS MAX RATE is displayed in the text field.
Note: Dashes are displayed whenever a default value is not specified by the protocol.
If the protocol specifies bolus prevention, NO BOLUS is displayed instead of
BOLUS SIZE .
If no defaults are specified by the protocol, ALL MODES is displayed when the protocol is initially selected.
Press CONFIRM to advance to the next protocol. When the last protocol in a library has been presented for viewing, the next VIEW (library name) prompt is displayed.
See the Auto Syringe AS50 Infusion Pump Technical Manual for instructions on creating and editing protocols.
Modes Group (
VIEW MODES
)
This option group displays the name of each enabled infusion mode. Each listing shows only the mode name (for example ML/HR ) in the text field. For a list of the supported infusion modes, see the “Configuration Review Table” on page 62.
Syringes Group (
VIEW SYRINGES
)
This option group displays the enabled syringe sizes for each enabled syringe manufacturer.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 63
Configuration
For each enabled manufacturer, the pump displays a VIEW (name) prompt, where (name) is the manufacturer’s name. A pending Y response is also displayed. To begin reviewing the enabled syringe sizes for that manufacturer, press CONFIRM to accept the Y response.
Syringe size descriptions are displayed in the text field. Press CONFIRM to view each additional syringe.
To skip a manufacturer, press either ▲ or ▼ to change the Y to N and press CONFIRM .
The available manufacturer names and sizes are listed in the “Syringe Data Tables” on page 25.
Defaults Group (
VIEW DFAULTS
)
This option group displays the default options that are enabled. The pump allows four default options: Default Library ( DFAULT LIBRARY ), Default Protocol ( DFAULT
PROTOCOL ), Default Mode ( DFAULT MODE ) and Default Manufacturer ( DFAULT MFR ).
The defaults enable the pump to recall the selection that was used for the previous infusion, and to provide that selection in response to the first
▼
or
▲
keypress. This feature is especially useful when a particular protocol library, protocol, infusion mode, or syringe manufacturer is likely to be used repeatedly.
Miscellaneous Group (
VIEW MISC
)
This displays the miscellaneous options that are enabled. The Miscellaneous group options are: Remote Control, Protocol Library, Size Override, Occlusion Sensitivity, Audio
Range, Rate Range, Auto Lock, Backlight, and Enable DB9 Port.
Remote Control
This option allows the operator to control the pump from a remote device, such as a computer. A standard 9-pin connector (DB9) located on the rear of the pump is provided for an RS-232(C) communications link between the pump and the computer. See the
“Auto Syringe AS50 Infusion Pump Programmer’s Manual” for details.
Protocol Library (
LIBRARY
)
This option allows the operator to select from one or more protocol libraries containing frequently used program data. Protocols are defined by authorized personnel according to procedures detailed in the “Auto Syringe AS50 Infusion Pump Service Manual.”
Size Override
SIZE OVERRIDE modifies the syringe programming process by allowing the operator to select a syringe size that differs from the size detected by the pump. Instead of a VERIFY xx ML prompt, the pump displays SELECT SIZE
▼ ▲
. When
▼
or
▲
is pressed, the pump displays the syringe size that best matches the syringe. However, the operator can press
▼
or
▲
to select a different size. When CONFIRM is pressed to complete the syringe selection, if the syringe size selected appears different from the detected syringe size, the pump displays a VERIFY AGAIN prompt and the operator must press CONFIRM again to complete the selection.
Page 64 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Configuration
Note: To restart the syringe selection process, press CLR before pressing CONFIRM to complete syringe selection.
Occlusion Sensitivity (
PSI RANGE
)
This option sets the occlusion detection system operating range. Three choices are available: H (Highest back pressure), M (Medium setting), L (Lowest back pressure). For more details see the “Auto Syringe AS50 Infusion Pump Technical Manual.”
Audio Range
The audio volume level can be configured for: H (High) or L (Low). The recommended setting is H . The quieter L setting should only be used in noise-sensitive environments where the infusion is constantly monitored.
Rate Range
RATE RANGE can be configured for H (High, up to 438 mL/hr), M (Medium, 120 mL/hr) or
L (Low, 15 mL/hr). The maximum delivery rate will be limited to the lesser of the Rate
Range option setting or the syringe maximum (see the “Syringe Data Tables” on page 25).
The minimum delivery rate is as shown in the table, regardless of the Rate Range setting.
Auto Lock
If an infusion has been programmed and the pump is in the Run state, the Auto Lock feature automatically turns on the Lock function two minutes after the last keypress.
Backlight (
VIEW BACKLITE
)
The status display backlight is lit whenever the charger is connected and the pump is powered on. The VIEW BACKLITE option turns the backlight on after a keypress when the charger is not connected. The backlight turns off automatically after a preset time period elapses (0 to 60 minutes). Press CONFIRM to view the current setting.
Enable DB9 Port
When this option is set to Y (yes), the standard 9-pin connector (DB9) located on the rear of the pump is enabled as the serial RS-232(C) communication link (to allow remote control operation). When this option is set to N (no), the RS-232(C) serial communication link is disabled.
Alerts Group (
VIEW ALERTS
)
This option group displays the alert and alarm options that are enabled. The Alerts group options are: IDLE ALARM, NEAR END ALERT, BOLUS COMPLETE ALERT, DOSE
COMPLETE ALARM , and VOLUME LIMIT ALARM .
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 65
Configuration
Idle Alarm
This option enables the pump to issue a continuous audible alarm when a PUMP IS IDLE condition occurs. If the option is disabled, the pump beeps 15 times at two minute intervals when a PUMP IS IDLE condition occurs.
Near End Alert (NEAR END)
This option enables the pump to issue an alert during an infusion prior to an expected
VOL LIMIT or EMPTY alarm. The option specifies the number of minutes prior to the alarm that the expected alert will be issued and the number of beeps in the alert.
The form of a Near End alert display is xx MIN VOL LIM , xx MIN EMPTY or xx MIN (protocol).
The value xx is the configured number of minutes. The first and second forms apply to impending VOLUME LIMIT or EMPTY conditions when no protocol has been selected. The third form applies to either condition when a protocol has been selected. Press CONFIRM to view the current settings.
Note: When an infusion is started, no NEAR END alert will occur if fewer than the configured number of minutes remain before an EMPTY or VOL LIMIT alarm is due to occur. For example, if the NEAR END alert is configured for 10 minutes and 9 minutes remains before the volume limit is reached, no NEAR END alert will occur when START is pressed.
Bolus Complete Alert (BOLUS COMPLETE)
This option enables the pump to issue an alert when a bolus completes.
Dose Complete Alarm (DOSE COMPLETE)
This option enables the pump to issue an alarm when a dose completes in SINGLE
DOSE mode.
Volume Limit Alarm (VOLUME LIMIT)
This option applies to mL/hr and mL/min modes only. During initial programming, the pump prompts the operator to enter a volume limit value in mL. The Volume Limit feature triggers an alarm after the pump has delivered the pre-programmed volume of drug. With this option enabled, a NEAR END alert is issued when the Volume Limit will be reached within a preconfigured time period at the current rate of infusion. See “Near End Alert” on page 66.
Identifier
The Identifier is not a configuration group, but is included here for reference purposes.
The Identifier is a short message (up to eight characters) that is displayed for three seconds following the Lamp Test. The Identifier can be created or edited during the pump configuration process.
Page 66 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Configuration
Terminating Configuration Review
When the last configuration group option has been reviewed, the pump automatically restarts the Lamp Test, as if it had just been turned on.
To cancel Configuration Review at any time, turn the pump off.
Custom Configuration
The pump can be reconfigured as often as desired, to select features that best suit the needs of the healthcare facility.
If there is any question regarding the pump’s current configuration or applicability for a particular clinical application, the operator and healthcare professionals should verify that the settings are appropriate.
While no special tools are needed, the configuration should be changed only by authorized personnel.
Factory Standard Configuration
The factory standard configuration is:
Protocol Library:
Modes:
See the chart on the following page.
ML/HR
SINGLE DOSE
MCG/MIN
MCG/KG/MIN
MCG/KG/HR
Manufacturers and Sizes: B-D 1, 3, 5, 10, 20, 30, 60 mL Plastic
Monoject 1, 3, 6, 12, 20, 35, 60 mL
Terumo 1,3,5,10, 20, 30, 60 mL
Defaults: All enabled
Misc:
Alerts:
PSI RANGE = M (Medium)
AUDIO RANGE = H (High)
RATE RANGE = H (High)
BACKLITE = 2 MINUTES
LIBRARY=N
IDLE ALARM enabled
MINALERT= 10, 5 (10 Minutes, 5 Beeps)
DOSE COMPLETE enabled
VOLUME LIMIT enabled
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 67
Configuration
Protocol Libraries
Library Name
ANALGESC
ANES-HYP
CARDIO
MUSRELAX
Drug Name
Alfentanil
Remifentanil
Sufentanil
Propofol
Dobutamine
Dopamine
Isoproterenol
Nitroglycerine
Nitroprusside
Norepinephrine
Atracurium
Mivacurium
Rocuronium
Succinylcholine
Vecuronium
Protocol
Name
Default
Mode
ALFENTAN mcg/kg/min
REMIFENT mcg/kg/min
SUFENTAN mcg/kg/hr
PROPOFOL mcg/kg/min
DOBUTAMI mcg/kg/min
DOPAMINE mcg/kg/min mcg/min ISOPROTE
NITROGLY mcg/min mcg/kg/min NITROPRU
NOREPINE
ATRACURI
MIVACURI mcg/min mcg/kg/min mcg/kg/min
ROCURONI
SUCCINYL
VECURONI mg/kg/min mg/min mcg/kg/min
Default
Concentration
(mg/mL)
Bolus
Allowed
0.05
No
0.05
0.05
10
5
3.2
0.004
0.4
0.2
0.004
0.5
2
10
2
0.2
No
No
No
No
No
No
No
No
No
No
No
No
No
No
NONE Protocol
Selection Option
Yes
Yes
Yes
Yes
For NONE Library Selection Option, see “Using Protocols” on page 32.
Page 68 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Pump Maintenance
Chapter 1
Charging the Battery
The battery charger cord plugs into the bottom of the pump.
Note: The plug and receptacle are shaped to fit only one way. DO NOT force the plug into place.
For the best battery operating time, the battery should be charged for at least 16 hours before using the pump. Disconnect the charger if the pump is not to be used within the next 72 hours. Prolonged overcharge and high temperature shorten the battery operating time and service life.
The battery voltage may be displayed by pressing CONFIRM briefly during the Lamp Test, or by pressing CONFIRM after the pump has been programmed.
!
Note: Battery voltage display is only an approximate indication of the battery’s remaining charge. If the battery voltage is less than 6.1 volts, or is dropping rapidly, assume that the charge remaining in the battery is minimal. Recharge the battery.
Do not charge or store the pump in direct sunlight, or under excessively warm conditions.
Replacement of the battery every 16 months is recommended for optimal pump performance.
Use only “900 Series” chargers or other accessories that are labeled specifically for use with Auto Syringe AS40A or AS50 Series Infusion
Pumps. The standard battery charger (USA) is Model #C-AS40-01-900.
V
OUTLET
CHARGER
CHARGER
PLUG
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 69
Pump Maintenance
IV Pole Clamp
Cautions-Federal (U.S.A.) law r e n
I.V
60015 e
5
The pump is supplied with a detachable, adjustable pole clamp which allows it to be mounted to a standard vertical IV pole. The pole clamp can also be adjusted to allow the pump to be used on a horizontal pole.
To attach the pole clamp to the pump, first slide the clamp in place, so that the MOUNTING PIN CUTOUTS (1) on the clamp engage the MOUNTING
PINS (5) built into the rear of the pump. Tighten the MOUNTING
SCREW (4) into the threaded
RECESS (6) just below the battery cover. Tools are not required for this operation.
The JAW (3) clamps the pump to the
IV pole. Turn the KNOB (2) counterclockwise to open the jaw, and clockwise to close it. Tighten the knob by hand only. Do not use tools to tighten the knob.
6
Page 70 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
IV Pole Loop
The pump includes an IV pole loop built into the top of the pump. This loop may be swiveled as required, or folded down and out of the way when not in use.
Cleaning
!
All the external moving parts must be kept clean of sticky films and other foreign substances.
If spillage onto external surfaces occurs, clean the surfaces immediately by the method described below.
Routine Cleaning and Disinfection Techniques
Clean or disinfect the exterior of the device with a soft cloth or cotton swabs, sparingly dampened with any of the recommended cleaning agents listed below. Remove the rubber bumpers and clean or disinfect them separately.
Remove residual cleaner or disinfectant from all accessible surfaces with a soft cloth or cotton swabs sparingly dampened with water, and allow to dry before replacing the rubber bumpers.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 71
Pump Maintenance
Acceptable cleaning or disinfecting agents are:
Cleaner
Soapy water
Isopropyl alcohol
10% solution of chlorine bleach and water
ASCEND germicidal detergent
Cidex Formula 7 sterilizing and disinfecting solution
LpHSE, Septisol germicidal detergents
TOR, Hi-Tor Plus germicidal detergents
Manufacturer
Various
Various
Commercially available
Huntington Professional
Products
Johnson & Johnson
Steris Corporation
Huntington Professional
Products
!
!
!
Note: Follow the manufacturer’s dilution instructions for concentrated cleaning and disinfecting agents.
Recommendation of the above cleaning or disinfecting agents is based on their compatibility with the materials of which the pump is constructed. No representation is intended as to the effectiveness of any cleaning agent.
Surfaces to be disinfected must be free of gross (visible) foreign matter. Cleaning the device before disinfection is recommended.
Some surfaces are not accessible with the above method. If spillage has occurred onto inaccessible surfaces or into the mechanism, the device should be Service Cleaned as soon as possible. Service Cleaning methods are described in the Auto Syringe AS50 Infusion Pump Service
Manual.
Attempts to clean or disinfect internal parts, autoclaving or sterilization by ethylene oxide gas will damage the pump and void the warranty.
Do not use the following chemicals on the device, as they will damage the front panel: acetone, acetoaldehyde ammonia, benzene, hydroxytoluene, methylene chloride, and ozone, Do not use cleaners containing n-alkyl dimethyl ethylbenzyl ammonium chloride unless they appear in the list of recommended cleaners in the Pump Maintenance section of this manual.
Page 72 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Conditions Requiring Service Cleaning
Service cleaning is a procedure for removing foreign matter from areas that are not accessible through routine cleaning methods. Service cleaning must only be performed by qualified individuals. Instructions for service cleaning may be found in the Auto Syringe
AS50 Infusion Pump Service Manual.
Conditions which require service cleaning include:
● Spillage into the pumping mechanism, or into the barrel clamp, flange clamp, or plunger driver mechanism.
● Accumulation of foreign matter in areas not accessible to routine cleaning.
● Excessive syringe-related alarms or alerts, such as CHECK SYRINGE, CHECK
PLUNGER, CHECK FLANGE , or CHECK BARREL .
● Sticking of the plunger driver, the barrel clamp, or flange clamp.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 73
Pump Maintenance
Periodic Maintenance
Daily Check
This Daily Check should be performed prior to programming an infusion. If the pump fails any part of the Daily Check, take it out of service.
1.
Before using the pump, check for slippage of the plunger driver by applying gentle back pressure.
2.
Check the plunger driver for free travel through its full range.
3.
Check the barrel clamping mechanism for proper locking, and check that it opens and closes freely when either barrel clamp release is pressed.
4.
Check that the plunger assembly opens freely to its full travel and returns to a fully closed position without assistance.
If connecting the charger, make sure that the charger plug and socket are clean and fully engaged, and that the ON CHARGE light is lit when the charger is plugged in.
Battery Replacement
In general, the service life of rechargeable batteries is based on many factors, including construction, size, usage, frequency of charging and discharging, and depth of discharge prior to recharge. To minimize the chances of experiencing a problem with rechargeable batteries, it is advisable to replace them before their service life is expended.
Baxter advises that pump battery replacement be made part of a regularly scheduled maintenance program. Baxter recommends that the AS50 Pump’s rechargeable battery be replaced every 16 months.
Page 74 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Display, Battery, and Syringe Sensor Check
SYRINGE m f r s i z e
INFUSE
RATE
OVER
BODY
WT
EVERY
CONC.
NEXT
DOSE
IN
T FINAL
VOL LIMITVOL
T DR UG
VOL m L m L min hr kg hr :min
Units mg m L hr :min mg U hr kg hr :min mgrams mLmcg
1.
Be sure that no syringe is mounted, and that the barrel clamp is closed.
2.
Press and hold the STOP key during the normal power-up Lamp Test. Compare the display to the illustration. Verify that all annunciators, decimal points and digits are legible.
3.
Be sure that all lights on the Status
Panel light up during the Lamp Test.
Note: If the charger is plugged in, the ON
CHARGE light will be on before, during and after the lamp test.
4.
Release the STOP key.
m L
Note: If CHECK BARREL, CHECK FLANGE or
CHECK PLUNGER is displayed, one of the syringe mounting sensors may be malfunctioning. Make sure the plunger rod, flange and barrel capture mechanisms are free.
Turn the pump off and back on again. If the error message is repeated, remove the pump from service.
Keypad Check
1.
After the pump completes the power-up Lamp Test, press every key except
▲
,
▼
, and
CONFIRM in any order, one key at a time. Each keypress should result in a short beep.
2.
Press any two (or more) keys simultaneously. There should be no response.
3.
Press and hold any key (except
▲
,
▼
, START, STOP, and CONFIRM ). There should be only one beep, no matter how long the key is held down.
4.
Press and hold any key (except
▲
,
▼
, START, STOP, and CONFIRM ), then press any other key. There should be no response to the second keypress.
5.
Press and briefly hold down START and STOP , one at a time. The pump should sound repeated beeps while each key is held down. Holding down START for a prolonged period of time will cause a failsafe alarm.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 75
Pump Maintenance
Battery Voltage Check
Note: Battery voltage display is only an approximate indication of the battery’s remaining charge. If the battery voltage is less than 6.1 volts, or is dropping rapidly, assume that the charge remaining in the battery is minimal. Recharge the battery.
To display the battery voltage, press CONFIRM after the pump is fully programmed. The battery voltage should be more than 6.1 V if the pump is to be operated on battery power.
A charged battery displays a battery voltage of approximately 7.0 volts.
In Case of Difficulty
Check the IV extension set carefully to be sure the tubing is not pinched, and that there are no obstructions to prevent proper fluid flow.
Problems associated with a particular syringe size or manufacturer are likely to be caused by variations in syringe dimensions. Typical symptoms are an inability to completely empty certain syringes, and LINE OCCLUDED alarms when the syringe becomes empty.
If you apply tape, paper stickers, or labels to the syringe, the apparent diameter of the syringe barrel may increase, which can impair the accuracy of the Syringe Recognition system.
Syringe Recognition problems may be caused by improperly mounted syringes. Excess clamping force can crush the syringe barrel, which results in an erroneous barrel diameter reading.
Frequent CHECK BARREL messages may be caused by improperly centering the syringe in the barrel clamp cradle.
If the pump fails to operate as described, contact the Baxter Healthcare Corporation
Andover Service Center for assistance at:
1-800-343-0366 - Customer Service
1-800-634-4602 - Technical Service
Page 76 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Alerts and Alarms
Chapter 2
07-19-B3-430
The pump notifies the operator when important conditions occur:
● Alert:
A condition has occurred which requires operator attention. The following indications occur:
• The pump continues to operate
• The ALERT light flashes
• The pump beeps
• The front panel and status lights may display additional information
● Alarm:
Immediate action is required in order to proceed with operation. The following indications occur:
• The pump stops delivering the infusion
• The ALERT light flashes
• The pump beeps
• The front panel and status lights may display additional information
● Failsafe Alarm:
A malfunction has occurred that could affect the safe operation of the device. The pump goes into a “failsafe shutdown.” The following indications occur:
• The pump stops delivering the infusion and cannot be restarted. The infusion must be reprogrammed after the problem is corrected.
• The SYSTEM light is on continuously
• The pump beeps
• The front panel and status lights may display additional information
• The pump must be turned off and back on to clear the alarm
If any of the aforementioned conditions occur, refer to the charts on the following pages for instructions on correcting the alert or alarm condition.
Auto Syringe AS50 Infusion Pump Operator’s Manual Page 77
Alerts and Alarms
Alerts
Message
BOLUS COMPLETE
(may be accompanied by 2 beeps and flashing ALERT light)
CHARGE IS LOW
(10 beeps, flashing BATTERY light)
Cause
The bolus has completed.
CHECK BARREL
(single beep)
CHECK FLANGE
(single beep)
CHECK PLUNGER
(single beep)
CHECK SYRINGE
(single beep and Standby light flashes)
DATA MISSING
(single beep)
EMPTY LIBRARY
(single beep)
FINISH EDITING
(single beep)
Correction
None required.
The battery needs charging. If a charger is not connected, or is not charging the battery, this alert repeats every 15 minutes.
The barrel clamp is not fully closed on the syringe barrel, or the syringe is not properly seated in the cradle.
The syringe flange is not seated in the flange slot.
Connect the charger. When the battery charge reaches the normal operating range, the alert self-cancels.
If the battery voltage is very low, the alert may repeat once or twice after the charger is connected.
Make sure the syringe is centered in the barrel clamp cradle, and that the syringe flange is in the barrel clamp flange slot. To ensure proper seating, use the heel of the hand to press the barrel clamp firmly against the syringe. Motor movement is not allowed until the syringe is properly mounted.
Remove the syringe and remount it.
Make sure the syringe is centered in the barrel clamp cradle, and that the syringe flange is in the barrel clamp flange slot. To ensure proper seating, use the heel of the hand to press the barrel clamp firmly against the syringe. Motor movement is not allowed until the syringe is properly mounted.
The plunger capture mechanism has not captured the syringe plunger.
At least two of the three syringe mounting points (flange, barrel, plunger) are not correctly positioned.
Grasp the finger grip on the plunger driver and pull it all the way out.
Slide the driver down until it contacts the syringe plunger and push the driver in firmly. Motor movement is not allowed until the syringe is properly mounted.
Check flange, barrel, and plunger, and ensure that they are mounted correctly. Motor movement is not allowed until the syringe is properly mounted.
Enter the additional data required. START, PURGE , or BOLUS was pressed when the program information is incomplete.
The selected protocol library contains no protocols.
Pressing START, PURGE , or
BOLUS while a field is open for editing.
Select a different library.
Complete the editing process.
Page 78 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Message
KEYPAD LOCKED
(single beep)
LOAD SYRINGE
(single beep)
NO LIBRARY
(single beep)
NO LIBRARY IS ENABLED
(single beep)
NOT ALLOWED
(single beep)
OUT OF RANGE
(single beep)
PUMP IS IDLE
(15 beeps and ALERT light flashes)
PUMP LIMIT
(single beep)
SIZE MISMATCH
(single beep)
SIZE UNKNOWN
(single beep)
USE DIGITS
(single beep)
The BOLUS range.
Cause
Pressing a key while the Lock function is active.
Attempt to start an infusion without a syringe in place.
All three syringe detection sensors have failed.
VIEW LIBRARY no libraries exist.
VIEW LIBRARY
was selected and was selected and no libraries are enabled.
key was pressed when the bolus feature is not allowed by the selected protocol.
No bolus is possible for the current combination of syringe, body weight, and drug concentration.
The pump has been in Standby state for at least 2 minutes. Also occurs if there is no keypress within
15 seconds after the infusion is stopped.
Attempting to program a value outside the pump’s acceptable
Correction
Press LOCK to unlock the keypad.
Mount a syringe after manually purging.
If alert continues to occur, have the pump serviced.
Press CONFIRM to proceed to
VIEW MODES prompt.
Press CONFIRM to proceed to
VIEW MODES prompt.
Use the NONE protocol library selection or the NONE protocol selection (if available) if the bolus feature is required.
Use a different syringe or drug concentration if the bolus feature is required.
The alert is self-cancelling. The
Lock function suppresses the audio portion of this alert. Note: If the keypad is locked, there will be no beeps, but the PUMP IS IDLE message will still appear.
The pump displays the nearest acceptable value. Press CONFIRM to accept displayed value, or enter a different value and press CONFIRM .
The detected syringe size is different from the programmed size.
The pump cannot identify the syringe size for the selected manufacturer.
Attempt to use
▲
or
▼
to initialize a field containing dashes.
Check to be sure the correct manufacturer has been programmed. Press CLR if necessary and reprogram the syringe manufacturer and size.
Check to be sure the correct manufacturer has been programmed. Ensure that the barrel clamp is fully closed. Press CLR if necessary and reprogram the syringe manufacturer and size.
Use digit keys instead.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 79
Alerts and Alarms
Message
VERIFY AGAIN
(single beep)
▲▼
IN USE
(single beep)
<xx MIN EMPTY or
<xx MIN VOLUME LIMIT or
<xx MIN (protocol)
(configurable number of beeps.
ALERT light flashes until syringe is empty or volume limit is reached)
Cause
The detected syringe size is different from the programmed size.
An attempt was made to use digit keys to edit a number that was modified with an arrow key.
At the present rate of infusion, approximately xx minutes remain until the syringe will be empty or the volume limit will be reached. If a protocol has been selected, the protocol name will be displayed instead of END or VOLUME LIMIT .
Correction
Press
▲ or
▼ to change size or
CONFIRM to complete size override.
Continue programming with the arrow keys, or clear the field and reprogram it with digit keys.
Pressing any key silences the audio portion of the alert. The alert is reset when a filled syringe is installed, or when the plunger driver is moved back.
Note: Since this function is sensitive to variations in syringe dimensions, the exact timing of this alert may vary.
Page 80 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Alerts and Alarms
Alarms
Message
CHECK BARREL
(continuous beeps and flashing
ALERT light)
CHECK FLANGE
(continuous beeps and flashing
ALERT light)
CHECK PLUNGER
(continuous beeps and flashing
ALERT light)
CHECK SYRINGE
(continuous beeps and flashing
ALERT light)
DOSE COMPLETE
(continuous beeps and flashing
ALERT light)
DOSE DUE
(continuous beeps and flashing
ALERT light)
Cause
During motor movement the pump has detected that the barrel clamp is not fully closed on the syringe barrel, or the syringe barrel is not properly seated in the cradle.
Correction
Pressing any key silences the audio portion of the alarm. Make sure the syringe is centered in the barrel clamp cradle, and that the syringe flange is in the barrel clamp flange slot. To ensure proper seating, use the heel of the hand to press the barrel clamp firmly against the syringe. Motor movement is not allowed until the syringe is properly mounted.
During motor movement the pump has detected that the syringe flange is not seated in the flange slot.
Pressing any key silences the audio portion of the alarm. Remove the syringe and remount it. Make sure the syringe is centered in the barrel clamp cradle, and that the syringe flange is in the barrel clamp flange slot. To ensure proper seating, use the heel of the hand to press the barrel clamp firmly against the syringe. Motor movement is not allowed until the syringe is properly mounted.
During motor movement the pump has detected that the plunger capture mechanism has not captured the syringe plunger.
At least two of the three syringe mounting points (flange, barrel, plunger) are not correctly positioned.
Pressing any key silences the audio portion of the alarm. Grasp the finger grip on the plunger driver and pull it all the way out. Slide the driver down until it contacts the syringe plunger. Push in firmly. Motor movement is not allowed until the syringe is properly mounted.
Pressing any key silences the audio portion of the alarm. Check flange, barrel, and plunger, and ensure that they are mounted correctly. Motor movement is not allowed until the syringe is properly mounted.
Pressing any key silences the alarm.
(Single Dose Infusion Mode only)
The programmed dose has been delivered.
(Manual Schedule Infusion Mode only) The NEXT DOSE IN timer has reached 0:00, and the next dose is due to be delivered.
Pressing any key silences the audio portion of the alarm. Press START to deliver the scheduled dose.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 81
Alerts and Alarms
Message
EMPTY or
EMPTY (protocol )
(continuous beeps and flashing
ALERT light)
FINISH EDITING
(continuous beeps and flashing
ALERT light)
LINE OCCLUDED
(continuous beeps and flashing
ALERT light)
PUMP IS IDLE
(continuous beeps and flashing
ALERT light)
VOLUME LIMIT
(continuous beeps and flashing
ALERT light)
Cause
The syringe is empty.
Correction
Pressing any key silences the audio portion of the alarm. Release and slide the plunger driver back at least
1.6 inches (4 cm) to reset the alarm.
Note: due to variability in syringe dimensions, a LINE OCCLUDED alarm may occur before the EMPTY alarm. Option: The infusion may be restarted without resetting EMPTY condition. The pump will continue to display EMPTY and ALERT light will continue flashing. The infusion will continue until STOP is pressed or a
LINE OCCLUDED alarm occurs.
Pressing any key silences the audio portion of the alarm. Complete the editing operation.
The Rate or Dose field is being edited while an infusion is in progress, and there has been no keypress for 15 seconds.
There is an occlusion, line constriction, or other condition that causes excess plunger force.
The pump has been in Standby state for at least 2 minutes. Also occurs if there is no keypress within
15 seconds after the infusion is stopped.
The programmed volume limit has been reached.
The alarm automatically stops the pump. Pressing any key silences the audio portion of the alarm. Check for empty syringe, kinked tubing, clogged catheter, etc. Check for foreign material preventing movement of the pump mechanism.
RELIEVE RESIDUAL SYRINGE
PRESSURE BY RELEASING THE
PLUNGER DRIVER. If the plunger driver is not released, an unintentional bolus may occur when the blockage is cleared. The message clears when the infusion is successfully restarted.
Pressing any key silences the audio portion of the alarm. The Lock function suppresses the audio portion of this alarm.
Note: If the keypad is locked, there will be no beeps, but the PUMP IS
IDLE message still appears.
Pressing any key silences the audio portion of the alarm. Restarting the pump (by pressing START or initiating a bolus) resets the Volume
Limit field.
Page 82 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Alerts and Alarms
Failsafe Alarms
Failsafe alarms occur only when proper operation cannot be assured. The infusion cannot be restarted, and must be reprogrammed after the problem has been corrected. The pump must be turned off to clear the alarm.
Message
( ERROR CODE )
(continuous two-tone audio,
SYSTEM light on)
CHARGE DEPLETED
(continuous two-tone audio,
SYSTEM and BATTERY lights on)
FUSE FAILURE (continuous two-tone audio, SYSTEM and
BATTERY lights on)
POWER FAULT (continuous, two-tone long beeps, SYSTEM and
BATTERY lights on)
Cause
Fuse self-test has detected a fuse failure.
Correction
An internal error has been detected. Repair is required. Make a note of all displayed error codes and the conditions prior to the error. Take the pump out of service.
Note: In some cases, there may not be a legible error code display.
The battery voltage is below minimum operating level.
Connect the charger and recharge the battery for a minimum of 16 hours.
Repair is required. Take the pump out of service.
The pump’s internal power supply is out of tolerance, or there is a fault in the charger.
Repair is required. Take the pump and charger out of service.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 83
Alerts and Alarms
This page is intentionally blank.
Page 84 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Warranty and Service
Information
Chapter 3
Warranty
Baxter Healthcare Corporation (“Baxter”) warrants that the equipment shall be free from defects in material and workmanship when delivered to the original purchaser. Baxter's sole obligation shall be limited to repair or replacement at Baxter’s option and expense, of the defective part or unit for a period of one year following the date of initial delivery. The applicable period for batteries is limited to a period of 12 months following the date of initial delivery.
The warranty extends only to the original purchaser and is not assignable or transferable, and shall not apply to auxiliary equipment or disposable accessories. Baxter’s commitment to repair or replace the product will be null and void if this product is used contrary to the directions for use contained in the labeling. Baxter will assume no responsibility for incidents which may occur if the product is not used in accordance with product labeling.
THERE ARE NO OTHER WARRANTIES INCLUDING ANY IMPLIED WARRANTY AND
ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE WHICH EXTEND BEYOND THE DESCRIPTION OF THE PRODUCT AND
THOSE EXPRESSLY SET FORTH IN ITS LABELING. In no event shall Baxter be responsible for incidental, consequential, or exemplary damages. Modification, alteration, recalibration or abuse, and service by other than a Baxter authorized representative may void the warranty.
Service Information
While under Baxter Warranty, Service Agreement (optional), or lease agreement, the instrument must not be opened by unauthorized personnel.
For service and repair information for this product, call the Baxter Healthcare Corporation
Andover Service Center at 1-800-343-0366 (Customer Service) or 1-800-634-4602
(Technical Service).
Shipping costs for all units returned to Baxter shall be paid for by the customer. The unit must be packed in its original container or in a container that will provide adequate protection during shipment. To ensure prompt return, the Baxter Healthcare Corporation
Andover Service Center must be notified before shipping any unit for repair. When calling, please be prepared to provide the product code number and serial number of the unit. A service authorization number will be issued and should accompany all communications. A brief written description of the problem should be attached to the instrument when it is returned for service.
Baxter will not be responsible for unauthorized returns or for units damaged in shipment due to improper packing.
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 85
Warranty and Service Information
This page is intentionally blank.
Page 86 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Index
Symbols
(hh:mm) in text , 1
▲▼
IN USE error , 80
<xxMIN (protocol) alert , 80
<xxMIN EMPTY alert , 80
<xxMIN VOL LIM alert , 80
A action keys , 18 alarms , 66 , 81 fail-safe , 83 alerts , 67 , 78 group , 65 light , 19
ALFENTAN protocol , 68
ANALGESC library , 68
ANES-HYP library , 68 arrow keys , 21 description , 18
ATRACURI protocol , 68 audio range , 65
Auto Lock function , 59
AUTO SCHEDULE mode , 45
B backlight , 60 , 65 barrel clamp illustration , 9 release lever illustration , 9 battery charging , 69 cover illustration , 9 light , 19 operating time , 15 replacement , 74 voltage check , 76
B-D Plastic deliverable volumes , 26 purge delivery , 54 rate ranges , 25 beeps , 66
07-19-B3-430 bolus complete alert , 66
CUSTOM DILUTION mode , 40 default size , 22 delivery , 57 description , 55 editing , 58 prevention , 21 programming , 56 rate , 56 repeating delivery , 58 review , 57 size limits , 55 units , 55
BOLUS COMPLETE alert , 78 details , 66 bolus modes , 34 , 35
C cancelling an edit , 50
CARDIO library , 68 catalog code , 15 changing the battery , 74 changing the syringe , 58
CHARGE DEPLETED fail-safe alarm , 83
CHARGE IS LOW alert , 78 charger port illustration , 9 charging the battery , 69
CHECK BARREL alarm , 81 alert , 78
CHECK FLANGE alarm , 81 alert , 78
CHECK PLUNGER alarm , 81 alert , 78
CHECK SYRINGE alarm , 81 alert , 78 cleaning and disinfection , 71 conditional message displays , 77
Auto Syringe AS50 Infusion Pump Operator’s Manual Index-1
configuration custom , 67 factory standard , 67 configuration review feature , 61 table , 62 terminating , 67 confirm key , 18 continuous infusion modes programming , 12 counting time , 43 , 46 cradle illustration , 9
CUSTOM DILUTION mode , 39
D dashes in display , 21 data entry , 48 testing , 20 , 51
DATA MISSING error , 78 date and time display , 61
DB9 port , 65 defaults , 67 bolus size , 22 concentration , 21 dose or rate , 21 group , 64 infusion mode , 21 deliverable volume , 15 description, general , 3 digit keys , 18 disinfection and cleaning , 71
DOBUTAMI protocol , 68
DOPAMINE protocol , 68
DOSE COMPLETE alarm , 81 details , 66
DOSE DUE alarm , 81 dose initiation , 46 drug concentration selection , 17
E
EDIT key , 20 editing cancellation , 50 in run state , 50 in standby state , 48
EDITING alarm , 82
EMPTY (protocol) alarm , 82
EMPTY alarm , 82
EMPTY LIBRARY error , 78 enable DB9 port , 65
Index-2 Auto Syringe AS50 Infusion Pump Operator’s Manual entering data , 20 epidural administration precautions , 7 error code fail-safe alarm , 83 error responses table , 78 expanded data display , 60
F factory standard configuration , 67 fail-safe alarms table , 83 field selection , 48
FINISH EDITING error , 78 flange slot , 9 flashing annunciators , 20 data , 20 flow rate accuracy , 15 range , 15 function keys , 18
FUSE FAILURE fail-safe alarm , 83
G general description , 3 information , 15
H hospital identifier , 66
I identifier , 66
IDLE ALARM alert , 66 infusion modes continuous , 23 overview , 23 selecting , 33 timed , 24 infusion rate ranges , 25 selection , 16 input/output port (RS232) , 9 intended use , 1
ISOPROTE protocol , 68
IV pole clamp , 70
IV pole loop details , 71 illustration , 9
07-19-B3-430
K key sounds , 21 keypad description , 18 illustration , 9
KEYPAD LOCKED error , 79
L
LCD panel description , 16 illustration , 9
LINE OCCLUDED alarm , 6 , 82
LOAD SYRINGE error , 79 lock light , 19 activating , 59 auto lock , 65
Lock function , 59
M maintenance cleaning and disinfecting , 71 daily check display and syringe sensor , 75 keypad , 75 mechanical , 74 service cleaning , 73 manual purge , 28
MANUAL SCHEDULE mode , 42 mfr codes , 16
MIVACURI protocol , 68 modes
AUTO SCHEDULE , 45 by unit per time frame , 33 factory standard configuration , 67
MANUAL SCHEDULE , 42
SINGLE DOSE , 41
Monoject deliverable volumes , 26 purge delivery , 54 rate ranges , 25
MUSRELAX library , 68
N
NEAR END alert , 66 next dose selection , 17
NITROGLY protocol , 68
NITROPRU protocol , 68
NO LIBRARY error , 79
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual
NO LIBRARY IS ENABLED error , 79
NOREPINE protocol , 68
NOT ALLOWED error , 79 number out of range , 51
O occlusion sensitivity , 65 on charge light , 19
ON/OFF switch , 30 illustration , 9 operating instructions battery time , 15 charging the battery , 69 checking lamp test , 31
IV pole clamp , 70
IV pole loop , 71 manual purge , 28 mounting the syringe , 28 purging , 53 simplified , 10 states , 20 see also maintenance, programming
OUT OF RANGE error , 79 over body wt selection , 17
P plunger assembly , 9 pole clamp attach recess , 9 mounting pins , 9
POWER FAULT fail-safe alarm , 83 power requirement , 15 precautions , 5 epidural administration , 7 programming
AUTO SCHEDULE mode , 45 bolus , 56 by unit per time frame , 33 continuous infusion modes , 12
CUSTOM DILUTION mode , 39 during run state , 50 during standby state , 48 field selection , 48 getting started , 11
MANUAL SCHEDULE mode , 42 modes , 35 overview , 32 , 48 selecting a protocol , 32 selecting a protocol library , 32 selecting an infusion mode , 33
Index-3
SINGLE DOSE mode , 41 timed infusion modes , 13
PROPOFOL protocol , 68 protocol library bypassing , 22 groups , 63 option , 64 overview , 21 selecting , 32 using , 32
PSI range , 65
PUMP IS IDLE alarm , 82 alert , 79
PUMP LIMIT error , 79 purging delivery table , 54 description , 53 instructions , 53 manual , 28
R rate range option , 65
REMIFENT protocol , 68 remote control , 64 reviewing configuration options , 61
ROCURONI protocol , 68
RS232 port , 9 run lights , 19 state , 20 editing in , 50
S serial communication , 22 service information , 85 number , 76 , 85
SINGLE DOSE mode , 41
SIZE MISMATCH error , 79 size override option , 64
SIZE UNKNOWN error , 79 specifications , 15 standby light , 19 state , 20 editing in , 48 status display description , 16 status panel description , 19 illustration , 9
Index-4 Auto Syringe AS50 Infusion Pump Operator’s Manual
SUCCINYL protocol , 68
SUFENTAN protocol , 68 syringe changing , 58 data tables , 25 factory standard configuration , 67 manufacturers , 15 group , 63 selection , 16 mounting , 28 sensor check , 75 size selection , 16 system light , 19
T tables alarms , 81 alerts , 78 bolus units , 55 configuration review , 62 error responses , 78 fail-safe alarms , 83 infusion rate , 25 programming continuous infusion modes , 12 timed infusion modes , 13 purge delivery , 54 technical specifications , 15 temperature range , 15
Terumo deliverable volumes , 26 purge delivery , 54 rate ranges , 25 testing data , 20 , 51 text conventions , 1 time counting , 43 , 46 data fields , 20 timed infusion modes programming , 13 total delivered display , 59 total drug vol display , 17 troubleshooting , 76
U
UNITS/HR, UNITS/MIN, MUNITS/HR, and MUNITS/
MIN Mode , 35
USE DIGITS error , 79 using protocols , 32
07-19-B3-430
V
VECURONI protocol , 68
VERIFY AGAIN error , 80 version display , 61
VIEW ALERTS?
details , 65 group selection , 62
VIEW DFAULTS?
details , 64 group selection , 62
VIEW LIBRARY?
details , 63 group selection , 62
VIEW MFRS?
details , 63 group selection , 62
VIEW MISC?
details , 64 group selection , 62
VIEW MODES?
details , 63 group selection , 62 voltage check , 76 volume accuracy , 15 limit , 33 volume limit option , 66 selection , 17
VOLUME LIMIT alarm , 82 volumes table , 26
W warranty , 85
07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Index-5
This page is intentionally blank.
Index-6 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
*0719B3430_*
Baxter Healthcare Corporation
Medication Delivery Division
Deerfield, IL 60015 USA
07-19-B3-430
© Copyright 1999, 2002, Baxter Healthcare Corporation. All Rights Reserved.
Revision B
May 2002