EMC - Manufacturer’s Declaration and guidance

Guidance and manufacturer’s declaration – electromagnetic emissions

The Midmark Dental Light is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the Midmark Dental Light should assure that it is used in such an environment.

Emissions Test

RF Emissions CISPR 11

RF Emissions CISPR 11

Harmonic emissions IEC 61000-3-2

Compliance

Group 1

Class A

Class A

Electromagnetic environment – guidance

The Midmark Dental Light uses RF energy only for its internal functions. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The Midmark Dental Light is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Voltage fl uctuations/fl icker emissions

IEC 61000-3-3

Complies

Recommended separation distances between portable and mobile RF communications equipment and the

Midmark Dental Light

The Midmark Dental Light is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Midmark Dental Light can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the Midmark Dental Light as recommended below, according to the maximum output power of the communications equipment.

radiated maximum output power of transmitter W

0.01

0.1

1

10

100

Separation distance according to frequency of transmitter m

150 kHz to 80 MHz d = [3.5/V1] √P

0.12

0.37

1.17

3.69

11.67

80 MHz to 800 MHz d = [3.5/E1] √P

0.12

0.37

1.17

3.69

11.67

800 MHz to 2.5 GHz d = [7/E1] √P

0.23

0.74

2.33

7.38

23.33

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from structures, objects and people.

003-2190-99

English - 16

TP200 20-42-FO-00012 Rev A1 C2169 © Midmark Corporation 2016

Guidance and manufacturer’s declaration – electromagnetic immunity

The Midmark Dental Light is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the Midmark Dental Light should assure that it is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level Electromagnetic environmental - guidance

Portable and mobile RF communications equipment should be used no closer to any part of the Midmark Dental Light including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Conducted RF

IEC 61000-4-6

3V rms

150kHz to 80MHz

3V

Recommended separation distance: d = [3.5/V1] √P

Radiated RF

IEC 61000-4-3

3V/m

80MHz to 2.5GHz

3V/m d = [3.5/E1] √P 80 MHz to 800 MHz d = [7/E1] √P 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts

(W) according to the transmitter manufacturer and d is the recommended separation in meters (m).

Field strength from fi xed RF transmitters, as determined by the electromagnetic site survey, a Should be less than the compliance level in each frequency range. b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from structures, objects and people.

a Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be considered. If the measured fi eld strength in the location is which the Midmark Dental Light is used exceeds the applicable

RF compliance level above, the Midmark Dental Light should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Midmark Ultra Chair.

b Over the frequency range 150kHz to 80MHz, fi eld strengths should be less than 3 V/m.

003-2190-99

English - 17

TP200 20-42-FO-00012 Rev A1 C2169 © Midmark Corporation 2016

Guidance and manufacturer’s declaration – electromagnetic immunity

The Midmark Dental Light is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the Midmark Dental Light should assure that it is used in such an environment.

Immunity Test

Electrostatic discharge (ESD)

IEC 61000-4-2

IEC 60601

Test Level

± 6 kV contact

± 8 kV air

Compliance Level

± 2, 4, 6 kV contact

± 2, 4, 8 kV air

Electromagnetic environmental - guidance

Floors should be wood, concrete or ceramic tile. If fl oors are covered with synthetic material, the relative humidity should be at least 30 %.

Electrical fast Transient / burst

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

± 2 kV

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment.

Surge

IEC 61000-4-5

±1kV line(s) to line(s)

±2kV lines(s) to earth

± 0.5, 1 kV

± 0.5, 1, 2 kV

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

for 0.5 cycle)

T

40% U

T

(60% dip in U

T for 5 cycles)

70% U

T

(30% dip in U for 25 cycles)

T

<5% U for 5 s)

T

(>95% dip in U

T

3 A/m

0 % UT

(100 % dip in UT) for 0.5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT) for 25 cycles

<5 % UT

(>95 % dip in UT) for 5 s

3 A/m Power frequency (50/60 Hz) magnetic fi eld

IEC 61000-4-8

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the Midmark Dental Light requires continued operation during power mains interruptions, it is recommended that the

Midmark Dental Light be powered from an uninterruptible power supply or a battery.

Power frequency magnetic fi elds should be at levels characteristic of a typical location in a typical commercial or hospital environment.

003-2190-99

English - 18

TP200 20-42-FO-00012 Rev A1 C2169 © Midmark Corporation 2016

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • Adjustable light intensity for customized illumination
  • Composite setting for working with composite materials
  • Adjustable focus for precise light projection
  • Auto-on function for hands-free operation

Related manuals

Frequently Answers and Questions

How do I turn on the light using the auto-on function?
Pull the lighthead down (below horizontal) to turn the light ON
How do I adjust the light intensity?
Press (and release) the Intensity Button to advance through the intensity settings: High/Medium/Low
How do I activate the composite setting?
Press (and release) the Composite Button to produce an extra, low-intensity light beam