The Thermo VT 6060 M and VT 6130 M vacuum drying ovens are designed for gentle drying of temperature-sensitive loads in a controlled environment. These ovens feature a vacuum system that allows for efficient removal of moisture, preventing damage to heat-sensitive materials. The ovens are equipped with various safety features, including a safety valve and overtemperature protection. These devices offer a wide range of applications, including drying pharmaceuticals, chemicals, and other delicate materials.
Advertisement
Advertisement
Vacuum Drying Oven VT 6060 M/ VT 6130 M Operating instructions BEE Q Heraeus [=== ofS = — — То › * | o | ===) | E... —— e | EE ен не неее 1 | = 1] La Ty | q. 1,1 Valid: 11.2005 / 50033393 H Thermo ELECTRON CORPORATION Below is a list of the international Thermo marketing organizations. Postal adress Germany Thermo Electron LED GmbH D — 63505 Langenselbold Robert-Bosch-Straße 1 Enquiries from Germany Telefon Vertrieb 0800 1 536376 Service 0800 1 112110 Fax Vertrieb/Service 0800 1 112114 E-Mail [email protected] Enquiries from Europe, Middle East and Africa Tel. + 49(0) 6184 / 90-6940 Fax + 49(0) 6184 / 90-7474 E-Mail [email protected] Internet: www.thermo.com Postal address USA Thermo Electron Corporation. 275 Aiken Road Asheville, NC 28804 USA Enquiries from North America Phone + 1 800-879 7767 Fax + 1 828-658 0363 E-Mail [email protected] Enquiries from Latin America Phone + 1 828-658 2711 Fax + 1 828-645 9466 E-Mail [email protected] Enquiries from Asia Pacific Phone +1 852-2711 3910 Fax + 1 852-2711 3858 E-Mail [email protected] Thermo Electron LED GmbH, 63505 Langenselbold, Germany If translated versions of this manual are used, the German version is the authoritative basis. Subject to technical modifications. The safety concerning the protection of persons, environment and material to be treated mainly depends on the behaviour of the operating personnel of these units. A Please read and observe the following instructions carefully before starting the unit in order to avoid faults and resulting damage, especially adverse health effects. 1a 50033393 Information according EN 61010 WARNING NOTES e (Caution: General hazard area! AN e Caution: Hot surface! AMBIENT OPERATING CONDITIONS For indoor use only For altitudes up to 2000 m above sea level Temperature range from 5 °C to 40 °C Maximum relative humidity: 80 %, for temperatures up to 31 °C, decreasing linearly to 50 % relative humidity at 40 °C e Main power supply: Voltage fluctuations not to exceed +/- 10 % of the nominal value ELEKTRICAL DATA * Overvoltage category: II * Contamination level: 2 50033393 1b WEEE Compliance: This product is required to comply with the European Union's Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Electron has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Electron’s compliance with these Directives, the recyclers in your country, and information on Thermo Electron products wich may assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEEROHS. 1c 50033393 Table of Contents 1. 1.1 2. 3. 3.1 3.2 3.3 3.4 3.5 3.5.1 3.5.2 3.5.3 3.6 3.7 3.8 3.9 3.10 3.11 , 3.12 3.12.1 3.12.2 Introduction Use | Installation and Connection A Description of the Unit Control Panel Analog, digital pressure readout (option) Shutoff valve Venting valve Fine metering valve for inert gas operation Temperature measurement Load temperature recorder (option) Load temperature indication, switchable (option) Pt 100 connection (option) Temperature control Timer (option) Master switch Connection for a central monitor (option) Inert gas connection Installation of load temperature sensor for: - digital temperature readout (option) - recorder (option) - Pt 100 connection (option) Safety features A Safety valve Overtemperature protection Start-Up Service and Maintenance Appendix Technical Specs. Dimensioned sketches Interface RS 232 i.connection w. Eurotherm controller oon Interface RS 485 i.connection w. Eurotherm controller (option 50033393 O 0 ©) 0 © 0 Y Ww 9-11 10 11 - 13 14 15 16 17 18,17 19 20, 21 22, 23 1. Introduction Your new Vacuum Drying Oven is characterized by a high degree of economic efficiency and outstanding quality. Before putting your unit into operation, please read the following instructions carefully so that you'll avoid mistakes which might damage the oven. Paragraphs dealing with safely precautions are identi- fied by A Paragraphs identified by TE deal with the proper operation of the unit. The safety and performance of the unit will be guaranteed only if required checkouts, service and repairs are carried out by our Customer Service or by experts authorized by us. Please indicate the data listed on the nameplate if you have complaints or if you need to order spare parts. The vacuum drying oven is in compliance with: * DIN128 80 Part 1 * DIN VDE 0700 Part 1, IEC 335-1 50033393 3. 1.1 Use Prior to installing and operating your vacuum drying oven, you should consult the guidelines, if any, issued by your local occupational trade associations. The units are especially suited for the gentle drying of temperature-sensitive loads. The vacuum drying ovens must not be used for drying processes or heat treatments releasing combustible vapors which might form explosive mixtures with the air. Finally, the ovens are not designed for heat treatments of dangerous dusts or fibrous materials. 50033393 _4- Installation and Comection Transportation Place of instal- lation Inert gas operation {option) For installation and connection, please consult the guidelines, if any, issued by your local occupational trade associations. Care should be taken not to lift the unit by the door or the door handle, The unit must be put on a level, non-combustible surface which should be free trom vibration. Don't operate or leave the unit in corners where air cannoi circulate. The location itself must be dry and the ambient temperature should not exceed 40 °C. The unit must be placed in a location which is free of drafis. This is particularly important when it is to operate at low drying temperatures. If the unit is operated with inert gas, it will be supplied with an oxygen-displacing gas (e.g. N,) Therefore, make sure that the gases released by the system are properly evacuated and the room is aired out accordingly. If you install several ovens or if you fit them undemeath a laboratory bench, check to see if you need to observe special rules regarding ventilation. For possible regulations concerning ventilation, you should consult the guidelines issued by the local occupational trade associations. The following illustration shows the required clearances between the oven and the walls/adjacent furniture/equipment: 50033393 - 5- Base cabinet installation Stacking Installation <> E Vacuum oven VT 6060 can be fitted into a base cabinet to save on space. For safety reasons and in acc.w. DIN 4102, these cabinet bases must be of non-combustible materials. The following needs to be observed: * + + Connect the unit to the power supply. Connect the supply lines to the vacuum oven. Push it into the cabinet and make sure not to damage the lines. There must be sufficient clearance between the unit and the rear wall of the cabinet for outlet and blind port connections. Clearance to the side walls must be at least 25 mm, i.e. 50 mm to the ceiling. Units cannot be stacked on top of one another. The required connections are located on the rear wall of the unit. The required connection accessories for options are included in the scope of delivery and placed inside the chamber. Remove them from the chamber prior to the first start-up. The following connections are located on the rear wall of the unit: и 50033393 = a TTT > 00] - LS Line cord (not shown) Vacuum pump connection (1) (small flange DN 25), hose nozzle & 20 mm (accessories in the chamber) Inert gas hose connection (2) outer diam.4 mm, accessories in the chamber, cf. pge.19. Load temperature measurement (3) (option) (accessories in the chamber) Connection for a central monitor (4) (option), cf. pg.18. DN-connection to install the optional equipment, or other vacuum fittings (5). Line connection Connection of the vacuum pump Parts connection Prior to connecting the vacuum drying oven to the line, make sure that the power supply ratings coincide with those stated on the nameplate. The vacuum drying oven comes with a permanently installed line cable and grounded plug. Required line fusing is 16 A, slow-action type. Suitable vacuum pumps are those with a pumping speed from 1 - 30 m3/h The exhaust port with small flange DN 25 (DIN 28403) and the removable hose nozzle (J 20 mm) are located inside the chamber when the unit is delivered. They will have to be mounted on the back wall of the unit. (cf. ill. pge.6, pos.1). The vacuum pump is connected either with the vacuum hose, 32 x 16 mm and the hose nozzle or, if the nozzle has been removed, directly with the flange. In order to avoid damage to the vacuum pump (from vapors, condensation, etc.), make sure to observe the instructions of the respective pump manufacturer. Consult pages 16 - 18. 50033393 7. Description of the Unit, pages 8 - 19 Control Panel, Basic Oven Model ® ® я. _— и, y LE i gd || = = С | 50033393 Press.read-out analog Vacuum shutoff valve Venting valve Temperature controller Overtemperature protection Master switch 3.1 3.1 32 3.3 34 3.5 Press-readout analog : Press.readout digital (Option) Vacuum shutoff valve cf.Start-Up page 20 Venting valve cf. Start-Up page 20 Fine metering valve for inert gas ct.page 17 Temperature measurement 3.5.1. Temperature recorder (Option) rea чат - wdabkrd ma aT sam ELE Fam =" 4a DLL LLC cf. separate instruction manual No. 50047444 50033393 OPEN 5 AU CLOSED Vertical mark OPEN Horizontal mark clockwise = CLOSED counterclockwise = OPEN 3.5.2 Load temp.display switchable or E Га = ©) © а @ «© Pt 100 Pt 100 connection w. switch. Function: - For digital display, key is not pressed - Pt 100 output, key is pressed Display is extinguished Pt 100 3.5.3 Pt 100 connection Refer to page 17 for the installation of the sensor Pr 300 [L-L ) weil / white gelb / yellow grau / grey grün / green A Make sure to only connect current supplies/amplifiers whose voltage supply comes from the same safety circuit (cf. VDE-regulations 0100/410). (Example: via a safety transformer and VDE-regulation 0551) 50033393 - 10 - 3.6 Temperature Control Electronic temp.controlier "Digicon” Note: 3.6.1 Interface for temp-sensor Output signal of the temp.controlier Connection of external set point program Module with pin assignment °C à electron. controller w. digital display for temp. range up © 10 200 °С i” The electronic controller features a digital display for the set and actual temperature values. | Press the control dial and select the desired working temperature. The set value appears in the digital display. if you release the control dial, the actual value will appear. The LED (top left) indicates that the heating system is "ON". Keep in mind that the temperature sensor is mounted on the heated oven casing. Therefore, the temperature indicated is higher than that of the chamber. When the diode is plugged in, internal switches to external set point because of the built-in bridge between pins 3 and 4. Temperature reads as voltage 0-10 V at both sockets "X ext." 0V=0°C 10 \ = 200°С Admissible int.resistance: В; = 1kQ Connect the diode plug with connection cord to socket "W ayy ” (0... 20 mA). OmA=0°C 20mA=200°C Admissible load: R; = 500 ka 1+0... 20 mA 2 -0... 20 mA 3 Set point 4 Ext. sel poinl 5 5+0... 10Y 7-0...10Y Make sure to connect only current supplies/amplifiers with voltage coming from the same safety circuit (cf. VDE-regulations 0100/410). ria via a safety transformer and VDE-regulation 0551 50033393 - 11 - Temperature controller Kelvitron O t Heragua КТО Тетрегашге- _ Light-emitting diode display TA "heating" | + Setpoint- T= oy TA °C > Up / down keys В ai + Start - Up (= = Input, > = Display) Tum on main switch of the unit. » Green pilot lamp lights up. » Digital displays light up for approx. 10 sec. (test phase). » After the test phase the display will indicate the actual temperature value, con- troller output indication shows the status of the output: active / non active (pi- lot lamp). Press "°C" key. » The actual set point appears on display (flashing). Change set point with "+" or "-" key . (e Ifthe "+" or "-" key will be pressed longer, the value of set point in the dis- play runs faster: Set points up to 99.9 °C are in 0.1 °C steps adjustable. Set points up to 200 “C arein 1 °C steps adjustable. Press "°C" key once more. » The desired set point will be stored, the display does not flash, the display shows the actual temperature value. » The unit heats up to the desired temperature. (2 If the "°C" key will not be actuated, the old set point is maintained. 50033393 - 12 - Temperature measurement Thermicon P (programmable) Eurotherm 2404/P4 (programmable) (prog af | [кодов prog.cxt|2 ry . + “.. +. rm = sam a. BE + L r и " + e E о КД +++ 0 ELE +” E E hmin = © rE ras a... LC "A ddr ... чт E; ca LL... =" ra. me FA Aaa For description and start-up, see enclosed operating instructions e AO ZN For description and start-up, see enclosed operating instructions For interface pin assignment, cf. pages 27 and 28 50033393 - 13 - 3.7 Timer: Daily program timer Synchronous timer 24 h - Clock Function: To switch the healing system on or off | Switch function: TT Operation w. timer fs manual operation (w/o timer) Setting: At the start of operation, set lhe actual time on the program dial by turning il to the right. Select the desired switching intervals by locking down the tabs The locked dial segments define the length of the process interval. Minimum switching frequency: 15 minutes. Function: Synchronous time switch for switching the oven on and off; 24 h duration at 50 Hz line frequency 20 h duration at 60 Hz line frequency Ca * + pee * + A - - - Li + - * + = + Settina: Function of the mode switch Position |: To switch on the oven after the set lime has expired, or for operation without a synchronous timer. Position Il: To switch off the oven after the set time has expired. 50033393 -14 - Caution: Power failure 3.8 Master switch The premounted dial (24 h) is valid for a line frequency oí 50 Hz. For a line frequency of 60 Hz, the enclosed 20 h dial must be substituted. The control button is released from the dial by impact from below (snug fit}. Mounting of the 20 h dial - Sel timer to O position - Mount 20 h dial in O position - Push on lhe control button. In case of a power failure, please note thal the unit has an operational reserve of approx. 90 h. Function: Turn the control knob to connect or disconnect the unit from all poles. O | O * Turn on the master switch, — The green pilot light comes on. 50033393 - 15 - 39 Connection for The unit is equipped so that it can be connected to a central warning systems (f.ex. telephone or building alarm monitor systems) (potential-free contact) Function: When the temperature safeguard (lemperature limit cut-out) and timer respond (connection/disconnection of the heating system), the appropriate message will be transmitted to the customer-installed alarm system. The potential-free contacts (1 make and 1 break contact) accommodate the following circuits: * power supply circuit: max. 250 V AC, 3 À * SELV/SELV-E circuits (cf. VDE 0100/410): cf. also connection diagram Po - - I Tn Flix | - == l 12% | | ! —© 9 о + ED Po dix КЛ Kx 7 Six: Kx | NC ( | \ 12 | \ | } | PS | = \ | , pi al | | ; e - Ax | ON — > > — : _ _ o Br Lh / Optical Accoustic / | signal signal AE A 50033393 - 16 - 3.10 Inert gas con- 3.11 nection A Installation of the load temp. sensor for: For the connection of inert gas, the customer is to provide a pressure relief valve set to 1 bar. For technical safety reasons, make sure not to change the 1 bar setting. Connect the gas supply to the unit. There is a gas nozzle (cf. ill.) on the back wall of the unit. Outer diam.: 4 mm. i и FT > И ча Ио \ Друо > a E Inert gas connection | | )) | en a) § |» CL y ve y ® - ; ms —" X - digital temperature display (option) - recorder (option) - Pt 100 connection (option) With the load temperature measurement module you've the possibility to connect an additional Pt 100 resistance thermometer (option), cf. ill. A temperature sensor is soldered into the small flange. The connection cord is equipped with a 5-pin plug which needs to be plugged into the 5-pin socket on the rear wall of the unit. (cf.ill.) WII Am La | 50033393 2 47- 3.12. Safety features 3.12.1 Safety valve The safety valve at the flange prevents an inad- missible rise in pressure when the unit operates with protective gas (rated excess pressure < 0.1 bar). The vacuum drying oven, provided with the fittings for inert gas connection, is equipped with a safety valve. The tee with mounted safety valve is inside the chamber in those units with option for an additional load temperature measurement. During assembly, make sure not to damage the safety valve. For assembly prior lo start-up, proceed as follows: Assembly of the Safety valve (cf.ill.): * Place O-Ring (Pos. 1) on centering ring (Pos.2) * Attach rings (Pos.1 & 2) to safety valve (Pos. 3) * Attach safety valve (Pos.3) with the rings to the socket (Pos. 5) " Install the ring segments (Pos.4) and fasten them with the clamp (Pos, 6) When operating the oven without load temperature sensor, mount the safety valve directly at the connection socket (cf.ill.) 50033393 - 18 - 3.12.2 Overtemperature protection A Protection of the oven Protection of the load The vacuum drying oven is equipped with a tempera- ture limit cut-out, Thermal Safety Class 2 in acc.w. DIN 12880 Part 1. When set correctly, the lemperature limit cut-out protects the vacuum drying oven, the surrounding area and the load against an inadmissible rise in temperature. The temperature limit cut-out is independent of the temperature controller. When it functions properly, it positively disconnects the healer from all poles. Setting: When the control knob is set to the limit stop, the temperature limit cut-out protects the oven. The temperature limit cut-out protects the load if it is set 20 °C higher than the temperature set on the controller. To ascertain the temperature at which the temperature limit cut-out responds, and thus to be able to adjust it properly, the unit must be put into operation. Set the control knob to the limit stop. After the selected working temperature has been reached, reset the temperature limit cut-out to the trigger position. If the cut-out responds, the red warning light comes on, and the heating system is permanently disconnected from all poles. Set the temperature limit cut-out 20 °C higher than the cut-off temperature indicated on the dial. Press the key to release the temperature limit cut-out. Should the temperature limit cut-out disconnect (red warning light comes on), inspite of being set properly, proceed as follows: 1. Check and correct, if necessary, the setting of the temperature controller and the trigger position of the temperature limit cut-out. 2. 1f you're unable to release the temperature limit cut-out, request a Service Technician. The optimum temperature limit cut-out adjustment is achieved if you turn back the control knob by at least two graduation marks. Then release it and finally set it approximately one graduation mark above the cut-off point. To change the setting, press down and release the temperature limit cut-out. The red warning light goes out. Operational test (cf. paragraph above "Setting": The temperature limit cut-out should, at reasonable intervals, be tested for proper operation. | 15 recommended that this test be carried out routinely by the authorized operating personnel, e.g. prior to the start of a longer work process. 50033393 - 19- Start-Up Securing the Shelves Removing the shelves Loading Note Turning the unit on A Evacuation Venting Controller setting When putting the oven into operation, familiarize yourself with the regulations issued by your local trade associations! Remove all accessories from the chamber. Place each shelf on its supports and push it into the oven until it touches the rear wall to secure it. When pulling the shelf out, make sure that the safety calches slide underneath the supports to prevent the shelf from tilting. Push the shelves all the way in so that they touch the rear wall. Lift them up and remove them. When loading the unit, make sure not to scratch the glass panes in the door. Observe the load limit per shell. (cf. section on Technical Data) Arrange the load on the shelves in a way that heat flow is not impaired. The temperatures which the controller indicates refer to the shelves only. Don't set anything on the floor of the chamber. * Shut the door * Turn on the master switch - the green pilot light comes on * Set the temperature controller to the desired working temperature * Set the temperature limit cut-out 20 °C above the working temperature Don't touch anything around the door areal You could get burns from the protruding rim of the inner casing * Close the air inlet valve {marking in vert.pos.). * Turn on the vacuum pump * Put the lever of the shutoff valve in the "OPEN" (vertical) position * Monitor the pressure inside the chamber at the pressure indicator * Put the lever of the shutoff valve in the "CLOSED" (horizontal) position * At the end of operation, vent the vacuum chamber. To do this, open the air inlet valve {marking in horiz.pos.). The lowest working temperature of the vacuum oven must be set to at least 10 ”C above ambient temperature. 50033393 - 20- Operation with inert gas A Note Shutdown General remarks on operation Following evacuation, nitrogen enters the chamber through the fine metering valve until pressure inside corresponds to the atmospheric pressure. Subsequently it is possible to evacuate the unit again and to vent it a second time with inert gas (depending on the type of application). If the fine metering valve is open following pressure equalization, approx. 10 [/Min of gas will be released into the area where the unit is placed. * Turn off the master switch * Shut down the inert gas supply * Vent the unit * Turn off the vacuum pump The heating system arrangement achieves optimum temperature uniformity at a vacuum < 10 mbar. At this vacuum, heat transfer occurs through radiation and the aluminum shelves. At a higher pressure (lower vacuum), the heat conductivity of the gas molecules in the chamber increases. However, improved heat conduction from the heated walls to the load is offset by an increased loss of heat across the glass door. Heat convection is caused if the pressure should rise until it matches atmospheric conditions. This would lead to higher temperatures in the upper portion of the chamber. 50033393 - 21 - 5. Maintenance A Proper operation and safety of the unit will be and repairs guaranteed only if maintenance, service, repairs and modifications are carried out by experts authorized by Thermo . The unit must be checked at appropriate intervals to make sure it functions correctly. Windows The vacuum oven is equipped with a pressure-proof pane against the risk of implosion. It prevents broken glass from {lying around the laboratory area if the inner window pane were to burst. The inner and the outer glass panes must be checked at regular intervals for scratches or other surface damages. If you detect any sort of damage to the glass panes, shut down the unit and have a service technician replace the window. Readjusting the If the door does not close properly or is no longer door leakproof, adjust the bearing pin and the catch: Adjusting the Unscrew the cover panel (cf.ill.above) bearing pin Remove the layer of insulation. Release the Phillips screw of the bearing pin. Use a small spindle to prevent the bearing pin from rotating also. After you've loosened the Phillips screw, turn the bearing pin with the spindle, and determine the desired setting tor height and depth. Retighten the Phillips screw and check if the door is tight. Adjusting the Release the catch nut M10 with a wrench (SW 17) catch and turn the catch clockwise for at least one complete т rotation. After this, retum the catch to the same position. (E=- Retighten nut M10. (cf.left ill.) qd | 50033393 - 22 - Replacing the gasket Hints on care Remove the defective gasket and install a replacement. Subsequently you need to submit the unit to a leak test. Inner casing: Commercially available detergents in small quantities can be used, but no acids, chloric solvents or saline solutions. Con | It's best to just wipe them off with a wet cloth. 50033393 - ?3- 6. APPENDIX ak, vn Fagin © Technical Data: >“ Materials: | aaa ERAN A A 6 eT Gad eins 18358 1h ве | -:::VT.8060 M =;| RIM nile TAN - VT 6130 M ax “ . | Unil 1 Innee casing Shelves Door gaskel Outer casing heat-reistant polyester varnish Galvanized sheet sleel-on outside coaled with Standard Option 1.4301 1.4571 Aluminium Sil.rupper Viton Standard Option 1.4301 1.4571 Aluminium Sil.rupper Vilon EE if 1; 2 TA Wi q oa 3 A A aa ME ME pp AT ‘Colours: lag : E ed So Hirt "e xh a a FR Artem nn Aes prit fée Ta LA od 87 ey PARADA ae A ] Outer casing, grey-white Modules, agate grey Operaling elements light, blue RAL 9002 RAL 7038 RAL 5012 RAL 9002 RAL 7038 RAL 5012 fee Не с PIATRA à Dimensions 11 x Sl e Sra nA; SH is AA ATA „м чая, eee cf. dimensioned sketch P- 26 mm Ext: dimensions: Lak and alone os Made meet Tue той Aal | wc Fi м mais Casing T, incl. door handle 630 810 mm Ba 744 895 mm H, 576 720 mm NÉ dimensions: «ét miodsirue m SSE Aw se E t, 371 529 mm b, 415 495 mm h, 345 489 mm height betw. shelves hi,=135 hi,=158 mm hi, =124 hi,=124 mm hi, =124 mm Chamber volume 53 128 | “Dimensions, connections: CAT za инс BL me syrah "il A AA dc EN. fe: = . A 1 щит Vakuum port DN 25 DN 25 + hose nozzle Z 20 2 20 mm add'l. connection (cf. options) DN 25 DN 25 Inern gas conn.: inner e 4 inner e 4 mm fiir hose 26x4x1 mm „Gewicht: SE за ar = ap eh es [Dia seis i e EC RE 2 мы. beled edi iene YO CI FA IDEA = io Egil Empty welg ht 82 153 kg max. load 40 60 kg max. load: per shelf 20 20 kg Electrical data: esa aka - 4 pe RE HE VERA nen A iz TL Зое, dr sos CE Rated voltage 230/120 230 VAC Rated freq. 50/60 50/60 Hz Rated current 1,7/1,5 2,2 kW Rated power 7,4/12,7 9,6 A Class of prozection | Type of protection IP 20 IP 20 50 033393 24 Technical Data VT 6060 M VT 6130 M Times: (minutes) Rise time (98 %) from 25 “C to: 200 °C 150 “C 70 °C Holding power (Wh/h) *) 200 °C 150 °С 70 °С Temperatures**: Rated temp. °C variation in space 200 °C 150 °C 70 °C DIN 12880 variation in time at rated temp. (*C) DIN 12880 Accuracy of digital temp. display in % of full scale reading Attainable ultimate vacuum mbar (hPa) Leak rate mbar 1/sec Pumping speed 75 min 75 min 40 min 550 370 200 200 °C H H HH == Lo -N UTN < 0.5 H HH 140 min 115 min 70 min 870 550 280 200 °C Pan <J 7N 7 A < 0.5 1 to 30 m°/h | *) In line with temp. of placement area ** Measured on the shelves in acc. w. DIN 12 880 Part 2. Times and temperatures under vacuum of < 10 mbar Gases released w. inert gas connection: < 10 I/min. 50033393 . 25 - Dimensioned Sketches — lá , ere rr | LI L Ne... aa EE OE TEC ec. EE [ IL e J Li J = Hz Ÿ Size 6130 Size 6060 - Rights reserved to make technical changes without express notice - - 26 - 50033393 Eurotherm Controller with Interface RS 232 Assignment of the 25-pin D-Sub socket: 13 ] Pin 1 5 1 A 2 0000©0000©00000 3 0009000000000 4 25 - 14 5 6 7 8 20 Shield RX ( reception ) Tx ( transmission ) ( connected to 5) ( connected to 4) ( connected to 8 and 20 ) COM ( common, ground) ( connected to 6 and 20) ( connected to 6 and 8 ) All unlisted pins are not wired The interface pin assignment is designed to permit direct connection to an IBM or IBM-compatible computer if the computer is equipped with a 25-pin plug. If the computer interface has a 9-pin plug, a commercially available adapter (9-pin socket and 25-pin plug) can be used. Upon delivery of the unit, the controller address is set to 01, and the Baud rate to 9600. For other addresses or other Baud rates, the controller must be reconfigured (see operating instructions for the EUROTHERM controller). 50033393 - 27 - Eurotherm Controller with interface RS 485 Assignment of the 25-pin D-Sub socket: - cl 0000000000000 000009000000 25 14 Pin Г) == md — OWN CO en A tU О — Shield ( connected to 12 › TX + ( transmission + ) ( connected to 5 ( connected to 4) ( connected to 8 and 20) COM ( common, ground) ( connected to 6 and 20) RX + ( reception +) Rx ~ ( reception - ) Tx - ( transmission - ) ( connected to 6 and 8) All unlisted pins are not wired Upon delivery of the unit, the controller address is set to 01, and the Baud rate to 9600. For other addresses or other Baud rates, the controller must be reconfigured (see operating instructions for the EUROTHERM controller). Caution: In the case of some RS 422/485 interface cards for personal computers, it may be necessary to change the polarity of the signals Rx and Tx. If the controller does not respond to the computer interface, connect Rx+ of the PC interface to Tx- of the controller, and Tx+ of the PC interface to Rx- of the controller. 20033393 - 28 - ">
Advertisement
Gentle drying of temperature-sensitive loads
High economic efficiency
Inert gas operation (optional)
Digital temperature control
Temperature limit cut-out for safety
Vacuum operation
Shelves for load arrangement
Pressure-proof window
Easy to use control panel
Compact design
The vacuum pump connection is located on the back wall of the unit. Connect it either with the vacuum hose and hose nozzle or directly with the flange.
The maximum load per shelf for both models is 20 kg.
The temperature limit cut-out protects the oven, the surrounding area, and the load against an inadmissible rise in temperature. It is independent of the temperature controller.
Open the air inlet valve (marking in horiz. pos.).
Where are the Kelvitron temperature controller instructions located?
The Kelvitron controller includes separate operating instructions.
How to remove the shelves from a vacuum drying oven?
To remove the shelves, push them all the way into the oven until they touch the rear wall, lift them up, and then pull them out of the chamber. Use the safety catches to prevent the shelf from tilting during removal.
What is the catalogue number or document part number?
The document part number is 50033393 H.