5 Installation

► Avoid deposits, accumulated gas and air in the pipe system�

The unit can be installed irrespective of the orientation if the following is ensured:

-No air bubbles can form in the pipe system�

-The pipes are always completely filled�

5.1 Recommended installation locations

Example of an optimised installation:

1

F

► Install the unit so that the measuring pipe is completely filled�

► Arrange for inlet and outlet pipe lengths� Disturbances caused by bends, valves, reductions, etc� are compensated for� It applies in particular: no shut-off and control devices are allowed directly in front of the unit�

2

5 x D

F

2 x D

S S

S = disturbance; D = pipe diameter; F = flow direction

10