14

Setup

Replace the Work Surface

1. Align the work surface rear pins with the openings on the brackets of the technology tray as shown.

2. Lower the front of the work surface and engage the latches at the front of the technology tray.

3. Turn the security key clockwise to lock the front latch.

4. Remove the security key.

Setup

Adjust the Manual Work Surface Lift System (Some Models)

1

If the work surface seems difficult to raise, or seems to fall too easily, the work surface manual lift system may need to be adjusted.

Note: This feature is not available on carts with the electronic work surface lift system.

1. Use a soft tool to pry and remove the rear column cap(1).

2. Using a 17mm (11/16") wrench rotate the exposed adjusting screw clockwise to increase the preload, counter-clockwise to decrease.

3. Raise and lower the work surface to test the adjustment.

Note: The lift is properly adjusted when the work surface will raise and lower with approximately the same effort.

4. Repeat the adjustment and test as necessary.

5. Install the rear column cap.

15

Setup

Install the Monitor

For Pivoting Monitor Mount:

1. Pivot the mount head to allow access to the mounting holes.

16

2. Use the mounting screws provided with the monitor to attach the monitor to the mount faceplate.

For Non-Pivoting Monitor Mounts:

Use the mounting screws provided with the monitor to attach the monitor mount to the monitor.

The mounting holes may need to be enlarged to accommodate the monitor mounting screws.

If mounting screws are not provided along with the monitor, refer to the monitor documentation for the correct mounting screw size.

It may be easier to connect the monitor cables while the monitor is off the cart.

Setup

Adjust the Monitor Lift System

If the monitor seems difficult to raise or creeps downward, the monitor lift system may need to be adjusted.

1. Using a 5mm hex wrench, insert it into the opening on the top of the column.

2. Rotate the hex wrench clockwise to increase the weight capacity of the adjustable height monitor mount, counter-clockwise to decrease.