Kobold TMU-W Coriolis Mass Flow Meter Installation and Operation Manual

Add to My manuals
118 Pages

advertisement

Kobold TMU-W Coriolis Mass Flow Meter Installation and Operation Manual | Manualzz
TMU-W
High-Pressure Coriolis Mass Flow Meter
Installation and operation manual
Please read the instructions carefully and store them in a safe place
Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
November 2020
Heinrichs Messtechnik GmbH
Contents
INTRODUCTION ................................................................................................................................ 5
I.
II.
III.
IV.
V.
VI.
Shipping and storage; product inspection ................................................................................................ 5
Warranty ................................................................................................................................................. 5
Maintenance, Repair and Hazardous substances ..................................................................................... 5
Disposal ................................................................................................................................................... 5
Supplementary operating instructions ..................................................................................................... 5
Operating manual of explosion-proof flowmeters ................................................................................... 5
1.
IDENTIFICATION ...................................................................................................................... 6
2.
STEPS PRIOR TO OPERATION................................................................................................... 7
2.1
2.2
2.3
2.4
2.5
2.6
2.7
3.
3.1
3.2
3.3
3.4
3.5
Safety advisory for the user ..................................................................................................................... 8
Hazard warnings ...................................................................................................................................... 8
Proper use of the device .......................................................................................................................... 9
Installation and servicing ......................................................................................................................... 9
Returning your flowmeter for servicing or calibration ............................................................................10
Replacement of the transmitter electronics ............................................................................................10
Maintenance...........................................................................................................................................11
Transmitter .............................................................................................................................................. 11
Coriolis mass flow sensor ........................................................................................................................ 11
THE TMU-W SENSOR ............................................................................................................ 12
Application domain of the TMU-W sensor ..............................................................................................12
Mode of operation..................................................................................................................................12
Measuring principle ................................................................................................................................ 12
System configuration .............................................................................................................................. 12
Acquisition ............................................................................................................................................... 12
Performance characteristics of the TMU-W sensor .................................................................................13
Reference conditions .............................................................................................................................. 13
TMU-W flow ranges ................................................................................................................................ 13
Density measurement ............................................................................................................................. 13
Accuracy .................................................................................................................................................. 14
Pressure loss in Hydrogen dispensers ..................................................................................................... 14
Environmental Conditions ....................................................................................................................... 14
Operating conditions ..............................................................................................................................15
Installation............................................................................................................................................... 15
Installation Orientation ........................................................................................................................... 16
Pressure surges ....................................................................................................................................... 18
Using the device with hazardous fluids ................................................................................................... 18
Vibration stability .................................................................................................................................... 18
Process conditions................................................................................................................................... 19
Gas content ............................................................................................................................................. 19
Process pressure range ........................................................................................................................... 19
Outlet pressure ....................................................................................................................................... 19
Connection to the transmitter ................................................................................................................ 19
Construction details ................................................................................................................................20
Installation lengths and weight ............................................................................................................... 20
Dimension drawing for the type TMU-W004 .......................................................................................... 20
Dimension drawing for the type TMU-W006 .......................................................................................... 21
Dimension drawing for the type TMU-W012 .......................................................................................... 21
Materials ................................................................................................................................................. 21
Page 2 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
4.
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
4.9
4.10
5.
5.1
6.
6.1
6.2
6.3
6.4
UMC4 TRANSMITTER ............................................................................................................ 22
Application domain of the UMC4 transmitter .........................................................................................22
Mode of Operation .................................................................................................................................22
System configuration ..............................................................................................................................22
UMC4 performance characteristics .........................................................................................................22
Reference conditions .............................................................................................................................. 22
Measured error ....................................................................................................................................... 22
Repeatability error .................................................................................................................................. 22
Influence of ambient temperature.......................................................................................................... 22
Operating conditions ..............................................................................................................................23
Installation conditions and cable glands ................................................................................................. 23
NPT cable glands ..................................................................................................................................... 23
Environmental conditions ....................................................................................................................... 24
Process conditions................................................................................................................................... 25
DSB data memory module ......................................................................................................................25
Input measured variables .......................................................................................................................26
Measured Values ..................................................................................................................................... 26
Measuring range ..................................................................................................................................... 26
Outputs ..................................................................................................................................................27
Output circuits......................................................................................................................................... 27
Current outputs ....................................................................................................................................... 28
Binary Outputs ........................................................................................................................................ 29
Construction details UMC4 .....................................................................................................................30
Mounting / Dimensions ........................................................................................................................... 30
Weight ..................................................................................................................................................... 32
Material ................................................................................................................................................... 32
Construction details UMC4-RM...............................................................................................................33
Dimensions .............................................................................................................................................. 33
Mounting bore pattern ........................................................................................................................... 34
Weight ..................................................................................................................................................... 34
Materials ................................................................................................................................................. 34
ELECTRICAL CONNECTIONS UMC4 ........................................................................................ 35
Mains connections and their protection classes......................................................................................35
Wiring diagrams ...................................................................................................................................... 35
The output terminals ............................................................................................................................... 36
Sensor connection ................................................................................................................................... 36
CERTIFICATES AND APPROVALS ............................................................................................ 41
Conformity to CE Directives ....................................................................................................................41
Explosion protection ...............................................................................................................................41
Custody transfer applications .................................................................................................................41
Patents ...................................................................................................................................................41
OPERATING MANUAL TMU-W / UMC4
Page 3 of 118
Heinrichs Messtechnik GmbH
7.
7.1
7.2
7.3
8.
8.1
8.2
8.3
9.
OPERATION ........................................................................................................................... 42
Control unit BE4 ......................................................................................................................................42
Introduction ............................................................................................................................................ 42
Operating modes ..................................................................................................................................... 43
Operator interface .................................................................................................................................. 43
The keys and their functions ................................................................................................................... 44
Functional classes, functions and parameters ........................................................................................ 45
Display ..................................................................................................................................................... 46
UMC4 transmitter functional classes ......................................................................................................47
MEASURED VALUES functional class ....................................................................................................... 48
PASSWORD functional class .................................................................................................................... 53
COUNTER functional class ....................................................................................................................... 55
MEASUREMENT PROCESSING functional class........................................................................................ 57
FLOW functional class ............................................................................................................................. 61
DENSITY functional class ......................................................................................................................... 66
TEMPERATURE functional class ............................................................................................................... 73
PULSE OUTPUT functional class .............................................................................................................. 76
STATUS functional class........................................................................................................................... 79
CURRENT OUTPUTS functional class ....................................................................................................... 82
SIMULATION functional class .................................................................................................................. 87
SELF-TEST functional class ....................................................................................................................... 91
UMC4 TRANSMITTER SETTINGS functional class .................................................................................... 94
SENSOR SETTINGS functional class.......................................................................................................... 97
CUSTODY TRANSFER functional class .................................................................................................... 100
Density calibration ................................................................................................................................105
Conditions ............................................................................................................................................. 105
Procedure .............................................................................................................................................. 105
TROUBLE SHOOTING ........................................................................................................... 107
Self-help Checklist.................................................................................................................................107
UMC4 transmitter error messages ........................................................................................................108
Standard operating mode ..................................................................................................................... 108
Custody transfer mode .......................................................................................................................... 108
List of error messages ........................................................................................................................... 109
Returning the Meter .............................................................................................................................113
Declaration of Decontamination ........................................................................................................... 114
DECLARATION OF CONFORMITY ......................................................................................... 115
Page 4 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Introduction
I.
Shipping and storage; product inspection
Shipping and storage
The device is to be safeguarded against dampness, contamination (especially the inside of the flow meter),
impact and damage. Open the packaging with caution to prevent unintentional damage.
Adhere to the temperature limits during storage.
Product inspection
Upon receipt of the product, check the contents of the box and the product particulars against the
information on the delivery slip and order form so as to ensure that all ordered components have been
supplied. Notify us of any shipping damage immediately upon receipt of the product. Any damage claim
received at a later time will not be honoured.
II.
Warranty
Your flowmeter was manufactured in accordance with the highest quality standards and was thoroughly
tested prior to shipment. However, in the event any problem arises with your device, we will be happy to
resolve the problem for you as quickly as possible under the terms of the warranty, which can be found in
the terms and conditions of delivery. Your warranty will only be honoured if the device was installed and
operated in accordance with the instructions for your device. Any mounting, commissioning and/or
maintenance work is to be carried out by qualified and authorized technicians only.
III.
Maintenance, Repair and Hazardous substances
When used in the intended manner no special maintenance is required. However, the flowmeter should be
checked within the context of routine maintenance of the facility and the pipelines. Should a repair,
calibration or maintenance become necessary, be sure to clean the device thoroughly and follow the steps in
section 0, “Returning the Meter” before returning the device to Heinrichs Messtechnik.
The operator is liable for any substance removal or personal damage costs arising from inadequate cleaning
of a device sent for repair.
IV.
Disposal
Observe the regulations applicable to disposal in the country of installation!
V.
Supplementary operating instructions
Supplement operating manuals are available for special features, interfaces and operations relating to your
device, request your copy from our service department.
VI.
Operating manual of explosion-proof flowmeters
For installation of the flowmeter within hazardous areas read the operation manual of explosion-proof
flowmeters. It contains all the EX-relevant information for your flowmeter.
Warning!
Only devices designated as EX-certified on their rating plates may be used in areas of
potentially explosive atmospheres!
The use of standard equipment in EX-hazardous areas is strictly prohibited.
OPERATING MANUAL TMU-W / UMC4
Page 5 of 118
Heinrichs Messtechnik GmbH
1. Identification
Manufacturer:
Heinrichs Messtechnik GmbH
Robert-Perthel-Strasse 9
D-50739 Cologne
Germany
Phone:
+49 221 49708-0
Fax:
+49 221 49708-178
Internet:
www.heinrichs.eu
Email:
[email protected]
Product type:
Mass flow-rate meter for liquid and gaseous products under high pressure
Product name:
Sensor type:
TMU-W
Transmitter type:
UMC4 or UMC4-RM
File name:
tmu-w_umc4_ba_20.02_en.docx
Version:.
20.02, dated November 5, 2020
Page 6 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
2. Steps prior to operation
It is essential that you read these operating instructions before installing and
operating the device. The device is to be installed and serviced by a qualified
technician only. The UMC4 transmitter is to be used exclusively to measure mass
and volume flow, as well as liquid and gas density and temperature, in conjunction
with a Heinrichs Messtechnik TM, TME, TMR, TMU, TM-SH or HPC sensor.
Downloading of the present document from our web site www.heinrichs.eu and printing out this
document is allowed only for the purposes of using our mass flowmeters. All rights reserved. No
instructions, wiring diagrams, and/or supplied software, or any portion thereof, may be produced, stored,
in a retrieval system or transmitted by any means, electronic, mechanical, photocopying or otherwise,
without the prior written permission of Heinrichs Messtechnik GmbH.
Although the materials in the present document were prepared with extreme care, errors cannot be
ruled out. Hence, neither the company, the programmer nor the author can be held legally or otherwise
responsible for any erroneous information and/or any loss or damage arising from the use of the
information enclosed.
Heinrichs Messtechnik GmbH extends no express or implied warranty in regard to the applicability of the
present document for any purpose other than that described.
We plan to optimize and improve the products described and in so doing will incorporate not only our
own ideas but also, and in particular, any suggestions for improvement made by our customers. If you
feel that there is any way in which our products could be improved, please send your suggestions to the
following address:
Company:
Heinrichs Messtechnik GmbH
HM-EE (R&D Department)
Robert-Perthel-Strasse 9
D-50739 Cologne
Germany
or:
via fax :
+49 (221) 49708-178
via email:
[email protected]
Note:
We reserve the right to change the technical data in this manual in the light of any
technical progress that might be made.
For updates regarding this product, visit our website at www.heinrichs.eu, where you
will also find contact information for the Heinrichs Messtechnik distributor nearest you.
For information regarding our own sales operations, contact us at [email protected].
OPERATING MANUAL TMU-W / UMC4
Page 7 of 118
Heinrichs Messtechnik GmbH
2.1
Safety advisory for the user
The present document contains the information that you need in order to operate the product described
herein properly. This document is intended for use by qualified personnel. This means personnel who are
qualified to operate the device described herein safely, including

electronics engineers,

electrical engineers, or

service technicians
who are conversant with the safety regulations pertaining to the use of electrical and automated
technical devices and with the applicable laws and regulations in their own country.
Such personnel must be authorized by the facility operator to install, commission and service the product
described herein, and must have read and understood the contents of this operating instructions before
working with the device.
2.2
Hazard warnings
The purpose of the hazard warnings listed below is to ensure that device operators and maintenance
personnel are not injured and that the flow meter and any devices connected to it are not damaged.
The safety advisories and hazard warnings in the present document that aim to avoid placing operators
and maintenance personnel at risk and to avoid material damage are prioritized using the terms listed
below, which are defined as follows in regard to these instructions herein and the advisories pertaining
to the device itself.
Warning
means that failure to take the prescribed precautions could result in injury, substantial
material damage or even death. Always comply to these warnings and proceed with
caution.
Caution
means that failure to take the prescribed precaution could result in material damage or
destruction of the device. We advice always to abide to these instructions!
Note
means that the accompanying text contains important information about the product,
handling the product or about a section of the documentation that is of particular
importance.
Page 8 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
2.3
Proper use of the device
The Coriolis Mass Flow Sensor is intended for the sole use of direct and continuous mass flow
measurement of liquids and gases.
To ensure safety for people and the environment adhere to the installation and operational instructions
and warning in this manual
Warning
The operator is responsible for ensuring that the material used in the sensor and
housing are suitable and that such material meets the requirements for the process
medium and the ambient site conditions.
The manufacturer accepts no responsibility for the selection of unsuitably materials.
Warning
Only sensors marked as EX-certified on their rating plates may be used in EX hazardous
locations. Standard equipment is not permitted for installation and use in EX hazardous
locations.
For installation within hazardous areas read the Ex-supplementary manual. It contains
all EX-relevant parameters for the sensor and the UMC4 transmitter.
Caution
To ensure the device performs correctly and safely, it must be shipped, stored, set up,
mounted, operated and maintained correctly.
2.4
Installation and servicing
The devices described in this manual are to be installed and serviced only by qualified technical
personnel such as a qualified Heinrichs Messtechnik electronics engineer or service technician.
Warning
Before servicing, the device must be completely de-energised and disconnected from all
peripheral devices. The technician must also ensure that the device is completely
disconnected from any live circuits.
Only original replacement parts are to be used.
Heinrichs Messtechnik GmbH accepts no liability for any loss or damage of any kind arising from
improper operation of any product, improper handling or use of any replacement part, or from
external electrical or mechanical effects, overvoltage or lightning. Any such improper operation, use
or handling shall automatically invalidate the warranty for the product concerned.
In the event a problem arises with your device, or if you need assistance in diagnosing a problem with
your device, please contact us at one of the following numbers to arrange to have your device repaired:
+49 (0)221 49708-0
OPERATING MANUAL TMU-W / UMC4
+49 (0)221 49708-178
Page 9 of 118
Heinrichs Messtechnik GmbH
2.5
Returning your flowmeter for servicing or calibration
Before sending your flowmeter back to us, for servicing or calibration, make sure it is completely clean.
Any residues of substances that could be hazardous to the environment or human health are to be
removed from all crevices, recesses, gaskets, and cavities of the housing before the device is shipped.
Warning
The operator is liable for any loss or damage of any kind, including personal injury,
decontamination measures, removal operations and the like that are attributable to
inadequate cleaning of the device.
Any device sent in for servicing is to be accompanied by a declaration of
Decontamination, a template of which is provided in section 8.3.1.
When returned, the device is to be accompanied by a document describing the problems encountered.
Please include in this document the name of a contact person whom our technical service department
can contact to enable us to repair your device as expeditiously as possible and minimize the repair costs.
2.6
Replacement of the transmitter electronics
Before replacing the transmitter electronics, read the safety instructions in Section 2.4, “Installation and
servicing” on page 9.
The data memory chip (DAB) with the calibrating data of the sensor is an integral component of the
control unit (display BE4). Removal and installation of the control unit is described in chapter 4.6, “DSB
data memory module” on page 25.
Should an exchange of the transmitter electronics become necessary, it is essential that the whole
electronic stack is replaced. This comprises of all circuit boards in the electronic compartment and in the
terminal compartment. The overall accuracy of the measurement up to the analogous outputs can only
be guaranteed when all circuit boards are replaced. Only the control unit with the integrated memory for
the calibrating data of the sensor shall remain with the device.
Caution
The complete stack is to be replaced with all of its printed circuit boards (with the
exception for the display unit containing the memory module). This is particularly
important for the explosion-proof transmitter. The specified precision of the
electronics is only guaranteed if the complete stack is replaced.
Page 10 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
2.7
Maintenance
Transmitter
The transmitter is maintenance-free.
We recommend cleaning the viewing-glass of the transmitter at regular intervals; check the enclosure for
corrosion damages and the solid seat of the cable glands.
Warning!
In the event an enclosure lid O-ring gasket is damaged, humidity may enter the
enclosure and cause damage to the internal electronic circuits.
Indications of such are:

Visible discolouration’s or condensation on the viewing-glass of the transmitter,

corrosion damages to the enclosure,
Coriolis mass flow sensor
The sensor is largely maintenance-free. When handled correctly its functionality will only be impaired by
corrosion or deposits inside of the measuring tubes. Therefore, both should be implicitly avoided. Remove
deposits in the tubes and in or around the splitter on a regularly basis by means of a suitably washing
method. Failure to do so may result in a loss of measurement precision.
Warning!
In the event of a tube rupture, e.g. due to corrosion or damage, medium will leak into
and fill the enclosures body, which can lead to subsequent damage to the external
housing, particularly at high process pressures!
OPERATING MANUAL TMU-W / UMC4
Page 11 of 118
Heinrichs Messtechnik GmbH
3. The TMU-W sensor
3.1
Application domain of the TMU-W sensor
The TMU-W sensor is intended solely for the direct and continuous mass flow measurement of liquids and
gases under high pressure, irrespective of their conductivity, density, temperature, pressure, or viscosity.
The sensor can be utilised for the direct and continuous mass flow measurement of chemical fluids,
suspensions, molasses, paint, varnish, lacquer, pastes and similar materials.
3.2
Mode of operation
Measuring principle
The mass flowmeter is based on the Coriolis
principle whereby in a rotating system a
force (known as the Coriolis force) is exerted
on a mass at a rotation point that is moving
towards or away from this point.
FC  2  m    
System configuration
The flowmeter consists of a sensor mounted in to a piping system and a remote installed transmitter (see
Section 4.1, “Application domain of the UMC4” on page 22), connected to the sensor via a special cable (e.g.
on a wall or to a pipe).
The transmitter oscillates the flow tubes of the sensor by means of an excitation coil and captures, via pickup coils, the measuring signal which is proportional to the mass flow. After temperature compensation, the
measuring signal is converted into an analogue output signal that is consistent with the measuring range
setting.
Acquisition
Measured variables:
Mass flow,
Temperature.
Page 12 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
3.3
Performance characteristics of the TMU-W sensor
Reference conditions

Established flow profile

Inlet section has to correspond to mounting length

Control valves always positioned downstream

Measurement is to be performed with a liquid containing no gas bubbles

Flow tubes are to be kept clean at all times

Process temperature is to be regulated as specified in Section 3.4.6, “Process conditions” on
page 19

Process pressure is to be regulated as specified in Section 3.4.8, “Process pressure range” on
page 19

Ambient temperature is to range from + 10 °C to + 30 °C (50 °F to 86 °F)

Warm-up period: 15 minutes

Standard calibration is to be performed at 20 %, 50 % and 100 % (two times each)

High-frequency interference is to be regulated according to the EMC standards stated in section
9, “Declaration of Conformity” on page 115
TMU-W flow ranges
Mass-flow
Model
min.
max.
Measuring range
Measuring range
kg/min [lbs/min]
kg/min [lbs/min]
kg/min [lbs/min]
Zero point stability
(of range)
TMU-W004
0.13 [0.3]
4 [8.8]
0.0004 [0.0009]
TMU-W006
0.5 [1.1]
16 [35.3]
0.0016 [0.0035]
TMU-W012
2.5 [5.5]
25 [55.1]
0.0025 [0.0055]
Reference conditions: in conformity with IEC 770:
Table 1: TMU-W flow ranges
Temperature: 20 °C, relative humidity: 65 %, air pressure: 101.3 kPa
Fluid: water
Density measurement
Due to its constructive properties, a density calibration for the TMU-W family of sensors is not foreseen,
thus also disabling the determination of volume flow.
OPERATING MANUAL TMU-W / UMC4
Page 13 of 118
Heinrichs Messtechnik GmbH
Accuracy
Mass flow
Fluids
TMU-W004 and TMU-W006
± 0.1 % of actual flow ± zero point stability
TMU-W012
± 0.15 % of actual flow ± zero point stability
Repeatability error
± 0.05 % of actual flow (sensor with transmitter) ± ½ zero point stability
Mass flow
Gases
TMU-W004, TMU-W006 and
± 0.5 % of actual flow ± zero point stability
TMU-W012
Repeatability error
(1)
(1)
(1)
(1)
± 0.25 % of actual flow (sensor with transmitter) ± ½ zero point stability
(1)
Additional measured values
Volume flow
Not available
Temperature
± 0.5 °C
Hysteresis
n/a
Settling time
1 to 15 seconds
Startup drift
15 minutes
Long-term drift
± 0.02 % of upper-range value per year
Influence of ambient temperature
± 0.005 % per K
Influence of fluid temperature
Compensated
Influence of fluid pressure
For fluids: too small to be relevant
(1) Refer to section 3.3.2, “TMU-W flow ranges” for detailed information on flow ranges
Table 2: Flow-Rate Accuracy
Pressure loss in Hydrogen dispensers
The pressure loss of gases with constant mass flow is highly dependent on the operating pressure in the
mass meter and the resulting flow rate. The TMU-W is designed that its pressure lost will not exceed 30 bar
during a standardized refuelling operation with hydrogen and an empty vehicle tank, as defined in the
SAEJ2601. With an almost full vehicle tank, the pressure loss shall be much less than 20 bar.
Environmental Conditions
Ambient temperature
− 40 °C to + 60 °C (-40 °F to 140 °F)
Storage temperature
− 40 °C to + 80 °C (-40 °F to 176 °F)
Climatic category
In conformity with IEC 654-1. Unsheltered class D locations with direct open-air climate.
Ingress protection
Sensor: IP 67 (NEMA 6), Transmitter: IP68 / 1 m for 24 hours (NEMA 6P) acc. to DIN EN 60529 with
mounted and sufficiently tightened approved cable glands.
Page 14 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
3.4
Operating conditions
Installation
The sensor is to be protected, wherever possible, against valves, manifolds and similar fittings that generate
turbulence. The sensor is to be installed in accordance with the following instructions.
Diagram showing flowmeter installation
Flowmeter installation: A = sensor, B = valve, C = pipe clamps and supports
The sensor is not to be used to support a pipe or
other pipe components.
Do not install the sensor in suspended pipes.
Do not adjust the position of a pipe by pulling or
grasping the sensor.
OPERATING MANUAL TMU-W / UMC4
Page 15 of 118
Heinrichs Messtechnik GmbH
Installation Orientation
Without compromising its accuracy, the TMU-W can be installed and operated in various orientations. The
following representations show the most common installation positions and provide tips on how the
operator can prevent installation-related influences on the measurement.
Standard installation position
Installation position A
Installation position B
Installation position C
Page 16 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Type of medium
Position
Assessment
Pure liquids
Standard installation
Self-draining flow tubes
position
Position A or B
OK
Position C
Liquid residue remains in pipe
Liquids containing
Standard installation
Self-draining flow tubes, gas bubbles do not accumulate
homogeneously
position
in flowmeter
dispersed gas
Position A
Not recommended owing to gas bubble accumulation in
flowmeter
Position B
Gas bubbles may accumulate in the presence of low flow
Position C
No gas bubble accumulation in flowmeter, liquid residues
velocities
may remain in device after discharge
Liquids containing
Standard installation
substances that
position
could form deposits
Self-draining flow tubes, no deposit formation
Position A
OK
Position B
Substances in the liquid could form deposits at low flow
velocities
Position C
Not recommended owing to presence in flowmeter of
substances that could form deposits
Liquids containing
Standard installation
homogeneously dispersed position
gas, which may contain
substances that could
form deposits
Position A
Self-draining flow tubes, no accumulation of gases or
substances that could form deposits
Not recommended owing to gas bubble accumulation in
flowmeter
Position B
Gas bubbles or substances that could form deposits at
low flow velocities
Position C
Not recommended owing to presence in flowmeter of
substances that could form deposits
Gases that do not form
Standard installation
condensate
position,
Any of these installations positions can be used
Position A, B or C
Gas, condensate-
Standard installation
Flow direction should be from top to bottom so that any
forming gas/liquid,
position
condensate that forms can flow out efficiently
moisture
Position A
OK
Position B
Condensate might form in flowmeter
Position C
Not recommended owing to condensate accumulation in
flowmeter
Table 3: Sensor Installation Positions
OPERATING MANUAL TMU-W / UMC4
Page 17 of 118
Heinrichs Messtechnik GmbH
Pressure surges
Pressure surges in a pipe could be provoked by a sudden decrease in flow caused by rapid closing of a valve
or similar factors. This change in pressure can lead to under-pressure downstream from a valve that has
been closed rapidly, and to outgassing. If the valve is mounted directly on the inlet section of the flowmeter,
a gas bubble can form in the flow tube that can cause a measuring signal disturbance that would shift the
zero point of the output signal. In extreme cases, a pressure surge could cause mechanical damage to the
sensors and/or flow tube.
Whenever possible, quick-closing valves should be mounted downstream from the sensor. If this is not
feasible, such valves are to be mounted a minimum of 10 x DIA (Φ
) ) from the nearest sensor. Alternatively,
valve closing speed can be reduced.
Using the device with hazardous fluids
The sealing technology used in the TMU-W mass flowmeters connector plug renders the device unsuitable
for use with hazardous fluids. Only sensors that meet the standards for safety instruments are suitable for
use with hazardous fluids.
In the case of welded components, a coloured liquid penetration test can be performed on the welds, or the
first seam of the weld can be x-rayed.
Vibration stability
The sensors are insensitive to vibration. Vibration stability has been validated in accordance with
DIN IEC 68-2-6, for up to 1 g at 10 to 150 Hz.
If pipe vibration is greater than 1 g in the 5-2000 Hz range, additional supports shall be mounted such as
depicted in the following images. Such supports will prevent vibration from affecting the device’s
mechanical configuration and/or measurement readings.
Image 1: Wall mounted supports
Page 18 of 118
Image 2: Floor-mounted supports
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Process conditions
Process temperature
− 40 °C to + 100 °C (-40 °F to 212 °F); rating plate specifications apply
Physical state
Liquid product
(maximum density 2 kg/l)
Gaseous product
(minimum density 0.002 kg/l in operating state)
Viscosity
0.3 up to 2,000 mPas (0.3 to 2,000 cP)
Gas content
The use of products containing gases is not permitted for custody transfer operations. In other applications,
the presence of gas will increase false readings. For the readings of products containing gas to be valid,
small gas bubbles must be homogeneously distributed in the fluid.
Large gas bubbles will automatically provoke extremely false readings and will shift the zero point. Thus, the
extent to which readings are false is determined by the process conditions. A rule of thumb in this regard is
as follows: A 1 % gas component will increase false readings by 1 %. The gas component is not to exceed 5 %.
Process pressure range
The applicable pressure range is dependent on the sensor and the selected process connections. It is
essential that the correct process connections for the intended process pressure is selected during the
ordering process.
TMU-W004
up to 1000 bar
TMU-W006
up to 500 bar
TMU-W012
up to 1000 bar
Outlet pressure
Outlet pressure must be greater than the vapour pressure ps of the measured product.
Connection to the transmitter
The sensors of the TMU-W family are only available as remote mount versions. The connection to the
sensor is achieved via a special ten-cored cable by means of a detachable screw-on connector.
During installation, regulations and applicable legal standards are to be adhered to.
The maximum cable length is 200 m (approx. 650 ft). See Section 5.1.3 on page 36 for information regarding
the connection and cable specifications.
OPERATING MANUAL TMU-W / UMC4
Page 19 of 118
Heinrichs Messtechnik GmbH
3.5
Construction details
Installation lengths and weight
Dimensions and weight:
Model
Sensor length: flange to flange
Process connection
6MF 9/16-18 UNF
TMU-W004
Sensor length
Sensor weight
mm [inch]
kg [lbs]
347
[13.66]
2.8
[6.2]
400
[15.75]
3.8
[8.4]
415
[16.34]
5.2
[11.5]
9/16 F562C
Hofer 7/8"
½ NPT(F),
6MF 9/16-18 UNF
TMU-W006
9/16 F562C
Hofer 7/8"
½ NPT(F),
6MF 9/16-18 UNF
12MF 3/4-16 UNF
TMU-W012
9/16 F562C
Hofer 7/8"
½ NPT(F),
Table 4: Dimensions and Weights
For further dimensions of the sensors, refer to the dimensions drawings in the following sections
Dimension drawing for the type TMU-W004
Remote mount configuration with IP67 Connector suitable for process temperatures up to 100 °C (212 °F):
For further dimensions and weight, see Section 3.5.1, “Installation lengths and weight” on page 20.
Page 20 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Dimension drawing for the type TMU-W006
Remote mount configuration with IP67 Connector suitable for process temperatures up to 100 °C (212 °F):
For further dimensions and weight, see Section 3.5.1, “Installation lengths and weight” on page 20.
Dimension drawing for the type TMU-W012
Remote mount configuration with IP67 Connector suitable for process temperatures up to 100 °C (212 °F):
For further dimensions and weight, see Section 3.5.1, Installation lengths and weight” on page 20.
Materials
Sensor housing:
1.4301 (304L)
Flow tubes:
1.4571 (316Ti)
Splitter:
1.4571 (316Ti)
Flange Connectors:
OPERATING MANUAL TMU-W / UMC4
1.4571 (316Ti)
Page 21 of 118
Heinrichs Messtechnik GmbH
4. UMC4 transmitter
4.1
Application domain of the UMC4 transmitter
The UMC4 transmitter (hereinafter referred to as UMC4) for use with TM, TME TMR, TMU, TM-SH and HPC
Coriolis mass-flow sensors, is a programmable transmitter designed to captured and processes
measurement data from its associated sensor for displaying on its built in display or for the transmission of
measurement results via various interfaces.
Via its BE4 control unit, the UMC4 can be customised for use to fit a variety of applications. Although basic
configuration settings such as transmitter calibration are factory performed, other parameters such as
those for measurement data processing, analysis, display and output are user definable.
4.2
Mode of Operation
The Coriolis mass flow meter works on the principle that in a rotating system a force (known as Coriolis
force) is exerted on a mass at a rotation point that is moving towards or away from this point. Utilising a
suitable sensor construction, this force can be exploited for the direct measurement of the mass flow. The
UMC4 transmitter measures and evaluates signals received from sensors positioned on the flow tubes (see
Section 3.2.1, “Measuring principle” on page 12).
4.3
System configuration
Transmitter:
The UMC4 transmitter regulates the excitation of the sensor vibration system and processes the sensor
signals. The standard model is equipped with two analogue, passive 4 to 20 mA outputs, an impulse or
frequency output and a status output, and is enabled for digital data transfer via the HART ® protocol.
Sensor:
Coriolis sensors measure mass-flow, density and temperature of fluids or gases. They can be used to
perform measurements with any liquid or gaseous product providing that the sensor material is suitable for
the product to be measured.
4.4
UMC4 performance characteristics
Reference conditions
In conformity with IEC 770
Temperature: 20 °C (68 °F), relative humidity: 65 %, air pressure: 101.3 kPa (14.7 psi)
Measured error
For measurement errors and zero point stability refer to the sensor data sheet in section 3.3.2, “TMU-W
flow ranges” on page 13.
Repeatability error
 0.05 % of the actual value (Sensor with transmitter).
For more details refer to the sensor data sheet or section 3.3.2, “TMU-W flow ranges” on page 13.
Influence of ambient temperature
Max.  0.05 % per 10 °C
Page 22 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
4.5
Operating conditions
Installation conditions and cable glands
The integral mount version of the UMC4 transmitter shall be installed in accordance with the sensor
installation requirements stated in section 3.
When selecting the position of installation, especially for the remote-mounted UMC4, ensure a vibrationfree mounting.
Warning:
Cable glands are not contained in the basic scope of supply. The operator is
responsible for ensuring that cable glands or plugs according to the enclosures type of
EX-protection and the provided threads are used. The enclosures thread type is stated
on the transmitters rating plate.
Caution:
Where applicable, metalized cable gland must be used for the sensor / transmitter
cable to ensure a sufficient conductive connection for the cable shield.
(See section 5.1.3, “Sensor connection” on page 36)
NPT cable glands
The transmitter housing is designed for use with M20x1.5 threaded cable glands. For NPT threaded cable
glands the manufacturer adds certified NPT adapters. These NPT adapters are mechanically fixed to the
thread of the transmitter housing.
Warning:
NPT adapters mounted by the manufacturer may never be removed by the customer! In
the event of a removal of these adapters, the protection class Ex-d can no longer be
guaranteed.
OPERATING MANUAL TMU-W / UMC4
Page 23 of 118
Heinrichs Messtechnik GmbH
Environmental conditions
Ambient temperature
− 20 °C to + 60 °C (-4 °F to 140 °F), below 0 °C (32 °F) readability of the LC display may diminish.
Storage temperature
−25 °C to + 60 °C (-13 °F to 140 °F)
Climatic category
In conformity with IEC 654-1. Unsheltered class D locations with direct open-air climate.
Vibration immunity
The UMC4 is insensitive to moderate vibration; vibration stability has been validated in accordance with
DIN IEC 68-2-6, for up to 1 g at 10 to 150 Hz.
Ingress protection
Standard UMC4 housing, IP 68 / 1 m for 24 hours (NEMA 6P)
Explosion-proof single compartment electronics enclosure (Ex-db)
Terminals: The process terminals are situated at the back-end of the single compartment Ex-d enclosure.
Caution:
Ingress protection IP 68 can only be guaranteed if suitable and sufficiently tightened
cable glands or conduits are used. If the cable glands are not tightened sufficiently,
water may leak into the terminal compartment of the enclosure and cause damage to
the electronics.
Caution:
Particular care must be taken if the enclosures viewing window becomes fogged or
discoloured, which may be an indication of moisture, water or product seeping through
the wire sheath into the transmitter’s housing.
Caution:
Electromagnetic compatibility can only be guaranteed if the lids of the enclosure are
securely tightened. Leaving the enclosure open may lead to electromagnetic
disturbances.
Warning
In Ex hazardous areas, only sensors and transmitters marked as such on their rating
plates may be used!
Page 24 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Process conditions
For detailed information on the process conditions and specifications, please refer to section 3.3,
“Performance characteristics of the TMU-W sensor”.
When compact mounted, the process conditions can influence the operation of the transmitter if the
manufactures installation and operational conditions are not abided to.
Remote mounted transmitters are not affected by the prevailing process conditions of the sensor
Note:
Compact-mounted transmitters are not available for all versions of sensors.
4.6
DSB data memory module
The memory device is integrated on the PCB of the control unit where all sensor data such as sensor
constants, model numbers, serial numbers, and so on are stored. Consequently, the memory module is
linked to the sensor. If the transmitter’s electronics are exchanged for any reason, the control unit BE4 must
be removed from the old and installed onto the new electronic stack.
Warning:
When replacing the transmitter electronics, ensure that the applicable standards
and regulations pertaining to electrical devices, device installation and process
technology are abided to. The highly integrated electronic components of the UMC4
may be damaged when exposed to ESD hazards. Only when installed in the
transmitters enclosure are the electronics compliant to EMC standards.
Exchanging the Control Unit
To exchange the BE4 control unit abide to the following instructions referring to the image on the following
page.
After loosening the 4 screws situated below the foil, the control unit with the display can be detached
together with the foil. For this purpose, the foil itself does not need to be removed.
When transferring the BE4 PCB to a new electronic stack, care should be taken to ensure the correct seating
of the 10 poled connector. An incorrectly plugged connector will lead to malfunction and may also damage
the electronics.
OPERATING MANUAL TMU-W / UMC4
Page 25 of 118
Heinrichs Messtechnik GmbH
Removal and exchange of the BE4 Control unit:
The DSB Memory device is situated
on the rear side of the control unit
When the flow meter is powered up, the device continues using the values stored in the memory device.
Thus, the DSB memory device provides maximum safety and comfort when device components are
replaced.
Due to the device specific parameters saved in the memory device, the control units are not arbitrarily
interchangeable between identically constructed transmitters. Replace circuit boards must be ordered
from Heinrichs using the transmitter’s serial number specification. The calibration data of the sensor are
programmed into the delivered BE4 control panel directly by the manufacture
4.7
Input measured variables
Measured Values
The UMC4 measures via the connected sensor the following variables:

mass flow rate,

temperature,

density
A fourth variable, the volume flow, is derived from the measured variables; mass flow rate and density.
Note:
Density and volume variables are not available for all sensor types.
Measuring range
The measuring range, which varies according to the used sensor (TM, TME, TMR, TMU or HPC), can be found
in the relevant data sheet and in sections 3.3.2, “TMU-W flow ranges” and 3.3.3, “Density measurement” on
page 13.
The specified measuring range of the delivered sensor is also printed on the devices rating plate.
Page 26 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
4.8
Outputs
Output circuits
All Signal output circuits of the UMC4 are configured as passive outputs and therefore require an auxiliary
power input for operation. They are isolated from the rest of the UMC4 circuitry via optocouplers.
Each output circuit is also galvanically isolated from each other as well as to ground.
Analogue outputs:
2x 4 to 20 mA passive current outputs
Current output 1:
Mass flow, volume flow, density, temperature
(Output 1 superimposed with HART® protocol )
Current output 2:
Mass flow, volume flow, density, temperature
Refer to section 4.8.2, “Current outputs” on page 28 for connection advice and section 7.2.10, “CURRENT
OUTPUTS functional class” on page 82 for programming of the current outputs
Binary outputs:
2x Binary outputs (NPN)
Pulse output:
Pulse duration: default value 50 ms
adjustable range is 0.1 to 2000 ms (0.0 = Automatically set)
Pulse-pause ratio is 1:1 if the set pulse duration is not exceeded.
As a frequency output max. 1 kHz
Pulse value: settable in decade increments of selected pulse unit
Status output:
For: forward and reverse flow, MIN flow rate, MAX flow rate, MIN
density, MAX density, MIN temperature, MAX temperature, alarm
Second pulse output (phase-shifted to Pulse1 by 90°)
Max. Electrical values:
Unom
= 30 V
Imax
= 200 mA
Refer to section 7.2.8, “PULSE OUTPUT functional class” on page 76 and 7.2.9, “STATUS functional class” on
page 79 for programming of the binary outputs.
Warning:
For connection to, as well as the maximum electrical output
parameters of the transmitter for use in potentially Explosive
Atmospheres refer to the relevant type examination certificate or the
applicable Ex-supplementary operating manual.
OPERATING MANUAL TMU-W / UMC4
Page 27 of 118
Heinrichs Messtechnik GmbH
Current outputs
4.8.2.1
Accuracy
Maximal error of the current outputs is ±0.1 % of the actual reading + 0.05 % full scale flow.
4.8.2.2
Load
Standard version:
 500 ohms
Explosion-proof version:
 500 ohms
HART minimum load:
250 ohms

The minimal voltage required at the passive current output terminals to ensure correct operation is 10 V.
The maximum voltage of 30 V must never be exceeded.
The maximum load is dependent on the applied terminal voltage and can be calculated using the following
equation:
R max 
U  10V 1
 500 
23mA
maximum load
600 Ohm
500 Ohm
load
400 Ohm
300 Ohm
200 Ohm
100 Ohm
0 Ohm
0V
5V
10 V
15 V
20 V
external power supply voltage
25 V
30 V
Rmax
Figure 1: Maximum Loads of the Current Outputs
4.8.2.3
HART®
A number of connection possibilities are available for HART® communication, with the condition that the
loop resistance is less than the maximum load specified in Section 4.8.2, “Current outputs” on page 28. The
HART® interface is connected via terminal 11 and 12 or 41 and 42 respectively. The minimum load
impedance of 250 Ω must be adhered to.
For information regarding operation of the transmitter using the HART hand-held terminal, see “Operation
of the UMC4 transmitter using the HART hand-held terminal.”
Page 28 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Communication via Siemens PDM®
PDM® is the configuration software of Siemens that is used to operate HART ® or Profibus PA compatible
devices.
To connect a desktop or laptop computer to the UMC4, a HART® interface is required in addition to
communication software such as PDM. The HART® interface, which has two connections, converts the levels
of the RS 232 interface or USB interface into an FSK signal (frequency-shift keying).
For further details see section 3.4.10, “Connection to the transmitter” on page 19
4.8.2.4
Damping
The damping of the output signals is programmable from 1 to 60 seconds.
The standard factory setting is 3 seconds.
4.8.2.5
Error indication
An error in, or failure of the meter can be indicated via the current outputs or the status output. The current
outputs can be set to a failure signal (alarm) of I < 3.8 mA or I > 22 mA. The status output can be configured
as a make or break contact.
4.8.2.6
Low flow cut-off
The low flow cut-off can be set to values between 0 and 20% via the menu. The set value is a percentual
value of the upper-range setting. If the measured value is lower than the set cut-off, the flow rate will be
set to 0.0 (kg/h). This results in the analogue output being set to 0/4 mA, and the pulse output will cease
generating pulses.
Binary Outputs
The binary outputs utilise NPN transistors for the transmission of their data. To collect this data (e.g.
Pulses), the connection of an external circuitry is necessary. Image 3 depicts an example of such a circuit.
For use in EX atmospheres, refer to the EX-supplement manual for additional connection requirements.
Image 3: Example for wiring a Binary output
OPERATING MANUAL TMU-W / UMC4
Page 29 of 118
Heinrichs Messtechnik GmbH
4.9
Construction details UMC4
The UMC4 with the standard field enclosure possesses a variety of mounting options, making it adaptable to
available local circumstances.
Mounting / Dimensions
Horizontal pipe mounting
Vertical pipe mounting
Page 30 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Wall mounting
Pipe mounting with a junction box
OPERATING MANUAL TMU-W / UMC4
Page 31 of 118
Heinrichs Messtechnik GmbH
Weight
Approx.:
2.4 kg (5.5 lbs) (separate UMC4 transmitter without mounting system)
Material
Enclosure:
Aluminium pressure die-casting, max. 0.5% Mg; yellow chromate.
Chemically resistant paint.
Terminal box:
Aluminium pressure die-casting. Chemically resistant paint
Connection neck:
1.4301 stainless steel
Mounting bracket:
Page 32 of 118
2 mm thick sheet stainless steel / (6 mm with ships approval)
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
4.10 Construction details UMC4-RM
The UMC4-RM is designed as a rack-mount transmitter designated for installation and operation in an area
protected against harsh environmental conditions.
Dimensions
OPERATING MANUAL TMU-W / UMC4
Page 33 of 118
Heinrichs Messtechnik GmbH
Mounting bore pattern
Note: The above depiction is not in scale and is therefore not suitable to be used as a mounting template.
Weight
Approx.:
1.1 kg (2.4 lbs)
Materials
Main electronic hood:
1 mm thick sheet stainless steel
Terminal hood:
1 mm thick sheet stainless steel
Base plate:
2 mm thick aluminium plate
Display Foil:
Polyester (Breakdown Voltage = 16.5 kV)
The aluminium base plate is conductively connected to the ground plane of the electronics and provides an
optimal ground plane for the electrical connection to the ground plane of the systems rack.
The enclosure possesses no Ingress protection and must be protected from environmental influences.
Page 34 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
5. Electrical Connections UMC4
5.1
Mains connections and their protection classes
Auxiliary power (1)
19 V to 36 V DC (nominal 24 V DC)
24 V AC
+ 5 %, − 20 %
50/60 Hz
90 V - 265 V AC
50/60 Hz
(1): For custody transfer operations, the transmitter is only available as a 24 VDC version.
Power input
4.5 VA
Main fuse:
5x20 mm IEC 60127-2
Main voltage
19V ... 36V DC
24V AC
90V ... 265V AC
rated Current
250mAT
250mAT
250mAT
rated voltage
250V AC
250V AC
250V AC
breaking capacity
1500A / 250V AC
1500A / 250V AC
1500A / 250V AC
Table 5: Fuse properties
Fuse Type: Little Fuse Series 0215.250 HXP
UMC4 Fuse holder
Wiring diagrams
Terminal Designation of the Power connections
Ex Type of protection
Transmitter Type
Pin designation
UMC4
L, N and PE
UMC4-RM
L, N and PE
Standard
None
Ex ec
(Protected by
(Non Ex)
x
x
x
Ex db)
x
Table 6: Power Terminals Designation
On DC Voltage devices, L represents the “+” and N the “-“ of the supply lines.
Power connection terminals UMC4
OPERATING MANUAL TMU-W / UMC4
Power connection terminals UMC4-RM
Page 35 of 118
Heinrichs Messtechnik GmbH
The output terminals
For connection in explosive environments, the output signals of the UMC4 are available in various-protection
classes.
The designations of the terminals varies depending on the ordered protection class.
The terminal designation and their protection class can be taken from the following table:
Terminal Designation of the Signal Outputs
Terminal designation
Designation
and (Polarity)
Current 1, 4 to 20mA with
HART
®
11(-) and 12(+)
Type of protection
Ex-device
Ex ia
13(-) and 14(+)
x
16(-) and 17(+)
(passive pulse)
46(-) and 47(+)
Binary output 2 (status or 2nd
19(-) and 20(+)
passive pulse output)
(Non-Ex)
x
x
x
x
43(-) and 44(+)
Binary output 1
device
non-Ex ia
41(-) and 42(+)
Current 2, 4 to 20mA
Standard
x
x
x
x
x
x
49(-) and 50(+)
x
Table 7: Signal Output Terminals Designation
UMC4
UMC4-RM
Output Signals shown in Ex-ia type of protection
Sensor connection
Depending on the purchased transmitter variant and its associated sensor type, the connection to the UMC4
can be configured either with a pigtail cable, an IP67 terminal box, or in the case of the UMC4-RM with external
terminal blocks.
The pigtail cable version is mounted to the UMC4 during construction, is considered an integral part of the
transmitter and may only be dismantled/replaced by trained personal.
Page 36 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Wire designation of the sensor cable
Type of protection
Designation
Terminal / Pin designation
Non-Ex
Ex ia
(Standard)
Sensor lines
SENSOR1 +
1
x
x
SENSOR1 -
2
x
x
SENSOR2 +
3
x
x
SENSOR2 -
4
x
x
TIk-
5
x
x
Temperature sensor -
6
x
x
Temperature sensor +
7
x
x
TIk+
8
x
x
EXCITER1
9
x
x
EXCITER2
10
x
x
Shield
x
x
Shield
Table 8: Wire designation of the sensor cable
For the connection between the sensor and transmitter, one of the following cables is supplied:

Non-explosion proof applications
SLI2Y(ST)CY 5 x 2 x 0.5 mm²
grey (max. 300 m)

Explosion-proof applications
SLI2Y(ST)CY 5 x 2 x 0.5 mm²
blue (max. 300 m)
(blue for explosion-proof applications, grey for non-explosion proof applications).
Customer specific cable lengths can be ordered during the ordering process.
Attention: Cables not certified by the manufacturer may impair the accuracy of the measurement as well as
EMC compliance.
Suitable alternative cables:
SLI2Y (ST) C11YÖ 5 x 2 x 0.5 mm or
SG [5(2 LiY 0.50)St]FStC11Y
OPERATING MANUAL TMU-W / UMC4
Page 37 of 118
Heinrichs Messtechnik GmbH
5.1.3.1
Connection via Terminal Box
When both sensor and transmitter are fitted with terminal boxes, the connection cable is prepared at both ends
for the connection to the terminals and cable gland of the terminal box.
Schirm / shield
10 rs / pink
9 gr / grey
8 sw / black
5 vio / violett
7 bl / blue
6 rt / red
4 ge / yellow
3 gn / green
2 ws / white
1 br / brown
Outer shield
Schirm / shield
rs / pink
- 10
gr / grey
- 9
sw / black
- 8
vio / violett
- 5
bl / blue
- 7
rt / red
- 6
ge / yellow
- 4
gn / green
- 3
ws / white
- 2
br / brown
- 1
The outer shield is connected to the terminal box cable glands at both ends, The inner shields of the wire-pairs
are connected to each other (drilled together) and connected to the “Schirm / Shield” terminal in the terminal
box.
5.1.3.2
Connection via Connector Plug
When the associated sensor is fitted with a connector, the connection cable is prepared with a
pre-confectioned connector plug at one end, and wires for the connection to the terminals and cable gland of
the of the UMC4 terminal box at the other end.
Outer shield
The outer shield is connected to the terminal box cable gland at one end, and the connector’s outer-case at the
other end. The inner shields of the wire-pairs are connected to each other (drilled together) and connected to
the “Schirm / Shield” terminal in the junction box.
Caution:
The colours of the sensor’s wires in the terminal box may differ to the colours of the
connection cable’s wires between terminal box and transmitter! Decisive for the
connection is the numbers of the terminal in the terminal box and transmitter.
Page 38 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
UMC4 with pigtail connection cable
For connector pin assignment, see the table on page 37
For advice on cable glands: See also 4.5.1, “Installation conditions and cable glands” on page 23.
UMC4 junction box and WAGO terminals
For terminal assignment, see the table on page 37
For advice on cable glands: See also 4.5.1, “Installation conditions and cable glands” on page 23.
OPERATING MANUAL TMU-W / UMC4
Page 39 of 118
Heinrichs Messtechnik GmbH
UMC4-RM (Rack-Mount) and terminal blocks
For terminal assignment, see the table on page 37
The connected cable requires a strain relief near the terminal blocks
Page 40 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
6. Certificates and Approvals
6.1
Conformity to CE Directives
The measuring system complies with the legal requirements of the following directives:

2014/30/EU:
Electromagnetic Compatibility Directive

2014/34/EU:
ATEX Directive

2014/35/EU:
Low Voltage Directive

2014/68/EU: Pressure Equipment Directive
The attachment of the CE mark indicates that the device complies with the aforementioned directives. Refer
to section 9 “Declaration of Conformity” on page 115 for detailed information.
6.2
Explosion protection
The transmitters and sensors are separately certified for use in explosive atmospheres. When designated for
use in such atmospheres, it is essential that the Ex-supplement manual of the relevant device is obtained to
supplement this manual. The Ex-supplement manual contains additional important information regarding
installation in potentially explosive atmospheres.
Certificate numbers EX
Certification Type
UMC4
UMC4-RM
TMU-W
ATEX (Europe)
BVS 10 ATEX E110 X
BVS 20 ATEX Exxx X
DMT 01 ATEX E149 X
IECEx (Int.)
IECEx BVS 11.0088X
IECEx BVS 20.xxxxX
IECEx BVS 11.0084X
KCS (S. Korea)
12-KB4BO-0117X
12 KB4BO-0116X
Table 9: Certifications Explosion Protection
Type Approval certificates are available on our website www.heinrichs.eu, or upon request.
6.3
Custody transfer applications
The Combination UMC4/TMU-W004 and UMC4RM/TMU-W004 possess an evaluation certification according
to the OIML-CS Scheme for conformance to the following recommendations:
Certificate numbers OIML
Certification Type
UMC4
UMC4-RM
TMU-W004
OIML R139 (Hydrogen)
PENDING
PENDING
PENDING
OIML R137
PENDING
N/A
PENDING
Table 10: Certifications OIML
The declarations of conformity certifying the Heinrichs Messtechnik UMC4 family of transmitters for custody
transfer operations can be downloaded from our website at www.heinrichs.eu, or is available upon request.
For further documentation, please contact Heinrichs Messtechnik sales or service department.
6.4
Patents
Patent Pending
OPERATING MANUAL TMU-W / UMC4
Page 41 of 118
Heinrichs Messtechnik GmbH
7. Operation
7.1
Control unit BE4
Introduction
The UMC4 transmitter can be operated either by means of the installed BE4 control unit, a desktop or by
laptop computer in conjunction with PDM® software, or via HART® Communicator.
In the following sections, transmitter operation and parameterization are described using the integrated
BE4 control unit. To use the control unit the window mounted screw-on lid must be removed.
In Ex hazardous areas, ensure that all relevant safety regulations have been observed before opening the
window lid.
Control unit BE 4
Image 4: Description of the Control unit BE4
The Control unit BE4 has an integrated alphanumeric display with two 16-character lines (format 15 x 52
mm). To improve the readability of the display at low ambient brightness, it is equipped with a backlight.
Measurement data and settings can be read directly from this display.
Page 42 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Operating modes
The UMC4 can be operated in the following modes:
1.
Display mode:
In display mode, measured values can be displayed in various
combinations and UMC4 settings can also be displayed. Parameter
settings cannot be changed in this mode. Display mode is the
standard (default) operating mode when the device is powered up.
2.
Programming mode:
In programming mode, UMC4 parameters can be redefined. After
entering the correct password, changes that are permissible for the
customer (customer password) or all functions (service password
for technicians) can be realized.
Operator interface
Functional classes are
Legend
displayed as headings
beneath which displays and
Headline
parameters are shown in
Main menu
logical groups.
Beneath this is the menu
level, which lists all measured
Functional class
Functional class
Functionial class
Functionial class
value displays or the headings
for their underlying
parameters (parameter
level).
Function with
Function
with
numerical
Input
numerical Input
Numerical Input
153.40
Parameter level
All functional classes are
interlinked horizontally, while
Display
Display
Menu level subpoint
all sub-menus that are
assigned to a functional class
are displayed beneath the
relevant class.
OPERATING MANUAL TMU-W / UMC4
Function
valueFunction
selected
value
from selected
list
from list
Selection
[no]
___________
no
yes
Page 43 of 118
Heinrichs Messtechnik GmbH
The keys and their functions
There are six keys available for navigating through the menus and amending settings.
Caution:
Using sharp or sharp-edged objects such as pencils or screwdrivers to press the keys
may cause irreparable damage.
Cursor keys:
Using the cursor keys, the operator can change numerical values, give YES/NO answers
and select parameters. Each key is assigned a symbol as stated in the following table:
Descriptor
Symbol
Cursor key, arrow to the right

Cursor key, arrow to the left

Cursor key, arrow to the top

Cursor key, arrow to the bottom

Table 11: Key Assignment Control Unit
Esc key:
The “Esc” key allows you to cancel the current action. Pressing “Esc“ moves you to the
next higher level.
Pressing “Esc” twice moves you directly to the MEASURED VALUES functional class.
ENTER key:
Pressing the “Enter“ key moves you from the menu level to the selected Sub-menu
parameter level..
All entries are acknowledged with the “Enter“ key.
Image 5: BE4 Control unit Push-button description
Page 44 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Functional classes, functions and parameters
Functional classes are written in all upper case letters (headings). The functions beneath each functional
class are written in upper and lower case.
The various functional classes and functions are describes in Section 7.2, “UMC4 transmitter function”
starting on page 47.
The lower line contains the following elements:
-
Informational texts
-
YES/NO answers
-
Alternative values
-
Numerical values (with dimensions, if applicable)
-
Error messages
If the user attempts to modify values for any of these parameters without entering the required password,
the message “Access denied” will be displayed (see also 7.1.2, “Operating modes” on page 43 and 7.1.5.3,
“Passwords” on page 46).
7.1.5.1
Selection window/make a selection
In the selection window, the first line of the LCD always contains the heading, while the second line displays
the current setting. This setting is shown in square brackets if the system is in Programming mode.
Function name
[Selection]
In Programming mode (see 7.1.2, “Operating modes” on page 43), i.e. after a password has been entered
(see 7.1.5.3, “Passwords” on page 46), the operator can navigate to the desired setting by using the key
or the key and the operator can then confirm your selection by pressing “Enter“. To retain the current
setting, press “Esc”.
OPERATING MANUAL TMU-W / UMC4
Page 45 of 118
Heinrichs Messtechnik GmbH
7.1.5.2
Input window/modify a value
In the input window, the first line of the LCD always shows the heading, while the second line shows the
current setting.
Example:
Function name
-4,567 Unit
These modifications can only be made in Programming mode (refer to 7.1.2, “Operating modes” on page 43),
which means that a correct password (see 7.1.5.3, “Passwords” on page 46) must be entered. To move the
cursor from one decimal place to the next, use the orkeys. To increase the value of the decimal place
just under the cursor by “1,” use thekey, and use key to lower the number by 1. To change the minus
and plus sign, place the cursor in front of the first digit. To confirm and apply the change, press “Enter“. To
retain the current value, press “Esc”.
7.1.5.3
Passwords
Programming mode is password protected. The customer password allows all changes to be made that are
permissible for customers. This password can be changed when the device is first put into operation. Should
the password be changed, retain the new password in a safe place.
The UMC4 customer password in the device when delivered is 0002.
The service password allows for modification of all UMC4 functions. This password is not given to
customers.
For further information on customer passwords, see Section 7.2.2, “PASSWORD functional class” on page
53.
Display
The LCD display is designed for operation at temperatures ranging from − 20 °C to + 60 °C (-4° F to 140 °F)
without incurring any damage. However, at freezing or near-freezing temperatures, the display becomes
slow and readability of the measured values is reduced. At temperatures below − 10 °C (14 °F), only static
values (parameter settings) can be displayed. At temperatures exceeding 60 °C (140 °F), contrast decreases
substantially. When submitted to higher temperatures for long periods of time, the liquid crystals may dry
out.
Page 46 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2
UMC4 transmitter functional classes
The software functions of the UMC4 transmitter are divided into functional classes, are arrayed in a circle and
can be navigated by using the orcursor keys. To go back to your starting point (the MEASURED VALUES
functional class) press “Esc”.
Function Overview
(Main Menue UMC4)
MEASURED VALUES
CUSTODY
TRANSFER
PASSWORD
SETTINGS
SENSOR
COUNTERS
SETTINGS
TRANSMITTER
MEASUREMENT
PROCESSING
SELF-TEST
FLOW
SIMULATION
DENSITY
CURRENT OUTPUTS
TEMPERATURE
STATUS
PULSE OUTPUT
In the following, all software functions that can be accessed using the customer password are described.
Functions that are only accessible to the vendor (service functions) are not described in the present document.
OPERATING MANUAL TMU-W / UMC4
Page 47 of 118
Heinrichs Messtechnik GmbH
MEASURED VALUES functional class
The MEASURED VALUES functional class contains all functions for displaying the measured values.
MEASURED VALUES
functional class
MEASURED VALUES
Page 48 of 118
Mass flow
(QM)
Mass flow
Counter (F)
Volume flow
(QV)
Mass flow
Density
Counter forward
(F)
Mass flow
Temperature
Counter reverse
(R)
QV
Counter (F)
Density
QV
Density
Temperature
Display mode
start -up
Elapsed time
d h min
Raw values
Display mode
[QM]
_______________
QM
QV
Counter F
Counter R
Density
Temperature
QM + Counter F
QM + Density
QM + temperature
QV + Counter F
QV + Density
Raw values
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.1.1
Mass flow
After selecting the Mass flow function, the following will be displayed:
Mass flow
XXX.X kg/h
The LCD shows the current mass flow. The operator can define the display unit in the FLOW functional class
using the Mass flow QM unit function.
7.2.1.2
Volume flow
After selecting the Volume flow function, the following will be displayed:
Volume flow
XXX.X m³/h
Volume flow can only be displayed if density measurement has been calibrated and activated. Otherwise, an
error message is displayed. The operator can define the display unit in the FLOW functional class using the
Volume flow QV unit function.
7.2.1.3
Counter forward
After selecting the Counter forward function, the current reading of the forward flow counter will be displayed.
Counter forward
XXXXXXXX.XX kg
The operator can define the display unit in the COUNTERS functional class using the Unit of counters function.
7.2.1.4
Counter reverse
After selecting the Counter reverse function, the current reading of the reverse flow counter will be displayed.
Counter reverse
XXXXXXXX.XX kg
The operator can define the display unit in the COUNTERS functional class using the Unit of counters function.
OPERATING MANUAL TMU-W / UMC4
Page 49 of 118
Heinrichs Messtechnik GmbH
7.2.1.5
Density
Depending on the settings in the DENSITY functional class, the process or reference density will be displayed.
Density can only be displayed if the sensor is suitable for density measurement and has been calibrated
accordingly.
Density
XXX.X g/l
The operator can define the display unit in the DENSITY functional class using the Density unit function.
7.2.1.6
Temperature
After selecting the Temperature function, the following will be displayed:
Temperature
XXX,XX °C
The LCD shows the current temperature of the measured fluid in degrees Celsius, Fahrenheit or Kelvin.
7.2.1.7
Elapsed time
The LCD shows the operating time that has elapsed in d(ays), h(ours) and min(utes) since the system was
initialized and commissioned by the vendor:
Elapsed time
256 d 18 h 06 min
7.2.1.8
Mass flow + Counter forward
After selecting the Mass flow + Counter forward function, the current mass flow will be displayed in the first line
of the LCD:
XXX.X kg/h
XXXXXXXX.XX kg
The second line shows the value of the counter forward. The operator can define the display unit in the FLOW
functional class using the Mass flow QM unit function and the counter unit using the Unit of counters function in
the COUNTERS functional class.
Page 50 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.1.9
Mass flow + Density
After selecting the Mass flow + Density function, the following will be displayed:
XXX.X kg/h
XXX.X g/cm³
The first line of the LCD shows the current mass flow and the second the density of the measured fluid. You
define the display unit in the FLOW functional class using the Mass flow QM unit function and the density unit
using the Density unit function in the DENSITY functional class.
7.2.1.10
Mass flow + Temperature
After selecting the Mass flow + Temperature function, the following will be displayed:
XXX.X kg/h
XXX °C
The first line of the LCD shows the current mass flow and the second line the temperature of the
measured fluid. You define the display unit in the FLOW functional class using the Mass flow QM unit
function.
7.2.1.11
Volume flow + Counter forward
After selecting the Volume flow + Counter forward function, the current mass flow will be displayed in the first line
of the LCD:
XXX.X m³/h
XXXXXXXX.XX m³
The second line shows the value of the counter forward. The operator can define the display unit in the FLOW
functional class using the Volume flow QV unit function and the counter unit using the Unit of counters function in
the COUNTERS functional class.
7.2.1.12
Volume flow + Density
After selecting the Volume flow + Density function, the following will be displayed:
XXX.X m³/h
XXX.X g/cm³
The first line of the LCD shows the current volume flow and the second line the density of the measured fluid.
The operator can define the display unit in the FLOW functional class using the Volume flow QM unit function
and the unit for density measurement in the DENSITY functional class using the Density unit function.
OPERATING MANUAL TMU-W / UMC4
Page 51 of 118
Heinrichs Messtechnik GmbH
7.2.1.13
Display mode during startup
By choosing the Display mode during startup function the operator can define the default display. After the
operator switched the device on and did not touch any keys for a longer period of time, the defined default
display will be shown:
Display mode
[QM]
One of the following default displays can be selected.

QM (Mass flow)

QV (Volume flow)

Counter f(orward)

Counter r(everse)

Density

Temperature

QM + Counter f

QM + Density

QM + Temperature

QV + Counter f

QV + Density

and Raw values
7.2.1.14
Raw values
The “Raw values display” supports fault diagnosis and trouble shooting. Please inform our service department
about the clear text error messages and the contents of this “Raw values display.”
xxx.xxx
fff.ffff
ttt.tttt
eee.aaa
The displayed values have the following meaning:
xxx.xxx: Measure for the phase displacement between the sensor signals.
ttt.ttt:
Indicates the measured sensor temperature.
fff.ffff:
Indicates the current oscillation frequency of the system.
eee.aaa: Indicates the value of the excitation current (eee) and the sensor voltage (aaa).
Page 52 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
PASSWORD functional class
The PASSWORD functional class is comprised of the functions for entering and changing the customer
password and entering the service password. To cancel the current action, press “Esc“.
PASSWORD
functional class
PASSWORD
7.2.2.1
Customer
password
Password ?
0000
Change customer
password
Enter New password
0000
Service
password
Password ?
0000
Custodytransfer
Reset errors
Reset errors
Code 0000
Customer password
After selecting the Customer password function and pressing “Enter“, the following will be displayed:
Password?
0000
The numbers 0000 are displayed and can be changed by toggling the arrow keys as description in section
7.1.5.2, “Input window/modify a value”.
Pressing the “Enter” key verifies the password.
OPERATING MANUAL TMU-W / UMC4
Page 53 of 118
Heinrichs Messtechnik GmbH
If the entered password is correct, the following message will be displayed:
Password
valid
If the entered password is incorrect, the following message will be displayed:
Password
invalid
The customer default password upon delivered is 0002.
A valid customer password allows all software parameter changes to be made that are permissible for
customers. After the operator switched the device off or did not touch any keys for about 15 minutes, the
authorization to change settings related to password entry will automatically be cancelled. If the operator does
not enter a valid password, all settings can be displayed but not changed. Parameter changes via HART or
Profibus PA may be carried out any time without entering password.
7.2.2.2
Change customer password
After entering a valid customer password, you may change the existing password and enter a new one. After
selecting the Change customer password function and pressing “Enter“, the following will be displayed.
Enter New password
0000
The numbers 0000 are displayed and can be changed by toggling the arrow keys as description in section
7.1.5.2, “Input window/modify a value”.
Pressing the “Enter” key confirms the new password.
Note:
Press “Enter“ to confirm and save the new password. Make sure that you entered the
desired password.
A copy of the password should be kept in a safe place.
Reactivation of a transmitter at the vendor’s site due to a lost password is not part of our
warranty.
7.2.2.3
Service password
You do not need the service password for setting the functions necessary for operation.
The service password is reserved for service technicians and not provided to customers. Correct settings are
essential for proper operation of the device (e.g. parameterization and calibration values).
Page 54 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
COUNTER functional class
The COUNTERS functional class is comprised of the following functions:
COUNTERS
functional class
COUNTERS
Unit of
counters
Accumulation of:
[kg]
_________________
g
kg
t
m³
cm³
l
USG
UKG
USB
lbs
shton
lton
ft³
acft
oz
floz
xxQM
xxQV
Reset
counters
Reset counters ?
[no]
________________
no
yes
To change the current settings, enter the customer password. Otherwise, the settings can only be displayed but
not changed. To cancel the current action, press “Esc”.
OPERATING MANUAL TMU-W / UMC4
Page 55 of 118
Heinrichs Messtechnik GmbH
7.2.3.1
Unit of counters
After choosing the Unit of counters function and pressing “Enter”, the current forward and reverse
counter unit will be displayed:
Accumulation of:
[kg]
One of the following units can be selected.

Mass units:
g, kg, t, lbs, shton, lton and oz

Volume units:
m³, cm³, l, USG, UKG, USB, ft³, acf and floz

Programmable mass unit:
xxQM,

Programmable volume unit:
xxQV.
When the unit is changed, the counters will be reset to 0.00 automatically.
The volume unit only makes sense if the sensor has been calibrated for density measurement. Press “Enter” to
confirm and save the selection. Forward and reverse counters will now show the selected unit.
The valence of the programmable units are defined by the settings of the flow units described in sections 7.2.5.2,
“Factor mass flow QM programmable unit” on page 62 and 7.2.5.8, “Factor volume flow QV programmable unit”
on page 65.
7.2.3.2
Reset counters
To reset the totalizing counters, you need to toggle to [yes]. Forward and reverse counters will be reset at the
same time (0.00).
Reset counters
[no]
By pressing “Esc” or toggling to [no] the operator can cancel the current action without changing the counter
readings.
Page 56 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
MEASUREMENT PROCESSING functional class
The MEASUREMENT PROCESSING functional class is comprised of all functions that affect the processing of
the measured values.
To change the current settings, enter the customer password. Otherwise, the settings can only be displayed but
not changed. To cancel the current action, press “Esc”.
MEASUREMENT PROCESSING
functional class
MEASUREMENT PROCESSING
Damping
Damping
03 s
Low flow cut-off
Low flow cut-off
01 %
Low flow cut -off
hysteresis
Low flow cut -off
Hysteresis 01 %
Zeropoint
calibration
x.xxx kg/h
cal. ? [no]
_________________
no
yes
Filter
OPERATING MANUAL TMU-W / UMC4
Filter
[standard]
_________________
weak
mid
standard
strong
special 1
special 2
special 3
special 4
Page 57 of 118
Heinrichs Messtechnik GmbH
7.2.4.1
Damping
The damping value is intended to dampen abrupt flow rate changes or disturbances. It affects the
measured value display and the current and pulse outputs. It can be set in intervals of 1 second from
1 to 60 seconds. After choosing the Damping value function and pressing “Enter“, the following
selection field will be displayed:
Damping
03 s
The current damping value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.4.2
Low flow cut-off
The value for low flow cut-off (low flow volume) is a limiting value stated as a percentage that relates to the
upper-range value of the flow rate. If the volume drops below this value (e.g. leakage), the displayed value and
the current outputs will be set to “ZERO.” The value for low flow cut-off can be set from 0 to 20 % in 1-percent
increments. After choosing the Low flow cut-off function and pressing “Enter“, the following selection field will
be displayed:
Low flow cut-off
00 %
The low flow volume will be displayed and can be changed by toggling the arrow keys . After setting the new
low flow volume, you confirm your entry with “Enter“.
For devices used in custody transfer operations, you need to deactivate the low flow cut-off function,
i.e. to set this value to 0 %.
Page 58 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.4.3
Low flow cut-off hysteresis
The hysteresis of the low flow volume is the flow rate expressed as a percentage of the upper range
value by which the volume must fall below or surpass the set low flow volume in order to activate or
deactivate the function. The hysteresis of the low flow volume can be set in 1-percent increments
from 0 to 10 %. After selecting the Low flow cut-off hysteresis function and pressing “Enter“, the
following selection field will be displayed:
Low flow cut-off
Hysteresis 00 %
The current hysteresis will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry
7.2.4.4
Zero point calibration
Using the Zero point calibration function the operator can recalibrate the zero point of your meter in
the measuring system. Zero point calibration is to be realized after any installation procedure or after
any type of work has been performed on in the pipes near the sensor.
Caution:
It is important that this function is only performed when certain that the medium in
the sensor is not moving (flowing). Otherwise, the subsequently measured flow rates
will provide an incorrect zero-point. The sensor may be completely empty or filled
with medium. A partially filled sensor or a fluid containing air bubbles will lead to an
incorrect zero point calibration.
Note:
Calibrating a sensor filled with a fluid is preferable than calibrating with empty
tubes!
After choosing the Zero point calibration function and pressing “Enter“, the current remaining flow
will be displayed:
QM = 0,00 kg/h
cal. ? [no]
The operator can toggle between [yes] and [no].
After setting the new value, press “Enter“ to confirm your entry. Entering [yes] initiates a new zero point
calibration.
OPERATING MANUAL TMU-W / UMC4
Page 59 of 118
Heinrichs Messtechnik GmbH
7.2.4.5
Filter
For a noise reduction of the actual readings and minimizing disturbances on the sensor lines signal filter are
used. Following settings are available:
 weak
 mid
 standard (factory setting)
 strong
 special
Selecting of “weak”, “mid” or standard influences the dynamics of the actual readings just less. The setting of
damping time (see section 7.2.4.1, “Damping” at page 58 above) determines the dynamic behaviour of sensor
and transmitter. Filter “strong” operates as an intense low pass filter to reduce the noise of the actual readings.
Then if setting the damping time shorter than 3 seconds, the low pass filter determines the dynamic behaviour
to actual reading’s variations.
Note:
Setting of the filters “special 1” to “special 4” is carried out by the manufacturer.
These filter characteristics are optimized for specific custom applications, defined by
the manufacturer. Using these filters in standard applications, can result in a
degrading of the measurement signal! To use of these filters, consult the
manufacturer.
The operator can toggle between noise filter types with the arrow keys. After setting the new filter type, press
“Enter“ to confirm your entry.
Page 60 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
FLOW functional class
The FLOW functional class is comprised of functions that affect lower- and upper-range values and the
processing of the measured flow rates. In Programming mode (see 7.1.2, “Operating modes”), i.e. after a
password has been entered (see 7.1.5.3, “Passwords” and 7.2.2, “PASSWORD functional class”), the operator
can change the settings regarding flow. To cancel the current action, press “Esc”.
FLOW
functional class
FLOW
Mass flow QM unit
Mass flow in
[kg/h]
_________________
kg / s
kg / min
kg / h
t/h
g / min
lbs / s
lbs /min
lbs / h
shton / h
lton / h
kd/d
t/s
t/min
t/d
g/s
g/h
g/d
lbs/d
st/min
ston/d
lt/min
lton/d
oz/s
oz/min
oz/h
oz/d
xx/s
xx/min
xx/h
xx/d
Mass flow limit
hysteresis
Mass flow limit
hysteresis 01 %
Volume flow in
[l / h]
________________
Volume flow unit
QV
m³ / h
l/ h
l / min
l/s
USG / h
USG / min
USG / s
UKG / h
UKG / min
UKG / s
USB / d
MG / d
m³ / s
cm³/min
ft³ / min
acft / s
cm³/min
cm³/s
cm³/h
m³/min
m³/d
ft³/s
ft³/d
USB/s
USB/min
USB/h
floz/s
floz/min
floz/h
xx/s
xx/min
xx/h
xx/d
Factor QM
prog. unit
F = 001.00 kg
mass flow
QM range
QM Range = 100%
00120.00 kg/h
Mass flow
QM limit MIN
Mass flow limit
MIN = 10 %
Factor QV
prog. unit
F = 001.00 l
Mass flow
QM limit MAX
Mass flow limit
MAX = 90 %
Volume flow
QV range
QV Range = 100 %
0120.00 l/h
OPERATING MANUAL TMU-W / UMC4
Page 61 of 118
Heinrichs Messtechnik GmbH
7.2.5.1
Mass flow QM unit
Using this function, the operator can define the physical unit for all display functions, limit values and the
upper-range value of mass flow. After choosing the Mass flow QM unit function and pressing “Enter”, the
following selection field will be displayed:
Mass flow QM unit
[kg/h]
One of the following units can be selected:

kg/s, kg/min, kg/h, kg/d,

t/s, t/min, t/h,t/d,

g/s, g/min, g/h, g/d,

lbs/s, lbs/min, lbs/h,lbs/d,

shton/min, shton/h, shton/d,

lton/h, lton/min, lton/d,

oz(s, oz/min, oz/h, oz/d,

xx/s, xx/min, xx/h, xx/d (programmable mass flow unit)
Press “Enter” to confirm and save the selection.
A conversion factor can be entered as a substitute for a not available mass flow unit as described in the afterfollowing chapter 7.2.5.2, “Factor mass flow QM programmable unit” on page 62. In this case the unity xx is
selected into combination with the desired time unit.
7.2.5.2
Factor mass flow QM programmable unit
To display another mass flow unity than one of the predefined standard units a factor can be entered for the
conversion of the reading.
F = 001.0 kg
The factor always refers to the unity of kg.
A new factor can be entered by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
Page 62 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.5.3
Mass flow QM range
This function allows the operator to set the upper-range value for mass flow. The upper-range value takes on
the unit defined using the Mass flow unit function. The upper-range value will scale the current and frequency
outputs assigned to mass flow. After choosing the Mass flow QM range function and pressing “Enter”, the
following selection field will be displayed:
QM range=100%
XXXXX.XX kg/h
The current upper-range value for mass flow will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.5.4
Mass flow QM limit MIN
The MIN limiting value for mass flow can be evaluated via the status output. You enter the value as a
percentage of the set upper-range value. If the mass flow is lower than that limit value, the status output will
be set in case the corresponding assignment has been made. If the alarm function has also been activated for
the assigned current output, the applied current will change to < 3.2 mA or > 20.5 mA / 22 mA. After choosing
the Mass flow QM limit MIN function and pressing “Enter”, the following selection field will be displayed:
Mass flow limit
MIN = 10 %
The current MIN upper-range value for mass flow will be displayed and can be changed by toggling the arrow
keys.
Pressing the “Enter” key confirms the entry.
7.2.5.5
Mass flow QM limit MAX
The MAX limiting value for mass flow can be evaluated via the status output. You enter the value as a
percentage of the set upper-range value. If the mass flow surpasses this limit value, the status output will be
set in case the corresponding assignment has been made. If the alarm function has also been activated for the
assigned current output, the applied current will change to < 3.2 mA or > 20.5 mA / 22 mA. After choosing the
Mass flow QM limit MAX function and pressing “Enter”, the following selection field will be displayed:
Mass flow limit
MAX = 90 %
The current MAX upper-range value for mass flow will be displayed and can be changed by toggling the arrow
keys.
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 63 of 118
Heinrichs Messtechnik GmbH
7.2.5.6
Mass flow QM limit hysteresis
The hysteresis of the QM limiting values is the flow rate in percent based on the upper-range value and
indicates the value which must fall below or surpass the set limiting values in order to activate or deactivate the
function. The hysteresis of the QM limiting values can be set in 1-percent increments from 0 to 10 %. After
choosing the Mass flow QM limit hysteresis function and pressing “Enter”, the following selection field will be
displayed:
Mass flow limit
Hysteresis 00 %
The current hysteresis value will be displayed and can be changed by toggling the arrow keys.
Confirm the entry with “Enter”.
7.2.5.7
Volume flow QV unit
This function allows the operator to define the physical unit for all display functions and the upper-range value
for volume flow. After choosing the “Volume flow QV unit” function and pressing “Enter”, the following
selection field will be displayed:
Volume flow QV unit
in [m³/h]
One of the following units can be selected:

m³/d, m³/h, m³/min, m³/s, cm³/h, cm³/min,cm³/s

l/h, l/min, l/s,

USG/h, USG/min, USG/s,

UKG/h, UKG/min, UKG/s

USB/d, USB/h, USB/min, USB/s,

MG/d

ft³/d, ft³/min, ft³/s

acft/s

floz/h, floz/min, floz/s

xx/h, xx/min, xx/h.
Press “Enter” to confirm and save the selection.
A conversion factor can be entered as a substitute for a not available mass flow unit as described in the afterfollowing chapter 7.2.5.8, “Factor volume flow QV programmable unit” on page 65. In this case the unity xx is
selected into combination with the desired time unit.
Page 64 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.5.8
Factor volume flow QV programmable unit
To display another volume flow unit than one of the predefined standard units a factor can be entered for the
conversion of the reading.
F = 001.0 l
The factor always refers to the unit of l.
The Factor value for volume flow will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.5.9
Volume flow QV range
This function allows the operator to set the upper-range value for volume flow. The upper-range value takes on
the unit defined using the Volume flow QV unit function. After choosing the Volume flow QV range function and
pressing “Enter”, the following selection field will be displayed:
QV range=100 %
XXXXX.XX m³/h
The current upper-range value for volume flow will be displayed and can be changed by toggling the arrow
keys.
Pressing the “Enter” key confirms the entry.
Output and display of the measured value is only possible for mass flowmeters for which received a density
calibration.
OPERATING MANUAL TMU-W / UMC4
Page 65 of 118
Heinrichs Messtechnik GmbH
DENSITY functional class
The functional class DENSITY is comprised of the functions that affect the lower- and the upper-range value
and the processing of the measured density values. The additional service functions regarding density
calibration will not be described in these instructions.
DENSITY
functional class
DENSITY
Density
measurement
on/off
Measurement
[on]
________________
on
off
fixed
Density unit
Density in
[g / l]
_______________
kg / l
g/l
g / cm³
lbs / ft³
lbs / USG
kg/m³
lb/USB
xxx
Pipe emty below
0500 g/l
Fixed density
setting
Fixed density
0998.1 g/l
Display ref./
Process dens.
Display of
[Process density]
________________
Reference density
Process density
Factor density
progr. unit
F =0998 g/l
Ref.-density
Temp. coeff.
Temp.coeff.
+00.000 E-5/K
Density
lower range val
Density 0 % =
0600 g/l
Ref.-density
Ref. temperature
Ref. temperature
015.00 °C
Density
upper range val
Density 100 % =
1100 g/l
Ref.-density
operat. pressure
operat. pressure
001.000 bar
Density
limit MIN
Density
MIN = 0979.0 g/l
Calibr. density
hot medium
Frequency / temp.
measure [no]
_________________
yes
no
Density
limit MAX
Density
MAX = 1200.0 g/l
Measured. Values
hot medium
060.50°C 166.409 Hz
Rho= 0994.1 g/l
Density limit
Hyst.= 000.5 g/l
Finish calibration
of density
Calculate?
[no]
______________
yes
no
Density limit
hysteresis
Page 66 of 118
Density limit
for empty pipe
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.6.1
Density measurement on/off
This function allows the operator to activate density measurement. After selecting the Density measurement
on/off function, press “Enter” to display the following selection field:
Measurement
[on]
The operator can choose between the following settings:

on
density measurement is switched on

off
density measurement is switched off

fixed
density measurement is switched off; a fixed replacement value will be displayed and
used for calculating the volume flow
To confirm and apply the selection, press “Enter”.
If density measurement is switched on and the message “Density not calibrated” is displayed, no density
calibration was carried out by the vendor.
Note:
Density measurement can only be activated if density calibration has been performed.
If no density calibration has been performed, the density and volume flow values will
be set to “0.0” in the MEASURED VALUES functional class and the message “Density
unknown” will be displayed
7.2.6.2
Density unit
This function allows the operator to define the physical unit for all display functions and the density lower- and
upper-range value. After selecting the Density unit function, press “Enter” to display the following selection
field:
Density unit
[g/l]
The operator can choose between the following units:

g/l, kg/m³

kg/l

g/cm³

lbs/ft³

lbs/USG, lbs/USB

xxx
Press “Enter” to confirm and apply the selection.
A conversion factor can be entered as a substitute for a not available density as described in the after-following
section 7.2.6.3, “Factor programmable density unit” on page 68.
OPERATING MANUAL TMU-W / UMC4
Page 67 of 118
Heinrichs Messtechnik GmbH
7.2.6.3
Factor programmable density unit
To display another density unity than one of the predefined standard units a factor can be entered for the
conversion of the reading.
F = 0998.0 g/l
The factor always refers to the unity of g/l.
The Factor value for density will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.6.4
Density lower-range value
This function allows the operator to define the lower-range value for density measurement in the selected unit.
If density is equal or below this value, the assigned current output will be set to its initial value of 0/4 mA.
After selecting the Density lower-range value function, press “Enter” to display the following selection field:
Density 0 % =
XXXXX g/l
The current lower-range value will be displayed and can be changed by toggling the arrow keys.
Confirm the entry with “Enter”.
7.2.6.5
Density upper-range value
This function allows the operator to define the upper-range value for density measurement in the selected unit.
For this density, the assigned current output will be set 20 mA. The applied current of the current output
assigned to the density value is linearly interpolated based on the ratio between the measured value and the
difference between lower- and upper-range value.
After selecting the Density upper-range value function, press “Enter” to display the following selection field:
Density 100 % =
XXXXX g/l
The current upper-range value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
Page 68 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.6.6
Density limit MIN
The MIN limiting value for density can be evaluated via the status output and thus triggers an external alarm.
This value is entered as an absolute value in the unit defined using the Density unit function.
After selecting the Density limit MIN function, press “Enter” to display the following selection field:
Density limit
MIN = 0000.0 g/l
The current MIN limiting value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.6.7
Density limit MAX
The MAX limiting value for density can be evaluated via the status output. This value is entered as an absolute
value in the unit defined using the Density unit function.
After selecting the Density limit MAX function, press “Enter” to display the following selection field:
Density limit
MAX = 0000.0 g/l
The current MAX limiting value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.6.8
Density limit hysteresis
The hysteresis of the density limiting values indicates the absolute density value in the unit defined using
Density unit function. The measured density must fall below or surpass the set limiting values by the set
hysteresis value in order to activate or deactivate the function.
After selecting the Density limit hysteresis function, press “Enter” to display the following selection field:
Density limit
Hysteresis 000.0 g/l
The current limit hysteresis value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 69 of 118
Heinrichs Messtechnik GmbH
7.2.6.9
Density limit for empty pipe
If the measured density or the fixed value falls below this limiting value, the message “Empty pipe” will be
displayed, and an alarm will be triggered.
Press “Enter” to display the following selection field:
Pipe empty below
0500.0 g/l
The current limiting value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.6.10
Fixed density
If the operator selected the fixed option described in Section 7.2.6.1, “Density measurement on/off” on page 67,
density measurement will be switched off. The replacement value defined in the following selection field will be
displayed.
Press “Enter” to display the following selection field:
Fixed density
0998.1 g/l
The current fixed density will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
The density unit can be defined for all settings and displays as described in Section 7.2.6.2, “Density unit”.
7.2.6.11
Reference/process density display
When measuring density in a mass flowmeter, usually process density is displayed. Process density is the density
of the fluid at the measured temperature. Reference density can also be displayed as an option. In this case the
measured process density will be converted based on a reference temperature. To do so, the reference
temperature, the volume temperature coefficient of the fluid and the pressure at reference density (for gases)
must be known and have been programmed.
Volume measurement also depends on this setting. If “Process density” is set, the measured volume flow will
be displayed. If “Reference density” is set, a volume standardized to the reference density will be displayed.
Display of
[Process density]
The current operating mode for density measurement will be displayed and can be changed between the two
modes by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
Page 70 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.6.12
Temperature coefficient
In order to calculate the reference density using the process density, the temperature coefficient of the fluid
density must be known. In order to improve the resolution and facilitate data entry, the unit of the temperature
coefficient is set to 10 -5 1/K.
Temp. coeffic.
00.00 E-5/K
The current density temperature coefficient value in 10 -5 1/K will be displayed and can be changed between
the two modes by toggling the arrow keys.
7.2.6.13
Reference temperature
In order to calculate the reference density, the temperature to which the density relates is needed. The
temperature for fuel oil usually is 15 °C.
Ref. temperature
015.00 °C
The reference temperature will be displayed in °C and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.6.14
Operating pressure
This function has been prepared for the consideration of gas equations for the measurement of reference
density and volume for gases. In this software version, it will not be used for calculations.
operat. pressure
001.00 bar
The current value process pressure will be displayed in bar and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 71 of 118
Heinrichs Messtechnik GmbH
7.2.6.15
Density calibration hot medium
A single point density calibration can be made with a suitable mass flow sensor by the operator. The procedure
is described in detail in chapter 7.3, “Density calibration” at page 105.
With this function the necessary measurement of the resonant frequency and the medium temperature is
made. The sensor must be filled with a liquid medium. At a temperature of e.g. 60 °C hot water can be used as
harmless medium or for optimal results the process medium under normal operating conditions.
Frequency/temp.
measure? [no]
After selecting [yes] with the arrow keys, press “Enter” and the measurement is executed.
7.2.6.16
Measured values hot medium
The values of the function “Density calibration hot medium” above are displayed in the upper line. Pressing 2
times the green “Enter” confirms them without any change. Afterwards the density of the measured medium
has to be entered as reference in the lower line.
60.50°C 166.409 Hz
Rho = 0994,1 g/l
The density is always entered in the unit g/l (equivalently too kg/m³) and if required can be amended by toggling
the arrow keys.
7.2.6.17
Finish density calibration
In order to finalise and store the density calibration by both preceding functions it is necessary to complete
some internal calculations.
Calculate?
[no]
by toggling the arrow keys switch the displayed option to “yes” and press “Enter”. The reference values for the
density measurement are then calculated and saved. Furthermore, to activate density measurement the
density measurement must be activated as described in section 7.2.6.1, “Density measurement on/off” on page
67.
Page 72 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
TEMPERATURE functional class
The TEMPERATURE functional class is comprised of the functions that affect the lower- and the upper-range
value and the processing of the measured temperature. The additional service functions shall not be described
in this manual. Modifications can only be made in Programming mode (see 7.1.2 Operating modes), which
means that a correct password (see 7.1.5.3, “Passwords” and 7.2.2, “PASSWORD functional class”) must be
entered.
TEMPERATURE
functional class
TEMPERATURE
OPERATING MANUAL TMU-W / UMC4
Temperature
unit
Temperature in
[°C]
_____________
°C
°F
K
Temperature
lower range val.
Temperature
0% = - 010 °C
Temperature
upper range val.
Temperature
100% = + 100 °C
Temperature limit
MIN
MIN. temperature
- 050 °C
Temperature limit
MAX
MAX. temperature
+250 °C
Max. measured
temperature
permissible +250 °C
measured +197°C
Page 73 of 118
Heinrichs Messtechnik GmbH
7.2.7.1
Temperature unit
This function allows the operator to set the unit for temperature measurement. Press “Enter” to display the
following selection field:
Temperature in
[°C]
The set unit will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
All display windows, measuring ranges and limiting values refer to the selected unit.
7.2.7.2
Temperature lower-range value
This function allows the operator to define the lower-range value for temperature measurement. Lower
temperatures will set the assigned current output to the minimum value of 0/4 mA. The temperature is entered
in the set temperature unit. After selecting the Temperature lower-range value function, press “Enter” to display
the following selection field:
Temperature
0% = + 005 °C
The current lower-range value for temperature measurement will be displayed and can be changed by toggling
the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.7.3
Temperature upper-range value
This function allows the operator to define the upper-range value for temperature measurement. For this
temperature, the assigned current output will be set to the upper-range value of 20 mA. The applied current of
the current output assigned to the temperature value is linearly interpolated based on the ratio of the
measured value to the difference between lower- and upper-range value.
The temperature is entered in the set temperature unit. After selecting the Temperature upper-range value
function, press “Enter” to display the following selection field:
Temperature
100 % = +090 °C
The current upper-range value for temperature measurement will be displayed and can be changed by toggling
the arrow keys.
Pressing the “Enter” key confirms the entry.
Page 74 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.7.4
Temperature limit MIN
The MIN limiting value for temperature can be evaluated via the status output. This value is entered in the set
temperature unit.
After selecting the Temperature limit MIN function, press “Enter” to display the following selection field:
MIN temperature
-010 °C
The current MIN limiting value will be displayed and can be changed by toggling the arrow keys. If the measured
value falls below the limiting value, the “Alarm” status message will be displayed
Pressing the “Enter” key confirms the entry.
7.2.7.5
Temperature limit MAX
The MAX limiting value for temperature can be evaluated via the status output. This value is entered in the set
temperature unit.
After selecting the Temperature limit MAX function, press “Enter” to display the following selection field
MAX temperature
+ 250 °C
The current MAX limiting value will be displayed and can be changed by toggling the arrow keys. If the
measured value falls below the limiting value, the “Alarm” status message will be displayed.
Pressing the “Enter” key confirms the entry.
7.2.7.6
Max. measured temperature
After selecting this display, the largest measured temperature will be displayed. For comparison, the set
maximum limiting value will be displayed in the first line.
permissible +250 °C
measured
+197 °C
This value cannot be reset since it stores the maximum measured process temperature.
OPERATING MANUAL TMU-W / UMC4
Page 75 of 118
Heinrichs Messtechnik GmbH
PULSE OUTPUT functional class
The PULSE OUTPUT functional class is comprised of the functions regarding the pulse output.
PULSE OUTPUT
functional class
PULSE OUTPUT
Page 76 of 118
Pulse or frequency
output
Output of
[Pulses]
_______________
Pulses
Frequncy
Pulse output unit
Accumulation of
1.0 kg
______________
g
kg
t
m³
cm³
l
USG
UKG
USB
lbs
ston
lton
ft³
acft
oz
flox
xxQM
xxQV
Pulse value
1 pulse per
[1.0] unit
________________
0.001
0.01
0.1
1.0
10.0
100.0
Pulse width
Pulse width
0050.0 ms
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.8.1
Pulse or frequency output
The Pulse or frequency output function allows the operator to define whether pulses per represent a unit of flow
or a frequency between 0 and 1 kHz that represents an analogue output over the measuring range.
After selecting the frequency setting, the maximum frequency of 1 kHz will be generated when the upper-
range value for mass or volume flow is reached (depending on the selected pulse unit). If the flow rate falls
below the low flow volume, the actual frequency is 0 Hz.
After selecting the pulse setting, pulse value and unit the transmitter will determine the number of pulses per
flow volume. When choosing a combination of these settings that cannot be fulfilled in real time for the upperrange value (e.g. the number of pulses per time unit cannot be generated due to the pulse width which is too
large), the error message “Pulse width too large” or “Inconsistent parameter” will be displayed.
Press “Enter” to display the current setting:
Output of
[Pulses]
The current setting will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.8.2
Pulse output unit
This function allows the operator to define the unit to be counted. After selecting the Pulse output unit function,
press “Enter” to display the following selection field:
Accumulation of
1.0 kg
The current output unit will be displayed and can be changed by toggling the arrow keys.
the operator can choose between the following units:

Mass units:
o

Volume units
o

m³, cm³, l, USG, UKG, USB, ft³, acft, floz
progr. mass unit:
o

g, kg, t, lbs, ston, lton, oz
xxQM
prog. volume unit
o
xxQV
Pressing the “Enter” key confirms the entry.
The valency of the programmable units are defined by the settings of the flow units described in sections 7.2.5.2,
“Factor mass flow QM programmable unit” on page 62 and 7.2.5.8, “Factor volume flow QV programmable unit”
on page 65.
OPERATING MANUAL TMU-W / UMC4
Page 77 of 118
Heinrichs Messtechnik GmbH
7.2.8.3
Pulse value
This function allows the operator to define how many pulses will be output per unit counted. After selecting the
Pulse value function, press “Enter” to display the current unit:
1 pulse per
[1.0] unit
The current Pulse value will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the pulse values:
0.001, 0.01, 0.1, 1.0, 10.0, 100.0
Pressing the “Enter” key confirms the entry.
7.2.8.4
Pulse width
This function allows the operator to change the width of the output pulse to be output. If the pulse width is too
large for the actual pulse number, it will be reduced automatically. In this case the warning “Pulse output
saturated” will be displayed.
After selecting the Pulse width function, press “Enter” to display the following selection field:
Pulse width
0050.0 ms
The current pulse width will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
The maximum output frequency can be calculated using the following formula:
f 
1
 _ 1000 Hz
2 * pulse width[ms ]
If connecting to electrical counter relays, we recommend pulse widths greater than 4 ms; for electromechanical
counter relays the preset value should be 50 ms.
Page 78 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
STATUS functional class
The functional class STATUS is comprised of the functions for setting the status output.
STATUS
functional class
STATUS
7.2.9.1
Status output
active state
Output active
[closed]
_________________
closed
open
Status output 1
assignment
Output 1 assigned to
[Alarm]
________________
Forw. flow
Rev. flow
MIN QM
MAX QM
MIN density
MAX density
Alarm
IMP2 90°
Status output active state
The status output can be compared to an electrical relay that can function as make or break contact. For safetyrelevant applications, the operator will choose the break contact setting so that a power failure or failure of the
electronics can be detected like an alarm. In standard applications, the output is used as make contact.
The Status output state active state function allows the operator to define the behaviour of the status output.
Status output active
[closed]
The current active state will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

closed

open
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 79 of 118
Heinrichs Messtechnik GmbH
7.2.9.2
Status output 1 assignment
This function allows the operator to define to which event the status output is to be assigned. The most general
assignment is the alarm assignment because all set limiting values and the self-test function are then
monitored via the status output.
After selecting the Status output assignment function, press “Enter” to display the current assignment.
Output 1 assigned to
[Alarm]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

Flow direction recognition
o
o

Forward flow
Reverse flow
Limiting values:
o
MIN QM
o
MAX QM
o
MIN density
o
MAX density

All limiting values and error detection

Pulse output 2 for custody transfer operations
o
o
Alarm
IMP2 90°,
Pressing the “Enter” key confirms the entry.
When selecting the IMP2 90° setting, a second pulse output will be realised via the status output that can be
used for custody transfer operations.
Page 80 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.9.3
Status output 2 assignment
Instead of current output 2 a second status output may be selected for custody transfer operations.
It has the same assignment possibilities as status output 1. However, it cannot be used as a pulse output.
After selecting the Status output assignment function, press “Enter” to display the current assignment.
Output 2 assigned to
[not available]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

Standard setting
o


Flow direction recognition:
o
Forward flow
o
Reverse flow
Limiting values
o

Not available
MIN QM
o
MAX QM
o
MIN density
o
MAX density
All limiting values and error detection
o
Alarm
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 81 of 118
Heinrichs Messtechnik GmbH
CURRENT OUTPUTS functional class
The CURRENT OUTPUT functional class allows the operator to perform the settings for the current outputs of
the transmitter.
CURRENT OUTPUTS
functional class
CURRENT OUTPUTS
Page 82 of 118
Curr. output I1
0/4 - 20 mA
Curr. output I1
[4 - 21,6mA]
_______________
4 - 21,6 mA
4 - 20,5 mA
Curr. output I1
alarm
I1: alarm
[>22mA]
_______________
<3.8mA
>22 mA
not used
Curr. output I1
assignment
I1 assigned to
[Mass flow.]
______________
Mass flow.
Volume flow.
Density
Temperature
Curr. output I2
0/4 - 20 mA
Curr. output I2
[4 - 21,6mA]
_______________
4 - 21,6 mA
4 - 20,5 mA
Curr. output I2
alarm
I2: alarm
[not used]
_______________
<3.8mA
>22 mA
not used
Curr. output I2
assignment
I2 assigned to
[Temperature]
______________
not used
Mass flow
Volume flow
Density
Temperature
using HART:
standard I1: mass flow
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.10.1
Current output I1 4 to 20 mA
The “Current output I1 4 to 20 mA” function allows the operator to define the range in which the current output
is to be operated. Within the range from 4 to 21.6 mA (= 0 ... 110 %) HART® communication is not possible. The
range from 4 to 20.5 mA follows the NAMUR recommendation and covers the range from 0 to 104 % of the
measuring range. The standard range from 4 to 21.6 mA allows for a control of the measuring range of up to
110 %.
Press “Enter” to display the current setting.
Current output I1
[4] – 21.6 mA
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the settings:


4 – 20.5 mA
4 – 21.6 mA
Pressing the “Enter” key confirms the entry.
The following diagram displays the impacts of the selected setting
Current Output
25,0 mA
20,0 mA
Output
15,0 mA
10,0 mA
5,0 mA
0,0 mA
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
110%
120%
Measured value
4 - 20.5mA
NAMUR
4 - 21.6 mA
Standard
Figure 2: Current Output as NAMUR or Standard
OPERATING MANUAL TMU-W / UMC4
Page 83 of 118
Heinrichs Messtechnik GmbH
7.2.10.2
Current output I1 alarm
This function allows the operator to define the state taken on by the current output when a state of alarm is
detected. This information can be analysed in the control system. Press “Enter” to display the current setting:
I1 : alarm
[>22mA]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

not used
no alarm function

> 22 mA
current rise in the case of an alarm

< 3.8 mA
current reduction in the case of an alarm
Pressing the “Enter” key confirms the entry
7.2.10.3
Current output I1 assignment
This function allows the operator to define the measured value to be output as an analogue signal via current
output I1. When devices with HART® communication capabilities are used, current output I1 is usually assigned
to mass flow. Press “Enter” to display the current setting.
I1 assigned to
[Mass flow]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

Mass flow

Volume flow

Density

Temperature
Pressing the “Enter” key confirms the entry
Page 84 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.10.4
Current output I2 4 to 20 mA
The “Current output I2 4 to 20 mA” function allows the operator to define the range in which the current output
is to be operated. The range of 4 to 20.5 mA follows the NAMUR recommendation and covers the range from 0
to 104 % of the measuring range. The standard range of 4 to 21.6 mA allows for a control of the measuring
range of up to 110 %.
Press “Enter” to display the current setting.
Current output I2
[4] – 21.6 mA
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

4 – 20.5 mA

4 – 21.6 mA
Pressing the “Enter” key confirms the entry
7.2.10.5
Current output I2 alarm
This function allows the operator to define the state taken on by the current output when a state of alarm is
detected. This information can be analysed in the control system. Press “Enter” to display the current setting.
I2 : alarm
[not used]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

not used
no alarm function

> 22 mA
current rise in the case of an alarm

< 3.8 mA
current reduction in the case of an alarm
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 85 of 118
Heinrichs Messtechnik GmbH
7.2.10.6
Current output I2 assignment
This function allows the operator to define the measured value to be output as an analogue signal via current
output I2. Press “Enter” to display the current setting.
I2 assigned to
[Temperature]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:

Mass flow

Volume flow

Density

Temperature

not available (in this case the vendor setting must not be changed)
Pressing the “Enter” key confirms the entry.
Page 86 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
SIMULATION functional class
The functional class SIMULATION is comprised of the functions for simulating the outputs. If simulation is
activated, all output signals will be generated based on the selected type of simulation. The peripherals
connected to the device can be tested without a flowing product.
Simulation will be deactivated automatically if the operator switched the device off or did not touch any control
unit keys for about 10 minutes. Simulation can also be activated and controlled via HART ® commands.
SIMULATION
functional class
SIMULATION
Simulation
on / off
Simulation
[off]
_________________
on
off
Simulation
direct / QMabs
Simulation
[direct]
_______________
direct
QM, D, T
Simulation
QMabs
Set QMabs
+0900.000 kg/h
Simulation
density
Set density
0500.000 g/l
Simulation
temperature
Simulation
status output
Status output
[on]
_________________
on
off
Simulation
pulse output
Set frequency
0009.0 Hz
Simulation
curr. output I1
Set I1
I1 = 12.00 mA
Simulation
curr. output I2
Set I2
I2 = 12.00 mA
Set temperature
+090 °C
OPERATING MANUAL TMU-W / UMC4
Page 87 of 118
Heinrichs Messtechnik GmbH
7.2.11.1
Simulation on/off
The Simulation on/off function allows the operator to activate or deactivate simulation. If simulation is activated,
all output signals will be generated based on the selected type of simulation. The peripherals connected to the
device can be tested without a flowing product. Press “Enter” to display the current status.
Simulation
[off]
The current setting will be displayed and can be changed by toggling the arrow keys.
The operator toggle between “on” and “off.”
Pressing the “Enter” key confirms the entry.
Note:
Simulation will be deactivated automatically if the operator switches the device off or
does not touch any control unit keys for approximately 10 minutes.
7.2.11.2
Direct simulation
This function allows the operator to define whether simulation is comprised of the measurement of the three
physical values mass flow, density and temperature or whether the outputs will be set directly. Press “Enter” to
display the selected type of simulation.
Simulation
[direct]
The current setting will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:


Direct
QM, D, T
pulse and current outputs are programmed directly
a measurement is simulated
Pressing the “Enter” key confirms the entry.
If “direct” simulation is activated, any output will perform based on the settings described in Section 7.2.11.4,
“Direct simulation of outputs”. It is therefore recommended that the settings be defined before starting
simulation. They can then be purposefully changed during simulation.
The status of the outputs during measured value simulation based on the setting “QM, D, T” depends on the
selected simulation values of these three variables, the measuring range settings and the assignment of the
outputs. If, for example, the pulse output is assigned to volume measurement, it will be affected by all three
simulation values at the same time [V ≈ QM (T) / D (T)].
Page 88 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.11.3
Measured value simulation
If the operator selected the setting “QM, D, T” described in Section 7.2.11.2 on page 88, the following three
possible settings will affect the output behaviour during measured value simulation, where all measured values
are simulated at the same time.
7.2.11.3.1 Simulation mass flow QM abs
In order to simulate mass flow, the operator can define a “measured value.” The flow rates will be simulated in
both directions. All outputs will perform based on the simulated measured value.
Set QM abs
±0900.0 kg/h
The current value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.11.3.2 Density simulation
In order to simulate density/volume measurement, the operator can define a “density measured value.” If
volume measurement is assigned to an output, it will change depending on mass flow and density simulation.
All outputs will perform based on the simulated measured value.
Set density
0500.0 g/l
The current value will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
7.2.11.3.3 Temperature measurement simulation
In order to simulate a temperature, the operator can define a “measured value.” All outputs will perform based
on the simulated measured value.
Set temperature
+090 °C
The current value will be displayed and can be changed by toggling the arrow keys
Pressing the “Enter” key confirms the entry.
OPERATING MANUAL TMU-W / UMC4
Page 89 of 118
Heinrichs Messtechnik GmbH
7.2.11.4
Direct simulation of outputs
If the operator selected the setting “Direct simulation” described in Section 7.2.11.2, “Direct simulation” on
page 88, the following four possible settings will affect the output behaviour during measured value simulation,
where all measured values are simulated at the same time.
7.2.11.4.1 Status output simulation
The Status output simulation function allows the operator to purposefully activate the status output.
Press “Enter” to display the current state.
Status output
[off]
The current setting will be displayed and can be changed by toggling the arrow keys
The operator can toggle between “on” and “off”.
Pressing the “Enter” key confirms the entry.
7.2.11.4.2 Pulse output simulation
The Pulse output simulation function allows the operator to define a frequency to be assigned to the pulse
output. After selecting this function and pressing “Enter”, the following selection field will be displayed:
Set frequency
0210.0 Hz
The current value will be displayed and can be changed by toggling the arrow keys
The frequency range of the pulse output can be set from 6 Hz to 1100 Hz.
Pressing the “Enter” key confirms the entry.
7.2.11.4.3 Simulation current output I1
This function allows the operator to define a current for current interface 1. Press “Enter” to display the set
current.
Set I1
I1 = 10.50 mA
The current value will be displayed and can be changed by toggling the arrow keys.
Values in the range from 3.8 mA to 22.6 mA are permitted.
Pressing the “Enter” key confirms the entry.
7.2.11.4.4 Simulation current output I2
As described in Section 7.2.11.4.3, current output 2 can also be configured.
Page 90 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
SELF-TEST functional class
The SELF-TEST function class is comprised of the functions relating to the self-test of the sensor.
The diagnostic functions of the transmitter, which monitor the proper functioning of the electronics and the
software, are always active and cannot be switched off. The excitation current can be monitored in addition.
SELF-TEST
functional class
SELF-TEST
Sensor test
on / off
Sensor test
[off]
_______________
off
on
Max. deviation of
excitation
Max. deviation
020 %
Self-test
calibration
Calibration
[no]
_______________
no
yes
Monitoring
sensor
S1 7749 S2 7812
E 1280
Display of
sensoramplitudes
S1 090 S2 090 mV
112.8 Hz 12.8 mA
OPERATING MANUAL TMU-W / UMC4
Page 91 of 118
Heinrichs Messtechnik GmbH
The excitation current of each sensor in the system individually depends on the sensor itself, the fluid and the
installation conditions. If the excitation currents changes while the fluid remains the same, conclusions may be
drawn for e.g. potential wear and tear, viscosity changes or air bubbles. The operator has the possibility of
defining a “normal state” (“Self-test calibration”) and setting the limit for a permissible deviation. This function
is deactivated in the device when delivered.
7.2.12.1
Sensor test on/off
The Sensor test on/off function allows the operator to activate or deactivate the monitoring function of the
excitation current.
Sensor test
[off]
The current setting is displayed and can be changed by toggling the arrow keys.
The operator can toggle between “on” and “off.” The standard factory setting is “off.”
Pressing the “Enter” key confirms the entry.
7.2.12.2
Max. deviation of excitation
This function allows the operator to define a limiting value in the form of a percentage deviation from the
normal value. The excitation current is electronically limited to 50 mA (display value 500) and may take on
larger values for only a limited period of time (transient reactions).
Max. deviation
020 %
The current value will be displayed and can be changed by toggling the arrow keys.
When changing the max. deviation value, permissible fluctuations shall be considered.
Pressing the “Enter” key confirms the entry.
7.2.12.3
Self-test calibration
The magnitude of the excitation current not only depend on the sensor itself but also on the installation
conditions and the viscosity and density of the fluid. Therefore, the “normal” value can only be determined at
the installation site during operation using the Self-test calibration function.
Calibration
[no]
The current setting is displayed and can be changed by toggling the arrow keys.
When set to [yes], the “normal” value is automatically calculated. No additional information is required for this
function.
Pressing the “Enter” key confirms the entry.
Page 92 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.12.4
Monitoring of sensor amplitude and excitation current
The first line of this window contains the amplitudes of the sensor signals S1 and S2 in 10 µV. Both values
should be close to each other or identical (ideal case). The second line shows the excitation current in 10 µA
units.
S1 7749 S2 7812
E 1280
Example: The sensors have amplitudes of 77.49 mV and 78.12 mV. The excitation current is 12.8 mA. These
values are used as reference values for the self-test function. They are measured by using the function
7.2.12.3, “Self-test calibration” on page 92. Afterwards they can be displayed or edited by this function.
7.2.12.5
Display of sensor amplitudes
The first line of this window contains the actual measured amplitudes of the sensor signals S1 and S2. Both
values should be close to each other or identical (ideal case). The second line shows the excitation frequency
and current.
S1 090 S2 089 mV
112.8 Hz 12.8 mA
Example: The sensors have amplitudes of 90 mV and 89 mV. The excitation current is 12.8 mA and the actual
resonance frequency is 112.8 Hz.
The combination with the raw value display (see chapter 7.2.1.14, “Raw values” on page 52) supports the
analysis of all electrical signals between mass flow sensor and transmitter.
OPERATING MANUAL TMU-W / UMC4
Page 93 of 118
Heinrichs Messtechnik GmbH
UMC4 TRANSMITTER SETTINGS functional class
This functional class is comprised of the general settings (e.g. language) affecting the behaviour of the
transmitter.
Transmitter SETTINGS UMC4
functional class
SETTINGS
TRANSMITTER
Page 94 of 118
Language
Language
[German]
________________
German
English
Serial number
Serial number
123456
SoftwareRevision
Rev. 05.24
EECS78E3 CS78E3
Reset
system error
Reset error
[no]
_______________
yes
no
Device address
Profibus / Modbus
Device address
126
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.13.1
Language
Two languages are presently available in the control unit BE4: German and English.
Language
[English]
The current value will be displayed and can be changed by toggling the arrow keys.
Other languages such as Russian, French, Italian or Spanish will be available in a special version of the
BE4 control unit
Pressing the “Enter” key confirms the entry.
7.2.13.2
Serial number
With the help of the Serial number function, the transmitter is assigned to an order. This number provides
access to internal vendor data if the device needs servicing. The serial number is printed on the rating plate of
the transmitter. After selecting this function, press “Enter” to display the following information field:
Serial number:
123456
To ensure that the sensor, the transmitter and the documents created within quality management are assigned
correctly, this entry may not be changed.
7.2.13.3
Software version
When the function Software version is selected, the software version of the transmitter software is displayed.
Rev.
05.24
EECSA3C9 CSA3C9
The second line contains the hexadecimal checksum that was calculated via the program storage created during
program development and the microcontroller checksum of the same storage. Both checksums must be identical,
should the program storage not been damaged.
OPERATING MANUAL TMU-W / UMC4
Page 95 of 118
Heinrichs Messtechnik GmbH
7.2.13.4
Reset system error
The integrated diagnostic system of the UMC4 transmitter distinguishes between two types of errors (see also
Section 8.2, “UMC4 transmitter error messages”). Self-test errors such as problems with a sensor line or
inconsistent parameter inputs are displayed as textual error messages. Once the error has been eliminated, the
message automatically disappears from the display. For further information, see Section 8.2.3.1, “Display of
self-test errors”.
Errors that are attributable to system memory or software, division by zero, or a fault in the electronics unit are
designated as system errors. These error messages are not reset automatically after the error (usually of very
brief duration) is eliminated. Before resetting a system error manually, we advise you to contact our technical
service department. For further information, see Section 8.2.3.1, “Display of self-test errors”.
Reset error
[no]
The setting “no” will be displayed and can be changed by toggling the arrow keys.
If the operator toggles to [yes] and confirms the action by pressing “Enter”, the error messages disappear from
the display. If the message reappears shortly after, contact our technical service department for help.
7.2.13.5
Profibus/Modbus device address
Note: This setting is reserved for future communication developments and is at present unused.
Before connecting fieldbus devices to a bus system, the operator must define a device address. This address is
a unique assignment to a participant device in a bus system (similar to a street number).
After selecting the Profibus/Modbus device address, press “Enter” to display the set address:
Device address
126
The current Address will be displayed and can be changed by toggling the arrow keys.
Pressing the “Enter” key confirms the entry.
Page 96 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
SENSOR SETTINGS functional class
The SENSOR SETTINGS functional class is comprised of the settings regarding the mass flow sensor.
SETTINGS SENSOR
functional class
SETTINGS SENSOR
set by manufacturer
change only after recalibration
Sensor
constant C
Sensor constant
0150.30
Flow tube material
Flow tube of
[1.4571]
_______________
1.4571
1.4301
HC 4
HB 2
Tantal
Nickel
Sonder
Flow direction
Flow direction
[forward]
_______________
forward
reverse
forw. & revers
OPERATING MANUAL TMU-W / UMC4
set by manufactorer
Page 97 of 118
Heinrichs Messtechnik GmbH
7.2.14.1
Sensor constant C
Sensor constant C is the sensor calibration value for mass flow. This constant is defined when the flowmeter is
calibrated at the factory and can be found on the rating plate.
Sensor constant
+0150.00 kg/h
Caution:
Changing the sensor constant C to a value that differs from the value on the rating
plate of the sensor will result in incorrect mass-flow measurements.
Normally, the sensor constant is changed only when the device is calibrated, e.g. for a validation measurement
for a custody transfer operation.
Note:
The sensor constant must always be preceded by a plus or minus sign. The delivery
default setting is a plus sign. If inlet and outlet section are interchanged when the
device is installed (the flow direction is indicated by an arrow on the sensor), the
transmitter will display a “forward flow” negative measurement value. If the (plus or
minus) sign of the sensor constant is then changed without changing the actual value,
a plus sign will again be displayed. No changes need be made in the disposition of the
electrical connections (wires).
7.2.14.2
Sensor material
The Sensor material function allows the flow tube material code to be entered. This material code can be found
on the sensor rating plate. This setting is defined by the vendor when the device is first put into operation at the
factory.
Flow tube material
[1.4571]
The information in this field is purely of informative nature.
Page 98 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.14.3
Flow direction
This function allows the operator to define the flow direction that the transmitter will evaluate. Only “forward”
should be selected so as to prevent reverse flow from being measured. The standard factory setting is “forward
& reverse.” After selecting the Flow direction function, press “Enter” to display the current setting.
Flow direction
[forward]
The current direction will be displayed and can be changed by toggling the arrow keys.
the operator can choose between:
 forward
 reverse
 forward & reverse
Pressing the “Enter” key confirms the entry.
Flow direction
Output
(current, pulse,
frequency)
120%
100%
80%
60%
40%
20%
0%
measured flow rate
forward flow
reverse flow
forward & reverse
Figure 3: Influence of the Flow Direction Setting
OPERATING MANUAL TMU-W / UMC4
Page 99 of 118
Heinrichs Messtechnik GmbH
CUSTODY TRANSFER functional class
The custody transfer functional class comprises of all parameters of the UMC4’s user interface required to
configure the custody transfer mode. Furthermore, the Alarm-Logger and Configuration Audit Logger are also
situated in this menu where each alarm can be individually viewed. The Configuration Audit Logger records the
last 20 parameter amendments or events such as the opening of the custody stamp switch or a renewed
setting of the zero-point.
CUSTODY TRANSFER
functional class
CUSTODY TRANSFER
Alarm Counter
Alarm logger
read
Alarm logger
reset
Config. counter
Config. logger
read
Config. logger
reset
Page 100 of 118
Sticky Alarms
Sticky Alarms
[ no ]
no
yes
Reset Errors
Reset Errors
code 0000
Num of reset
errors
Num of errors
00007
00110
Alarm if
reverse flow
Alarm
[ no ]
no
yes
Entry No. 113 >
17d 05h 32 Min
Limit for reverse
flow
Limit
-0%
Only measure in
custody mode
Only measure in
custody tx[ no ]
No
yes
Pulse outputs at
alarm
at alarm
[no Alarm]
no Alarm
Stop IMP2
Stop IMP1,2
00036
Entry No. 036 >
15d 22h 37 Min
Pressing > button
lists the errors
Reset logger
[ no ]
no
yes
Pressing > button
lists the errors
Reset logger
[ no ]
no
yes
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.15.1
Alarm counter
The Alarm counter counts each time an alarm occurs or an alarm disappears. It is also incremented each time
the device is powered up. The Alarm counter is stored in the EEPROM and can be displayed by pressing the
“Enter” key.
Alarm counter
00035
7.2.15.2
Alarm Logger read
The Alarm Logger stores the most recent 30 changes of the device status (alarm occurring, alarm disappearing).
The Alarm Logger is stored in the EEPROM.
The entries of the logger can be read by pressing the “Enter” key.
Entry No. 036 >
15d 22h 37 Min
By using the orcursor keys it is possible to navigate through all saved entries.
7.2.15.3
Alarm Logger reset
The Alarm Logger can be reset. To enable a reset to be performed, the customer password must first be
entered (See 7.2.2.1, “Customer password” on page 53).
Reset
[no]
7.2.15.4
Config. counter
The Configuration Audit counter counts each time a parameter is amended. It counts if the custody stamp is set
or opened. The Configuration Audit counter is stored in the EEPROM and can be displayed by pressing the
“Enter” key.
Config. counter
00035
OPERATING MANUAL TMU-W / UMC4
Page 101 of 118
Heinrichs Messtechnik GmbH
7.2.15.5
Config. logger read
The Configuration Audit Logger stores the most recent 20 amendments made to parameters or the execution
of e.g. zero point calibration. The Configuration Audit Logger is stored in the EEPROM.
The entries of the logger can be read by pressing the “Enter” key.
Entry No. 105 >
09d 42h 12 Min
By using the orcursor keys it is possible to navigate through all saved entries.
7.2.15.6
Config logger reset
The Configuration Audit Logger can be reset. To enable a reset to be performed, the customer password must
first be entered. (See 7.2.2.1, “Customer password” on page 53).
Reset
[no]
7.2.15.7
Sticky alarms
In some applications it may be a requirement to keep an alarm message displayed even after the alarm itself
has disappeared („sticky“). Sticky alarm messages need to be acknowledged by means of customer access.
With the “Sticky Alarm” parameter, this kind of behaviour can be activated or deactivated. When deactivated
[off], the alarm message is removed from the display once the reason for the alarm has ceased to exist.
However, the Alarm Logger stores the occurrence of the alarm and the alarm counter is incremented.
The default value for “sticky alarm” is [no]
Sticky Alarms
[no]
7.2.15.8
Reset errors
When Sticky Alarms are activated, the alarm messages requires customer access (Password: 5773) for them to
be removed. To reset the messages a predefined code must be entered and validated. This function is also
available in custody transfer mode without having to open the custody stamp. The error counter is incremented
each time the function is performed. This counter is stored in EEPROM. The Alarm Logger as well as the
Configuration Audit Logger recognise the execution of this function.
This function can also be accessed via the “PASSWORD“ menu. (Password: 5773)
(See 7.2.2.1, “Customer password” on page 53).
Reset Errors
Code 0000
Page 102 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.2.15.9
Num of reset errors
Displays the error reset counter, which is incremented each time an alarm message is reset by customer access
(see above).
Num of errors
00007
7.2.15.10
Alarm if reverse flow
In some applications, it may be required to receive an alarm if there is reverse flow. With this parameter, the
Alarm can be activated or deactivated. If activated [yes] a reverse flow below the set value “limit for reverse
flow“ (see 7.2.15.11, “Limit for reverse flow”) will result in an alarm.
The default value for “alarm if reverse flow” is [no].
Alarm
[no]
7.2.15.11
Limit for reverse flow
An alarm is generated If the “Alarm if reverse flow“ is activated and a reverse flow more negative than this
parameter is present. This parameter is given in % of URV.
The default value for “limit for reverse flow” is [-0%]
Limit
-0%
7.2.15.12
Only measure in custody mode
If this parameter is set to “yes“ an alarm is activated when the custody stamp is opened.
The default value for “only measure in custody mode” is [yes]
Only measure in
custody tx [yes]
OPERATING MANUAL TMU-W / UMC4
Page 103 of 118
Heinrichs Messtechnik GmbH
7.2.15.13
Pulse outputs at alarm
This parameter selects the behaviour of the pulse outputs in the event of an alarm during a flow delivery.
The following choices are available for selection:


No Alarm - Both outputs continue pulsing
Stop IMP2 - Pulse 2 is stopped. Pulse output 2 is disabled while pulse output 1 continues delivering
pulses.

Stop IMP1,2 - Both outputs are stopped, no pulses are delivered.
The default value for “pulse outputs at alarm” is [Stop IMP2]
At Alarm
[Stop IMP2]
Figure 4: Pulse behaviour in the event of an alarm
Figure 4 depicts the behaviour of the two pulse outputs in the event of an alarm during a flow delivery
Page 104 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
7.3
Density calibration
For continuous processes, where only small variations of temperature are experienced and where only liquid
medias of similar density are used, the user can perform a density calibration locally by at the place of installation.
Note:
Density measurement is not available for all sensor types.
Before proceeding with the density calibration described in this chapter ensure that,
your sensor is capable of density measurement.
Conditions
For a local density calibration the following conditions must be fulfilled:

The sensor must be suitable for a density measurement. Contact our service department to check the
suitability of your sensor for a density calibration. For all suitable sensors, the Heinrichs offers a 3point calibration with an accuracy of at least 5 g/l. Sensors, for which the manufacturer calibration is
not available, are not suitable for local single point calibrations.

For the local calibration the medium’s density must be known or be determined by a suitable procedure
using exactly the same process temperature present in the sensor during measurement.

The medium must be liquid. For gaseous media the density calibration is not suitable.
Procedure
The density calibration procedure is as follows:

The transmitter is powered up and the sensor is filled with the medium under operating conditions.

A waiting period of at least 15 minutes is to be given, to allow the sensor including its enclosure and
flanges can adopt the operating temperature.

After the input of the customer’s password select functional class DENSITY and “density calibration hot
medium” (See chapter 7.2.6.15, on page 72).

Independent of the type of liquid medium and its temperature the beginning of the calibration is
confirmed by selecting “yes”. Thereupon the transmitter captures the medium temperature and
current resonant frequency of the sensor.

Subsequently the function “measured values hot medium“ must be selected, the displayed measured
values of temperature and frequency are confirmed by pressing the Enter-key two times. In the next
field the density of the medium must be entered in unit of g/l or kg/m³. (See also chapter 7.2.6.16 on
page 72.)

Subsequently, the function “finish density calibration” is used to finish density calibration. (See also
chapter 7.2.6.17 on page 72.)

Finally the function “density measurement on/off” activates the density measurement. (See also chapter
7.2.6.1 on page 67.)

Now the measured density and also volume flows can be indicated or assigned to one of the outputs e.g.
current output 2.
OPERATING MANUAL TMU-W / UMC4
Page 105 of 118
Heinrichs Messtechnik GmbH
Single Point Density Calibration
Density calibration
Single point density calibration
without manufacturer
calibration
Warm up of sensor and
medium to operation
temperature
15 minutes minimum waiting
and operationg time
Determination of
medium's density
Determination of medium
density by external
measurement
(Quality of medium's density
determination should be 1g/l)
Enter customer's
password
DENSITY
Switch over to functional class
density
Measurement
hot Medium
Go to subitem measuremnt
hot medium
Select "yes" to measure
actual frequency and
temperature
Measured values
hot medium
confirm the measured
temperature and frequency
Enter te density of your
medium
Finish density
calibration
Density on
Switch on density measurment
(herewith volume flow
meeasuremnt is also possible)
End
Page 106 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
8. Trouble shooting
Should you experience problems with your flow meter after the meter has been installed and commissioned,
go through the following checklist of regularly encountered errors before contacting our service department.
8.1
Self-help Checklist
Observation
Possible causes and remedy
The display is blank and there

The supply voltage cables are not properly connected. Check!
are no output signals

The transmitters fuse has blown.
For position and approved replacement types refer to section 5.1
on page 35.
The display is blank but the

Measuring electronics are defect. Contact HM service

The BE4 display unit has been rotated and was incorrectly seated
output signals are delivering
back onto its connector. Remove the BE4 unit and replace it
signals
The display contains
carefully onto its connector.

The BE4 Unit is defect. Contact service

A foreign language is set.
undefinable characters
Refer to section 7.2.13.1 on page 95 to change the language

The communication to the display has been disrupted.
Restart the transmitter by removing its power.
If the problem persists, contact our service department.
A measured value is displayed

but is not being updated on a
The Ambient temperature is below -25 °C
The function of the LCD display cannot be guaranteed below -25
regular basis
°C, but will return to full function once the ambient temperature
rises.
An error message is displayed

Delete the error as described in section 7.2.13.4 and restart the
transmitter. If the error persists, note the error code and refer to
section 8.2.3 for an error description.

If the error could not be rectified in this way, contact our service
department for further assistance
Table 12: Self-help Checklist
OPERATING MANUAL TMU-W / UMC4
Page 107 of 118
Heinrichs Messtechnik GmbH
8.2
UMC4 transmitter error messages
The integrated UMC4 transmitter distinguishes between two types of errors. Self-test errors such as
problems with a sensor line or inconsistent parameter inputs are displayed as text error messages. Once the
error has been eliminated, the message automatically disappears from the display. For further information,
see Section 8.2.3.1, “Display of self-test errors”.
Errors that are attributable to system memory or software, division by zero, or a fault in the electronics unit
are designated as system errors. These error messages are not reset automatically after the error (usually of
very brief duration) is eliminated. Before resetting a system error manually, we advise that you contact our
technical service department. For further information, see Section 8.2.3.2, “Display of system error”.
If the cause of any of the error messages described below cannot be eliminated, contact the device vendor.
Standard operating mode
The transmitter operates as described above. After the cause of the error message has been eliminated, the
message automatically disappears. The self-test for monitoring the excitation current can be activated or
deactivated via the “Sensor test” function.
Custody transfer mode
When the device is placed in Custody transfer mode, any errors will remain on the display until the error itself
has disappeared or until the error is acknowledged via the “Enter” button.
All error occurrences and parameters amended whilst in Custody Transfer Mode are logged in an Audit log file,
which can be requested via HART®.
Page 108 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
List of error messages
8.2.3.1
Display of self-test errors
Self-test errors are displayed as plain text in the set language (German or English) on the second line of the
LCD.
Display
Description
Possible cause of error and remedy
Empty-pipe detection has been
Product contains air bubbles/pipe is empty. Bubble-
activated.
free filling must be ensured.
(English)
empty pipe
The density of the medium is less
than the set limit for the empty
pipe detec-tion, pipe is empty.
Partially filled?
The exciter cannot stimulate the
pipes sufficiently to achieve the
This error can be caused through air / gas bubbles in
the pipes or through a partial filling.
necessary vibration amplitude
Power fail?
Detects a power disruption in
Check power supply
custody transfer applications during
a mass-flow of Q > 0.5 % from the
measurement range end-value. By
no-flow conditions, the power can be
disconnected without error. Only
valid when “Sticky Alarms” are
activated
malfunction T
Wire breakage / short circuit in the
temperature measurement circuit
Check the lines between temperature sensor and
transmitter.
Measure resistance of PT1000
malfunction S1
malfunction S2
Wire breakage / short circuit of the
Check the lines between sensor coil and transmitter.
sensor coil 1 connection.
Measure coil resistance.
Wire breakage / short circuit of the
Check the lines between sensor coil and transmitter.
sensor coil 2 connection.
Measure coil resistance.
OPERATING MANUAL TMU-W / UMC4
Page 109 of 118
Heinrichs Messtechnik GmbH
Display
Description
Possible cause of error and remedy
(English)
Does not vibrate The measuring tubes are not
Possible causes:
exc. too large
Air bubbles inside the flow tubes
vibrating
A excitation current exceeding the
limit will be detected.
Non-symmetrical filling of the flow tubes.
Foam arising through air contents or decompression
effects.
Residues from media on the tube walls e.g. wax from
hydrocarbons, crystallized media, condensate when
measuring gas.
Possible electrical reasons:
Check the wires and terminals between exciter coil
and transmitter.
Check exciter wires for short circuit with sensor
housing.
Remedy:
Fill or empty the flow tubes completely – disconnect
transmitter from power for min. 30 sec. – reconnect
to power.
exc. too small
Detects a too small exciter current /
Check the lines between excitation coil and
exciter voltage
transmitter.
Obligatory in Custody Transfer Mode
QM > 110 %
The mass-flow is 10 % larger than
Reduce the flow rate and adjust the measuring range
the set measurement range end-
if necessary.
Forward or Revers flow totaliser
Reset totaliser; possibly change to a bigger totaliser
overflow
unit
value.
OVERFLOW !
curr. 1 saturated The output of current interface 1 is
overdriven. Based on the selected
settings and the currently assigned
Check the upper-range value and the flow rate
settings.
measured variable, the current to be
output is > 21.6 mA.
curr. 2 saturated The output of current interface 2 is
overdriven. Based on the selected
settings and the currently assigned
Check the upper-range value and the flow rate
settings.
measured variable, the current to be
output is > 21.6 mA.
pulse out satur.
The pulse output is overdriven. The
Check pulse duration, pulse value, and measuring
current measured value requires a
range.
pulse rate, which cannot be
Check the flow rate.
generated based on the set values
for pulse duration.
Page 110 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
Display
Description
Possible cause of error and remedy
The measured temperature exceeds
Product temperature is too high; adjust the
(English)
Temperature >
MAX
the set upper-range value for
temperature range and the limit values if necessary.
temperature.
Temperature <
The measured temperature is below
Product temperature is too low; adjust the
MIN
the set lower-range value for
temperature range and the limit values if necessary.
params
Set parameters are inconsistent.
Check the parameter settings.
temperature.
inconsist
The set parameters are conflicting.
Example: Upper-range value, pulse value and pulse
duration must be matched in such a way that the
combination fits for all possible measured values.
ext. EEPROM
The Data Memory Chip (DSB) with
The BE4 control unit is not correctly seated in its
missing
the sensor calibration data and the
connector socket.
customer settings is not accessible.
Check socket seating.
Table 13: Self-Test Error List
Note:
Error message: “Parameter is inconsistent” (system error 0x0400)?
To generate a list of the inconsistencies, first enter a valid password and then an
invalid password. The control unit will show a list of current errors (only once). The
operator can then correct the inconsistent settings after again entering a valid
password.
OPERATING MANUAL TMU-W / UMC4
Page 111 of 118
Heinrichs Messtechnik GmbH
8.2.3.2
Display of system error
System errors consist of the message text “system error” and a 4-digit number in hexadecimal code. The
meaning of the individual error codes is described in the following table. If several errors occur at the same
time, the hexadecimal sum of the individual errors will be displayed. The errors are coded in such a way that
the individual errors can be easily identified. The sums are unique.
Descriptor label
Constant/
(never displayed)
display
SystemfehlerDiv0
0x0001
Computing fault / Division through Null
SystemfehlerIntEEProm
0x0002
Check sum transmitter data incorrect, New initialisation
SystemfehlerPruefsumme
0x0004
Check sum Sensor data incorrect.
Systemfehlerleeres EEPROM
0x0008
Ext. EEPROM present but blank / unwritten
SystemfehlerEEPROM
0x0010
Save / Read Memory value failed
SystemfehlerZeitkonstante
0x0800
Initialisation of the time constants failed
SystemfehlerMesswert
0x0200
Error in the calculation of the measured values
SystemfehlerFrequenz
0x0040
Error in the frequency / density measurement
SystemfehlerParameter
0x0400
Settings inconsistent
Systemfeh-
0x0800
Check sum in the parameter of the Custody transfer
lerRAM_pruefsumme
Description
required
measured value
Systemfeh-
0x1000
Check sum error in the program memory (Flash-PROM:
Systemfeh-
0x0800
Check sum error during communication between DSP
lerFLASH_pruefsumme
lerDSP_pruefsumme
SystemfehlerZaehler
sector 0xD0000 – 0xDFFFF)
and M16
0x0800
Totalisators are two-fold saved. Fault after inconsistency
during power up.
SystemfehlerWDG
0x0800
Internal Watchdog: Time limit exceeded
SytemfehlerSchreibfehler
0x0800
Memory cell in working memory (RAM) defect.
Systemfehler DSP Version
0x0080
The firmware of the DSP is not the current issue (not
aligned with that of the transmitters operating system)
SystemfehlerDSPKommu
0x2000
Communication between the DSP and microcontroller is
disrupted, processing of the measured values does not
take place.
Table 14: System errors
Page 112 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
8.3
Returning the Meter
If all attempts to return the device to an operational condition have failed, contact our service department to
arrange the return of your device for repair.
Before sending the device back for repair or servicing, please ensure the following steps have been
performed:

Always enclose a fully completed declaration of decontamination. You will find a template in
section 8.3.1

Ensure all medium residues have been removed, be sure to clean the seal grooves and recesses
thoroughly.

Provide a description of the encountered problem, providing as much information as possible as well
as a contact person for following correspondence.

Inform us of any special handling requirements you or your processes may have.
OPERATING MANUAL TMU-W / UMC4
Page 113 of 118
Heinrichs Messtechnik GmbH
Declaration of Decontamination
Company name: .............................................
Address: ........................................................................................
Department: ...................................................
Name of contact person: .........................................................
Phone: ..............................................................
Information pertaining to the enclosed Coriolis flow meter
Model: …..............................................................................................
was operated using the following fluid: ...............................................................................................................................
In as much as this fluid is water-hazardous / toxic / corrosive / combustible / a health hazard /
environmentally hazardous
we have performed the following steps:
 Checked all cavities in the device to ensure that they are free of fluid residues*
 Washed and neutralized all cavities in the device*
 Cleaned all seals/gaskets and other components that came into contact with the measured medium*
 Cleaned the housings and all surfaces*
*tick applicable items
We hereby warrant that no health or environmental hazard will arise from any fluid residues on or in the
enclosed device.
Date: .............................................................
Signature: ...................................................................................
(Name in Block)
Company Stamp
Page 114 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik
9. Declaration of Conformity
Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
November 2020
Heinrichs Messtechnik GmbH
Page 116 of 118
OPERATING MANUAL TMU-W / UMC4
Heinrichs Messtechnik GmbH
OPERATING MANUAL TMU-W / UMC4
Page 117 of 118
Version / Printed:
Heinrichs Messtechnik GmbH
05.11.2020 / 05.11.2020
Robert-Perthel-Straße 9
We reserve the right to make
Germany
specifications without notice
D 50739 Cologne
amendments to the technical
File Name:
TMU-W_UMC4_BA_20.02_EN.DOCX
Telephone: +49 (221) 4 97 08 - 0
Telefax: +49 (221) 4 97 08 - 178
Internet: http://www.heinrichs.eu
Printed in Germany
Number of Pages 118
Email : [email protected]
Page 118 of 118
OPERATING MANUAL TMU-W / UMC4